Location via proxy:   [ UP ]  
[Report a bug]   [Manage cookies]                

Meter Testing and Standards Laboratory: Address

Download as xls, pdf, or txt
Download as xls, pdf, or txt
You are on page 1of 1

GOVERNMENT OF KERALA

DEPARTMENT OF ELECTRICAL INSPECTORATE


METER TESTING AND STANDARDS LABORATORY
THIRUVANANTHAPURAM - 695 016
Phone No - 0471 2591080

File Number : MTSL/463/2023-T Job Number :403/23 / I / T / ET


No of Pages : 1 Date : 16-06-2023
TESTING OF Earth TESTER

Address Sri. Mahesh Kumar N S


Powertech Electricals
Pangappara, Thiruvananthapuram

Details of Earth Tester under testing


Make Waco
Serial Number 341386
Model DET
Range 0-10Ω-100Ω-1000Ω

Details of Equipment used for Calibration


Name of Equipment Make Model S.No
High Precision Decade Resistance Box RA Instruments 55X-DRB 1540

Date Of Receipt: 6/7/2023 Date Of Test:16-06-2023

TEST RESULT

Resistance in Ω: Range 10Ω


Actual reading 1 2 3 4 5 6 7 8 9 10
Indicated reading 1 2 2.94 3.84 4.94 5.83 6.86 7.68 8.67 9.88
% error 0.00 0.00 -2.00 -4.00 -1.20 -2.83 -2.00 -4.00 -3.67 -1.20
Resistance in Ω: Range 100Ω
Actual reading 10 20 30 40 50 60 70 80 90 100
Indicated reading 10 19.9 30 40 49 59 70 79 89 100
% error 0.00 -0.50 0.00 0.00 -2.00 -1.67 0.00 -1.25 -1.11 0.00
Resistance in Ω: Range 1000Ω
Actual reading 100 200 300 400 500 600 700 800 900 1000
Indicated reading 99 198 293 391 482 596 688 795 880 964
% error -1.00 293.00 391.00 482.00 -3.60 -0.67 -1.71 -0.63 -2.22 -3.60
Remarks: : All errors are within tolerable limit.

(Note: This test result pertains to the tested samples only and shall not be reproduced
except in full without the written approval of this laboratory)