Section 1 - Introduction: Ge Medical Systems Signa 1.5T CTL Phased Array Coil

Download as pdf or txt
Download as pdf or txt
You are on page 1of 44

GE MEDICAL SYSTEMS SIGNA 1.

5T CTL PHASED ARRAY COIL


REV 1 DIRECTION 46–003014

SECTION 1 – INTRODUCTION

1-1 HOW THE CTL PHASED ARRAY COIL OPERATES


The Cervical Thoracic Lumbar (CTL) Multicoil Array is a receive only MRI antenna designed to be used in 1.5T Signa
MRI systems equipped with the Multicoil Compatibility Kit (M1090PA).
The CTL array was designed and optimized for use in imaging the entire human spine. The array provides the high
SNR of smaller MRI surface coils with the large field of view (FOV=48cm) of larger MRI surface coils. The array con-
sists of six individual antenna loop elements, each of which can be used simultaneously within the limits of the system
architecture.
The overall length of the antenna portion of the array is 74.42 cm. This allows for three modes of operation, each
providing a 48 cm FOV. The CTLTOP mode utilizes antenna elements 1 through 4. The CTLMID mode utilizes anten-
na elements 2 through 5. And finally, the CTLBOT mode utilizes antenna elements 3 through 6.
The Coil consists of four major assemblies: (See Illustration 1–1.)
 Phase Adjustment Circuit Board (1) 2121448
 CTL Input Daughter Board (6) 2121446, 2121446–2, 2121446–3, 2121446–4, 2121446–5, and
2121446–6
 T/L Antenna Circuit Board (1) 2121444
 Cervical Multicoil Flexible Circuit (1) 46–307883P1
Each antenna loop element consists of an input daughter board and an antenna loop circuit. The input daughter
boards contain the input circuitry for each element, while the loop circuit defines the antenna geometry for each ele-
ment.
The antenna Loop circuits for elements 1 & 2 are located primarily on the contoured Cervical Multicoil Flexible Circuit
Board. A small portion of the antenna loop circuit for element 2 is located on the rigid T/L Antenna Circuit Board.
The antenna loop circuit for elements 3 through 6 are entirely contained on the T/L Antenna Board.
Antenna elements 1 and 2 are contoured and irregularly shaped while elements 3 through 6 are flat and rectangular
in shape. Each individual loop element is capacitively loaded with lumped element chip capacitors to reduce patient
loading effects. The total capacitance value of all the series chip capacitors combined is chosen in conjunction with
the input capacitor Cin to allow for a matched impedance (Zin=50) at the antenna input. When the coil is loaded,
a tuning capacitor is placed in parallel with one of the lumped element chip capacitors to allow for final tuning of the
loop input impedance.
The input circuitry on the six Input Daughter boards includes an input matching network and a transmit blocking net-
work for each antenna element.
The input circuitry on each board is identical, with the exception of component values, for all six array elements. The
input circuitry was put on separate boards so the heat producing components (PIN diodes) could be distanced away
from the patient surface without moving the antenna loop.
The function of the input matching network on this board is to match the input impedance of the individual antenna
loop element to the value desired by the preamp. The match is accomplished by a simple L section circuit. The input
capacitor, Cin, in conjunction with the antenna itself transforms the loaded input impedance up to 50 ohms and the
inductor, Lin, removes any minor reactive component of the impedance.

1–1 INTRODUCTION
GE MEDICAL SYSTEMS SIGNA 1.5T CTL PHASED ARRAY COIL
REV 1 DIRECTION 46–003014

1–1 HOW THE CTL PHASED ARRAY COIL OPERATES (continued)


A second function of the daughter boards is that of transmit blocking, which is performed by both the input matching
network and the transmit blocking network. When the PIN diodes are in the “on” state, the networks form high imped-
ances which are placed in series with the coil loop. This minimizes circulating currents in the loop and therefore mini-
mizes the magnetic field distortion during body coil transmit conditions.
The success of the blocked condition is dependent on the magnitude of the reactive impedances of Cin & Lin being
equal and the total resistance of the diode – inductor combinations being low. Equal reactances are guaranteed by
tuning the inductors to maximize blocked impedances and low resistance is guaranteed by correct DC bias of the
diode and correct tuning of the inductor.
During receive mode operation, it is important to the overall performance of the CTL array to decouple the individual
antenna loop elements. This is accomplished through two different methods.
For adjacent antenna loop elements, the decoupling is accomplished geometrically. The magnetic flux linking two
adjacent loops is cancelled by adjusting the overlap spacing between them. For both adjacent and non–adjacent an-
tenna elements, a second method of decoupling is used. The 1.5T low noise preamplifier used in the Signa MRI sys-
tem has a very small real input impedance (less than 6). If the electrical length of the system between the preamp
input and the input matching network of the antenna loop is kept at a multiple of a half–wavelength, then the impedance
seen at the diode terminals of the antenna loop’s input matching network will also be small and real. This small real
impedance at the diode terminals combined with Lin and Cin of the input matching network form a “lossy” parallel LC
blocking network similar to the transmit blocking network. This ”receive mode” blocking impedance reduces the indi-
vidual antenna loop current which in turn reduces the magnetic flux produced by the loop. This reduction in magnetic
flux in turn reduces the coupling between antenna elements.
The function of the phase adjust circuit board is to insure that the electrical length between the preamp and the anten-
na loop input is a multiple of a half wavelength, so that the second method of receive decoupling is working. The phase
adjustment is made with sections of microstrip line in the board.
”COIL 1” COAX CTL MOTHER/DAUGHTER
46–317421P1 JUMPER STRIPS (2) CIRCUIT BOARD ASSEMBLY
J1 2133084

”6 COAX”
EXTERNAL J1 CTL
RIBBON CTL INPUT CERVICAL MULTICOIL
30 PIN BENDIX CONNECTOR CABLE COIL 1 J7 INPUT DAUGHTER BD. CIRCUIT BOARD
PIN ASSIGNMENT 46–317151P1 DAUGHTER BD. (FOR COIL 2) (COILS 1 AND 2)
B3 – COIL1 COIL 2 J8 (FOR COIL 1) 2121446–2
FLEXIBLE CIRCUIT
B5 – COIL2 2121446 46–307883P1
B7 – COIL3 COIL 3 J9 J1
B9 – COIL4 COIL 4 J10
TO B11 – COIL5
MULTI. B13 – COIL6 COIL 5 J11
COIL A3 – JUMPER
PREAMP A5 – COIL 6 J12 JUMPER STRIPS (2)
STRIPS (2)
PROTECT. A7 – J2
A9 – ”COIL 2” COAX
A11 – 46–317421P2
A13 – PHASE T/L ANTENNA CIRCUIT BOARD
ADJUSTMENT (COILS 3, 4, 5, AND 6)
CIRCUIT 2121444
BOARD
2121448 JUMPER
STRIPS
(2) JUMPER
Lin STRIPS
CTL INPUT (2)
DAUGHTER BD.
(FOR COIL 3) CTL INPUT
PIN 2121446–3 DAUGHTER BD.
DIOD
Cin J3 ”COIL 3” COAX CABLE 46–317421P3
J1 (FOR COIL 4)
2121446–4 JUMPER
”COIL 4” COAX CABLE 46–317421P9 J1 STRIPS
ZMatch
E ZBlock J4
(2) JUMPER
J5 ”COIL 5” COAX CABLE 46–317421P10 STRIPS
J1 (2)
”COIL 6” COAX CABLE 46–317421P11 J1 CTL INPUT
IMPEDANCE MATCHING & J6 CTL INPUT DAUGHTER BD.
DAUGHTER BD. (FOR COIL 5)
TRANSMIT BLOCKING NETWORKS 2121446–5
(FOR COIL 6)
2121446–6

  
      


  

1–2 INTRODUCTION
GE MEDICAL SYSTEMS SIGNA 1.5T CTL PHASED ARRAY COIL
REV 1 DIRECTION 46–003014

1-2 COMPATIBILITY
The CTL Phased Array Coil is compatible with the following hardware configurations:
D Signa Advantaget (4.5 or higher)
D Signa Advantaget 1.5T (1.5T only)
D Signa Horizont 1.5T (1.5T only)

1-3 RELATED DOCUMENTS


D Direction 2124201, MR Release 5.x Signa Service Methods MR CD–ROM

D Direction 15400, Signa Advantage 1.5T, 1.0T, & 0.5T System


D Direction 15300, Signa Advantage System

1-4 ORGANIZATION OF THIS DOCUMENT


This manual is divided into the following sections:
Section 1, INTRODUCTION, describes how the CTL Phased Array Coil operates and when and where the CTL
Phased Array Coil can be used.
Section 2, SET UP AND CALIBRATION, describes installation procedures.
Section 3, FUNCTIONAL CHECKS, describes the normal power-up sequence.
Section 4, REPLACEMENT/MAINTENANCE, describes field maintenance procedures.
Section 5, RENEWAL PARTS, lists field replaceable parts.

1-5 ENVIRONMENTAL REQUIREMENTS


Operate and store the CTL Phased Array Coil in the Scanner Room.

1-6 SPECIFICATIONS

PARAMETER 

COIL DESIGN Six element receive only antenna array for human spine imaging

TUNED FREQUENCY 63.86 MHz (factory set)

OPTIMUM FIELD OF VIEW Group of four antenna loops out of six providing 48cm FOV

OPERATING AND STORAGE ENVI- Operate and store the CTL Phased Array Coil in the Scanner Room
RONMENT

1–3 INTRODUCTION
GE MEDICAL SYSTEMS SIGNA 1.5T CTL PHASED ARRAY COIL
Blank DIRECTION 46–003014

1–4 INTRODUCTION
GE MEDICAL SYSTEMS SIGNA 1.5T CTL PHASED ARRAY COIL
REV 1 DIRECTION 46–003014

SECTION 2 – SET UP AND CALIBRATION

2-1 CHECKING THE SHIPPING LIST


Table 2-1 lists the M1087CT Signa Advantage 1.5T CTL Phased Array Coil parts. Check that all parts have been
shipped.

TABLE 2-1
CTL PHASED ARRAY COIL SHIPPING LIST

QTY.  PART NO.

1 CTL Phased Array Coil and Cable 2104799

1 Neck Coil Pad 46–307593P1

1 Back Coil Pad 46–301961P1

5 Forehead Strap 46–317677P1

5 Chin Strap 46–317676P1

1 Positioner and six (6) Spherical Phantoms 46–317605G1

1 Operator’s Manual 2144814

1 Service Manual 46–003014

2-2 INSTALLING THE CTL PHASED ARRAY COIL


At the Boot Terminal, install a new soft key. This key will be used by the operator to select CTL TOP (elements 1–4),
CTL MIDDLE (elements 2–5), or CTL BOTTOM (elements 3–6) imaging. Name the keys “CTL TOP”, “CTL MID” or
“CTL BOT”. See Illustration 2–1.

CTL TOP

CTL MIDDLE

CTL BOTTOM

RCV1
RCV2 RCV3 RCV0 RCV1 RCV2 CTL ARRAY
THIS END
TOWARDS
COIL CONNECTOR

COIL NUMBERS – 1 2 3 4 5 6
M4074A

CTL PHASED ARRAY COIL TO TPS RECEIVER CORRESPONDENCE


ILLUSTRATION 2–1

2–1 SET UP AND CALIBRATION


GE MEDICAL SYSTEMS SIGNA 1.5T CTL PHASED ARRAY COIL
REV 1 DIRECTION 46–003014

2–2 INSTALLING THE CTL PHASED ARRAY COIL (continued)


Refer to references in Section 1–3, RELATED DOCUMENTS, for information on installing soft keys (use Coil/Phased
Array default values in Table 2–2 and 2–3).

TABLE 2-2
COIL VALUES

SYS- COIL COIL EXTREM- CABLE COIL LOSS RECON SCALE XMIT MULTI-
TEM NAME TYPE ITY COIL LOSS FACTOR COIL COIL
TYPE
 CTLTOP        Head Coil Recon Scale  
Factor x 1.00
  
 
CTLMID        Head Coil Recon Scale  
Factor x 1.00
  
 
CTLBOT        Head Coil Recon Scale  
Factor x 1.00
  
 
CS12        Head Coil Recon Scale  
Factor x 1.00
  
 
LS45        Head Coil Recon Scale  
Factor x 1.00
  
 
LS56        Head Coil Recon Scale  
Factor x 1.00
  
 

TABLE 2-3
PHASED ARRAY VALUES

MULTI–COIL NUMBER OF START STOP PORT ENABLE ERROR ENABLE


NAME RECEIVERS RECEIVER RECEIVER MASK MASK

CTLTOP 4 0 3 7 7

CTLMID 4 0 3 6 6

CTLBOT 4 0 3 14 14

CS12 2 1 2 3 3

LS45 2 0 1 4 4

LS56 2 1 2 12 12

2–2 SET UP AND CALIBRATION


GE MEDICAL SYSTEMS SIGNA 1.5T CTL PHASED ARRAY COIL
REV 1 DIRECTION 46–003014

2-3 FUNCTIONAL CHECKS


1. Perform a Body coil scan SNR verification. Refer to Section 3–1, BODY COIL SNR VERIFICATION.

2. Perform a CTL Phased Array Coil SNR Verification. Refer to Section 3–2, CTL PHASED ARRAY COIL SNR

VERIFICATION.

2-4 PERIODIC QUALITY ASSURANCE CHECK


On a periodic basis, such as during planned maintenance, perform the quality assurance check outlined below to en-
sure that the CTL Phased Array Coil is operating properly with no appreciable degradation of image quality.

1. Perform a CTL Phased Array Coil SNR Verification. Refer to Section 3–2, CTL PHASED ARRAY COIL SNR

VERIFICATION.

2. Record the date and values calculated in Section 3–3, SNR IMAGE ANALYSIS in columns 2–7 under “SNR Data
QA Check” of the Data Table as is instructed.

3. As is instructed in the Data Table, divide the SNR value obtained in the periodic QA check (recorded in columns
2–7) by the original SNR value and record in column 8 of the Data Table.

4. If this ratio is not greater than 85%, then there may be a problem in the Multicoil system. Contact your local GE
Service Representative.

2–3 SET UP AND CALIBRATION


GE MEDICAL SYSTEMS SIGNA 1.5T CTL PHASED ARRAY COIL
Blank DIRECTION 46–003014

2–4 SET UP AND CALIBRATION


GE MEDICAL SYSTEMS SIGNA 1.5T CTL PHASED ARRAY COIL
REV 1 DIRECTION 46–003014

SECTION 3 – FUNCTIONAL CHECKS

3-1 BODY COIL SNR VERIFICATION


Note
An alternate proprietary procedure is available for GE use and to customers with a valid Advanced
Service Package Limited License. Refer to ”TLT PROCEDURE” calibration procedure located in
Direction 2124201, MR Release 5.x Signa Service Methods or Direction 2124202, Release 4.x/3.x
MR Signa Service Methods.

Phantom Required (shipped with each corresponding CTL Phased Array Coil)
 CTL Phased Array Coil Positioner with phantoms (CTL Phantom/Positioner), 46–317605G1

Setup Procedure



The Quad Head Coil must be completely removed from the cradle before performing any
body scans. Failure to do this may result in damage to the Head Coil T/R Network.

1. Remove Quad Head Coil (if present) from cradle.


2. Select [New Exam] to allow a new Landmark to be set.
3. Position the CTL Phantom/Positioner with phantoms in the center of the cradle. Landmark the center of the posit-
ioner and advance the positioner with phantoms to isocenter using the [ADVANCE TO SCAN] button. See Il-
lustration 3–1.

(VIEW FROM LEFT SIDE OF CTL POSITIONER)

CTL TOP

CTL MID

CTL BOTTOM

PHANTOM

CRADLE
THIS END
TOWARDS LANDMARK HERE
THE MAGNET BORE

CTL PHANTOM/POSITIONER LANDMARK SETUP


ILLUSTRATION 3–1

3–1 FUNCTIONAL CHECKS


GE MEDICAL SYSTEMS SIGNA 1.5T CTL PHASED ARRAY COIL
REV 1 DIRECTION 46–003014

3–1 BODY COIL SNR VERIFICATION (continued)


4. Setup Scan Prescription as shown in Table 3–1 (4.x) or Table 3–2 (5.x).

TABLE 3–1
BODY COIL SNR – SCAN PROTOCOL (4.x)

SCAN PRESCRIPTION (4.x)

MAIN MENU [NEW STUDY] SCAN TIMING


Number of Echoes [1]
PATIENT STUDY PARAMETERS Echo Time (TE) [30 msec]
ID: GESERVICE See Note 1 Rep Time (TR) [300 msec]
Monitor SAR? N [NEXT PAGE]
[NEXT PAGE] (twice)
SCANNING RANGE
PATIENT POSITION Field of View [48 cm]
Patient Entry [Head First] Scan Thickness [3 mm]
Patient Position [Supine] Scan Location (S/I) 0
Coil Type [Body Coil] FOV Center (R/L) RO (A/P) A0
Axial/Sag. Landmark [Sternal Notch] [NEXT PAGE]
[NEXT PAGE]
ACQUISITION TIME
IMAGING PARAMETERS Acq. Matrix [256 x 256]
Image Mode [Single Scan] Imaging Time [1 NEX 1:58]
Scan Plane [Sagittal] Frequency Direction [R/L]
Pulse Sequence [Multiple Echo] Contrast [NO]
Imaging Option [None] Table Delta: 0
Enter PSD Filename: [NEXT PAGE]
[NEXT PAGE]
REVIEW [NEXT PAGE]

SCAN OPERATIONS

Note 1: GESERVICE ID automatically activates the research mode.

3–2 FUNCTIONAL CHECKS


GE MEDICAL SYSTEMS SIGNA 1.5T CTL PHASED ARRAY COIL
REV 1 DIRECTION 46–003014

3–1 BODY COIL SNR VERIFICATION (continued)

TABLE 3–2
BODY COIL SNR – SCAN PROTOCOL (5.x)

SCAN PRESCRIPTION (5.x)


MAIN MENU [New Exam] SCAN SET–UP
Prescan Options none
PATIENT/EXAM INFORMATION Auto CF [Peak]
ID: geservice [Scanning Range]
Name: body snr
Weight: 300 SCANNING RANGE
[Patient Position] Field of View [48 cm]
Scan Thickness [3 mm]
PATIENT POSITION Interscan Spacing [Other] 0
Patient Entry [Head First] Start Loc: (I/S) 0
Patient Position [Supine] End Loc: (I/S) 0
Axial/Sag. Landmark [Sternal Notch] No. of Scan locations 1
Coil Type [Body Coil] FOV Center: (L/R) 0 (P/A) 0
Scan Plane [Sagittal] [ ] [Acq Time]
[Imaging Params]
ACQUISITION TIME
IMAGING PARAMETERS Acq. Matrix (freq.) [256]
Image Mode [2D] Acq. Matrix (phase) [256]
(K SAR must be “On” K) [Monitor SAR] Frequency Direction [A/P]
Pulse Sequence [Spin Echo] Phase FOV [1.00 48cm]
Imaging Option [None] Imaging Time [1 NEX 1:58]
Enter PSD Filename: Contrast [NO]
[Next Screen] [Scan Timing] Table Delta: 0 mm
[Scan Ops]
SCAN TIMING
Number of Echoes [1] SCAN OPERATIONS
Echo Time (TE) [30 msec]
Rep Time (TR) [300 msec]
[Scanning Range]

5. Select [Auto Prescan] to properly calibrate the RF power level for the 90 degree and 180 degree pulses.
6. Select [Scan]. Observe the resulting images. Ensure that there are no artifacts of any sort in the resulting image.
Record the Exam number and Series number for SNR Calculations.
7. Select [Scan] again. This second image will be used for determination of Body Coil mode SNR.
8. Select [Cancel]. Refer to Section 3–3 for SNR image analysis.

3–3 FUNCTIONAL CHECKS


GE MEDICAL SYSTEMS SIGNA 1.5T CTL PHASED ARRAY COIL
REV 1 DIRECTION 46–003014

3-2 CTL PHASED ARRAY COIL SNR VERIFICATION


Note
An alternate proprietary procedure is available for GE use and to customers with a valid Advanced
Service Package Limited License. Refer to ”TLT PROCEDURE” calibration procedure located in
Direction 2124201, MR Release 5.x Signa Service Methods or Direction 2124202, Release 4.x/3.x
MR Signa Service Methods.

Phantom Required (shipped with each corresponding Phased Array Coil)


 CTL Phased Array Coil Positioner with phantoms (CTL Phantom/Positioner), 46–317605G1

Setup Procedure
1. Select [New Exam] to allow a new Landmark to be set.



The Quad head Coil must be completely removed from the cradle before performing any body
scans. Failure to do this may result in damage to the Head Coil T/R Network.

2. Remove Quad Head Coil (if present) from cradle.


3. Place CTL Phased Array Coil to be tested onto table.
4. Connect CTL Phased Array Coil connector to its mating connector in the Carriage Assembly.
5. Place the phantom/positioner down on the CTL Coil. Ensure the two pegs on the bottom of the positioner mate
with two positioning holes on the top cover of the CTL Coil Array. See Illustration 3–2.

3–4 FUNCTIONAL CHECKS


GE MEDICAL SYSTEMS SIGNA 1.5T CTL PHASED ARRAY COIL
REV 1 DIRECTION 46–003014

3–2 CTL PHASED ARRAY COIL SNR VERIFICATION (continued)

At each end of the CTL phantom is a


CTL PHANTOM/POSITIONER ‘peg’ extending from the bottom of the
end bracket. This ‘peg’ mates with
alignment holes in the coil cover.

COIL CONNECTOR

ALIGNMENT PEG (2)

CTL PHASED ARRAY COIL

ALIGNMENT
HOLE (2)

COIL COVER

Place the phantom/positioner down


on the coil, carefully inserting the
pegs into the alignment holes.

CTL PHANTOM/POSITIONER SETUP


ILLUSTRATION 3–2

3–5 FUNCTIONAL CHECKS


GE MEDICAL SYSTEMS SIGNA 1.5T CTL PHASED ARRAY COIL
REV 1 DIRECTION 46–003014

3–2 CTL PHASED ARRAY COIL SNR VERIFICATION (continued)


6. Position the CTL Phased Array Coil and Positioner with phantoms in the center of the cradle. Landmark for CTL
MID and advance the positioner with phantoms to isocenter using the [ADVANCE TO SCAN] button. See Illus-
tration 3–3.

(VIEW FROM LEFT SIDE OF CTL POSITIONER) Note: Testing the CTL coil requires
two scans. Landmark in the center
CTL TOP
of the coil (between the 3rd and 4th
phantom in the mid region). First
CTL MID
scan the CTL Top (S120), this tests
the top (end towards the coil connec-
CTL BOTTOM
tor) half of the array. Then scan the
CTL Bot (I120), this tests the bottom
PHANTOM ARRAY 1
2 3 4 5 6 (end away from the coil connector)
half of the array. The phantom
spheres are on 120mm centers.
CTL PHASED ARRAY COIL

THIS END
TOWARDS S120 I120
COIL CONNECTOR LANDMARK HERE

Note: The CTL array contains six coils (only four are used at one time). Three regions are
designated (landmarks for each region are marked on the side of the array). The CTL Top array
uses coils 1 through 4 (Black Marking) (counting from the head end). CTL Mid uses coils 2
through 5 (Blue Marking). CTL Bot uses coils 3 through 6 (Black Marking). If CTL Coil is land-
marked as above, then Scan Location for CTL Top is S120 and CTL Bot is I120.
CTL PHANTOM/POSITIONER LANDMARK SETUP (CTL TOP)
ILLUSTRATION 3–3

Phased Array Coil Receiver Paths


The Phased Array Coils have multiple individual antenna loop elements inside the assembly; the TPS chassis has
multiple receivers (RCV0–RCV3). See Illustration 3–4.

CTL TOP

CTL MID

CTL BOTTOM

THIS END RCV1


TOWARDS RCV2 RCV3 RCV0 RCV1 RCV2
COIL CONNECTOR

CTL ARRAY

COIL NUMBERS – 1 2 3 4 5 6
M4074A

CTL PHASED ARRAY COIL TO TPS RECEIVER CORRESPONDENCE


ILLUSTRATION 3–4

3–6 FUNCTIONAL CHECKS


GE MEDICAL SYSTEMS SIGNA 1.5T CTL PHASED ARRAY COIL
REV 1 DIRECTION 46–003014

3–2 CTL PHASED ARRAY COIL SNR VERIFICATION (continued)


7. Setup Scan Prescription as shown in Table 3–3 (4.x) or Table 3–4 (5.x).

TABLE 3–3
CTL PHASED ARRAY COIL SNR – SCAN PROTOCOL (4.x)

SCAN PRESCRIPTION (4.x)

MAIN MENU [NEW STUDY] SCAN TIMING


Number of Echoes [1]
PATIENT STUDY PARAMETERS Echo Time (TE) [30 msec]
ID: GESERVICE See Note 1 Rep Time (TR) [300 msec]
Monitor SAR? N [NEXT PAGE]
[NEXT PAGE] (twice)
SCANNING RANGE
PATIENT POSITION Field of View [48 cm]
Patient Entry [Head First] Scan Thickness [3 mm]
Patient Position [Supine] Scan Location (S/I) S120
Coil Type [Other Coils] [CTL TOP] FOV Center (R/L) RO (A/P) A0
Axial/Sag. Landmark [Sternal Notch] [NEXT PAGE]
[NEXT PAGE]
ACQUISITION TIME
IMAGING PARAMETERS Acq. Matrix [256 x 256]
Image Mode [Single Scan] Imaging Time [1 NEX 1:58]
Scan Plane [Sagittal] Frequency Direction [R/L]
Pulse Sequence [Multiple Echo] Contrast [NO]
Imaging Option [None] Table Delta: 0
Enter PSD Filename: [NEXT PAGE]
[NEXT PAGE]
REVIEW [[NEXT PAGE]

SCAN OPERATIONS

Note 1: GESERVICE ID automatically activates the research mode.

3–7 FUNCTIONAL CHECKS


GE MEDICAL SYSTEMS SIGNA 1.5T CTL PHASED ARRAY COIL
REV 1 DIRECTION 46–003014

3–2 CTL PHASED ARRAY COIL SNR VERIFICATION (continued)

TABLE 3–4
CTL PHASED ARRAY COIL SNR – SCAN PROTOCOL (5.x)

SCAN PRESCRIPTION (5.x)


MAIN MENU [New Exam] SCAN SET–UP
Prescan Options None
PATIENT/EXAM INFORMATION Auto CF [Peak]
ID: geservice [Scanning Range]
Name: body snr
Weight: 300 SCANNING RANGE
[Patient Position] Field of View [48 cm]
Scan Thickness [3 mm]
PATIENT POSITION Interscan Spacing [Other] 0
Patient Entry [Head First] Start Loc: (I/S) S120
Patient Position [Supine] End Loc: (I/S) S120
Axial/Sag. Landmark [Sternal Notch] No. of Scan locations 1
Coil Type [Other Coils] [CTL TOP] FOV Center: (L/R) 0 (P/A) 0
Scan Plane [Sagittal] [ ] [Acq Time]
[Imaging Params]
ACQUISITION TIME
IMAGING PARAMETERS Acq. Matrix (freq.) [256]
Image Mode [2D] Acq. Matrix (phase) [256]
(K SAR must be “On” K) [Monitor SAR] Frequency Direction [R/L]
Pulse Sequence [Spin Echo] Phase FOV [1.00 48cm]
Imaging Option [None] Imaging Time [1 NEX 1:58]
Enter PSD Filename: Contrast [NO]
[Next Screen] [Scan Timing] Table Delta: 0 mm
[Scan Ops]
SCAN TIMING
Number of Echoes [1] SCAN OPERATIONS
Echo Time (TE) [30 msec]
Rep Time (TR) [300 msec]
[Scan Set–Up]

8. Select [Auto Prescan] to properly calibrate the RF power level for the 90 degree and 180 degree pulses.
9. Select [Scan]. Observe the resulting image of the spheres. Ensure that there are no artifacts of any sort in the
sphere images. Record the Exam number and Series number for SNR Calculations.
10. Select [Scan] again. This second image of the spheres will be used for determination of CTL TOP mode SNR.
11. Select [Cancel] and [New Series].

3–8 FUNCTIONAL CHECKS


GE MEDICAL SYSTEMS SIGNA 1.5T CTL PHASED ARRAY COIL
REV 1 DIRECTION 46–003014

3–2 CTL PHASED ARRAY COIL SNR VERIFICATION (continued)


12. Ensure the landmark is still for CTL MID. See Illustration 3–5.

(VIEW FROM LEFT SIDE OF CTL POSITIONER) Note: Testing the CTL coil requires
two scans. Landmark in the center
CTL TOP
of the coil (between the 3rd and 4th
phantom in the mid region). First
CTL MID
scan the CTL Top (S120), this tests
the top (end towards the coil connec-
CTL BOTTOM
tor) half of the array. Then scan the
CTL Bot (I120), this tests the bottom
PHANTOM ARRAY 1
2 3 4 5 6 (end away from the coil connector)
half of the array. The phantom
spheres are on 120mm centers.
CTL PHASED ARRAY COIL

THIS END
TOWARDS S120 I120
COIL CONNECTOR LANDMARK HERE

Note: The CTL array contains six coils (only four are used at one time). Three regions are
designated (landmarks for each region are marked on the side of the array). The CTL Top array
uses coils 1 through 4 (Black Marking) (counting from the head end). CTL Mid uses coils 2
through 5 (Blue Marking). CTL Bot uses coils 3 through 6 (Black Marking). If CTL Coil is land-
marked as above, then Scan Location for CTL Top is S120 and CTL Bot is I120.
CTL PHANTOM/POSITIONER LANDMARK SETUP (CTL BOTTOM)
ILLUSTRATION 3–5

13. Select [Patient Position], [Other Coils], [CTL BOT].


14. Select [Scanning Range], Scan Location as I120 (4.x) or Start & End Loc as I120 (5.x), and [Accept]. Advance
to SCAN OPERATIONS.
15. Select [Auto Prescan] to properly calibrate the RF power level for the 90 degree and 180 degree pulses.
16. Select [Scan]. Observe the resulting image of the spheres. Ensure that there are no artifacts of any sort in the
sphere images. Record the Exam number and Series number for SNR Calculations.
17. Select [Scan] again. This second image of the spheres will be used for determination of CTL BOT mode SNR.
18. Select [Cancel]. Refer to Section 3–3 for SNR image analysis.

3–9 FUNCTIONAL CHECKS


GE MEDICAL SYSTEMS SIGNA 1.5T CTL PHASED ARRAY COIL
REV 1 DIRECTION 46–003014

3-3 SNR IMAGE ANALYSIS

3–3–1 SNR Image Analysis (Release 4.X) (See Section 3–3–2 For Release 5.X)

Description
The CLIPS macro will ask for the first image in the series to analyze. The macro assumes you have done two back–to–
back scans, and will take the designated image and the next image in the series (i.e. if you give it image 3, it will take 3
and 4). It will also ask you to designate three points on the border of the phantom image to use in positioning the
analysis ROI. If your image is too big, too small, or the three points were selected too close together, the macro will
inform you.

Procedure
1. Touch [UTILITIES], then [Clips].
CLIPS
–––––––––––––––––––––––––
1 Run CLIPS
2 Save files
3 Restore files
4 Remove files
5 Exit
Enter the number of your selection 1 

1 = Operator console
2 = Remote1
3 = Remote2
4 = Exit
Which image processor would you like to use (1 2 or 3) ? 1 
IP selected is at the operator console.
Do you wish to boot the Image Processor? (Y or N)
Y     
      
   
Welcome to the Command Line Image Processing System (CLIPS)
CLIPS > LIST(STUDY); 

3–10 FUNCTIONAL CHECKS


GE MEDICAL SYSTEMS SIGNA 1.5T CTL PHASED ARRAY COIL
REV 1 DIRECTION 46–003014

3–3–1 SNR Image Analysis (Release 4.X)(continued)


STUDY LIST
========================================================================
STUDY # PATIENT NAME PATIENT I.D. DATE

–––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––
00001 BDY_SNR GESERVICE 9–JULY–89
00002 HD_SNR GESERVICE 9–JULY–89
––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––
CURRENT DEFAULTS ––– PATIENT ID = / /
N – NEXT PAGE
P – PREVIOUS PAGE
S – SELECT STUDY
C – CANCEL
choice ?
S 
ENTER STUDY NUMBER ––
1     
  
CLIPS > LIST(SERIES); 

SERIES LIST
========================================================================
SERIES # # OF IMAGES SERIES TYPE PULSE TYPE
––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––
001 002 AXIAL ME
002 002 SAG ME
003 002 COR ME
––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––––
CURRENT DEFAULTS ––– PATIENT ID = SNR 0000S/00S/001
N – NEXT PAGE
P – PREVIOUS PAGE
S – SELECT SERIES
C – CANCEL
choice ?
S 
ENTER SERIES NUMBER ––
1         
CLIPS > DIS;    
LIST(IMAGE);       


3–11 FUNCTIONAL CHECKS


GE MEDICAL SYSTEMS SIGNA 1.5T CTL PHASED ARRAY COIL
REV 1 DIRECTION 46–003014

3–3–1 SNR Image Analysis (Release 4.X)(continued)


CLIPS > EXECUTE(SNR); ENTER or just EXE(SNR); ENTER .
THIS MACRO CALCULATES THE SNR USING THE GE HEAD OR BODY PHANTOM.
ENTER THE NUMBER OF THE FIRST IMAGE OF THE TWO–IMAGE SET AND !
EXAMPLE INPUT> 1!
Please start entering data. Type an ! to end input mode.
CLIPS INPUT> 1! ENTER
CORRECT POSITIONING OF THE SNR ANALYSIS ROI DEPENDS ON
FINDING THE CENTER OF THE PHANTOM. YOU WILL BE PROMPTED
TO IDENTIFY THREE POINTS AT THE EDGE OF THE PHANTOM.
CARE SHOULD BE TAKEN TO PICK THREE POINTS AS FAR AWAY FROM
EACH OTHER AS POSSIBLE (I.E. 2, 6, AND 10 O CLOCK) AND
AVOID FLAT SPOTS AND IRREGULARITIES ON THE EDGE. See Illustration 3-6.
Note
The order of cursor placement must be 10, 6, then 2 o’clock positions. Using a different order may
give an error.

1st
3rd

2nd
M2136A

SNR CURSOR POSITIONS


ILLUSTRATION 3–6

POSITION CURSOR AT FIRST POINT AND ENTER !


Place cursor at 10:00 position on edge of image.
Please start entering data. Type an ! to end input mode.
CLIPS INPUT> ! ENTER
POSITION CURSOR AT SECOND POINT AND ENTER ! Place cursor at 6:00 position.
Please start entering data. Type an ! to end input mode.
CLIPS INPUT> ! ENTER

3–12 FUNCTIONAL CHECKS


GE MEDICAL SYSTEMS SIGNA 1.5T CTL PHASED ARRAY COIL
REV 1 DIRECTION 46–003014

3–3–1 SNR Image Analysis (Release 4.X)(continued)


POSITION CURSOR AT THIRD POINT AND ENTER ! Place cursor at 2:00 position.
Please start entering data. Type an ! to end input mode.
CLIPS INPUT> ! ENTER
PHANTOM POINTS AND RADIUS OK. BEGIN ANALYSIS.
The program will then perform the analysis and disĆ
play the final values on the image monitor. Record
Signal, Noise and Signal to Noise in Data Table.
DO YOU WANT TO SAVE THIS DATA?
ENTER A 1! FOR YES, A 0! FOR NO.
Please start entering data. Type an ! to end input mode.
CLIPS INPUT> 0! ENTER
DATA NOT SAVED
CLIPS > LIST(SERIES); ENTER Select the next series and repeat the above analysis
for each image pair. Type BYE; ENTER when comĆ
pleted to exit CLIPS.
2. When all analyses are complete, reset Auto Center Frequency mode by touching [SCAN MODES],
[Default Auto CF], then [EXECUTE].
Note
You do not have to reset the Auto Center Frequency mode if you are proceeding on to more image
quality scans.
3. Record the date and value calculated in column 2 under “SNR Data QA Check” of the Data Table on page DT–1
(following Section 5, RENEWAL PARTS) as is instructed.

3–13 FUNCTIONAL CHECKS


GE MEDICAL SYSTEMS SIGNA 1.5T CTL PHASED ARRAY COIL
REV 1 DIRECTION 46–003014

3–3–2 SNR Image Analysis (Release 5.X)

Description
The SNR tool retrieves two operator selected images. Signal value is computed as the mean pixel value in a ROI
covering 80% of the image. The image is analyzed to determine the center of the image for positioning the ROI. A
difference image is created by subtracting the second image from the first and the same ROI is used to calculate noise
from the subtracted image. The signal value, noise value, and signal to noise ratio are reported. There is an option to
save the difference image with the results annotated.

Procedure
1. Touch [UTILITIES], [MR Tools], [Image Quality], then [SNR Test]. The SNR Test screen is displayed; see
Illustration 3–7.
2. Enter first image exam, series, and image numbers. If exam, series, or image numbers are not known, select
[List Exams], [List Series], or [List Images] to display list to choose from.
Note
Image number selection must be back lit (highlighted) to be able to enter information. Use Switch key
on keyboard to transfer control from left to right side of Touch Screen.
3. Enter second image series and image numbers. If series or image numbers are not known, select [List Series]
or [List Images] to display list to choose from. The second image Exam Number is defaulted to the same as first
image.

NEMA STANDARD SNR TEST

Filter Image
Off Off

First List/Sel List/Sel List/Sel


Image Exams Series Images

xxx x x

Second List/Sel List/Sel


Image Series Images

x x

******************************
Body Axial SNR Data

Signal : xx.x
Noise : xx.x
SNR : xx.x

ÍÍÍ
******************************

ÍÍÍ
ÍÍÍ
Cancel Utility MR Tools Image Accept
Main Main Quality

ÍÍ
ÍÍ Note: Accept changes to continue only after an analysis has been performed.

SNR TEST SCREEN


ILLUSTRATION 3–7

3–14 FUNCTIONAL CHECKS


GE MEDICAL SYSTEMS SIGNA 1.5T CTL PHASED ARRAY COIL
REV 1 DIRECTION 46–003014

3–3–2 SNR Image Analysis (Release 5.X) (continued)


4. If high pass filtering is desired to be performed on data, touch [Filter Off] which will highlight and change to [Filter
On].
5. If the difference image annotated with data is to be created, touch [Image Off] which will highlight and change to
[Image On].
6. Touch [Accept] to begin analysis. The final values are displayed on the touch screen, see Illustration 3–7.
7. Touch [Continue] then select the next exam and repeat the above analysis for each image pair.
8. Record the date and value calculated in columns 2–7 under “SNR Data QA Check” of the Data Table on page
DT–1 (following Section 5, RENEWAL PARTS) as is instructed.

3–15 FUNCTIONAL CHECKS


GE MEDICAL SYSTEMS SIGNA 1.5T CTL PHASED ARRAY COIL
REV 1 DIRECTION 46–003014

3-4 CHECKING THE PROTECTION PIN DIODES WITH DIGITAL MULTIMETER (DMM)
Note
Due to the fact that there are two PIN diodes in parallel for each antenna element, this procedure will
not indicate if one of the PIN diodes is defective, only if both are defective.
1. Select the DIODE TEST function on the Digital Multimeter (DMM).
2. Connect the NEGATIVE lead of the DMM to the external cable connector row A. Connect the POSITIVE lead
to the external cable connector row B. Refer to Table 3–5.
3. A reading of 0.400 to 0.900 should be observed on the DMM.
4. If a reading below 0.400 is observed in either direction, either the output cable is shorted or bad PIN diode.
5. If a reading above 0.900 is observed in step 2, both PIN diodes are defective.
6. Connect the NEGATIVE lead of the DMM to the external cable connector row B. Connect the POSITIVE lead
to the external cable connector row A. Refer to Table 3–5.
7. A reading of INFINITY should be observed on the DMM.
8. If a reading of INFINITY is observed in both directions, either the output cable is open or both PIN diodes are
open.
9. If any of the above conditions fails, replace the coil. The Pin diodes are not FRU’s.

TABLE 3–5
DIODE TEST CONNECTIONS

POSITIVE LEAD CONNECTION NEGATIVE LEAD CONNECTION

COIL NUMBER FOR STEP 2 FOR STEP 4 FOR STEP 2 FOR STEP 4

1 B3 A3 A3 B3

2 B5 A5 A5 B5

3 B7 A7 A7 B7

4 B9 A9 A9 B9

5 B11 A11 A11 B11

6 B13 A13 A13 B13

3–16 FUNCTIONAL CHECKS


GE MEDICAL SYSTEMS SIGNA 1.5T CTL PHASED ARRAY COIL
REV 1 DIRECTION 46–003014

3-5 CHECKING THE EXTERNAL CABLE


1. Select the DIODE TEST function on the Digital Multimeter (DMM).
2. Connect the NEGATIVE lead of the DMM to the external cable connector row A. Connect the POSITIVE lead
to the external cable connector row B. Refer to Table 3–5.
3. Flex the external cable, especially near the connectors and the strain relief, and observe that a reading of 0.400
to 0.900 should remain on the DMM, with no instability or fluctuations.
4. Connect the NEGATIVE lead of the DMM to the external cable connector row B. Connect the POSITIVE lead
to the external cable connector row A. Refer to Table 3–5.
5. Flex the external cable, especially near the connectors and the strain relief, and observe that a reading of INFIN-
ITY should remain on the DMM, with no instability or fluctuations.
6. If the cable fails any of the above tests, it should be replaced.

3–17 FUNCTIONAL CHECKS


GE MEDICAL SYSTEMS SIGNA 1.5T CTL PHASED ARRAY COIL
Blank DIRECTION 46–003014

3–18 FUNCTIONAL CHECKS


GE MEDICAL SYSTEMS SIGNA 1.5T CTL PHASED ARRAY COIL
REV 1 DIRECTION 46–003014

SECTION 4 – REPLACEMENT/MAINTENANCE

Note
Each antenna loop element is tuned at the factory to provide the proper input impedance. Since an-
tenna elements 1 and 2 are contained on three separate circuit boards, none of these boards are field
replaceable. These circuits must remain together as a matched set.

4-1 DISASSEMBLY/ASSEMBLY OF CTL PHASED ARRAY COIL


1. Remove the two foam pads. See Illustration 4–1.
2. Remove the sixteen screws holding the top and bottom housing together.
3. Separate the two housings from each other.
4. To reassemble, put the two housings together and lightly tighten the sixteen screws.
Note
The screws are very susceptible to breakage from overtightening.

PAD

PAD

SCREWS
2104799
PAD
PAD

35

2132617
DISASSEMBLY/ASSEMBLY OF CTL PHASED ARRAY COIL
ILLUSTRATION 4–1

  
 
   
GE MEDICAL SYSTEMS SIGNA 1.5T CTL PHASED ARRAY COIL
REV 1 DIRECTION 46–003014

4-2 REPLACING THE EXTERNAL CABLE


1. Disassemble the top and bottom housings. Refer to Section 4–1.
2. Remove the two screws and remove the clamp holding the external cable in place.
3. Disconnect the six SMB connectors on the external cable from the Phase Adjustment Board. See Illustration
4–2.
4. Remove the bad external cable.
5. Install the new external cable in place. See Illustration 4–2.
6. Install the clamp and screws in place.
7. Reconnect the six SMB connectors to the Phase Adjustment Board in the proper order.
8. Assemble the top and bottom housings together. Refer to Section 4–1.

SCREW (2)
CLAMP
CABLE CONNECTIONS
EXTERNAL CABLE PHASE ADJ. BD.
SMB CONNECTORS (6)
EXTERNAL
CABLE J7 J7
J8 J8
J9 J9
PHASE ADJUSTMENT BOARD J10 J10
J11 J11
J12 J12

M301800A
SHT 2

REPLACEMENT OF EXTERNAL CABLE


ILLUSTRATION 4–2

  
 
   
GE MEDICAL SYSTEMS SIGNA 1.5T CTL PHASED ARRAY COIL
REV 1 DIRECTION 46–003014

4-3 REPLACING THE INTERNAL CABLES


Note
Cable lengths are tuned. Therefore, internal cable replacement must involve the correct replacement
cable of appropriate length.
1. Disassemble the top and bottom housings. Refer to Section 4–1.
2. Remove the cable ties and the clamps holding the internal cables in place.
3. Disconnect the bad cable from the Phase Adjustment Board and appropriate coil board. See Illustration 4–3.
4. Remove the bad internal cable.
5. Install the new internal cable to the Phase Adjustment Board and appropriate coil board. See Illustration 4–3.
6. Install the clamps and cable ties in place. Use non–magnetic cable ties (46–208758P8).
7. Assemble the top and bottom housings together. Refer to Section 4–1.

CABLE TIE

CLAMP

DAUGHTER
BOARD

REPLACEMENT OF INTERNAL CABLES


ILLUSTRATION 4–3

  
 
   
GE MEDICAL SYSTEMS SIGNA 1.5T CTL PHASED ARRAY COIL
REV 1 DIRECTION 46–003014

4-4 REPLACING THE PHASE ADJUSTMENT CIRCUIT BOARD


1. Disassemble the top and bottom housings. Refer to Section 4–1.
2. Disconnect the twelve SMB connectors from the Phase Adjustment Board. See Illustration 4–4.
3. Remove the two screws and spacers holding the Phase Adjustment Board in place.
4. Remove the bad Phase Adjustment Board (2121448).
5. Install the new Phase Adjustment Board in place. See Illustration 4–4.
6. Install the spacers and screws on the Phase Adjustment Board.
7. Reconnect the twelve SMB connectors to the Phase Adjustment Board.
8. Assemble the top and bottom housings together. Refer to Section 4–1.

J1
J2
J3
J4
J5
J6

SCREW
SPACER

PHASE ADJUSTMENT
SCREW BOARD
CLAMP

EXTERNAL
CABLE

REPLACEMENT OF PHASE ADJUSTMENT BOARD


ILLUSTRATION 4–4

   
 



GE MEDICAL SYSTEMS SIGNA 1.5T CTL PHASED ARRAY COIL
REV 1 DIRECTION 46–003014

4-5 REPLACING THE PROTECTION PIN DIODES



As coil tuning and decoupling tuning may be affected by PIN diode replacement, field re-
placement of the PIN diodes will not be performed.

4-6 REPLACING THE MECHANICAL HARDWARE


1. Disassemble the top and bottom housings, if necessary. Refer to Section 4–1.
2. Disconnect and remove the bad part. Refer to Section 5, Renewal Parts, for part numbers.
3. Install the new part in place.
4. Assemble the top and bottom housings together. Refer to Section 4–1.

4-7 CHECKING THE CABLES


Check the external cable for cracks or breaks once each week. Replace the external cable if any damage or wear
is found. Refer to Section 4–2, REPLACING THE EXTERNAL CABLE, for external cable replacement procedure.

4-8 CLEANING THE CTL PHASED ARRAY COIL



Avoid damaging sensitive electronic parts. Do not spray or pour dishwashing solution
directly onto the CTL Phased Array Coil, or external cable. Do not use alcohol to clean the
CTL Phased Array Coil. Never submerge the CTL Phased Array Coil in any liquid.

Clean the CTL Phased Array Coil and external cable with a mild dishwashing liquid and water solution. Wet a soft
cloth with the solution and proceed to clean.
The removable pads can be cleaned with a 10% bleach solution.

  
 
   
GE MEDICAL SYSTEMS SIGNA 1.5T CTL PHASED ARRAY COIL
Blank DIRECTION 46–003014

  
 
   
GE MEDICAL SYSTEMS SIGNA 1.5T CTL PHASED ARRAY COIL
REV 1 DIRECTION 46–003014

SECTION 5 – RENEWAL PARTS

g iNRAW n

BIO HAZARD! EQUIPMENT BEING RETURNED FROM USE IN A CLINICAL SETTING MUST
BE CLEAN AND FREE OF BLOOD AND OTHER INFECTIOUS SUBSTANCES. THE DEPART-
MENT OF TRANSPORTATION (DOT) HAS RULED THAT “ITEMS THAT WERE SATURATED
AND/OR DRIPPING WITH HUMAN BLOOD THAT ARE NOW CAKED WITH DRIED BLOOD; OR
WHICH WERE USED OR INTENDED FOR USE IN PATIENT CARE” ARE “REGULATED MEDI-
CAL WASTE” FOR TRANSPORTATION PURPOSES AND MUST BE TRANSPORTED AS A
HAZARDOUS MATERIAL. UNDER NO CIRCUMSTANCES SHOULD A PART OR EQUIPMENT
WITH VISIBLE BODY FLUIDS BE TAKEN OR SHIPPED FROM A CLINIC OR SITE (FOR EXAM-
PLE, SURFACE COILS).

1. Employees shall follow proper decontamination procedures for clean up of bloodborne pathogens. Refer to Sec-
tion 4–8, CLEANING THE CTL PHASED ARRAY COIL. It is the responsibility of the GEMS employee to insure the
part/equipment has been properly decontaminated prior to shipment.

5–1 RENEWAL PARTS


GE MEDICAL SYSTEMS SIGNA 1.5T CTL PHASED ARRAY COIL
REV 1 DIRECTION 46–003014

CTL PHASED ARRAY COIL 2132617 (SEE PAGE 5–4 FOR PARTS LIST)

25
2

84

72

29 27 35
28

BOTTOM VIEW

M2132617A
SHT 1

5–2 RENEWAL PARTS


GE MEDICAL SYSTEMS SIGNA 1.5T CTL PHASED ARRAY COIL
REV 1 DIRECTION 46–003014

CTL PHASED ARRAY ARRAY 2132617 (ILLUSTRATION CONT’D FROM PG 5–1)

34

33
5 32
6
22
31

37
30
77 8
36
36 31

36
30

L2132617A

5–3 RENEWAL PARTS


GE MEDICAL SYSTEMS SIGNA 1.5T CTL PHASED ARRAY COIL
REV 1 DIRECTION 46–003014

CTL PHASED ARRAY ARRAY 2132617

Item Part Number FRU Name Quantity Description (Remarks)


1 2132617 1 CTL COIL XXX CTL PHASED ARRAY COIL
2 46–301961P1 2 PAD 1 GREY NYLON FABRIC W/VELCRO HOOKS
5 2133085 2 CABLE ASSEM 1 CTL COIL 6–COAX, 30” EXTERNAL CABLE
6 2121448 2 CIRCUIT BD 1 PHASE ADJUSTMENT CIRCUIT BOARD

22 46–208758P8 2 STRAP 4 1/16” TO 1–1/4” DIA. WIRE BUNDLE


25 46–307593P1 2 PAD 1 NECK COIL PAD, NYLON W/VELCRO HOOKS
27 46–317475P1 2 VELCRO PAD 6 2”W X 2”L, WHITE, ADHESIVE BACKED
28 46–317475P2 2 VELCRO PAD 1 2”W X 4”L, WHITE, ADHESIVE BACKED
29 46–317476P2 2 VELCRO PAD 2 2”W X 4”L, WHITE, ADHESIVE BACKED

30 46–230007P9 2 SCREW,MACH 5 10–32 X 5/8” BIND HEAD SLOT, DELRIN


31 46–230007P8 2 SCREW,MACH 5 10–32 X 3/8” BIND HEAD SLOT, DELRIN
32 46–220184P40 2 SCREW MACH 2 10–32 X 0.375 LG RD HD SLOT BRASS
33 46–208913P12 2 SCREW,MACH 2 10–32 X 0.50 LG SLOT FH BRS
34 46–220184P45 2 SCREW MACH 2 10–32 X 1.000 LG RD HD SLOT BRASS

35 46–220184P41 2 SCREW,MACH 16 10–32 X 0.500 BRASS


36 46–230007P13 2 SCREW,MACH 4 10–32 X 1 DELRIN AF
37 46–220184P42 2 SCREW MACH 8 10–32 X 0.625 LG RD HD SLOT BRASS
41 46–313398P1 2 SOLDER AR TIN 63/LEAD 37, ROSIN CORE, 0.062” DIA., 1 LB. SPOOL
47 46–317676P1 2 CHIN STRAP 5 24” LONG, WHITE VELCRO LOOP
48 46–317677P1 2 FORHEAD STRP 5 24” LONG, 1” WIDE, WHITE VELCRO LOOP

72 46–317475P4 2 VELCRO LOOP 1 2”W X 18”L, WHITE, ADHESIVE BACKED


77 46–307011P1 2 ROUND SPACER 1 0.312 OD, 0.193 ID, 0.25” LONG, NYLON
79 46–282389P1 2 ADHESIVE AR ADHESIVE TRANSFER TAPE, TYPE A–25
84 46–317475P5 2 VELCRO LOOP 1 2”W X 7”L, WHITE, ADHESIVE BACKED

46–221735P1 2 DIODE 12 50V, 10W PIN DIODE


46–317605G1 2 CTL PHANTOM 1 CTL POSITIONER/PHANTOM ASSEMBLY
46–306629P2 2 FRICTION PAD 6 PART

5–4 RENEWAL PARTS


GE MEDICAL SYSTEMS SIGNA 1.5T CTL PHASED ARRAY COIL
REV 1 DIRECTION 46–003014

CTL PHASED ARRAY COIL 2104799 (SEE PAGE 5–7 FOR PARTS LIST)

25

84

72

29 27

28

35

BOTTOM VIEW

M2104799A
SHT 1

5–5 RENEWAL PARTS


GE MEDICAL SYSTEMS SIGNA 1.5T CTL PHASED ARRAY COIL
REV 1 DIRECTION 46–003014

CTL PHASED ARRAY ARRAY 2104799 (ILLUSTRATION CONT’D FROM PG 5–5)

34

33
32
5
6
22
31

37
30
77 8
36
36 31

36
30

L2104799A

5–6 RENEWAL PARTS


GE MEDICAL SYSTEMS SIGNA 1.5T CTL PHASED ARRAY COIL
REV 1 DIRECTION 46–003014

CTL PHASED ARRAY ARRAY 2104799

Item Part Number FRU Name Quantity Description (Remarks)


1 2104799 1 CTL COIL XXX CTL PHASED ARRAY COIL
2 46–301961P1 2 PAD 1 GREY NYLON FABRIC W/VELCRO HOOKS
5 2133085 2 CABLE ASSEM 1 CTL COIL 6–COAX, 30” EXTERNAL CABLE
6 2121448 2 CIRCUIT BD 1 PHASE ADJUSTMENT CIRCUIT BOARD

22 46–208758P8 2 STRAP 4 1/16” TO 1–1/4” DIA. WIRE BUNDLE


25 46–307593P1 2 PAD 1 NECK COIL PAD, NYLON W/VELCRO HOOKS
27 46–317475P1 2 VELCRO PAD 6 2”W X 2”L, WHITE, ADHESIVE BACKED
28 46–317475P2 2 VELCRO PAD 1 2”W X 4”L, WHITE, ADHESIVE BACKED
29 46–317476P2 2 VELCRO PAD 2 2”W X 4”L, WHITE, ADHESIVE BACKED

30 46–230007P9 2 SCREW,MACH 5 10–32 X 5/8” BIND HEAD SLOT, DELRIN


31 46–230007P8 2 SCREW,MACH 5 10–32 X 3/8” BIND HEAD SLOT, DELRIN
32 46–220184P40 2 SCREW MACH 2 10–32 X 0.375 LG RD HD SLOT BRASS
33 46–208913P12 2 SCREW,MACH 2 10–32 X 0.50 LG SLOT FH BRS
34 46–220184P45 2 SCREW MACH 2 10–32 X 1.000 LG RD HD SLOT BRASS

35 46–220184P41 2 SCREW,MACH 10 10–32 X 0.500 BRASS


36 46–230007P13 2 SCREW,MACH 4 10–32 X 1 DELRIN AF
37 46–220184P42 2 SCREW MACH 8 10–32 X 0.625 LG RD HD SLOT BRASS
41 46–313398P1 2 SOLDER AR TIN 63/LEAD 37, ROSIN CORE, 0.062” DIA., 1 LB. SPOOL
47 46–317676P1 2 CHIN STRAP 5 24” LONG, WHITE VELCRO LOOP
48 46–317677P1 2 FORHEAD STRP 5 24” LONG, 1” WIDE, WHITE VELCRO LOOP

72 46–317475P4 2 VELCRO LOOP 1 2”W X 18”L WHITE, ADHESIVE BACKED


77 46–307011P1 2 ROUND SPACER 1 0.312 OD, 0.193 ID, 0.25” LONG, NYLON
79 46–282389P1 2 ADHESIVE AR ADHESIVE TRANSFER TAPE, TYPE A–25
84 46–317475P5 2 VELCRO LOOP 1 2”W X 7”L, WHITE, ADHESIVE BACKED

46–221735P1 2 DIODE 12 50V, 10W PIN DIODE


46–317605G1 2 CTL PHANTOM 1 CTL POSITIONER/PHANTOM ASSEMBLY
46–306629P2 2 FRICTION PAD 6 PART

5–7 RENEWAL PARTS


GE MEDICAL SYSTEMS SIGNA 1.5T CTL PHASED ARRAY COIL
Blank DIRECTION 46–003014

5–8 RENEWAL PARTS


GE MEDICAL SYSTEMS SIGNA 1.5T CTL PHASED ARRAY COIL
REV 1 DIRECTION 46–003014

DATA TABLE

Use the space provided below to record the calculated signal to noise ratio (SNR) data obtained from Section 3,
FUNCTIONAL CHECKS. After recording the SNR data obtained during the initial Array installation, record subse-
quent SNR data in column 2 – 7 below and the date they were obtained in column number 1 as a periodic QA check.
If the ratio of any of the coils found in column 8 is not greater than 85%, then there is a problem in the Multicoil system.

Original SNR data obtained at initial coil installation:

Body Coil SNR Value: _________________

CTL Phased Array Coil SNR Value:


Coil 1 _________ Coil 2 _________ Coil 3 _________ Coil 4 _________
Coil 5 _________ Coil 6 _________

Date:

SNR DATA QA (QUALITY ASSURANCE) CHECK:

1 2 3 4 5 6 7 8

Coil 1 Coil 2 Coil 3 Coil 4 Coil 5 Coil 6 New values divided


Date: QA SNR QA SNR QA SNR QA SNR QA SNR QA SNR by original
Value: Value: Value: Value: Value: Value: values above: (avg.)

> 85% ?

> 85% ?

> 85% ?

> 85% ?

> 85% ?

> 85% ?

> 85% ?

> 85% ?

> 85% ?

> 85% ?

> 85% ?

> 85% ?

Make additional copies of this document as is needed.

DT–1 DATA TABLE


       

Blank 
 

DT–2 DATA TABLE


GE MEDICAL SYSTEMS SIGNA 1.5T CTL PHASED ARRAY COIL
REV 1 DIRECTION 46–003014

DEFECTIVE COIL RETURN FORM

Note
To allow for proper assessment of defective returned coils, this form must be completely filled out and
accompany all returned coils. Include films or prints of any image quality related complaints with a
description of scan protocol used.

Date

Site Name

Site Address

Service Engineer

Coil Serial Number

Date Coil Installed

Description of Coil Problem

ELECTRICAL CHECKS

PIN Diode Test (Section 3–4)


Diode Drop Forward Bias
Diode Drop Reverse Bias

DR–1 DEFECTIVE RETURN


GE MEDICAL SYSTEMS SIGNA 1.5T CTL PHASED ARRAY COIL
REV 1 DIRECTION 46–003014
Note
An alternate proprietary procedure is available for GE use and to customers with a valid Advanced
Service Package Limited License. Refer to ”TLT PROCEDURE” calibration procedure located in
Direction 2124201, MR Release 5.x Signa Service Methods or Direction 2124202, Release 4.x/3.x
MR Signa Service Methods. Use Table DR–1 and DR–2 to record the TLT results of the defective coil.

TABLE DR–1
TLT DATA FOR DEFECTIVE PHASED ARRAY COIL

SITE: NAME: DATE:


TLT FILE NUMBER:
SNR: NOISE (0)
NOISE (1)
NOISE (2)
NOISE (3)
NOISE (C)
SNR MEAN (0)
SNR MEAN (1)
SNR MEAN (2)
SNR MEAN (3)
SNR MEAN (C)
TR MAP: 89–91 (C) % %
(Combined Results) 85–95 (C) % %
65–115 (C) % %
FLIP MEAN (C)
FLIP SDV (C)
RCV MEAN (C)
RCV SDV (C)
DIFF MEAN (C)
DIFF SDV (C)

TABLE DR–2
TLT DATA FOR DEFECTIVE SURFACE COIL

SITE: NAME: DATE:


TLT FILE NUMBER:
SNR: NOISE
SNR MEAN
SNR SIGNAL
SNR AREA
TR MAP: 89–91 % %
85–95 % %
65–115 % %
FLIP MEAN
FLIP SDV
RCV MEAN
RCV SDV
DIFF MEAN
DIFF SDV

DR–2 DEFECTIVE RETURN