Professional Documents
Culture Documents
Ag Select B2-B3 (Uk)
Ag Select B2-B3 (Uk)
AG-SELECT-B2/B3 series
Table of contents
1. General informations....................................................................................................................3
2. Technical data ..............................................................................................................................3
3. B2 connection diagram ................................................................................................................4
3.1. TTL connection diagram for B2 instrument ........................................................................5
4. B3 connection diagram ................................................................................................................6
4.1. TTL connection diagram for B3 instrument ........................................................................7
5. Display description ......................................................................................................................8
6. B2-B3 Control panel description .................................................................................................9
6.1. Light signs..........................................................................................................................10
6.2. Keyboard............................................................................................................................10
7. Overall dimensions ....................................................................................................................11
8. Defaults parameter .....................................................................................................................12
9. Programming the controller .......................................................................................................13
9.1. Starting controller operation ..............................................................................................13
9.2. Programming procedure.....................................................................................................14
9.2.1. Selecting the measurement channel ...........................................................................14
9.2.2. Main menu .................................................................................................................15
9.2.3. Setting up the controller.............................................................................................16
9.2.4. Calibration menu........................................................................................................23
9.2.5. Setpoint settings .........................................................................................................24
9.2.6. Alarms configuration .................................................................................................27
9.3. Setpoint manual activation.................................................................................................29
9.4. Proximity sensor ................................................................................................................29
9.5. PT100 connection ..............................................................................................................29
9.6. MMC Card configuration...................................................................................................30
9.6.1. Inserting MMC Card..................................................................................................30
9.6.2. Removing MMC Card ...............................................................................................30
9.6.3. Current outputs menu.................................................................................................31
10. Firmware version ...................................................................................................................32
11. RESET procedure ..................................................................................................................33
12. Password restricted menu...........................................................................................................34
13. Priming of the pumps...................................................................................................................35
Pg. 2 di 35
1. General informations
Electronic instruments controlling electrochemical parameters such as pH, Redox or Chlorine are
widely used in swimming pools, waterworks and water treatment plants.
• Capability of performing the most possible measurements with just one type of electronic
board: pH, Redox (mV), Cl (ppm).
• Simple and easy to learn programming procedure providing two types of menu: an EXPERT
menu allowing the user to control indispensable functions, and a SIMPLE menu giving the
user the full capability of setting all functions.
2. Technical data
Parameter Value
Tensione di Alimentazione 24 - 230 Vac 50/60Hz. 20-48 Vdc -10 / +15% Voltage range
Level control – PT100 connection – Relay output 6A (resistite load) 1A (inductive load).
Pg. 3 di 35
3. B2 connection diagram
Fig 1
Pg. 4 di 35
3.1. TTL connection diagram for B2 instrument
Fig 2
Pg. 5 di 35
4. B3 connection diagram
Fig 3
Pg. 6 di 35
4.1. TTL connection diagram for B3 instrument
Fig 4
Pg. 7 di 35
5. Display description
Fig 6
Pg. 8 di 35
6. B2-B3 Control panel description
Fig 7
Fig 8
Pg. 9 di 35
6.1. Light signs
SET 1 Attivo
SET 2 Attivo
6.2. Keyboard
Shift right – Used to select the digit to modify when setting passwords or times.
Pg. 10 di 35
7. Overall dimensions
Fig 9
Fig 10
Pg. 11 di 35
8. Defaults parameter
4 SETPOINT 1 and 2 actuation delay 00:03 m:s 00:03 m:s 00:03 m:s
14 Temperature unit °C °C °C
Pg. 12 di 35
9. Programming the controller
All programming parameters and modes of operation of the instrument can be set by using its
keyboard and dedicated display.
• METER A - Ph
• METER B - RX
• METER C - Chlorine
To modify that selection, the initial configuration must be changed by setting the EXPERT mode
during SETUP; there is nothing to prevent the user from having the system work simultaneously
with three pH measurements or, for example, two pH and one Chlorine measurements.
LCD AG SELECT B2
operation mode
LCD AG SELECT B3
operation mode
Pg. 13 di 35
9.2. Programming procedure
METER button .
When setting the instrument, it is advisable to follow the programming procedure outlined below.
Conversely, if only a single parameter must be changed, it is advisable to go directly to the menu
comprising the concerned function, and perform the necessary changes or settings.
The MAIN menu is comprised of 6 submenus, by using which all controller’s functions can be set:
Pg. 14 di 35
9.2.2. Main menu
Pg. 15 di 35
9.2.3. Setting up the controller
keys
BEWARE!
The instrument can carry out three distinct types of measurement: pH, Redox or Chlorine; that
means that, depending on the requirements of the system and type of probe connected, the user can
decide what type of measurement to control. To outline the programming procedure, the example
shows the pH control setting, however the procedure is the same for controlling the Redox
potential or Chlorine.
Pg. 16 di 35
As soon as the type of programming
is defined, the type of measurement
that the instrument is to control can
be set: pH, Redox or Cl-ppm.
Use below keys to select the type of
measurement
Pg. 17 di 35
As soon as the password is set, the
unit of measure for temperature can
be selected (°C or °F). Press
keys &
to confirm the selection.
Pg. 18 di 35
If the manual function is selected,
the instrument’s display shows the
default temperature (25°C); by
pressing the
keys the temperature value can be set
in the range between 0 and 99.9°C.
to confirm settings.
The controller has a calendar and an internal clock for the management of the exits with timer and
the storage of the data recorded by the controller; in order to set up the clock the following settings
must be carried out.
Pg. 19 di 35
value by using the keys
Once the date and time are set, by pressing the key, the SETUP menu (main) can follow;
thanks to the clock with date setting, two functions have been integrated: Flocculant and Cleaning
which activate in timer and schedule mode, some auxiliary relays (one for each measure channel).
By using the Flocculant function it is possible to activate the flocculant dosing system, up to a
maximum of 4 interventions (timer and scheduled mode) during the day.
By using the Cleaning function it is possible to activate a dosing pump for the cleaning of the
electrode, up to a maximum of 4 interventions (timer and scheduled mode) during the day; the
difference from the flocculant mode, is that the Cleaning mode interrupts the tool’s operations
(disabling the set-point). At the end of such intervention the tool awaits the start up time (see Start-
up delay).
Using buttons
select the function that needs to be
activated
Pg. 20 di 35
By pressing the buttons select the days in which the auxiliary output needs to
be activated; scroll down all the possible options until the day or the combination of days that need
to be activated appears:
• Off.
• Whole week.
• 5 day week.
• 6 day week.
• Saturday and Sunday.
• Days odd numbered.
• Days even numbered.
• Monday.
• Tuesday.
• Wednesday.
• Thursday.
• Friday.
• Saturday.
• Sunday.
Pg. 21 di 35
After having selected the days or the day in which the intervention needs to be carried out, the
activation time needs to be set. Press the buttons to set the hours or the minutes
and use the button to select the field (hours/minutes) on which to carry out the settings.
Carry on with the programming to define the activation times by pressing the key
After having established the day or days of the intervention, the duration of it needs to be set.
Regarding the Cleaning function, it is possible to set the seconds of the auxiliary output activation,
where the message on the display is as follow:
In the case of the Flocculant function it is possible to set hours and minutes of activation of the
auxiliary output; the message appearing on the display is as follow:
Use buttons
to set the hours or the minutes of
activation.
Use the button to select the field (hours/minutes) on which to carry out the setting of the
activation time.
Pg. 22 di 35
9.2.4. Calibration menu
Select the CAL function from the main menu by pressing the key.
Press
Pg. 23 di 35
9.2.5. Setpoint settings
Choosing the SETPOINT to set (1 or 2) by pressing the key is the first operation to carry
out.
or keys.
Press keys to
define the type of actuation and
Pg. 24 di 35
After defining the type of actuation,
setting the hysteresis value is the
next step.
The Controller allows the user to define a delay time relating to SETPOINT actuation. Set the
following value to activate that
function.
Pg. 25 di 35
When choosing the proportional mode, three parameters must be set to allow the TTL output to
operate correctly, as follows:
keys.
Pg. 26 di 35
9.2.6. Alarms configuration
The Controller features various alarms that can be configured by the user; three types of alarm can
be set:
Pg. 27 di 35
The controller can raise an OVER
alarm when the measurement value
does not come back to the Setpoint
value within the established time
interval; to activate that function
proceed as follows:
set the hours or minutes, press the key to select the field to ch’ange (hours or minutes)
When the sign “ALARM” appears during the normal functioning of the tool, the corresponding
relay contacts commute, making it possible to signal the state of emergency from a distance; also, at
the moment that one of these emergency situations arising, the dosage of the corresponding section
METER 1, 2 or 3 (in the case of B3) will block it self and the indications represented in the
examples in the following table appear on the display:
Maximum alarm It is
activated, when the
maximum measure value
is exceeded, above which
the tool should block.
AUX output goes from
normally open to normally
closed
Minimum alarm It is
activated, when the
minimum measure value
is exceeded, above which
the tool should block.
AUX output goes from
normally open to normally
closed
Pg. 28 di 35
Over alarm It is
activated, when the
maximum foreseen time is
exceeded, within which
the measure should reach
the SETPOINT value.
AUX output goes from
normally open to normally
closed
In order to carry out such operation the SETPOINT menu that needs to be activated needs to be
As long as the buttons are kept pressed all the above outputs remain activated.
It is possible to define the functioning modes of the REMOTE entrances through the configuration
of the J43 jumper on the B2 and the J43 and J44 jumpers on the B3 tool: inserting the Clevis in
correspondence to the INDEP writing, the REMOTE entrances become independent from each
other and each activates or deactivates the outputs of the METER section associated to them,
instead, by inserting the clevis in correspondence to the SLAVE writing, only the REMOTE
entrance of the METER 1 remains active, which controls all the outputs on the tool.
In order to activate the tool (METER 1, 2 o 3) the proximity sensor should be Normally Closed.
Regarding the two poles PT100 it is necessary to short circuit the two terminals of the clamps
marked “C” with a clevis (fig.1 e 3) and connect the two wires of the PT100 between one of the
above poles “C” and the third pole which is still free; instead for the four wire one it is necessary to
connect both wires to one of the two pairs of twisted wires at the third pole and the other two wires
of the other twisted pair to the two poles marked “C”.
Pg. 29 di 35
9.6. MMC Card configuration
The Controller foresees a recording of the data collected on Memory Card of MMC type, the
memorized information is as follows: measured value of time unit, state of SETPOINT, state of
alarms.
Fig 11
In case of error the LED starts to flash rapidly, remove the Memory Card, format it by using a
personal computer in FAT16 or FAT32 mode and try to repeat the above inserting procedure.
WARNING
In order to avoid any damages or loss of data collected in the Memory Card it is indispensable to
insert two due AA batteries in the battery compartment indicated in Fig. 1
Pg. 30 di 35
9.6.3. Current outputs menu
The controller is equipped with two current outputs settable by the user; the measurement value
corresponding to 4 or 20 mA can be set for every output.
Press keys to
set the pH value for the first current
output, corresponding to 20 mA.
Pg. 31 di 35
Press key to confirm the choice.
It is possible to connect the current outputs to some proportional pumps current controlled, in this
case it is necessary to coincide the “low” value of the current (4 mA) with the SETPOINT set on the
tool and according to the system’s needs, the “high” value of the current (20 mA) to the value of the
measure from which the proportional intervention should start.
Press simultaneously
keys,
on display appears the Firmware
controller version.
Pg. 32 di 35
11. RESET procedure
Within 15 seconds from switching on the tool press the button and enter the main menu
the following writing appears on the display:
Simultaneously press the buttons , the following writing appears on the display:
Press twice consecutively the button , followed by the button . In order to carry
out a total RESET (return to factory setting).
Press twice consecutively the button , followed by the button . In order to carry
out a partial RESET (return to factory setting keeping the sensor calibration).
Pg. 33 di 35
12. Password restricted menu
while the controller is in measuring mode, the user gets into a special menu where he is
asked if he intends to perform a single point calibration or instead to enter the required password in
order to get access to regular setting menus.
On the other hand, if the user has selected to get access to regular setting menus, the relevant
password must be entered, as described in paragraph 8.3.1.
Pg. 34 di 35
13. Priming of the pumps
To facilitate the priming of the pumps, it is possible to manually activate the output of the
SETPOINT.
Such procedure is accessible even in presence of password by pressing simultaneously two keys (as
following described) while the visualization of the measure is in progress.
Until the keys stay pressed the whole outputs remain activated.
Pg. 35 di 35
COD. DMU 00129ML1-A (10-2007)