Astm E168
Astm E168
Astm E168
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
where:
Ai = total absorbance at wavenumber i,
ain = absorptivity at the wavenumber i of component n,
b = path length of the cell in which the mixture is sampled,
and
cn = concentration of component n in the mixture.
10.2 During calibration, concentrations cn are known, and
baseline corrected absorbances A are measured. The experi-
mental absorptivity-path length products ainb are then calcu-
lated (see Note 2). During analysis, the absorptivity-path length
products ainb are known, and the absorbances A are measured.
The unknown concentrations are then calculated (see Section
17). Therefore, accurate calibration generally requires that
experimental absorptivity values be obtained from at least n
standards. The following requirements must be met:
10.2.1 The number of standards must be equal to or greater
than the number of analytes, n, and
FIG. 1 An Analytical Working Curve
10.2.2 The number of analytical wavenumbers, i, must be
--`,,,,```,`,`,,,,,`,`,`-`-``,```,,,`---
equal to or greater than the number of independent
calculate a. The calculation of a where curvature is present will components, n.
be discussed in 18.1 and 18.2.
NOTE 2—All absorbance conversions use transmittance (that is, the
NOTE 1—In practice, the calibration curve may not have a y intercept of decimal value), not percent transmittance. Regardless of form (that is,
zero. This could be due to a variety of factors including, but not limited to, decimal or percent), the term transmittance refers to the term P/P0 of Eq
incompletely resolved analyte bands, reflection losses, and solvent inter- 2, and should not be called transmission. (See Terminology E131).
ferences. It is important that the method used to calculate the calibration
curve not force the y intercept to be zero. 10.3 The first requirement allows the analyst to use more
than the minimum number of standards. Over-determination of
9.1.4 For analysis, dissolve the unknown in the solvent, standards permits error estimation in the analytical result. The
measure the absorbance, A, and determine the concentration, c, second requirement allows the use of more than the minimum
of the analyte graphically or by calculation. Convert this number of peaks for specifying a chemical system, where at
concentration in solution to the concentration in the unknown least one distinctive band is selected for each component
sample. (7-10).
9.1.5 Both analysis time and chance of error are less if the
concentrations of the unknowns and the cell path length are 10.4 The procedures used in multicomponent analysis will
kept the same over a series of analyses, and the concentrations be discussed further in the following section which is also an
of the calibration solutions have bracketed the expected high introduction to general solution phase analyses.
and low values of the unknown solutions (6, 7).
11. Multicomponent Solution Analysis
10. Theory for Multicomponent Analysis
11.1 For the quantitative analysis of mixtures, Eq 4 is
10.1 Beer’s law is expressed for a mixture of n indepen- applicable. The absorptivities ain of the n components of the
dently absorbing components at a single path length and single mixture at the ith analytical wavenumber are determined from
wavenumber as: absorbance measurements made on each component taken
A 5 a 1 bc1 1a 2 bc2 1······1a n bcn (3) individually. These absorbances must be measured under
conditions (sample path length, temperature, pressure, and
Eq 3 defines an absorbance at a wavenumber as being due to solvent) identical to those used for the unknowns, and they
the sum of the independent contributions of each component. should be corrected for baselines as discussed in Sections 12 –
In order to solve for the n component concentrations, n 14. Absorbance measurements are made with concentrations of
independent equations containing n absorbance measurements the analyte bracketing the amounts expected in the unknown
at n wavenumbers are necessary. This is expressed for constant samples.
path length as follows:
11.2 Where possible, prepare samples as dilute solutions
A 1 5 a 11bc1 1a 12bc2 1····1a 1n bcn (4)
and place in cells of appropriate path lengths (typically 0.2 to
A 2 5 a 21bc2 1a 22bc2 1····1a 2n bcn 1.0 mm). Use lower concentrations in longer path length cells
rather than higher concentrations in shorter path length cells to
·· ········ ······ obtain absorbance values in the 0.3 to 0.8 range. Lower
reproducible baseline. The correction of raw data for baseline drawn from one side to the other as in Fig. 4. The absorbance
absorbance is important in some methods. The guiding factor of the band is calculated as the value at the peak maximum A1
in baseline selection is the reproducibility of the results. minus the baseline absorbance minimum A23. An inappropriate
Methods used for drawing baselines with computerized instru-
ments are similar in most ways to those for data recorded on
chart paper. Where differences exist, they will be explained in
Annex A1.
13. Single Wavenumber Measurement
13.1 A technique known as the “cell-in-cell-out” method is
often used in single-beam infrared work. In this method, a
blank (that is, solvent in cell, potassium bromide (KBr) pellet,
or other substrate) is measured at a fixed wavenumber and then
the analyte readings are recorded (7). In the simplest cell-in-
cell-out method, a zero absorbance baseline is used (see Fig.
2). If the spectrum cannot be obtained in absorbance, the
absorbance is calculated as in Eq A12.1 where T2 = 1.0 and T1
= transmittance at the analyte wavenumber (1, 6) (see Note 2).
14. Baseline Method (7)
14.1 The cell-in-cell-out technique was the method of
choice for early single-beam infrared instruments. After the
advent of double-beam dispersive spectrometers, the baseline
method has been the method of choice. Portions of the data
around the base of the bands are picked as baseline references.
There are two common variations.
14.2 When one baseline point is chosen, the value of an
absorbance minimum, A2, is subtracted from the absorbance FIG. 3 A One-Point Baseline
--`,,,,```,`,`,,,,,`,`,`-`-``,```,,,`---
thickness of each film and apply a proportional correction for
deviations from standard thickness.
NOTE 4—The spectra of films and pellets can be complicated by the
presence of a fringe pattern. For pellets and films, follow the suggestions
in A4.5.1.2 and Note A5.1, respectively. A fringe pattern is undesirable
because analyte absorbance values can be altered by its presence.
15.2.2 In cases where all components of a mixture are
determined to a total of 100 %, it is usually sufficient to
determine only the ratios of absorbances. In such cases, it is not
necessary to know the thickness of the sample layer; it is only
necessary to know the ratio of the components. However, a
FIG. 4 A Two-Point Baseline knowledge of the thickness is needed to determine the presence
of impurities because the total then will be less than 100 %.
15.2.3 The above procedure for films is also used with
choice of baseline in this situation may have deleterious effects powders prepared as mulls. Measurement of thickness can be
on the accuracy of the final calculation. accomplished by an internal standard technique as described in
A4.4.2. This involves the addition to the sample of a known
NOTE 3—The above baseline correction procedure should be performed weight ratio of a compound having an absorption band of
only if the spectrum is plotted in absorbance units. When the spectrum is
plotted in transmittance, the two baseline transmittances and the transmit- known absorptivity that does not overlap the bands of the
tance at the analytical wavenumber should be converted to absorbance. sample.
The corrected baseline absorbance can be calculated by Eq A12.1 in 15.2.4 When powders are measured as pressed plates or
Annex A12. Conversion to absorbance is required because a sloping linear pellets, analytical curves are prepared in the same manner as
baseline in transmittance becomes curved in absorbance.
solutions, see Sections 9 and 11.
15. Nonsolution Analyses 15.3 Gases:
15.1 Liquids: 15.3.1 All calibration measurements for a given analysis
15.1.1 Analyzing a liquid mixture without the use of a must be made at a fixed total pressure. This pressure must be
diluting solvent is sometimes complicated by intermolecular equal to the total pressure employed in the analysis. An
forces. An absorption band may undergo intensity changes or analysis may be set up in either of two ways:
frequency shifts, or both, relative to the same absorption band 15.3.1.1 Method 1—A fixed sample pressure is established
of the component in solution. The absorbance contribution of a that is a fraction of the total pressure obtained by addition of a
component in a mixture can seldom be calculated from its nonabsorbing diluent gas.
absorbance measured in the pure state. It is desirable to 15.3.1.2 Method 2—A fixed sample pressure is used as the
determine the absorptivities from known mixtures having total pressure. Analytical curves are prepared by introducing a
proportions near those of the samples. pure component at various measured pressures which bracket
15.1.2 Prepare mixtures having known concentrations of the the expected component pressures in the sample. A diluent gas
various components covering the expected ranges. Measure is then added to bring the total pressure up to the established
baseline corrected absorbances at each of the wavelengths value.
chosen for the analysis and substitute them (along with the 15.3.2 In Method 2, the analytical curve preparation does
known concentrations) in Eq 4. Solve for the absorptivity-path not allow for the possibility of band broadening for different
length products, ainb directly from the set of n simultaneous components. This factor is more properly addressed by follow-
equations, or use a multivariant method (see Annex A8) if ing Method 1 where the same diluent gas is employed for
sufficient data are available. sample preparation and calibration. Low molecular weight
15.1.3 If the concentrations in the unknowns vary widely, gases frequently produce very strong, sharp absorption fea-
calculation of a second set of the ainb products is recom- tures. Addition of a diluent gas and use of pressure less than
mended. A second set may be necessary due to the presence of atmospheric may be necessary. Absorbances are measured for
--`,,,,```,`,`,,,,,`,`,`-`-``,```,,,`---
absorbances, An, by matrix inversion (6). The inverted equa-
tions will have the following form:
C 1 5 A 1 F 11 1A 2 F 12 1····1 A n F 1n (5)
C 2 5 A 1 F 21 1A 2 F 22 1····1A n F 2n
C n 5 A 1 F n1 1A 2 F n2 1····1A n F nn
of replicate measurements; the bias represents the degree to where sT is the standard deviation of the transmittance
which the measured quantity approaches the true value. The measurement. The relative standard deviation of the concen-
sources of experimental error limiting bias or precision, or tration is:
both, are broadly classified as determinate or indeterminate
error (1, 14, 15).
sc
c
5S0.434
logT DS D
sT
T
(10)
19.2 Determinate error is systematic error which can be
attributed to definite causes. In quantitative infrared analyses, and the standard deviation of the transmittance is calculated
determinate error may arise from problems such as optical from Eq A6.6 for a series of n measurements of T. sT can be
misalignment, photometric inaccuracy, stray radiant power, determined from the noise in the 100 % line since generally sT
poor spectral resolution, improper sample handling, or devia- will be independent of T.
tions from Beer’s law. Quantitative bias depends upon mini-
mizing determinate error. 20. Keywords
19.3 Indeterminate, or random, error arises from uncontrol- 20.1 infrared spectroscopy; molecular spectroscopy; quan-
lable variables, and limits the precision with which measure- titative analysis
ANNEXES
--`,,,,```,`,`,,,,,`,`,`-`-``,```,,,`---
(Mandatory Information)
A1.1 Obtaining a Good Spectrum for Baseline Procedures by computing the negative logarithm of the transmittance
A1.1.1 There are two ways to get good (FT-IR) spectra. For spectrum. The background can be that of the open beam, or an
the first method, three steps are necessary. 1) Obtain both aperturing device, or an accessory, such as an ATR or gas cell
single-beam background and single-beam sample spectra. 2) or a liquid cell containing solvent used to dissolve the analyte.
Ratio the single-beam sample to the single-beam background NOTE A1.1—Dispersive duel-beam instruments perform the above by
spectrum. This provides a spectrum in transmittance and using the reference beam to obtain a background simultaneously. Hence,
requires conversion to absorbance. 3) Convert to absorbance the reference beam should contain similar beam limiting accessories as
A2.1 All data should be expressed in absorbance as a A2.4 When integrated area is used for quantitative analyses,
function of wavenumber. the reliability of the results frequently depend on the baseline
treatment selected. The accuracy by band area is often im-
A2.2 Band shape changes can cause peak-height data to be proved by limiting the range of absorbances. The wings
nonlinear. Band area, however, may remain essentially unaf-
contribute very little signal while contributing substantial
fected by the changes in shape of the band because band area
uncertainties to the total area. A useful guideline is to limit the
is a function of the total number of absorbing centers in the
integration limits to absorbance values which are no smaller
sample. If the shape change is caused by changes in intermo-
lecular forces, even band area may not be linear. than 20 to 30 % of the peak absorbance.
--`,,,,```,`,`,,,,,`,`,`-`-``,```,,,`---
A3.1 In reference to Fig. A3.1, when no baseline points are where I0 is given by Eq A3.1.
used for an area calculation, the area between lower and upper
wave-number limits is the following: A3.3 If a two-point baseline treatment is used with absor-
bances AW2 at wavenumber w2 and Aw3 at wavenumber w3, as
I 5 A w4 ∆1A w411 ∆1···1A w521 ∆ (A3.1) shown in Fig. A3.3, the formulation is as follows:
where: I 2 5 I 0 2 I 2b (A3.3)
I = integrated absorbance (area),
∆ = sampling interval in wavenumbers,
A = absorbance measured at the designated interval,
I 2b 5 F (S
np
j51
A w2 1
~ A w3 2 A w2 ! ~ w 4 2 w 2 1j∆ !
~ w 3 2 w 2! DG ∆ (A3.4)
I 1 5 I 0 2 ~ A w2 ! ~ w 5 2 w 4 !
I 1 5 I 0 2 @ ~ ∆ w2 ~ np 2 1 !! #
--`,,,,```,`,`,,,,,`,`,`-`-``,```,,,`---
A4.1 Where possible, solution techniques are recom- A4.4.3 Break up the mixture and distribute it over the
mended. However, other methods are discussed. mortar surface by gentle grinding with the pestle. Rub to an
extremely fine powder by vigorous back and forth grinding
A4.2 Liquids—Liquids may be measured either neat or until the mixture becomes caked and takes on a smooth, glossy,
preferably, as solutions in a sealed cell of suitable path length light-reflecting surface.
(15).
A4.4.4 Add one drop of mulling agent (see Note A4.5),
A4.3 Solution Techniques: grind with a rotary motion, using a smooth, hard mortar and
pestle, until all of the mixture is picked up and suspended. The
A4.3.1 For liquids and solids, the following procedures are
suspension should be viscous, not fluid.
recommended but may not be all encompassing:
A4.3.2 Choose a solvent that has minimal absorbance at the NOTE A4.5—Mineral oil, fluorinated hydrocarbon, or appropriate sub-
stances can be used as the mulling agent. Matching the refractive indices
analytical wavenumbers and that will completely dissolve the as closely as possible gives best results.
sample.
A4.4.5 Transfer the sample to a clean, flat salt plate with a
A4.3.3 Measure a specified amount of sample into a volu- clean, rubber policeman or clean, plastic spatula.
metric flask or other suitable container, and add some of the
solvent to be used. The analyte must be completely dissolved. NOTE A4.6—A discussion of salt plates and cell-window materials is
covered in Practice E1252.
A4.3.4 Allow for temperature equilibration to the same A4.4.6 Cover with a second clean, flat plate; squeeze and
temperature as that used for the standards. rotate to obtain the desired thickness and to remove all trapped
A4.3.5 Make up to volume or weight with solvent and mix air.
thoroughly. A4.4.7 Visually observe the scattering of light passing
A4.3.6 Aqueous solutions can be analyzed using flat surface through the sample. As a guideline, the sample may appear
or circular ATR techniques. slightly hazy, but objects on the far side should be distinguish-
able.
NOTE A4.1—For other difficult solutes, such as elastomers and tars, it is
frequently more convenient to roughly weigh the solute in a suitable
container, add solvent from a graduated cylinder, dissolve, and run the
A4.5 Solids By Pressed-Pellet Technique (8, 12):
analysis. The concentration is then obtained by doing a percent solids A4.5.1 Particle size of the sample should be reduced to less
content on an aliquot from the remaining solution. than the analytical wavelength. This is relatively easy for many
NOTE A4.2—Solvent influence or overlap on the absorption bands to be
measured must be recognized and taken into account in the calculation.
samples without incurring any change in the sample. However,
NOTE A4.3—For best results, measure unknown sample solutions at polymorphic changes, degradation, and other changes may
concentrations that will place the analyte absorbances in the range of those occur during grinding. If such changes do happen and are not
used for calibration. controlled, the method will not be valid.
A4.5.1.1 Grinding conditions must be established for the
A4.4 Solids by Mull Techniques (8, 12):
particular sample type since grinding severity can affect
A4.4.1 The following is the recommended procedure for absorption-band intensity.
preparing solids by mull techniques; however, other methods A4.5.1.2 Weigh the preground sample and powdered potas-
also may be appropriate. Average particle size of the sample sium bromide (KBr) or other alkali halide in specified amounts
should be reduced to less than the analytical wavelength. The (sample weight should be about 1 % of the KBr weight; about
particle size reduction is relatively easy for many samples 350 mg KBr is appropriate for a 13-mm disk to avoid
without incurring any change in the sample. However, poly- interference fringes), and mix by hand in a mortar. This should
morphic changes, degradation, and other changes may occur be a mixing step rather than a grinding step.
during grinding. If such changes do occur and are not A4.5.1.3 Place the mixture in an appropriate evacuable die,
controlled, the method will not be valid. If changes are evacuate to at least 15 mm Hg and press at sufficient pressure
suspected, other techniques such as attenuated total reflectance to produce a transparent disk. Follow the manufacturer’s
(12) or diffuse reflectance (11) should be investigated. recommendation as to pressure for a particular die. Other
A4.4.2 Weigh slightly more than the minimum amount of methods, such as minipress cells are also used, but these
sample required, and then weigh the desired amount of an methods may not be as satisfactory due to crazing caused by
appropriate internal standard. Mix thoroughly. trapped air and water vapor.
NOTE A4.4—An appropriate internal standard (7, 11, 12) is a substance NOTE A4.7—Precaution: During pressing operations, place the die
that (a) exhibits a band in a suitable region of the spectrum, and as close symmetrically in the press. Otherwise, the die may forcefully slip out of
to the analyte’s wavenumber as possible, (b) is not present in the sample, the press, causing personnel injury or damage to surrounding equipment.
and (c) does not react chemically with or dissolve in the sample or mulling Some laboratories require safety shields in front of presses.
agent. An alternative to a separate internal standard is to use a band in the NOTE A4.8—Since the purity of alkali halide powders are not all the
sample that does not change as the moiety of interest is varied. This same, the same alkali halide powder should be used for the sample and
approach is very useful in polymer analyses. blank.
--`,,,,```,`,`,,,,,`,`,`-`-``,```,,,`---
--`,,,,```,`,`,,,,,`,`,`-`-``,```,,,`---
the total pressure of the sample, and the total pressure within the cell and
m are readily available and other lengths can also be obtained. the temperature.
A4.6.1.2 Cells can be filled through the use of a vacuum NOTE A4.13—Mixing does not happen automatically. Attention must be
system coupled to a pressure gage for measurement of partial paid to assure complete transfer and equilibrium mixing.
pressures. NOTE A4.14—Calibration mixtures should be cross-checked by an
A4.6.1.3 An alternative method for filling and using long independent method.
path length cells is to have the cell incorporated into a
A5.1 Polymeric materials are studied by a variety of tech- desired value and released after a predetermined time. The
niques; the choice depends on the physical properties of the template and the pressed film are quickly quenched (and dried
particular polymer. Thermoplastics are hot pressed to form gently, if necessary). Film thickness is determined by averag-
films of a desired thickness. Cross-linked elastomers are often ing multiple micrometer measurements of the sample area in
prepared by microtoming, while others, particularly reinforced the infrared beam. These measurements can then be used to
materials, can be cold ground and pressed into a pellet using a calibrate the thickness with standard bands.
suitable sintering agent such as KBr or made into a mull using NOTE A5.1—Occasionally, a sinusoidal pattern is superimposed on a
a suitable suspending liquid such as mineral oil or fluorocar- spectrum. This is referred to as fringing. The fringe pattern is caused by
bons. Another method of forming a thin film is to cast from multiple internal reflections within the sample. This phenomenon can be
solvent. This is especially suitable for resins which either are reduced by placing fine (Grade 6/0) corundum paper between the first
sheet of aluminum foil and the backing plate. The corundum paper causes
not thermoplastic or break down at pressing temperatures. All a matte finish which scatters the reflected light diminishing the multiple
of these techniques can be successfully used for quantitative or internal reflections. This technique must be used cautiously because
semiquantitative analysis. If the above methods are not precision can be lessened by the matte finish (7).
satisfactory, either ATR (12) or diffuse reflectance (11) may be NOTE A5.2—Aluminum foil or other metal in contact with the molten
useful. polymer may need a mold-release agent. Many releasing agents contain
silicone polymers which can interfere with the peaks of interest. Wiping
A5.2 Preparation of polymers by the hot-pressing technique the surface with a solvent is sometimes effective, but use of cover plates
requires a suitable hydraulic press with heated platens capable coated with TFE-fluorocarbon or with a nonmigratory releasing agent is
better.
of reaching the softening point of the polymer. In some cases,
NOTE A5.3—For very thin films, templates are less useful, and
water-cooled platens are also necessary to obtain the desired
release-coated, flat metal plates can be used. Use smaller amounts of
results. The pressing zone for a number of common polymers polymer to achieve thinner films.
is given (18, 19, 20, 21). For quantitative results, the use of a NOTE A5.4—For quantitative determination of blended components, it
template with an opening to accommodate the infrared beam is is advisable to increase homogeneity by pressing a large sample, about 5
recommended. Typically templates from 0.3 to 6 mil (7.5 to g, into a film. Cut this film and stack the pieces so outer edges will lie on
150 µm) are easily cut from metal shimstock. Sufficient center portions, and then repress. Select smaller samples from this blend
powdered or pellets of polymer to fill the opening is placed in for final pressing.
the template between two layers of aluminum foil which are NOTE A5.5—Pressing techniques must be used with caution when
investigating crystalline polymers, or polymers with other order-
then placed between two larger smooth backing plates. A small producing structures. These polymers are temperature sensitive. Some
amount of pressure is applied while the polymer heats to the problems can be avoided by annealing under controlled conditions before
desired temperature, then pressure is increased rapidly to the quantitative measurements are made.
A6.1 The Gaussian Distribution and Confidence Intervals: visualize the utility of the standard deviation, the following
A6.1.1 Often random error has a normal (Gaussian) distri- substitution can be made:
bution (16,17). This is generally true when there is no single ~x 2 µ!
y5 (A6.3)
dominant source of error. We can define a population as a σ
hypothetical set of N observations from which the sample of ~ 2y 2 !
observations actually obtained are taken. In a normally distrib- p ~ y ! dy 5 ~ 2π ! exp dy (A6.4)
2
uted population, the probability that the measured quantity has
a value between x and x + dx is given by p(x) dx as follows: The area under the distribution curve represents the prob-
ability that a measured value will be within a particular range,
p ~ x ! dx 5
~ 2π !
σ
21/2
exp F 2~x 2 µ !
2σ 2
2
G dx (A6.1) and the confidence interval placed upon a measured value can
be given in terms of the standard deviation. For example, if the
where µ is the true value of the measured quantity and σ is error is normally distributed, 68.26 % of the time the value of
the true standard deviation of the measurements. σ is defined: the single measurement will lie in the range between −σ and σ;
F G 95.44 % of the time the measurement will lie between −2σ and
n 1/2
( ~x
i51
i 2 µ !2 2σ; and 99.74 % of the time the measurement will lie between
σ5 (A6.2) −3σ and 3σ.
N
where N is the number of measurements. The standard A6.2 Gaussian Distribution of a Sample:
deviation, σ, has the same units as the quantity being measured
A6.2.1 In practice it is necessary to deal with a limited
and expresses the degree of scatter among the measured
number of measurements, n (14, 24, 25) called a sample, which
quantities and is related to the precision of the measurements.
The quantity σ2 is called the variance and has additive are obtained from the population. From these n observations,
properties. The overall variance of a multistep process is the arithmetic mean, x̄, is taken as an estimate of the true
simply the sum of the variances of the individual steps if there population mean, µ, where:
n
is no correlation between the sources of variation of the
individual steps. The square root of the overall variance yields (x
i51
i
X̄ 5 (A6.5)
the standard deviation for the whole process. To more fully n
--`,,,,```,`,`,,,,,`,`,`-`-``,```,,,`---
F( G
n 1/2
2 fications are necessary for the second case as follows:
~ x i 2 x̄ !
s5
i51
n21
(A6.6)
t5
~ x̄ 1 2 x̄ 2 !
s S n 1n 2
n 1 1n 2 D 1/2
(A6.9)
A7.1 Detailed descriptions of multilinear regression (also analysis, and analyses using derivative techniques. Because of
called inverse least-squares), principal components regression the many possible variations of these methods and their relative
and partial least squares are discussed in Practices E1655. complexity, only the general nature of these methods is
Other methods of multivariate analyses are briefly discussed presented.
below as classical least squares, cross-correlation, factor
A8.1 When the quantitative equations relating A and c are A 5 KC1E (A8.1)
written in the form of Beer’s law (see Eq 4), then the where the A matrix is an i × m matrix of m calibration
least-squares analysis is a classical regression (often referred to spectra at i wavenumbers, and C is an n × m matrix of the n
in infrared spectroscopic literature as the K-matrix method) known component concentrations for each of the m mixtures. K
(28-35). Either univariate methods (that is, each reference is the i × n matrix of the n pure component spectra at unit
spectrum contains only spectral features of one component in concentration and unit relative path length for each of the i
the spectral region of interest (28, 29, 31) or multivariate wavenumbers. E is the i × m matrix of random noise or error
methods (that is, the calibration standards are mixtures of the in the absorbance values of the calibration spectra. K is
pure components) (30) may be used. The univariate method is estimated by least-squares methods as follows:
simply a special case of the multivariate technique. When the
multivariate methods are employed, the least-squares analysis K̂ 5 AC' ~ CC' ! 21 (A8.2)
is a two-step analysis involving both a calibration step and an where K̂is the least-squares estimate of K, the primes
analysis step. The calibration step takes the m known calibra- indicate transposed matrices and the superscript −1 indicates the
tion mixture spectra and estimates the pure component spectra inverse of the matrix. The estimated K matrix, K̂, is then used
by least-squares methods. The calibration equations are as in the least-squares prediction of the unknown sample concen-
follows: tration as follows:
--`,,,,```,`,`,,,,,`,`,`-`-``,```,,,`---
Ĉ 5 ~ K̂K̂' !
21
K̂' A (A8.3) independent of the number of wavenumbers used in the
where Ĉis the least-squares estimate of the unknown sample analysis, the complexity of matrix inversion is also indepen-
concentrations and A is now the unknown sample spectrum at dent of the number of wavenumbers used. This favors the use
i wavenumbers. In the univariate case, the K matrix is simply of a large number of wavenumbers in the analysis which
constructed directly from the measured pure component spec- improves bias and precision and makes analyses possible even
tra. Baseline fitting of the sample spectrum can also be when the signal-to-noise ratios are less than one (28). Note that
included in the solution of Ĉ in Eq A8.3 to allow for baseline the use of a larger number of wavenumbers also increases the
variations between the calibration and sample spectra (28, 29). chance that the analysis will be compromised by an unknown
A weighted least-squares analysis may also be used to give component. The effect of an unknown component can be
greater emphasis to those spectral data with greater signal-to- reduced if the algorithm is written to minimize the influence of
noise ratios (29, 30). In Eq A8.2 and A8.3, the final number of those regions of the spectra that contain the unknown interfer-
simultaneous equations to be solved can be reduced to the ences (29, 30). The minimum number of calibration standards
number of components used in the analysis. This means that and wavenumbers necessary for the analysis is equal to the
the dimension of the matrix which must be inverted is also the number of components.
same as the number of components. Since this dimension is
A9.1 The use of cross-correlation methods in infrared spec- mine quantitatively the amount of one contained in the other. If
troscopy has recently been described (20, 32, 33). Cross- Beer’s law is valid, then the value of the correlation function at
correlation techniques can evaluate the similarity between two zero displacement is a linear function of concentration. The
spectra (for example, the reference and the sample) to deter- mathematical details are given (20, 32, 33).
A10.1 Abstract factor methods utilize the concept of repre- Analysis of an unknown sample is then accomplished by
senting spectral data as a linear data set. Instead of using determining the scores of the principal factors for the sample
discrete data points representing changes in component spectrum. The relation between scores and concentrations
concentration, factor analysis uses the entire spectrum of each determined from the sample data can then be used to yield the
standard and transforms the data into a linear representation of sample concentrations. The linear regression analysis may be
quantitative information. Factor abstract methods involve find- carried out using any of a number of factor spaces (or basis
ing a set of principal factors (for example, eigenvectors or sets) including Fourier and Taylor series or more commonly,
loadings) required to reproduce the calibration spectra. The principal components analysis (PCA or PCR) and partial least
relation between the scores of these principal factors for each squares (PLS). Theories and application of factor analysis
calibration spectrum and the desired properties is determined methods are presented in the literature (9, 10, 21, 36) and
by the application of standard linear regression methods. Practices E1655.
--`,,,,```,`,`,,,,,`,`,`-`-``,```,,,`---
A11.1 Derivative methods are useful when the absorption tude of the negative-derivative lobe, or the difference between
band or bands being analyzed do not yield a suitable choice for the two lobes. The ratio of any of the above quantities between
a baseline as described in Sections 12 and 13. Derivative corresponding sample and reference bands is assumed to be
techniques have generally required that isolated spectral bands proportional to the ratio of sample and reference concentra-
be used in the analysis although one least-squares full- tions.
spectrum derivative method has been described which allows A11.1.2 Use of Second Derivatives—The second-derivative
for overlapping peaks (31). The derivative of a band or spectrum may be calculated from the first-derivative spectrum
spectrum can often be calculated by taking the difference in the same manner that the first-derivative spectrum was
between absorbances in the spectrum at equally spaced wave- calculated from the absorption spectrum. Generally the second-
numbers. However, because derivatives increase spectral noise, derivative spectrum has a positive lobe, a negative lobe, and a
mathematical smoothing (for example, Savitzky-Golay second positive lobe. The magnitude of any of these or their
smoothing) is often used either before or after derivatives are differences can be used in the quantitative analysis.
calculated.
--`,,,,```,`,`,,,,,`,`,`-`-``,```,,,`---
NOTE A11.1—Depending upon the specific algorithm used, the first
A11.1.1 Use of First Derivatives—There are three possible derivative of the first derivative may not be exactly identical to the second
methods used in first-derivative analyses. These include using derivative. This is not normally a problem, as long as a consistent
the magnitude of the positive first-derivative lobe, the magni- approach is used.
where:
T0 = transmittance with a very thick or concentrated sample
in the beam (or with the beam blocked),
T1 = transmittance obtained with the sample in the beam,
and
T2 = transmittance without the sample in the beam.
If T0 = 0, then Eq A12.1 reduces to the following: FIG. A12.1 Transmittance Zero Determination
A 5 log~ T 2 /T 1 ! (A12.2)
A13.1 Record the interval of the spectrum containing the A13.3.5 Convert these transmittance values to absorbance
band(s) to be measured. This interval should normally include and determine the analyte absorbance.
at least one maximum and one minimum transmittance.
A13.4 For computerized data, do the following:
A13.2 For each of the analytical wavenumbers that are A13.4.1 Record the spectrum of the analyte sample.
specified in the method make certain all peaks to be measured
are in the desired absorbance range. A13.4.2 Record the spectrum of a blank if possible.
A13.3 For instruments without a computer, do the following A13.4.3 Subtract the blank from the sample spectrum scal-
at the selected wavenumbers: ing appropriately.
A13.3.1 Measure the zero transmittance with a thick or A13.4.4 Using a preselected baseline treatment, determine
concentrated sample (see Annex A12). the analyte absorbance.
A13.3.2 Read the transmittance value T1 for the sample (see NOTE A13.1—For dispersive instruments, a reference beam attenuator
Fig. A12.1). is useful for shifting the apparent zero absorbance for highly absorbing
samples. For optical null instruments, changing the position of the
A13.3.3 If possible, prepare a blank that is free of analyte, attenuator will require readjustment of the gain for maximum sensitivity.
then measure T2. NOTE A13.2—Most computerized instruments have software for quan-
titative analysis which give step-by-step assistance in both calibration and
A13.3.4 Using a preselected baseline treatment, determine analysis. The general procedure is, however, as outlined in A13.1 –
--`,,,,```,`,`,,,,,`,`,`-`-``,```,,,`---
the analyte and baseline transmittances (see Note 3). A13.4.4.
REFERENCES
(1) Ingle, J.D. and Crouch, S.R., Spectrochemical Analysis, Prentice Hall, (18) Henniker, J.C., Infrared Spectrometry of Industrial Polymers, Aca-
Englewood Cliffs, NJ, 1988. demic Press, 1967.
(2) Ramsay, D.A., Journal of the American Chemical Society, Vol 74, (19) Brandrup, J., and Immergut, E.H., Polymer Handbook, 2nd edition,
1952, p. 72. John Wiley and Sons, New York, 1975, p. III-51.
(3) Anderson, R.J., and Griffiths, P.R., Analytical Chemistry, Vol 47, (20) Mann, C. K., Goleniewski, J.R., and Sismanidis, C.A., Applied
1975, p. 2339. Spectroscopy, Vol 36, 1982, p. 223.
(4) Zhu, C., and Griffiths, P.R., Applied Spectroscopy, Vol 47, 1988, pp. (21) Malinowski, E.R., and Howery, D.G., Factor Analysis in Chemistry,
1403 and 1409. John Wiley and Sons, Inc., New York, 1980.
(5) Hieftje, G.M., Dehonigs, D., Hirschfeld, T., Applied Spectroscopy, Vol (22) Seisler, H.W., and Holland-Moritz, K., Infrared and Raman Spec-
39, No. 2, p. 253. troscopy of Polymers, Vol 4, Marcel Deker, Inc., New York, 1980, p.
(6) Bauman, R.P., Absorption Spectroscopy, John Wiley and Sons, Inc., 116.
New York, 1962. (23) An Infrared Atlas for the Coatings Industry , Federation of Societies
for Coatings Technology, Philadelphia, 1980, p. 17.
(7) Potts, W.J., Jr., Chemical Infrared Spectroscopy: Volume 1,
(24) Meyer, S.L., Data Analysis for Scientists and Engineers, John Wiley
Techniques, John Wiley and Sons, Inc., New York, 1963.
and Sons, New York, 1975.
(8) Conley, R.T., Infrared Spectroscopy, 2nd edition, Allyn and Bacon,
(25) Fisher, R.A., The Design of Experiments, Hafner, New York, 1971.
Inc., Boston, 1972.
(26) Selby, S.M., ed., Standard Mathematical Tables, 15th edition,
(9) Kramer, R., Chemometric Techniques for Quantitative Analysis,
Chemical Rubber Co., Cleveland, 1967.
Marcel Decker, New York, NY, 1998.
(27) Fisher, R.A., and Yates, F., Statistical Tables, Hafner, Darien, CT,
(10) Beebe, K.R., Pell, R.J., Seasholtz, M.B., Chemometrics, A Practical 1970.
Guide, John Wiley and Sons, Inc., New York, NY, 1998. (28) Haaland, D.M., and Easterling, R.G., Applied Spectroscopy, Vol 34,
(11) Griffiths, P.R., and deHaseth, J.A., Fourier Transform Infrared 1980, p. 539.
Spectroscopy, John Wiley and Sons, Inc., New York, 1986. (29) Haaland, D.M., and Easterling, R.G., Applied Spectroscopy, Vol 36,
(12) Smith, A.L., Applied Infrared Spectroscopy, John Wiley and Sons, 1982, p. 665.
Inc., New York, 1979. (30) Haaland, D.M., Easterling, R.G., and Vopicka, D.A., Applied
(13) Cole, K.C., Thomas, Y., Pellerin, E., Dunoulin, M.M., and Paroli, Spectroscopy, Vol 39, 1985, p. 73.
R.M., Applied Spectroscopy , Vol 50, No. 6, 1996, pp. 774-780. (31) Kisner, H.J., Brown, C.W., Kavarnos, G.J., Analytical Chemistry,
(14) Skoog, D. A., and West, D.M., Principles of Instrumental Analysis, Vol 54, 1982, p. 1479.
2nd edition, Saunders, Philadelphia, 1980. (32) Tyson, L.L., Ling, Y.C., and Mann, C.K., Applied Spectroscopy, Vol
(15) Rothman, L.D., Crouch, S.R., and Ingle, J.D., Analytical Chemistry, 38, 1984, p. 663.
Vol 47, 1975. (33) Tyson, L.L., Vickers, T.J., and Mann, C.K., Applied Spectroscopy,
(16) Infrared Application Study, No. 10, Perkin-Elmer Corp. Vol 38, 1984, p. 697.
(17) Ferraro, J.R., and Basile, L.J., Fourier Transform Infrared (34) Antoon, M.K., Keonig, J.H., and Koenig, J.L., Applied Spectroscopy,
Spectroscopy, Vol 2, Academic Press, New York, Chapter 2, 1979. Vol 31, 1977, p. 518.
ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.
This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.
This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or service@astm.org (e-mail); or through the ASTM website
(www.astm.org). Permission rights to photocopy the standard may also be secured from the Copyright Clearance Center, 222
Rosewood Drive, Danvers, MA 01923, Tel: (978) 646-2600; http://www.copyright.com/
--`,,,,```,`,`,,,,,`,`,`-`-``,```,,,`---