Manual Q LAS - e
Manual Q LAS - e
Manual Q LAS - e
Q-LAS
PHOTODISRUPTOR
Q-Switch Nd:YAG Laser
With Slit Lamp
Version Number
PHOTODISRUPTOR
Q-Switch Nd:YAG Laser
With Slit Lamp
Table of Contents
1 Introduction ............................................................................................................................... 6
1.1 Copyright ................................................................................................................................... 6
1.2 Marking and Symbols ................................................................................................................ 6
1.3 Intended Use ............................................................................................................................. 7
Indications for Capsulotomy: .................................................................................................... 7
Contraindications ...................................................................................................................... 8
2 Theory and Technical Options.................................................................................................. 9
3 Transport / Storage ................................................................................................................. 10
4 Installation............................................................................................................................... 11
4.1 Installation Site ........................................................................................................................ 11
4.2 Room Requirement ................................................................................................................. 11
4.2.1 Warning Signs ........................................................................................................................ 11
4.2.2 Window Shielding ................................................................................................................... 11
4.2.3 Reflecting Surfaces ................................................................................................................ 12
4.3 Electrical Connection .............................................................................................................. 12
4.4 Transport and unpacking the device (Only Germany) ............................................................ 12
5 Safety Information and Technical Certification ....................................................................... 13
5.1 General.................................................................................................................................... 13
5.2 Eyewear Protection ................................................................................................................. 14
5.3 Electrical Protection ................................................................................................................ 14
5.4 Explosion and Fire Hazard ...................................................................................................... 15
5.5 Protection against Undesired Radiation .................................................................................. 15
5.6 NOHD Safety Distance ........................................................................................................... 16
5.7 CE-Regulations ....................................................................................................................... 16
5.8 External Interlock-Plug ............................................................................................................ 17
5.9 Console Protecting Housing.................................................................................................... 17
5.10 Safety Shutter / Aiming Beam ................................................................................................. 17
5.11 Manual Reset .......................................................................................................................... 17
5.12 Stickers, Labels and Markings ................................................................................................ 18
5.13 Operating Conditions: ............................................................................................................. 20
5.14 Electromagnetic Compability................................................................................................... 20
6 User Information and System Introduction ............................................................................. 21
6.1 Technical Introduction Training ............................................................................................... 21
6.2 Laser-Safety-Training .............................................................................................................. 21
6.3 Medical Introduction Training .................................................................................................. 22
6.4 List of Medical Device ............................................................................................................. 22
6.5 Medical Device Parts and Accessories ................................................................................... 22
6.5.1 Front Panel ............................................................................................................................. 23
6.5.2 Power Supply Rear ................................................................................................................. 23
6.6 Door Switch Installation ........................................................................................................... 24
6.7 Manual Remote (optional) ....................................................................................................... 24
7 Operation ................................................................................................................................ 25
7.1 Laser Console Adjustment ...................................................................................................... 25
7.2 DISPLAY ................................................................................................................................. 26
7.2.1 Energy Selection .................................................................................................................... 27
7.2.2 Burst Mode (Number of Pulses) ............................................................................................. 27
7.2.3 Reset Counter ........................................................................................................................ 27
File: Manual Q-Las EN_25_02_2014 4
UserManual Q-LAS
February 2014
1 Introduction
We appreciate your decision to purchase the Q-Las Laser and wishing you a productive and
successful usage with this multi-functional laser system.
The Q-Las Laser generates a high-intensity laser beam which may induce injuries in improper
handling. Therefore, this User Manual should be read carefully before using the device. Should
you have any further questions regarding safety, the use of the device, or concerning laser
and laser radiation, please contact A.R.C. Laser GmbH or your local authorized dealer (see
Section 10.3 "Sales and Service - Information").
ATTENTION!
We are obliged by local laws to sell this device only to licensed physicians, through our local
authorized dealer which holds a valid MDEL (Medical Device Establishment License).
1.1 Copyright
This user manual is copyright protected. The copyright prohibit any duplication of parts or of
the entire user manual without an explicit written consent by A.R.C. Laser GmbH.
Approved copy or parts extracted from this user manual, as licensed by A.R.C. Laser GmbH,
shall include references to the original document and author as stated in this original
document. Furthermore, this copyright applies to any translated copy of the user manual to
other languages.
We would like to point out that, this user manual was prepared based on A.R.C. Laser
proprietary data and to the best of our knowledge. A.R.C. Laser reserve the rights to
revise, renew or modify any of the included drawings, images or text without further notice.
The laser radiation sign warns the user against aiming laser beams to undesired
surfaces by improper handling. The laser beams of this device are not visible.
When using the protection goggles filtering such beams, the user can neither
identify nor see the beams.
Iridotomy
With several pulses will be made with puncture-like openings through the iris without the
removal of iris tissue. It is typically used to decrease intraocular pressure in patients with
angle-closure glaucoma.
Contraindications
Eyes that have no potential visual function
Eyes with corneal defects such as lesions, scars, edema or a cloudy cornea
Cloudy intra-ocular media or opacities within the eye. Focussing might become
impossible or the laser could be absorbed with harmful side effects.
Ocular inflammations
If the device is not focussing
Eyes that have a glass intraocular lens as there is a risk of fracture of the glass
Disability of the patient to sit in front of the slit lamp, open the eye and keep
the position and the eye open for minimum 2 minutes
CAUTION - The device may only be used by physicians that undergone appropriate training and
have extent knowledge of the medical effects and risks involved in using this device, as well as the
understanding for using this device in accordance with this manual.
This Laser is a Nd:YAG Laser with a Wavelength of 1064 nm. The radiation generated in the laser
head uses a flash lamp. The laser beam of the device has a small diameter, this less than 10μm in
the patients eye.
The word laser means „light amplification by stimulated emission of radiation“.
All laser types include three fundamental elements:
The lasing medium (providing atoms), or molecules allowing to amplify the light.
An excitation source (the abovementioned molecules or atoms are excited with), and an optical
resonator which, similar to a resonant circuit, supports or generates the radiation stimulated by the
internal reflection.
Atoms and molecules may have different energy states, which can generate electromagnetic
phosound radiation changing from one state to the next by means of absorption and consequently
by radiation. In their normal state, the electrons of an atom are in the so-called basic or normal
state.
By supplying energy through light – or though plasma formation, as inert gas lasers do –
electrons are transported into the higher state. The dwell time in the higher state is very short
though. and the electrons sink back into their original state. During this short phase, the
previously stored energy is released again by sending out photons.
The consequently emitting photons have the same wavelength and the same direction within
the laser. Inside the resonator cavity, the photons between the two mirrors are compelled to
pass through the cavity several times. During this time, they hit further photons thus generating
a monochromatic light. To make the light built up this way in the resonator also accessible
outside, the initial mirror of the laser is made permeable in parallel.
The Q-Las is a pulsed laser, which generates short pulses with high energy. It is equipped with
an internal Q-switch. This switch opens the resonator for a very short time only, as soon as
significantly more atoms has been stimulated by the excitation source than it would be the
case with a continuous laser operation. In this way, short and high-energy laser pulses are
generated.
The wavelength depends of the laser light emitting medium. In this case, it is an IR beam. A
special crystal cuts the emitted wavelength by half. So the pulses are modified into green
laser pulses with high energy.
The Q-Las laser beam path is superposed on the observation beam path of the slit lamp.
The desired position can be accurately determined in advance with the aid of the aiming
beam.
Utilising a knob at the Laser unit one can shift the laser Focus in two steps.
3 Transport / Storage
Prior to delivery this instrument has been packed with special care in order to minimize
transport risks.
However, before unpacking your laser, check the packaging for signs of damage. If you
discover any contact the shipping company immediately and only remove the packaging in the
presence of a representative of the company. Draw up a list of all damaged parts and make
sure it is signed by the transport company’s representative.
The device should never be exposed to temperatures below 2°C. or above 40°C The
ambient air must be dry and clean. A humidity above 80% can result in the destruction of
the Q-LAS when switching on the device.
WARNING
If the instrument is at a temperature below 15°C switching on the instrument may cause serious
damage. Unpack the instrument and leave it at normal temperature for at least half a day to ensure
that the internal components warm up gradually.
4 Installation
4.1 Installation Site
Prior to delivery, however, the user need to make sure a suitable location was prepared for the
installation.
The Q-Las Laser System should be installed at a readily accessible place not too close to a
heater or next to a sink. The laser should not be operated close to a heater as its air cooling
system works best when the ambient temperature does not exceed 21°C.
Higher ambient temperature may result in shorter working times as the device will be switching
off earlier to prevent overheating.
If the ambient temperature is too low (below 15°C), the device cannot be started, in order to
avoid possible condensation on its internal optics as this might cause a permanent damage.
The unit should be positioned that the laser beam of the slit lamp is not directed onto a
door, a window or reflective material.
The unit should be installed in a dust-free room. No carpets should be on the floor or
mounted on the walls.
If the unit is not in use, it should be covered to avoid contamination.
The cables have to be placed that as there occur no tripping hazards or other hazards.
The air humidity in the treatment room should be monitored and kept below 75%.
As a safety measure against direct or indirect laser radiation within the laser room, it is
necessary - clearly defined by the accident prevention regulations for laser radiation - that all
personnel in the room must wear laser safety goggles.
When using the laser only such laser safety goggles must be used, which are designed for the
wavelength of 532 nm and in which the CE mark and the Class L6 or DLB6 is noted. Goggles
according to approval are available at ARC Laser GmbH.
Goggles for the specific wavelength are included. For subsequent orders, please contact
A.R.C. Laser GmbH, indicating the wavelength or the article number of the Goggle. If you have
further questions regarding safety goggles, please contact us.
There are different types of goggles that also allow eyeglasses wearers to wear tight goggles
on all sides. In particular, scattered radiation, which does not directly hit the front of the eye,
may present risks resulting from internal reflections in the glasses themselves. We strongly
recommend wearing the goggles, which also guarantee full side protection.
The treatment room should not be opened from the outside during the treatment. Outside the
treatment room of a warning light should be installed. During the operation, this laser warning
light must be switched on. In addition, adequate safety goggles should be available in the room
to provide accompanying person with glasses. For the patient, a patient-compliant eye
protection must be applied.
A door interlock is required by the accident prevention regulations. The standard unit is
equipped with an interlock connector to which a door interlock switch can be connected
directly. Open the door and the unit switches off. Here we suggest as an alternative, to close
the door from inside so that accidental opening from the outside cannot be done.
ATTENTION!
Never work with the device if you notice any visible damage to the device.
ATTENTION!
Never work with the device if you notice any visible damage to outlet plug, or notice the wires have
become exposed because of improper handling.
The device needs to undergo a safety inspection every 24 months carried out by qualified
service personnel to ensure continued safe use of the laser. In the equipment book any faults
during service should be noticed.
When working with the laser, make sure the laser is switched from READY mode to
STAND-BY mode, in case the treatment has to be interrupted. This will assure that no
laser radiation will be emitted due to unintentional stepping on the pedal switch.
In operating rooms in which several pedal switches are available, it is particularly important to
make sure that the laser pedal switch is in within the operator vicinity.
ATTENTION!
During a laser treatment, the system is in "Ready" mode. Should the operator need to pause the
treatment for any reason; the laser must be switched back to "Stand-by" mode. The device must
be turned off when left without supervision, to prevent usage by unauthorized individuals. Installing
and/or operating the Laser in any other way differing from the one described herein may cause
hazardous exposure to radiation.
√ 4𝑃
– 𝐷𝑖𝑎𝑚𝑒𝑡𝑒𝑟 𝐵𝑒𝑎𝑚 𝐵𝑢𝑛𝑑𝑙𝑒
𝑀𝑍𝐵 ∗ 𝜋
𝑁𝑂𝐻𝐷 =
𝐵𝑒𝑎𝑚 𝐷𝑖𝑣𝑒𝑟𝑔𝑒𝑛𝑐𝑒
Burst II
NOHD: 3,06 m
Burst III
NOHD: 3,94 m
However, this hazard distance is irrelevant as the access to the laser radiation is blocked by
the marked operation room for the laser device.
(See also "Room Requirements" Section 4.2 and "Eyewear Protection" Section 5.2).
5.7 CE-Regulations
The Q-Las Lasersystem was accredited by the notified body in accordance with the
European directive 93/42 for medical equipment. Therefore the device is labeled with the
CE mark CE 0483.
The device was tested for electrical compliance as well as for mechanical safety. All parts
used by A.R.C. Laser for the Q-Las Laser comply with CE regulations.
Any additional equipment that needs to be attached to the device must require the official
approval of the local inspection authority. No modifications for the device are allowed as
these may have a serious risk potential and will void the regulatory approval as well as the
warranty.
An Equipment book and an regulatory approval number is included to the device.
Interlock-Connector
The plug is intended to be used as an additional safety switch by connecting it to the door of
the treatment room. Should the door be unexpectedly opened, laser radiation will immediately
drop to zero.
Please be advised that the laser can only operate when the plug is fully inserted into the socket.
Manufacturer
Date of manufacturing
Serial number
See manual
Applied Part Type BF Bild 2: Label under the switch
no disposal in domestic waste
Connections
O Interlock Connector
O Warning Light
O Foot switch/
Manual Remote
O Laser Connector
O Power supply
Bild 3: Labels at the rear
Warninglabel
6.2 Laser-Safety-Training
The Q-Las laser system is designed for medical applications. It may only be used by a
physician who received training by an authorized personal. In addition to the trainings offered
by A.R.C. Laser, we recommend to attend seminars offered by us. These seminars contain
topics like “working with different laser systems”. Moreover, introductions into the laser safety
and the use of lasers in general will be covered in these seminars. We recommend you to
make sure that only the person who was trained will use the laser. We recommend to ensure
that only the person which is trained on the device and noted into the equipment book is using
the laser.
The fact that even personnel not working directly with the laser should attend the "laser safety”
or "laser application” courses, has proven to be very useful. Laser safety, the basic principles
of the laser and of laser treatment is mainly discussed in these courses. It is particularly
important that the operating personnel attend the laser safety courses. In these courses,
extensive training is given to master precautionary measures when working with lasers (e.g.
caution in case of combustible material, the importance of laser protective goggles etc.) are
dealt with in detail.
Training for the accompanying personnel is offered in addition to the introductory, and is given
by the local A.R.C. authorized dealer service person when installing the device.
In this training, special attention is given for safety in general like the laser protective goggles.
It also covers clinical information as well as references to the indication but does not intend to
be complete and to be used as is, since each patient may require different parameters.
A.R.C. Laser GmbH holds a list of recommended courses as well as laser safety courses
available, and can be retrieved from us through our local dealer at any time.
Do not attempt to perform any type of service or maintenance work to the device. A service
technician trained by A.R.C. Laser GmbH should, only carry out any calibrations or
adjustments that require opening the protective housing. This includes also any type of optics
cleaning within the laser system.
User manual
For Information of availability of other accessories, please contact ARC Laser GmbH or
your local distributor.
ATTENTION!
Only the parts and applications as specified by A.R.C. Laser GmbH can be used with the
Laser. Any use of other non approved accessories may pose a serious risk hazard to the
operator and/or patient; As well as not provide the expected result. A.R.C. Laser GmbH
undertakes no liability in this case.
Key switch
Laser-STOP Ready
Key switch
The key switch is used to turn off on the laser and protects it against unauthorized third-
party use. The Ready light turns off when the unit is switched off with the key switch.
Laser-STOP, Panic switch
The large red STOP button will instantly stop all laser functions when pressed.
Main switch
7 Operation
ATTENTION!
The Q-Las Laser may only be operated by personnel that undergone relevant training for the
device and that proves the necessary knowledge for the laser's applications.
This part of the manual describes mainly the technical aspect of the device functionality without
providing extensive details on its medical use. A more elaborated medical use is detailed at
A.R.C. Laser application manuals, which can be obtained through your local authorized
distributer.
Device settings as well as various parameters adjustments should only be carried out in
compliance with the operating instructions manuals. Any modifications or settings not indicated
in the operation manuals may cause malfunctions.
ATTENTION!
A warning lamp/s outside the treatment room door/s must be switched on as soon as the laser
starts to operate; the doors have to be marked as laser rooms, visible from outside (with the
warning signs).
ATTENTION!
Since the aiming beam shares with the infrared laser beam, the same optical delivery system,
it provides a very good indication for checking the intactness of the laser optical delivery
system.
If the target spot does not appear at the distal end of the laser delivery system, or if it has low
intensity, or appears to be diffused, this may indicate a damaged or not properly installed
delivery system of the unit.
7.2 DISPLAY
Burst Mode 1 / 2 / 3
Selector
Ready Button
Reset Counter
Brightness
Aiming Beam
The LCD display of the control box displays different screens depending on the state of the
unit. The screen shows up as soon as the unit is turned ON (when the key switch is ON).
The Display will show:
100
During the Warm-up it counts down from 100 to 0
Then the laser will perform 3 simulation shots
The READY LED flashes
The Counter will show 000
The current output is displayed in the upper display.
Possible Errors:
Energy
The Slit lamp is not turned on (green button) selection
The main switch is not turned on
Use the Energy selection button to change the values. The energy ranges from 1.0 mJ to
maximum with steps of 0.1 mJ
Energy (mJ)
according to the Burst Mode (1,2 or 3 Pulses)
Note:
If the laser is in Ready Mode and the laser is not used for 3 minutes, the unit switches
automatically into standby mode.
7.2.4 Tristar-Focus
The Q-LAS is equipped with 3 aiming beams.
All 3 beams intersect in the focal plane to form a single spot. The laser exerts its effect
on the focal plane. An exception is the focus shift, see chapter 7.3
In the default state of the laser the three target have been determined. These can also
be put into rotation.
Rotation of the aiming beam
Simply press the reset button for at least 4 sec. The aiming beams start rotating as long
as you again push the Reset button (at least 4s). The Q-Las will memorize this setting
and will restart with the rotating aiming beam.
Note: The counter will not be affected/reset by this change.
Aiming Beam
The focus layer is The focus layer is Focus layer Push the reset button for a
inside the lens at the capsule behind the lens least 4s and toggle the
rotation of the aiming beam.
Power Selection
Use the energy knob to increase or
decrease the energy values.
Focus Shift
The Focus of the Nd:YAG will be moved posterior :
0: Focus Position 30 µm is behind
Aiming Beam
Turn the knob to increase or
decrease the brightness of the
Adjustments on the laser head Aiming Beam.
7.4 Footswitch
To trigger the laser press the Footswitch. The Remote control has to be in ready mode.
7.5 Treatment
Prepare and seat the patient, his chin and forehead resting on the chinrest.
Adjust the slit lamp position, illumination and focus as required.
Select your initial values for energy, pulse(s) level. It is always best to
begin at an energy level below the anticipated treatment level and then increase the energy
as required during the treatment
Fixate the patient’s eye with a lens and focus the aiming beam into the eye to be treated.
Press the „READY“ mode button to enable the laser: the yellow light turn ON.
Treatment can be started.
1. Turn the key switch to position „OFF“ (see Chapter 6.5.1) to turn the laser OFF
3. Switch OFF the unit with the main switch at the table.
ATTENTION!
REMEMBER TO TURN OFF THE SLIT LAMP and the system should be covered to
avoid dust setting on the optics.
WICHTIG:
The panic switch must be used only in case of emergency. It is located on the front of
the Power supply.
8 SPECIFICATIONS
8.1 General
Modell Q-Las
Cooling Air cooling
8.2 Laser-Data
Wavelength 1064 nm
Lasersource Nd:YAG
Energy 1.0 mJ to Maximum - Single Pulse
Energy Setting Adjustment in steps with 0.1 mJ from
1.0 mJ to its maximum
Pulse width 4 ns
Burst Mode 1, 2 or 3 Pulses
Spot size < 10 µm
Cone angle 16 º
Aiming beam 635 nm red < 1mW
8.5 Compliance:
- This laser product complies with 21 CFR 1040 as applicable
- Electrical leakage below UL 544 limits
- Power supply and enclosures designed per IEC 601-1
- Safety of laser products: IEC 825-1
- Safety of therapeutic laser equipment IEC 601-2-22
- European Directive 93/42/EWG
- Electromagnetic compatibility IEC 601-1-2
9 SERVICE
9.1 Introduction
The device was designed, developed and tested according to research based on state-of-
the-art technology. As a rule, similar devices have a service life exceeding 5 years. This
product service life is defined to be 10 years in order to guarantee the availability of its
spare parts during this period of time. However, to make sure the product is in good
working order, the device can display its internals working status through the display.
ATTENTION!
Do not attempt to perform any type of service or maintenance work to the device. Any
calibrations or adjustments that require to open the protective housing, should only be carried
out by a service technician trained by A.R.C. Laser GmbH. This includes also any type of
optics cleaning within the laser system.
ATTENTION!
Disconnect AC power completely before cleaning the case (turn off the main switch).
ATTENTION!
Never look directly into the laser beam emitted from the slit lamp once the foot switch has
been activated, as this will cause a severe eye injury. Always switch off the laser and
disconnect it from the power cable when cleaning the console and components.
1. Use alcohol-free, non-soaked cotton swab and try to clean the lens from the outside
with distilled water. If this is not successful, please continue with step number 2
2. Take a few drops of 100 % alcohol and sprinkle it gently the end of the cotton swab
and then proceed as in step number 1
3. Always wipe only in one direction of the optical surface. Never run with the same
swab back in the other direction. Any particles that have already been fixed in the
cotton swab would scratch the optical surface upon return again and this may lead
to irreparable damage to optics
1) Use the focus stick which is supplied with the unit and place it into the hole of the
slit lamp axis. Refer to the slit lamp manual which is supplied with the unit.
ATTENTION!
Press the oculars firmly into their mounting so they are fully inserted into the binocular
piece. Adjust each ocular to your accommodation so that the single slit on the target
paper is in sharp focus to avoid straining your eyes.
2) Turn on the laser. Do not press the foot switch to trigger the laser, leave in
STANDBY mode. Center the Aiming Beam in the visual field on the Focus stick.
3) Check the two aiming beams are superimposed when you are on the Focus stick
4) A disturbed Aiming beam may be an indication of a faulty system; the Aiming
beam can be a tool to check the integrity of the system.
F04 Optokoppler Permanently checked by test shots Please contact your local sales &
Safety Shutter during the system start-up and service-representative of A.R.C. Laser
Time Out during the ready phase
F06 Foot switch Checked during the system start Please contact your local sales &
short circuit – service-representative of A.R.C. Laser
pre / post
F11 Unable to set Checked during the system start Please contact your local sales &
power service-representative of A.R.C. Laser
F38 DA converter Checked during the system start Please contact your local sales &
service-representative of A.R.C. Laser
F47 Performance Checked during the system start Please contact your local sales &
check, service-representative of A.R.C. Laser
Checksum
error
F50 I2C-error Checked during the system start Please contact your local sales &
service-representative of A.R.C. Laser
F51 keys error User interface error Please contact your local sales &
service-representative of A.R.C. Laser
F58 Pulse Checking the maximum charge Please contact your local sales &
Threshhold voltage at the service-representative of A.R.C. Laser
Voltage charge controller
F0C Knob Position Permanently checked. Needs to be Please check knob position of the
Error placed in a locked position. Focus Shift.
9.6 Disposal
The unit should be disposed according to the valid laws, rules and
guidelines of your country. Contact your local government’s recycling or
solid waste management department to learn more about the services it
provides. This laser must not be disposed in your garbage. A.R.C. Laser
would be very happy to assist you in the question of how to dispose the
device properly. Costs and responsibility for the proper return of the laser
to ARC Laser has to be beard by the customer. Please contact our Service department for
further information.
10 Customer Service
10.1 Warranty-Information
A.R.C. Laser GmbH provides a two-year warranty. Within these two years, any parts showing
a defect will be replaced free of charge. This does not include any purchased parts acquired
from a third party. Our warranty covers the repair works and the replacement of defective parts.
However, we reserve the right to renew even complete units and to adjust them to the technical
progress.
Refer any claims or defects to
A.R.C. LASER GmbH Bessemerstr. 14, D-90411 Nuremberg
Phone: +49 911 21779-13 (-0), Fax: +49 911 21779-99, Email: service@arclaser.de
Repair works carried out by third parties or modifications of the device void the warranty claim.
The use of parts other than those approved together with the device or purchased from other
suppliers also voids the warranty for the complete device. Any parts, units or modifications of
the device require the explicit written consent of A.R.C. Laser GmbH.
Voltage dips, < 5 % UT Complies with IEC 60601 The supply voltage
short interruptions (>95 % dip in UT) quality should be that of a
and variations on for ½ cycle typical commercial or
power supply hospital environment.
input lines 40 % UT When the user of the
IEC 61000-4-11 (60 % dip in UT) laser requires continued
for 5 cycles operation during power
mains interruptions, it is
70 % UT recommended that the
(30 % dip in UT) laser is powered from its
for 25 cycles battery or an
uninterruptible power
<5 % UT supply or an external
>95 % dip in UT) battery..
for 5 s
Magnetic field 3 A/m Complies Magnetic fields in mains
with power frequency should be
frequency (50/60 those of a typical
Hz) commercial or hospital
IEC 61000-4-8 environment.
NOTE: UT is the A.C. mains voltage prior to the application of the test level.
d = [ 3,5 : V1 ] √P
d = [ 3,5 : E1 ] √P
for 80 MHz to 800 MHz
d = [ 7 : E1 ] √P
for 800 MHz to 2,5 GHz
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people
www.arclaser.de