Location via proxy:   [ UP ]  
[Report a bug]   [Manage cookies]                

RITS USA LLC - Medical Gas Master Alarms & Area Alarms

Download as pdf or txt
Download as pdf or txt
You are on page 1of 13

RITS - USA, LLC

PO Box 1015. Buford, GA 30515-1015, USA


Phone: +1- 770-831 1480 Fax: +1- 770-573 3712 E-mail: rits@rits-usa.com

Medical Gas Master Alarms

Features
The Tri-Tech Master Alarm Panel monitors and displays
normal and alarm conditions from up to 64 remote medical
gas source signals. Pressure switches and DISS union
check valve connectors are sold separately.

• ETL listed to UL-1069 and CSA-C22.2 No. 205


• Complies with NFPA 99. Made in the U.S.A.
• Self-contained unit – designed for ease of installation and
service.
• MCP circuit board available for interface to building
management system.
• Microprocessor controlled
• Self-diagnostic (test button) and error message display for
ease of maintenance.
• Audio and visual alarm indicators
• Bright easy to read L.E.D. displays – clearly visible in both
day and night lighting conditions
• Visual (green) indication of normal condition
• Constant display and monitoring of each source alarm signal (Master alarm shown above is 16 signals – part # DU16)
• Individual user programmable remote signal alarm points
(NC, NO or disabled) Each source signal module shall monitor 16 signals. The
• Dry contacts for remote monitoring of all alarm conditions alarm shall be capable of monitoring and displaying up to 64
• Alarm history display of previous alarm conditions medical gas source signals in increments of 16 source
• Hinged frame with lanyards for easy accessibility signals. The alarm is modular and may be initially installed
• Three year PC board warranty as a smaller unit (i.e. 16 signals) and later expanded to (i.e.
• Real time main line gas pressure readout modules available 32 signals).
Specification
The Master Alarm Panel shall be the Tri-Tech Medical In addition, each 16 signal Master Alarm Module shall
Master Alarm Panel. The panel shall be microprocessor incorporate the following features:
controlled and designed to comply with NFPA 99. The
• Individual user programmable remote signal alarm points to
panel shall be 100% digital and shall not require re-
accept NO or NC signals, or may be disabled. Factory preset to
calibration. The panel shall be able to interface with accept Normally Closed signals
building management systems with the use of an MCP • LED indicators (Green) confirms normal status, (Red) indicates
circuit board. The master alarm panel shall be compatible abnormal condition
with the T-Net PC based medical gas information • Each signal easily labeled and positioned to suit any
management system. The alarm panel shall be enclosed in requirement using self adhesive labels provided with the alarm
a steel box and shall be designed to accept an electrical • A set of dry contacts with relay to trigger an optional remote
input range of 120-240 volts AC – 50-60 hertz. The alarm in the event of an alarm condition
source voltage shall be stepped down with a self-contained
transformer. The panel shall contain audible and visual
alarm indicators. The audible alarm may be silenced by
pressing the alarm silence button, but the visual alarm
indicator can only be cancelled by fault correction. The
alarm shall detect and filter out transient signals (less than
0.6 seconds). The alarm shall be capable of displaying
alarm history for all possible alarm conditions.
RITS - USA, LLC
PO Box 1015. Buford, GA 30515-1015, USA
Phone: +1- 770-831 1480 Fax: +1- 770-573 3712 E-mail: rits@rits-usa.com

Medical Gas Master Alarms


RITS - USA, LLC
PO Box 1015. Buford, GA 30515-1015, USA
Phone: +1- 770-831 1480 Fax: +1- 770-573 3712 E-mail: rits@rits-usa.com

Medical Gas Master Alarms


Ordering Information:

Label Colors Area Alarm Gas Services |Master Remote Blank Slot for
Signals Future Expansion

U = USA (NFPA) O = Oxygen 16 = 16 points B = Blank Slot


C = Canada (CSA) V = Medical Vacuum 32 = 32 points
A = Medical Air 48 = 48 points
N = Nitrous Oxide 64 = 64 points
T = Nitrogen
C = Carbon Dioxide
W = WAGD/EVAC
S = AGSS
H = Hyperbaric Oxygen
U = Utility Air
L = Helium
I = Instrument Air (USA) Surgical Air (Canada)
D = Carbon Dioxide 80 psig
M = Gas Mixtures 50 psig
P = Gas Mixtures high pressure 180 psig
R = Tri-Gas
F = Future

Examples:

DU16 = 16 signal Master Alarm, USA colors, 2 slot box


DUOVB = 2 gas Area Alarm – Oxygen, Vacuum and Blank slot, USA colors, 2 slot box
DUOFB = 1 gas Area Alarm – Oxygen, Future and Blank slot, USA colors, 2 slot box
DUOVANT = 5 gas Area – OXY, VAC, AIR, N2O, N2 and Blank slot, USA colors, 2 slot box

DCOV16B = 2 gas Area – OXY, VAC & 16 signal Master & Blank slot, USA colors, 3 slot box
DU32 = 32 signal Master Alarm, USA colors, 3 slot box
DUOVANTB = 5 gas Area – OXY, VAC, AIR, N2O, N2 and Blank slot, USA colors, 3 slot box
DUOVANTCWMB = 8 gas Area, OXY, VAC, AIR, N2O, N2, CO2, WAGD, Gas Mixture and Blank
slot, USA colors, 3 slot box

DU64 = 64 signal Master, USA colors, 5 slot box


DUOVAB32 = 3 gas Area – OXY, VAC, AIR, Blank slot & 32 signal Master, USA colors, 5
slot box
DUOVANCTWU32 = 8 gas Area – OXY, VAC, AIR, N2O, CO2, N2, WAGD, Utility Air & 32
signal Master, USA colors, 5 slot box

*Note – Medical Gas Alarms come in one of the three configurations shown above - 2 slots , 3 slots or 5 slots.
RITS - USA, LLC
PO Box 1015. Buford, GA 30515-1015, USA
Phone: +1- 770-831 1480 Fax: +1- 770-573 3712 E-mail: rits@rits-usa.com

Medical Gas Master Alarms


Ordering Information:

Logic Module Logic Module Area Alarm Module Master Alarm Module Blank Module
With Buzzer With Buzzer (Choose 3 letters from 16 Signals per module
& Two Gas chart below – one for
(Area Alarm each gas service)
Modules)

Part Number Part Number Part Number Part Number


DU = USA (NFPA) A = Medical Air 16 = One 16-signal module B = Blank Module
DC = Canada (CSA) C = Carbon Dioxide 32 = Two 16-signal modules
E = EVAC/WAGD 48 = Three 16-signal modules
F = Future 64 = Four 16-signal modules
H = Hyperbaric Oxygen
N = Nitrous Oxide
O = Oxygen
T = Nitrogen
U = Utility Air
V = Vacuum
Examples:

DU16 = 16 signal Master Alarm


DUOVB = 2 gas Area Alarm – Oxygen, Vacuum and Blank Module
DUOFB = 1 gas Area Alarm – Oxygen, Future and Blank Module

DUOV16B = 2 gas Area Alarm – OXY, VAC & 16 signal Master Alarm and Blank Module
DU32 = 32 signal Master Alarm
DUOVANTB = 5 gas Area Alarm – Oxygen, Vacuum, Air, Nitrous Oxide, Nitrogen and Blank Module

DU64 = 64 signal Master Alarm


DUOVAB32 = 3 gas Area Alarm – OXY, VAC, AIR, Blank Module & 32 signal Master Alarm
DUOVANCTEU32 = 8 gas Area Alarm – OXY, VAC, AIR, N2O, CO2, N2, EVAC, Utility Air
& 32 signal Master Alarm

*Note – Medical Gas Alarms come in one of the three configurations shown above - 2 slots , 3 slots or 5 slots.
RITS - USA, LLC
PO Box 1015. Buford, GA 30515-1015, USA
Phone: +1- 770-831 1480 Fax: +1- 770-573 3712 E-mail: rits@rits-usa.com

Medical Gas Alarms - Area Alarm Panels


Features
The Tri-Tech Area Alarm Panel digitally displays gas
pressure (0.5 psi increments) and monitors and displays
normal and alarm conditions for up to 14 medical gases.
Transducers are included and may be mounted in the alarm
back box or remotely.

• ETL listed to UL-1069 and CSA-C22.2 No. 205


• Complies with NFPA 99. Made in the U.S.A.
• Self-contained unit - Designed for ease of installation and
service.
• Microprocessor controlled
• Self-diagnostic (test button) and error message display for ease
of maintenance.
• Audio and visual alarm indicators
• Bright easy to read L.E.D. displays – clearly visible in both day
and night lighting conditions
• Constant display and monitoring of each gas
• User programmable high/low set points
• Dry contacts for remote monitoring of all alarm conditions on (Area alarm shown is 3 gases – part # DUOAV)
each gas module and on the CPU module for the entire panel
• Alarm history display of previous alarm conditions In addition, each Area Alarm Module shall incorporate the
• Easy to read – color coded gas modules following features:
• Hinged frame with lanyards for easy accessibility • Does not require re-calibration
• Optional interface to the hospital TNET alarm information • Gas specific sensor with DISS nut & nipple. An error message
management system (area & master information). will be displayed if incorrect sensor or no sensor is attached.
• Three year PC board warranty • User programmable pressure limits (Programmed from factory at
60/40 psig and 12 in Hg)
Specification • Shall be capable of displaying gas readouts in PSI (in Hg), BAR
or kPa, button selected.
The alarm shall be the Tri-Tech Medical Area Alarm Panel. • Gas alarm repeat feature factory set at 10 minutes, adjustable
The panel shall be microprocessor controlled and designed from 1 minute to 999 minutes, or off
to comply with NFPA 99. The panel shall be 100% digital • Digital Transducers to be mounted inside the alarm for easy
and shall not require re-calibration. The alarm panel shall be access, or may be mounted remotely up to 5,000 ft (1,524 m)
enclosed in a steel box and shall be designed to accept an utilizing twisted pair wiring
electrical input range of 120-240 volts AC – 50-60 hertz. • Gas specific DISS risers with serviceable Frontall™ (front end
The source voltage shall be stepped down with a self- loaded) cartridge demand check valve
contained transformer. The panel shall contain audible and
visual alarm indicators. The audible alarm may be silenced
by pressing the alarm silence button, but the visual alarm
indicator can only be cancelled by fault correction. The
alarm shall detect and filter out transient (less than 0.6
seconds) signals created by R.F.I. The alarm shall be capable
of displaying alarm history for all possible alarm conditions.

Each vertical slot shall display up to three gases. The alarm


shall be capable of monitoring and displaying up to 14 gases
per alarm panel. Gas modules can be arranged in
accordance with the customer’s requirements.
RITS - USA, LLC
PO Box 1015. Buford, GA 30515-1015, USA
Phone: +1- 770-831 1480 Fax: +1- 770-573 3712 E-mail: rits@rits-usa.com

Medical Gas Alarms - Area Alarm Panels


RITS - USA, LLC
PO Box 1015. Buford, GA 30515-1015, USA
Phone: +1- 770-831 1480 Fax: +1- 770-573 3712 E-mail: rits@rits-usa.com

Medical Gas Alarms - Area Alarm Panels


Ordering Information:

Label Colors Area Alarm Gas Services |Master Remote Blank Slot for
Signals Future Expansion

U = USA (NFPA) O = Oxygen 16 = 16 points B = Blank Slot


C = Canada (CSA) V = Medical Vacuum 32 = 32 points
A = Medical Air 48 = 48 points
N = Nitrous Oxide 64 = 64 points
T = Nitrogen
C = Carbon Dioxide
W = WAGD/EVAC
S = AGSS
H = Hyperbaric Oxygen
U = Utility Air
L = Helium
I = Instrument Air (USA) Surgical Air (Canada)
D = Carbon Dioxide 80 psig
M = Gas Mixtures 50 psig
P = Gas Mixtures high pressure 180 psig
R = Tri-Gas
F = Future

Examples:

DU16 = 16 signal Master Alarm, USA colors, 2 slot box


DUOVB = 2 gas Area Alarm – Oxygen, Vacuum and Blank slot, USA colors, 2 slot box
DUOFB = 1 gas Area Alarm – Oxygen, Future and Blank slot, USA colors, 2 slot box
DUOVANT = 5 gas Area – OXY, VAC, AIR, N2O, N2 and Blank slot, USA colors, 2 slot box

DCOV16B = 2 gas Area – OXY, VAC & 16 signal Master & Blank slot, USA colors, 3 slot box
DU32 = 32 signal Master Alarm, USA colors, 3 slot box
DUOVANTB = 5 gas Area – OXY, VAC, AIR, N2O, N2 and Blank slot, USA colors, 3 slot box
DUOVANTCWMB = 8 gas Area, OXY, VAC, AIR, N2O, N2, CO2, WAGD, Gas Mixture and Blank
slot, USA colors, 3 slot box

DU64 = 64 signal Master, USA colors, 5 slot box


DUOVAB32 = 3 gas Area – OXY, VAC, AIR, Blank slot & 32 signal Master, USA colors, 5
slot box
DUOVANCTWU32 = 8 gas Area – OXY, VAC, AIR, N2O, CO2, N2, WAGD, Utility Air & 32
signal Master, USA colors, 5 slot box

*Note – Medical Gas Alarms come in one of the three configurations shown above - 2 slots , 3 slots or 5 slots.
RITS - USA, LLC
PO Box 1015. Buford, GA 30515-1015, USA
Phone: +1- 770-831 1480 Fax: +1- 770-573 3712 E-mail: rits@rits-usa.com

Medical Gas Alarms - Area Alarm Panels


Ordering Information:

Logic Module Logic Module Area Alarm Module Master Alarm Module Blank Module
With Buzzer With Buzzer (Choose 3 letters from 16 Signals per module
& Two Gas chart below – one for
(Area Alarm each gas service)
Modules)

Part Number Part Number Part Number Part Number


DU = USA (NFPA) A = Medical Air 16 = One 16-signal module B = Blank Module
DC = Canada (CSA) C = Carbon Dioxide 32 = Two 16-signal modules
E = EVAC/WAGD 48 = Three 16-signal modules
F = Future 64 = Four 16-signal modules
H = Hyperbaric Oxygen
N = Nitrous Oxide
O = Oxygen
T = Nitrogen
U = Utility Air
V = Vacuum
Examples:

DU16 = 16 signal Master Alarm


DUOVB = 2 gas Area Alarm – Oxygen, Vacuum and Blank Module
DUOFB = 1 gas Area Alarm – Oxygen, Future and Blank Module

DUOV16B = 2 gas Area Alarm – OXY, VAC & 16 signal Master Alarm and Blank Module
DU32 = 32 signal Master Alarm
DUOVANTB = 5 gas Area Alarm – Oxygen, Vacuum, Air, Nitrous Oxide, Nitrogen and Blank Module

DU64 = 64 signal Master Alarm


DUOVAB32 = 3 gas Area Alarm – OXY, VAC, AIR, Blank Module & 32 signal Master Alarm
DUOVANCTEU32 = 8 gas Area Alarm – OXY, VAC, AIR, N2O, CO2, N2, EVAC, Utility Air
& 32 signal Master Alarm

*Note – Medical Gas Alarms come in one of the three configurations shown above - 2 slots , 3 slots or 5 slots.
RITS - USA, LLC
PO Box 1015. Buford, GA 30515-1015, USA
Phone: +1- 770-831 1480 Fax: +1- 770-573 3712 E-mail: rits@rits-usa.com

Alarma de área Med Touch con Pantalla Táctil


 Sensor específico para gases con tuerca y boquilla DISS. Si falta
Características un sensor o se ha conectado uno incorrecto, aparecerá en pantalla
El panel de alarma de área Med Touch de Tri-Tech muestra de un mensaje de error
forma digital la presión de los gases (incrementos de 1 psi [6,9  Tendrá capacidad para mostrar lecturas de salida de los gases en
kPa]) y proporciona las condiciones de alarma exigidas por la PSI, kPa, BAR, inHg o mmHg
última edición de la norma NFPA 99 para hasta 16 gases  La función de repetición de alarma de gas está configurada de
medicinales. La alarma posee certificación ETL de cumplimiento fábrica en 10 minutos; se puede modificar o desactivar
con las normas UL 1069 y CSA C22.2 Nro. 205 de equipos de
señalización. Se incluyen los transductores. Se incluyen los tubos
verticales DISS; los transductores pueden instalarse remotamente o
en la caja posterior.

 Garantía de cinco años para las piezas y de un año para la


mano de obra
 Pantalla táctil LCD de 2,85” individual para cada servicio de
gas
 La pantalla táctil LCD permite la configuración y la
programación de todas las alarmas sin tener que utilizar
herramientas
 Una condición normal en color verde o de alarma en color rojo
para cada servicio de gas confirma el estado de cada servicio
de gas en particular
 Instrucciones de preparación para emergencias: los paneles de
alarma Med Touch permiten a los usuarios configurar
instrucciones personalizadas para cada señal de alarma que
aparece cuando la señal está en alarma
 Hasta 4 placas de área/gas en una caja de dos ranuras, 8 en un
panel de tres ranuras y 16 en un panel de cinco ranuras
 Historial de los últimos eventos (por placa de gas/punto de
señal) Especificación
 Fabricado en EE. UU. La alarma será la alarma de área Med Touch de Tri-Tech Medical. El
 “Etiquetas” de gas precargadas según normas NFPA e ISO panel se controlará mediante microprocesador y se diseñará conforme
 Idiomas precargados: inglés e inglés/francés a la norma NFPA 99. El panel incorporará una pantalla táctil LCD de
2,85” individual para cada servicio de gas que permitirá la
 Texto editable y “etiquetas” de alarma
configuración y la programación de todas las alarmas sin tener que
 Unidad compacta, diseñada para facilitar la instalación y el
utilizar herramientas. El panel será 100 % digital y no requerirá
servicio técnico
recalibración. La unidad estará dentro de una caja de acero, diseñada
 Pantalla de mensajes de error de autodiagnóstico para facilitar
para aceptar un rango de entrada eléctrica de 120 a 240 voltios CA,
el mantenimiento
de 50 a 60 hercios. La tensión de alimentación se reducirá con un
 Indicadores de alarma sonoros y visuales transformador autónomo. El panel contendrá indicadores de alarma
 Pantallas LCD luminosas, fáciles de leer: se ven con claridad sonoros y visuales. La alarma sonora puede silenciarse presionando el
tanto con iluminación diurna como nocturna botón de silencio de la alarma, pero el indicador de alarma visual
 Visualización y supervisión constantes de cada gas puede cancelarse solamente mediante la corrección de fallas. La
 Puntos de ajuste alto/bajo programables por el usuario alarma detectará y filtrará señales transitorias (de menos de 0,6
 Contactos secos para supervisión remota de todas las segundos) creadas por R.F.I. (Interferencia por radiofrecuencia). La
condiciones de alarma de cada módulo de gas y del módulo de alarma tendrá capacidad para mostrar en pantalla el historial de todas
la CPU (Unidad central de procesamiento) para todo el panel las condiciones posibles de alarma. La alarma tendrá capacidad para
 Fácil de leer: módulos de gases con código de colores supervisar y mostrar hasta 16 gases por cada panel de alarma. Los
 Marco abisagrado que facilita el acceso módulos de gases pueden disponerse según las necesidades del
 Pantallas con código de colores cliente. Se debe ingresar un código PIN de seguridad para poder
acceder al software de programación.
Además, cada módulo de alarma de área incluirá las siguientes
características:
 No requiere recalibración

Página 1 de 5
RITS - USA, LLC
PO Box 1015. Buford, GA 30515-1015, USA
Phone: +1- 770-831 1480 Fax: +1- 770-573 3712 E-mail: rits@rits-usa.com

Alarma de área Med Touch con Pantalla Táctil


Los siguientes opcionales se pueden agregar al panel en el momento de
realizar el pedido o después de su instalación: Pantalla de gas/área
 Notificación por mensaje de texto o correo electrónico opcional de los Se muestra en condición de alarma
eventos de alarmas con envío a 5 direcciones como máximo e indica las salas/áreas que
 Módulo Ethernet opcional. Este módulo se puede incorporar a controla. Cuando se presenta una
cualquier panel de alarma. Proporciona las siguientes funciones: condición de alarma, la pantalla
 Servidor web que permite el acceso remoto del usuario a la página parpadea llamando la atención y
web de la alarma para ver una imagen gráfica de la alarma con alterna las salas/áreas controladas
todas las condiciones de señal, información de registro de con las instrucciones de emergencia
historial/presión y eventos programadas por el usuario.
 Notificación por mensaje de texto de las condiciones de alarma
(envío a 5 direcciones como máximo)
 Registro de historial/eventos: mantiene una lista continua de las 100
condiciones de alarma más recientes y un archivo de los 1000 Se muestra en condición de alarma
eventos de alarma más recientes; puede descargarse a través de la e indica las instrucciones de
página web integrada emergencia. Cuando se presenta
 Interfaz Modbus una condición de alarma, la
 Es posible ver y descargar el registro de eventos a través de la página pantalla parpadea llamando la
web integrada que incluye los últimos 1000 eventos. atención y alterna las salas/áreas
 Conectividad por Ethernet opcional con página web integrada controladas con las instrucciones
 Interfaz con el sistema de automatización del edificio a través de de emergencia programadas por el
Ethernet usuario.

Se muestra en condición normal. El


código de colores específico y el
nombre del gas aparecen en la parte
superior en el idioma de la
instalación. La presión y las
unidades de medida del gas se
muestran en la parte media de la
pantalla. La condición normal o de
alarma se muestra en la parte
inferior central. Las salas/áreas
controladas se alternan con las
instrucciones de emergencia
programadas por el usuario en la
parte inferior.

Un menú de iconos fácil de utilizar


proporciona acceso rápido a las
opciones de programación.

La seguridad está incorporada en


cada tarjeta de circuitos. Se debe
ingresar el código PIN de cuatro
dígitos correcto para poder acceder
al software de programación.

Página 2 de 5
RITS - USA, LLC
PO Box 1015. Buford, GA 30515-1015, USA
Phone: +1- 770-831 1480 Fax: +1- 770-573 3712 E-mail: rits@rits-usa.com

Alarma de área Med Touch con Pantalla Táctil


Cómo hacer su pedido: Sistema de pedido modular fácil de utilizar.
Complete los espacios en blanco para especificar la alarma de área Med Touch que se ajuste a sus
necesidades.

Numeración del modelo:

T N C G G G G G G G G G G G G G G G G C

2 ranuras delanteras
3 ranuras delanteras
5 ranuras delanteras

C = Comunicación opcional
2 – Servidor web Ethernet, registro de eventos, notificaciones de
alarma (envío a 5 usuarios como máximo) e interfaz Modbus

G = Servicio de gases
A – Aire medicinal L – Helio
B – Ranura vacía – 4 posiciones (vacías) para ampliación vertical futura en una ranura N – Óxido nitroso
C – Dióxido de carbono – 50 psig (344,7 kPa) O - Oxígeno
D – Dióxido de carbono – 80 psig (551,6 kPa) R – Tri-gas
F – Ampliación futura S - Sistema de evacuación de gases anestésicos
H – Oxígeno hiperbárico T - Nitrógeno
I – Aire de instrumentación V – Vacío medicinal
J – Mezcla de oxígeno y dióxido de carbono DISS 1080 (CO2>7 %) W- Gas anestésico de desecho/Evacuación
K – Mezcla de oxígeno y helio DISS 1060 (He>80 %)

C = Tipos de gas
U – Etiquetado por Tri-Tech con código de colores según NFPA, inglés
I – Etiquetado por Tri-Tech con código de colores según ISO, inglés/francés

N = Cantidad de ranuras en el panel delantero


2 – Panel delantero con 2 ranuras 3 – Panel delantero con 3 ranuras
5 - Panel delantero con 5 ranuras

T = Tipo de alarma
T - Pantalla táctil digital

Página 3 de 5
RITS - USA, LLC
PO Box 1015. Buford, GA 30515-1015, USA
Phone: +1- 770-831 1480 Fax: +1- 770-573 3712 E-mail: rits@rits-usa.com

Alarma de área Med Touch con Pantalla Táctil


Ejemplos de n.o de pieza

T2UOVAF = alarma de área con 2 ranuras etiquetada por Tri-Tech, con código de colores según NFPA,
inglés. Tipos de gas oxígeno, vacío medicinal, aire medicinal, módulos de gas en 2. a ranura.
El último módulo en la 2.a ranura es para futura ampliación (vacío). Sin comunicación por
interfaz opcional.

T3IOVANTSFF2 = Alarma de área con 3 ranuras etiquetada por Tri-Tech, con código de colores según
ISO, inglés/francés. Tipos de gas oxígeno, vacío medicinal, aire medicinal y óxido
nitroso, en 2.a ranura, nitrógeno, sistema de evacuación de gases anestésicos y dos
módulos (vacíos) para futura ampliación en la 3.a ranura. Comunicación por interfaz
Modbus y servidor web Ethernet opcional.

T5UOVANHIFFBB2 = alarma de área con 5 ranuras etiquetada por Tri-Tech, con


código de colores según NFPA, inglés. Tipos de gas
oxígeno, vacío medicinal, aire medicinal y óxido nitroso,
en 2.a ranura, oxígeno hiperbárico, aire de
instrumentación y dos módulos (vacíos) para futura
ampliación en la 3.a ranura (las ranuras 4 y 5 están
vacías). Comunicación por interfaz Modbus y servidor
web Ethernet opcional. Comunicación opcional

Página 4 de 5
RITS - USA, LLC
PO Box 1015. Buford, GA 30515-1015, USA
Phone: +1- 770-831 1480 Fax: +1- 770-573 3712 E-mail: rits@rits-usa.com

Alarma de área Med Touch con Pantalla Táctil

Métodos habituales para la


instalación de transductores
remotos

En tubería con En válvula de zona con


una “T” y válvula de demanda específica de gas válvulas o kit de modernización E Z Find de Tri-Tech

Página 5 de 5

You might also like