Mindray A5 OM
Mindray A5 OM
Mindray A5 OM
Anesthesia System
A N E S T H E S I A S YS T E M
Anesthesia System
A N E S T H E S I A S YS T E M
Copyright © Shenzen Mindray Bio-Medical Electronics Co., Ltd, 2010 to 2018. All rights reserved. Contents of this
publication may not be reproduced in any form without permission of Shenzen Mindray Bio-Medical Electronics Co., Ltd.
Table of Contents
Table of Contents
Table of Contents ................................................................................................................................................................................................................ i
Foreword .............................................................................................................................................................................................................................vii
Indications For Use ..........................................................................................................................................................................................................vii
Responsibilities of Operators .......................................................................................................................................................................................vii
Warnings, Cautions, and Notes ...................................................................................................................................................................................vii
Warnings............................................................................................................................................................................................................................ viii
Cautions...............................................................................................................................................................................................................................xii
Notes.....................................................................................................................................................................................................................................xv
Intellectual Property Statement............................................................................................................................................................................... xvii
Warranty Statements ................................................................................................................................................................................................... xvii
Disclaimers...................................................................................................................................................................................................................... xviii
Phone Numbers and How To Get Assistance..................................................................................................................................................... xviii
Manufacturer’s Responsibility ................................................................................................................................................................................. xviii
Manufacturer and Address ...................................................................................................................................................................................... xviii
Symbols............................................................................................................................................................................................................................ xviii
Product Description............................................................................................................................................1 - 1
General System Overview .........................................................................................................................................................................................1 - 2
General Description..........................................................................................................................................................................................1 - 2
Key Features ........................................................................................................................................................................................................1 - 3
Fresh Gas Dosing ...............................................................................................................................................................................................1 - 3
Flow Control ........................................................................................................................................................................................................1 - 4
Vaporizer Mounting..........................................................................................................................................................................................1 - 4
Anesthesia Ventilator.......................................................................................................................................................................................1 - 4
Breathing System ..............................................................................................................................................................................................1 - 4
Active Anesthetic Gas Scavenging System ..............................................................................................................................................1 - 5
Passive Anesthetic Gas Scavenging System ............................................................................................................................................1 - 5
Power Management / Battery Supply........................................................................................................................................................1 - 5
Workplace Ergonomics....................................................................................................................................................................................1 - 7
Hook .......................................................................................................................................................................................................................1 - 7
Physical Views................................................................................................................................................................................................................1 - 8
Main Unit (Front View).....................................................................................................................................................................................1 - 8
Main Unit (Rear View).................................................................................................................................................................................... 1 - 10
Main Unit (Left View)..................................................................................................................................................................................... 1 - 12
Main Unit (Right View).................................................................................................................................................................................. 1 - 13
Main Unit (Top View)..................................................................................................................................................................................... 1 - 14
Breathing System (Top View) ..................................................................................................................................................................... 1 - 15
Breathing System (Left View) ..................................................................................................................................................................... 1 - 16
Active Anesthetic Gas Scavenging System (AGSS) (Top, Right, and Rear Views).................................................................... 1 - 18
Passive Anesthetic Gas Scavenging System (AGSS) (Right View) ................................................................................................. 1 - 20
Installation ..........................................................................................................................................................2 - 1
Unpacking.......................................................................................................................................................................................................................2 - 3
Initial Setup.....................................................................................................................................................................................................................2 - 4
Install the Vaporizer.....................................................................................................................................................................................................2 - 5
Filling and Draining the Vaporizer ..............................................................................................................................................................2 - 7
System Interface .................................................................................................................................................3 - 1
Main Screen Components.........................................................................................................................................................................................3 - 2
System Information Header .....................................................................................................................................................................................3 - 5
Elapsed / Countdown Timer..........................................................................................................................................................................3 - 5
Patient Size ..........................................................................................................................................................................................................3 - 6
Alarm and Prompt Messages ........................................................................................................................................................................3 - 6
Alarms............................................................................................................................................................................................................................ 9 - 21
Safety Specifications ................................................................................................................................................................................................ 9 - 22
ASTM F 1208 – 89 (2005) Disclosures................................................................................................................................................................. 9 - 23
Leakage of Breathing System..................................................................................................................................................................... 9 - 23
Resistance of Breathing Systems .............................................................................................................................................................. 9 - 23
CO2 Absorber Resistance............................................................................................................................................................................. 9 - 23
CO2 Absorber Capacity................................................................................................................................................................................. 9 - 23
Unidirectional Valve Opening Pressure.................................................................................................................................................. 9 - 24
Data Storage (Non-Volatile) and Recording .................................................................................................................................................... 9 - 24
Electromagnetic Compatibility ............................................................................................................................................................................ 9 - 24
Accessories ..........................................................................................................................................................A - 1
Accessory Kits ............................................................................................................................................................................................................... A - 2
AG Accessories ............................................................................................................................................................................................................. A - 2
CO2 Absorbent Canister............................................................................................................................................................................................ A - 2
Gas Cylinder Accessories .......................................................................................................................................................................................... A - 2
Gas Supply Hoses ........................................................................................................................................................................................................ A - 3
Manuals and Reference Cards ................................................................................................................................................................................ A - 3
Mounting Accessories ............................................................................................................................................................................................... A - 3
Networking and USB Storage ................................................................................................................................................................................. A - 4
Vaporizers....................................................................................................................................................................................................................... A - 4
Scavenging Accessories............................................................................................................................................................................................ A - 5
User Accessible Spare Parts ...............................................................................................................................B - 1
Active AGSS ................................................................................................................................................................................................................... B - 2
Breathing System ........................................................................................................................................................................................................ B - 2
CO2 Absorbent Canister............................................................................................................................................................................................ B - 2
Flow Sensor ................................................................................................................................................................................................................... B - 2
Gas Cylinder Accessories .......................................................................................................................................................................................... B - 3
O2 Sensor........................................................................................................................................................................................................................ B - 3
Battery ............................................................................................................................................................................................................................. B - 3
Parameters and Factory Defaults ......................................................................................................................C - 1
Waveform/Spirometry Tabs .................................................................................................................................................................................... C - 2
Alarm Limits................................................................................................................................................................................................................... C - 3
Setup Menu ................................................................................................................................................................................................................... C - 5
Alarm Volume and History....................................................................................................................................................................................... C - 8
Date and Time .............................................................................................................................................................................................................. C - 8
Demographics .............................................................................................................................................................................................................. C - 9
Ventilation Modes ....................................................................................................................................................................................................... C - 9
Linked Ventilation Parameter ...............................................................................................................................................................................C - 13
Ventilation Parameter Relationships..................................................................................................................................................................C - 15
Pneumatic Diagram ........................................................................................................................................... D - 1
Pneumatic Diagram of the A5 System.................................................................................................................................................................D - 2
Abbreviations, Symbols, and Units of Measure ............................................................................................... E - 1
Abbreviations ................................................................................................................................................................................................................E - 2
Symbols............................................................................................................................................................................................................................E - 4
Units of Measure...........................................................................................................................................................................................................E - 5
Attention Symbols .......................................................................................................................................................................................................E - 6
Preparation for Malignant Hyperthermia Susceptible Patients..................................................................... F - 1
Malignant Hyperthermia Causes, Effects and Treatment ..............................................................................................................................F - 2
Malignant Hyperthermia Washout ........................................................................................................................................................................F - 2
Washout Procedure for Malignant Hyperthermia Susceptible Patients with A5 Anesthesia Delivery Systems........................F - 2
References.......................................................................................................................................................................................................................F - 4
Foreword
WARNING: Do not operate the A5 Anesthesia System before reading these
instructions.
The operating instructions for the A5 Anesthesia Delivery System (hereinafter referred to as A5
Anesthesia System, A5 System, A5, or individual A5 is intended to provide information for proper
installation, operation, and general maintenance of the A5 System to the user.
General knowledge and understanding of the features and functions of the A5 System are
prerequisites for its proper use.
For servicing information or assistance, please contact an authorized representative in your area.
Rx only: U.S. Federal Law restricts this device to sale by or on the order of a
physician or other practitioner licensed by state law to use or order the
use of this device.
NOTE: Figures in this manual are provided for reference purposes only.
Screens may differ based on the system configuration and selected
parameters.
The A5 is intended for use by licensed clinicians, for patients requiring anesthesia within a health care
facility, and can be used for adult and pediatric populations.
Responsibilities of Operators
The proper function of the A5 System can only be guaranteed if it is operated and serviced in
accordance with the information provided in this manual and by an authorized Mindray service
representative. Non-compliance with this information voids all guarantee claims.
The A5 System must be operated by qualified and trained personnel only. All operators must fully
observe these operating instructions and relevant additional documentation. They must also comply
with the WARNINGS, CAUTIONS, and NOTES detailed in this manual.
WARNING — Indicates a potential hazard or unsafe practice that, if not avoided, could result in death
or serious injury to the patient or user.
CAUTION — Indicates a potential hazard or unsafe practice that, if not avoided, could result in
product/property damage or minor personal injury to the patient or user.
Warnings
WARNING: Do not operate the A5 Anesthesia System before reading these
instructions.
WARNING: This machine must only be operated by trained, skilled medical staff.
WARNING: Before putting the system into operation, the operator must verify that
the equipment, connecting cables, and accessories are in correct
working order and operating condition.
WARNING: Multiple AC power outlets are provided on the rear of the A5. These
outlets are intended to supply power to additional equipment that
form a part of the anesthesia system (i.e. vaporizers, gas analyzers,
etc.). Do not connect other equipment to these outlets, as patient
leakage current may be affected. Each outlet is rated 3 A; the total
current that may be drawn through all outlets is 10 A on the A5 System ;
do not attempt to exceed these load ratings. Do not connect additional
Multiple Portable Socket Outlets (i.e. Multiple outlet extension cords)
(MPSOs) or extension cords to these outlets.
WARNING: Do not open the equipment housings. All servicing and future
upgrades must be carried out only by trained and authorized Mindray
personnel.
WARNING: Do not rely exclusively on the audible alarm system for patient
monitoring.
WARNING: Adjustment of alarm volume to a low level may result in a hazard to the
patient.
WARNING: To avoid the possibility of explosion, do not use the equipment in the
presence of flammable anesthetic agents, vapors or liquids. Do not use
flammable anesthetic agents such as ether and cyclopropane for this
equipment. Use only non-flammable anesthetic agents that meet the
requirements specified in ISO 80601-2-13. The A5 Anesthesia System
can be used with halothane, enflurane, isoflurane, sevoflurane, and
desflurane. Only one anesthetic agent can be used at a time.
WARNING: Fresh gas flow must never be switched off before the vaporizer is
switched off. The vaporizer must never be left switched on without a
fresh-gas flow. Anesthetic agent vapor at a high concentration can get
into the machine lines and ambient air, causing harm to people and
materials.
WARNING: To avoid the risk of electric shock, this equipment must only be
connected to a supply mains with protective earth.
WARNING: Possible electric shock hazard. The machine may only be opened by
authorized service personnel.
WARNING: Always set the alarm limits so that the alarm is triggered before a
hazardous situation occurs. Incorrectly set alarm limits may result in
operating personnel not being aware of drastic changes in the patient’s
condition.
WARNING: Electric shock and fire hazard: Do not clean the machine while it is
powered on and/or plugged into an outlet.
WARNING: Disconnect the power plug from the mains supply before removing the
rear panels or servicing the A5 unit.
WARNING: Malfunction of the central gas supply system may cause more than one
or even all devices connected to it to stop their operation
simultaneously.
WARNING: The anesthesia system will cease to deliver gas at pressures below the
minimum specified gas pipeline supply pressure.
WARNING: Use extreme care while handling the absorbent as it is a caustic irritant.
WARNING: Use care in lifting and manipulating vaporizers during the mounting
process as their weight may be greater than expected, based on their
size and shape.
WARNING: Do not use talc, zinc stearate, calcium carbonate, corn starch, or similar
material to prevent sticking of the bellows, as these materials may
enter the patient's lungs or airway, causing irritation or injury.
WARNING: Single use respiratory hoses, face masks, sensors, sodalime, water
traps, sampling lines, airway adapters, and other single use items may
be considered potential biologically hazardous items and should not
be reused. Dispose of these items in accordance with hospital policy
and local regulations for contaminated and biologically hazardous
items.
WARNING: Review the performance specifications of the disposal system that the
transfer and receiving systems are intended to be used with, to ensure
compatibility.
WARNING: Ensure that the current alarm presets are appropriate before use on
each patient.
WARNING: A hazard can exist if different alarm presets are used for the same or
similar equipment in any single area.
WARNING: Due to the size and weight of the A5, it should only be moved by
qualified personnel.
WARNING: Excess load may cause a tip hazard while moving the A5. Before
moving, remove all equipment from the top shelf and all monitoring
equipment mounted to the side of the A5. Use care when moving the
A5 up or down inclines, around corners, and across thresholds. Do not
attempt to roll the A5 over hoses, cords, or other obstacles.
WARNING: Leaks or internal venting of sampled gas may affect accuracy. Perform
the proper preoperative tests to ensure that the device is performing
properly. Leaky circuits can not be used.
WARNING: Operation of the A5 below the minimum flow values may cause
inaccurate results.
WARNING: If the A5 is damaged in any way that compromises the safety of the
patient or user, discontinue use and attach a visible tag that marks the
A5 as unusable. Call Mindray Technical Support.
WARNING: Review the specifications of the AGSS transfer and receiving systems
and the specifications of the A5 System to ensure compatibility and to
prevent a mismatched receiving system.
WARNING: Avoid connecting two or more hose assemblies in series as this may
cause a loss of pressure and flow.
WARNING: A hazard may exist due to the use of improper connectors. Ensure all
assemblies use the proper connectors.
WARNING: Inspect all breathing system components carefully before each use.
Ensure all components do not contain any obstructions or debris that
can cause a potential hazard to the patient.
WARNING: Use breathing circuits and manual bags in accordance with ASTM F1208
and compatible with standard 22mm male conical fittings per ASTM
specifications F 1054.
WARNING: The mains plug is used to isolate the Anesthesia System circuits
electrically from the SUPPLY MAINS. Do not position the Anesthesia
System so that it is difficult to operate the plug.
WARNING: Do not touch the patient when connecting the peripheral equipment
via the I/O signal ports or replacing the oxygen cell to prevent patient
leakage current from exceeding the requirements specified by the
standard.
WARNING: If the Drive Gas Pressure Low alarm occurs when the gas supply
pressure is greater than 200 kPa, contact your service personnel or us.
WARNING: The anesthesia system shall not be serviced or maintained while being
connected on a patient.
Cautions
CAUTION: To ensure patient safety, use only parts and accessories specified in this
manual.
CAUTION: At the end of its service life, the equipment, as well as its accessories,
must be disposed of in compliance with the guidelines regulating the
disposal of such products, and in accordance with local regulations for
contaminated and biologically hazardous items.
CAUTION: Magnetic and electrical fields are capable of interfering with the proper
performance of the equipment. Ensure that all external devices
operating in the vicinity of the equipment comply with the relevant
EMC requirements. Mobile phones, x-ray equipment, and MRI devices
are possible sources of interference as they may emit higher levels of
electromagnetic radiation.
CAUTION: The A5 Anesthesia System may become unstable if the unit is tilted
beyond 10 degrees. Use extreme caution when moving or resting the
unit on surfaces exceeding a 10 degree slope. Do not hang articles on
the sides of the unit that would cause an excessive imbalance.
CAUTION: Perform the daily checks specified on the checklist. In case of a system
fault, do not operate the system until the fault has been corrected.
CAUTION: Before starting the machine, users must be familiar with the
information contained in these Operating Instructions and must have
been trained by an authorized representative.
CAUTION: If the machine does not function as described, it must be examined and
repaired as necessary by qualified service personnel before being
returned to use.
CAUTION: Handle the machine with care to prevent damage or functional faults.
CAUTION: Ensure that the gas supply of the machine always complies with the
technical specifications.
CAUTION: Before clinical use, the machine must be correctly calibrated and/or the
respective tests must be performed, as described in these Operating
Instructions.
CAUTION: If system faults occur during the initial calibration or testing, the
machine should not be operated until those faults have been corrected
by a qualified service person.
CAUTION: After servicing, functional, sensor, and system tests must be performed
before clinical use.
CAUTION: Use cleaning agent sparingly. Excess fluid could enter the machine,
causing damage.
CAUTION: Never immerse the oxygen sensor or its connector in any type of liquid.
CAUTION: The valve disc in each of the inhalation and exhalation valve assemblies
on the breathing system is fragile and must be handled with care while
removing the valve cage from the valve assembly.
CAUTION: If moisture remains in the bellows after cleaning, the bellows surface
folds may become tacky and prevent the bellows from properly
expanding. Ensure all moisture is removed from the bellows after
cleaning.
CAUTION: Do not connect any devices to the SB ports other than Mindray
approved USB storage devices and a supported USB mouse (see
‘‘Networking and USB Storage’’ on page A-4).
CAUTION: Users should monitor oxygen percentage (FiO2%) when using the
Auxiliary O2/Air Flow Meters. Unknown oxygen concentrations may be
delivered to the patient unless oxygen monitoring is used.
CAUTION: The A5 is NOT suitable for use in a magnetic resonance imaging (MRI)
environment.
CAUTION: Use the power cord provided with the product. If a substitute is
necessary, use only hospital grade power cords.
CAUTION: Unintended movement may occur if the casters are not locked. The
operator should lock casters during use of the machine.
CAUTION: Unsecured devices may slide off the top shelf. Devices should be
securely attached to the top shelf.
CAUTION: The voltage on the auxiliary outlets is the same voltage as the outlet
into which the A5 machine is plugged. Ensure that devices plugged into
the auxiliary outlets are rated for the same supply voltage as the A5.
CAUTION: During the transport and storage of the vaporizer, block the gas inlet
and outlet of the vaporizer with plugs to prevent foreign substances
from entering the vaporizer.
CAUTION: Do not use any flow outlets as handles for moving the A5. The flow
outlets may become damaged. Use the metal side bars on the main
body when moving the A5.
CAUTION: Do not push down on the bag arm forcefully or hang heavy objects
onto it. Excessive weight may bend and damage the bag arm.
CAUTION: Use care in lifting and manipulating the breathing system block during
removal from its mounting arm as handling may be awkward due to its
weight and shape.
CAUTION: Turn the flow controls slowly. To avoid damaging the control valves, do
not turn further when the flowmeter reading is outside the range.
When turning a flow control knob clockwise to decrease flow, the
flowmeter should reach zero before the knob reaches its most
clockwise mechanical stop (Off) position. Do not turn any further when
the knob has reached the Off position.
Notes
NOTE: Figures in this manual are provided for reference purposes only.
Screens may differ based on the system configuration and selected
parameters.
NOTE: Put the equipment in a location where you can easily see the screen and
access the operating controls.
NOTE: Keep this manual close to the equipment so that it can be obtained
conveniently when needed.
NOTE: The software was developed in compliance with IEC 60601-1. The
possibility of hazards arising from software errors is minimized.
NOTE: This manual describes all features and options. Your equipment may
not have all of them.
NOTE: Check the liquid level of the anesthetic agent before and during all
operations. When the liquid level is below the warning line, more
anesthetic agent needs to be added. Refer to the vaporizer Instructions
For Use for filling the vaporizer and other information.
NOTE: Areas designated for the servicing of oxygen equipment shall be clean,
free of oil and grease, and not used for the repair of other equipment.
NOTE: Opening the cylinder valve quickly may cause unexpected pressure
differentials and create a potential for fire or explosion arising from
oxygen pressure shocks. Open and shut the cylinder valve slowly.
NOTE: The power device, terminal units and pipeline system can be supplied
by one or several different manufacturers.
NOTE: The A5 can be equipped with one scavenger system to provide the best
match with the hospital’s waste-gas disposal system. The scavenger
system shall comply with ISO 80601-2-13.
NOTE: The Anesthesia System is compatible with gases (O2, N2O, and Air) and
anaesthetic agents (Halothane, Enflurane, Isoflurane, Sevoflurane, and
Desflurane).
NOTE: The Anesthesia System is compatible with gases (O2, N2O, and Air) and
anaesthetic agents (Halothane, Enflurane, Isoflurane, Sevoflurane, and
Desflurane).
Mindray intends to maintain the contents of this manual as confidential information. Disclosure of
the information in this manual in any manner whatsoever without the written permission of Mindray
is strictly forbidden.
Release, amendment, reproduction, distribution, rental, adaptation, translation or any other derivative
work of this manual in any manner whatsoever without the written permission of Mindray is strictly
forbidden.
This posting serves as notice under 35 U.S.C.§287(a) for Mindray patents: http://
www.mindrayna.com/patents.
Warranty Statements
Shenzen Mindray Bio-Medical Electronics Co., Ltd. warrants that its products will be free from defects
in workmanship and materials for a period of three (3) years from the date of purchase except that
disposable or one-time use products are warranted to be free from defects in workmanship and
materials up to a date one year from the date of purchase or the date of first use, whichever is sooner.
This warranty does not cover consumable items such as, but not limited to, batteries, external cables,
O2 sensors, CO2 absorbents, breathing circuits, hoses, or mounts.
Calibration may be performed without the need to disassemble the instrument. It is the responsibility
of the purchaser to perform calibration as necessary, in accordance with the instructions provided in
this manual.
Except as otherwise provided herein, the terms, conditions, and limitations of Shenzen Mindray Bio-
Medical Electronics Co., Ltd.’s standard warranty will remain in effect.
Shenzen Mindray Bio-Medical Electronics Co., Ltd. will not be liable for any incidental, special, or
consequential loss, damage, or expense directly or indirectly arising from the use of its products,
liability under this warranty and the buyer’s exclusive remedy under this warranty is limited to
servicing or replacing at Shenzen Mindray Bio-Medical Electronics Co., Ltd.’s option at the factory or at
an authorized distributor, any product which shall under normal use and service appear to the
Company to have been defective in material or workmanship.
No agent, employee, or representative of Shenzen Mindray Bio-Medical Electronics Co., Ltd. has any
authority to bind Shenzen Mindray Bio-Medical Electronics Co., Ltd. to any affirmation, representation,
or warranty concerning its products, and any affirmation, representation or warranty made by any
agent, employee, or representative shall not be enforceable by buyer.
This warranty is expressly in lieu of any other express or implied warranties, including any implied
warranty or merchantability or fitness, and of any other obligation on the part of the seller.
Disclaimers
Product Improvements — Shenzen Mindray Bio-Medical Electronics Co., Ltd. retains the right to
modify the machine and/or operating instructions without prior notification. These operating
instructions explain all features of the A5 System and are correct at time of manufacture. Instructions
and models produced at a later stage, may contain improvements or modifications that were not
included in previous models.
Please include the instrument model number, the serial number (located on the back of the A5), and
a description of the problem with all requests for service.
Warranty questions should be directed to a local representative. A list of offices, along with their
phone numbers, is provided at the end of this manual.
Manufacturer’s Responsibility
The effects on safety, reliability, and performance of the equipment are the manufacturer’s
responsibility only if:
Mindray Building, Keji 12th Road South, High-tech industrial park, Nanshan,
Address:
Shenzhen 518057, P.R. China
Symbols
The following table provides descriptions of symbols that are used on the device and/or within this
manual.
Environment:
Caution
Temperature Range
Electrical:
Environment:
Alternating Current
Pressure Range
(AC)
Electrical:
Gas Cylinder
Equipotentiality
Electrical:
Gas Inlet
Fuse or circuit breaker
Electrical:
Gas Outlet
Input Output
Electrical:
Gas Flow: Maximum
Light
Electrical:
Gas Flow: Minimum
Power On
Electrical:
Gas Flow Total
Power Standby
Electrical:
Gas:
Protective Earth
O2 Flush
(Ground)
Electrical:
WEEE (Waste of
E
Electrical and
Electronic
O2 Sensor Connector
Equipment) Marking.
Separate treatment
from general waste at
end of life.
Identifier:
Gas Pipeline Connection
Manufacturer
Identifier:
Manufacturer’s Material:
Reference/Catalog Polyphenyl-sulfone
Number
Identifier:
Material:
Serial Number
Polysulfone
Indicator
Lock/Unlock:
Touchpad
Direction
Lock/Unlock:
Manual ventilation via Breathing Bag
Lock
Lock/Unlock:
Automatic Ventilation
Unlock
No Heavy Objects
Water Trap
Do Not Crush
Protection against
IPX1 vertically falling water
drops
The A5 Anesthesia System consists of a main unit (includes an anesthetic ventilator and flowmeter
monitor assembly) and a detachable breathing system. The applied part of the anesthesia system are
breathing tubes and masks. Connect the patient to the anesthesia system via breathing tubes and
masks.
•
Electronic PEEP is available in all automatic ventilation modes. User control over inspiratory flow
(Tslope) is possible in PCV, SIMV, and PS modes. Automatic fresh gas compensation limits the effect
on the patient ventilation from changes in fresh gas flow rate by the operator. The traditional bellows
system is driven by oxygen and makes patient disconnections clearly visible.
FEATURE DESCRIPTION
Display 15 inch color LCD with touchscreen
Navigation Graphical user interface for easy navigation
Ventilation Manual and automatic ventilation modes and monitoring: VCV, SIMV-VC,
PCV, SIMV-PC , PS, and Manual
Fresh Gas Delivery Continuous and intermittent anesthesia flow, total flow rotameter,
virtual dual flow tubes, electronically displayed on screen for ease of use
3 cylinder mount locations on rear
Breathing System Heated, adjustable swivel, side hose ports, single turn APL valve
Ergonomics Large stainless steel work surface
Adjustable breathing system block via swivel up to 50 degrees
Electronic PEEP Positive End Expiratory Pressure (PEEP) is set and controlled electronically.
Clear Data Display Two large waveforms for pressure and flow or
Spirometry Loops
USB Mouse Support The A5 system supports a wired USB mouse, which can be plugged into one
of the two SB ports at the rear of the unit. A cursor appears when the mouse
is plugged. The cursor disappears if the user touches the screen or after 15
seconds of mouse inactivity.
The USB mouse can serve as a backup to both the touchscreen and
touchpad.
• Virtual On-Screen dual flow tube and numerical readouts to display the O2, N2O, and Air flows
• A knob guard to prevent inadvertent movement of the flow control knobs
• Gas supply gauges to indicate the gas pipeline supply pressures and gas cylinder pressures
• Mechanical total flowmeter to display the combined flow of O2, Air, and N2O
• An O2 flush button
• A single combined output of auxiliary O2 and Air with flowmeters
Safety systems within the A5 work to prevent hypoxic mixtures from being delivered to the patient.
Nitrous oxide will not be delivered unless oxygen flow is present. A pneumatic safety system assures
that at least 21% O2 is present when setting mixtures of O2 and N2O. Additionally, if the A5 is placed
in the standby mode, O2 fresh gas flow is not available.
WARNING: Ensure that both O2 and N2O flow controllers are turned OFF fully at the
start and at the end of each case.
All A5 units are designed to maintain a safe O2:N2O ratio by allowing nitrous oxide to be set to a flow
rate that is proportional to a previously adjusted flow of oxygen. The N2O flow is limited by the flow
of O2 so that a safe ratio of no less than 21% oxygen can be maintained. The A5 is designed to
maintain oxygen flow at its previously set level when N2O is decreased.
When adjusting N2O and O2 flow rates, always adjust the oxygen flow first to enable the nitrous oxide
flow. To add N2O to the fresh gas flow, open the N2O flowmeter valve, but only after opening the O2
flowmeter valve.
For the A5 model, a third, non-functional vaporizer parking spot on the side of the unit is provided as
part of the standard configuration.
The A5 offers additional ventilation modes, which include Pressure Control Ventilation (PCV) with and
without Volume Guarantee (VG), and Synchronized Intermittent Mandatory Ventilation-Pressure
Control (SIMV-PC).
The breathing system provides access to the APL valve and breathing bag along with a view of the
airway pressure gauge. The APL valve has a single turn knob that provides a clear view of the manual
breathing pressure setting. The absorber assembly incorporates a cam-lock device that opens and
closes to provide access to the absorber canister. Either a CO2 absorbent Pre-Pak or loose fill can be
used. Two water traps that can be drained are located on the CO2 absorber assembly and on the
breathing system block.
NOTE: Operating the A5 with a full water trap in the breathing system block
does not allow the water to condense appropriately. The trap should be
removed and emptied when filled with water.
Two (2) flow sensors in the breathing system measure inspired and expired gases for control and
monitoring. Inspired oxygen concentration is monitored via a fuel-cell type sensor. Breathing
pressure is monitored with both a PAW gauge (mechanical) and electronic gauge. The breathing
system can be swiveled for ease of positioning. A leak test port is provided to allow for leak testing
during startup.
The breathing system connects to the A5 main unit through the following ports:
• Drive gas port, designed for use with oxygen as the drive gas
• Fresh gas port
• Exhaust gas port
• Flow sensor pressure transmission pipeline port
The breathing system contains the following ports for end-user connections:
A recessed main switch is provided to power the system ON and to put the system on power standby
where the battery supply continues to charge as necessary when the A5 is plugged into an external
power source. The main switch also stops the O2 fresh gas supply when the A5 is placed in Power
Standby mode.
Auxiliary AC outlets on the rear of the machine operate independently of the main switch position.
The A5 provides four (4) auxiliary AC outlets. The auxiliary AC outlets are not powered when
operating the A5 on the internal battery supply.
NOTE: Use the battery supply in the A5 at least once every month to extend
battery life. Charge the battery supply before its power capacity is
depleted.
NOTE: Inspect and replace the battery supply at regular service intervals.
Long-term battery life depends on how frequent and how long the
battery supply is used. For a properly maintained and stored lithium-
ion battery, its long-term life expectancy is approximately three (3)
years. In more aggressive usage, life expectancy can be shortened.
Replacing lithium-ion batteries every three (3) years is recommended.
NOTE: In case of battery failure, contact Mindray service personnel for battery
supply replacement.
NOTE: When a battery has been stored for a long time, or the battery is
depleted, recharge the battery at once. Otherwise, the low battery may
not be sufficient to power the A5 if the AC power is unavailable.
CAUTION: Please replace your battery when it reaches the end of its service life.
Failure to replace the battery may cause serious damage to your device
from battery overheating.
When power is lost for less than or equal to 60 s, the alarm settings prior to the power loss shall be
restored automatically.
The on-screen battery symbol indicates the battery status (see FIGURE 1-1).
PART(S) DESCRIPTION
Battery supply is fully charged.
AC power is connected.
The A5 is being powered by AC power. The solid portion
represents the current charge level of the batteries in proportion
to its maximum charge level.
Battery supply is partially charged.
AC power is connected and charging battery supply.
The A5 is being powered by AC power.
PART(S) DESCRIPTION
Battery supply is fully charged.
AC power is not connected.
The A5 is being powered by internal battery supply.
If the battery capacity is too low, power supply failure will result. A high-level alarm will be triggered
and the message Low Battery Voltage! will be displayed in the technical alarm area. In this case,
apply AC power to the A5 Anesthesia System to resume operation and charge the battery supply.
The operator of the A5 should be positioned in front of the monitor at a comfortable distance to view
all displayed waveforms, text, and controls.
1.1.12 Hook
There is a hook located on the front of the breathing system that can be used to hang the tubes of
the breathing circuit.
A1 A18
A19
A2
A20
A3 A21
A22
A4
A23
A5 A24
A6
A25
A7
A26
A8
A9 A27
A10
A28
A11
A12
A29
A13
A14
A15
A16
A17
PART(S) DESCRIPTION
A1 Alarm Light Illuminates red, yellow, or cyan during an alarm condition
to indicate the alarm priority. Red = high priority,
Yellow = medium priority, cyan = low priority,
off = no alarm condition.
A2 LCD Touchscreen Display / See section ‘‘System Interface’’ on page 3-1
System Interface
A3 Auxiliary O2/Air Flowmeters Auxiliary O2/Air Flowmeters for auxiliary O2/Air output
PART(S) DESCRIPTION
A4 Auxiliary O2/Air Gas Outlet Nozzle (barbed connector) for auxiliary O2/Air output.
Combines the auxiliary O2/Air flowmeters into a single
output of O2 only, Air only, or O2/Air blend, depending
upon the O2 and Air flow adjustments.
A5 Auxiliary O2 Gas Power Outlet High pressure O2 outlet (DISS connector) for connecting
external devices such as a jet ventilator.
A6 Flow Control Knobs N2O, Air, and O2 gas dosing. Turn each knob
counterclockwise to increase flow.
A7 Pressure Gauges (pipeline) Indicate the pressure at pipeline inlets for O2, Air, and N2O.
A8 Pressure Gauges (cylinder) Indicate the pressure at cylinder inlets for O2, Air, and N2O.
A9 Total Flow Meter Displays the combined flow rate of O2, Air, and N2O.
A10 O2 Sensor Electrical Port Connects the O2 sensor cable on the breathing system to
the main A5 unit.
A11 O2 Flush Button Provides high flow O2 to the inspiratory limb of the
breathing system.
A12 Vacuum suction fixing clip Holds the tubes of the negative pressure suction device.
A13 Touchpad Allows alternate control of the touch screen. Pull out to use.
A18 Work Light Located under the top shelf to illuminate the work level
shelf and allow the user to read the vaporizer dial setting in
a darkened room.
A19 Work Light Switch Turns on/off the work light. Three settings: Off, Low, and
High. The user can turn on the work light only when the
main power switch is turned on.
A20 Vaporizer Mounting Manifold / An interface for two/three Selectatec-type vaporizers to
Mounting Bar mount in this location. Bar holds two/three (optional)
vaporizers. An interlock within the vaporizers provides for
use of one vaporizer to deliver one agent at a time.
A21 Vaporizer Mount Valve Vaporizer index and outlet ports.
Cartridge
A22 Vaporizer Parking Spot Holds a non-functional vaporizer for user convenience.
A23 Vaporizer Locking Device Vaporizer locking mechanism to secure against accidental
disconnection
A24 AC Status LED Illuminated when the system is connected to an AC power
source.
A25 Battery Charging LED Illuminated when the battery supply is charging.
A26 Main Power Switch Switch to turn the system on or off.
A27 Handle Metal bar used to assist moving the A5
PART(S) DESCRIPTION
A28 Key lock Key and lock for securing the drawers
A29 Storage Drawers Drawers (3) for storage (lockable)
B1 B2 B3
B4
B5
B6 B7
B9
B8
B10
B11
B13
B12
B14
PART(S) DESCRIPTION
B1 Circuit Breakers Breakers for each auxiliary outlet
3 A each (quantity 4), 10 A total (quantity 1)
PART(S) DESCRIPTION
B3 Exhaust Fan Forces air to cool electronics and prevent buildup of O2
concentration. Do not block.
B4 Communication Ports SP1, DP1, CS1, SB1, SB2 (see section 9.6.4 (page 9-8)
"Communication Ports").
The A5 outlets are covered with two (2) metal plates, and
require a tool to access. Only authorized personnel can
access these outlets.
B6 Equipotential stud / lug Provides a ground point. Eliminates the ground potential
difference between different devices to ensure safety.
B7 Hooks Allows user to hang or wrap cords
B8 Cylinder Supply Connections Interface connectors to high pressure supply tanks (O2, Air,
and N2O)
B9 Gas Pipeline Supply Connections for O2, Air, and N2O from a pipeline gas supply
Connections
B10 Sample Line Exhaust Gas Inlet Inlet for exhaust gas from gas module. Merges with the
AGSS connector that connects to the AGSS.
B11 AGSS Connector Connects the AGSS or waste gas disposal system
B12 AGSS Anesthetic Gas Scavenging System
B13 Cylinders Supply tanks (E-size) containing high pressure O2, Air, and
N2O to act as backup supply if the pipeline pressure is
removed. Note: Tanks not supplied by Mindray.
B14 AGSS Transfer Hose Routes exhaust gases from main unit to scavenger.
C4 C1
C2
C3
C5
PART(S) DESCRIPTION
C1 Auxiliary O2/Air Auxiliary O2/Air Flowmeters for auxiliary O2/Air output
Flowmeters
C2 Auxiliary O2/Air Gas Nozzle (barbed connector) for auxiliary O2/Air output. Combines the
Outlet auxiliary O2/Air flowmeters into a single output.
C3 Auxiliary O2 Gas High pressure O2 outlet (DISS connector) for connecting external
Power Outlet devices such as a jet ventilator.
C4 Rail Mount Enables mounting of patient monitors and most standard attachment
arms for other devices. Rail mounts are on both left and right sides of
the A5.
C5 Module slot AG module can be inserted into the slot and identified.
D1
D2
D3
D4
PART(S) DESCRIPTION
D1 Vaporizer Mounting An interface for two/three Selectatec-type vaporizers to
Manifold / Mounting Bar mount in this location. Bar holds two/three (optional)
vaporizers. An interlock within the vaporizers provides for use
of one vaporizer to deliver one agent at a time.
D2 Key Lock Key and lock for securing the drawers
D3 Storage Drawers Drawers (3) for storage (lockable)
D4 Rail Mount Enables mounting of patient monitors and most standard
attachment arms for other devices. Rail mounts are on both
left and right sides of the A5.
E1
Rear
E4
E3
E2
Front
PART(S) DESCRIPTION
E1 Top Shelf Top level surface
E2 Work Level Shelf Work Level surface (stainless steel)
E3 Handle Wrap-around metal bar used to assist moving the A5 device
E4 Mounting Holes Allows mounting of optional equipment to the top shelf
(i.e., DPM6 and DPM7 mounting plates and kits. See
section A.7 (page A-3) "Mounting Accessories")
F1
F9
F2
F3
F4 F8
F5
F6
F7
PART(S) DESCRIPTION
F1 Bellows (including bellows Bellows that separates the breathing system gases from the
dome)1 oxygen drive gas
F2 PAW Gauge2 Indicates the patient airway pressure
F3 Expiratory Limb Exhaled breathing circuit connection
F4 Inspiratory Limb Inhaled breathing circuit connection
F5 Expiration Valve Allows flow of expiratory gas from the patient to the re-
breathing system, and prevents reverse flow.
F6 Inspiration Valve Allows flow of inspiratory gas to the patient, and prevents
reverse flow.
F7 O2 Sensor Cable Assembly An electro-galvanic fuel cell device to measure the
concentration of O2. The assembly is composed of the O2
cable, O2 cell cover, and O2 sensor.
1
The bellows dome is a transparent cover with graduation marks from 300 to 1500 mL. These marks are for
reference only. Tidal volume (Vt) should be read exclusively from the display of the user interface. Delivered Vt
is a combination of bellows displacement and fresh gas flow.
2
The APL valve and Paw gauge numerics are for reference only. Calibrated patient airway pressure is displayed
on the user interface.
PART(S) DESCRIPTION
F8 APL (Airway Pressure Rotary regulator for setting the breathing system pressure
Limiting) Valve2 limit during manual ventilation. Its scale shows approximate
pressure. Set to SP during Spontaneous breathing.
or
Quick Release APL Valve2 Rotary regulator for setting the breathing system pressure
limit during manual ventilation. Its scale shows approximate
pressure. Set to SP during Spontaneous breathing. When
necessary, lift the APL valve upward to release pressure
quickly.
F19 Auto/Manual Bag Switch Enables switching between Automatic and Manual
ventilation modes
1 The bellows dome is a transparent cover with graduation marks from 300 to 1500 mL. These marks are for
reference only. Tidal volume (Vt) should be read exclusively from the display of the user interface. Delivered Vt
is a combination of bellows displacement and fresh gas flow.
2 The APL valve and Paw gauge numerics are for reference only. Calibrated patient airway pressure is displayed
on the user interface.
G7
G6
G5
G4
G1
G3
G2
FIGURE 1-8 Breathing System (Left View, the Flexible Bag Arm (optional))
G8
G6
G5
G4
G1
G3
G2
FIGURE 1-9 Breathing System (Left View, the Fixed Height Bag Arm (standard))
PART(S) DESCRIPTION
G1 CO2 Absorber Canister Container for CO2 absorbent material loose fill or Pre-Paks)
G2 Condensate Drain Valve Turn counter-clockwise (looking from bottom) to drain water
collected in the absorber canister.
G3 Absorber Canister Lock Lever-type locking mechanism to lock (horizontal position) or
unlock (vertical position) the absorber canister from the
canister assembly.
G4 Water Trap Accumulates condensate from the breathing system. Must be
removed and emptied periodically. To remove, turn clockwise
(looking from top).
G5 Absorber Bypass Assembly Maintains pressure in the breathing circuit when changing
the sodalime contents in the CO2 absorber canister.
G6 Hook Hang the tubes of the breathing system.
PART(S) DESCRIPTION
G7 Flexible Bag Arm Provides the interface for the manual ventilation bag. The
flexible bag arm can be adjusted to desired height and the
bag port can be rotated 360°.
G8 Fixed Height Bag Arm Provides the interface for the manual ventilation bag The
height of fixed bag arm cannot be adjusted and the bag port
is in a fixed direction.
Top View
H1
H2
Right View Rear View
H3
Max Max
H5
Min Min
H4
PART(S) DESCRIPTION
H1 Mounting Rail Attachment Allows the AGSS to be mounted on the side rail. Contains a
thumbscrew that must be tightened against the mounting
rail.
H2 Flow Adjust Knob Turn clockwise or counter-clockwise to adjust the flow in the
AGSS until the float is between Min and Max marks.
H3 Exhaust Port Exhaust port to the hospital’s waste gas scavenging system.
H4 Inlet Port Intake for exhaust gases from the breathing system. An AGSS
transfer hose connects the Inlet and AGSS ports (see
FIGURE 1-3) to transfer the exhaust gases.
H5 Float Indicates exhaust flow. Adjusted by turning the Flow Adjust
Knob (H2) until the float is between the Min and Max marks.
I1 I2
I3
Rear View
PART(S) DESCRIPTION
I1 Inlet Port Intake for exhaust gases from the breathing system
connecting with the AGSS ports.
I2 Exhaust Port Exhaust port to the hospital’s waste gas scavenging system.
I3 Manual Bag When the manual bag is inflated, it indicates that the passive
AGSS is blocked.
Unpacking.....................................................................................................................................................................................................2-3
Initial Setup...................................................................................................................................................................................................2-4
WARNING: Do not use masks or breathing tubes that are antistatic or conductive.
They can cause burns if they are used near high frequency
electrosurgical equipment.
WARNING: This A5 Anesthesia System has waste gas exhaust ports. The operator of
the machine should pay attention to the disposal of the residual
breathing gas scavenged.
CAUTION: The operational environment and the power source of the equipment
must comply with the requirements as specified in the A5 ‘‘Product
Specifications’’ on page 9-1.
2.1 Unpacking
When the A5 Anesthesia System is delivered, IMMEDIATELY inspect the box for any damage.
a. If there is NO damage and ALL tip indicators on the box exterior are intact, then sign and
date the bill of lading or airway bill to indicate safe receipt of the A5.
b. If there is DAMAGE or ANY of the tip indicators on the box exterior have been activated,
then conditionally accept the delivery and clearly describe the damages on the bill of
lading or airway bill. BOTH the carrier and recipient must sign and date the bill of lading
or airway bill. Save all damaged factory packaging until further instructed by Mindray.
The receiver should immediately contact Mindray Customer Service at +86 755
26582479 / 26582888.
WARNING: Use vaporizers compliant to ISO 80601-2-13. See section A.9 (page A-4)
"Vaporizers". Refer to the vaporizer manufacturer’s Instructions For Use
for mounting, filling, or draining the vaporizer and other information.
WARNING: Use care in lifting and manipulating vaporizers during the mounting
process as their weight may be greater than expected, based on their
size and shape.
NOTE: The barometric pressure may differ from the calibration pressure of the
anesthetic vaporizer. This may cause an inaccurate output of the
anesthetic agent. The operator should continuously monitor the
concentration of anesthetic agent during system use to determine if
the outputted concentration is accurate.
1. If replacing and removing the vaporizer, lift each vaporizer straight up off the manifold. Do
not pull the vaporizer forward. Do not rotate the vaporizer on the manifold.
2. Align the new vaporizer over the valve cartridges of the mounting bar, slightly tilting back
the vaporizer. Hang the vaporizer on the mounting bar as shown in FIGURE 2-2. Ensure that
the locking mechanism handle is in the unlocked position. Ensure that the dial is in the “0”
(Transport) position or equivalent, depending upon the vaporizer manufacturer’s
Instructions For Use.
Locking Mechanism
Handle in the
Unlocked Position
3. Rotate the locking mechanism handle clockwise into the locked position as shown in
FIGURE 2-3.
Locking Mechanism
Handle in the Locked
4. Final check:
1. Ensure that the top of the vaporizer is horizontal. If not, remove and reinstall the vaporizer.
2. If a vaporizer lifts off the manifold, repeat steps 1 through 3 to reinstall the vaporizer. If the
vaporizer lifts off a second time, do not use the system.
WARNING: For the A5 Anesthesia System, using or turning on more than one
vaporizer simultaneously is prohibited and prevented by a mechanical
interlock. Do not attempt to override this safety mechanism.
WARNING: Ensure that the correct anesthetic agent is used. The vaporizer is
designed with the specific anesthetic agent named on it and further
indicated by color coded labeling. The concentration of the anesthetic
agent actually output will vary if the vaporizer is filled with the wrong
agent.
WARNING: Do not reuse the agent drained from the vaporizer. Treat as a hazardous
chemical and follow local regulations for proper disposal.
15 16 13 14
MAIN SCREEN
NUMBER COMPONENT DESCRIPTION
1 Elapsed / Countdown Timer Displays elapsed time or countdown time. Select to start,
stop, or reset the timer.
2 Fresh Gas Flow Area Displays real-time flowmeter levels for N2O, Air, and O2.
3 Patient Size Displays the currently selected patient size (Adult,
Pediatric, or Infant). Select to change the patient size when
the A5 is in Standby mode, Manual mode or Monitorl*
mode
4 Current Ventilation Mode Displays the current ventilation mode (VCV, SIMV-VC, PCV,
SIMV-PC*, PS, Manual, Bypass**,Monitor or Standby.)
5 Waveforms Tab See “Waveforms Tab” on page 3-11.
6 Spirometry Tab See “Spirometry Tab” on page 3-13.
7 Demographics Tab See “Demographics Tab” on page 3-19.
* Monitor mode is only available with the AG module.
**SIMV-PC and Bypass are only available on A5.
MAIN SCREEN
NUMBER COMPONENT DESCRIPTION
8 Alarm / Prompt Message Displays physiological alarms, technical alarms, and
Area prompt messages. The most recent highest priority alarm is
displayed at the top.
See ‘‘Alarms and Messages’’ on page 6-1 for tables that list
the individual messages and their associated priority levels.
High priority messages are red. Medium priority messages
are yellow. Low priority messages are cyan. Prompt
messages are black text on white background.
9 Alarm Silence Icon Displays the alarm silence icon and Alarm Silence
countdown timer for 120 seconds when the Silence softkey
is selected.
10 Breathing System Warmer Icon Indicates the warmer is not active.
11 System Date and Time Displays the current system date and time. Select to adjust
the date and time. See ‘‘Date and Time’’ on page 3-8.
12 Main Power Supply and Displays the main power supply and battery state. See
Battery Status Icon “Power Management / Battery Supply” on page 1-5.
13 Ventilations Mode and Displays tabs for all ventilation modes (VCV, SIMV-VC, PCV,
Setting Parameters Area SIMV-PC*, PS, Manual/Bypass*or Monitor). Each tab
displays the ventilation mode and its parameters. Select a
tab and the “Set Mode” softkey to change the ventilation
mode. Select a parameter button to change the parameter
setting. See ‘‘Ventilation Modes’’ on page 5-8.
* Monitor mode is only available with the AG module.
**SIMV-PC and Bypass are only available on A5.
MAIN SCREEN
NUMBER COMPONENT DESCRIPTION
14 System Softkeys Setup – Select to open the Setup menu. The Setup menu
contains the General tab, Display tab, System tab, and
Service tab.
Alarms – Select to open the Alarms menu to set alarm
limits, set alarm volume, and view all active alarms.
Silence – Select Silence softkey to silence all currently
sounding alarm tones. The alarm will sound if a new alarm
occurs.
If the silenced alarms contain middle or high level alarms,
the alarm audio will be paused for 120 seconds.The alarm
silence icon and 120 second countdown time appear at the
top of the screen. Select again to resume the alarm audio.
Note, however, the alarm will sound if that a new alarm
occurs while the system is in an audio-paused state. If this
occurs, you can select the Silence softkey again to silence
the new alarm and reset the silence countdown timer to
120 seconds.
If the silenced alarms are only low level alarms, the alarm
audio will be turned off till there is a new alarm occurs.
Note, however, the alarm will sound if that a new alarm
occurs while the system is in an audio-off state. If the new
alarm is low level alarm, you can select the Silence softkey
again to turn off the new alarm audio. If the new alarm is
medium or high level alarm, you can select the Silence
softkey again to silence the new alarm for 120 seconds.
Elapsed Timer
Select the Start button to turn on the elapsed timer (see FIGURE 3-3). Select the Stop button to
pause the elapsed timer and the timer will flash. Select the Reset button to turn off the elapsed timer.
Input the time with the keypad and select the Start button to turn on the countdown timer (see
FIGURE 3-4). Select the Stop button to pause the countdown timer and the timer will flash. Select the
Reset button to turn off the countdown timer.
When the countdown timer is expired, the system will pop-up a warning dialog (see FIGURE 3-5) and
provide a notification sound at the same time, The sound will repeat until the Done button is
pressed.
Elapsed and Countdown Timer (software bundle version 02.06.00 and later)
Turn on both the elapsed and countdown timer, the figure below displays on the screen (see
FIGURE 3-6).
Elapsed Timer
Countdown Timer
Patient Size
The remaining alarms are displayed in the lower area and grouped by priority. The most recent of
these alarms are displayed first.
Select this area to display a list of all active alarms. See ‘‘Alarms and Messages’’ on page 6-1 for tables
that list the individual messages and their associated priority levels. High priority messages are red.
Medium priority messages are yellow. Low priority messages are cyan. Prompt messages are black
text on white background (see FIGURE 3-8).
1. Select the Date and Time icon. The Date/Time dialog is displayed (see FIGURE 3-11).
2. Use the dialog keypad and softkeys to adjust the date, time, 12/24 hour format, date format,
and daylight savings time.
NOTE: If applicable, select Daylight Savings Time first before all other
settings.
PART(S) DESCRIPTION
Battery supply is fully charged.
AC power is connected.
The A5 is being powered by AC power. The solid portion
represents the current charge level of the batteries in proportion
to its maximum charge level.
Battery supply is partially charged.
AC power is connected and charging batteries.
The A5 is being powered by AC power.
The flowmeter numerics display a precision to two decimal digits for flows < 1 L/min and one
decimal digit for flows ≥ 1 L/min.
For the A5, the size (height) of the fresh gas flow tubes changes depending on whether the AG
module is connected as shown in FIGURE 3-14.
Waveforms Tab
If the measured value of Paw or Flow is less than the boundary minus a margin at the end of two
continuous breath cycles, the A5 System will auto scale the Paw or Flow at the beginning of next
breath cycle.
Paw scale:
The margin will be 10 cmH2O if pressure ≥ 30 cmH2O.
Flow scale:
You can press the button (see FIGURE 3-16) to only view the spirometry loop (see FIGURE 3-17).
Spirometry Tab
You can press the button (see FIGURE 3-17) to see the spirometry loop and waveforms (see
FIGURE 3-16).
FIGURE 3-19 Spirometry: Pressure-Flow Loop (only for software bundle version 02.06.00 and
later)
Spirometry loops reflect patient lung function and ventilation. They also indicate other related
parameters such as compliance, over-inflation, breathing system leak, and airway blockage.
The system provides three spirometry loops: pressure - volume loop (see FIGURE 3-17), flow - volume
loop (see FIGURE 3-18) and pressure - flow loop (see FIGURE 3-19). Loops data comes from pressure
and flow data. Only one loop is displayed at a time.
The Spirometry tab displays four softkeys: Loop Type, Show Reference, Save Loop, and Review
Loops.
FIGURE 3-20 Spirometry Softkeys: Loop Type, Show Reference, Save Loop, and Review Loops
The Show Reference softkey (see FIGURE 3-20) is used to select and display a saved baseline loop,
reference loop, or no loop (Off ) in the spirometry loop window, overlapped with the currently
plotting loop. Only the four most recently saved reference loops are listed chronologically.
When a reference loop or baseline loop is selected to display in the spirometry loop window, the time
stamp will also be displayed.
The saved baseline or reference loop can be reviewed with its numeric data (via Review Loops
softkey) or displayed with the currently plotting loop on the same graph for comparison (via Show
Reference softkey).
1. From the main screen, select Spirometry tab > Save Loop softkey.
If there is no baseline loop saved in memory, the currently plotting loop will be saved
automatically as the baseline loop.
2. If a baseline loop is already saved in memory, a dialog box will appear with the choices of
“Baseline“ and ”Reference“. Select ”Baseline”. A confirmation dialog will be displayed with
the text “Selecting Yes will replace the currently saved Baseline loop. Do you want to
proceed?” If “Yes” is selected, the currently saved baseline loop will be replaced. If “No” is
chosen, the save will be aborted.
To save a reference loop:
1. From the main screen, select Spirometry tab > Save Loop softkey. If a baseline loop is
already saved in memory, a dialog box will appear with the choices of "Baseline" and
"Reference". Select "Reference".
A maximum of four (4) sets of reference loops plus one (1) Baseline loop and corresponding numeric
data can be saved.
When the maximum of four (4) loops is reached, and the user attempts another save, a confirmation
dialog will be displayed with the following text, “Selecting Yes will replace the oldest reference
loop. Do you want to proceed?” If “Yes” is chosen, the oldest data will be removed as the new data
is added. If "No" is chosen, the save will be aborted.
Small Loop Window: These small graphic windows show the baseline and reference loops. The
baseline loop (only one) is always located on the left and has a white border around its graph. The
reference loops (up to four) are located to the right of the baseline loop. The reference loops are
displayed from oldest (left) to newest (right).
The baseline loop information is displayed below the small baseline loop window. The reference loop
information is displayed in cyan highlight for the reference loop that is selected.
Large Loop Window: This graphic window shows an enlarged view of the selected reference loop
overlapped with the baseline loop.
Loop Type: The Loop Type softkey is used to choose the type of loop to review. The choices are:
Pressure - Volume, Flow - Volume and Pressure - Flow. Default loop type is Pressure - Volume
loop.
Delete Loop: The Delete Loop softkey is used to delete a selected Reference loop. When a reference
loop is deleted, the newer reference loops will shift to the left. The Delete Loop softkey will be
disabled (grayed out) if no reference loops have been saved. The baseline loop cannot be deleted. It
can only be replaced by another baseline loop.
Numeric Data Area: Displays the numerical data associated with a saved Baseline loop and saved
Reference loops. The parameters listed in column form include:
• Time
• Expiratory Minute Volume (Mv Exp)
• Inspiratory Minute Volume (Mv Insp)
• Expiratory Tidal Volume (Vt Exp)
• Inspiratory Tidal Volume (Vt Insp)
• Ratio of Inspiratory time to Expiratory time (I:E)
• Positive End Expiratory Pressure (PEEP)
• Rate
• Peak Inspiratory Pressure (PEAK)
• Plateau Pressure (PLAT)
• Mean Pressure (MEAN)
• Dynamic Airway Compliance (Compl)
• Airway Resistance (Raw)
Last Name
DOB (Date Of Birth) Enter the information from the virtual keypad. If the input is outside the
Age accepted range, a prompt message is displayed. If the age of the patient is less
than 1, the Age will display < 1. These fields are cleared when the case has
Weight (lbs/kg) ended or if the A5 is power cycled.
Bed Enter up to 20 characters per field. When the Restore default settings checkbox
is selected, these fields are NOT cleared when the cas has ended.
Room
Point of Care
Facility
TABLE 3-2 Demographic Tab Fields for Patient and Hospital Data
Demographics Tab
A5 ventilation modes:
• Volume Control Ventilation (VCV)
• Synchronized Intermittent Mandatory Ventilation with VC mode (SIMV-VC)
• Pressure Control Ventilation (PCV)
• Synchronized Intermittent Mandatory Ventilation with PC mode (SIMV-PC)
• Pressure Support ventilation (PS)
• Manual,
• Bypass
• Monitor (with AG module)
1. Select a desired ventilation mode tab. The Set Mode softkey begins to blink green.
2. Optionally, select one or more parameter buttons to change the parameter settings of the
desired ventilation mode. Select the “Accept” button to save each parameter change.
3. Select the “Set Mode” softkey to finalize and change the ventilation mode.
NOTE: If the Set Mode softkey is not selected after several seconds, an
audible reminder is sounded, and then the desired ventilation
mode is cancelled.
The Setup menu contains the General tab, Display tab, System tab, and Service tab. See
section 3.12 (page 3-34) "Display Tab".
If the silenced alarms contain middle or high level alarms, the alarm audio will be paused for 120
seconds.The alarm silence icon and 120 second countdown time appear at the top of the screen.
Select again to resume the alarm audio. Note, however, the alarm will sound if that a new alarm
occurs while the system is in an audio-paused state. If this occurs, you can select the Silence softkey
again to silence the new alarm and reset the silence countdown timer to 120 seconds.
If the silenced alarms are only low level alarms, the alarm audio will be turned off till there is a new
alarm occurs. Note, however, the alarm will sound if that a new alarm occurs while the system is in an
audio-off state. If the new alarm is low level alarm, you can select the Silence softkey again to turn off
the new alarm audio. If the new alarm is medium or high level alarm, you can select the Silence
softkey again to silence the new alarm for 120 seconds.
3.9.5 History
Select the History button on the main screen to access a patient’s historical physiological
parameters. The History dialog contains List Trends, Graphic Trends and an Event Log tab.
There is an interactive link among the three history tabs. When switching between tabs, the cursor
will automatically position itself on the corresponding record that was selected in the previous tab.
The List Trends displays a tabular list of the physiological parameters. Trend data automatically
displays in one minute intervals unless an alternate interval is selected.
BUTTON FUNCTION
Moves the cursor to the oldest record from its current position.
Moves the cursor one page back from its current position.
BUTTON FUNCTION
Moves the cursor one record back from its current position.
Moves the cursor one record forward from its current position.
Moves the cursor one page forward from its current position.
Moves the cursor to the newest record from its current position.
Previous Event Moves the cursor to the previous event from its current position.
Next Event Moves the cursor to the next event from its current position.
Moves the cursor up one parameter from its current position.
Moves the cursor down one parameter from its current position.
TABLE 3-3
Set Display Interval to 1Min, 5Min, 10Min, 15Min, 30Min, 1Hour, or 2Hour.
3.9.5.1.5 List Trend Export (software bundle version 02.06.00 and later)
The Export button on the List Trend tab will allow the contents of the history to be exported to a
USB mass storage device. The format of the data exported is a .html file which can be opened using
Internet Explorer version 6.0, 7.0 and 8.0. The Export button on the List Trend tab is only available
when the system is in Standby mode.
NOTE: If Internet Explorer greater than version 8.0 is used to view the
exported file, set it to compatibility mode.
Graphic trends display allows the user to observe the trend of the physiological parameters. The
trend is reflected through a curve. Every point on the curve corresponds to the parameter value at a
specific time point. Graphic Trends tab displays end case event, captured event and parameter alarm
event. Graphic trend data automatically displays in one minute intervals unless the zoom is selected.
NUMBER DESCRIPTION
1 Event marker. The Dotted,colored line indicate an event occurred at that time.
Events could be the followings: end case, capture an event or a physiological
alarm occurs. When end case or capture an event , the dotted line is white. When
a physiological alarm occurs, the dotted line is in the same color with alarm. If
multiple events occured, dotted line is in same color with the event of the highest
alarm level. The event level can be specified as:high alarm level event>medium
alarm level event>low alarm level event>capture event>end case event.An end
case event occured during this period.
2 Current cursor. The corresponding time displays above the cursor. If alarms or
events occured at that time, the corresponding alarm information will also display
above the cursor (hereinafter referred to as event bubble).
Found in software bundle version 02.06.00 and later:
Clicking on the event bubble will cause the event log tab (see FIGURE 3-36) to
open on that specific event.
3 An end case event occured during this period.
4 The parameter data of the time indicated by cursor.
TABLE 3-4
NOTE: The Graphic Trends will be cleared after the anesthesia machine
undergoes power failure or is turned off.
BUTTON FUNCTION
Moves the cursor to the oldest record from its current position.
Moves the cursor one page back from its current position.
Moves the cursor one record back from its current position.
Moves the cursor one record forward from its current position.
Moves the cursor one page forward from its current position.
Moves the cursor to the newest record from its current position.
Previous Event Moves the cursor to the previous event from its current position.
Next Event Moves the cursor to the next event from its current position.
Moves the cursor up one parameter from its current position.
Moves the cursor down one parameter from its current position.
TABLE 3-5
3.9.5.2.3 Zoom
Zoom allows for the trends to be displayed in a specified time interval between two neighboring
columns.
3.9.5.2.5 Graphic Trends Export (software bundle version 02.06.00 and later)
The Export button on the Graphic Trends tab will allow the contents of the history to be exported
to a USB mass storage device. The format of the data exported is a .html file which can be opened
using Internet Explorer version 6.0, 7.0 and 8.0. The Export button on the Graphic Trends tab is only
available when the system is in Standby mode.
NOTE: If Internet Explorer greater than version 8.0 is used to view the
exported file, set it to compatibility mode.
The Event Log tab logs such events as technical alarms, physiological alarms, capture events, delay
power off, end case, delay power off canceled and system time change. Events can be physiological
indicating that a patients physiological alarm thresholds have been violated or technical indicating
that a specific technical issue has occurred.
An alarm entry and captured events in the Event Log displays the time, date, event, priority and
additional information which includes the Ventilation Mode, Patient Size, and Monitored Parameters.
NOTE: The Event log will not be cleared after the anesthesia machine
undergoes power failure or is turned off.
NOTE: The system can store up to 500 records of Event Logbook. When
a new event occurs after 500 events are already stored, the new
event overwrites the earliest one.
BUTTON FUNCTION
Moves the scroll up one record.
TABLE 3-6
Set Filter to High, Medium, Low, Informational or All. The A5 will display the corresponding event
based on your setting.
NOTE: If Internet Explorer greater than version 8.0 is used to view the
exported file, set it to compatibility mode.
3.10 Setup
Select the Setup softkey (see FIGURE 3-33) to open the Setup menu (see FIGURE 3-37).
The Setup menu contains the General tab, Display tab, System tab, and Service tab. See
section 3.12 (page 3-34) "Display Tab".
NOTE: The Service tab is for use only by Mindray Technical Service.
Please contact Mindray Technical Support for details.
To perform a leak test, select the Test Leak/Compliance button. Follow the on-screen directions and
prompts. See ‘‘Leak and Compliance Tests’’ on page 4-9 for more information. Note that information
for the last Leak Test / Compliance is displayed next to the button.
After cycling power, the breathing system warmer will return to the default state.
NOTE: Before zeroing the flow meters, make sure to disconnect the
gas supply (N2O, Air, O2).
Screen Brightness
To adjust the screen brightness:
Clean Screen
To clean the LCD touch screen:
Calibrate Touch
To calibrate the LCD touch screen:
Pressure Display
To change the pressure display:
4. Select the Accept button to confirm the change, or select Cancel button to disregard the
change.
Plimit Line
The Plimit line function displays a dashed line in the Pressure waveform area to indicate the Plimit
position. The Plimit line can be displayed in VCV, SIMV-VC, and PCV with VG on mode. The Plimit line
function can be switched On or Off by the user. The default value for Plimit Line is On.
Plimit Line
NOTE: The Plimit line does not affect the auto-scaling algorithm. If the
Plimit line is turned on but not visible, it may be because the
line is positioned off the waveform scale.
4. Select the Accept button to confirm the change, or select Cancel button to disregard the
change.
5. If needed, select the Load Scales Defaults button and then select the Yes button to restore
the factory default configurations. Select the Accept button to confirm the change, or select
the Cancel button to disregard the change.
Waveform Display
1. Select Setup softkey > Display tab.
2. Select the Waveform Display button.
3. Select the desired waveform.
4. Select the Accept button to confirm the change, or select Cancel button to disregard the
change.
SYSTEM TAB
BUTTON CHOICES DESCRIPTION
Calibration External AG Module Select to calibrate the External AG
O2 Sensor Module or O2 sensor. Follow the
screen directions. The date and time
of the last calibration is displayed
next to the O2 Sensor or External
AG Module button.
Default Settings Default Patient Size Select to set the default patient size.
(default=Infant, Adult, Pediatric
SYSTEM TAB
BUTTON CHOICES DESCRIPTION
Time Settings Time Zone (Default = UTC-05:00) Select to set the UTC time zone offset.
DST Start (Default =First Sunday in Select to set the START of Daylight
April at 2:00 AM) Savings Time. This setting is not
available if DST is set to Manual.
DST End (Default =Last Sunday in Select to set the END of Daylight
October at 3:00 AM) Savings Time. This setting is not
available if DST is set to Manual.
Network See section 3.13.1 (page 3-41) "Network Configuration".
Set HL7 Most Recent (Default) Select the compatible HL7 version
Compatibility 02.00.00 according to the HL7 version of
01.05.02 clinical information system. If clinical
01.00.00 to 01.05.01 information system is not connected,
None keep the setting as default.
Change Password — Select to change the System tab
password. The authorized
administrator should change the
default password immediately after
the system is installed to prevent
unauthorized access to the System
tab. The password can be up to 6
digits in length containing numerals 0
to 9.
Units Pressure (default=cmH2O, hPa, Select to set the Pressure Unit of
mbar) measure.
SYSTEM TAB
BUTTON CHOICES DESCRIPTION
Configuration — Select to display the machine ID and
Information the status of system functions.
SETTINGS PARAMETERS
This Machine
Configure Ethernet Enter:
• IP Address (default = 192.168.23.250)
• Subnet (default = 255.255.255.0)
• Default Gateway (default = [blank])
SETTINGS PARAMETERS
Configure Serial Select:
• Protocol (None (default), MR-Link/HL7, MR-WATO,
Philips (software bundle version 02.09.00 and later))
• Baud Rate (57600, 11520 (default))
• Data Bits (8 (default), 7, 6, 5)
• Stop Bits (1(default), 2)
• Parity (Odd, Even, None (default))
• Interval:
Enabled when Protocol=None: Off (default);
Enabled when Protocol=HL7:10 Sec, 30 Sec, 1
Min (default), 5 Min, 30 Min, 1 Hour, 2 Hour, 6
Hour, 12 Hour, 24 Hour.
Network Protocol
TABLE 3-11 Network Configuration Settings and Parameters
SETTINGS PARAMETERS
Configure HL7 Interval(10 sec, 30 sec, 1 min (default), 5 min, 30 min,1
hour, 2 hour, 6 hour, 12 hour, 24 hour)
Destination IP (default = 192.168.23.200)
Port (default = 1550)
Set HL7 Compatibility(Most Recent (Default), 02.02.01
to 02.10.00, 02.00.00, 01.05.02, 01.00.00 to 01.05.01,
None)
Send Waveforms(Off(Default), On)
Send Alarms(Off(Default), On)
Send Alarm Ack.(Off(Default), On)
Preoperative Check List (software bundle version 02.09.00 and later)................................................................. 4-17
NOTE: It is recommended that the user check that N2O cutoff and O2/N2O ratio
are normal before use. Use an O2 concentration tester to monitor the O2
concentration in the gas output.
NOTE: Do not use the A5 anesthesia machine if a test failure occurs. Contact
Mindray Technical Support for assistance.
WARNING: Check if the gasket is properly installed in place while installing the
absorber canister. If the gasket is not properly installed (for example,
gasket is not evenly seated and centered) it may cause breathing
system leakage.
An effective method for detecting pneumatic circuit occlusions, leaks, and other system problems
can be found in the A5 pre-operative checkout procedures. In addition, it is recommended that the
breathing circuit be tested for the ability to effectively deliver positive pressure ventilation before
beginning each case. Testing the ability to properly ventilate a test lung can quickly identify an
occluded circuit limb and other breathing circuit problems.
Before starting each case, test the machine’s ability to ventilate the patient by removing the
breathing bag from the bag arm and connecting it to the patient connection (elbow or Y-piece on
the disposable circuit). Set the ventilator to deliver a specific tidal volume to the test lung and verify
the exhaled tidal volume monitor. Observe that the test lung (breathing bag) inflates as the bellows
descends, and that the test lung deflates during the exhalation phase. Observe that the measured
exhaled volume matches the tidal volume set on the ventilator. With the ventilator running, lower the
fresh gas flow to zero and observe if the bellows rapidly falls with each exhalation. If this occurs, then
a leak should be suspected, identified, and repaired.
This test should be performed before starting each case. By verifying that a test lung (breathing bag)
can be manually and mechanically ventilated, this indicates that the A5 is capable of ventilating a
patient with the attached breathing circuit.
WARNING: To ensure proper machine operation, user safety, and patient safety,
follow all checkout procedures established by the facility before
administering anesthesia to the patient.
1. With the anesthesia machine connected to AC Power, turn the Mains switch to ON and verify
that the unit is operating on AC. Follow the on-screen prompts to perform and complete the
automatic machine start-up tests.
2. a. Check the O2 Supply fail-safe message and alarm.
(See “O2 Pipeline Test” on page 4-18.)
b. Test low O2 concentration alarm.
(See “Test the O2 Concentration Monitoring and Alarms” on page 4-26.
c. Test high and low airway pressure alarms.
(See “Test the High Paw Alarm” on page 4-27.)
(See “Test the Low Paw Alarm” on page 4-28.)
d. Test low minute volume and apnea alarms.
(See “Test the Low Minute Volume (MV) Alarm” on page 4-27.)
(See “Test the Apnea Alarm” on page 4-27.)
3. Verify that the O2 sensor displays approximately 21% in room air and above 94% after
exposure to 100% O2 (see“Test the O2 Concentration Monitoring and Alarms” on page 4-26).
4. Check that the vaporizers are properly installed and sufficiently filled and that filler ports are
tightly closed. Verify that only one vaporizer turns ON at a time (see “Install the Vaporizer” on
page 2-5).
5. Perform a 40 cmH2O manual leak test. If present, set the left vaporizer to ON and perform a
40 cmH2O manual leak test. Set the vaporizer to OFF. Repeat for the right vaporizer if
installed (see “Manual Leak Test” on page 4-22).
6. Perform a vaporizer leak test for each vaporizer installed on the A5 system (see “Manual Leak
Test” on page 4-22).
7. Check that the function of Anesthetic Gas Scavenging System is normal (see “Inspect the
Active/Passive Anesthetic Gas Scavenging System” on page 4-29).
8. Drain any moisture from the breathing system water trap.
9. Drain and wipe with a soft cloth out any moisture from the condensation drain valve at the
bottom of the absorber canister assembly.
Prior to each patient, before administering anesthesia, the following should be done:
1. Inspect the A5 for damage or hazardous conditions; ensure all necessary equipment and
supplies are present, e.g., drugs, CO2 absorbent (not exhausted), breathing circuits and tank
wrench.
2. Check that central supply O2, N2O and Air pressures are each within the pipeline input range
specifications (i.e., 40 to 87 psi).
3. Check that O2, N2O and Air flowmeters operate properly: Check that all flow levels on the
monitor screen are at zero flow with flow-control valves closed. Adjust flow of all gases
through their full ranges and check for erratic movements of the gas levels.
4. Check that a hypoxic mixture of less than 21% O2 may not be administered: Attempt to
create an hypoxic O2/ N2O mixture by slowly opening the N2O flow control valve fully with
the O2 flow valve fully closed (no N2O gas should be flowing). Then, slowly open the O2 flow
valve and observe O2 and N2O rise in proportion to maintain a minimum concentration of
21% O2 in fresh gas.
5. Perform a vaporizer leak test for each vaporizer installed on the A5 system (see “Manual Leak
Test” on page 4-22).
6. Verify that Auxiliary O2 and Air are available and functioning.
7. Verify that a Self-inflating Manual Ventilation device is available and functioning.
8. Check that the O2, N2O, and Air cylinders (if present) are mounted on the A5, have adequate
pressure, and no high pressure leaks are present (see “Cylinder Tests” on page 4-20).
9. Check that valves on the O2, N2O, and Air cylinders (if present) are closed until needed to
prevent unintentional use of gas.
10. With a breathing circuit and reservoir bag attached, check that the unidirectional valves
operate by visual inspection.
11. Check ventilation capability in Standby, Manual, VCV and PCV ventilator modes.
12. Check that patient suction is adequate to clear the airway.
13. Verify ability of required monitors and check alarms.
The following step is recommended to be performed when prompted by the machine:
• Complete the 21% O2 Calibration (see “O2 Sensor Calibration” on page 7-6).
• Complete the 21% and 100% O2 Calibration (see “O2 Sensor Calibration” on page 7-6).
• Perform a vaporizer leak test (see “Manual Leak Test” on page 4-22).
After the system self-test is completed, the test results are displayed on the screen. Startup
alarm messages also may be displayed.
RESULT COMMENTS/OPTIONS
Pass condition The Pass condition indicates that the A5 has passed the
Example: System Self-Test. No errors have been detected. Alarms
and hardware are functioning properly.
All-Functional error condition The All-Functional error condition indicates that errors
Example: have been detected. However, all automatic ventilation,
manual, and bypass modes are still enabled.
Manual Only error condition The Manual Only error condition indicates that the A5 can
Example: be used in manual mode only.
RESULT COMMENTS/OPTIONS
Machine Non-Functional error condition The Machine Non-Functional error condition indicates
Example: that the A5 cannot be used.
Bundle Version – The Bundle Version is displayed in all System Self-Test results. The Bundle Version is
the version number of the package of software that is installed in the A5. If the Bundle Version
displays a fail status, contact Mindray Technical Support.
FIGURE 4-1 Automatic Circuit Leak Test (software bundle version 03.13.00 and later)
NOTE: If fresh gas is detected by the system before proceeding with the
Automatic Circuit Leak & Compliance Test, a message is displayed on
the screen to adjust all flowmeters to zero.
or
2. Ensure that the sample line port of the breathing circuit is occluded.
3. Install the manual bag.
4. Adjust all flowmeters to zero.
5. Set the Auto/Manual switch to the Auto position:
6. Ensure the CO2 Absorber Canister is closed and locked (software bundle version 03.13.00
and later).
NOTE: The “Continue” button can be selected only when the Auto/Manual
switch is set to the Auto position and when no fresh gas is detected.
3. Compare the test results with the information in TABLE 4-3, “Automatic Circuit Leak and
Compliance Test Results,” on page 4-11, and proceed accordingly.
RESULTS COMMENTS/OPTIONS
Automatic Circuit Leakage: Pass Leak rate ≤200 mL/min
Compliance Test: XX.X mL/cmH2O Compliance test results are displayed in green.
Example:
Select Continue to proceed to the Manual Circuit Leak
Test screen.
Automatic Circuit Leakage: XXX mL/min Leak rate >200 mL/min and ≤1000 mL/min
Compliance Test: Fail
Example: The results screen displays the compliance values and
time of the last compliance test that passed. If the
compliance test has never been performed successfully,
the compliance values and test time are displayed as ---.
RESULTS COMMENTS/OPTIONS
Automatic Circuit Leakage: Fail: Fresh gas Fresh gas is detected. Approximate threshold for fresh gas
flow detected detection is 0.15 L/min of gas flow.
Compliance Test: Fail
Example: Adjust all flowmeters to zero.
Select Retry to repeat the test.
RESULTS COMMENTS/OPTIONS
MACHINE NON-FUNCTIONAL Safety valve control test or pressure verification test failed.
Automatic Circuit Leakage: Pass Select Retry to repeat the Automatic Circuit Leak Test &
Compliance Test: XX.X mL/cmH2O Compliance Test.
Safety Valve Control: Fail or
Example: Contact service if this error condition persists.
or
or
NOTE: The “Continue” button can be selected only when the Auto/Manual
switch is set to the Manual position and when no fresh gas is detected.
3. Compare the test results with the information in TABLE 4-4, “Manual Circuit Leak Test
Results,” on page 4-15, and proceed accordingly.
RESULTS COMMENTS/OPTIONS
Pass Manual Circuit Leak Test passed.
Example:
Adjust the APL valve to SP position.
Select Continue to proceed to the main screen.
Fail: Fresh gas flow detected Manual Circuit Leak Test failed.
Example: Fresh gas is detected.
RESULTS COMMENTS/OPTIONS
Fail Manual Circuit Leak Test failed.
Example: Fresh gas is not detected.
WARNING: Incorrect gas mixtures can cause patient injury. If the O2:N2O ratio
system does not supply O2 and N2O in the correct proportions, do not
use the system.
CAUTION: Slowly open the cylinder valves to avoid damage. Do not use excessive
force on the flow controls. After performing the cylinder tests, close all
cylinder valves if cylinder supplies are not used.
CAUTION: Turn the flow controls slowly. To avoid damaging the control valves, do
not turn further when the flowmeter reading is outside the range.
When turning a flow control knob clockwise to decrease flow, the
flowmeter should reach zero before the knob reaches its most
clockwise mechanical stop (Off) position. Do not turn any further when
the knob has reached the Off position.
WARNING: To prevent damage, turn the flow controls fully clockwise (flow OFF)
before using the system.
Before the test, ensure that the vaporizers are correctly installed. For details about vaporizer
installation, see ‘‘Install the Vaporizer’’ on page 2-5.
NOTE: Do not perform this test on the vaporizer when the concentration
control is between “OFF” and the first graduation above “0” (zero) as
the amount of anesthetic drug outputted is very small within this
range.
3. Connect one end of the breathing circuit to the bag arm, one end to the inspiratory port and
the Y-piece to the test port:
4. Mount and lock the vaporizer onto the vaporizer mount. (Certain vaporizers need to be set
to at least 1% for correct testing. See the vaporizer manufacturer’s manual for details.)
5. Set the fresh gas flow to 200 mL/min.
6. Set the APL valve to 75 and verify that the pressure on the airway pressure gauge increases
above 30 cmH2O within 2 minutes.
7. Turn off the vaporizer.
8. Repeat Steps 4, 5, 6, and 7 for the other vaporizer.
WARNING: Do not use a test plug that is small enough to fall into the breathing
system.
1. Ensure that the breathing system is correctly connected and not damaged.
2. Ensure that the check valves in the breathing system work correctly:
1. The inspiratory check valve opens during inspiration and closes at the start of expiration.
2. The expiratory check valve opens during expiration and closes at the start of inspiration.
8. Connect the manual bag to the manual bag port. Push the O2 flush button to fill the manual
bag. Ensure that the sensor measures at least 90% O2.
9. Ensure that a high O2 alarm (“FiO2 Too High”) occurs.
10. Set the FiO2 high alarm limit to 100% and ensure that the alarm cancels.
NOTE: Do not block the active AGSS pressure compensation openings during
the inspection. If the float cannot rise, the possible reasons are:
1. The float surface is tacky. Turn over the active AGSS and check if the
float moves up and down freely.
2. The float is rising slowly. The filter may be blocked. Check if the filter
is blocked.
3. The waste gas disposal system is not working or the pump rate is less
than 50 L/min at which the active AGSS works normally. Check the
waste gas disposal system.
NOTE: The knob on the top of the scavenger is meant to adjust the flow from
the EVAC. When the knob is fully closed it does not need to completely
shut off flow.
WARNING: Before using the A5 Anesthesia System on the patient, ensure that the
system is correctly assembled and in good condition, and that all the
tests described in the Preoperative Test are already completed. In case
of test failure, do not use the system. Have a qualified Mindray service
representative repair the system.
• A prompt sound is given when user turns off the A5. If the power switch is turned off in
Standby mode, the A5 will immediately power off.
• If the power switch is turned off in Manual mode or in any of the Automatic ventilation
modes, the A5 will wait 12 seconds to power off completely. In the 12-second power off
delay period, the screen will display a 10 second countdown timer. If the A5 is performing
Automatic ventilation, the ventilator will continue ventilating the patient in the current
ventilation mode.
• A beep is sounded for each second of the countdown from 10 to 1 second, after which a
two-second shutdown sound is given when the timer reaches zero.
• The volume of power off delay sound can be adjusted in the System Alerts setting in the
Alarm Volume menu.
• When the user turns on the machine during the power off delay period, the countdown
timer will disappear, and the ventilator will resume its previous state.
NOTE: The powering off function is not implemented during Standby, only
when actively ventilating.
Ending the case changes the current patient size to the default patient size and loads the user
defaults for the system; clears the patient demographics; clears the User Alarm Log and
Spirometry Loops (including the currently plotting loop, reference loop, and baseline loop); and
places the system into Standby mode (see FIGURE 5-4).
After the End Case button is selected, a warning box with a Restore default settings checkbox
will be displayed. Selecting the Restore default settings checkbox reloads the user defaults, clears
patient demographics, clears the history, clears the spirometry reference loops, and places the
system into Standby mode (see FIGURE 5-4).
If the Restore Default settings checkbox is not selected, all the settings are retained.
In Standby, all system functions are idle. It is the default system startup mode and is used after
ending the case.
NOTE: The A5 system will not allow the End Case button to be selected until
the Auto/Manual ventilation switch is set to Manual, and system
detects the individual fresh gas flows are effectively turned off (i.e.,
flow < 0.05 L/min).
3. Select the End Case button in the Manual tab (see FIGURE 5-2).
4. Follow the screen prompts to end the case and enter Standby mode.
5. To exit Standby, set the Auto/Manual ventilation switch to Manual, then touch the screen
or turn on the fresh gas flow to more than 0.2 L/min of individual gas.
NOTE: To exit Standby by turning on the fresh gas flow, the flow must be
increased to more than 0.2 L/min.
NOTE: The End Case button can be selected only when the system is not in
Standby, all fresh gas flows are off, and the Auto/Manual switch is in the
Manual position.
NOTE: When the system is in Standby mode, the Bypass,Monitor and End Case
buttons in the Manual tab are disabled. However, the Alarms button
remains enabled and can be toggled to On or Off.
WARNING: Selecting End Case to enter Standby mode will stop ventilation and
parameter monitoring. Do not select Standby mode if the patient
requires continuous ventilation.
1. Select Manual mode or select the End Case button (in the Manual tab) to enter Standby
mode.
2. Select the Patient Size softkey at the top left of the main screen. The softkey displays
“Adult”, “Pediatric”, or “Infant”.
3. Select the Patient Size: Adult, Pediatric, or Infant.
4. Select the Accept softkey to finalize your selection.
NOTE: The A5 saves the latest patient parameter settings (VCV, PCV, PCV-VG ,
PS, SIMV-VC, SIMV-PC , and Alarms) for each patient type: Adult,
Pediatric, and Infant. Changing to another patient type does not erase
the parameter settings from the previous patient type. For example,
changing from Adult to Pediatric and back to Adult will result in the
Adult patient parameter settings still saved.
NOTE: You can adjust O2 concentration in the breathing system through the
O2 flow control.
NOTE: The total flowmeter is calibrated based on 100% O2. The accuracy of the
flowmeter may degrade with other gas or mixed gas.
NOTE: When viewing the readings on the total flowmeter, keep your visual
angle at the same level of the float. The reading of the scale may vary
when viewed at a different angle.
NOTE: If the readings shown on the electronic flowmeters differ from that on
the total flowmeter, the electronic flowmeter will prevail and the total
flowmeter is an approximate value.
NOTE: When the AC power supply is not connected and batteries are depleted,
the flow and the composition of the fresh gas are not affected. When
the individual N2O or Air supply fails, the corresponding fresh gas
cannot be achieved. When O2 supply fails, both O2 and N2O fresh gas
cannot be achieved.
NOTE: Install the vaporizers with a Selectatec interlock system that are
compliant to ISO 80601-2-13 on the A5 unit. Refer to the manufacturer’s
vaporizer Instructions For Use for filling or draining the vaporizer and
other information.
WARNING: Ensure that the correct anesthetic agent is used. The vaporizer is
designed with the specific anesthetic agent named on it and further
indicated by color coded labeling. The concentration of the anesthetic
agent actually output will vary if the vaporizer is filled with the wrong
agent.
2. Mount the vaporizer filled with anesthetic agent onto the A5 Anesthesia System. See “Install
the Vaporizer” on page 2-5.
NOTE: When the drive gas supply fails, mechanical ventilation cannot work
normally.
1. Select the tab of the desired ventilation mode. The “Set Mode” button (or “Preset Mode”
button in manual) will flash (see FIGURE 5-6).
2. Select the “Set Mode” button (or “Preset Mode” button in manual) to confirm.
If the “Set Mode” button is not selected after several seconds, an audio reminder will sound
for several seconds and then the system will return to the previous ventilation mode.
3. Optionally, select each available ventilation parameter to edit the parameter setting.
4. Move the Auto/Manual Bag switch to the Auto position.
NOTE: When the Auto/Manual switch is in Auto position, all the buttons in
Manual tab (Alarms, Bypass, Monitor and End Case) are disabled;
Alarms are set to On; and Bypass is set to Off.
When the Auto/Manual switch is set to Manual, and the Alarms button in the Manual mode
tab is set to Off, the alarm limit indicators on the main screen to the right of the measured values
related to Pressure and Volume (such as PEAK and MV) will change to Off (see FIGURE 5-7).
The Alarms button setting (On/Off) in the Manual mode tab is saved and restored when
toggling from Manual to Auto and back to Manual mode. For example, if the Alarms button is
set to Off, this setting will be saved and restored to Off after switching to Auto and back to
Manual mode.
The patient can be ventilated by hand using the breathing bag. The pressure will be limited to
the value set on the APL valve.
NOTE: In the manual ventilation mode, you can use the APL valve to adjust the
breathing system pressure limit and gas volume in the manual bag.
When the pressure in the breathing system reaches the pressure limit
set for the APL valve, the valve opens to release excess gas.
NOTE: The APL valve adjusts the breathing system pressure limit during
manual ventilation. Its scale shows approximate pressure.
NOTE: When Bypass mode is On, the Alarms button is disabled and set to Off.
Enter Cardiac Bypass mode by setting the Bypass softkey in Manual mode to On. When the
Bypass softkey is set to On, the Alarm softkey is disabled and set to Off automatically. When
Bypass is set to Off, the Alarm button returns to its setting before entering Bypass. When
exiting Manual mode or discharging a patient, Bypass will be set to Off.
NOTE: When Monitor mode is On, the Alarms button is disabled and set to Off.
Enter the Monitor mode by setting the Monitor softkey in Manual mode to On. When the
Monitor softeky is set to On, the Alarm softkey is disabled and set to Off automatically. When
Monitor is set to Off, the Alarm button restores to its settings before entering the Monitor mode.
When exiting Manual mode or discharging a patient, Monitor will be set to Off.
When the system is working in Monitor mode, the flow, volume and pressure waveforms and
measured values are removed from the Waveforms tab. Only the CO2 waveform and the CO2
parameters will remain on the Waveforms tab. The Rate as determined by the AG module is
displayed in the measured values area.
Setting Alarms
In Manual ventilation mode, when Bypass and Monitor are set to Off, the pressure, volume and
apnea alarms can be turned off by setting the Alarms softkey to Off. The related alarm limits are
then displayed as Off.
Pressure, volume and apnea alarms can be turned on by setting the Alarms softkey to On, which
returns the related alarm limits to their original settings.
The CO2 and the CO2 apnea alarms can be turned on by setting the CO2 Alarms softkey to On or
by switching to mechanical ventilation mode which returns the related alarm limits to their
original settings.
When the system exits standby mode and the CO2 Alarms softkey is On, the system will not
activate the CO2 and the CO2 apnea alarms until three continuous CO2 waves are monitored.
The CO2 and the CO2 apnea alarms are disabled for 30 seconds when the ventilation mode is
switched from Manual to Auto or when the CO2 Alarms softkey is set from Off to On. After 30
seconds, the CO2 and the CO2 apnea alarms would be enabled even if CO2 has not been
detected.
In VCV mode, you need to set Plimit to prevent high airway pressure from injuring the patient. In
this mode, you can select to set Tpause to improve patient pulmonary gas distribution and PEEP
to improve expiration of end-tidal carbon dioxide and to increase oxygenation of breathing
process.
To ensure the set tidal volume gas delivery, the ventilator adjusts gas flow based on the
measured inspiratory volume, dynamically compensates for the loss of tidal volume arising from
breathing system compliance and system leakage and eliminates the effect of fresh gas as well.
This is called tidal volume compensation.
In the VCV mode, if tidal volume compensation has failed, the A5 Anesthesia System can
continue delivering gas stably but cannot compensate for the effects of fresh gas flow and
breathing system compliance losses.
In VCV and SIMV-VC modes, when inspiration pressure reaches Plimit, respectively, the
inspiration pressure is held.
NOTE: The screen displays the calculated Tinsp when adjusting the I:E ratio
(software bundle version 02.06.00 and later).
NOTE: Before activating a new mechanical ventilation mode, ensure that all
related parameters are set appropriately.
In the PCV mode, you can set PEEP to improve expiration of end-tidal carbon dioxide and to
increase oxygenation of breathing process.
For the A5, in PCV mode, Tidal Volume Guarantee (VtG) can be enabled with the VtG setting.
When VtG is a value, then Pinsp is disabled. The ventilator attempts to deliver the set VtG while
maintaining the PAW at or below PlimVG. When VtG is Off, PlimVG is disabled and Pinsp is
enabled. Changing the value of Pinsp will automatically set PlimVG to the same value, but PlimVG
can be adjusted without affecting the value of Pinsp.
NOTE: In PCV mode, even when the PlimVG or Pinsp parameters are inactive,
they are restricted to the parameter relationship equations
PlimVG≥PEEP+5 and Pinsp≥PEEP+5. See section C.9 (pg. C-15)
"Ventilation Parameter Relationships".
NOTE: The screen displays the calculated Tinsp when adjusting the I:E ratio
(software bundle version 02.06.00 and later).
NOTE: Before activating a new mechanical ventilation mode, ensure that all
related parameters are set appropriately.
In VCV and SIMV-VC modes, when inspiration pressure reaches Plimit, the inspiration pressure is
held.
NOTE: The screen displays the calculated I:E ratio based on Rate and Tinsp
when adjusting the Tinsp (software bundle version 02.06.00 and later).
SIMV-PC parameters:
NOTE: The screen displays the calculated I:E ratio based on Rate and Tinsp
when adjusting the Tinsp (software bundle version 02.06.00 and later).
NOTE: Before activating a new mechanical ventilation mode, ensure that all
related parameters are set appropriately.
The user can set the Trigger flow, ΔP, PEEP, minimum allowed breathing frequency, and Slope
Time. If the Min Rate (bpm) is violated, the A5 will give an Apnea Ventilation breath to assure
ventilation is occurring.
NOTE: Apnea Ti permits the user to vary the inspiratory time of the apnea
backup breaths. Apnea backup breaths are only triggered when the
patient does not achieve the Min Rate that is set by the user. If the
patient’s spontaneous breaths meet or exceed the Min Rate, the apnea
backup is not used.
NOTE: Before activating a new mechanical ventilation mode, ensure that all
related parameters are set appropriately.
1. Ensure that the breathing system is set up and the APL valve is set properly before stopping
mechanical ventilation.
2. Set the Auto/Manual Bag switch to the Manual Bag position. This selects manual
ventilation and stops mechanical ventilation.
Ventilation parameter values that are non-linked are set according to relationship equations.
Section C.9 ‘‘Ventilation Parameter Relationships’’ on page C-15 includes a table of equations to
show how non-linked parameter values are set when changing ventilation modes.
5.10.1 Pressure
The Pressure parameter group consists of 3 parameters (see FIGURE 5-17):
NOTE: The high alarm limit for Airway Peak Pressure (PEAK) is displayed to the
top right of the reading. The low alarm limit for Airway Peak Pressure
(PEAK) is displayed to the bottom right of the reading.
NOTE: The display of either Plateau Pressure (PLAT) or Mean Pressure (MEAN)
is configured from the System menu tab.
5.10.2 Volume
The Volume parameter group consists of 3 parameters (see FIGURE 5-18):
NOTE: The high alarm limit for Minute Volume (MV) is displayed to the top
right of the reading. The low alarm limit for Minute Volume (MV) is
displayed to the bottom right of the reading.
• Fraction of inspired carbon dioxide and End-tidal carbon dioxide(FiCO2 and EtCO2)
• Fraction of inspired oxygen and End-tidal oxygen(FiO2 and EtO2)
• Fraction of inspired nitrous oxide and End-tidal nitrous oxide(FiN2O and EtN2O)
• Fraction of inspired anesthetic agent and End-tidal anesthetic agent(FiAA and EtAA, AA
stands for anesthetic agent)
• Minimum alveolar concentration(MAC)
• Age
If the parameter data is out of range, it is displayed as “– – –”.
NOTE: The high alarm limit is displayed to the top right of the reading. The low
alarm limit is displayed to the bottom right of the reading.
FiO2 values above 100%, although not realistic, are possible due to errors in calibration.
NOTE: The high alarm limit is displayed to the top right of the reading. The low
alarm limit is displayed to the bottom right of the reading.
The Y-axis of the Pressure vs. Time waveform is labeled Paw (which represents Airway Pressure).
The unit of measure is cmH2O, hPa, or mbar. The Y-axis can automatically adjust the
scalesThough the X-axis is not labeled, it represents a time scale of 0 to 15 seconds.
The auto-zeroing intervals are: startup, 5 mins, 15 mins, 30 mins, 60 mins, and every 120 mins
thereafter.
NOTE: The A5 will display the message “Auto-zeroing in process” during the
auto-zeroing intervals.
The Y-axis of the Flow vs. Time waveform represents Flow. The unit of measure is L/min. The Y-
axis can automatically adjust the scales. Though the X-axis is not labeled, it represents a time
scale of 0 to 15 seconds.
The Y-axis of the Volume vs. Time waveform is labeled Volume . The unit of measure is ml. The Y-
axis can automatically adjust the scales. Though the X-axis is not labeled, it represents a time
scale of 0 to 15 seconds.
The Y-axis of the CO2 vs. Time waveform is labeled CO2 . The unit of measure is mmHg, kPa, or %.
The Y-axis can automatically adjust the scales. Though the X-axis is not labeled, it represents a
time scale of 0 to 15 seconds.
NOTE: N2O, O2, and AA waveforms are available for software bundle version
02.06.00 and later.
The N2O vs. Time waveform can be displayed in the waveform area.
FIGURE 5-25 Example Simulated N2O vs. Time Waveform (available with the AG module)
The Y-axis of the N2O vs. Time waveform is labeled N2O . The unit of measure is %. You can adjust
the scales of the Y-axis (See “Gas Scales (software bundle version 02.06.00 and later, with an AG
module connected)” on page 3-37.). Though the X-axis is not labeled, it represents a time scale of
0 to 15 seconds.
FIGURE 5-26 Example Simulated O2 vs. Time Waveform (available with the AG module)
O2 vs. Time
The Y-axis of the O2 vs. Time waveform is labeled O2 . The unit of measure is %. You can adjust the
scales of the Y-axis (See “Gas Scales (software bundle version 02.06.00 and later, with an AG
module connected)” on page 3-37.). Though the X-axis is not labeled, it represents a time scale of
0 to 15 seconds.
FIGURE 5-27 Example Simulated AA vs. Time Waveform (available with the AG module)
AA vs. Time
The Y-axis of the AA vs. Time waveform is labeled AA . The unit of measure is %. You can adjust
the scales of the Y-axis (See “Gas Scales (software bundle version 02.06.00 and later, with an AG
module connected)” on page 3-37.). Though the X-axis is not labeled, it represents a time scale of
0 to 15 seconds. If no agent is detected, the system displays AA vs. Time waveform (see FIGURE 5-
27). If an anesthetic agent such as sevoflurane is detected, the system displays Sev vs. Time
waveform (see FIGURE 5-28)..
FIGURE 5-28 Example Simulated Sev vs. Time Waveform (available with the AG module)
If the measured values of Paw, Flow, or Volume are less than the boundary minus a margin (see
TABLE 5-1) at the end of two continuous breath cycles, the system will autoscale the Paw, Flow, or
Volume at the beginning of the next breath cycle.
SCALE MARGIN
Paw 10 cmH2O if PAW ≥ 30 cmH2O
3 cmH2O if PAW < 30 cmH2O
Flow 10 L/min if Flow ≤ 30 L/min
15 L/min if Flow > 30 L/min
Volume 25 mL if volume ≤ 100 mL
100 mL if volume > 100 mL
TABLE 5-1 Autoscaling Margins of Paw, Flow, and Volume
Currently plotting loop, reference loop, and baseline loop can be displayed in manual and
mechanical ventilation modes.
End the case will clear spirometry loops (baseline and reference loops).
NOTE: This will not occur when the Restore default settings box is unselected .
Restart the machine will clear spirometry loops (baseline and reference loops).
Spirometry is disabled in Bypass mode. If Bypass mode is entered when the Spirometry tab is
open, then the system will switch to the Waveforms tab.
The Y-axis of the Pressure-Volume Spirometry loop represents Volume. The X-axis is labeled Paw
(which represents Airway Pressure).
The Y-axis of the Flow-Volume Spirometry loop represents Flow. The X-axis represents Volume.
The Y-axis of the Pressure - Flow Spirometry loop labeled PAW represents airway pressure. The X-
axis represents Flow.
Introduction .................................................................................................................................................................................................6-2
Displaying Alarms.....................................................................................................................................................................................6-5
Silencing Alarms........................................................................................................................................................................................6-8
6.1 Introduction
The A5 System provides alarms and messages that are indicated to the user by visual and audible
alerts. Alarms and messages appear at the top of the Main Screen and in the Alarms window (see
FIGURE 6-1). Users can adjust alarm properties, which include setting alarm limits to trigger alarm
conditions, adjusting alarm volume, and silencing alarms.
FIGURE 6-1 Alarms and Messages On The Main Screen and In The Alarms Window
• During the self-test, the alarm LED will illuminate in sequence with the colors red, yellow, and
cyan for approximately 1 second each color.
• The system speaker will produce one tone after the alarm light is in self-test.
• Physiological Alarm:
This is an alarm caused by a patient-related variable. It requires a response from the user. It can
have the following priority: high, medium, or low.
• Technical Alarm:
This is an alarm caused by a machine-related variable. It requires a response from the user. It can
have the following priority: high, medium, or low.
• Prompt Message:
This is a message to the user. It does not require a response from the user. It always has the lowest
priority, below Physiological and Technical alarms. It is displayed in black text on white
background.
• An alarm LED located on top of the LCD monitor. The LED can illuminate red, yellow, cyan, or
OFF depending on the alarm condition.
TABLE 6-1 describes the alarm behavior of different alarm types and different alarm priority labels.
If multiple alarms occur simultaneously, the audio and LED behavior will follow the highest
priority active alarm.
• Colored alarm messages displayed on the Main Screen. High priority messages are red.
Medium priority messages are yellow. Low priority messages are cyan. Prompt messages are
black text, white background. Messages are displayed according to priority and time. (See
‘‘Displayed Order of Alarm Messages’’ on page 6-6.)
• Alarm audio through the system alarm speaker. TABLE 6-1 lists the audio behavior for each
type of alarm.
ALARM
ALARM LED
ALARM TYPE PRIORITY AUDIO BEHAVIOR MESSAGE BEHAVIOR COLOR
Physiological High Play high priority alarm white text red Red
Alarm sound, the interval between background, high
each play is 5 ± 1 sec. priority icon.
Low Play low priority alarm sound, white text cyan Cyan
the interval between each background, low
play is 17 ± 1 sec. priority icon.
Technical High Play high priority alarm white text red Red
Alarm sound, the interval between background, high
each play is 5 ± 1 sec. priority icon.
Low Play low priority alarm sound, white text cyan Cyan
the interval between each background, low
play is 17 ± 1 sec. priority icon.
• High priority
• Medium priority
• Low priority
1. On the Main Screen, select the Alarms softkey or touch the Alarm Message area at the top
of the screen.
The Alarms windows is displayed.
2. Select the Active tab.
A list of all active alarm messages is displayed (see FIGURE 6-2). Up to 15 current alarms can
be displayed on screen, after which a scroll bar is used to display the remaining alarms.
Alarms are displayed in order of priority and time. See section 6.2.1 (page 6-6) "Displayed
Order of Alarm Messages" for more information.
Alarm messages are displayed in Area A and Area B according to the following rules:
• To be in Area A, an alarm must be both the highest priority AND the most recent (Area A does not
cycle). The remaining active alarms and prompt messages cycle in Area B.
• New Alarms with less priority than alarms in Area A are displayed immediately in Area B, and the
cycle proceeds from that position in the list.
• Alarms cycling in Area B are grouped and displayed in the following order: highest priority,
medium priority, low priority, and prompt messages. In each group, the most recent alarm is
displayed first.
• If the alarm in Area A is removed, then the most recent and highest priority alarm from Area B is
moved to Area A.
The Alarms volume settings adjust the audio level of all High, Medium, and Low Priority sounding
alarms. The System Alerts volume settings adjust the audio level of all sounding pop-up prompts
and non-confirmed ventilation mode alerts.
WARNING: Do not rely exclusively on the audible alarm system when using the A5
Anesthesia System. Adjustment of alarm volume to a low level may
result in a hazard to the patient. Always keep the patient under close
surveillance.
NOTE: The auditory alarm signal A-weighted sound pressure level is within 45
to 85 dB.
Select Silence softkey to silence all currently sounding alarm tones. The alarm will sound if a new
alarm occurs.
If the silenced alarms contain middle or high level alarms, the alarm audio will be paused for 120
seconds.The alarm silence icon and 120 second countdown time appear at the top of the screen.
Select again to resume the alarm audio.
NOTE: The alarm will sound if that a new alarm occurs while the system is in an
audio-paused state. If this occurs, you can select the Silence softkey
again to silence the new alarm and reset the silence countdown timer to
120 seconds.
If the silenced alarms are only low level alarms, the alarm audio will be turned off till there is a new
alarm occurs.
NOTE: The alarm will sound if that a new alarm occurs while the system is in an
audio-off state. If the new alarm is low level alarm, you can select the
Silence softkey again to turn off the new alarm audio. If the new alarm
is medium or high level alarm, you can select the Silence softkey again
to silence the new alarm for 120 seconds.
NOTE: When using the A5 Anesthesia System, ensure that the alarm limits of
each parameter are set to the appropriate values for the patient.
1. On the main screen, select the Alarms softkey. The Alarms windows is displayed.
Alarm Softkey
or
When the monitoring value on the main screen is flashing , select the flashing area to open
the Alarms window with the currently alarming parameter selected.
2. Select the Limits tab or Agents tab. (see FIGURE 6-7, FIGURE 6-8 and FIGURE 6-9.)
3. Select a parameter softkey.
The softkey is highlighted when selected.
4. Use the on-screen keypad to enter the desired parameter value, or continuously press
FIGURE 6-7 Limits tab in the Alarms Window (without AG module connected)
FIGURE 6-8 Limits tab in the Alarms Window (with AG module connected)
FIGURE 6-9 Agents tab in the Alarms Window (with AG module connected)
FIGURE 6-10 Load Alarm Defaults button in the Alarms window (without AG module
connected)
FIGURE 6-11 Load Alarm Defaults button in the Alarms window (with AG module
connected)
FIGURE 6-12 Load Alarm Defaults button in the Agents window (with AG module
connected)
The Auto Alarm Limits button is disabled when the A5 is in Standby mode, Manual mode or Monitor
mode (with AG module). The Auto Alarm Limits button is also disabled when the current mode is PS,
SIMV-VC, or SIMV-PC.
If there is not a valid measured minute volume (MV), the corresponding MV alarm limits will not be
adjusted.
If the average value for PEAK, PLAT, and PEEP cannot be calculated, the corresponding alarm limits
will not be adjusted.
If the calculated alarm limit is more than the high threshold of setting range or less than the low
threshold, the corresponding threshold is used as the auto alarm limit.
For each alarm message, corresponding actions are given instructing you to troubleshoot problems.
If the problem persists, contact your service personnel.
NOTE: The Disable in Manual and Cardiac Bypass mode column indicates how
this alarm is controlled by the alarm on/off button and the cardiac
bypass mode button in manual mode.
DISABL
DISABLED DISABLED ED IN
WHEN IN MONIT
ALARM ALARM IS STANDBY OR
MESSAGE CAUSE PRIORITY OFF MODE MODE
EtCO2 Too EtCO2 > high alarm limit Medium No Yes No
High setting
EtCO2 Too Low EtCO2 < low alarm limit Medium No Yes No
setting
FiCO2 Too High FiCO2 > high alarm limit Medium No Yes No
setting
EtN2O Too EtN2O > high alarm limit Medium No Yes No
High setting
EtN2O Too Low EtN2O < low alarm limit Medium No Yes No
setting
FiN2O Too FiN2O > high alarm limit Medium No Yes No
High setting
FiN2O Too Low FiN2O < low alarm limit Medium No Yes No
setting
EtHAL Too EtHAL > high alarm limit Medium No Yes No
High setting
EtHAL Too Low EtHAL < low alarm limit Medium No Yes No
setting
FiHAL Too High FiHAL > high alarm limit Medium No Yes No
setting
FiHAL Too Low FiHAL < low alarm limit Medium No Yes No
setting
EtENF Too High EtENF > high alarm limit Medium No Yes No
setting
EtENF Too Low EtENF < low alarm limit Medium No Yes No
setting
FiENF Too High FiENF > high alarm limit Medium No Yes No
setting
FiENF Too Low FiENF < low alarm limit Medium No Yes No
setting
EtISO Too High EtISO > high alarm limit Medium No Yes No
setting
EtISO Too Low EtISO < low alarm limit Medium No Yes No
setting
FiISO Too High FiISO > high alarm limit Medium No Yes No
setting
FiISO Too Low FiISO < low alarm limit Medium No Yes No
setting
EtSEV Too High EtSEV > high alarm limit Medium No Yes No
setting
EtSEV Too Low EtSEV < low alarm limit Medium No Yes No
setting
FiSEV Too High FiSEV > high alarm limit Medium No Yes No
setting
* N/A - Not Applicable. This alarm message does not exist within this mode and therefore cannot be disabled
or enabled.
TABLE 6-3 Physiological Alarm Messages
DISABL
DISABLED DISABLED ED IN
WHEN IN MONIT
ALARM ALARM IS STANDBY OR
MESSAGE CAUSE PRIORITY OFF MODE MODE
FiSEV Too Low FiSEV < low alarm limit Medium No Yes No
setting
EtDES Too High EtDES > high alarm limit Medium No Yes No
setting
EtDES Too Low EtDES < low alarm limit Medium No Yes No
setting
FiDES Too High FiDES > high alarm limit Medium No Yes No
setting
FiDES Too Low FiDES < low alarm limit Medium No Yes No
setting
EtO2 Too High EtO2 > high alarm limit Medium No Yes No
setting
EtO2 Too Low EtO2< low alarm limit setting Medium No Yes No
Apnea CO2 No breath is detected and High Yes Yes No
Apnea time ≥ Apnea alarm
time.
* N/A - Not Applicable. This alarm message does not exist within this mode and therefore cannot be disabled
or enabled.
TABLE 6-3 Physiological Alarm Messages
NOTE: If an Apnea CO2 alarm occurs, the CO2 apnea elapse timer will display
on the CO2 waveform screen. The time displayed is the time since the
last breath and the time will reset once the CO2 Apnea alarm has
cleared (software bundle version 02.06.00 and later).
NOTE: Startup alarms priority is only used to display in the Service menu
alarm logbook.
NOTE: Startup Result if Fail column indicates the result when this startup
phase alarm is triggered, which may be ALL, only manual, and Non-
Functional.
NOTE: “All” indicates that all Automatic Ventilation, Manual Ventilation, and
Cardiac Bypass modes are enabled.
MACHINE
MODE STARTUP
ALARM WHEN RESULT IF
MESSAGE CAUSE PRIORITY CHECKED FAIL REMARK
Bundle Version Incompatible firmware High Startup Non- CPU Board
Error / version is installed. Functional
Incompatible
version found
Bundle Version: Self-test result cannot High Startup Non- CPU Board
Time out be obtained due to an Functional
internal communication
error.
Flowmeter DVCC, AVDD or VC High Startup Only Electronic
Voltage Error / voltage error Manual Flowmeter
Flowmeter Board
Voltage: Fail
Flowmeter Self CPU, Flash or WTD error High Startup Non- Electronic
Test Error / Functional Flowmeter
Flowmeter Self Board
Test Fail
Flowmeter Self Self-test result cannot High Startup Non- Electronic
Test: Time out be obtained due to an Functional Flowmeter
internal communication Board
error.
Aux Control 1. CPU, Flash or WTD High Startup Non- Aux Vent
Module Self Test error Functional Control
Error / Aux 2. After power on, CPU Board
Control Module board can't
Self Test: Fail communicate with the
Aux Control board.
Aux Control Self-test result cannot High Startup Non- Aux Vent
Module Self be obtained due to an Functional Control
Test: Time out internal communication Board
error.
TABLE 6-4 Startup Alarm Messages
MACHINE
MODE STARTUP
ALARM WHEN RESULT IF
MESSAGE CAUSE PRIORITY CHECKED FAIL REMARK
Ventilator Self 1. CPU, TIMER, RAM, High Startup Non- Ventilator
Test Error / WTD, EEPROM or AD Functional Control
Ventilator Self error Board
Test: Fail 2. After power on, CPU
board cannot
communicate with the
ventilator board.
Ventilator Self Self-test result cannot High Startup Non- Ventilator
Test: Time out be obtained due to an Functional Control
internal communication Board
error.
Ventilator 5V or 12V voltage error High Startup Only Ventilator
Voltage Error / Manual Control
Ventilator Board
Voltage: Fail
PEEP Valve 1. PEEP valve voltage Medium Startup Only Ventilator
Failure / PEEP error. Manual Control
Valve: Fail 2. PEEP valve pressure Board
error.
Insp Valve 1. Inspiratory valve Medium Startup Only Ventilator
Failure / Insp voltage error. Manual Control
Valve: Fail 2. Inspiratory valve flow Board
error.
Safety Valve PEEP safety valve Medium Startup Only Ventilator
Failure / Safety voltage error. Manual Control
Valve: Fail Board
Flow Sensor Ventilator flow is out of Low Startup Only Ventilator
Failure / Flow range. Manual Control
Sensor: Fail Board
Calibrate Flow 1. Calibration table isn't Low Startup Only Ventilator
Sensor and Insp found in EEPROM. Manual Control
Valve 2. Checksum of Board
Calibration table does
not match.
Calibrate 1. Calibration table isn't Low Startup Only Ventilator
Pressure Sensor found in EEPROM. Manual Control
and PEEP Valve 2. Checksum of Board
Calibration table does
not match.
Perform 100% 1. Calibration table isn't Low Startup All Ventilator
O2 Sensor found in EEPROM. Control
Calibration 2. Checksum of Board
Calibration table does
not match.
Ventilator After powering on, CPU High Startup Non- CPU Board
Initialization board cannot send the Functional
Error / parameter settings to
Ventilator the ventilator board.
Initialization:
Fail
TABLE 6-4 Startup Alarm Messages
MACHINE
MODE STARTUP
ALARM WHEN RESULT IF
MESSAGE CAUSE PRIORITY CHECKED FAIL REMARK
Ventilator Self-test result cannot High Startup Non- CPU Board
Initialization: be obtained due to an Functional
Time out internal communication
error.
Drive Gas Drive Gas Pressure is High Startup All Ventilator
Pressure Low low. Control
Board
O2 Supply O2 Supply Failure. High Startup All Ventilator
Failure / O2 Control
Supply: Fail Board
Power Supply 3.3V, 5V, 12V voltage High Startup Only Power
Voltage Error / error. Manual Board
Power Supply
Voltage: Fail
RT Clock Needs There is no button High Startup only All CPU Board
Battery battery cell available in
the system, or the
button battery cell
power is depleted.
RT Clock Failure RT chip malfunction. High Startup only All CPU Board
/ RT Clock: Fail
External AG Self If the module sends the Low Startup only All AG
Test Error ErrorMsg, except for Module
data limit error and
unspecified accuracy,
"External AG Self Test
Error" will be triggered.
External AG: External AG selftest Low Startup only All AG
Time out result can not be Module
obtained due to
communication error.
TABLE 6-4 Startup Alarm Messages
MACHINE DISABLE IN
ALARM MODE WHEN STANDBY
MESSAGE CAUSE PRIORITY CHECKED MODE
IP Address Conflict The IP address of the Medium Runtime No
machine is the same as
the IP address of another
device in the local
network.
Fan Failure Speed of the fan ≤ 20% of Medium Runtime No
normal speed
Fan Failure O2 Speed of Module Rack Medium Runtime No
fan < 3640
TABLE 6-5 CPU Board Runtime Alarm Messages
MACHINE DISABLE IN
ALARM MODE WHEN STANDBY
MESSAGE CAUSE PRIORITY CHECKED MODE
Power System Lost communication with High Runtime No
Comm Stop CPU board for 10
seconds.
Power Supply 3.3V, 5V, 12V voltage error High Runtime No
Voltage Error
Low Battery Battery voltage is less High Runtime No
Voltage! than 10.6V for 5 seconds.
System going Battery voltage is less High Runtime No
DOWN, Battery than 10.2V.
depleted!
Battery Battery undetected Medium Runtime No
Undetected
Battery in Use AC power fail Low Runtime No
Power Board High Power board temperature High Runtime No
Temp is greater than 95° C
Heating Module 1. Both resistance Low Runtime No
Failure temperatures are greater
than 105° C or less than 0°
C for 20 seconds.
2. One of the resistance
temperatures is greater
than 110° C for 15
seconds.
Breathing Circuit Breathing Circuit is not High Runtime No
Not Mounted mounted.
TABLE 6-6 Power Board Runtime Alarm Messages
NOTE: If the power board loses communication with the CPU board for 10
seconds, the alarm buzzer will be turned on.
NOTE: If the system restarts accidentally, the alarm buzzer will sound for 10
seconds to show notification.
MACHINE DISABLE IN
ALARM MODE WHEN STANDBY
MESSAGE CAUSE PRIORITY CHECKED MODE
Flowmeter DVCC, AVDD, or VC voltage High Runtime No
Voltage Error error
N2O Flow Too High N2O flow is greater than 15 Low Runtime No
L/min for 1 second.
O2 Flow Too High O2 flow is greater than 25 L/ Low Runtime No
min for 1 second.
Air Flow Too High Air flow is greater than 20 L/ Low Runtime No
min for 1 second.
TABLE 6-7 Electronic Flowmeter Board Runtime Alarm Messages
MACHINE DISABLE IN
ALARM MODE WHEN STANDBY
MESSAGE CAUSE PRIORITY CHECKED MODE
O2-N2O Ratio Error N2O flow is greater than 0.5 L/ High Runtime No
min and greater than 4 times
O2 flow for 1.6 seconds.
Flowmeter Comm Lost communication with High Runtime No
Stop CPU board for 10 seconds.
When this alarm is triggered,
the fresh gas flow value will
be displayed as '---'.
NO Fresh Gas Fresh gas flow is less than 50 Medium Runtime Yes
mL/min for 5 seconds when
the machine is not in
Standby mode or Monitor
mode..
Internal N2O Flow The I2C communication Low Runtime No
Failure between the CPU and N2O
flow sensor has failed.
Internal O2 Flow The I2C communication Low Runtime No
Failure between the CPU and O2 flow
sensor has failed.
Internal Air Flow The I2C communication Low Runtime No
Failure between the CPU and Air
flow sensor has failed.
TABLE 6-7 Electronic Flowmeter Board Runtime Alarm Messages
MACHINE DISABLE IN
ALARM MODE WHEN STANDBY
MESSAGE CAUSE PRIORITY CHECKED MODE
Aux Control Lost communication with High Runtime No
Module Comm CPU board for 10 seconds.
Stop
Ventilator Voltage 5V or 12V voltage error High Runtime No
Error
PEEP Valve Failure 1. PEEP valve voltage error Medium Runtime No
2. PEEP valve pressure
error
Insp Valve Failure 1. Inspiratory valve voltage Medium Runtime No
error
2. Inspiratory valve flow
error
Safety Valve PEEP safety valve voltage Medium Runtime No
Failure error
Flow Sensor 1. Inspiratory flow is out of Low Runtime No
Failure range.
2. Expiratory flow is out of
range.
* N/A - Not Applicable. This alarm message does not exist within this mode and therefore cannot be
disabled or enabled.
TABLE 6-8 Ventilator Control Board Runtime Alarm Messages
MACHINE DISABLE IN
ALARM MODE WHEN STANDBY
MESSAGE CAUSE PRIORITY CHECKED MODE
Check Flow 1. Inspiratory reverse flow High Runtime N/A *
Sensors 2. Expiratory reverse flow
Pinsp Not Pinsp does not reach the Low Runtime N/A *
Achieved Pinsp setting in pressure
mode.
Vt Not Achieved Vt does not reach the Vt Low Runtime N/A *
setting in volume mode.
Auto Ventilation When system is in the Auto High Runtime N/A*
Disabled Ventilation, Non functional
state
Automatic The machine is in the Low Runtime No
Ventilation automatic ventilation
Disabled disabled state.
Auto Ventilation Automatic Circuit Leak Low Runtime No
Disabled-Leak Test Test failed, and the result is
Failed "Manual Only".
Patient Circuit 1. Ppeak is less than Medium Runtime N/A
Leak 2cmH2O for continuously
30s during mechanical
ventilation.
2. Patient is not connected.
CO2 Absorber CO2 Canister is not High Runtime No
Canister Not mounted.
Locked
O2 Sensor O2 Sensor is not Low Runtime No
Disconnected connected.
Replace O2 sensor The O2 value is less than Medium Runtime No
5%.
Perform 100% O2 O2 value is greater than Low Runtime No
Sensor Calibration 110% or between 5% and
15% for 3 seconds.
Ventilator Comm Lost communication with High Runtime No
Stop the CPU board for 10
seconds.
Drive Gas Pressure Drive Gas Pressure is low. High Runtime No
Low
O2 Supply Failure O2 Supply Failure High Runtime No
Fresh Gas Flow In VCV and SIMV-VC Low Runtime N/A
Too High modes, the fresh gas flow
is greater than or equal to
the desired flow.
* N/A - Not Applicable. This alarm message does not exist within this mode and therefore cannot be
disabled or enabled.
TABLE 6-8 Ventilator Control Board Runtime Alarm Messages
DISABLE
WHEN
EXTERNAL AG
MACHINE IS IN
ALARM MODE WHEN STANDBY
MESSAGE CAUSE PRIORITY CHECKED MODE
AG Hardware Error AG module Hardware Medium Runtime Yes
Error.
O2 Sensor Error Paramagnetic O2 sensor Medium Runtime Yes
error.
External AG Self Module fault or Low Runtime Yes
Test Error communication failure
between the module and
anesthesia system.
AG Hardware AG module hardware High Runtime Yes
Malfunction malfunction. The AG
module enters Standby
and measurement stops.
AG Init Error The AG module was High Runtime Yes
installed improperly or
malfunctioned.
AG No Watertrap The AG module Low Runtime Yes
watertrap was installed
improperly or not
installed.
AG Watertrap Type When the patient type is Low Runtime Yes
Wrong infant, but the watertrap
type is adult/pediatric,
this alarm will be
triggered.
AG Change When the actual flow is Medium Runtime Yes
Watertrap less than 75% of the set
flow, the alarm indicates
that the watertrap is
gradually occluded and it
is necessary to replace
the water trap.
AG Comm Stop AG module malfunction High Runtime No
or communication
failure.
AG Airway Pump rate is lower than High Runtime Yes
Occluded 20ml/min for 1 second.
AG Zero Failed Gas measurements may Low Runtime Yes
have bad accuracy
during zeroing.
Mixed Agent and More than one Low Runtime Yes
MAC < 3 anesthetic gas and MAC
<3
Mixed anesthetic More than one Medium Runtime Yes
gas and MAC >= 3 anesthetic gas and MAC
>= 3
DISABLE
WHEN
EXTERNAL AG
MACHINE IS IN
ALARM MODE WHEN STANDBY
MESSAGE CAUSE PRIORITY CHECKED MODE
External AG When external AG High Runtime No
Module module is removed, this
Disconnected alarm will be triggered.
Incompatible AG The AG Version Limit is High Runtime No
Software Version On, and the AG module is
loaded while the AG
software version is lower
than 1.7.3.0.
CO2 Over Range The monitoring value Low Runtime Yes
N2O Over Range exceeds the measurable
range.
HAL Over Range
ENF Over Range
ISO Over Range
SEV Over Range
DES Over Range
O2 Over Range
Rate Over Range The monitoring value of Low Runtime Yes
Rate exceeds the
measurable range when
this type of alarm is
triggered, “--” will be
displayed.
Theory of Operation................................................................................................................................................................................7-3
Block Diagram.............................................................................................................................................................................................7-3
O2 Sensor Calibration.............................................................................................................................................................................7-6
WARNING: Do not use a malfunctioning A5 Anesthesia System. Have all repairs and
service done by an authorized service representative.
WARNING: Do not use talc, zinc stearate, calcium carbonate, corn starch, or similar
material to prevent sticking of the bellows, as these materials may
enter the patient's lungs or airway, causing irritation or injury.
WARNING: Do not use lubricants that contain oil or grease. They can burn or
explode in the presence of high O2 concentrations.
WARNING: Obey infection control and safety procedures. Used equipment may
contain blood and body fluids.
CAUTION: Never immerse the oxygen sensor or its connector in any type of liquid.
Dispose of the oxygen sensor per the manufacturer’s specification.
During the inspiratory phase, the microprocessor-controlled inspiratory valve produces the preset
drive gas inspiratory flow and the expiratory valve closes. The drive gas enters the bellows dome in
the patient circuit and depresses the bellows inside the dome to move downward. This forces the gas
inside the bellows to enter the patient's lungs until the end of the inspiratory phase.
During the expiratory phase, the inspiratory valve closes and the expiratory valve opens. The patient
can expire freely. The patient's expired gas, mixed with the fresh gas, enters and lifts the bellows
inside the dome. The drive gas outside the bellows is scavenged to the Anesthetic Gas Scavenging
System (AGSS) until the end of the expiratory phase.
During ventilation, the ventilator performs real-time monitoring over airway pressure and flow. If the
airway pressure or minute volume is outside the user-preset alarm limits, an audible and visible alarm
occurs. When the airway pressure is higher than the limit value determined by the PEAK high alarm
limit, the ventilator enters the expiratory phase automatically to avoid causing injury to the patient.
Additionally, the ventilator has a built-in pressure safety valve that opens at an approximate pressure
of 110 cmH2O (11kPa).
Breathing Patient
Anesthetic Gas Delivery System
System
NOTE: During cleaning and setup, inspect the parts and seals for damage.
Replace or repair as necessary.
MINIMUM
FREQUENCY MAINTENANCE
Daily Clean the external surfaces.
Every 72 hours Perform 21% O2 calibration (O2 sensor in breathing system).
The A5 will prompt the user for 21% O2 calibration (only for units with an galvanic
O2 cell).
Monthly Water trap on AG module.
Annually Periodic maintenance due, to be performed by a trained technician.
Gas Bench calibration.
Contact Mindray Technical Support for details.
Every three years Periodic maintenance due, to be performed by a trained technician.
Contact Mindray Technical Support for details.
As necessary • Perform 100% O2 calibration after replacing the O2 sensor.
• Replace the O2 sensor if it cannot be calibrated.
• Before installing the cylinder, use a new cylinder gasket on the cylinder yoke.
• Empty the water trap if there is water buildup.
• Replace the sodalime in the canister if sodalime color change is detected. Follow
the manufacturer’s instructions.
• Replace the flow sensor if the seal for the flow sensor is damaged, the membrane
inside the flow sensor is cracked or distorted, or the flow sensor is cracked or
distorted.
• Calibrate the flow sensor after re-installing the cleaned or disinfected flow sensor,
replacing with a new flow sensor, or when tidal volume measurement is inaccurate.
• Replace the transfer tube if it is damaged.
• Inspect the O2 flush button for normal movement. If not ,refer to the service
manual for the disassembling and cleaning.
NOTE: During calibration, do not operate the pneumatic parts. Do not move or
press the breathing tubes.
NOTE: Calibrate the flow sensor after re-installing the cleaned or disinfected
flow sensor, replacing with a new flow sensor, or when tidal volume
measurement is inaccurate.
The flow sensor must be calibrated whenever the flow volume is out of specification or after
changing the flow sensor.
FIGURE 7-2 Flow Sensor Calibration FIGURE 7-3 Flow Sensor Calibration
Begin Successful
For continued O2 sensor accuracy, the A5 checks for 21% O2 calibration approximately every 72
hours. The A5 prompts the user for 21% O2 calibration as follows:
• When the machine is powered on, if more than 72 hours have elapsed since the last
successful calibration, the prompt message "Calibrate O2 sensor for 21%" is displayed. The
message disappears after successful calibration.
• If the machine is kept powered on, the prompt message “Calibrate O2 sensor for 21%” is
displayed at the next Standby mode after 5am after 72 hours have elapsed since the last
successful calibration.
NOTE: If the alarm message "RT Clock Needs Battery" or "RT Clock Failure" is
displayed, the prompt message "Calibrate O2 sensor for 21%” is
disabled.
The O2 sensor must be removed from the breathing system before calibrating it at 21%. The O2
sensor can be reinstalled after verifying that there is no water build-up in the O2 sensor and its
installation part.
NOTE: The breathing system automatically seals off the O2 sensor port when
the O2 sensor is removed.
NOTE: The A5 system will not allow the End Case button to be selected until
the Auto/Manual ventilation switch is set to Manual, and system
detects the individual fresh gas flows are effectively turned off (i.e.,
flow < 0.05 L/min).
d. Follow the screen prompts to end the case and enter Standby mode.
2. Select Setup > General > Calibrate O2 Sensor.
Only 21% O2 sensor calibration is available in the General tab,
or
Select Setup > System (system password needed) > Calibration > O2 Sensor.
Both 21% and 100% O2 sensor calibrations are available in the System tab.
The 21% button is highlighted by default.
NOTE: In the System tab, 21% oxygen sensor calibration must be completed
before performing 100% calibration. The 100% button is disabled if a
21% oxygen sensor calibration has not been successfully completed
within 72 hours.
3. Remove the O2 sensor from the O2 sensor port on the breathing system.
Allow three (3) minutes for the sensor to acclimate to the environment.
4. Carefully follow the on-screen prompts to prepare for calibration.
5. Select the Begin button to start 21% O2 sensor calibration. The system will indicate the
calibration status when the process is completed.
6. When 21% O2 sensor calibration is successfully completed, reinstall the O2sensor into the O2
sensor port on the breathing system. If an error code in red (e.g., 00 00 00 10) is displayed,
see TABLE 7-2, “O2 Sensor Calibration Error Codes,” on page 7-8 for troubleshooting
information.
7. If you are in the Setup > General, select Done when 21% O2 sensor calibration is
completed. Skip the remaining steps below.
or
If you are in the Setup > System and wish to skip 100% O2 sensor calibration, select Done to
close the calibration window. Skip the remaining steps below.
8. Select the 100% button to perform 100% O2 sensor calibration.
9. Carefully follow the on-screen prompts to prepare for calibration.
10. Select the Begin button to start 100% O2 sensor calibration. The system will indicate the
calibration status when the process is completed. If an error code in red (e.g., 00 00 00 10) is
displayed, see TABLE 7-2, “O2 Sensor Calibration Error Codes,” on page 7-8 for
troubleshooting information.
11. After calibration, select Done to close the calibration window.
00 00 00 02 O2 supply pressure is low. . Check that the O2 sensor is connected to the cable
During 100% calibration correctly.
process, O2 supply pressure . Check the O2 supply pressure.
was not sufficient. . Check that the O2 sensor output voltage in the
calibration menu is steady.
. Replace the O2 sensor.
00 00 00 04 O2 sensor is disconnected. . Check that the O2 sensor is connected to the cable
Sampled data is greater correctly.
than 2900 (AD value). . Check that the O2 sensor output voltage in the
calibration menu is steady.
. Replace the O2 sensor.
00 00 00 08 21% calibration value is . Check that the O2 sensor is connected to the cable
outside of the expected correctly.
range (150~500) . Check that the O2 sensor is in 21% O2.
(AD value). . Check that the O2 sensor output voltage in the
calibration menu is steady.
. Replace the O2 sensor.
00 00 00 10 100% calibration value is . Check that the O2 sensor is connected to the cable
outside of the expected correctly.
range (800~2028) (AD . Check that the O2 sensor is in 100% O2.
value). . Check that the O2 sensor output voltage in the
calibration menu is steady.
. Replace the O2 sensor.
00 00 00 20 Error writing to EEPROM. . Repeat the calibration.
. Replace the O2 sensor.
. Replace the CPU board.
TABLE 7-2 O2 Sensor Calibration Error Codes
• More CO2 stays in the CO2 absorbent canister to react and produce water.
• More moist, exhaled gas stays in the breathing system and CO2 absorbent canister to
produce condensed water.
Check the inspiratory and expiratory flow sensors when abnormal flow waveform or unstable tidal
volume fluctuation is detected. Check the sensor for water. If there is water build-up, clear it
immediately before use.
• Use a filter between the flow sensor and the patient to limit water condensation in the flow
sensor.
• Check the water trap for water before using the A5 Anesthesia System. If there is water build-
up, clear it immediately.
WARNING: Check water build-up inside the flow sensor before every system use.
Pooled water in the flow sensor causes erroneous readings.
WARNING: Ensure that all breathing system parts are completely dried after the
breathing system is cleaned and disinfected.
For additional information about infection control, refer to the ASA’s Recommendations for Infection
Control for the Practice of Anesthesiology, second edition. For additional information on reprocessing
medical devices, refer to AAMI TIR 30:2003, A compendium of process, materials, test methods, and
acceptance criteria for cleaning reusable medical devices.
TABLE 7-3 lists the cleaning and disinfecting agents and autoclaving process that may be used on the
A5 Anesthesia System.
AGENT CLASSIFICATION
Water Cleaning agent
Green soap tincture Cleaning agent
Sodium hypochlorite solution, 10% available chlorine Disinfecting agent
Isopropyl alcohol (70%) Disinfecting agent
* All breathing system components are autoclavable except the Paw gauge, flow sensor, O2 sensor, and
bellows. The components can be autoclaved up to a maximum temperature of 134 ºC (273 ºF).
AGENT CLASSIFICATION
Super Sani-Cloth (0.5% Quaternary ammonium chloride Disinfecting agent
and 55% Isopropyl alcohol)
Cidex (Only for bellows, Inspiratory Pressure Gauge and Ins/ Exp Disinfecting agent
Flow sensors)
ALPET D2 Surface sanitizer wipes Disinfecting agent
Viraguard surface disinfectant towelette Disinfecting agent
Autoclaving process * Autoclaving
* All breathing system components are autoclavable except the Paw gauge, flow sensor, O2 sensor, and
bellows. The components can be autoclaved up to a maximum temperature of 134 ºC (273 ºF).
Read all content in this section before disassembling, cleaning, disinfecting, and re-assembling the
bellows to avoid equipment malfunction and patient injury.
1. The bellows dome is a transparent cover with graduation marks from 300 to 1500 mL.
Remove the bellows dome by turning it counterclockwise and lifting it away from the
breathing system (see FIGURE 7-5).
2. Detach the bellows from the base plate (see FIGURE 7-6).
3. Detach the top plate from the bellows (see FIGURE 7-7).
4. Remove the bellows adapter ring from inside the bellows (see FIGURE 7-8). Note the
orientation of the bellows adapter ring as it is being removed to ensure that it is properly
inserted during reassembly. (If the ring contains grooves, the ring should be oriented so that
the grooves are facing downward in the final reassembly.)
6. Cleaning
a. To prevent damage, wash each component gently using a recommended cleaning
agent (see section 7.10.2 (page 7-11) "Cleaning and Disinfecting Agents / Autoclaving").
Ensure that all bellows surfaces are cleaned. Do not autoclave the bellows.
b. Rinse with clean, hot water, and allow to dry.
NOTE: Dry the bellows by allowing it to hang so that it is fully expanded. This
will facilitate thorough drying and prevent it from sticking to itself.
CAUTION: If moisture remains in the bellows after cleaning, the bellows surface
folds may become tacky and prevent the bellows from properly
expanding. Ensure all moisture is removed from the bellows after
cleaning.
c. After all bellows components are completely dry, inspect them for damage before
disinfection or re-assembly and functional testing.
d. If disinfecting the bellows components, continue with step 7, otherwise skip to step 8.
7. Disinfection
NOTE: Ensure that all bellows components have been cleaned as described in
step 6 before disinfecting.
See section 7.10.2 (page 7-11) "Cleaning and Disinfecting Agents / Autoclaving" and use an
approved disinfecting agent for all bellows components while adhering to facility policies and
procedures.
8. Connect the bellows to the breathing system by reassembling all components in the reverse
order. Prior to use after cleaning or disinfecting, power up the system and follow the on-
screen prompts to perform the leak test and the compliance test (see section 4.5 (page 4-9)
"Leak and Compliance Tests").
Expiration
Valve
Inspiration
Valve
CAUTION: The valve disc in each of the inhalation and exhalation valve assemblies
on the breathing system is fragile and must be handled with care while
removing the valve cage from the valve assembly.
2. The valve cage will be removed in this step (see FIGURE 7-12). The six prongs of the valve
cage have tabs that secure it in the valve assembly. While noting the previous CAUTION, use
two hands to remove the valve cage by gently manipulating the prongs to release the tabs.
As the valve cage is lifted away from the assembly, ensure that the valve disc does not fall
out.
3. Remove the valve disc from the valve cage (see FIGURE 7-12).
4. Remove the O-ring from the bottom of the valve assembly (see FIGURE 7-12.
Valve Cage
Valve Disc
O-ring
CAUTION: The valve disc in each of the inhalation and exhalation valve assemblies
on the breathing system is fragile and must be handled with care while
removing the valve cage from the valve assembly.
5. Cleaning
a. Wash each component using a recommended cleaning agent (see section 7.10.2 (page
7-11) "Cleaning and Disinfecting Agents / Autoclaving").
b. Rinse with clean, hot water, and allow to dry.
c. After all components are completely dry, verify that the valve disc and the prongs of the
valve cage are undamaged before disinfection or re-assembly and functional testing.
d. If disinfecting the valve components, continue with step 6, otherwise skip to step 7.
6. Disinfection
NOTE: Ensure that all valve components have been cleaned as described in
step 5 before disinfecting.
See section 7.10.2 (page 7-11) "Cleaning and Disinfecting Agents / Autoclaving" and use an
approved disinfecting agent for all valve components while adhering to facility policies and
procedures.
7. Reassembly
Reassemble the valve components in the reverse order, noting any previously stated CAU-
TION. Prior to use after cleaning or disinfecting, power up the system and follow the on-
screen prompts to perform the leak test and the compliance test (see section 4.5 (page 4-9)
"Leak and Compliance Tests").
Oxygen Sensor
CAUTION: Never immerse the oxygen sensor or its connector in any type of liquid.
2. Clean the oxygen sensor exterior with a soft, lint-free cloth, and a recommended cleaning
agent (see section 7.10.2 (page 7-11) "Cleaning and Disinfecting Agents / Autoclaving").
Allow to dry thoroughly.
3. Inspect the oxygen sensor for damage and replace as necessary.
4. Re-insert the oxygen sensor if it had been removed.
2. Cleaning
a. Clean the APL valve with a soft, lint-free cloth and a recommended cleaning agent (see
section 7.10.2 (page 7-11) "Cleaning and Disinfecting Agents / Autoclaving"). Allow it to
dry thoroughly.
b. If disinfecting the APL valve, continue with step 3, otherwise skip to step 4.
3. Disinfection
NOTE: Ensure that the APL valve has been cleaned as described in step 2
before disinfecting.
See section 7.10.2 (page 7-11) "Cleaning and Disinfecting Agents / Autoclaving" and use an
approved disinfecting agent for the APL valve while adhering to facility policies and procedures.
4. Reassemble the APL valve by turning its base collar clockwise until it is securely tightened.
Prior to use after cleaning or disinfecting, power up the system and follow the on-screen
prompts to perform the leak test and the compliance test (see section 4.5 (page 4-9) "Leak
and Compliance Tests").
2. Clean the PAW gauge with a soft, lint-free cloth and a recommended cleaning agent (see
section 7.10.2 (page 7-11) "Cleaning and Disinfecting Agents / Autoclaving"). Allow it to dry
thoroughly.
3. Re-insert the PAW gauge if it was removed. Prior to use after cleaning or disinfecting, power
up the system and follow the on-screen prompts to perform the leak test and the
compliance test (see section 4.5 (page 4-9) "Leak and Compliance Tests").
Bag Arm
Retaining
Ring
2. Cleaning
a. Clean the bag arm with a soft, lint-free cloth and a recommended cleaning agent (see
section 7.10.2 (page 7-11) "Cleaning and Disinfecting Agents / Autoclaving"). Allow it to
dry thoroughly.
b. If disinfecting the bag arm, continue with step 3, otherwise skip to step 4.
3. Disinfection
NOTE: Ensure that the bag arm has been cleaned as described in step 2 before
disinfecting.
See section 7.10.2 (page 7-11) "Cleaning and Disinfecting Agents / Autoclaving" and use an
approved disinfecting agent for the bag arm while adhering to facility policies and procedures.
4. Reassemble the bag arm to the breathing system. Prior to use after cleaning or disinfecting,
power up the system and follow the on-screen prompts to perform the leak test and the
compliance test (see section 4.5 (page 4-9) "Leak and Compliance Tests").
Absorber Canister
Condensate Drain
2. While holding a small cup below the drain, turn the condensate drain valve clockwise to
open the drain and collect any water that may have gathered. Turn the drain valve counter-
clockwise to close the drain. After draining out moisture wipe out excess moisture with a
soft cloth. Discard any water collected.
WARNING: Use extreme care while handling the absorbent as it is a caustic irritant.
3. Rotate the locking mechanism handle clockwise into the unlocked position (see FIGURE 7-
19). This separates the absorber canister from the top of the assembly. While noting the
previous WARNING, remove the absorber canister. Then remove the Pre-Pak or loose fill
absorbent from the canisters. Dispose of the absorbent per the manufacturer’s specification.
FIGURE 7-19 Absorber Canister, Unlocked FIGURE 7-20 Absorber Canister, Locked
4. Cleaning
a. Clean the absorber canister with a soft, lint-free cloth and a recommended cleaning
agent (see section 7.10.2 (page 7-11) "Cleaning and Disinfecting Agents / Autoclaving").
Allow them to dry thoroughly.
b. If disinfecting the absorber canister,, continue with step 5, otherwise skip to step 7.
5. Disinfection
NOTE: Ensure that the absorber canister has been cleaned as described in step
4 before disinfecting.
6. See section 7.10.2 (page 7-11) "Cleaning and Disinfecting Agents / Autoclaving" and use an
approved disinfecting agent for the absorber canister, while adhering to facility policies and
procedures.Make sure that the gasket is correctly installed. The comparison between correct
installation and incorrect installation is shown below.
Correct
installation
Incorrect
installation
WARNING: Use extreme care while handling the absorbent as it is a caustic irritant.
WARNING: Check if the gasket is properly installed in place while installing the
absorber canister. If the gasket is not properly installed (for example,
gasket is not evenly seated and centered) it may cause breathing
system leakage.
NOTE: Ensure that the absorber canister is completely dry before adding
absorbent.
WARNING: The gasket on the absorber canister should be cleaned before adding
new absorbent.
7. While noting the previous WARNING, add new Pre-Pak or loose fill absorbent to the
absorber canister. Re-install the absorber canister into the assembly. Rotate the locking
mechanism handle clockwise into the locked position (see FIGURE 7-20). Prior to use after
cleaning or disinfecting, power up the system and follow the on-screen prompts to perform
the leak test and the compliance test (see section 4.5 (page 4-9) "Leak and Compliance
Tests").
Buckle
Bypass Assembly
CAUTION: Use care in lifting and manipulating the breathing system block during
removal from its mounting arm as handling may be awkward due to its
weight and shape.
CAUTION: The breathing system block is calibrated and matched with the
anesthesia machine at the factory. A label in the back of the machine
indicates the serial number of the matching breathing system block.
When reassembling, ensure that the breathing system block and
anesthesia machine are properly matched. Otherwise, the breathing
system must be recalibrated.
5. While holding the sides of the breathing system block, firmly separate and slide it away from
its mounting arm.
6. Cleaning
a. Clean the breathing system block exterior with a soft, lint-free cloth and a
recommended cleaning agent (see section 7.10.2 (page 7-11) "Cleaning and Disinfecting
Agents / Autoclaving"). Allow to dry thoroughly.
b. If disinfecting the breathing system block, continue with step 7, otherwise skip to step 8.
7. Disinfection
NOTE: Ensure that the breathing system block has been cleaned as described
in step 6 before disinfecting. High level disinfection of the breathing
system block can be performed through steam autoclaving up to a
maximum temperature of 134 ºC (273 ºF).
Using an autoclave, follow the manufacturer’s instructions for high level disinfection of the breathing
system block while adhering to facility policies and procedures.
8. Reassemble the breathing system components in reverse order. Prior to use after cleaning or
disinfecting, power up the system and follow the on-screen prompts to perform the leak test
and the compliance test (see section 4.5 (page 4-9) "Leak and Compliance Tests").
CAUTION: To ensure patient safety, use only parts and accessories specified in this
manual.
Transfer Hose
3. Clean the outer surface of the AGSS and Transfer Hose with a soft, lint-free cloth and a
recommended cleaning agent (see section 7.10.2 (page 7-11) "Cleaning and Disinfecting
Agents / Autoclaving"). Allow to dry thoroughly.
4. Remove the top of the AGSS. Inspect the AGSS filter and shake it over a waste container to
clean it as necessary (see FIGURE 7-27). If the filter must be replaced, dispose of the old filter
per local disposal regulations.
AGSS Filter
5. Reassemble the AGSS and Transfer Hose and reconnect them to the A5 in the reverse order.
Visual inspection should be performed every 30 days to ensure timely replacement of worn or
damaged parts.
Introduction .................................................................................................................................................................................................8-2
Make AG Settings......................................................................................................................................................................................8-6
Troubleshooting........................................................................................................................................................................................8-8
8.1 Introduction
The Anaesthetic Gas (AG) module measures the patient's anesthetic and respiratory gases, and
incorporates the features of the O2 module as well.
The AG (anesthesia gas) module determines the concentrations of certain gases using the infrared
(IR) light absorption measurement. The gases that can be measured by the AG module absorb IR
light. Each gas has its own absorption characteristic. The gas is transported into a sample cell, and an
optical IR filter selects a specific band of IR light to pass through the gas. For multiple gas
measurement, there are multiple IR filters. This means that higher concentration of IR absorbing gas
causes a lower transmission of IR light. The amount of IR light transmitted after it has been passed
though an IR absorbing gas is measured. From the amount of IR light measured, the concentration of
gas present can be calculated.
Oxygen does not absorb IR light as other breathing gases and is therefore measured relying on its
paramagnetic properties. Inside the O2 sensor are two nitrogen-filled glass spheres mounted on a
strong rare metal taut-band suspension. This assembly is suspended in a symmetrical non-uniform
magnetic field. In the presence of paramagnetic oxygen, the glass spheres are pushed further away
from the strongest part of the magnetic field. The strength of the torque acting on the suspension is
proportional to the oxygen concentration. From the strength of the torque, the concentration of
oxygen is calculated.
1. An EtCO2 waveform;
2. Measured parameters: O2, EtCO2, FiCO2, EtN2O, FiN2O, EtAA, FiAA and MAC,
where, AA stands for any of the five anesthetic agents: Des (desflurane), Iso (isoflurane), Enf
(enflurane), Sev (sevoflurane), or Hal (halothane),
MAC is a basic index indicating the depth of inhaled anesthesia. The ISO 80601-2-55 defines MAC as
follows: alveolar concentration of an inhaled anesthetic agent that, in the absence of other anesthetic
agents and at equilibrium, prevents 50% of subjects from moving in response to a standard surgical
stimulus.
Anesthetic
Des Iso Enf Sev Hal N 2O
agent
1 MAC 6.0% 1.15% 1.7% 2.1% 0.77% 105%*
* 1 MAC nitrous oxide can only be reached in a hyperbaric chamber.
TABLE 8-1 1 MAC of various inhaled anesthetic agents
NOTE: The data shown in this table is from ISO 80601-2-55, which are
published by the U.S. Food and Drug Administration for a
healthy 40-year-old male patient.
NOTE: In actual applications, although the A5 accounts for patient age, the
effects of weight and other factors on the inhaled anesthetic agent
should be considered.
When one or more than one anesthetic agents are used, the formula for calculating MAC is:
N −1
EtAgenti
MAC =
i =0 AgentVolagei
Where, N stands for the number of all anesthetic agents (including N2O) which the AG module can
measure, EtAgenti for the concentration of end-tidal anesthetic agent and AgentVolagei for the 1MAC
value corresponding to the anesthetic agent after age correction.
NOTE: The formula above is only suitable for patients who are older than one
year old. If the patient is less than one year old, the system will use one
year to do age correction.
For example, for a 60-year-old patient, if the AG module detects 0.9% Iso and 50% N2O in the patient
end-tidal mixed gas, the 1MAC of Iso is 1.01% and 1MAC of N2O is 92.7% of the 60-year-old patient
based on the above age correction formula. The MAC value is calculated as follows:
0 .9 % 50%
MAC = + = 1 .4
1.01% 92.7%
Indicator
NOTE: The AG module (see FIGURE 8-3) is configured with the function of
compensating barometric pressure automatically.
CAUTION: Position the airway adapter properly so that the part connecting to the
gas sampling tube is pointing upwards. This prevents condensed water
from entering the gas sampling tube and causing an occlusion as a
result.
CAUTION: The watertrap collects water drops condensed in the sampling tube and
therefore prevents them from entering the module. If the collected
water reaches a certain amount, you should drain it to avoid airway
blockage.
CAUTION: The watertrap has a filter preventing bacterium, vapor and patient
secretions from entering the module. After long-term use, dust or other
substances may compromise the performance of the filter or even
block the airway. In this case, replace the watertrap. Replacing the
watertrap once a month is recommended.
WARNING: Do not apply high volume watertraps to Infant patients. Otherwise, patient injury
could result.
WARNING: Make sure that all connections are reliable. Any leak in the system can result in
erroneous readings due to patient breathing gas mixed with ambient air.
NOTE: When using the A5 Anesthesia System, ensure that the alarm limits of
each parameter are set to the appropriate values for the patient.
NOTE: Gas data is reported as zero if the measured concentration is below the
defined threshold level during more than 3 s: CO2 - 0.1/0.3%; N2O - 3/
3%; O2 - 0/0%, Agents - 0.15/0.3% (Full/ISO accuracy).
8.7 Troubleshooting
If the gas inlet (including watertrap, sampling tube and airway adapter) is occluded by condensed
water, airway occlusion will be prompted on the screen.
If the expired O2 concentration is higher than the inspired O2 concentration, it is possible that the
pump rate is too low. Setting Gas Bench Flow Rate to High is recommended.
Metal lever
Exhaust tube
To scavenge the sample gas to the waste gas disposal system, depress the metal lever and then plug
the exhaust tube to the sample gas return port marked as shown in the above picture.
WARNING: When using the AG module to perform AG measurements on the patients who are
receiving or have recently received anesthetic agents, connect the outlet to the sample
gas return port to prevent the medical staff from breathing in the anesthetic agents.
• Gas cylinder, with a certain standard gas or mixture gas. Gas concentration should meet the
following requirements: AA≥1.5%, CO2≥1.5%, N2O≥40%, O2≥40%, of which AA represents an
anesthetic agent. a/c≤0.01 (a is the gas absolute concentration accuracy; c is the gas
concentration).
• T-shape connector
• Tubing
Follow this procedure to perform a calibration:
Tubing
Relief valve
AG module
Gas cylinder
2. Ensure that the system is in Standby mode. If not, select the End Case button in the Manual
tab and follow the on-screen prompts to end the case and enter Standby mode.
3. Select the Setup softkey> System tab (system password needed).
4. Select the Calibration button.
5. Select the External AG Module button.
6. Wait for the AG module to be completely warmed up
7. Enter the actual concentration of the calibration gas.
8. Turn on the calibration gas canister. The system displays the real-time concentration of
calibration gas.
9. Select the Calibrate button to start to calibrate the AG Module. The system will display the
results of the calibration status when the process is completed.
10. After calibration, select Done to close the Calibration window.
11. Select Accept to close the Setup window.
Safety Designations.................................................................................................................................................................................9-4
Physical Specifications...........................................................................................................................................................................9-5
Environmental Specifications............................................................................................................................................................9-6
Electrical Specifications.........................................................................................................................................................................9-7
Pneumatic Specifications.....................................................................................................................................................................9-8
The anesthesia workstation shall be used together with the monitoring devices, alarm system and
protective devices below:
• The pressure restriction device, expiratory volume monitor and breathing system with alarm
system also comply with ISO 80601-2-13.
• AG monitor in compliance with ISO 80601-2-55.
• O2 monitor in compliance with ISO 80601-2-55.
WARNING: Due to the size and weight of the A5, it should only be moved by
qualified personnel.
WARNING: To avoid tip hazards, use care when moving the A5 up or down inclines,
around corners and across thresholds. Remove all equipment from the
top shelf and mounted to the side of the A5 before moving. Do not
attempt to roll the A5 over hoses, cords or other obstacles.
Pipeline Input Pressure Range: N2O: 280 to 600 kPa (40 to 87 psi)
Air: 280 to 600 kPa (40 to 87 psi)
O2: 280 to 600 kPa (40 to 87 psi)
Pipeline Input Flow Rate Range: O2: Max. 190 L/min (Including maximum drive gas flow rate,
maximum flow rate to seal PEEP valve, maximum O2 Flow
meter and maximum O2 flush)
Air: Max. 20 L/min
N2O: Max. 20 L/min
Pipeline Connections: DISS threaded body as per CGA V-5
Gas Configuration: N2O, Air, O2
Cylinder Input Flow Rate Range: O2: Max. 190 L/min (Including maximum drive gas flow rate,
maximum flow rate to seal PEEP valve, maximum O2 Flow
meter and maximum O2 flush)
Air: Max. 20 L/min
N2O: Max. 20 L/min
Cylinder Connections: Pin-Index Safety System (PISS)
Yoke Configuration: N2O, Air, O2
9.7.6 O2 Controls
O2 supply failure alarm: 185.5 to 254.5 kPa (27 to 36 psi)
Flow (L/min)
Pressure drop(kPa)
Flow(L/min)
Flow (L/min)
Flow(L/min)
9.8.6 Resistance
Expiratory resistance:
Pressure drop(kPa)
Flow(L/min)
Inspiratory resistance:
Pressure drop(kPa)
Flow(L/min)
Note: The block heater does not operate while the system is being powered by the internal battery supply.
TABLE 9-18 Breathing System Temperature Controller
9.8.9 Materials
All materials in contact with the patient’s exhaled gas are autoclavable, except the flow sensors,
pressure gauge, bellows, and O2 cell. All materials in contact with the patient's gas comply with
ISO 10993-1, ISO 10993-5, ISO 10993-10.
1): 10% to 90%. Sample gas flow: 200 ml/min. DRYLINETM watertrap. Adult DRYLINETM sampling line (2.5 m).
2): 10% to 90%. Sample gas flow: 120 ml/min. DRYLINETM watertrap. Pediatric DRYLINETM sampling line (2.5 m).
TABLE 9-21 AG Module
Range: CO2 : 0 to 30 %
O2 : 0 to 100 %
N2O : 0 to 100 %
DES : 0 to 30 %
SEV : 0 to 30 %
ENF : 0 to 30 %
ISO : 0 to 30 %
HAL : 0 to 30 %
ISO accuracy mode As Full accuracy specifications, but derated as follows:
Add ±0.3%ABS to accuracy for CO2;
Add ±8%REL to accuracy for all agents;
N2O accuracy is ± (8%REL+2%ABS).
Full accuracy mode Gas Range (%REL) Accuracy (%ABS)
CO2 0 to 1 ±0.1
1 to 5 ±0.2
5 to 7 ±0.3
7 to 10 ±0.5
>10 Unspecified
N2O 0 to 20 ±2
20 to 100 ±3
O2 0 to 25 ±1
25 to 80 ±2
80 to 100 ±3
DES 0 to 1 ±0.15
1 to 5 ±0.2
5 to 10 ±0.4
10 to 15 ±0.6
15 to 18 ±1
>18 Unspecified
SEV 0 to 1 ±0.15
1 to 5 ±0.2
5 to 8 ±0.4
>8 Unspecified
ENF, ISO, HAL 0 to 1 ±0.15
1 to 5 ±0.2
>5 Unspecified
1)
Rise time@200ml/min CO2 : ≤250 ms
O2 :≤500 ms
N2O : ≤250 ms
ENF : ≤350 ms
DES, SEV, ISO, HAL: ≤300 ms
1): 10% to 90%. Sample gas flow: 200 ml/min. DRYLINETM watertrap. Adult DRYLINETM sampling line (2.5 m).
2): 10% to 90%. Sample gas flow: 120 ml/min. DRYLINETM watertrap. Pediatric DRYLINETM sampling line (2.5 m).
TABLE 9-21 AG Module
NOTE: Inaccuracy specifications are affected by the breath rate and I:E change.
The end-tidal gas reading is within specification for breath rate below
15BPM and I:E ratio smaller than 1:1 relative to the gas readings
without breath; Add ±6%REL to inaccuracy for HAL and O2 for breath
rate larger than 15 BPM; Add ±6%REL to inaccuracy for all gases for
breath rate larger than 30 BPM (inaccuracy for HAL and O2 are
unspecified in this case); inaccuracy is unspecified for breath rate larger
than 60 BPM.
NOTE: Data sample rate 25 Hz. Data presentation is 50 Hz, every second data
point is interpolated.
NOTE: Inspiratory and end tidal CO2 concentration readings are identified by
AIONTM Platinum Multigas Analyzers using the lowest and highest
values respectively of the temporal CO2-curve. Corresponding
readings of N2O and anesthetic agents are taken at the same point in
time. Inspiratory and end-tidal O2 concentration readings are
identified by the O2 mean value during the respiratory phase as
identified by the temporal CO2 curve. Once correctly identified, the
highest and lowest O2 concentration readings during each part of the
phase will be presented as inspiratory and end-tidal O2 respectively.
9.10.2 Alarms
AG Alarm Limits Range Default Step Unit
EtCO2 High Limit Off, 2 to 99 Adult/Ped: 50 mmHg 1 mmHg, %, kPa
Infant: 45 mmHg
EtCO2 Low Limit Off, 0 to 97 Adult/Ped: 25 mmHg
Infant: 30 mmHg
FiCO2 High Limit Off, 1 to 99 4 mmHg
EtO2 High Limit Off, low+2 to 100 88%
EtO2 Low Limit Off, 18 to high-2 18%
EtN2O High Limit Off, 2 to 100 55% 1 %
EtN2O Low Limit Off, 0 to 98 0%
FiN2O High Limit Off, 2 to 100 53%
FiN2O Low Limit Off, 0 to 98 0%
EtHal High Limit Off, 0.2 to 5.0 3% 0.1 %
EtHal Low Limit Off, 0.0 to 4.8 0%
FiHal High Limit Off, 0.2 to 5.0 2%
FiHal Low Limit Off, 0.0 to 4.8 0%
EtEnf High Limit Off, 0.2 to 5.0 3% 0.1 %
EtEnf Low Limit Off, 0.0 to 4.8 0%
FiEnf High Limit Off, 0.2 to 5.0 2%
FiEnf Low Limit Off, 0.0 to 4.8 0%
EtIso High Limit Off, 0.2 to 5.0 3% 0.1 %
EtIso Low Limit Off, 0.0 to 4.8 0%
FiIso High Limit Off, 0.2 to 5.0 2%
FiIso Low Limit Off, 0.0 to 4.8 0%
EtSev High Limit Off, 0.2 to 8.0 6% 0.1 %
EtSev Low Limit Off, 0.0 to 7.8 0%
FiSevHigh Limit Off, 0.2 to 8.0 5%
FiSev Low Limit Off, 0.0 to 7.8 0%
Ventilator Performance
Drive Pressure: 280 to 600 kPa
Inspiratory flow range: 2.4 to 110 L/min
Flow Valve Range: 1 to 110 L/min
Minute Volume Monitoring Accuracy: 0 to 30 L/min ±15% of the displayed value, repeatable to
±5% over a 1-hour period
* Specifications are applicable after warm-up time of the Breathing System (Section 9.8.6).
9.13 Alarms
Self-test: Self-testing of alarm system functions (alarm light, speaker,
and buzzer) is performed when A5 System is powered on.
Alarm Indicators: Audible: speaker / buzzer
Visual: alarm light and on-screen alarm messages
(Audible and visual alarms comply with the requirements of
IEC 60601-1-8.)
Alarm Categories: Physiological alarms: three levels (high, medium, low)
Technical alarms: three levels (high, medium, low)
Sound Levels: 10 alarm sound levels, adjustable (levels 1 to 10)
Alarm Status: Normal Status: all alarms are functioning properly
Silence Status: silenced alarms do not produce alarm audio;
only new alarms produce alarm audio
Sound Pressure levels (normal ≤ 60 dBA
operation without alarm): Measured from the patient's head location at 1 meter
height, 1 meter from the front of the unit, and 1 meter to the
left of the unit.
• Manual, Inspiratory flow: flow rate = 0.5 L/s @ 0.59 kPa resistance
• Manual, Inspiratory flow: flow rate = 1.0 L/s @ 0.24 kPa resistance
• Manual, Expiratory flow: flow rate = 0.5 L/s @ 0.21 kPa resistance
• Manual, Expiratory flow: flow rate = 1.0 L/s @ 0.43 kPa resistance
• Auto, Inspiratory flow: flow rate = 0.5 L/s @ 0.23 kPa resistance
• Auto, Inspiratory flow: flow rate = 1.0 L/s @ 0.58 kPa resistance
• Auto, Expiratory flow: flow rate = 0.5 L/s @ 0.44 kPa resistance
• Auto, Expiratory flow: flow rate = 1.0 L/s @ 0.20 kPa resistance
NOTE: Using accessories, sensors and cables other than those specified may
result in increased electromagnetic emission or decreased
electromagnetic immunity of the equipment.
NOTE: The anesthesia machine or its components should not be used adjacent
to or stacked with other equipment. If adjacent or stacked use is
necessary, the anesthesia machine or its components should be
observed to verify normal operation in the configuration in which it will
be used.
NOTE: The anesthesia machine needs special precautions regarding EMC and
needs to be installed and put into service according to the EMC
information provided below.
NOTE: Other devices may interfere with this equipment even though they
meet the requirements of CISPR.
NOTE: When the input signal is below the minimum amplitude provided in
technical specifications, erroneous measurements could result.
WARNING: Use of accessories, transducers and cables other than those specified or
provided by the manufacturer of this device could result in increased
electromagnetic emissions or decreased electromagnetic immunity of
this device and result in improper operation.
WARNING: Use of this device adjacent to or stacked with other device should be
avoided because it could result in improper operation. If such use is
necessary, this device and the other device should be observed to
verify that they are operating normally
3 .5 3 .5 7
d = P d = P d = P
V 1 E 1 E1
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.20 1.20 2.30
10 3.80 3.80 7.30
100 12.00 12.00 23.00
For transmitters at a maximum output power not listed above, the recommended separation distanced in
meters (m) can be determined using the equation applicable to the frequency of the transmitter, where P
is the maximum output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
NOTE: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE: At 80 MHz and 800 MHz, the separation distance for the higher
frequency range applies.
ESSENTIAL
PERFORMANCE
ANESTHESIA SYSTEM'S ESSENTIAL
PERFORMANCE TESTED DURING
EMC IMMUNITY CRITERIA DURING EMC
TESTS IMMUNITY
1. Oxygen flow under 1.1 Oxygen 1.1.1 Oxygen supply No false O2 supply failure
all conditions except supply failure failure protection alarm shall be activated and
the failure of the protection device device the fresh gas flow shall be
oxygen supply maintained when the O2
(pipeline or cylinder) supply pressure is within the
to the anaesthetic rated input pressure range.
workstation or the
generation of a 1.2 Interruption 1.2.1 Power The anesthesia system can run
technical alarm of the electrical management on AC power supply and
condition power supply battery supply, and
1. Battery in Use alarm of low
priority shall be indicated only
in case of AC power supply
failure.
2. The ventilation shall be
maintained, and the control
and monitoring accuracy shall
meet the requirements of the
specification.
3. The fresh gas flow shall be
maintained, and the accuracy
shall meet the requirements of
the specification.
1.3 Oxygen flush / /
ESSENTIAL
ANESTHESIA SYSTEM'S ESSENTIAL PERFORMANCE
TESTED DURING
PERFORMANCE EMC IMMUNITY CRITERIA DURING EMC
TESTS IMMUNITY
2. Delivery of a non- 2.1 Alarm / /
hypoxic gas mixture to condition for
the patient or power supply
generation of a failure
technical alarm 2.2 Internal 2.2.1 Battery power The residual capacity of
condition
electrical power supply battery power can be
source indicated normally when
battery power works.
2.3 Protection 2.3.1 Control and Control accuracy:
against monitoring accuracy Tidal volume: 30±10 ml
hazardous Breath rate: 30± 1 bpm or ± 10
output % of the set value, whichever is
greater
Monitoring accuracy:
Tidal volume: 30±10 ml
Breath rate: 30± 1 bpm or ± 10
% of the set value, whichever is
greater
WARNING: Use only accessories specified in this chapter. Using other accessories
may cause incorrect measured values or equipment damage.
WARNING: Check the accessories and their packages for damage. Do not use them
if any sign of damage is detected.
WARNING: At the end of its service life, the equipment, as well as its accessories,
must be disposed of in compliance with the guidelines regulating the
disposal of such products, and in accordance with local regulations for
contaminated and biologically hazardous items.
The following accessories are designed for the A5 Anesthesia System. The use of other accessories is
not recommended. To place an order for these or other accessories, contact Customer Service at +86
755 26582479 / 26582888 or order accessories online at www.mindray.com.
A.2 AG Accessories
PART NUMBER DESCRIPTION
125-000005-00 DRYLINE I, Watertrap (adult/pediatric, reusable, 3-slot)
125-000006-00 DRYLINE I, Watertrap (neonate, reusable, 3-slot)
115-043017-00 Sampling line (adult/pediatric, disposable)
115-043018-00 Sampling line (neonate, disposable)
115-043020-00 Airway adapter (straight, disposable)
115-043021-00 Airway adapter (elbow, disposable)
6800-30-50842 Multi-gas module with accessory kit (3-slot)
115-016612-00 O2 Port Cover Kit
A.9 Vaporizers
PART NUMBER DESCRIPTION
0992-00-0148 Sevoflurane Vaporizer with Quick Fill Adapter
0004-00-0100 Sevoflurane Quick Fill Bottle Adapter
0992-00-0149 Isoflurane Vaporizer with Fill Adapter
0004-00-0101 Isoflurane Fill Bottle Adapter
115-020218-00 Three vaporizer mount
040-001997-00 Desflurane Vaporizer
Breathing System......................................................................................................................................................................................B-2
O2 Sensor .......................................................................................................................................................................................................B-3
Battery..............................................................................................................................................................................................................B-3
The following spare parts are designed for the A5 Anesthesia System. The use of other spare parts is
not recommended. To place an order for these or other spare parts, contact Customer Service at +86
755 26582479 / 26582888 or order spare parts online at www.mindray.com.
B.6 O2 Sensor
PART NUMBER DESCRIPTION
040-001270-00 O2 Sensor, A Series
801-0631-00102-00 O2 Sensor Cable and Housing, A Series
801-0631-00091-00 O2 Sensor Cable, A Series
B.7 Battery
PART NUMBER DESCRIPTION
115-018012-00 Lithium-ion Battery
Waveform/Spirometry Tabs................................................................................................................................................................C-2
Setup Menu..................................................................................................................................................................................................C-5
Demographics ............................................................................................................................................................................................C-9
Current selection
saved when
OBJECT RANGE DEFAULT powered off
Waveform/Spirometry Waveform tab, Spirometry Waveform tab No
Tab tab
Spirometry Tab: Loop Pressure - Volume, Flow - Pressure - Volume No
Type Volume, Pressure- Flow
Spirometry Tab: Save Reference, Baseline Reference No
Loop
Spirometry Tab: Show Off, Baseline, [time] Off No
Reference
Spirometry Tab: Pressure - Volume, Flow - Pressure - Volume No
Review Loops: Loop Volume, Pressure- Flow
Type
Current selection
saved when
PARAMETER RANGE DEFAULT UNIT powered off
Peak High The greater of 10 and Adult: 50 cmH2O Yes
(Paw Low+1) to 100 Pediatric: 40
Step: 1 Infant: 40
Peak Low 0 to the lesser of 70 and Adult: 10 cmH2O Yes
(Paw High–1) Pediatric: 8
Step: 1 Infant: 8
MV High The greater of 0.2 and Adult: 12 L/min Yes
(MV Low+0.1) to 25 Pediatric: 6
Step: 0.1 Infant: 6
MV Low 0 to the lesser of 20 and Adult: 1 L/min Yes
(MV High–1) Pediatric: 1
Step: 0.1 Infant: 0.2
FiO2 High The greater of 21 and Off % Yes
(FiO2 Low+1) to 100,
Off
Step: 1
FiO2 Low 18 to the lesser of 98 18 % Yes
and (FiO2 High–1)
Step: 1
EtCO2 High Off, 2 to 99 Adult: 50 mmHg mmHg, % Yes
Step: 1 Pediatric: 50 kPa
mmHg
Infant: 45mmHg
EtCO2 Low Off, 0 to 97 Adult: 25 mmHg mmHg, % Yes
Step: 1 Pediatric: 25 kPa
mmHg
Infant: 30
FiCO2 High Off, 1 to 99 4 % Yes
Step: 1
EtN2O High Off, (Low+2) to 100 55 % Yes
Step: 1
EtN2O Low Off, 0 to (High-2) 0 % Yes
Step: 1
Current selection
saved when
PARAMETER RANGE DEFAULT UNIT powered off
EtEnf Low Off, 0.0 to (High-0.2) 0 % Yes
Step: 0.1
FiEnf High Off, (Low+0.2) to 5.0 2 % Yes
Step: 0.1
FiEnf Low Off, 0.0 to (High-0.2) 0 % Yes
Step: 0.1
Etlso High Off, (Low+0.2) to 5.0 3 % Yes
Step: 0.1
Etlso Low Off, 0.0 to (High-0.2) 0 Yes
Step: 0.1
Filso High Off, (Low+0.2) to 5.0 2 % Yes
Step: 0.1
Filso Low Off, 0.0 to (High-0.2) 0 % Yes
Step: 0.1
EtSev High Off, (Low+0.2) to 8.0 6 % Yes
Step: 0.1
EtSev Low Off, 0.0 to (High-0.2) 0 Yes
Step: 0.1
FiSev High Off, (Low+0.2) to 8.0 5 % Yes
Step: 0
FiSev Low Off, 0.0 to (High-0.2) 0 % Yes
Step: 0.1
EtDes High Off, (Low+0.2) to 18.0 8 % Yes
Step: 0.1
EtDes Low Off, 0.0 to (High-0.2) 0 % Yes
Step: 0.1
FiDes High Off, (Low+0.2) to 18.0 6 % Yes
Step: 0.1
FiDes Low Off, 0.0 to (High-0.2) 0 % Yes
Step: 0
EtO2 High Off, (Low+0.2) to 100 88 % Yes
Step: 1
EtO2 Low Off, 10 to (High-2) Off % Yes
Step: 1
CO2 Apnea 10 sec, 15 sec, 20 sec, 30 sec, Yes
Delay Time 25 sec, 30 sec, 35 sec,
40 sec
Infant watertrap:
Low(70 ml/min),
Med(90 ml/min),
High(120 ml/min)
Current selection
saved when powered
PARAMETER RANGE DEFAULT off
Display Tab: Waveform No module: Flow Yes
Display Volume, Flow
AG module:
Volume, Flow, AA, O2,
N2O
Current selection
saved when powered
PARAMETER RANGE DEFAULT off
System Tab: Network: None, HL7, MR-WATO, None Yes
This Machine: Configure Philips
Serial: Protocol
System Tab: Network: 10 sec, 30 sec, 1 min, 5 1 min Yes
This Machine: Configure min, 30 min, 1 hour, 2
Serial: Interval hour , 6 hour, 12 hour,
24 hour
System Tab: Network: 2, 1. 1 Yes
This Machine: Configure
Serial: Stop Bits
System Tab: Network: 10 sec, 30 sec, 1 min, 5 1 min Yes
Network Protocol: min, 30 min, 1 hr, 2 hr, 6
Configure HL7: Interval hr, 12 hr, 24 hr
System Tab: Network: — 192.168.23.200 Yes
Network Protocol:
Configure HL7:
Destination IP
System Tab: Network: 0 - 65535 1550 Yes
Network Protocol:
Configure HL7: Port
System Tab: Network: Most Recent, 02.02.01 Most Recent Yes
Network Protocol: to 02.10.00, 02.00.00,
Configure HL7: Set HL7 01.05.02, 01.00.00 to
Compatibility 01.05.01, None
System Tab: Network: On, Off Off Yes
Network Protocol:
Configure HL7: Send
Waveforms
System Tab: Network: On, Off Off Yes
Network Protocol:
Configure HL7: Send
Alarms
System Tab: Network: On, Off Off Yes
Network Protocol:
Configure HL7: Send
Alarms Ack.
System Tab: Network: On, Off Off Yes
Network Protocol: MD2
System Tab: Network: — 192.168.23.99 Yes
Network Protocol:
Configure MD2:
Destination IP
System Tab: Network: — 6678 Yes
Network Protocol:
Configure MD2: Port
System Tab: Network: 10 sec, 30 sec, 1 min, 5 Off Yes
SNTP Protocol: Interval min, 30 min, 1 hr, 2 hr, 6
hr, 12 hr, 24 hr
Current selection
saved when powered
PARAMETER RANGE DEFAULT off
System Tab: Network: 0 - 255 132.163.4.103 Yes
SNTP Protocol: Primary
Server IP
System Tab: Network: 0 - 255 210.72.145.44 Yes
SNTP Protocol:
Secondary Server IP
Current selection
saved when powered
PARAMETER RANGE DEFAULT off
Alarm Volume level 1-10 3 Yes
System Alerts level 1-10 3 Yes
Volume
Event Log: Filter High, Medium, Low, All, All Yes
Informational
Display Interval 1 min, 5 min, 10 min, 1 min, 30 1 min Yes
min, 1 hr, 2 hrs
Display Group Fresh Gas, Gas, Ventilation, All All Yes
Current selection
saved when powered
PARAMETER RANGE DEFAULT off
Day 1-31 1 Yes
Month 1-12 1 Yes
Year 1900-2099 2009 Yes
Hour — 00 (24 hr) Yes
12 am (12 hr)
Minute 00-60 00 Yes
AM/PM AM/PM AM Yes
12/24 hour 12, 24 12 Yes
Date format YYYY/MM/DD, MM/DD/YYYY, YYYY/MM/DD Yes
DD/MM/YYYY
Daylight Savings On, Off Off Yes
Time
C.6 Demographics
PARAMETER RANGE DEFAULT
Patient ID — —
Bed — —
First Name — —
Room — —
Last Name — —
Point of Care — —
DOB — —
Age — —
Weight(Lbs.) — —
Facility — —
Ventilation Modes
PARAMETER VCV SIMV-VC PCV SIMV-PC PS MANUAL
Vt Range: Range: — — — —
20 to 1500 mL 20 to 1500 mL
Step: 1 Step: 1
Defaults: Defaults:
Adult: 600 mL Adult: 600 mL
Pediatric: 120 mL Pediatric: 120 mL
Infant: 20 mL Infant: 20 mL
VtG — — Range: — — —
20 to 1500 mL
Step: 1
Default:
Off
VG — — Default: — — —
046-002773-00
Off
Rate Range: 4 to 100 bpm Range: 4 to 100 bpm Range: 4 to 100 bpm Range: 4 to 100 bpm — —
Step: 1 bpm Step: 1 bpm Step: 1 bpm Step: 1 bpm
Defaults:
Adult: 4 bpm
Pediatric: 6 bpm
Defaults: Defaults:
Adult: 2.0 sec Adult: 2.0 sec
Pediatric: 1.0 sec Pediatric: 1.0 sec
Infant: 1.0 sec Infant: 1.0 sec
Pinsp — — Range: Range: — —
PEEP+5 to 70 cmH2O PEEP+5 to 70 cmH2O
Step: 1 cmH2O Step: 1 cmH2O
Defaults: Defaults:
Adult: 15 cmH2O Adult: 15 cmH2O
Pediatric: 10 cmH2O Pediatric: 10 cmH2O
Infant: 10 cmH2O Infant: 10 cmH2O
Tpause Range: Range: — — — —
Off, 5% to 60% Off, 5% to 60%
Step: 1% Step: 1%
046-002773-00
Defaults: Defaults:
Adult: 50 cmH2O Adult: 50 cmH2O
Pediatric: 40 cmH2O Pediatric: 40 cmH2O
Infant: 20 cmH2O Infant: 20 cmH2O
PlimVG — — Range: 5 - 100 — — —
cmH2O
Step: 1 cmH2O
Default:
Pinsp
PEEP Range: Range: Range: Range: Range: —
Off, 3 to 30 cmH2O Off, 3 to 30 cmH2O Off, 3 to 30 cmH2O Off, 3 to 30 cmH2O Off, 3 to 30 cmH2O
Step: 1 cmH2O Step: 1 cmH2O Step: 1 cmH2O Step: 1 cmH2O Step: 1 cmH2O
Ventilation Modes
Default: Off Default: Off Default: Off Default: Off Default: Off
C - 11
C - 12
Ventilation Modes
PARAMETER VCV SIMV-VC PCV SIMV-PC PS MANUAL
ΔP — Range: — Range: Range: —
3 to 50 cmH2O 3 to 50 cmH2O 3 to 50 cmH2O
Step: 1 Step: 1 Step: 1
Default: 0.2 sec Default: 0.2 sec Default: 0.2 sec Default: 0.2 sec
PS — Range: On, Off — Range: On, Off — —
Step: — Step: —
Default: Off
Alarm — — — — — Range: On, Off
Step: —
Default: On
Default:
5.0 sec (adult)
3.0 sec (Pediatric)
2.0 sec (Infant)
* The Tslope setting is an approximation. The exact waveform shape may not be realized under certain clinical scenarios.
Parameters and Factory Defaults Linked Ventilation Parameter
4 ≤ Rate ≤ 100
Vt
20 ≤ Vt ≤ 1500
Plimit Plimit ≥ PEEP+5
10 ≤ Plimit ≤ 100
SIMV-VC Rate
4 ≤ Rate≤ 100
Vt
20 ≤ Vt ≤1500
ΔP ΔP ≤ Plimit-PEEP
3 ≤ ΔP ≤ 50
Plimit Plimit ≥ PEEP+5
Plimit ≥ ΔP+PEEP
10 ≤ Plimit ≤ 100
VENTILATION Parameter
MODE Parameter Relationship Equation (s)
PCV Rate
4 ≤ Rate ≤ 100
VtG If VtG is not Off.
20 ≤ Vt ≤ 1500
Pinsp Pinsp ≥ PEEP+5
5 ≤ Pinsp ≤ 70
PlimVG PlimVG ≥ PEEP+5
5 ≤ PlimVG≤ 100
SIMV-PC Rate
4 ≤ Rate ≤ 100
Pinsp Pinsp ≥ PEEP+5
5 ≤ Pinsp ≤ 70
NOTE: Even when the PlimVG, Pinsp, or ΔP parameters are inactive, they are
restricted to the parameter relationship equations.
$*66
)ORZ
)ORZ
3
2
12
1&
3
)ORZ
)ORZ
)ORZ
)ORZ
12
1&
E.1 Abbreviations
ABBREVIATION DESCRIPTION
AA anesthetic agent
AG anesthetic gas
AGSS anesthetic gas scavenging system
APL airway pressure limit
Apnea Ti inspiratory time for apnea backup breaths
BTPS body temperature and pressure, saturated
C compliance (Cdyn)
CO2 carbon dioxide
Des desflurane
ENF enflurane
Et end-tidal
EtAA end-tidal anesthetic agent
EtCO2 end-tidal carbon dioxide concentration at expiration
EtDES end-tidal desflurane concentration at expiration
EtENF end-tidal enflurane concentration at expiration
EtHAL end-tidal halothane concentration at expiration
EtISO end-tidal isoflurane concentration at expiration
EtN2O end-tidal nitrous oxide concentration at expiration
EtO2 end-tidal oxygen concentration at expiration
EtSEV end-tidal sevoflurane concentration at expiration
EUI extended unique identifier
Fi fractional concentration
FiAA fractional concentration of anesthetic agent in inspired gas
FiCO2 fractional concentration of carbon dioxide in inspired gas
FiDES fractional concentration of desflurane in inspired gas
FiENF fractional concentration of enflurane in inspired gas
FiHAL fractional concentration of halothane in inspired gas
FiISO fractional concentration of isoflurane in inspired gas
FiN2O fractional concentration of nitrous oxide in inspired gas
FiO2 fractional concentration of oxygen in inspired gas
FiSEV fractional concentration of sevoflurane in inspired gas
Flow flow
HAL halothane
I:E ratio of inspiration time to expiration time
ISO isoflurane
MAC mean aveolar concentration
MEAN mean pressure
Min Rate minimum breath rate
MV minute volume
N2O nitrous oxide
O2 oxygen
ABBREVIATION DESCRIPTION
Pinsp pressure control level of inspiration
Plimit pressure limit level
PlimVG pressure limit level of volume guarantee
PAW airway pressure
PCV pressure control ventilation
PEAK peak pressure
PEEP positive end-expiratory pressure
PLAT plateau pressure
PS pressure support
ΔP pressure support level added to PEEP
R resistance
Rate breath rate
SEV sevoflurane
SIMV-PC synchronized intermittent mandatory ventilation - pressure control
SIMV-VC synchronized intermittent mandatory ventilation - volume control
SP Spontaneous breathing
Tinsp time of inspiration
Tpause percentage of inspiratory plateau time in inspiratory time
Tslope time for the pressure to rise to target pressure
Trigger trigger sensitivity
Vt tidal volume
VtG tidal volume guarantee
VCV volume control ventilation
VG volume guarantee control
E.2 Symbols
SYMBOL DESCRIPTION SYMBOL DESCRIPTION
- minus > greater than
% percent ≤ less than or equal to
/ per, divide, or ≥ greater than or equal to
≈ approximately ± plus or minus
^ power × multiply
+ plus © copyright
= equal to ™ trademark
< less than ® registered trademark
02
01
ATTENTION!
NUMBER DESCRIPTION
01 Total Flowmeter: The total flowmeter is calibrated based on 100% O2. The accuracy of
the flowmeter may degrade with other gas or mixed gas.
When viewing the readings on the total flowmeter, keep your visual angle at the same
level of the float. The reading of the scale may vary when viewed at a different angle.
If the readings shown on the electronic flowmeters differ from that on the total
flowmeter, the electronic flowmeter will prevail and the total flowmeter is an
approximate value.
02 Only vaporizers with Selectatec Interlock-Systems may be used with the A5 unit.
Use vaporizers compliant to ISO 80601-2-13. See section A.9 (page A-4) "Vaporizers".
Refer to the manufacturer’s vaporizer Instructions For Use for filling or draining the
vaporizer and other information.
Use care in lifting and manipulating vaporizers during the mounting process as their
weight may be greater than expected, based on their size and shape.
03 04
05
ATTENTION!
NUMBER DESCRIPTION
03 Each auxiliary outlet is rated at 100 to 120 VAC @ 60 Hz.
04 Individual outlet current is limited to 3 A. Total mains output current is limited to 10 A.
05 Sample Line Exhaust Gas Inlet: Inlet for waste gas from an optionally attached gas
module. Merges with the AGSS connector that connects to the AGSS.
06
08
07
ATTENTION!
NUMBER DESCRIPTION
06 Auxiliary O2/Air Gas Outlet: Nozzle (barbed connector) for auxiliary O2/Air output.
Combines the auxiliary O2/Air flowmeters into a single output.
07 Maximum supporting weight: 25 kg at a maximum distance of 0.31 m
08 Warning: Hot
09
11
10
ATTENTION!
NUMBER DESCRIPTION
09 Maximum supporting weight: 25 kg at a maximum distance of 0.31 m
10 Maximum supporting weight: 25 kg at a maximum distance of 0.31 m
11 Maximum supporting weight: 25 kg at a maximum distance of 0.31 m
Rear
12
Front
ATTENTION!
NUMBER DESCRIPTION
12 Top Shelf: 40 kg MAX. 88 lbs MAX.
14
15
13
ATTENTION!
NUMBER DESCRIPTION
13 Do not push down on the bag arm forcefully or hang heavy objects onto it. Excessive
weight may bend and damage the bag arm.
14 Autoclavable up to 134°C. Polyphenylsulfone (PPSU).
15 APL Valve: The APL valve and PAW gauge numerics are for reference only. Calibrated
patient airway pressure is displayed on the user interface.
17
16
ATTENTION!
NUMBER DESCRIPTION
16 134°C >PPSU<. Autoclavable up to 134°C.
Operating the A5 with a full water trap in the breathing system block does not allow
the water to condense appropriately. The trap should be removed and emptied when
filled with water.
Operating without a water trap will cause the Leak Test to fail.
17 Bellows Dome: The bellows dome is a transparent cover with graduation marks from
300 to 1500. These marks are for qualitative purposes only. Tidal volume (VT) should
be read exclusively from the display of the user interface. Delivered tidal volume (VT)
is a combination of bellows displacement and fresh gas flow.
Washout Procedure for Malignant Hyperthermia Susceptible Patients with A5 Anesthesia Delivery
Systems ........................................................................................................................................................................................................... F-2
References..................................................................................................................................................................................................... F-4
MH has had a reported mortality rate decrease from 70%-80% to less than 5% if preventive measures
and effective management are adopted.2 The early therapy requires immediate discontinuation of all
the triggering agents, adequate oxygenation and ventilation, institution of aggressive cooling
measures, administration of dantrolene sodium, and appropriate treatment for hyperkalemia.
Ultimately, the only effective treatment for an MH crisis is the intravenous administration of
dantrolene sodium and supportive therapy to combat the symptoms.1,2
1. Turn off and remove all the vaporizers from the anesthesia system to prevent their
inadvertent use.
2. Remove the carbon dioxide absorbent, breathing bag and the entire patient breathing
circuit, filters, sampling line, water trap, and airway adapter and replace with new circuit and
parts, connect a new breathing bag or test lung to the patient Y-piece.
3. Ventilate for a minimum of 40 minutes using mechanical ventilation with the following
settings, 700 ml tidal volume, I:E ratio of 1:2, 12 breaths/minute, PEEP Off, and oxygen fresh
gas flow rate of 15 L/min.
4. Upon completion of the 40 minute flush, remove the patient breathing circuit. Allow the
bellows to deflate completely. Replace with a new patient breathing circuit, including bag
and new carbon dioxide absorbent. Perform the pre-operative checkout.
5. Maintain the oxygen fresh gas flow rate of 15 L/min throughout the case to functionally
create a non-rebreathing system and minimize rebound of volatile concentration at low
fresh gas flow rates.
Recommendations (4 alternatives):
F.4 References
1. Hopkins PM. Malignant hyperthermia: pharmacology of triggering. Br J Anaesth. 2011 Jul;
107 (1): 48-56.
2. Kim DC. Malignant hyperthermia. Korean J Anesthesiol. 2012 Nov; 63 (5): 391-401.
3. Kim TW, Nemergut ME. Preparation of modern anesthesia workstations for malignant
hyperthermia-susceptible patients: a review of past and present practice. Anesthesiology.
2011 Jan;114 (1):205-212.
4. Schuster F, Johannsen S, Schneiderbanger D, Roewer N. Evaluation of suspected malignant
hyperthermia events during anesthesia. BMC Anesthesiology 2013, 13: 24.
5. Gunter JB, Ball S, Tan-Win S. Preparation of the Drager Fabjus anesthesia machine for the
malignant -hyperthermia susceptible patient. Anesth Analg 2008; 107: 1936-45.
6. Reber A, Schumacher P, Urwyler A. Effects of three different types of management on the
elimination kinetics of volatile anaesthetics. Implications for malignant hyperthermia
treatment. Anaesthesia 1993; 48: 862-5.
7. Crawford MW, Prinzhausen H, Petroz GC. Accelerating the washout of inhalational
anesthetics from the Drager Primus anesthetic workstation. Anesthesiology 2007; 106:289-
94.
8. Prinzhausen H, Crawford MW, O'Rourke J, Petroz GC. Preparation of the Drager Primus
anesthetic machine for malignant hyperthermia-susceptible patients. Can J Anesth 2006; 53:
885-90.