Gas & Steam Flow Computer MODEL 415: 415-M-V3.2 June 2017
Gas & Steam Flow Computer MODEL 415: 415-M-V3.2 June 2017
Gas & Steam Flow Computer MODEL 415: 415-M-V3.2 June 2017
FLOW COMPUTER
MODEL 415
1. Introduction 7
2. Specification 10
3. Operation 14
4. Options 39
415-M-V3.2
5. Calibration 50
6. Input Circuits 62
7. Installation 71
7.1 General 71
7.2 Wiring Designations for the Mode 415 73
7.3 Ex 410 Enclosure Dimensions 75
Appendix 76
Index 78
415-M-V3.2
List of Symbols Used in this Manual
415-M-V3.2
Symbol Description SI Units US Units
415-M-V3.2
Symbol Description SI Units US Units
415-M-V3.2
415-M-V3.2
Introduction 7
1. INTRODUCTION
The Model 415 Gas Flow Computer incorporates compensation for gas
and vapours to the following equations:
1. Ideal Gas Law using temperature & pressure correction, but where
compressibility is ignored.
Options include a 4-20mA re-transmission, high and low flow alarms and
an RS232/422/485 output. A unique feature available with the
RS232/422/485 output is the ability to print flowrates and totals at
programmable time intervals.
This enables the instrument to function as a data logger when used in
conjunction with a printer, or other storage device.
Two versions of the instrument are available, the Model 415R with direct
RTD input, and the 415A with a 4-20mA temperature input. Both accept a
4-20mA pressure input.
The Model 415 is designed to supersede the Models 405GS and 405ST.
1 Redlich & Kwong. "An equation of State". Chem Rev, vol 44, p233, 1949.
2 Par Research Project NX-19. "Extension of Range of Supercompressibility
Tables", American Gas Association, 1962.
415-M-V3.2
8 Introduction
415-M-V3.2
Introduction 9
Model 415 R. 1 0 E C
Temperature Options
R for RTD input 0 for no option
A for 4-20mA 1 for 4-20mA output
2 for RS232/422/485
3 for Relay option
4 for 4-20mA and relays
5 for RS232/422/485 and
relays
Mounting
1 for panel mounting
2 for field mounting
3 for explosionproof
415-M-V3.2
10 Specification
2. SPECIFICATION
General
Frequency Input
415-M-V3.2
Specification 11
4-20mA Inputs
Temperature
Measurement Range: -100°C (-148°F) to 200°C (392°F),
Note: a wider temperature range can be
handled via a 4-20mA input.
Accuracy: 0.1°C
RTD Type: Platinum PT100.
Linearity: The non-linearity of the RTD is internally
compensated for.
Pressure Input
4-20mA Output
415-M-V3.2
12 Specification
Relay Output
RS232/422/485 Option
Pulse Output
Ideal Gas
415-M-V3.2
Specification 13
General Gas
Natural Gas
Steam
415-M-V3.2
14 Operation
3. OPERATION
The Model 415 uses a low power CMOS microprocessor to perform
all measurement and control functions.
The scaled pulse out, 4-20mA output option and high/low alarm option are
also based on the DEFAULT display selection. For example, the 4-20mA
output would be a re-transmission of the Mass flowrate, if the DEFAULT
display is set to Mass.
The DISPLAY key can be used to step through the data which can be displayed,
as follows:
Gas Flow
Corrected Volume (Rate & Total)
Mass (Rate & Total)
Temperature & Pressure
415-M-V3.2
Operation 15
Steam
Mass (Rate & Total)
Energy (Rate & Total)
Temperature & Pressure
Specific Weight & Enthalpy
If any value other than the default display values are selected, they will
remain displayed for 5 seconds, after which the display will automatically
revert to the default values.
SI Units
Rate Total
cm/h kcm
scm/h kscm
Nm3 kNm3
kg/h tonne
MJ/h GJ
US Units
Rate Total
cft/h kcft
scft/h kscft
lbs/h klbs
BTU x 1000/h MBTU
The RESET key can be used to reset the totals whenever one of the totals
is displayed. Both totals will be reset at the same time. The RESET
switch can be disabled during calibration to prevent front panel resetting.
415-M-V3.2
16 Operation
This section applies only to gas flow measurement and, if the Model 415 is to
be used for steam measurement, the reader can skip this section and go to
section 3.3.
The Model 415 will accept inputs from a wide range of flowmeters with
the flowrate calculated by the equations defined below. Both mass flow
and volume corrected flow to a base temperature and pressure are
calculated and displayed in either SI (metric) or US units. For an
explanation of the symbols used in the equations see the list at the
beginning of this manual.
In SI Units
3.4834 G P B
rB = ZB T B
kg/m3 .....(2)
In US Units
2.6988 G P B
rB = ZB T B
lbs/ft3 .....(3)
Standard Conditions
415-M-V3.2
Operation 17
QM =r B . Q VB ......(5)
TB TB ZB
Q VB = S V . TF . TF . ZF. .A
Q M= r B .Q VB
PF TB ZB
Q VB = S VB . PB
. TF . ZF
. A ......(7)
Q M= rB Q VB
415-M-V3.2
18 Operation
PF TB ZB
Q VB = S VB . PB
. TF
. ZF
.A ......(8)
Q M = r B . Q VB
Note that the pressure and temperature are still square rooted, even
though the flow signal A is not. This is because the output from the D.
P. transmitter is not truly volumetric, but will be affected by a change in
density of the gas being measured. Therefore, the equations
relating to differential pressure will apply.
415-M-V3.2
Operation 19
Example 1
415-M-V3.2
20 Operation
Example 2
= 260 m 3 /hr
415-M-V3.2
Operation 21
Example 3
101.325 30 + 273.2
S M1 = 1000 x 220 x 15 + 273.2
= 696.1 kg/hr
415-M-V3.2
22 Operation
Example 4
415-M-V3.2
Operation 23
Example 5
415-M-V3.2
24 Operation
The equations 1 to 9 are then used to calculate both corrected volume and
mass flow.
Example 6
T C = -239.9 C
P c = 1296.9 kPa
415-M-V3.2
Operation 25
Z ......(10)
F PV = ZFB
FPV is calculated in the flow computer using the NX-19 equation Z for natural
gas and (FPV)2 is substituted into equations 4 to 8 in place of B
ZF
Also, the temperature and pressure must be within the following ranges:
415-M-V3.2
26 Operation
Both the mass flow (in kg/hr or lbs/hr) and the heat content (enthalpy)
are calculated internally based o n the 1967 IFC Formulation (ASME)
Equations.
The equations use the pressure and temperature readings to determine:
Mass Flow
N. frequency (Hz) 1
SI Units : Q M(SI) = 1000 . KF
. nF ......(11)
N. frequency 1
US Units : Q M(US) = 62.435 . KF
. nF ......(12)
415-M-V3.2
Operation 27
Energy Flow
QM(SI) x h
SI Units : Q E(SI) = 1000
......(13)
Q M(US) x h
US Units : Q E(US) = 0.42992 1000
......(14)
Mass Flow
SV
SI Units : Q M(SI) = 1000 x nF xA ......(15)
......(16)
SV
US Units : Q M(US) = 62.447 x nF xA
Energy Flow
Mass Flow
nB
QM = SM . nF . A ......(17)
Energy Flow
415-M-V3.2
28 Operation
Mass Flow
nB
QM = SM . nF .A ......(18)
Energy Flow
Note that the Specific Weight (density) is still square rooted even
though the flow signal A is not. This is because the output from the
D.P. transmitter is not truly volumetric, but will be affected by a
change in the steam density. Therefore, the gas equations relating to
differential pressure must apply.
415-M-V3.2
Operation 29
Example 7
Units US Units
K-factor = 68.32
Timebase hours
415-M-V3.2
30 Operation
Example 8
From the steam tables at 1300 kPa (abs) and a specific weight
of 216.05 dm3/kg, the temperature can be calculated as 350°C,
and is in a superheated state. Hence, steam measurement via
a 4-20mA input with square law relationship is selected and
the following key parameters are programmed:
Units SI Units
Span 10,000
Base Temperature 350°C
Base Pressure 1300 kPa
Timebase hours
Steam Superheated
415-M-V3.2
Operation 31
3.4 FILTERING
The Flow Computer has a digital filter which will average out these
fluctuations and enable the Rate to be read to four digit accuracy. The
ability to select a suitable filtering level means that highly accurate and
stable readings can be obtained without excessive lag. When the Rate is
retransmitted via the 4-20mA output, the filtering will also average out any
fluctuations on the output.
The diagram below shows a pulsating signal input together with the effect
of filtering.
Rate
Filtered Response
Unfiltered Response
Time
415-M-V3.2
32 Operation
A 90% 99%
1 0 0
2 1 2
4 2 4
6 3 6
10 5 11
15 8 17
20 11 22
25 14 28
35 20 40
45 25 51
60 34 69
75 43 86
90 52 103
99 57 113
415-M-V3.2
Operation 33
The following diagram graphs the change in scaling factor with frequency for
a hypothetical flowmeter. The heavy black line represents the actual scaling
factor of the flowmeter, while the light black line is the approximation used
in the instrument.
S c a l i n g F a c t or
Fact 4
Fact 5 Fact 1
Fact 3
Fact 2
Fact 6
Frequency
415-M-V3.2
34 Operation
During Calibration, the program requires the user to input a frequency and
the Scaling Factor (K-factor of the flowmeter) at up to 10 points on the
curve.
Generally these points will correspond to those shown on the Certificate.
415-M-V3.2
Operation 35
For single analog flow inputs, an input table can be programmed to correct
for any non-linearities between the flow signal and the actual flowrate.
This feature is very useful when using the flow computer with some types of
VA flowmeters or laminar flow tubes which may exhibit slight non-linear
characteristics.
The user programs the table starting with 1.0000 and programs
corresponding values of A and AC. As soon as A is programmed as 0.0000,
no further input to the table is possible and the program will correct over
the number points that were programmed. A maximum of 20 points can
be programmed.
At the base temperature and pressure, the flow equation, with non-
linearity correction is defined as:
Q = Span . A C
415-M-V3.2
36 Operation
Example
Inches wg Inches wg
where A= Inches wg at 20mA
= 52.2007
flowrate flowrate
AC = flowrate at 20mA
= 7075.85
Q = 7075.89 x A C
415-M-V3.2
Operation 37
Note that due to the uneven pulse output spacing on this output, the pulse
output cannot be used to drive rate indicators.
415-M-V3.2
38 Operation
Relay or
Impulse Counter
Model 415
10
5.6 ohms
33V
Zener
12
DC Supply
Model 415
DC Supply 11
Out (8-24V)
External Load
Resistor 10K
Logic Input
5.6 ohms 10
415-M-V3.2
Options 39
DC to DC
T erminal
Convertor
+15V 24
23
Digital to
22
Analog
Convertor
Opto-Isolation 0V 20
-15V 21
415-M-V3.2
40 Options
4.1.2 Calculation
By being independently able to set the output range, the instrument can
effectively be programmed to amplify the input signal. In driving chart
recorders, for example, this enables the output to zoom in on a particular
operating area, instead of having to display the full operating range of the
transducer.
For example, 4mA may be set as 0 kg/min and 20mA as 200 kg/min.
However, the user could set 4mA as representing 100 kg/min and 20mA as
representing 120 litres/min.
For rates or displayed values above and below the maximum and
minimum values the output will remain at its 20mA or 4mA level
respectively.
It should be noted that the output will be updated every 0.5 seconds in
unison with the display and, between updates, the output value is constant.
415-M-V3.2
Options 41
Terminal
+15V 24
Link
23
I(+)
I(-) LOAD
22
500 ohm
maximum
0V 20
-15 V 21
*
* For driving impedence loads over
500 ohms, terminate to terminal 21
415-M-V3.2
42 Options
Terminal
+15V 24
23
I(+)
I(-)
22
LOAD
0V 20
-15V 21
External supply of 23
to 30 VDC
415-M-V3.2
Options 43
With this option installed, the circuits for both the RS232 and RS422/485 are
provided as standard. They can be used to interface to both printers and
computers, and a number of standard protocols are built into the
instrument.
4.2.1 Hardware
The RS232 interface is primarily used with printers or for simple communication
with a computer over a short distance. The RS422 and RS485 interfaces are
used for communication over a long distance or in applications requiring
multipoint communication.
24 (+)
RS422 Out
23 (-)
26 (+)
RS422 In
25 (-)
21 Data In
RS232
22 Data Out
27 CTS
20/2B G r o u n d
415-M-V3.2
44 Options
To covert the RS422 interface to an RS485 interface, the RS422 (-) Data In
Terminal must be connected to the RS422 (-) Data Out Terminal and the
RS422 (+) Data In Terminal must be connected to the RS422 (+) Data Out
Terminal.
These connections will convert the RS422 4 wire interface to the RS485 2
wire interface, as shown in figure 2.
Load
120 ohms T wisted Pair
Load
Host 120 ohms
Computer
415-M-V3.2
Options 45
T wisted Pair
+ Load
120 ohms
Host
Computer -
Gnd
- + Gnd - + - + G n d- - +
In Out In Out
400 Series 400 Series
Instrument Instrument
415-M-V3.2
46 Options
The Model 415 has a real time clock and enables the time and date to be set
and printed on tickets. The date format can be European
(days/months/years) or USA (months/days/hours), while the time is on a 24
hour clock.
Note that the clock will only retain its time for 3 days minimum if there is
no power connected to the instrument. After this period, the clock may
need to be reset.
Battery backup is provided. The battery will typically need replacing every
two years or more frequently if extended power downs are a feature of the
installation. Battery type is a CR2032 coin cell.
All new instruments are supplied with a ‘pullout battery life protection tab’
Please do not remove the tab until you are ready to install and apply
power to the instrument.
Printer
A CTS input is provided, and will prevent the instrument from transmitting
any further characters to a printer if the printer buffer is full. The CTS input
is usually connected to the "Data Buffer Full" output from the printer.
If the printer buffer is large enough to handle the messages output from
the instrument, then this input need not be used and should be left
unconnected.
Computer
For more information on the computer interface please consult the manual
"The RS232/422/485 Communications Option for the 400 Series, Version 2".
415-M-V3.2
48 Options
The totals can be programmed to reset manually via the front panel,
or automatically after each print is initiated, or at 24:00 after the
print is initiated.
Note that if manual reset is selected, it is still possible to prevent front panel
reset by inhibiting this function in General Setup (in Calibration).
415-M-V3.2
Options 49
The Relay output option consists of two Form C relays which can be
preset during calibration to energise when the Default flowrate exceeds
or drops below the preset values.
The "low" relay is energised whenever the rate is below the preset value, and
the "high" relay is energised whenever the rate exceeds the preset value. The
preset values are programmed during calibration as described in section 5.
34 Normally Open
Relay 1
35 Norm ally Clos ed
Low Alarm
36 Common
31 Normally Open
33 Common
415-M-V3.2
50 Calibration
5. CALIBRATION
The Calibration routine enables the Setup Parameters to be programmed, as
well as enabling the input signals to be checked.
2 By pressing the TOTAL key and, while still holding, pressing the
DISPLAY key. Both keys must then be held for approximately 6
seconds. This second method of access can be disabled during
the calibration so that it is only possible to enter the calibration
routine via the link across terminals 1 and 2.
Note that the arrows in the Rate and Total key switches indicate that
these switches can be used to change and increment digits respectively.
415-M-V3.2
Calibration 51
On first entering the Calibration routine, the display will show the Model
number followed by:
SELECT
(GENERAL SETUP)
1. GENERAL SETUP
2. GAS PARAMETERS
3. FLOW PARAMETERS
4. OPTIONS
5. TEST
6. EXIT
The user can toggle between these menus using the" " key. To enter a
menu, the DISPLAY key is then pressed.
In the following flowcharts, the options which can be selected using the
or keys are shown in brackets ( ), and values to be entered are shown as
xxxx's. The flowcharts show the program flow during the Calibration and a
List of Definitions is given immediately after the flowchart. This list covers
those terms which are not explained elsewhere in the text.
To exit Calibration, step through the Setup program until the end, and
press the DISPLAY switch when Exit is displayed, (ensure the calibration link
is removed).
To enter a value of, say, 101.325 when the display shows 0.0000, the
key would be pressed 9 times until 000.000 is displayed. Once the correct
position is reached, the and keys can then be used as normal to
enter data.
415-M-V3.2
52 Calibration
SELECT
CALIBRATION
(GENERAL SETUP)
DISPLAY CONTRAST
ADJUST
FLOW UNITS
(SI UNITS, US UNITS)
TOTAL UNITS
(UNITS X 1000
X 1)
FLOW TIMEBASE
(DAYS, HOURS, MINUTES, SECONDS)
FRONT ACCESS
(ENABLE, DISABLE)
FRONT RESET
(ENABLE, DISABLE)
SELECT
(EXIT)
415-M-V3.2
Calibration 53
SELECT
(GAS PARAMETERS)
GAS EQUATION
(STEAM, IDEAL GAS, GENERAL GAS, NATURAL GAS)
STEAM TYPE
(SATURATED, SUPERHEATED)
SELECT SELECT
(EXIT) (EXIT)
Only required for differential pressure flowmeters. With volumetric flowmeters, there is no need to enter base temperature &
pressure and these values can be left at the default values.
415-M-V3.2
54 Calibration
SELECT
(FLOW PARAMETERS)
LOW SPAN (at 20mA) SPAN (at 20mA) SPAN (at 20mA)
XXXX XXXX XXXX
1
For Steam measurement, the Span defaults to mass.
415-M-V3.2
Calibration 55
FILTER FACTOR
XX
FLOWRATE DECIMAL
XXX.XX
TOTAL DECIMAL
XXX.XX
TEMP OFFSET
415R ONLY
±X.XX
PRESSURE INPUT
(ABSOLUTE, GAUGE)
PRESSURE at 20mA
XXXX
SELECT
(EXIT)
415-M-V3.2
56 Calibration
SELECT
(OPTIONS)
OUTPUT TYPE
(4-20mA)
OUTPUT at 4mA IF
XXXXX 4-20mA
OUTPUT OPTION
IS INSTALLED
OUTPUT at 20mA
XXXXX
DATE FORMAT
(dd/mm/yy), mm/dd/yy)
DATE XX/XX/XX
TIME XX:XX
IF
BAUDRATE
RS232/RS422/RS485
(300 - 9600)
OPTION IS
INSTALLED
WORD LENGTH
(7, 8)
PARITY
(NONE, ODD, EVEN)
SIGNAL TYPE
(RS232, RS422, RS485)
415-M-V3.2
Calibration 57
IDENTIFICATION NO
XX
TYPE OF COMMS
(PRINTER, COMPUTER)
PRINTER ECHO
(STANDARD 80 COLUMN) (FULL DUPLEX, HALF DUPLEX)
(EPSON CTM 290)
(MODEL 624)
PRINT METHOD
(ON RESET, TIME INTERVAL)
IF
RS232/RS422/RS485
OPTION IS
INSTALLED
PRINT INTERVAL
(1, 10, 30 MINUTES,
1, 6, 12, 24 HOURS)
RESET TYPE
(FRONT PANEL RESET)
(RESET AT 24:00 HOURS)
(RESET EACH PRINT)
IF
ALARM OPTION
LOW ALARM SETPOINT IS INSTALLED
XXXXX
SELECT
(EXIT)
415-M-V3.2
58 Calibration
SELECT
(TEST)
PRESSURE INPUT
XX.X mA
DATE: XX/XX/XX
TIME: XX:XX
SELECT
(EXIT)
415-M-V3.2
Calibration 59
5.2 DEFINITIONS
GENERAL SETUP
GAS PARAMETERS
415-M-V3.2
60 Calibration
FLOW PARAMETERS
415-M-V3.2
Calibration 61
415-M-V3.2
62 Input Circuits
6. INPUT CIRCUITS
The Model 415 has a regulated output which can be used to power sensors.
A trimpot on the rear of the instrument allows the voltage to be adjusted
in the range of 8-24 Volts and the output can supply a maximum of 50mA.
The Model 415 has an input conditioning card which will accept signals
from most pulse or frequency producing flowmeters. An 8 position DIL
switch on the rear panel enables the input circuit to be configured for
different signal types:
- Turbine Flowmeters
- Open Collector Outputs
- Reed Switches
- Logic Signals
- Two Wire Proximity Switches.
415-M-V3.2
Input Circuits 63
Switch Settings
The following switch settings are recommended for different input signal
types.
Input
Terminals Switch Settings
Input Signal Type CH1
+ - 1 2 3 4 5 6 7 8
A. Logic Signal, Not Not
9 8 off off off on off off
CMOS,Pulse used used
B. Open
Not Not
Collector or Reed 9 8 off off off on on off
used used
Switch
C. Namur
Not Not
Proximity (set DC 11 9 off on on on off off
used used
out to 8 volts)
D. Switch or
Not Not
Reed Switch with 9 8 off off off on on on
used used
debounce circuit
E. Coil (20m V Not Not
9 8 on off off off off off
P-P minimum) used used
General Specification
Input Impedance
Input type a: 100K
Input types b & d: 10K
Input type c: 1K
Input type e: 100K
415-M-V3.2
64 Input Circuits
+5V
10K
S7 INPUT COMPARATOR
1K 100K
Pulse Input
(9)
S3 S8
+
100R .01 1uF
+5V
(8)
Common 33K
S2
S5 S4
33K 1K2
415-M-V3.2
Input Circuits 65
Model 415
on
9
Common
8 1 8
2. Open-Collector
Model 415
on
9
Common
8
1 8
3. Reed Switch
Model 415
on
9
1 8
8
415-M-V3.2
66 Input Circuits
4. Coils
on Model 415
9
8 1 8
Model 415
on
+8V 11
9 1 8
6. Opto-Sensors
Resistor on
Model 415
11
1 8
9 eg. pre-amplifiers
Common and opto-sensors.
8
415-M-V3.2
Input Circuits 67
Four wire RTD measurement is the most accurate form of measurement and
can be used for measurements with the RTD up to 100 meters from the
instrument. It is recommended to use shielded cable when interfacing to
RTD's.
Two or three wire RTD's can be used in place of 4 wire RTD's, but 4 wires
must be taken to the RTD and the signal and current wires joined as close to
the RTD as possible.
MODEL 415R
R T D
T emperature Sensor
4 I (+)
I (-)
14
5
Signal (+)
Signal (-)
6
2
T wisted Pair
Cable
Shield
415-M-V3.2
68 Input Circuits
When wiring the RTD, care must be taken to ensure the (+) of the Current
on terminal 4 is connected to the same side of the RTD as the (+) of the
signal on terminal 5. The RTD has no polarity and can be connected in
either direction.
4-20mA Inputs
Each 4-20mA has a 250 ohm input connected to the signal ground as
shown below. When the instrument is AC powered, there is sufficient
current from the DC output to power up to three current loops. If more
loops are to be powered,an external DC power source is required.
12
When using only one 4-20mA flow input, the signal must be connected
to terminal 3 as shown above
Shielding: When shielding the input signals, the shield should be connected
to the case earth and not connected at the transmitter end (ie.
Ground at one end only).
415-M-V3.2
Input Circuits 69
12-24 VDC
- +
MODEL 415A
+ - Flow 1 (CH3) 3
DP Hig h Flo w
+ - Fl o w 2 ( C H1 ) 14 250
DP Low Flow ohm
250
+ - Temperature 4
T ohm
(CH0)
250
ohm
+ - Pressure 7
P
(CH2)
250
ohm
2
Signal Ground
415-M-V3.2
70 Input Circuits
2
RESET
5
DISPLAY
6
415-M-V3.2
Installation 71
7. INSTALLATION
7.1 GENERAL
Terminal designations for the Model 415 Flow Computer are given on
the following pages. The cutout hole in the panel should be 5.5"
(139mm) wide x 2.6" (67mm) high. Two side clips are supplied to
secure the instrument into the panel.
A case earthing point is provided via an earth lug on the side of the case.
Note that this earthing point is for the case only and there is complete
electrical isolation between this point and all electronic circuits. For EMC
purposes, or when the instrument is connected to mains, this point must be
connected to a good earth using a multi-stranded, braided wire or strap. All
relay outputs are totally isolated from the case and from the internal
circuitry.
A Supply Output voltage is provided to power sensors. This output will provide
a regulated voltage of 8 to 24 volts and the voltage is adjustable by means of
the potentiometer on the rear panel. Maximum current is 65mA and the
instrument comes with the voltage factory set at 24 Volts. When the
instrument is powered from a DC power source, the maximum output voltage
on the Supply Output is the DC Input Voltage less 3.5 volts.
The instrument will operate from either 14-28 volts DC or from the mains.
The mains voltage is factory set to either 95-135 VAC (110 VAC nominal) or
190-260 VAC (220 VAC nominal). An internal mains transformer provides full
isolation between the mains and the electronic circuits.
It is good practice to use shielded cables for all signal connections to the
Model 415. Care must be taken to separate signal cables from power cables
so as to minimise interference.
Overall shields should be connected to the case earth at the instrument end
only.
This connection should be as short as possible and connected to the earthing
lug on the side of the case.
415-M-V3.2
72 Installation
415-M-V3.2
Installation 73
2B Signal Ground
415-M-V3.2
74 Installation
415-M-V3.2
Installation 75
Door
115.0
Hinge 288
302 221
Bottom View
Earth Point
70.0 70.0
71.5 71.5
Enclosure with 3 x M20 Gland holes Enclosure with 3 x ¾“ NPT Gland holes
415-M-V3.2
76 Appendix
APPENDIX
PROPERTIES OF SELECTED GASES
415-M-V3.2
Appendix 77
415-M-V3.2
78 Index
E K
Index Earthing Point, 71 K-factor Range, 10
External Supply, 69
4-20mA Inputs, 68 Ex 410 Enclosure, 75 L
4-20mA Output, 39
405GS, 7 F Laminar Flow Tubes, 35
405ST, 7 Linear Interpolation, 33
Filtering, 31 Logic Signals, 62
A Floating Point
Numbers, 51 M
Analog Inputs, 67 Flow Correction, 60
ASME, 26 Mass, 20
Flow Equations, 16
Atmospheric Model Number, 9
Flow Input, 62
Pressure, 61 Molecular Weight, 16
Flow Parameters, 51
Multipoint
Flow Signal Type, 60
B Fluctuations, 31
Communication, 44
Front Access, 59
Battery, 46 Front Reset, 59
N
Natural Gas, 25
C G Non-Linearity
Calibration Correction, 33
Gas Parameters, 51
Certificate, 33 NX-19, 7
Gases, 76
Calibration Routine, General Gas, 24
50
General Setup, 51
O
Coils, 66
Ground Terminal, Offset, 61
Compressibility, 24 71
Computer, 47 Open Collector, 62
Options, 51
Cutoff, 60 I Orifice Plates, 17, 27
D Ideal Gas Law, 22 Output Pulse, 37
IFC Formulation, 26 Output Voltage, 71
Data Logging, 48 Impulse Counter, 38
Date Format, 46 Input Circuits, 62 P
Decimal Points, 60 Installation, 71
Default Display, 14, 59 Pitot Tubes, 17, 27
Interference Power Requirements,
Density, 16 Suppression, 72 10
DIL Switch, 62 Isolation, 39, 71 Pressure Input, 11, 61
Dimensions, 10 Printer, 46
Display, 10
Display Contrast, 59
Dual Differential
Pressure, 18, 28
415-M-V3.2
Index 79
Properties, 76 Temperature
Protocol, 46 Measurement Range,
Protocols, 43 11
Proximity Switches, 62 Temperature Offset, 61
Pulse Output, 12 Test, 51
Timebase, 59
R Transducer Supply, 10
Turbine, 17
Redlich-Kwong, 7
Reed Switches, 62 V
Relay Output, 49
Remote Switches, 70 VA Flowmeters, 35
RS232, 43 Vortex, 17, 27
RS232/422/485
Interface Option, 43 W
RS422, 43
Wiring Designations, 73
RTD Input, 67
S
Saturated, 26
Setup Parameters, 50
Shielding, 68
Snubbers, 72
Span, 19
Specific Enthalpy, 26
Specific Gravity, 16
Specific Volume, 26
Specification, 10
Standard Conditions,
16
Steam Measurement,
26
Steam Turbines, 27
Supercompressibility
Factor, 25
Superheated, 26
Symbols, 5
T
Target
Flowmeters, 17, 27
415-M-V3.2