AMB SAMPLE PREPARATION Sample preparation is the single most important item for accurate results. The sample must be prepared so the whole of the sample is heated uniformly and the moisture is near the surface of the material. For powders no further preparation is necessary. For most other materials the sample should be ground, chopped, grated or otherwise made as fine as possible. Liquids or paste may need to be spread onto glass fibre pads to distribute them evenly and thinly. Some materials that may splatter as the y are heated (oils, butter) may need to be spread on the glass fibre pads or mixed with dry sand. Trial and error is often the only method to find the best method for the material. When the sample is placed on the weighing pan it should be as thin as possible, however enough material should be used to assure good repeatability of the results. Typically samples of 3 to 8 grams are used. Some materials will require special preparation. For example when testing cheese, if the cheese is very soft it may be necessary to use dry sand as a buffer material so it does not splatter or form a skin. To do this it would be necessary to prepare the sand by drying it in an oven until you were certain no moisture remains. Put the sand on the sample pan and tare the balance. Mix the sand with the cheese so you have smooth mixture, spread this mixture on the sample pan and then do the test. An alternative to this preparation is to spread the cheese on dry spun glass gauze to assist in spreading the heat and preventing splatter or skins forming. Remember to tare the balance with the gauze in the weighing pan for accurate results. Other materials will hold water within them in a way that is not suitable for this type of test. For example sugar will bind the water to the sugar molecules so strongly that heat will not remove the water before the sugar is burned. Most plastics also are not suitable for this type of method due to the very low percent moisture and the heat required to extract this moisture.
page 2
X:\COMPS\rich\web tech content\AMB Application revB.doc
12/12/2001
AMB PARAMETERS When initially setting the parameters for a material often it is a matter of trial and error to find the most suitable settings. Initially use mode 7. This will allow the test to run without interruption until you stop it. Thus if the material dries slowly the balance will not falsely determine the test is over. The strobe time can initially be set to 15 seconds. If you have some experience with an oven method or other moisture balances the temperature should be set to these same values. However if you do not have any experience then the initial temperature setting is a guess at first. Try using 120C initially and then increase it if there are no signs that the material is burned , or decrease it if the material burns or goes excessively brown. Some material need to be dried at lower temperatures if they form a skin over the surface during drying. We find it helpful to have the balance connected to a PC for data collection or to a printer during these test to help determine the best settings for test. Begin the test and watch the output from the scale. After a few minutes the moisture should increase. After some time the %moisture will become stable for a long period of time. Terminate the test at this time. Repeat the test at other temperatures, generally try to use the highest temperature possible that does not damage the material. After you are getting consistent results with 10-15g samples you can try reducing the size of the samples. Often good results are obtained using samples of 3-8g. By observing the results the time and temperatures needed to finish test it is possible to determine the typical time of a test, the typical %moisture and the strobe time that will be needed to determine the automatic stopping point. KNOW THE MATERIAL As described above some materials are not well suited for this method of moisture determination. In addition it is important to know enough about the materials so that you do not endanger or inconvenience others. For example some solve nts have very low flash points that are not suitable for drying with heat sources that can reach 100C. Other materials will cause odour or fumes that will be dangerous or objectionable, consider drying fertiliser slurry without using a vent to extract the odour.
page 3
X:\COMPS\rich\web tech content\AMB Application revB.doc
12/12/2001
TEMPERATURE OF THE SAMPLE When the samples are dried the temperature at the sample may need to be known. The AMB Moisture balances have a location to insert a second temperature probe with a 3mm shaft into the samples. When the samples are tested using Mode 7 as before the 2 nd temperature probe can be used to measure the temperature within the sample. The results from these test should be discarded as the probe touching the samples will affect the readings. When the temperature of the sample has been determined a correction factor can be stored in one of the 9 SubSt memories for future use. Refer to Appendix A in the User Manual for details of setting the values for SubSt1 to SubSt9. TEMPERATURE CONTROL within the AMB BALANCES The temperature of the weighing changer is measured by a sensor just above the samples. As the temperature approaches the final value the lamps will be cycled on/off to maintain the temperature at the set point. The initial increase in temperature will happen q uickly, it will usually take about 1 minute to gain the first 90% of the final temperature and then another 30 seconds to reach the final temperature. This method helps avoid the temperature overshooting the set point and thus avoids damaging the sample .
RS-232 DATA FROM THE AMB BALANCES The AMB Balances will output the current results at the same interval as the strobe time. These results can be used to plot the progress of a test or to keep records of a test. The data can be collected using many different RS-232 interface programs or a program that runs with Windows is available from Adam Equipment. Contact the Sales Department at Adam Equipment or your Dealer for more details. EFFECT OF CALIBRATION ON MOISTURE DETERMINATION The AMB moisture balance can be used as a laboratory balance if it is calibrated with a mass standard of OIML Class F2 or better. However the scales do not need to be calibrated if they are only used for moisture determination. The results obtained from a moisture test are not dependent upon the accuracy of the balance as the results are a ratio of 2 weights.
page 4
X:\COMPS\rich\web tech content\AMB Application revB.doc
12/12/2001
APPENDIX B
GUIDANCE NOTES The following guide notes are offered as examples of temperature, display update time and expected results when drying various substances. The preparation, distribution and amount of product can all significantly effect the results. It is important to find suitable procedures to process the material, usually by trial and error. In general it is recommended that a temperature of 100C, a time interval of 10 seconds, and a sample sized 2-5g be selected for initial test. After a number of tests with these conditions this will provide a basis for comparing other times, temperatures or sample preparations. The items in the following table are taken from actual test results. The user of the AMB balance will need to determine the best procedures for their samples.
page 5
X:\COMPS\rich\web tech content\AMB Application revB.doc
12/12/2001
BUTTER CEMENT CHARCOAL CHOCOLATE POWDER COCOA POWDER COFFEE CREAMER POWDER CORNFLAKES DRY APPLE PULP DUST FLOUR GROUND COFFEE MARGARINE MAYONNAISE MILK MILK CHOCOLATE MILK POWDER MUSTARD PAPER POLYAMIDE POTATO FLAKES POWDERED SOUP RED WINE SLIME SUGAR SUNFLOWER OIL WET APPLE WHITE GLUE YOGHURT
15 15 10 5 20 15
2-4 5-8 5-10 8-10 2-3 3-4 1-2 2-3 2-5 2-4 2-3 2-4 2-5 3-4 2-3 3-5 11-12 4-5 10-14 5-8 2-5 2-3
120 100 104 130 106 138 138 120 106 90 130 106 138 106 80 100 130 138 138 100 136 110
15 10 10 10 5 20 20 15 15 15 20 20 20 15 15 15 15 15 20 10 15 15
9.7 76.5 7.3 12.5 2.8 16 56.5 88 1.3 5 76.4 6.4 2 6.9 3 97.4 80 11.9 0.1 7.5 54.3 86.5
0 0.1 0.3 0.1 0.1 0.1 0.4 0.2 0.1 0.2 0.7 0.1 0.2 0.1 0.2 0.1 0.1 0 0 0.1 0
5-7 10-15 8-15 4-5 4 10 10 6-8 3.5 6 10 10 75 7.5 4.5-7 15-20 90 10 2 5-10 6-8 4.5-6.5
This list will be extended as information becomes available. If you wish to supply any information that may help others, please contact Adam Equipment or your dealer.