PPMSTTOmanual B0
PPMSTTOmanual B0
PPMSTTOmanual B0
Measurement System
Thermal Transport Option User’s Manual
Trademarks
All product and company names appearing in this manual are trademarks or registered trademarks of their respective holders.
U.S. Patents
4,791,788 Method for Obtaining Improved Temperature Regulation When Using Liquid Helium Cooling
4,848,093 Apparatus and Method for Regulating Temperature in a Cryogenic Test Chamber
5,311,125 Magnetic Property Characterization System Employing a Single Sensing Coil Arrangement to Measure AC
Susceptibility and DC Moment of a Sample (patent licensed from Lakeshore)
5,647,228 Apparatus and Method for Regulating Temperature in Cryogenic Test Chamber
5,798,641 Torque Magnetometer Utilizing Integrated Piezoresistive Levers
Foreign Patents
U.K. 9713380.5 Apparatus and Method for Regulating Temperature in Cryogenic Test Chamber
C O N T E N T S
Table of Contents
PREFACE ..............................................................................................................................................................vii
Contents and Conventions ...............................................................................................................................vii
P.1 Introduction ......................................................................................................................................................vii
P.2 Scope of the Manual.........................................................................................................................................vii
P.3 Contents of the Manual ....................................................................................................................................vii
P.4 Conventions in the Manual..............................................................................................................................viii
CHAPTER 1
Introduction ......................................................................................................................................................... 1-1
1.1 Introduction .................................................................................................................................................... 1-1
1.2 Overview of the Thermal Transport Option ................................................................................................... 1-1
1.2.1 Purpose of Measuring Thermal Transport Properties .............................................................................. 1-3
1.3 Measurement Modes....................................................................................................................................... 1-3
1.3.1 Continuous Measurement Mode .............................................................................................................. 1-3
1.3.2 Single Measurement Mode ...................................................................................................................... 1-3
1.4 Measured Thermal Properties......................................................................................................................... 1-5
1.4.1 Thermal Conductivity .............................................................................................................................. 1-5
1.4.2 Seebeck Coefficient ................................................................................................................................. 1-5
1.4.3 Electrical Resistivity ................................................................................................................................ 1-6
1.4.4 Figure of Merit......................................................................................................................................... 1-6
1.5 Theory of Operation ....................................................................................................................................... 1-7
1.5.1 Hardware ................................................................................................................................................. 1-7
1.5.2 Thermal and Electrical Circuit................................................................................................................. 1-7
1.5.3 Software Models...................................................................................................................................... 1-8
1.5.4 Estimating Errors in the Data................................................................................................................... 1-9
1.5.5 Correcting for Heat Loss........................................................................................................................ 1-11
1.5.6 Correcting for Seebeck Coefficient of Manganin Leads........................................................................ 1-11
1.6 Start-up Checklist for Secondary Installation ............................................................................................... 1-11
CHAPTER 2
Hardware .............................................................................................................................................................. 2-1
2.1 Introduction .................................................................................................................................................... 2-1
2.2 Thermal Transport Hardware.......................................................................................................................... 2-1
2.2.1 Thermal Transport Sample Puck ............................................................................................................. 2-2
2.2.2 User’s Kit................................................................................................................................................. 2-3
2.2.3 Nickel Calibration Samples ..................................................................................................................... 2-5
2.2.4 WaveROM EPROM ................................................................................................................................ 2-5
2.2.4.1 Replacing the WaveROM Chip ........................................................................................................ 2-6
2.2.5 Thermal Transport Connection Cable...................................................................................................... 2-6
2.2.6 User Bridge Board ................................................................................................................................... 2-7
2.3 ACT Hardware ............................................................................................................................................... 2-7
2.3.1 Model 7100 AC Transport Controller...................................................................................................... 2-7
2.3.2 AC Board................................................................................................................................................. 2-7
CHAPTER 3
Software ................................................................................................................................................................. 3-1
3.1 Introduction .................................................................................................................................................... 3-1
3.2 Overview of Thermal Transport Software ...................................................................................................... 3-1
3.2.1 Measurement Units.................................................................................................................................. 3-2
3.3 Thermal Transport Control Center ................................................................................................................. 3-3
3.3.1 Control Center Tabs................................................................................................................................. 3-3
3.3.2 Measurement Menu ................................................................................................................................. 3-5
3.3.2.1 Options for Advanced Users............................................................................................................. 3-8
3.3.3 System Status........................................................................................................................................... 3-9
3.4 Thermal Transport Data Files ....................................................................................................................... 3-10
3.4.1 Saving Raw Thermal Transport Data .................................................................................................... 3-10
3.4.2 Data File Header .................................................................................................................................... 3-10
3.4.3 Format of Measurement Data Files........................................................................................................ 3-11
3.4.4 Format of Raw Data Files...................................................................................................................... 3-13
3.5 Data Examination ......................................................................................................................................... 3-14
CHAPTER 4
Sample Preparation .......................................................................................................................................... 4-1
4.1 Introduction .................................................................................................................................................... 4-1
4.2 Sample-Mounting Considerations .................................................................................................................. 4-1
4.2.1 Geometry ................................................................................................................................................. 4-2
4.2.2 Lead-Mounting Epoxies .......................................................................................................................... 4-3
4.2.2.1 Silver-Filled H20E Epoxy ................................................................................................................ 4-3
4.2.2.2 Tra-Bond 816H01 Epoxy.................................................................................................................. 4-3
4.3 Two-Probe and Four-Probe Lead Configurations........................................................................................... 4-4
4.3.1 Two-Probe Lead Configuration ............................................................................................................... 4-4
4.3.2 Four-Probe Lead Configuration............................................................................................................... 4-5
4.4 Checking the Sample Contact......................................................................................................................... 4-6
4.5 Using the Puck-Mounting Station................................................................................................................... 4-6
CHAPTER 5
Measurements ..................................................................................................................................................... 5-1
5.1 Introduction .................................................................................................................................................... 5-1
5.2 Taking Thermal Transport Measurements...................................................................................................... 5-1
5.2.1 Connect Leads to the Sample................................................................................................................... 5-2
5.2.2 Measure the Sample Dimensions............................................................................................................. 5-2
5.2.3 Mount the Sample.................................................................................................................................... 5-3
5.2.4 Install the Sample .................................................................................................................................... 5-3
5.2.5 Start the High-Vacuum System ............................................................................................................... 5-3
5.2.6 Open the Data File ................................................................................................................................... 5-4
5.2.7 Define the Measurement.......................................................................................................................... 5-4
5.2.8 Run the Measurement .............................................................................................................................. 5-5
5.2.8.1 Running the Measurement Interactively ........................................................................................... 5-5
5.2.8.2 Running the Measurement in a Sequence ......................................................................................... 5-5
5.2.9 Scanning or Ramping the Temperature While Measuring....................................................................... 5-5
CHAPTER 6
Troubleshooting ................................................................................................................................................. 6-1
6.1 Introduction .................................................................................................................................................... 6-1
6.2 Jumps or Noise in the Data ............................................................................................................................. 6-1
6.2.1 Gaps in the Data Versus Temperature ..................................................................................................... 6-2
6.2.2 Steps in the Data ...................................................................................................................................... 6-2
6.3 Thermal Radiation “Tail” in the Thermal Conductivity Data......................................................................... 6-3
6.4 High-Vacuum Problems ................................................................................................................................. 6-3
CHAPTER 7
Maintenance ........................................................................................................................................................ 7-1
7.1 Introduction .................................................................................................................................................... 7-1
7.2 Using the Puck Adjustment Tool.................................................................................................................... 7-1
7.3 Greasing the Puck Fingers and the Coldfoot Clamp....................................................................................... 7-2
APPENDIX A
Installation............................................................................................................................................................A-1
A.1 Introduction................................................................................................................................................... A-1
A.2 Installing Thermal Transport Hardware........................................................................................................ A-1
A.3 Installing Thermal Transport Software ......................................................................................................... A-3
APPENDIX B
Status Codes and Error Messages ..............................................................................................................B-1
B.1 Introduction....................................................................................................................................................B-1
B.2 System Status Codes ......................................................................................................................................B-1
B.2.1 General PPMS System Status Codes ......................................................................................................B-1
B.2.2 Thermal Transport System Status Codes ................................................................................................B-3
APPENDIX C
Pinout Tables .......................................................................................................................................................C-1
C.1 Introduction....................................................................................................................................................C-1
C.2 Thermal Transport Pinouts.............................................................................................................................C-1
C.2.1 Sample Connections................................................................................................................................C-1
References ..............................................................................................................................................References-1
Figures
Figure 1-1. Thermal and Electrical Connections for an Idealized Sample ............................................................... 1-7
Figure 1-2. Heat Pulse and Temperature and Voltage Response at Hot and Cold Thermometer Shoes in an
Idealized Sample ........................................................................................................................... 1-8
Figure 5-1. Thermal Tab in Thermal Transport Measurement Dialog Box.............................................................. 5-6
Figure 5-2. Resistivity Tab in Thermal Transport Measurement Dialog Box .......................................................... 5-8
Tables
Table 1-1. System Requirements for the Thermal Transport System ....................................................................... 1-2
Table 1-2. Thermal Transport System Parameters ................................................................................................... 1-2
Table 1-3. Thermal Transport System Components ................................................................................................. 1-2
Table 1-4. Styles for Measurements Taken in Single Measurement Mode .............................................................. 1-4
Table 2-1. Recommended Sample Parameters for Nickel Calibration Samples....................................................... 2-5
Table 3-1. PPMS System Data Items That Can Be Saved to the TTO Measurement Data File ............................... 3-9
Table 3-2. Fields in Thermal Transport Measurement Data File............................................................................ 3-11
Table 3-3. Fields in Thermal Transport Raw File................................................................................................... 3-13
Table 4-1. Sample Geometries and Range of Measurable Thermal Conductivities ................................................. 4-2
Table 4-2. Approximate Thermal Conductance of Epoxies...................................................................................... 4-5
Table 5-1. Minimum and Maximum Parameter Limits for Continuous Mode Measurements 5-6
Table 5-2. General Settings for Continuous Mode Measurements ........................................................................... 5-7
Table 5-3. Resistivity Excitation Parameters for Continuous Mode Measurements ................................................ 5-8
Table B-1. Status Associated with Bits of General System Status Field ..................................................................B-1
Table B-2. TTO Status Codes...................................................................................................................................B-3
P.1 Introduction
This preface contains the following information:
• Section P.2 discusses the overall scope • Section P.4 illustrates and describes
of the manual. conventions that appear in the manual.
• Section P.3 briefly summarizes the
contents of the manual.
NOTE
CAUTION! Cautionary notes are preceded with the word CAUTION! This signals conditions
that could result in loss of information or damage to your equipment.
WARNING! Warnings are preceded with the word WARNING! This signals conditions that
could result in bodily harm or loss of life.
Introduction
1.1 Introduction
This chapter contains the following information:
• Section 1.2 presents an overview of • Section 1.5 explains the TTO system’s
the TTO system. theory of operation.
• Section 1.3 describes the TTO system • Section 1.6 contains the start-up checklist
measurement modes. for secondary installation of the TTO.
• Section 1.4 explains how the TTO
system measures thermal properties.
COMPONENT FUNCTION
PPMS Resistivity Option Provides user bridge board that reads two
(Model P400) thermometer shoes.
PPMS AC Transport Measurement System Outputs current to heater and sample while
(Model P600) providing low-noise, phase-sensitive detection.
PPMS High-Vacuum Option Provides thermal isolation for measurements.
(Model P640) Cryopump or Turbo Pump may be used.
PPMS MultiVu Software Provides single user interface for PPMS and
Version 1.1.6 or Later PPMS options.
* In addition to the requirements in Table 1-1, the PPMS Continuous Low-Temperature Control
(CLTC) option (Model P800) is highly recommended. CLTC provides extended low-temperature
control.
PARAMETER VALUE
Pressure High vacuum (~10-4 torr) If you require use of significant magnetic
fields (H > 0.1 T) at temperatures below T ~
Temperature 1.9−390 K 20 K, please inquire with Quantum Design.
Magnetic field 0−14 T when T > 20 K
Taken together with electrical resistivity ρ, the thermal conductivity and Seebeck coefficient also
provide a measure of the so-called thermoelectric figure of merit Z = α2/(κρ), which is a quantity of
practical significance because it quantifies a material’s ability to transport heat by the application of an
electric current (Peltier effect), or conversely, a material’s ability to generate an electric field by
passing a thermal current (Seebeck effect, described above). The figure of merit is usually expressed
as the dimensionless quantity Z × T, where Z × T ~ 1 is a common benchmark for viability of a
material for thermoelectric applications.
subtle curve-fitting calculations are required, so interpretation of the raw data is in principle more
straightforward. Researchers who study thermal transport properties usually employ this steady-state
technique because of its simplicity and robustness. In either style of single measurement⎯stability or
timed⎯described below in Table 1-4, data is first taken in the heater “off” state once the system
settles. After the user-specified heater power is applied, the system waits for the selected equilibrium
condi-tion before making the final measurement in the heater “on” state. You can view the live ∆T vs.
time data in the Waveform tab of the Thermal Transport control center to monitor measurement
progress.
MEASUREMENT
STYLE DEFINITION
Stability System takes first measurement in heater “off” state once temperature stability
at both hot and cold sample thermometers is within a specified window, stated
either as a percentage of T or as an absolute number (in kelvin). After heat is
applied, system waits for the same stability criterion to be met before taking
final measurement. Heater power is turned off after conclusion of this mea-
surement. User-specified timeout forces system to take a measurement at
timeout period even if stability criterion has not been met.
Timed Sends heat pulse of user-specified duration into sample. System takes a mea-
surement of temperatures and thermal voltages before applying heat, and then
takes final measurement at end of heat pulse.
• Thermal conductivity
• Seebeck coefficient
• Electrical resistivity
• Thermoelectric figure of merit
If thermal conductivity, Seebeck coefficient, and electrical resistivity are all measured, then the
thermoelectric figure of merit, which is the algebraic combination of these three measurements, can be
determined.
Separate measurement protocols are provided for thermal conductivity, Seebeck coefficient, and
electrical resistivity because these individual quantities may be more accurately measured by using
excitation currents and temperature differentials optimized for each situation. Limits for the param-
eters defining each measurement may be specified prior to running the measurement. Section 5.3
discusses the measurement parameters.
Each measured thermal transport property may be determined in either of the two measurement modes
(continuous or single) supported by the TTO system; refer to Section 1.3. You select a measurement
mode, and then you select the thermal properties to measure in that mode.
• Four-terminal geometry minimizes the effects of thermal and electrical resistance of the leads
• Continuous measurements while slewing in temperature provide high density of data
• Careful attention to the removal of effects of temperature drift, thermal radiation, and other
systematic errors
• Robust, easy-to-use, fully automated measurements
1.5.1 Hardware
When measuring in continuous mode, the DSP hardware in the Model 7100 AC Transport Controller
generates the heat pulse in the chip resistor heater on the sample, which can be described as an “on”
cycle of constant power followed by an “off” cycle of equal duration. The waveform for this pulse
was programmed specially for the TTO system in the waveROM EPROM on the AC board, so older
AC boards must have the old waveROM swapped for the new waveROM (labeled “THRMXPT
4201”) to run TTO. Section 2.2.4 discusses the waveROM EPROM in more detail.
Figure 1-2 on the following page illustrates the heat pulse as well as the temperature and voltage
response at the hot and cold thermometer shoes in an idealized sample.
screw down onto the leads. For thermal conductivity T hot HEAT
Figure 1-2. Heat Pulse and Temperature and Voltage Response at Hot and Cold
Thermometer Shoes in an Idealized Sample
Top panel: Time trace of hot and cold thermometers during an idealized
heat pulse; note that the PPMS base temperature is slewing.
Middle panel: Corresponding temperature ∆T and voltage ∆V differentials
across the sample, indicating thermal time constants τ1 and
τ2 and the estimate of the asymptotic differential ∆T∞.
Bottom panel: Heater power during square-wave heat pulse.
where ∆T∞ represents the asymptotic temperature drop across the sample if the heater is left on
indefinitely, and τ1 and τ2 are long and short empirical time constants, respectively, for the sample (see
Figure 1-2). The fitting routine performs an exhaustive search over the space of these three parameters,
reducing the space iteratively until the parameter values that yield the minimum in the residual of the
curve fit are identified satisfactorily. Equation 1-1 is appropriate to the data taken during the heating
pulse, while the data taken during the cooling pulse is simultaneously fit essentially by changing the
sign of the model equation: ∆Tmodel,cooling = A − ∆Tmodel,heating, where A is a constant. Due to long
thermal diffusion times (τ1), the thermal history of the sample must be accounted for in the model, and
this is achieved by including the remanent effects of the two previous pulses in modeling the current
pulse.
The fitting routine for Seebeck coefficient data is similar, yet it is less computationally intensive. The
∆V vs. time data is read back from the DSP buffer at the end of the measurement, and after the ∆T vs.
time data is fit to obtain τ1 and τ2, a linear least-squares routine fits the data to the following equation:
∆Vmodel = ∆V∞ × {1− [τ1 × exp(−t/τ1) ± τ2' × exp(−t/τ2')]/(τ1- τ2')} + bt + c (Equation 1-2)
where ∆V∞ is the asymptotic Seebeck voltage drop akin to ∆T∞ in equation 1-1, b and c are parameters
that describe linear drift and offset voltages, respectively, and τ2' = 0...τ1 is swept so that for each value
of τ2' a linear regression in ∆V∞, b, and c is performed. Note that “±” is used between the exponential
terms and signifies that a full search is done for each sign. The physical significance of this is that the
Seebeck coefficient of the material responsible for the short time constant τ2 (that is, the leads) may be
of the opposite sign as that for the material associated with the long time constant (that is, the sample).
This is in contrast to the case of the thermal conductivity, which is always positive. The parameter for
the linear voltage drift b is included here to account for varying thermal voltages in the wiring to the
sample and also the slow microvolt-level drift in the preamp electronics.
A similar measurement technique, previously published by Maldonado1, describes modeling of the
thermal and thermoelectric response of a sample to a low-frequency, square-wave heat pulse.
However, the thermal circuit considered in that work was considerably simpler than that which is
appropriate to TTO, and hence the modeling was done differently.
where N is the number of data points making up the curve. In the measurements of κ and α, N = 64,
while for ρ, N = 128. Since the thermal conductance K = P/∆T, errors in the heater power P (see the
next section) must also be taken into account. The standard deviation in the conductivity is then
calculated:
2 2 2 2
⎛R ⎞ ⎛ 2IRδI ⎞ ⎛ 0.2 × Ploss ⎞ ⎛ 0.1 × T∞ × K shoes ⎞
σ(κ) = κ × ⎜⎜ ∆T ⎟⎟ + ⎜ ⎟ +⎜ ⎟ +⎜ ⎟ (Equation 1-4)
⎝ ∆T∞ ⎠ ⎝ P ⎠ ⎝ P ⎠ ⎝ P ⎠
1
Maldonado, O. Pulse method for simultaneous measurement of electric thermopower and heat conductivity at low
temperatures. Cryogenics, vol. 32, (no. 10), 1992. 908−12.
The first term is the residual of the curve fit mentioned above, the second term propagates the error in
the heater current I (heater resistance is R) due to the digital-analog converter, the third term is the
error in the estimation of the sample radiation term where 20% combined error in the estimation of
sample surface area and emissivity is assumed, and the last term is the error in the thermal conductance
leak from the shoe assemblies Kshoes where a 10% error in this correction is assumed (see the next
section for details on heat losses).
The error in the measurement of the thermal voltage ∆V vs. time has a similar expression as equation
1-3, so the standard deviation in the Seebeck coefficient α = ∆V/∆T is the following:
2 2
⎛R ⎞ ⎛ R ∆T ⎞
σ(α) = α × ⎜⎜ ∆V ⎟⎟ + ⎜⎜ ⎟⎟ . (Equation 1-5)
⎝ ∆V∞ ⎠ ⎝ ∆T∞ ⎠
Resistivity measurements are made both preceding and following each thermal measurement so that
the average of the two ρ and σ(ρ) values is reported in the data file. The residual of the curve fits Rρ is
obtained from the stream of voltage V vs. time data as the following:
Rρ
σ(ρ) = ρ × (Equation 1-7)
VPP
where VPP is the peak-to-peak amplitude of the voltage vs. time signal.
The standard deviation in the figure of merit ZT is obtained by propagating the errors from each of the
measurements:
where the last term is the standard deviation of the sample temperature over the measurement.
is the radiation from the sample, S is the total sample surface area, ε is the infrared emissivity of the
radiating surface (see Section 5.2.2 for more information on estimating the emissivity), Thot/cold are the
average temperatures of the hot and cold thermometers during the measurement, and σT = 5.67 × 10-8
W m-2 K-4 is the Stefan-Boltzmann constant. The factor of ½ in the equation is due to the approxima-
tion that only half of the sample surface area is radiating at the hot temperature, while the other half is
at the cold temperature.
Since radiative heat losses are often very difficult to accurately estimate, you should expect errors in
measurements of thermal conductance above T ~ 300 K that are on the order of ±1 mW/K.
Hardware
2.1 Introduction
This chapter contains the following information:
• Section 2.2 discusses and illustrates the • Section 2.4 discusses the High-Vacuum
TTO hardware. option hardware that is used with TTO.
• Section 2.3 discusses the ACT option • Section 2.5 explains how to calibrate new
hardware that is used with TTO. shoe assemblies.
CAUTION! Use care when threading the radiation shield onto the puck. The copper metal is soft, so excessive
force or misthreading of the piece can easily damage the threads.
• Puck-mounting station
A pivoting, rotating socket is mounted to a heavy base (Figure 2-2) and holds the Thermal
Transport sample puck in a fixed position, giving you better access to the sample leads while you
are connecting the shoes or making other adjustments. Tighten the two thumbscrews on the
mounting station once the desired orientation for the puck is achieved.
• The 14-pin Lemo connector plugs into the gray, color-coded port on the PPMS probe head.
• The connector labeled “J1 (P1) User Bridge” plugs into the “P1−User Bridge” port on the Model
6000.
• The connector labeled “J2 (P1) Sample Current Out” plugs into the “P1−Sample Current Out”
port on the Model 7100.
• The connector labeled “J5 (P5) Sample Voltage In” plugs into the “P5−Sample Voltage In” port
on the Model 7100.
CAUTION! The Model 7100 provides as much as 200 mA of current when being controlled by the TTO system.
Although this is lower than the hardware limit of 2 A, this current can still damage samples in the
current path. Use only currents that can be safely handled by all hardware and samples in the circuit.
2.3.2 AC Board
The AC board is installed in the Model 6000 PPMS Controller and is located behind the “P3−Option”
port, which is the port connecting the Model 6000 to the Model 7100. The waveROM EPROM
(Section 2.2.4) plugs into the AC board. The AC board includes a DSP, digital-to-analog converter
(DAC), current drivers, and other control electronics that are necessary to synthesize excitation signals
and process sample response signals. The DSP provides the excitation waveform and processes the
sample signal.
A. Charcoal Holder Removed for Puck Calibration B. Charcoal Holder Installed for Normal Operation
Figure 2-8. Close-up View of Contact Fingers and Charcoal Holder on Contact Baffle Assembly
The vertical plate that is usually mounted between the sample and the plugs for the shoe assemblies
must be removed before plugging in the calibration fixture. Unscrew the two Phillips-head screws at
NOTE the base of the plate only enough to remove the plate, and then retighten the screws to hold the PC
board. Use caution so that you do not strain the wiring on the bottom side of the PC board: do not lift
or turn the board or pinch any wires when retightening the screws.
If a heater shoe is being calibrated, plug it into the left-hand socket (the socket closest to the marking
“PCB 3084-576” –see Figure 2-9). If calibrating thermometer shoe assemblies, plug them into the
other two sockets, with “Thermometer A” in the middle socket and “Thermometer B” in the right-hand
socket. Note that wiring in the shoe assemblies is symmetric, so plugging in the connectors in either of
the two possible orientations will make the proper electrical connections.
Next, locate a gold-plated copper calibration sample from the TTO User’s Kit. Bend the leads on the
sample so that you can mount it as shown in Figure 2-9. Then mount the copper sample to the cold
foot and mount the copper shoes to the sample. When mounting the heater shoe, make sure the copper
sample does not touch the solder pad of the heater resister, which could cause an electrical short.
Note the serial numbers of each shoe assembly. Then screw the shield onto the TTO puck. Unscrew
the shield cap and make sure none of the copper shoes are touching any part of the puck or the shield.
Replace the shield cap; then insert the puck into the PPMS. Remove the charcoal carrier on the contact
baffle assembly (see Section 2.4.1) to ensure exchange gas is not cryopumped at low temperatures.
Place the baffle set inside the sample chamber. Then purge and seal the sample chamber.
In the calibration wizard window, check the box for the thermometers and/or heaters you wish to
calibrate and enter their serial numbers. The default temperature range should be 1.8 to 400 K. The
heater parameters have been selected to optimize the signal for the 2kΩ heater resistors supplied.
Press the “Start” button to begin calibration, which will last approximately 16 hours.
After calibration is complete, you must make the appropriate changes to TTO initialization file
TTO.INI if you wish to use the newly calibrated shoe assemblies immediately. See Section 3.2.
THERM B
THERM A
HEATER
Figure 2-9. Calibration Fixture Plugged into TTO Puck and Illustrating Sockets for
Each Shoe Assembly
Software
3.1 Introduction
This chapter contains the following information:
• Section 3.2 presents an overview • Section 3.4 discusses the TTO data
of the TTO system software. files.
• Section 3.3 discusses the Thermal • Section 3.5 explains how to examine
Transport control center. data saved to a TTO data file.
The data file for TTO measurements (sample.dat, where sample is the file name you select)
contains the results for each measurement that was completed successfully. The raw data from all
measurements can also be saved in a separate file (sample.raw).
As soon as the TTO software is activated in PPMS MultiVu, the Thermal Transport Log window
(Figure 3-1) opens and indicates which thermometer and heater calibration files will be used. These
settings, as well as several other parameters, can be edited in the Tto.ini initialization file located in
the QdPpms\ThermalTransport\System directory. The thermometer and heater calibration
files are located in the QdPpms\ThermalTransport\Calibration directory. The TTO log
can be viewed by using the View¾TTO Status Log menu command.
Clicking the right mouse button inside the graph window in the Waveform tab opens a menu that
allows selection of thermal conductivity, Seebeck coefficient, or resistivity results, and whether to
plot the fitted curve along with raw data. The title of the graph indicates which data is being
displayed. For thermal conductivity data, “Temperature Delta” refers to the difference between hot
and cold thermometers, while for Seebeck measurements “Seebeck Voltage” refers to the voltage
difference between the hot and cold shoes.
To the right of the graph are listed three parameters that briefly summarize the curve fitting results:
the total amplitude obtained by the curve-fitting routine (equivalent to Delta Temp. or Seebeck
Volt. in the data file), the long time constant in the measurement (that is, τ1 ⎯this is not relevant to
resistivity data), and the residual of the curve fit (that is, the error estimate for the reported total
amplitude). Section 1.5 contains more information on the AC measurements and error estimation in
TTO data.
In the Waveform tab, you can also zoom to examine details of the data by dragging the mouse from
the upper left to the lower right corner of the graph while holding down the left mouse button. To
zoom out, drag the mouse in the opposite direction while holding down the button, or select Zoom All
in the pull-down menu, which is accessed by clicking the right mouse button in the graph.
Other features on the Advanced tab include a heater test in which you select the desired heater current
to apply, as well as an option to swap the software’s assignment of hot and cold thermometers.
Settings and limits for thermal measurements (thermal conductivity and Seebeck coefficient) are
determined in the Thermal tab (Figure 3-8). These settings and limits include limits of heater period
and power and expected limits of Seebeck readback voltage, as well as target amplitude for the heat
pulse (expressed as percentage of sample temperature) and target value of period ratio, which is
defined as measurement period divided by the time constant (tαυ1) of the sample (see Section 3.4.3 for
more information on these quantities). Any changes made in this tab are saved only if the Set button in
the tab is selected. A ToolTip displays hardware limits for the heater power if the cursor is placed over
the panels showing the heater power limits. Note that a period ratio of at least 8 is recommended
because choosing a value lower than this has been shown to produce artifacts in the thermal transport
measurements due to insufficient data for the software model.
Due to the complexity of making an adaptive resistivity measurement, a separate tab is devoted to
resistivity (Figure 3-9) so that you can have maximum freedom in setting limits such as min/max
excitation and min/max frequency, as well as measurement duration. The measurement excitation
frequency is allowed to vary so that a 90% resistive (that is, in-phase) signal is obtained. Autoranging
parameters may also be changed, with the default being Sticky Autorange. Changes made in this tab
are saved only if the Set button in the tab is selected. See the Physical Property Measurement System:
AC Transport Option User’s Manual for more information on resistivity measurement parameters.
Table 3-1. PPMS system data items that can be saved to the TTO measurement data file.
Items in bold and all capital letters are always written to the data file.
ITEM DEFINTION
• The Status bar at the bottom of the Thermal Transport control center succinctly describes the
progress of an on-going measurement and summarizes the results of the last measurement.
Color-coded warning and error messages in the Status bar alert you to possible problems.
Warning messages appear on a yellow background. Error messages appear on a red background.
• The Waveform tab in the Thermal Transport control center (Figure 3-5) displays raw data
waveforms and curve fits for any of the measurements, and the tab indicates the fitting
parameters of signal amplitude, time constant τ1 (where relevant) and the residual for the fit.
• The measurement Progress bar and countdown timer in the bottom right corner of the Thermal
Transport Measurement dialog box indicate the time remaining before the measurement is
complete. Refer to Figure 3-7. In the Progress bar, yellow indicates the heater “on” state while
blue indicates “off.”
• The right side of the Thermal Transport Measurement dialog box summarizes the most recent
measurement and the parameters (period and power) used for the current measurement. Refer to
Figure 3-8. Note that if the cursor is placed over the displayed results, a ToolTip shows the
estimated error for the results.
• The Thermal Transport Log window (Figure 3-1) keeps a record of all messages that appear in
the Thermal Transport control center Status bar. The TTO status log is saved in the file
TtoLog.txt, which is in the directory C:\QdPpms\ThermalTransport\LogFiles.
Because the data file header identifies a particular sample, changing samples without changing data
files can destroy the validity of the data in the file. Therefore, you are encouraged to use the
NOTE automated routines in the Thermal Transport control center. These automated routines prompt you
for new data file(s) and new sample information after you install a different sample.
ITEM DEFINITION
(continues)
ITEM DEFINITION
ITEM DEFINITION
Sample Preparation
4.1 Introduction
This chapter contains the following information:
It is important that the resistance⎯either thermal or electrical⎯at the interface between the leads and
the sample be minimized. This is especially important when a two-probe measurement (Section 4.3.1)
is performed, because any contact resistance is directly reflected in the measured sample thermal and
electrical resistance. In addition, you are advised to minimize the thermal diffusion time in the leads
by keeping them short (2−3 millimeters, if possible), because this expedites the measurement process.
4.2.1 Geometry
The geometry of the sample is constrained due to a variety of considerations, the most obvious of
which is the Thermal Transport sample puck. Mounted vertically on the puck, a sample cannot be
much longer than 20 mm, while the minimum convenient sample length is typically 3 mm.
Another aspect to consider is the thermal diffusion time in the sample, defined as τ ~ C/K, where
C(Joule/K) is the heat capacity and K(Watt/m) is the thermal conductance of the sample. This places
an operational lower limit on K so that the measurement time does not become excessively long (one
measurement is typically designed to be ~ 8 × τ; see Section 1.5). Since K = κ × A/l, where κ(W/m -
K) is the thermal conductivity, A is the cross-sectional area of the sample, and l is the length, this
implies a lower limit on A/l for a given value of κ. Another relation that can be easily derived from the
above equations is τ ~ cp × l2/κ, where cp(J/m3 - K) is the specific heat of the material. This implies
that, for a given material, the time constant simply scales quadratically with the length, with a typical
practical upper bound of lmax ~ 10 mm on samples.
While the thermal diffusion time τ places a lower limit on A/l, the heater power P(W) = κ × ∆T × (A/l),
where ∆T ~ 0.03 × T (typical value) is the temperature drop across the sample, sets the upper limit on
A/l because the heater is limited by the 10-V compliance limit of the Model 7100 current source. For
an R = 2 - kΩ chip heater, Pmax = V2/R = 50 mW.
Keep in mind that the constraints mentioned here are most stringent at high T, where τ is generally
longer and the required ∆T is larger. Table 4-1 gives some examples of sample geometries and the
range of measurable thermal conductivities based on the above considerations and using a 2-kΩ heater.
Brick 8 × (2 × 2) 2−30
Needle 10 × (1 × 1) 10−150
Pill 3 × (5 × 5) 0.1−1.5
Use a generous (but not excessive) amount of epoxy when attaching leads so that the bond is strong
and provides very good thermal contact.
NOTE
The sample-mounting kit contains starter samples of epoxies, including Silver-Filled H20E from
Epoxy Technology, Inc., and nonconductive Tra-Bond 816H01 from Tra-Con, Inc. Product
specifications and Materials Safety Data Sheets (MSDS’s) for both epoxies are included in the epoxy
kit.
Note that the thermal conductance of the epoxies decreases very rapidly below 100 K, so the thermal
contact resistance may be significant at low temperature even if it is not at room temperature. If you
know the cross-sectional area and the approximate thickness of the epoxy used, then you can estimate
the contact resistance due to the thermal resistance of the epoxy using the data in Table 4-2 and the
equation for the thermal resistance:
Use a four-probe lead configuration method when sample resistivity (thermal or electrical) is too low
to allow you to neglect the contribution of lead contact resistance. Thus, the four probes are attached
individually and you avoid the problem of contact resistance at the Thot /V+ and Tcold /V− probes. This
is because very little thermal or electrical current passes into the copper shoes from the sample and
hence they approximate much better the ideal of passive probes of the sample’s temperature and
voltage profile. Note that care must be taken to keep separate the epoxy pads on the sample, or else
this advantage is compromised due to thermal/electrical currents that may shunt through the epoxy
pads at the T/V probes. Figure 4-2 shows a sample mounted in the four-probe configuration.
.
Figure 4-2. Example of Leads Mounted in Four-Probe Configuration
Measurements
5.1 Introduction
This chapter contains the following information:
• Section 5.2 explains how to take • Section 5.4 discusses the TTO
measurements with the TTO system. measurement process.
• Section 5.3 discusses the measurement
mode parameters.
You are encouraged to use the Thermal Transport control center to perform all normal TTO system
operations. The automated routines in the control center help ensure that you complete the necessary
NOTE
procedures when you install new samples and create data files. The examples of immediate-mode
measurements in this chapter illustrate use of the control center.
An artifact from thermal radiation can be seen as T3 “tail” in the thermal conductance that is
visible at temperatures above ~ 200 K. Radiation from the sample and the shoe assemblies is
NOTE
corrected for in the data from the Conductance (W/K) and Conductivity (W/m - K)
columns in the data file. However, no corrections are made in the Raw Conductance
(W/K) data. Section 1.5.5 explains how to correct for heat losses in TTO measurements.
While attaching the shoes, use care not to pull on the thin wires running to the probe shoes, and
avoid scratching the areas near solder pads, especially where the wire is soldered to the back wall
NOTE
of a shoe.
4. Check carefully that neither the shoes nor their wires are touching each other or any part of the
puck, and that the sample is contacting the puck only at the clamp of the coldfoot.
5. Place the radiation shield on the puck, remove the shield cap, and inspect to verify that no wires,
shoes, or that the sample touches the shield.
6. Replace the shield cap before you insert the sample into the PPMS probe.
• Slowly ramp the temperature (rate ~ 0.5 K/min.) and measure continuously. This method uses
the Set Temperature sequence command and is usually the more expedient technique.
• Scan in temperature (or magnetic field) and take single measurements after the system has
stabilized at the new temperature or field. This method uses the Scan Temperature/Scan Field
commands.
Data from any measurement is automatically saved if a data file is open.
The Physical Property Measurement System: PPMS MultiVu Application User’s Manual discusses
sequence files and all standard system sequence commands in detail.
The Clear button in the Thermal tab restores either the values that were in memory after the last time
the Set button in the tab was selected, or the default values if Set has not yet been selected.
Table 5-1. Minimum and Maximum Parameter Limits for Continuous Mode Measurements
(continues)
Table 5-1. Minimum and Maximum Parameter Limits for Continuous Mode Measurements (Continued)
Temp. Rise Target value of temperature Defined as rise in Thot due to the heater, divided by
rise during heating, average temperature during this time. This parameter
expressed in percent of can be decreased in order to minimize uncertainty in
absolute temperature. temperature during a pulse, such as in cases where
Heater power is adaptively physical properties are changing very rapidly with
adjusted to achieve the temperature. Errors in data increase sharply if Temp.
desired Temp. Rise. Rise < 1% is chosen, due to small magnitude of
thermal signal.
Seebeck Determines initial gain Entering an expected maximum value determines
Voltage settings, in µV, for DSP initial gain settings for DSP voltage readback
voltage readback preamp. preamp. Software uses “sticky autorange” algorithm
to rescale preamp if initial guess was far off.
Period Ratio Provides feedback for Defined as ratio of the period to long time-constant
heater period. tau1 of sample. Default period ratio value of 8 has
been found empirically to be near minimum needed
by curve-fitting algorithm in order to converge on
correct result.
PARAMETER FUNCTION
Save Marginal Results Prompts software to write to data file results for which
fitting algorithm was able to converge on a value but
encountered significant errors from one or more sources.
Results are deemed marginal if the error on the curve fit is
between 50% and 200%.
Discard First N Results Prompts software to discard user-specified number of first
results. This setting is useful because first several (~ 3)
data points are usually of poor quality because parameters
were still in the process of being optimized.
PARAMETER FUNCTION
The parameters for resistivity measurements are adjusted adaptively in the continuous measurement
mode, in the same way as the heater power and period are allowed to vary. Refer to the Physical
NOTE
Property Measurement System: AC Transport Option User’s Manual for resistivity settings.
Troubleshooting
6.1 Introduction
This chapter contains the following information:
1. Make sure that the condition Period Ratio (= Period/τ1) > 8 is met at all temperatures.
Considering that the maximum period is 1430 seconds, your sample must be designed so that the
thermal diffusion time τ1 is not too long. If Period Ratio is too small, the curve-fitting software is
not able to adequately fit the data.
2. Make sure that an adequate heat pulse can be applied (default is Temp. Rise = 3%) across the
sample; that is, the thermal conductance of the sample is not above about 20 mW/K. If Temp.
Rise falls below the 1% level, data can become noisy.
3. Verify that the leads are attached to the sample with a generous amount of epoxy that is well
cured and that the lengths of the leads are kept to a minimum.
4. If temperature or field is being slewed while measurements are made, verify that the slew rate is
slow (typically less than 1 K/min.) and uniform over the course of a measurement.
• Check to see if the jump exists only in the magnitudes of thermal conductance and electrical
resistance, but is absent in Seebeck coefficient data. This is evidence that the sample geometry
(A/l) has changed due to internal cracking of the sample (this can often occur under temperature
cycling) or breaking of the epoxy bonds to the leads. Weak epoxy bonds are evidence of poor
epoxy strength or poor thermal matching of the epoxy and the sample.
• Look at the standard deviation in the measured quantities and also at the .raw data to see if the
software is adequately modeling the data. A step in the data can occur when the data fitting is
poor enough that several distinct solutions have comparable curve fit errors. See the Error
column in the .dat file and consult Appendix B to interpret the TTO status code.
Maintenance
7.1 Introduction
This chapter contains the following information:
• Section 7.2 explains how to use the puck • Section 7.3 explains how to grease the
adjustment tool. puck fingers and the coldfoot clamp.
Installation
A.1 Introduction
This appendix contains the following information:
• Section A.2 explains how to install the • Section A.3 explains how to install the
TTO hardware. TTO software.
Note that in order to run, TTO requires that the AC Transport (ACT) option be installed because
TTO and ACT share the hardware configuration file
NOTE
C:\QdPpms\ACTrans\Calibration\Actcal.cfg. For information about installing
the ACT option, refer to the Physical Property Measurement System: AC Transport Option
User’s Manual.
B.1 Introduction
This appendix contains the following information:
0 Status unknown.
1 Normal stability at target temperature.
2 Stable.
5 Within tolerance; waiting for equilibrium.
6 Temperature not in tolerance, not valid.
0−3
7 Filling/emptying reservoir.
10 Standby mode invoked.
13 Temperature control disabled.
14 Request cannot complete, impedance not functioning.
15 General failure in temperature system; contact Quantum Design.
(continues)
Table B-1. Status Associated with Bits of General System Status Field:
Description of General System Status Measure Codes
(Continued)
0 Status unknown.
1 Persistent mode, stable.
2 Persist switch warming.
3 Persist switch cooling.
4 Driven mode, stable at final field.
5 Driven mode, final approach.
6 Charging magnet at specified voltage.
4−7 7 Discharging magnet.
8 Magnet reset.
9 Current error, incorrect current in magnet.
10 Persistent switch heater error.
11 Magnet quench.
12 Magnet charging error.
14 Power supply error.
15 General failure in magnet control system.
0 Status unknown.
1 Purged and sealed.
2 Vented and sealed.
3 Sealed, condition unknown.
4 Performing purge/seal routine.
5 Performing vent/seal sequence.
8−11
6 Pre-pump (turbo pump) / High vacuum evacuate (cryopump).
7 High vacuum.
8 Pumping continuously.
9 Pre-vent / Flooding continuously.
14 High vacuum error.
15 General failure in gas control system.
(continues)
Table B-1. Status Associated with Bits of General System Status Field:
Description of General System Status Measure Codes
(Continued)
0 Status unknown.
1 Sample stopped at target value.
5 Sample moving toward set point.
12−15
8 Sample hit limit switch.
9 Sample hit index switch.
15 General failure.
In Table B-2, “poor curve fit” means that the residual of the fit was greater than 50% or that the Period
Ratio was less than 4, and “no curve fit” means that the software was unable to fit the data to the
model.
The five bits for each error category represent, respectively, errors in thermal conductivity, Seebeck
coefficient, first resistivity, second resistivity, or figure of merit. At a glance, you can interpret the
TTO status code as the following:
Pinout Tables
C.1 Introduction
This appendix contains the following information:
SAMPLE
PUCK /
GREY P1 PORT ON P5 PORT ON P1 USER PORT ON
FUNCTION SYMBOL LEMO MODEL 7100 MODEL 7100 MODEL 6000
5 6 11 6 5
RED
PCB 4084-575
REV A
3 9 13 10
PIN
BETWEEN
CONNECTORS
4 8 14 7
BLUE
Figure C-2. Top View of Pinout of Connector Sockets on Thermal Transport Sample Puck
Maldonado, O. 1992. Pulse method for simultaneous measurement of electric thermopower and heat conductivity
at low temperatures. Cryogenics 32:10, 908−12.
Quantum Design. 2001. Physical Property Measurement System: AC Transport Option User’s Manual.
⎯⎯⎯. 2000. Physical Property Measurement System: Cryopump High-Vacuum Option User’s Manual.
⎯⎯⎯. 2000. Physical Property Measurement System: PPMS MultiVu Application User’s Manual.
⎯⎯⎯. 1999. Physical Property Measurement System: Resistivity Option User’s Manual.
⎯⎯⎯. 2000. Physical Property Measurement System: Turbo Pump High-Vacuum Option User’s Manual.
Weast, R. C. 1988. CRC Handbook of Chemistry and Physics. Boca Raton, FL: CRC Press.