CERTIFICATE OF CALIBRATION

Issued By EMICS Ltd Certificate Number 340822


Date of Issue 19 September 2012
0331

EMICS Ltd
248 Radford Bv. Page 1 of 2 Pages
Radford Approved Signatory
Nottingham
NG7 5QG

Customer : Melton Power Services


Leicester, Leicester

Date Received : 28 August 2012

Instrument - System ID : ID375


Description : Digital Multimeter Ref. Number : MPS 029/WS
Manufacturer : Fluke
Model Number : 179
Serial Number : 81490641
Procedure Version : S10.1/N

Environmental Conditions
Temperature : 20°C +/- 5°C Mains Voltage : 240V +/- 10V
Relative Humidity : 50% +/- 20% Mains Frequency : 50Hz +/- 1Hz

Comments

The calibration was performed in accordance with Laboratory procedure CPE 1


No adjustments were made.

The quoted uncertainties include the instruments display resolution.

Calibration Information
The instrument was calibrated against laboratory standards whose values are traceable to recognised National
Standards. The uncertainty limits quoted refer to the measured values only, with no account being taken of the
instruments ability to maintain its calibration.

The reported expanded uncertainty is based on a standard uncertainty multiplied by a coverage factor k=2, providing
a level of confidence of approximately 95%. The uncertainty evaluation has been carried out in accordance with
UKAS requirements.

Calibrated By : Sam Cooke Date of Calibration : 19 September 2012


This certificate is issued in accordance with the laboratory accreditation requirements of the United Kingdom Accreditation Service. It provides
traceability of measurement to recognised national standards, and to the units of measurement realised at the National Physical Laboratory or other
recognised national standards laboratories. This certificate may not be reproduced other than in full, except with the prior written approval of the
issuing laboratory.
CERTIFICATE OF CALIBRATION
Certificate Number
340822
UKAS Accredited Calibration Laboratory No. 0331
AS FOUND RESULTS
Page 2 of 2 Pages

Test Title Applied Value Reading Uncertainties


DC Voltage Linearity
6V DC Range 1.000V 1.000V 580.1uV
-1.000V -1.000V 580.8uV
2.000V 2.000V 765.6uV
-2.000V -2.000V 761.6uV
3.000V 2.999V 948.8uV
-3.000V -3.000V 948.8uV
4.000V 3.999V 1.1mV
-4.000V -4.000V 1.1mV
5.000V 4.999V 1.3mV
-5.000V -5.000V 1.3mV
6.000V 5.998V 1.5mV
-6.000V -6.000V 1.5mV

DC Voltage
600mV Range 0.0mV 0.0mV 40.1uV
100.0mV 100.0mV 58.1uV
60V Range 10.00V 10.00V 5.8mV
600V Range 100.0V 100.0V 40.3mV
1000V Range 1 000V 999V 580.1mV

AC Voltage
1000V Range @ 200Hz 700V 703V 3.4V
600V Range @ 200Hz 100.0V 100.1V 261.1mV
60V Range @ 200Hz 10.00V 10.01V 27.1mV
6V Range @ 200Hz 1.000V 1.000V 2.6mV
600mV Range @ 200Hz 100.0mV 100.0mV 270.4uV

DC Current
60mA Range 10.00mA 9.99mA 26uA
400mA Range 100.0mA 100.0mA 260uA
6A Range 1.000A 0.999A 2.6mA
10A Range 10.00A 10.02A 28.1mA

AC Current
10A Range @ 60Hz 10.00A 10.02A 37.2mA
6A Range @ 60Hz 1.000A 1.000A 3.7mA
400mA Range @ 60Hz 100.0mA 100.1mA 402.6uA
60mA Range @ 60Hz 10.00mA 10.00mA 26.5uA

Resistance
600Ω Range 0.0Ω 0.2Ω 40mΩ
100.1Ω 100.1Ω 220.1mΩ
6kΩ Range 1.000kΩ 1.001kΩ 2Ω
60kΩ Range 10.00kΩ 9.99kΩ 20Ω
600kΩ Range 100.0kΩ 100.0kΩ 200Ω
6MΩ Range 1.000MΩ 1.001MΩ 2kΩ
50MΩ Range 10.00MΩ 10.03MΩ 36kΩ

End of results