Controlling Quality of Radiological Examination of Electronic Devices
Controlling Quality of Radiological Examination of Electronic Devices
Controlling Quality of Radiological Examination of Electronic Devices
This standard has been approved for use by agencies of the Department of Defense.
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959, United States.
1
E 801 – 01
NOTE 1—Use additional layers of shim material as required. The layers shall be 1 3 2 in. (25.4 3 50.8 mm). The addition shall be identified by the
placement of lead numbers which denote the thickness immediately adjacent to the penetrameter numbers during exposure.
NOTE 2—Tolerance is 60.001 in. (0.025 mm) where dimensions are 0.000 and 60.003 in. (0.076 mm) where dimensions are 0.00.
NOTE 3—Bond materials together with cyanoacrylic or equivalent fast-drying epoxy.
NOTE 4—Particle holes are 0.031 in. (0.787 mm) nominal diameter.
NOTE 5—Tolerance on particle diameter is + 0.0003 in. (0.0076 mm).
NOTE 6—Wire grooves are 0.007 in. (0.178 mm) depth with 90° inclusive angle.
NOTE 7—The number hole is 0.25 in. (0.635 mm) nominal diameter and 0.125 in. (0.318 mm) deep.
6.2 The IQI used shall be the one whose image has a of film, or by pixel grey level value in the case of radioscopic
radiological density or grey level equal to that of the electron examination. If the IQI having the greatest density (number
device 610 %. The density or grey level is measured on an eight) does not produce a resultant image density within
area of the IQI image that contains no wire or particle images. 610 % of the electron device, additional shim stock shall be
It shall be measured using a calibrated densitometer in the case used. The shim shall be of the same type of steel as in the IQI
2
E 801 – 01
8.1.1 When image density or grey level varies more than 6
10 % from that adjacent to the IQI, then an IQI to show
acceptable sensitivity at the most dense part of the specimen
image and a second IQI to show acceptable sensitivity at the
least dense portion of the specimen image, will serve to qualify
the image.
8.1.2 Simultaneous Exposures—In radiography, when a part
or parts of the same design are exposed simultaneously under
the same geometrical conditions in a 360° radiation beam, a
minimum of one IQI shall be required in each quadrant. Where
two IQI’s are required, it is acceptable to use the radiographi-
cally equivalent IQI and one other IQI having the same wire
size.
8.1.3 Two IQI’s shall be used for each lot radiographed as
NOTE 1—Permanently etched or engraved serial number. shown in Fig. 3. The IQI’s shall be located at diagonally
NOTE 2—1⁄8 in. lead markers corresponding to serial number. opposite corners of the film as shown in Fig. 3. In radioscopic
FIG. 2 Arrangement of IQI and Lead Serial Number for Calibration
examinations, similar qualification of opposite sides of the field
Image
of view (FOV) is required once during examination of similar
parts (that is, within 610 % tolerance).
and shall be placed under the IQI. The additional shim shall NOTE 3—Radiographic densities shall be measured with a calibrated
exceed the IQI outside dimensions by at least 0.125 in. (3.18 densitometer. When films are exposed simultaneously in one film holder,
mm) on all sides. The thickness of the shims shall be noted density variations should be determined on the single or combined films,
with lead numerals placed adjacent to the shimmed IQI. depending on the manner in which they are read and interpreted.
6.3 The distortion shall not exceed 10 % unless a higher or Radiographic film density shall be in the range of 1.0 to 3.5.
NOTE 4—For parts of irregular geometry or widely varying thickness, it
lower percentage is agreed upon between the purchaser and the
may be necessary to radiograph the first unit of a given design to
supplier. Measurements for computation of distortion shall be determine proper placement of IQI’s for subsequent radiography. Simi-
made in the grid formed by the crossing wires. larly, in radioscopic examinations, a dry run will assist with IQI position-
6.3.1 Determine percent distortion (Note 2) as follows: ing and scan plan development.
D 5 @~SM 2 SA!/SA# 3 100
9. Part Identification
9.1 In radiography, the image of identification markers for
where: the coordination of the part with the film shall appear on the
D = percent distortion, film, so as not to interfere with the interpretation of the
SM = wire spacing as measured on the radiograph, and radiograph and so that it is evident that the complete part order
SA = actual wire spacing. was obtained. These marker positions shall coincide with the
NOTE 2—Distortion may be positive or negative. part number.
6.4 Particles G through L shall be visible in the image when 9.2 In radioscopy, the image header for each part image
IQI’s 1 through 7 are used and particles G through K shall be shall include a part serial number for traceability and to ensure
visible in the image when IQI number 8 is used. that the complete part order was obtained.
6.4.1 If shims are required, the number and size of particles
that can be resolved shall be agreed on between the purchaser 10. Identification of the Image
and the supplier. 10.1 In both radiography and radioscopy, a system of
positive identification of the image shall be provided in
7. Positioning of IQI’s addition to Section 8. Any or all of the following may appear:
7.1 The IQI’s shall be positioned on the source side of the the name of the examining laboratory, the date, the part
electronic device being imaged. They shall be placed so that number, the view, and whether original or subsequent expo-
the plane of the IQI is normal to the radiation beam if the same sure.
provision is made for the device(s) being examined.
7.1.1 The IQI’s used for each exposure or field of view shall 11. Multiple-Film Techniques
be positioned at the outer edges of the film or field of view but 11.1 Film techniques with two or more films of equal or
not impeding the area of interest being examined. different speeds in the same holder will be permitted, provided
that the appropriate wires and particles in the penetrameter for
8. Number of IQI’s a specific area are demonstrated.
8.1 One IQI shall represent an area within which the
radiological densities or grey fields do not vary more than 12. Non-Film Techniques
610 % (Note 3). At least one IQI per field of view or film shall 12.1 The use of non-film imaging techniques will be per-
be exposed simultaneously with the specimen, except as noted mitted, provided that the applicable penetrameter wires and
in 8.1.1-8.1.3 and (Note 4). particles are demonstrated in the resultant image.
3
E 801 – 01
13. Image Quality 15.2 A calibration image shall be made by exposing the
13.1 The radiological image shall be free of blemishes that individual semiconductor IQI at the proper distance from the
interfere with its interpretation. focal spot. The IQI shall be centered in the beam during this
exposure. The certificate shall show the actual dimensions of
14. Source-Film Distance the grids formed by the cross wires and the parallel variances
14.1 Any source-film distance will be satisfactory provided within the grid. It shall also state the actual size of each
that the required quality level is attained. particle.
ASTM International takes no position respecting the validity of any patent rights asserted in connection with any item mentioned
in this standard. Users of this standard are expressly advised that determination of the validity of any such patent rights, and the risk
of infringement of such rights, are entirely their own responsibility.
This standard is subject to revision at any time by the responsible technical committee and must be reviewed every five years and
if not revised, either reapproved or withdrawn. Your comments are invited either for revision of this standard or for additional standards
and should be addressed to ASTM International Headquarters. Your comments will receive careful consideration at a meeting of the
responsible technical committee, which you may attend. If you feel that your comments have not received a fair hearing you should
make your views known to the ASTM Committee on Standards, at the address shown below.
This standard is copyrighted by ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959,
United States. Individual reprints (single or multiple copies) of this standard may be obtained by contacting ASTM at the above
address or at 610-832-9585 (phone), 610-832-9555 (fax), or service@astm.org (e-mail); or through the ASTM website
(www.astm.org).