ELI 150c - 250c Szervice Manuals-EnGLITPDF
ELI 150c - 250c Szervice Manuals-EnGLITPDF
ELI 150c - 250c Szervice Manuals-EnGLITPDF
ELI® 150c/250c
12-LEAD RESTING ELECTROCARDIOGRAPH
SERVICE MANUAL
CAUTION: Federal law restricts this device to sale by or on the order of a physician.
©2019 Welch Allyn This document contains confidential information that belongs to Welch Allyn, Inc. No part of this
document may be transmitted, reproduced, used, or disclosed outside of the receiving organization without the express
written consent of Welch Allyn, Inc. Welch Allyn is a registered trademark of Welch Allyn, Inc. E-Scribe, ELI, and
VERITAS are trademarks of Welch Allyn, Inc. Cisco® is the registered trademark of Cisco Systems, Inc. DICOM® is
the registered trademark of the National Electrical Manufacturers Association for its standards publications relating
to digital communications of medical information.
9516-177-50-ENG Rev N
Revision Date: 2019-20
EU IMPORTER
Manufacturer’s Responsibility
Welch Allyn, Inc. is responsible for the effects on safety and performance only if:
• Assembly operations, extensions, readjustments, modifications, or repairs are carried out only by persons
authorized by Welch Allyn, Inc.
The user of this device is responsible for ensuring the implementation of a satisfactory maintenance schedule.
Failure to do so may cause undue failure and possible health hazards.
Equipment Identification
Welch Allyn, Inc. equipment is identified by a serial and reference number on the back of the device. Care should
be taken so that these numbers are not defaced.
This document contains information that is protected by copyright. All rights are reserved. No part of this
document may be photocopied, reproduced, or translated to another language without prior written consent of
Welch Allyn, Inc.
Welch Allyn, Inc. makes no warranty of any kind with regard to this material including, but not limited to,
implied warranties of merchantability and fitness for a particular purpose. Welch Allyn, Inc. assumes no
responsibility for any errors or omissions that may appear in this document. Welch Allyn, Inc. makes no
commitment to update or to keep current the information contained in this document.
Any serious incident that has occurred in relation to the device, should be reported to the manufacturer and the
competent authority of the Member State in which the user and/or patient is established.
Disposal
This product and its accessories must be disposed of according to the local laws and regulations. Do not dispose of this
product as unsorted municipal waste. For more specific disposal information see www.welchallyn.com/weee.
1
WARRANTY INFORMATION
b) Parts and/or accessories of the Product/s not obtained from or approved by Welch Allyn;
c) Misapplication, misuse, abuse, and/or failure to follow the Product/s instruction sheets
and/or information guides;
f) Other events outside of Welch Allyn’s reasonable control or not arising under normal operating conditions.
THE REMEDY UNDER THIS WARRANTY IS LIMITED TO THE REPAIR OR REPLACEMENT WITHOUT
CHARGE FOR LABOR OR MATERIALS, OR ANY PRODUCT/S FOUND UPON EXAMINATION BY
WELCH ALLYN TO HAVE BEEN DEFECTIVE. This remedy shall be conditioned upon receipt of notice by Welch
Allyn of any alleged defects promptly after discovery thereof within the warranty period. Welch Allyn’s obligations
under the foregoing warranty will further be conditioned upon the assumption by the purchaser of the Product/s (i) of
all carrier charges with respect to any Product/s returned to Welch Allyn’s principal place or any other place as
specifically designated by Welch Allyn or an authorized distributor or representative of Welch Allyn, and (ii) all risk
of loss in transit. It is expressly agreed that the liability of Welch Allyn is limited and that Welch Allyn does not
function as an insurer. A purchaser of a Product/s, by its acceptance and purchase thereof, acknowledges and agrees
that Welch Allyn is not liable for loss, harm, or damage due directly or indirectly to an occurrence or consequence
therefrom relating to the Product/s. If Welch Allyn should be found liable to anyone under any theory (except the
expressed warranty set forth herein) for loss, harm, or damage, the liability of Welch Allyn shall be limited to the
lesser of the actual loss, harm, or damage, or the original purchase price of the Product/s when sold.
EXCEPT AS SET FORTH HEREIN WITH RESPECT TO REIMBURSEMENT OF LABOR CHARGES, A
PURCHASER’S SOLE EXCLUSIVE REMEDY AGAINST WELCH ALLYN FOR CLAIMS RELATING TO THE
PRODUCT/S FOR ANY AND ALL LOSSES AND DAMAGES RESULTING FROM ANY CAUSE SHALL BE
THE REPAIR OR REPLACEMENT OF DEFECTIVE PRODUCT/S TO THE EXTENT THAT THE DEFECT IS
NOTICED AND WELCH ALLYN IS NOTIFIED WITHIN THE WARRANTY PERIOD. IN NO EVENT,
INCLUDING THE CLAIM FOR NEGLIGENCE, SHALL WELCH ALLYN BE LIABLE FOR INCIDENTAL,
SPECIAL, OR CONSEQUENTIAL DAMAGES, OR FOR ANY OTHER LOSS, DAMAGE, OR EXPENSE OF
ANY KIND, INCLUDING LOSS OF PROFITS, WHETHER UNDER TORT, NEGLIGENCE OR STRICT
LIABILITY THEORIES OF LAW, OR OTHERWISE. THIS WARRANTY IS EXPRESSLY IN LIEU OF ANY
OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING, BUT NOT LIMITED TO THE IMPLIED
WARRANTY OF MERCHANTABILITY AND THE WARRANTY OF FITNESS FOR A PARTICULAR
PURPOSE.
2
USER SAFETY INFORMATION
WARNING(S)
This manual gives important information about the use and safety of this device. Deviating from operating
procedures, misuse or misapplication of the device, or ignoring specifications and recommendations could
result in increased risk of harm to users, patients and bystanders, or damage to the device.
Device captures and presents data reflecting a patient’s physiological condition that when reviewed by a trained
physician or clinician can be useful in determining a diagnosis; however, the data should not be used as a sole
means for determining a patient’s diagnosis.
Users are expected to be licensed clinical professionals knowledgeable about medical procedures and patient
care, and adequately trained in the use of this device. Before attempting to use this device for clinical
applications, the operator must read and understand the contents of the user manual and other accompanying
documents. Inadequate knowledge or training could result in increased risk of harm to users, patients and
bystanders, or damage to the device. Contact Welch Allyn service for additional training options.
To ensure that electrical safety is maintained during operation from AC (~) power, the device must be plugged
into a hospital-grade outlet.
Only use parts and accessories supplied with the device and available through Welch Allyn, Inc.
Patient cables intended for use with the device include series resistance (9 Kohm minimum) in each lead for
defibrillation protection. Patient cables should be checked for cracks or breakage prior to use.
Conductive parts of the patient cable, electrodes, and associated connections of type CF applied parts, including
the neutral conductor of the patient cable and electrodes, should not come into contact with other conductive
parts including earth ground.
ECG electrodes could cause skin irritation; patients should be examined for signs of irritation or inflammation.
To avoid the possibility of serious injury or death during patient defibrillation, do not come into contact with
device or patient cables. Additionally, proper placement of defibrillator paddles in relation to the electrodes is
required to minimize harm to the patient.
3
USER SAFETY INFORMATION
This device was designed to use the electrodes specified in the user manual. Proper clinical procedure must be
employed to prep the electrode sites and to monitor the patient for excessive skin irritation, inflammation, or
other adverse reactions. Electrodes are intended for short-term use and should be removed from the patient
promptly following testing.
To avoid potential for spread of disease or infection, single-use disposable components (e.g., electrodes) must
not be reused. To maintain safety and effectiveness, electrodes must not be used beyond their expiration date.
A possible explosion hazard exists. Do not use the device in the presence of a flammable anesthetic mixture.
Where the integrity of external protective earth conductor arrangement is in doubt, the device shall be operated
from its internal electrical power source.
To improve immunity to potential interfering electromagnetic signals, shielded cabling is recommended when
connecting the device to a network.
Medical devices have been designed to have a higher degree of protection against electric shock than, for
instance, information technology equipment because patients often are connected to multiple devices and
also may be more prone to the adverse effect of electric currents than healthy persons. All equipment that is
connected to the patient, can be touched by the patient, or can be touched by another person while that
person touches the patient at the same time, should have the same level of protection against electric shock
as medical equipment. The ELI 150c and ELI 250c are medical devices that have been designed to be
connected to other devices for the purpose of receiving and transmitting data. Certain measures must be
taken to prevent the risk of excessive electric current flow through the operator or patient when connected:
All electrical equipment that is not medical electrical equipment must be placed outside of the “patient
environment,” defined by applicable safety standards to be at least 1.5 meters (5 feet) from the patient.
Alternatively, non-medical equipment may be provided with additional protection such as an additional
protective earth connection.
All medical electrical equipment that has a physical connection to the ELI 150c/ELI 250c or the patient
or is in the patient environment must comply with applicable safety standards for medical electrical
devices.
All electrical equipment that is not medical electrical equipment and has a physical connection to the
ELI 150c/ELI 250c must comply with applicable safety standards, such as IEC 60950 for information
technology equipment. This includes information network equipment connected through the LAN
connector.
Conductive (metal) parts that can be touched by the operator in normal use and that are connected to non-
medical equipment should not be brought into the patient environment. Examples are connectors for
shielded Ethernet or USB cables.
If multiple devices are connected to each other or to the patient, device chassis and patient leakage currents
may be increased, and should be measured for compliance with applicable standards for medical electrical
systems.
Avoid the use of portable multiple socket outlets. If used and not compliant with medical electrical
device standards, an additional protective earth connection is required.
4
USER SAFETY INFORMATION
To prevent electric shock due to unequal ground potentials that may exist between points of a distributed
network system or fault conditions in external network connected equipment, network cable shielding (where
used) must be connected to protective earth ground appropriate to the area where the device is used.
The device has not been designed for use with high-frequency (HF) surgical equipment and does not provide a
protective means against hazards to the patient.
When the 40 Hz filter is used, the frequency response requirement for diagnostic ECG equipment cannot be
met. The 40 Hz filter significantly reduces high-frequency components of the ECG and pacemaker spike
amplitudes, and is recommended only if high-frequency noise cannot be reduced by proper procedures.
The quality of the signal produced by the device may be adversely affected by the use of other medical
equipment, including but not limited to defibrillators and ultrasound machines.
For proper operation and the safety of users or patients and bystanders, equipment and accessories must be
connected only as described in this manual. Do not connect a telephone line cable to the LAN connector.
Some Welch Allyn electrocardiographs can be equipped with a GPRS (cellular modem) or wireless LAN
(WLAN) module for transmitting ECG records. Device labeling and the presence of an antenna port will
indicate if your device is equipped with such a module. If so equipped, the following notices apply:
The GPRS module operates in allocated frequency bands depending on the model. Identification of
the installed GPRS module can be found on a label on the bottom of the device.
MultiTech Systems, Inc. Model MTSMC-H5: 850/900/1700(AWS)/1900/2100 MHz
The WLAN identification can be found on a label on the bottom of the device.
B&B Electronics Model WLNN-AN-MR551: 2400 MHz
(model subject to change without notice)
5
USER SAFETY INFORMATION
Use of the GPRS or WLAN module may interfere with other equipment operating in the vicinity. Check with
local authorities or spectrum management officials in your facility to determine if restrictions apply to the use
of this feature in your area.
Do not transmit via the GPRS or WLAN module with a missing or damaged antenna. Replace a damaged
antenna immediately.
Use only the antenna supplied for use with this device. Unauthorized antennas, modifications, or attachments
could damage the GSM module and may contravene local RF emission regulations or invalidate type approval.
To ensure compliance with current regulations limiting both maximum RF output power and human exposure
to radio frequency radiation, a separation distance of at least 20 cm must be maintained between the device's
antenna and the head and body of the user and any nearby persons at all times. To help prevent degradation of
RF signal and to avoid excess RF energy absorption, do not touch the antenna during data transmission.
The GPRS and WLAN modules comply with applicable RF safety standards including standards and
recommendations for the protection of public exposure to RF electromagnetic energy that have been established
by governmental bodies and other qualified organizations, such as the following:
Federal Communications Commission (FCC)
Directives of the European Community
Directorate General V in Matters of Radio Frequency Electromagnetic Energy
Proper functioning backup items such as spare lead wires, front-end device, and other equipment are
recommended on hand to prevent delayed treatment due to an inoperable device.
Caution(s)
To prevent possible damage to the keyboard, do not use sharp or hard objects to depress keys, only
use fingertips.
Do not attempt to clean the device or patient cables by submersing into a liquid, autoclaving, or steam cleaning
as this may damage equipment or reduce its usable life. Wipe the exterior surfaces with a warm water and mild
detergent solution and then dry with a clean cloth. Use of unspecified cleaning/disinfecting agents, failure to
follow recommended procedures, or contact with unspecified materials could result in increased risk of harm to
users, patients and bystanders, or damage to the device.
No user-serviceable parts inside. Screw removal by qualified service personnel only. Damaged or suspected
inoperative equipment must be immediately removed from use and must be checked/repaired by qualified
service personnel prior to continued use.
The rechargeable internal battery is a sealed lead-acid type and it is totally maintenance free. If the battery
appears to become defective, refer to Welch Allyn Service Department.
Do not pull or stretch patient cables as this could result in mechanical and/or electrical failures. Patient cables
should be stored after forming them into a loose loop.
No calibration or special equipments are needed for the proper operation or maintenance of the device.
● When necessary, dispose of the device, its components and accessories (e.g., batteries, cables, electrodes),
and/or packing materials in accordance with local regulations.
Use only No. 26 AWG or larger telecommunication line cord.
6
USER SAFETY INFORMATION
Note(s)
Patient movements may generate excessive noise that may affect the quality of the ECG traces and the proper
analysis performed by the device.
Proper patient preparation is important to proper application of ECG electrodes and operation of the device.
The algorithm detecting electrode misplacements is based on normal physiology and ECG lead order, and tries
to identify the most likely switch; however, it is advisable to check the other electrode positions in the same
group (limb or chest).
There is no known safety hazard if other equipment, such as pacemakers or other stimulators, is used
simultaneously with the device; however, disturbance to the signal may occur.
If electrode is not properly connected to the patient, or one or more of the patient cable lead wires is damaged,
display will indicate a lead fault for the lead(s) where the condition is present and if the signal is being printed,
the respective lead(s) will print out as a square wave.
As defined by IEC 60601-1 and IEC 60601-2-25, the device is classified as follows:
Class I equipment or internally powered.
Type CF defibrillation-proof applied parts.
Ordinary equipment.
Equipment not suitable for use in the presence of a flammable anesthetic mixture.
Continuous operation.
NOTE: From a safety perspective, per IEC 60601-1 and derivative standards/norms, this device is
declared to be “Class I” and uses a three-prong inlet to ensure an earth connection is made along with
mains. The ground terminal on the mains inlet is the only protective earth point in the device. Exposed
metal accessible during normal operation is double insulated from mains. Internal connections to earth
ground are functional earth.
This device is intended to be used in a hospital or doctor’s office setting, and should be used and stored
according to the environmental conditions specified below:
Device must be configured at the factory for use with the WAM.
After operating the device using battery power, always reconnect the power cord. This ensures that the batteries
will be automatically recharged for the next time you use the device.
7
USER SAFETY INFORMATION
The device is a member of the ELI 1xx or ELI 2xx Series 2 electrocardiograph family.
Some Welch Allyn electrocardiographs can be equipped with an optional wireless data transmission module
(WLAN or GSM/GPRS mobile). Both these technologies use radios to transmit data to a Welch Allyn
receiving application. Due to the nature of radio transmissions, it’s possible that, due to the characteristics of
the environment where the device is located, some other RF sources may interfere with the transmission
generated by the device. Welch Allyn has tested the coexistence of the device with other devices that can
interfere such as devices using WLAN, Bluetooth radio, and/or cell phones. Although the current technology
allows a very successful rate of transmission, it’s possible that in some rare occurrences, the system may not
perform at its best resulting in a “failed transmission.” When this occurs, patient data will not be erased from
the device nor stored in the receiving application, ensuring that partial or corrupted data are not made available
to the receiving station. If the failure mode persists the user should move to a position where the RF signals
may propagate better and allow successful transmissions.
WLAN Option
Wireless options transmit in the 2.4 GHz or 5ghz range. Other nearby wireless devices may cause interference.
If possible, move or turn off other devices to minimize potential interference.
The Wireless LAN module used is compliant with the IEEE 802.11 a, b, g and n standards.
Access Points used should respect IEEE 802.11 standards as well as local Radio Frequency regulations. The
device will scan the available channels and connect to the Access Point on the channel where the SSID that is
configured on the device is available.
8
USER SAFETY INFORMATION
The following table shows the radio channels allocated in different geographic areas in the world. For bands
802.11b and g, only channels 1, 6, 11 and 14 (Japan only) are non-overlapping; for band 802-11a, channels
shown represent non-overlapping channel numbers.
9
USER SAFETY INFORMATION
The following table lists the frequency allocated for each channel used by the WLAN option.
In order to achieve the best transmission rate, it is necessary that the facility where the device is operated can
provide good area coverage. Please consult the IT personnel of the facility to verify the proper WLAN
availability in the area where the device will be used.
RF wave propagation may be blocked or reduced by the environment where the device is used. Most common
areas where this may occur are: shielded rooms, elevators, underground rooms. In all the above situations, it is
recommended to move the device to a proper location and verify with the IT personnel of the facility the areas
where the WLAN signals are available.
10
EQUIPMENT SYMBOLS AND MARKINGS
Symbol Delineation
Alternating current
Protective earth
Network (LAN)
USB port
Input
ON/OFF (power)
11
EQUIPMENT SYMBOLS AND MARKINGS
Antenna
Medical Device
Model Identifier
Reorder Number
12
ELECTROMAGNETIC COMPATIBILITY (EMC)
Electromagnetic compatibility with surrounding devices should be assessed when using the device.
An electronic device can either generate or receive electromagnetic interference. Testing for electromagnetic
compatibility (EMC) has been performed on the device according to the international standard for EMC for medical
devices (IEC 60601-1-2). This IEC standard has been adopted in Europe as the European Norm (EN 60601-1-2).
The device should not be used adjacent to, or stacked on top of other equipment. If the device must be used adjacent
to or stacked on top of other equipment, verify that the device operates in an acceptable manner in the configuration
in which it will be used.
Fixed, portable, and mobile radio frequency communications equipment can affect the performance of medical
equipment. See the appropriate table for recommended separation distances between the radio equipment and the
device.
The use of accessories, transducers, and cables other than those specified by Welch Allyn may result in
increased emissions or decreased immunity of the equipment.
13
ELECTROMAGNETIC COMPATIBILITY (EMC)
RF Emissions CISPR 11 Group 1 The equipment uses RF energy only for its internal function.
Therefore, its RF emissions are very low and not likely to cause
any interference in nearby electronic equipment.
RF Emissions CISPR 11 Class A The equipment is suitable for use in all establishments other
than domestic and those directly connected to the public low-
voltage power supply network that supplies buildings used for
Harmonic Emissions Complies domestic purposes.
IEC 61000-3-2
Electrostatic +/- 6 kV contact +/- 6 kV contact Floors should be wood, concrete, or ceramic
discharge (ESD) +/- 8 kV air +/- 8 kV air tile. If floors are covered with synthetic
IEC 61000-4-2 material, the relative humidity should be at
least 30%.
Electrical fast +/- 2 kV for +/- 2 kV for Mains power quality should be that of a typical
transient/burst power supply lines power supply lines commercial or hospital environment.
IEC 61000-4-4 +/- 1 kV for +/- 1 kV for
input/output lines input/output lines
Surge +/- 1 kV differential +/- 1 kV differential Mains power quality should be that of a typical
IEC 61000-4-5 mode mode commercial or hospital environment.
+/- 2 kV common +/- 2 kV common
mode mode
Voltage dips, <5% UT <5% UT Mains power quality should be that of a typical
short (>95% dip in UT) (>95% dip in UT) commercial or hospital environment.
interruptions, for 0.5 cycle for 0.5 cycle
and voltage 40% UT 40% UT
variations on (60% dip in UT) (60% dip in UT)
power supply for 5 cycles for 5 cycles
input lines
IEC 61000-4-11
Power frequency 3 A/m 3 A/m Power frequency magnetic fields should be at
(50/60 Hz) levels characteristic of a typical location in a
magnetic field typical commercial or hospital environment.
14
ELECTROMAGNETIC COMPATIBILITY (EMC)
The equipment is intended for use in the electromagnetic environment specified in the table below. The customer or
the user of the equipment should ensure that it is used in such an environment.
d
3.5
Conducted RF 3 Vrms 3 Vrms P
3Vrms
IEC 61000-4-6 150 kHz to 150 kHz to
80 MHz 80 MHz
3.5 80 MHz to 800 MHz
d P
3V / m
Radiated RF 3 V/m 3 V/m
IEC 61000-4-3 80 MHz to 80 MHz to
7 800 MHz to 2.5 GHz
2.5 GHz 2.5 GHz d P
3V / m
a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios,
amateur radios, AM and FM radio broadcast, and TV broadcast cannot be predicted theoretically with accuracy. To assess
the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the
measured field strength in the location in which the equipment is used exceeds the applicable RF compliance level above, the
equipment should be observed to verify normal operation. If abnormal performance is observed, additional measures may be
necessary, such as reorienting or relocating the equipment.
b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [3] V/m.
15
ELECTROMAGNETIC COMPATIBILITY (EMC)
The equipment is intended for use in the electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the equipment can help to prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the
equipment as recommended in the table below, according to the maximum output power of the communications
equipment.
Rated Maximum Output Power Separation Distance According to Frequency of Transmitter (m)
of Transmitter W
d 1.2 P d 2.3 P
0.01 0.1 m 0.2 m
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in
meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by the
absorption and reflection from structures, objects, and people.
16
ELECTROMAGNETIC COMPATIBILITY (EMC)
Regulatory Radio Compliance
WLAN
B&B electronics1 Radio Module 9373 with part number WLNN-AN-MR551
FCC ID: F4AWLNN551
1
Manufacturer also called B+B SmartWorx
17
ELECTROMAGNETIC COMPATIBILITY (EMC)
Industry Canada (IC) Emissions
RF Radiation Hazard Warning
Using higher gain antennas and types of antennas not certified for use with this product is not allowed. The
device shall not be co-located with another transmitter.
Cet avertissement de sécurité est conforme aux limites d'exposition définies par la norme CNR-102 at relative
aux fréquences radio.
Operation is subject to the following two conditions: (1) this device may not cause interference, and (2) this
device must accept any interference, including interference that may cause undesired operation of this device.
L’utilisation de ce dispositif est autorisée seulement aux conditions suivantes: (1) il ne doit pas produire de
brouillage et (2) l’ utilisateur du dispositif doit étre prêt à accepter tout brouillage radioélectrique reçu, même
si ce brouillage est susceptible de compromettre le fonctionnement du dispositif.
WLAN
B&B electronics1 Radio Module 9373 with part number WLNN-AN-MR551
IC ID: 3913A-WLNN551
1
Manufacturer also called B+B SmartWorx
18
ELECTROMAGNETIC COMPATIBILITY (EMC)
European Union
Czech Welch Allyn tímto prohlašuje, ze tento WLAN device je ve shodě se základními
požadavky a dalšími příslušnými ustanoveními směrnice 2014/53/ES.
Danish Undertegnede Welch Allyn erklærer herved, at følgende udstyr WLAN device
overholder de væsentlige krav og øvrige relevante krav i direktiv 2014/53/EF
Dutch Bij deze verklaart Welch Allyn dat deze WLAN device voldoet aan de essentiële
eisen en aan de overige relevante bepalingen van Richtlijn 2014/53/EC.
English Hereby, Welch Allyn, declares that this WLAN device is in compliance with the
essential requirements and other relevant provisions of Directive 2014/53/EC.
Estonian Käesolevaga kinnitab Welch Allyn seadme WLAN device vastavust direktiivi
2014/53/EÜ põhinõuetele ja nimetatud direktiivist tulenevatele teistele asjakohastele
sätetele.
Finnish Welch Allyn vakuuttaa täten että WLAN device tyyppinen laite on direktiivin
2014/53/EY oleellisten vaatimusten ja sitä koskevien direktiivin muiden ehtojen
mukainen.
French Par la présente, Welch Allyn déclare que ce WLAN device est conforme aux
exigences essentielles et aux autres dispositions de la directive 2014/53/CE qui lui
sont applicables
German Hiermit erklärt Welch Allyn die Übereinstimmung des Gerätes WLAN device mit den
grundlegenden Anforderungen und den anderen relevanten Festlegungen der
Richtlinie 2014/53/EG. (Wien)
Greek ΜΕ ΤΗΝ ΠΑΡΟΥΣΑ Welch Allyn ΔΗΛΩΝΕΙ ΟΤΙ WLAN device
ΣΥΜΜΟΡΦΩΝΕΤΑΙ ΠΡΟΣ ΤΙΣ ΟΥΣΙΩΔΕΙΣ ΑΠΑΙΤΗΣΕΙΣ ΚΑΙ ΤΙΣ ΛΟΙΠΕΣ
ΣΧΕΤΙΚΕΣ ΔΙΑΤΑΞΕΙΣ ΤΗΣ ΟΔΗΓΙΑΣ 2014/53/ΕΚ
Hungarian Alulírott, Welch Allyn nyilatkozom, hogy a WLAN device megfelel a vonatkozó
alapvetõ követelményeknek és az 2014/53/EC irányelv egyéb elõírásainak.
Italian Con la presente Welch Allyn dichiara che questo WLAN device è conforme ai
requisiti essenziali ed alle altre disposizioni pertinenti stabilite dalla direttiva
2014/53/CE.
Latvian Ar šo Welch Allyn deklarē, ka WLAN device atbilst Direktīvas 2014/53/EK
būtiskajām prasībām un citiem ar to saistītajiem noteikumiem.
Lithuanian Šiuo Welch Allyn deklaruoja, kad šis WLAN device atitinka esminius reikalavimus ir
kitas 2014/53/EB Direktyvos nuostatas.
Malti Hawnhekk, Welch Allyn, jiddikjara li dan WLAN device jikkonforma mal-htigijiet
essenzjali u ma provvedimenti ohrajn relevanti li hemm fid-Dirrettiva 2014/53/EC
Portuguese Welch Allyn declara que este WLAN device está conforme com os requisitos
essenciais e outras disposições da Directiva 2014/53/CE.
Slovak Welch Allyn týmto vyhlasuje, ze WLAN device spĺňa základné požiadavky a všetky
príslušné ustanovenia Smernice 2014/53/ES.
Slovene Šiuo Welch Allyn deklaruoja, kad šis WLAN device atitinka esminius reikalavimus ir
kitas 2014/53/EB Direktyvos nuostatas.
Spanish Por medio de la presente Welch Allyn declara que el WLAN device cumple con los
requisitos esenciales y cualesquiera otras disposiciones aplicables o exigibles de la
Directiva 2014/53/CE
Swedish Härmed intygar Welch Allyn att denna WLAN device står I överensstämmelse med
de väsentliga egenskapskrav och övriga relevanta bestämmelser som framgår av
direktiv 2014/53/EG.
19
ELECTROMAGNETIC COMPATIBILITY (EMC)
20
ELECTROMAGNETIC COMPATIBILITY (EMC)
Philippines National Telecommunications B&B: ESD - 1818097C
Commission
South Korea Korea Communications This equipment is Industrial (Class A) electromagnetic wave
Commission (대한민 국 방송통 suitability equipment and seller or user should take notice of it,
and this equipment is to be used in the places except for home.
신위원 회) – KCC 이 기기는 업무용(A급) 전자파적합기기로서 판 매자
Certification number: 또는 사용자는 이 점을 주의하시기 바라 며,
B&B: R-C-BVT-9373 가정외의 지역에서 사용하는 것을 목적으로 합니다.
UAE
B&B ER65768/18
21
MAINTENANCE & CLEANING
□
Clean lint free cloth
□
Mild detergent
□
Isopropyl Alcohol (80-99%)
□
Cyanoacrylate Adhesive (e.g., Loctite 444 or equivalent product)
□
Anaerobic Thread locker
□
Smart Thermal Paper
□
10% Household bleach and water solution (Sodium Hypochlorite solution consisting of a
minimum 1:500 dilution and maximum of 1:10 dilution for disinfecting use only)
□
Multi-Meter
□
ECG Simulator (10 Lead)
□
AM12 Patient Input Module (9293-048-5X)
□
WAM Patient Input Module (30012-019-5X)
□
WAM/AM12 Lead Wire Set (9293-046-70)
□
Patient Cable Snap Adapter Set (9281-002-50)
□
Phillips #2 Screwdriver
□
10 Lead Shorting Block (or equivalent)
□
Lead Test Failure Box (or equivalent)
□
FAT 32 USB Memory Device
□
PC with ELI-Link v3.10 or later
□
USB Cable Type A to B (6400-012)
□
Phone Cable (6400-004)
□
PC with ELI-Link or Escribe and Modem (or equivalent)
□
PC with NIC and ELI-Link v4.5 or later
□
Wireless Router 802.11 (a, b, g, n)
□
Electrical Safety Analyzer
22
MAINTENANCE & CLEANING
Preventive Maintenance
2.1 Turn unit on and print the device configuration per configuration section of this
manual. Attach a copy to the Preventive Maintenance Report.
2.4.1 Ensure battery is fully charged before performing these tests, voltage and current
limits are based on a fully charged battery.
2.4.2 Ensure there is no power connected to the UUT AC inlet.
2.4.3 Remove upper housing and writer assembly. Disconnect battery by pulling
battery cable off of the red terminal.
23
MAINTENANCE & CLEANING
2.7.1 AC LED/Display
Connect AC power cord to the unit and verify that the green AC LED (located to
the left of the display) illuminates continuous.
NOTE: The battery indicator will be clear when charging and will illuminate white
when fully charged.
Verify text on display is clear and legible and there are no flickering or missing
lines/pixels. Record result on PMR.
2.7.2 Writer
Open and close the writer door to verify smooth operation. Verify that the door
unlatches without sticking and that it latches completely. From the main screen,
simultaneously press shift+alt+RHY. Verify that a test page is printed and the
writer stops on the cue mark. The perforation of the paper should line up with the
tear edge on the writer. Assure there are no gaps in the printing and the print
darkness is uniform across the entire page. Verify the writer gears do not skip
and paper is tracking properly (you may need to print multiple pages to observe
this). Record result on PMR.
24
MAINTENANCE & CLEANING
The receiving station for modem, LAN and WLAN transmissions should be
running Welch Allyn ELI-Link software. Refer to the ELI-Link user manual for
proper configuration.
Successful transmission of the test records can be verified by viewing the ECG
records in the unit directory after transmission and confirming they are marked as
“transmitted” (as defined in the product user manual). Record result on PMR
Modem
LAN
WLAN
GSM/GPRS
USB host (USB memory device needed)
USBD
2.8 Clean unit per the instructions provided on a later page of this section of the service manual.
25
MAINTENANCE & CLEANING
If the cardiograph housing was opened for repair or inspection work, the following safety tests
should be performed in accordance with the IEC 60601 standards.
The ELI150C and ELI250C are considered a Class 1 Type CF devices, intended to only be
utilized with the Welch Allyn AM12 or WAM patient input modules. Defibrillation isolation from
the patient is provided by the patient input modules, which are tested separately as part of
the manufacturing process (they are considered non-serviceable devices), therefore Hi-pot
testing is not required for these cardiograph models.
Earth Leakage
Enclosure Leakage
Non-conductive (fully insulated) chassis testing should be performed utilizing 200
cm2 conductive foil or equivalent, earth ground on AC input is utilized for
functional earth (not safety grounding).
Patient Leakage
Applied part – patient input (utilize Welch Allyn AM12 patient cable)
Patient Auxiliary Current
Applied part – patient input (utilize Welch Allyn AM12 patient cable)
26
MAINTENANCE & CLEANING
1. Disconnect the power source. Remove cables and lead wires from device before cleaning.
2. For general cleaning of cables and lead wires, use a soft, lint-free cloth lightly moistened
with a mild soap and water solution. Wipe and air dry.
3. For disinfecting the exterior surfaces of the device, patient acquisition module, cables,
and lead wires, wipe exterior using:
a soft, lint-free cloth with a solution of Sodium Hypochlorite (10% household bleach and
water solution) minimum 1:500 dilution (minimum 100 ppm free chlorine) and maximum
1:10 dilution as recommended by the APIC Guidelines for Selection and Use of
Disinfectants
CAUTION:
Prevent liquid from penetrating the device and do not attempt to clean/disinfect the
device or patient cables by submerging into a liquid, autoclaving, or steam cleaning.
Do not sterilize the device or lead wires with Ethylene Oxide (EtO) gas.
Do not immerse cable ends or lead wires; immersion can cause metal corrosion. Use
caution with excess liquid as contact with metal parts may cause corrosion.
Improper cleaning products and processes can damage the device, produce brittle lead
wires and cables, corrode the metal, and void the warranty. Use care and proper
procedure whenever cleaning or maintaining the device.
27
MAINTENANCE & CLEANING
Unit Serial #: _
Power Testing
Note Battery Age (If Possible) / _ (week/year)
Battery (Open Circuit) Voltage VDC (>12.5vdc)
Battery (with Load) Voltage VDC (>11.7vdc)
Off Current Ua (<100uA)
On Current mA (<250mA)
* Based upon customer usage and age of main battery, replace as needed.
□ Verify all power cables are properly reconnected and reassemble unit
Functional testing
AC LED/Display Functionality PASS / FAIL (Circle One)
Writer Test PASS / FAIL (Circle One)
ECG & Keyboard Matrix Testing PASS / FAIL (Circle One)
ECG Noise Test PASS / FAIL (Circle One)
Lead failure Test PASS / FAIL (Circle One)
Communication Option(s)
USB host PASS / FAIL / NA (Circle One)
USBD PASS / FAIL / NA (Circle One)
MODEM PASS / FAIL / NA (Circle One)
GSM/GPRS PASS / FAIL / NA (Circle One)
LAN PASS / FAIL / NA (Circle One)
WLAN PASS / FAIL / NA (Circle One)
□ Device Cleaning
28
DEVICE CONFIGURATION
Configuration Menus
The configuration pages define all operational conditions that do not change on a daily or patient-to-patient basis.
Once you set these default conditions, you will rarely need to use the configuration screens again. To access the
configuration menus:
To print the device’s configuration settings, select F6 (More) from real-time ECG view. Select F6 (More) again
followed by F1 (Print Configuration). The configuration printout captures every configuration setting: the
software version, the cart number of the device, and the date and time that the configuration printout occurred.
29
DEVICE CONFIGURATION
30
DEVICE CONFIGURATION
31
DEVICE CONFIGURATION
32
DEVICE CONFIGURATION
Configuration Settings
Software Version
Cart Number
Site Number
Identifies the site of your device. Site numbers designate the hospital, clinic, or institution for ECG records stored in
an E-Scribe system and must be defined for transmitting and retrieving ECGs from that system. You can use up to
four digits for the site number. Numbers from 0 – 4095 are supported.
Site Name
Defines your clinic, hospital, or office name. You can enter up to 30 alphanumeric characters. The site name prints
at the bottom, left edge of the ECG printout.
Telephone Number
Specifies the telephone number for internal modem transmission to another unit or to an E-Scribe system. Enter up
to 45 numeric characters.
You may need to dial a 9 to get an outside line. To wait for an additional dial tone, use the letter W.
EXAMPLE: 9W14145554321
EXAMPLE: P14145554321
(If necessary, you can use both the letter W and the letter P in the same phone number.)
NOTE: It is not necessary to use alpha characters in the telephone number with GSM/GPRS mobile
connectivity.
TIP: To quickly delete or modify a phone number, use a shortcut. From the application screen,
simultaneously press (SHIFT) + ALT + P. To edit an existing telephone number, use the Tab key.
Language
CAUTION: Function labels are immediately translated upon selecting a new language and exiting the
configuration screen.
33
DEVICE CONFIGURATION
If an unknown language is visible, use the following steps to revert to the language of your country:
Alphabets of specific languages may require use of special characters in demographic fields. This is accomplished
by using the SYM key on the keyboard.
Volume
Defines the keyboard click loudness. Available settings range from 0 (off) to 8 (loud).
Determines when the electrocardiograph will switch off in order to conserve the battery life of the device. The
battery time out will only occur if the keyboard has not been depressed for the time specified. The battery time out
setting is ignored if an active ECG signal is detected during transmission or while rhythm printing.
Flash Size
Indicates ECG storage capacity. Normal indicates standard memory capacity. Expanded indicates the optional
expanded memory has been installed.
ID Format
Defines the format for the patient demographic information prompts. There are three standard formats: short,
standard, or long. A custom ID format can be downloaded from ELI Link or an E-Scribe system. See Appendix A
to download a custom ID.
The short format includes the patient's last and first name, patient ID number, date of birth (automatically calculates
the age), and gender.
The standard format includes the patient's last name, patient ID number, age, height, weight, gender, race,
medication 1, medication 2, and a location field.
The long format is identical to the standard format except that it includes the patient's first name, room, and
comment fields.
Auto-Fill ID
When enabled, the device will automatically populate last name, first name, date of birth, age, and gender in the ID
screen if records with matching patient ID are found in the ECG directory.
34
DEVICE CONFIGURATION
AC Filter
The device removes 60 Hz or 50 Hz interference. The setting you select depends on the line frequency in your
country. Always use the 60 Hz setting in the U.S. If AC interference is present, check to see that the proper AC
filter is selected.
Paper Speed
Configure to 25 mm/s or 50 mm/s for default ECG printouts. For rhythm printouts and display, speeds of 5 mm/s or
10 mm/s are also available. See Device Configuration section to change speeds for display or rhythm printing. Paper
speed is printed at the bottom right corner of the ECG printout.
Filter
The ECG plot-frequency filter (or print filter) can be set to 0.05 to 40 Hz, 0.05 to 150 Hz, or 0.05 to 300 Hz. The
plot-frequency filter does not filter the acquired digital record. A 40 Hz plot-filter setting will reduce the noise
(40 Hz and higher frequencies) on the printed ECG, and a 150 Hz plot-filter setting will reduce the noise (150 Hz
and higher frequencies) on the printout; a 300 Hz plot-filter setting will not filter the printed ECG. The filter setting
is printed at the bottom right corner of the ECG printout.
Height/Weight Units
Defines the units of weight and height to either pounds/inches (lb/in) or kilograms/centimeters (kg/cm).
Date Format
Select either U.S. or European format for entering and displaying the patient’s date of birth.
NOTE: The date format option does not modify the acquisition date printed on each ECG.
Interpretation
The device automatically analyzes ECGs and prints the optional interpretation on the ECG printout. This setting
allows you to select or suppress the “interpretive” text on the ECG report.
NOTE: The ECG interpretations offered by the device are only significant when used in conjunction with
a physician over-read as well as consideration of all other relevant patient data.
Reasons
The reasons statements indicate why a particular interpretive statement was printed. Reasons statements print
enclosed in [square brackets] within the interpretive text if the interpretation option is turned on. Turning the
reasons statement function on or off does not affect the measurements performed or the interpretive statements
selected by the analysis program.
For Example:
Anteroseptal Infarct [40+ ms Q WAVE IN V1-V4]
Where “Anteroseptal Infarct” is the interpretive statement,
and “40+ ms Q WAVE IN V1-V4” is the reason statement or explanation as to why the
interpretive statement was printed.
35
DEVICE CONFIGURATION
Append
A status or statement phrase can be appended to the ECG and printed under the interpretive text printout. Either
“unconfirmed report” or “reviewed by” can be selected; however, if you wish to have nothing appended to the ECG,
select “blank”.
Number of Copies
Defines the number of printed copies when an ECG is taken. (V1.x) A zero (0) setting prints the original only; one
(1) prints the original plus 1 copy; two (2) prints the original plus 2 copies, and so on. Up to 9 copies may be
selected. (V2.x) A zero (0) setting does not print the ECG; one (1) prints the original; two (2) prints the original plus
1 copy, and so on. Up to 9 copies may be selected.
Defines whether or not printed copies will include interpretation. The clinician may request the first ECG printout
with the interpretation included. Additional copies may be printed with or without the interpretation.
Defines the number of ECGs retrieved from an E-Scribe system. The ECGs are retrieved by ID number. A zero (0)
setting retrieves the most current ECG for that ID number. Settings from one (1) to seven (7) retrieve the most
current ECG plus “X” number of ECGs identified by the entered value. EXAMPLE: If you enter the number 5, you
will retrieve the most current ECG plus the five preceding ECGs for that ID number. ECGs retrieved from the
E-S cribe are only printed at the device and not saved.
Delete Rule
Defines the rule to mark ECGs as deleted in the ECG directory. ECGs that are marked for deletion will be
automatically removed or erased based on their acquisition date (a first-in/first-out philosophy) to make room for the
new ECG record. ECGs are only erased from the directory when they are marked for deletion and if the directory
becomes full. More than one ECG may be removed from the directory in order to make room for the new incoming
record. The delete rule selections are:
Storage Sensitivity
Dictates the resolution of all stored ECG records. The sensitivity setting is either Normal or High. If the value is set
to High, the stored ECG will have a high resolution. As a result, the record size will be large and will reduce the
storage capacity in the ECG directory.
Defines whether or not a newly acquired ECG will be automatically saved to the directory once it is acquired and
printed. If the auto-save configuration option is set to No and the record is printed, the device will prompt you to
“Save ECG?” F1 (Save) will store the ECG in the directory.
36
DEVICE CONFIGURATION
Defines whether or not the device will automatically print the ECG after acquisition. If the selected configuration
option is set to No, a manual printout is possible.
Caps Lock
Use A4 Paper
The ELI 250c accommodates use of Z-fold thermal paper in either letter size (8.5 x 11 inches; 216 x 279 mm) or A4
size (8.27 x 11.69 inches; 210 x 297 mm). The provided paper tray spacer is required for use with A4 size paper.
Rhythm Formats
Defines the default values for rhythm printing. It is possible to set a 3 or 6-channel default rhythm format for the
ELI 150c. For the ELI 250c, a 3, 6, or 12-channel default rhythm format is possible. Define rhythm leads one
through three to customize a 3-channel rhythm printout or define rhythm leads one through six to customize the
6-channel rhythm printout.
Plot Format
Defines the default for one of the available plot formats in either standard or Cabrera presentation. Please note that
regardless of the plot format selected, 10 seconds of 12 leads are always stored.
37
DEVICE CONFIGURATION
Rhythm Leads
Displays continuous rhythm of selected ECG leads and permits printing of selected leads. User may toggle between
selected leads, system set leads, or I, II, III, aVR, aVL, and aVF followed by V1, V2, V3, V4, V5, and V6.
Average RR
Enabling this option will display an averaged RR value to appear on the report.
QTcB
Enabling this option will display a Bazett’s corrected QT value on the report along with the default linear
QTc value.
QTcF
Enabling this option will display a Fridericia corrected QT value on the report along with the default linear
QTc value.
ECG Capture
Up to 5 minutes accumulated ECG data can be acquired internally for use with the Best 10 feature. The device
automatically selects the best 10 seconds from within the 5-minute buffer.
Users can switch between BEST 10 or LAST 10 by selecting F5 (More) followed by F5 (Last) or F5 (Best)
depending on the current view.
Band Mode
Use 850/1900 MHz (US) or 900/1800 MHz (EU). (Applies to ELI 150c only.)
Sync Media
Defines the default transmission setting. Select None, Modem, LAN, WLAN, or GSM/GPRS (GSM/GPRS option
applies to ELI 150c only). Optional connectivity options which have been purchased and installed will be available
for default selection.
An ELI x50c communicating over GPRS can be configured to automatically set its clock to match the time on a time
sync server. The time sync server must return a time stamp in the ELI x50c’s local time zone via the daytime
protocol (RFC 867). The time sync server must have a public IP address, and the standard port is 13. The server
must return the time in one of the following formats:
Format 1
day mon dd HH:mm:ss yyyy
Example
Wed Jul 15 17:05:49 2010
38
DEVICE CONFIGURATION
Format 2
hh.mm.ss tt mm/dd/yyyy
Example
02:38:51 PM 07/18/2011
Time sync servers running the Dimension 4 (http://www.thinkman.com/dimension4/index.htm) time sync software
support Format 1.
DHCP
Defines whether the Dynamic Host Communication Protocol (DHCP) will be used to obtain an IP address. If DHCP
is Yes, the network will automatically and dynamically assign an IP address. If DHCP is No, you must enter the IP
address, def gateway, and sub net mask.
NOTE: All parameters related to network connection must be entered under the direction of the IT
Manager of the facility where the device is installed.
IP Address
Enter the fixed IP address for network transmissions (if DHCP is not selected).
Def Gateway
Enter the address of the default gateway (if DHCP is not selected).
Host IP
NOTE: Addresses are always entered as 4 sets of 3 digits; therefore, an address of 192.168.0.7 must be
entered as 192.168.000.007.
Port Number
LAN MAC
Security (WEP)
Wired Equivalent Privacy (WEP) is an encrypted security protocol (part of the 802.11 standard). Access points can
have multiple WEP keys stored. Each one of them is identified by a number (e.g., 1, 2, 3, 4).
WEP Key
39
DEVICE CONFIGURATION
WEP Key ID
Enter the 128-bit WEP key ID value (26 digits in 13 sets of two digits).
WLAN MAC
Shows the MAC address of the device’s wireless module for configuring access points.
SSID
Service Set Identifier (SSID) is the name of the wireless network. All ELI 150c electrocardiographs that will
transmit to the same network must have the same SSID name. This field is case sensitive.
WPA-PSK/WPA2-PSK
Allows for implementation of the “personal mode” of WPA. This mode of encryption employs Temporal Key
Integrity Protocol (TKIP) which dynamically changes keys as the system is used.
PSK Passphrase
The passphrase may be from eight to 63 ASCII characters or 64 hexadecimal digits (256 bits).
WPA-LEAP
Cisco® LEAP (Light Extensible Authorization Protocol) enables use of the device with wireless networks
employing the LEAP encryption protocol.
LEAP Password
WPA2-PEAP
Enables use of the device with wireless networks employing the PEAP encryption protocol.
PEAP Password
40
DEVICE CONFIGURATION
Worklist Management
The device can download and process ECG order lists from the E-Scribe or another compatible information
management system which identifies the ECGs (or ECG orders) needed for particular patients. Implementation of
an order-based workflow can significantly reduce demographic data entry errors at the electrocardiograph. Orders
are deleted from the list when the ordered ECG is acquired.
When set to Standard, new order lists are appended to the remaining list. When set to Refresh, each new order list
will override the previously downloaded one.
Comm. Protocol
Select UNIPRO32, DICOM32, OR DICOM32ext for software v1.x.x. DICOM32 and DICOM32ext are only
available if the DICOM option has been installed. Select UNIPRO64 OR DICOM64 FOR SOFTWARE V2.x.x.
DICOM64 is only available if the DICOM option has been installed.
NOTE: This parameter must be entered under the direction of the IT Manager of the facility where the
device is installed.
NOTE: Units ship by default with Comm Protocol set to UNIPRO32. The UNIPRO32 setting is not
supported by E-Scribe versions prior to V8.10 or ELI Link versions prior to V3.00. For questions about
compatibility of your device with E-Scribe or ELI Link and UNIPRO32, contact Welch Allyn Technical
Support. Units shipped with v2.x.x software will not connect with E-Scribe and need Eli Link v4.0 or
above.
Sync Mode
Select None, XMT, or XMT+Orders. None requires a manual transmission of reports and then a second manual
request to receive orders from the cardiology management system. XMT will automatically transmit the report;
XMT+Orders will both transmit the report and retrieve the orders.
Sync Date/Time
Select Yes or No. Yes will synchronize the date/time with the approved cardiology management system. With No,
there will be no date/time synchronization. Date/time synchronization is done through ELI Link V3.10 or later.
Defines fields required for ECG transmission to the cardiology management system. None will allow data
transmission without limitation; Last Name requires the technician to enter a minimum of the Last Name; Last
Name and ID requires the technician to enter a minimum of the Last Name and the patient’s ID.
41
UNIT DISASSEMBLY
ELI 150c
ELI 250c
42
UNIT DISASSEMBLY
Remove two thumbscrews from underneath the cart platform by turning counterclockwise. *Cart may not be exact
model as pictured.
*The ELI 150c and ELI 250c are very similar in design, so only the ELI 150c will be shown, unless an important
difference needs to be identified.
Item# 31
43
UNIT DISASSEMBLY
Open the writer drawer then lift the upper housing while rotating it counterclockwise according to the picture below.
This will allow the writer cover to pass through the housing opening to free the upper housing.
CAUTION: Remove battery power to the unit by disconnecting the battery connector BEFORE
removing the keyboard assembly. If battery power is not removed prior to disconnecting the keyboard
cable, damage to the motherboard may result.
Writer Removal
To remove the writer assembly, remove the 4 chassis screws shown below. Once they are removed, carefully turn
the unit over supporting the loose writer to ensure the cable connections are not stressed during the process.
Item# 36
44
UNIT DISASSEMBLY
Disconnect the writer interface cables and the motor cable shown below, then remove the writer assembly.
45
UNIT DISASSEMBLY
NOTE: The writer assembly can be obtained as a complete assembly for service purposes, or a specific
part or subassembly can be obtained to repair a specific writer related issue. The entire writer door with
the platen roller, latch assembly, and instruction label attached is available as an assembly; and the
thermal print head, print head mount, anti-static brush, and associated cables are also available as an
assembly. (Refer to the SERV ASSY item numbers on the item listing at the end of this section of the
manual).
To remove the writer door assembly, slide the door open and press downward on the latching tabs to allow the writer
door assembly to be slid out.
46
UNIT DISASSEMBLY
The writer latch mechanism can be disassembled to gain access to the writer latch bar and spring by removing the 4
screws shown below.
Item # 34
To remove the Gearbox assembly, remove the three screws as shown in the illustration below.
Item #36
47
UNIT DISASSEMBLY
To remove the writer motor, the pinion gear must be removed by loosening the pinion set screw as shown above.
NOTE: The set screw is installed at the factory with a Vibra-Tite coating to prevent loosening due to
vibration; ensure a coated set screw is used when reinstalling a new motor. The set screw should be
tightened to a torque 3.5 pound inches to ensure a proper connection to the motor shaft.
Item #34
The motor can then be removed by the removal of the two mounting screws shown above (actual screws are TORX
head for this product).
48
UNIT DISASSEMBLY
To remove the thermal print head assembly, remove the rubber O-ring as indicated in the illustration below.
Next, flip the writer assembly over and lift the thermal print head assembly out, taking care to feed the wires through
the writer base slots (shown below) as it is removed.
49
UNIT DISASSEMBLY
If the Print head assembly is disassembled further into the individual parts, care should be taken to re-torque the
ground screw to 3.5 pound inches during reassembly; the shoulder screw can be tightened completely as it allows
the Print head and mount to expand and contract when the device is exposed to abrupt temperature changes. When
replacing the Print head, keep in mind that the anti-static brush is a separate item and can not be reused from the old
Print head.
NOTE: When repairing units with symptoms of light or uneven print darkness, Welch Allyn recommends
that the entire Print head assembly be replaced to ensure the problem resolved completely. Slight
variances in the shape of the Print head and/or the print head mount can result in the unit exhibiting
these symptoms. Replacement of the entire assembly will ensure the problem is completely resolved.
(Refer to the SERV ASSY item numbers on the item listing at the end of this section of the manual).
50
UNIT DISASSEMBLY
Keyboard Removal
CAUTION: Remove battery power to the unit by disconnecting the battery connector BEFORE
removing the keyboard assembly. If battery power is not removed prior to disconnecting the keyboard
cable, damage to the motherboard may result.
Remove the 3 keyboard mounting screws which fasten the keyboard to the unit’s lower housing (150c only).
Item # 36
Next, press firmly and outward on the keypad connector hold-downs (as shown below) to release the keyboard
ribbon cable; then remove the keypad assembly from the bottom housing.
51
UNIT DISASSEMBLY
To remove the LCD Display, use a small flat blade screwdriver to lift up the latch mechanism that holds the LCD
ribbon cable into the connector; then slide the cable out of the connector.
Be careful not to apply excessive force, as this connector mechanism is very small and fragile.
IMPORTANT ELI250c: A new ELI250C keyboard PCBA was introduced in 2019 that has compatibility
requirements. The original part #36025-108-150 (item 107a) can use any software version, the newer part #36025-108-
400 (item 107b) must operate using v1.3.2 or newer (series 1.x), or v2.1.1 or newer (series 2.x) software.
IMPORTANT ELI150c: A new ELI150C keyboard PCBA was introduced for 4th edition compliance in 2020 that has
compatibility requirements. The original part #36025-102-150 (item 10a) can use any software version, the newer part
#36025-102-400 (item 10b) must operate using v1.3.3 or newer (series 1.x), or v2.1.2 or newer (series 2.x) software.
The rubber keypad and LCD can then be separated from the keyboard PCB assembly.
NOTE: When troubleshooting key press related issues, careful inspection and cleaning of the key pucks
and PCB contacts should be performed prior to any electrical troubleshooting.
52
UNIT DISASSEMBLY
IMPORTANT ELI150c Keypad: A new ELI150C keypad was introduced for 4th edition compliance in
2020 that has compatibility requirements. Exchanging the identical part for servicing purposes is
recommended, however if the older part (8359-003-50) is not available, the newer keypad
(413252) can be used to service the older models. The older keypad CANNOT be used on the newer 4th
edition compliant version. Refer to the table below for compatibility information.
8359-003-50 Yes No
The visual difference between the two keypads is shown below, the new part has a portion removed.
8359-003-50 413252
When servicing an ELI150c 4th edition device, there is also an additional item #92 used to reduce
electromagnetic interference (EMI). When servicing the keyboard/keypad items, it may be necessary to
replace the gasket material to ensure connectivity between the two surfaces. The gasket positioning is
shown below.
Remove the adhesive liner on the gasket, then mount the gasket to the pad within the cutout of the keypad.
53
UNIT DISASSEMBLY
The Main Processor is located on the Keypad Motherboard. When installing the ribbon cable onto the circuit board,
it is imperative to install using the following process. If care is not taken when pressing the ribbon cable into the
connector, damage to the solder connections to the Ball Grid Array of the processor due to flexing of the printed
circuit board assembly could occur.
To install, ensure that the cable connector is correctly aligned with the board connector. Grasp the keypad side of
the board with your forefingers, and press with equal pressure with both thumbs on the ribbon connector. Some
force may be required to fully seat this connector. See picture below:
54
UNIT DISASSEMBLY
Be sure the battery connector is disconnected, then remove the four screws as shown in illustration below.
Item #30
IMPORTANT ELI150c I/O Board: A new ELI150C I/O Board was introduced for 4th edition compliance
in 2020 that has compatibility requirements. Exchanging the identical part for servicing purposes is
recommended, however if the older part (26025-105-15x) is not available, the newer I/O Board can be used
to service the older models. The older I/O Board CANNOT be used on the newer 4th edition compliant
version. Refer to the table below for compatibility information.
26025-105-15x Yes No
55
UNIT DISASSEMBLY
If the unit is equipped with an optional Wireless Acquisition Module (WAM), the Universal Transceiver Key
(UTK) will be inserted into the secondary USB connector and held in place by the plastic housing ensure it
does not become dislodged (see image below). The UTK may need to be transferred to the new I/O board if
replaced during servicing activity.
Refer to the Item Identification Table for the correct item (#90 a/b), as there are two versions of
WAM/UTK pairs that must match (v1 to v1 or v2 to v2) for the wireless interface to operate properly.
Units utilizing the v2 UTK will have a round “2” label (item # 91) near the ECG input connector on the
housing.
To remove the Power Supply, disconnect the battery cables and then remove the 4 mounting screws
as shown in illustration below.
Item #32
56
UNIT DISASSEMBLY
Battery Replacement
CAUTION: Be careful not to short the positive and negative terminals of the battery with a metal
tool when removing the battery from the unit; as this could result in damage to the unit, or personal injury
to the repair technician.
The battery is installed at the factory by adhering it to the lower housing with two pieces of double stick foam tape.
To remove the battery, insert a small prying tool underneath the battery in the area shown below; then carefully lift
the battery edge to separate the battery from the lower housing. Removal of the battery will require a fair amount of
force, as the foam tape will typically tear in the center leaving the adhesive portions on the battery and lower
housing.
57
UNIT DISASSEMBLY
The adhesive left on the lower housing (as shown below) will need to be removed with a small scraping tool prior to
installation of the new battery.
NOTE: Welch Allyn recommends that a battery assembly that includes the pre-mounted foam tape be
obtained for battery replacements that are not performed in a Welch Allyn Service Center. This will ensure
that the battery is properly adhered to the lower housing to avoid it being dislodge during use, and will
allow for proper removal for routine maintenance. (Refer to the SERV ASSY item numbers on the item
listing at the end of this section of the manual).
Reassembly of the ELI 150c/250c can be performed by reversing the sequence of the previous disassembly
procedure.
When installing screws for all circuit boards, use a torque driver set to 3.5 in/lbs.
When installing screws for Upper Housing and Writer Assembly, use a torque driver set to 5.0 in/lbs.
58
UNIT DISASSEMBLY
The items highlighted in grey that are listed in the ELI 150c item description listing identify the serviceable level of
the device.
Some subcomponents of assemblies listed are not available as individual service items from Welch Allyn, the
assembly level item must be used for servicing purposes.
Item numbers below 100 are also used with the ELI 150c. Item numbers above 100 are specific to ELI 250c.
59
UNIT DISASSEMBLY
60
UNIT DISASSEMBLY
61
UNIT DISASSEMBLY
The items highlighted in grey that are listed in the ELI 250c item description listing identify the serviceable level of
the device.
Some subcomponents of assemblies listed are not available as individual service items from Welch Allyn, the
assembly level item must be used for servicing purposes.
Item numbers below 100 are also used with the ELI 150c. Item numbers above 100 are specific to ELI 250c.
62
UNIT DISASSEMBLY
63
UNIT DISASSEMBLY
64
UNIT DISASSEMBLY
Item numbers below 100 are also used with the ELI 150c. Item numbers above 100 are specific to ELI 250c.
2 25018-034-50
3 25020-060-50
4 25020-067-50
5 25020-076-50
6 26025-045-151
65
UNIT DISASSEMBLY
7 26025-073-50
8 26025-074-50
9 26025-077-170
9910-023-03
(radio only)
10 36025-102-150
66
UNIT DISASSEMBLY
107 36025-108-150
11 26025-110-150
128 26025-111-150
12 3171-009
13 3171-010
67
UNIT DISASSEMBLY
14 3225-003
15 3225-008
16 3375-004
17 3375-006
68
UNIT DISASSEMBLY
18 3600-008
19 3600-009
20 3600-016
21 4027-001
22 4027-002
69
UNIT DISASSEMBLY
23 4027-003
24 4800-006
25 5400-019
108 5450-004
26 5450-005
70
UNIT DISASSEMBLY
28 600-0515
29 6001-002-01
30 6020-060
31 6020-061
32 6020-062
33 6020-430-02
34 6020-735-02
35 6020-835
71
UNIT DISASSEMBLY
36 6020-835-02
37 6030-025
38 6100-004
39 6125-004
40 6125-017
41 6140-003
42 6141-003
72
UNIT DISASSEMBLY
43 6160-003
44 6320-003
45 6520-003
46 6545-007-01
47 6570-420-01
73
UNIT DISASSEMBLY
48 7400-019
49 7401-003
50 7403-001
51 7480-090
52 7495-001
58 8342-004-52
114 8342-004-53
59 8342-008-02
74
UNIT DISASSEMBLY
60 8342-009-01
61 8342-018-01
62 8342-019-01
63 8342-020-01
64 8347-004-51
65 8347-005-51
66 8347-006-51
75
UNIT DISASSEMBLY
67 8347-007-51
68 8347-009-50
69 8359-001-50
70 8359-002-50
71 8359-003-50
71b 413252
76
UNIT DISASSEMBLY
72 8359-004-51
119 8360-001-50
120 8360-002-50
121 8360-003-50
77
UNIT DISASSEMBLY
73 9025-049-02
9042-073-01 (ELI)
74 9042-073-02 (BUR)
9042-073-12 (MLBUR)
75 9050-059-05
76 9050-059-06
78
UNIT DISASSEMBLY
76 9050-059-09
76 9050-059-10
77 9050-059-07
9050-086-01 (ELI)
9050-086-02
(INMETRO)
78 9050-086-11 (MLBUR)
79 9910-017
79
UNIT DISASSEMBLY
80 9910-022
81 26025-099-151
26025-105-151 (w/o
COMM)
26025-105-152
(w/MODEM)
26025-105-153
(w/GSM)
26025-105-154
82 (w/LAN+WLAN) Example: 26025-105-151
83 SERV-ASSY-177-01
80
UNIT DISASSEMBLY
85 41000-028-53
92 728427
126 41000-028-52
81
DEVICE SPECIFICATIONS UNIT DISASSEMBLY
Feature Specifications
Instrument Type 12-lead electrocardiograph
Input Channels Simultaneous acquisition of all 12 leads
Standard Leads Acquired I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6
Waveform Display Backlit, ¼ VGA color LCD (320 x 240);
3, 4+4, or 6+6 lead presentation
Input Impedance Meets or exceeds requirements of ANSI/AAMI EC11
Input Dynamic Range
Electrode Offset Tolerance
Common Mode Rejection
Patient Leakage Current Meets or exceeds requirements of ANSI/AAMI ES1
Chassis Leakage Current
Digital Sampling Rate 40,000 s/sec/channel used for pacemaker spike detection;
1000 s/sec/channel used for recording and analysis
Special Functions Optional Welch Allyn VERITAS resting ECG interpretation with age and
gender specific algorithm; connectivity options for bidirectional
communication
Paper Type Perforated double Z-fold thermal paper; 108 mm (4”) wide, 200 sheets
Thermal Printer Computer-controlled dot array; 8 dots/mm
Thermal Printer Speeds 5, 10, 25, or 50 mm/s
Gain Settings 5, 10, or 20 mm/mV
Report Print Formats Standard or Cabrera; 3, 3+1, 3+3, or 6 channel
Rhythm Print Formats 3 or 6 channel with configurable lead groups
Keyboard Type Elastomeric keyboard with complete alphanumeric keys, soft-key menu, and
dedicated function keys
Frequency Response 0.05 to 300 Hz
Filters High-performance baseline filter; AC interference filter 50/60 Hz; low-pass
filters 40 Hz, 150 Hz, or 300 Hz
A/D Conversion 20 bits (1.17 microvolt LSB)
Device Classification Class I, Type CF defibrillation-proof applied parts
ECG Storage Internal storage up to 100 ECGs; optional expanded up to 200 ECGs
Weight 7.2 lbs. (3.3 kg) including battery (without paper)
Dimensions 11.25 x 11.5 x 3.75” (29.2 x 30.5 x 10.2 cm)
Power Requirements Universal AC power supply (100-240 VAC at 50/60 Hz) 110 VA; internal
rechargeable battery
82
SECTION 6
Feature Specifications
Instrument Type 12-lead electrocardiograph
Input Channels Simultaneous acquisition of all 12 leads
Standard Leads Acquired I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6
Waveform Display Backlit, ¼ VGA color LCD (320 x 240);
3, 4+4, or 6+6 lead presentation
Input Impedance Meets or exceeds requirements of ANSI/AAMI EC11
Input Dynamic Range
Electrode Offset Tolerance
Common Mode Rejection
Patient Leakage Current Meets or exceeds requirements of ANSI/AAMI ES1
Chassis Leakage Current
Digital Sampling Rate 40,000 s/sec/channel used for pacemaker spike detection;
1000 s/sec/channel used for recording and analysis
Special Functions Optional Welch Allyn VERITAS resting ECG interpretation with age and
gender specific algorithm; connectivity options for bidirectional
communication
Paper Type Perforated Z-fold thermal paper; A4 or 8.5 x 11” wide, 250 sheets
Thermal Printer Computer-controlled dot array; 8 dots/mm
ECG Storage Internal storage up to 100 ECGs; optional expanded up to 200 ECGs
Weight 11.25 lbs. (5.1 kg) including battery (without paper)
Dimensions 15.5 x 17 x 4” (39.4 x 43.2 x 10.2 cm)
Power Requirements Universal AC power supply (100-240 VAC at 50/60 Hz) 110 VA; internal
rechargeable battery
83
DEVICE SPECIFICATIONS
TROUBLESHOOTING
84
TROUBLESHOOTING
SECTION 6
Test Menu
Press F6 then Shift+ALT+x and enter the admin password.
1. Auto Test
a. This will test the LCD, Flash, ADC Readings and SDRAM.
2. Speaker
a. Tests the speaker
4. Writer Test
a. Tst P - Prints a single sheet test
b. Rhy P – prints a test Rhythm page
c. Load – Prints all the interpretation statements – 9 pages
7. Configure Modem
a. Allows configuration of the modem if installed
8. Set Configuration
85
TROUBLESHOOTING
9. Stats
a. F1 – CLR F – Clear all Flags. This will remove the Print, Transmit and Delete flags on all records
b. F2 – DUMP – This will copy all records to a USB flash drive without changing any flags
c. F3 – AQ LP – Acquire Forever Test. – Caution – This will acquire a ECG every 5 minutes. The
only way to stop it is to unplug the unit from AC and remove the battery.
d. F4 – Fill – This will fill the directory with records based on one record in the directory.
Media – F2
This is used when the WLAN module is replaced.
86
CONFORMANCE TESTING TROUBLESHOOTING
Conformance Testing
Conformance testing is to be performed by Authorized Welch Allyn Service Representatives to verify the device is
functioning correctly after repair operations have been performed. Testing results should be documented on the test
data record (TDR) at the end of this section of the manual. Include the following printouts: Configuration page,
Writer Test, ECG, and Noise Test.
Print the device configuration (attach to the Test Data Record (TDR)
Power Testing
Ensure battery is fully charged before performing these tests, voltage and current limits are based on a fully
charged battery.
Ensure there is no power connected to the UUT AC inlet.
Remove upper housing and writer assembly. Disconnect battery by pulling battery cable off of the red terminal.
NOTE: Based upon customer usage and age of battery, replace as needed.
This information can be found on the white “date code” sticker located on the battery (use the earliest date that
is not crossed out). Record Date on TDR
Measure battery voltage using a voltage meter; verify the meter reads greater than 12.5vdc.
Record result on TDR
Battery (load)
Measure the battery voltage using a volt meter and a power resistor load (10ohm, 20watt) in parallel with the
battery. After approximately 5 seconds, verify the meter reads greater than 11.7vdc. Record result on TDR
Off current
Connect a current meter in line with battery. With the UUT power off, verify the current meter reads less
than 100 uA. Record result on TDR
On current
Turn on the unit and verify the current meter reads less than 250 mA. Record result on TDR
AC charging current
Apply AC power to the unit and verify that the current draw from the battery reverses polarity and the value
starts decreasing as time increases. Record result on TDR
Disconnect the current meter and measure the battery charger output voltage between the red disconnected
battery cable and the negative terminal on the battery. It should read between 13.0vdc and 14.0vdc. Record
result on TDR
87
CONFORMANCE TESTING
Verify all power cables are reconnected properly and reassemble unit.
Functional Testing
AC LED/Display
Connect AC power cord to the unit and verify that the green AC LED (located to the left of the display)
illuminates continuous.
NOTE: The battery indicator will be clear when charging and will illuminate white when fully charged.
Turn the unit on and verify the text on display is clear and legible and there are no flickering or missing
lines/pixels. Record result on TDR
NOTE: This test should be performed with the AC power turned on. The test point, trim pot, and ground
are found on the front, upper left on the keyboard (see picture below).
1) Install paper into the unit with the cue mark approx. 1 – 2 inches away from the tear bar. Make sure that the
cue sensor is seeing white and not any markings on the paper.
2) Use a DMM to measure the DC voltage at test point (TP2/QSNS) on the keyboard with respect to ground
(GND1). Adjust R1 V_ADJ on the keyboard to between 1.95 V-DC and 2.05 V-DC at test point P2. Set this as
close to 2.0 V-DC as possible. Record result on TDR
3) Perform either the Writer test or print a test ECG. The paper should cue to the next sheet of paper, print and
then advance to the beginning of the next sheet of paper.
4) If the Writer test is performed the results should be compared with the test printout in this manual.
88
CONFORMANCE TESTING
Writer
Open and close the writer door to verify smooth operation. Verify that the door unlatches without sticking and
that it latches completely. From the main screen, simultaneously press shift+alt+RHY. Verify that a test page
is printed and the writer stops on the cue mark. The perforation of the paper should line up with the tear edge
on the writer. Assure there are no gaps in the printing and the print darkness is uniform across the entire page.
Verify the writer gears do not skip and paper is tracking properly (you may need to print multiple pages to
observe this). Record result on TDR
Connect an ECG simulator to the AM12 or WAM patient interface. Set the simulator to a known heart rate and
amplitude. Press the ECG key to capture an ECG. Verify there is an audible beep with each key press. Enter
Last name “PARCFL8” (Note: “PARCFL8” ensures the keyboard matrix is fully tested), then press F6 (Done).
Verify that 12 ECG traces print correctly and assess the printout quality. Ensure uniform darkness across entire
printout. Record result on TDR
Connect a Shorting Block (TF-0063) and adapter or equivalent to the AM12 or WAM patient interface. Set the
ECG gain on the unit to 20mm/mV. Print a rhythm strip (approx. 1 page). Verify that no channels have more
than 0.5mm of noise as measured by using Welch Allyn thermal paper (Smallest grid line = 1mm).
Record result on TDR.
Connect an AM12 or WAM patient cable to the patient input of the unit, with the other end connected to a lead
failure box (TF-0620 or equivalent) or patient simulator. Using the lead fail box, momentarily press each push
button to open the patient leads one at a time or disconnect one lead at a time from the simulator and verify the
display indicates an open lead condition for the corresponding lead. Record result on TDR
Lead Message
Right Arm ‐ LEADS OFF
Left Arm ‐ LA OFF
Left Leg ‐ LL OFF
V1 ‐ V1 OFF
V2 ‐ V2 OFF
V3 ‐ V3 OFF
V4 ‐ V4 OFF
V5 ‐ V5 OFF
V6 ‐ V6 OFF
All Leads off ‐ LEADS OFF
89
CONFORMANCE TESTING
The receiving station for modem, LAN and WLAN transmissions should be running Welch Allyn ELI-
Link software.
Refer to the ELI-Link user manual for proper configuration.
Verify successful transmission of all applicable communication options by acquiring ECG records that include
the
transmission method in the “Patient Name” field (such as Last Name = USBD) then subsequently transmitting
the
ECG record stored to a compatible receiving device. Consult the product user manual if needed to properly
configure the communication settings for each option present on the unit under test.
Successful transmission of the test records can be verified by viewing the ECG records in the unit directory
after
transmission and confirming they are marked as “transmitted” (as defined in the product user manual).
Record results on the TDR
- Modem
- LAN
- WLAN
- GSM/GPRS
- USB host (USB memory device needed)
-USBD
Device Cleaning
Clean unit per the instructions provided in the Maintenance & Cleaning section of the service manual.
Safety Testing
If the cardiograph housing was opened for repair or inspection work, the following safety tests should
be performed in accordance with the IEC 60601 standards.
The ELI150C and ELI250C are considered a Class 1 Type CF devices, intended to only be utilized
with the Welch Allyn AM12 or WAM patient input modules. Defibrillation isolation from the patient
is provided by the patient input modules, which are tested separately as part of the manufacturing
process (they are considered non-serviceable devices), therefore Hi-pot testing is not required for
these cardiograph models.
Earth Leakage
Enclosure Leakage
Non-conductive (fully insulated) chassis testing should be performed utilizing 200 cm2
conductive foil or equivalent, earth ground on AC input is utilized for functional earth (not
safety grounding).
Patient Leakage
Applied part – patient input (utilize Welch Allyn AM12 patient cable)
Patient Auxiliary Current
Applied part – patient input (utilize Welch Allyn AM12 patient cable)
90
CONFORMANCE TESTING
Unit Serial #:
Power Testing
Note Battery Age (If not possible enter N/A) / (week/year)
Battery (Open Circuit) Voltage VDC (>12.5vdc)
Battery (with Load) Voltage VDC (>11.7vdc)
Off Current uA (<100uA)
On Current mA (<250mA)
AC Charging Current PASS / FAIL (Circle One)
Battery Charger Output Voltage VDC (13.0-14.0vdc)
* Based upon customer usage and age of main battery, replace as needed.
□ Verify all power cables are properly reconnected and reassemble unit
Functional testing
AC LED/Display Functionality PASS / FAIL (Circle one)
Cue Sensor Calibration VDC (1.95-2.05VDC) PASS / FAIL (Circle one)
Writer Test PASS / FAIL (Circle one)
ECG & Keyboard Matrix Testing PASS / FAIL (Circle one)
ECG Noise Test PASS / FAIL (Circle one)
Lead Failure Test PASS / FAIL (Circle one)
Communication Option(s)
Modem PASS / FAIL / NA (Circle one)
LAN PASS / FAIL / NA (Circle one)
WLAN PASS / FAIL / NA (Circle one)
GSM/GPRS PASS / FAIL / NA (Circle one)
USB host PASS / FAIL / NA (Circle one)
USBD PASS / FAIL / NA (Circle one)
□ Device Cleaning
91
CONFORMANCE TESTING
Communication Options
LAN
WLAN
Modem
GSM/GPRS Mobile transmission (ELI150c Only)
Transmission to USB Thumb-drive
USB Mount to Windows PC (USB Device option)
Sync Media will attempt a connection of the preferred method first, and if it fails or times out, will attempt the
secondary mode of communication.
DPAC Failure
This error message occurs if the DPAC fails initialization. This likely indicates a hardware problem preventing
communication with the Module or a hardware problem with the Module itself.
DHCP Failure
This error message occurs if the Module failed to acquire an IP address via DHCP.
NOTE: It is possible that the Module can still transmit data with an incorrect gateway or subnet mask
depending on the path through the network the data is being transmitted.
92
Communication Options (Software only)
- LAN
- USB (USB Device option)
Modem
Shortcuts
F6 (More)
Shift/Alt/X
Enter ‘admin’ password, Press Enter key
F6 (More)
F1/#1 (Directory of Stored ECG’s)
Shift/Alt/D
93