Clsi Ila23 A
Clsi Ila23 A
Clsi Ila23 A
Vol. 24 No. 16
Replaces I/LA23-P
Vol. 23 No. 24
This guideline addresses components for harmonizing and assessing the quality of
immunoassay systems for several commonly used dose-response indicator categories,
e.g., radioisotopes, enzymes, fluorescence, luminescence, reagents, and experimental
components criteria essential to characterizing an immunoassay.
A guideline for global application developed through the NCCLS consensus process.
NCCLS...
Serving the World’s Medical Science Community Through Voluntary Consensus
NCCLS is an international, interdisciplinary, nonprofit, the need for field evaluation or data collection, documents
standards-developing, and educational organization that may also be made available for review at an intermediate
promotes the development and use of voluntary (i.e., “tentative”) consensus level.
consensus standards and guidelines within the healthcare
Proposed An NCCLS consensus document undergoes the
community. It is recognized worldwide for the
first stage of review by the healthcare community as a
application of its unique consensus process in the
proposed standard or guideline. The document should
development of standards and guidelines for patient
receive a wide and thorough technical review, including an
testing and related healthcare issues. NCCLS is based on
overall review of its scope, approach, and utility, and a line-
the principle that consensus is an effective and cost-
by-line review of its technical and editorial content.
effective way to improve patient testing and healthcare
services. Tentative A tentative standard or guideline is made
available for review and comment only when a
In addition to developing and promoting the use of
recommended method has a well-defined need for a field
voluntary consensus standards and guidelines, NCCLS
evaluation or when a recommended protocol requires that
provides an open and unbiased forum to address critical
specific data be collected. It should be reviewed to ensure its
issues affecting the quality of patient testing and health
utility.
care.
Approved An approved standard or guideline has achieved
PUBLICATIONS
consensus within the healthcare community. It should be
An NCCLS document is published as a standard, reviewed to assess the utility of the final document, to
guideline, or committee report. ensure attainment of consensus (i.e., that comments on
earlier versions have been satisfactorily addressed), and to
Standard A document developed through the consensus
identify the need for additional consensus documents.
process that clearly identifies specific, essential
requirements for materials, methods, or practices for use NCCLS standards and guidelines represent a consensus
in an unmodified form. A standard may, in addition, opinion on good practices and reflect the substantial
contain discretionary elements, which are clearly agreement by materially affected, competent, and interested
identified. parties obtained by following NCCLS’s established
consensus procedures. Provisions in NCCLS standards and
Guideline A document developed through the
guidelines may be more or less stringent than applicable
consensus process describing criteria for a general
regulations. Consequently, conformance to this voluntary
operating practice, procedure, or material for voluntary
consensus document does not relieve the user of
use. A guideline may be used as written or modified by
responsibility for compliance with applicable regulations.
the user to fit specific needs.
COMMENTS
Report A document that has not been subjected to
consensus review and is released by the Board of The comments of users are essential to the consensus
Directors. process. Anyone may submit a comment, and all comments
are addressed, according to the consensus process, by the
CONSENSUS PROCESS
NCCLS committee that wrote the document. All comments,
The NCCLS voluntary consensus process is a protocol including those that result in a change to the document when
establishing formal criteria for: published at the next consensus level and those that do not
result in a change, are responded to by the committee in an
• the authorization of a project appendix to the document. Readers are strongly encouraged
• the development and open review of documents to comment in any form and at any time on any NCCLS
document. Address comments to the NCCLS Executive
• the revision of documents in response to comments Offices, 940 West Valley Road, Suite 1400, Wayne, PA
by users 19087, USA.
• the acceptance of a document as a consensus VOLUNTEER PARTICIPATION
standard or guideline.
Healthcare professionals in all specialties are urged to
Most NCCLS documents are subject to two levels of volunteer for participation in NCCLS projects. Please
consensus—“proposed” and “approved.” Depending on contact the NCCLS Executive Offices for additional
information on committee participation.
I/LA23-A
ISBN 1-56238-533-X
Volume 24 Number 16 ISSN 0273-3099
Assessing the Quality of Immunoassay Systems: Radioimmunoassays
and Enzyme, Fluorescence, and Luminescence Immunoassays;
Approved Guideline
W. Harry Hannon, Ph.D.
Mark A. Atkinson, Ph.D.
Dorothy J. Ball, Ph.D.
Robin G. Lorenz, M.D., Ph.D.
Per N.J. Matsson, Ph.D.
Deborah M. Moore, M.T.(ASCP)
Ronald J. Whitley, Ph.D.
Abstract
NCCLS document I/LA23-A—Assessing the Quality of Immunoassay Systems: Radioimmunoassays and Enzyme, Fluorescence,
and Luminescence Immunoassays; Approved Guideline addresses components for harmonizing and assessing the quality of
immunoassay systems for several commonly used dose-response indicator categories, (e.g., radioisotopes, enzymes, fluorescence,
luminescence, reagents, and experimental components criteria) essential to characterizing an immunoassay.
The Area Committee on Immunology and Ligand Assays merged NCCLS documents LA1-A2—Assessing the Quality of
Radioimmunoassay Systems; Approved Guideline—Second Edition and DI4-T—Enzyme and Fluorescence Immunoassays;
Tentative Guideline into one document assimilating the residual segments of LA1-A2, and updating information in DI4-T into a
more generic model, along with the addition of new information for each topic. I/LA23-A has broader utility and applicability
while providing resource information previously available in the other two documents.
This new guideline describes the iterations in the development, performance characterization, and certification from sample
collection to method transferability. Specific nuances of each of the different dose-response systems for immunoassays are
addressed while placing emphasis on mechanisms to assess the quality of the different immunoassay systems—factors that
contribute to reliable and reproducible results. This guideline is particularly useful for specific details on optimization and
harmonization of immunoassays, especially for those measurands (analytes) that are measured only by quantitation of antigen-
antibody reactions (e.g., protein hormones, IgG, serum specific proteins).
NCCLS. Assessing the Quality of Immunoassay Systems: Radioimmunoassays and Enzyme, Fluorescence, and Luminescence
Immunoassays; Approved Guideline. NCCLS document I/LA23-A (ISBN 1-56238-533-X). NCCLS, 940 West Valley Road,
Suite 1400, Wayne, Pennsylvania 19087-1898 USA, 2004.
THE NCCLS consensus process, which is the mechanism for moving a document through two or more levels of review by the
healthcare community, is an ongoing process. Users should expect revised editions of any given document. Because rapid
changes in technology may affect the procedures, methods, and protocols in a standard or guideline, users should replace
outdated editions with the current editions of NCCLS documents. Current editions are listed in the NCCLS Catalog, which is
distributed to member organizations, and to nonmembers on request. If your organization is not a member and would like to
become one, and to request a copy of the NCCLS Catalog, contact the NCCLS Executive Offices. Telephone: 610.688.0100;
Fax: 610.688.0700; E-Mail: exoffice@nccls.org; Website: www.nccls.org
Number 16 NCCLS
This publication is protected by copyright. No part of it may be reproduced, stored in a retrieval system,
transmitted, or made available in any form or by any means (electronic, mechanical, photocopying,
recording, or otherwise) without prior written permission from NCCLS, except as stated below.
NCCLS hereby grants permission to reproduce limited portions of this publication for use in laboratory
procedure manuals at a single site, for interlibrary loan, or for use in educational programs provided that
multiple copies of such reproduction shall include the following notice, be distributed without charge,
and, in no event, contain more than 20% of the document’s text.
Permission to reproduce or otherwise use the text of this document to an extent that exceeds the
exemptions granted here or under the Copyright Law must be obtained from NCCLS by written request.
To request such permission, address inquiries to the Executive Director, NCCLS, 940 West Valley Road,
Suite 1400, Wayne, Pennsylvania 19087-1898, USA.
Suggested Citation
Proposed Guideline
September 2003
Approved Guideline
May 2004
ISBN 1-56238-533-X
ISSN 0273-3099
ii
Volume 24 I/LA23-A
Committee Membership
Acknowledgement
The Area Committee on Immunology and Ligand Assays would like to extend its appreciation to Jeffrey
Warren, M.D. (University of Michigan) and Marvin Goldsmith (Goldsmith Graphics), past committee
members, for their contributions. We also recognize the contributions of Peter E. Maxim, Ph.D.
(deceased), a past committee member representing the Food and Drug Administration. These individuals
contributed substantially to the development of this important NCCLS protocol.
iii
Number 16 NCCLS
iv
Volume 24 I/LA23-A
Contents
Abstract ....................................................................................................................................................i
1 Scope..........................................................................................................................................1
2 Introduction................................................................................................................................1
3 Standard Precautions..................................................................................................................2
4 Terminology...............................................................................................................................2
4.1 Definitions ....................................................................................................................2
4.2 Acronyms and Abbreviations .......................................................................................6
5 Primary Components for Immunoassays ...................................................................................7
5.1 Antibody .......................................................................................................................7
5.2 Antigen .......................................................................................................................10
5.3 Separation Procedures.................................................................................................11
6 Sample Collection, Handling, and Storage ..............................................................................13
6.1 Sample Collection.......................................................................................................13
6.2 Sample Requirements .................................................................................................13
6.3 Sample Handling.........................................................................................................13
6.4 Sample Storage ...........................................................................................................14
7 Radioimmunoassays ................................................................................................................14
7.1 Radioisotopic Systems ................................................................................................14
7.2 Reagents......................................................................................................................14
7.3 Assay Description .......................................................................................................15
7.4 Detection and Quantitation .........................................................................................16
7.5 Limitations and Precautions........................................................................................16
7.6 Radiolabeled Waste Products .....................................................................................17
8 Enzyme Immunoassays............................................................................................................17
8.1 Enzyme Systems .........................................................................................................17
8.2 Reagents......................................................................................................................18
8.3 Assay Description .......................................................................................................18
8.4 Detection and Quantitation .........................................................................................18
8.5 Limitations and Precautions........................................................................................19
9 Fluorescent Immunoassays ......................................................................................................19
9.1 Fluorescent Systems ...................................................................................................19
9.2 Reagents......................................................................................................................20
9.3 Assay Description .......................................................................................................20
9.4 Detection and Quantitation .........................................................................................21
9.5 Limitations and Precautions........................................................................................21
v
Number 16 NCCLS
Contents (Continued)
10 Luminescent Immunoassays ....................................................................................................21
10.1 Luminescent Systems .................................................................................................21
10.2 Reagents......................................................................................................................22
10.3 Assay Description .......................................................................................................22
10.4 Detection and Quantification ......................................................................................22
10.5 Limitations and Precautions........................................................................................23
11 Assay Performance Characteristics..........................................................................................23
11.1 Accuracy, Trueness, and Precision .............................................................................23
11.2 Sensitivity and Specificity ..........................................................................................24
11.3 Comparison of Quantitative Tests ..............................................................................24
11.4 Reference Intervals .....................................................................................................25
11.5 Reference Values (Expected Values)..........................................................................25
11.6 Results in Test Comparisons.......................................................................................26
12 Quality Assurance and Control................................................................................................26
12.1 General........................................................................................................................26
12.2 Quality Control Enhancement Parameters..................................................................26
12.3 Technical Considerations............................................................................................27
13 Necessary Product Insert Information for Immunoassay Systems ..........................................27
References.............................................................................................................................................29
Additional References...........................................................................................................................30
vi
Volume 24 I/LA23-A
Foreword
A comprehensive coverage of the field of immunoassays is too large for the scope of this document. The
area committee, during development of this guideline, focused on the core quality management issues.
For detailed information, other publications cited in the general references should be consulted. I/LA23-A
replaces NCCLS documents LA1-A2—Assessing the Quality of Radioimmunoassay Systems; Approved
Guideline—Second Edition and D14-T—Enzyme and Fluorescence Immunoassays; Tentative Guideline.
A Note on Terminology
In keeping with NCCLS’s commitment to align terminology with that of ISO, the following describes the
metrological terms and their uses in I/LA23-A:
The term accuracy refers to the “closeness of the agreement between the result of a (single) measurement
and a true value of a measurand” and comprises both random and systematic effects. Trueness is used in
this document when referring to the “closeness of the agreement between the average value from a large
series of measurements and to a true value of a measurand”; the measurement of trueness is usually
expressed in terms of bias. Precision is defined as the “closeness of agreement between independent
test/measurement results obtained under stipulated conditions.” As such, it cannot have a numerical value,
but may be determined qualitatively as high, medium, or low. For its numerical expression, the term
imprecision is used, which is the “dispersion of results of measurements obtained under specified
conditions.” In addition, different components of precision are defined in ILA23-A, primarily
repeatability, i.e., “the closeness of the agreement between results of successive measurements of the
same measurand carried out under the same conditions of measurement”; while reproducibility describes
the “closeness of agreement of results of measurements under changed conditions.”
The NCCLS Harmonization Policy recognizes ISO terms as the preferred terms. When appropriate,
alternative terms are included parenthetically to help avoid confusion.
vii
Number 16 NCCLS
The term measurand (a particular quantity subject to measurement) is used in combination with the term
analyte (component represented in the name of a measurable quantity) when its use relates to a biological
fluid/matrix; and the term measuring range in combination with reportable range when referring to “a set
of values of measurands for which the error of a measuring instrument (test) is intended to lie within
specified limits.”
The term diagnostic sensitivity is combined with the term clinical sensitivity, and correspondingly the
term diagnostic specificity is combined with the term clinical specificity, because in Europe, the term
“clinical” often refers to clinical studies of drugs under stringent conditions.
Users of I/LA23-A should understand, however, that the fundamental meanings of the terms are identical
in many cases, and to facilitate understanding, terms are defined in the Definitions section of this
guideline.
All terms and definitions will be reviewed again for consistency with international use, and revised
appropriately during the next scheduled revision of this document.
Key Words
viii
Volume 24 I/LA23-A
1 Scope
This document presents guidelines for immunoassays of macromolecular analytes. The factors likely to be
important in achieving reliable and reproducible results are emphasized. Use of this document should
promote greater reliability and comparability in immunoassay results.
The definitions, information, and procedures necessary to properly assess the quality of immunoassay
systems are described. Awareness of the evaluation process allows laboratory personnel to better assess
systems that meet the specific needs of the patient population.
Immunoassays are widely used to quantitate specific measurands (analytes) in complex mixtures such as
clinical samples. Immunoassays using enzymes or fluorescers as labels are recent developments. Enzyme
immunoassays (EIA), fluorescence immunoassays (FIA), and luminescence immunoassays (LIA) were
developed to provide a simple, sensitive immunoassay technique that does not use unstable and
potentially dangerous radioisotopes. At present, enzyme, fluorescence, and luminescence immunoassays
are typically less sensitive than radioimmunoassays (RIA). However, high sensitivity is not necessary in
many applications, and there are reasons to expect that sensitivity comparable to radioimmunoassay can
and will be achieved by EIA and FIA in the near future. There are no criteria on whether RIA, EIA, FIA,
or LIA is the best method for a particular analyte measurement. When radioisotopes cannot be used or
when radioisotope decay counters are not available, techniques such as EIA, FIA, or LIA are obligatory.
In practice, EIA, FIA, and LIA systems have exhibited other advantages, including high specific activity,
reagent stability, and applicability to simple instrumentation. Immunoassays using luminescent
technologies are now among the most sensitive, with analytical detection limits as low as one zeptomole
(10-21 moles).
2 Introduction
Immunoassays have become essential tools for the analytic operation of clinical diagnostic and research
laboratories. Numerous advances in immunoassay techniques continue to drive new technologies,
especially for application to research in proteomics and the human genome: highly sensitive dose-
response indicators, methods for reduction in nonspecific binding and background signal, simultaneous
analyte measurements, improved automation, and miniaturized analytic systems.
At the scheduled review of several immunoassay documents, the Area Committee on Immunology and
Ligand Assay decided to develop one new document rather than expand the older ones. The area
committee combined the most relevant parts of these existing documents on radioimmunoassay and
enzyme and fluorescence immunoassays. A new section for luminescence was added to reflect its
popularity and wide use by manufacturers of automated instruments. The sections on antigen-antibody
components, sample requirements, quality assurance, and assay performance were enhanced for improved
utility of the guideline for developers and users of immunoassays. Also, the sections on antibody
components were provided in greater detail, because the antibody is probably the most important element
in the development and performance of a high-quality and low-bias immunoassay. This guideline will
provide information critical to the understanding of immunoassays to the manufacturer, the researcher,
and the healthcare professional.
3 Standard Precautions
Because it is often impossible to know what might be infectious, all human specimens are to be treated as
infectious and handled according to “standard precautions.” Standard precautions are guidelines that
combine the major features of “universal precautions and body substance isolation” practices. Standard
precautions cover the transmission of any pathogen and thus are more comprehensive than universal
precautions, which are intended to apply only to transmission of blood-borne pathogens. Standard
precaution and universal precaution guidelines are available from the U.S. Centers for Disease Control
and Prevention (Guideline for Isolation Precautions in Hospitals. Infection Control and Hospital
Epidemiology. CDC. 1996;Vol 17;1:53-80), (MMWR 1987;36[suppl 2S]2S-18S), and (MMWR
1988;37:377-382, 387-388). For specific precautions for preventing the laboratory transmission of blood-
borne infection from laboratory instruments and materials and for recommendations for the management
of blood-borne exposure, refer to the most current edition of NCCLS document M29—Protection of
Laboratory Workers from Occupationally Acquired Infections.
4 Terminology
4.1 Definitions
Accuracy – Closeness of the agreement between the result of a measurement and a true value of the
measurand (VIM93)1; NOTES: a) “Accepted reference value” may be used in place of “true value”; See
Trueness, below.
Activity (of a radioactive material) – The number of radioactive transitions taking place in a sample per
unit time; NOTE: See Specific activity.
Adjuvant – A substance admixed with an immunogen to elicit a more marked immune response
(RHUD1.7CD). 2
Affinity – 1) The force by which atoms, ions, molecules, prosthetic groups, and particles are attracted or
held together in chemical compounds; 2) In Immunology, a measure of the attraction or force of
association between a single antigenic site and a single antibody to that site.
Analyte – Component represented in the name of a measurable quantity; NOTES: a) In the type of
quantity “mass of protein in 24-hour urine,” “protein” is the analyte. In “amount of substance of glucose
in plasma,” “glucose” is the analyte. In both cases, the long phrase represents the Measurand (ISO
17511)3; b) In the type of quantity “catalytic concentration of lactate dehydrogenase isoenzyme 1 in
plasma,” “lactate dehydrogenase isoenzyme 1” is the analyte. (ISO 18153)4; c) In this document, the term
Analyte is combined with the term Measurand when its use relates to a biological fluid/matrix.
Analytical method – Set of operations, described specifically, used in the performance of particular
measurements according to a given method (VIM93)1; NOTE: The term Analytical method (U.S.) is
equivalent to Measurement procedure (Europe).
Analytical specificity – Ability of a measurement procedure to determine solely the measurable quantity
it purports to measure; NOTES: a) In quantitative testing, the ability of a measurement procedure to
determine only the component it purports to measure or the extent to which the assay responds only to all
subsets of a specified analyte and not to other substances present in the sample; b) For qualitative or
semiquantitative tests, the method’s ability to obtain negative results in concordance with negative results
obtained by the reference method; c) In Immunology, specificity is an antiserum quality defining its
reactivity with defined antigens and lack of specificity is the inaccuracy introduced by cross-reacting
and/or interfering substances, because cross-reacting substances compete with the analyte for antibody-
binding sites.
2 An NCCLS global consensus guideline. ©NCCLS. All rights reserved.
Volume 24 I/LA23-A
Antigen – Any substance that can stimulate the production of antibodies by an organism and combine
specifically with them.
Avidity – The net affinity of all binding sites of all antibodies in the antiserum, under specified
physicochemical reaction conditions. NOTES: a) It is a function of the affinities of the antibody-
combining sites on all antibodies present in an antiserum and all of the antigenic determinants of available
macromolecules; b) See Affinity.
Bias – Difference between the expectation of the test results and an accepted reference value; NOTES: a)
Bias is the total systematic error as contrasted to random error. There may be one or more systematic
error components contributing to the bias. A larger systematic difference from the accepted reference
value is reflected by a larger bias value; b) The measure of Trueness is usually expressed in terms of bias.
(ISO 3534-1) 5
Calibrator – Reference material whose value is used for the independent variable in a calibration
function.
Clinical sensitivity – The proportion of patients with a well-defined clinical disorder whose test values
are positive or exceed a defined decision limit (i.e., a positive result and identification of the patients who
have a disease); NOTES: a) It is the fraction of clinically true positive classification divided by the sum
of clinically true positive and clinically false negative; b) The clinical disorder must be defined by criteria
independent of the test under consideration; c) The term Clinical sensitivity (U.S.) is equivalent to
Diagnostic sensitivity (Europe).
Clinical specificity – The proportion of subjects who do not have a specified clinical disorder and whose
test results are negative or within the defined decision limit; NOTES: a) It is the fraction of clinically true
negative classifications divided by the sum of clinically true negative plus clinically false positive
classifications; b) The term Clinical Specificity (U.S.) is equivalent to Diagnostic specificity (Europe).
Competitive assay – An assay based on the competition of labeled and unlabeled analytes for a receptor.
Cross-reactivity – In Immunology, the reaction of an antibody with an antigen other than that which
elicited its formation, as a result of shared, similar, or identical antigenic determinants.
Cut-off values – The quantitative value of a measured analyte that is used to decide whether the result is
considered above or below a clinical or analytical decision point (usually positive or negative).
Detection limit//Limit of detection – The lowest amount of analyte in a sample which can be detected
but not quantified as an exact value (WHO-BS/95.1793).6
An NCCLS global consensus guideline. ©NCCLS. All rights reserved.
3
Number 16 NCCLS
Dynamic range – The total span over which an analysis can provide results; NOTES: a) Analytically,
the functional range of an assay over which concentrations of an analyte can be measured with accuracy
and precision; b) Physiologically, the full range of analyte levels to be expected in patient samples.
Efficiency – In Immunoassay, the percentage (number fraction multiplied by 100) of results that are true
results, whether positive or negative.
Epitope – Any site on an antigen molecule at which an antibody can bind; the chemical structure of the
site determining the specific combining antibody.
False-negative result – A negative test result for a patient or specimen that is positive for the condition or
constituent in question.
False-positive result – A positive test result for a patient or specimen that is negative for the condition or
constituent in question.
Gold standard – A nonspecific term that indicates that a process or material(s) is the best available
approximation of the truth.
Heterogeneous immunoassay – An immunoassay that requires the physical separation of free labeled
antigen (or antibody) from the labeled antigen (or antibody) bound in an immune complex, prior to
measurement of the quantity of label.
Homogeneous immunoassay – An immunoassay that requires only the mixing of a sample (analyte) and
immunochemical reagents (antibodies or antibody conjugates) with no wash step(s) to disturb the binding
equilibrium before the bound fraction is measured; NOTE: The measurand (analyte) must produce a
detectable dose-response signal upon binding that distinguishes it from unbound analyte.
Idiospecificity – A characteristic that describes the reactivity of an antiserum with a unique subset of a
broad antigen class.
Immunogen – Any substance that elicits a cellular and/or humoral immune response and the production
of antibody in a biological system; NOTE: See also Antigen.
Measurand – Particular quantity subject to measurement (VIM93)1; NOTES: a) This term and
definition encompass all quantities, while the commonly used term “analyte” refers to a tangible entity
subject to measurement. For example, “substance” concentration is a quantity that may be related to a
©
4 An NCCLS global consensus guideline. NCCLS. All rights reserved.
Volume 24 I/LA23-A
particular analyte; b) In this document, the term Analyte is combined with the term Measurand when its
use relates to a biological fluid/matrix.
Negative predictive value (NPV) – The likelihood that an individual with a negative test result does not
have the disease, or other characteristic which the test is designed to detect; NOTE: This varies with
prevalence of the disease unless the test is 100% sensitive.
Noncompetitive assay – An immunoassay in which the analyte (antigen) is captured by the receptor (or
first antibody), and a second labeled antibody is used to measure the amount of analyte bound or captured
by the binding site.
Positive predictive value (PPV) – The likelihood that an individual with a positive test result has a
particular disease, or characteristic, which the test is designed to detect; NOTE: This varies with
prevalence of the disease unless the test is 100% specific.
Postzone effect – The increased solubility of immune complexes resulting from the presence of a marked
excess of antigen in relationship to antibody concentration; NOTE: See also Prozone effect.
Potency – In Immunologic Testing, 1) The characteristic of an antibody that represents the concentration
of antibody and the avidity for a given substrate antigen in a defined method; 2) The characteristic of an
antigen solution that represents the concentration of the antigen in a defined method.
Precision – Closeness of agreement between independent test results obtained under stipulated
conditions; NOTES: a) Precision depends only on the distribution of random errors and does not relate to
the true value or the specified value; b) The measure of precision usually is expressed in terms of
imprecision and computed as a standard deviation of the test results. Less precision is reflected by a larger
standard deviation. (ISO 3534-1).5
Prevalence – The extent of occurrence expressed as a fraction of the numbers affected by the disease or
condition compared to the total number of members in the specified group.
Prozone effect – The result of a suboptimal antigen-antibody reaction in which either the antibody or
antigen is in excess, incomplete, or blocks an optimal reaction.
Quencher – Any molecular species that reduces the radiance measured from an emitting molecular
species (generally applied to fluorescence emission).
Reference value (accepted) – A value that serves as an agreed-upon reference for comparison and which
is derived as: a theoretical or established value based on scientific principles; an assigned value based on
experimental work of some national or international organization; or a consensus value based on
collaborative experimental work under the auspices of a scientific or engineering group (ISO 5725-1).7
Repeatability (of results of measurements) – Closeness of the agreement between results of successive
measurements of the same measurand carried out under the same conditions of measurement (VIM93).1
Reportable range – A set of values of measurands for which the error of a measuring instrument is
intended to lie within specified limits; NOTE: In this document, the term Reportable range (U.S.) is
equivalent to Measuring range (Europe).
Reproducibility (of results of measurements) – Closeness of the agreement between the results of
measurements of the same measurand carried out under changed conditions of measurement (VIM93).1
Sensitivity – In quantitative testing, the change in response of a measuring {system or} instrument
divided by the corresponding change in the stimulus (modified from VIM93)1; NOTE: In the context of
QC, the power of error detection of a QC system.
Specific activity – A measure event per unit mass per unit time; NOTE: See Activity.
Specificity – The ability of a measurement procedure to measure solely the measurand; NOTES: a)
Specificity has no numerical value in this context; b) See Measurand.
Titer – 1) The reciprocal of the dilution factor required to produce a defined outcome in a defined system;
NOTE: The titer is usually proportional to the analyte concentration; 2) In Radioimmunoassay, the
dilution of the antibody at which a specified percentage of the radiolabeled analyte is bound under defined
conditions.
Tracer – A radiolabeled analyte for use in dose-response for radioimmunoassay; NOTE: The production
of the tracer may use the replacement of one or more atoms in the analyte with a radioisotope or the
covalent labeling of the analyte with an isotope tag, most commonly 125I.
Trueness – The closeness of agreement between the average value obtained from a large series of test
results and an accepted reference value (ISO 3534-1) 5; NOTE: See also Accuracy, Bias.
Turnover number – The number of substrate molecules converted by an enzyme to product molecules
per unit time when the enzyme is saturated with substrate (under optimum conditions).
RIA radioimmunoassay
SRM standard reference material
TMB 3, 3′, 5, 5′ - tetramethylbenzidine
3
H tritium
WHO World Health Organization
5.1 Antibody
The principle function of the antibody is to specify and aid in the quantification of the analyte being
measured. Antibodies used for the detection of antigen or of other antibodies may be monoclonal, mixed
monoclonal, polyclonal, or combined monoclonal-polyclonal. High affinity and avidity usually are
associated with higher sensitivity of the tests derived from such antibodies; however, in some cases, high
affinity can be associated with reduced specificity. Therefore, the antibodies used in a particular
immunoassay should be matched to the criteria necessary for the performance of the particular method.
• immunization of an animal with an antigen followed by the subsequent collection and purification of
serum (i.e., polyclonal antibodies).
Monoclonal antibodies can be powerful immunochemical tools due to the attributes of single epitope-
binding specificity, homogeneity, and potentially unlimited supply. Due to their high degree of
specificity, monoclonal antibodies can be used successfully, under some conditions, when polyclonal
antibodies cannot. For example, screening and cloning techniques can permit the production of specific
monoclonal antibodies when sufficiently pure antigen is not available for in vivo production of specific
polyclonal antibodies. However, production of a “good” monoclonal antibody is often a difficult,
laborious, and costly task.
Depending on the application, the use of monoclonal antibodies is not always advantageous, especially if
affinity-purified polyclonal antibodies are available. Monoclonal antibodies often show low affinity and
avidity. Recognition of a single epitope can limit the usefulness of a monoclonal antibody for species- or
genus-wide detection of infectious agents and for test methods dependent on immunoprecipitation if
epitope density on the antigen is low. In addition, the antigen epitope might be shared by other infectious
agents, thus making the monoclonal antibody nonspecific.
The pooling of two or more monoclonal antibodies having different epitope specificities can overcome
these problems. Also, screening procedures can select for higher-affinity monoclonal antibodies. Because
these antibodies can behave differently from one assay protocol to another, screening of monoclonal
antibodies should employ a method as close as possible to the proposed application. The immunologic
assay design should include careful evaluation of the many factors that influence the choice of an
antibody reagent.
5.1.2 Immunogen
By its definition, an immunogen is any substance that can elicit an immune response. For the purpose of
designing an immunoassay, the immunogen will generally be the chosen analyte itself or a substance
sharing critical structural features with the analyte. For optimal reagent development, the immunogen will
preferably be pure and well characterized. Methods to demonstrate purity and characterization include:
• comparison of assay results using an assay specific for the immunogen (e.g., an immunoassay) and
an assay for total mass (e.g., total protein assay)
5.1.3 Production
The following immunization and collection parameters are important in antibody production:
• use of adjuvant
• immunization schedule
• bleeding schedule and technique for withdrawal (e.g., blood removal, plasmapheresis).
For the production of monoclonal antibodies, the following methods are also important:
• immunization
• hybridization
• cloning
• antibody production.
©
8 An NCCLS global consensus guideline. NCCLS. All rights reserved.
Volume 24 I/LA23-A
Serum is the preferred choice for the antibody preparation and may be obtained from plasma by standard
procedures. Some minimal processing is required and includes:
• preservation against:
— bacterial growth through addition of antibacterial agent
— thermal denaturation by storing at low temperature or freezing
— proteolysis through the addition of agents that inactivate proteases.
It may also be necessary to process the antibody to remove specific interferences or an undesired antibody
fraction. This processing may involve:
In other situations, it may be necessary to alter the specificity of the antibody preparation. Examples of
such alterations include:
5.1.5 Characterization
5.1.5.2 Titer
In general, the titer of an antibody is less important than its specificity, provided that the titer lies within
the range of values that permit the preparation of useful reagents. In certain situations, it may be
necessary to raise or lower the titer of an antibody preparation through concentration or with the addition
of nonimmune material.
Manufacturers of antibodies (and kits that utilize them) should indicate the form(s) of antibody supplied
with immunoassay (e.g., IgM, IgG). The manufacturer should also document reactivity with the isotype
and subclass of immunoglobulins (e.g., anti-µ or anti-ν chain or anti-F(c) fragment) for the reagent, for
kits designed to identify immunoglobulins (either a single class or subclass of antibodies). The lack of
reactivity with light chains must be documented. In addition, document that antibodies to IgG react with
all subclasses unless one or more subclasses are not important in detection of disease or immune
response.
Indicate known assay cross-reactivities in the package insert. Provide measures to counteract interference
when known (e.g., addition of nonreacting immunoglobulins to inactivate heterophilic antibodies). If
these measures are used in the assay, the manufacturer should state this fact.
5.1.5.5 Specificity
Conduct specificity studies in the presence of the measurand (analyte) in biological samples. A test for
undesired functions, such as cross-reactivity, should be at least as sensitive as the actual immunoassay.
Ideally, determine the level of cross-reactant needed to change the assay response by a stated amount
throughout the normal analyte range. Many interfering substances do not give response curves parallel to
that of the analyte.
• impact of common pathological states, physiological states, and sample mishandling that might alter
the specificity
5.2 Antigen
An important difference exists in operational definitions between an immunogen (see Section 5.1.2) and
an antigen, with the latter being a substance capable of binding to a specific antibody. Whereas all
immunogens are antigens, not all antigens are immunogenic. In the design of an immunoassay, antigens
are used as reagents in the assay system and as compounds to develop the antibodies that serve as the
source for immunodetection. Depending on the purpose of the assay, the antigen reagent may be
composed of discrete molecules (e.g., purified proteins) or a defined complex of molecules (e.g., virus).
Antigens may be obtained from a natural source or prepared from an in vitro system. To obtain antigen
that is representative of all variants, a pool of materials may be required.
Unwanted or nonspecific reactivity in immunoassays can result from the presence of contaminant
materials in antigen preparations. In general, the more highly purified the antigen, the smaller potential
for contaminant reactivity, and the better the performance in the assay. For direct detection of
contaminants in an antigen preparation, analyze several concentrations of the preparation using sieving-
gel electrophoresis, isoelectric focusing, two-dimensional electrophoresis, or similar methods. Also, use a
sensitive staining method, such as direct silver staining or immunoblotting, followed by a sensitive
detection method.
Choose the purification techniques to prevent antigenic changes caused by the removal or destruction of
some particular antigen or changes in the physical form of the antigen (e.g., proteolytic digestion, degree
of glycosylation, dissociation, aggregation, denaturation, small molecule binding) that may be relevant to
assay performance.
©
10 An NCCLS global consensus guideline. NCCLS. All rights reserved.
Volume 24 I/LA23-A
Purification of antigens from a microbiological system (e.g., viruses grown in tissue culture) requires the
following additional considerations:
• Contaminating microorganisms must be excluded from the tissue culture so that their presence does
not complicate purification.
• Harvest the antigens using a technique that minimizes contaminants from the culture and culture
medium.
• If a crude or partially purified antigen (e.g., virus preparation) is used, reactivity caused by
contaminants can be partially addressed by using a control antigen. A control antigen is prepared
from a culture (as well as the control sample) not infected with antigen microorganism but one that is
otherwise identical in composition and purification protocol. Immunological reactivity with
contaminants can be accounted for by subtracting the control antigen reactivity. However, this
process will not compensate for new host antigens arising from the infection.
5.2.2 Characterization
Test the antigenic preparation by physical, chemical, and immunological methods. Analyze the
preparation to ensure that each of the required antigens is present in the sample.
The ideal separation technique will completely separate free analyte from immunochemically bound
analyte. The separation technique should not significantly distort the equilibrium that exists between free
and bound entities. Good separation techniques are rapid, easy to perform, not strongly subject to minor
variations in process, and not affected by the sample matrix.
In this technique, one of the reagents is linked to a solid support that can be readily manipulated during
the assay to cause separation. The types of solid supports can differ considerably in design and
composition. The solid phase can be some device (e.g., dipstick or bead) that can be moved from reagent
to reagent; or it can be the reaction vessel itself, so that various assay reagents are added to the vessel.
General considerations on the solid support include:
An NCCLS global consensus guideline. ©NCCLS. All rights reserved.
11
Number 16 NCCLS
• The composition of the solid-phase surface should be controlled, so that binding of assay reagent is
reproducible and stable.
• Conditions for binding of assay reagents (e.g., pH, temperature) may need to be controlled to ensure
reproducibility and stability.
• Because the final assay system is not homogeneous, a bulk, which is representative of the whole lot of
reagents, cannot be sampled. Consequently, the final quality control of the product should be
carefully designed to provide sufficient statistical sampling.
• The method of binding should be designed so that reagent loss from the solid support during the assay
is minimal, because reagent loss can affect reproducibility and sensitivity.
• Binding may alter the immunochemical reactivity of the antibody or antigen reagent.
• Care should be taken to ensure that all antigens of interest bind to the solid phase in the proper
proportions.
• Where light absorption or fluorescence measurements are made in the presence of the solid support,
special precautions are recommended to assure sufficient optical uniformity.
The reagent is typically bound to the solid support either through adsorption or covalent binding.
Adsorption is easier to accomplish but is less reproducible and allows for loss of adsorbed reagent during
the immunoassay.
Covalent linkage is easier to monitor and control than adsorption. In addition, there is less potential for
loss of reagent during the assay. There are numerous reagents available for covalently linking a
compound to a solid support. A linking reagent should be selected which minimizes any damage to the
immunochemical properties of the assay reagent. After linking the desired reagent, remaining functional
linking groups are usually rendered nonreactive.
A second antibody directed against one of the primary assay reactants can be used to precipitate the
complex. Alternatively, the second antibody can be bound to a solid support. These techniques have the
potential advantage of increasing specificity of the assay by the additional immunochemical binding.
There is, however, the disadvantage that an additional biological reagent must be carefully controlled.
Antibody used to achieve separation should be of sufficient specificity to bind mainly the protein to be
separated. It should be free of interfering impurities and of adequate titer. When used for
immunoprecipitation, a broad antibody concentration range is preferable, whereby all immune complexes
are precipitated between the prozone and postzone regions.
Other binding proteins (e.g., Staph protein-A, biotin-avidin, and biotin-streptavidin) may be used to bind
the primary antibody.
There are various other techniques for separation which should be chosen on the basis of individual
circumstances, including salt (e.g., ammonium sulfate) or solvent (e.g., ethanol, polyethylene glycol)
precipitation. These techniques rely on the differential solubility of proteins in the presence of these
compounds. Because the efficiency of precipitation depends on the amount of total protein present,
protein concentration must be controlled. In addition, the precipitations are often strongly temperature
©
12 An NCCLS global consensus guideline. NCCLS. All rights reserved.
Volume 24 I/LA23-A
dependent. These techniques also have the potential for disrupting complexes and shifting the
immunochemical equilibrium.
Proper sample selection and collection techniques are essential for optimal test performance. Before
specimens are collected, specific instructions should be available either from a test manufacturer or
provided in a laboratory procedure manual.
Immunoassays are applicable to a variety of matrices. The user should identify the appropriate matrices,
(e.g., blood spots, serum, plasma, urine) with the necessary patient preparation, collection techniques, and
devices that are known to work for the assay. Any anticoagulants that may be used should be indicated.
(Refer to NCCLS documents H3—Procedures for the Collection of Diagnostic Blood Specimens by
Venipuncture, H4—Procedures and Devices for the Collection of Diagnostic Blood Specimens by Skin
Puncture, H11—Procedures for the Collection of Arterial Blood Specimens, GP16—Routine Urinalysis
and Collection, Transportation, and Preservation of Urine Specimens and LA4—Blood Collection on
Filter Paper for Newborn Screening Programs.) Provide the storage conditions and stability for the
samples. Note any container restrictions.
Assay performance must be proven on all sample types for which it is intended. Minimum sample types
include both fresh and frozen sera which have been centrifuged to remove particulates. Additional sample
types may include whole blood, cerebrospinal fluid, amniotic fluid, urine, and saliva.
Samples should be transported to the laboratory as expeditiously as possible to minimize the growth of
contaminants and the release of proteases from any cells that are present. The effects of transportation on
sample integrity vary with the type of sample and the nature of the assay.
Within the laboratory, certain treatments, (e.g., heating) might be necessary for one methodology, but
preclude use of the sample for further tests. If necessary, samples should be split before treatment.
The manufacturer has the responsibility for specifying required conditions for optimal test performance,
but it is the laboratory’s responsibility to inform its users about appropriate collection and transport of
samples. Manufacturers are responsible to notify users of the effect of high lipid or hemolyzed samples on
the test performance. Samples that arrive in unsuitable condition for accurate testing should be rejected
after notifying the healthcare professionals (e.g., physicians) responsible for care of the patients. In
general, laboratories should not offer tests for which samples cannot realistically be obtained and stored in
the manner recommended for the test methods.
Appropriate sample storage can be critical in obtaining accurate test results. If the assay cannot be run
immediately, processed samples (e.g., serum/plasma separated and remove from cells) should be stored at
4 °C or per assay manufacturer's instructions. If a longer delay is expected, freezing of samples may be
necessary. The manufacturer should recommend the preferred temperature for freezing. However, the
excessive freeze/thaw cycles should be avoided, and the user may consider the use of stored control
samples with the test samples. Storage conditions can also alter reactivity in some assay methods but not
in others.
7 Radioimmunoassays
The most commonly used radioisotope for radioimmunoassays (RIA) is iodine (125I). Tritium (3H) is also
used for some applications. These assays are referred to as “gamma” or “beta” immunoassays,
respectively. Competitive assays involve the use of radiolabeled analyte as a tracer to estimate the bound
antibody and the free fractions, whereas immunometric assays use tracer-labeled antibody for measuring
the amount of bound analyte. The level of radioactivity (ionizing radiation) is measured (disintegrations
per minute) using a scintillation counter. The higher the specific activity of the tracer, the more sensitive
the assay. However, with protein analytes, radiolysis can damage the proteins. Although radiolabeled
immunoassays have high sensitivity and the tracer can be similar to the analyte, the assay kits have a short
shelf life, and safety regulations and radioactive waste disposal create additional concerns relative to other
immunoassays.
Specific information about the intended use and a summary explanation of the principle of the test for the
user should be provided. The method of radiolabeling should be specified, and any chemical
modifications to the analyte in order to make radiolabeling possible should be described. The information
should be sufficient to ensure the authorized user has a generic appreciation of the test applications and an
adequate understanding of the chemistry involved.
7.2 Reagents
The quantity of reagent supplied (mass, volume, or number of units, as appropriate) should be indicated.
The user should also note if the separation reagent was specifically selected or pretreated, because similar
reagents from different manufacturers, or different lots from the same manufacturer should not be
interchanged.
The biological source (e.g., animal species) in which a second antibody was produced, as well as the
general characteristics of immunogen (e.g., purified rabbit gamma globulin) in immunoprecipitant
separation (see Sections 5.1.1 through 5.1.4), should be identified. The buffer used, its concentration, and
pH should be indicated, and all other reagents identified. Stability, under carefully defined conditions of
storage, should be stated. Preservatives used in reagents (e.g., sodium azide) should be identified, together
with any necessary precautions. All shelf-life and storage conditions for reconstituted or freshly prepared
reagents should be noted.
A variety of techniques and reagents may be used to separate the bound antibody from free fractions. All
reagents used to separate bound fraction from the free fraction should be identified. Reagents should be
identified by chemical name, if no proprietary information is involved (e.g., charcoal, polyethylene
glycol, or ammonium sulfate), as well as by general chemical composition (e.g., anion-exchange resin,
silicate, immunoprecipitant in the solution or attached to a solid phase [inside of a tube, polymer beads,
©
14 An NCCLS global consensus guideline. NCCLS. All rights reserved.
Volume 24 I/LA23-A
etc.]) and by physical separation techniques (e.g., ultrafiltration, dialysis, gel filtration, or
electrophoresis). Any effect of interfering substances on the separation process should be described, and
pertinent references listed. No additional information other than that required by regulatory bodies is
recommended to assess the quality of separation reagents for use in RIAs.
7.2.2 Conjugates
The chemical process of the radiolabeling techniques for preparation of the tracer should be described.
The specific sites of the isotope label and the number of labels for the tracer should be identified, if
known. For labeled proteins, the specific sites may not be known. The specific activity should be listed,
and the conjugation process used for production of the specific antibodies should be described (see
Section 5.1).
7.2.3 Calibrators
Calibrators should show traceability to certified reference materials, if available. If certified reference
materials are not available, the source of relevant information and physical properties for the substance
used should be specified. Calibrators prepared from substances of attested chemical and physical purity
should be identified with their sources.
For substances with unknown chemical and physical purity (or which are unstable in highly purified
form), other types of calibrators are necessary. The source, and any physical properties and stability
concerns, should be identified. Many substances exist in numerous forms, and suitable reference materials
may be difficult to select. For biological reference material, dose-response relationships parallel to the
substance to be quantified in the assay should be given. The assay should measure the reference material
in the same biological matrix as the unknown samples (see Section 11). Although this procedure may not
give absolute concentrations, the relative concentrations so determined may be clinically useful. Potency,
in terms of the appropriate reference material (including confidence limits) and the number and type of
independent assays performed to determine this potency, should be provided. A sufficient supply of
calibrators should be maintained for repeated use over extended periods of time. Recalibration of new
calibrator preparations at frequent time intervals should be avoided.
The book, A National Understanding for the Development of Reference Materials and Methods for
Clinical Chemistry,8 discusses many generalities, as well as specific aspects of calibration and reference
materials. Additional literature about calibration and reference materials should be reviewed. (See
NCCLS document NRSCL13—The Reference System for the Clinical Laboratory: Criteria for
Development and Credentialing of Methods and Materials for Harmonization of Results.)
The classic RIA is a competitive binding (displacement) immunoassay in which the analyte from the
sample displaces the radiolabeled species of the analyte or its close analog. A dose-response curve for
measurements is produced by measuring the ratio of radioactivity bound in the absence of analyte (B0)
versus the bound radioactivity (B) in the presence of known quantities of analyte. Sandwich immunoassay
configurations for RIA are used also where one antibody binds the analyte to a solid phase and a second
radiolabeled antibody (dose-response indicator) is added.
For both assay types, the separation of bound and free radiolabeled materials is critical for measurements.
A total count and blank sample are important components of the RIA (see Section 7.4). Separation may
be accomplished by precipitation of immune complexes, adsorption of the free label, or by washing the
solid phase (see Section 7.2.1). For some beta isotope assays, the RIA may use proximity to the
scintillation chemicals embedded in the solid phase to produce a dose-response curve. Radioactivity is
generally measured as counts-per-minute (CPM), and at least 100 total counts for a measurement should
be accumulated to avoid imprecision due to stochastic counting errors. Radiolabeled reagents should be
stored as specified by the manufacturer, and they must be handled and disposed of according to national
and regional regulatory guidelines for radioisotope use (see Sections 7.5 and 7.6). Since denaturation may
alter the immunoreactivity of analytes, especially for proteins, the radiolabeled reagents should not be
used beyond their expiration dates, even if they are still providing adequate CPM rates.
The radioactivity should be measured using a gamma or beta scintillation counter, and the level of bound
and free fractions relative to each level of calibrator used should be determined. Background, nonspecific
binding, and total count tubes should be measured in the process and used to determine quenching and
other assay performance characteristics.
As an example, a tabular worksheet showing the sample number, other identification, count rate, dose-
response variable (y-axis), and analyte concentration (x-axis) should be prepared. The method of
calculation should be noted, including the computer algorithm, if appropriate. Written and graphic
examples of typical calculations should be given, e.g., documentation of numerical steps, count-rate
conversions, percent-binding calculations, or a computer-generated curve. A typical calibration curve
(dose-response curve), including the effective assay range, should be shown.
Data on standard deviation or coefficients of variation over the clinically significant parts of the
calibration curve should be included. The method of calculation should be stated and references cited.9
The RIA kit has a short and defined shelf life. The activity of the radioactive isotope gradually decreases
as it decays. The rate of decay is specific to the isotope. Iodine125, commonly used for assays, has a half-
life of 60 days. One or more atoms are replaced in the analyte to produce the tracer.
Sources of error, interferences, and inherent limitations of the assay should be identified. A
comprehensive listing of the interfering substances (drugs that test patients might be taking and structural
analogs of the analyte) and the relative percent cross-reactivity to the specific analyte measured should be
provided. The appropriate literature reference should be included.
RIAs can be used only in approved institutions or by individuals operating under a general license for
radioisotope use.10 Radioactive materials may be received, acquired, and used only by authorized
laboratories and then only for in vitro clinical or laboratory tests not involving internal or external
administration of radioisotopes to humans. Appropriate national and local regulations must be adhered to
with the use of radioisotopes. Storage of isotopes must be limited to specifically designated areas.
Access to radioactive materials must be limited to authorized personnel only.
Containers must be labeled with the type and actual amount of radioactivity contained at the calibration
date of the isotope; and the package insert must indicate the theoretical radioactivity in the kit. All known
items that might influence the validity of the assay results should be delineated.
Good laboratory practices imply safe laboratory operations. Diligence must always be exercised to ensure
that one is working in a safe environment and not subject to unsafe conditions. RIAs include many steps
that require caution by the analyst to avoid personal injury. Personal hazards are commonplace in the
laboratory and should not be taken lightly (see the current version of NCCLS document M29—Protection
of Laboratory Workers from Occupationally Acquired Infections).
©
16 An NCCLS global consensus guideline. NCCLS. All rights reserved.
Volume 24 I/LA23-A
The disposal of radioactive materials is subject to the regulation and conditions of the user license and in
accordance with the recommendations of any local regulatory authority. (See NCCLS document GP5—
Clinical Laboratory Waste Management.) Approved users must ensure complete internal controls for
receipt, storage, use, transfer, and ultimate disposal of all radioactive materials. Regardless of the method
of disposal, records of radioactive waste should be maintained, even if two or three methods of disposal
exist for the laboratory. Complete records of the receipt, use, and disposal of all radioactive materials
should be maintained, even if disposal is handled by a radioactive waste broker. Small-quantity generators
using RIA kits only may be exempt.10
8 Enzyme Immunoassays
Enzyme immunoassays (EIA) use an enzyme to label an antigen or antibody.11,12 The high turnover
number of enzymes provides a significant amplification, and this catalytic property makes enzymes an
obvious choice as labels. Immunoassays that use enzyme labels have several advantages, including:
• they have strong signal strength, high sensitivity, and wide applicability
• enzymes are less hazardous and have a longer shelf-life than most radioactive tracers
• qualitative assays can be performed with minimal equipment and at low cost
• assays can be used by untrained personnel in point-of-care and home testing applications.
There are some disadvantages. Measurement of enzyme activity usually requires an extra incubation step
in the assay procedure. In addition, enzymes are susceptible to interferences and/or changes in assay
conditions (e.g., time, temperature, pH, inhibitory substances). Enzymes also have practical limitations
related to their large size, and they tend to bind nonspecifically to reaction vessels.
The two most commonly used enzyme labels are horseradish peroxidase and alkaline phosphatase.
Horseradish peroxidase is popular due to its high turnover number, its small molecular size (44 kDa
compared to 140 kDa for alkaline phosphatase), and large number of hydrogen donors that can reduce
hydrogen peroxide to generate colored, fluorescent, or luminescent products. Alkaline phosphatase is
often favored because of its simple reaction kinetics and the variety of assay systems available. Glucose-
6-phosphate dehydrogenase is a very common label for homogenous EMIT assays.
Increasingly, enzymes are being used as secondary labels. For example, antigen (or antibody) labeled
with the low molecular weight cofactor biotin may be used as a primary label, and the biotin binding
protein streptavidin labeled with enzyme may be used as a secondary label. Enzymes are also being used
in assays with labels that contain enzyme substrates, enzyme inhibitors, coenzymes, or enzyme cofactors.
8.2 Reagents
Enzyme immunoassays, other than homogenous immunoassays,13 require a procedure for physically
separating the antibody-antigen complex from other reactants. Several methods are available, but solid
phases with antigen or antibody attached have become dominant. Solid phases are usually plastics (e.g.,
polystyrene) that adsorb enzyme proteins noncovalently.
8.2.2 Conjugates
Enzymes are covalently linked to the antigen or antibody to form a conjugate. The optimal ratio of
enzyme to antigen or antibody varies and is determined empirically. The bifunctional reagent
glutaraldehyde is often used for alkaline phosphatase and horseradish peroxidase. Conjugation methods
using the periodate method may be used with glycoproteins such as horseradish peroxidase. The most
common methods for coupling haptens and enzymes are the mixed anhydride and active-ester procedures.
Molecular biological techniques are also available for sensitive proteins that may be damaged during
conjugation reactions.
Enzyme-labeled antibodies or antigens first react with the ligand of interest, and enzyme substrate is
subsequently added. Measurement of the resultant decrease in substrate concentration or increase in
product concentration is then used either to detect or quantitate the antigen-antibody reaction. Essentially
all enzyme immunoassays can be classified as competitive (reagent-limited) or noncompetitive (reagent-
excess). In a typical competitive immunoassay, enzyme-labeled analyte competes with unlabeled analyte
in the sample for a limited number of antibody binding sites. After incubation, the unbound material is
removed, leaving the bound enzyme attached to the solid phase. Substrate is added, and at the end of an
incubation period, the enzyme reaction is stopped by addition of an inhibitor or by changing the pH. The
strength of the signal depends on the amount of enzyme present, which in a competitive assay is inversely
proportional to the amount of measurand (analyte) in the original sample.
Most enzyme immunoassays are heterogeneous and require a physical separation system. Homogeneous
(separation-free) assays use kinetic measurements of the enzyme activity. When antibody reacts with the
enzyme-labeled antigen, the activity of the enzyme is diminished. Enzyme activity is directly related to
the concentration of the measurand (analyte) in the sample.
Enzymes are easily detected using convenient substrates that give rise to colored, fluorescent, or
luminescent compounds. Colored products can be monitored visually with the naked eye or measured
accurately in a photometer. The most common colorimetric substrate for alkaline phosphatase is p-
nitrophenyl phosphate. ABTS, OPD, and TMB are common peroxidase substrates. TMB is often chosen,
since it gives the highest absorbance values, low backgrounds, and is not mutagenic. Colorimetric signals
generated with horseradish peroxidase are an order of magnitude greater than alkaline phosphatase.14
Fluorescent-labeled substrates or products are inherently more sensitive than colorimetric measurements,
since fluorescent compounds can be repeatedly excited by incident radiation. The most common substrate
is 4-methylumbelliferyl phosphate with alkaline phosphatase as the enzyme label. Luminescent enzyme
immunoassays can achieve levels of sensitivity that are several orders of magnitude better than
colorimetric or fluorometric assays. Substrates that yield luminescent end points are available for the
common enzyme labels. A very sensitive assay for alkaline phosphatase, for example, uses a
©
18 An NCCLS global consensus guideline. NCCLS. All rights reserved.
Volume 24 I/LA23-A
Colorimetric enzyme assays, particularly noncompetitive immunometric assays, are limited by the
working range of the spectrophotometer (about 0.1 to 1.5 absorbance units). Extension of this dynamic
range is possible by monitoring the absorbance at a wavelength that is offset from the peak maximum, or
by monitoring color development kinetically.
Temperature, polarity, pH, and dissolved oxygen content affect fluorescent enzyme assays. Interferences
may also arise from light scattering (particulate matter), natural background fluorescence (proteins,
bilirubin, and NADH), and quenching effects from molecules in the sample.
Some luminescent enzyme immunoassays, particularly those using horseradish peroxidase, have low
quantum yields and produce a weak light emission that rapidly decays. Addition of certain enhancer
compounds can substantially increase the light output, reduce the background light emission from the
substrates, and improve the signal-to-noise ratio.
9 Fluorescent Immunoassays
There are several types of immunoassay systems based on fluorescence technology. Fluorescent methods
use a fluorescent molecule to produce an emission as it passes from an excited state to the ground state.
Emissions are captured and quantified with a fluorescence spectrophotometer. Fluorescein, green
fluorescent protein, methylumbelliferone, and phycoerythrin are examples of fluorescers. Flow cytometry
is an application of fluorescence technology to read streams of cells or particles labeled with multiple,
unique fluorochromes. Fluorescence polarization methods use polarized light to distinguish bound and
unbound antibody-analyte complexes. Time-resolved fluorescent methods use fluorescers with long decay
times (typically lanthanide ions, such as europium) and instruments that excite with microsecond-pulsed
light, and record fluorescence emission only during the dark phase of each pulse.
• Fluorescence intensity. A fluorescer with high quantum yield will maximize the sensitivity of
detection.
• Fluorescer attachment. The fluorescer will usually have a functional group that will form a covalent
bond with the compound of interest. The fluorescer conjugate must be soluble.
• Stability. The fluorescer system must be stable during the assay and when stored.
An NCCLS global consensus guideline. ©NCCLS. All rights reserved.
19
Number 16 NCCLS
• Linearity and dynamic range. Fluorescence intensity readings should be linear with respect to the
concentration of the fluorescer unless confounding factors interfere.
9.2 Reagents
The chemical and physical characteristics of fluorescer and/or quencher must be determined and should
include purity, molecular weight, excitation, and emission spectra. If the fluorescer is conjugated to
protein, the purity and stoichiometry of the conjugate should be determined. When conjugating
fluorescers and proteins, the following should be considered:
• The fluorescence of the conjugate peaks at a defined ratio of label to protein, beyond which the
fluorescence may decrease.
• Conjugation may change the solubility, stability, size, shape, and net charge of the protein.
• The excitation and emission spectra of the conjugate may differ from that of the free fluorescer.
• Quenching effects cause nonlinear responses at higher fluorescer concentrations, while noise and
photobleaching effects contribute to nonlinearity at lower concentrations.
9.2.2 Calibrators
Reference materials may be patient samples of known composition or a sample matrix to which a
quantitative amount of reference material has been added. The reference material may be:
• a pure, well-defined material, such as the fluorescein reference material (Fluorescein SRM)15,a ;
• an impure material of unknown composition but defined origin, obtained by a reproducible procedure.
Certain analytes have differing and complex characteristics in different samples (e.g., rheumatoid factor
may be IgG, IgA, IgM, or a mixture). In this case, the reference material is defined by its
immunochemical and functional property, rather than by a unique feature. Such a biological reference
material must give a dose-response relationship that correlates to the substance to be quantified in the
assay, and it should be stable in the biological matrix to be studied.
a
Available from NIST.
©
20 An NCCLS global consensus guideline. NCCLS. All rights reserved.
Volume 24 I/LA23-A
but the result is a fluorescent emission that is quantitatively proportional to the amount of analyte in the
test sample.
Fluorescent signal detection may be accomplished by several instrumental techniques: filter fluorometers,
spectrofluorometers, fluorescence polarization analyzers, and time-resolve fluorometers.14 Appropriate
excitation and emission wavelengths are characteristic of the assay system under development and are
optimized for the detection system chosen.
Quantitation of signal and correlation to measurand (analyte) concentration requires the development of a
dose-response curve. To develop a dose-response curve, two fundamental factors are required:
The dynamic range of the assay must be within the capability of the detector. A series of equal-volume
serial dilutions of the calibrator material across a defined range will demonstrate the limits of assay
linearity. Linear regression of the log dilution factor (X) against log instrument reading (Y) should give a
best-fit line with residuals at each point no greater than 10% of the expected dilution value. Multiple runs
will allow an analysis of the accuracy and precision of the assay. The three biggest contributions to non-
linearity are quenching, photodegradation, and noise. Most systems are generally capable of linear
responses across at least two decades of fluorescer concentration (a hundredfold range). Quantitative
measurements are less reliable outside the linear range, and assays should be optimized to use the linear
response range for the most clinically relevant measurements. Because of reduced noise and resistance to
quenching, time-resolved fluorescence has the largest linear dynamic range (a thousandfold) among the
various platforms in common use.
There are several limitations that are unique to fluorescent systems. Interferences can produce either
positive or negative bias: substances in the sample that fluoresce at the same wavelengths as the analyte-
fluor will result in a falsely elevated signal. The presence of interfering substances may quench the
fluorescence (e.g., by competing for antibody), resulting in a reduced signal. Fluorescence emission is
sensitive to pH, temperature, ionic strength, and solvent viscosity. All of these parameters must be
characterized, optimized, and controlled.
10 Luminescent Immunoassays
Chemiluminescence, electroluminescence, and bioluminescence are all biochemical systems that produce
light.16 In a chemiluminescent system, a chemical reaction (usually oxidation) generates an organic
molecule in an electronically excited state, which emits photons upon return to the ground state. The
chemical reaction is initiated by the addition of an oxidizing agent. Light emission by
electroluminescence is accomplished by generating an electronically excited state of an appropriate
molecule through the application of an electric potential. A bioluminescent reaction utilizes a biological
system (e.g., luciferin/luciferase) to catalyze a series of reactions, which results in light emission. Each of
these systems offers the following advantages:
• Because sample radiation is not required, the high backgrounds due to light scattering and nonspecific
excitation are eliminated, allowing greater sensitivity and dynamic range.
An NCCLS global consensus guideline. ©NCCLS. All rights reserved.
21
Number 16 NCCLS
• Improved signal detection coupled with the power of immunoassay technology provides better
specificity.
• Luminescent assays often occur on solid phases which capture the analyte and wash away interfering
substances, enhancing sensitivity and specificity.
10.2 Reagents
10.2.1 Lumophores
A luminescent molecule from one of the following classes typically provides the signal in
chemiluminescent assays: acylhydrazides (luminol), 1,2-dioxetanes (AMPPD), acridinium esters and
related analogs, or pyridopyridazines; the luciferin/luciferase pair is used in bioluminescent systems.17
One application of electroluminescence employs an electronically excited state of a ruthenium complex.18
10.2.2 Conjugates
Assay formats are typically monoclonal/polyclonal sandwich assays on solid phases. Both direct and
indirect competitive immunoassay protocols predominate in chemiluminescent systems. Quantitative light
emission may result from direct chemical cleavage of an acridinium ester for example; or from the action
of an enzyme or enzymatic system on a luminescent substrate (e.g., luciferin/luciferase or a
luminol/enzyme pair); or from an electronically excited state of an appropriate molecule, resulting from
the application of an electrical potential difference to the reaction.
10.3.1 Calibrators
There is no reference material for luminescence. The detector in use can be calibrated with a defined set
of dilutions of an appropriate material of proven stability. For example, a quantitative set of dilutions of
an acridinium conjugate would provide an assessment of the dynamic range of the system for an
acridinium-based assay. Manufacturers of luminometers also use isotopes (14C and 3H) to standardize an
instrument population. Assay calibration may be accomplished by strategies previously discussed for
fluorescence assays.b
The signal generated by a luminescent reaction is detected as photons by a photomultiplier tube for a
defined period, and the count data are transferred to a central processing unit to yield a relative light unit
measurement. A background measurement is subtracted from the reaction measurement to yield the final
measurement. Immunoassays require nonlinear algorithms to fit curves and establish performance criteria.
b
Calibrator materials should be NIST or WHO reference materials when available.
©
22 An NCCLS global consensus guideline. NCCLS. All rights reserved.
Volume 24 I/LA23-A
Because photon detection is a very sensitive method, contaminating signals may result from a variety of
sources. Biological organisms produce luminescent materials; cleaning solutions and laboratory dust may
contain substances that produce light; etc. The validity criteria for background measurements and assay
measurements are safeguards against many interfering substances. The maintenance procedures provided
by the manufacturer address system decontamination and cleanliness, but the importance of Good
Laboratory Practices cannot be overstated when working with this technology.
Even those familiar with laboratory terminology are often confused by the international community’s use
of metrological terms. (See the Note on Terminology beneath the Foreword) To narrow the
communication gap, it is important to have a general understanding that accuracy, trueness, and precision
are typically qualitative terms, and therefore, have no numerical value. To express these terms
quantitatively, the terms error, bias, and imprecision are used.
Accuracy (of measurement) refers to how close a single measurement result (test method result) comes to
a true, expected, or accepted reference value. But it is usually expressed in terms of error (of
measurement), or how far away one’s result is from one’s expectations (i.e., mathematically, the result of
a single measurement minus a true, expected, or accepted reference value).
Trueness refers to the degree of agreement between an average value of a large series of measurements
and a true, expected, or accepted reference value. Bias is the quantitative expression of trueness and refers
to the degree of systematic difference from the accepted reference or true value, whether positive or
negative. A larger systematic difference will mean a larger bias.
Precision refers to the degree of agreement between independent test/measurement results obtained under
stipulated conditions. It depends only on the distribution of random errors, and it does not relate to a
reference or true value. Determined qualitatively as “high,” “medium,” or “low,” it is expressed
mathematically in terms of imprecision (i.e., the dispersion of results of measurements obtained under
specified conditions) and is measured as the standard deviation (SD) or coefficient of variation (CV). A
lower CV value reflects higher precision.
The “stipulated conditions” are further broken down into “repeatability (within-run) conditions” (i.e.,
independent test results obtained with the same method, on identical test material, in the same laboratory,
by the same operator, using the same equipment, within a short interval of time) and “reproducibility
(total precision) conditions” (i.e., where test results are obtained with the same method, on identical test
items, under different settings (e.g. in different laboratories), with different operators using different
equipment.5
Traceability to an accepted reference value for an immunoassay should take into account that different
reagents may react to different extents to the various epitopes of the reference material and yield different
although related quantities.3
The recommended trueness and precision of an assay for clinical decision may depend on the seriousness
of the condition, the degree of separation of results obtained from affected and unaffected persons, and
the assay methods available. (See NCCLS document I/LA21—Clinical Evaluation of Immunoassays.)
The analytical evaluation of a new immunoassay or comparison to existing immunoassays should include
adequate documentation of trueness and precision (including repeatability and reproducibility) at all
levels of diagnostic (clinical) performance.
An NCCLS global consensus guideline. ©NCCLS. All rights reserved.
23
Number 16 NCCLS
Several interrelated concepts are of vital importance in the interpretation of laboratory tests. These
include detection limits, cut-off values, sensitivity, and specificity (analytical and diagnostic [clinical]),
positive predictive value, negative predictive value, false-positive results, false-negative results,
prevalence, efficiency, and “gold standards.” These terms are defined in Section 4.
The lower immunochemical detection limit for an assay represents the lowest level of an analyte that can
be reproducibly detected. The laboratory or test sensitivity is identical to the detection limit. Even
laboratory sensitivity can be defined two ways: the detection limit (the lowest value that can be measured
different from zero), and the smallest difference that can be measured reliably between any two values.
Detection limits are expected to vary with sample type, e.g., an assay can show significantly different
detection limits in the testing of saliva and serum.
Other terms can also be defined from both laboratory and clinical standpoints. From a clinical standpoint,
the positive predictive value of a test is the percentage of patients who have a positive test result and are
indeed positive for the trait or analyte. Conversely, the negative predictive value is the percentage of
persons with a negative test result who are indeed negative. Efficiency (the percentage of results that are
true results, whether positive or negative) is of clinical importance if false-positive and false-negative
results are equally undesirable. Diagnostic (clinical) sensitivity, in contrast, is the percentage of patients
who are positive for a trait and test positive; and diagnostic (clinical) specificity is the percentage that do
not have the trait and test negative.
The comparison of quantitative data from different immunoassays is commonly performed by simple
linear regression, with a high correlation coefficient (r) taken to indicate comparability of results. This
approach is often misused. First, a nonzero intercept suggests differences in behavior of the two assays,
but is usually ignored. Secondly, the r value is unduly influenced by the highest values; if testing includes
a few very high results, a high r may be associated with poor comparison for the lower values more
commonly seen in patient samples. Finally, simple linear regression implies superiority of the method
used as the independent variable. Deming regression can be used to overcome the last problem.19
However, more sophisticated approaches are necessary to overcome others. The logarithmic
transformation method of Finney,20,21 which recommends similar behavior of samples at all levels (or of
dilutions of samples) for colinearity, is one suggested approach. The correlation of results can be
demonstrated visually in package inserts by scatter plots and/or residual plots covering the entire assay
range, but without correlation coefficients.
To compare similar reagents (e.g., different antigen preparations, calibrators, controls, etc.) within an
immunoassay, establish the functional linear relationship between the materials (rather than simple
regression). NCCLS document EP6—Evaluation of the Linearity of Quantitative Measurement
Procedures: A Statistical Approach provides a method for evaluating linearity. A plot of test results for
two materials, assayed in multiple dilutions, should give a straight line with a zero intercept. The absence
of linearity implies antigen or antibody excess at the assay limits; a nonzero intercept implies matrix or
antigenic differences between the two materials. An assessment of the linearity of a method allows the
user to establish the dynamic range or range of clinical utility for that method.
NCCLS document EP9—Method Comparison and Bias Estimation Using Patient Samples provides
guidelines for designing experiments to evaluate the bias between two methods that measure the same
measurand (analyte). Ideally, a test method should be compared with a reference method. For the clinical
laboratory, the comparative method is often the current method. The evaluation should determine whether
the two methods yield equivalent results within the statistical limits of the experiment. NCCLS document
EP9 allows the estimation of the bias (expected difference) between two methods at various
©
24 An NCCLS global consensus guideline. NCCLS. All rights reserved.
Volume 24 I/LA23-A
concentrations of analyte. If the comparative method is the same one used by the manufacturer in the
statement of claims, it is possible to statistically compare the experimental results to the manufacturer’s
claims in order to verify acceptable performance.
Reference interval, also known as “reference range” (historically as “normal range”), is the span of test
values expected for a designated (reference) population of individuals (i.e., the range of values
representing 95% of the individuals presumed to be healthy [or normal]). A variety of factors determine
the reference interval for the measurand (analyte). The size and characteristics should be provided for the
selected reference population contributing to the reported reference interval. Diagnostic laboratories, test
manufacturers, research test developers, and users of clinical immunoassays need to establish their own
appropriate reference intervals to assess and understand the clinical applications of the immunoassay.
Intervals provided by manufacturers may be useful as a performance example and guide. Reference
intervals are necessary to meet requirements of reliability and usefulness for the quantitative test. (See
NCCLS document C28—How to Define and Determine Reference Intervals in the Clinical Laboratory
for detailed methodological approaches and recommended procedures.) Reference interval should not be
confused with dynamic range, diagnostic cutoff, and measuring range (reportable range). Immunoassays
may involve the use of medical decision limits, such as cutoff values, especially for screening
applications. (See the current edition of NCCLS document I/LA21—Clinical Evaluation of
Immunoassays.) A diagnostic cutoff is often based upon diagnostic (clinical) sensitivity and specificity
for a specific medical disorder.
A reference value provides a means of comparing or relating an observed data element to a reference
database from a characterized population of subjects. Reference values may be from correlation data that
have been determined by assessment of the immunoassay against a more specific measurement procedure,
or against a reference method or a reference material. Expected values are produced by traceability and
verification to a certified reference material to support the performance of the immunoassay (e.g., the
WHO International Reference Preparation (IRP) for human Thyroid-Stimulating Hormone - IRP hTSH
80/558) and to establish traceable values for the calibrators. Reference materials prepared in a human
serum matrix are most useful for immunoassay calibration and harmonization. Most international
reference preparations are supplied as purified protein extracts, but a few are provided in a human serum
matrix. An example of such a human serum reference preparation for immunoassays is the serum protein
standard certified for 14 analytes.22
Definitive and reference methods are well-established, certified methods that are used for comparison and
calibration of other methods for reference value assessments. For analytes measured by immunoassays,
reference methods are considered the highest category achievable. However, the only proposed
immunoassay for this category is for digoxin. NCCLS document I/LA9—A Candidate Reference Method
for Serum Digoxin: A Model for Radioimmunoassay Reference Methods describes a working model that
delineates and exemplifies the extent of documentation required for the development, evaluation,
validation, and transferability of an immunoassay reference method. A few independent chemical
reference methods (e.g., isotope-dilution gas chromatography-mass spectrometry method for cortisol) are
available for some analytes that are measurable by immunoassays. These reference methods are very
useful to evaluate method bias and matrix influences among immunoassays and to guide method
improvements. When reference methods are not available, especially for proteins, pure analytes or
certified reference preparations are used for calibration.
The management of equivocal data from test comparisons is currently a matter of debate. Of particular
concern is the question of whether the equivocal results should be included as false-positive results, false-
negative results, or both, in the data analysis. A decision about this point is beyond the purview of this
document. (See NCCLS document I/LA21—Clinical Evaluation of Immunoassays) However, it is crucial
that a manufacturer indicate the percentage of equivocal results to be expected in actual patient testing
and how such results were interpreted in the data analysis used for product licensure, clearance, or
approval (e.g., inclusion or exclusion in calculations of positive and negative predictive values, sensitivity
and specificity, efficiency, etc.). If this information is not included in the package insert, the laboratorian
should request it from the manufacturer. Obviously, a test that has a high percentage of equivocal results
should be viewed with caution.
12.1 General
All quality control (QC) parameters must be determined within the laboratory in which they are to be
used. Carry out at least ten laboratory runs to establish temporary confidence limits for the QC samples.
Establish working limits after 20 runs and re-establish them after each additional 20 runs or a defined
interval. Include at least two QC samples at different levels that bracket the decision point in each
analytical run, in a matrix simulating the unknown sample to the extent possible, without contributing to
the variance of the assay (e.g., human serum matrix). (See NCCLS document C24—Statistical Quality
Control for Quantitative Measurements: Principles and Definitions.) The impact of the matrix on the
performance of the immunoassay should be determined. (See NCCLS document EP14—Evaluation of
Matrix Effects.) The matrix of any provided QC samples should be identified in the product insert.
For the pathological situations that the immunoassay is designed to study, statistically adequate sampling
is required in a specified and (preferably) matched reference sample group to permit reliable
differentiation from normal. State the distribution of the assay results from normal and disease reference
sample groups with a 95% confidence limit. Clearly state the overlap (if any) of disease and reference
interval sample groups.
In addition to the routine quality control sample system and plots, other parameters for immunoassays
should be monitored. Some suggested QC parameters for immunoassays include:
• the slope from the linear regression equation for the calibration curve;
• the y-intercept from the linear regression equation for the calibration curve;
• the corrected CPM for the total count tubes used for calibrators (RIA); and
• the coefficient of determination from the linear regression equation of the calibration curve.
For acceptable performance of the immunoassay, the mean value for the suggested parameters must fall
within designated and established limits. If the calculated value for any of these parameters falls outside
the defined limits, take appropriate remedial action based on the particular parameter. The mean value
and upper and lower 95% confidence limits are determined for each parameter during the process of
©
26 An NCCLS global consensus guideline. NCCLS. All rights reserved.
Volume 24 I/LA23-A
setting the QC sample limits. The use of additional parameters in quality control of immunoassays targets
the source of the problem and helps focus the corrective action for rapid response.
Several technical problems warrant special consideration for quality assurance of immunoassays:
heterophile antibody interference; high-dose hook effects; and the quality of calibration curve-fitting. In
contrast to the predictable and controlled impact of most technical errors, heterophilic antibodies can
cause significant interference problems in most immunoassays and are difficult to detect. Circulating
human heterophile antibodies have the ability to bind to immunoglobulins of other species, including the
animal species used to generate immunoassay reagent antibodies. Immunometric “sandwich” assays for
antigen detection are particularly susceptible to this interference. Heterophilic antibodies may either
bridge the capture and signal antibodies, thereby producing a false-positive signal and a falsely elevated
result; or they may bind exclusively to the capture or detection antibody, resulting in loss of signal and a
falsely low result. The inclusion of serum or immunoglobulins from the same species as the reagent
antibody is often effective in minimizing interference of this type. Unless an immunoassay is confirmed
to be free of this interference, heterophilic antibodies should be suspected if test results are not consistent
with the clinical picture. The laboratory should rerun questionable specimens using a different assay,
employ blocking steps or serial dilution, or separate immunoglobulins from analyte by chromatography
before the immunoassay.
Interference with antibody binding can also occur when extremely high concentrations of analyte in the
patient sample exceed the binding capacity of the capture and signal antibodies, making these antibodies
unavailable to form antibody-antigen complexes. This parameter is known as the “high-dose hook effect,”
and the phenomenon results in a severe underestimation of the measurand (analyte) concentration. This
problem mainly affects immunometric assays where the range of analyte concentrations is very high, such
as in assays for choriogonadotropin and many tumor markers. It is common practice to analyze samples in
such assays at several dilutions to check on the validity of the result.
Linear transforms of dose-response curves for immunoassays may indicate large dynamic ranges, but
caution should be used without a thorough understanding of these transformed calibration curves. The
reliability of immunoassay results depends on the accurate interpretation of assay data using an
appropriate calibration curve. Curve-fitting—the process of matching calibration data to a defined
mathematical function (e.g., a straight line)—has the potential to introduce significant bias or imprecision.
For example, severe nonuniformity of variance is introduced by the logit-log transformation, unless a
weighted regression technique is used to provide protection against poor replicates or outliers. Many
different curve-fitting algorithms are available, and the quality of the fit should be checked whenever a
new immunoassay is introduced. A simple procedure to check for curve-fitting bias is to back-calculate
calibrator values from their signal levels (using the equation for the fitted curve) and compare with
preassigned values. Stability and analyzer-to-analyzer variation are important problems associated with
master calibration curves that have been established by manufacturers and adjusted by users. Appropriate
QC parameters (e.g., goodness-of-fit residuals, small and random) should be monitored to ensure that
calibration integrity is maintained.
• assay procedure
• expected values
• bibliography
©
28 An NCCLS global consensus guideline. NCCLS. All rights reserved.
Volume 24 I/LA23-A
References
1
ISO. International Vocabulary of Basic and General Terms in Metrology. Geneva: International Organization for Standardization; 1993.
2
Flexer SB, ed. Random House Unabridged Dictionary. 2nd ed. CD-ROM; 1994.
3
ISO. In vitro diagnostic medical devices – Measurement of quantities in biological samples – Metrological traceability of values assigned
to calibrators and control materials. ISO 17511. Geneva: International Organization for Standardization; 2003.
4
ISO. In vitro diagnostic medical devices – Measurement of quantities in biological samples – Metrological traceability of values assigned
to catalytic concentration of enzymes in calibrators and control materials. ISO 18153. Geneva: International Organization for
Standardization; 2003.
5
ISO. Statistics – Vocabulary and symbols – Part 1: Probability and general statistical terms. ISO 3534-1. Geneva: International
Organization for Standardization; 1993.
6
WHO. Expert Committee on Biological Standardization. Glossary of Terms for Biological Substances Used for Texts of the Requirements.
WHO unpublished document BS/95.1793. World Health Organization: Geneva; 1995.
7
ISO. Accuracy (trueness and precision) of measurement methods and results – Part 1: General principles and definitions. ISO 5725-1.
Geneva: International Organization for Standardization; 1994.
8
Boutwell JH, ed. A National Understanding for the Development of Reference Materials and Methods for Clinical Chemistry. Washington,
DC: American Association of Clinical Chemistry; 1978.
9
IFCC. Provisional recommendation on quality control in clinical chemistry. Clin Chem. 1976; 22:1922-1932.
10
10 CFR 31. General domestic licenses for byproduct material. September 1, 1982.
11
Schuurs AHWM, Van Wecman BK. Enzyme immunoassays. Clin Chim Act. 1997;81:1-40.
12
Yolkin RH. Enzyme immunoassays for the detection of infectious agents in body fluids: Current limitations and future prospects. Rev Infect
Dis. 1982:4;35-68.
13
Ullman EF. Homogeneous immunoassays: EMIT and beyond. J Clin Ligand Assay. 1999:22;221-227.
14
Kricka LJ, Wild D. Signal generation and detection systems (excluding homogenous assays). In: Wild D, ed. The Immunoassay Handbook.
London: Nature Publishing Group. 2001:159-175.
15
Wang L, Gaigalas AK, Abbassi F, Marti GE, Vogt RF, Schwartz A. Quantitating fluorescence intensity from fluorophors: Practical use of
MESF values. J Res Natl Inst Stand Technol. 2002;107:339-353.
16
Kricka LJ. Chemiluminescent and bioluminescent techniques. Clin Chem. 1991:37;1472-1481.
17
Rongen HAH, Hoetelmans RMW, Bult A, Van Bennekom WP. Chemiluminescence and immunoassays. J Pharm Biomed Anal.
1994;12:433-62.
18
Blackburn GF, Haresh PS, Kenten JH, et al. Electrochemiluminescence detection for development of immunoassays and DNA probe
assays for clinical diagnostics. Clin Chem. 1991;37:1534-1539.
19
Martin RF. General Deming regression for estimating systematic bias and its confidence interval in method comparison studies. Clin
Chem. 2000;46:100-104.
20
Finney DJ. Statistical Methods in Biological Assay. 2nd ed. London: Griffin; 1964.
21
Dudley RA, Edwards P, Eklins RE, et al. Guidelines for immunoassay data processing. Clin Chem. 1985;31:1264-1271.
22
Reimer CB, Smith SJ, Hannon WH, et al. Progress towards international reference standards for human serum proteins. J Biol Stand.
1978;6:133-158.
23
21 CFR 809. In vitro diagnostic products for human use. April 1, 1997.
24
Directive 98/79/EC of The European Parliament and of the Council. Directive on in vitro diagnostic medical devices. Official Journal of the
European Communities. L331;October 27, 1998.
Additional References
42 CFR 493. Clinical Laboratory Improvement Amendments of 1988 (Revised). October 1, 2000;65:878-999.
Ashihara Y, Kasahara Y, Nakamura RM. Immunoassays and immunochemistry. In: Henry JB, ed. Clinical Diagnosis and Management by
Laboratory Methods. Philadelphia: W.B. Saunders Company. 2001:821-849.
Chard T. An Introduction to Radioimmunoassay and Related Techniques. 4th ed. Amsterdam: Elsevier; 1990.
Gosling JP, ed. Immunoassays: A Practical Approach. Oxford: Oxford University Press; 2000.
Price CP, Newman DJ. Principles and Practices of Immunoassays. 2nd ed. London: Macmillan; 1997.
Soini E, Hemmila I. Fluoroimmunoassays: Present status and key problems. Clin Chem. 1979;25:353-361.
Ullman F. Homogeneous fluorescence and enzyme immunoassays for proteins. Tokai J Exp Clin Med Supplement. 1979;4:7-32.
Wild D, ed. The Immunoassay Handbook. 2nd ed. New York: Nature Publishing Group. 2001:3-247.
NCCLS consensus procedures include an appeals process that is described in detail in Section 8 of
the Administrative Procedures. For further information, contact the Executive Offices or visit our
website at www.nccls.org.
General
1. This document is not useful as written. The document provides a broad and incomplete overview that lacks
clear, concise explanations and descriptions for a novice; plus, it is clearly inadequate for a knowledgeable user.
• The committee, during development of this guideline, focused on the core quality management issues for
immunoassays and does not agree with this general comment. The development of I/LA23-A was
implemented to replace NCCLS documents LA1-A2 and DI4-T by incorporating their pertinent
information and updating them to include new detection systems. I/LA23-A presents guidelines primarily
for immunoassays of macromolecular analytes. The factors likely to be important in achieving reliable
and reproducible results are emphasized. Use of this document should promote greater reliability and
comparability in immunoassay results. The definitions, information, and procedures necessary to
properly assess the quality of immunoassay systems are described. The range of applications for
immunoassays is extensive. The degree of variations in configurations is large and may involve a
hierarchy of antibodies used with different specificities for capture, separation, measurement, and dose
amplifications. A comprehensive coverage of the field of immunoassays is too large for the scope of this
document.
2. All parts of the document dealing with calibrators tell about traceability of the products to standards (reference
materials or methods); however, the document never mentions calculation of the components of uncertainty. For
this reason, maybe the word “calibrator” should be changed to the word “adjustor.”
• The committee believes (based on the definition of “calibrator”) that it is the correct term. The definition
for “calibrator” was added to Section 4.1 (formerly Section 3.1).
• Definitions of “qualitative assay,” “quantitative assay,” and “semiquantitative assay” have been added to
the definition of “assay” in Section 4.1.
4. Add a note following the definition of “clinical sensitivity” stating that it is the fraction of clinically true
positive classification divided by the sum of clinically true positive and clinically false negative.
• The suggested change has been made in Section 4.1 for the definition of “clinical sensitivity.”
5. Fourth bullet, second dashed sentence: What if there is more than one contaminant?
• The second dashed statement has been revised to read: “- antibody(s) specific for likely contaminant(s).”
6. Third bullet: A control antigen is prepared from a culture as well as the control specimen.
• The second sentence in the third bullet in Section 5.2.1 has been revised to read: “A control antigen is
prepared from a culture (as well as the control sample) not infected with antigen microorganism but one
that is otherwise identical in composition and purification protocol. Immunological reactivity with
contaminants can be accounted for by subtracting the control antigen reactivity.”
• The following sentence has been added to Section 6: “Standard and universal precautions should be
followed (see Section 3, Standard Precautions).
8. Manufacturers are responsible to notify users of the effect of high lipid or hemolysed samples on the test
performance.
• The commenter’s statement has been added to the third paragraph in Section 6.3.
• The third bullet in Section 8.1 has been revised to read: “quantitative and qualitative assays may be
rapid and can be automated.”
11. The first paragraph of this section refers to reference materials, and the second part tells about reference
methods. However, the last part of this paragraph refers, again, to reference materials. So, the sentence
beginning, “Since reference methods are not available, especially for proteins…” up to the end should be moved
to the end of the first paragraph.
Section 12, Quality Assurance and Control (Formerly Section 11, Quality Assurance)
12. The term “quality assurance” usually reflects how to guarantee that the product is satisfactory for the intended
use, and involves all phases (processes) of the medical laboratory. This document seems entirely focused on
detecting and minimizing analytical errors. Consequently, the title of this section should be “Quality Control.”
• For clarity, the title of Section 12 has been changed to “Quality Assurance and Control” and appropriate
revisions have been made in Section 12.2.
Section 13, Necessary Product Insert Information for Immunoassay Systems (Formerly Section 12)
13. Bullet “Specimen collection and preparation.” Are safety guidelines and universal precautions to be included?
• The following statement has been added to Section 13: “Standard and universal precautions, and safety
guidelines.”
14. Reagents, calibrators, and controls should also be identified with master lot and batch lot for traceability from
manufacturer through to end user. Reagents are used internationally and therefore a consensus date format
would be an advantage as the labeling can cause confusion between day-month-year and month-day-year.
• For clarity, Section 13 has been revised to include a sentence reading: “Date labeling can cause confusion
between day-month-year and month-day-year; therefore, the way that it is written should be clearly
explained.” A bulleted statement reading, “Identify reagents, controls, and calibrators by lot” has also
been added.
I/LA23-A addresses the quality system essentials (QSEs) indicated by an “X.” For a description of the other NCCLS
documents listed in the grid, please refer to the Related NCCLS Publications section at the end of the document.
Purchasing &
Improvement
Organization
Management
Management
Information
Satisfaction
Assessment
Facilities &
Occurrence
Documents
Equipment
& Records
Service &
Personnel
Inventory
Control
Process
Process
Safety
X X
C24 GP5
C28 M29
D12
D13
EP6
EP9
I/LA18
I/LA21 I/LA21 I/LA21 I/LA21
Adapted from NCCLS document HS1—A Quality System Model for Health Care.
Path of Workflow
A path of workflow is the description of the necessary steps to deliver the particular product or service that the
organization or entity provides. For example, GP26-A2 defines a clinical laboratory path of workflow which
consists of three sequential processes: preanalytic, analytic, and postanalytic. All clinical laboratories follow these
processes to deliver the laboratory’s services, namely quality laboratory information.
I/LA23-A addresses the clinical laboratory path of workflow steps indicated by an “X.” For a description of the
other NCCLS documents listed in the grid, please refer to the Related NCCLS Publications section at the end of the
document.
Preanalytic Analytic Postanalytic
Interpretation
Management
Test Request
Assessment
Laboratory
Collection
Specimen
Specimen
Specimen
Specimen
Transport
Post-test
Receipt
Review
Testing
Results
Patient
Report
X X X X X X
H3 H3
Adapted from NCCLS document HS1—A Quality System Model for Health Care.
C28-A2 How To Define and Determine Reference Intervals in the Clinical Laboratory; Approved Guideline –
Second Edition (2000). This document provides guidance for determining reference values and reference
intervals for quantitative clinical laboratory tests.
DI2-A2 Immunoprecipitin Analyses: Procedures for Evaluating the Performance of Materials—Second Edition;
Approved Guideline (1993) (Reaffirmed 1999). This guideline provides a description of and procedures for
evaluating the performance of materials used in immunoprecipitin analyses. It also includes a discussion on
specificity.
DI3-A Agglutination Analyses: Antibody Characteristics, Methodology, Limitations, and Clinical Validation;
Approved Guideline (1993) (Reaffirmed 1999). This guideline describes the specificity of antibodies and
antigens for agglutination techniques, guidance labeling information, and characteristics and limitations of
agglutination methods.
EP6-A Evaluation of the Linearity of Quantitative Measurement Procedure: A Statistical Approach; Approved
Guideline (2003). This document provides guidelines for characterizing the linearity of a method during a
method evaluation; for checking linearity as part of routine quality assurance; and for determining and stating a
manufacturer’s claim for linear range.
EP9-A2 Method Comparison and Bias Estimation Using Patient Samples; Approved Guideline—Second Edition
(2002). This document addresses procedures for determining relative bias between two clinical methods or
devices; and for the design of a method comparison experiment using split patient samples and data analysis.
GP5-A2 Clinical Laboratory Waste Management; Approved Guideline—Second Edition (2002). Based on U.S.
regulations, this document provides guidance on safe handling and disposal of chemical, infectious, radioactive,
and multihazardous wastes generated in the clinical laboratory.
HS1-A A Quality System Model for Health Care; Approved Guideline (2002). This document provides a model for
healthcare service providers that will assist with implementation and maintenance of effective quality systems.
H3-A5 Procedure for the Collection of Diagnostic Blood Specimens by Venipuncture; Approved Standard—
Fifth Edition (2003). This document provides procedures for the collection of diagnostic specimens by
venipuncture, including line draws, blood culture collection, and venipuncture in children. It also includes
recommendations on order of draw.
I/LA18-A2 Specifications for Immunological Testing for Infectious Diseases; Approved Guideline – Second Edition
(2001). This guideline outlines: specimen requirements; performance criteria; algorithms for the potential use
of sequential or duplicate testing; recommendations for intermethod comparisons of immunological test kits for
detecting infectious disease; and specifications for development of reference materials.
I/LA21-A Clinical Evaluation of Immunoassays; Approved Guideline (2002). This guideline provides
recommendations on designing trials that are appropriate for evaluating both the safety and effectiveness of
immunoassays. It is a valuable resource in determining the necessary steps in designing an evaluation for new
methods, new applications for existing methods, or variations on existing methods.
M29-A2 Protection of Laboratory Workers from Occupationally Acquired Infections; Approved Guideline—
Second Edition (2001). This document provides guidance on: the risk of transmission of hepatitis viruses and
human immunodeficiency viruses in any laboratory setting: specific precautions for preventing the laboratory
transmission of blood-borne infection from laboratory instruments and materials; and recommendations for the
management of blood-borne exposure.
NRCSL13-A The Reference System for the Clinical Laboratory: Criteria for Development and Credentialing of
Methods and Materials for Harmonization of Results; Approved Guideline (2000). This document
provides procedures for developing and evaluating methods and materials to provide a harmonized clinical
measurement system.
*
Proposed- and tentative-level documents are being advanced through the NCCLS consensus process; therefore, readers should
refer to the most recent editions.
An NCCLS global consensus guideline. ©NCCLS. All rights reserved. 35
Number 16 NCCLS
NOTES
NOTES
NOTES
NOTES
Donna M. Meyer, Ph.D., Susan Blonshine, RRT, RPFT, FAARC Carolyn D. Jones, J.D., M.P.H.
President TechEd AdvaMed
CHRISTUS Health
Wayne Brinster J. Stephen Kroger, M.D., MACP
Thomas L. Hearn, Ph.D., BD COLA
President Elect
Centers for Disease Control and Prevention Kurt H. Davis, FCSMLS, CAE Willie E. May, Ph.D.
Canadian Society for Medical Laboratory Science National Institute of Standards and Technology
Emil Voelkert, Ph.D.,
Secretary Mary Lou Gantzer, Ph.D. Gary L. Myers, Ph.D.
Roche Diagnostics GmbH Dade Behring Inc. Centers for Disease Control and Prevention