Ie MLG m1xxx 10
Ie MLG m1xxx 10
Ie MLG m1xxx 10
com
Instruction manual
M1140
M1200
M1220
M1300
M1320
M1321
M1322
M1323
M1440
Multilogger
© Copyright: COMET SYSTEM, s.r.o.
It is prohibited to copy and make any changes in this manual, without
explicit agreement of company COMET SYSTEM, s.r.o. All rights reserved.
2 IE-MLG-M1xxx-10
Content
INTRODUCTION ........................................................................................ 4
GENERAL SAFETY PRECAUTIONS ......................................................... 6
GETTING STARTED .................................................................................. 7
Multilogger in portable application ................................................. 7
Multilogger in stationary application .............................................. 8
INSTALLATION INSTRUCTION ................................................................. 9
Recommendations for mounting ................................................. 12
INPUTS OF MULTILLOGER .................................................................... 15
Input connectors ......................................................................... 16
ALARM OUTPUT ..................................................................................... 20
POWER SUPPLY OF MULTILOGGER .................................................... 22
KEYBOARD OPERATION ........................................................................ 25
Control and indication parts ........................................................ 25
Basic display operation ............................................................... 26
Main menu .................................................................................. 28
CONNECTING WITH THE COMPUTER .................................................. 32
SOFTWARE COMET VISION .................................................................. 33
DEVICE SETUP USING SOFTWARE ...................................................... 35
Device configuration basics ........................................................ 35
Device configuration items .......................................................... 35
APPLICATIONS NOTES .......................................................................... 45
OPERATION AND MAINTENANCE RECOMMENDATIONS.................... 51
TECHNICAL PARAMETERS .................................................................... 54
Power supply .............................................................................. 54
ALARM OUT output .................................................................... 56
Communication interfaces ........................................................... 57
Measuring, data memory and real time clock .............................. 58
Electromagnetic compatibility ...................................................... 58
Operation and storage conditions ............................................... 59
Mechanical characteristics .......................................................... 59
Technical parameters of the inputs ............................................. 60
APPENDIXES .......................................................................................... 71
Appendix 1: Selected error messages......................................... 71
Appendix 2: Output and power supply circuits............................. 73
Appendix 3: Input circuits ............................................................ 74
Appendix 4: Connecting of Pt1000 RTD probe ............................ 75
IE-MLG-M1xxx-10 3
Introduction
The Multilogger has been designed for measuring and recording physical
and electric quantities with an adjustable logging interval from 1 second to
24 hours. The device is equipped with 4 inputs (connectors) for external
probes or signals. Moreover, some versions also include an internal
atmospheric pressure and/or CO2 concentration sensor. The production
line consists of several models with various combinations of the input
ranges of the measured quantities. The Multilogger can be used either as
stationary or portable device.
4 IE-MLG-M1xxx-10
Multilogger drawing:
IE-MLG-M1xxx-10 5
General safety precautions
6 IE-MLG-M1xxx-10
Getting started
• Connect the appropriate probes and signals to the device - for the
DigiS/M, DigiL/M and Pt1000/M probes, follow the optical key for correct
orientation of the connector. Probes of the DigiL/M-xx line with a cable
can placed in the holder on the side of device (optional accessory). For
more information, see Chapter “INPUTS OF MULTILOGGER”.
• Switch on the Multilogger - push the switch button on the device. The
display lights and the instrument starts working. If you will communicate
with PC via the Ethernet interface, the power adapter +5V DC is
necessary (supplied accessories). For more information, see Chapter
“POWER SUPPLY OF MULTILOGGER”.
• Install Comet Vision software to your computer - it is available at
www.cometsystem.com. If you purchased a license for full version,
register it.
• Connect Multilogger with your PC - you can use the USB, Ethernet or
RS232 port (special cable necessary). The external power supply is
needed for Ethernet interface communication. For more information, see
Chapter “CONNECTING WITH THE COMPUTER”.
• Device settings - use SW to set up the device in accordance with the
connected probes and your requirements. If you are not going to use the
device soon, turn it off and place it in its transportation case. For more
information, see Chapter “DEVICE SETUP USING SOFTWARE”.
• Measurements - remove the Multilogger from the transportation case,
connect probes and external power adapter (if needed) and turn the
device on. Some device settings can be changed from its keyboard.
When measurement ends, move the device to your computer room and
download the recorded data to PC. For more information, see Chapter
“KEYBOARD OPERATION”.
• Charging batteries - if the Multilogger includes rechargeable batteries,
you can recharge them when needed. For more information, see also
Chapter “POWER SUPPLY OF MULTILOGGER”.
• Maintenance and periodic inspections - In order to secure reliable
operation of the device, it is recommended to make regular inspections. For
more information, see part “OPERATION AND MAINTENANCE
RECOMMENDATIONS”.
IE-MLG-M1xxx-10 7
Multilogger in stationary application
• Connect the appropriate probes and signals to the device - for the
DigiS/M, DigiL/M and Pt1000/M probes, follow the optical key for correct
orientation of the connector. Probes of the DigiL/M-xx line with a cable
can be placed in the holder on the side of device (optional accessory).
For more information, see Chapter “INPUTS OF MULTILOGGER”.
• Switch on the Multilogger - push the switch button on the device. The
display lights and the instrument starts working. If you will communicate
with PC via the Ethernet interface, the power adapter +5V DC is
necessary (supplied accessories). For more information, see Chapter
“POWER SUPPLY OF MULTILOGGER”.
• Install the COMET Vision software to your computer - it is available
at www.cometsystem.com. If you purchased a license for the extended
SW, register it.
• Connect Multilogger with your PC - you can use the USB, Ethernet or
RS232 port (special cable necessary). For Ethernet interface
communication the external power supply is needed. For more
information, see Chapter “CONNECTING WITH THE COMPUTER”.
• Device settings - use COMET Vision software to set up the device in
accordance with the connected probes and your requirements. If
needed, check the alarms functionality. For more information, see
Chapter “DEVICE SETUP USING SOFTWARE”.
• Maintenance and periodic inspections - In order to secure reliable
operation of the device, it is recommended to make regular checks. For
more information, see part “OPERATION AND MAINTENANCE
RECOMMENDATIONS”.
8 IE-MLG-M1xxx-10
Installation instruction
Device dimensions:
IE-MLG-M1xxx-10 9
The way of mounting the holder to Multilogger and mounting holes
dimensions:
MLP001
MLP002
10 IE-MLG-M1xxx-10
MLP004
IE-MLG-M1xxx-10 11
Recommendations for mounting
12 IE-MLG-M1xxx-10
• Do not use cables longer than 30 m generally; for RTD sensors is
recommended max. 15 m; cables must be installed only in interior areas.
• Use shielded cables for the voltage and current inputs - connect the
shielding to the Multilogger but not to the end devices if these do not
have a special terminal for it.
In this case, shielding must
not be connected to external
metal parts of this device or
to any other device. Never
use shielding as a signal
wire.
IE-MLG-M1xxx-10 13
• Do not use common leads for several inputs.
14 IE-MLG-M1xxx-10
Inputs of Multilloger
Manufactured models differ from each other by their input ranges and by
input connectors. Prior to connecting voltage and current input signals, you
should check their parameters in the following part: “TECHNICAL
PARAMETERS OF THE MULTILOGGER INPUTS”.
IE-MLG-M1xxx-10 15
Input connectors
The Multilogger does not have galvanically isolated inputs. For this reason,
only devices that are not connected to any other circuit may be connected
to the inputs of Multilogger. Simplified circuit diagram of inputs is shown in
Appendix 3.
16 IE-MLG-M1xxx-10
Two voltage ranges are available on each
universal thermocouple connector for measuring
small voltages:
• range -60 mV to +140 mV
• range -18 mV to +18 mV
Use the white uncompensated connector
connected according to the picture.
Terminals ______________________________________
Inputs (0 to 20) mA, (0 to 10) V, binary inputs and the counter are
equipped with a two-part, self-locking WAGO terminal block. Insert
flat-bladed screwdriver to
rectangle terminal hole and push
screwdriver towards away from
you - contact is released. Insert
wire to released terminal (circular
hole behind the rectangular one)
and close the terminal by
removing the screwdriver.
The terminal block can be removed from the device without the need to
disconnect individual wires!
IE-MLG-M1xxx-10 17
Connection of the transmitter with voltage output:
18 IE-MLG-M1xxx-10
Connection of the transmitter with the two current outputs (passive):
IE-MLG-M1xxx-10 19
Alarm output
ATTENTION - the External audio alarm unit will not work if the
Multilogger does not have an external power supply.
• External alarm with its own power supply - make the connection
according to the following picture. Be careful do not exceed the
maximum allowed output current from OUT terminal (see the part
Technical Parameters). Use an external relay for higher currents.
20 IE-MLG-M1xxx-10
IE-MLG-M1xxx-10 21
Power supply of Multilogger
The Multilogger is powered from three AA batteries and optionally also from
an external +5 V DC power supply, which can be also used for charging the
batteries. A simplified schematics of the internal supply circuits is showed in
Appendix 2.
Batteries _______________________________________
• The new Multilogger is equipped with high-quality NiMH
rechargeable batteries; the device is preset to intelligent algorithm
to their continuous charging if external power is present.
• Charging batteries:
a) batteries can be charged if an external power source is
connected
22 IE-MLG-M1xxx-10
b) the charging process will start if the NiMH charging
accumulators are selected in the device configuration
c) when the Multilogger is turned off, the batteries cannot be
charged
d) the Multilogger allows for two battery charging modes -
high current fast charging mode and low current slow
charging mode
• Slow low current charging mode is mainly used for stationary
installations. Batteries work as a backup source in case of
external power failure. Charging is automatically activated
after restoring of external power supply. During slow
charging device operates normally. Recharging of
completely discharged battery takes several days, however,
in normal operation, do not expect a completely discharged
battery. The charging process is indicated on the device
display using the corresponding symbol .
IE-MLG-M1xxx-10 23
• Battery replacement:
If you want to keep the correct time setting in your device,
do not let the battery discharge completely.
When replacing the battery, do it quickly. The timer circuit
runs without battery for max. 5 minutes if the device is
switched off.
a) Switch the Multilogger off.
b) Remove the battery cover, take out the old batteries and
insert new batteries (take care for correct polarity). Battery
polarity is shown on its holder.
c) Switch the Multilogger on and make sure the set time is
correct (device menu View settings - General settings).
d) If you change the battery chemistry (rechargeable <=>
non- rechargeable), you have to make this change in the
device configuration also.
24 IE-MLG-M1xxx-10
Keyboard operation
The front panel of the Multilogger includes 6 keys, 3 indication LED diodes
and a graphic display.
• The red key with the switch symbol is used for Multilogger turning on/
off. The MIN/MAX keys (with the arrows) are used for quick show of the
minimal and maximal measured values and also for moving through
menu items. Function of keyboard upper row buttons is defined by actual
description of each button on the lower LCD line display.
• The indication LED diodes inform us about the current state of alarms
and allows the operation personnel to observe this important information
even under low light and from a greater distance.
• The LEDs show an overview about all active alarms in the Multilogger. In
the device configuration, we can assign individual alarms to the LED
diodes for customizing the indication to our needs.
IE-MLG-M1xxx-10 25
Basic display operation
Current measured values (or min/max) and their units are shown on the
device display. To make it easier, the short description of measured
quantity, the corresponding input number or a user description of the input,
can be optionally displayed. Using the VIEW multifunction key, you can
switch among these views on the display. You can choose from 3 different
display options: input number + short description of measured quantity,
input number without the short description of measured quantity and
channel number + short description of measured quantity. If the Multilogger
setup allows it, audio signaling of the currently active alarms can be
temporarily muted using the MUTE multifunction key.
The bell icon on the line of the measured value indicates an active alarm for
this quantity.
The status line of the display shows icons that indicate the current status or
device mode:
26 IE-MLG-M1xxx-10
active recording of analog values (in the “during alarm
only” or “controlled by an external input” recording modes)
inactive recording of analog values (in the “during alarm
only” or “controlled by an external input” recording modes)
external power supply present (if the status line is
occupied by icons of a higher priority, it is not displayed)
minimal measured values are displays
low voltage of the real time clock circuit detected, the time
could be incorrect!
alarm at least on one channel (the particular channel is
marked with a bell by the given value)
system alarm (alarm sources can be found in the System
information menu)
IE-MLG-M1xxx-10 27
Main menu
28 IE-MLG-M1xxx-10
c) Real time clock circuit state: tick - OK, exclamation mark
- low voltage of the real time clock circuit detected, the
time could be incorrect!
d) Activities of the Alarm out signal: tick - active, field
without a tick - inactive.
e) MUTE activated: tick - was activated (alarm is still active),
field without a tick - was not activated (or the alarm is
inactive).
f) Activities of EN signal: tick - active, field without a tick -
inactive.
g) battery voltage: range from 3.4 V (completely discharged
battery) to 4.5 V (new or just charged battery)
h) device internal temperature (affects the battery charging)
i) cold junction temperature of the thermocouples (only for
models with a thermocouple connector)
j) Multilogger firmware version.
IE-MLG-M1xxx-10 29
3. Display settings _______________________________
Allows to set the parameters of the displaying:
• Font size (large, medium, small)
• Displaying input description of the measured quantity (yes, no)
• Order of the displayed channels (based on the input number,
based on the channel number, selection)
30 IE-MLG-M1xxx-10
6. Deactivation of Latched alarm ____________________
If the “Latched alarms” function is enabled in the device configuration,
the active alarms are not reset when the measured values return
back to the allowed limits. Instead, they remain active until they are
deactivated by the user. They thus warn us that “something was
happening” during the monitored period, i.e. that the alarm limits were
exceeded. This function can be protected using PIN2.
IE-MLG-M1xxx-10 31
Connecting with the computer
You can connect the Multilogger to the computer through the following
communication interfaces:
32 IE-MLG-M1xxx-10
Software COMET Vision
• the Windows 7 and later or the Windows Server 2008 R2 and later
operating system
• 1.4 GHz processor
• RAM 1 GB
IE-MLG-M1xxx-10 33
How to work with the program ______________________
The main menu of the program is on the right side with items:
“Home”, “Devices”, “Files” and “Online Displays”. These components
control appearing bookmarks. Each item has its bookmarks, one of
them named “Home” is similar for all items. The other options are
specific for each item.
34 IE-MLG-M1xxx-10
Device setup using software
Upon being the Multilogger switched on using the red button, the device
begins to take measurements and to record and evaluate alarms according
its user configuration. Complete configuration of the device can be done
using the COMET Vision software. Some parameters can be changed from
the device keyboard or from its web pages. When editing the Multilogger
configuration, the device normally works, however, access to some of its
functions is limited (downloading records, editing of the device configuration
by other users).
IE-MLG-M1xxx-10 35
General - Preferences ____________________________
Here you can select language localization Language on the device
terminal, Default temperature unit °C/°F and, for some models,
Atmospheric pressure unit. If your model does not measure
atmospheric pressure but it is necessary to know it (for calculated
humidity values and CO2 concentration), you need to specify Ambient
average pressure.
For barometers, you can enter a correction constant for sea level
recalculation as the actual sea level pressure or altitude. Actual sea
level pressure can be obtained from, for example, nearby observatory
or nearby weather station.
36 IE-MLG-M1xxx-10
General - Advanced - Device start ___________________
Here you can disable the “Off key” of the device.
The Ethernet interface does not work without external power supply.
However, this interface can send a warning e-mails and records to
the database using DataLink protocol in the battery mode if you turn
on this feature. These messages will be sent immediately after power
failure or on new alarm. Additionally, they can be sent repeatedly at
the specified interval.
IE-MLG-M1xxx-10 37
Record ________________________________________
This options are used for controlling the record functions of the
Multilogger. Recording can be turned on and off as needed
(Recording On/Off). However, if the record is turned off and then you
turn it on again, this takes up 0.1% of the free memory capacity. Use
the Cyclic record if you require the device to continue the recording
even when the data memory is completely full. In this case, the oldest
data are replaced by new ones. If you do not use this option, the
recording will stop once the memory is full. Another important
parameter is the Recording interval. It can be set from 1 second to 24
hours (see the Technical Information). The recording is performed in
integral multiples of a set interval. For example, if you turn the device
on at 5:05 a.m. and if you have set the logging interval at 1 hour, the
first data will be recorded at 6:00 a.m., next ones at 7:00 a.m., etc.
By default, the device records the actual values at the time of the
recording interval. However, if continuous recording is on (see
above), you can select recording of average, minimal and maximal
values during recording interval. These settings can be performed in
the section Recorded values.
38 IE-MLG-M1xxx-10
Alarms can be signaled using the LED diodes directly on the
Multilogger (Optical signalization - LED). Since the Multilogger can
work in the battery mode, additional options are available for
minimizing the current consumption. For a state without an alarm can
be selected green LED indication. When the device is powered from
batteries, you can select more energy economical mode, under which
the LED diodes only shortly blink in the case of an alarm.
Alarm signaling can be also performed using the ALARM OUT output.
Upon an alarm, the Multilogger can switch on the open collector
transistor of this output in accordance with the description in Chapter
“ALARM OUTPUT”. This transistor can be optionally switched on or
switched off when there is an alarm. The Mute function can be also
used for this output.
Here you can also allow remote erasing of the alarm memory and
remote deactivation of alarm signalization from the software
COMET Vision or from the web (Remote Mute & Erase of Latched
alarms); moreover, you can permit or forbid alarm deactivation from
the device keyboard (Local Mute enabled). To protect access to these
functions, you can use the PIN2 code (Security panel).
IE-MLG-M1xxx-10 39
Alarm events - Mute of signalization _________________
The Mute function is used for deactivating alarm signalization by the
user. It can be optionally applied to audio signalization, the ALARM
OUT output or both. It can be performed from the keyboard, from the
web or from the software COMET Vision. To protect access to these
functions, you can use the PIN2 code, which the operation personnel
must enter. This function deactivates enabled alarm signalizations
until a new alarm occurs. The deactivation can be permanent (until a
new alarm occurs) or temporary (if the alarms remain active even
after selected time out, the signalization is restored).
40 IE-MLG-M1xxx-10
Security ________________________________________
The Multilogger allows an easy security option by the means of PIN
codes:
PIN1: protection against unauthorized changes of the
device configuration (from a PC or device keyboard),
erasing of data memory and counter erasing
PIN2: protection against unauthorized alarm deactivation
(Mute), confirmation of alarm memory (Latched
alarms) and erasing of the Min/Max values
IE-MLG-M1xxx-10 41
Ethernet - Email - General _________________________
Here you can do the basic email setup. If you want to use the sending
of warning emails on alarms, switch this function on (Email sending
ON). Next, set the IP address or SMTP server or its domain name
(SMTP Server), port (SMTP Server port) and, if the server requires
authentication, enter your login name and password (Autentication).
Continue by entering addresses of up to three recipients of the
warning emails (Email recipient #1 - #3) and, finally, enter the
address of the email sender (Sender). This is usually the same as the
user name in the SMTP authentication.
You can use shortened emails (Short email) especially for forwarding
them as SMS messages to mobile phones. You can also insert your
own comments to the emails: if the alarm appears (Alarm note), if the
alarm expires (Clear alarm note), in the case of system alarm
(System alarm note) and in the case of exceeding the memory
occupation limit (Memory overflow note).
42 IE-MLG-M1xxx-10
Ethernet - Web __________________________________
The Multilogger has its own web pages, where you can monitor
currently measured values and alarm states, and where you can
change some configurations. First of all, you have to allow the
website option (Web pages On). If necessary, you can change the
port number (Web server port), which is set at 80 by default. If you
want to make configuration changes through the website, you have to
allow this function (Use web for device configuration).
Inputs _________________________________________
Here you can specify the quantities that you will measure on every
one inputs, their recording and alarm states. You can see a simple list
of all inputs and measured quantities on the screen.
First of all, specify the probe or signal types connected to each input
(Settings button). Here you also choose the measured quantities that
you will acquire from this probe or input. You can acquire multiple
quantities at the same time from some probes (for example,
temperature, relative humidity, dew point, etc.). However, you can
also decide not to use a specific input, in this case click on Clear
button or remove all set measured quantities. You can name the input
by your own name (Input name), which you will then see by all
measured values. Once you are done with this configuration, you can
return back to the list of inputs and measured quantities.
Now, when you have correctly set your inputs and selected measured
quantities, you can configure them in more details using the Edit
button. For each measured quantity, you can choose if it should be
recorded (Recording) and if you wish to use the default setting (Basic
settings - default) of the physical unit (6 characters), short description
of the measured quantity (4 characters) and number of decimals.
However, you can also opt for your own setup (Basic settings -
custom) and enter the stated items manually. If you need the convert
IE-MLG-M1xxx-10 43
the measured values to different values using a two-point linear
transformation, turn on the User calibration function (User calibration).
For example, you can use it for converting current values measured
using the current loop to a different physical quantity or for applying
recalculation to the measured values.
In the next step, set alarms for individual measured quantities. You
can set up to 2 alarms for each measured quantity (Alarm 1 and
Alarm 2 buttons). Each alarm has to be enabled first (On/Off).
Next, select if the alarm occurs upon exceeding or dropping under the
given limit value (value is greater than / value is lower than) and enter
this limit. For binary inputs only define the state of the input. Next, you
need to specify the alarm delay (for the duration of), which is used for
eliminating short-term exceeding of the limit value. Hysteresis has a
similar function (Alarm hysteresis) - it is used to prevent alarm
oscillation when the measured value fluctuates nearby the alarm limit.
We recommend not to set it at zero. We have hereby completed the
setup for alarm occurrence. Now we just need to specify if the alarm
will be indicated optically on the Multilogger (Optical signalization -
LED), acoustically (Activate internal audio signalization), by the
ALARM OUT output (Activate ALARM OUT output) or by submitting a
warning email (Send email message ...).
44 IE-MLG-M1xxx-10
Applications notes
IE-MLG-M1xxx-10 45
We recommend to use the Multilogger with NiMH accumulators;
specify the following device configuration:
• NiMH accumulators (if they are installed).
• Do not turn on the Energy saving mode. The Multilogger will
perform measurement with a fast sampling (1 second for all
quantities with the exception of CO2 concentrations, which will be
measured every 2 minutes).
• Switch on the display backlight as needed. If it is necessary to
make the measured values on the Multilogger well visible all the
time, turn the backlight on only when the device is powered from
an external power supply.
• Use the optical alarm signalization using the LED diodes as
needed.
• Set the audio alarm signalization as needed.
• If you want to operate the Multilogger “remotely”, disable “Off key”
on its keyboard (COMET Vision software, option General -
Advanced - Device start). You will thus prevent unauthorized or
unwanted device turn-off by unqualified personnel. In these cases,
it is also advisable to limit access to the device configuration from
the keyboard using PIN codes.
• If your LAN is backed up and works correctly during mains failure,
it is possible to send warning e-mails and records to COMET
database (DataLink). See part Device setup using software,
General - Advanced – Peripherals, Ethernet in battery mode
options.
• Attention - if you use an external audio unit powered from the
Multilogger, it will not work during external power supply failure!
After an external power failure, when the device was running from the
batteries for a while, the batteries are automatically recharged by low
current as soon as the external power is restored.
46 IE-MLG-M1xxx-10
Recommended settings for a long time battery powered
application ______________________________________
In this application, the Multilogger is powered only from batteries and
a fast response to changing measured quantities is not required,
however, it is required that the device runs from the batteries as long
as possible. Recorded data are downloaded after measurement.
IE-MLG-M1xxx-10 47
If the Multilogger is not equipped with an internal pressure sensor,
you have to enter this value using the SW in the device configuration,
item General - Preferences. You can enter either the current
pressure value at the given location (without conversion to the sea
level) or the current altitude.
48 IE-MLG-M1xxx-10
How to set a variable alarm limit _____________________
In some applications, you need the alarm limit to depend on another
measured value.
IE-MLG-M1xxx-10 49
A similarly for sensors with a voltage or current output - unwanted
couplings can cause incorrect measurements.
If the Multilogger reports an error state, you can find more information
in Appendix 1 - Selected error messages.
50 IE-MLG-M1xxx-10
Operation and maintenance recommendations
IE-MLG-M1xxx-10 51
Recommendations for metrological verification _________
Metrological verification is performed in accordance with application
requirements specified by the user. In some cases, calibrations must
be performed by the accredited laboratory. The manufacturer
recommends to perform these verifications once a year.
• metrological verifications
• regular revisions and inspections subject to the intervals specified
in the corresponding standards
• assessments of all problems that occur since the last inspection
• visual inspection of the device and checking of the connectors
• device functionality inspections (i.e. the functions used for a given
application):
a) verification of data transfer to computer
b) verification of alarms - change input value to activate
alarm and check display message and alarm
signalization (if used)
c) check the battery voltage (in device menu)
52 IE-MLG-M1xxx-10
• Verification of cabling - check the connection quality of cables,
check visually entire cable length for damage and the route of
cables for interference, especially whether some parallel power
wires are not near.
Placing out of operation after the end of the device life ___
Disconnect the power, remove batteries and return data logger to the
supplier or specialized company as electric waste.
IE-MLG-M1xxx-10 53
Technical parameters
Power supply
Batteries _______________________________________
Recommended battery type:
3x AA NiMH rechargeable batteries Sanyo ENELOOP
1,900 mAh, 1.2 V
Alternative batteries:
3 x AA alkaline battery 1.5 V
________________________________________________________________________________________________
54 IE-MLG-M1xxx-10
Backing up of date and time in the Multilogger:
< 5 minutes after removing the batteries and with the
device turned off
________________________________________________________________________________________________
IE-MLG-M1xxx-10 55
Fast charging mode current:
approximately 350 mA
________________________________________________________________________________________________
This output has been designed, for example, for external acoustic
signalization or a telephone dialer. The way of its activation can be
programmed when configuring the Multilogger. The open connector
transistor (OUT), external power voltage (+5 V) and common terminal
(GND) are available on the device terminals. The wiring of the output
circuits is shown in Appendix 2.
56 IE-MLG-M1xxx-10
Communication interfaces
RS232C _______________________________________
Communication signals:
RxD, TxD, GND
Galvanic isolation:
Interface is not galvanically isolated
Connector RS232:
Terminal block
Maximum cable length:
15 m, only in interior areas
Baudrate:
115,200 Bd, optionally 230,400 Bd
USB ___________________________________________
Compatibility:
USB1.1 and USB 2.0
Connector USB:
mini USB
Vendor ID:
0403
Product ID:
6001
Baudrate:
460,800 Bd
IE-MLG-M1xxx-10 57
Ethernet _______________________________________
Compatibility:
10/100 MBit Ethernet, galvanically isolated
Connector for Ethernet:
RJ45
Electromagnetic compatibility
The device is tested according to standard EN 61326-1: 2006,
Article 6, Table 1.
Emission:
EN 55011: class B
58 IE-MLG-M1xxx-10
Immunity:
EN 61000-4-2: class A (4/8 kV)
EN 61000-4-3: class A (3V/m)
EN 61000-4-4: class A (0.5/1 kV)
EN 61000-4-5: class A (external power supply only)
EN 61000-4-6: class A (3 V)
Mechanical characteristics
Dimensions (h x w x d):
178 x 95 x 37 mm without cables and connectors
Weight:
approximately 380 g, including batteries
Protection:
IP20
Connecting terminal block
two-part, maximum cross-section of the conductor:
1.5 mm2
Included mounting accessories:
MLP001 - wall holder for Multilogger
IE-MLG-M1xxx-10 59
Optionally mounting accessories:
MLP002 - lockable wall holder for Multilogger
MLP003 - magnetic holder for Multilogger
MLP004 - 18 mm probe holder for Multilogger
The Multilogger allows measure temperature not only in °C, but also in °F.
If you use °F, the below stated ranges and accuracy have to be
recalculated to °F.
60 IE-MLG-M1xxx-10
The following devices can be connected to the input with
the MiniDIN connector:
• Digi /M digital probe for measuring temperature and
humidity.
Measured quantities:
Temperature
Relative humidity
Calculated quantities:
Dew point
_______________________________________________________________________
IE-MLG-M1xxx-10 61
• Custom Pt1000 RTD probe for temperature
measuring
The input parameters are identical to the above
described Pt1000/M. For these probes, you have
to compensate the cable length using the User
calibration manual settings. The connector wiring
is shown in Appendix 4. The manufacturer
recommends to preferably use the Pt1000/M
intelligent probes.
62 IE-MLG-M1xxx-10
The following devices can be connected to this input:
• Thermocouple of the “K” type (NiCr-Ni)
Measured quantity:
temperature
Range:
(-200 to 1,300) °C
Accuracy (without probes)
± (0.3% from the measured value + 1.5 °C)
Resolution:
better than 0.1 °C *)
Cold junction:
compensated in the temperature range of (-10
to 60) °C
______________________________________________________________________________________
IE-MLG-M1xxx-10 63
Resolution:
better than 0.1 °C *)
Cold junction:
compensated in the temperature range of (-10
to 60) °C
______________________________________________________________________________________
64 IE-MLG-M1xxx-10
• Thermocouple of the “N” type (NiCrSi-NiSiMg)
Measured quantity:
temperature
Range:
(-200 to 1,300) °C
Accuracy (without probes)
± (0.3% from the measured value of + 1.5 °C)
Resolution:
better than 0.1 °C *)
Cold junction:
compensated in the temperature range of (-10
to 60) °C
______________________________________________________________________________________
IE-MLG-M1xxx-10 65
Resolution:
better than 0.5 uV *)
Input impedance
approximately 107 Ω
Minimum allowable input voltage:
-0.2 V
Maximum allowable input voltage:
+3 V
Connector:
miniature uncompensated thermocouple
connector (white color)
______________________________________________________________________________________
66 IE-MLG-M1xxx-10
Universal input with a terminal ______________________
IE-MLG-M1xxx-10 67
• DC current in the range of 0 to 20 mA
Range:
(0 to 20) mA DC
Accuracy:
± 20 uA
Resolution:
better than 0.1 uA *)
Input impedance
approximately 100 Ω
Minimum allowable input current:
0 mA (disconnected circuit)
Maximum allowable input current:
limited to approximately 40 mA
Maximum cable length:
maximum of 30 meters, use shielded cables
______________________________________________________________________________________
The binary input will not work correctly if the source has
a high impedance at its “switched on” state. The problem
can be solved by connecting an external resistor of an
impedance of approximately (1 to 10) kΩ connected in
parallel to the input.
Input voltage for level “switched off”:
>2V
Minimum input voltage:
0V
Maximum input voltage:
+30 V DC
68 IE-MLG-M1xxx-10
Impedance of the “switched on” contact:
< 10 kΩ
Impedance of the “switched off” contact:
> 300 kΩ
Voltage on the disconnected contact:
approximately 3 V
_________________________________________________________________________________
IE-MLG-M1xxx-10 69
Internal barometric sensor_________________________
Models equipped with this sensor:
Parameters:
Measured quantity:
barometric pressure (absolute)
Range:
M1321: 600 hPa to 1,100 hPa
M1323: 700 hPa to 1,100 hPa
Accuracy:
±1.3 hPa at 23 °C
Other options:
recalculation to the sea level
Supported physical quantities:
hPa, kPa, mbar, mmHg, inHg, inH2O, PSI, oz/in2
Parameters:
Measured quantity:
CO2 concentration in the air
Range:
0 to 2,000 ppm
Accuracy:
± (50 ppm + 2% of the measured value)
at 23 °C and 1,013 hPa
70 IE-MLG-M1xxx-10
Appendixes
IE-MLG-M1xxx-10 71
Error Description and solution
Error 12 The CO2 concentration sensor is not available or does not
: work correctly.
Error 16 Send the Multilogger for repair.
The source value for the calculated channel is not available.
Error 19
Check calculated channel settings and its source values.
Error 20 Calculation error, incorrectly calibrated device.
Send the Multilogger for repair.
Error 21 Counter overflow.
Erase the counter using the SW and repeat measurement.
If you do not want to be considered the counter overflow as
error you can change this setting of the input.
Error 22 Internal voltage error. Undesirable coupling among the
inputs, incorrect connection, damaged cables, damaged
device.
Check the input connections and the connected cables.
Error 50, Device configuration error, damaged configuration.
Error 53 Download the configuration to the SW, check of the settings
: of individual items and save them back to the Multilogger.
Error 56
Error 51 The probe of the Digi/M line contains invalid calibration
constants.
Send the probe for repair or contact the manufacturer.
Error 52 The probe of the Pt1000/M line contains invalid calibration
constants.
Send the probe for repair.
Error 90 The Ethernet interface - the measured value is out of
Error 91 numerical range.
Verify the measured value on the device display, contact
the manufacturer.
Error 128 Communication error.
: Repeat the communication, contact the manufacturer.
Error 255
72 IE-MLG-M1xxx-10
Appendix 2: Output and power supply circuits
IE-MLG-M1xxx-10 73
Appendix 3: Input circuits
74 IE-MLG-M1xxx-10
Appendix 4: Connecting of Pt1000 RTD probe
IE-MLG-M1xxx-10 75