Jurnal Cohst Scrib
Jurnal Cohst Scrib
Jurnal Cohst Scrib
RESEARCHARTICLE OPENACCESS
Manuscript received May 18, 2022; revised August 20, 2022; accepted August 12, 2022; date of publication August
25, 2022
Digital Object Identifier (DOI): https://doi.org/10.35882/ijahst.v2i4.96
Copyright © 2022 by the authors. This work is an open-access article and licensed under a Creative Commons
Attribution-ShareAlike 4.0 International License(CC BY-SA 4.0)
How to cite: Deby Anggraeni, Isnanto, and Ida Chairanna Mahirawatie, “The Relationship Between Knowledge Of
Dental And Oral Health And The Number Of Functioning Teeth In Indonesian Elderly People ”, International
Journal of Advanced Health Science and Technology, vol. 2, no. 4, pp. 232–237, Augustus. 2022.
ABSTRACTThe ventilator is a supporter of respiratory needs which is very important for the patient so that there are
several parameters that must be monitored specifically, such as the measurement of pressure and flow rate used in the
ventilator system, the accuracy of which must be in accordance with the accuracy of the respirator. One of the important
parameters to monitor is PIF (Peak Inspiratory Flow) which is the peak inspiratory flow rate given through the ventilator.
PIF that is too high or too low can cause adverse effects on the patient. PIF monitoring can be seen through the PIF value
and waveform on the PIF. Monitoring the waveform of the PIF will be very useful to improve the results of using the
ventilator. The purpose of this research is to get the accuracy and precision of the sensor to display the waveform of the
ventilator output. The procedure carried out is to use the F1031V sensor to detect the flow generated by the ventilator and
then detect the PIF value and PIF waveform. From this research, the measurement of accuracy and precision of the F1031V
sensor to detect PIF and generate a waveform graph is said to be good. This is because the highest error value is ±2.04% at
the 20 LPM setting. While the value of the largest standard deviation at the 30 LPM setting is 1.517 and the greatest
uncertainty value at the 30 LPM setting is 0.061. Then, the largest correction value is found in the setting of 20 LPM and 30
LPM, namely 0.4. PIF monitoring is carried out to maximize patient care and reduce the breakdown time on the ventilator.
232
International Journal of Advanced Health Science and Technology
Multidisciplinary : Rapid Review : Open Access Journal e-ISSN:2808-6422; p-ISSN:2829-3037
pressure (PEEP) etc. . PIF is the peak inspiratory flow rate, made previously. This study aims to obtain the accuracy and
PIF is the maximum flow given through the ventilator. Most precision of the sensor to display the waveform of the
modern ventilators can provide flow rates between 60 and ventilator output so that the data obtained can be used as a
120 L/min. Flow flow should be titrated to meet the patient's support for making calibration tools with sensors that have
inspiratory needs[8][9]. If the PIF is too low for the patient, been studied. pressure sensor to measure and display PIP and
the patient may experience dyspnea, patient-ventilator PIF waveforms.
asynchrony, and increased work of breathing. High PIF This study aims to analyze the waveform generated from
increases PIF pressure and lowers mean line pressure, this the flow sensor output as a flow sensor and the appearance of
can lead to decreased oxygenation. In most patients, a peak the waveform on the Ventilator Tester Design with PIP and
flow rate of 60 L/min is sufficient. Higher flow rates are PIF Waveform Displays as Validation (PIF).
required in patients with higher ventilator requirements.
Higher PIF may also be required in patients with obstructive II. METHODHOLOGY
pulmonary disease to reduce inspiratory time [8], thereby Data retrieval on the Ventilator Tester module was
increasing expiratory time and reducing the risk of auto- carried out using Ventilator 1 which was carried out at
PEEP. Lumajang Hospital. The research design used in making the
In the appearance of the PIF parameter there is a waveform module is Pre-experimental with the After Only Design type.
to determine the timing of respiration on mechanical In this design the researcher only uses one group of subjects
ventilation, in this waveform the timing of the patient's and only sees the results without measuring and knowing the
breathing will be known according to the settings given by initial conditions, but there is already a comparison group.
the nurse, monitoring the waveform from PIF will be very The independent variable in this study is the PIF module
useful to provide further treatment for patients, this has the used. the dependent variable is the flow sensor used, namely
potential to improve the results of the use of the ventilator P1031V. and the controlled variable is the microcontroller
[10], thus making the waveform an important parameter for board used, namely arduino.
This study uses a P1031V flow sensor as a flow input,
monitoring in order to determine the accuracy of the
then the data flow will be processed using an Arduino
waveform detected by the ventilator, in the last few studies
microcontroller, and the flow results will be displayed on the
discussing the ventilator tester module it provides
LCD Display [21].
information about the ventilator output but not the waveform
A. DATA ANALYSIS
which is output by the sensor [11], with these shortcomings
The measurements of each parameter, flow from 20 – 40
the researcher wants to discuss the appearance of the PIF
LPM, and pressure were all repeated 5 times. The average
parameter waveform as a ventilator output validation.
value of the measurement is obtained by using the mean or
In 2014, Fred Duprez et al., conducted a study entitled
the average by applying equation (1).:
"Accuracy of Medical Oxygen Flowmeters: A Multicentric
Field Study", this study aims to determine the accuracy of B. (1)
medical flow to determine the error rate generated from
medical flow meters, and obtain conclusion if the number of
use of flow meters will affect the accuracy of the output [12]. where x denotes the mean (mean) for the n-measurements,
x1 denotes the first measurement, x2 denotes the second
Then, in 2020, Natsumi T. Hamahata et al., conducted a flow
measurement, and xn denotes n measurements. Standard
curve study when patients received assistance from
deviation is a value that indicates the degree (degree) of
mechanical ventilation. In this study, the researchers
variation in a data set or a measure of the standard
concluded that flow waveform shapes need to be considered
deviation of the mean. The standard deviation (SD) formula
when giving mechanical ventilation, giving attention to flow can be shown in equation (2):
waveforms can To improve the outcome of therapy,
researchers hope that further research on flow waveforms
when providing mechanical ventilation to patients [10]. C. (2)
Then, in 2021, Tomy Abuzairi et al., conducted a study on
the manufacture of an accessible ventilator tester so that
ventilator checks can be carried out to maintain the accuracy where xi indicates the number of desired values, x indicates
the average of the measurement results, n indicates the
of the device. Researchers expect further research on certain
number of measurements. Uncertainty (UA) is a doubt that
sensors so that the design of the Low Cost Ventilator Tester
appears in each measurement result. The uncertainty
can be used accurately [11].
formula is shown in equation (3):
Based on the research that has been done previously, the
author will make a study entitled "Design of a Ventalitor
Tester With PIP and PIF Waveform Displays as Validation
(PIF)" which is the development of research that has been
233
International Journal of Advanced Health Science and Technology
Multidisciplinary : Rapid Review : Open Access Journal e-ISSN:2808-6422; p-ISSN:2829-3037
234
International Journal of Advanced Health Science and Technology
Multidisciplinary : Rapid Review : Open Access Journal e-ISSN:2808-6422; p-ISSN:2829-3037
235
International Journal of Advanced Health Science and Technology
Multidisciplinary : Rapid Review : Open Access Journal e-ISSN:2808-6422; p-ISSN:2829-3037
236
International Journal of Advanced Health Science and Technology
Multidisciplinary : Rapid Review : Open Access Journal e-ISSN:2808-6422; p-ISSN:2829-3037
REFERENCES
[1] D. J. Dries and J. J. Marini, “Mechanical Ventilation,” Crit. Care
Nephrol. Third Ed., vol. 196, no. April, pp. 10-21.e2, 2019, doi:
10.1016/B978-0-323-44942-7.00003-0.
[2] D. Giustivi, F. Bottazzini, and M. Belliato, “Respiratory
monitoring at bedside in covid-19 patients,” J. Clin. Med., vol. 10,
no. 21, pp. 1–7, 2021, doi: 10.3390/jcm10214943.
[3] P. A. H. Organization, “Flowchart for the Management of
Suspected COVID-19 Patients at the First Level of Care and in
Remote Areas in the Region of the Americas, July 2020,” no.
July, 2020.
[4] D. A. Suherlim, H. Permana, and L. Lubis, “Correlation between
haemoglobin concentration and oxygen saturation (SpO2) in
elderly professors,” J. thee Med. Sci. (Berkala Ilmu Kedokteran),
vol. 50, no. 2, pp. 157–162, 2018, doi:
10.19106/jmedsci005002201804.
[5] A. Canu, M. Canu, S. Marinkovic, S. Faul, and E. Popovici,
“Respiration rate calculation using low power DSP processor and
SpO 2 sensor,” MeMeA 2011 - 2011 IEEE Int. Symp. Med. Meas.
Appl. Proc., no. July, pp. 517–520, 2011, doi:
10.1109/MeMeA.2011.5966757.
[6] M. V. P. Charles et al., “Ventilator-associated pneumonia,”
Australas. Med. J., vol. 7, no. 8, pp. 334–344, 2014, doi:
10.4066/AMJ.2014.2105.
[7] S. Kharel, A. Bist, and S. K. Mishra, “Ventilator-associated
pneumonia among ICU patients in WHO Southeast Asian region:
A systematic review,” PLoS One, vol. 16, no. 3 March, pp. 1–13,
2021, doi: 10.1371/journal.pone.0247832.
[8] M. S. Volpe, J. M. Naves, G. G. Ribeiro, G. Ruas, and M. R.
Tucci, “Effects of manual hyperinflation, clinical practice versus
expert recommendation, on displacement of mucus simulant: A
laboratory study,” PLoS One, vol. 13, no. 2, pp. 1–11, 2018, doi:
10.1371/journal.pone.0191787.
[9] M. Elizabeth, C. Yoel, M. Ali, M. S. Loebis, H. Arifin, and P.
Sianturi, “Comparison of ventilation parameters and blood gas
analysis in mechanically-ventilated children who received chest
physiotherapy and suctioning vs. suctioning alone,” Paediatr.
Indones., vol. 56, no. 5, p. 285, 2017, doi:
10.14238/pi56.5.2016.285-90.
[10] N. T. Hamahata, R. Sato, and E. G. Daoud, “Go with the flow -
Clinical importance of flow curves during mechanical ventilation:
A narrative review,” Can. J. Respir. Ther., vol. 56, no. July, pp.
11–20, 2020, doi: 10.29390/cjrt-2020-002.
[11] T. Abuzairi, A. Irfan, and Basari, “COVENT-Tester: A low-cost,
open source ventilator tester,” HardwareX, vol. 9, p. e00196,
2021, doi: 10.1016/j.ohx.2021.e00196.
[12] F. Duprez et al., “Accuracy of Medical Oxygen Flowmeters: A
Multicentric Field Study,” Health (Irvine. Calif)., vol. 06, no. 15,
pp. 1978–1983, 2014, doi: 10.4236/health.2014.615232.
237