What Is Measurement Risk
What Is Measurement Risk
What Is Measurement Risk
Henry A. Zumbrun
Morehouse Instrument Company, Inc.
Apr • May • Jun 2017 27 Cal Lab: The International Journal of Metrology
What Is Measurement Risk?
Henry A. Zumbrun
So, what does this mean for a metrology laboratory? Uncertainty are as follows:
It means that any lab making a statement of compliance,
calling an instrument “in tolerance,” must consider Tolerance
___________________
TUR =
measurement uncertainty and properly calculate T.U.R. Expanded Uncertainty
considering the location of the measurement. In simplistic
(USL − LSL)
terms, T.U.R. = Tolerance Required/Uncertainty of the TUR = ___________
4 * u
Measurement (at a 95% confidence interval). If the
Uncertainty of the Measurement is not less than the where:
tolerance required, there will be a significant risk of false TUR = Test Uncertainty Ratio,
accept. In simplistic terms, a TUR that produces less than USL = Upper Specification Limit,
+/- 2 % upper and lower risk would be required to ensure LSL = Lower Specification Limit, and
the measurement is valid. u = standard uncertainty.
Keys to lowering measurement risk include having Note: We are using 4 assuming k=2, the proper formula
your calibration provider replicate how the instrument is would be 2 times the actual k value is for a 95 % confidence
used in the field, having competent technicians, using the interval.
right equipment, and lowering overall uncertainties by
the calibration provider. There is quite a bit of difference
between force measurement labs with CMCs of 0.1 percent,
The Calculation of TUR for Tolerances
0.05 percent, 0.02 percent, 0.01 percent, 0.005 percent and
0.002 percent of applied force. Not using the laboratory • ((Upper Specification Limit - Lower Specification
with the right capability to meet your requirements is like Limit))/(4 * Standard Uncertainty)
using a ruler to calibrate a gauge block. • Combined Uncertainty (u) – The square root of
Table 1 shows the Test Uncertainty Ratios (TUR) that the sum of the squares of all the input quantity
force calibration labs with different calibration capabilities uncertainty components.
can provide for various levels of required tolerances. The __________________________
√ ( ) ( ) ( )
far-left column represents the calibration standard required CMC 2 ______ Res 2 ____ Rep 2
u = ______
+
3.464 + 1
for force measurements. Deadweight primary standards k
are often required to achieve CMCs of better than 0.01
• CMC = Calibration and Measurement Capability.
% of applied force. A high-end load cell calibrated by
This should be found on the calibration report.
deadweights would be required to achieve CMCs of better
than 0.05 %. This table indicates the best TUR that the labs • Res = This is the resolution of the Unit Under Test
can provide for the same load cell at similar conditions. Per (UUT) The divisor for resolution will either be
this table, only calibration labs with CMCs around 0.02 % 3.464 or 1.732 (depending on how the UUT least
or better can calibrate devices with a tolerance of 0.1 %. significant digit resolves).
They may still need to adjust the device to read closer to • Rep = Repeatability of the Unit Under Test (UUT).
the nominal value. We will discuss guard banding later. Repeatability of UUT must be used if repeatability
The table was derived from TUR and uncertainty studies were not previously accounted for in the
formulas found in JCGM 100:2008 and ANSI/NCSLI CMC. If accounted for in the CMC, this would
Z540.3-2006. The formulas used to determine TUR and not be required.
Cal Lab: The International Journal of Metrology 28 Apr • May • Jun 2017
What Is Measurement Risk?
Henry A. Zumbrun
Apr • May • Jun 2017 29 Cal Lab: The International Journal of Metrology
What Is Measurement Risk?
Henry A. Zumbrun
Figure 4.
How to Lower the Risk (PFA) By Lowering This road is not more difficult; it’s just different from the
the Uncertainty current way you may be doing things. Choosing to consider
the impact of not doing things right—and making the
decision to select the best calibration provider—will make
1. Use better equipment with a lower resolution and/ all the difference. The rest is just putting formulas in place
or better repeatability; e.g., higher quality load cell to report and know your measurement risk.
for force measurement.
2. Use a better calibration provider with a Calibration References
and Measurement Capability (CMC) low enough
to reduce the measurement risk.
[1] JCGM 100:2008 Evaluation of measurement data — Guide
3. Pay attention to the uncertainty values listed in
to the expression of uncertainty in measurement, BIPM.
the calibration report issued by your calibration
provider. Make sure to get proper T.U.R. values [2] ILAC P14:01/2013 Policy For Uncertainty in Calibration.
for every measurement point (but pay attention
to the location of the measurement). [3] ANSI/NCSLI Z540.3-2006 Requirements for the Calibra-
The last graph (Figure 4) shows the same test instrument tion of Measuring and Test Equipment.
with a lower Standard Uncertainty. This was a real scenario [4] ISO/IEC 17025:2005: General requirements for the compe-
where an instrument was modified from a 10 lbf resolution
tence of testing and calibration laboratories.
to a 2 lbf resolution. The total risk is now 0 and the device
will be “in tolerance” with less than 2 % total risk from [5] AS9100C Quality Management Systems - Requirements
reading of 9,996 through 10,006 lbf. There are several
for Aviation, Space, and Defense Organizations.
acceptable methods for applying a guard band to obtain
what the measured value needs to be in order to maintain
less than 2 % total risk.
These graphs comply with Method 5: Guard Bands Henry A. Zumbrun (hzumbrun@morehouse.com),
Based on Expanded Uncertainty in the ANSI/NCSLI Z540.3 Morehouse Instrument Company, Inc., York, Pennsylvania.
Handbook and is described in ISO 14253-1, and included in
ILAC G8, and various other guidance documents.
After reading this paper, you may be standing at a
crossroads and wondering if any of this extra work is
necessary. To the left is the same rough path you’ve
been travelling all along. This is the path that says, “If
it’s not broken, why fix it?” You might be thinking that
measurement risk has not been an issue before, or you’ll
just wait until an auditor questions you about it (or there is
a train wreck). Yet, to the right is the road that fewer people
realize will help solve their measurement problems today.
Cal Lab: The International Journal of Metrology 30 Apr • May • Jun 2017