ELI 150c - 250c Szervice Manuals-EnGLITPDF

Download as pdf or txt
Download as pdf or txt
You are on page 1of 97

Welch Allyn®

ELI® 150c/250c
12-LEAD RESTING ELECTROCARDIOGRAPH
SERVICE MANUAL

Manufactured by Welch Allyn, Inc., Skaneateles Falls, NY U.S.A.

CAUTION: Federal law restricts this device to sale by or on the order of a physician.
©2019 Welch Allyn This document contains confidential information that belongs to Welch Allyn, Inc. No part of this
document may be transmitted, reproduced, used, or disclosed outside of the receiving organization without the express
written consent of Welch Allyn, Inc. Welch Allyn is a registered trademark of Welch Allyn, Inc. E-Scribe, ELI, and
VERITAS are trademarks of Welch Allyn, Inc. Cisco® is the registered trademark of Cisco Systems, Inc. DICOM® is
the registered trademark of the National Electrical Manufacturers Association for its standards publications relating
to digital communications of medical information.

For patent information, please visit www.welchallyn.com/patents


For information about any Welch Allyn product, visit: https://www.welchallyn.com/en/about-us/locations.html
Customer Service and Technical Support: https://www.welchallyn.com/en/other/contact-us.html 1.888.667.8272,
mor_tech.support@hillrom.com

9516-177-50-ENG Rev N
Revision Date: 2019-20

(150c) 901129 ELECTROCARDIOGRAPH


(250c) 901131 ELECTROCARDIOGRAPH

EU IMPORTER

Welch Allyn, Inc. Welch Allyn Limited


4341 State Street Road Navan Business Park, Dublin Road,
Skaneateles Falls, NY 13153 USA Navan, Co. Meath C15 AW22
Ireland
www.welchallyn.com
TABLE OF CONTENTS
NOTICES ........................................................................................................................................................................... 1
MANUFACTURER’S RESPONSIBILITY .............................................................................................................................................. 1
RESPONSIBILITY OF THE CUSTOMER .............................................................................................................................................. 1
EQUIPMENT IDENTIFICATION....................................................................................................................................................... 1
COPYRIGHT AND TRADEMARK NOTICES ......................................................................................................................................... 1
OTHER IMPORTANT INFORMATION .............................................................................................................................................. 1
NOTICE TO EU USERS AND/OR PATIENTS ...................................................................................................................................... 1
DISPOSAL ................................................................................................................................................................................ 1
WARRANTY INFORMATION.............................................................................................................................................. 2
YOUR WELCH ALLYN WARRANTY................................................................................................................................................. 2
USER SAFETY INFORMATION............................................................................................................................................ 3
WARNING(S) ........................................................................................................................................................................ 3
CAUTION(S)............................................................................................................................................................................. 6
NOTE(S) ................................................................................................................................................................................. 7
WIRELESS DATA TRANSMISSION .................................................................................................................................................. 8
WLAN OPTION ....................................................................................................................................................................... 8
EQUIPMENT SYMBOLS AND MARKINGS ......................................................................................................................... 11
SYMBOL DELINEATION ............................................................................................................................................................. 11
ELECTROMAGNETIC COMPATIBILITY (EMC) .................................................................................................................... 13
GUIDANCE AND MANUFACTURER’S DECLARATION: ELECTROMAGNETIC EMISSIONS ............................................................................. 14
GUIDANCE AND MANUFACTURER’S DECLARATION: ELECTROMAGNETIC IMMUNITY ............................................................................. 14
GUIDANCE AND MANUFACTURER’S DECLARATION: ELECTROMAGNETIC IMMUNITY ............................................................................. 15
RECOMMENDED SEPARATION DISTANCES BETWEEN PORTABLE AND MOBILE RF COMMUNICATIONS EQUIPMENT AND THE EQUIPMENT ....... 16
REGULATORY RADIO COMPLIANCE ............................................................................................................................................. 17
Federal Communications Commission (FCC) ................................................................................................................. 17
Industry Canada (IC) Emissions ..................................................................................................................................... 18
Declaración de conformidad Mexico ............................................................................................................................. 18
European Union ............................................................................................................................................................. 19
RADIO COMPLIANCE TABLE ...................................................................................................................................................... 20
MAINTENANCE & CLEANING .......................................................................................................................................... 22
RECOMMENDED TOOLS AND SUPPLIES ........................................................................................................................................ 22
PREVENTIVE MAINTENANCE ..................................................................................................................................................... 23
DEVICE CLEANING & DISINFECTING ............................................................................................................................................ 27
ELI 150/250C PREVENTIVE MAINTENANCE REPORT ..................................................................................................................... 28
DEVICE CONFIGURATION ............................................................................................................................................... 29
SETTING TECHNICIAN PASSWORD .............................................................................................................................................. 29
CONFIGURATION MENUS ......................................................................................................................................................... 29
SUMMARY OF CONFIGURATION MENUS ...................................................................................................................................... 30
CONFIGURATION SETTINGS ....................................................................................................................................................... 33
UNIT DISASSEMBLY ........................................................................................................................................................ 42
ELI 150c ........................................................................................................................................................................ 42
REMOVAL OF THE UNIT FROM CART ........................................................................................................................................... 43
COVER ASSEMBLY REMOVAL..................................................................................................................................................... 43
WRITER REMOVAL .................................................................................................................................................................. 44
KEYBOARD REMOVAL .............................................................................................................................................................. 51
INSTALLATION OF THE RIBBON CABLE TO THE KEYPAD ASSEMBLY ..................................................................................................... 54
I/O BOARD REMOVAL ............................................................................................................................................................. 55
BATTERY REPLACEMENT ........................................................................................................................................................... 57
DEVICE SPECIFICATIONS ................................................................................................................................................. 82
ELI 150C SPECIFICATIONS ........................................................................................................................................................ 82
ELI 250C SPECIFICATIONS ........................................................................................................................................................ 83
TROUBLESHOOTING ....................................................................................................................................................... 84
SYSTEM TROUBLESHOOTING CHART ........................................................................................................................................... 84
ECG TROUBLESHOOTING CHART ............................................................................................................................................... 84
TRANSMISSION TROUBLESHOOTING CHART ................................................................................................................................. 84
TRANSMISSION TROUBLESHOOTING CHART (CONTINUED) ........................................................................................................... 85
PAPER CUE FAULT ................................................................................................................................................................... 85
TEST MENU........................................................................................................................................................................... 85
CONFORMANCE TESTING ............................................................................................................................................... 87
CONFORMANCE TESTING ......................................................................................................................................................... 87
POWER TESTING..................................................................................................................................................................... 87
FUNCTIONAL TESTING.............................................................................................................................................................. 88
DEVICE CLEANING................................................................................................................................................................... 90
SAFETY TESTING ..................................................................................................................................................................... 90
ELI 150C/250C TEST DATA RECORD ........................................................................................................................................ 91
ELI 150C/250C COMMUNICATION OPTIONS .................................................................................................................. 92
COMMUNICATION OPTIONS ..................................................................................................................................................... 92
COMMUNICATION ERROR MESSAGES ......................................................................................................................................... 92
COMMUNICATION OPTIONS (SOFTWARE ONLY) ............................................................................................................................ 93
COMMUNICATION OPTIONS (HARDWARE + SOFTWARE) ................................................................................................................ 93
Modem ........................................................................................................................................................................... 93
NOTICES

Manufacturer’s Responsibility

Welch Allyn, Inc. is responsible for the effects on safety and performance only if:

• Assembly operations, extensions, readjustments, modifications, or repairs are carried out only by persons
authorized by Welch Allyn, Inc.

• The device is used in accordance with the instructions for use.

Responsibility of the Customer

The user of this device is responsible for ensuring the implementation of a satisfactory maintenance schedule.
Failure to do so may cause undue failure and possible health hazards.

Equipment Identification

Welch Allyn, Inc. equipment is identified by a serial and reference number on the back of the device. Care should
be taken so that these numbers are not defaced.

Copyright and Trademark Notices

This document contains information that is protected by copyright. All rights are reserved. No part of this
document may be photocopied, reproduced, or translated to another language without prior written consent of
Welch Allyn, Inc.

Other Important Information

The information in this document is subject to change without notice.

Welch Allyn, Inc. makes no warranty of any kind with regard to this material including, but not limited to,
implied warranties of merchantability and fitness for a particular purpose. Welch Allyn, Inc. assumes no
responsibility for any errors or omissions that may appear in this document. Welch Allyn, Inc. makes no
commitment to update or to keep current the information contained in this document.

Notice to EU Users and/or Patients

Any serious incident that has occurred in relation to the device, should be reported to the manufacturer and the
competent authority of the Member State in which the user and/or patient is established.

Disposal

This product and its accessories must be disposed of according to the local laws and regulations. Do not dispose of this
product as unsorted municipal waste. For more specific disposal information see www.welchallyn.com/weee.

1
WARRANTY INFORMATION

Your Welch Allyn Warranty


WELCH ALLYN, INC. (hereafter referred to as “Welch Allyn”) warrants that components within Welch Allyn
products (hereafter referred to as “Product/s”) will be free from defects in workmanship and materials for the
number of years specified on documentation accompanying the product, or previously agreed to by the purchaser
and Welch Allyn, or if not otherwise noted, for a period of twenty-four (24) months from the date of shipment.
Consumable, disposable or single use products such as, but not limited to, PAPER or ELECTRODES are warranted
to be free from defects in workmanship and materials for a period of 90 days from the date of shipment or the date
of first use, whichever is sooner.
Reusable product such as, but not limited to, BATTERIES, BLOOD PRESSURE CUFFS, BLOOD PRESSURE
HOSES, TRANSDUCER CABLES, Y-CABLES, PATIENT CABLES, LEAD WIRES, MAGNETIC STORAGE
MEDIUMS, CARRY CASES or MOUNTS, are warranted to be free from defects in workmanship and materials for
a period of 90 days. This warranty does not apply to damage to the Product/s caused by any or all of the following
circumstances or conditions:
a) Freight damage;

b) Parts and/or accessories of the Product/s not obtained from or approved by Welch Allyn;

c) Misapplication, misuse, abuse, and/or failure to follow the Product/s instruction sheets
and/or information guides;

d) Accident; a disaster affecting the Product/s;

e) Alterations and/or modifications to the Product/s not authorized by Welch Allyn;

f) Other events outside of Welch Allyn’s reasonable control or not arising under normal operating conditions.
THE REMEDY UNDER THIS WARRANTY IS LIMITED TO THE REPAIR OR REPLACEMENT WITHOUT
CHARGE FOR LABOR OR MATERIALS, OR ANY PRODUCT/S FOUND UPON EXAMINATION BY
WELCH ALLYN TO HAVE BEEN DEFECTIVE. This remedy shall be conditioned upon receipt of notice by Welch
Allyn of any alleged defects promptly after discovery thereof within the warranty period. Welch Allyn’s obligations
under the foregoing warranty will further be conditioned upon the assumption by the purchaser of the Product/s (i) of
all carrier charges with respect to any Product/s returned to Welch Allyn’s principal place or any other place as
specifically designated by Welch Allyn or an authorized distributor or representative of Welch Allyn, and (ii) all risk
of loss in transit. It is expressly agreed that the liability of Welch Allyn is limited and that Welch Allyn does not
function as an insurer. A purchaser of a Product/s, by its acceptance and purchase thereof, acknowledges and agrees
that Welch Allyn is not liable for loss, harm, or damage due directly or indirectly to an occurrence or consequence
therefrom relating to the Product/s. If Welch Allyn should be found liable to anyone under any theory (except the
expressed warranty set forth herein) for loss, harm, or damage, the liability of Welch Allyn shall be limited to the
lesser of the actual loss, harm, or damage, or the original purchase price of the Product/s when sold.
EXCEPT AS SET FORTH HEREIN WITH RESPECT TO REIMBURSEMENT OF LABOR CHARGES, A
PURCHASER’S SOLE EXCLUSIVE REMEDY AGAINST WELCH ALLYN FOR CLAIMS RELATING TO THE
PRODUCT/S FOR ANY AND ALL LOSSES AND DAMAGES RESULTING FROM ANY CAUSE SHALL BE
THE REPAIR OR REPLACEMENT OF DEFECTIVE PRODUCT/S TO THE EXTENT THAT THE DEFECT IS
NOTICED AND WELCH ALLYN IS NOTIFIED WITHIN THE WARRANTY PERIOD. IN NO EVENT,
INCLUDING THE CLAIM FOR NEGLIGENCE, SHALL WELCH ALLYN BE LIABLE FOR INCIDENTAL,
SPECIAL, OR CONSEQUENTIAL DAMAGES, OR FOR ANY OTHER LOSS, DAMAGE, OR EXPENSE OF
ANY KIND, INCLUDING LOSS OF PROFITS, WHETHER UNDER TORT, NEGLIGENCE OR STRICT
LIABILITY THEORIES OF LAW, OR OTHERWISE. THIS WARRANTY IS EXPRESSLY IN LIEU OF ANY
OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING, BUT NOT LIMITED TO THE IMPLIED
WARRANTY OF MERCHANTABILITY AND THE WARRANTY OF FITNESS FOR A PARTICULAR
PURPOSE.

2
USER SAFETY INFORMATION

WARNING: Means there is the possibility of personal injury to you or others.

Caution: Means there is the possibility of damage to the device.

Note: Provides information to further assist in the use of the device.

WARNING(S)

 This manual gives important information about the use and safety of this device. Deviating from operating
procedures, misuse or misapplication of the device, or ignoring specifications and recommendations could
result in increased risk of harm to users, patients and bystanders, or damage to the device.

 Device captures and presents data reflecting a patient’s physiological condition that when reviewed by a trained
physician or clinician can be useful in determining a diagnosis; however, the data should not be used as a sole
means for determining a patient’s diagnosis.

 Users are expected to be licensed clinical professionals knowledgeable about medical procedures and patient
care, and adequately trained in the use of this device. Before attempting to use this device for clinical
applications, the operator must read and understand the contents of the user manual and other accompanying
documents. Inadequate knowledge or training could result in increased risk of harm to users, patients and
bystanders, or damage to the device. Contact Welch Allyn service for additional training options.

 To ensure that electrical safety is maintained during operation from AC (~) power, the device must be plugged
into a hospital-grade outlet.

 Only use parts and accessories supplied with the device and available through Welch Allyn, Inc.

 Patient cables intended for use with the device include series resistance (9 Kohm minimum) in each lead for
defibrillation protection. Patient cables should be checked for cracks or breakage prior to use.

 Conductive parts of the patient cable, electrodes, and associated connections of type CF applied parts, including
the neutral conductor of the patient cable and electrodes, should not come into contact with other conductive
parts including earth ground.

 ECG electrodes could cause skin irritation; patients should be examined for signs of irritation or inflammation.

 To avoid the possibility of serious injury or death during patient defibrillation, do not come into contact with
device or patient cables. Additionally, proper placement of defibrillator paddles in relation to the electrodes is
required to minimize harm to the patient.

3
USER SAFETY INFORMATION

 This device was designed to use the electrodes specified in the user manual. Proper clinical procedure must be
employed to prep the electrode sites and to monitor the patient for excessive skin irritation, inflammation, or
other adverse reactions. Electrodes are intended for short-term use and should be removed from the patient
promptly following testing.

 To avoid potential for spread of disease or infection, single-use disposable components (e.g., electrodes) must
not be reused. To maintain safety and effectiveness, electrodes must not be used beyond their expiration date.

 A possible explosion hazard exists. Do not use the device in the presence of a flammable anesthetic mixture.

 Where the integrity of external protective earth conductor arrangement is in doubt, the device shall be operated
from its internal electrical power source.

 To improve immunity to potential interfering electromagnetic signals, shielded cabling is recommended when
connecting the device to a network.

 Medical devices have been designed to have a higher degree of protection against electric shock than, for
instance, information technology equipment because patients often are connected to multiple devices and
also may be more prone to the adverse effect of electric currents than healthy persons. All equipment that is
connected to the patient, can be touched by the patient, or can be touched by another person while that
person touches the patient at the same time, should have the same level of protection against electric shock
as medical equipment. The ELI 150c and ELI 250c are medical devices that have been designed to be
connected to other devices for the purpose of receiving and transmitting data. Certain measures must be
taken to prevent the risk of excessive electric current flow through the operator or patient when connected:

 All electrical equipment that is not medical electrical equipment must be placed outside of the “patient
environment,” defined by applicable safety standards to be at least 1.5 meters (5 feet) from the patient.
Alternatively, non-medical equipment may be provided with additional protection such as an additional
protective earth connection.

 All medical electrical equipment that has a physical connection to the ELI 150c/ELI 250c or the patient
or is in the patient environment must comply with applicable safety standards for medical electrical
devices.

 All electrical equipment that is not medical electrical equipment and has a physical connection to the
ELI 150c/ELI 250c must comply with applicable safety standards, such as IEC 60950 for information
technology equipment. This includes information network equipment connected through the LAN
connector.

 Conductive (metal) parts that can be touched by the operator in normal use and that are connected to non-
medical equipment should not be brought into the patient environment. Examples are connectors for
shielded Ethernet or USB cables.

 If multiple devices are connected to each other or to the patient, device chassis and patient leakage currents
may be increased, and should be measured for compliance with applicable standards for medical electrical
systems.

 Avoid the use of portable multiple socket outlets. If used and not compliant with medical electrical
device standards, an additional protective earth connection is required.

4
USER SAFETY INFORMATION

 After defibrillation pulse, electrocardiograph has a maximum recovery time of 5 seconds.

 To prevent electric shock due to unequal ground potentials that may exist between points of a distributed
network system or fault conditions in external network connected equipment, network cable shielding (where
used) must be connected to protective earth ground appropriate to the area where the device is used.

 The device has not been designed for use with high-frequency (HF) surgical equipment and does not provide a
protective means against hazards to the patient.

 When the 40 Hz filter is used, the frequency response requirement for diagnostic ECG equipment cannot be
met. The 40 Hz filter significantly reduces high-frequency components of the ECG and pacemaker spike
amplitudes, and is recommended only if high-frequency noise cannot be reduced by proper procedures.

 The quality of the signal produced by the device may be adversely affected by the use of other medical
equipment, including but not limited to defibrillators and ultrasound machines.

 For proper operation and the safety of users or patients and bystanders, equipment and accessories must be
connected only as described in this manual. Do not connect a telephone line cable to the LAN connector.

 Some Welch Allyn electrocardiographs can be equipped with a GPRS (cellular modem) or wireless LAN
(WLAN) module for transmitting ECG records. Device labeling and the presence of an antenna port will
indicate if your device is equipped with such a module. If so equipped, the following notices apply:

 The GPRS module operates in allocated frequency bands depending on the model. Identification of
the installed GPRS module can be found on a label on the bottom of the device.
 MultiTech Systems, Inc. Model MTSMC-H5: 850/900/1700(AWS)/1900/2100 MHz
 The WLAN identification can be found on a label on the bottom of the device.
 B&B Electronics Model WLNN-AN-MR551: 2400 MHz
(model subject to change without notice)

5
USER SAFETY INFORMATION

 Use of the GPRS or WLAN module may interfere with other equipment operating in the vicinity. Check with
local authorities or spectrum management officials in your facility to determine if restrictions apply to the use
of this feature in your area.

 Do not transmit via the GPRS or WLAN module with a missing or damaged antenna. Replace a damaged
antenna immediately.

 Use only the antenna supplied for use with this device. Unauthorized antennas, modifications, or attachments
could damage the GSM module and may contravene local RF emission regulations or invalidate type approval.

 To ensure compliance with current regulations limiting both maximum RF output power and human exposure
to radio frequency radiation, a separation distance of at least 20 cm must be maintained between the device's
antenna and the head and body of the user and any nearby persons at all times. To help prevent degradation of
RF signal and to avoid excess RF energy absorption, do not touch the antenna during data transmission.

 The GPRS and WLAN modules comply with applicable RF safety standards including standards and
recommendations for the protection of public exposure to RF electromagnetic energy that have been established
by governmental bodies and other qualified organizations, such as the following:
 Federal Communications Commission (FCC)
 Directives of the European Community
 Directorate General V in Matters of Radio Frequency Electromagnetic Energy

 Proper functioning backup items such as spare lead wires, front-end device, and other equipment are
recommended on hand to prevent delayed treatment due to an inoperable device.

Caution(s)

 To prevent possible damage to the keyboard, do not use sharp or hard objects to depress keys, only
use fingertips.

 Do not attempt to clean the device or patient cables by submersing into a liquid, autoclaving, or steam cleaning
as this may damage equipment or reduce its usable life. Wipe the exterior surfaces with a warm water and mild
detergent solution and then dry with a clean cloth. Use of unspecified cleaning/disinfecting agents, failure to
follow recommended procedures, or contact with unspecified materials could result in increased risk of harm to
users, patients and bystanders, or damage to the device.

 No user-serviceable parts inside. Screw removal by qualified service personnel only. Damaged or suspected
inoperative equipment must be immediately removed from use and must be checked/repaired by qualified
service personnel prior to continued use.

 The rechargeable internal battery is a sealed lead-acid type and it is totally maintenance free. If the battery
appears to become defective, refer to Welch Allyn Service Department.

 Do not pull or stretch patient cables as this could result in mechanical and/or electrical failures. Patient cables
should be stored after forming them into a loose loop.

 No calibration or special equipments are needed for the proper operation or maintenance of the device.

● When necessary, dispose of the device, its components and accessories (e.g., batteries, cables, electrodes),
and/or packing materials in accordance with local regulations.
 Use only No. 26 AWG or larger telecommunication line cord.

6
USER SAFETY INFORMATION

Note(s)

 Patient movements may generate excessive noise that may affect the quality of the ECG traces and the proper
analysis performed by the device.

 Proper patient preparation is important to proper application of ECG electrodes and operation of the device.

 The algorithm detecting electrode misplacements is based on normal physiology and ECG lead order, and tries
to identify the most likely switch; however, it is advisable to check the other electrode positions in the same
group (limb or chest).

 There is no known safety hazard if other equipment, such as pacemakers or other stimulators, is used
simultaneously with the device; however, disturbance to the signal may occur.

 If electrode is not properly connected to the patient, or one or more of the patient cable lead wires is damaged,
display will indicate a lead fault for the lead(s) where the condition is present and if the signal is being printed,
the respective lead(s) will print out as a square wave.

 As defined by IEC 60601-1 and IEC 60601-2-25, the device is classified as follows:
 Class I equipment or internally powered.
 Type CF defibrillation-proof applied parts.
 Ordinary equipment.
 Equipment not suitable for use in the presence of a flammable anesthetic mixture.
 Continuous operation.

NOTE: From a safety perspective, per IEC 60601-1 and derivative standards/norms, this device is
declared to be “Class I” and uses a three-prong inlet to ensure an earth connection is made along with
mains. The ground terminal on the mains inlet is the only protective earth point in the device. Exposed
metal accessible during normal operation is double insulated from mains. Internal connections to earth
ground are functional earth.

 This device is intended to be used in a hospital or doctor’s office setting, and should be used and stored
according to the environmental conditions specified below:

Operating temperature: +10° to +40°C (+50° to +104°F)


Operating humidity: 10% to 95% RH, non-condensing

Storage temperature: -40° to +70°C (-40° to +158°F)


Storage humidity: 10% to 95% RH, non-condensing

Atmospheric pressure: 500 hPa to 1060 hPa

 WAM™ (wireless acquisition module) must be paired to electrocardiograph before operation.

 Device must be configured at the factory for use with the WAM.

 After operating the device using battery power, always reconnect the power cord. This ensures that the batteries
will be automatically recharged for the next time you use the device.

7
USER SAFETY INFORMATION

 The device is UL classified:

WITH RESPECT TO ELECTRIC SHOCK, FIRE AND MECHANICAL


HAZARDS ONLY IN ACCORDANCE WITH UL60601-1, IEC60601-1,
CAN/CSA C22.2 No. 60601-1, IEC 60601-1-1, CAN/CSA C22.2 No.
60601-1-1-02, IEC60601-2-25 AND CAN/CSA C22.2 No. 601.2.25-94.

 The device is a member of the ELI 1xx or ELI 2xx Series 2 electrocardiograph family.

Wireless Data Transmission

 Some Welch Allyn electrocardiographs can be equipped with an optional wireless data transmission module
(WLAN or GSM/GPRS mobile). Both these technologies use radios to transmit data to a Welch Allyn
receiving application. Due to the nature of radio transmissions, it’s possible that, due to the characteristics of
the environment where the device is located, some other RF sources may interfere with the transmission
generated by the device. Welch Allyn has tested the coexistence of the device with other devices that can
interfere such as devices using WLAN, Bluetooth radio, and/or cell phones. Although the current technology
allows a very successful rate of transmission, it’s possible that in some rare occurrences, the system may not
perform at its best resulting in a “failed transmission.” When this occurs, patient data will not be erased from
the device nor stored in the receiving application, ensuring that partial or corrupted data are not made available
to the receiving station. If the failure mode persists the user should move to a position where the RF signals
may propagate better and allow successful transmissions.

WLAN Option
 Wireless options transmit in the 2.4 GHz or 5ghz range. Other nearby wireless devices may cause interference.
If possible, move or turn off other devices to minimize potential interference.

 The Wireless LAN module used is compliant with the IEEE 802.11 a, b, g and n standards.

 Access Points used should respect IEEE 802.11 standards as well as local Radio Frequency regulations. The
device will scan the available channels and connect to the Access Point on the channel where the SSID that is
configured on the device is available.

8
USER SAFETY INFORMATION

 The following table shows the radio channels allocated in different geographic areas in the world. For bands
802.11b and g, only channels 1, 6, 11 and 14 (Japan only) are non-overlapping; for band 802-11a, channels
shown represent non-overlapping channel numbers.

Band Typical Power Region Frequency No. of Channel numbers


Range (GHz) channels
15 dBm / 32 mW USA/Canada 2.401 - 2.473 11 1 – 11
802.11b Europe 2.401 - 2.483 13 1 – 13
Japan 2.401 - 2.495 14 1 – 14
13 dBm / 18 mW USA/Canada 2.401 - 2.473 11 1 – 11
802.11g Europe 2.401 - 2.483 13 1 – 13
Japan 2.401 - 2.483 13 1 – 13
17 dBm / 50 mW USA/Canada 5.15 - 5.35, 13 36,40,44,48,52,56,60,64,149,
5.725 - 5.825 153,157,161,165
Europe 5.15 - 5.35, 19 36,40,44,48,52,56,60,64,100,
5.47 - 5.725 104,108,112,116,120,124,
128,132,136,140
802.11a Japan 4.91 – 4.99, 23 36,40,44,48,52,56,60,64,100,
5.15 - 5.35, 104,108,112,116,120,124,
5.47 - 5.725 128,132,136,140,184188,
192,196
China 5.725 - 5.825 5 149,153,157,161,165

9
USER SAFETY INFORMATION

 The following table lists the frequency allocated for each channel used by the WLAN option.

Channel Center Frequency Frequency Spread


1 2412 MHz 2399.5 MHz - 2424.5 MHz
2 2417 MHz 2404.5 MHz - 2429.5 MHz
3 2422 MHz 2409.5 MHz - 2434.5 MHz
4 2427 MHz 2414.5 MHz - 2439.5 MHz
5 2432 MHz 2419.5 MHz - 2444.5 MHz
6 2437 MHz 2424.5 MHz - 2449.5 MHz
7 2442 MHz 2429.5 MHz - 2454.5 MHz
8 2447 MHz 2434.5 MHz - 2459.5 MHz
9 2452 MHz 2439.5 MHz - 2464.5 MHz
10 2457 MHz 2444.5 MHz - 2469.5 MHz
11 2462 MHz 2449.5 MHz - 2474.5 MHz
12 2467 MHz 2454.5 MHz - 2479.5 MHz
13 2472 MHz 2459.5 MHz - 2484.5 MHz
14 2484 MHz 2471.5 MHz – 2496.5 MHz

 In order to achieve the best transmission rate, it is necessary that the facility where the device is operated can
provide good area coverage. Please consult the IT personnel of the facility to verify the proper WLAN
availability in the area where the device will be used.

 RF wave propagation may be blocked or reduced by the environment where the device is used. Most common
areas where this may occur are: shielded rooms, elevators, underground rooms. In all the above situations, it is
recommended to move the device to a proper location and verify with the IT personnel of the facility the areas
where the WLAN signals are available.

10
EQUIPMENT SYMBOLS AND MARKINGS

Symbol Delineation

CAUTION The caution statements in this manual identify conditions or practices


that could result in damage to the equipment or other property, or loss of data.

WARNING The warning statements in this manual identify conditions or


practices that could lead to illness, injury, or death. In addition, when used on a
patient applied part, this symbol indicates defibrillation protection is in the
cables. Warning symbols will appear with a grey background in a black and
white document.

Alternating current

Protective earth

Telephone line (modem)

Network (LAN)

Defibrillator-proof type CF applied part

USB port

Input

ON/OFF (power)

Stop (of action)

Shift key (to enter upper case text)

Enter key (accept data/return)

11
EQUIPMENT SYMBOLS AND MARKINGS

Initiate printing of 12-lead ECG

Initiate printing of continuous rhythm strip

Transmit, receive and time sync operation depending upon configuration


settings

Do not dispose as unsorted municipal waste. Requires separate handling


for waste disposal according to local requirements.

Antenna

Indicates compliance to applicable European Union directives

Follow instructions/directions for use (DFU) – mandatory action. A copy of


the DFU is available on this website. A printed copy of the DFU can be
ordered from Welch Allyn for delivery within 7 calendar days.

Medical Device

Model Identifier

Reorder Number

Non-ionizing electromagnetic radiation

12
ELECTROMAGNETIC COMPATIBILITY (EMC)

Electromagnetic compatibility with surrounding devices should be assessed when using the device.

An electronic device can either generate or receive electromagnetic interference. Testing for electromagnetic
compatibility (EMC) has been performed on the device according to the international standard for EMC for medical
devices (IEC 60601-1-2). This IEC standard has been adopted in Europe as the European Norm (EN 60601-1-2).

The device should not be used adjacent to, or stacked on top of other equipment. If the device must be used adjacent
to or stacked on top of other equipment, verify that the device operates in an acceptable manner in the configuration
in which it will be used.

Fixed, portable, and mobile radio frequency communications equipment can affect the performance of medical
equipment. See the appropriate table for recommended separation distances between the radio equipment and the
device.

The use of accessories, transducers, and cables other than those specified by Welch Allyn may result in
increased emissions or decreased immunity of the equipment.

13
ELECTROMAGNETIC COMPATIBILITY (EMC)

Guidance and Manufacturer’s Declaration: Electromagnetic Emissions


The equipment is intended for use in the electromagnetic environment specified in the table below. The customer or
the user of the equipment should ensure that it is used in such an environment.

Emissions Test Compliance Electromagnetic Environment: Guidance

RF Emissions CISPR 11 Group 1 The equipment uses RF energy only for its internal function.
Therefore, its RF emissions are very low and not likely to cause
any interference in nearby electronic equipment.
RF Emissions CISPR 11 Class A The equipment is suitable for use in all establishments other
than domestic and those directly connected to the public low-
voltage power supply network that supplies buildings used for
Harmonic Emissions Complies domestic purposes.
IEC 61000-3-2

Voltage Fluctuations/ Complies


Flicker Emissions
IEC 61000-3-3

Guidance and Manufacturer’s Declaration: Electromagnetic Immunity


The equipment is intended for use in the electromagnetic environment specified in the table below. The customer or
the user of the equipment should ensure that it is used in such an environment.

Emissions Test Compliance Compliance Level Electromagnetic Environment: Guidance

Electrostatic +/- 6 kV contact +/- 6 kV contact Floors should be wood, concrete, or ceramic
discharge (ESD) +/- 8 kV air +/- 8 kV air tile. If floors are covered with synthetic
IEC 61000-4-2 material, the relative humidity should be at
least 30%.
Electrical fast +/- 2 kV for +/- 2 kV for Mains power quality should be that of a typical
transient/burst power supply lines power supply lines commercial or hospital environment.
IEC 61000-4-4 +/- 1 kV for +/- 1 kV for
input/output lines input/output lines
Surge +/- 1 kV differential +/- 1 kV differential Mains power quality should be that of a typical
IEC 61000-4-5 mode mode commercial or hospital environment.
+/- 2 kV common +/- 2 kV common
mode mode
Voltage dips, <5% UT <5% UT Mains power quality should be that of a typical
short (>95% dip in UT) (>95% dip in UT) commercial or hospital environment.
interruptions, for 0.5 cycle for 0.5 cycle
and voltage 40% UT 40% UT
variations on (60% dip in UT) (60% dip in UT)
power supply for 5 cycles for 5 cycles
input lines
IEC 61000-4-11
Power frequency 3 A/m 3 A/m Power frequency magnetic fields should be at
(50/60 Hz) levels characteristic of a typical location in a
magnetic field typical commercial or hospital environment.

NOTE: UT is the AC Mains voltage prior to application of the test level.

14
ELECTROMAGNETIC COMPATIBILITY (EMC)

Guidance and Manufacturer’s Declaration: Electromagnetic Immunity

The equipment is intended for use in the electromagnetic environment specified in the table below. The customer or
the user of the equipment should ensure that it is used in such an environment.

IEC 60601 Test Compliance


Emissions Test Level Level Electromagnetic Environment: Guidance

Portable and mobile RF communications equipment


should be used no closer to any part of the equipment,
including cables, than the recommended separation
distance calculated from the equation applicable to the
frequency of the transmitter.

Recommended separation distance

d  
3.5 
Conducted RF 3 Vrms 3 Vrms  P
3Vrms 
IEC 61000-4-6 150 kHz to 150 kHz to
80 MHz 80 MHz
 3.5  80 MHz to 800 MHz
d   P
3V / m 
Radiated RF 3 V/m 3 V/m
IEC 61000-4-3 80 MHz to 80 MHz to
 7  800 MHz to 2.5 GHz
2.5 GHz 2.5 GHz d   P
3V / m 

Where P is the maximum output power rating of the


transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in meters (m).

Field strengths from fixed RF transmitters, as


determined by an electromagnetic site surveya, should
be less than the compliance level in each frequency
rangeb.

Interference may occur in the vicinity of equipment


marked with the following symbol:

a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios,
amateur radios, AM and FM radio broadcast, and TV broadcast cannot be predicted theoretically with accuracy. To assess
the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the
measured field strength in the location in which the equipment is used exceeds the applicable RF compliance level above, the
equipment should be observed to verify normal operation. If abnormal performance is observed, additional measures may be
necessary, such as reorienting or relocating the equipment.

b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [3] V/m.

15
ELECTROMAGNETIC COMPATIBILITY (EMC)

Recommended Separation Distances Between Portable and Mobile RF


Communications Equipment and the Equipment

The equipment is intended for use in the electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the equipment can help to prevent electromagnetic interference by
maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the
equipment as recommended in the table below, according to the maximum output power of the communications
equipment.

Rated Maximum Output Power Separation Distance According to Frequency of Transmitter (m)
of Transmitter W

150 KHz to 800 MHz 800 MHz to 2.5 GHz

d  1.2 P d  2.3 P
0.01 0.1 m 0.2 m

0.1 0.4 m 0.7 m


1 1.2 m 2.3 m
10 4.0 m 7.0 m

100 12.0 m 23.0 m

For transmitters rated at a maximum output power not listed above, the recommended separation distance d in
meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.

NOTE 1: At 800 MHz, the separation distance for the higher frequency range applies.

NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by the
absorption and reflection from structures, objects, and people.

16
ELECTROMAGNETIC COMPATIBILITY (EMC)
Regulatory Radio Compliance

Federal Communications Commission (FCC)


This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions:
 This device may not cause harmful interference.
 This device must accept any interference received, including interference that may cause undesired
operation.
This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to
Part 15 of FCC Rules. These limits are designed to provide reasonable protection against harmful interference
in a residential installation. This equipment generates, uses, and can radiate radio frequency energy. If not
installed and used in accordance with the instructions, it may cause harmful interference to radio
communications. However, there is no guarantee that interference will not occur in a particular installation. If
this equipment does cause harmful interference to radio or television reception, which can be determined by
turning the equipment off and on, the user is encouraged to try and correct the interference by one or more of
the following measures:
 Reorient or relocate the receiving antenna
 Increase the distance between the equipment and the receiver
 Connect the equipment to an outlet on a circuit different from that to which the receiver is connected
 Consult the dealer or an experienced radio/TV technician for help
The user may find the following booklet prepared by the Federal Communications Commission helpful: The
Interference Handbook This booklet is available from the U.S. Government Printing Office, Washington, D.C.
20402. Stock No. 004-000-0034504. Welch Allyn is not responsible for any radio or television interference
caused by unauthorized modification of the devices included with this Welch Allyn product, or the
substitution or attachment of connecting cables and equipment other than specified by Welch Allyn. The
correction of interference caused by such unauthorized modification, substitution, or attachment will be the
responsibility of the user.

WLAN
B&B electronics1 Radio Module 9373 with part number WLNN-AN-MR551
FCC ID: F4AWLNN551
1
Manufacturer also called B+B SmartWorx

17
ELECTROMAGNETIC COMPATIBILITY (EMC)
Industry Canada (IC) Emissions
RF Radiation Hazard Warning
Using higher gain antennas and types of antennas not certified for use with this product is not allowed. The
device shall not be co-located with another transmitter.

Cet avertissement de sécurité est conforme aux limites d'exposition définies par la norme CNR-102 at relative
aux fréquences radio.

This device complies with RSS 210 of Industry Canada.

Operation is subject to the following two conditions: (1) this device may not cause interference, and (2) this
device must accept any interference, including interference that may cause undesired operation of this device.

L’utilisation de ce dispositif est autorisée seulement aux conditions suivantes: (1) il ne doit pas produire de
brouillage et (2) l’ utilisateur du dispositif doit étre prêt à accepter tout brouillage radioélectrique reçu, même
si ce brouillage est susceptible de compromettre le fonctionnement du dispositif.

This Class B digital apparatus complies with Canadian ICES-003.

Cet appareil numérique de la classe B est conform à la norme NMB-003 du Canada.

WLAN
B&B electronics1 Radio Module 9373 with part number WLNN-AN-MR551
IC ID: 3913A-WLNN551
1
Manufacturer also called B+B SmartWorx

Declaración de conformidad Mexico


La operación de este equipo está sujeta a las siguientes dos condiciones:
1. es posible que este equipo o dispositivo no cause interferencia perjudicial y
2. este equipo o dispositivo debe aceptar cualquier interferencia, incluyendo la que pueda causar su
operación no deseada.

18
ELECTROMAGNETIC COMPATIBILITY (EMC)
European Union
Czech Welch Allyn tímto prohlašuje, ze tento WLAN device je ve shodě se základními
požadavky a dalšími příslušnými ustanoveními směrnice 2014/53/ES.
Danish Undertegnede Welch Allyn erklærer herved, at følgende udstyr WLAN device
overholder de væsentlige krav og øvrige relevante krav i direktiv 2014/53/EF
Dutch Bij deze verklaart Welch Allyn dat deze WLAN device voldoet aan de essentiële
eisen en aan de overige relevante bepalingen van Richtlijn 2014/53/EC.
English Hereby, Welch Allyn, declares that this WLAN device is in compliance with the
essential requirements and other relevant provisions of Directive 2014/53/EC.
Estonian Käesolevaga kinnitab Welch Allyn seadme WLAN device vastavust direktiivi
2014/53/EÜ põhinõuetele ja nimetatud direktiivist tulenevatele teistele asjakohastele
sätetele.
Finnish Welch Allyn vakuuttaa täten että WLAN device tyyppinen laite on direktiivin
2014/53/EY oleellisten vaatimusten ja sitä koskevien direktiivin muiden ehtojen
mukainen.
French Par la présente, Welch Allyn déclare que ce WLAN device est conforme aux
exigences essentielles et aux autres dispositions de la directive 2014/53/CE qui lui
sont applicables
German Hiermit erklärt Welch Allyn die Übereinstimmung des Gerätes WLAN device mit den
grundlegenden Anforderungen und den anderen relevanten Festlegungen der
Richtlinie 2014/53/EG. (Wien)
Greek ΜΕ ΤΗΝ ΠΑΡΟΥΣΑ Welch Allyn ΔΗΛΩΝΕΙ ΟΤΙ WLAN device
ΣΥΜΜΟΡΦΩΝΕΤΑΙ ΠΡΟΣ ΤΙΣ ΟΥΣΙΩΔΕΙΣ ΑΠΑΙΤΗΣΕΙΣ ΚΑΙ ΤΙΣ ΛΟΙΠΕΣ
ΣΧΕΤΙΚΕΣ ΔΙΑΤΑΞΕΙΣ ΤΗΣ ΟΔΗΓΙΑΣ 2014/53/ΕΚ
Hungarian Alulírott, Welch Allyn nyilatkozom, hogy a WLAN device megfelel a vonatkozó
alapvetõ követelményeknek és az 2014/53/EC irányelv egyéb elõírásainak.
Italian Con la presente Welch Allyn dichiara che questo WLAN device è conforme ai
requisiti essenziali ed alle altre disposizioni pertinenti stabilite dalla direttiva
2014/53/CE.
Latvian Ar šo Welch Allyn deklarē, ka WLAN device atbilst Direktīvas 2014/53/EK
būtiskajām prasībām un citiem ar to saistītajiem noteikumiem.
Lithuanian Šiuo Welch Allyn deklaruoja, kad šis WLAN device atitinka esminius reikalavimus ir
kitas 2014/53/EB Direktyvos nuostatas.
Malti Hawnhekk, Welch Allyn, jiddikjara li dan WLAN device jikkonforma mal-htigijiet
essenzjali u ma provvedimenti ohrajn relevanti li hemm fid-Dirrettiva 2014/53/EC
Portuguese Welch Allyn declara que este WLAN device está conforme com os requisitos
essenciais e outras disposições da Directiva 2014/53/CE.
Slovak Welch Allyn týmto vyhlasuje, ze WLAN device spĺňa základné požiadavky a všetky
príslušné ustanovenia Smernice 2014/53/ES.
Slovene Šiuo Welch Allyn deklaruoja, kad šis WLAN device atitinka esminius reikalavimus ir
kitas 2014/53/EB Direktyvos nuostatas.
Spanish Por medio de la presente Welch Allyn declara que el WLAN device cumple con los
requisitos esenciales y cualesquiera otras disposiciones aplicables o exigibles de la
Directiva 2014/53/CE

Swedish Härmed intygar Welch Allyn att denna WLAN device står I överensstämmelse med
de väsentliga egenskapskrav och övriga relevanta bestämmelser som framgår av
direktiv 2014/53/EG.

19
ELECTROMAGNETIC COMPATIBILITY (EMC)

Radio Compliance Table

Argentina Ente Nacional de las


Comunicaciones
(ENACOM) C-22663 (B&B)
Australia Australian Communications and
Media Authority (ACMA) Radio
Compliance Mark (RCM).
Brazil Agência Nacional de Este produto contém a placa 9373
Telecomunicações (ANATEL) código de homologação ANATEL
Modelo: B&B B&B: 02432-19-10488.
02432-19-10488 Este equipamento não tem direito à
proteção contra interferência
prejudicial e não pode causar
interferência em sistemas devidamente
autorizados
EAC Products meet all requirements of the
corresponding technical regulations
and have passed all conformity
assessment procedures.
Indonesia Keterangan Identification
a. a. [61733/I/SDPPI/2019] (B&B) is a
[61733/I/SDPPI/2019] number of certificate issued for
(B&B) adalah nomor certified telecommunication equipment
sertifikat yang b. [8620] (B&B) is a number of PLG
diterbitkan untuk ID based on one Certification Body
setiap alat dan database
perangkat
telekomunikasi
b. [8620] (B&B)
adalah nomor PLG ID
(identitas pelanggan)
berdasarkan database
Lembaga Sertifikasi

Mexico Instituto Federal de Telecomunicaciones This product contains


(Federal Telecommunications Institute— and Approved module,
IFETEL) Model No. 9373,
IFETEL No.
RCPBB9319-0533
(B&B)
Morocco AUTHORIZED BY MOROCCO ANRT
B&B: Approval number: MR 17490 ANRT 2018
Date of approval: 13-SEP-2018

Oman Telecommunications Regulatory Authority B&B R/6162/18


D172249

Paraguay Comisión Nacional de NR: 125/2019


Telecomunicaciones

Pakistan Pakistan Telecom Authority

20
ELECTROMAGNETIC COMPATIBILITY (EMC)
Philippines National Telecommunications B&B: ESD - 1818097C
Commission

Singapore Info-Communications Media


Development Authority (IMDA)

South Korea Korea Communications This equipment is Industrial (Class A) electromagnetic wave
Commission (대한민 국 방송통 suitability equipment and seller or user should take notice of it,
and this equipment is to be used in the places except for home.
신위원 회) – KCC 이 기기는 업무용(A급) 전자파적합기기로서 판 매자
Certification number: 또는 사용자는 이 점을 주의하시기 바라 며,
B&B: R-C-BVT-9373 가정외의 지역에서 사용하는 것을 목적으로 합니다.

Class A Equipment (Industrial Broadcasting & Communication


Equipment)
A급 기기 (업무용 방 송통신기자재)

UAE
B&B ER65768/18

21
MAINTENANCE & CLEANING

Recommended Tools and Supplies

Device Cleaning / Consumables:


Clean lint free cloth

Mild detergent

Isopropyl Alcohol (80-99%)

Cyanoacrylate Adhesive (e.g., Loctite 444 or equivalent product)

Anaerobic Thread locker

Smart Thermal Paper

10% Household bleach and water solution (Sodium Hypochlorite solution consisting of a
minimum 1:500 dilution and maximum of 1:10 dilution for disinfecting use only)

Preventive Maintenance / Conformance Testing


Multi-Meter

ECG Simulator (10 Lead)

AM12 Patient Input Module (9293-048-5X)

WAM Patient Input Module (30012-019-5X)

WAM/AM12 Lead Wire Set (9293-046-70)

Patient Cable Snap Adapter Set (9281-002-50)

Phillips #2 Screwdriver

10 Lead Shorting Block (or equivalent)

Lead Test Failure Box (or equivalent)

FAT 32 USB Memory Device

PC with ELI-Link v3.10 or later

USB Cable Type A to B (6400-012)

Phone Cable (6400-004)

PC with ELI-Link or Escribe and Modem (or equivalent)

PC with NIC and ELI-Link v4.5 or later

Wireless Router 802.11 (a, b, g, n)

Electrical Safety Analyzer

22
MAINTENANCE & CLEANING

Preventive Maintenance

Preventive maintenance is recommended to be performed on the ELI150c/250c once every 12


months.

WARNING: Preventive maintenance is to be performed by Welch Allyn authorized service


only. personnel

2.0 Maintenance Procedure

2.1 Turn unit on and print the device configuration per configuration section of this
manual. Attach a copy to the Preventive Maintenance Report.

2.2 Remove the unit cover.

2.3 Perform a visual inspection of the following items:

2.3.1 Enclosure/Housing – Look for damage or cracks in the external housing or


enclosure that could possibly expose the device to the introduction of foreign
objects or fluids. Attention should also be paid to areas that could expose an
operator or patient to internal circuitry of the device.
2.3.2 Contamination – Look for any contamination that may have occurred over time
that could not be seen with the housing in place.
 Fluid damage (perhaps caused during device cleaning)
 Debris on or behind display shield
 Battery leakage (lithium and main battery)
2.3.3 Internal Cabling – Look for cracked, pinched or partially disconnected cable
connections.
2.3.4 Fuse Ratings – Verify PCB mounted fuses (items 23 and 24) the meet the
specifications defined in the item description listing.
2.3.5 Markings and Labeling – Verify all labels and device markings are clearly visible
and legible to the device user and have not been worn off or rendered
unreadable through the use of harsh cleaning agents.
2.3.6 Integrity of Mechanical Parts – Verify the following items are properly secured to
the device and have not become loose or damaged through usage over time.
 AC Inlet
 Patient Input Connector
 Communication ports and antenna
 Writer mechanics/latching mechanism

2.4 Power Testing

* Based upon customer usage and age of battery, replace as needed.

2.4.1 Ensure battery is fully charged before performing these tests, voltage and current
limits are based on a fully charged battery.
2.4.2 Ensure there is no power connected to the UUT AC inlet.
2.4.3 Remove upper housing and writer assembly. Disconnect battery by pulling
battery cable off of the red terminal.

23
MAINTENANCE & CLEANING

2.4.4 Note battery age (if possible)


This information can be found on the white “date code” sticker located on the
battery (use the earliest date that is not crossed out). Record date on
Preventative Maintenance Report (PMR).
2.4.5 Battery (open circuit)
Measure battery voltage using a voltage meter; verify the meter reads greater
than 12.5vdc. Record result on PMR.
2.4.6 Battery (load)
Measure the battery voltage using a volt meter and a power resistor load
(10ohm, 20watt) in parallel with the battery. After approximately 5 seconds,
verify the meter reads greater than 11.7vdc. Record result on PMR.
2.4.7 Off current
Connect a current meter in line with battery. With the UUT power off, verify the
current meter reads less than 100 micro amps. Record result on PMR.
2.4.8 On current
Turn on the unit and verify the current meter reads less than 250 milli amps.
Record result on PMR.
2.4.9 AC charging current
Apply AC power to the unit and verify that the current draw from the battery
reverses polarity and the value starts decreasing as time increases. Record
result on PMR.
2.4.10 Battery charger output voltage
Disconnect the current meter and measure the battery charger output voltage
between the red disconnected battery cable and the negative terminal on the
battery. It should read between 13.0vdc and 14.0vdc. Record result on PMR.

2.5 Verify all power cables are reconnected properly

2.6 Reassemble unit in reverse order of disassembly

2.7 Functional Testing

2.7.1 AC LED/Display
Connect AC power cord to the unit and verify that the green AC LED (located to
the left of the display) illuminates continuous.
NOTE: The battery indicator will be clear when charging and will illuminate white
when fully charged.
Verify text on display is clear and legible and there are no flickering or missing
lines/pixels. Record result on PMR.

2.7.2 Writer
Open and close the writer door to verify smooth operation. Verify that the door
unlatches without sticking and that it latches completely. From the main screen,
simultaneously press shift+alt+RHY. Verify that a test page is printed and the
writer stops on the cue mark. The perforation of the paper should line up with the
tear edge on the writer. Assure there are no gaps in the printing and the print
darkness is uniform across the entire page. Verify the writer gears do not skip
and paper is tracking properly (you may need to print multiple pages to observe
this). Record result on PMR.

24
MAINTENANCE & CLEANING

2.7.3 ECG & Keyboard Matrix


Connect an ECG simulator to the AM12 or WAM patient interface. Set the
simulator to a known heart rate and amplitude. Press the ECG key to capture an
ECG. Verify there is an audible beep with each key press. Enter Last name
“PARCFL8” (Note: “PARCFL8” ensures the keyboard matrix is fully tested), then
press F6 (Done). Verify that 12 ECG traces print correctly and assess the
printout quality. Ensure uniform darkness across entire printout. Record result on
PMR.

2.7.4 ECG Noise Test


Connect a Shorting Block (TF-0629-*) and adapter or equivalent to the AM12 or
WAM patient interface. Set the ECG gain on the unit to 20mm/mV. Print a
rhythm strip (approx. 1 page). Verify that no channels have more than 0.5mm of
noise as measured by using Welch Allyn thermal paper. (Smallest grid line =
1mm) Record result on PMR

2.7.5 Communication Option Testing (as applicable)

The receiving station for modem, LAN and WLAN transmissions should be
running Welch Allyn ELI-Link software. Refer to the ELI-Link user manual for
proper configuration.

Verify successful transmission of all applicable communication options by


acquiring ECG records that include the transmission method in the “Patient
Name” field (such as Last Name = USBD) then subsequently transmitting the
ECG record stored to a compatible receiving device. Consult the product user
manual if needed to properly configure the communication settings for each
option present on the unit under test.

Successful transmission of the test records can be verified by viewing the ECG
records in the unit directory after transmission and confirming they are marked as
“transmitted” (as defined in the product user manual). Record result on PMR

 Modem
 LAN
 WLAN
 GSM/GPRS
 USB host (USB memory device needed)
 USBD

2.8 Clean unit per the instructions provided on a later page of this section of the service manual.

25
MAINTENANCE & CLEANING

2.9 Safety Testing

If the cardiograph housing was opened for repair or inspection work, the following safety tests
should be performed in accordance with the IEC 60601 standards.

The ELI150C and ELI250C are considered a Class 1 Type CF devices, intended to only be
utilized with the Welch Allyn AM12 or WAM patient input modules. Defibrillation isolation from
the patient is provided by the patient input modules, which are tested separately as part of
the manufacturing process (they are considered non-serviceable devices), therefore Hi-pot
testing is not required for these cardiograph models.

 Earth Leakage
 Enclosure Leakage
Non-conductive (fully insulated) chassis testing should be performed utilizing 200
cm2 conductive foil or equivalent, earth ground on AC input is utilized for
functional earth (not safety grounding).
 Patient Leakage
Applied part – patient input (utilize Welch Allyn AM12 patient cable)
 Patient Auxiliary Current
Applied part – patient input (utilize Welch Allyn AM12 patient cable)

26
MAINTENANCE & CLEANING

Device Cleaning & Disinfecting

1. Disconnect the power source. Remove cables and lead wires from device before cleaning.
2. For general cleaning of cables and lead wires, use a soft, lint-free cloth lightly moistened
with a mild soap and water solution. Wipe and air dry.
3. For disinfecting the exterior surfaces of the device, patient acquisition module, cables,
and lead wires, wipe exterior using:

Clorox Healthcare® Bleach Germicidal Wipes (use according to instructions on product


label), or

a soft, lint-free cloth with a solution of Sodium Hypochlorite (10% household bleach and
water solution) minimum 1:500 dilution (minimum 100 ppm free chlorine) and maximum
1:10 dilution as recommended by the APIC Guidelines for Selection and Use of
Disinfectants

CAUTION:

Prevent liquid from penetrating the device and do not attempt to clean/disinfect the
device or patient cables by submerging into a liquid, autoclaving, or steam cleaning.

Do not expose cables to strong ultra-violet radiation.

Do not sterilize the device or lead wires with Ethylene Oxide (EtO) gas.

Do not immerse cable ends or lead wires; immersion can cause metal corrosion. Use
caution with excess liquid as contact with metal parts may cause corrosion.

Do not use excessive drying techniques such as forced heat.

Improper cleaning products and processes can damage the device, produce brittle lead
wires and cables, corrode the metal, and void the warranty. Use care and proper
procedure whenever cleaning or maintaining the device.

27
MAINTENANCE & CLEANING

ELI 150/250c Preventive Maintenance Report

Unit Serial #: _

□ Print device configuration (attach to this report)

□ Remove the units upper housing

Perform Visual Inspection PASS / FAIL

Power Testing
Note Battery Age (If Possible) / _ (week/year)
Battery (Open Circuit) Voltage VDC (>12.5vdc)
Battery (with Load) Voltage VDC (>11.7vdc)
Off Current Ua (<100uA)
On Current mA (<250mA)

AC Charging Current PASS / FAIL (Circle One)

Battery Charger Output Voltage VDC (13.0-14.0vdc)

* Based upon customer usage and age of main battery, replace as needed.

□ Verify all power cables are properly reconnected and reassemble unit

Functional testing
AC LED/Display Functionality PASS / FAIL (Circle One)
Writer Test PASS / FAIL (Circle One)
ECG & Keyboard Matrix Testing PASS / FAIL (Circle One)
ECG Noise Test PASS / FAIL (Circle One)
Lead failure Test PASS / FAIL (Circle One)
Communication Option(s)
USB host PASS / FAIL / NA (Circle One)
USBD PASS / FAIL / NA (Circle One)
MODEM PASS / FAIL / NA (Circle One)
GSM/GPRS PASS / FAIL / NA (Circle One)
LAN PASS / FAIL / NA (Circle One)
WLAN PASS / FAIL / NA (Circle One)

□ Device Cleaning

Safety Testing PASS / FAIL (Circle One)


Earth Leakage
Enclosure Leakage
Patient Leakage
Patient Leakage Current

Performed by: _ Date: / /

28
DEVICE CONFIGURATION

Setting Technician Password

1. From real-time ECG view, select F6 (More) followed by F5 (Set Time/Date).


2. While holding down  (SHIFT), depress ALT and P simultaneously.
3. If required, enter password. This will automatically advance you to the set passwords display.

NOTE: The factory default password is “admin” (lowercase, no quotation marks); it is


suggested that the password be changed after installation of the unit.
4. Enter a technician password followed by a second entry to confirm.

NOTE: Password is case sensitive and alphanumeric.

5. From this display, select F6 (Exit) to return to real-time ECG view.

Configuration Menus

The configuration pages define all operational conditions that do not change on a daily or patient-to-patient basis.
Once you set these default conditions, you will rarely need to use the configuration screens again. To access the
configuration menus:

1. From real-time ECG view, select F6 (More) followed by F5 (Set Time/Date).


2. While holding down  (SHIFT), depress ALT and C simultaneously.
3. If required, enter password. The first configuration screen will appear. Notice the page indicator in the
upper right-hand corner.

To navigate the configuration menus:

 Use F4 (Page) to toggle through the configuration pages.


 Use F1 (▲) and F2 (▼) to move back and forth through each configuration option.
 Use F3 (►) to toggle through pre-programmed available settings per configuration field.
 Use F6 (Exit) to return to real-time ECG view. Any changes you have made will be saved.
 Use BKSP to erase entry errors.

To print the device’s configuration settings, select F6 (More) from real-time ECG view. Select F6 (More) again
followed by F1 (Print Configuration). The configuration printout captures every configuration setting: the
software version, the cart number of the device, and the date and time that the configuration printout occurred.

29
DEVICE CONFIGURATION

Summary of Configuration Menus

Configuration Parameter Definition

Software Version Displays software version on printout and display


Cart Number Numeric field 0 to 65535
Site Number Numeric field 0 to 4095
Site Name Alphanumerical field (30 digits)

Telephone Number Alphanumerical field (45 digits)


Language Available software languages
Volume Numerical field 0 to 8
Battery Timeout 10 min, 30 min, 60 min
Flash Size Normal or expanded (optional)
ID Format Standard, Short, Long, Custom
Auto-Fill ID YES/NO
AC Filter 50 Hz, 60 Hz, None
Paper Speed 25 or 50 mm/sec
Filter Frequency response for printouts: 40 Hz, 150 Hz, 300 Hz
Height/Weight Units lb/in or kg/cm
Date Format US (mm/dd/yyyy) or European (dd.mm.yyyy)
Interpretation YES/NO
Reasons YES/NO
Append Unconfirmed Report, Reviewed by
# of Copies 0–9
Copies with Interp. YES/NO
# ECGs Retrieved 0–7
Delete Rule Post Plot, Post Transmit, Post Plot/Xmt
Storage Sensitivity Normal or High
Auto-Save ECG YES/NO
Auto-Print ECG YES/NO
Cap Lock YES/NO
Use A4 paper
YES/NO
(ELI 250c only)
Rhythm Format 3 or 6 channel (ELI 150c); 3, 6, or 12 channel (ELI 250c)
3 Rhythm Lead 1 V1-V6, I, II, III, aVR, aVL, aVF
3 Rhythm Lead 2 V1-V6, I, II, III, aVR, aVL, aVF
3 Rhythm Lead 3 V1-V6, I, II, III, aVR, aVL, aVF
6 Rhythm Lead 1 V1-V6, I, II, III, aVR, aVL, aVF

30
DEVICE CONFIGURATION

Summary of Configuration Menus (continued)

Configuration Parameter Definition


6 Rhythm Lead 2 V1-V6, I, II, III, aVR, aVL, aVF
6 Rhythm Lead 3 V1-V6, I, II, III, aVR, aVL, aVF
6 Rhythm Lead 4 V1-V6, I, II, III, aVR, aVL, aVF
6 Rhythm Lead 5 V1-V6, I, II, III, aVR, aVL, aVF
6 Rhythm Lead 6 V1-V6, I, II, III, aVR, aVL, aVF
3, 3+1, 3+3, 6 channel; Cabrera or standard (ELI 150c)
Plot Format
3+1, 3+3, 6, 6+6, 12 channel; Cabrera or standard (ELI 250c)
3+1 Rhythm Lead V1-V6, I, II, III, aVR, aVL, aVF
3+3 Rhythm Lead 1 V1-V6, I, II, III, aVR, aVL, aVF
3+3 Rhythm Lead 2 V1-V6, I, II, III, aVR, aVL, aVF
3+3 Rhythm Lead 3 V1-V6, I, II, III, aVR, aVL, aVF
Bar Code Scanner YES/NO
Avg RR YES/NO
QTcB YES/NO
QTcF YES/NO
ECG Capture Last 10 or Best 10
Band Mode
(GSM/GPRS only) 850/1900MHz (US) or 900/1800MHz(EU)
(ELI 150c only)
None, Modem, LAN, WLAN, GSM/GPRS
Sync Media
(GSM/GPRS option applies to ELI 150c only)
DHCP
YES/NO
(active for LAN or WLAN)
IP Address
XXX.XXX.XXX.XXX
(active for LAN or WLAN)
Def Gateway
XXX.XXX.XXX.XXX
(active for LAN or WLAN)
Sub Net Mask
XXX.XXX.XXX.XXX
(active for LAN or WLAN)
Host IP
XXX.XXX.XXX.XXX
(active for LAN or WLAN)
Port Number
Numeric field (9 digits)
(active for LAN or WLAN)
None, WEP128, WEP64, WPA-PSK, WPA-LEAP, WPA-PSK64, WPA-PSK128,
Security
WPA-LEAP 64, WPA-LEAP128, WPA2-PSK, WPA2-PEAP
LAN MAC XX XX XX XX XX XX
WLAN MAC XXXXXXXXXXXX
SSID Alphanumerical field (30 digits) (not on printout)

31
DEVICE CONFIGURATION

Summary of Configuration Menus (continued)

Configuration Parameter Definition


WEP Key Numeric (1 digit) (not on printout); valid range 1-4
WEP Key ID Alphanumerical field (26 digits) A-F, 0-9 (not on printout)
PSK Passphrase Alphanumeric field (64 digits) (not on printout)
LEAP User Name Alphanumeric field (32 digits) (not on printout)
LEAP Password Alphanumeric field (32 digits) (not on printout)
PEAP User Name Alphanumeric field (63 digits) (not on printout)
PEAP Password Alphanumeric field (63 digits) (not on printout)
Worklist Management Standard or Refresh
Comm Protocol UNIPRO32, DICOM32, DICOM32ext OR UNIPRO64, DICOM64 (V2.x software)
Sync Mode None, XMT, XMT+Orders
Sync Date/Time YES/NO
XMT Mandatory Fields None, Last Name, ID, Last Name+ID

32
DEVICE CONFIGURATION

Configuration Settings

Software Version

Identifies the software version of your electrocardiograph.

Cart Number

Indicates which electrocardiograph acquired or transmitted a particular ECG.

Site Number

Identifies the site of your device. Site numbers designate the hospital, clinic, or institution for ECG records stored in
an E-Scribe system and must be defined for transmitting and retrieving ECGs from that system. You can use up to
four digits for the site number. Numbers from 0 – 4095 are supported.

Site Name

Defines your clinic, hospital, or office name. You can enter up to 30 alphanumeric characters. The site name prints
at the bottom, left edge of the ECG printout.

Telephone Number

Specifies the telephone number for internal modem transmission to another unit or to an E-Scribe system. Enter up
to 45 numeric characters.

You may need to dial a 9 to get an outside line. To wait for an additional dial tone, use the letter W.

EXAMPLE: 9W14145554321

To insert a pause, use a comma (,).


To change tone dialing to pulse dialing, use the letter P.

EXAMPLE: P14145554321

(If necessary, you can use both the letter W and the letter P in the same phone number.)

NOTE: It is not necessary to use alpha characters in the telephone number with GSM/GPRS mobile
connectivity.

TIP: To quickly delete or modify a phone number, use a shortcut. From the application screen,
simultaneously press  (SHIFT) + ALT + P. To edit an existing telephone number, use the Tab key.

Language

There are several languages available on the electrocardiograph.

CAUTION: Function labels are immediately translated upon selecting a new language and exiting the
configuration screen.

33
DEVICE CONFIGURATION

If an unknown language is visible, use the following steps to revert to the language of your country:

1. F6 (More) from real-time ECG view.


2. Select F5 (Set Time/Date).
3. Simultaneously press  (SHIFT) + ALT + C.
4. Enter password (“admin”)
5. Press F2 (▼) four times.
6. Press F3 (►) until the desired language appears.
7. F6 (Exit) to return to real-time ECG view.

Alphabets of specific languages may require use of special characters in demographic fields. This is accomplished
by using the SYM key on the keyboard.

Volume

Defines the keyboard click loudness. Available settings range from 0 (off) to 8 (loud).

Battery Time Out

Determines when the electrocardiograph will switch off in order to conserve the battery life of the device. The
battery time out will only occur if the keyboard has not been depressed for the time specified. The battery time out
setting is ignored if an active ECG signal is detected during transmission or while rhythm printing.

Flash Size

Indicates ECG storage capacity. Normal indicates standard memory capacity. Expanded indicates the optional
expanded memory has been installed.

ID Format

Defines the format for the patient demographic information prompts. There are three standard formats: short,
standard, or long. A custom ID format can be downloaded from ELI Link or an E-Scribe system. See Appendix A
to download a custom ID.

The short format includes the patient's last and first name, patient ID number, date of birth (automatically calculates
the age), and gender.

The standard format includes the patient's last name, patient ID number, age, height, weight, gender, race,
medication 1, medication 2, and a location field.

The long format is identical to the standard format except that it includes the patient's first name, room, and
comment fields.

Auto-Fill ID

When enabled, the device will automatically populate last name, first name, date of birth, age, and gender in the ID
screen if records with matching patient ID are found in the ECG directory.

34
DEVICE CONFIGURATION

AC Filter

The device removes 60 Hz or 50 Hz interference. The setting you select depends on the line frequency in your
country. Always use the 60 Hz setting in the U.S. If AC interference is present, check to see that the proper AC
filter is selected.

Paper Speed

Configure to 25 mm/s or 50 mm/s for default ECG printouts. For rhythm printouts and display, speeds of 5 mm/s or
10 mm/s are also available. See Device Configuration section to change speeds for display or rhythm printing. Paper
speed is printed at the bottom right corner of the ECG printout.

Filter

The ECG plot-frequency filter (or print filter) can be set to 0.05 to 40 Hz, 0.05 to 150 Hz, or 0.05 to 300 Hz. The
plot-frequency filter does not filter the acquired digital record. A 40 Hz plot-filter setting will reduce the noise
(40 Hz and higher frequencies) on the printed ECG, and a 150 Hz plot-filter setting will reduce the noise (150 Hz
and higher frequencies) on the printout; a 300 Hz plot-filter setting will not filter the printed ECG. The filter setting
is printed at the bottom right corner of the ECG printout.

Height/Weight Units

Defines the units of weight and height to either pounds/inches (lb/in) or kilograms/centimeters (kg/cm).

Date Format

Select either U.S. or European format for entering and displaying the patient’s date of birth.

U.S. Date Format: MM/DD/YYYY


European Date Format: DD.MM.YYYY

NOTE: The date format option does not modify the acquisition date printed on each ECG.

Interpretation

The device automatically analyzes ECGs and prints the optional interpretation on the ECG printout. This setting
allows you to select or suppress the “interpretive” text on the ECG report.

NOTE: The ECG interpretations offered by the device are only significant when used in conjunction with
a physician over-read as well as consideration of all other relevant patient data.

Reasons

The reasons statements indicate why a particular interpretive statement was printed. Reasons statements print
enclosed in [square brackets] within the interpretive text if the interpretation option is turned on. Turning the
reasons statement function on or off does not affect the measurements performed or the interpretive statements
selected by the analysis program.

For Example:
Anteroseptal Infarct [40+ ms Q WAVE IN V1-V4]
Where “Anteroseptal Infarct” is the interpretive statement,
and “40+ ms Q WAVE IN V1-V4” is the reason statement or explanation as to why the
interpretive statement was printed.

35
DEVICE CONFIGURATION

Append

A status or statement phrase can be appended to the ECG and printed under the interpretive text printout. Either
“unconfirmed report” or “reviewed by” can be selected; however, if you wish to have nothing appended to the ECG,
select “blank”.

Number of Copies

Defines the number of printed copies when an ECG is taken. (V1.x) A zero (0) setting prints the original only; one
(1) prints the original plus 1 copy; two (2) prints the original plus 2 copies, and so on. Up to 9 copies may be
selected. (V2.x) A zero (0) setting does not print the ECG; one (1) prints the original; two (2) prints the original plus
1 copy, and so on. Up to 9 copies may be selected.

Copies with Interpretation

Defines whether or not printed copies will include interpretation. The clinician may request the first ECG printout
with the interpretation included. Additional copies may be printed with or without the interpretation.

Number of ECGs Retrieved

Defines the number of ECGs retrieved from an E-Scribe system. The ECGs are retrieved by ID number. A zero (0)
setting retrieves the most current ECG for that ID number. Settings from one (1) to seven (7) retrieve the most
current ECG plus “X” number of ECGs identified by the entered value. EXAMPLE: If you enter the number 5, you
will retrieve the most current ECG plus the five preceding ECGs for that ID number. ECGs retrieved from the
E-S cribe are only printed at the device and not saved.

Delete Rule

Defines the rule to mark ECGs as deleted in the ECG directory. ECGs that are marked for deletion will be
automatically removed or erased based on their acquisition date (a first-in/first-out philosophy) to make room for the
new ECG record. ECGs are only erased from the directory when they are marked for deletion and if the directory
becomes full. More than one ECG may be removed from the directory in order to make room for the new incoming
record. The delete rule selections are:

Post Plot = ECG is automatically marked for deletion after printing


Post Transmit = ECG is automatically marked for deletion after transmission
Post Plot/Transmit = ECG is automatically marked for deletion after transmission and printing

Storage Sensitivity

Dictates the resolution of all stored ECG records. The sensitivity setting is either Normal or High. If the value is set
to High, the stored ECG will have a high resolution. As a result, the record size will be large and will reduce the
storage capacity in the ECG directory.

Auto-Save ECG (V1.x only)

Defines whether or not a newly acquired ECG will be automatically saved to the directory once it is acquired and
printed. If the auto-save configuration option is set to No and the record is printed, the device will prompt you to
“Save ECG?” F1 (Save) will store the ECG in the directory.

Auto-Print ECG (V1.x only)

36
DEVICE CONFIGURATION

Defines whether or not the device will automatically print the ECG after acquisition. If the selected configuration
option is set to No, a manual printout is possible.

Caps Lock

All character entry is translated to uppercase.

Use A4 Paper

The ELI 250c accommodates use of Z-fold thermal paper in either letter size (8.5 x 11 inches; 216 x 279 mm) or A4
size (8.27 x 11.69 inches; 210 x 297 mm). The provided paper tray spacer is required for use with A4 size paper.

Rhythm Formats

Defines the default values for rhythm printing. It is possible to set a 3 or 6-channel default rhythm format for the
ELI 150c. For the ELI 250c, a 3, 6, or 12-channel default rhythm format is possible. Define rhythm leads one
through three to customize a 3-channel rhythm printout or define rhythm leads one through six to customize the
6-channel rhythm printout.

Plot Format

Defines the default for one of the available plot formats in either standard or Cabrera presentation. Please note that
regardless of the plot format selected, 10 seconds of 12 leads are always stored.

The ECG plot options are:

Format Option ECG Data


2.5 seconds of 12 leads in a 3-channel format, plus 10-
second rhythm strip of one user-selectable lead in a 1-
3+1 channel format.

Cabrera also available.


2.5 seconds of 12 leads in a 3-channel format.
3
(ELI 150c only)
Cabrera also available.
5 seconds of 12-leads in a 6-channel format.
6
Cabrera also available.
2.5 seconds of 12 leads in a 3-channel format, plus 10-
second rhythm strip of user-selectable leads in a 3-
3+3 channel format.

Cabrera also available.


12 10 seconds of 12 leads in a one page printout.
(ELI 250c only)
5 seconds of 6 leads in a 6-channel format, plus 10-
second rhythm strip of user-selectable leads in a 6-
6+6
channel format.
(ELI 250c only)
Cabrera also available.

37
DEVICE CONFIGURATION

Rhythm Leads

Displays continuous rhythm of selected ECG leads and permits printing of selected leads. User may toggle between
selected leads, system set leads, or I, II, III, aVR, aVL, and aVF followed by V1, V2, V3, V4, V5, and V6.

NOTE: Rhythm acquisition is not stored in memory, only printed.

NOTE: See Unit Configuration section to acquire a rhythm printout.

Average RR

Enabling this option will display an averaged RR value to appear on the report.

QTcB

Enabling this option will display a Bazett’s corrected QT value on the report along with the default linear
QTc value.

QTcF

Enabling this option will display a Fridericia corrected QT value on the report along with the default linear
QTc value.

ECG Capture

Up to 5 minutes accumulated ECG data can be acquired internally for use with the Best 10 feature. The device
automatically selects the best 10 seconds from within the 5-minute buffer.

Users can switch between BEST 10 or LAST 10 by selecting F5 (More) followed by F5 (Last) or F5 (Best)
depending on the current view.

Band Mode

Use 850/1900 MHz (US) or 900/1800 MHz (EU). (Applies to ELI 150c only.)

Sync Media

Defines the default transmission setting. Select None, Modem, LAN, WLAN, or GSM/GPRS (GSM/GPRS option
applies to ELI 150c only). Optional connectivity options which have been purchased and installed will be available
for default selection.

An ELI x50c communicating over GPRS can be configured to automatically set its clock to match the time on a time
sync server. The time sync server must return a time stamp in the ELI x50c’s local time zone via the daytime
protocol (RFC 867). The time sync server must have a public IP address, and the standard port is 13. The server
must return the time in one of the following formats:

Format 1
day mon dd HH:mm:ss yyyy
Example
Wed Jul 15 17:05:49 2010

38
DEVICE CONFIGURATION

Format 2
hh.mm.ss tt mm/dd/yyyy
Example
02:38:51 PM 07/18/2011

Time sync servers running the Dimension 4 (http://www.thinkman.com/dimension4/index.htm) time sync software
support Format 1.

DHCP

Defines whether the Dynamic Host Communication Protocol (DHCP) will be used to obtain an IP address. If DHCP
is Yes, the network will automatically and dynamically assign an IP address. If DHCP is No, you must enter the IP
address, def gateway, and sub net mask.

NOTE: All parameters related to network connection must be entered under the direction of the IT
Manager of the facility where the device is installed.

IP Address

Enter the fixed IP address for network transmissions (if DHCP is not selected).

Def Gateway

Enter the address of the default gateway (if DHCP is not selected).

Sub Net Mask

Enter the sub net address (if DHCP is not selected).

Host IP

Enter the IP address of the host server.

NOTE: Addresses are always entered as 4 sets of 3 digits; therefore, an address of 192.168.0.7 must be
entered as 192.168.000.007.

Port Number

Enter the port number used by the host server.

LAN MAC

Shows the MAC address of the LAN.

Security (WEP)

Wired Equivalent Privacy (WEP) is an encrypted security protocol (part of the 802.11 standard). Access points can
have multiple WEP keys stored. Each one of them is identified by a number (e.g., 1, 2, 3, 4).

WEP Key

Enter the WEP key number.

39
DEVICE CONFIGURATION

WEP Key ID

Enter the 128-bit WEP key ID value (26 digits in 13 sets of two digits).

WLAN MAC

Shows the MAC address of the device’s wireless module for configuring access points.

SSID

Service Set Identifier (SSID) is the name of the wireless network. All ELI 150c electrocardiographs that will
transmit to the same network must have the same SSID name. This field is case sensitive.

WPA-PSK/WPA2-PSK

Allows for implementation of the “personal mode” of WPA. This mode of encryption employs Temporal Key
Integrity Protocol (TKIP) which dynamically changes keys as the system is used.

PSK Passphrase

The passphrase may be from eight to 63 ASCII characters or 64 hexadecimal digits (256 bits).

WPA-LEAP

Cisco® LEAP (Light Extensible Authorization Protocol) enables use of the device with wireless networks
employing the LEAP encryption protocol.

LEAP User Name

User name can be up to 32 characters in length.

LEAP Password

LEAP password can contain up to 32 characters.

WPA2-PEAP

Enables use of the device with wireless networks employing the PEAP encryption protocol.

PEAP User Name

User name can be up to 63 characters in length.

PEAP Password

Password can contain up to 63 characters.

40
DEVICE CONFIGURATION

Worklist Management

The device can download and process ECG order lists from the E-Scribe or another compatible information
management system which identifies the ECGs (or ECG orders) needed for particular patients. Implementation of
an order-based workflow can significantly reduce demographic data entry errors at the electrocardiograph. Orders
are deleted from the list when the ordered ECG is acquired.

When set to Standard, new order lists are appended to the remaining list. When set to Refresh, each new order list
will override the previously downloaded one.

Comm. Protocol

Select UNIPRO32, DICOM32, OR DICOM32ext for software v1.x.x. DICOM32 and DICOM32ext are only
available if the DICOM option has been installed. Select UNIPRO64 OR DICOM64 FOR SOFTWARE V2.x.x.
DICOM64 is only available if the DICOM option has been installed.

NOTE: This parameter must be entered under the direction of the IT Manager of the facility where the
device is installed.

NOTE: Units ship by default with Comm Protocol set to UNIPRO32. The UNIPRO32 setting is not
supported by E-Scribe versions prior to V8.10 or ELI Link versions prior to V3.00. For questions about
compatibility of your device with E-Scribe or ELI Link and UNIPRO32, contact Welch Allyn Technical
Support. Units shipped with v2.x.x software will not connect with E-Scribe and need Eli Link v4.0 or
above.

Sync Mode

Select None, XMT, or XMT+Orders. None requires a manual transmission of reports and then a second manual
request to receive orders from the cardiology management system. XMT will automatically transmit the report;
XMT+Orders will both transmit the report and retrieve the orders.

Sync Date/Time

Select Yes or No. Yes will synchronize the date/time with the approved cardiology management system. With No,
there will be no date/time synchronization. Date/time synchronization is done through ELI Link V3.10 or later.

XMT Mandatory Fields

Defines fields required for ECG transmission to the cardiology management system. None will allow data
transmission without limitation; Last Name requires the technician to enter a minimum of the Last Name; Last
Name and ID requires the technician to enter a minimum of the Last Name and the patient’s ID.

41
UNIT DISASSEMBLY

ELI 150c

ELI 250c

42
UNIT DISASSEMBLY

Removal of the Unit from Cart

Remove two thumbscrews from underneath the cart platform by turning counterclockwise. *Cart may not be exact
model as pictured.

*The ELI 150c and ELI 250c are very similar in design, so only the ELI 150c will be shown, unless an important
difference needs to be identified.

Cover Assembly Removal


Turn the unit upside down and use a T10 Torx driver to remove the 6 housing screws shown below. Once the
screws are removed, carefully flip the Device back over so that it is upright.

Item# 31

43
UNIT DISASSEMBLY

Open the writer drawer then lift the upper housing while rotating it counterclockwise according to the picture below.
This will allow the writer cover to pass through the housing opening to free the upper housing.

CAUTION: Remove battery power to the unit by disconnecting the battery connector BEFORE
removing the keyboard assembly. If battery power is not removed prior to disconnecting the keyboard
cable, damage to the motherboard may result.

Writer Removal
To remove the writer assembly, remove the 4 chassis screws shown below. Once they are removed, carefully turn
the unit over supporting the loose writer to ensure the cable connections are not stressed during the process.

Item# 36

44
UNIT DISASSEMBLY

Disconnect the writer interface cables and the motor cable shown below, then remove the writer assembly.

45
UNIT DISASSEMBLY

NOTE: The writer assembly can be obtained as a complete assembly for service purposes, or a specific
part or subassembly can be obtained to repair a specific writer related issue. The entire writer door with
the platen roller, latch assembly, and instruction label attached is available as an assembly; and the
thermal print head, print head mount, anti-static brush, and associated cables are also available as an
assembly. (Refer to the SERV ASSY item numbers on the item listing at the end of this section of the
manual).

To remove the writer door assembly, slide the door open and press downward on the latching tabs to allow the writer
door assembly to be slid out.

46
UNIT DISASSEMBLY

The writer latch mechanism can be disassembled to gain access to the writer latch bar and spring by removing the 4
screws shown below.

Item # 34

(View below with screws removed)

To remove the Gearbox assembly, remove the three screws as shown in the illustration below.

Item #36

47
UNIT DISASSEMBLY

To remove the writer motor, the pinion gear must be removed by loosening the pinion set screw as shown above.

NOTE: The set screw is installed at the factory with a Vibra-Tite coating to prevent loosening due to
vibration; ensure a coated set screw is used when reinstalling a new motor. The set screw should be
tightened to a torque 3.5 pound inches to ensure a proper connection to the motor shaft.

Item #34

The motor can then be removed by the removal of the two mounting screws shown above (actual screws are TORX
head for this product).

48
UNIT DISASSEMBLY

To remove the thermal print head assembly, remove the rubber O-ring as indicated in the illustration below.

Next, flip the writer assembly over and lift the thermal print head assembly out, taking care to feed the wires through
the writer base slots (shown below) as it is removed.

49
UNIT DISASSEMBLY

If the Print head assembly is disassembled further into the individual parts, care should be taken to re-torque the
ground screw to 3.5 pound inches during reassembly; the shoulder screw can be tightened completely as it allows
the Print head and mount to expand and contract when the device is exposed to abrupt temperature changes. When
replacing the Print head, keep in mind that the anti-static brush is a separate item and can not be reused from the old
Print head.

NOTE: When repairing units with symptoms of light or uneven print darkness, Welch Allyn recommends
that the entire Print head assembly be replaced to ensure the problem resolved completely. Slight
variances in the shape of the Print head and/or the print head mount can result in the unit exhibiting
these symptoms. Replacement of the entire assembly will ensure the problem is completely resolved.
(Refer to the SERV ASSY item numbers on the item listing at the end of this section of the manual).

50
UNIT DISASSEMBLY

Keyboard Removal

CAUTION: Remove battery power to the unit by disconnecting the battery connector BEFORE
removing the keyboard assembly. If battery power is not removed prior to disconnecting the keyboard
cable, damage to the motherboard may result.

Remove the 3 keyboard mounting screws which fasten the keyboard to the unit’s lower housing (150c only).

Item # 36

Next, press firmly and outward on the keypad connector hold-downs (as shown below) to release the keyboard
ribbon cable; then remove the keypad assembly from the bottom housing.

51
UNIT DISASSEMBLY

To remove the LCD Display, use a small flat blade screwdriver to lift up the latch mechanism that holds the LCD
ribbon cable into the connector; then slide the cable out of the connector.

Be careful not to apply excessive force, as this connector mechanism is very small and fragile.

IMPORTANT ELI250c: A new ELI250C keyboard PCBA was introduced in 2019 that has compatibility
requirements. The original part #36025-108-150 (item 107a) can use any software version, the newer part #36025-108-
400 (item 107b) must operate using v1.3.2 or newer (series 1.x), or v2.1.1 or newer (series 2.x) software.

IMPORTANT ELI150c: A new ELI150C keyboard PCBA was introduced for 4th edition compliance in 2020 that has
compatibility requirements. The original part #36025-102-150 (item 10a) can use any software version, the newer part
#36025-102-400 (item 10b) must operate using v1.3.3 or newer (series 1.x), or v2.1.2 or newer (series 2.x) software.

The rubber keypad and LCD can then be separated from the keyboard PCB assembly.

NOTE: When troubleshooting key press related issues, careful inspection and cleaning of the key pucks
and PCB contacts should be performed prior to any electrical troubleshooting.

52
UNIT DISASSEMBLY

IMPORTANT ELI150c Keypad: A new ELI150C keypad was introduced for 4th edition compliance in
2020 that has compatibility requirements. Exchanging the identical part for servicing purposes is
recommended, however if the older part (8359-003-50) is not available, the newer keypad
(413252) can be used to service the older models. The older keypad CANNOT be used on the newer 4th
edition compliant version. Refer to the table below for compatibility information.

Original version 4th Edition Compliant


ELI150c ELI150c

8359-003-50 Yes No

413252 Yes Yes

The visual difference between the two keypads is shown below, the new part has a portion removed.

8359-003-50 413252

ELI150c 4th Edition gasket:

When servicing an ELI150c 4th edition device, there is also an additional item #92 used to reduce
electromagnetic interference (EMI). When servicing the keyboard/keypad items, it may be necessary to
replace the gasket material to ensure connectivity between the two surfaces. The gasket positioning is
shown below.

Remove the adhesive liner on the gasket, then mount the gasket to the pad within the cutout of the keypad.

53
UNIT DISASSEMBLY

Installation of the Ribbon Cable to the Keypad Assembly

The Main Processor is located on the Keypad Motherboard. When installing the ribbon cable onto the circuit board,
it is imperative to install using the following process. If care is not taken when pressing the ribbon cable into the
connector, damage to the solder connections to the Ball Grid Array of the processor due to flexing of the printed
circuit board assembly could occur.

To install, ensure that the cable connector is correctly aligned with the board connector. Grasp the keypad side of
the board with your forefingers, and press with equal pressure with both thumbs on the ribbon connector. Some
force may be required to fully seat this connector. See picture below:

54
UNIT DISASSEMBLY

I/O Board Removal

Be sure the battery connector is disconnected, then remove the four screws as shown in illustration below.

Item #30

IMPORTANT ELI150c I/O Board: A new ELI150C I/O Board was introduced for 4th edition compliance
in 2020 that has compatibility requirements. Exchanging the identical part for servicing purposes is
recommended, however if the older part (26025-105-15x) is not available, the newer I/O Board can be used
to service the older models. The older I/O Board CANNOT be used on the newer 4th edition compliant
version. Refer to the table below for compatibility information.

Original version 4th Edition Compliant


ELI150c ELI150c

26025-105-15x Yes No

26025-105-40x Yes Yes

55
UNIT DISASSEMBLY

If the unit is equipped with an optional Wireless Acquisition Module (WAM), the Universal Transceiver Key
(UTK) will be inserted into the secondary USB connector and held in place by the plastic housing ensure it
does not become dislodged (see image below). The UTK may need to be transferred to the new I/O board if
replaced during servicing activity.

Refer to the Item Identification Table for the correct item (#90 a/b), as there are two versions of
WAM/UTK pairs that must match (v1 to v1 or v2 to v2) for the wireless interface to operate properly.
Units utilizing the v2 UTK will have a round “2” label (item # 91) near the ECG input connector on the
housing.

To remove the Power Supply, disconnect the battery cables and then remove the 4 mounting screws
as shown in illustration below.

Item #32

56
UNIT DISASSEMBLY

Battery Replacement

CAUTION: Be careful not to short the positive and negative terminals of the battery with a metal
tool when removing the battery from the unit; as this could result in damage to the unit, or personal injury
to the repair technician.

The battery is installed at the factory by adhering it to the lower housing with two pieces of double stick foam tape.
To remove the battery, insert a small prying tool underneath the battery in the area shown below; then carefully lift
the battery edge to separate the battery from the lower housing. Removal of the battery will require a fair amount of
force, as the foam tape will typically tear in the center leaving the adhesive portions on the battery and lower
housing.

57
UNIT DISASSEMBLY

The adhesive left on the lower housing (as shown below) will need to be removed with a small scraping tool prior to
installation of the new battery.

NOTE: Welch Allyn recommends that a battery assembly that includes the pre-mounted foam tape be
obtained for battery replacements that are not performed in a Welch Allyn Service Center. This will ensure
that the battery is properly adhered to the lower housing to avoid it being dislodge during use, and will
allow for proper removal for routine maintenance. (Refer to the SERV ASSY item numbers on the item
listing at the end of this section of the manual).

Reassembly of the ELI 150c/250c can be performed by reversing the sequence of the previous disassembly
procedure.

When installing screws for all circuit boards, use a torque driver set to 3.5 in/lbs.

When installing screws for Upper Housing and Writer Assembly, use a torque driver set to 5.0 in/lbs.

58
UNIT DISASSEMBLY

The items highlighted in grey that are listed in the ELI 150c item description listing identify the serviceable level of
the device.

Some subcomponents of assemblies listed are not available as individual service items from Welch Allyn, the
assembly level item must be used for servicing purposes.

Item numbers below 100 are also used with the ELI 150c. Item numbers above 100 are specific to ELI 250c.

ELI 150c Item Description Listing


Item # Part # Description
1 22500-150-51 WRITER ASSEMBLY ELI/BUR 150c - NO LABEL
(See Item # 74 for ELI or BUR label part number)
2 25018-034-50 CABLE ASSEMBLY ELI 200+ PRINTHEAD TO PCB
3 25020-060-50 CABLE ASSY CUE SENSOR TO MTHRBD ELI 200+
4 25020-067-50 GROUND WIRE FOR ELI 230 PRINTHEAD
5 25020-076-50 CABLE POWER INTERNAL ELI 150c
6 26025-045-151 ELI 200+ CUE SENSOR PCB ASSEMBLY
7 26025-073-50 SIM SIMULATOR PCB ASSEMBLY
8 26025-074-50 REMOTE SIM CONNECTOR PCB ASSEMBLY
9 26025-077-170 B&B-N WLAN MODULE PCB ASSY
10a 36025-102-150 ELI 150c KEYBOARD PCB ASSEMBLY -TESTED
10b 36025-102-400 * ELI150c KEYBOARD PCB ASSY 4th ED-TESTED
11 26025-110-150 ELI 150c FLEX CABLE INTERCONNECT
12 3171-009 CABLE ASSY COAX 6" SMA-F BLKHD to MMCX-M
13 3171-010 CABLE COAX U.FL TO RP-SMA BLKHD 100mm
14 3225-003 CONN, MOD PHONE, 4 PIN, RA, LO PRO
15 3225-008 CONN RJ-45 8 PIN SHIELDED w/GREEN LEDS
16 3375-004 CONN USB RECEPTACLE TYPE B
17 3375-006 CONN USB RECEPT TYPE A UPRIGHT HIGH RET
18 3600-008 ANTENNA F1 900/1800 MHz HINGED RA
19 3600-009 ANTENNA F2 850/1900 MHz RIGHT ANGLE
20 3600-016 ANTENNA DUAL BAND 2400/5000 MHz
21 4027-001 FUSE POLYSWITCH TR 600V 150mA
22 4027-002 FUSE 1A 250V TIME LAG RADIAL 8x8.5x4mm
23 4027-003 FUSE 5A 250V TIME LAG RADIAL 8x8.5x4mm
24 4800-006 BATTERY RECHARGEABLE SLA 12V 2.2/2.3Ah
25 5400-019 LCD 3.5" TFT ACTIVE MATRIX 320 x 240
26 5450-005 PRINTHEAD THERMAL 108mm 4.25"
28 600-0515 COMMON MODE CHOKE 2A 4 PIN SM
29 6001-002-01 SCREW, SHOULDER HEX M3 x 0.5 STAINLESS
30 6020-060 SCREW THD-FORM PAN HD TORX 4-20x1/4"
31 6020-061 SCREW THD-FORM PAN HD TORX 4-20x1/2"
32 6020-062 SCREW THD-FORM PAN HD TORX 4-20x3/8"
33 6020-430-02 SCREW PHILLIPS PAN HEAD M3 X 6mm COATED
34 6020-735-02 SCREW FLAT HD TORX M3 x 6 COATED
35 6020-835 SCREW PAN HD TORX M3 x 8
36 6020-835-02 SCREW PAN HD TORX M3 x 8 COATED

* Item has a software compatibility requirement.

59
UNIT DISASSEMBLY

ELI 150c Item Description Listing


Item # Part # Description
37 6030-025 SET SCREW SOCKET M2.5 x 4
38 6100-004 WASHER, WAVE, .006/.030 x .18 x .25
39 6125-004 SPACER .19 x .25 x .125
40 6125-017 SPACER .19 X .25 X .063
41 6140-003 E-RING FOR 0.187 SHAFT ZN-PLATED STEEL
42 6141-003 O-RING BUNA-N 1/2 OD X 5/16 ID
43 6160-003 STANDOFF NYLON SNAP BOTH ENDS 0.062 BD
44 6320-003 FOOT BLACK .64 OD X .115 ADHESIVE
45 6520-003 BEARING BALL .1875ID SS
46 6545-007-01 MOTOR STEPPER PM 35mm 24V 40Ohm WINDINGS
47 6570-420-01 PLATEN / SHAFT 4.200 x 0.551 DIA
48 7400-019 TAPE POLYESTER FILM 1" X .05mm
49 7401-003 TAPE 2SIDED ADHESIVE 0.031 THK x.50 WIDE
50 7403-001 VIBRA-TITE 1oz
51 7480-090 BRUSH ANTI-STATIC 90mm FLEXIBLE
52 7495-001 CABLE TIE LOCKING 3.9 x .10
58 8342-004-52 GEAR BOX ELI 200+ IML/OPM
59 8342-008-02 LATCH RELEASE ELI 250c
60 8342-009-01 GEAR SPUR 22 TEETH WITH STAINLESS HUB
61 8342-018-01 BAR RELEASE PIVOT 3.950 X .118 DIA.
62 8342-019-01 SPRING COMPRESSION .5 OD X .85 L
63 8342-020-01 PIVOT BAR RESTRAINING PLATE
64 8347-004-51 PAPER TRAY ELI 150c
65 8347-005-51 PAPER TRAY COVER ELI 150c
66 8347-006-51 PRINTHEAD MOUNT ELI 150c
67 8347-007-51 ACCESS COVER ELI 150c
68 8347-009-50 SPRING BAR 6.1 x 0.093 DIA
69 8359-001-50 HOUSING UPPER ELI 150c
70 8359-002-50 HOUSING LOWER ELI 150c
71a 8359-003-50 KEYPAD ELASTOMERIC ELI 150c UNIVERSAL
71b 413252 KEYPAD ELASTOMERIC ELI 150c 4th ED
72 8359-004-51 LCD BEZEL ELI/BUR 150c
73 9025-049-02 LABEL ELI 2XX MULTITECH MODEM ID
9042-073-01 LABEL ELI 150c USER INSTRUCTIONS
74 9042-073-02 LABEL BUR 150c USER INSTRUCTIONS BURDICK
9042-073-12 LABEL BUR 150c USER INST BURDCK/McKESSON
75 9050-059-05 LABEL REG MT GSM F4
9050-059-06 LABEL REG WLAN DPAC-G2
9050-059-09 LABEL REG WLAN QUATECH-G2
76 9050-059-10 LABEL REG WLAN B&B
77 9050-059-07 LABEL REGULATORY UTK
LABEL ELI 150c NAMEPLATE
9050-086-01 LABEL ELI 150c INMETRO NAMEPLATE
78 9050-086-02 LABEL ELI 150c NAMEPLATE McKESSON
9050-086-11

60
UNIT DISASSEMBLY

ELI 150c Item Description Listing


Item # Part Description
#
79 9910-017 MODEM MULTITECH MT5600 V.92 5V SERIAL
80 9910-022 MODEM GSM MT SCKTMOD F4 QUAD-BAND US/EU
81a 26025-099-151 AC POWER SUPPLY 16VDC PCB ASSY w/UL
81b 26025-099-400 AC POWER SUP 16VDC wUL_4THED
82a 26025-105-151 ELI 150c/250c I/O CONNECTOR PCB ASSEMBLY w/o COMM
26025-105-152 ELI 150c/250c I/O CONNECTOR PCB ASSY w/MODEM
26025-105-153 ELI 150c/250c I/O CONNECTOR PCB ASSY w/GSM
26025-105-154 ELI 150c/250c I/O CONNECTOR PCB ASSY w/LAN+WLAN
82b 26025-105-400 ELI 150c/250c I/O CONN PCB ASSY w/o COM
26025-105-401 ELI 150c/250c I/O CONN PCB ASSY w/MODEM
26025-105-402 ELI 150c/250c I/O CONN PCB ASSY w/GSM
26025-105-403 ELI150c/250c I/O CON PCB ASSY w/LAN+WLAN
83 SERV-ASSY-177-01 BATTERY 12V 2.2/2.3Ah WITH FOAM TAPE
84 SERV-ASSY-177-02 ELI/BUR 150C WRITER LID ASSEMBLY – NO LABEL
(See Item # 74 for ELI or BUR label part number)
85 41000-028-53 ELI150C PRINTHEAD ASSEMBLY

90a 26025-092-151 USB TRANSCEIVER KEY (UTK) ASSEMBLY v1.x

90b 26025-092-404 UTK w/Software v2.x

91 728940 v2 UTK Label for Cardiographs

92 728427 Gasket EMI Foam 3mm H x 0.33" L

61
UNIT DISASSEMBLY
The items highlighted in grey that are listed in the ELI 250c item description listing identify the serviceable level of
the device.

Some subcomponents of assemblies listed are not available as individual service items from Welch Allyn, the
assembly level item must be used for servicing purposes.

Item numbers below 100 are also used with the ELI 150c. Item numbers above 100 are specific to ELI 250c.

ELI 250c Item Description Listing


Item # Part # Description
101 22500-250-50 WRITER ASSEMBLY ELI 250c – NO LABEL
(See Item # 122 for ELI or BUR label part number)
102 25018-046-50 CABLE ASSY BLK PRINTHEAD TO PCB ELI 250c
103 25018-047-50 CABLE ASSY WHT PRINTHEAD TO PCB ELI 250c
104 25020-074-50 CABLE ASSY CUE SENSOR TO PCB ELI 250c
105 25020-075-50 GROUND WIRE FOR ELI 250c PRINTHEAD
106 25020-077-50 CABLE POWER INTERNAL ELI 250c
6 26025-045-151 ELI 200+ CUE SENSOR PCB ASSEMBLY
9 26025-077-170 B&B-N WLAN MODULE PCB ASSY
107a 36025-108-150 ELI 250c KEYBOARD PCB ASSEMBLY - TESTED
107b 36025-108-400 * ELI 250c KEYBOARD PCB ASSY 4TH - TESTED
128 26025-111-150 ELI 250c FLEX CIRCUIT ASSEMBLY
13 3171-010 CABLE COAX U.FL TO RP-SMA BLKHD 100mm
14 3225-003 CONN, MOD PHONE, 4 PIN, RA, LO PRO
15 3225-008 CONN RJ-45 8 PIN SHIELDED w/GREEN LEDS
16 3375-004 CONN USB RECEPTACLE TYPE B
17 3375-006 CONN USB RECEPT TYPE A UPRIGHT HIGH RET
20 3600-016 ANTENNA DUAL BAND 2400/5000 MHz
21 4027-001 FUSE POLYSWITCH TR 600V 150mA
22 4027-002 FUSE 1A 250V TIME LAG RADIAL 8x8.5x4mm

* Item has a software compatibility requirement.

62
UNIT DISASSEMBLY

ELI 250c Item Description Listing


Item # Part # Description
23 4027-003 FUSE 5A 250V TIME LAG RADIAL 8x8.5x4mm
24 4800-006 BATTERY RECHARGEABLE SLA 12V 2.2/2.3Ah
25 5400-019 LCD 3.5" TFT ACTIVE MATRIX 320 x 240
108 5450-004 PRINTHEAD THERMAL 216mm 8.50"
28 600-0515 COMMON MODE CHOKE 2A 4 PIN SM
29 6001-002-01 SCREW, SHOULDER HEX M3 x 0.5 STAINLESS
30 6020-060 SCREW THD-FORM PAN HD TORX 4-20x1/4"
31 6020-061 SCREW THD-FORM PAN HD TORX 4-20x1/2"
32 6020-062 SCREW THD-FORM PAN HD TORX 4-20x3/8"
34 6020-735-02 SCREW FLAT HD TORX M3 x 6 COATED
35 6020-835 SCREW PAN HD TORX M3 x 8
36 6020-835-02 SCREW PAN HD TORX M3 x 8 COATED
37 6030-025 SET SCREW SOCKET M2.5 x 4
38 6100-004 WASHER, WAVE, .006/.030 x .18 x .25
39 6125-004 SPACER .19 x .25 x .125
40 6125-017 SPACER .19 X .25 X .063
41 6140-003 E-RING FOR 0.187 SHAFT ZN-PLATED STEEL
42 6141-003 O-RING BUNA-N 1/2 OD X 5/16 ID
44 6320-003 FOOT BLACK .64 OD X .115 ADHESIVE
45 6520-003 BEARING BALL .1875ID SS
46 6545-007-01 MOTOR STEPPER PM 35mm 24V 40Ohm WINDINGS
109 6570-842-02 PLATEN / SHAFT 8.421 x 0.551 DIA
49 7401-003 TAPE 2SIDED ADHESIVE 0.031 THK x.50 WIDE
50 7403-001 VIBRA-TITE 1oz
51 7480-090 BRUSH ANTI-STATIC 90mm FLEXIBLE
52 7495-001 CABLE TIE LOCKING 3.9 x .10
110 7495-012 CABLE TIE, LOCKING, 5.6 x .10
113 8342-003-51 PAPER TRAY COVER ELI 250c
114 8342-004-53 GEAR BOX ELI 200+ IML/IPM
115 8342-005-51 PAPER TRAY ELI 250c
116 8342-006-03 PRINTHEAD MOUNT ELI 250c
59 8342-008-02 LATCH RELEASE ELI 250c
60 8342-009-01 GEAR SPUR 22 TEETH WITH STAINLESS HUB
117 8342-017-01 SPRING BAR 10.125 X .156 DIA.
61 8342-018-01 BAR RELEASE PIVOT 3.950 X .118 DIA.
62 8342-019-01 SPRING COMPRESSION .5 OD X .85 L
63 8342-020-01 PIVOT BAR RESTRAINING PLATE
118 8342-025-50 RETAINER CLIP ELI 2XX WRITER SPRING BAR
119 8360-001-50 HOUSING UPPER ELI 250c
120 8360-002-50 HOUSING LOWER ELI 250c
121 8360-003-50 KEYPAD ELASTOMERIC ELI 250c
73 9025-049-02 LABEL ELI 2XX MULTITECH MODEM ID

63
UNIT DISASSEMBLY

ELI 250c Item Description Listing


Item # Part # Description
122 9042-074-01 LABEL ELI 250c USER INSTRUCTIONS
9042-074-02 LABEL BUR 250c USER INSTRUCTIONS-BURDICK
9050-059-06 LABEL REG WLAN DPAC-G2
9050-059-09 LABEL REG WLAN QUATECH-G2
76 9050-059-10 LABEL REG WLAN B&B
77 9050-059-07 LABEL REGULATORY UTK
123 9050-087-01 LABEL ELI 250c NAMEPLATE
9050-087-02 LABEL ELI 250c INMETRO NAMEPLATE
124 9326-002 ADHESIVE CYANOACRYLATE ESTER
79 9910-017 MODEM MULTITECH MT5600 V.92 5V SERIAL
81 26025-099-151 AC POWER SUPPLY 16VDC PCB ASSY w/UL
26025-105-151 ELI 150c/250c I/O CONNECTOR PCB ASSEMBLY w/o
COMM
26025-105-152 ELI 150c/250c I/O CONNECTOR PCB ASSY w/MODEM
82 26025-105-154 ELI 150c/250c I/O CONNECTOR PCB ASSY w/LAN+WLAN
83 SERV-ASSY-177-01 BATTERY 12V 2.2/2.3Ah WITH FOAM TAPE
125 SERV-ASSY-178-02 ELI/BUR 250C WRITER LID ASSEMBLY – NO LABEL
(See Item # 122 for ELI or BUR label part number)
126 41000-028-52 ELI250C PRINTHEAD ASSEMBLY
127 8360-004-50 LCD BEZEL ELI/BUR 250c
90a 26025-092-151 USB TRANSCEIVER KEY (UTK) ASSEMBLY v1.x
90b 26025-092-404 UTK w/Software v2.x
91 728940 v2 UTK Label for Cardiographs

64
UNIT DISASSEMBLY

Item numbers below 100 are also used with the ELI 150c. Item numbers above 100 are specific to ELI 250c.

(The item ID table below is sorted by part #)

ELI 150c/250c Item Identification Table


Item # Part # Picture

1 22500-150-51 (Label not included)

101 22500-250-50 (Label not included)

2 25018-034-50

3 25020-060-50

4 25020-067-50

5 25020-076-50

6 26025-045-151

65
UNIT DISASSEMBLY

ELI 150c/250c Item Identification Table


Item # Part # Picture

7 26025-073-50

8 26025-074-50

9 26025-077-170

9910-023-03
(radio only)

10 36025-102-150

66
UNIT DISASSEMBLY

ELI 150c/250c Item Identification Table


Item # Part # Picture

107 36025-108-150

11 26025-110-150

128 26025-111-150

12 3171-009

13 3171-010

67
UNIT DISASSEMBLY

ELI 150c/250c Item Identification Table


Item # Part # Picture

14 3225-003

15 3225-008

16 3375-004

17 3375-006

68
UNIT DISASSEMBLY

ELI 150c/250c Item Identification Table


Item # Part # Picture

18 3600-008

19 3600-009

20 3600-016

21 4027-001

22 4027-002

69
UNIT DISASSEMBLY

ELI 150c/250c Item Identification Table


Item # Part # Picture

23 4027-003

24 4800-006

25 5400-019

108 5450-004

26 5450-005

70
UNIT DISASSEMBLY

ELI 150c/250c Item Identification Table


Item # Part # Picture

28 600-0515

29 6001-002-01

30 6020-060

31 6020-061

32 6020-062

33 6020-430-02

34 6020-735-02

35 6020-835

71
UNIT DISASSEMBLY

ELI 150c/250c Item Identification Table


Item # Part # Picture

36 6020-835-02

37 6030-025

38 6100-004

39 6125-004

40 6125-017

41 6140-003

42 6141-003

72
UNIT DISASSEMBLY

ELI 150c/250c Item Identification Table


Item # Part # Picture

43 6160-003

44 6320-003

45 6520-003

46 6545-007-01

47 6570-420-01

73
UNIT DISASSEMBLY

ELI 150c/250c Item Identification Table


Item # Part # Picture

48 7400-019

49 7401-003

50 7403-001

51 7480-090

52 7495-001

58 8342-004-52
114 8342-004-53

59 8342-008-02

74
UNIT DISASSEMBLY

ELI 150c/250c Item Identification Table


Item # Part # Picture

60 8342-009-01

61 8342-018-01

62 8342-019-01

63 8342-020-01

64 8347-004-51

65 8347-005-51

66 8347-006-51

75
UNIT DISASSEMBLY

ELI 150c/250c Item Identification Table


Item # Part # Picture

67 8347-007-51

68 8347-009-50

69 8359-001-50

70 8359-002-50

71 8359-003-50

71b 413252

76
UNIT DISASSEMBLY

ELI 150c/250c Item Identification Table


Item # Part # Picture

72 8359-004-51

119 8360-001-50

120 8360-002-50

121 8360-003-50

77
UNIT DISASSEMBLY

ELI 150c/250c Item Identification Table


Item # Part # Picture

73 9025-049-02

9042-073-01 (ELI)
74 9042-073-02 (BUR)
9042-073-12 (MLBUR)

122 9042-074-01 (ELI)


9042-074-02 (BUR)

75 9050-059-05

76 9050-059-06

78
UNIT DISASSEMBLY

ELI 150c/250c Item Identification Table


Item # Part # Picture

76 9050-059-09

76 9050-059-10

77 9050-059-07

9050-086-01 (ELI)
9050-086-02
(INMETRO)
78 9050-086-11 (MLBUR)

79 9910-017

79
UNIT DISASSEMBLY

ELI 150c/250c Item Identification Table


Item # Part # Picture

80 9910-022

81 26025-099-151

26025-105-151 (w/o
COMM)
26025-105-152
(w/MODEM)
26025-105-153
(w/GSM)
26025-105-154
82 (w/LAN+WLAN) Example: 26025-105-151

83 SERV-ASSY-177-01

80
UNIT DISASSEMBLY

ELI 150c/250c Item Identification Table


Item # Part # Picture

84 SERV-ASSY-177-02 (Label not included)

85 41000-028-53

92 728427

(Label not included)


125 SERV-ASSY-178-02

126 41000-028-52

81
DEVICE SPECIFICATIONS UNIT DISASSEMBLY

ELI 150c Specifications

Feature Specifications
Instrument Type 12-lead electrocardiograph
Input Channels Simultaneous acquisition of all 12 leads
Standard Leads Acquired I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6
Waveform Display Backlit, ¼ VGA color LCD (320 x 240);
3, 4+4, or 6+6 lead presentation
Input Impedance Meets or exceeds requirements of ANSI/AAMI EC11
Input Dynamic Range
Electrode Offset Tolerance
Common Mode Rejection
Patient Leakage Current Meets or exceeds requirements of ANSI/AAMI ES1
Chassis Leakage Current
Digital Sampling Rate 40,000 s/sec/channel used for pacemaker spike detection;
1000 s/sec/channel used for recording and analysis
Special Functions Optional Welch Allyn VERITAS resting ECG interpretation with age and
gender specific algorithm; connectivity options for bidirectional
communication
Paper Type Perforated double Z-fold thermal paper; 108 mm (4”) wide, 200 sheets
Thermal Printer Computer-controlled dot array; 8 dots/mm
Thermal Printer Speeds 5, 10, 25, or 50 mm/s
Gain Settings 5, 10, or 20 mm/mV
Report Print Formats Standard or Cabrera; 3, 3+1, 3+3, or 6 channel
Rhythm Print Formats 3 or 6 channel with configurable lead groups
Keyboard Type Elastomeric keyboard with complete alphanumeric keys, soft-key menu, and
dedicated function keys
Frequency Response 0.05 to 300 Hz
Filters High-performance baseline filter; AC interference filter 50/60 Hz; low-pass
filters 40 Hz, 150 Hz, or 300 Hz
A/D Conversion 20 bits (1.17 microvolt LSB)
Device Classification Class I, Type CF defibrillation-proof applied parts
ECG Storage Internal storage up to 100 ECGs; optional expanded up to 200 ECGs
Weight 7.2 lbs. (3.3 kg) including battery (without paper)
Dimensions 11.25 x 11.5 x 3.75” (29.2 x 30.5 x 10.2 cm)
Power Requirements Universal AC power supply (100-240 VAC at 50/60 Hz) 110 VA; internal
rechargeable battery

82
SECTION 6

ELI 250c Specifications

Feature Specifications
Instrument Type 12-lead electrocardiograph
Input Channels Simultaneous acquisition of all 12 leads
Standard Leads Acquired I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6
Waveform Display Backlit, ¼ VGA color LCD (320 x 240);
3, 4+4, or 6+6 lead presentation
Input Impedance Meets or exceeds requirements of ANSI/AAMI EC11
Input Dynamic Range
Electrode Offset Tolerance
Common Mode Rejection
Patient Leakage Current Meets or exceeds requirements of ANSI/AAMI ES1
Chassis Leakage Current
Digital Sampling Rate 40,000 s/sec/channel used for pacemaker spike detection;
1000 s/sec/channel used for recording and analysis
Special Functions Optional Welch Allyn VERITAS resting ECG interpretation with age and
gender specific algorithm; connectivity options for bidirectional
communication
Paper Type Perforated Z-fold thermal paper; A4 or 8.5 x 11” wide, 250 sheets
Thermal Printer Computer-controlled dot array; 8 dots/mm

Thermal Printer Speeds 5, 10, 25, or 50 mm/s


Gain Settings 5, 10, or 20 mm/mV
Report Print Formats Standard or Cabrera; 3+1, 3+3, 6, 6+6, or 12 channel

Rhythm Print Formats 3, 6, or 12 channel with configurable lead groups


Keyboard Type Elastomeric keyboard with complete alphanumeric keys, soft-key menu,
and dedicated function keys
Frequency Response 0.05 to 300 Hz
Filters High-performance baseline filter; AC interference filter 50/60 Hz; low-pass
filters 40 Hz, 150 Hz, or 300 Hz
A/D Conversion 20 bits (1.17 microvolt LSB)
Device Classification Class I, Type CF defibrillation-proof applied parts

ECG Storage Internal storage up to 100 ECGs; optional expanded up to 200 ECGs
Weight 11.25 lbs. (5.1 kg) including battery (without paper)
Dimensions 15.5 x 17 x 4” (39.4 x 43.2 x 10.2 cm)
Power Requirements Universal AC power supply (100-240 VAC at 50/60 Hz) 110 VA; internal
rechargeable battery

83
DEVICE SPECIFICATIONS

TROUBLESHOOTING

System Troubleshooting Chart


LCD Message Problem Correction
BATTERY LOW – CHARGE UNIT Unable to acquire ECG or Charge the battery with AC power.
unable to print.
LEAD FAULT, NO ECG CAPTURE Lead fail or noisy ECG data. Correct faulty lead or noise.
NO ANSWER Unable to transmit ECG. Check for correct phone number.
Ensure modem and E-SCRIBE
are online.
Date/Time will not Save Defective Battery/Blown Fuse Test Fuse and Battery, correct defect.

ECG Troubleshooting Chart


Affected Leads Problem Correction
LEADS OFF OR ONE OR MORE OF Lead fail. Indication of
THE FOLLOWING: RA, LA, LL, V1, RL/RA/LA/LL/V1/V2/V3/V4/V5/V6.
V2, V3, V4, V5, V6 Check limb leads.

Correct faulty lead(s).


Lead I Missing/Noisy RA/LA. Check patient prep; re-prep if
necessary with new electrode.
Lead II Missing/Noisy RA/LL. Check patient prep; re-prep if
necessary with new electrode.
Lead III Missing/Noisy LA/LL. Check patient prep; re-prep if
necessary with new electrode.
All High Freq. Noise. Notch down filter from 300 Hz to 150
Hz; check proximity to power cables.

Transmission Troubleshooting Chart


LCD Message Problem Correction
TRANSMIT FAILED Unable to transmit ECG. Check phone line. Ensure site number
is valid. Try again.
ERROR-DICOM Not Enabled A DICOM communication was Configure the system to DICOM and
attempted, but the unit is not reboot.
configured for DICOM.
UNABLE TO SAVE ECG No available memory. Press stop to continue. Transmit or
mark records for deletion in the
ECG data too noisy to store. directory. Correct noise and try
acquisition/storage again.
DHCP FAILURE The WLAN module failed to get Contact Welch Allyn Technical Service.
an address from DHCP.
DPAC FAILURE WLAN failed to initialize. Contact Welch Allyn Technical Service.
CAN’T CONNECT TO ACCESS A link to the access point could Ensure the IP address is correct. If
POINT not be established. problem persists, contact Welch Allyn
Technical Service.
CAN’T CONNECT TO REMOTE A link to the access point was Ensure the IP address is correct. If
LINK established, but the link to the problem persists, contact Welch Allyn
destination failed. Technical Service.

84
TROUBLESHOOTING

SECTION 6

Transmission Troubleshooting Chart (continued)

LCD Message Problem Correction


UNABLE TO SAVE ORDER Order storage failed. Attempt to retransmit orders.
UNABLE TO SAVE WORK ITEM DICOM order storage failed. Directory full; mark records for deletion
or delete records.
INCORRECT RESPONSE Connection established, then Connection started but failed; attempt
failed. to reconnect.
NO CUSTOM ID Received orders failed. Previous Custom ID not compatible
with current Custom ID, or no Custom
ID.
PAPER QUEUE FAULT Unable to print. Add paper; manually advance page
Paper queue mark not evenly past closure point of writer and
detected as expected. close writer cover and press STOP.
CONNECTION FAILED Unable to transmit or receive Check for correct baud rate, phone
ECGs. number, and cable connections or site
number.
None File not successfully Check share permissions on host
transmitted via LAN. device.
None Unable to connect with LAN Implement hub vs. crossover cable.
with crossover cable.

Paper cue Fault


See Conformance Testing Section for Adjusting the Writer Cue Sensor.

Test Menu
Press F6 then Shift+ALT+x and enter the admin password.

1. Auto Test
a. This will test the LCD, Flash, ADC Readings and SDRAM.

2. Speaker
a. Tests the speaker

3. LCD Memory Test


a. Tests the LCD

4. Writer Test
a. Tst P - Prints a single sheet test
b. Rhy P – prints a test Rhythm page
c. Load – Prints all the interpretation statements – 9 pages

5. Full Keyboard Test


a. Tests each key individually

6. Reduced Keyboard Test


a. Short keyboard test

7. Configure Modem
a. Allows configuration of the modem if installed

8. Set Configuration

85
TROUBLESHOOTING

a. F1 - Set to Default – Caution – erases all settings


b. F2 - Set to Test – Caution – erases all settings
c. F3 – Set Custom Config. Enters configuration menu
d. F5 – Erase stored ECGs

9. Stats
a. F1 – CLR F – Clear all Flags. This will remove the Print, Transmit and Delete flags on all records
b. F2 – DUMP – This will copy all records to a USB flash drive without changing any flags
c. F3 – AQ LP – Acquire Forever Test. – Caution – This will acquire a ECG every 5 minutes. The
only way to stop it is to unplug the unit from AC and remove the battery.
d. F4 – Fill – This will fill the directory with records based on one record in the directory.

0. Auto Test List


a. SDRAM Test
b. Flash Test
c. ADC Reading
d. Checkdsk

Media – F2
This is used when the WLAN module is replaced.

86
CONFORMANCE TESTING TROUBLESHOOTING

Conformance Testing

Conformance testing is to be performed by Authorized Welch Allyn Service Representatives to verify the device is
functioning correctly after repair operations have been performed. Testing results should be documented on the test
data record (TDR) at the end of this section of the manual. Include the following printouts: Configuration page,
Writer Test, ECG, and Noise Test.

Print the device configuration (attach to the Test Data Record (TDR)
Power Testing

 Ensure battery is fully charged before performing these tests, voltage and current limits are based on a fully
charged battery.
 Ensure there is no power connected to the UUT AC inlet.
 Remove upper housing and writer assembly. Disconnect battery by pulling battery cable off of the red terminal.

NOTE: Based upon customer usage and age of battery, replace as needed.

 Note battery age (if possible)

This information can be found on the white “date code” sticker located on the battery (use the earliest date that
is not crossed out). Record Date on TDR

 Battery (open circuit)

Measure battery voltage using a voltage meter; verify the meter reads greater than 12.5vdc.
Record result on TDR

 Battery (load)

Measure the battery voltage using a volt meter and a power resistor load (10ohm, 20watt) in parallel with the
battery. After approximately 5 seconds, verify the meter reads greater than 11.7vdc. Record result on TDR

 Off current

Connect a current meter in line with battery. With the UUT power off, verify the current meter reads less
than 100 uA. Record result on TDR

 On current

Turn on the unit and verify the current meter reads less than 250 mA. Record result on TDR

 AC charging current

Apply AC power to the unit and verify that the current draw from the battery reverses polarity and the value
starts decreasing as time increases. Record result on TDR

 Battery charger output voltage

Disconnect the current meter and measure the battery charger output voltage between the red disconnected
battery cable and the negative terminal on the battery. It should read between 13.0vdc and 14.0vdc. Record
result on TDR

87
CONFORMANCE TESTING

 Verify all power cables are reconnected properly and reassemble unit.

Functional Testing

 AC LED/Display

Connect AC power cord to the unit and verify that the green AC LED (located to the left of the display)
illuminates continuous.

NOTE: The battery indicator will be clear when charging and will illuminate white when fully charged.

Turn the unit on and verify the text on display is clear and legible and there are no flickering or missing
lines/pixels. Record result on TDR

 Adjusting the Writer Cue Sensor

NOTE: This test should be performed with the AC power turned on. The test point, trim pot, and ground
are found on the front, upper left on the keyboard (see picture below).

1) Install paper into the unit with the cue mark approx. 1 – 2 inches away from the tear bar. Make sure that the
cue sensor is seeing white and not any markings on the paper.
2) Use a DMM to measure the DC voltage at test point (TP2/QSNS) on the keyboard with respect to ground
(GND1). Adjust R1 V_ADJ on the keyboard to between 1.95 V-DC and 2.05 V-DC at test point P2. Set this as
close to 2.0 V-DC as possible. Record result on TDR
3) Perform either the Writer test or print a test ECG. The paper should cue to the next sheet of paper, print and
then advance to the beginning of the next sheet of paper.
4) If the Writer test is performed the results should be compared with the test printout in this manual.

88
CONFORMANCE TESTING

 Writer

Open and close the writer door to verify smooth operation. Verify that the door unlatches without sticking and
that it latches completely. From the main screen, simultaneously press shift+alt+RHY. Verify that a test page
is printed and the writer stops on the cue mark. The perforation of the paper should line up with the tear edge
on the writer. Assure there are no gaps in the printing and the print darkness is uniform across the entire page.
Verify the writer gears do not skip and paper is tracking properly (you may need to print multiple pages to
observe this). Record result on TDR

 ECG & Keyboard Matrix

Connect an ECG simulator to the AM12 or WAM patient interface. Set the simulator to a known heart rate and
amplitude. Press the ECG key to capture an ECG. Verify there is an audible beep with each key press. Enter
Last name “PARCFL8” (Note: “PARCFL8” ensures the keyboard matrix is fully tested), then press F6 (Done).
Verify that 12 ECG traces print correctly and assess the printout quality. Ensure uniform darkness across entire
printout. Record result on TDR

 ECG Noise Test

Connect a Shorting Block (TF-0063) and adapter or equivalent to the AM12 or WAM patient interface. Set the
ECG gain on the unit to 20mm/mV. Print a rhythm strip (approx. 1 page). Verify that no channels have more
than 0.5mm of noise as measured by using Welch Allyn thermal paper (Smallest grid line = 1mm).
Record result on TDR.

 Lead Failure Test

Connect an AM12 or WAM patient cable to the patient input of the unit, with the other end connected to a lead
failure box (TF-0620 or equivalent) or patient simulator. Using the lead fail box, momentarily press each push
button to open the patient leads one at a time or disconnect one lead at a time from the simulator and verify the
display indicates an open lead condition for the corresponding lead. Record result on TDR

Lead Message
Right Arm ‐ LEADS OFF
Left Arm ‐ LA OFF
Left Leg ‐ LL OFF
V1 ‐ V1 OFF
V2 ‐ V2 OFF
V3 ‐ V3 OFF
V4 ‐ V4 OFF
V5 ‐ V5 OFF
V6 ‐ V6 OFF
All Leads off ‐ LEADS OFF

89
CONFORMANCE TESTING

 Communication Option Testing (as applicable)

The receiving station for modem, LAN and WLAN transmissions should be running Welch Allyn ELI-
Link software.
Refer to the ELI-Link user manual for proper configuration.
Verify successful transmission of all applicable communication options by acquiring ECG records that include
the
transmission method in the “Patient Name” field (such as Last Name = USBD) then subsequently transmitting
the
ECG record stored to a compatible receiving device. Consult the product user manual if needed to properly
configure the communication settings for each option present on the unit under test.
Successful transmission of the test records can be verified by viewing the ECG records in the unit directory
after
transmission and confirming they are marked as “transmitted” (as defined in the product user manual).
Record results on the TDR

- Modem
- LAN
- WLAN
- GSM/GPRS
- USB host (USB memory device needed)
-USBD

Device Cleaning

Clean unit per the instructions provided in the Maintenance & Cleaning section of the service manual.

Safety Testing

If the cardiograph housing was opened for repair or inspection work, the following safety tests should
be performed in accordance with the IEC 60601 standards.

The ELI150C and ELI250C are considered a Class 1 Type CF devices, intended to only be utilized
with the Welch Allyn AM12 or WAM patient input modules. Defibrillation isolation from the patient
is provided by the patient input modules, which are tested separately as part of the manufacturing
process (they are considered non-serviceable devices), therefore Hi-pot testing is not required for
these cardiograph models.

 Earth Leakage
 Enclosure Leakage
Non-conductive (fully insulated) chassis testing should be performed utilizing 200 cm2
conductive foil or equivalent, earth ground on AC input is utilized for functional earth (not
safety grounding).
 Patient Leakage
Applied part – patient input (utilize Welch Allyn AM12 patient cable)
 Patient Auxiliary Current
Applied part – patient input (utilize Welch Allyn AM12 patient cable)

90
CONFORMANCE TESTING

ELI 150C/250C Test Data Record

Unit Serial #:

□ Print device configuration (attach to this report)

Power Testing
Note Battery Age (If not possible enter N/A) / (week/year)
Battery (Open Circuit) Voltage VDC (>12.5vdc)
Battery (with Load) Voltage VDC (>11.7vdc)
Off Current uA (<100uA)
On Current mA (<250mA)
AC Charging Current PASS / FAIL (Circle One)
Battery Charger Output Voltage VDC (13.0-14.0vdc)

* Based upon customer usage and age of main battery, replace as needed.

□ Verify all power cables are properly reconnected and reassemble unit

Functional testing
AC LED/Display Functionality PASS / FAIL (Circle one)
Cue Sensor Calibration VDC (1.95-2.05VDC) PASS / FAIL (Circle one)
Writer Test PASS / FAIL (Circle one)
ECG & Keyboard Matrix Testing PASS / FAIL (Circle one)
ECG Noise Test PASS / FAIL (Circle one)
Lead Failure Test PASS / FAIL (Circle one)

Communication Option(s)
Modem PASS / FAIL / NA (Circle one)
LAN PASS / FAIL / NA (Circle one)
WLAN PASS / FAIL / NA (Circle one)
GSM/GPRS PASS / FAIL / NA (Circle one)
USB host PASS / FAIL / NA (Circle one)
USBD PASS / FAIL / NA (Circle one)

□ Device Cleaning

Safety Testing PASS / FAIL (circle result)


Earth Leakage
Enclosure Leakage
Patient Leakage
Patient Auxiliary Current

Performed by: Date: / /

91
CONFORMANCE TESTING

ELI 150c/250c COMMUNICATION OPTIONS

Communication Options

The following Communications Options are available on the ELI150c/ELI250c:

LAN
WLAN
Modem
GSM/GPRS Mobile transmission (ELI150c Only)
Transmission to USB Thumb-drive
USB Mount to Windows PC (USB Device option)

Sync Media (in Settings menu):

The Sync Media can be set to the following:

LAN + WLAN / WLAN + LAN


LAN + GSM / GSM + LAN
LAN + GPRS / GPRS + LAN

Sync Media will attempt a connection of the preferred method first, and if it fails or times out, will attempt the
secondary mode of communication.

Communication Error Messages

DPAC Failure
This error message occurs if the DPAC fails initialization. This likely indicates a hardware problem preventing
communication with the Module or a hardware problem with the Module itself.

Can't Connect To Access Point


This error message occurs if the module cannot associate with an access point. This likely indicates one of the
following things: the Module’s SSID is incorrect, no network with that SSID is available, the wrong security method
is chosen, or the AP’s configuration is incompatible with the Module.

DHCP Failure
This error message occurs if the Module failed to acquire an IP address via DHCP.

Can't Connect To Remote Link


This error message occurs if the module is able to communicate with the access point, but unable to establish a route
through the network to communicate with the Remote Host. This error message may indicate the Remote Host is
not ready to receive data. If the Remote Host appears to be ready, this message may indicate invalid entry of one of
the following network parameters: IP address, Def. Gateway, Sub Net Mask, Host IP, or Port Number. It may also
indicate that the correct security method was chosen but one or more of its required parameters were set incorrectly.

NOTE: It is possible that the Module can still transmit data with an incorrect gateway or subnet mask
depending on the path through the network the data is being transmitted.

92
Communication Options (Software only)

- LAN
- USB (USB Device option)

Communication Options (Hardware + Software)

WLAN GSM (150c Only)

Modem

Shortcuts

Configure Transmit Media

F6 (More)
Shift/Alt/X
Enter ‘admin’ password, Press Enter key

Directory Dump (To dump entire directory to USB thumb drive)

F6 (More)
F1/#1 (Directory of Stored ECG’s)
Shift/Alt/D

93