Operating Instructions: Modular Flow Meter System Ts
Operating Instructions: Modular Flow Meter System Ts
Operating Instructions: Modular Flow Meter System Ts
GB Operating Instructions
MODULAR
FLOW METER SYSTEM TS
with and without volume preselection
2
GB
3
GB
The operator must read the operating instructions before starting the flow meter and follow these instructions
during operation.
1 . Note the temperature and pressure limits. The pressure surges produced when closing shutoff valves
(valves, slides, etc.) must not exceed the maximum operating pressure.
2 . Disconnect the mains power supply and batteries before assembling and dismantling the various modules
and the data line.
3 . Ensure that all connections and fittings are tight and free from tension.
4 . Only alkaline-manganese batteries may be used.
5 . Repairs may only be carried out by the manufacturer.
The national accident prevention regulations must be observed without fail!
When metering flammable liquids according to Directive 67/548/EEC the following must be additionally
observed:
1 . Only use explosion proof flow meter modules.
2 . Before operation, ensure equipotential bonding of the delivery system.
3 . Use conductive hoses or pipes.
4 . Only use explosion proof relay modules.
5 . In explosion hazard areas only use explosion proof plugs.
6 . Electrostatic danger due to friction - to be wiped with wet cloth only. Do not use solvents for cleaning.
7 . Change of battery only outside of hazardous location with appropriate tools. Only use approved battery
types.
4
GB
Table of Contents
A. General 8
1. Hazard 9
2. Safety information 9
3. Scope of supply 9
C. Operating unit 18
1. Proper use 18
1.1 Technical data 18
1.2 Application in explosion hazard areas 19
1.2.1 Electrical data 20
2. Installation 21
2.1 Insert the batteries 21
2.2 Turning the operating unit 21
2.3 Separate installation of the operating unit 22
2.4 Hose installation of the operating unit 23
3. Operation 24
3.1 Manual mode 24
3.1.1 Resetting the part volume meter 24
3.1.2 Displaying totalizer 24
3.1.3 Displaying flow rate 24
3.1.4 Key lock in manual mode 25
5
GB
Table of Contents
6
GB
Table of Contents
F. Repairs 58
G. Traceability 59
H. Annex 59
1. Terminal assignment of cables and connectors 59
2. Trouble shooting 60/61
Certificates of Conformity 63
Spare-part list accessories for modular flow meter system 64-67
Spare part list flow meter program 68-71
7
GB A. General
A. General
The flow meter operates according to the positive displacement principle. In a housing fitted with inlet and outlet
is a measuring chamber through which the fluid is forced to flow on its way through the housing. In the measuring
chamber, a nutating disc blocks the fluid‘s clear path. As it flows through the measuring chamber, the fluid
pushes the nutating disc to one side. It is the action of displacing the impediment which produces the wobble
movement of the disc and circular motion of the nutating disc shaft. By means of a driver, a magnet with several
poles is set into rotation, which is able to activate a magnetically switched contact through the lid of the housing.
Every revolution of the nutating disc is assigned to a constant volume and a fixed number of pulses, so permitting
the electonic evaluating system to determine the volumetric flow through the number of received pulses.
For practically-orientated application, the flow meter system is made up of various modules:
À Operating unit
Electronic evaluating unit with keypad and LCD display for operation of
the flow meter.
Á Protection cap
Protects the modules basic flow meter, relay module or mains unit when
the operating unit is decentrally installed.
 Relay module
Switches the external electrical appliances such as pump motor and
solenoid valve and is controlled by the evaluating electronics of the
operating unit.
Mains unit
Supplies power to the operating unit when battery-powered operation is
not desired and when no relay module is needed.
8
A. General GB
1. Hazard
The modular flow meter system is developed and built in compliance with the relevant fundamental health- and
safety requirements and important EC Directives. Nevertheless, this product can cause danger if it is not used
as advised.
In case of maloperation or improper use guarantee is excluded and there is danger for
• life and health of the user,
• the flow meter and other material assets of the user,
• the efficiency with the flow meter.
All persons who are dealing with the installation/assembly, starting, operation, servicing and maintenance of
the flow meter have
• to be accordingly qualified,
• to observe the local safety and accident preventive rules,
• to exactly follow these operating instructions.
It is your safety!
2. Safety information
In these operating instructions the following symbols are used:
Danger!
Refers to a direct danger.
Non-observance of the precaution will cause death or extremely serious injury.
Warning!
Defines a possibly dangerous situation.
Non-observance of the precaution can cause death or extremely serious injury.
Caution!
Defines a possibly harmful situation.
The product or something in its environment may be damaged if this instruction is not observed.
3. Scope of supply
When unpacking the goods, check the articles that:
• there is no visible damage from transporting,
• the shipped articles, including accessories, are the articles you ordered,
• no screws have loosened during transportation.
9
GB B. Basic flow meter
Warning!
The metering of non-compatible liquids damages the flow meter.
Splashing liquid or flying parts can cause injury.
Check by means of the materials indicated in the technical data and a resistance table (e. g. Lutz
resistance table) whether the basic flow meter is suitable for the delivered liquid.
Warning!
Exceeding the temperature range and the operating pressure causes damage to the basic flow meter.
Splashing liquid or flying parts can cause injury.
The ranges indicated in the technical data for temperature of liquid and operating pressure have to be
observed.
Any other use as well as any modifications of the product have to be considered as improper use. The
manufacturer will not be held responsable for any damages resulting from such improper use, the risk lies
solely with the user.
10
B. Basic flow meter GB
Type VA I, II
Diagram 1
Maximum pressure depending on
temperature of the medium T / (°C)
11
GB B. Basic flow meter
Warning!
The brief pressure peaks or surges due to water hammer can cause serious damage and destroy the
flow meter, pump or other components in the system!
Damage by uncontrolled splashing liquid. Depending on the medium concerned, this can lead to
caustic burns, explosion hazards, toxic vapours, electric shocks, danger of slipping and other dangers!
Example: A valve closing time of 20 ms can cause a pressure surge of up to 10 bar in a 10 metres long 1" pipe
at a flow rate of 50 l/min.
Diagram 2
Pressure drop depending on flow rate.
12
B. Basic flow meter GB
Q: Flow rate
Krel: Relative correction factor out of diagram 3
K1: Recent correction factor out of main memory
K2: New correction factor
Diagram 3
Correction factor depending on flow rate
13
GB B. Basic flow meter
2. Installation
2.1 Connection for
drum pump operation
Mounting steps flow meter
type ST, SL, LM and UN:
1. Mount the check valve (option)
and filter screen into the inlet
(arrow). The non-return valve
may be only installed together
with a strainer.
2. Insert the o-rings into the collar
bush. Wet the sealing surfaces
and o-rings with a suitable
lubricant.
3. Place the wing nut onto the
collar bush
4. Screw the collar bush onto the
outlet of the pump tube
5. Screw the wing nut with the flow Connection for drum pump operation type ST, SL, LM and UN
meter
6. Screw the hose fitting onto the
flow meter outlet G 1 1/4
14
B. Basic flow meter GB
15
GB B. Basic flow meter
Valve connection
16
B. Basic flow meter GB
4. Repairs
When returning the device to the supplier it is compulsary to attach the
decontamination certificate duly filled out and signed by the operator (see
service area www.lutz-pumpen.de).
C. Operating unit
1. Proper use
Caution!
Device failure by corrosion!
Use the operating unit BE1 V or Ex-BE1 V for applications in aggressive environments.
Caution!
The electromagnetic compatibility of the device is compatible with the radiated interference and
interference immunity for residential and industrial areas. To avoid interfering influences in a
decentralised installation of individual modules, only use data cables from the Lutz company and route
the data cables separate from low-voltage lines.
18
C. Operating unit GB
Danger!
Danger of explosion! Friction causes electrostatic charges!
Blast wave: Flying parts can cause death.
Operating unit to be wiped with wet cloth only.
Danger!
Danger of explosion due to inadmissible warming up!
Blast wave: Flying parts can cause death.
Following electrical data and terminal assignment must be observed.
19
GB C. Operating unit
IIB
Lo (mH)
Co (µF)
2 16
1 19
0.5 24
IIB
Lo (mH)
Co (µF)
2 6.9
1 9.9
0.5 12
20
C. Operating unit GB
2. Installation
Mounting steps:
1. Remove cap
2. Release screws
3. Carefully lift the operating unit, paying attention to cable length!
4. Turn the operating unit to the required position
5. Mount the operating unit
6. Tighten the screws
7. Mount the cap
21
GB C. Operating unit
Mounting steps:
1. Remove batteries out of the
operating unit
2. Remove the cap and release
the screws
3. Carefully lift the operating
unit, paying attention to cable
length!
4. Release the plug
5. Remove the cap from the
intermediate plate
6. Screw the intermediate plate
onto a base (screws not
included in the scope of
supply)
7. Mount the cap
8. Screw the operating unit onto
the intermediate plate and
mount the cap
9. Connect the ribbon cable to
the plug in the protection cap
10. Screw the protection cap onto
the flow meter and mount the
cap
11. Connect the operating unit
and protection cap using the
data line
12. Insert the batteries
22
C. Operating unit GB
23
GB C. Operating unit
3. Operation
+ Following described key sequences must be pressed step by step.
Operation Display
+ The total flow rate appears as long as the key SPEED is depressed.
The flow rate is displayed in the selected volume unit per minute.
24
C. Operating unit GB
Operation Display
+ The key lock appears in the display L. The key lock also remains if there is no power supply (battery).
Releasing the key lock, press the key RESET and repeat the combination above.
+ Switching from automatic to manual mode is carried out in the same way.
The display AUTO disappears.
+ Switching between manual and automatic mode is only possible if a relay module is connected.
25
GB C. Operating unit
+ Complete the input by pressing the ENTER key. To correct the filling volume, the input must have
been completed (by pressing ENTER). It is then possible to enter a new filling volume.
26
C. Operating unit GB
Operation Display
Enter filling volume e.g. 50 LITRES
Display static
Enter filling volume e.g. 30 LITRES
27
GB C. Operating unit
Operation Display
+ The correction process is only successful if the new filling volume is greater than the volume which has
already been transported.
+ The entered filling volume flashes The filling process can be started (see chapter C 3.2.3).
28
C. Operating unit GB
Operation Display
+ The cycle counter is displayed as long as the key is depressed. It cannot be displayed while filling is
in process.
Operation Display
29
GB C. Operating unit
Operation Display
Enter slow volume, e.g. 1 litre (for deleting enter 0)
30
C. Operating unit GB
Operation Display
31
GB C. Operating unit
Operation Display
+ The key lock appears in the display L. The key lock also remains if there is no power supply.
The above combination must be entered again in order to deactivate the locking function.
3.3 Calibration
In the factory the meter is calibrated for the medium water transported at a mean flow rate of 40 l/min. Various
operating conditions such as slower flow rates or the transportation of media with greater viscosity can result in
measurement inaccuracy (see chapter B 1.6).
To adjust to these operating conditions, a changeable correction factor can be used to recalibrate the flow meter
in manual mode. The correction factor is either entered directly or the electronic evaluating unit calculates the
right correction factor according to a comparative measurement between the displayed and the actual filled
volume.
The flow meter is also equipped with nine addressable memory locations (memory numbers 1-9), in which
already ascertained correction factors can be stored and copied to the main memory (memory number 0) if
required.
32
C. Operating unit GB
Read the value on the measurement vessel and enter at numerical keypad
33
GB C. Operating unit
35
GB C. Operating unit
3.3.6 Copying the correction factor from the main memory into the
addressable memory location
Operation Display
36
C. Operating unit GB
Operation Display
37
GB C. Operating unit
Operation Display
• Change into manual mode at the end of filling process (see chapter C 3.2.1)
• Have correction factor displayed from the main memory (see chapter C 3.3.3)
VM k 1
k2 =
VV
• Enter calculated correction factor into the main memory (see chapter C 3.3.2)
38
C. Operating unit GB
39
GB C. Operating unit
3.4.1.4 Conversion of the flow volume in kilograms via the correction factor
For special applications, the flow meter is able to display the measured volume in a unit of mass such as kilograms.
The correlation between volume V and mass m is ascertained using density r.
m=ρ•V
If the correction factor for volume measurement is known (see chapter C 3.3.1 or correction factor diagram 3),
this is offset with the density r. Convert the correction factor according to the below formula
Example
Operation Display
Enter the calculated correction factor k2
+ The unit of volume (litres or gallons) displayed on the flow meter is now no longer valid.
Recommendation: Cover over the displayed unit of volume using a foil indicating the valid unit of mass.
40
C. Operating unit GB
Operation Display
41
GB C. Operating unit
Danger!
When changing the batteries in explosion hazard areas danger of explosion occurs!
Blast wave: Flying parts can cause death.
Change of battery only outside of hazardous location with appropriate tools. Only following battery
types are approved:
• VARTA High Energy; No. 4901; Size LR1
• VARTA Electronic Alkaline; No. 4001; LR1; LADY; N; AM5
• DURACELL ALKALINE; SIZE N; MN9100; LR1
• Energizer Alkaline; E90; LR1; N
Mounting steps:
1. Unscrew the battery compartment lid 20
2. Remove the batteries
3. Insert new batteries, checking for correct polarity 19
4. Screw the battery compartment lid 20
5. Press key RESET
+ If the flow meter is not used for longer periods, the batteries must be removed!
42
D. Relay module RM1/RM3 and mains unit NG1 GB
The relay module RM1/RM3 or the mains unit NG1 includes a power supply
for the operating unit. The batteries only then have a buffer function in the
event of a power failure.
The relay module RM1 and RM3 are fitted with three sockets permitting
the connection of external appliances À,Á and the power supply Â.
The mains unit NG1 is fitted only with a connection for the power supply Â.
Relay module
Danger!
Live mains connection!
Electric shocks can cause death.
An operating unit or a protection cap must be mounted to the relay modules RM1, RM3 or to the
mains unit NG1 as protection against touching and contact with water.
Caution!
The electromagnetic compatibility of the device is compatible with the emitted interference and
interference immunity for residential and industrial areas. To avoid interfering influences in a
decentralised installation of individual modules, only use data cables from the Lutz company and route
the data cables separate from low-voltage lines.
43
GB D. Relay module RM1/RM3 and mains unit NG1
Danger!
Danger of explosion!
Combustion hazard. Blast wave: Flying parts can cause death.
The relay module RM1, the mains unit NG1 and the connecting cables are not suitable for use in explosion
hazard areas!
If the relay module itself is not located in an explosion hazard area the relay module RM1 230 V may
supply the power and signals to the operating unit in zone 1.
For this application the relay module must be used with an ambient temperature of -20 up to +40°C. This
temperature range is also applied to the measured liquid when the relay module is mounted on the basic
flow meter.
44
D. Relay module RM1/RM3 and mains unit NG1 GB
Outputs:
Output 1 (Pump): U = 220 - 240 V ~
(ST4 and ST5) Imax = 4A
cos ϕ > 0.7
Pmax = 880 - 960 W
45
GB D. Relay module RM1/RM3 and mains unit NG1
2. Installation
46
D. Relay module RM1/RM3 and mains unit NG1 GB
47
GB D. Relay module RM1/RM3 and mains unit NG1
Relay module RM1: Connection of pump motor, main valve and SLOW valve
48
D. Relay module RM1/RM3 and mains unit NG1 GB
Relay module RM3: Connection of pump motor (AC) and solenoid valve
Relay module RM3: Connection of pump motor (3-phase current) and solenoid valve
49
GB D. Relay module RM1/RM3 and mains unit NG1
2.3 Flow meter with separate installation of relay module/mains unit and operating unit
Danger:
Live mains connection!
Electric shocks can kill you.
Disconnect the relay
module/mains unit by
pulling the plug from the
power supply before
dismantling!
Mounting steps:
1. Remove batteries out of the
operating unit
2. Remove the cap and release
the screws
3. Carefully lift the operating unit
and relay module/mains unit,
paying attention to cable
length!
4. Loosen the connector from the
intermediate plate
5. Unscrew the intermediate plate
6. Carefully remove the
intermediate plate, paying
attention to cable length!
7. Loosen the connector from the
intermediate plate
8. Remove the cap from the
intermediate plate
9. Screw the intermediate plate
onto a base (screws not
included in the scope of
supply)
10. Mount the cap
11. Insulate the intermediate plate
using the cap
12. Screw the operating unit and
relay module/mains unit onto
the intermediate plate and
mount the cap
13. Connect the ribbon cable to the
plug in the protection cap
14. Screw the protection cap onto the
flow meter and mount the cap
15. Connect the operating unit and Separate installation of relay module/mains unit and operating unit
protection cap using the data line
16. Insert the batteries
50
D. Relay module RM1/RM3 and mains unit NG1 GB
2.4 Flow meter with relay module/mains unit and separate installation of operating unit
Danger!
Live mains connection!
Electric shocks can kill you.
Disconnect the relay
module/mains unit by
pulling the plug from the
power supply before
dismantling!
Mounting steps:
1. Remove batteries out of the
operating unit
2. Remove the cap and release
the screws
3. Carefully lift the operating unit,
paying attention to cable
length!
4. Loosen the connector from the
circuit board of the operating
unit
5. Remove the cap from the
intermediate plate
6. Screw the intermediate plate
onto a base (screws not
included in the scope of supply)
7. Mount the cap
8. Mount the operating unit on the
intermediate plate using the
short screws supplied with the
protection cap, and then mount
the cap
9. Connect up the cable between
the relay module/mains unit
and the protection cap
10. Screw the protection cap and
relay module/mains unit onto
the flow meter using the long
screws and replace the cap
11. Connect the operating unit and
protection cap using the data
line
12. Insert the batteries
51
GB D. Relay module RM1/RM3 and mains unit NG1
Danger!
Live mains connection!
Electric shocks can kill you.
Disconnect the relay module / mains unit by pulling the plug from the power supply before dismantling!
Danger!
Danger of explosion from short circuit!
Combustion hazard. Blast wave: Flying parts can kill you.
Only use the types of fuses listed below. The unit can be damaged and the explosion protection is at
risk if higher fusing ratings are used.
As the blowing of a fuse generally precedes a fault, have the unit checked by an electrical specialist before
changing a fuse.
The units are adequately protected when 5 x 20 mm IEC 127 finewire fuses are used:
RM1 230 V RM1 120 V RM3 NG1 230 V
Power supply SI3: 0.1 A semi 0.1 A semi 0.1 A semi 0.1 A semi
time-lag (mT) time-lag (mT) time-lag (mT) time-lag (mT)
Motor connection SI1: 4 A time-lag (T) 6.3 A time-lag (T) fuse protection –
by mains supply
with max. 16 A
Valve output SI2: 0.5 A semi 0.5 A semi 0.5 A semi –
time-lag (mT) time-lag (mT) time-lag (mT)
52
E. Relay module Ex-RM1m GB
53
GB E. Relay module Ex-RM1m
If the pump and the valve are protected in the connecting compartment of the relay module from short circuits,
then the design voltage for the fuse must be equal to or greater than the given voltage rating for the pump or
valve. The breaking capacity of the fuse used must be equal to or greater than the maximum short-circuit current
(normally 1500 A) that can be assumed for the location of installation. If the breaking capacity of the installed
fuse differs from this, then the value for the maximum admissible short-circuit current shall be indicated on the
relay module.
2. Installation
Caution!
The electromagnetic compatibility of the device is compatible with the emitted interference and
interference immunity for residential and industrial areas. To avoid interfering influences in a
decentralised installation of individual modules, only use data cables from the Lutz company and route
the data cables separate from low-voltage lines.
54
E. Relay module Ex-RM1m GB
2.2 Dismantling/Assembly
2.2.1 Dismantling
1. Remove the cap
2. Release the screws
3. Carefully lift off the top module À and pull off the connector Á
4. Carefully lift off the relay module à and pull off the connector Ä
5. Release the screws
6. Carefully lift off the intermediate plate Å and pull off the connector Æ
7. Take out the screws with washers from the groove in the
intermediate plate Å
2.2.2 Assembly
Caution!
Pinching the electrical leads can damage the unit.
When assembling the unit, do not route electrical connections at
points where these can be pinched.
56
E. Relay module Ex-RM1m GB
57
GB E. Relay module Ex-RM1m / F. Repairs
À Pump:
Fusing element Ex 4 A
semi time-lag
 Solenoid valve:
Fusing element Ex 500 mA
semi time-lag
Danger!
Live mains connection!
Electric shocks can kill you.
Replacing the fuse may only be carried out by qualified personnel – a trained electrician. The power
supply shall be disconnected and prevented from accidental re-connection throughout duration of the
work. As the blowing of a fuse generally precedes a fault, have the unit checked before changing the
fusing elements.
Danger!
Fire and explosion hazard due to electric current.
Combustion hazard. Blast wave: Flying parts can kill you.
Use only original fusing elements.
1. To replace the fuses, dismantle the relay module in accordance with steps 1-4 of chapter E 2.2.1
2. Disconnect the defective fusing element and remove from the top hat rail.
3. Snap a new fusing element onto the top hat rail and connect.
4. Mount the relay module in accordance with steps 11-14 of chapter E 2.2.2
F. Repairs
Repairs should only be made by the manufacturer or authorized Lutz-dealers.
Only use genuine Lutz spare parts.
58
G. Traceability / H. Annex GB
G. Traceability
Products manufactured by Lutz-Pumpen for potentially explosives atmospheres are identified by an individual
batch number which allows them to be traced. This number provides the year of construction and the design of
the equipment.
This product is an appliance for potentially explosive atmospheres. In this regard and in compliance with the
EC ATEX 94/9 Directive, provisions must be made to ensure ascending and descending traceability.
Our ATEX notified quality system ensures this traceability up to the initial point of delivery.
Except as otherwise agreed in writing, anyone that guarantees to redeliver said equipment undertakes to put in
place a system that allows for equipment that is not conform to be recalled if necessary.
H. Annex
1. Terminal assignment of cables and connectors
RM1 Ex-RM1m
1 Neutral conductor
1+2 110-230V / 50-60 Hz 2 Pump motor / SLOW-Valve
3 Not in use 3 Main valve
Earth conductor 4 Not in use
PE Earth conductor
Relay module RM 1- mains connection Relay module RM 1 - connection of external electrical appliances
59
60
Operating mode Fault Cause Remedy
General No liquid registered Correction factor 0 in the main memory Calibrate flow meter (chapter C 3.3)
GB
Measured quantity differs significantly Correction factor does not match Calibrate flow meter (chapter C 3.3)
from the actual quantity operating conditions
Wrong unit of measure set Set correct unit of measure
(chapter C 3.4.1)
Leaks in hose system Install check valve if necessary
2. Trouble shooting
Radio interference on the data line Place data line separate from
low-voltage conduits
Liquid leaks from the housing Seal defective Replace seal
Hairline cracks in housing and/or on Check installation conditions for
screw connection due to mechanical mechanical stresses and correct them
influences
Screw connection not tightened Check that screw connections are
securely secure
Operating pressure exceeded, pressure Check pressure conditions
surges e.g. due to water hammer
Inadequate chemical resistance Contact manufacturer
Throughput decreases Filter screen soiled Remove and clean filter screen
Measuring chamber soiled Flow meter must be flushed and cleaned
regularly after handling aggressive,
tacky, crystallizing or very
contaminating liquids
Manual mode „Batt“ indication flashes or remains End of battery service life Replace batteries (chapter C 4.1)
permanently on
No indication after changing batteries Batteries fitted incorrectly Open battery compartment and check
polarity
Battery contacts corroded Clean contacts
Automatic mode No liquid registered Dry running Prevent air pockets in connecting lines,
Filling process interrupted by flow meter use a solid hose system
„STOP“ appears on display Solenoid valve does not open Check solenoid valve
Time required to fill the connecting lines Increase monitoring time
exceeds the set monitoring time (chapter C 3.2.10)
Operating mode Fault Cause Remedy
Automatic mode Swash plate blocked Clean flow meter, check resistance
61
GB
62
GB
Lutz - Pumpen GmbH
Erlenstraße 5-7
D-97877 Wertheim
Declaration of Conformity
Herewith we declare that the below-mentioned machine in it‘s conception and design and in the execution
marketed by us fully complies with the EC directives.
This declaration ceases to be valid if the machine is modified in any way without prior consultation with us.
Wertheim, 08.04.2008
64
GB
65
GB
When ordering spare parts always indicate the corresponding order No. and production No. or order No. of the unit to
be repaired.
66
GB
67
GB
68
GB
69
GB
70
GB
71
GB