Proeazy
Proeazy
Proeazy
DGLYLVLRQRI
Scientific Limited
Instruction manual
3&96HUXP)DHFHV
6SHUP &\WRORJ\
IMPORTANT Read before using
• Check the Centrifuge and the Packaging for any shipping damage.
Inform the shipping company and Centurion Scientific immediately
• Check the order for completeness, if not contact Centurion Scientific
Location
The Centrifuge should only be operated indoors.
The set up location must fulfil the following requirements
• A safety zone of at least 30Cm must be maintained around the Centrifuge
• People and hazardous substances must be kept out of this zone
during centrifugation
• The bench or supporting structure must be stable and free from resonance
• The Bench or supporting structure must be suitable for horizontal use of the
centrifuge
• The Centrifuge should Not be exposed to heat or strong sunlight
• Ultra violet rays can damage the stability of plastics and coatings
• Do not subject the Polypropylene, Delrin rotors or accessories to direct
sunlight
• The Zone location must be well ventilated at all times
• Store the original packaging, Keep this for future service to base , if required.
2
Fixing of rotor
Motor shaft
3
Rotor balancing. Very Important
4
Tube placement micro rotors
Haematocrit Rotor
✔ ✘ ✘
ALWAYS load your tubes opposite to each other
ALWAYS weigh samples to within0.1 Gram. For best separations. (see page 4)
ALWAYS fit the lid supplied
CHECK the Rim seal gasket and replace every 2 months (Part number 8011)
CHECK the rotor screw fixing tightness regularly
...........................................................................................................................................................................
✔ ✘ ✘
ALWAYS load your tubes opposite to each other
ALWAYS weigh samples to within0.1 Gram For best separations. (seep page 4)
ALWAYS fit the lid supplied
CHECK the two top seals and replace regularly (Part number 8044)
CHECK the rotor screw fixing tightness regularly 5
Tube placement Fixed angle rotors
✔ ✘ ✘
ALWAYS load your tubes opposite to each other
ALWAYS weigh samples to within 0.1 Gram. For best separations. (see page 4)
CHECK The rotor screw fixing tightness regularly
...................................................................................................................................................................................
✔ ✘ ✘
✔ ✘ ✘
ALWAYS load your tubes opposite to each other
ALWAYS weigh samples to within 0.1 Gram. For best separations. (see page 4)
CHECK The rotor screw fixing tightness regularly 6
Tube & bucket placements Swing out rotors
✔ ✘ ✘
ALWAYS load your tubes opposite to each other
ALWAYS weigh samples to within 0.1 Gram. For best separations
ALWAYS fit ALL the Tube Holders and any supports
CHECK The rotor screw fixing tightness regularly
.................................................................................................................................................................................
✔ ✘ ✘
ALWAYS load your tubes opposite to each other
ALWAYS Weigh Bucket and all samples within to 1.0 Gram
ALWAYS fit ALL 4 Buckets and 4 Adaptors supplied
CHECK the rotor screw fixing tightness regularly
.................................................................................................................................................................................
Microtitor Rotor
✔ ✘
ALWAYS load your plates opposite to each other
ALWAYS weigh samples to within 1.0 Grams. For best separations
ALWAYS fit both Buckets supplied (2)
CHECK the rotor screw fixing tightness regularly
7
Tube placements Swing out rotor
Wrong
Wrong
Even though there are two samples in two buckets opposite to each other it is not balanced.
NEVER run swing out rotors with only two buckets
8
A B
Tube placement Cytology & Cell wash rotors
Cytology Rotors
✔ ✘ ✘
ALWAYS load your holders opposite to each other
ALWAYS weigh samples to within 0.5 Gram. For best separations
ALWAYS fit the lid supplied
Check the two top Gaskets and replace regularly (Part Number 8055)
CHECK the rotor screw fixing tightness regularly
................................................................................................................................................................................
✔ ✘ ✘
ALWAYS load your tubes opposite to each other
ALWAYS weigh samples to within 0.2 Gram. For best separations Regularly smear a small amount of Silicon grease over the O rings
Check the inner and outer O rings and replace regularly (Part Number 8066)
CHECK the rotor screw fixing tightness regularly
9
Index. Instruction manuals for.
Micro's
EZ915 Micro tube Centrifuge. Complete with BRK5424 (A), BRK5436 (B),
BRK5448 (C), BRK5494 (D)
EZ 906 Fixed angle rotor Centrifuge. Complete with BRK5408 (A), BRK5404 (B)
...........................................................................................................................................................
Small to .5Litre
EZ706 Fixed angle rotor Centrifuge. Complete with BRK5412 (A), BRK5406 (B) rotor
..........................................................................................................................................................
Medium to 1Litre
EZ504 Swing out rotor Centrifuge. Complete with BRK1001 rotor & B5250 Bckts
EZ506 Fixed angle rotor Centrifuge. Complete with BRK5324 (A), BRK5308 (B)
Instructions of use
Program Fault
1 Lid open
2 Imbalance
3 None revolve
4 Inverter
Start.
Stop Function Lid open
................................................................................................................................................................................
3
Once running, your selected program will flash
4 When the run is finished it will flash faster
Function
Once your program has been selected, load your tubes, Start.
Stop
evenly and same weight, close the lid and press start.
Selected program light will Flash whilst running, when
finished A buzzer will sound (can be switched off)
Fault
Lid open. Close lid firmly (check that nothing is fouling closure)
None revolve Rotor will not turn, open lid, check if it turns freely, are your tubes
too tall?
If problem persists contact Service
Inverter A fault has occurred. Turn off your Centrifuge wait 3 minutes
Turn on again.
If problem persists contact Service
Programs per Model. Micro's
EZ913 1 2 3 4
EZ904 1 2 3 4
Program Serum 1 Serum 2 Urine Sperm
Speed (Rpm) 4,000 4,000 2,100 3,300
Time (Seconds) 420 600 600 900
Accelleration 9 8 8 8
Deceleration 9 8 8 8
EZ915
1 2 3 4
Program 15 mins 10 mins 5 mins Pellet
Speed(Rpm) 15,000 15,000 15,000 15,000
Time (Seconds) 900 600 300 30
Accelleration 9 8 8 9
Deceleration 9 8 8 9
EZ906
Program Serum Faeces Urine Sperm
Speed(Rpm) 6,000 1,780 2,080 3,250
Time (seconds) 300 300 600 900
Accelleration 9 8 8 8
Deceleration 9 8 8 8
Rcf(g) 3421 301 411 1004
Rotor. BRK5408 (A), BRK5404 (B)
Programs per Model Small
EZ704 1 2 3 4
EZ706 1 2 3 4
EZ504
1 2 3 4
Program Serum Faeces Urine Sperm
Speed(Rpm) 4,000 1400 1600 2,500
Time (Seconds) 300 600 300 900
Accelleration 9 8 8 8
Deceleration 9 8 8 8
EZ506 1 2 3 4
Program Serum Faeces Urine Sperm
Speed(Rpm) 5,000 1,500 1,750 2,750
Time (seconds) 240 600 300 900
Accelleration 9 8 8 8
Deceleration 9 8 8 8
Rcf(g) 3350 301 411 1010
Rotor BRK5324, (A) BRK5308 (B).rotor
Fault codes
Prime and Dual models only. Ambient
Fault codes are shown in the Speed section of the LED display (shown on previous page
The following List show what each Code means and what to do.
Noisy
. Unbalanced tubes, see Rotor pages
. Lid seal needs replacing
. Rotor Lid not fitted
Warning
The rotor can still be spinning at a high speed, if touched it can cause serious injuries.
Always wait till the rotor has stopped.
Proceed as follows
• Make sure the Rotor has stopped (view via the port in the Lid)
• Once stopped turn off the power via the inlet Switch
• Pull the Centrifuge slightly over the bench (as shown in (a)
• You will notice one (b) or some models two plastic plug (c). Plug *
• Pop open with a screwdriver (tool) and pull downwards .
• The lid will pop open You can now retrieve your samples
b
a
c c
After replace the Plastic plugs back into their original position.
Reconnect the centrifuge power line once the mains power has been restored.
Check all is working correctly by closing the lid, wait 5 seconds then press lid open.
If not working refit or Contact your Distributor
16
Care & maintenance. Centrifuge
Always keep your Centrifuge clean and free from moisture at all
! times, this will help prevent corrosion.
! It is recommended that your centrifuge is switched off and the lid left
open overnight to ensure that any moisture build up can evaporate.
Use only official Centurion Scientific Ltd accessories with your centrifuge.
!
! Yearly servicing is required to validate your warranty and increase the life
of your Centrifuge and accessories. See PrO-Care pages
10
Care & maintenance. Rotor
Ensure that all buckets are present in the rotor when running your
! centrifuge, as using the unit with an unbalanced rotor can cause distortion
and excessive wear. Read tube placements section in the previous section
! Do not use any kind of scourers or abrasive material to clean the buckets
and ensure that a pH neutral cleaner is always used.
! Aluminium buckets should be kept clean and moisture free to keep them in a
good, safe operating condition.
! Yearly servicing is required to validate your warranty and increase the life
of your rotor and accessories
11
PrO-Care
Service, Warranty & Maintenance pakages
WWW.ServicePrO-Care.com