F 2013 - 00 - Rjiwmtmtukve
F 2013 - 00 - Rjiwmtmtukve
1. Scope
1.1 This test method covers a gas chromatographic procedure for the determination of the ppm residual acetaldehyde (AA)
present in poly(ethylene terephthalate) (PET) homo-polymers and co-polymers which are used in the manufacture of beverage
bottles. This includes sample types of both amorphous and solid-stated pellet and preform samples.
1.2 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility
of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory
limitations prior to use.
2. Referenced Documents
2.1 ASTM Standards:
D 4509 Test Method for Determining the 24-Hour Gas (Air) Space Acetaldehyde Content of Freshly Blown PET Bottles2
1
This test method is under the jurisdiction of ASTM Committee F02 on Flexible Barrier Materials and is the direct responsibility of Subcommittee F02.30 on Test Methods.
Current edition approved April 10, 2000. 2001. Published A June 2001. Originally published as F 2013 – 00. Last previous edition F 2013 – 00.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.
1
F 2013 – 001
E 691 Practice for Conducting an Interlaboratory Study to Determine the Precision of a Test Method3
3. Terminology
3.1 The terms employed in this test method are commonly used in normal laboratory practice and require no special comment.
6. Sources of Error
6.1 A bias is known to exist if the ratio of sample mass (mg) to head-space vial volume (mL) exceeds a value of ten.
6.2 Acetaldehyde is very volatile and must be handled carefully to avoid sample loss during the calibration procedure. Storing
the standard vials in a refrigerator is a must to minimize the error due to volatility.
6.3 Failure to achieve a tight seal on the head-space vial will result in the loss of acetaldehyde during storage and desorption,
producing a false low value.
6.4 Failure to grind the sample to the appropriate particle size may lead to a false low value for residual AA due to the increased
path length for desorption.
6.5 Samples submitted for 8residual AA measurement’ should be stored in a freezer until they are tested. Failure to do so can
result in lower than expected results.
6.6 Excessive grinding of samples can cause residual AA contained therein to be desorbed. Extensive excessive grinding can
lead to actual melting of the polymer and AA generation. Samples which have been chilled in liquid nitrogen properly should only
be in the grinder for ;30 s or less.
7. Apparatus
7.1 Gas Chromatograph, Hewlett-Packard 6890 series4
7.2 Integrator, Hewlett-Packard Model 6890 series, H-P ChemStation, H-P LAN.4
7.3 Head-Space Sampler, Hewlett-Packard 7694.4
7.4 Column, 30-m by 0.53-mm i.d. inside diameter (megabore capillary column).5
7.5 Vials, 20-mL, head-space, with 20-mm septa, 20-mm aluminum caps, and crimper for 20-mm caps
7.5.1 Vials:
Perkin-Elmer Part Numer 0105–0129, or Kimble Glass Inc. Part Number 60827A-23756, or Kimble Glass Inc. Part Number
60827A-23757, or Hewlett-Packard Part Number 5182–08374.
7.5.2 Septa:
Pierce Part Number 12720, or Hewlett-Packard Part Number 9301–0807.4
7.5.3 Caps:
2
Annual Book of ASTM Standards, Vol 08.03.
3
Hewlett-Packard apparatus, website: hewlett-packard.com, was used in the development
3
Annual Book of this standard. If you are aware of alternative suppliers, please provide this information to ASTM headquarters. Your comments will receive careful
consideration at a meeting of the responsible technical committee, which you may attend. Standards, Vol 14.02.
4
J & W Scientific GS-Q (Catalogue Number 115–3435), available from 91 Blue Ravine Rd., Folsom, CA,
4
Hewlett-Packard apparatus, website: JWscientific.com, hewlett-packard.com, was used in the development of this standard. If you are aware of alternative suppliers,
please provide this information to ASTM headquarters. Your comments will receive careful consideration at a meeting of the responsible technical committee, which you may
attend.
5
AJ & W Scientific GS-Q (Catalogue Number 115–3435), available from Perkin-Elmer, 91 Blue Ravine Rd., Folsom, CA, website: JWscientific.com, was used in the
developmerknt of this stan-dard. If you are aware of alternative suppliers, please provide this information to ASTM headquarters. Your comments will receive careful
consideration at a meeting of the responsible technical committee, which you may attend.
6
Available from Kimbel/Kontes, Vineland, NJ, e-mail: cs@kimble-kontes.com. Perkin-Elmer, website: perkin-elmer.com.
7
Wheaton Part Number 224303, available
7
Available from 1501 North 10th St., Millville, NJ 08332, website: algroupwheaton.com, was used in the development of this standard. If you are aware of alternative
suppliers, please provide this information to ASTM headquarters. Your comments will receive careful consideration at a meeting of the responsible technical commitee, which
you may attend. Kimbel/Kontes, Vineland, NJ, e-mail: cs@kimble-kontes.com.
2
F 2013 – 001
Pierce Part Number 13214, or Hewlett-Packard Part Number 9301–0721.4
7.6 Crimper, 20-mm, Wheaton Part Number 224303.8
7.7 Decrimper, 20-mm, Kimble Part Number 69903–20.7
7.8 Wiley Mill, equipped with an 800 to 1000-µm screen, or equivalent
7.9 Syringe, calibrated, with certificate of calibration.9
7.10 Small Vacuum Cleaner, with host attachment for cleaning.
7.11 Analytical Balance, capable of accurately weighing to at least 60.0001 g.
7.12 Hammer.
8
Hamilton Digital with 2–µL capacity, Model
8
Wheaton Part Number 7002KH, 224303, available from 1501 North 10th St., Millville, NJ 08332, website: algroupwheaton.com, was used in the development of this
standard. If you are aware of alternative suppliers, please provide this information to ASTM headquarters. Your comments will recieive careful consideration at a meeting
of the responsible technical committee, which you may attend.
9
Protocol Analytical Supplies, Inc., Middlesex, NJ 08846 (R-3, S-2).
9
Hamilton Digital with 2–µL capacity, Model Number 7002KH, was used in the development of this standard. If you are aware of alternative suppliers, please provide
this information to ASTM H headquarters. Your comments will receieve careful consideration at a meeting of the responsible technical committee, which you may attend.
10
A research report is available from
10
Protocol Analytical Supplies, Inc., Middlesex, NJ 08846 (R-3, S-2). If you are aware of alternative suppliers, please provide this information to ASTM headquarters.
Request RR:F02–1015. Headquarters. Your comments will receive careful consideration at a meeting of the responsible technical committee, which you may attend.
3
F 2013 – 001
10.2 Pellets—May be cryogenically ground in a small grinding mill using liquid nitrogen. The final sample should be
thoroughly homogenized before sampling for analysis.
NOTE 5—Samples, either preforms, plaques, or pellets, should be chilled in the liquid nitrogen for several minutes until the liquid nitrogen stops boiling
and then dropped immediately into the grinder. Sample should be sufficiently ground in a few seconds. The grinder should not be allowed to operate more
than 20 to 30 s as in such cases undesirable sample heating can occur.
11. Procedure
NOTE 6—Refer to the general operating manual for the Hewlett-Packard 68904 gas chromatograph, the Hewlett-Packard 76944 head-space sampler,
and the Hewlett-Packard 68904 series integrator for instructions in performing steps in this procedure.
11.1 Adjust the H-P 68904 gas chromatograph to the conditions specified in Appendix X1. Adjust the H-P 76944 head-space
sampler to the conditions in Appendix X2. Set the H-P 68904 series integrator to the conditions in Appendix X3.
11.2 Sample Analysis:
11.2.1 Place 2 to 3 of polymer pellets (or crushed preform) into a small Dewar flask.
11.2.2 Cover the polymer with 20 to 40 mL of liquid nitrogen.
11.2.3 Allow the polymer to chill under the liquid nitrogen for approximately 3 min (or until most of the liquid N2 has
evaporated).
11.2.4 Turn on the Wiley mill equipped with a 800 to 1000-µm screen.
11.2.5 Slowly pour the remaining liquid nitrogen from the Dewar flask through the Wiley mill, followed by the chilled polymer
sample (tapping the sample may be required).
11.2.6 Collect the ground polymer in a small glass jar or small manila envelope.
11.2.7 Turn off the Wiley mill and clean it with a vacuum cleaner.
11.2.8 Allow the ground polymer sample to come to room temperature (approximately 10 min).
11.2.9 Weigh approximately 0.2000 (6 0.0200) g, recorded to the nearest 0.0001 g, into a 20-mL head-space vial.
11.2.10 Place a septum (with Teflon side down towards the inside of the vial) on the vial. Place an aluminum cap on top of the
septum, and crimp the cap with a crimper UNTIL THE CAP CANNOT BE TURNED. Remove the center piece of the aluminum
cap (if it exists).
11.2.11 Place the vial in the appropriate position in the head-space sampler.
11.2.12 Press the “Vial Parameter” key on the H-P 79644 head-space sampler, and enter the starting and stopping vial positions.
11.2.13 Press the “START” button on the H-P 76944 head-space sampler.
11.2.14 The head-space sampler will heat the sample for 60 min at 150°C and then automatically inject the head-space gas and
start the gas chromatograph and integrator.
11.2.15 The final report will appear on the H-P 68904 integrator or the data system when the GC is finished. This report will
give the micrograms of AA.
11.2.16 To determine the concentration in ppm of AA in the polymer sample, divide the micrograms of AA (reported in 11.2.15)
by the sample weight in the vial (recorded in 11.2.9 as grams of polymer).
12. Calculation
12.1 The AA response factor is calculated as described in 9.11 and 9.12. The ppm of AA can be calculated manually by
multiplying the response factor and the area of the AA peak, and then dividing this number by the sample weight in the vial (in
grams).
13. Report
13.1 Report the ppm of AA to two decimal places.
14. Precision and Bias
14.1 To be written
14.1 The following was taken from work done in completed by the International Society of Beverage Technologists (ISBT)
subcommittee concerning standardization of method to determine residual AA in PET.
14.2 The number of laboratories, materials, and determinations in this study meets the minimum requirements for determining
precision in accordance with Practice E 691. A complete report is on file at ASTM Headquarters.11
14.3 This round robin was conducted by having one laboratory mold PET preforms on a 48-cavity injection molding machine
and selecting 6 of those cavities as the sample set. Even though these preforms all came from one PET sample (material), each
cavity has its own unique AA value, and thus, were treated as six different materials. Also, two different types of precision and
bias were calculated, one based on each laboratory using their own calibration standard solution and another when each laboratory
calibrated with a “common” calibration standard.
11
This test method is under the jurisdiction of ASTM Committee F02 on Flexible Barrier Materials and is the direct responsibility of Subcommittee F02.30 on Test
Methods.
Current edition approved April 2001. Published June 2001. Originally published as F 2013 – 00. Last previous edition F 2013 – 00.
4
F 2013 – 001
Practice E 691 Study Minimum
Laboratories: 6 6
Materials: 6 4
Determinations: 3 2
14.4 Precision and Bias With Each Laboratory Using Their Own Calibration Standard—Precision, characterized by
repeatability, Sr and r, and reproducibility, SR and R, has been determined for the materials to be as follows:
Materials Average Sr SR r R
Material A 5.21 0.1812 0.6403 0.5074 1.7928
Material B 6.25 0.4060 0.7464 1.1368 2.0899
Material C 6.37 0.2880 0.6713 0.8066 1.8796
Material D 7.21 0.3285 0.7743 0.9198 2.1680
Material E 7.01 0.4217 0.8350 1.1808 2.3380
Material F 5.88 0.3930 0.7168 1.1003 2.0071
14.4.1 Since the materials used in this study are all from one specific type of material (PET), but have different AA levels
because they are from different cavities, it makes more sense to have one set of precision values rather than one for each cavity.
This will be derived by squaring each Sr and SR, averaging each of Sr2 and SR2 across materials and taking the square root.
Sr SR r R
0.3466 0.7335 0.9705 2.0538
14.4.1.1 Standard Deviation (Sr)—Sr is the square root of the average within laboratory variance.
14.4.1.2 Standard Deviation (SR)—SR is the square root of the sum of the within laboratory variance and between laboratory
variance of the laboratory means.
14.4.1.3 Repeatability—r is the interval representing the largest expected difference between two test results for the same
material, obtained by the same operator using the same equipment on the same day in the same laboratory. A difference larger than
8r’ indicates more variation is present than expected.
14.4.1.4 Reproducibility—R is the interval representing the largest expected difference between two test results for the same
material, obtained by different operators using different equipment in different laboratories, not necessarily on the same day. A
difference larger than 8R’ indicates more variation is present than expected.
14.5 Precision and Bias When Each Laboratory Uses a Common Calibration Standard—Precision, characterized by
repeatability, Sr and r, and reproducibility, SR and R, has been determined for the materials to be as follows:
5
F 2013 – 001
Materials Average Sr SR r R
Material A 5.42 0.1849 0.4128 0.5178 1.1596
Material B 6.47 0.4123 0.6438 1.1545 1.8026
Material C 6.59 0.2703 0.5020 0.7567 1.4057
Material D 7.45 0.3113 0.6333 0.8716 1.7732
Material E 7.26 0.4014 0.5747 1.1240 1.6090
Material F 6.10 0.3854 0.5085 1.0792 1.4237
14.5.1 Since the materials used in this study are all from one specific type of material (PET), but have different AA levels
because they are from different cavities, it makes more sense to have one set of precision values rather than one for each cavity.
This will be derived by squaring each Sr and SR, averaging each of Sr2 and SR2 across materials and taking the square root.
Sr SR r R
0.3376 0.5518 0.9453 1.5450
14.6 Bias—There are no recognized polymer standards by which to estimate bias of this test method. Testing a known liquid
standard with all laboratories using common calibration did not show any laboratory bias between laboratories or between the
average of all laboratories and the known value.
15. Keywords
15.1 acetaldehyde; carbonated soft drink; ground parison AA; PET bottles; 24 hours headspace; AA test; water
APPENDIXES
(Nonmandatory Information)
6
F 2013 – 001
7
F 2013 – 001
8
F 2013 – 001
FIG. X3.1 Hewlett Packard 6890 Series4 Integrator Method File (continued)
9
F 2013 – 001
FIG. X3.1 Hewlett Packard 6890 Series4 Integrator Method File (continued)
10
F 2013 – 001
11
F 2013 – 001
X5.1 Introduction
X5.1.1 The acetaldehyde (AA) calibration standards are used for calibrating gas chromatographs in the determination of AA
present in PET preforms which are used in the food and beverage industry.
X5.2 Materials
X5.2.1 The materials for AA calibration standards are as follows:
X5.2.1.1 Distilled water.
X5.2.1.2 1-L Class A volumetric flask.
X5.2.1.3 Sylon CT.12
X5.2.1.4 Acetaldehyde.
X5.2.1.5 Syringe (1 mL).
X5.2.1.6 Amber caps (4 mL).13
X5.2.1.7 Hole caps14 and TFE-fluorocarbon-faced septa.15
X5.2.1.8 Glass tray.
X5.2.1.9 Balance (capable of 1-mg readings).
X5.2.1.10 Refrigerator, set at 4°C.
X5.3 SetUp
X5.3.1 Use distilled water that is stored in a container other than PET (that is, high-density polyethylene or polyvinyl chloride
bottles)
X5.3.2 If this is the first time a 1-L Class A volumetric flask is being used or the flask has been used for something other than
AA preparation, then proceed to X5.3.3. If the flask has been used previously for AA solution preparation, then proceed to X5.3.4.
X5.3.3 The flask must be deactivated which is done with Sylon CT, or equivalent.12
X5.3.4 Using the distilled water described in X5.3.1, fill the deactivated or previously used 1-L flask to the fill line using the
meniscus.
X5.3.5 Place the flask in a refrigerator and allow it to cool and equilibrate to ;4°C.
X5.3.6 The following items should also be stored in the refrigerator at ;4°C including the acetaldehyde, a 1-mL syringe, a box
of unopened 4-mL amber vials,13 and a glass tray.
X5.3.7 Prepare the caps by placing the TFE-flurocarbon-faced septa15 in the aluminum 8hole-in-cap’ lids14 (Supelco cat.
# 2-712OU, or equivalent). Make sure that the TFE-flurocarbon facing is on the inside so that it is between the rubber septum and
the glass vial. Please note that caps do not have to be refrigerated.
X5.4 Procedure
X5.4.1 Remove the AA, syringe, and 1-L flask and place them beside the balance.
X5.4.2 Fill the syringe with AA, handling the syringe by the top, taking care not to touch the body of the syringe, thereby
causing it to heat up. Place the syringe on the balance and tare it.
X5.4.3 Remove the glass stopper from the 1-L flask, then insert the tip of the syringe into the water and discharge the AA.
Remove the syringe and touch the inside sidewall of the flask to remove any drops. Recap the flask and weight the empty syringe
immediately and record the weight loss.
X5.4.4 Invert the flask and shake at least ten times to ensure proper mixing of the AA.
X5.4.5 Remove the glass tray and the box of vials from the refrigerator. Place no more than 10 vials in the tray. The purpose
of the tray is to catch overfill material.
X5.4.6 Using the 1-L AA solution in the flask, remover the stopper and fill the vials to above the brim, then re-stopper the flask.
X5.4.7 Cap all of the vials with the caps prepared in X5.3.7. It is important to complete X5.4.6 and X5.4.7 as quickly as possible
to prevent AA loss.
X5.4.8 Repeat X5.4.6 and X5.4.7 until the desired amount of vials have been filled, never filling more than 10 at a time.
X5.4.9 For every five vials prepared, prepare one control vial of distilled water. This can be done at room temperature. This
should be done after X5.5.1 and X5.5.3 have been completed as to prevent mislabeling.
X5.4.10 Check all vials to ensure they have no air bubbles (discard vials with air bubbles).
12
Available from Supelco and other suppliers, Catalogue # 3-3065-U. It comes with instructions on deactivating glass.
13
Available from Supelco, Catalogue # 2-7115U, or equivalent.
14
Available from Supelco, Catalogue # 2-712OU, or equivalent.
15
Available from Supelco, Catalogue # 2–7144, or equivalent.
12
F 2013 – 001
X5.5 Labeling
X5.5.1 Prepare labels containing the concentration listed as XXX milligrams of AA/litre of solution using the value obtained
in X5.4.3 as well as the current date.
X5.5.2 Prepare labels for the control vials saying distilled water.
X5.5.3 Place labels on the appropriate vials.
13
F 2013 – 001
ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.
This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.
This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or service@astm.org (e-mail); or through the ASTM website
(www.astm.org).
14