TC 305
TC 305
TC 305
SOFTWARE PACKAGES
AUTOMATIC TEST FUNCTION
The Beamex QCAL System calibrators communicate with
Fully automatic test function for transmitters and other electrical PC's with either DOS or Windows based true relational calibration
converters. The automatic step function tests the instrument under database software. The powerful calibration software allows com-
test and stores the calibration results - just connect the instrument plete easy-to-use computer-aided ISO 9000 compatible calibration
to the calibrator. management.
APPLICATIONS
TC
RTD mA
Calibration of temperature
transmitters
The upper display indicates the simulated T/C or RTD
signal.
The lower display indicates the transmitter output error
in % of span (or in selected eng. unit).
mV mV
V V
mA mA
Display for measurement/ Keys (yellow) for selecting Keys (orange) for selecting mV, V, mA or
simulation section temperature/electrical mode, sensor ohm range for measurement/simulation
(T/C, RTD, mV, V, mA, ohm) type, °C or °F, cold junction section, display mode (eng. unit / %)
compensation mode (INT, EXT, MAN) and for setting the transmitter input
range values (0%, 100%)
Display for measurement Keys (blue) for selecting mV, V, mA range or Keys for selecting
section (mV, V, mA, switch) switch test function for measurement and setting
section, display mode (eng. unit / error % / special functions
%) and for setting the transmitter output and configuration
range values (0%, 100%)
OPERATING CONDITIONS
Storage temperature: -20 to +60 oC, -4 to +140 °F
Operating temperature: -10 to +50 oC, +14 to +122 °F
Humidity range: 0 to 80 % R.H. non-condensing
Temperature coefficient: ± 0.002 % RDG/ oC, ± 0.001 % RDG/°F
{ }
E -270...+1000 °C 0.1 °C
J -210...+1200 °C 0.1 °C
Measurement K -270...+1372 °C 0.1 °C
and Simulation N -270...+1300 °C 0.1 °C ± (0.02 % RDG + 0.003 mV) 1)
R -50 ...+1768 °C 0.1 °C
S -50 ...+1768 °C 0.1 °C
T -270... +400 °C 0.1 °C
Measurement RTD (Pt100) -200... +850 °C 0.01/0.05 °C ± (0.02 % RDG + 0.01 ohm) 2) 3)
Simulation RTD (Pt100) -200... +850 °C 0.01 °C ± 0.05 ohm 3) 4) 5)
Measurement mV -10 ... +120 mV 0.002 mV, 0.005 mV ± (0.02 % RDG + 0.003 mV)
Output mV -10 ... +120 mV 0.001 mV ± (0.02 % RDG + 0.003 mV)
Measurement V -1,3... +12 V 0.0002 V, 0.0005 V ± (0.02 % RDG + 0.0004 V)
Output V -1,3... +12 V 0.0001 V ± (0.02 % RDG + 0.0004 V)
Measurement mA -5 ... +60 mA 0.001 mA, 0.002 mA ± (0.02 % RDG + 0.002 mA)
Output mA 0 ... +60 mA 0.001 mA ± (0.02 % RDG + 0.002 mA)
Measurement ohm 0 ... 4000 ohm 0.005 to 0.2 ohm ± (0.02 % RDG + 0.010 ohm) 2)
Output ohm 5 ... 392 ohm 0.01 ohm ± 0.05 ohm 4) 5)
392...4000 ohm 0.1 ohm ± 0.5 ohm 4) 6)
1) Linearity error ± 0.05 °C, for internal ref. junction: ± 0.1°C STANDARDS:
2) 4 wire connection (add 0.010 ohm if 3 wires are used) B, E, J, K, N, R, S, T conform to international
3) Linearity error ± 0.02 °C standard IEC 584-1 and Pt-100 conforms
4) At I meas. = + 1 mA ± 10% to IEC 751.
5) I meas. range = ± (0.1 to 11 mA) Both temperature scalings IPTS-68 and ITS-90
6) I meas. range = ± (0.1 to 1.1 mA) are included (selectable).
MEASUREMENT SECTION
FUNCTION RANGE RESOLUTION UNCERTAINTY
Measurement mV -40 ... +130 mV 0.005 mV ± (0.02 % RDG + 0.01 mV)
Measurement V -5 ... +48 V 0.2, 0.5, 0.002 V ± (0.02 % RDG + 0.0004 V)
Measurement mA -5 ... +60 mA 0.001 mA, 0.002 mA ± (0.02 % RDG + 0.002 mA)
Note: Specifications are subject to change without prior notice. Complete specifiations are available from Beamex.
ORDERING INFORMATION
OPTIONS
• D: temp. switch function • E: 24 V DC loop supply
• S: customer defined sensors • TCAL Expansion
ACCESSORIES
• RS1A: RS232 Interface (non isolated) • RS2A: RS232 interface (isolated)
• FM1: frequency module (0...50,000 Hz) • CALDB3/HISDB3, QM6
Beamex's powerful calibration database software packages CALDB3/HISDB3 (DOS) and QM6 (Windows) allow ISO 9000
compatible calibration management and documentation. The history function of these software packages make trend
analyzing over the whole intrument lifetime easy.
PRESSURE/ELECTRICAL
TEMPERATURE
PRESSURE
ELECTRICAL
Specifications subject to change without notice. Ask your local sales representative or Beamex office for more technical details.
Representative:
Oy Beamex Ab Beamex, Inc. Beamex UK
P.O. Box 5 2220 Northwest Parkway Highcliffe Road
FIN-68601 PIETARSAARI Suite 185 Hamilton Industrial Park
Finland Marietta, GA 30067, USA Leicester LE5 1TY, UK
Phone +358 6 784 0111 Phone (770) 951-1927 Phone (0116) 246 1635
Fax +358 6 784 0404 (800) 888-9892 Facsimile (0116) 276 1874
E-mail sales@beamex.com Fax (770) 951-1928
Internet http://www.beamex.com E-mail beamex@mindspring.com