Specifications: Features Applications
Specifications: Features Applications
Specifications: Features Applications
ANALYTICAL SYSTEMS
• CE, CSA, RoHS and GOST-R compliance 1076 JOHNSON STREET, THETFORD MINES, QUEBEC CANADA G6G 5W6
(mm) • Internet/Web enable PHONE: (418) 334-0990 • FAX: (418) 334-0660 E-MAIL: info@cai-ca.com • WEB SITE: www.cai-ca.com
Copyright 2006 Contrôle Analytique Copyright 2006 Contrôle Analytique
ns
k2001
Embedded software:
Industrial I/O:
The K2001ns has a very simple and intuitive user interface that displays real-time measurements,
system status and other important data. The K2001ns provides a 4-20 mA contact that outputs the impurity measurement in realtime,
providing a way to transmit process values to a remote system. Moreover, dry contact digital
Wide measurement range and increased accuracy: outputs are available to provide external devices with the alarm and range state.
The K2001ns has many interesting features such as its 3 configurable ranges. For example, range 3 can measure
from 0 to 100 ppm, range 2, from, 0 to 10 ppm and range 1, from 0 to 1 ppm. The analyzer’s hardware gain auto-
matically changes to fit the range in use, increasing the resolution and therefore, the precision of measurement.
These ranges are configurable to your needs. For example, range 2 could be changed to 0 to 5 ppm to increase
the precision for this range of values. Moreover, the K2001ns offers an automatic mode for the range selection. Auto-calibration option (patented):
If the ppm value drops below the maximum value of a range, the K2001ns automatically adjust its range to ensure The K2001ns can be purchased with an automatic calibration option. With this feature on,
an accurate measurement. Furthermore, the automatic range can be overrun by the “Lock range” feature that the K2001ns toggle between 3 inlets (span gas, zero gas and sample gas) in order to automatically
forces the K2001ns to stay on a desired operating range, avoiding improper use of the analyzer. perform a calibration. The time between each calibration and its duration are user configurable.
The K2001 new series has also been upgraded with a chemical-interference-free patent pending design which This option avoids gas contamination, purge time and human errors due to manipulation.
avoids erroneous measurement due to undesired impurities.
Self-diagnosis analyzer:
The embedded software includes different
features that help in keeping track of
system abnormalities. Errors such as
under-scale, over-scale, low flow and
plasma off are managed automatically
by the software and reported in a
“Historic alarm” menu with the time
at which they occurred. All of these
errors trigger a System status relay
providing a way to inform the operator
of a malfunctioning. Moreover, the
K2001ns features 2 configurable alarms
in option. Set Alarm 1 to 11 ppm. If the
K2001ns’s measure is higher than
11ppm, it triggers the corresponding
dry contact relay.
Printed in Canada
frequency, intensity and electrodes which gives it the stability and the selectivity you need.
It also has a dual beam configuration and an interference free system (patent pending).
It is the only system on the market having no interference.
Copyright 2006 Contrôle Analytique Copyright 2006 Contrôle Analytique
SPECIFICATIONS
DETECTOR TYPE: Electromagnetically induced plasma cell (dual beam)
DETECTOR MATERIAL: Quartz, single element, vacuum tight to 10 PSI (69kPA)/New Patent Pending Detector
RANGES: 0-1 ppm, resolution of 10 ppb, user configurable up to 10000 ppm
0-10 ppm, resolution of 0.1 ppm, user configurable up to 10000 ppm
0-100 ppm, resolution of 1 ppm, user configurable up to 10000 ppm
ACCURACY: Better than ± 1% full scale
SAMPLE FLOW REQUIREMENTS: 20 to 150 sccm
GAS CONNECTIONS: Sample: 1/8” Swagelok or 1/8” VCR tube fitting
VENT: 1/8’’ swagelock tube fitting
RECOMMENDED CALIBRATION GAS: Zero: GP-200 purified gas (Getter)
FLOW ACCURACY: 0 to 200 sccm ± 1% full scale
RECOMMENDED MAXIMUM OPERATING PRESSURE: 20 PSIG (138 kPAG)
RECOMMENDED MINIMUM OPERATING PRESSURE: 4 PSIG (28 kPAG)
OPERATING TEMPERATURE: 10 oC to 35 oC
SUPPLY: 115 VAC, 50 – 60 Hz or 220 VAC, 50 – 60 Hz
POWER CONSUMPTION: 35 watts (maximum)
RESPONSE TIME: 20 seconds for 90% of a step change at 75 sccm (typically 10 seconds)
DRIFT: ± 1% over 24 hours (typically ± .5% over long term) ns
DIMENSIONS:
WEIGHT:
5.25’’ high, 18’’ deep, 19’’ wide
15 kg (33lbs)
k2001
CONNECTORS: • RJ-45 for Ethernet 10Base-T K2001 NEW SERIES TRACE NITROGEN IN ARGON
• DB-9 for RS-232 communication
• Curtis GTFX-10 screw terminal block for I/Os OR HELIUM ON LINE GAS ANALYZER
• Corcom 5EFM4S power entry module
STANDARD FEATURES: • Microprocessor controlled with 4x40 LCD display and a 4x4 keypad Contrôle Analytique, a leading manufacturer of trace gas The K2001 new series is now up to date with a new micro-
• Self diagnostic system software analyzers since 1995, is now offering the K2001 new series processor platform that allows a remote control of the analy-
• Manual or auto-ranging (user selectable) with its patent pending plasma emission detector. The K2001 zer through a local network or even through Internet. This
• Internet Web Enable new series offers the legendary reliability of its predecessor connectivity feature allows you to control it from anywhere
• Real Time Clock plus a state of the art connectivity and an upgraded chemical you want and to have Contrôle Analytique technicians assist
• Alarm Historic
• CE conform and CSA on demand interference free design. you in the analyser’s start-up or in any phase of diagnostic,
• Digital ouputs for remote monitoring: without even being on site.
(all dry relay contacts, SPST 1 amp 250 Vac resistive load) The K2001, designed for a continuous measurement of nitro-
- System status (1 output) gen in argon, helium or both, is known for its unique stability Contrôle Analytique is also proud to offer an exclusive auto-
- Range in use (3 output) over time that assures a nice repeatability for customers in calibration features that eliminates the need for the user
- 2 alarm outputs (user programmable set point) need of a precise nitrogen quantification. These satisfied to toggle calibration gases, avoiding time wastes and gas
OPTIONS: • Auto-zero and auto-span calibration at user defined intervals customers are located all around the world in different kinds contamination (patented).
• 4-20 mA isolated output: of industries such as Air Separation Plants.
- 600 ohms at 20 mA dc
- Fuse protection
- Transient protected
- Isolated from power ground and analyzer’s electrical circuit
FEATURES APPLICATIONS
- Programmable for over- and under-drive in case of a system alarm • Interference free • Air separation plants
- Manual software control for calibration, monitoring and troubleshooting • Guaranteed 10 ppb resolution / 2 ppb noise • Argon purification plants
• Isolation: 750 VDC or AC peak maximum • Measuring system based on spectral emission • Cryogenic truck loading station
• Serial port: RS-232/422/485/modem of plasma maintained in a quartz chamber, • Speciality gas laboratories
• Custom option: contact factory controlled by a microprocessor system • Process control
• Easy to operate • Steel industries
Dimensions: • Solid state mass flow controller • Chemical plants
• Low sample flow (works with as low as 20 sccm) • Gas management system
Standard rack mount 3U • Low power (35 watts) • Welding gas control
• No valves, columns, or carrier gas • Helium liquification plants
• No radioactive source
5.25” (133.3) high • No moving parts, all solid state
18” (457.0) deep • No sensor or detector to replace
19” (482.6) wide • Maintenance free
• Real Time Clock
• Alarm Historic