1.1 Application This series of temperature and humidity test chamber is suitable for reliability test
of industrial products, with the characteristics of temperature and humidity control
accuracy and wide control range. The performance index complies with the
requirements of GB5170.1.2.5.18-2017 《Test methods for basic parameters of
environmental test equipment for electric and electronic products at low
temperature, high temperature, constant humidity and heat, alternating humidity
and heat test equipment》.
* Note that other uses may cause personal injury and equipment damage!
1.2 Sample Testing and storage of flammable, explosive and volatile substance samples.
restrictions Testing and storage of corrosive substance samples.
Testing or storage of biological samples.
Testing and storage of strong electromagnetic emission source samples.
1.3 Sample In order to make your test data more real and effective, it is necessary to meet the
requirements following principles to use the test chamber reasonably:
The total mass of the load is not more than 80Kg per cubic meter of studio
volume.
The total volume of the load is not more than 1/5 of the volume of the working
room.
In any section perpendicular to the prevailing wind direction, the sum of the load
area should not be greater than 1/3 of the cross-sectional area of the working
chamber there. Do not block the flow of airflow when the load is placed.
KMH-150R
2 alison@komegtech.com
www. komegtech. com
3. Technical parameters
3.1 Test Conditions Equipment cooling method: air-cooled
Measured at room temperature + 25℃ with no load. The temperature and
humidity performance test is measured according to the relevant regulations of
GB / T2424.5 or IEC60068 -3-5 standards. The sensor is placed at the air outlet of
the air processing unit.
3.2 Temperature -20℃ ~ +150℃
range
3.3 Temperature ±0.5℃
fluctuation
3.4 Temperature ≤2.0℃
uniformity
3.5 Temperature ≤±2.0℃
deviation
3.6 Temperature Heating rate:
change rate -20℃~+150℃, average 3℃/min (no load)
Cooling rate:
+150℃~-20℃, average 1℃/min (no load)
3.7 Maximum load N/A
3.8 Humidity range 10 ~ 98%R.H
KMH-150R
3 alison@komegtech.com
www. komegtech. com
3.9 Temperature and
humidity range
4. Chamber structure
4.1 Structural Integral chamber structure
KMH-150R
4 alison@komegtech.com
www. komegtech. com
features The test equipment consists of three parts: the main insulation chamber, the
independent refrigeration unit and the electrical control cabinet.
4.2 Insulation Outer layer is plastic-sprayed anticorrosive electrolytic plate -- Middle insulation
maintenance layer is heat-resistant foam insulation material -- Inner cabinet is SUS304 stainless
structure steel plate
4.3 Exterior Material High quality anticorrosive electrolytic plate, surface electrostatic powder baking
paint, the color is komeg standard color.
4.4 Interior Material SUS304 stainless steel plate, the whole tank is fully welded
4.5 Thermal Rigid polyurethane foam insulation layer, thickness =100mm, flame retardant
insulation layer grade B2
4.6 Door Full size single door, open to the left.
The door frame is equipped with two silicon rubber sealing strips and
anti-condensation electric heating device to prevent external dew and frost.
4.7 Observation There is an observation window on the door. (W210 x H270mm)
window Multi - layer vacuum glass window with electronic defogging film for heat - proof
and anti - condensation.
4.8 Control panel Temperature (humidity) control touch display, start switch, emergency stop
switch, buzzer.
4.9 Unit parts Contains:
Refrigerating unit, pipeline of connection and drainage device, cooling fan,
distribution control cabinet.
Waterway control system for humidification and replenishment.
4.10 Power Switchboard
distribution cabinet Small cooling fan
Main power leakage circuit breaker
4.11 Standard Lead hole: diameter 50mm 2 with silicone plug, 1 on left and right sides (after the
configuration contract is signed).
Sample rack: 2 layers of stainless steel sample rack, bearing 30KG/ layer.
Window light: 1 (dc 24VLED light) (installed on the window, the switch is set on
the external PLC touch screen, and has the function of time-delay automatic
KMH-150R
5 alison@komegtech.com
www. komegtech. com
shutdown).
Move caster (with foot cup) 4pcs
5.4 Centrifugal wind Multi - wing centrifugal fan, aluminum alloy blades
wheel
5.5 Heater Open wire heater, SSR control, equipped with independent overheat protection
temperature switch.
The surface temperature will rise when the heater is energized.
The temperature of the convective air increases after the heating wire, and the heat
is extended to the air in the chamber and the specimen, which plays the role of
heating.
The heating power is precisely controlled by PID algorithm, and the output power
is regulated by solid state relay.
KMH-150R
6 alison@komegtech.com
www. komegtech. com
6. Cooling system
KMH-150R
7 alison@komegtech.com
www. komegtech. com
6.1 Specification This machine adopts mechanical compression refrigeration mode.
Intelligent refrigeration control: according to the temperature and load demand in
the chamber PID control solenoid valve switch output cooling capacity or PID
control heater (refrigeration does not heat, heating does not cool).
Traditional refrigeration control Intelligent energy-saving control
Refrigeration compressor start-stop According to the temperature
control temperature (large temperature requirements in the box, PID control
fluctuations, severely affecting solenoid valve switch output cooling
compressor life, technology that has capacity or PID control heating beeper
been eliminated) constant operation of (cooling without heating, heating
refrigeration compressor + heating without cooling).
output balance control (resulting in In the low-temperature working state,
refrigeration capacity and heating the heater is not involved in the work,
offsetting to achieve temperature and the PID is used to adjust the liquid
dynamic balance, waste a lot of supply of the refrigerant, and the
Electrical energy). three-way flow rate of the cooling
pipe, cold bypass pipe, and hot bypass
pipe is adjusted to achieve automatic
constant temperature of the working
chamber.
6.2 Refrigerant R404A
6.3 Cooling way Air-cooled condenser
6.4 Compressor U.S. Copeland compressor or French Tecumseh compressor
6.5 Wind condenser Air-cooled high-efficiency copper tube fin type forced convection heat exchanger
condenser
KMH-150R
8 alison@komegtech.com
www. komegtech. com
6.6 Evaporator Efficient multi-segment fin evaporator with hydrophilic film
6.7 Auxiliary device High-precision expansion valve, solenoid valve, oil separator, desiccant and other
components are imported from famous international brands.
6.8 Refrigeration The refrigeration system is designed with fully automatic protection measures.
process The liquid refrigerant is injected into the compressor suction line to prevent the
compressor from overheating during the high temperature cooling stage.
Fully implemented nitrogen protective welding and double-stage rotary vane
pump to ensure that the inside of the refrigeration system is clean and reliable.
The bottom of the compressor is designed with a water receiving tray, and the
condensed water is discharged to the outside of the box through a drain pipe at the
rear of the cabinet.
7. Control System
7.1 Specification Regulation and control: forced convection temperature and humidity control,
independent cold and hot side PID adjustment, heat and cooling can be
continuously adjusted, avoiding energy waste caused by hedging of cooling
capacity and heating capacity.
7.2 Controller Komeg 7-inch color touch screen intelligent fuzzy controller
* Operating system: Komeg KM-5166 cold output version
7.3 Screen display Temperature and humidity setting (SV) actual (PV) value is directly displayed.
Can display the execution program number, stage, remaining time and cycle
times, running time display.
Program editing and graphic curve display.
Fixed point or program action status display.
KMH-150R
9 alison@komegtech.com
www. komegtech. com
7-inch LED backlit display screen.
7.4 Resolution Temperature: + 0.01℃;Humidity: + 0.1%;Time: 0.01min.
7.5 Set up range Temperature settable range: -100 ~ 200℃ (note that it is not the equipment
performance range),
Adjust according to the temperature working range of the equipment (upper limit
+5℃, lower limit -5℃),
Humidity condition: 0 ~ 100% RH.
7.6 Program capacity Constant value operation time can be set to unlimited time or timing mode is 99 h
59 m.
Usable program capacity: 50 groups at most.
Usable memory capacity: 30 steps per group (step).
Commands can be executed repeatedly: each command can be cycled up to 999
times.
7.7 Communication Can be connected to a computer to display curves and data collection.
Interface Can be used as a monitoring and remote control system.
Can do multiple machines synchronous control.
RS-485 and LAN
7.8 Control switch 1) Emergency stop switch
2) Start (power) switch
KMH-150R
10 alison@komegtech.com
www. komegtech. com
8.5 Humidification Humidification heating tube over temperature protection (dry burning), abnormal
system water supply, abnormal drainage protection.
8.6 Power supply Phase sequence protection, phase loss protection, equipment leakage protection,
overload and short circuit protection.
8.7 Control circuit Overload and short circuit protection
8.8 Alarm When the above-mentioned protection occurs, the device stops running and emits
an audible and visual alarm. At the same time, the fault location and its cause and
solution are displayed on the instrument.
9. Use conditions
9.1 Operating 1. Ambient temperature: 5℃-35℃
environment 2. Relative humidity: not greater than 85%R.H
3. Atmospheric pressure: 80kPa~106kPa
4. Flat ground without vibration
5. Choose well-ventilated, without direct sunlight or direct radiation from other
heat sources
6. No strong airflow around: when the surrounding air is forced to flow, the
airflow should not blow directly onto the chamber.
7. No strong electromagnetic field around.
8. There is no high concentration of dust and corrosive substances around.
9.2 Power source 1. Connect 380V AC (±10%)
specification Three-phase line + protection ground, grounding resistance ≤ 4Ω.
Power switch uses 4P 40A air switch
2. Maximum power: 5 KW
3.Maximum working current: 13 A
4. Power frequency: 50±0.5Hz
9.3 Ground protection Ground resistance ≤ 4Ω.
9.4 Drain connection Pagoda connector plug φ10 inner diameter silicone hose
Note: Guide the condensed water out of the cabinet.
9.5 Power supply 1. This chamber have standard 5 meters power cable line.
wiring 2. Customers should provide a dedicated fusion-free switch for the equipment.
KMH-150R
11 alison@komegtech.com
www. komegtech. com
Condenser Yongqiang
Evaporator Yongqiang
Phase sequence
Swiss Carlo
relay
Intermediate
Omron or Swiss Carlo
relay
KMH-150R
12 alison@komegtech.com
www. komegtech. com
KMH-150R
13 alison@komegtech.com
www. komegtech. com
KMH-150R
14 alison@komegtech.com