Location via proxy:   [ UP ]  
[Report a bug]   [Manage cookies]                

Surtron300 400

Download as pdf or txt
Download as pdf or txt
You are on page 1of 4

AMBULATORY SURGERY

GENERAL SURGERY
PEDIATRIC SURGERY
PLASTIC SURGERY
THORAX SURGERY
VASCULAR SURGERY
DERMATOLOGY
ENDOSCOPY
GASTROENTEROLOGY
GYNECOLOGY
NEUROSURGERY
ORTHOPEDICS
OTORHINOLARYNGOLOGY
PNEUMOLOGY
UROLOGY
VETERINARY

www.surtron.com
High performance
capabilities in a
multipurpose generator

Reference code
300HP 400HP
10400.801 10400.901
Maximum output power monopolar pure CUT 300 W - 300 Ω 400 W - 300 Ω
Maximum output power monopolar cut-coag CUT80% 250 W - 300 Ω 300 W - 300 Ω
Maximum output power monopolar cut-coag CUT60% 200 W - 300 Ω 250 W - 300 Ω
Maximum output power monopolar cut-coag CUT40% 150 W - 300 Ω 200 W - 300 Ω
Maximum output power monopolar cut ENHANCED 200 W - 500 Ω 250 W - 500 Ω
Maximum output power monopolar cut-coag BLEND 200 W - 300 Ω 250 W - 300 Ω
Maximum output power monopolar coag SPEEDY 100 W - 500 Ω 120 W - 500 Ω
Maximum output power monopolar coag DEEP 100 W - 200 Ω 120 W - 200 Ω
Maximum output power monopolar coag SPRAY 70 W - 2000 Ω 100 W - 2000 Ω
Maximum output power monopolar coag SPRAY ARGON 70 W - 2000 Ω 100 W - 2000 Ω
Maximum output power bipolar CUT 95 W - 150 Ω 95 W - 150 Ω
Maximum output power bipolar cut-coag CUT80% 95 W - 150 Ω 95 W - 150 Ω
Maximum output power bipolar cut-coag CUT60% 95 W - 150 Ω 95 W - 150 Ω
Maximum output power bipolar cut-coag CUT40% 60 W - 150 Ω 60 W - 150 Ω
Maximum output power bipolar cut-coag BLEND 90 W - 150 Ω 90 W - 150 Ω
Maximum output power bipolar coag COAG 70 W - 100 Ω 70 W - 100 Ω
Working frequency Monopolar / Bipolar 425 kHz / 525 kHz 425 kHz / 525 kHz
Patient circuit F F
Selectable input voltage 115 - 230 Vac 115 - 230 Vac
Mains frequency 50 - 60 Hz 50 - 60 Hz
Electrical input power 1000 VA 1000 VA
Size WxHxD mm 470x150x400 470x150x400
Weight 17,5 Kgs 17,5 Kgs
TWIN OPERATIVE ACTIVITY
• Two output channels for monopolar cutting
• Two independent setting and output channels for
the most frequent monopolar coagulation
CONTROLLED CUTTING PROCEDURES
• Enhanced cutting current for eschar free incision
• Cutting current with selectable coagulating property
for bleeding controlled dissection
• Low capacitive coupling by slow BLEND current for
endoscopic procedures
TWIN CONTROLLED COAGULATIVE PROCEDURES
Multiprogrammable • SPEEDY* coagulation and DEEP**coagulation
available to two operators
by the user • SPRAY coagulation available by hand switch
or foot switch controls
WIDE BIPOLAR FEATURES
• Bipolar cutting current with selectable bleeding control
• Slow BLEND cutting current
• Separately activable Automatic Start/Stop coagulation
SURGEON AND PATIENT TOTAL SAFETY
• Acceptability of contact impedance between patient
body and return electrode continuously watched
by Skin Plate Electronic Control (SPEC)
• H.F. Leakage currents continuously monitored
• Mains fluctuation automatically compensated
before to start the delivery of HF power
Double generator PERSONALIZED PROCEDURES
• Independent choise of coagulative currents
for twin surgical and power levels by each of two operators

procedures • Ten personal configurable and immediately


recallable procedures
EXTENTION DRIVER AND INTERLOCK SIGNALS
• Drived output connection to external Argon Plasma
unit and Smoke Evacuator unit
AUTOMATIC SELF DIAGNOSIS
• Automatic diagnostic control of the actual complete
functionality of the electronic unit and the connected
accessories with coded report of failures
*SPEEDY coagulation also said Fulgurate or Forced coagulation
**DEEP coagulation also said Pin Point, Dessicate or Soft coagulation

SAFETY CONTROLS
EN60601-1 Patient/Plate circuit monitoring
EN60601-1-2 Output power monitoring
EN60601-2-2 Self check control
Electrical Class: I CF H.F. Leakage currents control
MDD 93/42/EC Class: II b Compensation of mains fluctuation
© 2009 LED SpA
PRINTED IN ITALY
LED SpA
PROGETTAZIONI E PRODUZIONI ELETTRONICHE

DP346GBa
via Selciatella, 40 - 04011 Aprilia(LT) - Italy
www.led.it - info@led.it - www.surtron.com

Data, descriptions, dimensions, weights and illustrations are purely indicative. Manufacturer reserved the right to modify them without warning.
Size - mm Standard Composition:
MONOPOLAR 1 MONOPOLAR 2 BIPOLAR
- Disposable handle with switches (5pcs) 755VL
Cut
60 70 80 100

Cut %
Coag 1 Coag 2 Cut
60 70 80 100

Cut %
Coag

Start

Stop
- Reusable handle with switches F4243
OK
CUT/COAG 1 CUT/COAG 2 BIPOLAR

- Blade electrode 7 cm (3 pcs) 152-110


152-115
OC

- Blade electrode 16 cm (3 pcs)


- Needle electrode 7 cm (3 pcs) 152-120
- Ball electrode 6 cm (3 pcs) 152-150
- Cable for neutral electrode 00404.08
- Disposable split neutral electrode (5pcs) F7920
- Electrode cleaning sponge F7520
- Double water-proof foot-switch (2 pcs) 00301.04
- Power supply cable 5 mt 00100.01

SURTRON and LED logo are registered trademarks of LED SpA