Astm D6241 14
Astm D6241 14
Astm D6241 14
Designation: D6241 − 14
iTeh Standards
priate safety and health practices and determine the applica-
bility of regulatory limitations prior to use. 4. Summary of Test Method
(https://standards.iteh.ai)
4.1 A test specimen is clamped without tension between
2. Referenced Documents
2
circular plates and secured in a tensile or compression testing
2.1 ASTM Standards: machine, or both. A force is exerted against the center of the
Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States
1
D6241 − 14
the testing is begun. If a bias is found, either its cause must be securing the ring clamp assemblage is suggested to be 11 mm
found and corrected, or the purchaser and the supplier must (7/16 in.) and equally spaced at a diameter of 220 mm (8.7 in.).
agree to interpret future test results in the light of the known The surfaces of these plates can consist of grooves with rubber
bias. O-rings or coarse sandpaper bonded onto opposing surfaces. It
5.4 This test method is not applicable to materials that are is suggested that 9.5-mm (3/8-in.) bolts be welded to the
manufactured in sizes that are too small to be placed into the bottom plate so that the top plate can be placed over the bolts
test apparatus in accordance with the procedures in this test and nuts easily tightened. A guide block may be used to help
method. Furthermore, it is not appropriate to separate plies of seat the material being clamped. Other clamps that eliminate
a geosynthetic or geocomposite for use in this test method. slippage are acceptable. See Fig. 2 and Fig. 3.
6. Apparatus 7. Sampling
6.1 Testing Machine, must be constant-rate-of extension 7.1 Lot Sample—In the absence of other guidelines, divide
(CRE) type, with autographic recorder conforming to the the product into lots and take lot samples as specified in
requirement of Specification D76/D76M. Practice D4354.
6.2 Plunger, with a flat diameter of 50 6 1 mm with a radial 7.2 Laboratory Sample—Consider the units in the lot
edge of 2.5 6 0.5 mm. See Fig. 1. sample as the units in the laboratory sample. For the laboratory
6.3 Clamping Apparatus, consisting of concentric plates sample, take a full width sample of sufficient length along the
with an internal diameter of 150 mm (5.9 in.), capable of selvage or edge of the roll so that the requirements of 7.3
clamping the test specimen without slippage (limit slippage of through 7.5.2 can be met. Exclude the inner and outer wraps of
test specimen to 5 mm). The external diameter is suggested to the roll or any material containing folds, crushed areas, or other
be 250 mm (9.8 in.). The diameter of the holes used for distortions not representative of the sample lot.
iTeh Standards
(https://standards.iteh.ai)
Document Preview
ASTM D6241-14
https://standards.iteh.ai/catalog/standards/sist/fcf51602-05b4-4e5e-9374-13672ded2c94/astm-d6241-14
2
D6241 − 14
iTeh Standards
(https://standards.iteh.ai)
Document Preview
ASTM D6241-14
https://standards.iteh.ai/catalog/standards/sist/fcf51602-05b4-4e5e-9374-13672ded2c94/astm-d6241-14
7.3 Remove test specimens from the laboratory sample in a 7.4 Test Specimens—From each unit in the laboratory
randomly distributed pattern across the width with no specimen sample, cut the specimens so that the edge of specimen will
taken nearer the selvage of fabric edge than 1/20 of the fabric extend beyond the edge of the clamp by 10 mm (0.39 in.) in all
width or 150 mm (6 in.), whichever is the smaller, unless directions.
otherwise specified.