MPS 002JK - Fluke 77
MPS 002JK - Fluke 77
MPS 002JK - Fluke 77
EMICS Ltd
248 Radford Bv. Page 1 of 2 Pages
Radford Approved Signatory
Nottingham
NG7 5QG
Environmental Conditions
Temperature : 20°C +/- 5°C Mains Voltage : 240V +/- 10V
Relative Humidity : 50% +/- 20% Mains Frequency : 50Hz +/- 1Hz
Comments
Calibration Information
The instrument was calibrated against laboratory standards whose values are traceable to recognised National
Standards. The uncertainty limits quoted refer to the measured values only, with no account being taken of the
instruments ability to maintain its calibration.
The reported expanded uncertainty is based on a standard uncertainty multiplied by a coverage factor k=2, providing
a level of confidence of approximately 95%. The uncertainty evaluation has been carried out in accordance with
UKAS requirements.
DC Voltage
320mV Range 0.0mV 0.0mV
100.0mV 99.7mV
32V Range 10.00V 10.00V
320V Range 100.0V 99.8V
1000V Range 1 000V 1 000V
AC Voltage
750V Range @ 200Hz 700V 703V
320V Range @ 200Hz 100.0V 100.5V
32V Range @ 200Hz 10.00V 10.04V
3.2V Range @ 200Hz 1.000V 1.003V
DC Current
32mA Range 10.00mA 9.97mA
320mA Range 100.0mA 99.7mA
10A Range 10.00A 9.97A
AC Current
10A Range @ 60Hz 10.00A 10.01A
320mA Range @ 60Hz 100.0mA 100.1mA
32mA Range @ 60Hz 10.00mA 10.01mA
Resistance
320Ω Range 0.0Ω 0.1Ω 110mΩ
100.1Ω 100.3Ω
3.2kΩ Range 1.000kΩ 1.001kΩ
32kΩ Range 10.00kΩ 10.00kΩ
320kΩ Range 100.0kΩ 100.0kΩ
3.2MΩ Range 1.000MΩ 1.001MΩ
32MΩ Range 10.00MΩ 10.02MΩ
End of results