Brochure Edan M3B
Brochure Edan M3B
Brochure Edan M3B
M3B
M3B
Capnograph Monitor
With SpO2 (Nellcor optional) and Respironics CO2 highly advanced monitoring technologies, EDAN M3B
ensures effective capnography monitoring for intubated and non-intubated patients for continuous long-term
monitoring. It is tailored for mechanically ventilated and non-intubated patients.
PR measurement
Respironics Loflo sidestream and CAPNOSTAT R 5
TM
Sidestream:
Compact Design
Unique bracket for CO2 module
No need for water trap
Low sample velocity of flow which is especially suitable for
neonatal patients (50ml/min)
One-Button-Switch
No need to calibrate at a regular basis
Fast and easy access by a single
Filter tip (Patent) to obstruct the water vapor from contamination
pushbutton:
Suitable for Adult, Pediatric and Neonatal patients
CO2 measurement can be turned
Long operation life to save your costs
on/off by just one button. Doctors
Special designed sensors which could support
may earn more time with the patients
Immersion Disinfection / Lens Hood feature Mainstream:
and less time with the machine.
Nellcor OximaxTM as an option. No need to calibrate at a regular basis
Suitable for most clinical tests covering adult, Suitable for any traditional breathing machine
pediatric and neonatal patients. Short warm-up time
M3B
Capnograph Monitor
Classification Accuracy
Anti-electroshock type Class I equipment Adult (including Pediatric)
and internal powered equipment 2% (70%~100% SpO2) 0 to 150 mm Hg, 0 to 19.7%, 0 to 20 kPa
EMC typeClass A Undefined(0%~70% SpO
2) (at 760 mm Hg). Barometric
Anti-electroshock degree Neonate 3% (70%~100% SpO2) Pressure supplied by Host
SpO2; BF Defibrillation type; TEMP; CF type. Undefined(0%~70% SpO
2) CO2 Resolution 0.1 mm Hg 0 to69 mm Hg
Ingress Protection IPX1(W/O Temp module) Pulse Rate 0.25 mm Hg 70 to 150 mmHg
IPX0 (withemp
T module) Measuring and Alarm Range CO2 Accuracy 0 -40 mm Hg 2 mm Hg
Specifications 20~300bpm 41-70 mm Hg 5% of reading
Size and Weight Resolution 1bpm 71 -100 mm Hg 8% of reading
Size 200.8mm(L) 41mm(H) 89mm(D) Accuracy 3bpm 101 -150 mm Hg 10% of reading
Weight 3kg Above 80 bpm 12% of reading
Display Respironics CO2 CO2 Stability Short Term Drift: Drift over four hours shall
5.7 color TFT resolution: 640 80 Mainstream not exceed 0.8 mm Hg max. Long Term
Power Supply CO2 Measurement Range Drift: Accuracy specification will be
100-240 VAC, 50/60Hz 0 to 150 mm Hg, maintained over a 120-hour period
Pmax=70VA FUSE T 1.6AL 0 to 19.7%, 0 to 20 kP
a (at Respiratory Rate Range
Battery 760 mm Hg) 2 to 150bpm
Type: Lithium-ion Rise Time Less than 60 ms -Adult / infant Respiratory Rate Accuracy 1 breath
Voltage:14.8 V DC Capacitance: 4,400 mAh Sample Cell/Filter
Reusable or Single Patient Use
Working period Proprietary single patient use sample cell
Airway Adapter
Color TFT: 480mm and inline filter are integra
ted with the
CO2 Resolution 0.1 mm Hg 0 to69 mm Hg
Rechargeable period < 360min sample line which eliminates cont amination
0.25 mm Hg 70 to 150 mm
Hg
Thermal Recorder (Optional) of the internal system
CO2 Accuracy 0 -40 mm Hg 2 mm Hg
Built-in thermal array 3 channels Nasal Sampling Kits for Non-intubatedPatients
41 -70 mm Hg 5% of reading
Paper Width 48 mm Adult, pediatric and infant nasal CO2
71 -100 mm Hg 8% of reading
Paper Speed 25mm/s sampling, nasal CO 2 sampling and O2
101-150 mm Hg 10% of reading
delivery Adult and pediatric nasal/oral
Above 80bpm 12% of reading
SpO2 (EDAN) CO2 sampling, nasal/oral CO2 sampling
CO2 Stability Short TermDrift: Drift over
Measuring Range 0 ~ 100 % and O 2 delivery
four hours shall not ex
ceed 0.8 mm
Alarm Range 0 ~ 100 % On-Airway Adapter KITS for Intubated Patients
Hg max.Long TermDrift: Accuracy
Resolution 1% Adult/Pediatric with and without dehumidification
specification will be maintained ov
er
Accuracy tubing Pediatric/Infant, low deadspace,
Adult (including Pediatric) a 120-hour period. with and without dehumidification tubing
2% 70%~100% SpO2 Respiration Rate Accuracy 1 breath Taper meets ISO5356-1
Undefined 0~70% SpO2 Calibration No routine user calibration required. Sample Kit Hours of Use
Neonate 3% 70%~100% SpO2 An airway adapter zerois required Nasal Cannula (all styles) -up to12 hours
Undefined 0~70% SpO2 when changing to a different
style of On-Airway Adapter Kits without dehumidi fication
Pulse Rate airway adapter
. tubing -up to 12 hours
Measuring and Alarm Range 30 ~ 300bpm Water Resistance Sample Cell Detection
Resolution 1bpm IPX4-Splash-proof (sensor head only) Insertion automatically turns sampling pump
Accuracy 3bpm Shock Impact EN60068-2-6 Sinusoidal Vibration on. Removal automatically turnssampling pump off.
Data update period 2s EN60068-2-27 Shock Water Resistance IPX4 -Splash-proof (When sample cell is
ISO 9919 EN60068-2-64 Random Vibration inserted in sample cell receptacle)
Able to withstand repeated6-foot Shock Impact IEC TR 60721-4-7 Class 7M3 (designed
SpO2 (optional, by Nellcor OxiMaxTM ) drops onto tiled floor while oper
ating. to withstand environments subjectto
Measuring Range 1~100% Sidestream significant vibrations or high shocklevels)
Alarm Range 1~100% Sample Rate 50 mL/min EN60068-2-27 Shock
Resolution 1% CO2 Measurement Range EN60068-2-64 Random vibration