Location via proxy:   [ UP ]  
[Report a bug]   [Manage cookies]                

1.micros 60

Download as pdf or txt
Download as pdf or txt
You are on page 1of 2

ABX Micros 60

Technical specifications

PHYSICAL SPECIFICATION SOFTWARE SPECIFICATIONS


• Dimensions & Weight: Height Width Length Weight • Data Processing :
Analyzer 42 cm 36 cm 32cm 14 Kg - LCD screen with 2 lines of 40 alphanumerical characters
16.5 in 14.2 in 12.6 in 30.9 lb - Capacity: Last result + graphics or 78 results
• Printer: 80 column dot matrix printer with optional memory smart card
• Throughput: - Mother board: 68HC11 microprocessor
- Up to 60 samples/hour in Open Tube - Transmit results & QC to LIS
- Up to 55 samples/hour in Close Tube - Unidirectional connection RS 232
• Sound Pressure Level: <60 dBa • Quality control management :
• Operating Temperature & Humidity: - Optional smart card based QC (31 controls x 3 levels)
16 - 34°C (61 - 93°F) room temperature
Maximum relative humidity 80% for temperature up to 31°C (88°F) PARAMETERS & PERFORMANCE DATA

Subject to technical modifications / HAN726CEN / © 2012 HORIBA Medical - HORIBA ABX SAS - France - RCS Montpellier 328 031 042 - © Shutterstock
decreasing linearity humidity at 40°C (104°F) • 18 Parameters
• Specimen Volume: CBC Mode - 10 µl / CBC 3 part DIFF - 10µl WBC RBC PLT
• Power Requirements: LYM# & LYM% HGB MPV
Power supply from 90 Vac to 240 Vac MON# & MON% HCT PCT*
50 Hz to 60 Hz GRA # & GRA% MCV PDW*
Power consumption Maximum 230 VA MCH
MCHC
METHODS & TECHNOLOGIES RDW
• RBC/PLT detection principles • Linearity:
Method
Ruby diameter
Impedance
50 µm
Depression of count 200 mb
Duration of count 2 (or 3) x 6 seconds
Parameter
WBC
RBC
HGB
Linearity Limit
0 - 100 x103/mm3
0 - 8 x106/mm3
0 - 26 g/dL
Visible Range
100 - 150 x103/mm3
8 - 18 x106/mm3
26 - 30 g/dL
ABX Micros 60
Dilution ratio 1/15 000 HCT 0 - 80 % 80 - 90%
Process efficiency in Hematology
• HGB measurement PLT (Whole blood) 0 - 2200 x103/mm3 2200 - 6000 x103/mm3
Method Photometry PLT (PLT concentrate) 0 - 4000 x103/mm3 4000 - 6000 x103/mm3
Wavelenght 555 nm • Precision:
Dilution ratio 1/250 Parameter %CV at
• HCT measurement WBC ≤2.5 10 x103/mm3
Method Numeric integration RBC ≤2 5 x106/mm3
• WBC measurement HGB ≤1.6 15 g/dl
Method Impedance HCT ≤2.0 45%
Ruby diameter 80 µm PLT ≤5.0 300 x103/mm3
Depression of count 200 mb LYM% ≤5.0 40%
Duration of count 2 (or 3) x 6 seconds MON% ≤10 10%
Dilution ratio 1/250 GRA% ≤3 50%
• 3 part Differentiation • Carry-over:
Method Impedance + specific lyse WBC <0.5%
action RBC <0.5%
• MCV,MCH,MCHC,RDW,PCT*,PDW* HGB <0.5%
Calculation HCT <0.5%
PLT <2%
NORMES *(RUO parameter, Research Use Only.)
• CE UL CSA
NF EN 60601-1. UL 3101-1 C22.2 n°1010-1
NF EN 61326. B
NF EN 61000-3-2
NF EN 61000-3-3

FRANCE +33 (0)4 67 14 15 15 - BENELUX +32 (0)3 281 49 08 - ITALY +39 / 06 51 59 22 1 - SPAIN +34 / 91- 353 30 10 - PORTUGAL +351 / 2 14 72 17 70
UK +44 (0) 1604 542650 - POLAND +48 / 22 6732022 - USA +1 / 949 453 0500 - BRAZIL +55 / 11 5545 1500 - THAILAND +66 / 2 861 59 95
CHINA +86 / 21 3222 1818 - INDIA +91 / 11 4646 5000 - GERMANY AXON LAB AG +49 / 7153 92260 - DISTRIBUTORS NETWORK +33 (0)4 67 14 15 16
HORIBA Medical online : http://www.horiba.com/medical
Solutions adapted to your needs

ABX Micros 60 > Concept and Technology


• CBC
> Flexible with a list of options
• Quality control management
Impedance, selective lysing Smart card based QC
(LMG) (31 controls x 3 levels)
Complete haematology profile • Liquid valves • Sampling mode
18 parameters Precise volumes and reliability Open tube, or
Closed Tube (4 position tube holder)
• No compressor
Stepper motor and low noise • Identification mode
Alphanumerical (with internal keyboard)
> Automatic Maintenance Barcode reader (for patient’s ID number)
Automatic cleaning of the • Integrated Reagents
sampling probe Bottles, or
ABX Minipack LMG (slide-in / slide-out reagent
pack, integrating reagents and waste)
• 60 samples per hour > 18 parameters
• Memory
RBC, WBC, HGB, HCT, MCV,
• Open or Closed tube Last result, or
MCH, MCHC, PLT, PCT, MPV,
Smart card reader
• Compact and lightweight PDW, RDW, LYM (% and #),
(storage of 78 results with graphics on
MON (% and #), GRAN
• Micro-sampling: rewritable memory smart card)
(% and #), WBC, RBC and
10µl of whole blood per cycle PLT histograms

l Simple and easy menu l Fully-automatic analysis cycle l Only 3 reagents (diluent, lyse, cleaner)

Smart Card Sampling Sampling Barcode reader Keyboard ABX Minipack LMG ABX Micros 60
on Open tube on Closed tube Open Tube

You might also like