Toxins: A Review: Sample Preparation and Chromatographic Technologies For Detection of Aflatoxins in Foods
Toxins: A Review: Sample Preparation and Chromatographic Technologies For Detection of Aflatoxins in Foods
Toxins: A Review: Sample Preparation and Chromatographic Technologies For Detection of Aflatoxins in Foods
Review
A Review: Sample Preparation and Chromatographic
Technologies for Detection of Aflatoxins in Foods
Kai Zhang 1, * and Kaushik Banerjee 2
1 US Food and Drug Administration/Center for Food Safety and Applied Nutrition, 5001 Campus Drive,
College Park, MD 20740, USA
2 National Reference Laboratory, ICAR-National Research Centre for Grapes, Pune 412307, India;
kbgrape@yahoo.com
* Correspondence: kai.zhang@fda.hhs.gov
Received: 29 July 2020; Accepted: 19 August 2020; Published: 21 August 2020
Abstract: As a class of mycotoxins with regulatory and public health significance, aflatoxins (e.g., aflatoxin
B1 , B2 , G1 and G2 ) have attracted unparalleled attention from government, academia and industry due
to their chronic and acute toxicity. Aflatoxins are secondary metabolites of various Aspergillus species,
which are ubiquitous in the environment and can grow on a variety of crops whereby accumulation
is impacted by climate influences. Consumption of foods and feeds contaminated by aflatoxins are
hazardous to human and animal health, hence the detection and quantification of aflatoxins in foods
and feeds is a priority from the viewpoint of food safety. Since the first purification and identification
of aflatoxins from feeds in the 1960s, there have been continuous efforts to develop sensitive and
rapid methods for the determination of aflatoxins. This review aims to provide a comprehensive
overview on advances in aflatoxins analysis and highlights the importance of sample pretreatments,
homogenization and various cleanup strategies used in the determination of aflatoxins. The use of
liquid-liquid extraction (LLE), supercritical fluid extraction (SFE), solid phase extraction (SPE) and
immunoaffinity column clean-up (IAC) and dilute and shoot for enhancing extraction efficiency and
clean-up are discussed. Furthermore, the analytical techniques such as gas chromatography (GC),
liquid chromatography (LC), mass spectrometry (MS), capillary electrophoresis (CE) and thin-layer
chromatography (TLC) are compared in terms of identification, quantitation and throughput. Lastly,
with the emergence of new techniques, the review culminates with prospects of promising technologies
for aflatoxin analysis in the foreseeable future.
1. Introduction
Aflatoxins are a class of toxic metabolites generated by certain fungi that are commonly found
in the environment. Aspergillus flavus and Aspergillus parasiticus are the dominant fungal species
that produce aflatoxins under drought, warm and humid conditions [1–3]. In the 1960s, Turkey “X”
disease resulted in the death of over 100,000 turkeys and other poultry in England [4–7]. The outbreak,
associated with contaminated feed, led to the identification of aflatoxins and commenced research on
aflatoxin toxicity. Since then aflatoxin contamination has been gradually recognized as a threat to global
food safety and trade [8–11]. Continuous and collaborative efforts from government, academia and
industry have been dedicated to understanding and mitigating the impact of aflatoxins on public
health and economy [12–14].
Aflatoxins are well known to contaminate crops in the field and during storage [15]. Commonly
consumed food commodities such as cereals, peanuts and dairy products are a major concern for
aflatoxin contamination because current agricultural practices and food processing techniques cannot
completely remove aflatoxins [1,16,17]. Consequently, aflatoxins can enter the food chain and give
rise to health and trade related issues [18–20]. Among various types of aflatoxins, the predominant
ones—aflatoxin B1 , B2 , G1 and G2 and aflatoxin M1 , a hydroxylated metabolite of aflatoxin B1 —can
pose a severe threat to human and animal health via dietary exposure.
Toxicity of aflatoxins is well–documented. As reported by the International Agency for Research
on Cancer (IARC), aflatoxins are carcinogenic and hepatotoxic to humans. Aflatoxin B1 , B2 , G1 and
G2 are classified as human carcinogens (group 1B). Aflatoxin B1 is the most potent carcinogen within
the group. Although aflatoxin M1 is less hepatotoxic and immunotoxic than its parent compound,
aflatoxin B1 , it is stable under pasteurization temperature used to process milk. It is classified as a
possible human carcinogen (group 2B) by the IARC. [21]. In response, regulatory bodies worldwide
have set regulatory levels for aflatoxins, especially for aflatoxin B1 and aflatoxin M1 , in various
commodities [22]. For example, in peanuts, regulatory levels range from 2 µg/kg for aflatoxin B1
and 4 µg/kg for total aflatoxins (sum of aflatoxin B1 + B2 + G1 + G2 ) by the European Union [23]
to 20 µg/kg for total aflatoxins by the United States Food and Drug Administration [24]. Similarly,
the Food Safety and Standards Authority of India [25] has set the limit of total aflatoxins at 10 µg/kg
for ready-to-eat nuts and 15 µg/kg for processed nuts. In the US, the action level of aflatoxin M1 in
raw milk is 0.5 µg/kg, while in the EU the maximum level of aflatoxin M1 in milk is 0.05 µg/kg and
0.025 µg/kg in infant formulas and dietary foods for special medical purposes intended specifically for
infants. Large variability among these regulatory levels exists because of the requirements (feasibility
and economic impact) and approaches to risk analysis are defined in national/regional legislation [26].
To comply with established regulatory levels and ensure food and feed are safe for consumption
and trade, development of analytical methods that are suitable for monitoring and surveillance of
aflatoxins are critical. A well-established aflatoxin method is based on a thorough understanding
of the target analytes (polarity, water solubility, thermostability, partitioning coefficient, pKa, etc.),
food composition (e.g., fat, water and sugar content), extraction conditions, sample clean-up and
instrumentation. Figure 1 shows the important physicochemical properties of aflatoxins B1 , B2 , G1 , G2
and M1Toxins
. 2020, 12, x FOR PEER REVIEW 3 of 39
Development of sample preparation and detection techniques has been influenced by these
properties. For example, selection of extraction solvents should be based on the polarity and
solubility of aflatoxins. Detection and quantitation of aflatoxins could be achieved using their UV and
fluorescence spectrum. Molecular ions and structurally specific fragments of aflatoxins are often used
to achieve unambiguous identification in mass spectrometry analysis. Over the years, numerous
methods have been developed to address challenges including heterogeneous distribution,
Toxins 2020, 12, 539 3 of 39
Development of sample preparation and detection techniques has been influenced by these
properties. For example, selection of extraction solvents should be based on the polarity and solubility
of aflatoxins. Detection and quantitation of aflatoxins could be achieved using their UV and fluorescence
spectrum. Molecular ions and structurally specific fragments of aflatoxins are often used to achieve
unambiguous identification in mass spectrometry analysis. Over the years, numerous methods have
been developed to address challenges including heterogeneous distribution, difficulties in cleanup
strategies for diverse matrices, laborious steps during extraction and matrix effects. By retrospectively
comparing different techniques and evaluating their potentials and limitations, we can understand
and appreciate how important discoveries in the past have led to our current aflatoxin analysis
procedures and the challenges we need to address now and the directions we want to take in the future.
So, it would be worthwhile to provide a comprehensive summary of developed sample preparation
and analytical technologies for the determination of aflatoxins. In brief, this review not only outlines the
significant advances in sample preparation, and analytical strategies in the area of aflatoxin analysis in
a wide range of food matrices, namely cereals, peanut, spices and dairy products, but also illuminates
existing and emerging challenges that should be addressed presently and in the foreseeable future
from a practical standpoint.
2. Sample Preparation
Aflatoxin samples are generally collected from the field following predefined sampling
plans [27–29]. After sample collection, determination of aflatoxins relies on how samples
are homogenized, followed by the removal of test portions (analytical samples) to ensure
representative qualitative and quantitative analytical data can be generated to support risk assessment,
exposure studies and compliance activities. Furthermore, due to the limitation of available detection
technologies, the majority of foods that are prone to mycotoxin contamination cannot be directly
analyzed for aflatoxins without extraction and clean-up, a fact that has been incentivizing researcher
to use different extraction and clean-up methods to separate aflatoxins from matrix components.
The purpose of sample preparation is to minimize the impact of heterogeneous distribution of
aflatoxins in various food commodities [30–32], selectively enrich for aflatoxins based on their
physicochemical properties and reduce bulk background matrix to make final extracts more suitable
for instrumental analysis.
homogenous sample particle size. The same strategy of grinding (and sieving) was applied in red
chili [39], and dry date [40] for the determination of aflatoxin B1 . Dhanshetty et al., 2019 [41] also
demonstrated the use of dry grinding for a range of cereal matrices providing satisfactory homogeneity
and recoveries of detected aflatoxins (84–106%). In another study, Zahra et al. [42], who performed dry
grinding as per the AOAC method [34] for various spices (e.g., black pepper and red chilies) and dry
fruits (e.g., dried apricots and figs). Spanjer et al., 2008 [43] successfully used dry milling to process
peanut, pistachio, wheat, maize, cornflakes, raisins and figs for the analysis of multiple mycotoxins
including aflatoxin B1 . However, due to capacity limitations of their device, the drying milling protocol
was only developed for matrices with sample sizes less than 1.5 kg.
Numerous heavy-duty and batch-type milling and mixing devices (e.g., Hobart VCM,
Waring Blendor and Polytron) are commercially available and have been tested in cereal grains,
tree nuts, botanicals and feeds [44–47]. Although, dry grinding has been widely used by mycotoxin
labs, when sample sizes are large or food products contain high sugar or fat content, dry grinding
could suffer clogging or buttering due to excessive heat generated by fast moving blades. In such
situations, sample grinding results in insufficient homogenization (e.g., particle sizes are too large) [48].
Therefore, despite its laborious and time-consuming nature, wet grinding can serve as an alternative
milling approach, providing better homogenization with smaller particle sizes and more uniform
distribution of aflatoxins.
determined in some matrices using wet grinding could be higher than those done by dry milling.
Similar observations were reported in other studies [36,53], underlying the effects of sample sizes,
matrix compositions, as well as grinding devices/conditions should be carefully evaluated. In a recent
study, Kumphanda et al. [54] demonstrated the analytical variances associated with testing a 25.0 g
slurry (1:1, sample:water), 50.0 g dry grind and 12.5 g dry grind sample portion in the ratio of 1:5:15,
suggesting that the wet grinding could provide improved homogeneity for cereals.
In recent years, LC-MS-based mycotoxin analysis has been increasingly used. Due to its
superior sensitivity, test portions (0.5–5 g) for LC-MS analysis are much smaller than those of
LC-Ultraviolet/Fluorescence detection (LC-UV/FLD) (25–50 g). To ensure test portions would give
representative information of the sample, Zhang et al., 2017 [55] developed a secondary fine-milling
procedure for LC-MS/MS-based multimycotoxin analysis. A subsample (25 g) of corn or wheat flour
was taken and mixed with water (sample to water mass ratio, 1/1) and 2 g of the resulting slurry
was taken for extraction and LC-MS/MS analysis. Using this procedure, the impact of potential
homogeneity concerns is minimized. Furthermore, this approach builds on the selection of a subsample
size (e.g., 25–50 g) previously shown to be representative of the initial sample, an approach that would
allow scaling to small analytical test portions. An additional benefit of small test portions is the
significant reduction in solvents used and waste generated, making sample analysis more cost-effective.
one of the continuing challenges is to characterize the correlation between homogeneity of the final
sample and grinding conditions (devices, milling time, sample type, sample size, subsample size and
test portion size). Use of sample sizes too small can lead to biased analysis of aflatoxins with high
variability. On the other hand, using unnecessary large sample sizes could increase operation cost and
impact throughput when a large number of samples need to be processed. Therefore, in the course
of routine sample analysis or evaluation of new homogenization devices/protocols, a homogeneity
test and quality control should be performed to ensure that the homogenization process could meet
the predefined degree of fineness in the final sample. In some situations, conventional homogeneity
protocols may be cumbersome [60], so it is important to explore and adopt existing technologies such
as light scattering to monitor the homogeneity of processed samples in an efficient manner [59,61].
Diversity of food commodities that need to be analyzed for aflatoxins makes it impossible to
establish a universal milling protocol for all sample matrices. Instead, food commodities can be
classified according to composition complexity including fat, protein, water, pigments and other
interferences. Products that are similar in composition can often be homogenized by the same or
slightly modified methods. Based on the class of the product, past experience and the literature record
of milling applications, the most appropriate milling practice can be selected and developed in a
practical manner.
2.2. Extraction
The diverse nature of aflatoxin sample matrices demands the use of a combination of extraction
and clean-up techniques in treating a sample prior to instrumental analysis. To achieve satisfactory
recoveries of aflatoxins, extraction needs to be applied to samples first, followed by clean-up, if necessary.
For dry matrices, hydration is generally required to wet and swell samples, ensuring efficient extraction
and separation of aflatoxins absorbed within the sample. For fatty matrices, additional defatting
steps are critical to eliminate co-extracted fatty interferences (e.g., lipid and cholesterol). In this
section, commonly used extraction solvents (e.g., acetonitrile, methanol, chloroform and hexane) and
techniques are reviewed and compared for their applications in aflatoxin analysis.
in the acetone/water phase was precipitated as insoluble lead salt. Aflatoxins were then partitioned into
chloroform from the acetone/water phase by repetitive liquid-liquid extractions prior to TLC analysis.
Since the late 1960s, liquid-liquid extraction has been gradually coupled with other developed
clean-up techniques (e.g., silica gel chromatography) for aflatoxin analysis [67–72]. For example,
to determine aflatoxins in peanut butter, Tarter et al. [71] used methanol to extract samples, followed by
partitioning fatty interference into hexane. The remaining aflatoxins in the aqueous methanol phase
were partitioned into chloroform. Then the chloroform extracts were further purified using an in-house
developed silica gel chromatography method. Two laboratories tested the method in eight different
matrices, including peanut butter, and achieved consistent results. The mean recoveries of spiked
aflatoxins (15–34 ng/g) ranged from 86 to 94% with coefficients of variation less than 7% for spiked
samples and 15% for naturally contaminated samples.
Sizoo and van Egmond [72] used liquid-liquid extractions to prepare samples for a total diet study,
focusing on aflatoxin B1 and M1 , and ochratoxin A. The samples (10 g each) were first extracted with
chloroform (100 mL), followed by filtration and evaporation of 50 mL of the extracts. The remaining
residue was purified by a liquid-liquid extraction using a mixture of methanol (10 mL), pentane (50 mL)
and PBS solution (1 mL). After phase separation, the upper layer (pentane) was re-extracted with 4 mL
of PBS solution. Combined PBS and methanol extracts were further mixed with pentane (50 mL) and
PBS solution (35 mL) for the second liquid-liquid extraction. With clear phase separation, the lower
layer was filtered and an aliquot (40 mL) of the filtrate (equivalent to 4 g test portion) was mixed with
4 mL of 0.1 M citric acid and 36 mL of water and processed using immunoaffinity columns prior to the
LC-FLD analysis.
Bourais et al. [73] developed a two-phase clean-up system to analyze aflatoxins in barley and milk
based on partitioning of aflatoxins between toluene and an in-house aqueous solvent, PBS-methanol
(1:1, v/v). Briefly, the aflatoxins were extracted from samples with methanol or chloroform. The resulting
extracts were evaporated and reconstituted in toluene, a hydrophobic solvent, followed by the
partitioning of aflatoxins to the PBS/methanol phase, while interferences remained in the organic phase.
The partitioned aflatoxins were directly determined by FLD. However, such an approach required
exhaustive optimization of partitioning conditions (e.g., combinations of different solvents at various
ratios, extraction time) for different matrices, a common drawback of liquid-liquid extraction [74].
To minimize solvent consumption of conventional liquid-liquid extraction and increase
concentration factors, liquid-liquid microextraction or dispersive liquid-liquid microextraction
(DLLME) [75] has been increasingly used for the determination of aflatoxins in dairy and oil
products [76–79]. As a miniaturization of liquid-liquid extraction, DLLME allows extracting and
preconcentrating target aflatoxins into a very small amount (microvolume) of the extraction phase.
For example, Campone [80] used a DLLME procedure for aflatoxin M1 determination in whole,
skimmed and powdered milk with LC-MS. Aflatoxin M1 was extracted and salted out by the addition
of sodium chloride and acetonitrile. After centrifugation, the acetonitrile supernatant was used as
the dispersive solvent and extractant (chloroform, 1.5 mL) and water (5 mL) were added sequentially
to perform DLLME. Hand shaking and centrifugation were used to facilitate extraction and achieve
phase separation. The developed method using DLLME for aflatoxin detection in milk gave a method
LOQ of 0.6 ng/kg, which is sufficient to measure aflatoxin M1 below the EU regulations (25–50 ng/kg).
To improve extraction efficiency and accurately control the volume of extraction solution, Sun et al. [81]
used a drop-on-demand inkjet device to inject extraction solvent (10 µL, chloroform) as ultrafine
droplets (20 µm diameter) at high frequency into wheat extracts. The extraction could be finished
within 1 min and recoveries of the aflatoxins (spiked at 0.05–0.5 µg/L) ranged from 83% to 93% with
RSDs < 5%. However, evaporation and solvent exchange had to be performed to remove chloroform
prior to the LC-MS analysis, which is a disadvantage when coupling DLLME with LC-MS.
Another interesting derivative of the liquid-liquid extraction is centrifugal partition
chromatography, in which both stationary and mobile phases are liquids and a strong centrifugal
force is used to immobilize a solvent on a stacked and interconnected rotating-discs as the stationary
Toxins 2020, 12, 539 8 of 39
phase. Just like conventional liquid chromatography, whereby solvent (the mobile phase) and sample
partition into a stationary phase, different analytes are eluted according to their partition coefficients
in the two solvents. In a recent study, Endre et al. [82] reported the separation and purification
of aflatoxins using centrifugal partition chromatography. After testing different biphasic systems,
a toluene–acetic acid–water (30:20:50, v/v/v/%) solvent system was selected to provide optimized
purification of aflatoxin B1 , B2 , G1 and G2 from an Aspergillus parasiticus fermented material in a
250 mL centrifugal partition chromatography column. Mass spectrometry was then used to confirm
the identity of the purified aflatoxins and their purity was determined to be greater than 96% using
LC-UV. Unfortunately, there was no sufficient method performance data regarding centrifugal partition
chromatography-based aflatoxin analysis for naturally contaminated samples.
While liquid-liquid extractions possess the advantage of a simple operation and ease of apparatus
handling, the methodology is often a cumbersome process requiring a large amount of organic solvent,
especially when processing multiple sample extractions. To optimize extraction conditions, one has to
exhaustively test various combinations of different solvents [83,84]. Liquid-liquid extractions are also
prone to error resulting in analyte dilution from handling large solvent volumes and reduced recovery
from the formation of sample emulsions. Due to the use of more effective clean-up (solid phase
extraction (SPE) or immunoaffinity columns) and detection (LC-FLD and LC-MS) technologies with
better sensitivity and specificity, liquid-liquid extraction has been significantly simplified and mainly
used as an extraction technique. In general, a one-step extraction is performed for aflatoxins, followed by
SPE, immunoaffinity clean-up or direct injection of the extraction [85–90].
To address insufficient sensitivity and selectivity issues of early SPE applications [91,98],
Sobolev and Dorner [99] developed an SPE clean-up procedure using basic aluminum oxide as
the adsorbent. Samples were extracted with methanol-water (80:20, v/v). An aliquot of the extract was
mixed with acetonitrile (1:1, v/v) and pipetted onto an Alltech Extract-Clean cartridge packed with
basic aluminum oxide where aflatoxins were eluted by gravity. The purified extract was injected onto
an LC system with fluorescence detection. Recoveries of aflatoxins B1 , B2 , G1 and G2 from peanuts
spiked at 5.0–2.5 mg/kg were between 80 and 87% with RSDs < 5%. The basic aluminum oxide sorbent
retained compounds of medium to high polarity in polar solvents like methanol, acetonitrile and water,
while significantly removing low polarity and nonpolar (both neutral and acidic) interferences from the
extract. Since food matrices that are prone to aflatoxin contamination contain similar components such
as water, carbohydrates, lipids and proteins, a careful selection of SPE sorbent chemistry is required to
promote the purification of aflatoxins from co-extracted food matrix components prior to downstream
analysis. Wilson and Romer [100] developed the Mycosep multifunctional clean-up column to
retain co-extracted compounds that could create interferences in aflatoxin analysis but allow targeted
aflatoxins to pass through. Test samples were extracted with acetonitrile–water (9:1, v/v) and resulting
extracts were passed through the column. Aflatoxins were analyzed by LC-FLD. The column was
tested using extracts from a variety of representative commodities such as corn, tree nuts, cereal grains,
seeds and mixed feeds. The average recoveries of total aflatoxins were greater than 95% total aflatoxins
and coefficients of variation were less than 3%, providing a selective one-step clean-up strategy for
sample preparation.
Successful developments of SPE applications using different adsorbents are also reported in
other studies. Ventura et al. [101] developed an SPE procedure for aflatoxin analysis in medicinal
herbs. Samples were extracted with methanol-water (8:2, v/v) and passed through an Oasis HLB
cartridge (packed with polymeric adsorbents) conditioned with methanol and equilibrated with water.
After washing the cartridge with methanol-water (3:7, v/v), aflatoxins were eluted with methanol.
The eluted extract was analyzed by LC-MS. Park and Kim [102] used Sep-Pak silica SPE tubes to clean
up rice samples for aflatoxin B1 analysis. Rice samples were extracted using 80% methanol and the
purified extracts were analyzed using LC-FLD. Recoveries were between 80 and 90% with RSDs < 6%.
Using SPE packed with Florisil, Sobolev [103] purified aflatoxins from corn, brown and white rice,
cottonseed, almonds, Brazil nuts, pistachios, walnuts and hazelnuts. Samples were extracted using
methanol-water (8:2, v/v), followed by an SEP clean-up step. Eluted aflatoxins were quantified by
LC-FLD. Recoveries of aflatoxins B1 , B2 , G1 and G2 were between 90 and 100% with RSDs < 5% in
tested matrices. Romero-González et al. [104] developed a multimycotoxin method, in which aflatoxins
B1 , B2 , G1 , G2 and M1 , fumonisin B1 and B2 , deoxynivalenol, ochratoxin A, HT-2 and T-2 toxins and
zearalenone were extracted from beer using C18 SPE cartridges. Key parameters such as the type of
adsorbent, elution solvents and dilution factors were evaluated. After sonication, beer samples were
directly loaded onto a C18 cartridge conditioned with a mixture of acetonitrile–methanol (6:4, v/v) and
water. Mycotoxins were eluted by the same mixture of acetonitrile–methanol and analyzed by LC-MS.
Recoveries of all targeted mycotoxins ranged from 76 to 106% with RSDs < 21%.
To simplify SPE, researchers have developed matrix solid-phase dispersion (MSPD) clean-up,
a technique that does not need cartridges but directly mixes samples and adsorbents together.
The mixture is then packed in a glass column, followed by the elution of target analytes [105].
The approach is generally applied to small test portions (e.g., 0.5–2 g) so it is better suited for
instruments with sufficient sensitivity, otherwise additional evaporation and concentration steps are
needed. Blesaa et al. [106] developed an MSPD procedure for the analysis of aflatoxin B1 , B2 , G1 and G2
in peanuts using C18 bonded silica as the dispersive adsorbents and acetonitrile as the eluting solvent.
Eluted aflatoxins were concentrated via evaporation and quantified using LC-FLD and identity of
aflatoxins were further confirmed using LC-MS. Recoveries of spiked aflatoxins at 2.5 and 5.0 ng/g
ranged between 78 and 86% with RSDs < 10%. The LOQs were 0.125–2.5 ng/g.
Toxins 2020, 12, 539 10 of 39
(11%) containing 0.5% (Tween-20® ) and water, followed by the elution of aflatoxins using methanol.
Sample eluants (in methanol) were evaporated and reconstituted in a hexane–acetone–methanol
solution for TLC analysis. This immunoaffinity column clean-up procedure successfully addressed the
insufficient elimination of matrix interferences associated with some classical clean-up procedures
such as SPE used for TLC [133,134].
Due to the co-occurrence of aflatoxins and ochratoxin A in some food (e.g., dried fruit,
spices and cocoa, nutmeg and botanical roots), it was a logical extension from single class mycotoxin
immunoaffinity columns to multimycotoxin immunoaffinity columns so that both aflatoxins and
ochratoxin A could be purified on the same column, an approach that is more efficient than using
two individual columns. Chan et al. [135] used a commercial immunoaffinity column (AflaOchraTM
HPLC immunoaffinity column) to develop a procedure that can simultaneously purify aflatoxins
and ochratoxins from maize cereals, whole corn, maize snacks, corn flour, polenta and peanut butter.
Average recoveries (within-batch) were 72–94% for ochratoxin A and 73–101% for aflatoxins spiked at
5 µg/kg. RSDr and RSDR of aflatoxins and ochratoxin A in maize cereals were 7.6–10.1% and 10.2–13.8%,
respectively. In this work, the estimated method detection limits of aflatoxins and ochratoxin A were
around 0.2 µg/kg. Trucksess et al. [131] compared the method performance using a commercial
multimycotoxin column (for both aflatoxins and ochratoxin A) and one single-mycotoxin column
(for aflatoxins only). The two types of columns were operated in a very similar manner, and aflatoxins
and ochratoxin A were measured using same LC-FLD. Results for aflatoxins (2–16 ng/g) using the single
mycotoxin IAC clean-up were comparable (80% recovery) for ginseng and other botanicals such as
ginger, liquorice and kava-kava. Using IAC clean-up for both aflatoxins and ochratoxin A, recoveries of
aflatoxins in ginseng were about 70% but around 60% in ginger, liquorice and kava-kava. However,
recoveries of spiked ochratoxin A (4 ng/g) in these botanicals were about 55%. Based on the results,
another single-laboratory study [136] and a multilaboratory collaborative study were conducted to
evaluate the performance of the multimycotoxin immunoaffinity column [137]. Similar applications of
the multimycotoxin column have been extended to other food matrices such as peanut butter, maize,
polenta, maize snacks, corn flour [135], bee pollen [138], olive oil [139] and cocoa product [140].
In recent years, multimycotoxin immunoaffinity columns have been developed and coupled with
LC-MS to simultaneously screen for aflatoxins, deoxynivalenol, ochratoxin A, fumonisins, HT-2 and
T-2 toxin and zearalenone in one sample clean-up step and instrumental analysis. Maria et al. [141],
Vaclavikova et al. [142], Wilcox et al. [143] and Ye et al. [144] demonstrated successful applications of
commercial multimycotoxin immunoaffinity columns in various foods. Although the immunoaffinity
columns could achieve simultaneous clean-up of multiple mycotoxins, it has been a challenge to
achieve satisfactory extraction efficiency of all targeted mycotoxins with different physicochemical
properties in a single extraction, where the use of multiple extractions would defeat the purpose of
a multimycotoxin analysis. Another practical issue with multimycotoxin immunoaffinity columns
is the capacity of mycotoxins purification. Clean-up of more mycotoxins on one column means less
flexibility and more stringent column conditions without comprising of clean-up efficiency for multiple
mycotoxins. The complexity and cost of column production increases with the number of different
types of antibodies that need to be immobilized. Furthermore, it was expected that immunoaffinity
columns could provide a clean extract with marginal matrix effects on LC-MS. Unfortunately, that is
not the case and making matrix-match calibration standards using immunoaffinity columns is not a
practical or economical practice. Therefore, in the future, study efforts may be dedicated to address
these challenges to better utilize immunoaffinity column clean-up and LC-MS for aflatoxin analysis.
A noteworthy attempt associated with immunoaffinity columns is to automate the clean-up
procedure using robotic devices. Sharman and Gilbert [145,146], Urano et al. [147] and
Carman et al. [148] developed and evaluated in-house automated immunoaffinity column clean-up
devices. In terms of clean-up efficiency and sample throughput, their results were comparable to
manual procedures. These early devices could replace human operators to perform sample clean-up
to some extent, but the degree of automation and robustness was low. These devices lacked the
Toxins 2020, 12, 539 13 of 39
to co-extraction and additional steps are required to clean-up the resulting extracts due to the
non-selective and aggressive nature of these extraction techniques. It is not surprising that only a
handful of applications displaying microwave-assisted extraction and accelerated solvent extraction
have been developed for aflatoxin analysis [163–166]. Compared to conventional liquid extraction,
these technologies demonstrate limited advantages other than efficiency. The cost and maintenance of
the instruments for accelerated solvent extractions and microwave-assisted extractions are significantly
higher than conventional lab glassware used for liquid extraction. Unless better selectivity or
online sample clean-up features could be incorporated with these energy-assisted extraction devices,
they would remain as research-oriented devices but adopted for the routine aflatoxin analysis.
3. Chromatography
and further confirmation tests are required, especially for compliance samples. Andrellos and Reid [180]
developed a confirmation procedure based on the fluorescent reaction products of aflatoxin B1 with
formic acid-thionyl chloride, acetic acid-thionyl chloride and trifluoroacetic acid. The procedure
required a large on plate mass of aflatoxin B1 , yet it did not work for other aflatoxins. Attempts to couple
TLC with mass spectrometry for more specific identification were reported by Haddon et al. [181,182],
Brumley et al. [183], Park et al. [184] and Tripathi et al. [185]. However, aflatoxins were unable to be
directly analyzed by MS requiring extensive clean-up steps to remove aflatoxins from the plate and
purified prior to MS analysis. These procedures were time consuming and instrument setups were
cumbersome and expensive.
In terms of quantitation, TLC visual estimation cannot generate accurate quantitative results.
A collaborative study also reported large variability among participating labs using TLC [186]. In some
cases, the measured errors could go up to 56% and precisions ranged from 20 to 28% [187,188].
The development of fluorodensitometric measurements of TLC spots can improve the quantitative
analysis of aflatoxins in terms of precision and accuracy [189], however, such experiments sacrifice
throughput because only one spot can be scanned at a time. One of the advantages of TLC is that
multiple samples could be analyzed in a parallel manner, while even modern LC-MS has to run samples
sequentially. Other issues related to TLC such as variability in sample spotting, silica gel adsorption
activity and silica gel particle size also lead to unreliable results. Furthermore, temperature and
humidity in the lab could affect TLC performance and aflatoxins could be photochemically degraded
under excess UV lights [190,191]. It is not surprising that these inherent disadvantages of TLC gradually
contributed to the replacement of TLC with LC-FLD and other advanced techniques.
quench the fluorescence of aflatoxin B1 and B2 . The method could detect aflatoxins in foods as low
as 10–20 ppb. Alternatively, Zimmerli [198] and Panalaks and Scott [199] used a flow cell of the
fluorometric detector filled with silica-gel particles to extract aflatoxins from the mobile phase after LC
separation. In this work, FLD detection had the same physical characteristics as that on a TLC plate.
Aflatoxins were absorbed and detected on a silica-gel particles solvent, thereby minimizing quenching
of fluorescence. Although the addition of the silica gel particles into the flow cell slightly impacted
chromatography (e.g., peak broadening), the approach was commercialized and widely integrated
with LC-FLD as an indispensable feature for aflatoxin analysis in foods [69,200–202].
In the 1980s, reversed phase LC rapidly replaced normal phase LC for aflatoxin analysis.
LC separation of aflatoxins on reversed phase columns was straightforward. Generally, mixtures of
water, methanol and/or acetonitrile were used as the mobile phase, promoting aflatoxin separation
using isocratic conditions. Fluorescence quenching in mobile phases remained a challenge for the
detection of aflatoxins at sub ppb levels in foods. Thee fluorescence emission of aflatoxin B1 and G1
could be significantly quenched by aqueous mobile phases utilized in reverse phase LC. To overcome
this limitation, a common practice is to derivatize aflatoxins, minimizing the impact of signal quenching.
Either precolumn derivatization with trifluoroacetic acid (TFA) or post-column derivatization with a
halogen (bromine or iodine), are commonly employed, adding one hydroxyl group or two atoms of the
halogen at the 8,9-double bond of the dihydrofuran moiety of aflatoxins. Alternatively, post-column
photochemical derivatization can be used for aflatoxin analysis.
The findings that aflatoxin B1 can react with TFA, forming a fluorescing product, aflatoxin B2 a in
aqueous solvents [203,204] paved the way for TFA-based derivatization, which was mainly used prior
to LC separation. Pons et al. [205] thoroughly investigated the hydration of aflatoxin B1 and G1 in acidic
aqueous solutions in terms of pH, reaction times/rates and product stability. The study confirmed
that the double bonds in the dihydrofuran moieties of the aflatoxins B1 and G1 could be hydrated,
generating reaction products with sufficient fluorescence for detection. However, the study also
demonstrated these products were not stable and gradually degraded into non-fluorescing compounds.
Takahashi [188] developed and optimized a TFA-based derivatization for the analysis of aflatoxins
in wine. Aflatoxins were extracted from spiked wine using methylene chloride and the concentrated
sample extracts were cleaned up on a silica gel-alumina column. After evaporation, the remaining
extracts were reacted with TFA for a few seconds, followed by the addition of 10% acetonitrile and
LC-FLD analysis. Derivatization products, aflatoxin B2a , G2a and two native aflatoxins, aflatoxin B2
and G2 were separated on a 25 cm × 3.2 mm I.D., 10 µm ODS-bonded silica gel column within 15 min.
A mixture of methanol and water was used as the mobile phase. Recoveries of the four aflatoxins
spiked at 1 ppb ranged from 88%-106%. The method could detect aflatoxins in wine as low as 0.02 µg/L,
8× more sensitive than TLC [206]. The procedure has been extended and modified for different food
matrices [207,208]. For example, using the same derivatization procedures, Beebe [209] developed an
LC-FLD method for aflatoxins analysis in green coffee and peanut butter. Recoveries of aflatoxins
spiked at 1–15 ppb ranged from 72 to 104%; Diebold and Zare [210] replaced TFA with hydrochloric
acid for derivatization and achieved the detection of aflatoxins at subpicogram levels in corn using a
laser fluorescence detector; Beebe and Takahashi [211] derivatized aflatoxin M1 extracts from milk and
dairy products using TFA, followed by an LC-FLD analysis. Aflatoxin M1 could be detected at as low as
0.3 ppb and recoveries were 93–125% at 1 ppb in spiked matrices. An AOAC official method was also
developed for aflatoxin analysis in corn and peanut butter using TFA precolumn derivatization, a C18
analytical column and water–methanol–acetonitrile as the mobile phase [212]. TFA derivatization
was companied by considerable disadvantages. These included peak broadening caused by residual
TFA and instability of the derivatization products, aflatoxin B2 a and G2 a in methanol [205,208,213].
Additionally, TFA derivatization had to be conducted offline prior to LC separation, making it less
desirable compared to post-column derivatization, which could be integrated with LC-FLD as an
online operation.
Toxins 2020, 12, 539 17 of 39
Similar to precolumn derivatization, post-column derivatization started with the finding that
iodination of aflatoxin B1 and G1 exhibit strong fluorescence [214]. Tuinstra and Haasnoot [215]
developed a prototype post-column derivatization device, in which an iodine solution was pumped
from a reservoir and mixed with eluted aflatoxins in a reaction coil (3000 mm × 0.5 mm, 60 ◦ C)
inserted between the LC column and FLD. This led to a 50× increase in sensitivity of aflatoxins.
Encouraged by the promising results, researchers conducted a series of studies to optimize this
post-column derivatization method, focusing on compositions of the mobile phase, iodine reagent
flowrates, reaction temperatures, iodine concentrations, confirmation of derivatization products using
MS, etc. [216–223].
Despite the wide acceptance of iodine-based post column derivatization, Kok and his
colleagues [224–226] developed another online post-column derivatization based on bromination of
aflatoxins to simplify derivatization steps and lower instrument cost. In their approach, bromine is
generated from bromide (e.g., potassium bromide) added into the mobile phase and reacted with eluted
aflatoxins in a post-column electrochemical cell, namely KOBRA-cell, prior to FLD analysis. The bromine
concentration can be carefully controlled by the generating current, maximizing fluorescence intensity.
Using this approach, aflatoxins could be detected at an on-column mass at 20–40 pg.
Other valuable signal-enhancement techniques were also developed [227–230]. Garner et al. [227]
developed a more cost effective bromination method using pyridinium bromide perbromide as the
derivatization agent. To further increase sensitivity, Chiavaro et al. [228] and Maragos et al. [229]
evaluated the addition of cyclodextrins as a fluorescence enhancer. Joshua [230] reported the first
photochemical derivatization method of aflatoxins. Compared to chemical derivatization, this approach
does not need the addition of any chemical reagents. Aflatoxins were derivatized using UV generated
by a photochemical reactor and could be detected in corn below 1 ppb. In recent years, there have
been notable efforts to evaluate and compare photochemical derivatization to chemical derivatization.
Studies by Papadopoulou-Bouraoui et al. [231] and Waltking et al. [232] demonstrate photochemical
derivatization is as efficient as chemical derivatization using bromine or iodine.
With the advent of commercial UHPLC, there are some applications of UHPLC-FLD for aflatoxin
analysis. Due to the use of columns packed with submicron (<2.0 µm) particles, rapid and efficient
separations could be achieved. The speed and high resolution of UHPLC made it an attractive tool
for analyzing mycotoxins in foods [233,234]. Ibanez et al. [235] developed an UHPLC-FLD method to
simultaneously determine aflatoxins, OTA and zearalenone in barley. Photochemical derivatization was
performed to enhance signal intensity of aflatoxins without impacting ochratoxin A and zearalenone.
Iodination or bromination, however, is not suited for multimycotoxin analyses as ochratoxin A and
zearalenone are unable to survive the derivatization process. The author also mentioned that the
impact of the void volume (0.25 mL) introduced by the photochemical coil on chromatography should
be monitored, though it is not an issue for conventional columns with bigger particle sizes.
Addel-Azeem et al. [236] used UHPLC-FLD to screen aflatoxins in food samples collected from
local markets. Without any pre- or post-column derivatization, the method could detect individual
aflatoxins as low as 0.02 ppb. An isocratic program was used to separate aflatoxins on a C18 column
(2.1 mm × 100 mm, 1.7 µm). Their results suggest there were significant retention time shits of aflatoxins
in solvent and sample matrices. Interestingly, the method seems more sensitive for aflatoxin B1 and
B2 than aflatoxin G1 and G2 . Banerjee and his colleagues [41,237,238] extended UHPLC-FLD-based
aflatoxin analysis to both food and feed matrices. Using a FLD flowcell (13 µL) and a large injection
volume (10 µL), these methods achieve the detection of aflatoxins below 1 ppb even in feed matrices
without performing conventional derivatization. Due to its sensitivity capabilities and ease of use,
these newly developed LC-FLD methods continue to attract mycotoxin labs to adopt them for routine
sample analysis.
Toxins 2020, 12, 539 18 of 39
cost [256].
Dilution seems to be a convenient approach to mitigate matrix effects, but it requires sensitive
LC-MS [257] and dilution factors are matrix-, instrument- and analyte-dependent. Additionally,
regulatory levels of aflatoxins should not be overlooked. For example, the current FDA action
level of aflatoxin M1 is 0.5 ppb in milk. If a 10× dilution is necessary to eliminate matrix effects,
the LC-MS instrument must be able to quantitate the aflatoxin below 0.05 ppb [89]. When factoring
in EU’s regulatory levels of aflatoxin M1 (0.025–0.05 ppb), more sensitivity is required if samples
are diluted prior to LC-MS analysis. While such sensitivity is not possible for the majority of
commercial LC-MS instruments, concentration and clean-up steps are necessary to ensure the detection
of aflatoxin M1 at and below the regulatory levels [258,259]. In-depth sample clean-up could be used
to eliminate matrix-components/interferences but very often the process is lengthy and laborious.
Sometime, even with immunoaffinity columns, LC-MS analysis was not free from matrix effects [142].
Matrix-matched calibration standards are commonly used to achieve satisfactory quantitative results.
In this approach, known amounts of aflatoxins are spiked into “clean” matrix and used as calibration
standards. This is expected to bring same/similar impact of matrix effects on the response of
aflatoxins in both calibration standards and samples. The preparation of matrix-matched calibration
standards is tedious, especially when one needs to deal with a wide variety of food matrices [57,260].
Standard addition would be straightforward only if one analyte needs to be quantitated. When using
the approach to quantitate the four aflatoxins at various concentrations in a sample, it becomes a
challenging operation for analysts. The ideal solution for matrix effects would be stable isotope
dilution assay (SIDA). Using commercially available 13 C-aflatoxins or deuterated aflatoxins as internal
standards (IS), one could quantitate native aflatoxins with solvent only calibration standards [256].
Very often extraction solvents and cleanup procedures have to be optimized depending on the matrix
to ensure extraction efficiency. The incorporation of labeled IS into the sample prior to extraction and
clean-up provides greater flexibility in sample extraction conditions across multiple sample matrices.
Vahl and Jørgensen [261] developed an LC-MS method for the determination of aflatoxins B1 , B2 ,
G1 and G2 in food and interestingly, they used aflatoxin M1 as the internal standard. Samples were
extracted with methanol and further cleaned by column chromatography, previously developed for
LC-LFD. On the LC-MS, recoveries for the aflatoxins spiked at 2 ppb in figs, peanut butter, peanuts,
chili, pepper and curry ranged between 40 and 270%. Variability of individual aflatoxins was also
unusually high. This study is a reminder of the challenges associated with LC-MS, especially matrix
effects, demonstrating how clean-up procedures that work well for LC-FLD might not work for LC-MS.
Cavaliere et al. [262] used aflatoxin M1 as IS for the determination of aflatoxins in corn. Samples were
extracted using acetonitrile–water (80:20, v/v), followed by dilution and SPE cleanup. Aflatoxins were
detected by LC-MS/MS. Quantitation was achieved by matrix-matched standards. Recoveries ranged
from 81 to 101% with RSDs < 12% and the method LODs were from 0.1 to 0.6 ppb. Using internal
Toxins 2020, 12, 539 20 of 39
standards is a good practice, but whether the selected internal standard would behave similarly to
target analytes in different matrices should be carefully evaluated. Without an appropriate internal
standard or matrix-matched calibration standard, one would not be able to effectively minimize the
impact of matrix effects on quantitation.
Cervino et al. [263] synthesized and characterized deuterated aflatoxins B2 and G2 and used them
to compensate the loss of analytes in the course of extraction and matrix effects for LC-MS-based
aflatoxin analysis. The two deuterated IS were spiked into samples (almonds and wheat flour) prior to
extraction so excellent recoveries and precisions were achieved. Recoveries were between 90% and
105% for all the four aflatoxins with coefficients of variation (CV) of 3.6%, 4.8%, 12% and 14% for
aflatoxin B1 , B2 , G1 and G2 , respectively. As expected, labeled B2 and G2 were more suited for the
quantitation of aflatoxin B2 and G2 than B1 and G1 due to similar physicochemical properties between
the labeled IS and the native aflatoxins (B2 and G2 ) and retention times on LC.
Li et al. [264] developed an isotope dilution LC-MS method for the determination of aflatoxins
in feeds. Samples were extracted using methanol-water (70:30, v/v) or acetonitrile–water (84:16, v/v)
depending on feed matrices, followed by filtration and dilution. Prior to LC-MS analysis, the extracts
were spiked with 13 C-aflatoxin B1 and aflatoxins were separated on a C18 (2.1 mm × 50 mm, 1.7 µm)
column. Detection was performed using a triple quadruple MS. Identification of aflatoxins was
achieved by comparing the retention times and relative ion ratios of the two MRM transitions in
standards and samples [265]. Quantitation was based on linear least squares calibration of the relative
response ratio of an aflatoxin and 13 C-aflatoxin B1 plotted versus concentration. Recoveries of aflatoxins
range from 78 to 122% with intra-day and inter-day precisions of 2–15% and 3–17%, respectively. In this
study due to the concern of cost, only one 13 C-aflatoxin was used and spiked into the sample extracts
right before LC-MS analysis. Had 13 C-IS been spiked into the samples prior to extraction, probably
only one extraction solvent would have been used. As to the cost of 13 C-IS, people often overlook the
savings in labor cost and increase in sample throughput [255].
To take advantage of the sensitivity of LC-MS, Xavier and Scussel [266] used a “dilute-and-shoot”
procedure to prepare samples for aflatoxin analysis in Brazil nut. Homogenized samples (25 g) were
first extracted using 100 mL of acetonitrile–water (80:20, v/v) and further diluted 4× using water prior
to LC-MS analysis. Matrix-matched calibration standards were used for quantitation. Recoveries of
spiked aflatoxins at 1–10 ng/g ranged between 92 and 100% with RSDs < 10%. The method combined a
simple sample preparation procedure and an effective solution to matrix effects together in a practical
manner. Similar approaches have been reported in other studies [267,268]. Breidbach and Ulberth [269]
developed a highly accurate quantitation method for aflatoxin B1 using 2D-LC and double isotope
dilution MS. Aflatoxin B1 was carefully separated from matrix-components using heart-cutting 2D-LC
and measurement uncertainty related to extraction efficiency and matrix effects was minimized
using 13 C-IS. In tested matrices (cereal-based baby food, maize and maize-based feed), the expanded
uncertainty of aflatoxin B1 at 0.197 ppb is 0.017 µg/kg (8.9%). The application of 2D-LC-MS might
not be suited for routine sample analysis but is appreciated for measurements that require high
accuracy such as characterization of certified reference materials. Recently, Campone et al. [270]
reported determination of aflatoxins, ochratoxin A, fumonisins B1 and B2 in beer using 2D-LC-MS.
For fumonisins and ochratoxin A, the 2D-LC separation eliminated co-eluted interfering compounds,
permitting quantification by solvent-only calibration standards. In contrast, matrix suppression was
still observed for aflatoxins, requiring matrix matched calibration standards for quantitation.
Co-contamination of aflatoxins and other mycotoxins in foods and feeds is well known. How to
efficiently monitor co-occurrence of aflatoxins and other mycotoxins has been a challenge for mycotoxin
researchers [271–273]. When multiple mycotoxins are needed to be analyzed, it is clear that LC-MS
would be a better analytical tool as it can achieve identification, quantitation and confirmation for
multiple mycotoxins in one analysis, a desirable feature that other techniques lack. For example,
the survey work performed by Lombaert et al. and Zhang et al. report mycotoxins in infant
foods from the retail market, and provide a good comparison of the application of LC-MS-based
Toxins 2020, 12, 539 21 of 39
multimycotoxin analysis vs. LC-FLD-based single mycotoxin analysis [274,275]. Without LC-MS,
multiple LC-FLD methods had to be used to collect information of individual mycotoxins such as
aflatoxins, deoxynivalenol and ochratoxin A, while using LC-MS, these mycotoxins could be identified
and quantitated by one method. For multimycotoxin analysis, using multiple single mycotoxin
methods would be much less practical than using a multimycotoxin method. As LC-MS could
simultaneously detect a large number of mycotoxins in one analysis, the current trend is to develop
multimycotoxin procedures, taking advantage of the analytical capacity of LC-MS, hence improving
throughput. A variety of LC-MS-based multimycotoxin methods have been developed. In these
methods, aflatoxins and other mycotoxins of interest are extracted from samples, followed by cleanup
procedures, if necessary, and then analyzed by LC-MS.
In a series of studies [276–281], the number of mycotoxins that could be simultaneously detected
in one LC-QQQ-MS analysis has been expanded from 39 to 300. Each mycotoxin is monitored by two
unique MRM transitions and retention time, which is sufficient to achieve confident identification [282].
It is evident that for multimycotoxin analyses, the capacity of conventional TLC or LC-UV/FLD is
unrivaled to modern LC-QQQ technologies. With increasing number of mycotoxins, researchers need
to find appropriate extraction and sample clean up procedures for target mycotoxins with different
physicochemical properties. Practically, it is also a QA/QC and logistical challenge to trace, store and
prepare calibration standards for a large number of analytes.
Spanjer et al. [43] extracted 33 mycotoxins including aflatoxins in various products (peanuts,
pistachios, wheat, maize, cornflakes, raisins and figs) using acetonitrile–water (80:20, v/v) and
methanol-water (70:30, v/v), followed by dilution with water and LC-MS/MS analysis. Although
matrix-matched calibration standards were prepared for quantitation, the practice is considered less
laborious than using SPE or immunoaffinity columns for cleanup. Compared to LC-FLD methods,
the LC-MS/MS method not only generated comparable results concerning aflatoxin B1 in FAPAs
proficiency test samples but also quantitated other mycotoxins in one analysis. Mol et al. [283]
proposed and evaluated a generic extraction approach using water-acetonitrile-1% formic acid as an
extraction solvent to analyze 136 pesticides, 36 natural toxins (including aflatoxins) and 86 veterinary
drugs in foods and feeds. After extraction, no further clean-up was performed, extracts were directly
injected into an LC-QQQ. Recoveries for 80% of the analytes were between 70% and 120% with
RSDs < 10%. Matrix effects were minimized by using matrix-matched calibration standards and
a small injection volume, 5 µL. It is clear that with reasonable performance criteria, LC-MS-based
multianalyte methods could significantly improve operation efficiency, replacing many matrix- and
compound-dependent single mycotoxin methods [241,284,285].
The intended use of multimycotoxin methods should be clearly defined so that a practical balance
between method performance and number of target mycotoxins could be reached. For example,
Zhang et al. [55] developed a stable isotope dilution LC-MS/MS method to analyze aflatoxins,
fumonisins, ochratoxin A, deoxynivalenol, HT-2/T-2 toxin and zearalenone in foods. As the method
was intended for the FDA regulatory program, only FDA regulated mycotoxins (e.g., aflatoxins) and a
few mycotoxins of health significance (zearalenone, HT-2/T-2 toxins) are included. Each mycotoxin
was quantitated using a 13 C-IS and the method could identify and quantitate these mycotoxins at
concentrations ranging from 1.0 to 1000 ng/g in peanut butter, wheat flour and corn. The majority of
recoveries ranged from 80 to 120% with RSDs < 20%. Greater than 90% of the average recoveries were
in the range of 90–110%, with repeatability RSDr (within laboratory) < 10% and reproducibility RSDR
(among laboratory) < 15%. Although such an approach would not be realistic for the quantification of
191 mycotoxins [279] or 295 bacterial and fungal metabolites in foods [284], a similar platform may be
utilized for screening mycotoxin occurrence in different matrices.
To date, the majority of LC-MS-based mycotoxin methods have been developed using LC-QQQ,
however, LC–high resolution mass spectrometry (HRMS; e.g., LC-Orbitrap-MS and LC-Q-time of
flight-MS) has gradually attracted attention [285,286]. LC–HRMS offers the acquisition of high
resolution full scan data with accurate m/z and the option of targeted or non-targeted sample acquisition
Toxins 2020, 12, 539 22 of 39
with retrospective data analysis. Although it is still a niche application, LC–HRMS has been evolving
rapidly in the field of mycotoxin analysis and provides a powerful tool to screen for aflatoxins and
other mycotoxins in food and feed products. Early HRMS applications were stymied by the lack of
impactful understanding and agreement among mycotoxin researchers regarding the role of LC–HRMS
in mycotoxin analysis. The lack of software tools and identification criteria created severe technical
impediments to the application of LC–HRMS, but rapid improvements of hardware (e.g., sensitivity,
resolution power and data acquisition rate), data mining software and data storage systems supported
by identification criteria guidance will likely facilitate the implementation of LC–HRMS moving
forward [282,287,288].
Like LC-QQQ, LC–HRMS, especially LC-Orbitrap and LC-Q-Orbitrap MS is commonly used
for multimycotoxin and multianalyte analysis. Due to limited use of LC-TOF in mycotoxin research,
very few studies use LC-TOF or LC-Q-TOF for aflatoxin analysis. Toth et al. [289] used a LC-Q-TOF
and studied fragmentation patterns of aflatoxins under different collision energies. No quantitative
analysis was performed. In an early LC–HRMS study, Vaclavik et al. [290] coupled a direct analysis
in real time (DART) ionization source with an Orbitrap MS for the quantitative analysis of multiple
mycotoxins in wheat and maize. Sample extracts were directly analyzed by DART-Orbitrap MS
without LC separation. The impact of resolution power (10,000–50,000 full width at half maximum,
FWHM, at m/z 200) on quantitation and mass accuracy was evaluated. For aflatoxins, the sensitivity of
the DART method was poor as compared to LC-QQQ. Lattanzio et al. [291] developed an LC–HRMS
method for the determination of EU regulated mycotoxins (aflatoxins, ochratoxin A, fumonisins,
deoxynivalenol and zearalenone) in cereals. A single stage LC-Orbitrap was operated in full-scan
mode to monitor each mycotoxin by its molecular ion and fragments generated using high energy
collision dissociation (HCD). At a resolving power of 100,000 FWHM, mass accuracy of targeted
mycotoxins ranged from 0.1 to 3.9 ppm. In terms of sensitivity (e.g., LODs), recoveries and repeatability,
the method offered analogous performance as compared to a previously validated LC-QQQ method.
Later, Liao et al. [57,262] compared quantitation on an LC-Q-Orbitrap operated in full scan and
data-dependent acquisition mode, to an LC-QQQ and concluded that both LC-MS methods had
sufficient sensitivity to identify and quantitate multiple mycotoxins, albeit different identification
criteria were used. Using an LC-Q-Orbitrap MS, Renaud and Sumarah [292] developed an LC-data
independent acquisition method. In this approach, both quantitative and qualitative analyses could
be conducted using collected data. For mycotoxins including aflatoxins with available standards,
they were identified and quantitated similar to conventional LC-MS analyses. Using a similar approach,
Castaldo et al. [293] developed an LC-Q-Orbitrap method, which combined target analysis and
non-target screening of mycotoxins in pet food. With better resolution power, data acquisition rates
and accurate mass spectra libraries, this approach is considered as a more advanced and efficient
method for mycotoxin analysis.
This review focuses on aflatoxins but not multimycotoxin analysis, so the above discussion aims
to use a few representative studies to illustrate major trends and challenges concerning LC-MS based
mycotoxin analysis and the impact of LC-MS on aflatoxin analysis. In summary, the introduction
of LC-QQQ-MS has transformed the conventional single/single class-mycotoxin-based approach to
multimycotoxin identification and quantitation. LC–HRMS enables both targeted and non-target data
analysis, leading to a new mycotoxin testing paradigm. As aflatoxin analysis is a multifaced challenge,
existing sampling plans, sample preparation techniques and method performance criteria (validation)
should be carefully evaluated and revised if necessary so that the potential of LC-MS-based technologies
could be fully reached in the coming years to increase the sensitivity, accuracy, specificity and efficiency
of aflatoxin analysis. Noteworthy, rapid assays (e.g., ELISA) and LC-FLD methods are expected to
compete with LC-MS, as there are substantial economic interests associated with mycotoxin analysis in
the industrial market [286], stymying the existing efforts to increase the use of LC-MS that has been
gaining increasing popularity among mycotoxin laboratories. Several studies compared LC-MS/MS to
other analytical techniques used for aflatoxin analysis, suggesting LC-MS/MS generated comparable
Toxins 2020, 12, 539 23 of 39
results compared to ELISA and LC-FLD in terms of precision and accuracy [294–296]. The choice
of a method depends on many factors, including but not limited to data quality of quantitation.
If mycotoxin molecular identification is required, LC-MS has been shown to be the method of choice.
Jiang et al. [308] used integrated plastic microfluidic devices with ESI-MS to separate and detect
aflatoxins. In the study, an in-house microfabricated affinity dialysis and concentration system was
developed. Specific detection of aflatoxin B1 was achieved by optimizing the molar binding ratios of
aflatoxin B1 antibody to each aflatoxin and a flow rate around 100 nL/min. The study shows potential
of miniaturization of CE but follow-up studies are needed to demonstrate its applicability in foods for
aflatoxin analysis. Furthermore, when CE is coupled with MS, cost and complexity of the system is no
less than that of LC-MS. Between a well-established technique (LC-MS) and an analytical technique with
certain potentials (CE-MS) for aflatoxin analysis, preference is most commonly given to LC-MS-based
technologies. It is expected that CE especially CE microchips will remain as a research-oriented
technique in the near future.
4. Conclusions
Aflatoxin contamination will remain a global public health concern yet challenges that each
country/region faces are very diverse. Global trading, climate changes and diverse regulatory policies
have further complicated the issue. Therefore, the need for analytical tools to combat mycotoxin
contamination varies, prompting the development of broad ranges of chromatographic and mass
spectrometric methods for the determination of aflatoxins. This review highlights trends and novel
techniques in the course of the history of aflatoxin methodology, with emphasis on applications
of sample preparation, chromatography, and mass spectrometry to food analysis. Aflatoxins are
heterogeneously distributed in foods and feeds so each step including sampling, sample preparation,
extraction, clean-up and instrumental determination plays a crucial role in the analysis of aflatoxins.
The trend is obvious any new techniques can outperform existing ones in terms of specificity, sensitivity
and throughput will be appreciated. Following the transition from TLC, LC-UV/FLD to LC-MS,
advances in detection technologies are complimented by the simplification of sample preparation,
transforming multistep extraction/clean-up to one step “dilute-and-shoot”. Similarly, identification has
shifted from non-specific properties (e.g., visual inspection, UV adsorption or fluorescence emission)
to intrinsic properties (e.g., m/z, ion ratios, molecular fragmentation) and traditional single-/single-class
mycotoxin analysis has been gradually replaced by multimycotoxin analysis using LC-MS. In the
future development of analytical methods, these are the areas where research will focus on. Our ability
to replace and upgrade current techniques in these areas will, to a great measure, determine our success
to monitor aflatoxins in foods.
Author Contributions: K.Z. and K.B. proposed the scope and structure of the review, performed literature search
and wrote the review. All authors have read and agreed to the published version of the manuscript.
Funding: This review received no external funding.
Acknowledgments: We are extremely grateful to the FDA Library staff for their excellence assistance in providing
references (journal articles and book chapters). During the COVID-19 pandemic, their professionalism and
support are very much appreciated. We thank Christine Parker at FDA, Center for Food Safety and Applied
Nutrition for proofreading the manuscript and providing helpful comments. We also thank Manisha Dhanshetty,
M.Sc., Ankita Lakade, PhD at National Reference Laboratory, ICAR-National Research Centre for Grapes for
their assistance.
Conflicts of Interest: The authors declare no conflict of interest.
References
1. Jelinek, C.F.; Pohland, A.E.; Wood, G.E. Worldwide occurrence of mycotoxins in foods and feeds—An update.
J. Assoc. Off. Anal. Chem. 1989, 72, 223–230. [CrossRef] [PubMed]
2. Payne, G.A.; Brown, M.P. Genetics and physiology of aflatoxin biosynthesis. Annu. Rev. Phytopathol. 1998,
36, 329–362. [CrossRef] [PubMed]
3. Joint FAO/WHO Expert Committee on Food Additives (JECFA). Safety evaluation of certain food additives
and contaminants. In WHO Technical Report Series, Proceedings of the Prepared by the 55th JECFA Meeting,
Geneva, Switzerland, 6–15 June 2000; WHO: Geneva, Switzerland, 2001.
Toxins 2020, 12, 539 25 of 39
28. European Commission. Commission Regulation (EC) No. 401/2006 of 23 February 2006. Laying down the
methods of sampling and analysis for the official control of the levels of mycotoxins in foodstuffs. Off. J.
Eur. Union 2006, 70, 12–34.
29. Food and Drug Administration (FDA). Investigations Operations Manual 2017 6-Aflatoxin Sample Sizes.
Available online: https://www.fda.gov/media/97223/download (accessed on 18 April 2020).
30. Shotwell, O.L.; Goulden, M.L.; Hesseltine, C.W. Aflatoxin: Distribution in contaminated corn. Cereal Chem.
1974, 51, 492–499.
31. Shotwell, O.L.; Goulden, M.L.; Bothast, R.J.; Hesseltine, C.W. Mycotoxins in hot spots in grains I. aflatoxins
and zearalenone occurrence in stored corn. Cereal Chem. 1975, 52, 687–697.
32. Castells, M.; Marin, S.; Sanchis, V.; Ramos, A.J. Distribution of fumonisins and aflatoxins in corn fractions
during industrial cornflake processing. Int. J. Food Microbiol. 2008, 123, 81–87. [CrossRef] [PubMed]
33. Association of Official Analytical Chemists (AOAC). AOAC Official Method 977.16 Sampling of Aflatoxins,
Preparation of samples. In Official Methods of Analysis of AOAC International, 17th ed.; AOAC International:
Rockville, MD, USA, 2000.
34. Francis, O.J.; Ware, G.M.; Carman, A.S.; Kirschenheuter, G.P.; Kuan, S.S. Use of Ten Gram Samples of Corn
for Determination of Mycotoxins. J. Assoc. Off. Anal. Chem. 1988, 71, 41–43. [CrossRef]
35. Whitaker, T.B.; Dickens, J.W.; Monroe, R.J. Variability associated with testing corn for aflatoxin. J. Am. Oil
Chem. Soc. 1979, 56, 789–794. [CrossRef]
36. Velasco, J.; Morris, S.L. Use of water slurries in aflatoxin analysis. J. Agric. Food Chem. 1976, 24, 86–88.
[CrossRef]
37. Trucksess, M.; Scott, P.M. Mycotoxins in botanicals and dried fruits: A review. Food Addit. Contam. Part A
2008, 25, 181–192. [CrossRef]
38. Asghar, M.A.; Iqbal, J.; Ahmed, A.; Khan, M.A. Occurrence of Aflatoxins Contamination in Brown Rice from
Pakistan. Iran. J. Public Health 2014, 43, 291–299.
39. Ozkan, A.; Bindak, R.; Erkmen, O. Aflatoxin B 1 and aflatoxins in ground red chilli pepper after drying.
Food Addit. Contam. Part B 2015, 8, 227–233. [CrossRef]
40. Iqbal, S.Z.; Asi, M.R.; Selamat, J. Aflatoxins in dates and dates products. Food Control. 2014, 43, 163–166.
[CrossRef]
41. Dhanshetty, M.; Banerjee, K. Simultaneous Direct Analysis of aflatoxins and ochratoxin A in cereals and
their processed products by ultra-high-performance liquid chromatography with fluorescence detection.
J. AOAC Int. 2019, 102, 1666–1672. [CrossRef] [PubMed]
42. Zahra, N.; Khan, M.; Mehmood, Z.; Saeed, M.K.; Kalim, I.; Ahmad, I.; Malik, K.A. Determination of Aflatoxins
in Spices and Dried Fruits. J. Sci. Res. 2018, 10, 315–321. [CrossRef]
43. Spanjer, M.C.; Rensen, P.M.; Scholten, J.M. LC-MS/MS multi-method for mycotoxins after single extraction,
with validation data for peanut, pistachio, wheat, maize, cornflakes, raisins and figs. Food Addit. Contam.
Part A 2008, 25, 472–489. [CrossRef] [PubMed]
44. Francis, O.J. Sample Preparation of Some Shelled Treenuts and Peanuts in a Vertical Cutter-Mixer for
Mycotoxin Analysis. J. Assoc. Off. Anal. Chem. 1979, 62, 1182–1185. [CrossRef] [PubMed]
45. Davis, N.D.; Dickens, J.W.; Freie, R.L.; Hamilton, P.B.; Shotwell, O.L.; Wyllie, T.D.; Fulkerson, J.F. Protocols for
Surveys, Sampling, Post-Collection Handling, and Analysis of Grain Samples Involved in Mycotoxin
Problems. J. Assoc. Off. Anal. Chem. 1980, 63, 95–102. [CrossRef] [PubMed]
46. Campbell, A.D.; Whitaker, T.B.; Pohland, A.E.; Dickens, J.W.; Park, D.L. Sampling, sample preparation,
and sampling plans for foodstuffs for mycotoxin analysis. Pure Appl. Chem. 1986, 58, 305–314. [CrossRef]
47. Patil, U.S.; King, S.; Holleran, S.; White, K.; Stephenson, C.; Reuther, J. Identifying Challenges and Risks
Associated with the Analysis of Major Mycotoxins in Feed and Botanicals. J. AOAC Int. 2019, 102, 1689–1694.
[CrossRef]
48. Spanjer, M.C.; Scholten, J.M.; Kastrup, S.; Jörissen, U.; Schatzki, T.F.; Toyofuku, N. Sample comminution for
mycotoxin analysis: Dry milling or slurry mixing? Food Addit. Contam. 2006, 23, 73–83. [CrossRef]
49. Whitaker, T.B.; Dickens, J.W.; Giesbrecht, F.G. Effects of Methanol Concentration and Solvent: Peanut Ratio
on Extraction of Aflatoxin from Raw Peanuts. J. Assoc. Off. Anal. Chem. 1984, 67, 35–36. [CrossRef]
50. Whitaker, T.B.; Dickens, J.W.; Giesbrecht, F.G. Optimum Methanol Concentration and Solvent/Peanut Ratio
for Extraction of Aflatoxin from Raw Peanuts by Modified AOAC Method II. J. Assoc. Off. Anal. Chem. 1986,
69, 508–510. [CrossRef]
Toxins 2020, 12, 539 27 of 39
51. Scholten, J.M.; Spanjer, M.C. Determination of Aflatoxin Bi in Pistachio Kernels and Shells. J. AOAC Int.
1996, 79, 1360–1364. [CrossRef] [PubMed]
52. Schatzki, T.F.; Toyofuku, N. Sample Preparation and Presampling of Pistachios. J. Agric. Food Chem. 2003,
51, 6068–6072. [CrossRef] [PubMed]
53. Whitaker, T.B.; Dickens, J.W.; Monroe, R.J. A water slurry method of extracting aflatoxin from peanuts. J. Am.
Oil Chem. Soc. 1980, 57, 269–272. [CrossRef]
54. Kumphanda, J.; Matumba, L.; Whitaker, T.; Kasapila, W.; Sandahl, J. Maize meal slurry mixing: An economical
recipe for precise aflatoxin quantitation. World Mycotoxin J. 2019, 12, 203–212. [CrossRef]
55. Zhang, K.; Schaab, M.R.; Southwood, G.; Tor, E.R.; Aston, L.S.; Song, W.; Eitzer, B.; Majumdar, S.; Lapainis, T.;
Mai, H.; et al. A Collaborative Study: Determination of Mycotoxins in Corn, Peanut Butter, and Wheat
Flour Using Stable Isotope Dilution Assay (SIDA) and Liquid Chromatography-Tandem Mass Spectrometry
(LC-MS/MS). J. Agric. Food Chem. 2017, 65, 7138–7152. [CrossRef]
56. Balasubramanian, S.; Gupta, M.; Singh, K.K. Cryogenics and its Application with Reference to Spice Grinding:
A Review. Crit. Rev. Food Sci. Nutr. 2012, 52, 781–794. [CrossRef]
57. Liao, C.-D.; Wong, J.W.; Zhang, K.; Hayward, D.G.; Lee, N.S.; Trucksess, M.W. Multi-mycotoxin Analysis of
Finished Grain and Nut Products Using High-Performance Liquid Chromatography–Triple-Quadrupole
Mass Spectrometry. J. Agric. Food Chem. 2013, 61, 4771–4782. [CrossRef]
58. Kujawski, M.; Mischke, C.; Stroka, J. 2014 Report on the 2014 Proficiency Test of the European Union
Reference Laboratory for Mycotoxins, for the Network of National Reference Laboratories. Available
online: https://ec.europa.eu/jrc/en/publication/eur-scientific-and-technical-research-reports/report-2014-prof
iciency-test-european-union-reference-laboratory-mycotoxins-network-national (accessed on 18 April 2020).
59. Phillips, M.M.; Seal, T.M.L.; Ness, J.M.; Zhang, K. Development and Characterization of a Multimycotoxin
Reference Material. J. AOAC Int. 2019, 102, 1642–1650. [CrossRef]
60. International Organization for Standardization (ISO). ISO-35 ISO GUIDE 35:2017 Reference
Materials—Guidance for Characterization and Assessment of Homogeneity and Stability. Available online:
https://www.iso.org/standard/60281.html (accessed on 20 August 2020).
61. Weinert, D.W.; Cleary, T.G.; Mulholland, G.W. Size Distribution and Light Scattering Properties of Test
Smokes (NIST SP 965). 2011. Available online: https://nvlpubs.nist.gov/nistpubs/Legacy/SP/nistspecialpubli
cation965.pdf (accessed on 18 April 2020).
62. Robertson, J.A.; Pons, W.A.; Goldblatt, L.A. Stability of Individual Aflatoxin B1, B2, G1, and G2 Standards in
Benzene and Chloroform Solutions. J. Assoc. Off. Anal. Chem. 1970, 53, 299–302. [CrossRef]
63. Garcia, M.; Blanco, J.L.; Suarez, G. Aflatoxins B1 and G1 solubility in standard solutions and stability during
cold storage. Mycotoxin Res. 1994, 10, 97–100. [CrossRef]
64. Fishbein, L.; Falk, H. Chromatography of mold metabolites. Chromatogr. Rev. 1970, 12, 42–87. [CrossRef]
65. Parker, W.A.; Melnick, D. Absence of aflatoxin from refined vegetable oils. J. Am. Oil Chem. Soc. 1966,
43, 635–638. [CrossRef] [PubMed]
66. Pons, W.A.; Goldblatt, L.A. The determination of aflatoxins in cottonseed products. J. Am. Oil Chem. Soc.
1965, 42, 471–475. [CrossRef] [PubMed]
67. Eppley, R.M. Screening Method for Zearalenone, Aflatoxin, and Ochratoxin. J. Assoc. Off. Anal. Chem. 1968,
51, 74–78. [CrossRef]
68. Thean, J.E.; Lorenz, D.R.; Wilson, D.M.; Rodgers, K.; Gueldner, R.C. Extraction, Cleanup, and Quantitative
Determination of Aflatoxins in Corn. J. Assoc. Off. Anal. Chem. 1980, 63, 631–633. [CrossRef] [PubMed]
69. Francis, O.J.; Lipinskl, L.J.; Gaul, J.A.; Campbell, A.D. High Pressure Liquid Chromatographic Determination
of Aflatoxins in Peanut Butter Using a Silica Gel-Packed Flowcell for Fluorescence Detection. J. Assoc. Off.
Anal. Chem. 1982, 65, 672–676. [CrossRef] [PubMed]
70. Pons, W.A.; Franz, A.O. High Pressure Liquid Chromatographic Determination of Aflatoxins in Peanut
Products. J. Assoc. Off. Anal. Chem. 1978, 61, 793–800. [CrossRef]
71. Tarter, E.J.; Hanchay, J.-P.; Scott, P.M. Improved Liquid Chromatographic Method for Determination of
Aflatoxins in Peanut Butter and Other Commodities. J. Assoc. Off. Anal. Chem. 1984, 67, 597–600. [CrossRef]
72. Sizoo, E.A.; Van Egmond, H.P. Analysis of duplicate 24-hour diet samples for aflatoxin B1, aflatoxin M1and
ochratoxin A. Food Addit. Contam. 2005, 22, 163–172. [CrossRef]
Toxins 2020, 12, 539 28 of 39
73. Bourais, I.; Amine, A.; Venanzi, M.; Micheli, L.; Moscone, D.; Palleschi, G. Development and Application of a
Two-Phase Clean-Up System in Food Samples Prior to Fluorescence Analysis of Aflatoxins. Microchim. Acta
2005, 153, 101–108. [CrossRef]
74. Maia, P.P.; De Siqueira, M.E.P.B. Occurrence of aflatoxins B1, B2, G1 and G2 in some Brazilian pet foods.
Food Addit. Contam. 2002, 19, 1180–1183. [CrossRef]
75. Zgoła-Grześkowiak, A.; Grześkowiak, T. Dispersive liquid-liquid microextraction. TrAC Trends Anal. Chem.
2011, 30, 1382–1399. [CrossRef]
76. Quigley, A.; Cummins, W.; Connolly, D. Dispersive Liquid-Liquid Microextraction in the Analysis of Milk
and Dairy Products: A Review. J. Chem. 2016, 2016, 1–12. [CrossRef]
77. Hamed, A.M.; Abdel-Hamid, M.; Gámiz-Gracia, L.; García-Campaña, A.M.; Arroyo-Manzanares, N.
Determination of Aflatoxins in Plant-based Milk and Dairy Products by Dispersive Liquid-liquid
Microextraction and High-performance Liquid Chromatography with Fluorescence Detection. Anal. Lett.
2018, 52, 363–372. [CrossRef]
78. Hamed, A.M.; Gonzalez, D.M.; García-Campaña, A.M.; Gámiz-Gracia, L. Determination of Aflatoxins in
Yogurt by Dispersive Liquid-liquid Microextraction and HPLC with Photo-Induced Fluorescence Detection.
Food Anal. Methods 2016, 10, 516–521. [CrossRef]
79. Wang, N.; Duan, C.; Geng, X.; Li, S.; Ding, K.; Guan, Y. One step rapid dispersive liquid-liquid micro-extraction
with in-situ derivatization for determination of aflatoxins in vegetable oils based on high performance liquid
chromatography fluorescence detection. Food Chem. 2019, 287, 333–337. [CrossRef] [PubMed]
80. Campone, L.; Piccinelli, A.L.; Celano, R.; Russo, M.; Rastrelli, L. Rapid analysis of aflatoxin M1 in milk using
dispersive liquid-liquid microextraction coupled with ultrahigh pressure liquid chromatography tandem
mass spectrometry. Anal. Bioanal. Chem. 2013, 405, 8645–8652. [CrossRef]
81. Sun, M.; Xu, D.; Wang, S.; Uchiyama, K. Inkjet-Based Dispersive Liquid-liquid Microextraction Method
Coupled with UHPLC-MS/MS for the Determination of Aflatoxins in Wheat. Anal. Chem. 2019, 91, 3027–3034.
[CrossRef]
82. Endre, G.; Hegedüs, Z.; Turbat, A.; Škrbić, B.D.; Vágvölgyi, C.; Szekeres, A. Separation and Purification of
Aflatoxins by Centrifugal Partition Chromatography. Toxins 2019, 11, 309. [CrossRef]
83. Cole, R.J.; Dorner, J.W. Extraction of Aflatoxins from Naturally Contaminated Peanuts with Different Solvents
and Solvent/Peanut Ratios. J. AOAC Int. 1994, 77, 1509–1511. [CrossRef]
84. Nyberg, M. Efficiency of different extraction solvent mixtures used in analyses of aflatoxins from a certified
peanut meal reference material. Food Addit. Contam. 2004, 21, 781–785. [CrossRef]
85. Bao, L.; Liang, C.; Trucksess, M.W.; Xu, Y.; Lv, N.; Wu, Z.; Jing, P.; Fry, F.S. Determination of Aflatoxins B1, B2,
G1, and G2 in Olive Oil, Peanut Oil, and Sesame Oil Using Immunoaffinity Column Cleanup, Postcolumn
Derivatization, and Liquid Chromatography with Fluorescence Detection: First Action 2013.05. J. AOAC Int.
2013, 96, 1017–1018. [CrossRef] [PubMed]
86. Bao, L.; Liang, C.; Trucksess, M.W.; Xu, Y.; Lv, N.; Wu, Z.; Jing, P.; Fry, F.S.; Anderson, S.; Carter, L.;
et al. Determination of Aflatoxins B1, B2, G1, and G2 in Olive Oil, Peanut Oil, and Sesame Oil Using
Immunoaffinity Column Cleanup, Postcolumn Derivatization, and Liquid Chromatography/Fluorescence
Detection: Collaborative Study. J. AOAC Int. 2012, 95, 1689–1700. [CrossRef]
87. Bao, L.; Trucksess, M.W.; White, K.D. Determination of Aflatoxins B1, B2, G1, and G2 in Olive Oil, Peanut Oil,
and Sesame Oil. J. AOAC Int. 2010, 93, 936–942. [CrossRef]
88. Williams, I.O.; Ugbaje, S.A.; Igile, G.O.; Ekpe, O.O. Occurrence of Aflatoxin in Some Food Commodities
Commonly Consumed in Nigeria. J. Food Res. 2015, 4, 81. [CrossRef]
89. Zhang, K.; Wong, J.W.; Hayward, D.G.; Vaclavikova, M.; Liao, C.-D.; Trucksess, M.W. Determination of
Mycotoxins in Milk-Based Products and Infant Formula Using Stable Isotope Dilution Assay and Liquid
Chromatography Tandem Mass Spectrometry. J. Agric. Food Chem. 2013, 61, 6265–6273. [CrossRef]
90. Zhang, K.; Xu, D. Application of Stable Isotope Dilution and Liquid Chromatography Tandem Mass
Spectrometry for Multi-Mycotoxin Analysis in Edible Oils. J. AOAC Int. 2019, 102, 1651–1656. [CrossRef]
91. Holaday, C.E. Minicolumn chromatography: State of the art. J. Am. Oil Chem. Soc. 1981, 58, A931–A934.
[CrossRef]
92. Scott, P.M. Recent developments in methods of analysis for mycotoxins in foodstuffs. TrAC Trends Anal. Chem.
1993, 12, 382–386. [CrossRef]
Toxins 2020, 12, 539 29 of 39
93. Van der Zijden, A.; Koelensmid, W.; Bolding, J.; Barrett, J.; Ord, C.B.; Philp, J. Aspergillus Flavus and
Turkey X Disease: Isolation in Crystalline Form of a Toxin responsible for Turkey X Disease. Nature 1962,
195, 1060–1062. [CrossRef]
94. Hartley, R.D.; Nesbitt, B.F.; O’Kelly, J. Toxic Metabolites of Aspergillus Flavus. Nature 1963, 198, 1056–1058.
[CrossRef]
95. Robertson, J.A.; Pons, W.A.; Goldblatt, L.A. Preparation of aflatoxins and determination of their ultraviolet
and fluorescent characteristics. J. Agric. Food Chem. 1967, 15, 798–801. [CrossRef]
96. Velasco, J. Detection of aflatoxin using small columns of florisil. J. Am. Oil Chem. Soc. 1972, 49, 141–142.
[CrossRef] [PubMed]
97. Holaday, C.E.; Lansden, J. Rapid screening method for aflatoxin in a number of products. J. Agric. Food Chem.
1975, 23, 1134–1136. [CrossRef] [PubMed]
98. Shantha, T. Methods for estimation of aflatoxins: A critical appraisal. J. Food Sci. Technol. 1994, 31, 91–103.
99. Sobolev, V.S.; Dorner, J.W. Cleanup Procedure for Determination of Aflatoxins in Major Agricultural
Commodities by Liquid Chromatography. J. AOAC Int. 2002, 85, 642–645. [CrossRef]
100. Wilson, T.J.; Romer, T.R. Use of the Mycosep Multifunctional Cleanup Column for Liquid Chromatographic
Determination of Aflatoxins in Agricultural Products. J. Assoc. Off. Anal. Chem. 1991, 74, 951–956. [CrossRef]
101. Ventura, M.; Gómez, A.; Anaya, I.; Díaz, J.; Broto, F.; Agut, M.; Comellas, L. Determination of aflatoxins B1,
G1, B2 and G2 in medicinal herbs by liquid chromatography–tandem mass spectrometry. J. Chromatogr. A
2004, 1048, 25–29.
102. Park, J.W.; Kim, Y.-B. Effect of Pressure Cooking on Aflatoxin B1in Rice. J. Agric. Food Chem. 2006,
54, 2431–2435. [CrossRef]
103. Sobolev, V.S. Simple, Rapid, and Inexpensive Cleanup Method for Quantitation of Aflatoxins in Important
Agricultural Products by HPLC. J. Agric. Food Chem. 2007, 55, 2136–2141. [CrossRef]
104. Romero-Gonzalez, R.; Martinez Vidal, J.L.; Aguilera-Luiz, M.M.; Frenich, A.G. Application of
conventional solid-phase extraction for multimycotoxin analysis in beers by ultrahigh-performance liquid
chromatography-tandem mass spectrometry. J. Agric. Food Chem. 2009, 57, 9385–9392. [CrossRef] [PubMed]
105. Barker, S.A.; Long, A.R.; Short, C.R. Isolation of drug residues from tissues by solid phase dispersion.
J. Chromatogr. A 1989, 475, 353–361. [CrossRef]
106. Blesa, J.; Soriano, J.; Moltó, J.C.; Marínb, J.R.; Mañes, J. Determination of aflatoxins in peanuts by matrix
solid-phase dispersion and liquid chromatography. J. Chromatogr. A 2003, 1011, 49–54. [CrossRef]
107. McCullum, C.; Tchounwou, P.; Ding, L.-S.; Liao, X.; Liu, Y.-M. Extraction of Aflatoxins from Liquid Foodstuff
Samples with Polydopamine-Coated Superparamagnetic Nanoparticles for HPLC-MS/MS Analysis. J. Agric.
Food Chem. 2014, 62, 4261–4267. [CrossRef]
108. Hashemi, M.; Taherimaslak, Z.; Rashidi, S. Application of magnetic solid phase extraction for
separation and determination of aflatoxins B1 and B2 in cereal products by high performance liquid
chromatography-fluorescence detection. J. Chromatogr. B 2014, 960, 200–208. [CrossRef]
109. Hashemi, M.; Taherimaslak, Z.; Rashidi, S. Enhanced spectrofluorimetric determination of aflatoxin M1 in
liquid milk after magnetic solid phase extraction. Spectrochim. Acta Part A Mol. Biomol. Spectrosc. 2014,
128, 583–590. [CrossRef] [PubMed]
110. Lord, H.L. Strategies for interfacing solid-phase microextraction with liquid chromatography. J. Chromatogr. A
2007, 1152, 2–13. [CrossRef] [PubMed]
111. Pawliszyn, J. Solid Phase Microextraction: Theory and Practice; Wiley-VCH: Hoboken, NJ, USA, 1997;
ISBN 0-471-19034-9.
112. Nonaka, Y.; Saito, K.; Hanioka, N.; Narimatsu, S.; Kataoka, H. Determination of aflatoxins in food samples
by automated on-line in-tube solid-phase microextraction coupled with liquid chromatography–mass
spectrometry. J. Chromatogr. A 2009, 1216, 4416–4422. [CrossRef]
113. Quinto, M.; Spadaccino, G.; Palermo, C.; Centonze, D. Determination of aflatoxins in cereal flours
by solid-phase microextraction coupled with liquid chromatography and post-column photochemical
derivatization-fluorescence detection. J. Chromatogr. A 2009, 1216, 8636–8641. [CrossRef]
114. Groopman, J.D.; Croy, R.G.; Wogan, G.N. In vitro reactions of aflatoxin B1-adducted DNA. Proc. Natl. Acad.
Sci. USA 1981, 78, 5445–5449. [CrossRef]
115. Groopman, J.D.; Haugen, A.; Goodrich, G.R.; Wogan, G.N.; Harris, C.C. Quantitation of aflatoxin B1-modified
DNA using monoclonal antibodies. Cancer Res. 1982, 42, 3120–3124. [PubMed]
Toxins 2020, 12, 539 30 of 39
116. Wu, S.; Yang, G.; Sun, T. Studies on immuno-concentration and immunoassay of aflatoxins. Chin. J. Oncol.
1983, 5, 81–84.
117. Scott, P.M.; Trucksess, M.W. Special report: Application of Immunoaffinity Columns to Mycotoxin Analysis.
J. AOAC Int. 1997, 80, 941–950. [CrossRef] [PubMed]
118. Shephard, G.S. Aflatoxin analysis at the beginning of the twenty-first century. Anal. Bioanal. Chem. 2009,
395, 1215–1224. [CrossRef] [PubMed]
119. Holcomb, M.; Thompson, H.C. Analysis of aflatoxins (B1, B2, G1, and G2) in rodent feed by HPLC using
postcolumn derivatization and fluorescence detection. J. Agric. Food Chem. 1991, 39, 137–140. [CrossRef]
120. Hu, Y.-Y.; Zheng, P.; Zhang, Z.-X.; He, Y.-Z. Determination of Aflatoxins in High-Pigment Content Samples
by Matrix Solid-Phase Dispersion and High-Performance Liquid Chromatography. J. Agric. Food Chem. 2006,
54, 4126–4130. [CrossRef]
121. Trucksess, M.; Weaver, C.; Oles, C.; D’Ovidio, K.; Rader, J. Determination of Aflatoxins and Ochratoxin A in
Ginseng and Other Botanical Roots by Immunoaffinity Column Cleanup and Liquid Chromatography with
Fluorescence Detection. J. AOAC Int. 2006, 89, 624–630. [CrossRef]
122. Chen, C.-Y.; Li, W.-J.; Peng, K.-Y. Determination of Aflatoxin M1in Milk and Milk Powder Using High-Flow
Solid-Phase Extraction and Liquid Chromatography−Tandem Mass Spectrometry. J. Agric. Food Chem. 2005,
53, 8474–8480. [CrossRef]
123. Candlish, A.A.G.; Stimson, W.H. Chromatography of Mycotoxins: Techniques and Applications; Betina, V., Ed.;
Elsevier: Amsterdam, The Netherlands, 1993; pp. 99–123.
124. Uchigashima, M.; Saigusa, M.; Yamashita, H.; Miyake, S.; Fujita, K.; Nakajima, M.; Nishijima, M. Development
of a Novel Immunoaffinity Column for Aflatoxin Analysis Using an Organic Solvent-Tolerant Monoclonal
Antibody. J. Agric. Food Chem. 2009, 57, 8728–8734. [CrossRef]
125. Ertekin, Ö.; Kaymak, T.; Pirinçci, S.S.; Akçael, E.; Öztürk, S. Aflatoxin-specific monoclonal antibody selection
for immunoaffinity column development. BioTechniques 2019, 66, 261–268. [CrossRef]
126. Groopman, J.D.; Donahue, K.F. Aflatoxin, A Human Carcinogen: Determination in Foods and Biological
Samples by Monoclonal Antibody Affinity Chromatography. J. Assoc. Off. Anal. Chem. 1988, 71, 861–867.
[CrossRef] [PubMed]
127. Groopman, J.D.; Trudel, L.J.; Donahue, P.R.; Marshak-Rothstein, A.; Wogan, G.N. High-affinity monoclonal
antibodies for aflatoxins and their application to solid-phase immunoassays. Proc. Natl. Acad. Sci. USA 1984,
81, 7728–7731. [CrossRef] [PubMed]
128. Carvajal, M.; Mulholland, F.; Carner, C.R. Comparison of the EASI-EXTRACT Immunoaffinity Concentration
Procedure with the AOAC CB Method for the Extraction and Quantitation of Aflatoxin B1 in Raw Ground
Unskinned Peanuts. J. Chromatogr. A 1990, 511, 379–383. [CrossRef]
129. Stroka, J.; Van Otterdijk, R.; Anklam, E. Immunoaffinity column clean-up prior to thin-layer chromatography
for the determination of aflatoxins in various food matrices. J. Chromatogr. A 2000, 904, 251–256. [CrossRef]
130. Medina, M.S.; Martínez, A.J. Mold occurrence and aflatoxin B(1) and fumonisin B(1) determination in corn
samples in Venezuela. J. Agric. Food Chem. 2000, 48, 2833–2836. [CrossRef]
131. Trucksess, M.W.; Stack, M.E.; Nesheim, S.; Page, S.W.; Albert, R.H.; Hansen, T.J.; Donahue, K.F. Immunoaffinity
Column Coupled with Solution Fluorometry or Liquid Chromatography Postcolumn Derivatization for
Determination of Aflatoxins in Corn, Peanuts, and Peanut Butter: Collaborative Study. J. Assoc. Off.
Anal. Chem. 1991, 74, 81–88. [CrossRef]
132. AOAC Official Method: AOAC 991.31-1994(2002), Aflatoxins in Corn, Raw Peanuts, and Peanut.
Available online: http://www.aoacofficialmethod.org/index.php?main_page=product_info&products_id=1
723 (accessed on 20 August 2020).
133. Takahashi, T. Aflatoxin Contamination in Nutmeg: Analysis of Interfering TLC Spots. J. Food Sci. 1993,
58, 197–198. [CrossRef]
134. Tuinstra, L.; Verhulsdonka, C.; Bronsgeest, J.; Paulsch, W. Aflatoxin B1 in compound feedstuffs containing
citrus pulp. Procedures for screening and semi-quantitative determination. Neth. J. Agric. Sci. 1975, 23, 10–17.
[CrossRef]
135. Chan, D.; Macdonald, S.; Boughtflower, V.; Brereton, P. Simultaneous determination of aflatoxins
and ochratoxin A in food using a fully automated immunoaffinity column clean-up and liquid
chromatography–fluorescence detection. J. Chromatogr. A 2004, 1059, 13–16. [CrossRef]
Toxins 2020, 12, 539 31 of 39
136. Trucksess, M.; Weaver, C.M.; Oles, C.J.; Rump, L.V.; White, K.D.; Betz, J.M.; Rader, J.I. Use of Multitoxin
Immunoaffinity Columns for Determination of Aflatoxins and Ochratoxin A in Ginseng and Ginger.
J. AOAC Int. 2007, 90, 1042–1049. [CrossRef]
137. Trucksess, M.; Weaver, C.M.; Oles, C.J.; Fry, F.S.; Noonan, G.O.; Betz, J.M.; Rader, J.I. Determination of
Aflatoxins B1, B2, G1, and G2 and Ochratoxin A in Ginseng and Ginger by Multitoxin Immunoaffinity
Column Cleanup and Liquid Chromatographic Quantitation: Collaborative Study. J. AOAC Int. 2008,
91, 511–523. [CrossRef] [PubMed]
138. Garcia-Villanova, R.J.; Cordon, C.; González-Paramás, A.M.; Aparicio, P.; Rosales, M.E.G. Simultaneous
Immunoaffinity Column Cleanup and HPLC Analysis of Aflatoxins and Ochratoxin A in Spanish Bee Pollen.
J. Agric. Food Chem. 2004, 52, 7235–7239. [CrossRef] [PubMed]
139. Ferracane, R.; Tafuri, A.; Logieco, A.; Galvano, F.; Balzano, D.; Ritieni, A. Simultaneous determination
of aflatoxin B1and ochratoxin A and their natural occurrence in Mediterranean virgin olive oil.
Food Addit. Contam. 2007, 24, 173–180. [CrossRef] [PubMed]
140. Turcotte, A.-M.; Scott, P.M.; Tague, B. Analysis of cocoa products for ochratoxin A and aflatoxins. Mycotoxin Res.
2013, 29, 193–201. [CrossRef] [PubMed]
141. Lattanzio, V.; Solfrizzo, M.; Powers, S.; Visconti, A. Simultaneous determination of aflatoxins, ochratoxin A
and Fusarium toxins in maize by liquid chromatography/tandem mass spectrometry after multitoxin
immunoaffinity cleanup. Rapid Commun. Mass Spectrom. 2007, 21, 3253–3261. [CrossRef]
142. Vaclavikova, M.; MacMahon, S.; Zhang, K.; Begley, T.H. Application of single immunoaffinity clean-up
for simultaneous determination of regulated mycotoxins in cereals and nuts. Talanta 2013, 117, 345–351.
[CrossRef]
143. Wilcox, J.; Donnelly, C.; Leeman, D.; Marley, E. The use of immunoaffinity columns connected in tandem for
selective and cost-effective mycotoxin clean-up prior to multi-mycotoxin liquid chromatographic–tandem
mass spectrometric analysis in food matrices. J. Chromatogr. A 2015, 1400, 91–97. [CrossRef]
144. Ye, Z.; Cui, P.; Wang, Y.; Yan, H.; Wang, X.; Han, S.; Zhou, Y. Simultaneous Determination of Four Aflatoxins
in Dark Tea by Multifunctional Purification Column and Immunoaffinity Column Coupled to Liquid
Chromatography Tandem Mass Spectrometry. J. Agric. Food Chem. 2019, 67, 11481–11488. [CrossRef]
145. Sharman, M.; Gilbert, J. Automated aflatoxin analysis of foods and animal feeds using immunoaffinity
column clean-up and high-performance liquid chromatographic determination. J. Chromatogr. A 1991,
543, 220–225. [CrossRef]
146. Sharman, M.; Macdonald, S.; Gilbert, J. Automated liquid chromatographic determination of ochratoxin A in
cereals and animal products using immunoaffinity column clean-up. J. Chromatogr. A 1992, 603, 285–289.
[CrossRef]
147. Urano, T.; Trucksess, M.W.; Page, S.W. Automated affinity liquid chromatography system for on-line isolation,
separation, and quantitation of aflatoxins in methanol-water extracts of corn or peanuts. J. Agric. Food Chem.
1993, 41, 1982–1985. [CrossRef]
148. Carman, A.S.; Kuan, S.S.; Ware, G.M.; Umrigar, P.P.; Miller, K.V.; Guerrero, H.G. Robotic Automated Analysis
of Foods for Aflatoxin. J. AOAC Int. 1996, 79, 456–464. [CrossRef] [PubMed]
149. Zhang, K. Evaluation of Automated Sample Preparation for Mycotoxin Analysis in Foods. J. AOAC Int. 2020,
044. [CrossRef]
150. Span, R.; Wagner, W. A New Equation of State for Carbon Dioxide Covering the Fluid Region from the
Triple-Point Temperature to 1100 K at Pressures up to 800 MPa. J. Phys. Chem. Ref. Data 1996, 25, 1509.
[CrossRef]
151. Anklam, E.; Berg, H.; Mathiasson, L.; Sharman, M.; Ulberth, F. Supercritical fluid extraction (SFE) in food
analysis: A review. Food Addit. Contam. 1998, 15, 729–750. [CrossRef] [PubMed]
152. Engelhardt, H.; Haas, P. Possibilities and Limitations of SFE in the Extraction of Aflatoxin B1 from Food
Matrices. J. Chromatogr. Sci. 1993, 31, 13–19. [CrossRef]
153. Taylor, S.L.; King, J.W.; Richard, J.L.; Greer, J.I. Analytical-scale supercritical fluid extraction of aflatoxin B1
from field-inoculated corn. J. Agric. Food Chem. 1993, 41, 910–913. [CrossRef]
154. Taylor, S.L.; King, J.W.; Greer, J.I.; Richard, J.L. Supercritical Fluid Extraction of Aflatoxin M1 from Beef Livert.
J. Food Prot. 1997, 60, 698–700. [CrossRef]
155. Selim, M.I.; Tsuei, M.-H. Development and Optimization of a Supercritical Fluid Extraction Method for the
Analysis of Aflatoxin B1 in Grain Dust. Am. Ind. Hyg. Assoc. J. 1993, 54, 135–141. [CrossRef]
Toxins 2020, 12, 539 32 of 39
156. Wu, P.-Y.; Yen, Y.-H.; Chiou, R.Y.-Y. Extraction of Aflatoxin from Peanut Meal and Kernels by CO2-Methanol
Using a Highly Pressurized Fluid Extraction Process. J. Food Prot. 1995, 58, 800–803. [CrossRef]
157. Holcomb, M.; Thompson, H.C.; Cooper, W.M.; Hopper, M.L. SFE extraction of aflatoxins (B1, B2, G1, and G2)
from corn and analysis by HPLC. J. Supercrit. Fluids 1996, 9, 118–121. [CrossRef]
158. Liau, B.-C.; Jong, T.-T.; Lee, M.-R.; Chang, C.-M.J. Supercritical fluid extraction and quantification of aflatoxins
in Zizyphi Fructus by liquid chromatography/atmospheric pressure chemical ionization tandem mass
spectrometry. Rapid Commun. Mass Spectrom. 2007, 21, 667–673. [CrossRef] [PubMed]
159. Riekkola, M.L.; Manninen, P.; Hartonen, K. SFE, SFE/GC and SFE/SFC: Instrumentation and Applications,
Hyphenated Techniques in Supercritical Fluid Chromatography and Extraction; Journal of Chromatography
Library Series; Jinno, K., Ed.; Elsevier Science Publishers: Amsterdam, The Netherlands, 1992; Volume 53,
pp. 275–304.
160. Purchase, F.H.; Steyn, P.S. Estimation of Aflatoxin M1 in Milk. J. AOAC Int. 1967, 50, 363–366. [CrossRef]
161. Stubblefield, R.D.; Shannon, G.M.; Shotwell, O.L. Aflatoxin M1 in Milk: Evaluation of Methods. J. Assoc. Off.
Anal. Chem. 1973, 56, 1106–1110. [CrossRef]
162. Richter, B.E.; Jones, B.A.; Ezzell, J.L.; Porter, N.L.; Avdalovic, N.; Pohl, C.A. Accelerated Solvent Extraction:
A Technique for Sample Preparation. Anal. Chem. 1996, 68, 1033–1039. [CrossRef]
163. Han, Z.; Ren, Y.; Zhu, J.; Cai, Z.; Chen, Y.; Luan, L.; Wu, Y. Multianalysis of 35 Mycotoxins in Traditional Chinese
Medicines by Ultra-High-Performance Liquid Chromatography–Tandem Mass Spectrometry Coupled with
Accelerated Solvent Extraction. J. Agric. Food Chem. 2012, 60, 8233–8247. [CrossRef] [PubMed]
164. Chen, S.; Zhang, H. Development of a microwave-assisted-extraction-based method for the determination
of aflatoxins B1, G1, B2, and G2 in grains and grain products. Anal. Bioanal. Chem. 2013, 405, 1623–1630.
[CrossRef] [PubMed]
165. Li, C.; Xie, G.; Lu, A.; Ping, H.; Ma, Z.; Luan, Y.; Wang, J. Determination of Aflatoxins in Rice and Maize by
Ultra-High Performance Liquid Chromatography–Tandem Mass Spectrometry with Accelerated Solvent
Extraction and Solid-Phase Extraction. Anal. Lett. 2014, 47, 1485–1499. [CrossRef]
166. Du, L.-J.; Chu, C.; Warner, E.; Wang, Q.-Y.; Hu, Y.-H.; Chai, K.-J.; Cao, J.; Peng, L.-Q.; Chen, Y.-B.; Yang, J.;
et al. Rapid microwave-assisted dispersive micro-solid phase extraction of mycotoxins in food using zirconia
nanoparticles. J. Chromatogr. A 2018, 1561, 1–12. [CrossRef] [PubMed]
167. Nesbitt, B.F.; O’Kelly, J.; Sargeant, K.; Sheridan, A. Aspergillus flavus and turkey X disease. Nature 1962,
195, 1062–1064. [CrossRef] [PubMed]
168. Betina, V. Chapter 7 Thin-Layer Chromatography of Mycotoxins. J. Chromatogr. Libr. 1993, 54, 141–251.
169. Trucksess, M.W. Natural toxins. In Official Methods of Analysis, 17th ed.; Horwitz, W., Ed.; AOAC International:
Gaithersburg, MD, USA, 2000; pp. 1–64.
170. Campbell, A.D. Report on Mycotoxins. J. Assoc. Off. Anal. Chem. 1967, 50, 343–346. [CrossRef]
171. Stoloff, L. Collaborative Study of a Method for the Identification of Afiatoxin B1 by Derivative Formation.
J. Assoc. Off. Anal. Chem. 1967, 50, 354–360. [CrossRef]
172. Otta, K.H.; Papp, E.; Bagócsi, B. Determination of aflatoxinsin food by over-pressured layer chromatography.
J. Chromatogr. A 2000, 882, 11–18. [CrossRef]
173. Mincsovicsa, E.; Tyihákb, E. Overpressured Thin-Layer Chromatography (OPTLC)—A Flexible Tool of
Analysis and Isolation. Nat. Prod. Commun. 2011, 6, 719–732.
174. Zlatkis, A.; Kaiser, R.E. High Performance Thin-Layer Chromatography; Zlatkis, A., Kaiser, R.E., Eds.; Elsevier:
Amsterdam, The Netherlands, 1977; pp. 85–128.
175. Lee, K.Y.; Poole, C.F.; Zlatkis, A. Simultaneous multi-mycotoxin determination by high performance thin-layer
chromatography. Anal. Chem. 1980, 52, 837–842. [CrossRef]
176. Scussel, V.M. Comparison of methods by TLC and HPTLC for determination of aflatoxin M1 in milk and B1
in eggs. Food Sci. Technol. 2003, 23, 46–52. [CrossRef]
177. Tyihák, E.; Mincsovics, E.; Kalász, H. New planar liquid chromatographic technique: Overpressured
thin-layer chromatography. J. Chromatogr. A 1979, 174, 75–81. [CrossRef]
178. Papp, E.; H-Otta, K.; Záray, G.; Mincsovics, E. Liquid chromatographic determination of aflatoxins.
Microchem. J. 2002, 73, 39–46. [CrossRef]
179. Móricz, Á.M.; H.-Otta, K.; Ott, P.G.; Tyihák, E. Separation and detection of aflatoxins using overpressured-layer
chromatography and bioautography. J. Planar Chromatogr. Mod. TLC 2003, 16, 417–420. [CrossRef]
Toxins 2020, 12, 539 33 of 39
180. Andrellos, P.J.; Reid, G.R. Confirmatory Tests for Aflatoxin B1. J. Assoc. Off. Agric. Chem. 1964, 47, 801–803.
[CrossRef]
181. Haddon, W.F.; Wiley, M.; Waiss, A.C. Aflatoxin detection by thin-layer chromatography-mass spectrometry.
Anal. Chem. 1971, 43, 268–270. [CrossRef] [PubMed]
182. Haddon, W.F.; Masri, M.S.; Randall, V.G.; Elsken, R.H.; Meneghelli, B.J. Mass Spectral Confirmation of
Aflatoxins. J. Assoc. Off. Anal. Chem. 1977, 60, 107–113. [CrossRef]
183. Brumley, W.C.; Nesheim, S.; Trucksess, M.; Trucksess, E.W.; Dreifuss, P.A.; Roach, J.A.G.; Andrzejewski, D.;
Eppley, R.M.; Pohland, A.E.; Thorpe, C.W.; et al. Negative ion chemical ionization mass spectrometry of
aflatoxins and related mycotoxins. Anal. Chem. 1981, 53, 2003–2006. [CrossRef]
184. Park, D.L.; DiProssimo, V.; Abdel-Malek, E.; Trucksess, M.W.; Nesheim, S.; Brumley, W.C.; Sphon, J.A.;
Barry, T.L.; Petzinger, G. Negative Ion Chemical Ionization Mass Spectrometric Method for Confirmation of
Identity of Aflatoxin B: Collaborative Study. J. Assoc. Off. Anal. Chem. 1985, 68, 636–640. [CrossRef]
185. Tripathi, D.N.; Chauhan, L.R.; Bhattacharya, A. Separation and Identification Chromatography/Fast Atom of
Mycotoxins Bombardment by Thin-Layer Mass Spectrometry. Anal. Sci. 1991, 7, 423–425. [CrossRef]
186. Beckwith, A.C.; Stoloff, L. Fluorodensitometric Measurement of Aflatoxin Thin Layer Chromatograms.
J. Assoc. Off. Anal. Chem. 1968, 51, 602–608. [CrossRef]
187. Shotwell, O.L.; Stubblefield, R.D. Collaborative Study of the Determination of Aflatoxin in Corn and Soybeans.
J. Assoc. Off. Anal. Chem. 1972, 55, 781–788. [CrossRef]
188. Takahashi, D.M. Reversed-phase high-performance liquid chromatographic analytical system for aflatoxins
in wines with fluorescence detection. J. Chromatogr. A 1977, 131, 147–156. [CrossRef]
189. Maggon, K.K.; Gopal, S. Fluorometric Determination of Aflatoxins B2a and G2a. Agric. Biol. Chem. 1974,
38, 681–684. [CrossRef]
190. Van Duuren, B.L.; Chan, T.-L.; Irani, F.M. Luminescence characteristics of aflatoxins B1 and G1. Anal. Chem.
1968, 40, 2024–2027. [CrossRef] [PubMed]
191. Lillard, D.A.; Lantin, R.S. Some Chemical Characteristics and Biological Effects of Photomodified Aflatoxins.
J. Assoc. Off. Anal. Chem. 1970, 53, 1060–1063. [CrossRef]
192. Carnaghan, R.B.A.; Hartley, R.D.; O’Kelly, J. Toxicity and Fluorescence Properties of the Aflatoxins. Nature
1963, 200, 1101. [CrossRef] [PubMed]
193. Seiber, J.N.; Hsieh, D.P.H. Application of High-Speed Liquid Chromatography to the Analysis of Aflatoxins.
J. Assoc. Off. Anal. Chem. 1973, 56, 827–830. [CrossRef]
194. Rao, G.H.R.; Anders, M. Aflatoxin detection by high-speed liquid chromatography and mass spectrometry.
J. Chromatogr. A 1973, 84, 402–406. [CrossRef]
195. Garner, R. Aflatoxin separation by high-pressure liquid chromatography. J. Chromatogr. A 1975, 103, 186–188.
[CrossRef]
196. Seitz, L.M. Comparison of methods for aflatoxin analysis by high-pressure liquid chromatography.
J. Chromatogr. A 1975, 104, 81–89. [CrossRef]
197. Manabe, M.; Ooto, T.; Matsuur, S. High-performance Liquid Chromatography of Aflatoxins with Fluorescence
Detection. J. Agric. Biol. Chem. 1978, 42, 2003–2007.
198. Zimmerli, B. Improvement in the limits of analysis of aflatoxins with high pressure liquid chromatography
with the use of silicone filled fluorescence detectors. J. Chromatogr. 1977, 131, 458–463. [CrossRef]
199. Panalaks, T.; Scott, P.M. Sensitive Silica Gel-Packed Flowcell for Fluorometric Detection of Aflatoxins by
High Pressure Liquid Chromatography. J. Assoc. Off. Anal. Chem. 1977, 60, 583–589. [CrossRef]
200. Pons, W.A. High Pressure Liquid Chromatographic Determination of Aflatoxins in Corn. J. Assoc. Off.
Anal. Chem. 1979, 62, 586–594. [CrossRef] [PubMed]
201. Awe, M.J.; Schranz, J.L. High Pressure Liquid Chromatographic Determination of Aflatoxins in Spices.
J. Assoc. Off. Anal. Chem. 1981, 64, 1377–1382. [CrossRef] [PubMed]
202. Ferguson-Foos, J.; Warren, J.D. Improved Cleanup for Liquid Chromatographic Analysis and Fluorescence
Detection of Aflatoxins M1 and M2 in Fluid Milk Products. J. Assoc. Off. Anal. Chem. 1984, 67, 1111–1114.
[CrossRef] [PubMed]
203. Buchi, G.; Foulkes, D.M.; Kurono, M.; Mitchell, G.F. The Total Synthesis of Racemic Aflatoxin B1. J. Am.
Chem. Soc. 1966, 88, 4534–4536. [CrossRef]
204. Maggon, K.K.; Gopal, S.; Viswanathan, L.; Venkitasubramanian, T.A. Fluorescence characteristics of aflatoxins
B 2a & G 2a. Indian J. Biochem. Biophys. 1972, 9, 195–198.
Toxins 2020, 12, 539 34 of 39
205. Pons, W.A.; Cucullu, A.F.; Lee, L.S.; Janssen, H.J.; Goldblatt, L.A. Kinetic study of acid-catalyzed conversion
of aflatoxins B1and G1to B2aand G2a. J. Am. Oil Chem. Soc. 1972, 49, 124–128. [CrossRef]
206. Takahashi, D.M. Thin Layer Chromatographic Determination of Aflatoxins in Wine. J. Assoc. Off. Anal. Chem.
1974, 57, 875–879. [CrossRef]
207. Kok, W. Derivatization reactions for the determination of aflatoxins by liquid chromatography with
fluorescence detection. J. Chromatogr. B Biomed. Sci. Appl. 1994, 659, 127–137. [CrossRef]
208. Holcomb, M.; Wilson, D.; Trucksess, M.; Thompson, H. Determination of aflatoxins in food products by
chromatography. J. Chromatogr. A 1992, 624, 341–352. [CrossRef]
209. Beebe, R.M. Reverse Phase High Pressure Liquid Chromatographic Determination of Aflatoxins in Foods.
J. Assoc. Off. Anal. Chem. 1978, 61, 1347–1352. [CrossRef]
210. Diebold, G.J.; Zare, R.N. Laser fluorimetry: Subpicogram detection of aflatoxins using high-pressure liquid
chromatography. Science 1977, 196, 1439–1441. [CrossRef]
211. Beebe, R.M.; Takahashi, D.M. Determination of aflatoxin M1 by high-pressure liquid chromatography using
fluorescence detection. J. Agric. Food Chem. 1980, 28, 481–482. [CrossRef]
212. Tarter, E.J.; Hanchay, J.; Scott, P.M. Changes of Official Methods of Analysis, Association of Official Analytical
Chemists, Arlington, VA, 1990 15th ed. Suppl. 1, Method 990.33. J. Assoc. Off. Anal. Chem. 1984, 67, 565–597.
[PubMed]
213. Roch, O.G.; Blunden, G.; Coker, R.D.; Nawaz, S. The development and validation of a solid phase
extraction/HPLC method for the determination of aflatoxins in groundnut meal. Chromatographia 1992,
33, 208–212. [CrossRef]
214. Davis, N.D.; Diener, U.L. A fluorometric-iodine (Fl-I) method for measuring aflatoxin in corn. J. Appl. Biochem.
1979, 1, 115–122.
215. Tuinstra, L.; Haasnoot, W. Rapid determination of aflatoxin B1 in dutch feeding stuffs by high-performance
liquid chromatography and post-column derivatization. J. Chromatogr. A 1983, 282, 457–462. [CrossRef]
216. Gilbert, J.; Shepherd, M.J. A survey of aflatoxins in peanut butters, nuts and nut confectionery products by
HPLC with fluorescence detection. Food Addit. Contam. 1985, 2, 171–183. [CrossRef] [PubMed]
217. Shepherd, M.J.; Gilbert, J. An investigation of HPLC post?column iodination conditions for the enhancement
of aflatoxin B1 fluorescence. Food Addit. Contam. 1984, 1, 325–335. [CrossRef]
218. Thiel, P.G.; Stockenström, S.; Gathercole, P.S. Aflatoxin Analysis by Reverse Phase HPLC Using Post-Column
Derivatization for Enhancement of Fluorescence. J. Liq. Chromatogr. 1986, 9, 103–112. [CrossRef]
219. Hurst, W.J.; Snyder, K.; Martin, R. Determination of aflatoxins in peanut products using disposable
bonded-phase columns and post-column reaction detection. J. Chromatogr. A 1987, 409, 413–418. [CrossRef]
220. Paulsch, W.E.; Sizoo, E.A.; Van Egmond, H.P. Liquid Chromatographic Determination of Aflatoxins in
Feedstuffs Containing Citrus Pulp. J. Assoc. Off. Anal. Chem. 1988, 71, 957–961. [CrossRef] [PubMed]
221. Chamkasem, N.; Cobb, W.Y.; Latimer, G.W.; Salinas, C.; Clement, B.A. Liquid Chromatographic Determination
of Aflatoxins, Ochratoxin A, and Zearalenone in Grains, Oilseeds, and Animal Feeds by Post-Column
Derivatization and On-Line Sample Cleanup. J. Assoc. Off. Anal. Chem. 1989, 72, 336–341. [CrossRef]
[PubMed]
222. Beaver, R.W.; Wilson, D.M.; Trucksess, M.W. Comparison of Postcolumn Derivatization-Liquid
Chromatography with Thin-Layer Chromatography for Determination of Aflatoxins in Naturally
Contaminated Corn. J. Assoc. Off. Anal. Chem. 1990, 73, 579–581. [CrossRef]
223. Holcomb, M.; Korfmacher, W.A.; Thompson, H.C. Characterization of Iodine Derivatives of Aflatoxin B1 and
G1 by Thermospray Mass Spectrometry. J. Anal. Toxicol. 1991, 15, 289–292. [CrossRef] [PubMed]
224. Kok, W.; Brinkman, U.; Frei, R. On-line electrochemical reagent production for detection in liquid
chromatography and continuous flow systems. Anal. Chim. Acta 1984, 162, 19–32. [CrossRef]
225. Kok, W.; Van Neer, T.; Traag, W.; Tuinstra, L. Determination of aflatoxins in cattle feed by liquid
chromatography and post-column derivatization with electrochemically generated bromine. J. Chromatogr. A
1986, 367, 231–236. [CrossRef]
226. Traag, W.; Van Trijp, J.M.; Tuinstra, L.; Kok, W. Sample clean-up and post-column derivatization for the
determination of aflatoxin B1 in feedstuffs by liquid chromatography. J. Chromatogr. A 1987, 396, 389–394.
[CrossRef]
Toxins 2020, 12, 539 35 of 39
227. Garner, R.C.; Whattam, M.M.; Taylor, P.J.L.; Stow, M.W. Analysis of United Kingdom spices for aflatoxins
using an immunoaffinity column clean-up procedure followed by high-performance liquid chromatographic
analysis and post-column derivatization with pyridinium bromide perbromide. J. Chromatogr. 1993,
648, 485–490. [CrossRef]
228. Chiavaro, E.; Dall’Asta, C.; Galaverna, G.; Biancardi, A.; Gambarelli, E.; Dossena, A.; Marchelli, R.
New reversed-phase liquid chromatographic method to detect aflatoxins in food and feed with cyclodextrins
as fluorescence enhancers added to the eluent. J. Chromatogr. A 2001, 937, 31–40. [CrossRef]
229. Maragos, C.M.; Appell, M.; Lippolis, V.; Visconti, A.; Catucci, L.; Pascale, M. Use of cyclodextrins as modifiers
of fluorescence in the detection of mycotoxins. Food Addit. Contam. Part A 2008, 25, 164–171. [CrossRef]
230. Joshua, H. Determination of aflatoxins by reversed-phase high-performance liquid chromatography with
post-column in-line photochemical derivatization and fluorescence detection. J. Chromatogr. A 1993,
654, 247–254. [CrossRef]
231. Papadopoulo-Bouraoui, A.; Stroka, J.; Anklam, E. Comparison of two post-column derivatization systems,
ultraviolet irradiation and electrochemical determination, for the liquid chromatographic determination of
aflatoxins in food. J. AOAC Int. 2002, 85, 411–416. [PubMed]
232. Waltking, A.E.; Wilson, D.; Chan, D.; Dunn, E. Liquid chromatographic analysis of aflatoxin using post-column
photochemical derivatisation: Collaborative study. J. AOAC Int. 2006, 89, 678–692. [CrossRef] [PubMed]
233. Macnair, J.E.; Lewis, K.C.; Jorgenson, J.W. Ultrahigh-Pressure Reversed-Phase Liquid Chromatography in
Packed Capillary Columns. Anal. Chem. 1997, 69, 983–989. [CrossRef]
234. Naushad, M.; Khan, M.R.; Abdullah Alothman, Z. History and introduction of UPLC/MS. In Ultra
Performance Liquid Chromatography Mass Spectrometry Evaluation and Applications in Food Analysis; Summarizes
Basic Chromatography Theory and Advancements in Ultrahigh Performance Liquid Chromatography
(UHPLC)-MS; Naushad, M., Khan, M., Eds.; CRC Press: Boca Raton, FL, USA, 2014; pp. 1–15.
235. Ibáñez-Vea, M.; Corcuera, L.A.; Remiro, R.; Murillo-Arbizu, M.T.; González-Peñas, E.; Lizarraga, E.
Validation of a UHPLC-FLD method for the simultaneous quantification of aflatoxins, ochratoxin A
and zearalenone in barley. Food Chem. 2011, 127, 351–358. [CrossRef]
236. Abdel-Azeem, S.; Diab, M.; El-Shahat, M.F. Ultra-high-pressure liquid chromatography–solid-phase clean-up
for determining aflatoxins in Egyptian food commodities. J. Food Compos. Anal. 2015, 44, 18–24. [CrossRef]
237. Oulkar, D.; Goon, A.; Dhanshetty, M.; Khan, Z.; Satav, S.; Banerjee, K. High-sensitivity direct analysis of
aflatoxins in peanuts and cereal matrices by ultra-performance liquid chromatography with fluorescence
detection involving a large volume flow cell. J. Environ. Sci. Health Part B 2017, 53, 255–260. [CrossRef]
238. Kumar, A.; Dhanshetty, M.; Banerjee, K. Development and Validation of a Method for Direct Analysis
of Aflatoxins in Animal Feeds by Ultra-High-Performance Liquid Chromatography with Fluorescence
Detection. J. AOAC Int. 2020, 103, 940–945. [CrossRef]
239. Kebarle, P.; Tang, L. From ions in solution to ions in the gas phase: The mechanism of electrospray mass
spectrometry. Anal. Chem. 1993, 65, 972A–986A. [CrossRef]
240. Di Stefano, V.; Avellone, G.; Bongiorno, D.; Cunsolo, V.; Muccilli, V.; Sforza, S.; Dossena, A.; Drahos, L.;
Vékey, K. Applications of liquid chromatography—Mass spectrometry for food analysis. J. Chromatogr. A
2012, 1259, 74–85. [CrossRef]
241. Tittlemier, S.; Cramer, B.; Dall’Asta, C.; Iha, M.; Lattanzio, V.; Maragos, C.; Solfrizzo, M.; Stranska, M.;
Stroka, J.; Sumarah, M. Developments in mycotoxin analysis: An update for 2018-19. World Mycotoxin J.
2020, 13, 3–24. [CrossRef]
242. Blakley, C.R.; Vestal, M.L. Thermospray interface for liquid chromatography/mass spectrometry. Anal. Chem.
1983, 55, 750–754. [CrossRef]
243. Vestal, M.L. High-performance liquid chromatography-mass spectrometry. Science 1984, 226, 275–281.
[CrossRef] [PubMed]
244. Hurst, W.J.; Martin, R.A.; Vestal, C.H. The Use of HPLC/Thermospray Ms for the Confirmation of Aflatoxins
in Peanuts. J. Liq. Chromatogr. 1991, 14, 2541–2550. [CrossRef]
245. Cappiello, A.; Famiglini, G.; Tirillini, B. Determination of aflatoxins in peanut meal by LC/MS with a particle
beam interface. Chromatographia 1995, 40, 411–416. [CrossRef]
246. Nesheim, S.; Brumley, W.C. Confirmation of identity of aflatoxins. J. Am. Oil Chem. Soc. 1981, 58, A945–A949.
[CrossRef]
Toxins 2020, 12, 539 36 of 39
247. Cai, Y.X.; McLaughlin, M.; Zhang, K. Advancing the FDA/Office of Regulatory Affairs Mycotoxin Program:
New Analytical Method Approaches to Addressing Needs and Challenges. J. AOAC Int. 2020, 103, 705–709.
[CrossRef]
248. Armbrecht, B.H.; Fitzhugh, O.G. Mycotoxins II: The biological assay of aflatoxin in Peking white ducklings.
Toxicol. Appl. Pharmacol. 1969, 6, 421–426. [CrossRef]
249. Scott, P.M.; Kennedy, B.P.C. Analysis and Survey of Ground Black, White, and Capsicum Peppers for
Aflatoxins. J. Assoc. Off. Anal. Chem. 1973, 56, 1452–1457. [CrossRef]
250. Davis, N.D.; Diener, U.L. Confirmatory Test for the High Pressure Liquid Chromatographic Determination
of Aflatoxin B1. J. Assoc. Off. Anal. Chem. 1980, 63, 107–109. [CrossRef]
251. Trucksess, M.W.; Brumley, W.C.; Nesheim, S. Rapid Quantitation and Confirmation of Aflatoxins in
Corn and Peanut Butter, Using a Disposable Silica Gel Column, Thin Layer Chromatography, and Gas
Chromatography/Mass Spectrometry. J. Assoc. Off. Anal. Chem. 1984, 67, 973–975. [CrossRef] [PubMed]
252. Zhang, K.; Wong, J.W.; Krynitsky, A.J.; Trucksess, M.W. Perspective on Advancing FDA Regulatory Monitoring
for Mycotoxins in Foods using Liquid Chromatography and Mass Spectrometry (Review). J. AOAC Int. 2016,
99, 890–894. [CrossRef] [PubMed]
253. Sforza, S.; Dall’Asta, C.; Marchelli, R. Recent advances in mycotoxin determination in food and feed by
hyphenated chromatographic techniques/mass spectrometry. Mass Spectrom. Rev. 2005, 25, 54–76. [CrossRef]
[PubMed]
254. Souverain, S.; Rudaz, S.; Veuthey, J.L. Matrix effect in LC-ESI-MS and LC-APCI-MS with off-line and on-line
extraction procedures. J. Chromatogr. A 2004, 1058, 61–66. [CrossRef]
255. Rychlik, M.; Asam, S. Stable isotope dilution assays in mycotoxin analysis. Anal. Bioanal. Chem. 2008,
390, 617–628. [CrossRef]
256. Trufelli, H.; Palma, P.; Famiglini, G.; Cappiello, A. An overview of matrix effects in liquid
chromatography-mass spectrometry. Mass Spectrom. Rev. 2011, 30, 491–509. [CrossRef]
257. Stahnke, H.; Kittlaus, S.; Kempe, G.; Alder, L. Reduction of Matrix Effects in Liquid
Chromatography–Electrospray Ionization–Mass Spectrometry by Dilution of the Sample Extracts: How Much
Dilution is Needed? Anal. Chem. 2012, 84, 1474–1482. [CrossRef]
258. Desmarchelier, A.; Tessiot, S.; Bessaire, T.; Racault, L.; Fiorese, E.; Urbani, A.; Chan, W.-C.; Cheng, P.;
Mottier, P. Combining the quick, easy, cheap, effective, rugged and safe approach and clean-up by
immunoaffinity column for the analysis of 15 mycotoxins by isotope dilution liquid chromatography
tandem mass spectrometry. J. Chromatogr. A 2014, 1337, 75–84. [CrossRef]
259. Bessaire, T.; Mujahid, C.; Mottier, P.; Desmarchelier, A. Multiple Mycotoxins Determination in Food by
LC-MS/MS: An International Collaborative Study. Toxins 2019, 11, 658. [CrossRef]
260. Liao, C.-D.; Wong, J.W.; Zhang, K.; Yang, P.; Wittenberg, J.B.; Trucksess, M.W.; Hayward, D.G.; Lee, N.S.;
Chang, J.S. Multi-mycotoxin Analysis of Finished Grain and Nut Products Using Ultrahigh-Performance
Liquid Chromatography and Positive Electrospray Ionization—Quadrupole Orbital Ion Trap High-Resolution
Mass Spectrometry. J. Agric. Food Chem. 2015, 63, 8314–8332. [CrossRef]
261. Vahl, M.; Jørgensen, K. Determination of aflatoxins in food using LC/MS/MS. Zeitschrift für
Lebensmitteluntersuchung und Forschung A 1998, 206, 243–245. [CrossRef]
262. Cavaliere, C.; Foglia, P.; Guarino, C.; Nazzari, M.; Samperi, R.; Laganà, A. A sensitive confirmatory method
for aflatoxins in maize based on liquid chromatography/electrospray ionization tandem mass spectrometry.
Rapid Commun. Mass Spectrom. 2007, 21, 550–556. [CrossRef] [PubMed]
263. Cervino, C.; Asam, S.; Knopp, D.; Rychlik, M.; Niessner, R. Use of Isotope-Labeled Aflatoxins for LC-MS/MS
Stable Isotope Dilution Analysis of Foods. J. Agric. Food Chem. 2008, 56, 1873–1879. [CrossRef] [PubMed]
264. Li, W.; Herrman, T.J.; Dai, S.Y. Determination of aflatoxins in animal feeds by liquid chromatography/tandem
mass spectrometry with isotope dilution. Rapid Commun. Mass Spectrom. 2011, 25, 1222–1230. [CrossRef]
[PubMed]
265. European Commission. Commission Decision 2002/657/EC implementing Council Directive 96/23/EC
concerning the performance of analytical methods and the interpretation of results. Off. J. Eur. Communities
2002, 50, 8–36.
266. Xavier, J.J.M.; Scussel, V.M. Development of an LC-MS/MS method for the determination of aflatoxins B1, B2,
G1, and G2in Brazil nut. Int. J. Environ. Anal. Chem. 2008, 88, 425–433. [CrossRef]
Toxins 2020, 12, 539 37 of 39
267. Huang, B.; Han, Z.; Cai, Z.; Wu, Y.; Ren, Y. Simultaneous determination of aflatoxins B1, B2, G1, G2, M1 and
M2 in peanuts and their derivative products by ultra-high-performance liquid chromatography–tandem
mass spectrometry. Anal. Chim. Acta 2010, 662, 62–68. [CrossRef]
268. Grio, S.J.L.; Frenich, A.G.J.; Vidal, J.L.M.; Romero-Gonzalez, R. Determination of aflatoxins B1, B2, G1, G2 and
ochratoxin A in animal feed by ultrahigh-performance liquid chromatography—Tandem mass spectrometry.
J. Sep. Sci. 2010, 33, 502–508. [CrossRef]
269. Breidbach, A.; Ulberth, F. Two-dimensional heart-cut LC-LC improves accuracy of exact-matching double
isotope dilution mass spectrometry measurements of aflatoxin B1 in cereal-based baby food, maize,
and maize-based feed. Anal. Bioanal. Chem. 2015, 407, 3159–3167. [CrossRef]
270. Campone, L.; Rizzo, S.; Piccinelli, A.L.; Celano, R.; Pagano, I.; Russo, M.; Labra, M.; Rastrelli, L. Determination
of mycotoxins in beer by multi heart-cutting two-dimensional liquid chromatography tandem mass
spectrometry method. Food Chem. 2020, 318, 126496. [CrossRef]
271. Rodrigues, I.; Naehrer, K. A Three-Year Survey on the Worldwide Occurrence of Mycotoxins in Feedstuffs
and Feed. Toxins 2012, 4, 663–675. [CrossRef] [PubMed]
272. Smith, M.-C.; Madec, S.; Coton, E.; Hymery, N. Natural Co-Occurrence of Mycotoxins in Foods and Feeds
and Their in vitro Combined Toxicological Effects. Toxins 2016, 8, 94. [CrossRef] [PubMed]
273. Lee, H.J.; Ryu, D. Worldwide Occurrence of Mycotoxins in Cereals and Cereal-Derived Food Products:
Public Health Perspectives of Their Co-occurrence. J. Agric. Food Chem. 2017, 65, 7034–7051. [CrossRef]
[PubMed]
274. Lombaert, G.A.; Pellaers, P.; Roscoe, V.; Mankotia, M.; Neil, R.; Scott, P.M. Mycotoxins in infant cereal foods
from the Canadian retail market. Food Addit. Contam. 2003, 20, 494–504. [CrossRef] [PubMed]
275. Zhang, K.; Flannery, B.M.; Oles, C.J.; Adeuya, A. Mycotoxins in infant/toddler foods and breakfast cereals in
the US retail market. Food Addit. Contam. Part B 2018, 11, 1–8. [CrossRef] [PubMed]
276. Sulyok, M.; Berthiller, F.; Krska, R.; Schuhmacher, R. Development and validation of a liquid
chromatography/tandem mass spectrometric method for the determination of 39 mycotoxins in wheat and
maize. Rapid Commun. Mass Spectrom. 2006, 20, 2649–2659. [CrossRef]
277. Sulyok, M.; Krska, R.; Schuhmacher, R. A liquid chromatography/tandem mass spectrometric multi-mycotoxin
method for the quantification of 87 analytes and its application to semi-quantitative screening of moldy food
samples. Anal. Bioanal. Chem. 2007, 389, 1505–1523. [CrossRef]
278. Vishwanath, V.; Sulyok, M.; Labuda, R.; Bicker, W.; Krska, R. Simultaneous determination of 186
fungal and bacterial metabolites in indoor matrices by liquid chromatography/tandem mass spectrometry.
Anal. Bioanal. Chem. 2009, 395, 1355–1372. [CrossRef]
279. Varga, E.; Glauner, T.; Berthiller, F.; Krska, R.; Schuhmacher, R.; Sulyok, M. Development and validation of
a (semi-)quantitative UHPLC-MS/MS method for the determination of 191 mycotoxins and other fungal
metabolites in almonds, hazelnuts, peanuts and pistachios. Anal. Bioanal. Chem. 2013, 405, 5087–5104.
[CrossRef]
280. Malachová, A.; Sulyok, M.; Beltrán, E.; Berthiller, F.; Krska, R. Optimization and validation of a quantitative
liquid chromatography–tandem mass spectrometric method covering 295 bacterial and fungal metabolites
including all regulated mycotoxins in four model food matrices. J. Chromatogr. A 2014, 1362, 145–156.
[CrossRef]
281. Sulyok, M.; Stadler, D.; Steiner, D.; Krska, R. Validation of an LC-MS/MS-based dilute-and-shoot approach
for the quantification of >500 mycotoxins and other secondary metabolites in food crops: Challenges and
solutions. Anal. Bioanal. Chem. 2020, 412, 2607–2620. [CrossRef] [PubMed]
282. SANTE/11813/2017. Guidance Document on Analytical Quality Control and Method Validation Procedures
for Pesticide Residues and Analysis in Food and Feed. Available online: https://ec.europa.eu/food/sites/food
/files/plant/docs/pesticides_mrl_guidelines_wrkdoc_2017-11813.pdf (accessed on 6 June 2020).
283. Mol, H.G.J.; Bolaños, P.P.; Zomer, P.; De Rijk, T.C.; Stolker, A.A.M.; Mulder, P.P.J. Toward a Generic Extraction
Method for Simultaneous Determination of Pesticides, Mycotoxins, Plant Toxins, and Veterinary Drugs in
Feed and Food Matrixes. Anal. Chem. 2008, 80, 9450–9459. [CrossRef] [PubMed]
284. Stroka, J.; Maragos, C.M. Challenges in the analysis of multiple mycotoxins. World Mycotoxin J. 2016,
9, 847–861. [CrossRef]
Toxins 2020, 12, 539 38 of 39
285. Malachova, A.; Stránská, M.; Vaclavikova, M.; Elliott, C.T.; Black, C.; Meneely, J.; Hajslova, J.; Ezekiel, C.N.;
Schuhmacher, R.; Krska, R. Advanced LC-MS-based methods to study the co-occurrence and metabolization
of multiple mycotoxins in cereals and cereal-based food. Anal. Bioanal. Chem. 2017, 410, 801–825. [CrossRef]
286. Renaud, J.B.; Miller, J.D.; Sumarah, M.W. Mycotoxin Testing Paradigm: Challenges and Opportunities for
the Future. J. AOAC Int. 2019, 102, 1681–1688. [CrossRef]
287. SANTE/12682/2019. Main Changes Introduced in Document Nº SANTE/12682/2019 with Respect to the
Previous Version (Document Nº SANTE/11813/2017). Available online: https://ec.europa.eu/food/sites/food/f
iles/plant/docs/pesticides_mrl_guidelines_wrkdoc_2019-12682.pdf (accessed on 6 June 2020).
288. FDA. Acceptance Criteria for Confirmation of Identity of Chemical Residues using Exact Mass Data within
the Office of Foods and Veterinary Medicine. 2015. Available online: https://www.fda.gov/media/96499/do
wnload (accessed on 16 June 2020).
289. Tóth, K.; Nagy, L.; Mándi, A.; Kuki, Á.; Mézes, M.; Zsuga, M.; Kéki, S. Collision-induced dissociation of
aflatoxins. Rapid Commun. Mass Spectrom. 2013, 27, 553–559. [CrossRef]
290. Vaclavik, L.; Zachariasova, M.; Hrbek, V.; Hajslova, J. Analysis of multiple mycotoxins in cereals under
ambient conditions using direct analysis in real time (DART) ionization coupled to high resolution mass
spectrometry. Talanta 2010, 82, 1950–1957. [CrossRef]
291. Lattanzio, V.; Della Gatta, S.; Godula, M.; Visconti, A. Quantitative analysis of mycotoxins in cereal foods
by collision cell fragmentation-high-resolution mass spectrometry: Performance and comparison with
triple-stage quadrupole detection. Food Addit. Contam. Part A 2011, 28, 1424–1437. [CrossRef]
292. Renaud, J.B.; Sumarah, M. Data independent acquisition-digital archiving mass spectrometry: Application
to single kernel mycotoxin analysis of Fusarium graminearum infected maize. Anal. Bioanal. Chem. 2016,
408, 3083–3091. [CrossRef]
293. Castaldo, L.; Graziani, G.; Gaspari, A.; Izzo, L.; Tolosa, J.; Rodríguez-Carrasco, Y.; Ritieni, A. Target Analysis
and Retrospective Screening of Multiple Mycotoxins in Pet Food Using UHPLC-Q-Orbitrap HRMS. Toxins
2019, 11, 434. [CrossRef]
294. Stefanović, S.M.; Spiric, D.; Petronijevic, R.; Nedeljkovic-Trailovic, J.; Milićević, D.; Nikolic, D.; Janković, S.
Comparison of two Analytical Methods (ELISA and LC-MS/MS) for Determination of Aflatoxin B1 in Corn
and Aflatoxin M1 in Milk. Procedia Food Sci. 2015, 5, 270–273. [CrossRef]
295. Kos, J.; Hajnal, E.J.; Jajić, I.; Krstović, S.; Mastilović, J.; Šarić, B.; Jovanov, P. Comparison of ELISA, HPLC-FLD
and HPLC-MS/MS Methods for Determination of Aflatoxin M1 in Natural Contaminated Milk Samples.
Acta Chim. Slov. 2016, 63, 747–756. [CrossRef] [PubMed]
296. Lattanzio, V.; Von Holst, C.; Lippolis, V.; De Girolamo, A.; Logrieco, A.; Mol, H.G.J.; Pascale, M. Evaluation
of Mycotoxin Screening Tests in a Verification Study Involving First Time Users. Toxins Basel 2019, 11, 129.
[CrossRef] [PubMed]
297. Friedli, F. Fused silica capillary GC/MS coupling: A new, innovative approach. J. High Resolut. Chromatogr.
1981, 4, 495–499. [CrossRef]
298. Rosen, R.T.; Rosen, J.D.; DiProssimo, V.P. Confirmation of aflatoxins B1 and B2 in peanuts by gas
chromatography/mass spectrometry/selected ion monitoring. J. Agric. Food Chem. 1984, 32, 276–278.
[CrossRef]
299. Goto, T.; Matsui, M.; Kitsuwa, T. Analysis of Aspergillus mycotoxins by gas chromatography using fused
silica capillary column. Proc. Jpn. Assoc. Mycotoxicol. 1990, 1990, 43–47. [CrossRef]
300. Goto, T.; Matsui, M.; Kitsuwa, T. Determination of aflatoxins by capillary column gas chromatography.
J. Chromatogr. 1988, 447, 410–414. [CrossRef]
301. Maragos, C.M. Analysis of mycotoxins with capillary electrophoresis. Semin. Food Anal. 1998, 3, 353–373.
302. Maragos, C.M. Measurement of Aflatoxins Using Capillary Electrophoresis. In Methods in Molecular Biology;
Trucksess, M.W., Pohland, A.E., Eds.; Humana Press Inc.: Totowa, NJ, USA, 2001; Volume 157, Chapter 6;
pp. 51–58.
303. Dolník, V.; Liu, S.; Jovanovich, S. Capillary electrophoresis on microchip. Electrophoresis 2000, 21, 41–54.
[CrossRef]
304. Li, P.; Zhang, Z.; Zhang, Q.; Zhang, N.; Zhang, W.; Ding, X.; Li, R. Current development of microfluidic
immunosensing approaches for mycotoxin detection via capillary electromigration and lateral flow technology.
Electrophoresis 2012, 33, 2253–2265. [CrossRef]
Toxins 2020, 12, 539 39 of 39
305. Holland, R.D.; Sepaniak, M.J. Qualitative Analysis of Mycotoxins Using Micellar Electrokinetic Capillary
Chromatography. Anal. Chem. 1993, 65, 1140–1146. [CrossRef] [PubMed]
306. Cole, R.O.; Holland, R.D.; Sepaniak, M.J. Factors influencing performance in the rapid separation of aflatoxins
by micellar electrokinetic capillary chromatography. Talanta 1992, 39, 1139–1147. [CrossRef]
307. Maragos, C.M.; Greer, J.I. Analysis of Aflatoxin B1in Corn Using Capillary Electrophoresis with Laser-Induced
Fluorescence Detection. J. Agric. Food Chem. 1997, 45, 4337–4341. [CrossRef]
308. Jiang, Y.; Wang, P.-C.; Locascio, L.E.; Lee, C.S. Integrated plastic microfluidic devices with ESI-MS for drug
screening and residue analysis. Anal. Chem. 2001, 73, 2048–2053. [CrossRef] [PubMed]
© 2020 by the authors. Licensee MDPI, Basel, Switzerland. This article is an open access
article distributed under the terms and conditions of the Creative Commons Attribution
(CC BY) license (http://creativecommons.org/licenses/by/4.0/).