Mammomat Novationdr
Mammomat Novationdr
Mammomat Novationdr
SP
Installation Instructions
FFDM-S Service Manual DRC
© Siemens AG 2004
The reproduction, transmission or
use of this document or its contents
is not permitted without express
written authority. Offenders will be
liable for damages. All rights,
including rights created by patent
grant or registration of a utility
model _or_ design,_are_ reserved.
English
Print No.: SPB7-250.812.02.01.02 Doc. Gen. Date: 02.04
Replaces: N/A
DirectRay® for Siemens
Service Manual
Chapter 1: Introduction
Chapter 3: Installation
Chapter 8: Replacement
Appendix A: Specifications
DirectRay Detector, PXCM, and Interface Card General Specifications ........................ A-1
Electrical......................................................................................................................... A-2
DirectRay Detector .................................................................................................. A-2
DirectRay PXCM...................................................................................................... A-2
Environment ................................................................................................................... A-2
Transit/Storage......................................................................................................... A-2
DirectRay Detector ............................................................................................ A-2
Operating ................................................................................................................. A-2
DirectRay Detector ............................................................................................ A-2
PXCM Connector Locations and Pinouts....................................................................... A-3
DirectRay Detector-to-PXCM Connection ............................................................... A-3
PXCM-to-Generator Connection.............................................................................. A-4
February 5, 2004 v
THIS PAGE IS INTENTIONALLY BLANK.
Intended Use
Digital images can be viewed, stored, or printed. The system provides a digital
image capture capability for radiographic examinations for mammography.
Organization
Chapter 1: Introduction
Provides a description of the DirectRay FFDM-S components.
Chapter 2: Theory of Operation
Explains the sequence of internal operations that occur with system components
during exposures. It also describes the DirectRay technology used in the DirectRay
Detector.
Chapter 3: Installation
Provides guidelines for installing the hardware components of the DirectRay
FFDM-S system
Chapter 4: Calibrating the AEC
Describes how to calibrate the AEC by generating a goal table for a desired
target/filter combination and loading it into the DirectRay Power and X-ray
Control Module (PXCM).
Chapter 5: PXCM Web Tool
Describes Detector- and PXCM-related service tasks using the DirectRay PXCM
Web Tool.
Chapter 6: Maintaining Image Quality
Contains information to help maintain acceptable image quality.
Chapter 7: Troubleshooting
Provides guidelines on how to troubleshoot the DirectRay components.
Chapter 8: Replacement
Provides replacement guidelines for the DirectRay system components, and
explains how to pack the Detector for shipment back to Direct Radiography Corp.
NOTE: Notes provide additional information, such as expanded explanations, hints, or reminders.
CAUTION Cautions point out procedures that you must follow precisely to avoid
damage to equipment, loss of data, or corruption of files in software applications.
WARNING Warnings point out procedures that you must follow precisely to avoid injury
to yourself or others.
GS20_419.cdr
Safety Precautions
• Observe all warnings and cautions in the procedures.
• Follow all safety labels on the equipment.
CAUTION Federal (USA) law restricts this device to sale by or on the order of a physician (or
properly licensed practitioner).
WARNING The DirectRay FFDM-S system includes no user serviceable parts. For service
assistance, contact the vendor who supplied the system.
IEC Regulations
Software used in the DirectRay FFDM-S system was developed using
IEC 60601-1-4 as a guide.
Manufacturer’s Responsibility
This equipment is sold with the understanding that Direct Radiography Corp., its
agents and representatives, do not accept any responsibility for overexposure of
patients or personnel to X-radiation.
Furthermore, Direct Radiography Corp. does not accept any responsibility for
overexposure of patients or personnel to X-radiation generated by this equipment
as a result of poor operating techniques or procedures.
Also, no responsibility is assumed for any unit that has not been serviced and
maintained in accordance with the technical service manual, or which has been
modified or tampered with in any way.
Restrictions on Use
The customer is responsible for assuring that accessories used with the DirectRay
FFDM-S system do not degrade the safety of the system.
Accessories and interfacing equipment used in conjunction with the DirectRay
FFDM-S system and located outside the patient environment may be exempt from
EN60601-1. Such equipment must, however, conform to EN60601-1-1 and
therefore to standards applicable to information technology equipment such as
EN60950 and to applicable electromagnetic compatibility requirements.
Environment
The operating environment is 20° to 35°C (68° to 95°F), 10 to 80% Relative
Humidity (non condensing). Make sure the ventilation openings at the rear of the
detector are not obstructed. For storage and transportation, the temperature range
must stay within 10° to 40°C (50° to 104°F) or permanent damage may occur. Such
damage could be considered misuse and may not be covered under warranty.
Do not autoclave this equipment.
Certification
UL Recognized Component certified to be in compliance with applicable
requirements of:
• UL 2601-1, Second Edition including Amendments 1 and 2
• CAN/CSA C222 No. 601.1-M909
Use of the FFDM-S with equipment not complying with equivalent safety
requirements may lead to a reduced level of safety in the resulting system. There
must be evidence that added equipment has been certified for safety in accordance
with the appropriate IEC 601-1 and/or IEC 601-1-1 harmonized national standard.
Otherwise use within the defined Patient Vicinity is not approved. Testing of the
FFDM-S in combination with the end use equipment shall include leakage current
test, ground continuity tests, temperature tests and other tests as required by the
appropriate standards. The maximum operating temperatures of certain safety
components, as defined in the applicable safety standards, must not be exceeded in
the end use installation to maintain the intended safety. The FFDM-S is intended
to be installed as part of x-ray equipment and serviced only by persons trained and
approved to work on that specific equipment, using the appropriate tools.
Classification
With reference to subclause 5 of IEC 60601-1:
• Protection against electric shock (5.1, 5.2) Class I.
• Applied Parts: Type B Applied Parts.
• Protection against harmful ingress of water (5.3) Ordinary.
• Degree of safety in the presence of flammable anesthetics or oxygen (5.5). Not
suitable for use in the presence of flammable anesthetics or oxygen.
• Mode of Operation (5.6): Continuous.
WARNING Dangerous Voltages are present on some components, printed wiring traces,
and heatsinks. A blown fuse is an indication of a catastrophic failure of circuit
components. Unauthorized replacement may result in a risk of fire.
CAUTION Some of the electronic components used in this product may be damaged
by static electricity or electrostatic discharge. Unpack and handle using appropriate
precautions.
CAUTION Each image detector is provided with a CD-ROM containing data for initial
gain calibration and defect mapping. The gain calibration process must be performed
again when installed at the end user site to compensate for variations in the specific
x-ray tube, breast tray, compression paddle, grid, and all other parts of the image path.
Any dirt, debris, or foreign objects in the path between X-ray source and image
detector elements at time of calibration may adversely affect subsequent images.
Calibration should be repeated at regular intervals to maintain image quality. Defect
mapping can be verified and updated by authorized service personnel.
NOTE: This product must be returned to the manufacturer for proper disposal of the selenium and
related electronic materials.
Environment
The operating environment is 10° to 50°C (50° to 122°F), 10 to 80% Relative
Humidity (non condensing).
Certification
UL Recognized Component certified to be in compliance with applicable
requirements of:
• UL 2601-1, Second Edition including Amendments 1 and 2
• CAN/CSA C222 No. 601.1-M909
Use of the PXCM-S with equipment not complying with equivalent safety
requirements may lead to a reduced level of safety in the resulting system. There
must be evidence that added equipment has been certified for safety in accordance
with the appropriate IEC 601-1 and/or IEC 601-1-1 harmonized national standard.
Otherwise use within the defined Patient Vicinity is not approved. Testing of the
PXCM-S in combination with the end use equipment shall include leakage current
test, ground continuity tests, temperature tests and other tests as required by the
appropriate standards. The maximum operating temperatures of certain safety
components, as defined in the applicable safety standards, must not be exceeded in
the end use installation to maintain the intended safety. The PXCM-S is intended
to be installed within X-ray equipment and serviced only by persons trained and
approved to work on that specific equipment, using the appropriate tools.
Classification
With reference to subclause 5 of IEC 60601-1:
• Protection against electric shock (5.1, 5.2) Class I.
• Protection against harmful ingress of water (5.3) Ordinary.
• Degree of safety in the presence of flammable anesthetics or oxygen (5.5). Not
suitable for use in the presence of flammable anesthetics or oxygen.
• Mode of Operation (5.6): Continuous.
WARNING Dangerous Voltages are present on some components, printed wiring traces,
and heatsinks. A blown fuse is an indication of a catastrophic failure of circuit
components. Unauthorized replacement may result in a risk of fire.
CAUTION Dielectric strength test caution: (Statement from Condor DC Supplies Inc.
regarding their GSM28-5 power supply) "When performing Dielectric Strength Tests,
catastrophic failure of the unit may result if a Dielectric Strength test voltage greater
than 1800 Vac is applied between primary and secondary circuits. The components
providing isolation from primary to secondary cannot be tested while installed in the
power supply without overstressing basic (primary to ground) insulation. All isolating
components are individually 100% tested at 4800 Vac prior to installation."
Technologist Identifies
Patient in System
Technologist Positions
Patient
Technologist
Adjusts Image
Technologist Reject
Adjust Accepts Image?
Yes Is the No
Procedure
Listed? Accept Really No
Reject?
No Yes
Procedure Done?
No
Technologist Closes
Procedure
GS20_410.cdr
NOTE: The PXCM is also referred to as the “brick.” Therefore, you will see the terms “PXCM” and
“brick” used interchangeable throughout this topic.
GEN_RDY_FOR_DR_EXP_ t_rotor_ok
t_reset_dr_det_rdy
t_det_rdy
DR_DET_RDY_FOR_EXP_ t_det_rdy_from_boost
t_grid_dly
X_RAY_ON_ t_gen_rdy_disable_to_xray_off
t_expose
DR_SYST_RDY_
t_det_cycle
t_dly_from_exp_finished
DR_DET_ERR_
GEN_POWERED_ON_
t_boost_to_query t_gen_rdy_to_query Brick Request
post exposure data msg
Code A* Code C
Brick Transmit
Message
t_gen_msg_dly Generator sends
Generator Transmit Code X* Code Z exposure data to brick.
Message
t mess exch
X-ray button
activated
Acc. anode t_boost t_prep_end_dly
GEN_RDY_FOR_DR_EXP_ t_rotor_ok
t_disable_dr_det_rdy
t_det_rdy
DR_DET_RDY_FOR_EXP_ t_det_rdy_from_boost
t_gen_rdy_disable_to_ray_off
X_RAY_ON_ t_grid_dly
pre exp.
t_expose
DR_SYST_RDY_
t_det_cycle
t_dly_from_exp_finished
DR_DET_ERR_
GEN_POWERED_ON_
t_boost_to_query t_gen_rdy_to_query Brick Request
post exposure data msg
Code A* Code B Code C
Brick Transmit
Message t_gen_msg_dly
Generator sends
Generator Transmit Code X* Code S Code Y Code Z exposure data to brick.
Message Code C
t aec setup
t mess exch
PREP_DR_DET_ Generator to PXCM The Generator has started the preparations to perform an
exposure.
Activation:
The Generator activates the signal when the exposure
preparations have started.
Deactivation:
The Generator deactivates the signal when:
DETECTOR (TFT)
MATRIX FRAME
The storage capacitor and charge-collection electrode are configured in such a way
that the electrode forms a “mushroom head” over much of the detector element
area. This unique design is illustrated in Figure 2-7 on page 2-11. The “mushroom”
electrode:
• Shields the FET from the high electric field that is generated by the high
voltage applied to the top electrode
• Shields the FET from signal charges that would collect on the insulation above
the FET
• Increases the charge collection efficiency of each detector element by
maximizing its area
In the configuration shown in Figure 2-7, the mushroom electrode covers 86% of
the total detector element area. The mushroom electrode material blocks the
electrons from injecting into the X-ray absorber above.
The layer above the detector element array is the actual X-ray conversion material.
To achieve sufficient X-ray absorption, a thickness of 500 µm of a-Se is used. This
material has been widely used in xeroradiography, so its electrical properties and
vacuum deposition processes are well understood.
The next layer is an insulating material with a high dielectric strength. Thickness
uniformity, good adhesion to selenium, optical transparency, and a low defect count
to avoid high field breakdowns are some key requirements for this dielectric
blocking layer.
A thin layer of metal is subsequently deposited on the dielectric material, serving
as an electrode to which several kV can be applied to separate the electron-hole
pairs. An additional material is deposited on top of the electrode material to
environmentally isolate the detector.
Electronic Readout
The readout of the charges starts immediately at the end of the X-ray exposure.
While all FETs (which are operated here as simple on–off switches) are not
conducting during the X-ray exposure, a positive gate pulse applied to the first gate
line simultaneously turns on all FETs connected to the first gate line. This causes
the image charge to flow to a bank of charge amplifier modules connected to the
data lines. The net X-ray signal is multiplexed and shifted out to a 14-bit
analog-to-digital converter (ADC) and subsequently stored in computer memory.
The readout of the charges continues row by row until the digital value of the
charge in each storage capacitor has been transferred to memory. At the end of the
readout cycle, a charge erase cycle is initiated to refresh the detector panel for the
next exposure.
The current commercial 139-µm pitch DirectRay Detector has approximately
8 million pixels in the detector element matrix and a readout time of 4 seconds.
Once in computer memory, the image data can be processed and displayed on a
video monitor or routed into a network system.
GS20_414.cdr
2. Remove all of the other packing material covering the inner box in which the
Detector is packaged.
GS20_415.cdr
GS20_416.cdr
5. Remove the packing material inside the inner box, which includes the
temperature logger that is used to ensure that the Detector maintained the
proper temperature will being shipped and stored.
CAUTION Handle the Detector with extreme care! Avoid touching the thin film top
cover that protects the imaging area. The imaging electronics under the front cover
are extremely sensitive. The actual image area starts at the front edge and is
approximately 25 mm inside the other three edges. Any dust or fingerprints in this
area may affect final image quality. Temperature extremes may cause permanent
damage.
G
G IN
A A
IM RE
A
Figure 3-2 shows the Detector connector used for the cable that connects the
Detector to the DirectRay PXCM.
GS20_411.cdr
NOTE: To ensure reliable communications, keep the fiber optic connections covered by protective
caps when not in use.
Figure 3-4 shows the PXCM connectors used for the cable that connects the
Detector to the DirectRay PXCM, and the fiber optic cable to the DirectRay
Interface card.
Figure 3-4. DirectRay PXCM Connectors (for cable to Detector and fiber optic cables to
Interface card)
Connectors
for fiber optic cables
to DirectRay PXCM
(shown with cap
on left connector) GS20_413.cdr
Figure 3-5. DirectRay Interface Card Fiber Optic Connectors (for fiber optic cable to
PXCM)
Handling
• Always handle fiber optic cables delicately and by the connector housing.
• Avoid acute bends in the cable, as this could break the fiber optic link inside
the cable.
Connecting
• Always perform one connection at a time.
• Remove the dust caps on the fiber optic connectors (you may have to rotate the
cap to loosen it).
• Holding the cable by the connector housing, insert the cable connector into the
connector on the device, making sure to align the keying features so the
connectors are properly mated, then turning the connector clockwise to lock it.
• To disconnect the cable connector, turn it counterclockwise, then pull it from
the connector on the device.
Cleaning
• Clean the connectors on both the cable and the devices with a lint-free,
alcohol-dampened cloth.
• Thoroughly wipe the side and end of the ferrule.
• Blow the ferrule with compressed air, or allow it to air dry, before reconnecting
the cable.
• Visually inspect the ferrule for lint.
• When not connected, replace the protective cap onto connector ends.
To exit the tool, open the File menu and choose Quit.
Procedure Options You choose what procedure you want to perform from these three options. Only one can be
selected at a time. When the tool starts, no procedure is selected. You must select one of the
three procedures to continue.
• Determine HVL. Select this procedure to determine the Half Value Layer (HVL) of the X-ray
Generator for a given target/filter combination. For the procedure for determining the HVL,
refer to “Determining the Half Value Layer (HVL)” on page 4-5.
• Determine ESE. Select this procedure to determine the Entrance Skin Exposure (ESE).
This step may be performed without determining the HVL of the X-ray Generator. However,
a valid HVL value must be known. For the procedure for determining the ESE, refer to
“Determining the Entrance Skin Exposure (ESE)” on page 4-8.
• Generate Goal. Select this procedure to generate the goal table for the selected target/filter
combination. This step may be performed without determining the HVL of the X-ray
Generator or the ESE. However, a valid mAs value must be known. For the procedure for
generating the goal, refer to “Generating the Goal Image” on page 4-11.
Target/Filter The desired target/filter combination. The goal table will be named to match (Mo/Mo is
goaltab1, Mo/Rh is goaltab2, W/Rh is goaltab3). Make sure that the Generator setting matches
this selection. Since this selection determines future instructions and calculations, this selection
will be disabled once the first HVL exposure is taken.
AGD Use this to select the desired Average Glandular Dose to be used in calculating the mAs value
for calibration. In a future version, this value will be logged on the end of the goal table file. The
default is 1.5 mGy (150 mR).
kV Use this to indicate the kV to be used during the calibration. This selection in the AEC
Calibration Tool window does not actually change the Generator settings, but is intended to
serve as a reminder to the user what the current Generator settings are. The default is 28 kV.
mAs Use this to indicate the mAs to be used during the calibration. This selection in the AEC
Calibration Tool window does not actually change the Generator settings, but is intended to
serve as a reminder to the user what the current Generator settings are.
Goal Images Use this to select the number of goal images to be taken and averaged together to create the
goal table. Select 1 to 4 images.
Enable Capture button Use this to enable the DRAPI for image capture. If the Status Window is green and displays
“DRAPI Armed”, it is not necessary to use this button. This button will only work when the status
window is yellow and displays “DRAPI Ready”. Clicking on this button will clear the Radiation
Entry field.
Units Use this to select the radiation units to be used. Choices are R/mR or Gy/mGy. When this
selection is changed, the AGD setting will change to convert the value to the current units. The
display in the instructions field after each exposure will also report in the selected units.
Reading Use this field to enter the dosimeter reading as instructed following each exposure for HVL or
ESE determination. Enter the numerical characters in the field and click the Submit button. The
last entry will be cleared whenever the Enable Capture button is pressed.
Status • “DRAPI Not Ready” (white). The DRAPI is processing the previous exposure.
• “DRAPI Armed” (green). The DRAPI is ready for exposure. Press the Generator exposure
buttons to expose.
NOTE: If you already know the HVL and want to proceed with determining the ESE value, go to
“Determining the Entrance Skin Exposure (ESE)” on page 4-8.
NOTE: You may want to make a note of the HVL value, so that you can skip the HVL procedure and
go right to the ESE procedure in future calibrations.
NOTE: If you already know the valid mAs to use and want to proceed to generating the goal image,
go to “Generating the Goal Image” on page 4-11.
The ESE procedure requires the HVL. If the HVL procedure has already been
performed, the tool will already have the HVL value. If you are skipping the HVL
procedure and starting with the ESE procedure, a dialog box displays, prompting
you to enter the HVL value.
The ESE procedure involves taking three exposures. The dosimeter readings for
each exposure are averaged together to determine the ESE for the selected Average
Glandular Dose (AGD). This in turn suggests the mAs that should be used to
generate the goal table.
To determine the mAs value:
1. In the tool’s main window, select the desired AGD value before starting the
ESE determination.
Once the first ESE exposure has been performed, the AGD value cannot be
changed.
2. Select the Determine ESE procedure option.
If the tool has the HVL value stored internally (for example, you just
performed the HVL procedure), go to step 5.
If the tool does not have the HVL value, the HVL Popup dialog box displays.
3. If you know the HVL value, enter the fractional value (for example, 0.3611),
then click OK or press the ENTER key.
If you do not know the HVL value, refer to “Determining the Half Value
Layer (HVL)” on page 4-5.
4. After entering the HVL value, select the Determine ESE procedure option
again.
In the main window, the settings change, based on the HVL and the selected
AGD, and instructional text is displayed.
Note that the Status field turns green and displays the message “DRAPI
Armed.”
NOTE: In most cases, the suggested mAs will not exactly match an mAs value to which the Generator
can be set. In this case, it is recommended that you use the closest lower mAs setting so that the
desired AGD will not be exceeded during future AEC exposures.
NOTE: You can select the Generate Goal option without performing the Determine HVL and
Determine ESE procedures, provided you know the proper mAs to configure the Generator.
In the main window, the settings change and instructional text is displayed.
NOTE: The text messages may be interspersed with DRAPI image frame messages. Also, the sample
messages shown below represent a selected AGD of 1.5 mGy.
Binning the goal image to make goaltab file goaltab1 for Mo/Mo target/filter.
Reading image file... done.
Generating binned image
done.
Writing AEC image (448x512) file... done.
Writing mem image (28x32) file... done.
Image binning complete.
Opening goaltab1 for goaltab input
Opening brick:/tmp/goaltab1 for goaltab output
Reading goaltab1 for goaltab input
Read 896 goaltab entries.
Closing goaltab1
Writing brick:/tmp/goaltab1 for goaltab output
Wrote 3584 bytes via rfs.
Writing the value “15” to brick:/tmp/goaltab1 for AGD output…
Wrote 4 bytes via rfs.
Closing brick:/tmp/goaltab1
Installing goaltab1 to long-term Brick configuration…
(/usr/local/bin/xrayconfig goaltab1 save /tmp/goaltab1)
“save” OK.
Installing goaltab1 to effective and short-term Brick configuration…
(/usr/local/bin/xrayconfig goaltab1 load /tmp/goaltab1)
“load” OK.
Goal Table goaltab1 successfully loaded on Brick.
***************************************
AEC Calibration procedure is complete.
***************************************
If you see the “save OK” message but an error instead of the “load OK” message, be
assured that the goaltab has been stored in the PXCM’s long-term storage and will
be auto-installed on reboot. Simply reboot the PXCM before attempting to take
AEC exposures. For information on rebooting the PXCM, refer to “Rebooting the
PXCM” on page 5-21.
Enter the copy’s name in the Copy file goal tab to: field, then click OK. The
goaltab for Mo/Mo will be copied to a new file on the PXCM, using the file name you
entered.
Any time a user attempts to use a function assigned moderate or high security, a
login prompt will be displayed, to verify that they have authority to use the
function.
Detector Functions
NOTE: The DirectRay Detector is also referred to as the “Array.” Therefore, you will see the term
“Array” used for “Detector” throughout the web tool and the topics in this chapter.
The Detector functions are accessible from the Array Control Panel page.
To display the Array Control Panel page, click either of the Array links on the
home page.
To view Detector log files, click the Log File(s) link on the Array Control Panel
page, or the Logs link on any of the other Array pages. The Advanced Search and
Detailed Log Info page displays. The operations you can perform from this page are
described in Table 5-1.
NOTE: For security purposes, you may be prompted to enter a login name and password before
executing this function. For more information, refer to “Function Security Levels and Login Prompts”
on page 5-2.
Search for a text string in the current Enter a search string in the adjacent field, then click
active log, and display the line in the log the Search button.
file in which it appears.
Display the first or last lines in the Enter the number of lines you want to display in the
current active log. adjacent field, then click either the Beginning or End
button.
View the current and all previous log Click the All Log Files button.
files.
To return to the Advanced Search and Detailed Log Info page from the active log
listing page, click the Detailed button.
NOTE: For security purposes, you may be prompted to enter a login name and password before
executing this function. For more information, refer to “Function Security Levels and Login Prompts”
on page 5-2.
3. Click the Browse button, then navigate to and select the appropriate software
component file from your Acquisition Workstation’s file system.
The file you selected is displayed in the Please select the file to transfer
field
CAUTION The PXCM Web Tool does not check that the appropriate file type has been
selected. If you select the wrong file to upload, you could cause the PXCM to stop
operating.
4. When you are ready to start the software file transfer, click the Upgrade
button.
The file transfer status is displayed on the page.
5. After the transfer is complete, reboot the Detector to have it start up using the
software you just transferred. For information about rebooting the Detector,
refer to “Powering the Detector On and Off” on page 5-9.
NOTE: For security purposes, you may be prompted to enter a login name and password before
executing this function. For more information, refer to “Function Security Levels and Login Prompts”
on page 5-2.
NOTE: The PXCM is also referred to as the “brick.” Therefore, you will see the terms “PXCM” and
“brick” used interchangeable throughout the web tool and the PXCM topics in this chapter.
To display the Brick Control Panel page, click either of the Brick links on the
home page.
To view more detailed PXCM information, click the Detailed button. A sample of
the PXCM detailed information is shown on the following page.
To view PXCM log files, click the Log File(s) link on the Brick Control Panel
page, or the Logs link on any of the other PXCM pages. The Advanced Search and
Detailed Log Info page displays. The operations you can perform from this page are
described in Table 5-1 on page 5-6.
NOTE: For security purposes, you may be prompted to enter a login name and password before
executing this function. For more information, refer to “Function Security Levels and Login Prompts”
on page 5-2.
To return to the Advanced Search and Detailed Log Info page from the active log
listing page, click the Detailed button.
NOTE: For security purposes, you may be prompted to enter a login name and password before
executing this function. For more information, refer to “Function Security Levels and Login Prompts”
on page 5-2.
CAUTION The Brickman page also includes an Execute Command button and an
Advanced button. These functions are intended only as an aid to Direct Radiography
Corp. Support staff in troubleshooting PXCM problems. Use of these functions could
adversely affect PXCM operation.
CAUTION The Brick AEC page also includes a Modify Settings button. This function
is intended only as an aid to Direct Radiography Corp. Support staff in troubleshooting
AEC problems. Use of this function could adversely affect AEC operation.
NOTE: For security purposes, you may be prompted to enter a login name and password before
executing this function. For more information, refer to “Function Security Levels and Login Prompts”
on page 5-2.
To view MCB firmware information, click the Brick Firmware link on the Brick
Control Panel page, or the MCB link on any of the other PXCM pages. The MCB
page displays.
NOTE: For security purposes, you may be prompted to enter a login name and password before
executing this function. For more information, refer to “Function Security Levels and Login Prompts”
on page 5-2.
CAUTION The MCB page includes an Advanced button. This function is intended
only as an aid to Direct Radiography Corp. Support staff in troubleshooting AEC
problems. Use of this function could adversely affect MCB operation.
2. Click the Browse button, then navigate to and select the appropriate MCB
firmware file from your Acquisition Workstation’s file system.
The file you selected is displayed in the Please select the Firmware file to
transfer field
CAUTION The PXCM Web Tool does not check that the appropriate file type has been
selected. If you select the wrong file to upload, you could cause the PXCM to stop
operating.
NOTE: To clear your file selection, click the Reset Form button.
3. When you are ready to start the firmware file transfer, click the Upgrade Now
button.
The file transfer status is displayed on the page.
4. After the transfer is complete, reboot the PXCM to have it start up using the
firmware you just transferred. For information, refer to “Rebooting the PXCM”
on page 5-21.
NOTE: For security purposes, you may be prompted to enter a login name and password before
executing this function. For more information, refer to “Function Security Levels and Login Prompts”
on page 5-2.
If you select the Save reserved files option, specify the files you want to
protect from being overwritten in the Enter Configuration files to be
retained field.
CAUTION The PXCM Web Tool does not check that the appropriate file type has been
selected. If you select the wrong file to upload, you could cause the PXCM to stop
operating.
5. For the Select Boot Mode option, select normal, unless directed to select a
different mode by Direct Radiography Corp. Support.
6. Select the Sync Safe & Normal Modes? option.
NOTE: To clear your entries or selections, click the Reset Form button.
7. When you are ready to start the software file transfer, click the Upgrade
button.
The file transfer status is displayed on the page.
8. After the transfer is complete, reboot the PXCM to have it start up using the
software you just transferred. For information about rebooting the PXCM,
refer to “Powering the Detector On and Off” on page 5-9.
NOTE: For security purposes, you may be prompted to enter a login name and password before
executing this function. For more information, refer to “Function Security Levels and Login Prompts”
on page 5-2.
NOTE: For security purposes, you may be prompted to enter a login name and password before
executing this function. For more information, refer to “Function Security Levels and Login Prompts”
on page 5-2.
To view more detailed Generator information, click the Detailed button. A sample
of the Generator detailed information is shown on the next page.
NOTE: The Generator Information page also includes an Advanced button for displaying even more
detailed information. However, this advanced information is intended only as an aid to Direct
Radiography Corp. Support staff in troubleshooting Generator problems. For this reason, you will
typically be using this function only at the direction of Direct Radiography Corp. Support.
NOTE: For security purposes, you may be prompted to enter a login name and password before
executing this function. For more information, refer to “Function Security Levels and Login Prompts”
on page 5-2.
2. Select the Duration of Change option to specify how long you want the new
scan timeout value to be in effect.
NOTE: To clear your selection and entry, click the Reset button.
3. In the Scan Timeout Value field, enter the timeout value (in seconds).
4. To save your modifications, click the Change Scan Timeouts button.
NOTE: The log files are intended as an aid to Direct Radiography Corp. Support staff in
troubleshooting system problems. For this reason, you will typically be using the Factory Service
functions only at the direction of Direct Radiography Corp. Support.
The Factory Service functions are accessible from the Factory Service Control
Panel page.
To display the Factory Service Control Panel page, click the Factory Service
link on the home page.
The summary information is displayed on the page. The page also includes a listing
of the available logs.
To save the PXCM run-time statistics to an archive (tar) and compressed
file, click the Brick Run-Time Stats link. From the dialog box that displays, you
can name and save the file in the Acquisition Workstation’s file system.
To save a log to a file, right-click on the link, then choose Save Target As from
the pop-up menu. From the dialog box that displays, you can name and save the log
file in the Acquisition Workstation’s file system.
To view a log, click its link. The log contents display on a new page.
NOTE: Calibration must not be performed for at least one hour after the Detector has been powered
on to allow it to warm up.
• When the room temperature has varied more than ±3°C (±5.4°F) from the last
calibration.
• Seam correction algorithm not turned on; Perform DirectRay Detector calibration
wrong algorithm selected. procedure (refer to “When to Calibrate the
DirectRay Detector” on page 6-1).
• Detector gain not calibrated properly.
Line Parallel to the Long Axis Line ends at seam or line partially defective
(gate lines).
Line Runs Through Seam • Read out problem; HV setting time too • Follow applicable problem solving.
short.
Increase in Defective Pixels Defective pixel file requires updating. Contact Direct Radiography Corp. Support
and Lines for information on bad pixel correction.
Appearance of 24 “256 blocks” Gain calibration not turned on. Follow calibration procedure.
of different Densities or
Detector not calibrated properly.
Counts
Garbled, scrambled, or Poor cable connection. Check pins and Detector connection to
truncated (from edges of long PXCM.
axis in) images
Electronic interference with or along cable. Identify and isolate sources of interference;
examine the cable pathway.
Fine Linear Pattern, Parallel to Electromagnetic noise interfering with Shield Motors, power sources, and so on,
Long Axis, in One or More Detector Read Out. located in proximity to the DirectRay
“256 Blocks” (usually more Detector.
pronounced on an edge)
Density Image is too dark or light. Verify LUT curves are properly set.
No Image or Very High HV setting incorrect. Check pac.cfg for correct setting.
Exposure Required to Obtain
HV contact failed. Replace DirectRay Detector.
an Image
High Contrast Objects Appear DirectRay Detector erase not working Check pac.cfg for correct settings.
as Ghost Image correctly.
Check for DirectRay Detector component
failure.
White lines towards the center DirectRay Detector defective. Replace DirectRay Detector.
of panel (Bar Code)
Dark lines / weak lines Lines need to be mapped out. Perform a procedure with X-ray and submit
the file to Direct Radiography Corp. Support
for mapping.
Poor image quality DirectRay Detector not calibrated to optimal Allow proper warm-up time to normal
operating temperature. operational temperature.
Distorted image with 256 DirectRay Detector defective. Replace DirectRay Detector.
blocks
GS20_418.cdr
Replacement Guidelines
The field replaceable units for the DirectRay FFDM-S system include:
• DirectRay Detector
• DirectRay PXCM
• DirectRay Interface card
• Cables
There are no field-reparable units. If one of these components fail, you should
replace the component with a new one and ship the failed component back to Direct
Radiography Corp. For information on how the DirectRay components are secured
and connected, refer to Chapter 3.
You may also have to reinstall DirectRay Detector or Power and X-ray Control
Module (PXCM) software if it becomes corrupted. For software installation
instructions, refer to Chapter 5.
2. Place the Detector in its bag, the temperature logger, the packing insert, and
the phase change material into the inner box, as shown below. Slide the lifting
band over the box.
Detector Pixel: X µm
Y x Y µm active element size
Y x Y detector element matrix
Y% geometric fill factor
DirectRay Detector
TBD
DirectRay PXCM
100-240 V, 60/50 Hz, 1 A, 1 phase
Environment
Transit/Storage
DirectRay Detector
Maximum gradient: X°C (X°F) per hour, assuming the unit is stored in an approved
shipping container
Operating
DirectRay Detector
+24vdc W
24com X
+24vdc T
24com U
GND L
GND H
Data Shield C
Data + D
Data - J
Serial Shield N
Xmit + A
Xmit - E
Recv + M
Recv - K
Bot EL ON F
PREP_DR_DET_ DB25P-2
GEN_RDY_FOR_DR_EXP_ DB25P-3
DR_DET_RDY_FOR_EXP_ DB25P-4
DR_SYST_RDY_ DB25P-5
DR_DET_ERR_ DB25P-6
GEN_POWERED_ON_ DB25P-7
X_RAY_ON_ DB25P-8
(spare) DB25P-9
RxD DB25P-10
TxD DB25P-11
GND_CTRL DB25P-14 to 21
(spare) DB25P-22
GND_232 DB25P-23 to 25