Skanmobile Datasheet Rev 2

Download as pdf or txt
Download as pdf or txt
You are on page 1of 12

Datasheet for Skanmobile High Frequency Diagnostic X-Ray System TDS-000015-0-20 Rev 2 Page 1 of 12

Tube-head & Collimator Specifications

Generator Type : High Frequency, DSP Controlled, Constant Potential (DC)


Control of High Voltage : Closed Loop
High Voltage Range : 40 kV – 100 kV Settable (Step size 1 kV)
Accuracy of High Voltage : < ± 5%
Generator Frequency : Upto 200kHz
High Voltage Ripple : Low frequency ripple < 4%
High frequency ripple < 10%
High Voltage Rise Time : < 3 mS
Control of Tube Current : Closed Loop
Tube-head current range : 10 mA – 100 mA
Maximum Exposure Time : 5 seconds
Minimum Exposure Time : 10 mS
mAs range : 0.1mAs to 250mAs
mAs accuracy : < ± (5%+0.2mAs) for >1mAs and
< ± (10%+0.2mAs) for ≤ 1mAs
Maximum Electrical Output : 4kW
Duty Cycle : 1:30 Duty cycle for exposure ≥ 1 sec and minimum of 30sec off
/ rest time & auto limit based on temperature.
Mode of Operation : Continuous operation with intermittent loading with rest time
of minimum 30sec and 1:30 duty cycle (for exposures ≥1sec)
adaptive and auto limit based on temperature.
Type of cooling : Forced cooling, 8200RPM,10.8CFM
X-ray field (at collimator tip) rectangle
Collimator Rotation : +/- 90 deg
Light intensity : >160 Lux at 1 meter
X-ray field accuracy : < 2% of SID
Additional X-Ray filtration : Minimum 2.0 mm Al equivalent
Total X-Ray filtration : ≥ 3.6mm AL equivalent @ 100kV
Minimum focus to skin distance : 300 mm
Maximum X-Ray field : 480 mm X 480 mm at 1 meter from focal spot.
Leakage radiation @ 1meter : < 50 mR / Hour
Leakage radiation technique : 100 kV, 4 mAs (Max power)
Ingress Protection Classification : IP2X
Cassette Storage Box : Yes, Standard Storage box provided
Tube Head Outer Covers : Aluminium sheet metal
*Anatomical Programming : Yes
*Feature available with External Console

Datasheet for Skanmobile High Frequency Diagnostic X-Ray System TDS-000015-0-20 Rev 2 Page 2 of 12
maximum mA vs KV

110
100
90
80
Voltage in kV

70 KV
60
50
40
30
30 40 50 60 70 80 90 100
current in mA

X-Ray Tube Insert Specifications

Tube Insert Model : OX105-2


Focal Spot (IEC60336) : 1.8 mm
Anode material : Tungsten
Anode angle : 19°
Insert Inherent filtration : 0.5 mm Al equivalent
Anode thermal capacity : 30 kJ
Nominal anode input power at 0.1s(DC) : 4200 W

Datasheet for Skanmobile High Frequency Diagnostic X-Ray System TDS-000015-0-20 Rev 2 Page 3 of 12
X-Ray Tube Insert Rating Chart

X-Ray Tube Insert Thermal Data

Datasheet for Skanmobile High Frequency Diagnostic X-Ray System TDS-000015-0-20 Rev 2 Page 4 of 12
Control Console Features (Tube Head)

LED Indications

1) Yellow colour indicates system is getting prepared to take X-Ray by


PREP / READY LED preheating filament.
2) Green colour indicates system is ready to take X-Ray.

X-RAY ON LED 1) X-Ray on Indication

FAULT LED 1) When there is a fault in the system, FAULT LED will glow in Red colour.

KEY Functions

1) Increments the set kV in steps of 1 and jumps to steps of 5 for


kV INCREMENT KEY continuous pressing.
2) Inactive if kV is preset at 100.

1) Decrements the set kV in steps of 1 and jumps to steps of 5 for


kV DECREMENT
continuous pressing.
KEY
2) Inactive if kV is preset at 40.

1) Increments the set mAs in steps of 0.1 and jumps to steps of 1 and
mAs INCREMENT
then to 10 for continuous pressing.
KEY
2) Inactive if mAs is preset at maximum setting.

1) Decrements the set mAs in steps of 0.1 and jumps to steps of 1 and
mAs DECREMENT
then to 10 for continuous pressing.
KEY
2) Inactive if mAs is preset at 0.1.

1) In fault state, this key resets the generator to standby mode. Some
faults cannot be reset by this key. They are reset only by switching off the
power once and switching on again. These are called as “Power on reset
faults”.
RESET KEY
2) In standby mode, displays Input voltage as Uin. Press KV INCREMENT
KEY for scrolling through other system parameters, DC bus voltage BUS,
heat sink temperature HSt, tank oil temperature Olt, PFC software version
PFC and Exposure count(number exposures delivered by tube)

Datasheet for Skanmobile High Frequency Diagnostic X-Ray System TDS-000015-0-20 Rev 2 Page 5 of 12
1) Switches on and off the collimator lamp. Internal timer switches off the
LAMP KEY
light after 30seconds.

DISPLAY Indications

1) Displays default/set KV value in standby mode.


KV DISPLAY
2) Displays Error message Exx(where xx stands for code) in fault mode

1) Displays default/set mAs value in standby mode.


mAs DISPLAY 2) Displays Error code in fault mode. Refer generator fault list section for
list of faults.

LCD Display

MODE Key SET Key


PREP / READY / X-RAY ON LED
kV Increment Key mAs Increment Key

kV Decrement Key mAs Decrement Key


LAMP ON LED FAULT LED
LAMP ON Key RESET Key

Control Console Features(External)

LED Indications

1) Yellow colour indicates system is getting prepared to take X-Ray by


PREP / READY / X- preheating filament.
RAY ON LED 2) Green colour indicates system is ready to take X-Ray.
3) X-Ray on Indication

FAULT LED 1) When there is a fault in the system, FAULT LED will glow in Red colour.

1) When the lamp is switched ON, LAMP ON LED will glow in yellow
LAMP ON LED
colour

KEY Functions

kV INCREMENT KEY 1) Increments the set kV in steps of 1 and jumps to steps of 5 for
continuous pressing.
2) Inactive if kV is preset at 100.
3) When MODE key is pressed, kV increment key is used to scroll various

Datasheet for Skanmobile High Frequency Diagnostic X-Ray System TDS-000015-0-20 Rev 2 Page 6 of 12
program modes

1) Decrements the set kV in steps of 1 and jumps to steps of 5 for


continuous pressing.
kV DECREMENT KEY 2) Inactive if kV is preset at 40.
3) When MODE key is pressed, kV decrement key is used to scroll various
program modes

1) Increments the set mAs in steps of 0.1 and jumps to steps of 1 and
mAs INCREMENT
then to 10 for continuous pressing.
KEY
2) Inactive if mAs is preset at maximum setting.

1) Decrements the set mAs in steps of 0.1 and jumps to steps of 1 and
mAs DECREMENT
then to 10 for continuous pressing.
KEY
2) Inactive if mAs is preset at 0.1.

1) In fault state, this key resets the generator to standby mode. Some
faults cannot be reset by this key. They are reset only by switching off
RESET KEY
the power once and switching on again. These are called as “Power on
reset faults”.

1) Switches on and off the collimator lamp. Internal timer switches off
LAMP KEY
the light after 30seconds.

MODE KEY 1) Used to enter various programming Modes

DISPLAY Indications

The display interface is a graphic LCD display of 128×64 resolution with


RGB backlight.

The LCD menu has the following main parameters.

1) Start up Menu
2) Self Test Menu
LCD DISPLAY 3) Mode Selection
4) Anatomy selection (Medium / Thick / Thin)
5) Exposure History
6) Fault History
7) Exposure counter
8) Ready for Exposure with anatomy
9) Software version of Control & PFC firmware

Datasheet for Skanmobile High Frequency Diagnostic X-Ray System TDS-000015-0-20 Rev 2 Page 7 of 12
Mains Power Requirements
Line voltage range : 230 Vac ± 10% / 110 Vac ± 10%
Line frequency : 50 Hz ± 5% / 60 Hz ± 5%
Maximum / Momentary Input current : ≤ 16 Amps for 230 Vac / < 35 Amps for 110 Vac
Standby input current : 0.6A max for 230 Vac input / 2.5A max for 110 Vac input
Electrical Connection : Line, Neutral and Earth (Earth is Mandatory)
Input Power Factor > 0.9
Electrical Classification Class I, Type B

Electrical Line Diagram


Tubehead
Tubehead Power Cable
High Voltage
power and control X-Rays

Exposure Handswitch

Mains Input

Console

Console Wire

Datasheet for Skanmobile High Frequency Diagnostic X-Ray System TDS-000015-0-20 Rev 2 Page 8 of 12
Diagrammatic Representation

Datasheet for Skanmobile High Frequency Diagnostic X-Ray System TDS-000015-0-20 Rev 2 Page 9 of 12
Mechanical Dimensions and Weight

Total Weight of System : 85 kg max for 230 V unit and 97 kg max for 110 V
unit
Weight of Tube Head including collimator : 25 kg max
Weight of collimator : 5 kg max
Weight of trolley including SLS and Gas : 60 kg max for 230 V unit and 62 kg max for 110 V
spring unit
Type of painting : Semi Glossy
Vertical travel at minimum position : 384 −10
+30
mm @ 23º −1º +4º

Vertical travel at maximum position : 2050 ± 50 mm @ 139º ± 4°


Horizontal Reach with tube head : 1334 ± 50 mm
Rotation of Tube head on horizontal arm : 360°
Swivel angle of Tube head on horizontal arm : < 290°

Datasheet for Skanmobile High Frequency Diagnostic X-Ray System TDS-000015-0-20 Rev 2 Page 10 of 12
Environmental Conditions

Operating conditions Conditions for transport and storage


Temperature : +10°C to +40°C Temperature : -30°C to +70°C
Humidity : 25% to 75% Humidity : 95 % non condensing
Altitude : 1500m Altitude : 3500m

Regulatory Compliance Statements


This product is designed and manufactured to meet the following standards and regulations:

Market Type Applicable Standards


IEC 60601-1, IEC 60601-1-3, IEC 60601-2-28, IEC 60601-2-54,
General Safety / Performance
ANSI/AAMI ES 60601-1, CAN/CSA C22.2No.60601-1:2008
Europe Safety / Performance MDD 93/42/EEC: 1993, class IIB device, CE Marked
India Radiation Safety AERB/SC/MED-2
Global QMS ISO 13485:2003, ISO 9001:2008

Datasheet for Skanmobile High Frequency Diagnostic X-Ray System TDS-000015-0-20 Rev 2 Page 11 of 12
Contact Details
Corporate Office & Radiology Division
Skanray Technologies Pvt. Ltd.
Plot No. 15-17, Hebbal Industrial Area,
Mysore 570016, Karnataka, India
Phone : +91 821 2415559
Fax : +91 821 2403344
e-mail : office@skanray.com
Web : www.skanray.com
24 Hour Hotline : +91 9901144411 (Customer Complaint and Incident Reporting Only)

Registered Office & Critical Care Division


Skanray Healthcare Pvt Ltd.
Plot # 360, KIADB Indl. Area, Hootagalli,
Mysore 570 018, Karnataka, INDIA
Phone : +91 821 2407000
Fax : +91 821 2407001
e-mail : inquiries@skanrayhealthcare.com
Web : www.skanrayhealthcare.com
Customer Interaction Centre : 1800-200-5858 and
1800-233-5858
(Toll-free: Monday to Saturday: 8:00AM to 8:00PM)

Datasheet for Skanmobile High Frequency Diagnostic X-Ray System TDS-000015-0-20 Rev 2 Page 12 of 12