Technical Information
Cerabar PMP23
Process pressure measurement
Application
The Cerabar is a pressure transducer for the measurement of absolute and gauge
pressure in gases, vapors, liquids and dust for applications with hygienic
requirements. The Cerabar can be used internationally thanks to a wide range of
approvals and process connections.
Your benefits
• High reproducibility and long-term stability
• Reference accuracy: up to 0.3%
• Customized measuring ranges
– Turn down up to 5:1
– Sensor for measuring ranges up to 40 bar (600 psi)
• Housing and process isolating diaphragm made of 316L
• IP69 protection
• Fully-welded process connections
• Suitable for CIP/SIP cleaning
Cerabar PMP23
Table of contents
2 Endress+Hauser
Cerabar PMP23
Endress+Hauser 3
Cerabar PMP23
Document information
Document function The document contains all the technical data on the device and provides an overview of the
accessories and other products that can be ordered for the device.
Symbol Meaning
DANGER!
DANGER This symbol alerts you to a dangerous situation. Failure to avoid this situation will result in
serious or fatal injury.
WARNING!
WARNING This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in
serious or fatal injury.
CAUTION!
CAUTION This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in
minor or medium injury.
NOTE!
NOTICE This symbol contains information on procedures and other facts which do not result in
personal injury.
Electrical symbols
Symbol Meaning
Permitted
Procedures, processes or actions that are permitted.
Forbidden
Procedures, processes or actions that are forbidden.
Tip
Indicates additional information.
Reference to documentation
A Reference to page
Reference to graphic
Visual inspection
Symbols in graphics
Symbol Meaning
1. , 2. , 3. … Series of steps
A, B, C, ... Views
4 Endress+Hauser
Cerabar PMP23
Brief Operating Instructions (KA): getting the 1st measured value quickly
The Brief Operating Instructions contain all the essential information from incoming acceptance to
initial commissioning.
The nameplate provides information on the Safety Instructions (XA) that are relevant for the
device.
Endress+Hauser 5
Cerabar PMP23
p
0
LRL LRV URV URL MWP OPL
A0029505
1 OPL The OPL (over pressure limit = sensor overload limit) for the measuring device
depends on the lowest-rated element, with regard to pressure, of the selected
components, i.e. the process connection has to be taken into consideration in
addition to the measuring cell. Also observe pressure-temperature dependency. For
the relevant standards and additional notes, see the "Pressure specifications" section
→ 19 .
The OPL may only be applied for a limited period of time.
2 MWP The MWP (maximum working pressure) for the sensors depends on the lowest-
rated element, with regard to pressure, of the selected components, i.e. the process
connection has to be taken into consideration in addition to the measuring cell.
Also observe pressure-temperature dependency. For the relevant standards and
additional notes, see the "Pressure specifications" section → 19 .
The MWP may be applied at the device for an unlimited period.
The MWP can also be found on the nameplate.
p - Pressure
6 Endress+Hauser
Cerabar PMP23
A0029545
1 Calibrated/adjusted span
2 Zero point-based span
3 URL sensor
Example
URL
TD =
|URV - LRV|
Endress+Hauser 7
Cerabar PMP23
1
2
3
4
p
A0016448
1 2 3
A0021926
Device features
PMP23
Measuring ranges From –400 to +400 mbar (–6 to +6 psi) to –1 to +40 bar (–15 to +600 psi)
OPL (depends on the measuring range) Max. 0 to +160 bar (0 to +2 400 psi)
8 Endress+Hauser
Cerabar PMP23
PMP23
Supply voltage 10 to 30 V DC
Output 4 to 20 mA
Options • Ex approvals
• Min. alarm current setting
• 3.1 Material certificates
• EHEDG/3A approvals
• Certificate of calibration
• Weld-in adapter
• IP69
Product design
B Cable
C- 1 M12 plug
Housing cap made of plastic
C- 2 M12 plug
A0027231 A0027232 A0021987 A0027289 For Ex ec and IP69: metal housing cap
Metal housing cap can also be ordered as an option.
D D Housing
E Process connection (sample illustration)
A0027227
System integration The device can be given a tag name (max. 8 alphanumeric characters).
1)
Description Option
Endress+Hauser 9
Cerabar PMP23
Input
Measured variable Measured process variable
PMP23: gauge pressure or absolute pressure
[bar (psi)] [bar (psi)] [bar (psi)] [bar (psi)] [bar (psi)]
40 bar (600 psi) 4) PMP23 –1 (–15) +40 (+600) 8 (120) 100 (1500) 160 (2400) 0 to 40 bar (0 to 600 psi) 1S
400 mbar (6 psi) 4) PMP23 0 (0) 0.4 (+6) 0.4 (6) 1 (15) 1.6 (24) 0 to 400 mbar (0 to 6 psi) 2F
4)
1 bar (15 psi) PMP23 0 (0) 1 (+15) 0.4 (6) 2.7 (40.5) 4 (60) 0 to 1 bar (0 to 15 psi) 2H
4)
2 bar (30 psi) PMP23 0 (0) 2 (+30) 0.4 (6) 6.7 (100.5) 10 (150) 0 to 2 bar (0 to 30 psi) 2K
4)
4 bar (60 psi) PMP23 0 (0) 4 (+60) 0.8 (12) 10.7 (160.5) 16 (240) 0 to 4 bar (0 to 60 psi) 2M
4)
10 bar (150 psi) PMP23 0 (0) 10 (+150) 2 (30) 25 (375) 40 (600) 0 to 10 bar (0 to 150 psi) 2P
4)
40 bar (600 psi) PMP23 0 (0) +40 (+600) 8 (120) 100 (1500) 160 (2400) 0 to 40 bar (0 to 600 psi) 2S
1) Highest turn down that can be set at the factory: 5:1. The turn down is preset and cannot be changed.
2) Other measuring ranges (e.g. –1 to +5 bar (–15 to 75 psi)) can be ordered with customer-specific settings (see the Product Configurator, order
code for "Calibration; Unit" option "J"). It is possible to invert the output signal (LRV = 20 mA; URV = 4 mA). Prerequisite: URV < LRV
3) Product Configurator, order code for "Sensor range"
4) Vacuum resistance: 0.01 bar (0.145 psi) abs
Maximum turn down which can be ordered for absolute pressure and gauge pressure sensors
Device Range 400 mbar (6 psi) 1 bar (15 psi) 2 bar (30 psi)
6 bar (90 psi) 4 bar (60 psi)
16 bar (240 psi) 10 bar (150 psi)
25 to 40 bar (375 to 600 psi)
10 Endress+Hauser
Cerabar PMP23
Output
Output signal Description Option 1)
4 to 20 mA (2-wire) 1
Load (for 4 to 20 mA In order to guarantee sufficient terminal voltage in two-wire devices, a maximum load resistance RL
devices ) (including line resistance) must not be exceeded, depending on the supply voltage UB of the supply
unit.
R L max
[W ]
1068
10 20 30 UB
[V]
1
A0029452
1 Power supply 10 to 30 V DC
2 RLmax maximum load resistance
UB Supply voltage
Signal on alarm 4 to 20 mA The response of the output to error is regulated in accordance with NAMUR NE43.
Factory setting MAX alarm: >21 mA
alarm current
Dead time, time constant Presentation of the dead time and the time constant:
I
100 %
90 %
63 %
t1 t2 t
t3
A0019786
Endress+Hauser 11
Cerabar PMP23
Dead time (t1) [ms] Time constant (T63), t2 [ms] Time constant (T90), t3 [ms]
6 ms 10 ms 15 ms
12 Endress+Hauser
Cerabar PMP23
Power supply
LWARNING
Limitation of electrical safety due to incorrect connection!
‣ In accordance with IEC/EN61010 a separate circuit breaker must be provided for the device .
‣ When using the measuring device in hazardous areas, installation must comply with the
corresponding national standards and regulations and the Safety Instructions or Installation or
Control Drawings.
‣ All explosion protection data are given in separate documentation which is available upon
request. The Ex documentation is supplied as standard with all devices approved for use in
explosion hazardous areas.
‣ Protective circuits against reverse polarity, HF influences and overvoltage peaks are integrated.
‣ The device must be operated with a 500 mA fine-wire fuse (slow-blow).
1 brown = L+
2 blue = L-
3 green/yellow = ground connection
(a) reference air hose
4 to 20 mA output PMP23 10 to 30 V DC
Current consumption and Number of wires Device Normal operation Alarm signal 1)
alarm signal
2 PMP23 ≤ 26 mA > 21 mA
Endress+Hauser 13
Cerabar PMP23
Cable specification For valve plug: < 1.5 mm2 (16 AWG) and Ø3.5 to 6.5 mm (0.14 to 0.26 in)
Residual ripple The device operates within the reference accuracy up to ±5 % of the residual ripple of the supply
voltage, within the permitted voltage range.
Overvoltage protection The device does not contain any special elements to protect against overvoltage ("wire to ground").
Nevertheless the requirements of the applicable EMC standard EN 61000-4-5 (testing voltage 1kV
EMC wire/ground) are met.
14 Endress+Hauser
Cerabar PMP23
Measuring uncertainty for The smallest extended uncertainty of measurement that can delivered by our standards is:
small absolute pressure • in range 1 to 30 mbar (0.0145 to 0.435 psi): 0.4 % of reading
measuring ranges • in range < 1 mbar (0.0145 psi): 1 % of reading.
Reference accuracy The reference accuracy contains the non-linearity [DIN EN 61298-2 3.11] including the pressure
hysteresis [DIN EN 61298-23.13] and non-repeatability [DIN EN 61298-2 3.11] in accordance with
the limit point method as per [DIN EN 60770].
Thermal change of the zero Measuring cell –10 to +85 °C (+14 to +185 °F) +85 to +100 °C (+185 to +212 °F)
output and the output span
% of the calibrated span for TD 1:1
% of URL
Switch-on time ≤2 s
Installation
Installation conditions • No moisture may enter the housing when installing or operating the device, or when establishing
the electrical connection.
• Point the cable and connector downwards where possible to prevent moisture from entering (e.g.
rain or condensation water).
Endress+Hauser 15
Cerabar PMP23
Influence of the installation Any orientation is possible. However, the orientation may cause a zero point shift i.e. the measured
position value does not show zero when the vessel is empty or partially full.
A B C
A0024708
Type Process isolating diaphragm Process isolating diaphragm Process isolating diaphragm
axis is horizontal (A) pointing upwards (B) pointing downwards (C)
PMP23 Calibration position, no effect Up to +4 mbar (+0.058 psi) Up to –4 mbar (–0.058 psi)
A0021904
1 Device
2 Shutoff device
3
A0024395
1 Device
2 Shutoff device
3 Siphon
16 Endress+Hauser
Cerabar PMP23
2 1
A0024399
1 Device
2 Shutoff device
Level measurement
A0024405
Endress+Hauser 17
Cerabar PMP23
Environment
Ambient temperature range Device Ambient temperature range 1)
PMP23 Devices for hazardous areas: –40 to +70 °C (–40 to +158 °F)
PMP23 Class 4K4H Air temperature: –20 to +55 °C (–4 to +131 °F),
relative humidity: 4 to 100 %
satisfied according to DIN EN 60721-3-4 (condensation is possible)
18 Endress+Hauser
Cerabar PMP23
Process
Process temperature range Device Process temperature range
for devices with metallic
process isolating diaphragm PMP23 –10 to +100 °C (+14 to +212 °F)
PMP23 At +135°C (+275 °F) for a maximum of one hour (device in operation but not within
Sterilization in place (SIP) measuring specification)
Endress+Hauser 19
Cerabar PMP23
Mechanical construction
Design, dimensions Device height
The device height is calculated from
• the height of the electrical connection
• the height of the housing and
• the height of the individual process connection.
The individual heights of the components are listed in the following sections. To calculate the device
height simply add up the individual heights of the components. Where applicable also take into
consideration the installation distance (space that is used to install the device). You can use the
following table for this purpose:
A0027268
Electrical connection
A B C 35 (1.38) D
25 (1.0)
25 (1.0)
28 (1.1)
45 (1.77)
M12x1 M12x1
r ³ 102 (4.02)
52 (2.05)
A0024426 A0024427
A0024428
A0024429
1)
Position Description Material Weight kg (lbs) Device Option
A M12 plug IP65/67 Housing cap made of plastic 0.012 (0.03) PMP23 M
(Additional dimensions → 35) Plug connector with cable
can be ordered as an accessory → 35
B M12 plug IP66/67 Housing cap made of metal 0.030 (0.07) PMP23 In the case of Ex ec type of protection,
the housing cap is made of metal.
Can be ordered separately via option "N".
20 Endress+Hauser
Cerabar PMP23
Housing
64 (2.52)
A0025115
Endress+Hauser 21
Cerabar PMP23
A B
36 36
7 (0.28)
48 (1.89)
48 (1.89)
7 (0.28)
29.5 (1.16)
26.8 (1.06)
ø25.4 (1)
22.3 (0.88)
10.9 (0.43)
ø29 (1.14)
8.8 (0.35)
Engineering unit mm (in). Diameter of process isolating diaphragm: 17.2 mm (0.68 in)
kg (lbs)
A Thread ISO 228 G 1" A Metal joint 316L 0.270 (0.60) CRN WQJ
2)
B Thread ISO 228 G 1" A Seal established via O-ring. 316L 0.270 (0.60) EHEDG, 3A , CRN WSJ
VMQ O-ring is enclosed with QJ and QK accessories.
22 Endress+Hauser
Cerabar PMP23
7 (0.28)
ø24 (0.94)
32
41 (1.61)
4 (0.16)
13.8 (0.54)
M24x1.5
17 (0.67)
2 ø31.5 (1.24)
A0021969
Engineering unit mm (in). Diameter of process isolating diaphragm: 17.2 mm (0.68 in)
kg (lbs)
M24 x 1.5 EPDM O-ring (2), pre-mounted 316L 0.150 (0.33) EHEDG, 3A, CRN X2J
M24 x 1.5 FKM O-ring (2), pre-mounted 316L 0.150 (0.33) EHEDG, 3A, CRN X3J
Endress+Hauser 23
Cerabar PMP23
Hygienic connections
28 (1.1)
28 (1.1)
10 (0.39)
10 (0.39)
3 (0.12)
3 (0.12)
ø27 (1.06) ø39 (1.54)
ø43.4 (1.71) ø55.4 (2.18)
A0021956
A0021957
C ø61 (2.4) D
4 (0.16)
28 (1.1)
ø4 (0.16)
28 (1.1)
11 (0.43)
3 (0.12)
A0021958
Engineering unit mm (in). Diameter of process isolating diaphragm: 17.2 mm (0.68 in)
PN kg (lbs)
24 Endress+Hauser
Cerabar PMP23
Hygienic connections
A B
28 (1.1)
28 (1.1)
ø50.5 (1.99) ø64 (2.52)
A0021976 A0021979
28 (1.1)
28 (1.1)
12.3 (0.48)
19 (0.75)
12.3 (0.48)
ø84 (3.31)
19 (0.75)
ø66 (2.6)
A0021981 A0021980
Engineering unit mm (in). Diameter of process isolating diaphragm: 17.2 mm (0.68 in)
PN kg (lbs)
A Tri-Clamp ISO 2852 DN 25 – DN 38 (1 ½"), DIN32676 DN25-38 3A, EHEDG, CRN 40 316L 0.160 (0.35) 3CJ
B Tri-Clamp ISO 2852 DN 40 – DN 51 (2"), DIN32676 DN50, 3A, EHEDG, CRN 40 316L 0.230 (0.51) 3EJ
EHEDG, 3A
C Varivent F pipe DN25-32 3A, EHEDG, CRN 40 316L 0.350 (0.77) 41J
D Varivent N pipe DN40-162 3A, EHEDG, CRN 40 316L 0.630 (1.39) 42J
Endress+Hauser 25
Cerabar PMP23
Hygienic connections
ø43.8 (1.72)
ø34 (1.34)
7 (0.28)
18.8 (0.74)
57 (2.24)
45 (1.77)
ø43.5 (1.71)
A0021986
Engineering unit mm (in). Diameter of process isolating diaphragm: 17.2 mm (0.68 in)
PN kg (lbs)
A Universal adapter 44 mm EPDM form seal, pre-mounted. 3A, EHEDG, CRN 10 316L 0.730 (1.61) 52J
26 Endress+Hauser
Cerabar PMP23
Process connections
• Endress+Hauser supplies a threaded connection made of stainless steel in accordance with AISI
316L (DIN/ EN material number 1.4404 or 1.4435). With regard to their stability-temperature
property, the materials 1.4404 and 1.4435 are grouped together under 13E0 in EN 1092-1: 2001
Tab. 18. The chemical composition of the two materials can be identical.
• "Clamp connections" and "Hygienic process connections": AISI 316L (DIN/EN material number
1.4435)
Description Material
Metal process isolating diaphragm AISI 316L (DIN/EN material number 1.4435)
Seals
See the specific process connection.
Endress+Hauser 27
Cerabar PMP23
1 2
3
3 6
5
A0027299
4
A0025117
7 8 9
A0025116
4 Connection PPSU
Filling oil
28 Endress+Hauser
Cerabar PMP23
Operability
Plug-on display PHX20 No display or other operation facility is required to operate the device. However, devices with a valve
(optional) plug can be fitted with the optional local display PHX20.
Description Option 1)
A 1-line liquid crystal display (LCD) is used. The local display shows measured values, fault messages
and information messages. The device display can be turned in 90° steps. Depending on the
orientation of the device, it is therefore easy to read the measured values.
Technical data
Electrical connection: To transmitter with 4 to 20 mA output and elbow plug DIN 43 650,
reverse polarity protection
Display power supply: Not required, powered automatically from the power loop
Programming: Via 2 keys, menu-guided, display range scaling, decimal point, damping,
error message
Degree of protection: IP 65
Endress+Hauser 29
Cerabar PMP23
RoHS The measuring system complies with the substance restrictions of the Restriction on Hazardous
Substances Directive 2011/65/EU (RoHS 2).
RCM-Tick marking The supplied product or measuring system meets the ACMA (Australian Communications and Media
Authority) requirements for network integrity, interoperability, performance characteristics as well
as health and safety regulations. Here, especially the regulatory arrangements for electromagnetic
compatibility are met. The products are labelled with the RCM- Tick marking on the name plate.
A0029561
EAC conformity The PMP21 and PMP23 devices meet the legal requirements of the applicable EAC directives. These
are listed in the corresponding EAC Declaration of Conformity along with the standards applied.
Endress+Hauser confirms successful testing of the device by affixing to it the EAC mark.
Safety Instructions (XA) Safety Instructions (XA) are supplied with the device depending on the approval. These instructions
are an integral part of the Operating Instructions.
The nameplate provides information on the Safety Instructions (XA) that are relevant for the
device.
Suitability for hygienic All materials in contact with foodstuffs comply with framework regulation (EC) 1935/2004. The
processes device is available with hygienic process connections (overview: see order code).
30 Endress+Hauser
Cerabar PMP23
LCAUTION
Contamination in the process!
Risk of contamination if incorrect seals and parts are used!
‣ To avoid the risk of contamination, when installing the device comply with the design principles
of EHEDG, Guideline 37 "Hygienic Design and Application of Sensors" and Guideline 16 "Hygienic
Pipe Couplings".
‣ Suitable assemblies and seals must be used to ensure hygienic design in accordance with 3-A SSI
and EHEDG specifications.
‣ The leak-proof connections can be cleaned with the cleaning methods typical of this industry
(CIP and SIP). Attention must be paid to the pressure and temperature specifications of the
sensor and process connections for CIP and SIP processes (clean in place/sterilize in place).
The seamless connections can be cleaned of all residue using any of the typical cleaning
methods within this industry.
74 -
TYPE EL
CLASS I
OCTOBER 2014
A0025304
Pressure Equipment Pressure equipment with allowable pressure ≤ 200 bar (2 900 psi)
Directive 2014/68/EU (PED)
Pressure equipment (with a maximum allowable pressure PS ≤ 200 bar (2 900 psi)) can be classified
as pressure accessories in accordance with Pressure Equipment Directive 2014/68/EU. If the
maximum allowable pressure is ≤ 200 bar (2 900 psi) and the pressurized volume of the pressure
equipment is ≤ 0.1 l, the pressure equipment is subject to the Pressure Equipment Directive (cf.
Pressure Equipment Directive 2014/68/EU, Article 4, point 3). The Pressure Equipment Directive
only requires that the pressure equipment shall be designed and manufactured in accordance with
the "sound engineering practice of a Member State".
Reasons:
• Pressure Equipment Directive (PED) 2014/68/EU Article 4, point 3
• Pressure equipment directive 2014/68/EU, Commission's Working Group "Pressure", Guideline
A-05 + A-06
Note:
A partial examination shall be performed for pressure instruments that are part of safety equipment
for the protection of a pipe or vessel from exceeding allowable limits (safety accessory in accordance
with Pressure Equipment Directive 2014/68/EU, Article 2, point 4).
Manufacturer declarations Depending on the desired configuration, the following documents can be ordered additionally with
the device:
• FDA conformity
• TSE-free: materials free from animal origin
• Regulation (EC) No. 2023/2006 (GMP)
• Regulation (EC) No. 1935/2004 on materials and articles intended to come into contact with food
Endress+Hauser 31
Cerabar PMP23
1
3
A0031778
Other standards and The applicable European guidelines and standards can be found in the relevant EU Declarations of
guidelines Conformity. The following were also applied:
DIN 16086:
Electrical pressure measuring instruments, pressure sensors, pressure transmitters, pressure
measuring instruments, concepts, specifications on data sheets
Procedure for writing specifications in data sheets for electrical pressure measuring instruments,
pressure sensors and pressure transmitters.
EN 61326-X:
EMC product family standard for electrical equipment for measurement, control and laboratory use.
EN 60529:
Degrees of protection provided by enclosures (IP code)
CRN approval A CRN approval is available for some device versions. A CRN-approved process connection with a
CSA approval must be ordered for a CRN-approved device. The CRN-approved devices are assigned
the registration number 0F18141.5C.
Ordering information: Product Configurator, order code for "Process connection" (the CRN process
connections are indicated appropriately in the "Mechanical construction" section.)
32 Endress+Hauser
Cerabar PMP23
Sensor range; % A
PMP23 3.1 Material documentation, wetted metal parts, EN10204-3.1 inspection certificate JA
PMP23 Surface finish measurement ISO4287/Ra, wetted metal parts, inspection certificate KB
Ordering information
Detailed ordering information is available from the following sources:
• In the Product Configurator on the Endress+Hauser website: www.endress.com -> Click "Corporate"
-> Select your country -> Click "Products" -> Select the product using the filters and search field ->
Open product page -> The "Configure" button to the right of the product image opens the Product
Configurator.
• From your Endress+Hauser Sales Center: www.addresses.endress.com
Product Configurator - the tool for individual product configuration
• Up-to-the-minute configuration data
• Depending on the device: Direct input of measuring point-specific information such as
measuring range or operating language
• Automatic verification of exclusion criteria
• Automatic creation of the order code and its breakdown in PDF or Excel output format
• Ability to order directly in the Endress+Hauser Online Shop
Endress+Hauser 33
Cerabar PMP23
Accessories
Weld-in adapter Various weld-in adapters are available for installation in vessels or pipes.
PMP23 Weld-in adapter G1, 316L, 3.1, conical metal joint, QF 52010171
EN10204-3.1 material, inspection certificate
PMP23 Weld-in adapter G1, 316L, 3.1, silicone O-ring seal, QK 52011896
EN10204-3.1 material, inspection certificate
If installed horizontally and weld-in adapters with a leakage hole are used, ensure that the leakage
hole is pointing down. This allows leaks to be detected as quickly as possible.
Process adapter M24 The following process adapters can be ordered for the process connections with order option X2J and
X3J:
Device Description Order number Order number with inspection certificate 3.1 EN10204
34 Endress+Hauser
Cerabar PMP23
(0.8)
ø20 53 (2.09)
A0024475
³40 (1.57)
A0024476
28
(1.1)
(1.38)
35
20 (0.79)
A0024478
³40 (1.57)
A0024477
Endress+Hauser 35
Cerabar PMP23
Documentation
Field of Activities Pressure measurement, powerful instruments for process pressure, differential pressure, level and
flow:
FA00004P/00/EN
Safety Instructions (XA) Safety Instructions (XA) are supplied with the device depending on the approval. These instructions
are an integral part of the Operating Instructions.
The nameplate provides information on the Safety Instructions (XA) that are relevant for the
device.
*71353707* 71353707
www.addresses.endress.com