GC AFTGuide
GC AFTGuide
GC AFTGuide
Jul. 2018
LabSolutions
Instruction Manual
AFT (Advanced Flow Technology) Guide
Read this manual thoroughly before you use the product.
Keep this manual for future reference.
This page is intentionally left blank.
Introduction
Read this Instruction Manual thoroughly before using
the product.
Thank you for purchasing this product.
This manual describes how to use this product. Read this manual thoroughly before using
the product and operate the product in accordance with the instructions in this manual.
Keep this manual for future reference.
IMPORTANT
• If the user or usage location changes, ensure that this Instruction Manual is always
kept together with the product.
• If this manual or a product warning label is lost or damaged, immediately contact
your Shimadzu representative to request a replacement.
• To ensure safe operation, contact your Shimadzu representative if product
installation, adjustment, re-installation (after the product is moved), or repair is
required.
Settings During 1
1 Installation
1.1 Settings for Analytical Line configuration ..............
2 Select the AFT that are used at [Available Modules], click , and move them
to confirm the carrier gas flowrate as it flows in the opposite direction during back flush.
This makes it easy to develop a method for performing a back flush.
2 Click [Load Data File] to load the file of data acquired from the analysis in step
1.
3 Set [APC Pressure] and [Inj Pressure] in the [Back Flush] area.
Dead time refers to the time required for the carrier gas to travel from the splitter to the
injector. Ordinarily, substances remaining inside the column will be discharged within about
1.5 times this dead time. Use this as a guide when developing methods. In this case, enter
the temperature during the back flush as the column temperature. Caution is necessary
because if the back flush starts while the temperature is rising, the column flowrate will
be reduced as the column temperature rises. Accordingly, all of the substances remaining
in the column may not be discharged unless the back flush time is set to longer than 1.5
times the dead time.
• Setting the value in [APC Pressure} too high may have an adverse impact, such
as extinguishing the flame in the flame ionization detector.
• Setting the value in [Inj Pressure] too low may prevent the purge flow.
Ordinarily, use the default value.
4 Double-click the reference chromatogram display area to set [Back Flush Time].
Even when AFT elements other than the back flush element are used, back flush is
possible by controlling the APC pressure (switching pressure/splitting pressure) of each
element. This section describes the back flush setting procedure when AFT elements other
than the backflush element are used.
2 Select the [Back Flush] checkbox, and set the back flush parameters.
The following shows the correlation between the APC pressure and the restrictor carrier
gas flowrate when the restrictor (0.15 mm × 0.5 m) provided as standard with the
Shimadzu back flush element is used. Although it depends on the column oven
temperature, it is recommended that you ordinarily set the APC pressure to 100 kPa or
less during analysis.
3
2 Click [(AFT name)] in the [System Configuration] window. (In this picture, click
[HTCT].)
By setting [1stDetector] and [2ndDetector], it is possible to correctly calculate gas flow rate
when using a detector that is not at atmospheric pressure.
For the back flush settings, refer to"2.2.3 Back Flush by Using Other Types of AFT Systems"
1 Configure the ordinary analysis conditions. This means that [Switching Program] is
not Created.
2 Configure the switching pressure and the injector pressure so that the value in
[Recovery] reaches "100" %. Then perform the analysis.
2 Click [Load Data File] to load the file of data acquired from the analysis in step
1.
1 In the reference chromatogram display area, double-click two time positions each
to start and end the heart-cut feature.
The switching program will be created. Components for peaks in the region shown
with diagonal lines will be injected into the 2nd column for separation and
detection during the next analysis.
To delete or edit the switching program, display the window for editing the
program by double-clicking the region shown with diagonal lines.
Ordinarily, configure the settings so that the value in [Recovery] reaches "100" %. Note
that the switching recovery rate varies significantly (several dozen %) depending on errors
in the column dimensions. Only use it as a guide.
• The smaller the difference between the injector pressure and the switching pressure, the
higher the recovery rate.
• If the difference between injector pressure and the switching pressure is constant, the
larger the switching pressure, the higher the recovery rate.
For the details of the back flush settings, see "2.2.3 Back Flush by Using Other Types of
AFT Systems".
2 Click [Load Data File] to load the file of data acquired from the analysis in step
1.
The switching program will be created. Components for peaks in the region shown with
diagonal lines will be injected to the 2nd detector for detection during the next analysis.
• In order to always load the sample into the second detector throughout the analysis,
click [Switching Program Setting] to display the [Switching Program Setting] window, and
set the start time to 0.01 min and the end time to the analysis finishing time.
• To delete or edit the switching program, display the window for editing the program by
double-clicking the region shown with diagonal lines.
*The splitting pressure is not displayed for systems in which an APC is not used at the
splitting unit.
For the details of the back flush settings, see "2.2.3 Back Flush by Using Other Types of
AFT Systems".