Balometer Ebt721 Alnor
Balometer Ebt721 Alnor
Balometer Ebt721 Alnor
Model EBT721
Manometer
Model EBT720
The EBT720 is one of the most advanced, versatile, and easy to use manometers on the market today. Auto-zeroing allows you to make measurements throughout the day. The velocity matrix accessory is useful in measuring face velocity through filters, coils, and other specialized spaces.
Balometer
Model EBT721
The EBT721 Capture Hood is a multipurpose electronic air balancing instrument for reading air volume flow at diffusers and grilles. It is ideally suited for commissioning agents, test and balance contractors, facilities managers, health and safety specialists, and ventilation installers. This light weight, ergonomically designed kit saves time and money while helping to create a healthy and energy efficient environment.
Specifications
Balometer Models EBT720 and EBT721
Velocity Range Pitot probes Air flow probe Velocity matrix Accuracy Units Resolution Pressure Differential pressure 15 in. H2O (3735 Pa); 150 in. H2O (37.5 kPa), maximum safe operating pressure 15 to 40 in. Hg (356 to 1016 mm Hg) 2% of reading 0.001 in. H2O (0.25 Pa) static and differential; 2% of reading absolute in. H2O, in. Hg, Pa, hPa, kPa, mm Hg, cm Hg, mm H2O, cm H2O, 0.00001 in. H2O (0.001 Pa) static and differential; 0.01 in. Hg (1 mm Hg) absolute 25 to 2,500 ft3/min (42 to 4250 m3/h) capture hood 3% of reading 7 ft3/min (12 m3/h) at flows >50 ft3/min (>85 m3/h) ft3/min, m3/h, m3/min, l/s 1 ft3/min (1 m3/h) 0 to 95% RH temperature/RH probe 3% RH 0.1% RH 25 to 8,000 ft/min (0.125 to 40 m/s) 25 to 5,000 ft/min (0.125 to 25 m/s) 25 to 2,500 ft/min (0.125 to 12.5 m/s) 3% of reading 7 ft/min (0.04 m/s) at velocities >50 ft/min (>0.25 m/s) ft/min, m/s 1 ft/min (0.01 m/s)
Model EBT720
(Shown with optional accessories)
Applications
HVAC commissioning Clean room certification Troubleshooting HVAC systems Testing and balancing HVAC systems
Temperature Sensor in base 40 to 140F (4.4 to 60C) Temperature probe -40 to 250F (-40 to 121C) Temperature/RH probe 14 to 140F (-10 to 60C) Accuracy 0.5F (0.3C) from 32 to 160F (0 to 71C) Units F, C Resolution 0.1F (0.1C) Instrument Temperature Range Operating 40 to 140F (4.4 to 60C) Storage -4 to 160F (-20 to 71C) Statistics min, max, average up to 1000 readings Data Storage 1,000 readings, time and date stamped
Model EBT721
(Shown with optional accessories)
Hood Sizes Available (EBT721) Standard Hood Kits 801097 2 ft x 2 ft (610 mm x 610 mm) Optional Hood Kits 801201 801200 801202 801203 BSC Hood Kit 801204 801205 2 ft x 4 ft (610 mm x 1220 mm) 1 ft x 4 ft (305 mm x 1220 mm) 1 ft x 5 ft (305 mm x 1525 mm) 3 ft x 3 ft (915 mm x 915 mm) 8 in. x 22 in. (205 mm x 560 mm) 10 in. x 22 in. (255 mm x 560 mm)
The BSC hood kits are used to certify Class II bio-safety cabinets by taking direct in-flow measurements for NSF compliance. EBT720
Air capture hood, frame and base Measures air volume/flow rate Static/Differential Pressure (air) Air velocity, temperature, relative humidity probes (optional) Pressure sensor Automatic density correction Backpressure compensation Data logging (download/recall) Field calibration Statistics (minimum, maximum, average) Certificate of Calibration
Specifications subject to change without notice.
The BSC hood kits are used to certify Class II bio-safety cabinets by taking direct in-flow measurements for NSF compliance.
Recommended Accessories 800187 Air flow probe, 18 in. (46 cm) 800188 Temperature probe 800189 Humidity and temperature probe 800190 Velocity matrix, telescopic handle, (2) 8 ft. (2.4 m) neoprene tubing sections
EBT721
USA
Contact your local Alnor Distributor or visit our website www.alnor.com for more detailed specifications.