Basic Sensors and Principles
Basic Sensors and Principles
2
BASIC SENSORS AND
PRINCIPLES
Robert A. Peura and John G. Webster
This chapter deals with basic mechanisms and principles of the sensors used in
a number of medical instruments. A transducer is a device that converts energy
from one form to another. A sensor converts a physical parameter to an electric
output. An actuator converts an electric signal to a physical output. An electric
output from the sensor is normally desirable because of the advantages it gives
in further signal processing (Pallas-Areny and Webster, 2001). As we shall see
in this chapter, there are many methods used to convert physiological events to
electric signals. Dimensional changes may be measured by variations in
resistance, inductance, capacitance, and piezoelectric effect. Thermistors
and thermocouples are employed to measure body temperatures. Electro-
magnetic-radiation sensors include thermal and photon detectors. In our
discussion of the design of medical instruments in the following chapters,
we shall use the principles described in this chapter (Togawa et al., 1997).
The physician and biomedical researcher are interested in measuring the size,
shape, and position of the organs and tissues of the body. Variations in these
parameters are important in discriminating normal from abnormal function.
Displacement sensors can be used in both direct and indirect systems of
measurement. Direct measurements of displacement are used to determine
the change in diameter of blood vessels and the changes in volume and shape of
cardiac chambers.
Indirect measurements of displacement are used to quantify movements of
liquids through heart valves. An example is the movement of a microphone
diaphragm that detects the movement of the heart indirectly and the resulting
heart murmurs.
Here we will describe the following types of displacement-sensitive
measurement methods: resistive, inductive, capacitive, and piezoelectric
(Nyce, 2004).
45
c02_1 12/02/2008 46
POTENTIOMETERS
Figure 2.1 shows three types of potentiometric devices for measuring displace-
ment. The potentiometer shown in Figure 2.1(a) measures translational dis-
placements from 2 to 500 mm. Rotational displacements ranging from 108 to
more than 508 are detected as shown in Figure 2.1(b) and (c). The resistance
elements (composed of wire-wound, carbon-film, metal-film, conducting-
plastic, or ceramic material) may be excited by either dc or ac voltages. These
potentiometers produce a linear output (within 0.01% of full scale) as a function
of displacement, provided that the potentiometer is not electrically loaded.
The resolution of these potentiometers is a function of the construction. It
is possible to achieve a continuous stepless conversion of resistance for low-
resistance values up to 10 V by utilizing a straight piece of wire. For greater
variations in resistance, from several ohms to several megohms, the resistance
wire is wound on a mandrel or card. The variation in resistance is thereby not
continuous, but rather stepwise, because the wiper moves from one turn of
wire to the next. The fundamental limitation of the resolution is a function of
the wire spacing, which may be as small as 20 mm. The frictional and inertial
components of these potentiometers should be low in order to minimize
dynamic distortion of the system.
STRAIN GAGES
When a fine wire (25 mm) is strained within its elastic limit, the wire’s resistance
changes because of changes in the diameter, length, and resistivity. The
resulting strain gages may be used to measure extremely small displacements,
on the order of nanometers. The following derivation shows how each of these
parameters influences the resistance change. The basic equation for the
rL
R¼ (2.1)
A
The differential change in R is found by taking the differential
rdL dr
dR ¼ rA2 L dA þ L (2.2)
A A
We shall modify this expression so that it represents finite changes in the
parameters and is also a function of standard mechanical coefficients. Thus
dividing members of (2.2) by corresponding members of (2.1) and introducing
incremental values, we get
DR DL DA Dr
¼ þ (2.3)
R L A r
DR=R Dr=r
G¼ ¼ ð1 þ 2mÞ þ (2.5)
DL=L DL=L
Table 2.1 gives the gage factors and temperature coefficient of resistivity of various
strain-gage materials. Note that the gage factor for semiconductor materials is
approximately 50 to 70 times that of the metals.
Also note that the gage factor for metals is primarily a function of
dimensional effects. For most metals, m ¼ 0:3 and thus G is at least 1.6,
whereas for semiconductors, the piezoresistive effect is dominant. The desir-
able feature of higher gage factors for semiconductor devices is offset by their
greater resistivity–temperature coefficient.
Designs for instruments that use semiconductor materials must incorpo-
rate temperature compensation.
Strain gages can be classified as either unbonded or bonded. An unbonded
strain-gage unit is shown in Figure 2.2(a). The four sets of strain-sensitive wires
c02_1 12/02/2008 48
Temperature
Coefficient of
Resistivity
Material Composition (%) Gage Factor (˚ Cⴚ1ⴚ10ⴚ5)
SOURCE: From R. S. C. Cobbold, Transducers for Biomedical Measurements, 1974, John Wiley &
Sons, Inc.. Used with permission of John Wiley & Sons, Inc., New York.
Figure 2.3 Typical bonded strain-gage units (a) resistance-wire type, (b) foil
type, (c) helical-wire type. Arrows above units show direction of maximal
sensitivity to strain. [Parts (a) and (b) are modified from Instrumentation in
Scientific Research, K. S. Lion. Copyright # 1959 by McGraw-Hill, Inc. Used
with permission of McGraw-Hill Book Co.]
c02_1 12/02/2008 50
(a)
p Top view
A n A
p-type
diffused
layer
p
A– A
n Cross-sectional view
(b)
Clamp
(+) T1 n-type Si
plane T2 Q1
Q1(–) T1 Q2
(+) (+) Silicon
S1 S2
R1 (–) (–) R2 P1 P2
Q2 (–) S1 R2
(+) T2 S2 R1
Diffused
p region (c)
R1 p-type diffused layer
n Side view
R2
p F
R3
n Top view
R1
p
Figure 2.4 Typical semiconductor strain-gage units (a) unbonded, uniformly
doped, (b) diffused p-type gage, (c) integrated pressure sensor, (d) integrated
cantilever-beam force sensor. (From Transducers for Medical Measurements:
Application and Design, R. S. C. Cobbold. Copyright # 1974, John Wiley &
Sons, Inc. Reprinted by permission of John Wiley & Sons, Inc.)
c02_1 12/02/2008 51
unit. Long-term creep is a property of the rubber tubing. This is not a problem
for dynamic measurements.
Operational problems include maintaining a good contact between the mer-
cury and the electrodes, ensuring continuity of the mercury column, and controll-
ing the drift in resistance due to a relatively large gage temperature coefficient.
In addition, accurate calibration is difficult because of the mass-elasticity and
stress–strain relations of the tissue–strain-gage complex. The low value of resist-
ance means more power is required to operate these strain-gage units.
Lawton and Collins (1959) determined the static and dynamic response of
elastic strain gages. They found that the amplitude and phase were constant up
to 10 Hz. Significant distortion occurred for frequencies greater than 30 Hz.
Cobbold (1974) indicated that a problem not fully appreciated is that the gage
does not distend fully during pulsations when diameter of the vessel is being
measured. The mass of the gage and its finite mechanical resistance can cause it
to dig into the vessel wall as the vessel expands, so it can give a reading several
times lower than that measured using ultrasonic or cineangiographic methods.
Hokanson et al. (1975) described an electrically calibrated mercury-in-
rubber strain gage. Lead-wire errors are common with these devices because of
the low resistance of the strain gage. In Hokanson’s design, the problem was
eliminated by effectively placing the strain gage at the corners of the mea-
surement bridge. A constant-current source causes an output that is linear for
large changes in gage resistance. Figure 2.6 shows the device and its output
when applied to the human calf.
The Wheatstone bridge circuit is ideal for measuring small changes in resist-
ance. Figure 2.2(b) shows a Wheatstone bridge with an applied dc voltage of v i
and a readout meter Dv o with internal resistance Ri. It can be shown by the
voltage-divider approach that Dv o is zero—that is, the bridge is balanced—
when R1 =R2 ¼ R4 =R3 .
Resistance-type sensors may be connected in one or more arms of a bridge
circuit. The variation in resistance can be detected by measuring Dv o with a
differential amplifier feeding an analog-to-digital converter (ADC), which
feeds a computer.
Assume that all values of resistance of the bridge are initially equal to R0
and that R0 R1 . An increase in resistance, DR, of all resistances still results in
a balanced bridge. However, if R1 and R3 increase by DR, and R2 and R4
decrease by DR, then
DR
Dv o ¼ vi (2.6)
R0
Because of the symmetry a similar expression results if R2 and R4 increase by
DR and R1 and R3 decrease by DR. Note that (2.6), for the four-active-arm
bridge, shows that Dv o is linearly related to DR. A nonlinearity in DR=R0 is
present even when R0 =R1 ¼ 0.
It is common practice to incorporate a balancing scheme in the bridge
circuit [see Figure 2.2(b)]. Resistor Ry and potentiometer Rx are used to
change the initial resistance of one or more arms. This arrangement brings the
bridge into balance so that zero voltage output results from ‘‘zero’’ (or base-
level) input of the measured parameter.
To minimize loading effects, Rx is approximately 10 times the resistance of
the bridge leg, and Ry limits the maximal adjustment. Strain-gage applications
normally use a value of Ry ¼ 25 times the resistance of the bridge leg.
Alternating-current (ac) balancing circuits are more complicated because a
reactive as well as a resistive imbalance must be compensated.
Figure 2.7 shows (a) self-inductance, (b) mutual-inductance, and (c) differ-
ential transformer types of inductive displacement sensors. It is usually possible to
convert a mutual-inductance system into a self-inductance system by series of
parallel connections of the coils. Note in Figure 2.7 that the mutual-inductance
device (b) becomes a self-inductancedevice (a) whenterminals b–c are connected.
An inductive sensor has an advantage in not being affected by the
dielectric properties of its environment. However, it may be affected by
external magnetic fields due to the proximity of magnetic materials.
The variable-inductance method employing a single displaceable core is
shown in Figure 2.7(a). This device works on the principle that alterations in
the self-inductance of a coil may be produced by changing the geometric form
factor or the movement of a magnetic core within the coil. The change in
inductance for this device is not linearly related to displacement. The fact that
these devices have low power requirements and produce large variations in
inductance makes them attractive for radiotelemetry applications.
The mutual-inductance sensor employs two separate coils and uses the varia-
tion in their mutual magnetic coupling to measure displacement [Figure 2.7(b)].
Cobbold (1974) describes the application of these devices in measuring cardiac
dimensions, monitoring infant respiration, and ascertaining arterial diameters.
Van Citters (1966) provides a good description of applications of mutual
inductance transformers in measuring changes in dimension of internal organs
(kidney, major blood vessels, and left ventricle). The induced voltage in the
secondary coil is a function of the geometry of the coils (separation and axial
alignment), the number of primary and secondary turns, and the frequency and
amplitude of the excitation voltage. The induced voltage in the secondary coil
is a nonlinear function of the separation of the coils. In order to maximize the
output signal, a frequency is selected that causes the secondary coil (tuned
circuit) to be in resonance. The output voltage is detected with standard
demodulator and amplifier circuits.
The linear variable differential transformer (LVDT) is widely used in
physiological research and clinical medicine to measure pressure, displacement,
and force (Kesavan and Reddy, 2006). As shown in Figure 2.7(c), the LVDT is
composed of a primary coil (terminals a–b) and two secondary coils (c–e and d–e)
connected in series. The coupling between these two coils is changed by the
c02_1 12/02/2008 55
Figure 2.8 (a) As x moves through the null position, the phase changes 1808,
while the magnitude of v o is proportional to the magnitude of x. (b) An
ordinary rectifier demodulator cannot distinguish between (a) and (b), so a
phase-sensitive demodulator is required.
c02_1 12/02/2008 56
magnitude of output voltage results from two very different input displace-
ments. The direction of displacement may be determined by using the fact that
there is a 1808 phase shift when the core passes through the null position. A
phase-sensitive demodulator is used to determine the direction of displace-
ment. Figure 3.17 shows a ring-demodulator system that could be used with the
LVDT.
A
C ¼ e0 er (2.8)
x
where e0 is the dielectric constant of free space (Appendix A.1) and er is the
relative dielectric constant of the insulator (1.0 for air) (Bowman and Meindl,
1988). In principle it is possible to monitor displacement by changing any of
the three parameters er , A, or x. However, the method that is easiest to
implement and that is most commonly used is to change the separation
between the plates.
The sensitivity K of a capacitive sensor to changes in plate separation Dx is
found by differentiating (2.8).
DC A
K¼ ¼ e0 er 2 (2.9)
Dx x
Note that the sensitivity increases as the plate separation decreases.
By substituting (2.8) into (2.9), we can develop an expression showing that
the percent change in C about any neutral point is equal to the per-unit change
in x for small displacements. Thus
dC C
¼ (2.10)
dx x
or
dC dx
¼ (2.11)
C x
q ¼ kf (2.13)
where k is the piezoelectric constant, C=N. The change in voltage can be found
by assuming that the system acts like a parallel-plate capacitor where the
voltage v across the capacitor is charge q divided by capacitance C. Then, by
substitution of (2.8), we get
kf kfx
v¼ ¼ (2.14)
C e0 er A
Tables of piezoelectric constants are given in the literature (Lion, 1959; and
Cobbold, 1974).
Typical values for k are 2.3 pC=N for quartz and 140 pC=N for barium
titanate. For a piezoelectric sensor of 1 cm2 area and 1 mm thickness with an
applied force due to a 10 g weight, the output voltage v is 0.23 mV and 14 mV
for the quartz and barium titanate crystals, respectively.
There are various modes of operation of piezoelectric sensors, depending
on the material and the crystallographic orientation of the plate (Lion, 1959).
These modes include the thickness or longitudinal compression, transversal
compression, thickness-shear action, and face-shear action.
Also available are piezoelectric polymeric films, such as polyvinylidene
fluoride (PVDF) (Hennig, 1988; Webster, 1988). These films are very thin,
lightweight and pliant, and they can be cut easily and adapted to uneven
surfaces. The low mechanical quality factor does not permit resonance appli-
cations, but it permits acoustical broadband applications for microphones and
loudspeakers.
Piezoelectric materials have a high but finite resistance. As a consequence,
if a static deflection x is applied, the charge leaks through the leakage resistor
(on the order of 100 GV). It is obviously quite important that the input
impedance of the external voltage-measuring device be an order of magnitude
higher than that of the piezoelectric sensor. It would be helpful to look at the
c02_1 12/02/2008 59
q ¼ Kx (2.15)
where where
K ¼ proportionality constant, C/m
x ¼ deflection
dq dx
is ¼ ¼K (2.16)
dt dt
c02_1 12/02/2008 60
The modified circuit is shown in Figure 2.10(b), where the resistances and
capacitances have been combined. Assuming that the amplifier does not draw
any current, we then have
is ¼ iC þ iR (2.17)
Z
1
vo ¼ vC ¼ iC dt (2.18)
C
dv o dx v o
is iR ¼ C ¼K (2.19)
dt dt R
or
Vo ð jvÞ Ks jvt
¼ (2.20)
Xð jvÞ jvt þ 1
where
Ks ¼ K=C (sensitivity, V/m)
t ¼ RC (time constant)
EXAMPLE 2.2 A piezoelectric sensor has C ¼ 500 pF. The sensor leakage
resistance is 10 GV. The amplifier input impedance is 5 MV. What is the low-
corner frequency?
Note that by increasing the input impedance of the amplifier by a factor of 100,
we can lower the low-corner frequency to 0.64 Hz.
EXAMPLE 2.3 For a piezoelectric sensor plus cable that has 1 nF capaci-
tance, design a voltage amplifier (not a charge amplifier) by using only one
noninverting amplifier that has a gain of 10. It should handle a charge of 1 mC
generated by the carotid pulse without saturation. It should not drift into
saturation because of bias currents. It should have a frequency response from
0.05 to 100 Hz. Add the minimal number of extra components to achieve the
design specifications.
2.8 THERMOCOUPLES 63
control systems are used to maintain a desirable ambient temperature for the
infant.
In the study of arthritis, physicians have shown that temperatures of joints
are closely correlated with the amount of local inflammation. The increased
blood flow due to arthritis and chronic inflammation can be detected by
thermal measurements.
The specific site of body-temperature recording must be selected carefully
so that it truly reflects the patient’s temperature. Also, environmental changes
and artifacts can cause misleading readings. For example, the skin and oral-
mucosa temperature of a patient seldom reflects true body-core temperature.
The following types of thermally sensitive methods of measurement will be
described here: thermocouples, thermistors, and radiation and fiber-optic
detectors (Samaras, 2006). The voltage across a p–n junction changes about
2 mV/ C so temperature sensors that use this principle are available (Togawa,
2006).
2.8 THERMOCOUPLES
1
E ¼ aT þ bT 2 þ (2.21)
2
where T is in degrees Celsius and the reference junction is maintained at 0 8C.
Figure 2.13(a) is a thermocouple circuit with two dissimilar metals, A and
B, at two different temperatures, T1, and T2. The net emf at terminals c–d is a
function of the difference between the temperatures at the two junctions and
the properties of the two metals. In the practical situation, one junction is held
at a constant known temperature (by an ice bath or controlled oven) for a
reference in order to determine the desired or unknown temperature.
c02_1 12/02/2008 64
Figure 2.13 Thermocouple circuits (a) Peltier emf, (b) law of homogeneous
circuits, (c) law of intermediate metals, (d) law of intermediate temperatures.
dE
a¼ ¼ a þ bT þ (2.22)
dT
c02_1 12/02/2008 65
2.8 THERMOCOUPLES 65
Figure 2.14 The LT1025 electronic cold junction and the hot junction of the
thermocouple yield a voltage that is amplified by an inverting amplifier.
Note that a is not a constant but varies (usually increases) with temperature.
The sensitivities of common thermocouples range from 6.5 to 80 mV/ C at
20 C, with accuracies from ¼% to 1%.
For accurate readings, the reference junction should be kept in a triple-
point-of-water device the temperature of which is 0:01 0:0005 C (Doebelin,
1990). Normally the accuracy of a properly constructed ice bath, 0:05 C with a
reproducibility of 0:001 C is all that is necessary. Temperature-controlled
ovens can maintain a reference temperature to within 0:4 C. Figure 2.14
shows that modern thermocouple signal conditioners contain an electronic
cold junction (Tompkins and Webster, 1988; Sheingold, 1980).
Increased sensitivity may be achieved by connecting a number of ther-
mocouples in series, all of them measuring the same temperature and using the
same reference junction. An arrangement of multiple-junction thermocouples
is referred to as a thermopile. Parallel combinations may be used to measure
average temperature.
It is easy to obtain a direct readout of the thermocouple voltage using a
digital voltmeter. Chart recordings may be secured by using a self-balancing
potentiometer system. The linearity of this latter device is dependent only
on the thermocouple and potentiometer; it is independent of the other
circuitry.
Thermocouples have the following advantages: fast response time (time
constant as small as 1 ms), small size (down to 12 mm diameter), ease of
fabrication, and long-term stability. Their disadvantages are small output
voltage, low sensitivity, and the need for a reference temperature.
Numerous examples of the use of thermocouples in biomedical re-
search are given in the literature (Wren, 2006). Thermocouples can be
made small in size, so they can be inserted into catheters and hypodermic
needles.
c02_1 12/02/2008 66
2.9 THERMISTORS
2.9 THERMISTORS 67
where
b ¼ material constant for thermistor, K
T0 ¼ standard reference temperature, K
The value of b increases slightly with temperature. However, this does not
present a problem over the limited temperature spans for biomedical work
(10 C to 20 C). b, also known as the characteristic temperature, is in the range
of 2500 to 5000 K. It is usually about 4000 K.
The temperature coefficient a can be found by differentiating (2.23) with
respect to T and dividing by Rt. Thus
1 dR1 b
a¼ ¼ 2 ð%=KÞ (2.24)
Rt dT T
EXAMPLE 2.4 Sketch typical thermistor v–i characteristics with and with-
out a heat sink. Explain why there is a difference.
c02_1 12/02/2008 68
ANSWER
P V 2 =R
DT ¼ ¼
D:C: D:C:
V2 52
R¼ ¼ ¼ 125; 000 V
DTðD:C:Þ 0:1 Cð0:002 W= CÞ
artery occlusions, and so forth) (Qi, 2006). Here we shall deal with the basic
principles of thermal radiation and detector systems.
Every body that is above absolute zero radiates electromagnetic power,
the amount being dependent on the body’s temperature and physical propert-
ies. For objects at room temperature, the spectrum is predominantly in the far-
and extreme-far-infrared regions.
A blackbody is an ideal thermal radiator; as such, it absorbs all incident
radiation and emits the maximal possible thermal radiation. The radiation
emitted from a body is given by Planck’s law multiplied by emissivity e. This
expression relates the radiant flux per unit area per unit wavelength Wl at a
wavelength l ðmmÞ and is stated as
eC1
Wl ¼ ðW/cm2 mmÞ (2.25)
l5 ðeC2 =lT 1Þ
where
T ¼ blackbody temperature, K
e ¼ emissivity, the extent by which a surface deviates from a blackbody
(e ¼ 1)
Figure 2.16(a) shows a plot of (2.25), the spectral radiant emittance versus
wavelength for a blackbody at 300 K.
Wien’s displacement law gives the wavelength lm for which Wl is a
maximum. It can simply be found by differentiating (2.25) and setting this
to zero.
2898
lm ¼ ðmmÞ (2.26)
T
Figure 2.16(a) indicates lm ¼ 9:66 mm ðT ¼ 300 KÞ. Note from (2.25) that the
maximal level of spectral emittance increases with T, and from (2.26) that lm is
inversely related to T.
The total radiant power Wt, can be found by integrating the area under the
curve. This expression is known as the Stefan–Boltzmann law.
Figure 2.16 (a) Spectral radiant emittance versus wavelength for a blackbody
at 300 K on the left vertical axis; percentage of total energy on the right vertical
axis. (b) Spectral transmission for a number of optical materials. (c) Spectral
sensitivity of photon and thermal detectors. [Part (a) is from Transducers for
Biomedical Measurements: Principles and Applications, R. S. C. Cobbold.
Copyright # 1974, John Wiley & Sons, Inc. Reprinted by permission of John
Wiley & Sons, Inc. Parts (b) and (c) are from Measurement Systems: Applica-
tion and Design, E. O. Doebelin. Copyright # 1990 by McGraw-Hill, Inc. Used
with permission of McGraw-Hill Book Co.]
c02_1 12/02/2008 72
Shutter
Ambient sensor
Ear
Micro
IR Ta processor
MUX A/D
Tb
Amp.
Sensor Digital
Waveguide Window Shutter display
switch
Figure 2.18 The infrared thermometer opens a shutter to expose the sensor
to radiation from the tympanic membrane. [From J. G. Webster (ed.), Bio-
instrumentation, New York: John Wiley & Sons, 2004.]
body’s main thermostat, which regulates the core body temperature. This
approach has advantages over using mercury thermometers, thermocouples,
or thermistors. The standard temperature-measuring techniques measure the
temperature of the sensor, not that of the subject. The sensor must be in
contact with the patient long enough for its temperature to become the same
as, or close to, that of the subject whose temperature is being measured.
However, the infrared thermometry device detects emitted energy that is
proportional to the actual temperature of the subject. There is negligible
thermal time constant for the pyroelectric sensor (Fraden, 1997). The infrared
tympanic temperature-monitoring system has a response time in the order of
0.1 s and an accuracy of approximately 0:1 C. A disposable sanitary probe
cover is used to prevent cross-contamination from patient to patient. Ear
thermometry offers several clinical benefits over taking sublingual (oral) or
rectal measurements. Response is rapid, and readings can be obtained
c02_1 12/02/2008 74
Figure 2.19 Details of the fiber-sensor arrangement for the GaAs semi-
conductor temperature probe.
Optical systems are widely used in medical diagnosis. The most common use
occurs in the clinical-chemistry lab, in which technicians analyze samples of
blood and other tissues removed from the body. Optical instruments are also
used during cardiac catheterization to measure the oxygen saturation of
hemoglobin and to measure cardiac output.
c02_1 12/02/2008 75
Figure 2.20 (a) General block diagram of an optical instrument. (b) Highest
efficiency is obtained by using an intense lamp and lenses to gather and focus
the light on the sample in the cuvette, and a sensitive detector. (c) Solid-state
lamps and detectors may simplify the system.
Figure 2.20(a) shows that the usual optical instrument has a source, filter,
and detector. Figure 2.20(b) shows a common arrangement of components.
Figure 2.20(c) shows that in some cases, the function of source, filter, sample,
and detector may be accomplished by solid-state components.
The remainder of this chapter is divided into sections that deal with
sources, geometrical optics, filters, detectors, and combinations thereof.
TUNGSTEN LAMPS
Incandescent tungsten-wire filament lamps are the most commonly used
sources of radiation. Their radiant output varies with temperature and wave-
length, as given by (2.25). For l < 1 mm, tungsten has an emissivity of about
0.4 and thus emits about 40% of what it would if the emissivity were 1.0.
The relative-output spectrum shown in Figure 2.21(a) is only slightly altered.
For higher temperatures, lm , the maximal wavelength of the radiant-output
c02_1 12/02/2008 76
ARC DISCHARGES
The fluorescent lamp is filled with a low-pressure Ar–Hg mixture. Electrons
are accelerated and collide with the gas atoms, which are raised to an excited
level. As a given atom’s electron undergoes a transition from a higher level to a
lower level, the atom emits a quantum of energy. The energy per quantum
E ¼ hv ¼ hc=l, where h ¼ Planck’s constant, v ¼ frequency, c ¼ velocity of
light, and l ¼ wavelength.
Because the strongest transition of the mercury atom corresponds to
about 5 eV, Figure 2.21(e) shows the resulting wavelength to be about
250 nm. A phosphor on the inside of the glass bulb absorbs this ultraviolet
radiation and emits light of longer, visible wavelengths. The fluorescent
lamp has low radiant output per unit area, so it is not used in optical
instruments. However, it can be rapidly turned on and off in about 20 ms,
so it is used in the tachistoscope (which presents brief stimuli to the eye) used
in measurements of visual perception. Other low-pressure discharge lamps
include the glow lamp (such as the neon lamp), the sodium-vapor lamp, and
the laser.
High-pressure discharge lamps are more important for optical instruments
because the arc is compact and the radiant output per unit area is high. The
carbon arc has been in use for the longest time, but it has largely been replaced
by the mercury lamp (bluish-green color), the sodium lamp (yellow color), and
the xenon lamp (white color). These lamps usually have a clear quartz bulb
with electrodes at both ends of the spherical bulb. The zirconium arc lamp
provides an intense point source.
c02_1 12/02/2008 78
Figure 2.22 Forward characteristics for p–n junctions. Ordinary silicon diodes
have a band gap of 1.1 eV and are inefficient radiators in the near-infrared
range. GaAs has a band gap of 1.44 eV and radiates at 900 nm. GaP has a band
gap of 2.26 eV and radiates at 700 nm.
LASERS
Laser (light amplification by stimulated emission of radiation) action can occur
in GaAs. The end faces that are perpendicular to the p–n junction are polished
c02_1 12/02/2008 79
to serve as partial mirrors, thus forming a resonant optical cavity. The forward
current pumps a large population of the molecules to an excited energy level.
Radiation incident on the molecules causes the production of additional
radiation that is identical in character. This phenomenon, known as stimulated
emission, is produced by the feedback from the mirrors. Laser output is highly
monochromatic, collimated (parallel), and phase coherent. However, p–n
junction lasers are not widely used because they operate in the infrared
and require current densities of 103 A/cm2 or more, thus necessitating pulsed
(10–100 ns) operation rather than continuous wave (CW).
The most common laser is the He–Ne laser that operates at 633 nm in the
red region, as shown in Figure 2.21(a). The laser is operated by a low-pressure
arc similar to a neon sign and provides up to 100 mW. Partially reflective
mirrors at each end provide the resonant optical cavity and laser action.
Argon lasers provide the highest continuous-power levels (1–15 W) in the
visible part of the spectrum at 515 nm [Figure 2.21(a)]. This high-power output
permits photocoagulation of blood vessels in the eyes of patients suffering
from diabetic retinopathy.
CO2 lasers provide 50–500 W of CW output power and are used for cutting
plastics, rubber, and metals up to 1 cm thick.
Two solid-state lasers—both usually operated in the pulsed mode—are
widely used. The lasers are pumped by firing a flash tube that is wound around
them. The ruby laser has a moderate (1 mJ) output in the red region of the
spectrum at 693 nm, as shown in Figure 2.21(a). The neodymium in yttrium
aluminum garnet (Nd: YAG) laser has a high (2 W/mm2) output in the infrared
region at 1064 nm, as shown in Figure 2.21(a).
The most important medical use of the laser has been to mend tears in the
retina. A typical photocoagulator uses a pulsed ruby laser with a controllable
output. It is focused on a tear in the retina. The heat dissipated by the pulse forms
a burn, which, on healing, develops scar tissue that mends the original tear.
Section 13.10 provides further information on therapeutic applications of lasers.
Safety to the eye should be considered with respect to some light sources. It is
safe to look at a 100 W frosted light bulb for long periods of time. However,
looking at clear incandescent lamps, the sun, high-pressure arc sources, or lasers
can cause burns on the retina. Protective eyewear worn by the physician to protect
against lasers usually consists of a set of filters that attenuate at the specific
wavelengths emitted by the laser but transmit as much visible radiation as possible.
GEOMETRICAL OPTICS
There are a number of geometric factors that modify the power transmitted
between the source and the detector. In Figure 2.20(b), the most obvious
optical elements are the lenses. The lamp emits radiation in all directions. The
c02_1 12/02/2008 80
first lens should have as small an f number (ratio of focal length to diameter) as
practical. Thus it collects the largest practical solid angle of radiation from the
lamp. The first lens is usually placed one focal length away from the lamp, so
that the resulting radiation is collimated (that is, the rays are parallel). Thus, for
a point source, the second lens can be placed at any distance without losing any
radiation. Also, some interference filters operate best in collimated rays.
The second lens focuses the radiation on a small area of sample in the
cuvette. Because the radiation now diverges, third and fourth lenses are used to
collect all the radiation and focus it on a detector. Some spectrophotometers
[Figure 2.20(c)] transmit collimated radiation through the sample section. The
lenses can be coated with a coating that is a quarter-wavelength thick to
prevent reflective losses at air–glass surfaces. Full mirrors may be used to fold
the optical path to produce a compact instrument. Half-silvered mirrors enable
users to split the beam into two beams for analysis or to combine two beams for
analysis by a single detector. Curved mirrors may function as lenses for
wavelengths that are absorbed by normal glass lenses.
Scattered radiation must be prevented from reaching the detector. Inter-
nal support structures and mechanical components of optical instruments are
internally painted flat black to prevent scattered radiation. Stops (apertures
that pass only the desired beam size) may be placed at several locations along
the instrument’s optical axis to trap scattered radiation.
FIBER OPTICS
Fiber optics are an efficient way of transmitting radiation from one point to
another (Modell and Perelman, 2006). Transparent glass or plastic fiber with a
refractive index n1 is coated or surrounded by a second material of a lower
refractive index n2. By Snell’s law,
n2
sin uic ¼ (2.29)
n1
A ray is internally reflected for all angles of incidence greater than uic . Because
rays entering the end of a fiber are usually refracted from air ðn ¼ 1:0Þ into
glass (n ¼ 1:62 for one type), a larger cone of radiation ðu3 Þ is accepted by a
fiber than that indicated by calculations using 90 uic . Rays entering the end
of the fiber at larger angles ðu4 Þ are not transmitted down the fiber; they escape
through the walls.
c02_1 12/02/2008 81
Figure 2.23 Fiber optics The solid line shows refraction of rays that escape
through the wall of the fiber. The dashed line shows total internal reflection
within a fiber.
LIQUID CRYSTALS
Liquid crystals change their state in such a way that they modify passive
scattering or absorption of light. As the crystals melt, the three-dimensional
order becomes a two-dimensional or one-dimensional order. Layers or strands
form that can be seen as a clarification of the previously turbid melt.
In one medical application, the patient’s body is painted with a black water-
soluble varnish to show up the color of the liquid crystals better. Liquid crystals
are painted over the varnish, and any inflammation causes a rise in temperature
that is indicated by a color pattern. Liquid crystals are also used, in disposable
thermometers, in the measurement of oral temperatures. They are widely used
in wristwatches, because a low-voltage (1–15 V), low-power ð1 mW/cm2 Þ
electric field causes observable changes in digital-display elements.
FILTERS
Filters are frequently inserted in the optical system to control the distribution
of radiant power or wavelength. To reduce radiant power only, neutral-density
filters are used. When glass is partially silvered, most of the power is reflected,
and the desired fraction of the power is transmitted. When carbon particles are
suspended in plastic, most of the power is absorbed and the desired fraction of
the power is transmitted. Two Polaroid filters may also be used to attenuate the
light. Each filter transmits only that portion of the light that is in a particular
state of polarization. As one is rotated with respect to the other, the optical
transmission of the combination varies.
Color filters transmit certain wavelengths and reject others. Gelatin filters
are the most common type of absorption filters. An organic dye is dissolved in
an aqueous gelatin solution, and a thin film is dried on a glass substrate. An
example shown in Figure 2.21(b) is the infrared Kodak 87 Wratten filter. Glass
filters, made by combining additives with the glass itself in its molten state, are
extensively used. They provide rather broad passbands, as illustrated by the
blue Corning 5-56 filter shown in Figure 2.21(b).
Interference filters are formed by depositing a reflective stack of layers on
both sides of a thicker spacer layer. This sandwich construction provides multiple
reflection and interference effects that yield sharp-edge high, low, and bandpass
filters with bandwidths from 0.5 to 200 nm. Interference filters are generally used
with collimated radiation and cost more than those just mentioned. Interference
coatings are used on dichroic mirrors (cold mirrors), which reflect visible
radiation from projection lamps. The nonuseful infrared radiation is transmitted
through the coating and mirror to the outside of the optical system. This reduces
heat within the optical system without sacrificing the useful light.
Diffraction gratings are widely employed to produce a wavelength spec-
trum in the spectrometer. Plane gratings are formed by cutting thousands of
c02_1 12/02/2008 83
THERMAL SENSORS
The thermal sensor absorbs radiation and transforms it into heat, thus causing a
rise in temperature in the sensors. Typical thermal sensors are the thermistor and
the thermocouple. The sensitivity of such a sensor does not change with (is flat
with) wavelength, and the sensor has slow response [Figure 2.21(c)]. Changes in
output due to changes in ambient temperature cannot be distinguished from
changes in output due to the source, so a windmill-shaped mechanical chopper is
frequently used to interrupt the radiation from the source periodically.
The pyroelectric sensor (Fraden, 1997) absorbs radiation and converts it
into heat. The resulting rise in temperature changes the polarization of the
crystals, which produces a current proportional to the rate of change of
temperature. As it is for the piezoelectric sensor, dc response is zero, so a
chopper is required for dc measurements.
QUANTUM SENSORS
Quantum sensors absorb energy from individual photons and use it to release
electrons from the sensor material. Typical quantum sensors are the eye, the
phototube, the photodiode, and photographic emulsion. Such sensors are
sensitive over only a restricted band of wavelengths; most respond rapidly.
Changes in ambient temperature cause only a second-order change in sensi-
tivity of these sensors.
c02_1 12/02/2008 84
PHOTOEMISSIVE SENSORS
Photoemissive sensors—an example is the phototube—have photocathodes
coated with alkali metals. If the energy of the photons of the incoming
radiation is sufficient to overcome the work function of the photocathode,
the forces that bind electrons to the photocathode are overcome, and it emits
electrons. Electrons are attracted to a more positive anode and form a current
that is measured by an external circuit. Photon energies below 1 eV are not
large enough to overcome the work function, so wavelengths longer than
1200 nm cannot be detected. Figure 2.21(c) shows the spectral response of the
most common photocathode, the S4, which has lower sensitivity in the ultra-
violet region because of absorption of radiation in the glass envelope.
The photomultiplier shown in Figure 2.24 is a phototube combined with an
electron multiplier (Lion, 1975). Each accelerated electron hits the first dynode
with enough energy to liberate several electrons by secondary emission. These
electrons are accelerated to the second dynode, where the process is repeated,
and so on. Time response is less than 10 ns. Photomultipliers are the most
sensitive photodetectors. When they are cooled (to prevent electrons from
being thermally generated), they can count individual photons. The eye is
almost as sensitive; under the most favorable conditions, it can detect six
photons arriving in a small area within 100 ms. Photodiodes have replaced
photomultipliers in many applications.
PHOTOCONDUCTIVE CELLS
Photoresistors are the simplest solid-state photoelectric sensors. A photo-
sensitive crystalline material such as CdS or PbS [Figure 2.21(c)] is deposited
c02_1 12/02/2008 85
PHOTOJUNCTION SENSORS
Photojunction sensors are formed from p–n junctions and are usually made of
silicon. If a photon has sufficient energy to jump the band gap, hole–electron pairs
are produced that modify the junction characteristics, as shown in Figure 2.25. If
the junction is reverse-biased, the reverse photocurrent flowing from the cathode
to the anode increases linearly with an increase in radiation. The resulting
photodiode responds in about 1 ms. In phototransistors, the base lead is not
connected, and the resulting radiation-generated base current is multiplied by the
current gain (beta) of the transistor to yield a large current from collector to
emitter. The radiation–current characteristics have a nonlinearity of about 2%
because beta varies with collector current. The response time is about 10 ms.
PHOTOVOLTAIC SENSORS
The same silicon p–n junction can be used in the photovoltaic mode. Figure
2.25 shows that there is an open-circuit voltage when the junction receives
radiation. The voltage rises logarithmically from 100 to 500 mV as the input
radiation increases by a factor of 104. This is the principle of the solar cell that
is used for direct conversion of the sun’s radiation into electric power.
SPECTRAL RESPONSE
All of the aforementioned silicon sensors have the spectral response shown in
Figure 2.21(c). There is no response above 1100 nm because the energy of the
photons is too low to permit them to jump the band gap. For wavelengths
shorter than 900 nm, the response drops off because there are fewer, more-
energetic photons per watt. Each photon generates only one hole–electron
pair.
Because none of the common sensors is capable of measuring the radiation
emitted by the skin (300 K), which has a peak output at 9000 nm, special
sensors have been developed, such as the InSb sensor shown in Figure 2.21(c).
where
Sl ¼ relative source output
Fl ¼ relative filter transmission
Dl ¼ relative sensor responsivity
c02_1 12/02/2008 87
PROBLEMS 87
Figure 2.21(d) shows several results of this type of calculation. One of the
examples shown is an efficient system capable of making measurements in
the dark without stimulating the eye. Such a device can be used for tracking
eye movements. It can be formed from a tungsten source, a Kodak 87
Wratten filter, and a silicon sensor. If GaAs provides enough output, it can
replace both the tungsten source and the Kodak 87 Wratten filter (Borah,
2006).
PROBLEMS
2.1 For Figure P2.1, plot the ratio of the output voltage to the input voltage
v o =v i as a function of the displacement xi of a potentiometer with a total
displacement xt for ranges of Rm, the input resistance of the meter. Show that
the maximal error occurs in the neighborhood of xi =xt ¼ 0:67. What is the
value of this maximal error?
Figure P2.1
REFERENCES
Alihanka, J., K. Vaahtoranta, and S.-E Bj€orkqvist, Apparatus in medicine for the monitoring and/
or recording of the body movements of a person on a bed, for instance of a patient. United
States Patent 4,320,766, 1982.
Anonymous, Manual on the Use of Thermocouples in Temperature Measurement. Publication
470A. Philadelphia: American Society for Testing and Materials, 1974.
Borah, J., ‘‘Eye movement, measurement techniques for.’’ In J. G. Webster (ed.), Encyclopedia of
Medical Devices and Instrumentation, 2nd ed. New York: Wiley, 2006, Vol. 3, pp. 263–286.
Bowman, L., and J. D. Meindl, ‘‘Capacitive sensors.’’ In J. G. Webster (ed.), Encyclopedia of
Medical Devices and Instrumentation. New York: Wiley, 1988, pp. 551–556.
Cobbold, R. S. C, Transducers for Biomedical Measurements. New York: John Wiley & Sons, Inc., 1974.
Conway, J. M., K. H. Norris, and C. E. Bodwell, ‘‘A new approach for the estimation of body
composition: Infrared interactance.’’ Am. J. Clin. Nutr., 1984, 40, 1123–1130.
Dechow, P. C., and Q. Wang, ‘‘Strain gages.’’ In J. G. Webster (ed.), Encyclopedia of Medical
Devices and Instrumentation. 2nd ed. New York: Wiley, 2006, Vol. 6, pp. 282–290.
Doebelin, E. O., Measurement Systems: Application and Design, 4th ed. New York: McGraw-Hill,
1990.
Fraden, J., ‘‘Noncontact temperature measurements in medicine.’’ In D. L. Wise (ed.), Bio-
instrumentation and Biosensors. New York: Marcel Dekker, 1991, pp. 511–550.
c02_1 12/02/2008 89
REFERENCES 89
Fraden, J., Handbook of Modern Sensors: Physics, Designs, and Applications. 2nd ed. Woodbury,
NY: American Institute of Physics, 1997.
Geddes, L. A., and L. E. Baker, Principles of Applied Biomedical Instrumentation, 3rd ed. New
York: Wiley, 1989.
Hennig, E. M., ‘‘Piezoelectric sensors.’’ In J. G. Webster (ed.), Encyclopedia of Medical Devices
and Instrumentation. New York: Wiley, 1988, pp. 2310–2319.
Hokanson, D. E., D. S. Sumner, and D. E. Strandness, Jr.. ‘‘An electrically calibrated plethysmo-
graph for direct measurement of limb blood flow.’’ IEEE Trans. Biomed. Eng., 1975, BME-
22, 25–29.
Hooper, B. A., ‘‘Endoscopes.’’ In J. G. Webster (ed.), Encyclopedia of Medical Devices and
Instrumentation, 2nd ed., New York: Wiley, 2006, Vol. 3, pp. 177–189.
Kesavan, S. K., and N. P. Reddy, ‘‘Linear variable differential transformers.’’ In J. G. Webster
(ed.), Encyclopedia of Medical Devices and Instrumentation. 2nd ed., New York: Wiley, 2006,
Vol. 4, pp. 252–257.
Korites, B. J., Microsensors. Rockland, MA: Kern International, 1987.
Lawton, R. W., and C. C. Collins, ‘‘Calibration of an aortic circumference gauge.’’ J Appl. Physiol.,
1959, 14, 465–467.
Li, Y., and X.-L. Su, ‘‘Piezoelectric sensors,’’ in J. G. Webster (ed.), Encyclopedia of Medical
Devices and Instrumentation, 2nd ed., New York: Wiley, 2006, Vol. 5, pp. 359–367.
Lion, K. S., Elements of Electronic and Electrical Instrumentation. New York: McGraw-Hill, 1975.
Lion, K. S., Instrumentation in Scientific Research. New York: McGraw-Hill, 1959.
Mendelson, Y., ‘‘Optical sensors.’’ In J. G. Webster (ed.), Encyclopedia of Medical Devices and
Instrumentation, 2nd ed. New York: Wiley, 2006, Vol. 5, pp. 160–175.
Melo, P. L., ‘‘Thermistors.’’ In J. G. Webster (ed.), Encyclopedia of Medical Devices and
Instrumentation, 2nd ed. New York: Wiley, 2006, Vol. 6, pp. 320–340.
Modell, M. D., and L. T. Perelman, ‘‘Fiber optics in medicine.’’ In J. G. Webster (ed.),
Encyclopedia of Medical Devices and Instrumentation, 2nd ed. New York: Wiley, 2006,
Vol. 3, pp. 301–315.
Northrop, R. B., Noninvasive Instrumentation and Measurement in Medical Diagnosis. Boca
Raton: CRC Press, 2002.
Nyce, D. S., Linear Position Sensors Theory and Application, Hoboken: Wiley, 2004.
Pallás-Areny, R., and J. G. Webster, Sensors and Signal Conditioning, 2nd ed. New York: Wiley,
2001.
Patel, A., M. Kothari, J. G. Webster, W. J. Tompkins, and J. J. Wertsch, ‘‘A capacitance pressure
sensor using a phase-locked loop.’’ J. Rehabil. Res. Devel., 1989, 26(2), 55–62.
Qi, H., ‘‘Thermography’’ In J. G. Webster (ed.), Encyclopedia of Medical Devices and Instrumen-
tation, 2nd ed. New York: Wiley, 2006, Vol. 6, pp. 346–355.
Re, T. J., and M. R. Neuman, ‘‘Thermal contact-sensing electronic thermometer.’’ Biomed.
Instrum. Technol., 1991, 25, 54–59.
Samaras, T., ‘‘Thermometry.’’ In J. G. Webster (ed.), Encyclopedia of Medical Devices and
Instrumentation, 2nd. ed. New York: Wiley, 2006, Vol. 6, pp. 355–362.
Sapoff, M., Thermistors for Biomedical Use. Fifth Symposium on Temperature, Proceedings.
Washington, DC: Instrument Society of America, 1971, 2109–2121.
Servais, S. B., J. G. Webster, and H. J. Monotoye, ‘‘Estimating human energy expenditure using an
accelerometer device.’’ J. Clin. Eng., 1984, 9, 159–171.
Sheingold, D. H. (ed.), Transducer Interfacing Handbook. Norwood, MA: Analog Devices, Inc.,
1980.
Sirohi, R. S., and M. P. Kothiyal, Optical Components, Systems, and Measurement Techniques. New
York: Marcel Dekker, 1991.
Stimson, A., Photometry and Radiometry for Engineers. New York: Wiley, 1974.
Sydenham, P. H., N. H. Hancock, and R. Thorn, Introduction to Measurement Science and
Engineering, New York; Wiley, 1992.
Togawa, T., ‘‘Integrated circuit temperature sensor.’’ In J. G. Webster (ed.), Encyclopedia
of Medical Devices and Instrumentation 2nd ed. New York: Wiley, 2006, Vol. 4, pp.
157–162.
c02_1 12/02/2008 90
Togawa, T., T. Tamura, and P. A. Oberg, Biomedical Transducers and Instruments. Boca Raton:
CRC Press, 1997.
Tompkins, W. J., and J. G. Webster (eds.), Interfacing Sensors to the IBM PC. Englewood Cliffs,
NJ: Prentice-Hall, 1988.
Tsoukalas, D., S. Chatzandrouroulis, and D. Goustouridis, ‘‘Capacitive microsensors for bio-
medical applications.’’ In J. G. Webster (ed.), Encyclopedia of Medical Devices and
Instrumentation, 2nd ed. New York: Wiley, 2006, Vol. 2, pp. 1–12.
Udd, E. (ed.), Fiber Optic Sensors: An Introduction for Engineers and Scientists. New York: Wiley,
1991.
vanCitters, R. L., ‘‘Mutual inductance transducers.’’ In R. F. Rushmer (ed.), Methods in Medical
Research. Vol. XI. Chicago: Year Book, 1966, 26–30.
Webster, J. G. (ed.), Tactile Sensors for Robotics and Medicine. New York: Wiley, 1988.
Wren, J., ‘‘Thermocouples.’’ In J. G. Webster (ed.), Encyclopedia of Medical Devices and
Instrumentation, 2nd ed. New York: Wiley, 2006, Vol. 6, pp. 340–346.