REV.

2021 Feb

Ultrasonic Sludge Density Meter_ ENV200

The  ENV200  is  an  ultrasonic  instrument  that  measures  the  density  of  suspended  solid  in 
liquid. The ENV200 utilizes the EEA (Envelope Energy Average) method that saves reception 
signal  envelop  and  then  calculates  its  energy,  rather  than  using  the  reception  signal’s 
amplitude change. ENV200 comprises of a controller and two types of sensors, such as Spool‐
piece and Clamp‐on type to accommodate all field demands at installation. 

Measuring Algorithm _ EEAM  

 
Conventional  ultrasonic  attenuation 
density  meter  just  determines  density 
with amplitude of received signals. Unlike 
this, ENV200 is able to measure changes of 
concentration  in  a  more  sophisticated 
manner  by  adopting  the  patented  EEAM 
(envelope  energy  averaging  method), 
which measures not only the amplitude of 
received  signals  but  also  observes  the 
shape  of  signal.  It  takes  all  energy  as 
envelope and then convert it into density 
 
 
 

Ultrasonic Sludge Density Meter_ENV200


Features

■ Continuous and real‐time measurement 
■ Reliable signal control EEAM(Envelope Energy Average Method) algorithm
■ Various types of sensors to accommodate all field demands at installation
■ Offer several density units, %, mg/l, g/l, ppm, kg/m³, g/cm³
■ Maximum 400 days data logging and monitoring
■ In‐situ measurement and calibration 
■ Automatic sensor sensitivity control 

Applications

■ Water / Wastewater Treatment
■ Pulp and Paper 
■ Food and Beverage 
■ Power Plant 
■ Chemical 
■ Mining 

Ultrasonic Sludge Density Meter_ENV200


Product Dimensions (mm)

Controller– C2S

Sensor

Spool-piece Clamp-on
(S2S2) (S2C)

Ultrasonic Sludge Density Meter_ENV200


Product Specification

Controller C2S

 
Measuring Principle  : Ultrasonic Attenuation and EEAM(Envelope Energy Average   
Method) 
Measuring Range  : 0 ~ 90% (0~900,000 ppm) 
  0 ~ 9999kg/m3    (0 ~ 9.999 g/cm3) 
Resolution  : 0.1% or 0.01% (Selectable),   
  1 kg/m3 or 0.001 g/cm3 
Display  : Graphic LCD 
(Density, Time, Echo profile, mA, Etc.) 
Accuracy  : ±1%   
Operating Temp.  : ‐20 ~ +70°C (‐4 to 158°F) 
Outputs  Current output : 4 to 20mA, nom. Load 250Ω   
                                (Load range: 100 to 750Ω) 
Relay output : 1 SPDT (5A, 250VAC)   
Digital output : RS232(STD.), RS485(Option)
Power Supply  Standard : 100V ~ 240V AC, 50~60Hz, < 14W 
  Option : 24V DC 
Enclosure Material  : Body/Cover ‐ ABS 
Dimension  : 252(W) x 238(H) x 125(D)mm 
Mounting  : Center Hole 205(W) x 246(H) mm (M6 x 4ea) 
Weight  : 2 kg 
IP Rating  : IP67 
Data Saving  : Max. 400 days data logging and trend 
Screen  : Numeric, Data trend, Echo profile, parameter 
Certificate  : CE 

Ultrasonic Sludge Density Meter_ENV200


Product Specification

Sensor

Spool-piece Type – S2S


Pipe Size  : DN50 ~ DN600
Operating Temp.  : ‐10 ~ +60°C (14 to 140°F) 
Frequency  : 1.4 MHz 
Cable Length  : 10m(STD.) 
IP Rating  : IP68 
Material  : Body: SUS304 (Opt. SUS316), Acoustic window: Epoxy 
Dimension  : Total outline dim._ depending on requested pipe size 
Sensor_ Φ 25mm x L 45.3mm * 2pcs 
Weight  : Sensor_ 85g x 2pcs 
 
 
Clamp-on Type – S2C
Pipe Size  : DN50 ~ DN300
Operating Temp.  : ‐10 ~ +70°C (14 to 158°F) 
Frequency  : 0.699~2.041 MHz 
Cable Length  : 10m(STD.) 
IP Rating  : IP68 
Material  : Body: Polycarbonate, Acoustic window: MC Nylon 
Dimension  : Incl. clamp parts_L 63 x W 135.5 x H 49(mm)* 2pcs 
Sensor _ Φ 39mm x H 31mm * 2pcs   
Weight  : Incl. Clamp parts_ 144 g x 2pcs / Sensor _36g x 2pcs   

Ultrasonic Sludge Density Meter_ENV200


Ordering Code

ENV200 CODE Description


C2S ENV200 Controller AC100~240V
Controller
Unit : g/l, mg/l, % , ppm or kg/m3, g/ m3    (select)

S2S2 Spool-piece type 2 sensor


Sensor
S2C Clamp-on type sensor
D_XXX DIN Standard, XXX = Pipe diameter in mm
Pipe J_XXX JIS Standard, XXX = Pipe diameter in mm
F_XXX Flange to Flange distance (Unit: mm)
DC DC 24V
RS4 RS485 (STD. RS232)
Option MOD Modbus communication
C_XX Total sensor cable length (Unit: m), (STD. 10m)
SMK Sensor mounting kit
Ex.) C2S-S2S2-D_100-F_450-C_10
Note controller and Spool-piece sensor, Pipe size DN100, Pipe length 450mm,
Total sensor cable length 10m
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 