Location via proxy:   [ UP ]  
[Report a bug]   [Manage cookies]                

Micro GT Heli-Deck User Manual: Findlay Irvine LTD

Download as pdf or txt
Download as pdf or txt
You are on page 1of 28

Micro GT Heli-deck

User Manual

Findlay Irvine Ltd


381-58 Issue 1
Introduction ................................................................... 1
Overview of the micro GripTester .............................................. 1
Theory of Operation ................................................................. 1
Major Components................................................................... 1
Equipment Supplied ................................................................. 2
micro GripTester Features .............................................. 3
micro GripTester Assemble ...................................................... 3
Carrying the micro GripTester once assembled .......................... 3
Surveys and Runs ................................................................... 3
Units of measurement .............................................................. 3
Wet/Dry Testing ....................................................................... 3
Survey Test Speed .................................................................. 3
Touch-screen LCD ................................................................... 4
GPS ........................................................................................ 4
Buzzer .................................................................................... 4
Setting the Time and Date ........................................................ 4
Storing Run Data ..................................................................... 4
Transferring Data to a PC ......................................................... 5
Analysing and Reports ............................................................. 5
Software Menus ............................................................. 6
Menus ..................................................................................... 6
Start Survey Menu ................................................................... 6
File Management Menu ............................................................ 7
Check Calibration .................................................................... 7
System Menu .......................................................................... 7
Quick Run Menu ...................................................................... 7
Help Menu............................................................................... 7
Preparing for a Survey ................................................... 8
Assemble Instructions .............................................................. 8
Lifting and Carrying ................................................................. 12
Pushing Handle ...................................................................... 12
Battery ................................................................................... 12
Checking Tyre Pressures ........................................................ 12
Switching On and Off .............................................................. 12
Setting the Software Defaults .................................................. 13
Checking the Zero .................................................................. 13
Creating Heli-deck Survey Files ....................................14
Guide to survey file creation and management ......................... 14
Preparing the Heli-deck ........................................................... 16
Run Error/Warning messages .................................................. 16
No Event Marker ..................................................................... 16
Centre-Line Warning ............................................................... 17
Performing the Heli-deck Survey ..................................17
Position of the Pushing Handle ................................................ 17
Scrub the Measuring Wheel..................................................... 17
Starting the Heli-deck Survey................................................... 17
Maintenance and Storage .............................................19
Maintenance ........................................................................... 19
Annual Calibration .................................................................. 19
Storage .................................................................................. 19
Heli-Deck Grip Tester Analyser .....................................20
Overview ................................................................................ 20
Installation .............................................................................. 20
Survey Folder ......................................................................... 21
Copying Survey Files from Grip Tester ..................................... 22
Survey Analysis ...................................................................... 22
Starting the Analysis ............................................................... 24
Options .................................................................................. 25
What Could Go Wrong? .......................................................... 26
Contact Us...................................................................26
Contact Details ....................................................................... 26
Introduction

Overview of the micro GripTester

The micro GripTester is a three-wheel manually pushed device which measures friction by the braked wheel,
fixed slip principle. The device is known as a "Continuous" friction measuring equipment (CFME) as it
measures continuously along the test path, rather than just taking a single spot reading.

The device carries its own water supply and has an inbuilt water control system to maintain a known water
film thickness where testing of the surface in wet condition is required. Its single measuring wheel, fitted with
a special smooth tread tyre made to an American Society for Testing and Materials (ASTM) standard, is
mounted on an instrumented axle.

As the device is pushed over the surface to be surveyed, the dynamic friction reading is automatically
calculated and stored. As well as friction, other parameters such as speed, distance travelled, water flow,
GPS (Global Positioning Satellite) readings and ambient temperature are also taken and recorded.

Theory of Operation

The majority of the micro GripTester’s weight is distributed over two drive wheels, which are fitted with
patterned tread tyres and mounted on a drive axle. The drive axle carries a sprocket; the stub axle, on which
the measuring wheel is mounted, carries a second sprocket with a different number of teeth than the drive
axle sprocket. The two axles are linked by a transmission chain; this transmission system continuously
brakes the measuring wheel, forcing it to slip.

This slipping wheel and the weight of the micro GripTester cause small bending movements in the stub axle
which are measured by strain gauges. The signals from these strain gauges are processed by an onboard
computer and stored either on an internal disk or a USB memory stick when a heli-deck survey is carried out.

The computer also receives signals from a proximity sensor activated by a wheel with 16 magnets mounted
on the drive axle. This allows distance and hence speed to be calculated.

Major Components

The major components of the micro GripTester are as follows:

two drive wheels, connected by a drive axle,


a measuring wheel mounted on an instrumented stub-axle and connected by a drive chain to the drive axle,
a Signal Processing Unit (SPU) mounted on the drive axle,
a User Interface unit containing a touch-screen LCD, On/Off and Start/Stop switches,
a Camel Pack Watering System to deliver water to the measuring wheel,
a Pushing Handle that can be stowed on top of the micro Griptester.

1
Equipment Supplied

The following equipment comprises the micro GripTester delivery package:

micro GripTester
Protective Carrying bag
Transportation Crate
Two unprotective Carry bags (for hand carry transportation only)
Battery Charger
Camel Pack Watering System
Tyre Pressure Gauge
USB Flash Memory Pen with User Manual and GripTester Run Viewer software installed
USB Cable for connection of micro GripTester to PC or Laptop
Two LCD Stylus Pens
LCD Cleaning Kit
Calibration Certificate

2
micro GripTester Features

micro GripTester Assembly

The micro GripTester is easily assembled/disassembled into two smaller lighter packages of approximately
12Kg each for transportation by hand to and from the heli-decks. The design is such that no hand tools are
required to assemble the micro GripTester when on-site.

Carrying the micro GripTester once assembled

When the micro GripTester folding handle is in the stowed position, the device may be lifted and carried by
the two cross-pieces on the pushing handle, one in each hand. Follow correct manual handling procedures
when lifting or carrying the unit taking the weight and lifting position into account.

Surveys and Runs

The micro GripTester software organises skid resistance surveys into Surveys and Runs.

A heli-deck survey will consist of two survey files and file extensions A and B will be added automatically by
the software when initially creating the survey. Each survey file will contain all the runs made in either
direction A (perpendicular to the Chevron) or Direction B (parallel to the Chevron).

A facility known as a "Quick-Run" exists to allow the user to carry out a run without having to first set up a
Survey and Run. This is not normally used for heli-deck testing, but can be used to check the GN numbers
prior to starting the survey and ensure the tyre has been scrubbed and wet before the first run.

Units of measurement

The micro GripTester measures friction in GripNumber (GN). In heli-deck mode GripNumbers equal friction
(u) GripNumbers are widely used in road and airport surface friction measurements and standards are
defined laying down minimum acceptable readings for different surface types. Correlation with other
measurement techniques is available.

Please contact Findlay Irvine Ltd. for further details.

Wet/Dry Testing

Heli-deck surveys must be conducted on a wet surface. Where wet surveys are required, the on-board water
control system allows a controlled depth of water to be delivered onto the measuring wheel. The standard
water film depth is 0.25mm. Other film depths may be selected if required. Heli-deck surveys should not be
conducted on a dry surface as this will show artificially high friction readings.

Survey Test Speed

Surveying is carried out at a slow walking pace. The measurements of friction are not particularly dependant
on speed and the User may set the target test speed for each Survey, however the normal speed of 0.4
meters per second is recommended for consistency unless other test speeds are required for specific
applications.

During a test run, a speedometer type dial will show the actual speed and a bleeper will indicate over-speed
or under-speed conditions.

3
Touch-screen LCD

The micro GripTester uses a daylight readable touch sensitive colour liquid crystal display (LCD) for display
and User input.

Two stylus pens are supplied in the compartments either side of the User Interface box. Simply tap the end
of the stylus on the LCD to make an entry.

A small keyboard will be displayed when numerical or text entry is required. Use the stylus to select the
required numbers or text.

Do not use any implement other than a stylus on the LCD as it may mark or damage the touchscreen.

GPS

A GPS (Global Positioning Satellite) receiver is built into the micro GripTester. The latitude and longitude are
displayed during a test run and also stored into the run data file along with altitude and various other GPS
parameters.

When the micro GripTester is switched on it may take several minutes for the GPS to obtain a position fix.

The GPS is also used to set the real time clock in the micro GripTester. As the GPS transmits UTC time
(which is the same as Greenwich Mean Time), the User may set (in the System parameters menu) the time
offset from UTC for the time zone in which the survey is being conducted.

Buzzer

The micro GripTester incorporates a buzzer that is used to indicate various states.

During a Run, a continuous tone indicates that the micro GripTester is being pushed too fast, at above the
preset test speed. An intermittent tone indicates that the micro GripTester is being pushed too slow, at
below the preset test speed.

The buzzer may also indicate fault conditions. A continuous bleeping when not conduction a Run indicates a
communications problem between the SPU and the User Interface.

Setting the Time and Date

Under normal circumstances, the GPS receiver automatically sets the real time clock in the micro GripTester.
As the GPS transmits UTC, the User may set (in the System parameters menu) the time offset from UTC for
the time zone in which the survey is being conducted.

If the GPS receiver does not have adequate reception, or if it is not fitted or defective, the User may set the
internal real time clock using the System menu.

When a valid GPS connection is restored, the real time clock will be reset to the GPS time (taking into
account any time zone offset).

Storing Run Data

When a heli-deck survey is complete, the data will be stored onto either the internal disk or a USB memory
pen (supplied) which may be plugged into the underside of the User Interface box.

The option of where the data will be stored (internal disk external USB memory pen) is selected by the User
in the System menu. Data and other files may be copied from the internal disk to external USB using the File
Management menu.

4
Transferring Data to a PC

Data may be transferred from micro GripTester to a laptop or PC either via a USB Flash Memory pen, or by
using ActiveSync, which is a program developed by Microsoft for use with the Windows products to transfer
information between a desktop PC and mobile devices such as a handheld computer, mobile phone etc.

When conducting a run, the User may set whether data is stored to the internal disk or the external USB
memory pen. If data is stored to the USB memory pen, the pen may be removed from the micro GripTester
and inserted in the PC and the file copied to the PC hard disk. If the run data file was stored to the internal
disk, copy it to the USB memory pen using the File Management menu then proceed as above.

Please refer to the chapter Copying data to a PC for further information.

Analysing and Reports

The small screen on the micro GripTester allows only an overview of run data. For most analysis purposes
the data would be transferred to a laptop or PC. An analysis software package (Helideck Griptester
Analyser) is provided on the USB memory pen supplied with the micro GripTester.

The Helideck GripTester Analyser software may be installed (licence free) on any laptop or PC. Instructions
are included. The software allows both survey files and associated data to be displayed and printed either
individually or as an average of the two surveys.

5
Software Menus

Menus

When the micro GripTester is switched on a "splash-screen" will be displayed for a short time and then the
display will show the main Menu screen. Behind each menu page are further screens

Many of the screens show time and date and also the Battery status indicator in the top right corner of the
LCD.

There are six main menus as follows:

Start Survey
File Management
Check Calibration
Quick Run
System
Help

Start Survey Menu

The Start Survey menu allows the User to setup a new Survey or select an existing Survey.

If New Survey is selected, use the displayed keyboard to enter a Survey name. Two Folders with file
extentions A and B are created with the same name and Run data will be stored as sub-folders in the Survey
folders. The Folders will be created on either the internal disk, or the external USB memory pen, whichever
is selected as default in the System menu (see later).

Also enter an Operator name and a Survey description. By tapping on More further parameters such as
Notes, Weather, test speed, water film depth, run length and surface type may be set.

Note that until these fields are set, it will not be possible to Save the Survey.

If Existing Survey is selected, the disk Folder structure is displayed. Surveys are stored in the Survey folder.
Double tap the Survey folder and select the Survey from the Folder list displayed.

After setting up a Survey or selecting an Existing Survey, tap Setup Run to setup the first run of the Survey.
Now enter the Run details, comprising Run Name (which is also the Folder name into which run data will be
stored) and Run Notes. Now tap Start Run.

The screen now displayed is the "pre-run" screen. The speedometer displays push speed with the centre of
the dial the target speed. GPS data, ambient temperature and distance are shown along with a graph of
GripNumber versus distance travelled.

As the run start point is passed, the Start/Stop switch is pressed or the Log button tapped. The GripNumber
graph will reset back to the start and the Distance will reset either to the run length if a run length was set, or
zero is no run length was set.

When the set Run distance is complete, or when the Start/Stop switch is pressed or Stop button tapped, a
Run Review screen will be displayed after a short delay. This screen shows a summary of the run data,
including the GripNumber average, maximum, minimum and standard deviation. At this point it is possible to
add further Run Notes if required.

Abandon or Save the Run data as required, and the screen will prompt for a choice of either another Run
within the same Survey or End Survey.

A detailed explanation of how to carry out a Heli-deck survey can be found in the ‘Carrying out a survey’
Section of this manual.

6
File Management Menu

The File Management menu allows the User to see the folders and files on the internal disk and external
USB memory pen (if fitted), and to copy files to or from these two disks.

All Surveys are stored onto the Surveys folder as described in the Chapter ‘Data Folder Structure’. Runs are
stored in individual folders within the Survey folder. To copy from internal disk to USB flash memory or vice
versa, select the folder of file to be copied, tap Copy then select the destination folder then tap Paste.

Check Calibration

The Check Calibration menu allows the User to verify that the measurement system is working satisfactorily
and that the "zero" measurement is accurate. To carry out the check, the measuring wheel is lifted just off
the ground by tilting the pushing handle downwards. Now very gently drop the wheel back onto the ground
without moving the micro GripTester backwards or forwards.

The measured GripNumber is displayed. Readings of 0.00, 0.01 or 0.02 are acceptable. If the reading is
above 0.03 this may indicate a problem and the micro GripTester may require calibration.

System Menu

The System Menu allows the User to:

Setup default Survey parameters


View System settings
View Device constants

The Survey parameters that may be configured are:

Test Speed (normal test speed is 0.4 metres per second)


Water depth (normal depth is 0.25mm for a "wet" survey)
Run Length (to suit application, normally zero for Helideck Surveys)
Log Data to (external USB memory pen or internal Disk)

System settings comprise:

Units of measurement (default GripNumber (GN))


Language
Time Zone offset from UTC

Device Constants comprise:

Factory calibration Date (factory set)


Software Revision
Windows CE Revision

Further options are available; however these are for factory use only and are password protected.

Quick Run Menu

The Quick Run menu is used when you require to carry out a quick check to verify that the equipment is
operating correctly or to quickly test a particular surface without first having to set up a Survey and Run.

When carrying out a Quick Run, the GripNumber, graph and associated information is displayed on the same
style page as when carrying out a normal Run as part of a Survey, however Default run parameters (test
speed, water depth etc.) are used.

After completing a Quick Run you have the option to save the data.

Help Menu

The Help menu is the text that you are reading. As well as Chapter titles and Topic pages, index and search
facilities are provided.

7
Preparing for a Survey
Assemble Instructions

The micro GripTester can be disassembled into two separate packages, each weighing less than 12Kg for
hand carrying purposes to and from the heli-deck to be surveyed, furthermore the micro GripTester can be
assembled/disassembled without the use of Hand tools.

The following is a step by step guide to assembling the micro GripTester.

1. The micro GripTester can be transported in two separate bags (See Fig 1 & 2) reducing the weight
to below 12Kg in each. One bag contains the Main GT Body; the second bag will contain the Handle,
wheels, Camel pack and accessories bag, containing Battery charger, tyre pressure gauge, UBB
Pen, USB cable, and LCD Cleaning kit.

Fig 1: Main GT Body Fig 2: Handle, Wheels and Accessories

2. First Remove the GT body from the bag then align and fit both Drive wheels, these are hand
tightened with a single nut as shown in Fig 3 and 4.

Fig 3: Drive Wheels Fig 4: Hand tighten the nut.

8
3. The next step, is to fit the Measuring Wheel, this is secured in place by had tightening the three
spring loaded nuts as shown in Fig 5.

Fig 5: Drive wheel fitted with 3 nuts.

4. When the micro GripTester is not being used, being transported or when carrying out routine
maintenance, the chain locking mechanism should be engaged, this is a safety feature, to prevent
accidental damage to the user or indeed the micro Grip Tester. Figure 6, shows the chain locking
mechanism, with Red indicating the chain is locked.

Fig 6: Chain locking mechanism

5. The next step in the assemble procedure is to fit the handle, there are 4 locking pins, to be fitted at
this stage and care must be taken when aligning the handle with the main body. Ensure the locking
pins are fitted in the correct positions and orientation as shown in Figs 7, 8 and 9

Fig 7: Locking pin fitted lower hole on Outside Fig 8: Locking pin fitted to upper hole on Inside

9
6. Once the micro GripTester handle has been secured to the main GT body correctly using the 4
locking pins, connect the power cable as shown in Fig 9. It is important to line up the two white
markers, before pushing home the cable and tightening.

Fig 9: Line up the two markers before fitting.

7. The push handles, which are positioned flat against the main GT handle for transportation, can now
be locked into position as shown.

Fig 10: Push handles for transportation Fig 11: Push handles in survey position

8. The micro GripTester display screed can be pivoted on its axis, to aid viewing, fig 12, also shows
where the USB pen and cables are connected.

Fig 12: Screen and USB connection

10
9. The Camel pack is filled with water prior to the survey, worn as shown in figure 14 and connected to
the micro GripTester main body in preparation for survey.

Fig 13: Fill the camel pack with water. Fig 14: Shows how the camel pack is worn.

10. The water hose is connected to the micro GripTester as shown, it is important to make sure the hose
is not kinked in any way when conducting the survey as this can restrict water flow significantly as
seen in figure 16.

Fig 15: Showing hose attachment. Fig 16: Water hose, incorrectly connected.

11. The micro GripTester is now fully assembled and ready to be used for heli-deck surveying, the chain
locking mechanism can now be disengaged, by rotating the switch as shown.

Fig 17: Locking mechanism, rotated to show Green.

11
Lifting and Carrying

When the micro GripTester folding handle is in the stowed position, the device may be carried using the two
handle cross-pieces. Place one hand on each cross-piece and follow correct manual handling procedures
when lifting or carrying the unit.

The handle may be unlocked from the stowed position and moved to the pushing position.

Pushing Handle

Before using the micro GripTester the handle must be set in pushing position. To unfold the handle,
withdraw the two pins on the handle hinges by pressing the pips on the pin-ends and pulling the pins out.
Rotate the handle on the hinge and re-insert the pins with the handle in the pushing position.

When finished surveying, reverse the procedure to move the pushing handle back to the stowed position.
The micro GripTester may then be lifted using the two cross-arms on the pushing handle. The weight will all
be taken on the hinge pins - the pins are chosen so that this is perfectly acceptable.

To aid manoeuvring of the micro GripTester, two jockey wheels are fitted on the base of the pushing handle.
When the pushing handle is pushed downwards, the weight of the device will be taken on the jockey wheels
and the micro GripTester may be rotated and re-positioned.

Battery

A 12V sealed lead-acid battery is included in the Signal Processing


Unit. The battery is charging using the charger supplied which is plugged into the Signal Processing Unit
(SPU). Charging the battery from flat takes about 4 hours. A fully charged battery will provide around 6
hours of continuous use.

The battery voltage is displayed on the System menu screen. In addition, the battery status is displayed as a
coloured box on the top-right hand corner of the LCD.

The colour of the icon will change as the battery voltage falls as follows. When the indicator turns red,
testing should be suspended until the battery has been charged.

Green - Fully charged


Blue
Yellow
Red - Low battery

If the micro GripTester is left idle for more than 30 minutes, it will switch itself off. Follow the
Normal procedure to switch back on.

Checking Tyre Pressures

A pump to inflate the tyres is included with the micro GripTester.

NOTE! Due to the small size of the tyres it is dangerous to use a compressor to inflate the tyres.

Before conducting a survey, check that the pressures of the drive tyres and measuring tyre is as follows:

Drive tyres 10 psi / 0.67bar


Measuring tyre 5 psi / 0.33 bar

Switching On and Off

On the User Interface box, the switch marked I/O is used to switch the micro GripTester on and off. To
switch on, press and hold down the switch for about one second.

The micro GripTester will switch on and after a few seconds the LCD will display a splash-screen. It will then

12
run the application and display the main software menu.

To switch the micro GripTester off, hold down the I/O switch for up to 6 seconds. Release the switch and
the device will switch off.

Setting the Software Defaults

There are a number of options that may be set by the User when carrying out surveys. Many of these will
not change from survey to survey and so may be set as Defaults so that the values are already set on
Survey and Run setup screens.

To set the defaults, tap the System menu.

The parameters that may have defaults set are:


Test Speed
Water Depth
Run length
Log Data to (Internal/USB)

You may also set the following:


Units
Time Zone
Language

Checking the Zero

Before undertaking micro GripTester surveys, it is good practice to check that the micro GripTester correctly
reads zero GripNumber when no drag is applied to the measuring wheel. If the zero has shifted very slightly
(between +0.05 and -0.05) the User can reset the zero. Larger shifts in zero indicate damage to the
measuring axle and the micro GripTester should be returned to the manufacturer for inspection.

To check the Zero, tap the ‘Check Zero’ menu option. Now press down on the pushing handle so that the
measuring wheel comes off the ground, then very gently drop the wheel back onto the ground without
moving the micro GripTester backwards or forwards, whilst keeping the handle raised. If the display reads
above 0.03, the unit may require re-calibration, if this is the case contact the manufacturers.

13
Creating Heli-deck Survey Files
Guide to survey file creation and management

1. From the microGripTester main menu, select ‘Start Survey’, then as shown below:

Fig: 1a Fig: 1b

2. The next stage is to enter details about the new survey you are creating, giving the survey a name,
entering the operators name and a short survey description are all possible on this screen.

Note: it is also important to check the Transverse box, if you intend doing a survey of a helideck,
aluminium or similarly ridged deck. As shown in Fig 2b.

Fig 2a. Fig 2b. Ridged Aluminium deck

3. Upon completion of this page, select the ‘More’ icon, this will take you to the screen shown below
where you can add notes, weather conditions and more importantly set the speed and water depths
for the survey.
Please Note: The water depth should always be set to 0.25mm and it is recommended the survey
speed be set to 0.4m/s as shown here:

Fig 3

On this screen, the user can also comment on the surface condition and select whether the data
should be logged to the internal storage card or USB memory stick.
Once the user has completed the data on this page, select ‘Back’ to return to the new survey file
main page.

14
4. The ‘Save Survey’ icon, which had previously been greyed out, can now be selected and the user is
presented with the following dialogue screen:

Fig 4

Essentially, the user is being asked in which direction he intends to survey first:

Yes, DIRECTION A – Across the helideck, parallel to the Chevron.


No, DIRECTION B – Across the helideck, perpendicular to the Chevron.

At this point two survey files will be generated with the Survey name entered previously, with
extension A or B, which is related to the direction of travel on deck and stored either on in the
storage card or memory stick for future use.

5. The user is now ready to begin ‘Run01’ of the survey of the helideck, in the direction chosen
previously. Assuming the user selected Yes, the first survey will take place in direction A, parallel to
the Chevron and the user will see a screen similar to the one shown below:

Fig 5

6. Pressing the ‘Start Run’ icon brings up the following dialogue box, which contains detailed
information on where to start the first run of the survey, pressing X brings up the a diagram with
further information on how to perform the survey in direction A. Select ‘Back’ and ‘OK’ when you are
ready to begin the survey:

Fig 6a Fig 6b

15
Preparing the Heli-deck

As previously mentioned the full heli-deck survey is performed in two directions, Direction A – Parallel to the
Chevron and Direction B – Perpendicular to the Chevron.
However there are several factors to consider when performing the survey regardless of the direction.

Heli-deck Map with Direction A indicated.

In order to map the heli-deck accurately the following must be done:


1. The heli-deck should be mapped in grid like fashion and the runs spaced evenly apart at a distance
on 1 metre between runs and the heli-deck should be marked up accordingly using chalked
lines or equivalent.
2. Each run will start where the previous run ended, in boustrophedonic fashion, as shown by the light
blue line for Direction A and brown lines for Direction B in the above diagram.
Note: If a run has to be abandoned, the user will have the choice from which direction to repeat the
run.
3. The ‘Event Marker’ button should be pressed each time the user traverses the centre line in direction
of travel, as indicated by the X in the above diagram.
4. When carrying out the run on a centre line, the ‘Event Marker’ should be pressed three times, once
as you enter the landing circle, once in the centre and once as you exit the landing circle as shown
above.

Run Error/Warning messages

The software has been designed to warn the user if there have been too few or too many event markers
during the run. This is a warning and it gives the user the opportunity to repeat the run if a mistake has been
made.

No Event Marker
If the user completes a run and forgot to press the event marker on the centre line, the following
screen will be displayed:

Fig 7a Fig 7b

Select ‘OK’, the user has the option to repeat the run in the same direction as just completed or in
the opposite direction by selecting ‘Yes’ or ‘No’ from the screen shown in Fig 7b.

16
Centre-Line Warning

When surveying the centre-line the user is expected to press the event marker 3 times, upon
entering the landing circle, in the centre and on leaving the landing circle. The following confirmation
message is displayed:

Performing the Heli-deck Survey


Position of the Pushing Handle

If the pushing handle is in the "stowed" position, remove the two locking pins, rotate the handle to the
pushing position and re-insert the locking pins.

Note: It is very important that when pushing the micro GripTester, the pushing handle is neither fully up nor
fully down. It must be free to move both upwards and downwards or the load on the measuring wheel will be
artificially altered.

Grasp the two rubber hand grips then raise the pushing handle to a comfortable pushing height.

Scrub the Measuring Wheel

It is important to ensure the measuring wheel is clean and free of any contamination prior to use, therefore if
is essential the micro Griptester be ‘scrubbed’ in prior to the survey. To do this use the ‘Quick Run’ option
and carry out two or three wet runs of approximately 20 metres in length.

Starting the Heli-deck Survey

1. From the position identified previously, the user is now ready to start the first run of the survey in
direction A and the following screen will be displayed:

Start the run by selecting the ‘Start/Stop’ button on the Right hand side of the display console. Press
the ‘Event Marker’ button on the Left hand side of the console, when crossing the centre line, and
then press the ‘Start/Stop’ button again, to end the first run.

17
2. At the end of the run the following screen will be displayed, assuming there have been no problems
with the run, this screen shows the Speed, water and Grip number in graphical form over the length
of the run.

Select ‘Save Run’ and then ‘Another Run’ to continue with the survey. The user will now be ready to
start ‘Run02’ of the survey.
3. Run 02 should now be performed in the opposite direction to the first run, one metre apart. Pressing
the ‘Event Marker’ button as the centre line is traversed.
4. The survey is continued in this fashion until the all the runs for Direction A are complete, with the
exception of the central run, which as described above require the user to press the ‘Event Button’
three times.
5. On conclusion of all the runs in Direction A, Select the ‘End Survey’ Button. This will take you back to
the Main Menu.
6. Once the user is ready to Survey the heli-deck in the opposite direction, Direction B, from the main
menu, select ‘Start Survey’ ‘Existing Survey’ and select the appropriate file as shown below and
select ‘Load’ followed by ‘Set-up Run’:

7. This time the Survey will be perpendicular to the chevron and the user will see the following
instructions when preparing for the first run, as before, pressing X shows a detailed map:

8. Upon completion of the survey, select the ‘End Survey’ icon, which will take the user back to the
main menu. From the main menu, the user can navigate to the survey files by selecting the ‘File
Management’ icon.

18
9. Files listed on the left are survey files stored on the internal storage card, files listed on the right are
survey files stored on the external USB stick. As can be seen below, the survey files are stored by
date under the particular survey conducted.

10. The ‘Findlay Irvine Helideck Grip Tester Analyser’ software should be used to interpret the results of
the survey. This software is supplied separately and runs on Windows.

Maintenance and Storage

Maintenance

The design of the micro GripTester is simple and modular and individual parts which are worn or damaged
can easily be replaced.

After use it is particularly important to remove any contamination by salt or runway de-icing chemicals
(particularly acetates and formates) as soon as possible. If these are allowed to dry on the measuring tyre,
it may be necessary to scrub the tyre clean to prevent the deposits from altering the readings.

The micro GripTester’s transmission chain should be oiled with a low viscosity chain oil (taking care not to
get oil on the tyres) at regular intervals.

The touchscreen LCD may be gently cleaned as required using the LCD cleaning kit supplied.

Annual Calibration

All measuring instruments should be calibrated at regular intervals and the micro GripTester is no different.
As well as calibrating the GripTester against standard devices and issuing a calibration certificate, the micro
GripTester is fully serviced and components with fixed lifetimes, such as battery, tyres etc. are replaced.

When a service is due, a pop-up message will be displayed when the micro GripTester is switched on.

Always ensure that the micro GripTester is adequately protected when returning it for maintenance or
calibration. It should be placed in the carry bag provided and placed in the padded transportation crate.

To arrange for an annual calibration and service, please contact Findlay Irvine at sales@findlayirvine.com or
telephone +44 1968 671200 or fax +44 1968 671237.

Storage

When the micro GripTester is being stored, it can either be left sitting in its normal position with all three
wheels on the ground, or it can be carefully tilted back to rest on the lower pushing handle with the wheels in
the air.

To prevent accidental damage, the micro GripTester should be stored in the carry-bag supplied whenever it
is not being used.

19
The battery should not be left discharged for long periods as this can reduce battery life. If the micro
GripTester is not used for extended periods then charging the battery say once every one or two months will
extend the battery life.

Heli-Deck Grip Tester Analyser


Overview

The Heli-deck GripTester Run Viewer is a software application for use on a laptop or personal computer that
provides an easy way to view graphs of a surveyed Heli-decks either in one direction or both directions
combined.

Installation

The Helideck Grip Tester analysis software is provided as a standard Windows .msi installation file. Copy the
file to your Windows PC, click on the file, follow the on-screen instructions and accept the default settings –

At this point, Windows will pop up a ‘User Access’ control box which will ask you to confirm that you want to
install the software.

20
The software has now been installed and an icon has been placed on your desktop –

Click on the icon to start using the software.

Survey Folder

Before you start using the software, you need to install the software as described in section, ‘Installation’

You then need to create a survey directory on your PC which will contain the helideck survey files from the
Grip Tester.

The name and location of the directory is entirely up to you but we recommend you name it something like
Helideck Surveys and create it in your user directory.

You should then create a sub-directory for each individual helideck as shown below:

21
Copying Survey Files from Grip Tester

The Grip Tester automatically creates survey files each time you perform a survey and these files need to be
copied to the survey sub- folder you created above.

Note: the Grip Tester can be configured to save the files to either an internal Storage Card or directly to an
external USB Stick. It’s probably better to configure the Grip Tester to save surveys to the Storage Card but
you’ll then need to use the Grip Tester ‘File Management’ menu to copy the files to a USB stick.

Once you have the survey files on a USB Stick, simply drag them to the appropriate sub-folder :

Survey Analysis

To analyse a survey, start the analysis software by double-clicking on the icon which was placed on the
desktop during the installation process:

The software will then display the initial screen, where the version number is shown in the corner:

22
You then need to configure the software to use the survey folder you created earlier.

Click on the Settings button, navigate to the survey folder and click OK.

Note: you must select the ‘root’ folder (and not one of the helideck sub-folders):

Now click on the Save button to save the folder path for future analyses:

The available surveys should now be visible in the Helideck dropdown box:

Click on the helideck you want to analyse and then the appropriate date from the Date dropdown box.

23
The software will then display the names of the survey files for each of Direction A and Direction B and will
automatically include each direction in subsequent analysis:

On occasion, the Direction A and Direction B part of the surveys may have been performed on different
dates.
If this is the case, the analysis software will only automatically detect one of the directions.

You will need to use the ‘Browse’ button next to the appropriate direction to navigate to and select the
missing direction.

Starting the Analysis

To start the analysis, click on the Analyse Data button:

The analysis page shows a grid representation of the helideck (aligned with the helideck name) comprising
average grip number readings averaged over one square metre.

Each square on the grid is colour coded to make it easy to see the areas which comply with CAA
requirements.

In addition, several other metrics are displayed including the deck width and length as determined by the
Grip Tester.

The Print button allows you to either obtain a printed copy of the results page or to save it as an image to file.

24
Options

Normally, you should use the data gathered in both Direction A and Direction B in the analysis.

However, you can elect to analyse in only one of the directions by ‘unchecking’ the appropriate Direction
checkbox.

It’s also possible to change the grip number ranges associated with colour coding but this should only be
used for temporary analysis and also requires entry of the administrator username and password.

Hover your mouse over the boundary between the ranges you want to change and then click and drag
accordingly. The then need to select the Accept button and enter the administrator credentials –

You can revert to the CAA default ranges by selecting the ‘Reset to CAA defaults’ button. Note that the
administrator credentials are not required.

25
What Could Go Wrong?
The analysis software performs validity checks on the run data and may pop-up warning messages if it
detects problems. If this occurs, you may have to perform another Grip Tester survey. However, the software
will allow you to continue and analyse the existing data. Typical error messages are:

Contact Us

Contact Details

For further information on any aspect of the micro GripTester, Viewer software or to arrange a calibration or
service, please contact Findlay Irvine by post, email, telephone or fax as follows:

Findlay Irvine Ltd.


42-44 Bog Road
Penicuik
Midlothian EH26 9BU
UK

Tel. +44 1968 671200


Fax +44 1968 671237

email: sales@findlayirvine.com

website: www.findlayirvine.com

26

You might also like