Manual OMC-118 PDF
Manual OMC-118 PDF
instruments
OMC-118 Anemometer
________________________________________________________________________________________________
WindObserver II
Page 2
Issue 17
Doc. No. 1390-PS-0004
January 2007
Contents
1.
2.
3.
4.
5.
FOREWORD............................................................................................6
INTRODUCTION......................................................................................6
FAST TRACK SET-UP ............................................................................6
SPECIFICATION .....................................................................................7
PRE-INSTALLATION ..............................................................................8
5.1.
5.2.
6.
INSTALLATION.....................................................................................10
6.1.
6.2.
6.3.
6.4.
6.5.
6.6.
6.7.
7.
8.
8.1.
8.2.
8.3.
8.4.
8.5.
8.6.
8.7.
Averaging ......................................................................................................... 28
Options ............................................................................................................. 28
8.8.
________________________________________________________________________________________________
WindObserver II
Page 3
Issue 17
Doc. No. 1390-PS-0004
January 2007
8.9.
8.10.
8.11.
NMEA Format................................................................................................... 32
Analogue outputs and options .......................................................................... 33
Output modes ......................................................................................................... 33
Channel outputs...................................................................................................... 33
Status ...................................................................................................................... 33
Temperature ........................................................................................................... 33
Tunnel mode ........................................................................................................... 33
Scaling .................................................................................................................... 33
Polar mode direction wraparound........................................................................... 34
Low wind speeds .................................................................................................... 34
9.
CONFIGURING......................................................................................35
9.1.
9.2.
9.3.
9.4.
9.5.
9.6.
9.7.
9.8.
9.9.
9.10.
9.11.
9.12.
9.13.
9.14.
10.
10.1.
10.2.
10.3.
10.4.
10.5.
10.6.
Cleaning ........................................................................................................... 42
Servicing........................................................................................................... 42
Fault-finding...................................................................................................... 42
Returning unit ................................................................................................... 43
Status (error) codes.......................................................................................... 43
Bench test......................................................................................................... 43
________________________________________________________________________________________________
WindObserver II
Page 4
Issue 17
Doc. No. 1390-PS-0004
January 2007
11.
APPENDICES........................................................................................46
11.1.
11.2.
11.3.
11.4.
________________________________________________________________________________________________
WindObserver II
Page 5
Issue 17
Doc. No. 1390-PS-0004
January 2007
1.
FOREWORD
Thank you for purchasing the WindObserver II manufactured by Gill Instruments Ltd.
The unit has no customer serviceable parts and requires no calibration or maintenance.
To achieve optimum performance we recommend that you read the whole of this manual
before proceeding with use. Do NOT remove black rubber transducer caps.
Gill products are in continuous development and therefore specifications may be subject to
change and design improvements without prior notice.
The information contained in this manual remains the property of Gill Instruments and
should not be copied or reproduced for commercial gain.
2.
INTRODUCTION
The Gill WindObserver II wind sensor is a very robust, lightweight unit with no moving
parts, outputting wind speed and direction. The units of wind speed, output rate and
formats are all user selectable.
The WindObserver II is available with or without de-icing heating (recommended if icing
is likely), and with or without analogue outputs.
The WindObserver II can be used in conjunction with a PC, datalogger or other device,
provided it is compatible with the RS422 output or the analogue outputs. Multiple units
can be networked if required.
Alternatively, the WindObserver II is designed to connect directly to the Gill WindDisplay
unit to provide a complete wind speed direction system without any configuration required
by the user.
The output message format can be configured in Polar, UV (2-axis), NMEA (0183
Version 3), Tunnel formats, and as either a Continuous output or Polled (requested by
host system) , detailed in full in Section 8 Message Formats & Options.
WindObserver II may be configured using WindCom software which is available, free of
charge, from the Gill website www.gill.co.uk, or by using standard communications
software on a PC. This is explained in Section 9 Configuring.
3.
If you are in a hurry to try out the WindObserver II, are familiar with Gill equipment, and
coupling to a PC using RS422, go to the following sections :
Section 6 Installation
Section 8 Message Formats & Options
Section 9 Configuring
After you have successfully set up the WindObserver II, we strongly advise that you then
go back and read the rest of the manual to ensure that you get the best results from the
WindObserver II.
________________________________________________________________________________________________
WindObserver II
Page 6
Issue 17
Doc. No. 1390-PS-0004
January 2007
4.
SPECIFICATION
Measurement
Output
Parameters
Units
Averaging
Wind Speed
Range
Accuracy
Resolution
Offset
Direction
Range
Accuracy
Resolution
Sonic Temperature Output
Range
Anemometer Status
Starting Threshold
Dead Band Wind Direction
Power Requirement
Anemometer only
Heating - Optional
Digital Output
Communication
Baud rates
Formats
Analogue Output - Optional
Quantity
Scale
Type
V output resistance
4-20mA loading
Dimensions
Size
Weight
Materials
External Construction
Environmental
Moisture protection
Operating temperature
Humidity
Precipitation
EMC
Icing
Standards
Site Calibration
1, 2, 4, 5, 8, 10Hz
UV, Polar, NMEA, Tunnel
m/s, Knots, MPH, KPH ft/min
Flexible 1-3600 seconds
0 - 65m/s (0-145mph)
2%@12m/s
0.01m/s
0.01m/s
0 - 359
2
1
-40C to + 70C
Supplied as part of standard message
0.01 m/s
None
9-30 V DC (40mA @ 12VDC)
3A @ 24V AC or DC
RS422, full duplex, network facility
1200, 2400, 4800, 9600, 19200, 38400
8 data, odd, even or no parity
3 (speed, direction, status or sonic temperature)
Multiples of 10m/s up to 70m/s
2.5V, 0-5V or 4-20mA
60 Ohms
10-300 Ohms
405mm x 210mm
1.5kg
Stainless Steel 316
IP66 (NEMA4X)
-55C to +70C (heated version)
5% to 100% RH
300mm/hr
BS EN 50081-1: 1992 (Emissions class B)
BS EN 50082-2: 1992 (Immunity)
FCC class A
MILSTD810E Method 521.1 Procedure 1
Traceable to UK national standards
None required. (See Section 10.6 Zero Wind Check)
________________________________________________________________________________________________
WindObserver II
Page 7
Issue 17
Doc. No. 1390-PS-0004
January 2007
5.
PRE-INSTALLATION
5.1.
Equipment supplied
or
or
or
and
and
and
and
5.2.
Installation requirements
________________________________________________________________________________________________
WindObserver II
Page 8
Issue 17
Doc. No. 1390-PS-0004
January 2007
Mounting Bracket
A mounting bracket is required (not provided) which will attach to the WindObserver II
via the four tapped holes in the base of the unit. Always ensure that the gasket supplied is
fitted to the base of the anemometer.
It is important that the gasket supplied forms a water tight seal on the base of the
anemometer.
In the event that the mounting bracket supplied by the customer is not flat or does not
form complete support for the anemometer base; the customer must fit a mounting ring as
shown in Figure 2.
See Section 6.7 for mounting details.
________________________________________________________________________________________________
WindObserver II
Page 9
Issue 17
Doc. No. 1390-PS-0004
January 2007
6.
INSTALLATION
Do NOT remove the black rubber transducer caps. Warranty is void if the
blue security seal is damaged or removed.
All the time the WindObserver II is not in its final location, it should be
protected from damage by keeping it in the original packaging as long as
possible, treating it as a delicate instrument.
If an Intrinsically Safe (IS) device is required, the I.S. WindObserver must be
used - the WindObserver II is NOT an IS device.
6.1.
Installation Guidelines
The WindObserver II has been designed to meet and exceed the stringent standards listed
in its specification. Operating in diverse environments all over the world, WindObserver II
requires no calibration or adjustment whatsoever.
As with any sophisticated electronics, good engineering practice should be followed to
ensure correct operation.
Always check the installation to ensure the WindObserver II is not affected by
other equipment operating locally, which may not conform to current standards,
e.g. radio/radar transmitters, boat engines, generators etc.
Guidelines
Avoid mounting in the plane of any radar scanner a vertical separation of
at least 2m should be achieved.
Radio transmitting antennas, the following minimum separations (all round)
are suggested
VHF IMM 1m
MF/HF 5m
Satcom 5m (avoid likely lines of sight)
Ensure the product is correctly earthed in accordance with this manual
Use cables recommended by Gill, keeping the length below the maximum allowed
(SeeSection 6.3) Where the cables are cut and re-connected (junction boxes, plugs
and sockets) the cable screen integrity must be maintained, to prevent the EMC
performance being compromised.
Earth loops should not be created earth the system in accordance with the
installation guidelines. (SeeSection 6.4)
Ensure the power supply operates to the WindObserver II specification at all times.
Avoid turbulence caused by surrounding structures that will effect the accuracy of the
WindObserver II such as trees, masts and buildings. The World Meteorological
Organisation make the following recommendation:
The standard exposure of wind instruments over level open terrain is 10m above
the ground. Open terrain is defined as an area where the distance between the
sensor and any obstruction is at least 10 times the height of the obstruction.
________________________________________________________________________________________________
WindObserver II
Page 10
Issue 17
Doc. No. 1390-PS-0004
January 2007
6.2.
Note : Prior to physically mounting the WindObserver II in its final location, we strongly
recommend that a bench system test is carried out to confirm the system is configured
correctly, is fully functional and electrically compatible with the selected host system and
cabling (preferably utilising the final cable length). The required data format, units,
output rate, and other options should also all be configured at this stage. If an Integrity
Check Chamber (ICC) has been purchased refer to Section 10.6.
6.3.
Cabling
Cable type
A RS422 compatible cable should be used, with the number of twisted pairs matching the
application. If both the (optional) Heating and Analogue outputs are utilised, then a 9 pair
cable is required.
Generic description Twisted pairs with drain wire, screened with aluminised tape,
with an overall PVC sheath. Wire size 7/0.2mm (24 AWG)
The table shows some suitable manufacturers references; other manufactures equivalents
can be used.
Table 1
Belden
Batt electronics
No. of pairs Gill ref.
ref.
ref.
2
9729
3
026-02660
9730
91030
4
026-03156
9728
91199
9
026-02663
8774
91009
Cable length
The typical maximum length at 9600 baud is 1km (3200ft), using the recommended cable.
If any problems of data corruption are experienced (due to, for example, a high local
electrical noise level), then a lower baud rate should be used. Alternatively, a thicker or
higher specification cable can be tried. See also Section 6.1
Cabling and junction box
The WindObserver II unit is fitted with a flying 9 pair cable attached (1.5m or 10m long).
This must be terminated in a suitable terminal box to IP66 or better, fitted with glands to
prevent moisture ingress.
The cable type from the terminal box to the host system must be as specified above. If any
cable is likely to be exposed to mechanical damage, it must be enclosed in a suitable
conduit or cable tray. The cable must be securely fixed with cable clamps or equivalent,
such that the cable is not under stress at the cable glands.
The gland area at the base of the WindObserver II should not be directly exposed to
moisture, as whilst the gland is sealed when mated, the anemometer is vented to air at the
base to avoid pressure build up. If an IP66 rating is essential or the unit is mounted other
than right way up use the gasket provided in the mounting kit.
________________________________________________________________________________________________
WindObserver II
Page 11
Issue 17
Doc. No. 1390-PS-0004
January 2007
6.4.
Connections
Important
Do NOT join any of the cores of the cable together. Joining cores
could damage the unit permanently. Any cores not used should be isolated.
Do NOT connect the units 0V, heating ve, analogue output 0V or digital 0V to the
screen or earth.
The screens of each pair are joined together inside the anemometer - these should be
joined to the cable screen(s) of any further cable run. Avoid long grounding loops. Digital
OV should be used in conjunction with RS422 TX RX lines in order to improve noise
immunity. Each pair in the cable is labelled with its pair number.
Table 2
Conductor colour
Designation
Pair 1
green
RS422_TXB (+)
black
RS422_TXA (-)
Pair 2
white
RS422_RXB (+)
black
RS422_RXA (-)
Pair 3
red
Supply V+
black
Supply VPair 4
blue
Digital 0v
black
Not used (isolate)
Pair 5
yellow
Heater + *
black
Heater *
Pair 6
brown
Analogue Ch 1 voltage output *
black
Analogue Ch 1 current output *
Pair 7
orange
Analogue Ch 2 voltage output *
black
Analogue Ch 2 current output *
Pair 8
white
Analogue Ch 3 voltage output *
red
Analogue Ch 3 current output *
Pair 9
green
Analogue 0 volt reference/current return *
red
Analogue 2.5 volt reference *
Screens
Chassis ground
Earthing or grounding
To ensure correct operation, and for maximum protection against lightning, the
anemometer MUST be correctly earthed (grounded) via its mountings. Inadequate
earthing will degrade anemometer performance, particularly in the presence of radio
frequency interference.
See Fig 1 Suggested mounting bracket and earthing (grounding) arrangements
The unit MUST be connected to an appropriate grounding point with a minimum of 6mm
copper wire, via the M5 base screws. The cable screens must be joined with any cable
screen continuing from the units cable via a junction box. The primary earth for the
anemometer must be provided via the base screws and not via the cable screens.
________________________________________________________________________________________________
WindObserver II
Page 12
Issue 17
Doc. No. 1390-PS-0004
January 2007
WindObserver II
9 pair cable
Pair
Signal names
colour
no.
TXB (+)
green
1
TXA ()
black
CONNECTED INTERNALLY
Supply V+
Supply V-
Heater + *
Heater - *
WindDisplay
Cable 2 or 3
twisted pairs
Screen and
drain wires
Signal
names
TXD +
TXD
Terminal
nos.
RS422+
RS422-
Ground (Earth)
red
black
+
O
yellow
black
+
-
2
1
Heater power
supply
see Section 6.6
* If fitted
Default Settings
The factory default settings are:
A0 B3 C1 E1 F1 G0000 H1 J1 K1 L1 M2 NA O1 P1 T1 U1 V1 X1 Y1 Z1
See Section 9 Configuring for further details
________________________________________________________________________________________________
WindObserver II
Page 13
Issue 17
Doc. No. 1390-PS-0004
January 2007
WindObserver II
9 pair cable
PC with RS422
or RS485 input
Cable - 3 or 4
twisted pairs
Signal names
Pair
no
Colour
TXB (+)
TXA (-)
green
black
RXB (+)
RXA (-)
RXB (+)
RXA (-)
white
black
TXB (+)
TXA (-)
Digital OV
blue
CONNECTED INTERNALLY
Supply V+
Supply V-
Heater + *
Heater - *
Signal names
Signal Ground
Screen and
drain wires
Ground (Earth)
red
black
DC Power
+
supply 9-30V
see section 6.6
yellow
black
+
-
Heated power
supply
see section 6.6
________________________________________________________________________________________________
WindObserver II
Page 14
Issue 17
Doc. No. 1390-PS-0004
January 2007
Networking units
Before coupling units into a network:
Each device must be configured with a unique Unit Identifier (letter B to Z)
It must be configured to a tri-state polled mode M3 or M4 .
See Section 9 - Configuring
WindObserver II
9 pair cable
PC with RS485
card
Signal names
Pair
no
Colour
TXB (+)
RXB (+)
1
2
green
white
TXA (-)
RXA (-)
1
2
black
black
Digital OV
blue
CONNECTED INTERNALLY
Supply V+
Supply V-
Heater +
Heater -
Cable - 2 or 3
twisted pairs
Signal names
.. ..
.
T / RXB (+)
T / RXA (-)
Signal Ground
Screen and
drain wires
Ground (Earth)
red
black
DC Power
+
supply 9-30V
see section 6.6
yellow
black
+
-
Heated power
supply
see section 6.6
To Next Unit
Note : Each unit in the network will require its own power supplies.
Please refer to section 8.3.
________________________________________________________________________________________________
WindObserver II
Page 15
Issue 17
Doc. No. 1390-PS-0004
January 2007
Signal names
TXB(+)
TXA()
Cable Pair
Pair 1
Pair 1
Colour
RXB(+)
RXA(-)
Pair 2
Pair 2
White
Black
3
10
Digital 0v
Pair 4
Blue
POWER V+
Pair 3
Red
POWER V+
POWER 0V
Pair 3
Black
11
POWER 0V
Green
Black
Pin Nos.
2
9
RXB(+)
RXA()
TXB(+)
TXA(-)
Signal Ground
Externally accessible
Internally accessible
________________________________________________________________________________________________
WindObserver II
Page 16
Issue 17
Doc. No. 1390-PS-0004
January 2007
DC Supply 4 way
Pin
Designation
1
Interface V+
2
Interface V3
Anemometer V4
Anemometer V+
________________________________________________________________________________________________
WindObserver II
Page 17
Issue 17
Doc. No. 1390-PS-0004
January 2007
________________________________________________________________________________________________
WindObserver II
Page 18
Issue 17
Doc. No. 1390-PS-0004
January 2007
6.5.
Analogue connections
When using the analogue outputs it is advisable to use a low pass filter to remove any high
frequency noise present.
The analogue signals can be susceptible to external interference if unprotected. The
anemometer cable has individual screens covering each channels signal wires. The
continuing cable connected to the anemometer must be of equal or greater cross sectional
area as the anemometer cable and must be individually screened or screened pairs.
Note that connections can also be made to a PC or other device as described in the
preceding section.
Connections for 5V output
Anemometer
Pair 9
Green
0v Ref
Anemometer
Pair 9
Red
2.5V Ref
________________________________________________________________________________________________
WindObserver II
Page 19
Issue 17
Doc. No. 1390-PS-0004
January 2007
Anemometer
Pair 9
Green
0v
6.6.
300 max
Power supplies
9 to 30v DC
60mA max. 50mA average
Heating
If fitted, the heating module requires a separate power supply.
Either
22 30V DC @ 3A
or
24V AC RMS 10% @ 3A AC RMS
NOTE: AC Supply must be isolated from Mains Supply
________________________________________________________________________________________________
WindObserver II
Page 20
Issue 17
Doc. No. 1390-PS-0004
January 2007
6.7.
Mechanical installation
Before installing, a bench system test is recommended (see Section 6.2).
All the time the WindObserver II is not in its final location, it should be
protected from damage by keeping it in the original packaging as long as
possible, treating it as a delicate instrument.
Location
It is important to ensure that the WindObserver II is mounted in a position clear of any
structure which may obstruct the airflow or induce turbulence.
Do NOT mount the WindObserver II in close proximity of high powered radar or radio
transmitters. A site survey may be required if there is any doubt about the strength of
external electrical noise. ( See Section 6.1 )
Orientation
Normally the WindObserver II is mounted vertically with the cable exit at the bottom.
Alternatively, the unit may be mounted in any orientation, but note that if the unit is
mounted with the cable uppermost ,the gasket provided in the mounting kit must be used.
Mounting bracket
A suggested mounting bracket is shown in Fig 1 below.
It is recommended that the WindObserver II mounting bracket is manufactured from
stainless steel 316, to prevent possible galvanic corrosion.
It is also recommended that the supplied gasket is always fitted to ensure IP66 rating and
to prevent galvanic corrosion if the mounting bracket is not manufactured from stainless
steel 316.
Stainless steel 316 fixings must be used.
________________________________________________________________________________________________
WindObserver II
Page 21
Issue 17
Doc. No. 1390-PS-0004
January 2007
________________________________________________________________________________________________
WindObserver II
Page 22
Issue 17
Doc. No. 1390-PS-0004
January 2007
Alignment
The anemometer should be set to point North (or to another known reference direction),
using the North Spar, which can be identified from the alignment indicator on the base of
the instrument. See Figure 2a.
If the unit is used in a tunnel, (and the Tunnel format is used), the unit would normally be
aligned with the N S axis in the horizontal plane.
Figure 2a North marker and UV polarity definition (UV format) and Mounting Ring
SEE FIG.
DEFINITION
________________________________________________________________________________________________
WindObserver II
Page 23
Issue 17
Doc. No. 1390-PS-0004
January 2007
________________________________________________________________________________________________
WindObserver II
Page 24
Issue 17
Doc. No. 1390-PS-0004
January 2007
7.
The WindObserver II is designed to link directly to the Gill WindDisplay unit to provide a
complete wind speed and direction system.
When coupled to a WindDisplay, the WindObserver II can be used as supplied,
however if a fault occurs, the WindDisplay may lock into the last valid reading. Reconfiguring the Wind Observer to Fixed Field Output (O2) will ensure that any fault
is flagged on the WindDisplay.
After coupling to a WindDisplay, the Wind Speed units and the Averaging period can be
selected using the WindDisplay controls. See the WindDisplay User Manual.
Note that although the WindDisplay can display wind speed in various units, these are
calculated within the WindDisplay. The data coming to the WindDisplay must be in
metres/sec (the factory default output setting).
Figure 3 WindDisplay
The WindDisplay has been designed to provide anemometer power (approximately 15V
@40mA).
If the WindObserver II has a de-icing heater fitted, this will need a separate power supply.
( See Section 6.6 ).
NOTE:- If the WindDisplay is configured for NMEA mode then the WindObserver II
must also be configured for NMEA mode and 4800-baud operation (configuration settings
M5 and B2).
________________________________________________________________________________________________
WindObserver II
Page 25
Issue 17
Doc. No. 1390-PS-0004
January 2007
8.
On first applying power to the WindObserver II, it will be in Measurement Mode, and it
will output wind measurement information in one of nine configurations as described
below.
Setting the output format, units, other parameters, options and the communication settings
are all carried out in the alternative Configuration Mode.
See Section 9 CONFIGURING for details of how this is done.
The factory default settings are shown here in bold, and for convenience some
Configuration codes (as used to set the configuration) are shown in blue boxes.
For example M3 .
8.1.
The wind speed measurements can be output in one of the following formats:
UV, Polar, Customer formats (NMEA and Tunnel).
8.2.
Output formats
The UV and Polar wind speed parameters are output in either ASCII or binary.
These parameters can be transmitted continuously or polled from the user.
Polar is also available in continuous NMEA format.
Figure 4 Output formats
Output format
ASCII
UV
ASCII
Polar
Binary
Binary
Tunnel
NMEA
UV
Polar
8.3.
Output
Continuous
Polled
Continuous
Polled
Continuous
Continuous
Continuous
Continuous
Tri-state o/p
No
Yes
No
Yes
No
No
No
No
Configuration code
M1
M3
M2
M4
M7
M8
M6
M5
Networking
________________________________________________________________________________________________
WindObserver II
Page 26
Issue 17
Doc. No. 1390-PS-0004
January 2007
RS485 I/F
(USB)
SWITCH
HYPER-TERMINAL
B (Full-D)
WObsII
RS422 I/F
(USB)
b (4-wire)
One PC (with two USB ports) can be used for both Terminals, although two may be less
confusing.
Note: Power must not be disconnected from the anemometer throughout this procedure.
2) Assuming the unit is set for Full Duplex, connect 9-way at [b] disconnect 9-way at
[a]. Set Switch to position [b] (4 wire).
3) On Terminal B, enter CONFIGURATION MODE. Set Half Duplex (E2). The
CONFIRM> prompt will appear.
4) Disconnect 9-way [b] and connect 9-way [a]. Set Switch to position [a] (2 wire).
5) On Terminal A, type E <ENTER> to confirm.
6) The unit should now be configured in Half Duplex mode.
Configuring WindObserver II for Full Duplex Operation.
1) Assuming the unit is set for Half Duplex, disconnect 9-way [b] and connect 9-way
[a]. Set Switch to position [a] (2 wire).
2) On Terminal A, type E1 to select Full Duplex. The CONFIRM> prompt will
appear.
3) Disconnect 9-way [a] and connect 9-way [b]. Set Switch to position [b] (4 wire).
4) On Terminal B type E <ENTER> to confirm.
8.4.
Units
8.5.
Output rate
The output frequency can be set between 1 and 10 per second. The wind speed is sampled
every 25ms, each axis being sampled sequentially. The following Table lists the possible
outputs per second and the corresponding number of averaged samples. A 25ms sample
period is used to process and output the results.
Outputs per Second
1
2
4
5
8
10
8.6.
Sample Average
39
19
9
7
4
3
Averaging
The Averaging Period can be set from zero to 3600 secs. (1 hour). The default setting is
zero. When averaging is enabled, data is output at a rate determined by the averaging
period. The data is an average of valid data collected during the averaging period.
8.7.
Options
________________________________________________________________________________________________
WindObserver II
Page 28
Issue 17
Doc. No. 1390-PS-0004
January 2007
8.8.
ASCII UV format
U axis wind speed
Units
<STX>A,+000.04,-000.02,M,+345.01,+022.36,00,<ETX>25
Start of
string
Unit
identification
Speed of
sound
Temperature
Status
End of
string
Check
sum
where
U axis wind speed
V axis wind speed
Units
Units
Metres per second (m/s)
Knots (knots)
Miles per hour (MPH)
Kilometres per hour (kph)
Feet per minute (fpm)
Start of string
Unit identification
Speed of sound
Temperature
ASCII value 2
Letter A Z (default A)
In metres per second (if enabled)
Sonic temperature in degrees C (if enabled)
Status
Value
0
60
Any other value
End of string
Check sum
ASCII value 3
Exclusive OR of all characters between Start of String
and End of String reported as ASCII hex.
Status
OK
Heating enabled and OK if enabled
Warning or fault condition
- see Section 10.5
________________________________________________________________________________________________
WindObserver II
Page 29
Issue 17
Doc. No. 1390-PS-0004
January 2007
Wind speed
Units
<STX>B,121,000.08,M,+344.66,+021.76,00,<ETX>06
Start of
string
Unit
identification
Speed of
sound
Temperature
Status
End of
string
Check
sum
where
8.9.
Wind direction
Wind speed
Binary UV format
U axis wind speed
Units
Speed of sound
Temperature
Status
where
U axis wind speed
V axis wind speed
Units
Units
Metres per second (m/s)
Knots (knots)
Miles per hour (MPH)
Kilometres per hour (kph)
Feet per minute (fpm)
________________________________________________________________________________________________
WindObserver II
Page 30
Issue 17
Doc. No. 1390-PS-0004
January 2007
Start of string
Speed of sound
Temperature
Status
Value
0
60
Any other value
Status
OK
Heating enabled and OK
Warning or fault condition
- see Section 10.5
Wind direction
Units
Speed of sound
Temperature
Status
where
Wind direction
Wind speed
Direction
Units
Status
Other parameters
________________________________________________________________________________________________
WindObserver II
Page 31
Issue 17
Doc. No. 1390-PS-0004
January 2007
Wind direction
Wind speed
Units
Checksum
delimiter
Start of string
Instrument
type
Wind
measurement
Status
Check
sum
where
Wind direction
Wind speed
Units
M
N
P
K
F
Start of string
Instrument type
Wind measurement
Status
$
IIMWV
R
Relative wind measurement
A
Valid measurement
V
Invalid measurement
Exclusive OR of all characters between $ and *
reported as ASCII hex.
Check sum
________________________________________________________________________________________________
WindObserver II
Page 32
Issue 17
Doc. No. 1390-PS-0004
January 2007
Channel 1
U
Magnitude
Magnitude
U
Channel 2
V
Direction
Direction
U direction*
Channel 3
Status or Temperature
Status
If channel 3 is configured for Status, the output will be 0v (or 4mA) if the unit is an OK
condition. In a fault condition, error code 01, 02, or 04, the output will be +5v (or 0mA).
See Section 10.5 for the list of Error Codes.
Temperature
If channel 3 is configured for Temperature, it reports the temperature if it is between 40
and +70C, scaled from +0.7V to +5V (6.25mA to 20mA). Temperatures outside this
range are reported as errors by setting the output to 0V (4mA).
Tunnel mode
In Tunnel mode, Channel 2 gives the direction of the wind along the U axis, where a U
vector (=wind from the N direction) is represented by 0v (or 4mA), and a +U vector
(=wind from the S direction) by 5v (or 20mA).
Scaling
All analogue outputs are in m/s and can be scaled to have a full scale deflection of 10 to
70 units.
In Polar mode the sign should be ignored. Only the + sign is effective.
In all cases zero Wind Speed is represented by 0V or 4mA.
________________________________________________________________________________________________
WindObserver II
Page 33
Issue 17
Doc. No. 1390-PS-0004
January 2007
Standard
Wraparound
Standard
Wraparound
0
2.5
5.0
n/a
0
1.67
3.33
5.0
4.00
12.00
20.00
n/a
4.00
9.33
14.67
20.00
________________________________________________________________________________________________
WindObserver II
Page 34
Issue 17
Doc. No. 1390-PS-0004
January 2007
9.
CONFIGURING
WindCom may be used to configure WindObserver II. This is an easy to use menudriven package, available as a free download at www.gill.co.uk .
Alternatively, you may use terminal emulator software such as Hyperterminal or
Terminal.
This section describes the commands used to change settings when using terminal
emulator software. For a fuller descriptions of the settings and options see Section 8 Message Formats & Options
Keystrokes are shown thus :- D 3 ENTER
9.1.
9.2.
________________________________________________________________________________________________
WindObserver II
Page 35
Issue 17
Doc. No. 1390-PS-0004
January 2007
9.3.
We strongly recommend that, as a standard procedure, you use this command ( D3 ) prior
to, and after, changing any settings. It shows the current settings for all the alterable
settings. We suggest you note down your settings, so that you can easily return to them.
Enters Configuration Mode (from Continuous mode)
or
*N
Enters Configuration Mode (from Polled mode)
D 3 ENTER
The WindObserver II responds with the current settings.
The factory default settings are
A0 B3 C1 E1 F1 G0000 H1 J1 K1 L1 M2 NA O1 P1 T1 U1 V1 X1 Y1 Z1
*
9.4.
Q ENTER
Changing settings
To change a setting, first go into Configuration mode and then refer to the sections below.
Enter the Configuration code of the new setting required, followed by ENTER .
If successful, the new setting will be echoed back as a message by the WindObserver II.
For example, to change the message format to NMEA, enter M 5 ENTER .
The WindObserver II will reply with M5. When the unit is returned to the Measurement
mode, it will be in NMEA format.
Note : The factory-set (default) settings are shown in bold in the following sections.
9.5.
Output format
Output format
ASCII UV Continuous
ASCII UV Polled (tri-state) *
ASCII Polar Continuous
ASCII Polar Polled(tri-state) *
Binary UV Continuous
Binary Polar Continuous
Tunnel Continuous
NMEA Continuous
Configuration code
M1
M3
M2
M4
M7
M8
M6
M5
________________________________________________________________________________________________
WindObserver II
Page 36
Issue 17
Doc. No. 1390-PS-0004
January 2007
9.6.
Unit Identifier
9.7.
Units
Units
metres/sec
(m/s)
knots
(knots)
miles / hour
(MPH)
kilometre/hour
(kph)
feet / minute
(fpm)
U1
U2
U3
U4
U5
Configuration
code
9.8.
Output rate
9.9.
10
P1
P3
P2
P4
P5
P6
Averaging
Setting
Configuration code
Averaging (Default)
G0000
Enter the required averaging period in seconds as a four figure number between 0000 and
3600.
9.10. Options
Speed of Sound (SOS) and Temperature
Setting
Configuration code
Report SOS
A1
Report Sonic temperature
A2
Report SOS and Sonic temperature
A3
Disable both
A0
Note that neither of these parameters is available in Tunnel or NMEA formats.
Heating (If fitted)
Heating is autonomous and requires no set-up once activated.
Setting
Configuration code
Disabled
H1
Activated
H2
Each transducer is heated independently and will be active when ambient temperature
drops below approximately +15C each transducer will de-activate when +25C threshold
is reached.
________________________________________________________________________________________________
WindObserver II
Page 37
Issue 17
Doc. No. 1390-PS-0004
January 2007
NMEA
Setting
NMEA string IIMWV
NMEA string WIMWV
Configuration code
K1
K2
Configuration code
V1
V2
45 Offset
Setting
Align U axis with
transducer axis
Align U axis +45
to transducer axis
Reverses Polar
Direction
Align @ 45 from
North
Configuration
code
Notes
X1. Aligns U axis with North/South axis.
X1
X2
X3
X4
Figure 4 shows the polarity of U and V if the wind components along the U and V axis are
blowing in the direction of the respective arrows.
-U
N
-V
+V
S
+U
Figure 4 UV Polarity
________________________________________________________________________________________________
WindObserver II
Page 38
Issue 17
Doc. No. 1390-PS-0004
January 2007
Configuration code
B1
B2
B3
B4
B5
B6
B7
Configuration code
Full duplex
E1
Half duplex
E2
Note:- Refer also to Para 8.3 Networking Configuration details.
Data and parity options
Setting
8 bits, no parity, 1 stop bit
8 bits, even parity, 1 stop bit
8 bits, odd parity, 1 stop bit
7 bits, no parity, 1 stop bit
7 bits, even parity, 1 stop bit
7 bits, odd parity, 1 stop bit
Configuration code
F1
F2
F3
F4
F5
F6
________________________________________________________________________________________________
WindObserver II
Page 39
Issue 17
Doc. No. 1390-PS-0004
January 2007
Configuration code
CR LF
LF
L1
L2
Configuration code
Configuration code
Status
Temperature
Y1
Y2
Scaling
Setting
10
20
30
40
50
60
70
Z1
Z2
Z3
Z4
Z5
Z6
Z7
Configuration code
Configuration code
C1
C2
________________________________________________________________________________________________
WindObserver II
Page 40
Issue 17
Doc. No. 1390-PS-0004
January 2007
Configuration code
J0
J1
J2
J3
Item
Command
code
D1
Unit configuration
D3
Analogue ID
(if present)
Anemometer
power supply
voltage
Integrity check
D2
Typical response
E000694
2.07
current configuration :
A0 B3 C1 E1 F1 G0000 H1 J1 K1
L1 M2 NA O1 P1 T1 U1 V1 X1 Y1
Z1
D4
GI1390 fitted.
D5
+29.8
D6
10.2. Servicing
There are no moving parts or user-serviceable parts requiring routine maintenance.
Opening the unit or breaking the security seal will void the warranty and the calibration.
In the event of failure, prior to returning the unit to your authorised Gill distributor, it is
recommended that :
All cables and connectors are checked for continuity, bad contacts, corrosion etc.
A bench test is carried out as described in Section 10.6.
You contact your supplier for advice
10.3. Fault-finding
Symptom
No output
Corrupted output
One way
communication
Failed / Incorrect
WindObserver II
ouput, data
invalid flag
Solution
Check DC power to WindObserver II, cable and connections.
Check comms settings of WindObserver II (as detailed in Section
9) and host system match, including correct Com port
Check unit is in Continuous mode
Check that in-line communication devices are wired correctly.
NOTE: It is usual for Anemometer TX + to be connected to
converter device RX +
Check comms settings of WindObserver II and host system
match.
Try a slower baud rate.
Check cable lengths and type of cable.
Check wiring is in accordance with the manual.
Check that transducer path is not blocked
________________________________________________________________________________________________
WindObserver II
Page 42
Issue 17
Doc. No. 1390-PS-0004
January 2007
Status
Condition
OK
Sufficient samples in average period
OK and heating
Sufficient samples in average period
enabled
OK
NMEA data valid
Axis 1 failed
Insufficient samples in average period on U axis
Axis 2 failed
Insufficient samples in average period on V axis
Axis 1 and 2 failed
Insufficient samples in average period on both axes
NVM error
NVM checksum failed
ROM error
ROM checksum failed
System gain at max.
Inaccurate results likely
Marginal system gain
Results OK, but marginal operation
Measurement average building
Heating current tripped or electronic failure
Thermistor open circuit
Heating element open circuit
NMEA data invalid
________________________________________________________________________________________________
WindObserver II
Page 43
Issue 17
Doc. No. 1390-PS-0004
January 2007
If there has been no significant changes to The WindObserverII head configuration then
*PASS* will confirm correct operation.
Alterations to the head geometry will result in a *FAIL* message. If this occurs please
contact Gill Instruments.
________________________________________________________________________________________________
WindObserver II
Page 45
Issue 17
Doc. No. 1390-PS-0004
January 2007
11. APPENDICES
11.1. Glossary & Abbreviations
Item
CAL
CR
CRLF
CSV
ENG
ESC
ETX
FAC
fpm
GND
HEX
I/P
IP66
KPH
LF
m/s
MAG
MAX
MPH
NEMA
NMEA 0183
(version 3)
No:
NVM
O/P
PC
PCB
POR
PROCOMM
ROM
RS232
Meaning
Calibration
Carriage Return
Carriage Return Line Feed
Comma Separated Variable
Engineering
ESCape key on keyboard used to stop any process that is being
executed
End of string character
Factory
Feet per minute
GrouND
HEXadecimal
InPut
Protection Classification
Kilometres per Hour
Line Feed
Metres per second
MAGnitude - scalar reference to wind speed
MAXimum
Miles per Hour
National Electrical Manufacturers Association
National Marine Electronics Association standard for interfacing
marine electronic navigational devices
Number
Non-Volatile Memory
Output
IBM compatible Personal Computer
Printed Circuit Board
Power On Reset
Terminal emulator software package
Read Only Memory
Communications standard
________________________________________________________________________________________________
WindObserver II
Page 46
Issue 17
Doc. No. 1390-PS-0004
January 2007
Item
RS422
RS485
RTS
RX
RXD
S/W
SOS
SEC
STX
TERM
TX
TXD
+VE
-VE
WindCom
WRT
Meaning
Communications standard
Communications standard
Request To Send
Receive
Received Data
SoftWare
Speed Of Sound
SECond
Start of string character
TERMinal
Transmit
Transmitted Data
Positive
Negative
Gill software used to configure WindObserver II
With Respect To
11.2. Guarantee
For terms of guarantee contact your supplier.
Warranty is void if the red security seal covering base nuts is damaged or broken, or the
transducer caps have been damaged.
________________________________________________________________________________________________
WindObserver II
Page 47
Issue 17
Doc. No. 1390-PS-0004
January 2007
________________________________________________________________________________________________
WindObserver II
Page 48
Issue 17
Doc. No. 1390-PS-0004
January 2007
to which this declaration relates, are in conformity with the protection requirements of Council Directive
89/336/EEC on the approximation of the laws relating to electromagnetic compatibility.
This Declaration of Conformity is based upon compliance of the product with the following harmonized
standards:
EN50081 - 1
EN50082 - 2
Signed by:
A.C.R. Stickland Director
Date of issue:
28/11/2000
Place of issue:
________________________________________________________________________________________________
WindObserver II
Page 49
Issue 17
Doc. No. 1390-PS-0004
January 2007