MVX User Manual en
MVX User Manual en
MVX User Manual en
User Manual
MVX
User Manual
www.acoemgroup.com
Copyright ©
This document is the property of 01dB-Metravib. No part of this publication may be duplicated,
reprinted or transmitted by electronic, mechanical, photographic or other means, or recorded,
translated, edited, abridged or expanded without the prior written consent of the owner
Table of contents
1 Introduction ....................................................................................................... 7
1.1 Hazardous areas .......................................................................................... 7
1.2 Installation .................................................................................................... 7
1.3 Version compatibility .................................................................................... 8
1.4 Hardware identification ................................................................................. 9
2 Evolutions ........................................................................................................ 10
2.1 Main evolutions of firmware version 5.4.0-7 ............................................... 10
2.2 Main evolutions of MVX version 5.1 ........................................................... 10
2.3 Main evolutions of MVX version 5.0 ........................................................... 10
2.3.1 Monitoring on 100% of the signal ........................................................ 10
2.3.2 “Time waves on event” signals recording ............................................ 13
2.3.3 DAT option performance optimisation ................................................. 14
2.3.4 BGI: for the monitoring of structures ................................................... 14
2.3.5 GCI: for particle counting..................................................................... 14
2.4 Main evolutions of MVX version 4 .............................................................. 15
2.4.1 New « Shock FinderTM » indicator (SFI) ............................................. 15
2.4.2 Increase of the efficiency..................................................................... 16
2.5 Main evolutions of MVX version 3.1 ........................................................... 16
2.5.1 Compatibility with the ESA analysis by ONEPROD-System ............... 16
2.5.2 Short-term memory ............................................................................. 16
2.5.3 OPC parameters via XCOM ................................................................ 17
2.5.4 Priority management ........................................................................... 17
2.5.5 Alarm triggering delay ......................................................................... 17
2.5.6 Vector tracking .................................................................................... 18
3 MVX Operating ................................................................................................. 19
3.1 Principle ..................................................................................................... 19
3.2 Operating conditions .................................................................................. 19
3.3 Operating in “Monitoring” mode .................................................................. 19
3.4 Operating in “Predictive Maintenance” mode ............................................. 20
4 Set-up ............................................................................................................... 21
4.1 Setup with OneProd CSM .......................................................................... 21
4.2 Set-up with ONEPROD XPR ...................................................................... 21
5 Indicator lights on the front panel.................................................................. 22
6 Connecting MVX on an Ethernet network ..................................................... 23
6.1 Ethernet port selection ............................................................................... 23
6.2 IP addressing mode ................................................................................... 24
6.2.1 Default address mode ......................................................................... 24
6.2.2 MVX dynamic address mode .............................................................. 25
6.2.3 MVX fixed address mode .................................................................... 25
6.3 How to check the Ethernet connection ....................................................... 26
6.4 Examples ................................................................................................... 27
6.4.1 Connecting one MVX with one PC in a point to point connection through
the MVX right port.............................................................................................. 27
6.4.2 Two MVX and two PC connected together as a “local network” .......... 28
6.5 MVX Web interface .................................................................................... 29
7 Using the MODBUS output interface ............................................................. 30
7.1 General points ............................................................................................ 30
7.2 Communication principles .......................................................................... 30
7.2.1 Serial MODBUS with RS485 connection ............................................. 30
7.2.2 MODBUS-TCP with Ethernet connection ............................................ 31
7.3 Data format................................................................................................. 31
7.3.1 Available indicators ............................................................................. 31
7.3.2 Modbus requests ................................................................................. 32
7.3.3 Number of indicators ........................................................................... 32
7.3.4 Values of indicators ............................................................................. 32
7.3.5 Status of indicators .............................................................................. 33
7.3.6 Units of indicators................................................................................ 34
7.3.7 Threshold values ................................................................................. 35
7.3.8 Values of operating parameters .......................................................... 37
8 Maintenance ..................................................................................................... 38
8.1 Preventive maintenance operations ........................................................... 38
8.2 Replacement of the battery ........................................................................ 38
8.3 1st level diagnosis ....................................................................................... 38
8.3.1 Normal boot sequence ........................................................................ 38
8.3.2 Regular working operations................................................................. 39
8.3.3 Bias voltage monitoring ....................................................................... 39
8.3.4 Offset compensation ........................................................................... 39
8.3.5 Defects indicated by the MVX leds ...................................................... 39
8.3.6 Summary ............................................................................................. 41
8.4 MVX Hard Reset ........................................................................................ 42
8.5 MVX Firmware upgrade ............................................................................. 42
7
1 Introduction
Thank you for purchasing the ONEPROD MVX monitoring and predictive
maintenance system.
This performing system allows for the monitoring of industrial machines using 8, 16,
24 or 32 measurement channels, depending on the selected model.
ONEPROD MVX can be connected to many transducers, even though its prime
objective remains vibration measurement and analysis.
1.2 Installation
See the dedicated manual for MVX Installation:
DOC3012
Brand of ACOEM
8
MVX
Release date / MVX type CAST CSM VIO XPR
(firmware)
ONLY FOR :
5.1.0-2 3.5.4 5.0.4 5.0.4 4.6.5
MVX1027000
ONLY FOR :
MVX1027000A
MVX1027000B 5.4.0-7 3.7.3 - - 4.6.7
MVX1029000
MVX1031000
Brand of ACOEM
9
Compatibility
Maximum
Original operating
Type number of ATEX Zone 2 RAM
system
channels
(or Cl.1-Div2)
(2) (1)
MVX1001000 16 NO 256 Mo WinCE
(2) (1)
MVX1002000 32 NO 256 Mo WinCE
(2) (1)
MVX1001000A 16 YES 256 Mo WinCE
(2) (1)
MVX1002000A 32 YES 256 Mo WinCE
(1)
: Upgrade of WinCE is not supported anymore.
(2)
: Upgrading RAM is NOT possible
Brand of ACOEM
10
2 Evolutions
Brand of ACOEM
11
Illustration:
“Standard” indicators are achieved at each cycle by MVX.
Thus, a very short phenomenon can be ignored if it happens during a processing
phase.
Thus the smallest transient phenomena can be detected (even if it happens only
once and if it lasts some milliseconds), according adapted parameters (i.e. pick
detection).
Besides it is obVIOusly possible to choose another set of parameters, which erases
transient phenomena in order to avoid false alerts (i.e. RMS detection with several
seconds time constant).
As a complement to such real time monitoring, MVX can record a new “Time waves
on event” signal category (Please see next §).
The following processing parameters can be applied to define real time indicators
which are monitored on 100% of the signals.
High pass filter : 2Hz or 10Hz
Signal integration : 0 or 1
Low pass filter : 1000Hz or no filter (i.e.20kHz)
Processing : RMS, pk or pk-pk
Averaging : continuous exponential with time constant between 1s and 25s
Averaged DC level (for process and GAP signals)
BGI indicator (Blade Guard Index)
GCI indicator (Gearbox Condition Index, see next §)
Broad band and narrow band extraction on real time FFT :
o FFT 400 pts, 800 pts, 1600 pts ou 3200 pts
o FFT 1kHz, 2kHz, 5kHz, 10kHz ou 20kHz,
o FFT with 50% fixed overlapping
Brand of ACOEM
12
Real time indicators are achieved in the limit of the available MVX processing
capability.
An automatic control of the parameters embeds a checking of MVX processing load
and cautions the operator when launching if overload is predictable.
Brand of ACOEM
13
Event
Alarm threshold
Real Time Parameter
t
Time wave on event Pre-trigger
Total time wave duration
(*) For first generation MVX which have been updated with V5.xx, the following
limitations (due to internal RAM) are to be considered:
To know if your product is affected please have a look on §12
Brand of ACOEM
14
Maximal duration (in s) with pre-trigger = 0 30 40 60 120 240 320 480 480
Maximal duration (in s) with pre-trigger = full length 9 13 19 39 78 117 156 156
Brand of ACOEM
15
In addition to the new functions described in the preVIOus chapter, Version #4.1 of
the MVX firmware includes the following evolutions:
It is especially suitable for low speed machineries, where standard indicators (i.e.
filtered RMS vlalues) are not always completely pertinent
Shock FinderTM also incorporates a smooth function dedicated for decreasing false
alarm risk.
Please consult XPR300 V4.3 user manual (chapter 14.8) for more details about SFI
set up.
Brand of ACOEM
16
The main observable improvement concerns the processing times that are lower vs
preVIOus firmware versions.
Version 3.1xxx of the MVX firmware includes mainly the following new functions:
MVX is now compatible with the electric signal analysis (ESA) functions available in
ONEPROD NEST ANALYST (XPR).
The analysis of electric signals consists in acquiring and analysing the power supply
voltages and currents of a motor and detecting the distinctive characteristics thereof
that indicate anomalies.
Electric or mechanical anomalies on the motor can thus be evidenced (unbalance,
misalignment, broken bar, etc.).
MVX stored in a FIFO buffer all or part of the indicators that were worked out during
the monitoring measurements.
MVX can store up to 1000 values for each indicator, with a periodicity ranging from a
few seconds (*) to 600 seconds.
This new function is available from ONEPROD NEST ANALYST (XPR) only.
Brand of ACOEM
17
(*): for a given application, the periodicity cannot be shorter than that for the
calculation of the indicators by MVX.
Example:
When an indicator goes into alarm mode, one can know whether this indicator
increased suddenly or progressively before reaching the alarm threshold.
(*): MVX does not have a local OPC connection. Parameters called “original OPC
parameters” are retrieved by the OPC client of the XCOM component, which is
physically located on the same PC as ONEPROD NEST ANALYST (XPR)
MVX can now take into account three possible priority levels for each task.
This function allows for a better monitoring of machines with complex or random
operating conditions.
Brand of ACOEM
18
MVX keeps it operating cycle and must confirm the presence of the alarm on all
measurements collected over the defined time.
If the alarm level remains present during the defined time, MVX triggers the alarm
status: relay, MODBUS status, acquisition on status change. If the level disappears,
the alarm is no longer effective.
MVX can now, and on request from ONEPROD NEST ANALYST (XPR), calculate
phased spectra.
The phase reference that is taken into account is that of the trigger input
corresponding to the machine under study.
Brand of ACOEM
19
3 MVX Operating
3.1 Principle
Unlike most predictive maintenance system on the market, ONEPROD MVX
performs sequences of simultaneous acquisitions on all measurement channels.
Thus, MVX alternates acquisition and processing phases and a set of updated
“results” indicators is available after each processing phase.
Brand of ACOEM
20
Spectra calculated for the extraction of indicators can neither be viewed nor stored by
MVX.
In this case, all signals (spectra and/or time signals) and indicators described in the
configuration are stored by MVX in order to be integrated into the predictive
maintenance database, either at once, or later on.
Between two scheduled operations, MVX can permanently calculate scalar indicators
described in the settings and compare them to alarm thresholds, like in the
monitoring mode.
In case of status change (threshold VIOlation), all measurements (spectra and/or
time signals) and indicators described in the parameter settings are stored by MVX to
be later integrated to the predictive maintenance database.
Short-term memory:
Furthermore, and for each calculated parameter, MVX will store in a FIFO buffer the
short-term history of scalar values.
The size of this FIFO buffer can be set up to 1000 values per indicators.
The periodicity for storage in the short term memory can also be programmed and
can range from a few seconds (i.e., MVX stores all calculated values) to 600 s.
By storing 1000 values with a periodicity of 600s, the capacity of this short-term
memory can thus reach 600 000s, i.e., about 1 week.
Brand of ACOEM
21
4 Set-up
When only monitoring tasks (to the exclusion of any predictive maintenance tasks)
are to be assigned to MVX, the OneProd CSM set-up interface should be used.
This interface is used to set up MVX; it does not allow for visual display or processing
of measurement results.
To assign predictive maintenance tasks to MVX, one should use MVX jointly with the
ONEPROD NEST ANALYST (XPR) predictive maintenance software program.
This platform will perform the set up of MVX, as well as the processing of
measurement results.
In both cases, the communication between MVX and the computer hosting OneProd
CSM or ONEPROD NEST ANALYST (XPR) used an Ethernet connection: existing
network or point-to-point connection.
CAUTION
Although set-up structures are compatible, we strongly recommend that the use of
XPR and CSM never be combined to set up a MVX module.
Brand of ACOEM
22
Power:
Valid
supply voltage
Danger:
At least one channel with
violation of a danger
threshold
Alarm:
At least one channel with
violation of an alarm
threshold
Sensor:
Measurement problem
on a least one channel
(sensor, cable, etc.)
System:
Detection of an
internal system
defect
485 in
Signal being
received
on line RS-485
485 out
Signal being
emitted
On lige RS-485
A
Acquisition in progress (*)
B
Processing in progress (*)
C
Communication in progress (*)
Brand of ACOEM
23
These two connectors may be used but their pin assignment are not identical:
Addressing mode led: fixed or dynamic addressing mode netwok led indicators
Brand of ACOEM
24
Please note that some recent Ethernet switch and portable PC automatically cross or
uncross their port.
IMPORTANT - Rules for the use of Ethernet ports:
It is recommended to use preferably the port B. This port can handle the
communication with the software OneProd (XPR, CAST, MSC, VIO) and PLC
via Modbus TCP.
If the Modbus devices are on a different network then connect them on port A
When the connection is OK, a green indicator lights on the RJ45 socket (high & right
corner, see on the upper scheme)
You have to reboot the MVX after any change on the network connection
This addressing mode can be restored using the Reset button: cf. § 8.4
Brand of ACOEM
25
Brand of ACOEM
26
C:\>ping 192.168.103.240
Brand of ACOEM
27
6.4 Examples
6.4.1 Connecting one MVX with one PC in a point to point
connection through the MVX right port
MVX: dynamic IP mode
Default address: serial number
used as “Hostname”
(IP = 169.254.x.y)
Standard
MVX : B Straight cable
Ethernet port
PC, dynamic IP
mode
In this example the PC and the MVX may be let in the dynamic IP address mode.
Brand of ACOEM
28
MVX1: Fixed IP
192.168.103.240 MVX2: Fixed IP
192.168.103.241
SWITCH
PC2: Fixed IP
192.168.103.101
PC1: Fixed IP
192.168.103.100
Brand of ACOEM
29
https://serialnumberofmymvx
it is important to use https
(exemple : https://lcf042563-322/)
Login : usermvx
Password : MVX
Brand of ACOEM
30
Whether it runs with XPR, with VIO or in stand-alone mode, MVX computes and
updates as often as possible a number of scalar levels (called “overall levels” or
“indicators”).
All these levels can be consulted using the dialogue function in MODBUS format.
The status of these indicators (OK, Alarm, etc.) is also available, along with their unit.
In any event, MVX is a MODBUS slave that answers requests issued by a Master.
This Master is usually a programmable automaton, a centralized management
system or a monitoring system.
It is possible to adjust RS485 configuration (baud rate, parity …) with CAST software.
Each MVX is assigned a MODBUS slave number ranging from 1 to 32. The
MODBUS slave address of each MVX is set from the ONEPROD NEST ANALYST
(XPR) software or from the OneProd CSM software.
Brand of ACOEM
31
Conventions:
Data in MODBUS-TCP format are available through the standard 502 port.
The MODBUS slave address must be set to 255 in requests (in this case, the
IP address is used to select each MVX).
CAUTION
The slave address had to be fixed at 1 (instead of 255) in the previous versions of
MVX.
This change can result in malfunction after upgrading MVX firmware to V5.0
Adress #255 allows for a better compatibility vs MODBUS-TCP standard..
Each indicator is assigned a number (ranging from 1 to 255), which is part of the
indicator’s parameters and is entered upon programming.
For more details, refer to the user manuals of XPR or CSM.
Brand of ACOEM
32
At address 1000 (& 1001), π (PI) value “3.1416” is displayed permanently, which
allows to check the proper operating of the connection.
Indicator values are available in an address range from 1002 to 1513 and sorted
according to the table below:
1000
3.1416, set value
1001
1002
Value of indicator #1
1003
1004
Value of indicator #2
1005
...
....
...
1512
Value of indicator #256
1513
Brand of ACOEM
33
0 Normal (OK)
1 Alarm
2 Danger
Defect (transducer or other defect making measurement
3
impossible)
At address 2000, one can read hexadecimal 5555 value permanently, which allows
checking the proper operating of the transfer connection.
Logical statuses of indicators are available in an address range from 2000 to 2256
and sorted according to the table below:
MODBUS memory address Content
Brand of ACOEM
34
At address 3000, one can read hexadecimal value AAAA permanently, which allows
checking the proper operating of the transfer connection.
Indicator units are available in the address range from 3000 to 3256, and sorted
according to the table below:
MODBUS memory address Content
Brand of ACOEM
35
Alarm and Danger threshold values are available provided that the corresponding
indicators are achieved in one operating condition only.
In other words, this function is not suitable when indicators are calculated in several
operating conditions and monitored with different sets of thresholds,.
As well as indicators, data are stored in IEEE floating format over 32 bits.
Two consecutive addresses are then required for each indicator value.
There are 4 thresholds for each indicator, even if all of them are not used in all cases.
At address 4000 (& 4001), π (PI) value “3.1416” is displayed permanently, which
allows to check the proper operating of the connection.
Indicator threshold values are available in an address range from 4008 to 6055 and
sorted according to the table below:
Brand of ACOEM
36
4000
3.1416 (set value)
4001
4002
Not used
4003
4004
Not used
4005
4006
Not used
4007
4008
Value of threshold #1 of indicator #1
4009
4010
Value of threshold #2 of indicator #1
4011
4012
Value of threshold #3 of indicator #1
4013
4014
Value of threshold #4 of indicator #1
4015
4016
Value of threshold #1 of indicator #2
4017
4018
Value of threshold #2 of indicator #2
4019
4020
Value of threshold #3 of indicator #2
4021
4022
Value of threshold #4 of indicator #2
4023
4024
Value of threshold #1 of indicator #3
4025
...
....
...
6054
Value of threshold #4 of indicator #256
6055
Brand of ACOEM
37
7000
Vit Rot Equipement 1
7001
7002
DC1 Equipement 1
7003
7004
DC2 Equipement 1
7005
7006
TOR1 Equipement 1
7007
7008
TOR2 Equipement 1
7009
7010
TOR3 Equipement 1
7011
7012
Vit Rot Equipement 2
7013
7014
DC1 Equipement 2
7015
…
…
Brand of ACOEM
38
8 Maintenance
The MVX system does not require any specific operation for its maintenance, except
for dust cleaning on the ventilation inlets.
In case of a particularly dusty atmosphere, it can be useful to get rid of the dust
accumulated inside MVX.
The theoretical lifetime of the data back-up battery is several dozen years for a MVX
module under constant voltage.
Also, we recommend that the battery be replaced as soon as the MVX module has
remained totally unused for more than a year.
To: The “power” green led lights. All the other leds are “ON” too.
Brand of ACOEM
39
~To + 60 seconds: Red leds are flashing on the two RJ45 Ethernet
connectors.
~To +70 seconds: All the leds are flashing. Then A & B leds are staying lighted
for a moment.
~To + 100 seconds: Integrity relay becomes energised; A & B leds become
OFF: the MVX is ready to work.
MVX goes into an initialisation phase, during which the two LEDs A and B blink
quickly and simultaneously.
Brand of ACOEM
40
In each case it is possible to know the defect details using the “Read MVX states”
button of the CAST Toolbox.
As well as before it’s possible to know the defect details using the “Read MVX status”
button of the CAST Toolbox.
Brand of ACOEM
41
8.3.6 Summary
In case of a problem, MVX LEDs allow performing a first-level diagnosis, which
makes troubleshooting easier.
Status of integrity
Power Sensor System A B C MVX status
relay
Blinks
yes no General analogue defect Disabled
quickly
Brand of ACOEM
42
Reset :
Push button
Go to Update tab
Browse for a .firm file
Upgrade the MVX
Brand of ACOEM
43
www.acoemgroup.com
Asia
Tel. +66 (2) 7112 293 – Fax +66 (2) 7112 293
Brand of ACOEM
South America