GF-9900 Transmitter-Manual
GF-9900 Transmitter-Manual
GF-9900 Transmitter-Manual
Quick Start
Your new Signet 9900 Transmitter needs to be
calibrated and the sensor needs to be initialized
prior to use. The following steps outline the
recommended procedure to start up a new system.
Description
The 9900 Transmitter, a member of Signet's line of SmartPro™
instruments, provides a single-channel interface for all Flow, pH/
ORP, Conductivity/Resistivity, Salinity, Pressure, Temperature,
Level and other applications. The 9900 is available in either
Panel Mount or Field Mount versions. Both versions can run on
12 to 32 VDC power (24 VDC nominal), and can power certain
sensors on loop power.
Field Mount
Safety
WARNING!
● Follow instructions carefully to avoid personal injury.
● This unit is designed to be connected to equipment which can
be hazardous to persons and property if used incorrectly. Read
and understand all associated equipment manuals and safety
warnings before using with this product.
● Remove power to unit before wiring connections.
● Wiring connections to this product should only be performed by
qualified personnel.
● Do not use unit if display is cracked or broken.
Contents
Table of Contents
New Features .................................................................................3
Compatibility ...................................................................................3
Dimensions .....................................................................................4
Module Installation..........................................................................5
Plug-In Modules........................................................................6
Relay Module............................................................................6
Direct Conductivity/Resistivity Module......................................7
H COMM Module ......................................................................7
Installation ......................................................................................8
Georg Fischer Signet
Wiring .............................................................................................9
Signal Type: Frequency ..........................................................10
Signal Type: Digital (S3L) ........................................................10
Signal Type: 4 to 20 mA..........................................................11
Terminal Identification .............................................................11
Sensor Wiring .........................................................................12
Power Wiring ..........................................................................15
Relay and Open Collector Wiring ...........................................16
Relay and Open Collector Outputs .........................................17
Relay Functions ............................................................................21
Operation ......................................................................................22
Menu System................................................................................24
System Setup Menu ...............................................................24
Common Menus ...........................................................................26
Loop Menu..............................................................................26
Relay Menu ............................................................................27
Option Menu ...........................................................................29
Sensor-Specific Menus.................................................................30
Flow ........................................................................................30
Advanced Fluid Measurement
pH ...........................................................................................32
ORP ........................................................................................34
Conductivity/Resistivity ...........................................................35
Pressure .................................................................................37
Level/Volume ..........................................................................38
Temperature ...........................................................................40
4 to 20 mA ..............................................................................41
Salinity ....................................................................................43
Troubleshooting ............................................................................45
Appendix.......................................................................................46
Averaging................................................................................46
LOG Current Loop Output ......................................................47
Custom Measurements ..........................................................48
Calibration Procedure - pH .....................................................52
Calibration Procedure - ORP ..................................................55
Calibration Procedure - Conductivity/Resistivity .....................56
Calibration Procedure - Flow ..................................................57
Calibration Error Messages ....................................................58
USP Limits ..............................................................................59
Specifications ...............................................................................60
Ordering Information.....................................................................64
2 9900 Transmitter
Overview
New Features
Features of the 9900 include:
● Large auto-sensing backlit display, “at a glance” visibility; displays
separate lines for the units, main and secondary measurements
● “Dial-type” digital bar graph
● Intuitive and “user-friendly” interface consistent with legacy Signet
ProPoint® and ProcessPro® devices
● Field upgradable relays
● Selectable current outputs (3.6 mA or 22 mA) to indicate error
● 4 to 20 mA input (with optional 8058 i-Go Signal Converter)
● Warning LED indicator
● Custom 13-character label capabilities for the channel type
● Factory default reset capability
● Optional H COMM Module for two-way communication
● Optional PC COMM configuration tool
Base units, plug-in modules, accessories and spare parts are sold, packaged Panel Field
and shipped separately from the factory. Unpack all components carefully. Mount Mount
The following items accompany every 9900 base unit: Mounting
X
● 9900 Base Unit Bracket
● Quick-clip panel mounting bracket (Panel Mount only) Wiring
X X
● Adhesive template for panel cutout (Panel Mount only) Connectors
● Wiring connectors for sensor and power inputs Adhesive
X
● Hardcopy Instruction Manual – English Template
● CD-ROM containing instruction manuals in English, German, French, Instruction
X X
Spanish, Chinese and Korean. Manual
CD-ROM X X
Compatibility
The 9900 is compatible with all GF Signet Flow, pH/ORP, Conductivity/ Flow
Resistivity, Salinity, Level, Temperature, Pressure and Turbidity products 515*, 525*, 2000, 2100, 2507,
2536*, 2537, 2540*, 2551, 2552
listed in the column to the right.
pH/ORP
● pH and ORP electrodes require the Signet 2750 DryLoc® Sensor 2724-2726 with 2750
Electronics (sold separately). 2756-WTx–2757-WTx with 3719
2764-2767 with 2750
● Conductivity/Resistivity or Salinity measurement requires either 2774-2777 with 2750
the optional Direct Conductivity/Resistivity Module (part number
3-9900.394) or the Signet 2850 Conductivity/Resistivity Sensor Conductivity/Resistivity
Electronics (sold separately). (NOTE: If using the 2850, use ONLY Salinity
2819-2823 with 2850 or Cond/Res
the one-channel digital (S3L) models, or the 2-channel with only one Module
channel connected. The 2-channel and 4 to 20 mA models 3-2850-52, 2839-2842 with 2850 or Cond/Res
-53, -62 and -63 are incompatible with the 9900.) Module
● Turbidity measurement using Signet 4150 requires Signet 8058 i-Go™
Level, Temperature, Pressure
Signal Converter (sold separately). 2250*, 2350*, 2450*
NOTE: Loop Power can be used ONLY for the following sensors: 515, 525,
Turbidity
2250, 2350, 2450, 2536, 2540 and 8058; all other measurement sensors 4150 with 8058
require DC power.
* Can be run on Loop Power
9900 Transmitter 3
Dimensions
Dimensions
99.06 mm
Panel Mount
(3.90 in.)
3-9900-1P
99.06 mm
(3.90 in.)
29.97 mm
(1.18 in.)
54.10 mm
91.44 mm
(2.13 in.)
(3.60 in.)
8.13 mm
Top View (0.32 in.) Side View
Field Mount
3-9900-1
44.45 mm
(1.75 in.)
106 mm 0.75
3.50
(4.16 in.) 25° Ref.
0.08
24.13 mm 0.19
(0.95 in.) 3.50
88.90 mm 3-9900.396 Angle Adjustment Adapter Kit
(3.50 in.)
(3-8051 shown)
NOTE: 3-8051 Integral Mount Kit sold separately.
4 9900 Transmitter
Plug-In Modules
Module Installation
If the 9900 Base Unit will be mounted in a panel, the plug-in modules may
be installed either before or after the base unit is mounted. If the 9900 Base
Unit will be mounted using the accessory wall mount bracket, install plug-in
modules first.
If installing both the Conductivity/Resistivity (Cond/Res) and the H COMM
Modules, install the H COMM Module first, then the Cond/Res Module on top
of it (see illustration on page 7).
The Relay and Cond/Res Modules attach with screws. The H COMM Module H COMM
simply plugs in. Base Unit
Module
er
DC Pow
To install modules:
5A 30 VDC
5A VAC
Rating:
Remove power from the 9900. Carefully align pins and connectors (do not
~
H COM
bend connecting pins) and push module firmly into place, then attach with 94
3-9900.3 ule
es Mod
To remove modules:
Remove power from the 9900. For Relay and Cond/Res Modules, unplug
connectors, remove screw(s), and carefully pull module straight out from the
base unit. Do not bend the connecting pins. For H COMM Module, squeeze
the tabs on the bottom edge, grasp the module using pliers and pull straight
out. Do not bend the connecting pins.
CAUTION
Avoid Electrostatic Discharge (ESD)
9900 Transmitter 5
Plug-In Modules
Plug-In Modules
Optional modules and accessories are available for the 9900:
● Relay Module (not available on field mount)
● Direct Conductivity/Resistivity Module
● H COMM Module
Each item is ordered separately.
DC Po
wer Modules are field-replaceable at any time.
ltage
Loop Vo
a.
See Installation and Ordering Information sections for more details.
a. Base Unit (required)
b. Slot for optional Cond/Res Module
c. Slot for optional H COMM Module
95
3-9900.3 dule
Mo
H COMM
Relay Module
wer
DC Po
(Panel Mount installations only)
ltage
Loop Vo
6 9900 Transmitter
Plug-In Modules
SHIELD
TEMP (WHT)
odule
inputs.)
H CO
H COMM Module
The H COMM Module enables communication between the 9900 and Loop
Volta
ge
HART® is a registered trademark of the HART Communication Foundation, Austin, Texas, USA.
Any use of the term HART hereafter in this document implies the registered trademark.
9900 Transmitter 7
Installation
Installation
System Start-up: Step 1
Prepare the transmitter installation location. If the back of the transmitter is
difficult to access when installed, wire the removable terminal blocks first,
then install it completely.
Loop Volta
ge mounting
Panel Mount Installation
bracket Tools and Equipment Required
3-9900.395 le
H COMM
Modu
● File (fine)
● ¼ DIN punch or jigsaw suitable for cutting panel opening to within
1 mm (0.04 in) tolerance.
● ¼ DIN punches are available and recommended for creating clean,
Quick-clip precise openings quickly and easily in most instrument panels.
● If a punch is not available, a jigsaw or other cutting tool can be used. An
adhesive template is provided to help guide the cutting process.
De-burr and smooth the opening with a file.
8 9900 Transmitter
Wiring
Wiring
System Start-up: Step 2
Wire the transmitter for all connections with the power off. Keep any
4 to 20 mA and relay-actuated output devices that are connected to it
offline at this time. Connect the sensors (pg. 12), power (pg. 15) and
relay(s) (pg. 16).
Wiring Tips:
Tools Required
● Do not route the sensor, DC power, or 4 to 20 mA cables in conduit
containing AC power wiring. Electrical noise may interfere with sensor ● Phillips screwdriver
● Flat-head screwdriver
signal. ● Wire strippers
● Routing the sensor cable in grounded metal conduit can help prevent
electrical noise and mechanical damage. For Field Mount
● Seal the cable entry points to prevent moisture damage. installations, refer to the
wiring diagram inside the
● Only one wire should be inserted into a terminal.
Field Mount housing.
● Splice double wires outside the terminal or use appropriate wire ferrule,
not to exceed 2 mm (0.08 in) diameter.
All wiring connections to the 9900 are made via removable terminals.
In general:
● Terminals accept 12 to 24 AWG wire (Cond/Res Module plug accepts
16 to 28 AWG wire).
● Strip 7 mm (0.28 in.) of insulation from wire tips and tin bare ends to
eliminate fraying.
● Insert wire tip or ferrule completely into the terminal and secure with the
screw.
● Do not allow any AC leads that may be connected to the internal relays
to come in contact with low voltage wiring.
Electrical noise may interfere with sensor signals:
● Do not route the sensor cable in conduit containing AC power wiring.
● Route the sensor cable in grounded metal conduit to help prevent
electrical noise and mechanical damage.
● Seal the cable entry points to prevent moisture damage.
CAUTION:
Avoid Electrostatic Discharge (ESD)
9900 Transmitter 9
Wiring
Maximum
Signet flow sensors 515, 525, 2000, 2100, 2507, 2536 and 2540 provide
Flow sensor
models with Cable Length a frequency output. (Flow sensors 2551 and 2552 can be configured with
Frequency 60 m 305 m either digital (S3L) or Frequency outputs, see pg. 13.)
Output (200 ft) (1000 ft)
515 X The maximum allowable cable length for sensors with frequency output is
525 X dependent upon the output signal strength of the sensors themselves, and
the degree to which the signals are susceptible to EMI or "noise." This is
2000 X
largely a function of whether the sensors are self-powered (515 and 525), or
2100 X
powered by an external source.
2507 X
2536 X ● The Input terminals on the 9900 carry Frequency data signals from
2537 X the sensor.
2540 X ● Do not route sensor or output cables in conduit containing AC power
2551 X wiring. Electrical noise may interfere with sensor signal.
2552 X ● Routing cable in grounded metal conduit will help prevent electrical
noise and mechanical damage.
● Seal cable entry points to prevent moisture damage.
● Only one wire should be inserted into a terminal. Splice double wires
outside the terminal.
In case of noise ● In case of noise interference, ground the sensor SHIELD wire to a local
interference, connect earth ground at a point near the sensor.
the cable shield to
earth ground. ● Consult the sensor manual for additional wiring information.
10 9900 Transmitter
Terminal Identification
Signal Type: 4 to 20 mA
When connecting a non-Signet sensor to the 9900, the sensor’s 4 to 20 mA
signal must be converted to digital (S3L). The 8058 i-Go Signal Converter
accepts any 4 to 20 mA signal and converts it into digital (S3L).
1. Wire the 8058 between the 4 to 20 mA loop source and the 9900
digital (S3L) input terminals (see pg. 14). NOTE: If connecting 8058-2,
use Channel 1 ONLY.
2. In the 9900 INPUT TYPE menu (see System Setup Menu discussion),
specify 4 to 20 mA INPUT.
3. Set additional labels and abbreviations as described on pgs. 41-42.
Red )
d ite
iel Wh
+GF+ Sh e to
4-to-20 mA
Black
S3L
White
Red
4-20 mA input S3L Output
1
AX
CM
2
35 VD
Loop1 +
3
iel
d k)
Loop1 -
Bla
Sh lac ck
Loop2+ 4
N/C
oB
+G
Loop2 - 5
For the 8058-2,
F
6
3 +
4-20 mA
et
8058-1
Input
L Converter
3-8058.610D
2
8058 i-Go™
1
N/C
( Ti
6
5
GND
4
+5VDC
S3 L
connect Channel 1 ONLY
Lo
9
op
3
PW 1
8
4-20mA to S
R
7
Output
Lo
S3 L
op
PW 2
R
+GF+
S3
DA L
Sign TA
et
80
58
N /C
7
8
9
8058-2
Terminal Identification
SHLD
DATA
GND
Terminals 3-4: Loop Power (may also be used for system power)
V+
12 to 24 VDC ±10%
● Autoswitches to loop power if DC power is lost.
NOTE: Backlight, LEDs and optional Relay Module do not
operate on loop power. Any connected sensors or sensor
electronics that cannot operate on loop power will also 7 8 9 10
be inoperative.
Terminals 5-6: Open Collector Power
● Software selectable for Normally Open or Normally Closed.
● May be disabled (Off) if not used. 1 PWR+
Terminals 7-10: Digital (S L)/Frequency Input 3
2 PWR–
7: V+: +5 VDC out to sensor (black wire)
3 Loop+
8: DATA: Input signal from sensor (red wire)
9: GND: Sensor ground (white wire) 4 Loop–
10: SHLD: Cable shield
5 6
OC- OC+
9900 Transmitter 11
Sensor Wiring
Sensor Wiring
er
DC Pow
NOTE: Loop Power cannot be used to power Signet models 2000, 2100,
2507, 2537, 2551 or 2552 Flow sensors.
Technical Notes:
• See corresponding product
manuals for maximum cable length. Wiring for:
• Maintain cable shield through cable
splice.
• Route sensor cable away from AC
power lines. 515 525 2536 2540
• 515 and 525 installations, connect
FLOW
Frequency
the silver (shield) wire to earth
ground in case of EMI noise
SHLD
DATA
interference.
GND
Black
Red
FLOW
Shield
12 9900 Transmitter
Sensor Wiring
SHLD
DATA
GND
4 frequency signal. When the
V+
jumper is removed (or placed
over one pin for storage) the
3 2551-XX-11 outputs a digital
(S3L) signal (recommended).
2551-XX-11
• The frequency output will
Blind Magmeter
be displayed as positive
flow regardless of the flow
direction. Signet recommends
Black configuring the 2551 with the
Red S3 L digital (S3L) output because it
is more accurate and will also
White display reverse flow (negative
4
Shield numbers).
• 5 VDC power is provided to the
3 2551 Magmeter by the 9900. No
additional power is required.
• Connect the silver (shield) wire to
earth ground in case of EMI noise
interference.
Wiring for:
Technical Notes:
2552 • The 2552 outputs an open
collector frequency signal
S3 L Frequency that can be connected to the
9900. Signet recommends
configuring the 2552 with the
SHLD
SHLD
DATA
DATA
GND
GND
9900 Transmitter 13
Sensor Wiring
Wiring for:
Technical Notes:
• Use 3-conductor shielded cable
for sensor cable splices up to
2250 2350 2450
300 m (1000 ft) max. S3L
• Maintain cable shield through
SHLD
DATA
cable splice.
GND
V+
• Route sensor cable away from
AC power lines.
• Connect the silver (shield) wire
to earth ground in case of EMI
noise interference.
• To work correctly with the 9900, 2551* 2750 2850
the 2850 must be set for a 1.0
cell constant (the actual probe
cell constant is set in the 9900). Black
Red
White
Shield
* 2551-XX-21, -41
Display Magmeter
S 3L
SHLD
60 m (200 ft).
4-20 mA input S3L Output
DATA
GND
X
1
MA
V+
2
DC
Loop1 +
3
35V Loop1 -
8058-1
Loop2+ 4
N/C
+G
Loop2 - 5
3
Input
L Converter
3-8058.610D
2
8058 i-Go™
1
N/C
6
5
GND
4
+5VDC
S3 L
Loo
9
p1
3
PW
8
4-20mA to S
R
7
Output
Loo
S3 L
p2
S 3
L
DAT
Sig A
net
805
8
N/
C
8058-2
7
8
9
close-up Red
S3L Output
White S3L
SHLD
DATA
GND
Shield
V+
Black
7 +5VDC
8058-2 Red
Technical Notes: close-up 8 S3 L
14 9900 Transmitter
Power Wiring
Power Wiring
CAUTION!
ProcessPro
(for reference only) 9900
Transmitter 9900
Terminals Terminals
Black Red
System Pwr Loop - 4 – Power PWR+ + Power
+ Supply – Supply
System Pwr Loop + 3 PWR–
Red 12 to 24 VDC Loop+ Black 12 to 32 VDC
AUX Power - 2 Loop–
AUX Power + 1
ProcessPro
(for reference only) 9900
Transmitter 9900
Terminals Terminals
PLC or Recorder
Blue
+ Loop Input Red Power
– 4 to 20 mA PWR+ +
PWR– – Supply
System Pwr Loop - 44 12 to 32 VDC
Loop+ Black
System Pwr Loop + 33 Loop–
Blue
AUX Power - 22 Black Power PLC or Recorder
–
Supply + Loop Input
AUX Power + 11 Red + 12 to 24 VDC – 4 to 20 mA
ProcessPro
Loop Powered 9900
(for reference only)
Transmitter 9900
Terminals Terminals
–
Power PWR+ Red
Supply PWR– – 32 VDC
System Pwr Loop - 4 Black Loop+
+ 12 to 24 VDC + ± 10%
Loop–
System Pwr Loop + 3 Black
AUX Power - 2
PLC or Recorder PLC or Recorder
AUX Power + 1 – Loop Input
Red – Loop Input
+ 4 to 20 mA + 4 to 20 mA
9900 Transmitter 15
Relay Wiring
OC – OC + 9WR9
Input OC + Power
PLC Supply
Gnd Gnd
If PLC needs 0 logic input when relay is Pull-Down Resistor
not energized, set NORMAL to CLOSED ..ȍ:
in the RELAY menu when using the Open
Collector (R1) with NPN style wiring,
Alarm
PWR +
PWR –
Loop + V+
Loop – Power with NORMAL
Supply set to OPEN.
OC – OC + Gnd
The 9900 Open Collector (R1) output provides high-speed switching capability. Signal frequencies can reach
400 pulses per minute.
The 9900 Open Collector (R1) output connection is dependent upon the type of circuit being controlled by
the output.
Most indicating instruments or control system inputs require a signal voltage of 0 to 5 V (TTL or CMOS logic
levels) or 0 to 24 V. Therefore, the 9900 Open Collector output circuits must be equipped with a pull-up or pull-
down resistor (not supplied), and a quality regulated 5 to 24 V (depending on the application) power supply (not
supplied) is recommended to function properly.
NO
operation, and will go ON when NC
relay energizes according to 9900 C
Relay settings. AC or DC
power
ALARM!
The valve is ON during normal
RELAY 3
Relay settings.
AC or DC
power
NO = normally open (closes when energized)
NC = normally closed (opens when energized) Valve
Flow
16 9900 Transmitter
Relay Outputs
If the 9900 is set to operate in RELAY LOW mode, when the user-defined condition for the activation is
met (e.g. exceeding an alarm limit) the Open Collector switch is turned “on.” If wired as standard “NPN-
style” output (see previous page) the logic level of the attached control system or PLC input consequently
becomes “low” logic level (when NORMAL is set to OPEN).
If a high input logic level is required for activation, it can be accomplished in one of three ways.
In order of preference,
1. Change the Open Collector (Relay 1) output function to “high” in the instrument's RELAY menu, or
2. Wire the Open Collector (R1) output “PNP” style as described on the previous page, or
3. Set the Open Collector (R1) to NORMAL CLOSED in the RELAY menu.
Fail-Safe Behavior
No matter the setting, the Open Collector output turns off if the 9900 loses power. This must be taken into
account when evaluating system failure consequences. If the system layout requires a “closed” or “on”
condition for the output in case of power loss, a mechanical dry-contact relay (NC contacts) must be used
instead of the Open Collector (R1) output.
The current through the Open Collector switch also must be limited. Typical Open Collector outputs allow
only for 10 to 50 mA switch current. Exceeding this current limit can burn out the Open Collector output
components immediately.
9900 Transmitter 17
Relay Outputs
The 9900 open collector and relays are selectable Open Collector Output
and configurable and can be used as switches that ● Longer life than a mechanical relay
respond when the process value moves above or ● No moving parts
below a user-defined setpoint or it can be used ● Faster ON/OFF switching capabilities than
to generate a pulse at a rate proportional to the mechanical relays
process value. They can be used for Low Alarm, ● Can switch DC voltage only (< 30 VDC)
High Alarm or Proportional Pulse triggering related ● Not recommended for use with inductive loads.
to the process value. All relay functions are set up
in the RELAY menus.
Relay energized
Relay de-energized
Process
Low Setpoint:
Hysteresis Relay is on when the measured value is less than
Low Setpoint
the setpoint.
Time
Process
High Setpoint
High Setpoint:
Hysteresis
Relay is on when the measured value is higher than
the setpoint.
Time
18 9900 Transmitter
Relay Outputs
Window In/Out:
Relay is on when the value is higher or lower than Process
:
in. ate
100 pulses
s/m ut R
100 pulses
by the settings in the CAL menu and the sensor
lse utp
input. The maximum pulse output from the relays is
Pu O
0 or
400 pulses per minute. Example usage would be to
10 ct
Pulse rate
to lle
0 Co
control solenoid-operated dosing pumps.
n
pe
O
For example: As the process value drops below
the setpoint, the output will start pulsing in relation
0 pulses
to the process value, the maximum pulse endpoint
and the programmed pulses/minute. The pulse 5 10
rate will change as the process value changes Starting point Endpoint
and approaches the programmed endpoint. This
functionality can be used to precisely control
the process.
The starting point, endpoint and maximum pulse rate In the example:
are selectable in the RELAY menus. • The output will be 0 pulses/min. when value is
less than 5.
• The output will be 50 pulses/min. when value
is 7.5.
• The output will be 100 pulses/min. when value
is greater than 10.
9900 Transmitter 19
Relay Outputs
PWM
Maximum • Pulse Width Modulation
Range PWM automatically varies the ratio of ON time to
100% OFF time proportional to minimum and maximum
Relay is range settings.
(as a % of Relay period)
80% always ON
Relay Pulse Width
• Volumetric Pulse
A pulse is generated each time a specified volume of
fluid is registered. For flow inputs only.
• Totalizer Volume
Relay activates and latches when a specified volume
of fluid is registered. For Flow inputs only.
Total Volume mode counts the TOTALIZER Units until
the setpoint volume is reached, then turns on the relay
until the resettable totalizer is reset.
If the Resettable Totalizer reading is greater than the
setpoint, the relay will be turned on immediately. The
relay will be off when the totalizer is reset to zero.
This mode is useful to trigger a reminder when a
process is due, as for a backwash cycle or filter
change.
20 9900 Transmitter
Relay Outputs
Relay Functions
System Start-up: Step 3
Set your relay functions to your own application
requirements.
Next step: System Setup (see pg. 24).
9900 Transmitter 21
Operation
Operation
Open Collector (R1) Indicator LED Backlight Sensor
(do not block)
Mechanical Relay (R2) Mechanical Relay (R3) Indicator LED
Indicator LED
Error Indicator
1 2 3 !
Bar Graph
Units of Measure
(GPM, pH, sec, %, etc.)
Value
Label
Menu Indication
ENTER
All possible segments shown in this illustration. The instrument's software controls
which segments are shown at any particular time. Only the bar graph and +GF+ logo
are visible when the unit is turned off.
RIGHT key
Select item or character to edit
22 9900 Transmitter
Operation
Keypad Functions
The four buttons of the keypad are used to navigate display modes
according to the descriptions in this table. Notice that the function
3s of each button may change depending on the display mode.
ENTER
System Setup: Menu Navigation
ENTER
or
Saves Changes
+
+ Or Select another
2x Menu Item
Return to
View Mode
9900 Transmitter 23
Operation
Menu System
24 9900 Transmitter
Menu System
MENU Mode is entered by pressing and holding ENTER for three seconds.
The ► button is used to change the position of the blinking cursor. When the
desired menu is blinking, press ENTER.
In the selected menu, use the ▲ and ▼ keys to navigate through the menu.
Use the ▲, ▼ and ► keys to edit the selected item (see Menu System
discussion, pg. 24).
To save the new selection, press the ENTER key. A message displaying
“Saving…” will be displayed for 3 seconds. After this message is displayed,
the newly selected value will be displayed, if applicable.
Password Overview
The password is often required to start editing. Once entered correctly, this
password will not be needed for subsequent edits. However, once the menu
system is exited, the password will be again be required when edit mode is In the MENU mode, if the wrong
code or password is entered, an
re-entered.
ERROR message is displayed.
● STD
The standard (STD) password is ▲▲▲▼, pressed in sequence. This
password is designed to protect the 9900 from unintentional changes.
It is best suited for systems where a group of people need to be able to
change settings.
● CODE To change the password, go to
The CODE default setting is 0000, adjustable to any 4-digit numerical OPTIONS mode, enter your desired
code up to 9999. Using a personal code provides the maximum degree code and press ENTER. (The STD
of security. This code can be modified in the Options mode. password cannot be changed.)
9900 Transmitter 25
Menu System
Common Menus
Common Menus
The menu system shares certain modes between sensor types. The following describes the EDIT Mode
menus found in common between most sensor types.
LOOP Menu
(ALL) Set value corresponding to desired 4 mA output. 5 digits max. Default = 0 (ORP = -999).
(ALL) Set desired Loop output value when sensor error (e.g., bad sensor, broken wire) is detected.
Select (3.6 mA, 22 mA). Default = 22.
(ALL) Allows fine-tuning to compensate for errors in other equipment connected to the 9900. Adjust
the minimum and maximum current output. The display value represents the precise current output.
Adjustment limits: from 3.80 mA minimum to 5.00 mA maximum. Default = 4.00 mA.
(ALL) Allows fine-tuning to compensate for errors in other equipment connected to the 9900. Adjust
the minimum and maximum current output. The display value represents the precise current output.
Adjustment limits: from 19.00 mA minimum to 21.00 mA maximum. Default = 20 mA.
(ALL) Press ▲ or ▼ to manually order any output current value from 3.6 mA to 21.00 mA to test the
output loop.
26 9900 Transmitter
Menu System
RELAY Menu
(pH, LEVEL/VOL, COND/RES and SALINITY only.) Select source for each of R1, R2 and R3 outputs.
Choose pH/TEMP, LEVEL/VOLUME, COND/RES or SAL/TEMP. Defaults = pH, COND, LEVEL, SAL.
(ALL) Set Open Collector (R1) as Normally Open or Normally Closed. Default = OPEN.
(ALL) Select the desired mode of operation for the open-collector (R1) output (OFF, LOW, HIGH, WINDow
IN, WINDow OUT, CYC LOW (except FLOW), CYC HIGH, PROP PuLSe, VOL PuLSe, PWM, TOTAL,
USP, ERROR mode (See chart on pg. 28). Default = OFF. Continue stepping through to select R2 and R3
output modes. When MODE is set to ERROR, delays energizing relay until after ON DELAY time expires if
sensor problem is detected. See Cycle High/Low discussion on pg. 18.
(ALL) Relay turns on if process measurement goes lower than this value. Set desired value.
(Shown if LOW, WIND IN/OUT or CYC LOW mode.)
NOTE: The corresponding indicator lights do not light up in PROP PLS and PWM modes. The LEDs light
up only when the Test Relay options are selected.
(ALL) Relay turns on if process measurement goes higher than this value. Set desired value.
(Shown if HIGH or WIND IN/OUT mode.)
NOTE: The corresponding indicator lights do not light up in PROP PLS and PWM modes. The LEDs light
only when the Test Relay options are selected.
(FLOW only) Amount of accumulated flow that must be counted before a pulse is sent out. Relay
turns on if flow volume exceeds this value. Set desired value. (Shown if CYC HIGH or VOL PLS mode.)
Default = 100.00.
(ALL) Hysteresis prevents the system from chattering around the set point. Set amount (in units of
measure from INPUT Mode) to add to SET LOW or SET HIGH values.
(Shown if LOW, HIGH, WIND IN/OUT, CYC LOW/HIGH or USP mode)
(COND/RES only) Relay turns on if USP value drifts by this value away from USP limit.
(Shown only in USP mode) See USP Limits discussion in the Appendix.
(ALL except PRESSURE) Set minimum setpoint value for proportional pulsing.
(Shown if PROP PLS mode.)
(ALL except PRESSURE) Set maximum setpoint value for proportional pulsing.
(Shown if PROP PLS mode.)
(ALL except PRESSURE) Set desired maximum pulse rate (400 max)
(Shown if PROP PLS mode.)
NOTE: Pulse width fixed at 100 ms.
(ALL except PRESSURE and FLOW) Set minimum value for pulse width modulation.
(Shown if PWM mode)
(ALL except PRESSURE and FLOW) Set maximum value for pulse width modulation.
(Shown if PWM mode.)
NOTE: Defaults for most relay functions are dependent upon sensor type and are not listed here.
9900 Transmitter 27
Menu System
(ALL) Set time in seconds (up to 99999) for relay to remain on. See discussion on pg. 18.
(Shown if CYC LOW/HIGH mode.)
(FLOW only) Amount of accumulated flow that must be counted before a pulse is sent out.
Set value. (Shown only if VOL PULS.)
(FLOW only) Set time value for one pulse width. (Shown only if VOL PULS.)
(ALL except PRESSURE and FLOW) Set time value for one complete pulse cycle (relay ON time + relay
OFF time). (Shown if PWM mode)
(FLOW only) Resettable value that, when exceeded, turns relay on. Must reset Totalizer
(in VIEW Mode) to clear relay. Set maximum value. (Shown only if TOTAL.)
28 9900 Transmitter
Menu System
OPTION Menu
Adjust the LCD contrast for best viewing. A setting of 1 is lowest contrast, 5 is highest. In general,
select lower contrast if the display is in warmer ambient light surroundings.
Default = 3.
(ALL) Set the decimal to the best resolution for your application. The display will automatically scale up
to this resolution. Select -----., ----.-, ---.-- or --.---.
Default = ----.-.
(FLOW only) Set the decimal to the best resolution for the Permanent Totalizer display. The display will
automatically scale up to this resolution. Select -----., ----.-, ---.-- or --.---.
Default = ----.-.
(FLOW only) Locks the TOTALIZER output. Select OFF, ON (Does not affect Permanent Totalizer).
Default = OFF.
(ALL) Enter desired password code. 4-character entry not displayed, ---- displayed instead.
(Shown if type = CODE.)
Enables Remote Setup to configure the 9900 via a computer and the PC COMM tool.
Press ► and select YES to enable. REMOTE SETUP flashes when mode is enabled.
NOTE: Communication with PC COMM tool is automatic when 9900 is in FACTORY state
(EntEr flashing).
Refer to the PC COMM Configuration/Diagnostic Tool manual, 3-0251.090, included with your
PC COMM tool.
9900 Transmitter 29
Flow
Sensor-Specific Menus
The following pages list the sensor-specific settings for each sensor type.
Flow
FLOW Setup Checklist
1. Make sure FLOW sensor type is selected (see System Setup
Menu, pg. 24).
2. Set the Units of Measure.
3. Set Sensor Type (Freq or S3L).
4. If Loop is used, set the minimum and maximum 4 to 20 mA
setpoints.
5. Set K-Factor (pulses per Unit Volume) from Flow Sensor manual.
6. Set Totalizer factor.
7. Set Last Cal Date and initials.
This is the normal display 8. If desired, set up relay functions for your own application.
and does not time out.
CAL Menu
Prevents relays from activating while making adjustments. Relays in Pulse mode will suspend pulsing
while HOLD OUTPUTS is set to YES. Select YES/NO.
Default = NO.
Set K-Factor (pulses per unit volume) from Flow Sensor manual. Min: 0.0001, max 999999.
Cannot be zero.
Default = 60.0000.
Sets the volume of each count of the Totalizer as a multiple of the volume unit of the K-Factor.
Min: 0.0001, max 999999. Cannot be zero.
Default = 1.0000.
Select to calibrate using Rate method (see Appendix).
30 9900 Transmitter
Flow
INPUT Menu
If your flow sensor is configured for frequency output, select FREQ. If configured for digital (S3L) output
(recommended), select S3L.
Default = FREQ.
Dampens display, output and relay response rates. Select Low (4 sec), Med (8 sec), High (32 sec), OFF
(near instantaneous). (See discussion in Appendix.)
Default = OFF.
Acts as a threshold for flow measurement response. A lower sensitivity setting gives a fast measurement
response, a higher setting gives a slower response. Value expressed in units of measurement; response
dependent on units of measurement being exceeded. (See discussion in Appendix.)
Default = 100.
9900 Transmitter 31
pH
pH
pH Setup Checklist
1. Make sure pH sensor type is selected (see System Setup Menu,
pg. 24).
2. Set the Temperature Units (°C or °F).
3. If Loop is used, set the minimum and maximum 4 to 20 mA
setpoints.
4. Perform calibration (EasyCal, Standard or Standard and Slope).
5. Set Last Cal Date and initials.
6. Select source for Open Collector and Relay output (pH or Temp).
7. If desired, set up relay functions for your own application.
This is the normal display
and does not time out.
INPUT Menu
Select °F or °C.
Default = °C.
Dampens display, output and relay response rates. Select Low (4 sec), Med (8 sec), High (32 sec), OFF
(near instantaneous) (see discussion in Appendix).
Default = OFF.
NOTE: Signet strongly recommends leaving averaging OFF for pH and Pressure measurements (see
discussion in Appendix).
32 9900 Transmitter
pH
CAL Menu
Select AT SENSOR to perform calibration using the Signet 2750 sensor electronics. Select AT
INSTRUMENT to perform calibration at the 9900 via EasyCal or manual calibration. (See pH Calibration
procedures in the Appendix.)
Default = AT INSTRUMENT.
Prevents relays from activating while making adjustments. Relays in Pulse mode will suspend pulsing
while HOLD OUTPUTS is set to YES.
Default = NO.
Applies a linear offset to the pH measurement. The ideal value is the average pH of your application. (A
sample of your application at process temperature is recommended.) (See pH Calibration procedures in
the Appendix.) Shows error message if offset too high.
Applies a slope to the pH measurement. The slope value and the standard value must be at least 2 pH
units apart. The ideal values are the minimum and maximum values of your process. (See pH Calibration
procedures in the Appendix.) Shows error message if slope is too low or high.
Applies a linear offset to the temperature measurement. The ideal value is the average temperature of
your application. "SAVING" will appear if offset is acceptable, "ERR TOO LARGE TO CALIBRATE" if
offset is outside of range.
9900 Transmitter 33
ORP
ORP
ORP Setup Checklist
1. Make sure ORP sensor type is selected (see System Setup Menu,
pg. 24).
2. If Loop is used, set the minimum and maximum 4 to 20 mA
setpoints.
3. Set Averaging.
4. Perform calibration or set Standard (and Slope if desired).
5. Set Last Cal Date and initials.
6. If desired, set up relay functions for your own application.
CAL Menu
Select AT SENSOR to perform calibration using the Signet 2750 sensor electronics. Select AT
INSTRUMENT to perform calibration at the 9900 via EasyCal or manual calibration. (See ORP
Calibration procedures in the Appendix.). Default = AT INSTRUMENT.
Prevents relays from activating while making adjustments. Relays in Pulse mode will suspend pulsing
while HOLD OUTPUTS is set to YES. Default = NO.
Applies a linear offset to the ORP measurement. For single point calibrations, assign the average value
of your process to ORP STANDARD. For two-point calibrations, assign the min or max value of your
process to ORP STANDARD. (See ORP Calibration procedures in the Appendix.).
Applies a slope to the ORP measurement. The ORP SLOPE is used for two-point calibration along with
the ORP STANDARD. If you applied the min value of your process to the ORP STANDARD, then apply
the max value to the ORP SLOPE. Else, apply the min value to the ORP SLOPE. The slope value and
the standard value must be at least 30 mV apart. (See pH Calibration procedures in the Appendix.)
Resets calibration to factory settings. After pressing ►, select YES/NO. (Shown if CAL AT INSTR)
INPUT Menu
Dampens display, output and relay response rates. Select Low (4 sec), Med (8 sec), High (32 sec),
OFF (near instantaneous). Default = OFF. (See discussion in Appendix.)
34 9900 Transmitter
Conductivity/Resistivity
Conductivity/Resistivity
Cond/Res Setup Checklist
1. Make sure COND/RES sensor type is selected (see System
Setup Menu, pg. 24).
2. Set Cell Constant.
3. Set the Temperature Units (°C or °F).
4. Set Conductivity units.
5. If Loop is used, set the minimum and maximum 4 to 20 mA
setpoints.
6. Set Temperature Compensation.
7. Set Last Cal Date and initials.
This is the normal display 8. Select source for Open Collector and Relay output (primary
and does not time out. measurement or temperature).
9. If desired, setup relay functions for your own application.
CAL Menu
Freezes measurement value; loop value continues normally. Relays in Pulse mode will suspend pulsing
while HOLD OUTPUTS is set to YES. Select (No/Yes).
Default = NO.
Shows real-time value and selected standard. “PLACE SENSOR IN STANDARD”. Unit waits until
reading is stable; if bad cal, returns “ERROR, CANNOT DETERMINE STANDARD”. Refer to buffer
values and AUTO CAL Procedure in the Appendix.
Shows “CONDUCTIVITY” on bottom line; when user presses any button the live value is frozen and the
user edits that value. If bad cal, returns “ERR TOO LARGE TO CALIBRATE”. See Manual Cal procedure
in Appendix.
Shows “TEMPERATURE” on bottom line; when user presses any button the live value is frozen and the
user edits that value. If bad cal, returns “ERR TOO LARGE TO CALIBRATE”.
9900 Transmitter 35
Conductivity/Resistivity
INPUT Menu
Enter cell constant of sensor. Select 20.0, 10.0, 1.0, 0.1, 0.01, or CUSTOM.
Default = 1.0.
Enter the precise cell constant from the certificate provided with your sensor, or from the information
label on the sensor. Shown if CELL CONSTANT = CUSTOM.
If the COND UNITS selection is PPM or PPB, set the ratio of Total Dissolved Solids to μS.
Default = 0.50.
Dampens display, output and relay response rates. Select Low (4 sec), Med (8 sec), High (32 sec), or
OFF (near instantaneous). (See discussion in Appendix.)
Default = “OFF”.
For LINEAR or PURE H2O temperature compensation, select a % per ºC slope. Maximum slope setting
is 9.99 % per ºC. Default = 2.0.
(If Temperature Compensation setting is NONE, this item will not be displayed.)
Factory-Set Span:
0.01 cell (2819, 2839) ......... 0 to 100 μS
0.10 cell (2820, 2840) ......... 0 to 1000 μS
1.0 cell (2821, 2841) ........... 0 to 10,000 μS
10.0 cell (2822, 2842) ......... 0 to 200,000 μS
20.0 cell (2823) ................... 0 to 400,000 μS
36 9900 Transmitter
Pressure
Pressure
PRESSURE Setup Checklist
1. Make sure PRESSURE sensor type is selected (see System
Setup Menu, pg. 24).
2. If Loop is used, set the minimum and maximum 4 to 20 mA
setpoints.
3. Set Units of Measurement (PSI, BAR, KPa).
4. Set Last Cal Date and initials.
5. If desired, set up relay functions for your own application.
CAL Menu
Prevents relays from activating while making adjustments. Relays in Pulse mode will suspend pulsing
while HOLD OUTPUTS is set to YES.
INPUT Menu
Dampens display, output and relay response rates. Select Low (4 sec), Med (8 sec), High (32 sec), OFF
(near instantaneous). (See discussion in Appendix.) Signet strongly recommends leaving averaging OFF
for pH and pressure measurements (see discussion in Appendix).
Default = OFF.
9900 Transmitter 37
Level/Volume
Level/Volume
LEVEL/VOLUME Setup Checklist
1. Make sure LEVEL/VOLUME sensor type is selected (see System
Setup Menu, pg. 24).
2. Select Main Measurement (Level or Volume).
3. Set Units of Measurement for LEVEL display (FT, IN, M, CM).
4. If desired, set Units of Measurement for VOLUME display.
5. Set the minimum and maximum 4 to 20 mA setpoints.
6. Set Specific Gravity.
7. Set Sensor Offset.
8. If VOLUME is used, set Shape.
This is the normal display 9. Set Last Cal Date and initials.
and does not time out. 10. If desired, set up relay functions for your own application.
Displays the Volume value on the bottom line of the screen when LVL is the MAIN MEAS selection in
INPUT menu.
Displays the Level value on the bottom line of the screen when VOL is the MAIN MEAS selection in
INPUT menu.
Bottom line shows one of three states (OFF, ON, PLS) for each of the three relays. Displays remaining
time for CYC LOW or CYC HIGH mode. The relay(s) will remain ON while counting down.
NOTE: To reset the timer: In the RELAY menu, select TEST RELAY function. The timer will reset to 0 if
the condition no longer exists when the TEST is performed.
The timer will restart if the condition still exists.
CAL Menu
Prevents relays from activating while making adjustments. Relays in Pulse mode will suspend pulsing
while HOLD OUTPUTS is set to YES.
Shows SET LEVEL on bottom line. When user presses any key, the live value is frozen and the user
edits that value. Returns either GOOD CAL or LEVEL OFFSET TOO LARGE.
Resets calibration to factory default. After pressing ►, select YES/NO.
38 9900 Transmitter
Level/Volume
INPUT Menu
Select ON = Measurement will be displayed as a percentage of full scale. OFF = Measurement will be
displayed in unit of measure selected in previous setting.
Default = OFF.
If PERCENT LEVEL = ON, set the desired full scale (100%) value in units of measure.
Default = 10.00.
Select unit of measure for VOLUME display (GAL, LIT, Lb, KG, FT3,, in3, M3, cm3).
Default = GAL.
Select ON = Measurement will be displayed as a percentage of full scale. OFF = Measurement will be
displayed in unit of measure selected in previous setting.
Default = OFF.
If PERCENT VOLUME = ON, set the full scale value (100%) in units of measure.
Default = 100.00.
Enter the specific gravity of the fluid at normal operating temperature. This setting is required only if the
level measurement is made by a pressure sensor or if kg or lb volume units are selected.
Default = 1.0000 (water).
Enter the distance from sensor location to the Zero reference point in the vessel (see discussion in
Appendix). Displayed in units of measure chosen in LEVEL UNITS.
Default = 0.
Dampens display, output and relay response rates. Select Low (4 sec), Med (8 sec), High (32 sec),
OFF (near instantaneous). (See discussion in Appendix.)
Default = OFF.
Select the shape of the vessel where the level sensor is located. VERT CYLINDER, HORIZ
CYLINDER, RECTANGLE, or CUSTOM. (To define a custom tank shape, see Appendix, Defining a
Custom Tank.)
Default = VERT CYLINDER.
If VERT CYLINDER or HORIZ CYLINDER is selected, enter the diameter of the cylinder. Displayed in
units of measure chosen in LEVEL UNITS.
Default = 2.0000.
9900 Transmitter 39
Temperature
Temperature
TEMPERATURE Setup Checklist
1. Make sure TEMPERATURE sensor type is selected (see System
Setup Menu, pg. 24).
2. If Loop is used, set the minimum and maximum 4 to 20 mA
setpoints.
3. Set Units of Measurement (°C or °F).
4. Set Last Cal Date and initials.
5. If desired, set up relay functions for your own application.
CAL Menu
Prevents relays from activating while making adjustments. Relays in Pulse mode will suspend pulsing
while HOLD OUTPUTS is set to YES. The temperature reading does not update while the 'HOLD
OUTPUT' setting is set to 'ON'. The display will read "OUTPUTS HELD" and the temperature reading
that is displayed is the last one that was shown prior to the 'HOLD OUTPUT' being set to 'ON'.
INPUT Menu
Select °C or °F.
Default = °C.
Dampens display, output and relay response rates. Select Low (4 sec), Med (8 sec), High (32 sec),
OFF (near instantaneous). (See discussion in Appendix.)
Default = “OFF”.
40 9900 Transmitter
4 to 20 mA
4 to 20 mA
4 to 20 mA Setup Checklist
1. Make sure 4-20 mA INPUT sensor type is selected (see
System Setup Menu, pg. 24).
2. Set 4 mA value (refer to your 3rd-party sensor manual).
3. Set 20 mA value (refer to your 3rd-party sensor manual).
4. If Loop is used, set the minimum and maximum 4 to 20 mA
setpoints.
5. Set Last Cal Date and initials.
6. If desired, set up relay functions for your own application.
Bottom line shows one of three states (OFF, ON, PLS) for each of the three relays. Displays remaining
time for CYC LOW or CYC HIGH mode. The relay(s) will remain ON while counting down.
NOTE: To reset the timer: In the RELAY menu, select TEST RELAY function. The timer will reset to 0 if
the condition no longer exists when the TEST is performed.
The timer will restart if the condition still exists.
CAL Menu
Prevents relays from activating while making adjustments. Relays in Pulse mode will suspend pulsing
while HOLD OUTPUTS is set to YES.
9900 Transmitter 41
4 to 20 mA
INPUT Menu
Dampens display, output and relay response rates. Select Low (4 sec), Med (8 sec), High (32 sec),
or OFF (near instantaneous). (See discussion in Appendix.)
Default = “OFF”.
42 9900 Transmitter
Salinity
Salinity
SALINITY Setup Checklist
1. Make sure SALINITY sensor type is selected (see System
Setup Menu, pg. 24).
2. Set Cell Constant.
3. Set the Temperature Units (°C or °F).
4. If Loop is used, set the minimum and maximum 4 to 20 mA
setpoints.
5. Set Last Cal Date and initials.
6. Select source for Open Collector and Relay output (primary
measurement or temperature).
This is the normal display 7. If desired, set up relay functions for your own application.
and does not time out.
CAL Menu
Prevents relays from activating while making adjustments. Relays in Pulse mode will suspend pulsing
while HOLD OUTPUTS is set to YES.
9900 Transmitter 43
Salinity
INPUT Menu
Enter the precise cell constant from the certificate provided with your sensor, or from the information
label on the sensor. Shown if CELL CONSTANT = CUSTOM.
Select °C or °F.
Default = °C.
Dampens display, output and relay response rates. Select Low (4 sec), Med (8 sec), High (32 sec),
OFF (near instantaneous). (see discussion in Appendix.) Default = OFF.
For LINEAR temperature compensation, select a % per ºC slope. Maximum slope setting is 9.99 %
per ºC. (If Temperature Compensation setting is NONE, this item will not be displayed.)
44 9900 Transmitter
Troubleshooting
Troubleshooting
9900 Transmitter 45
Appendix
Averaging
NO AVERAGING, NO SENSITIVITY
With SENSITIVITY set to 0 (zero) and AVERAGING set to OFF (0 seconds), the 9900 responds immediately to every
shift in the process. The dashed red line represents the actual output of the sensor in varying conditions.
AVERAGING ONLY
With SENSITIVITY still set to zero and AVERAGING set to MED or HIGH the rate is stabilized, but a sharp change in
rate is not represented for 8 to 32 seconds or longer.
0s 5s 10 s 15 s 20 s 25 s 30 s 35 s
46 9900 Transmitter
Appendix
Only two parameters need to be set up, the starting or base conductivity value (4 mA SETPNT) and the ending or
maximum conductivity value (20 mA SETPNT). The 4 mA setpoint may be larger than the 20 mA point
(reverse span).
Conductivity = 10n
[n = (mA value – 4.0) × ((Log10(ending value) – Log10(starting value) / 16 mA) + Log10(starting value))]
If only fixed thresholds are of interest, they can be calculated in mA and then the mA value can be checked
directly. Inside the 9900 the following equation is used:
Notes: If ADJUST 4 mA or ADJUST 20 mA is used, the mA value can be affected. To prevent any problems the
adjust function should only be used to get exactly 4.0 and 20.0 at the PLC. The 9900 is accurate and the adjust
functions are only needed to compensate for an offset due to noise or a not-so-accurate PLC input card.
The error value of either 3.6 mA or 22 mA should be tested first before applying the conductivity equation.
9900 Transmitter 47
Appendix
Custom Measurements
Offset (O):
The distance from the zero reference point (Z) to
the sensor location point (SLoc).
• Enter a positive value in the Calibrate menu if
Z(ero) the sensor is located above Z.
L<0 • Enter a negative value in the Calibrate menu if
O(ffset) < 0
the sensor is located below Z.
L>0
L>0 • Enter 0 in the Calibrate menu if the sensor is
SLoc located at Z.
Z(ero) SLoc
O(ffset) > 0 O(ffset) < 0
SLoc Level (L):
Z(ero)
The distance from zero reference point (Z) to
surface of fluid (displayed as “Level” by 9900.)
48 9900 Transmitter
Appendix
level in the vessel. Each successive point must Custom point #7 700 gal.
be greater than the preceding point.
● The last point must be equal to or greater than Custom point #6 380 gal.
the highest fluid level in the vessel.
Custom point #5 250 gal.
● A custom point should be located at all transition
points in the vessel shape (for example, at Custom point #4 160 gal.
custom point #9, where the shape changes from
a cylinder to a cone). Custom point #3 118 gal.
9900 Transmitter 49
Appendix
In the LEVEL/VOLUME INPUT menu (see page 39), if SHAPE is set to HORIZ CYLINDER, RECTANGLE or
CUSTOM, the tank shape can be defined with the following screens:
If Horiz Cylinder or Rectangle shape is selected, enter the length of the vessel in LEVEL
UNITS. 0.0000 to 99999.
If Rectangle shape is selected, enter the width of the vessel in LEVEL UNITS.
0.0000 to 99999.
If Custom shape is selected, enter the number of measurement points to be used to define
the vessel shape (see Level and Volume Calculation in Custom Shaped Vessels discussion).
Minimum 3 points, maximum 10 points. A larger number of points improves accuracy.
Select (AUTO, MAN). Manual allows you to edit both the Level and the corresponding Volume
for your custom tank. Automatic allows you to edit the Volume measurement (while displaying
an automatically calculated Level value). See example below.
Enter the Level (if MAN measurement is selected) at each custom point in your vessel. If
AUTO is selected, indication will read actual tank level in LEVEL UNITS at that point
in your tank.
Set the Volume (if manual measurement is selected) at each custom point in your vessel.
Where (X) is number of custom points
Where (X) is number of custom points
For example, in a 25-gallon conical tank set for three custom points:
1. Pour 10 gallons of fluid into the conical tank. POINT 1 LEVEL will indicate actual tank level.
2. In POINT 1 VOL, enter 10.
3. Pour another 10 gallons into the tank. POINT 2 LEVEL will indicate actual tank level.
4. In POINT 2 VOL, enter 10.
5. Pour the final 5 gallons into the tank. POINT 3 LEVEL will indicate actual tank level.
6. In POINT 3 VOL, enter 5.
50 9900 Transmitter
Appendix
Mass Conversion
m = D(water) × SG × V
m (kg) = 1000 × SG × V
Volume Calculations
Vertical cylinder:
r
V = π × r2 × h
where r = radius of cylinder
A h h = height of fluid
Rectangular vessel:
V=w×l×h
where w = width
l = length
h = height
Horizontal cylinder:
V = L × [(r2 × cos-1 (r-h/r)) - sqrt(2rh - h2) × (r-h)] × A
where A = area of segment
L = Length of vessel
r = radius of vessel
Length h = height of segment
9900 Transmitter 51
Appendix
Calibration Procedures
EasyCal Procedure - pH
EasyCal is the fastest and simplest periodic calibration method.
Requires prepared 4, 7 or 10 pH buffers (any two).
3 ................... +237
4 ................... +177 2
5 ....................+118
6 ..................... +59 To exit menus and return to +
7 ....................... +0 VIEW press ▲ and ▼ Display returns to VIEW
buttons at the same time. Mode in 10 minutes
8 ......................–59
9 ....................–118
10 ..................–177
11 ..................–237
12 ..................–296
NOTE: The solutions can be used for calibrating more than one
sensor; however, the solution must remain free of debris and must
not be diluted by rinse water from previous calibrations.
52 9900 Transmitter
Appendix
Quick Manual
Calibration Procedures:
ENTER
Allow for stabilization
30 seconds to
30s several minutes ENTER
to accept
1-Point Calibration:
to accept 1. Set solution standard.
ENTER
Allow for stabilization
30 seconds to
30s several minutes ENTER
to accept
to accept
+
Display returns to
VIEW Mode
9900 Transmitter 53
Appendix
ENTER to accept
Place electrode tip in
saturated pH 7.0 buffer.
30s
Allow for stabilization
pH 7.0 = 87 mV 30 seconds
to accept second
Place electrode tip in
30s ENTER
buffer calibration
saturated pH 4.0 buffer. Allow for stabilization
30 seconds*
pH 4.0 = 264 mV
54 9900 Transmitter
Appendix
Quick Manual
Allow for stabilization
ENTER
Calibration Procedures:
30 seconds to
30s several minutes
ENTER
to accept 1-Point Calibration:
to accept
1. Set solution standard.
ENTER
Allow for stabilization
30 seconds to
30s several minutes ENTER to accept
to accept
pH 4.0 = 264 mV
+ Display returns to
VIEW Mode
9900 Transmitter 55
Appendix
Available buffer values are: 1. Reset the sensor to factory calibration (refer to sensor manual for
● 10 procedure).
● 100
● 146.93
2. On the 9900, select AUTO CAL from the CAL menu. Press ►.
● 200 3. Place the electrode/sensor assembly into the conductivity test solution
● 500 appropriate to your operating range. Shake the electrode to dislodge
● 1000 any air bubbles visible on the surface of the electrode.
● 1408.8
● 5000
4. Allow at least 2 minutes for the electrode response to stabilize.
● 10,000 5. When the display stabilizes, press ENTER.
● 12856 6. If calibration is successful, 9900 will display "SAVING". If error is too
● 50,000 large, "OUT OF RANGE USE MANUAL CALIBRATION" will display.
● 100,000
(all values in μS)
Calibration is complete. Return the system to service.
56 9900 Transmitter
Appendix
9900 Transmitter 57
Appendix
58 9900 Transmitter
Appendix
USP Limits
USP (United States Pharmacopoeia) has defined a set of conductivity values Temperature USP limit
(limits) to be used for pharmaceutical water. The standard requires that Range (ºC) (μS)
conductivity measurement without temperature compensation be used for 0 to < 5 0.6
these applications. The limits vary according to the temperature of the sample. 5 to < 10 0.8
The 9900 has the USP limits stored in memory. It will automatically determine 10 to < 15 0.9
the proper USP limit based on the measured temperature.
15 to < 20 1.0
20 to < 25 1.1
Using the USP function
USP setpoints are defined as a percentage below the USP limit, so a USP 25 to < 30 1.3
alarm is always a HIGH alarm. The 9900 can be set to warn you if the 30 to < 35 1.4
conductivity approaches within a set percentage of the USP limit. 35 to < 40 1.5
40 to < 45 1.7
The following settings and conditions are required for a USP relay function: 45 to < 50 1.8
50 to < 55 1.9
1. In the RELAY menu: 55 to < 60 2.1
• RELAY MODE must be set to USP. 60 to < 65 2.2
2. In the INPUT menu: 65 to < 70 2.4
9900 Transmitter 59
Specifications
Specifications
General Environmental Requirements
Input channels: ....... One Ambient operating temperature:
Backlit LCD: .............. -10 to 70 °C
Enclosure and Display (14 to 158 °F)
Case Material: ........ PBT Storage Temp:........... -15 to 80 °C
Window: .................. Shatterproof glass (5 to 176 °F)
Keypad: .................. 4 buttons, injection-molded Relative Humidity: ..... 0 to 100% condensing for
silicone rubber seal Field Mount and Panel
Display: ................... Backlit, 7- and 14-segment Mount front only;
Indicators ................ “Dial-type” digital bar graph 0 to 95% non-condensing
Update rate: ............ 1 s for Panel Mount back side.
LCD Contrast: ......... 5 settings Maximum Altitude:..... 4,000 m (13,123 ft); use
only DC power supply to
Enclosure maintain UL safety standard
Size .................. ¼ DIN up to this altitude
Color ................ Black (Panel Mount), Enclosure Rating:...... Designed to meet NEMA
Yellow and black (Integral Mount) 4X/IP65 (front face only on
panel mount; field mount is
Mounting 100% NEMA 4X/IP65).
Panel ............... ¼ DIN, ribbed on four sides for
panel mounting clip inside panel
Field ................. Mounts to standard Signet field Performance Specifications
mount junction boxes. Optional System Accuracy
angle adjustment adapter is ● Primarily dependent upon the sensor.
available
Wall .................. Large enclosure (sold as an System Response
accessory) that encases the panel ● Primarily dependent upon the sensor.
mount transmitter Controller adds a maximum of 150 ms
processing delay to the sensor electronics.
● Minimum update period is 100 ms
● System response is tempered by the display
rate, output averaging and sensitivity feature.
60 9900 Transmitter
Specifications
No DC power input
Max. loop impedance
@ 12 V loop power........... 50 Ω max.
@ 18 V loop power...........325 Ω max.
@ 24 V loop power...........600 Ω max.
9900 Transmitter 61
Specifications
Input Specifications
Digital (S3L):................. Serial ASCII, TTL level,
9600 bps
Frequency:
Sensitivity: ........... 80 mV @ 5 Hz, gradually
increasing with frequency
Span: ................... 0.5 Hz to 1500 Hz @ TTL
level input
Accuracy: ............ ± 0.5% of reading max error
@ 25 °C
Range .................. 0.5 to 1500 Hz
Resolution: .......... 1 μs
Repeatability: ...... ± 0.2% of reading
Power Supply
Rejection: .......... No Effect ± 1 μA per volt
Short Circuit ...... Protected
Reverse Polarity Protected (no isolation when
using loop power only)
Update Rate: ....... (1/frequency) + 150 ms
62 9900 Transmitter
Specifications
9900 Transmitter 63
Ordering Information
Ordering Information
9900 Transmitter Base Unit - Single Channel, Multi-Parameter, 4 to 20 mA, Open Collector, DC Power
Mfr. Part No Code Description
3-9900-1P 159 001 695 9900 Base Unit, Panel Mount
3-9900-1 159 001 696 9900 Base Unit, Field Mount
Optional Modules
Mfr. Part No Code Description
3-9900.393 159 001 698 Relay Module - 2 DCR (Dry Contact Relays)
3-9900.394 159 001 699 Direct Conductivity/Resistivity Module
3-9900.395 159 001 697 H COMM Module
Accessories
Mfr. Part No Code Description
6682-0204 159 001 709 Conductivity Module Plug, 4 Pos, Right Angle
6682-1102 159 001 710 DC Power Plug, 2 Pos, Right Angle
6682-1103 159 001 711 Relay Module Plug, 3 Pos, Right Angle
6682-1104 159 001 712 Loop Power Plug, 4 pos, Right Angle
6682-3004 159 001 725 Terminal Block Plug
6682-3104 159 001 713 Freq/S3L Plug, 4 pos, Right Angle
7300-7524 159 000 687 24 VDC power supply 7.5 W, 300 mA
7300-1524 159 000 688 24 VDC power supply 15 W, 600 mA
7300-3024 159 000 689 24 VDC power supply 30 W, 1.3 A
7300-5024 159 000 690 24 VDC power supply 50 W, 2.1 A
7300-1024 159 000 691 24 VDC power supply 100 W, 4.2 A
3-0251 159 001 724 PC COMM Configuration/Diagnostic tool
3-8050 159 000 184 Universal Mount Kit
3-8050.396 159 000 617 RC Filter Kit (for relay use), 2 per kit
3-8051 159 000 187 Flow Sensor Integral Mount Kit
3-8052 159 000 188 ¾ in. Integral Mount Kit
3-8058-1 159 000 966 i-Go™ Signal Converter, wire-mount
3-8058-2 159 000 967 i-Go™ Signal Converter, DIN rail mount
3-9900.390 159 001 714 Standard Connector Kit, Right Angle, 9900 Transmitter
3-9900.391 159 001 715 Connector Kit, In-Line, 9900 Transmitter
3-9900.392 159 001 700 Wall Mount Accessory for 9900
3-9000.392-1 159 000 839 Liquid Tight Connector Kit, NPT (1 pc.)
3-9900.396 159 001 701 Angle Adjustment Adapter Kit (for Field Mounting)
Georg Fischer Signet LLC, 3401 Aerojet Avenue, El Monte, CA 91731-2882 U.S.A. • Tel. (626) 571-2770 • Fax (626) 573-2057
For Worldwide Sales and Service, visit our website: gfsignet.com • or call (in the U.S.): (800) 854-4090
For the most up-to-date information, please refer to our website at gfsignet.com