Common Uses and Cited Complications of e
Common Uses and Cited Complications of e
Common Uses and Cited Complications of e
Energy in Surgery
Ganesh Sankaranarayanan1, Rajeswara R. Resapu1, Daniel B. Jones2, Steven
1
Center for Modeling, Simulation and Imaging in Medicine, Rensselaer Polytechnic
Institute, Troy, NY
2
Beth Israel Deaconess Medical Center,
Funding: NIBIB/NIH grants # R01EB010037, # R01 EB009362, # R01 EB005807 and # R01
EB014305
Corresponding Author:
Prof. Suvranu De
JEC 2049
Email: des@rpi.edu
ABSTRACT
Background: Instruments that apply energy to cut, coagulate and dissect tissue with
minimal bleeding facilitate surgery. The improper use of energy devices may increase patient
morbidity and mortality. The current article reviews various energy sources in terms of their
M ethods: For the purpose of this review, a general search was conducted through
NCBI, SpringerLink and Google. Articles describing laparoscopic or minimally invasive surgeries
using a single or multiple energy sources are considered, as are the articles comparing various
Results and Conclusion: A review of the literature shows that the performance of
the energy devices depends upon the type of procedure. There is no consensus as to which
device is optimal for a given procedure. The technical skill level of the surgeon and the
knowledge about the devices are both important factors in deciding safe outcomes. As new
energy devices enter the market increases, surgeons should be aware of their indicated use in
argon beam coagulation, laser, radio frequency ablation, thermal damage, embolism, fire.
INTRODUCTION
The vast majority of surgical procedures performed throughout the world involve the use
of some device that applies energy to the tissue for cutting, coagulation, desiccation or
fulguration for the destruction or manipulation of the tissue. While various energy sources
including electricity, ultrasound, laser, argon gas, microwaves or radiofrequency waves may be
used, the fundamental principle involves tissue necrosis and hemostasis by heating. The
and the unraveling of collagen helices around 60ºC. Protein denaturation occurs between 70ºC
and 80ºC resulting in coagulation. Further heating to 90ºC results in dehydration or dessication.
Beyond 100ºC, the intercellular water boils, eventually vaporizing the cell allowing tissue cutting.
standardized curriculum for surgeons that address the physics, safe use and complications
associated with these devices that promote the best outcomes for patients. In a recent study [2],
it was found that many surgeons have knowledge gaps in the safe use of widely used energy-
based devices. To address this issue, the Society of American Gastrointestinal and Endoscopic
Surgeons (SAGES) has recently initiated the Fundamental Use of Surgical Energy (FUSE)
program to develop an educational curriculum that will cover both didactic and hands-on training
approaches to the use of energy in interventional procedures in the operating room and
endoscopic procedure areas. The goals are to prevent untoward events such as operating room
M ETHODS
The review includes a thorough search of literature concerning the physics, applications,
success, complications and comparison of various energy sources in surgery. Priority was given
to human studies and laparoscopic procedures. However, the studies involving animals (both in
vivo and in vitro) and open procedures which were unique and relevant to the assessment of the
energy sources were considered. Emphasis was placed on more recent studies covering the
latest techniques and studies conducted on live patients in the United States. The keywords
used in the search were numerous and the websites such as Medline, PubMed, SpringerLink
and Google were extensively covered along with various books published in the field.
Electrosurgery
Electrosurgery was developed by Dr. William T. Bovie in the 1920s, where a spark gap
generator was used to build the first electrosurgery tool, commonly known as RF knife or ‘Bovie’
[3]. In the 1950’s, the first bipolar unit was built by Dr. Leonard Malis, wherein two electrodes
Mechanism of Electrosurgery
In electrosurgery, heat is generated in the tissue by the flow of radio frequency (RF)
electric current unlike electrocautery where the heat is transferred directly from the tool to the
tissue. The use of RF current (voltage in the range of 300 to 500 KHz) eliminates
neuromuscular stimulation, which ceases above 100 KHz. When the RF electrical energy is
made to concentrate in a very small area in the tissue, typically by applying the energy through
pointed or hooked tool tips, the resulting high concentration of current flow in a narrow area
increases the cellular temperature which leads to various effects on the tissue including,
The RF energy can be applied to tissue by using either monopolar or bipolar tools. In
monopolar electrosurgery, the electrical circuit is completed by the passage of current from the
active electrode at the surgical site to the dispersive electrode (or the return electrode) attached
to the body of the patient. The active electrode can be of any form (usually a point, hook or a
blade) with sharp edges and/or blunt edges. The sharp edges increase the current density (the
amount of current per unit area) and used for cutting whereas the blunt edges are used for
coagulation. The return electrode is usually a wide pad, attached to the skin of the patient,
which disperses the heat and safely leads the current out of the body. In bipolar electrosurgery
both active and return electrodes are located in the same tool and the electrical circuit is closed
by the small area of tissues that are grasped or manipulated by the tool. Bipolar tools are, thus,
usually designed as grippers or forceps. Since the current only has to travel short distances in
bipolar surgery, the voltage required for the surgery is low. Lower voltage is better for uniform
drying of the tissue which minimizes the chance of re-bleeding. Thus the bipolar devices are
wave form generators with different duty cycles (percentage of time the energy is applied).
Different tissue effects such as cutting, coagulation can be achieved by selecting different duty
cycles with 100 % duty cycle enables cutting whereas lower duty cycles can be effectively used
different, they usually provide selections such as pure cutting, coagulation and blended
waveforms.
The waveforms with different duty cycles can be used to produce four main effects in
electrosurgery namely, cutting, coagulation, desiccation and fulguration. The cutting is achieved
by using a continuous waveform (100 % duty cycle) applied through the active electrode of a
monopolar electrosurgical tool with a pointed tip. The narrow tip allows for large current
concentration and when placed near the tissue but not in contact, generates an arc through
which the current rushes to the tissue generating large amount of heat (greater than 100°C)
which leads to rapid tissue vaporization and induces cutting. When a blunt instrument tip is used
with contact on the tissue, the decreased current concentration due to increased surface area
leads to increase in the tissue temperature but not to point of vaporization and creates a
perform coagulation or desiccation, a lower duty cycle high voltage waveform is used but can
also be performed with 100 % duty cycle lower voltage cutting waveform as well. Finally in
fulguration, a lower duty cycle high voltage waveform is applied through the active electrode of a
pointed monopolar electrosurgical tool tip in non contact mode close to the tissue. With high
voltage and low duty cycle ( usually 6 %), the heat generated by the current flowing through the
arc from the tool tip heats the tissues to form coagulum and with repeated application, increases
The most commonly used electrosurgery devices are listed in Table 1. The bipolar
instruments provide compression force in addition to thermal energy, which helps in better
sealing of blood vessels and attaining better burst pressures. Thus these instruments seal larger
blood vessels as opposed to monopolar devices which are limited to smaller vessels - usually
less than 2mm [4]. The earlier generation of electrosurgical tools did not have temperature
control and hence, resulted in thermal injuries. The latest instruments (e.g., Ethicon
today. According to the Association of periOpertaive Registered Nurses (AORN), there are
around 40,000 patient burn cases annually due to faulty electrosurgical devices and in 1999
alone, nearly $600 million was paid in claims for those injuries [5]. Care should be taken when
operating the electrosurgical devices, particularly monopolar devices. The sparking effect at the
tool tip may cause an explosion when it comes into contact with inflammable gases that are
often used for anesthesia during the operation. Further, the current travelling through the body
can interfere with any implanted medical devices such as pacemakers [6] and defibrillators. A
metal instrument or implant that comes in the way of the current passing through the body may
create a different loop causing tissue damage in unwanted regions. Other mechanisms through
which injuries can occur during electrosurgery include insulation failure [7], direct and capacitive
coupling. Insulation of the electrosurgical tool may break due to repetitive use of the equipment,
high intensity of current flowing through the wire and in repeated sterilization process. Burns
occur at places of insulation defect and can be fatal especially when the defect is small which
leads to high current concentration. Moreover the insulation defect is often very small and will
be difficult to detect using the naked eye and an active electrode monitoring system may abate
injury [8]. Direct coupling occurs when the active electrode is either intentionally or inadvertently
touched by another tool or scope, in which case, energy can be transmitted through the other
procedures, it is often dangerous since the view of surrounding organs is limited. Capacitive
coupling may occur in a laparoscopic surgery when the tools, tissues and trocars are in close
proximity creating capacitance effect (build up of charge between two conductors separated by
an insulator). This stored charge may discharge causing unintended tissue damage in the
immediate vicinity. If possible, the use of metal and hybrid (with plastic) cannula should be
avoided to eliminate injuries. This may be a particular hazard in the emerging technique of
single port laparoscopic surgery where all of the instruments are in close proximity. Though the
spread of current through the body is eliminated by the use of bipolar electrosurgery, chance of
damage to adjacent tissues still exists [9]. Apart from the common laparoscopic complications
that arise due to surgical error, the other major complication from electrosurgery is the thermal
injuries to adjacent organs. However, the complications can be both intra-operative and post-
operative and are specific to the type of procedure performed. Complicating the matter further is
that the maximum temperature and thermal spread when using energy based devices varies
based on the types of target tissue and the type of energy sources used [10] . Monopolar
electrosurgery was shown to have higher temperature and thermal spread [11-13]. Various
studies on laparoscopic procedures have shown complications while using monopolar and
bipolar electrosurgical instruments that includes conversions, failures, and recurrences [14-19].
There are cases in which death has been reported when using monopolar electrosurgical device
[20, 21]. Specifically Agarwal et al. [22] mentioned that the use of an energy source, especially
monopolar electrosurgery is the culprit for many of the injuries during laparoscopic
cholecystectomy.
Electrosurgery has the highest number of thermal injuries, but it still is one of the most
popular techniques used in laparoscopy. Out of the two electrosurgical modalities, monopolar
electrosurgery causes the most thermal damage but bipolar devices have also shown to
produce thermal damage [11]. Studies have shown that the bipolar device has the least amount
of thermal spread among the various energy devices [11, 12, 23] and provides safe sealing and
cutting quality that are similar to other energy based devices [20]. Bipolar electrosurgery devices
have shown to have shorter dissection time, provided better seal quality, lesser blood loss,
fewer conversion rates and are more cost effective than monopolar electrosurgery [24-26]. In
laparoscopic cholecystectomy (LC), a common laparoscopic procedure which involves the use
of energy, bipolar modality is shown in one study to have performed better than monopolar
electrosurgery [27]. In a number of other studies comparing different energy devices, the
bipolar electrosurgery devices have performed better than devices using other forms of energy
sources [26, 28-30]. High success rate have also been reported in surgical outcome reports
using bipolar energy devices [14, 15, 18]. Hence the cost of bipolar electrosurgery devices may
be justified for complex surgical cases. Monopolar technique might be preferable for simpler
surgical procedures when adequate care is taken while operating in the vicinity of critical
organs. For example, making incisions on the skin before inserting the laparoscopic instruments
is a job for the monopolar tool. The risk from electrosurgery can further be reduced by limiting
the thermal damage during electrosurgery. For example, Dodde et al. [31] reported a novel
thermal management system to reduce the thermal spread during monopolar electrosurgery.
One common safety measure is the optimization of the voltage application time. Most of the
electrosurgical units come with various power settings for cutting and coagulation. The right
setting should be chosen for the specific procedure. Tissue damage is further reduced by the
introduction of a hydrating medium to keep the surgical area wet and moist. As long as the
above mentioned precautions are taken and the common risk factors are considered,
Ultrasonic energy
The use of ultrasonic energy in medicine has been reported as early as 1960, where it
was used to treat Meniers’s disease. It has been used in the cutting and coagulation of tissue in
the late 1980s [33] where Amaral [34] popularized the technique and used it successfully in over
activated scalpel (UAS) is to use the low frequency mechanical vibrations (ultrasonic energy in
the range of 20-60 kHz) of the tool tips or the blades for tissue cutting and coagulation [32]. The
mechanical vibrations when transferred to the tissues on contact induces protein denaturation
by breaking down the hydrogen bonds in tissues due to the internal cellular friction caused by
the vibrations [33]. The mechanical vibrations are produced by the piezoelectric transducers
embedded in the tools which convert the applied electrical energy to mechanical vibrations
which are then transferred to the active blades for cutting or coagulation. The HS operate at a
frequency of 55.5 kHz whereas devices operating at various other frequencies also exist [35].
The cutting using an Ultrasonic surgical instrument is achieved by two methods. For
tissues and muscles with high protein densities, the mechanical stretching of the tissues beyond
its elastic limit due to the longitudinal motion of the sharp blades between 60 to 100 µm at 55.5
kHz is used for cutting. For tissues with low protein densities, such as liver, cavitation effect in
which intercellular water is vaporized at lower temperatures due to mechanical vibrations, there
by rupturing the cells is used for cutting. In general the cutting and coagulation in UAS depends
on various factors such as grip pressure, the shape and area of the blades in contact with the
The major advantage of using UAS is that it produces less heat compared to other
energy devices (less than 80° C compared to 100° C for electrosurgery) thereby reducing the
risk of thermal injury [35]. Due to lesser heat generation, charring and desiccation is also greatly
reduced. Since no smoke is produced, except for the mist produced due to cavitation effect
which dissipates much faster, UAS offers unobstructed view for endoscopic/ laparoscopic
procedures. The UAS does not transmit active current in the tissues and thereby eliminate any
risk of electric shock. The most commonly used ultrasonic energy devices are listed in Table 1.
Not many complications were reported in the use of harmonic scalpel in laparoscopy.
electrosurgery, altering of the frequency or impedance of the surgical system itself due to blade
fatigue, temperature elevation, excessive applied pressure, or improper use. Ultrasonic energy
causes atomization of fluid, which may create a transient mist. However, overall dissection time
may be shorter with ultrasonic cavitation aspirators (UCA) or ultrasonic-activated scalpels after
the initial learning curve [36]. It has been shown in many studies that the ultrasonic devices are
not as efficient in sealing medium to large sized blood vessels [26, 37, 38]. It is also shown that
UAS produces higher average temperatures [10] and is not reliable in sealing vessels larger
than 3mm [39]. In [40] it was shown that though there is no visible injury to naked eye during
dissection experiments on a swine using a UAS, histological examination had revealed serious
laparoscopic surgery includes injury to sigmoid colon [41], postoperative bleeding [42] and an
ischemic lesion [28]. Even the overheating of tissue (the non active blade of a UAS) after a
continuous activation of more than 10s resulted in histological damage to the intestinal mucosa
in a porcine [43].
The general conclusion of most studies advocating the use of ultrasonic energy is that
the minimal thermal spread leads to minimal thermal injury. However, there are studies that
contradict the statement by measuring the temperature of the tool tip instead of thermal spread
in the tissue. In [44] it was shown that at higher power settings, the ultrasonic devices
(Ultracision and Autosonix) created large thermal spread (up to 25.7 mm) and high
temperatures (140ºC at 10 mm distance) in porcine organs. In another study, the harmonic ACE
[10] is shown to have taken twice as long to cool down when compared to the other devices and
the temperatures generated by the ACE were inversely proportional to the thickness of the
tissue. Kinoshita et al. [48] says that the temperatures (150ºC) and thermal spread (10mm)
caused by ultrasonic device is far less than those caused by electrosurgery (350ºC and 22mm
respectively) in the porcine blood vessel cutting and coagulation. There are various reports of
Ultrasonic scalpel (Harmonic ACE) was also successfully used in division of pulmonary vessels
in video assisted lung resection [58]. Though proven to be an effective tool in gynecological
procedures, it is not a good tool for delicate reconstructive surgery for fertility due to the
cavitational effect [47]. Janssen et al. [51] showed that in laparoscopic cholecystectomy, the
learning curve of the surgeons using HS was very short compared to electrosurgery. The
general conclusion that can be drawn from the observations is that ultrasonic energy has more
advantages than disadvantages. Given the fact that more recent literature is available in the
field than other energy methods and the increasing market share of ultrasonic medical devices
[59], it can be said that ultrasonic energy may have an increasing role in the field of surgery.
Lasers
The first use of lasers in laparoscopic surgery was recorded in 1979 [60] and the regular
use of laparoscopic laser surgery has been reported as early as 1982 [61]. In a very short span
of time, lasers became very widespread in the medical field - ranging from cosmetic treatments
to highly complicated surgeries such as atrial fibrillation (AF) treatment [62, 63]. Today, lasers
electromagnetic or light waves are amplified multiple fold in an optical resonator (which contains
mirrors and a gain medium) and passed out in the form of high intensity light waves. The
amount of amplification in the resonator determines the amount of energy transmitted by the
light waves which are then absorbed by the tissue. This energy absorbed by the tissue then
manifests itself into heat which cuts and coagulates the tissue. The frequency of the laser beam
determines the width of the laser beam that can be generated (the higher the frequency of the
wave, the lower the diameter of the beam) and most commercial lasers use infra-red to
ultraviolet frequencies for medical applications. The power or intensity of a laser is measured in
terms of 'irradiation' defined as the ratio of power applied to the spot-size (cross sectional area)
of the laser beam (W/m2). Though only two variables (power and spot-size) are required to
calculate the value of irradiation, two other important variables need to be considered when
using lasers in surgery – time of exposure and wave length (or frequency). The use of lasers to
generate heat for destroying tumors is known as 'photo-thermal' therapy. The changes caused
by non-thermal mechanisms while using a laser are called 'photo-chemical' processes. Usually,
in photo-chemical processes, the amount of irradiance is so low that, instead of generating heat,
it induces chemical reactions in the cells thus causing inactivity. When the amount of irradiation
is too high (108 to 109 W/cm2), plasma formation takes place and leads to a 'photo-plasmal'
process. If the electric field is too strong, the result is a small region of plasma which is
associated with dielectric breakdown, strong electric fields, shockwave formation and tissue
rupture. Once the plasma forms, the tissue properties become immaterial as all the laser energy
Lasers can be classified into contact or non-contact types – based on their interaction
with the tissue. In non-contact laser mechanism, the tip of the laser delivery device remains at a
distance from the tissue to where it is being focused. When this kind of mechanism is used in a
liquid medium, an explosive vapor bubble is formed at the tip of the tool which carries energy to
the tissue. This type of energy delivery is preferred in a few ablation processes [65]. The contact
laser is the direct contact of the tool tip with the tissue. This causes the direct transfer of energy
from the tool to the tissue. The selection of a contact or non-contact laser depends upon the
sensitivity and accessibility of the tissue being operated on. The most commonly used Laser
need for advanced training in laser and laparoscopic surgery, risk of fire from flammable
materials ignited by lasers and increased operative time. The increased sedation period due to
the length of the operative time also leads to longer recovery time. Cellular damage around the
area of laser impingement can also be expected depending upon the size of the laser tip. One
of the major complications using laser as the energy source is the air embolism which can be
fatal [66-77]. Another complication with laser laparoscopic cholecystectomy is the injury to the
hepatic artery with pseudoaneurysm formation and hemobilia [78]. Hemorrhage has been
reported in other studies as well. In a review of 2344 laser laparoscopies over an 11-year period
(1982 - 1993), Ewen et al. [79] reported nine significant complications in which three cases of
The efficiency of lasers in laparoscopy were reported as early as 1989, when Reddick et
al. [80] studied 25 cases of laparoscopic cholecystectomy (in the US) and found no major
used in the treatment of infertility [81-86]. It is also a common source of energy in cosmetic and
eye surgeries. Success rates of more than 90% have also been reported in review of cases that
used CO2 laser along bipolar forceps in laparoscopic surgery [87, 88]. Advantages of
laparoscopic laser surgery over open techniques include minimal surgical morbidity, decreased
laparoscopic surgery include air embolism, hemorrhage and surgical emphysema. Moreover a
Ward et al. [90] first reported the use of argon beam coagulation (ABC) in head and neck
surgery in 1989. The use of ABC in MIS is first reported by Low et al. [91] in 1993. Numerous
studies describing the efficiency as well as the dangers of ABC have been well documented [92-
94].
Mechanism of argon beam coagulation:
bleeding. In ABC, a directed beam of argon gas from the electrode tip aids in conduction of the
radio frequency current to the tissue by ionization. Like laser, this is a non-contact method
where the argon gas - which is a good conductor of electricity - acts as a means of
transportation of the current from the tool to the tissue. ABC performs faster than conventional
coagulation systems and provides a more uniform and shallower coagulation region which
results in faster dispersion thus minimizing tissue damage. It also produces less smoke than the
conventional system. Since the argon has higher density, a jet of argon gas typically move the
blood away from the surface for effective coagulation and resulting in lesser eschar. The
protection of the active electrode tip from exposure to oxygen also results in less charring [95].
The ABC system is usually connected together with an electrosurgical system where argon gas
is released from the tip of the tool to achieve hemostasis. The most commonly used ABC
A major limitation to the use of ABC system is the potential danger of argon gas
embolism. Numerous instances of cardiac arrest were reported during the use of ABC due to
gas embolism [93, 94, 96]. Embolism (blockage of blood vessels) occurs due to the insolubility
of the argon gas in blood. The gas forms bubbles or cavities that can travel through the blood
stream and cause blockages in blood vessels. Death due to argon gas embolism has also been
reported [97-99]. Cases of non fatal argon gas embolism have been reported in [100-104].
Despite the risks, ABC continues to be used in surgery. A number of successful cases in
the use of ABC have been reported [105-109]. Dowling et al. [110] reported that the ABC was
for the safe use of ABC during laparoscopy have been tabulated by various researchers and
commercial manufacturers. Some important guidelines are, The flow rate for argon should be
chosen as low as possible to reduce the risk of argon gas embolism [111], direct contact of the
tool tip on the organ should be avoided and the electrode tip should be held at an oblique angle.
It was mentioned in a study that even at the point of longest application, the temperature
developed while using ABC was never higher than 100˚C in complete coagulation [112]. There
are other studies such as the one by Bobbio et al. [113] wherein argon beam technique is
compared with traditional surgery in the treatment of primary spontaneous pneumothorax (PSP)
using video assisted thoracic surgery (VATS) where no significant benefits of ABC were found.
The use of ABC has resulted in numerous cases of fatal and non fatal embolisms.. So, the use
of ABC must solely depend on the skill and discretion of the surgeon.
Radio frequency or RF which ranges from 3 kHz to 300 MHz is the type of electromagnetic (EM)
radiation that is commonly used in electrosurgery. RF has the lowest frequency of all the EM
waves and hence takes longer than other EM waves to generate heat in the tissue. It has been
shown in a number of studies that EM in the RF range is the most effective form of radiation.
Mechanism of RF
There are two mechanisms of RF that are used in MIS. One of them is the laparoscopic
electrosurgical usage of RF. The mode of operation is the same as that of electrosurgery where
the current applied to the tissue through the scalpel falls is the RF range. RF in electrosurgery
can be used in both monopolar and bipolar modes. Apart from laparoscopy, RF can also be
used in a percutaneous setting. Percutaneous treatment involves the insertion of a needle into
the organ to be operated on via a catheter inserted through the skin of the patient. RF current is
then applied to the tissue through the needle. Though it is different from laparoscopy, it is also a
MIS procedure where the needle tip is usually guided into the body by an ultrasound positioning
system. The most commonly used RF energy devices are listed in Table 1.
The most common usage of the RF in is radio-frequency ablation (RFA), which is also
number of studies show the application of LRFA in various procedures [17, 19, 114]. Beyer et al.
[114] studied minimally invasive bipolar radiofrequency ablation of lone atrial fibrillation (AF) in
100 patients at 3 North American Institutes between 2005 and 2007 and reported postoperative
complications that includes pacemaker requirement (5%), phrenic nerve palsy (3%),
hemothorax (3%), transient ischemic attack (1%) and pulmonary embolism (1%). Case of
deaths and high morbidity has also been reported in laparoscopic cholecystectomy procedure
2003, it was reported that RF ablation treatment of atrial fibrillation (arrhythmias) performed
using a catheter can severely narrow the pulmonary veins due to the formation of scar tissues
[116]. In rare cases, acute renal failure associated with radiofrequency liver ablation has been
observed [117]. Most complications of RF seem to arise when it is used in the vicinity of the
heart. This is due to the interference of the electrical with the electrical activity of the heart.
radiofrequency (RF) ablation of the fast atrioventricular (AV) nodal pathway [118]. In a study of
management of hepatic malignancies using RFA of malignant liver tumors in 608 patients
cartilage reshaping (LCR), it was shown that RF method allowed more uniform heating of larger
tissue samples but the lack of precise control of spatio-temporal distribution of heat limits the
usage of RF use in LCR [116]. The advantages of RF energy and the necessity to reduce
complications have resulted in the introduction of several new commercial devices [120, 121] .
In [122], the Gyrus Plasma Trissector (GPT), a novel bipolar RF system in laparoscopic radical
prostatectomies was shown to improve coagulation, reduce or eliminate sticking, seal large
vessels and allow secure grasping and dissecting of tissue. Ligasure, a radio-frequency-
energy-driven bipolar fusion device is used in many laparoscopic procedures [123-127]. For the
minimally invasive treatment of localized renal tumor, RFA along with cryoablation is shown to
be the most used and potentially promising therapies [128]. For the treatment of renal cell
carcinoma, it was shown in [129] that RFA assisted laparoscopic partial nephrectomy was
effective in providing hemostasis and in short term cancer control and also shown in [130] that
RFA treatment for small renal cell carcinoma found have significantly improved the quality of
life in most of the patients. In [131], it is also reported the reduction of intraoperative bleeding
and blood transfusion using RF assisted laparoscopic liver resection. However, Hompes et al.
[132] states that the laparoscopic liver resection itself is a procedure with minimal blood loss
between saline infused RFA and dry RFA [133], it was shown that wet RFA caused larger lesion
sizes in 10 porcine kidneys and one cycle of wet RFA was sufficient to cause irreversible cell
death compared to two cycles required using dry RFA. Although, percutaneous RFA in liver
treatments is less invasive and is considered the first choice, RFA with laparoscopic guidance is
highly recommended for patients with a relative contraindication for percutaneous RFA, such as
lesions adjacent to the gastrointestinal tract, gallbladder, bile duct and heart [134]. Similarly,
LRFA was also highly recommended for the treatment of hepatocellular carcinoma [135]. In
general, the most usage of RFA either laparoscopically or percutaneously is observed in the
liver and renal tumor ablation. However, a variety of studies featuring RFA in various surgeries
also exists. RFA is reported in the treatment of lower extremity varicosities (a minimally invasive
cosmetic procedure) where it fared better than stripping and foam sclerotherapy, although not
as effective as laser therapy [136]. A study of bipolar RF in the treatment of plantar fasciosis
[137] in patients who couldn’t be treated with conservative methods showed an improvement in
all 10 patients within a time span of 6 months to 1 year without any postoperative complications.
(LC) procedure an example (Table 2). In LC the common energy sources used are laser,
electrosurgery and ultrasonic energy. Since many of the case studies don’t explicitly state
injuries occurred while using the energy device, for LC, parameters such as mean operating
time, mean hospital stay, mean recovery time along with complications such as bile duct injury
(BDI), bile leak, conversion rate to open surgery and any reported death were collected for each
study and represented in Table 2. From the data, HC and Ultrasonically Activated Scalpel
(UAS) performed better than clip and electrosurgery methods based on operating time and bile
duct injuries. Huscher et al. [138] sates that the main advantage of UAS is the low-risk
dissection in the proximity of biliary structures. More over with UAS both cystic duct and vessels
can be separated without the need for ligature. The relative bloodless field of view when using
UAS also helps in discriminating anatomical structures. Redwan et al. [139] states that when
comparing HS and clip and electrosurgery for LC, HS performed better with lesser operating
time and the absence of major or minor bile leaks. No clear conclusion can be made regarding
laser since the operating time and injuries when compared to clip and electrosurgery varied
among the different studies and no direct comparison study exists between laser and
ultrasonically activated devices. Between monopolar and bipolar electrosurgery, though both
Discussion
A comparative study of the literature shows that the preferred source of energy in
laparoscopic surgery in the early 1990s was the monopolar electrosurgery, while bipolar and
laser was used much less frequently. However, in recent times, even with the advent of new
tools in electrosurgery, the preference is gradually shifting towards ultrasonic energy due to its
procedure, is rarely performed and harmonic scalpel (HS) has been used more often in recent
times. Fiber optic cables have made lasers more readily available in laparoscopy. The use of
coagulation (ABC) is a very effective technique to attain hemostasis, and despite the large
In spite of significant developments, the search for an ideal energy device that will result
in perfect hemostasis with minimum damage to surrounding tissue in the most efficient manner
posing the minimum threat to the patient in terms of short and long term complications remains
elusive. Each energy method has advantages and disadvantages and a thorough knowledge of
each devices is essential in deciding which energy source for be used for a specific procedure.
Relative advantages and disadvantages of existing energy methods are presented in Table 3.
In a 2004 paper, Harrell et al. [140] mention that in a survey of 500 surgeons in 1993,
18% (of over 500 respondents) reported to have caused electrosurgical injury during
laparoscopy and 54% reported to have known of another surgeon who has caused similar
injury. Deaths were more common during early in era of laparoscopy. A brief search of the
literature shows numerous instances of deaths in the 1980s using various energy sources. The
number of deaths and even the complications in laparoscopy, have greatly reduced in recent
around 2-4% in 1994 [141], came down to about 0.4% by 2005 [142]. Table 4 shows the various
In general, most of the studies suggest that the effect of any laparoscopic procedure
depends on the skill and familiarity of the surgeon with the surgical tools. An interesting
observation while reviewing studies performed to test various instruments or methods is that
some of the results seem contradicting. Devices which have been rated high by some studies
have been rated low in others. Surgical skill and familiarity with the particular device may
certainly be a possible contributor to this apparent paradox. Further, the devices which are
shown to work well in the laboratory may not work the same way in an actual laparoscopic
procedure. Hence, there is a need for developing uniform training regimens across surgical
specialties under clinical conditions. Particularly, the fundamental understanding of how each of
the energy devices work and their effect on the tissues is very important. For example, in
electrosurgery, the understanding of different power settings and its effects on tissues is very
important. More over the knowledge of safety issues with each of the devices should also be
known so that appropriate precautions could be used to minimize injury. Solid communication
and team coordination in high fire risk setting must be introduced into practice. Examples are
the high temperature and low cooling rate of ultrasonically activated devices even after switched
off, risk of air embolism in laser from high flow rate gas cooling and venous gas embolism while
using ABC. A standardized curriculum or manual with working principles of various energy
devices and their safety issues as envisioned by the FUSE program would be a valuable tool in
Each of the energy devices reviewed in this work had its own advantages and
greatest amount of thermal damage to adjacent tissue while ultrasonic energy results in the
least. In terms of their performance, ultrasonic devices provide the highest seal strength in
smaller vessels, while electrosurgery is more efficient for larger vessels. Argon beam
coagulation results in the most effective hemostasis on irregular surfaces, however, it also leads
used followed by microwave radiation. Lasers are very expensive and are mostly limited to
gynecological treatments in laparoscopy today, though at one time, they were widely used in
effectiveness of an energy device is dependent on the size of the blood vessel. Though there is
no clear winner in terms of operating time, in the series of study that were reviewed in this work,
the harmonic scalpel is shown to have reduced overall time compared to other energy sources
in MIS. In terms of death from complications, lasers and ABC have more reported cases than
the other methods and surgeons should be familiar in their use. Electrosurgery is still very
popular in MIS due to its familiarity with surgeons and the development of various enhanced
DISCLOSURE
1. Wu M-P, Ou C-S, Chen S-L, Yen EYT, Rowbotham R (2000) Complications and
recommended practices for electrosurgery in laparoscopy. The American Journal of Surgery
179(1):67-73.
2. Feldman L, Fuchshuber P, Jones D, Mischna J, Schwaitzberg S, Force tFT (2012)
Surgeons don't know what they don't know about the safe use of energy in surgery. Surgical
Endoscopy 26(10):2735-9. doi: 10.1007/s00464-012-2263-y.
3. Cushing H, Bovie WT (1928) Electrosurgery as an aid to the removal of intracranial
tumors. Surgery Gynecology And Obstetrics 47:751-84.
4. Albert M, Ellis CN, Fleshman J, Margolin D, Ng PC, Podnos YD. The Effective Use of
the ENSEAL System in Colorectal Surgery. 2010.
5. Lee J (2002) Update on Electrosurgery. Outpatient Surgery 3(2).
6. Govekar H, Robinson T, Varosy P, Girard G, Montero P, Dunn C, et al. (2012) Effect of
monopolar radiofrequency energy on pacemaker function. Surgical Endoscopy 26(10):2784-8.
doi: 10.1007/s00464-012-2279-3.
7. Montero P, Robinson T, Weaver J, Stiegmann G (2009) Insulation failure in laparoscopic
instruments. Surgical Endoscopy 24(2):462-5. doi: 10.1007/s00464-009-0601-5.
8. Vancaillie TG (1998) Active electrode monitoring. Surgical Endoscopy 12(8):1009-12.
doi: 10.1007/s004649900769.
9. Valleylab. (1999) Basics of Bipolar Electrosurgery. Clinical Information Hotline News.
10. Kim F, Chammas M, Gewehr E, Morihisa M, Caldas F, Hayacibara E, et al. (2008)
Temperature safety profile of laparoscopic devices: Harmonic ACE (ACE), Ligasure V (LV), and
plasma trisector (PT). Surgical Endoscopy 22(6):1464-9.
11. Box GN, Lee HJ, Abraham JB, Deane LA, Elchico ER, Abdelshehid CA, et al. (2009)
Comparative Study of In Vivo Lymphatic Sealing Capability of the Porcine Thoracic Duct Using
Laparoscopic Dissection Devices. The Journal of Urology 181(1):387-91.
12. Sutton PA, Awad S, Perkins AC, Lobo DN (2010) Comparison of lateral thermal spread
using monopolar and bipolar diathermy, the Harmonic Scalpel™ and the Ligasure™. British
Journal of Surgery 97(3):428-33.
13. Tulikangas PK, Smith T, Falcone T, Boparai N, Walters MD (2001) Gross and histologic
characteristics of laparoscopic injuries with four different energy sources. Fertility and Sterility
75(4):806-10.
14. Gol M, Kizilyar A, Eminoglu M (2007 ) Laparoscopic hysterectomy with retroperitoneal
uterine artery sealing using LigaSure: Gazi hospital experience. Arch Gynecol Obstet
276(4):311-4.
15. Kriplani A, Garg P, Sharma M, Lal S, Agarwal N (2008) A review of total laparoscopic
hysterectomy using LigaSure uterine artery-sealing device: AIIMS experience. J Laparoendosc
Adv Surg Tech A 18(6):825-9.
16. Polychronidis A, Tsaroucha AK, Karayiannakis AJ, Perente S, Efstathiou E, Simopoulos
C (2005) Delayed Perforation of the Large Bowel due to Thermal Injury during Laparoscopic
Cholecystectomy. The Journal of International Medical Research 33:360-3.
17. Siperstein A, Garland A, Engle K, Rogers S, Berber E, Foroutani A, et al. (2000) Local
recurrence after laparoscopic radiofrequency thermal ablation of hepatic tumors. Ann Surg
Oncol 7(2):106 - 13.
18. Darai E, Ackerman G, Bazot M, Rouzier R, Dubernard G (2007) Laparoscopic
segmental colorectal resection for endometriosis: limits and complications. Surgical Endoscopy
21(9):1572-7.
19. Berber E, Siperstein A (2008) Local recurrence after laparoscopic radiofrequency
ablation of liver tumors: an analysis of 1032 tumors. Ann Surg Oncol 15(10):2757-64.
20. Peterson HB, Ory HW, Greenspan JR, Jr. TCW (1981) Deaths associated with
laparoscopic sterilization by unipolar electrocoagulating devices, 1978 and 1979. Am J Obstet
Gynecol 139(2):141-3.
21. Willson PD, McAnena OJ, Peters EE (1994) A fatal complication of diathermy in
laparoscopic surgery. Minimally Invasive Therapy & Allied Technologies 3(1):19-20. doi:
doi:10.3109/13645709409152989.
22. Agarwal B, Gupta M, Agarwal S, Mahajan K (2007) Anatomical footprint for safe
laparoscopic cholecystectomy without using any energy source: a modified technique. Surgical
Endoscopy 21(12):2154-8.
23. Matthews B, Nalysnyk L, Estok R, Fahrbach K, Banel D, Linz H, et al. (2008) Ultrasonic
and Nonultrasonic Instrumentation: A Systematic Review and Meta-analysis. Arch Surg
143(6):592-600. doi: 10.1001/archsurg.143.6.592.
24. Hubner M, Demartines N, Muller S, Dindo D, Clavien PA, Hahnloser D (2008)
Prospective randomized study of monopolar scissors, bipolar vessel sealer and ultrasonic
shears in laparoscopic colorectal surgery. British Journal of Surgery 95(9):1098-104.
25. Targarona EM, Balague C, Marin J, Neto RB, Martinez C, Garriga J, et al. (2005) Energy
Sources for Laparoscopic Colectomy: A Prospective Randomized Comparison of Conventional
Electrosurgery, Bipolar Computer-Controlled Electrosurgery and Ultrasonic Dissection.
Operative Outcome and Costs Analysis. Surgical Innovation 12(4):339-44. doi:
10.1177/155335060501200409.
26. Hruby GW, Marruffo FC, Durak E, Collins SM, Pierorazio P, Humphrey PA, et al. (2007)
Evaluation of Surgical Energy Devices for Vessel Sealing and Peripheral Energy Spread in a
Porcine Model. The Journal of Urology 178(6):2689-93.
27. Edelman DS, Unger SW (1995) Bipolar Versus Monopolar Cautery Scissors for
Laparoscopic Cholecystectomy: A Randomized, Prospective Study. Surgical Laparoscopy
Endoscopy & Percutaneous Techniques 5(6):459-62.
28. Campagnacci R, de Sanctis A, Baldarelli M, Rimini M, Lezoche G, Guerrieri M (2007)
Electrothermal bipolar vessel sealing device vs. ultrasonic coagulating shears in laparoscopic
colectomies: a comparative study. Surgical Endoscopy 21(9):1526-31.
29. Demirturk F, Aytan H, Caliskan AC (2007) Comparison of the use of electrothermal
bipolar vessel sealer with harmonic scalpel in total laparoscopic hysterectomy. The Journal of
Obstetrics and Gynaecology Research 33:341-5.
30. Lamberton GR, Hsi RS, Jin DH, Lindler TU, Jellison FC, Baldwin DD (2008) Prospective
comparison of four laparoscopic vessel ligation devices. JOURNAL OF ENDOUROLOGY
22(10):2307–12.
31. Dodde R, Gee JS, Geiger JD, Shih AJ (2010) Monopolar Electrosurgical Thermal
Management System to Reduce Lateral Thermal Damage During Surgery. Journal of Medical
Devices 4(2):027505-1.
32. O'Daly BJ, Morris E, Gavin GP, O'Byrne JM, McGuinness GB (2008) High-power low-
frequency ultrasound: A review of tissue dissection and ablation in medicine and surgery.
Journal of Materials Processing Technology 200(1-3):38-58.
33. Smith R, Pasic R (2008) The role of vessel sealing technologies in laparoscopic surgery.
Surg Technol Int 17:208 - 12.
34. Lee SJ, Park KH (1999) Ultrasonic energy in endoscopic surgery. Yonsei Med J
40(6):545-9.
35. Liano C. Curbing laparoscopy burns. The Washington Times Web site. 2004.
36. Ghosh P (1999) Ultrasonic Devices in Cardiac Surgery. Asian Cardiovasc Thorac Ann
7(4):333-8.
37. Clements R, Palepu R (2007) In vivo comparison of the coagulation capability of
SonoSurg and Harmonic Ace on 4 mm and 5 mm arteries. Surgical Endoscopy 21(12):2203-6.
38. Harold KL, Pollinger H, Matthews BD, Kercher KW, Sing RF, Heniford BT (2003)
Comparison of ultrasonic energy, bipolar thermal energy, and vascular clips for the hemostasis
of small-, medium-, and large-sized arteries. Surgical Endoscopy 17(8):1228-30.
39. Wolfe BM, Gardiner BN, Leary BF, Frey CF (1991) Endoscopic Cholecystectomy: An
Analysis of Complications. Arch Surg 126(10):1192-8. doi:
10.1001/archsurg.1991.01410340030005.
40. Kadesky KM, Schopf B, Magee JF, Blair GK (1997) Proximity injury by the ultrasonically
activated scalpel during dissection. Journal of Pediatric Surgery 32(6):878-9.
41. Awwad JT, Isaacson K (1996) The harmonic scalpel: an intraoperative complication.
Obstet Gynecol 88(4):718-20.
42. McNally ML, Erturk E, Oleyourryk G, Schoeniger L (2001) Laparoscopic Cyst
Decortication Using the Harmonic Scalpel for Symptomatic Autosomal Dominant Polycystic
Kidney Disease. Journal Of Endourology 15(6):597-9.
43. Kim JS, Hattori R, Yamamoto T, Yoshino Y, Gotoh M (2010) How can we safely use
ultrasonic laparoscopic coagulating shears? International Journal of Urology 17:377-81.
44. Emam TA, Cuschieri A (2003) How Safe is High-Power Ultrasonic Dissection? Annals of
Surgery 237(2):186-91.
45. Schwartz RO (1994) Total Laparoscopic Total Hysterectomy with the Harmonic Scalpel.
Journal of Gynecologic Surgery 10(1):33-4.
46. Giannopoulos T, Chipchase J, Tailor A, Butler-Manuel S (2005) The use of harmonic
shears (Ultracision) for laparoscopic lymphadenectomies in women with gynaecological
malignancies. Gynecological Surgery 2(2):97-100.
47. Jansen FW, Trimbos-Kemper T, Baptist Trimbos J (2002) Ultrasonic scalpel in
laparoscopic gynaecological surgery: an observational study in 354 cases. Gynaecological
Endoscopy 11(1):47-51.
48. Lin J, Zhang X, Xu K (2001) Application of ultrasonic scalpel in gynecologic operative
laparoscopy. Chinese Medical Journal 114(12):1283-5.
49. Amarin NS (2008) Harmonic Scalpel and Clipless Cholecystectomy. World Journal of
Laparoscopic Surgery 1(2):6-8.
50. Gelmini R, Franzoni C, Zona S, Andreotti A, Saviano M (2010 ) Laparoscopic
cholecystectomy with Harmonic scalpel. JSLS, Journal of the Society of Laparoendoscopic
Surgeons 14(1):14-9.
51. Janssen IMC, Swank DJ, Boonstra O, Knipscheer BC, Klinkenbijl JHG, van Goor H
(2003) Randomized clinical trial of ultrasonic versus electrocautery dissection of the gallbladder
in laparoscopic cholecystectomy. British Journal of Surgery 90(7):799-803.
52. Elattar OM, Naga AAE, Maged H (2005) Laparoscopic appendectomy by ultrasonically
activated scalpel: a prospective study. Egyptian Journal of Surgery 24(3):164-7.
53. Lin PC, Thyer A, Soules MR (2004) Intraoperative ultrasound during a laparoscopic
myomectomy. Fertility and Sterility 81(6):1671-4.
54. Stringer NH (1994) Laparoscopic Myomectomy with the Harmonic Scalpel: A Review of
25 Cases. Journal of Gynecologic Surgery 10(4):241-5.
55. Msika S, Deroide G, Kianmanesh R, Iannelli A, Hay J-M, Fingerhut A, et al. (2001)
Harmonic scalpel in laparoscopic colorectal surgery. Diseases of the Colon & Rectum
44(3):432-6.
56. Vetere PF, Apostolis C (2010) Ureteral Injury Due to a Harmonic Scalpel During
Laparoscopic Salpingo-oophorectomy. JSLS, Journal of the Society of Laparoendoscopic
Surgeons 14(1):115-9.
57. Chachan S, Waters N, Kent A (2009) Laparoscopic management of cornual heterotopic
pregnancy with the use of Harmonic ACE®—a case report. Gynecological Surgery:1-4.
58. Tanaka T, Ueda K, Hayashi M, Hamano K (2009) Clinical application of an ultrasonic
scalpel to divide pulmonary vessels based on laboratory evidence. Interact CardioVasc Thorac
Surg 8(6):615-8. doi: 10.1510/icvts.2008.200584.
59. Driscoll P. Energy-Based Devices in General Surgery. MedMarket Diligence, LLC, 2006.
60. Bruhat M, Mage, C. & Manhes, M. ,(Year) editor Use of carbon dioxide laser via
laparoscopy. Laser Surgery III Proceedings of the Third Congress for the International Society
for Laser Surgery 1979: Kaplan; Published.
61. Sutton C (1986) Initial experience with carbon dioxide laser laparoscopy. Lasers in
Medical Science 1(1):25-31.
62. Hamman BL, Theologes TT,(Year) editors. Surgical treatment of atrial fibrillation with
diode-pumped laser. Baylor University Medical Center Proceedings; 2009 Published.
63. Williams MR, Casher JM, Russo MJ, Hong KN, Argenziano M, Oz MC (2006) Laser
Energy Source in Surgical Atrial Fibrillation Ablation: Preclinical Experience. The Annals of
Thoracic Surgery 82(6):2260-4.
64. Welch AJ, Torres JH, Cheong W-F (1989) Laser Physics and Laser-Tissue Interaction.
Texas Heart Institute Joumal 16(3):141–9.
65. Gerber GS (1995) Lasers in the treatment of benign prostatic hyperplasia. Urology
45(2):193-9.
66. Baggish MS, Daniell JF (1989) Death caused by air embolism associated with
neodymium:yttrium-aluminum-garnet laser surgery and artificial sapphire tips. American journal
of obstetrics and gynecology 161(4):877-8.
67. Challener RC, Kaufman B (1990) Fatal venous air embolism following sequential
unsheathed (bare) and sheathed quartz fiber Nd:YAG laser endometrial ablation.
Anesthesiology 73(3):548-51.
68. Greville AC, Clements EAF, Erwin DC, McMillan DL, Wellwood JM (1991) Pulmonary air
embolism during laparoscopic laser cholecystectomy. Anaesthesia 46:113-4.
69. Lang NP, Wait GM, Read RR (1991) Cardio-cerebrovascular complications from
Nd:YAG laser treatment of lung cancer. American journal of surgery 162(6):629-32.
70. Messiant F, Marquette C, Neviere R, Ramon P, Duverger D, Mathieu D (1995) Systemic
air embolism after laser resection of a tracheal tumor. Intensive Care Medicine 21(2):192-3.
71. Peachey T, Eason J, Moxham J, Jarvis D, Driver M (1988) Systemic air embolism during
laser bronchoscopy. Anaesthesia 43(10):872-5.
72. Ross DJ, Mohsenifar Z, Potkin RT, Roston WL, Shapiro SM, Alexander JM (1988)
Pathogenesis of cerebral air embolism during neodymium-YAG laser photoresection. Chest
94(3):660-2.
73. Yuan HB, Poon KS, Chan KH, Lee TY, Lin CY (1993) Fatal gas embolism as a
complication of Nd-YAG laser surgery during treatment of bilateral choanal stenosis.
International Journal of Pediatric Otorhinolaryngology 27(2):193-9.
74. Baggish MS, Daniell JF (1989) Catastrophic injury secondary to the use of coaxial gas-
cooled fibers and artificial sapphire tips for intrauterine surgery: A report of five cases. Lasers in
Surgery and Medicine 9(6):581-4.
75. Kelly M, Mathews HML, Weir P (1997) Carbon dioxide embolism during laser
endometrial ablation. Anaesthesia 52(1):65-7.
76. Schröder TM, Puolakkainen PA, Hahl J, Rämö OJ (1989) Fatal air embolism as a
complication of laser-induced hyperthermia. Lasers in Surgery and Medicine 9(2):183-5.
77. Tellides G, Ugurlu BS, Kim RW, Hammond GL (1998) Pathogenesis of Systemic Air
Embolism During Bronchoscopic Nd:YAG Laser Operations. Ann Thorac Surg 65(4):930-4.
78. Genyk YS, Keller FS, Halpern NB (1994) Hepatic artery pseudoaneurysm and hemobilia
following laser laparoscopic cholecystectomy. Surgical Endoscopy 8(3):201-4.
79. Ewen S, Sutton CJG (1995) Complications of laser laparoscopy: Eleven years
experience. Minimally Invasive Therapy & Allied Technologies 4(1):27-9. doi:
doi:10.3109/13645709509152749.
80. Reddick EJ, Olsen DO (1989) Laparoscopic laser cholecystectomy. Surgical Endoscopy
3(3):131-3.
81. Goldenberg M, Bider d, al. e (1994) Pregnancy: Laparosocopic laser surgery of primary
ovarian pregnancy Human Reproduction 9(7):1337-8
82. Langebrekke A, Kirschner R, Skår OJ, Sørnes T, Urnes A (1991) Treatment of tubal
pregnancy by laparoscopic laser surgery. Acta Obstetricia et Gynecologica Scandinavica 70(4-
5):331-4. doi: doi:10.3109/00016349109007882.
83. Langebrekke A, Sarnes T, Urnes A (1993) Fertility outcome after treatment of tubal
pregnancy by laparoscopic laser surgery. Acta Obstetricia et Gynecologica Scandinavica
72(7):547-9. doi: doi:10.3109/00016349309058161.
84. Sutton CJG, Ewen SP, Jacobs SA, Whitelaw NL (1997) Laser laparoscopic surgery in
the treatment of ovarian endometriomas. The Journal of the American Association of
Gynecologic Laparoscopists 4(3):319-23.
85. Weather LJ (1988) Carbon Dioxide Laser Laparoscopy in Treatment of Infertility and
Disorders Associated with Pelvic Pain. J Natl Med Assoc 80(2):185–7.
86. Shumalinsky D, Lobik L, Cytron S, Halpern M, Vasilyev T, Ravid A, et al. (2004)
Laparoscopic Laser Soldering for Repair of Ureteropelvic Junction Obstruction in the Porcine
Model. Journal of Endourology 18(2):177-81. doi: doi:10.1089/089277904322959833.
87. Karaman Y, Bingol B, Günenç Z (2006) Prevention of complications in laparoscopic
hysterectomy: Experience with 1120 cases performed by a single surgeon. Journal of Minimally
Invasive Gynecology 14(1):78-84.
88. Liu CY (1992) Laparoscopic hysterectomy. A review of 72 cases. J Reprod Med
37(4):351-4.
89. Wishnow KI, Johnson DE, Cromeens DM, Ro JY (1989) Laser photoirradiation of the
canine ureteral orifice: Comparison between contact and noncontact techniques. Lasers in
Surgery and Medicine 9(5):485-9.
90. Ward PH, Castro DJ, Ward S (1989) A Significant New Contribution to Radical Head and
Neck Surgery: The Argon Beam Coagulator as an Effective Means of Limiting Blood Loss. Arch
Otolaryngol Head Neck Surg 115(8):921-3. doi: 10.1001/archotol.1989.01860320031013.
91. Low RK, Moran ME (1993) Laparoscopic argon beam coagulation: Implications for
reconstructive urology. Minimally Invasive Therapy & Allied Technologies 2(2):75-8. doi:
doi:10.3109/13645709309152671.
92. Bowling DM (2002) Argon beam coagulation for post-tonsillectomy hemostasis.
Otolaryngology - Head and Neck Surgery 126(3):316-20.
93. Cornejo A, Liao L, Kenneth W (2010) Argon gas embolism with the use of argon beam
coagulation during open hepatic resection. The Internet Journal of Surgery 22(2).
94. Ikegami T, Shimada M, Imura S, Nakamura T, Kawahito S, Morine Y, et al. (2009) Argon
gas embolism in the application of laparoscopic microwave coagulation therapy. Journal of
Hepato-Biliary-Pancreatic Surgery 16(3):394-8.
95. Bateman SN, Noorily AD, McGuff HS (1996) Sharp Dissection, Electrosurgery, and
Argon-Enhanced Electrosurgery in Porcine Skin Flaps. Otolaryngology -- Head and Neck
Surgery 114(3):435-42.
96. Kono M, Yahagi N, Kitahara M, Fujiwara Y, Sha M, Ohmura A (2001) Cardiac arrest
associated with use of an argon beam coagulator during laparoscopic cholecystectomy.
European Journal of Cardio-thoracic Surgery 87(4):644-6. doi: 10.1093/bja/87.4.644.
97. Sezeur A, Partensky C, Chipponi J, Duron J-J (2008) Death During Laparoscopy: Can 1
Gas Push Out Another? Danger of Argon Electrocoagulation. Surg Laparosc Endosc Percutan
Tech 18(4):395-7
98. (1994) Fatal gas embolism caused by overpressurization during laparoscopic use of
argon enhanced coagulation. Health Devices 23(6):257 - 9.
99. Ousmane ML, Fleyfel M, Vallet B (2002) Venous gas embolism during liver surgery with
argon-enhanced coagulation. European Journal of Anaesthesiology 19(3):225-.
100. Kizer N, Zighelboim I, Rader JS (2009) Cardiac Arrest During Laparotomy With Argon
Beam Coagulation of Metastatic Ovarian Cancer. International Journal of Gynecological Cancer
19(2):237-8 10.1111/IGC.0b013e31819c54db.
101. Mastragelopulos N, Sarkar, M.R., Kaissling, G., Bähr, R., Daub, D. (1992) Argon gas
embolism in laparoscopic cholecystectomy with the Argon Beam One coagulator. Chirurg
63(12): 1053-4.
102. Reddy C, Majid A, Michaud G, Feller-Kopman D, Eberhardt R, Herth F, et al. (2008) Gas
Embolism Following Bronchoscopic Argon Plasma Coagulation. Chest 134(5):1066-9. doi:
10.1378/chest.08-0474.
103. Stojeba N, Mahoudeau G, Segura P, Meyer C, Steib A (1999) Possible venous argon
gas embolism complicating argon gas enhanced coagulation during liver surgery. Acta
Anaesthesiologica Scandinavica 43(8):866-7.
104. Veyckemans F, Michel I (1996) Venous Gas Embolism from an Argon Coagulator.
Anesthesiology 85(2):443-4.
105. Kwon AH, Inui H, Kamiyama Y (2001) Successful laparoscopic haemostasis using an
argon beam coagulator for blunt traumatic splenic injury. Eur J Surg 167(4):316-8.
106. Lurie S, Gur D, Glezerman M (2002) Argon beam coagulation during caesarean section.
BJOG: An International Journal of Obstetrics & Gynaecology 109(12):1415-.
107. Lucioni A, Orvieto MA, Zorn KC, Lotan T, Gong EM, Steinberg GD, et al. (2008) Efficacy
of the argon beam coagulator alone in obtaining hemostasis after laparoscopic porcine
heminephrectomy: a pilot study. Can J Urol 15(3):4091-6.
108. Meyer-Junghaenel L, Giest H, Waldschmidt Jr (2001) Laparoscopic Argon-Beam
Coagulation in Relapsing Congenital Dysontogenic Splenic Cysts. Pediatric Endosurgery &
Innovative Techniques 5(3):315-8. doi: doi:10.1089/10926410152634466.
109. Okada T, Sasaki F, Kurauchi N, Kubota M, Itoh T, Honda S, et al. (2007) Laparoscopic
liver biopsy using cup-shaped punch biopsy forceps and argon beam coagulator in children.
Pediatric Surgery International 23(10):947-51.
110. Dowling RD, Ochoa J, Yousem SA, Peitzman A, Udekwu AO (1991) Argon Beam
Coagulation is Superior to Conventional Techniques in Repair of Experimental Splenic Injury
The Journal of Trauma 31(5):717-21.
111. Palmer M, Miller CW, III CWVW, Orton EC (1993) Venous Gas Embolism Associated
with Argon-Enhanced Coagulation of the Liver. Journal of Investigative Surgery 6(5):391-9.
112. Falbo B, Vene M, Visini R, Grottola T, Di Paolo S, Picardi N (2004) [Experimental study
of the effect of the Argon Beam Coagulator on organic tissues from the viewpoint of surgical
utilization]. Annali Italiani Di Chirurgia 75(1):59-62.
113. Bobbio A, Ampollini L, Internullo E, Caporale D, Cattelani L, Bettati S, et al. (2006)
Thoracoscopic parietal pleural argon beam coagulation versus pleural abrasion in the treatment
of primary spontaneous pneumothorax. European Journal of Cardio-thoracic Surgery 29(1):6-8.
doi: 10.1016/j.ejcts.2005.10.034.
114. Beyer E, Lee R, Lam B-K. Point: Minimally invasive bipolar radiofrequency ablation of
lone atrial fibrillation: Early multicenter results. 2009. p. 521-6.
115. Currò G, Iapichino G, Melita G, Lorenzini C, Cucinotta E (2005) Laparoscopic
Cholecystectomy in Child-Pugh Class C Cirrhotic Patients JSLS, Journal of the Society of
Laparoendoscopic Surgeons 9(3):311-5.
116. Saad EB, Marrouche NF, Saad CP, Ha E, Bash D, R.D. White, et al. (2003)
Radiofrequency Treatment of Abnormal Heart Rhythm Can Damage the Vessels That Return
Blood from the Lungs to the Heart. Annals of Internal Medicine
117. Keltner JR, Donegan E, Hynson JM, Shapiro WA (2001) Acute Renal Failure After
Radiofrequency Liver Ablation of Metastatic Carcinoid Tumor Anesthesia & Analgesia 93:587-9.
118. Kocovic DZ, Harada T, Shea JB, Soroff D, Friedman PL (1993) Alterations of heart rate
and of heart rate variability after radiofrequency catheter ablation of supraventricular
tachycardia. Delineation of parasympathetic pathways in the human heart. Circulation
88(4):1671-81.
119. Curley SA, Marra P, Beaty K, al. E (2004 ) Early and Late Complications After
Radiofrequency Ablation of Malignant Liver Tumors in 608 Patients. Ann Surg 239(4):450–8.
120. Koffron AJ, Stein JA (2008 ) Laparoscopic liver surgery: parenchymal transection using
saline-enhanced electrosurgery. HPB (Oxford) 10(4):225–8.
121. Yim APC, Rendina EA, Hazelrigg SR, Chow LTC, Lee T-W, Wan S, et al. (2002) A new
technological approach to nonanatomical pulmonary resection: saline enhanced thermal
sealing. The Annals of Thoracic Surgery 74(5):1671-6.
122. Gozen AS, Teber D, Rassweiler JJ (2007) Principles and initial experience of a new
device for dissection and hemostasis. Minim Invasive Ther Allied Technol 16(1):58-65. doi:
doi:10.1080/13645700701191537.
123. Zarebczan B, Mohanty D, Chen H (2010) A Comparison of the LigaSure and Harmonic
Scalpel in Thyroid Surgery: A Single Institution Review. Annals of Surgical Oncology:1-5.
124. Munver R, Del Pizzo J, Sosa R (2003) Adrenal-preserving minimally invasive surgery:
The role of laparoscopic partial adrenalectomy, cryosurgery, and radiofrequency ablation of the
adrenal gland. Current Urology Reports 4(1):87-92.
125. Smulders J, de Hingh I, Stavast J, Jackimowicz J (2007) Exploring new technologies to
facilitate laparoscopic surgery: creating intestinal anastomoses without sutures or staples, using
a radio-frequency-energy-driven bipolar fusion device. Surgical Endoscopy 21(11):2105-9.
126. Soon PSH, Yeh MW, Sywak MS, Sidhu SB (2006) Use of the Ligasure vessel sealing
system in laparoscopic adrenalectomy. ANZ Journal of Surgery 76(9):850-2.
127. Hope WW, Burns JM, Newcomb WL, Heniford BT, Sing RF (2009) Safety and efficacy of
the electrothermal bipolar vessel sealer in trauma. JOURNAL OF ENDOUROLOGY 40(5):564-
6.
128. Kimura M, Baba S, Polascik TJ (2010) Minimally invasive surgery using ablative
modalities for the localized renal mass. International Journal of Urology 17(3):215-27.
129. Oefelein MG (2006) Delayed Presentation of Urinoma after Radiofrequency Ablation-
Assisted Laparoscopic Partial Nephrectomy. Journal of Endourology 20(1):27-30. doi:
doi:10.1089/end.2006.20.27.
130. Onishi T, Nishikawa K, Hasegawa Y, al. e (2007) Assessment of Health-related Quality
of Life after Radiofrequency Ablation or Laparoscopic Surgery for Small Renal Cell Carcinoma:
a Prospective Study with Medical Outcomes Study 36-Item Health Survey (SF-36) Jpn J Clin
Oncol 37(10):750-4.
131. Bachellier P, Ayav A, Pai M, Jean-Christopher W, Edoardo R, Daniel J, et al. (2007)
Laparoscopic liver resection assisted with radiofrequency. American journal of surgery
193(4):427-30.
132. Hompes D, Aerts R, Penninckx F, Topal B (2007) Laparoscopic liver resection using
radiofrequency coagulation. Surgical Endoscopy 21(2):175-80.
133. Tan BJ, El-Hakim A, Nora M, Yavor S, Arthur S, Benjamin R LEE (2004) Comparison of
laparoscopic saline infused to dry radio frequency ablation of renal tissue: evolution of
histological infarct in the porcine model. The Journal of urology 172(5):2007-12.
134. Asahina YN, Hiroyuki; Izumi, Namiki (2009) Laparoscopic radiofrequency ablation for
hepatocellular caecinoma. Digestive Endoscopy 21(2):67-72.
135. Sadiq TS, Shrestha R, Weeks S, Gerber DA (2004) Laparoscopic radiofrequency
ablation provides local control of hepatocellular carcinoma in patients awaiting liver transplant.
Journal of the American College of Surgeons 199(3):16.
136. van den Bos R, Lidia A, Michael K, Martino N, Tamar N (2009) Endovenous therapies of
lower extremity varicosities: A meta-analysis. Journal of vascular surgery : official publication,
the Society for Vascular Surgery [and] International Society for Cardiovascular Surgery, North
American Chapter 49(1):230-9.
137. Weil L, Glover JP, Weil LS. A New Minimally Invasive Technique for Treating Plantar
Fasciosis Using Bipolar Radiofrequency: A Prospective Analysis. 2008. p. 13-8.
138. Hüscher CGS, Lirici MM, Di Paola M, Crafa F, Napolitano C, Mereu A, et al. (2003)
Laparoscopic cholecystectomy by ultrasonic dissection without cystic duct and artery ligature.
Surgical Endoscopy 17(3):442-51.
139. Redwan AA (2010) Single-Working-Instrument, Double-Trocar, Clipless
Cholecystectomy Using Harmonic Scalpel: A Feasible, Safe, and Less Invasive Technique.
Journal of Laparoendoscopic & Advanced Surgical Techniques 20(7):597-603. doi:
doi:10.1089/lap.2009.0375.
140. Harrell AG, Kercher KW, Heniford BT (2004) Energy Sources in Laparoscopy. Surgical
Innovation 11(3):201-9. doi: 10.1177/107155170401100310.
141. Berry SM, Ose KJ, Bell RH, Fink AS (1994) Thermal injury of the posterior duodenum
during laparoscopic cholecystectomy. Surgical Endoscopy 8(3):197-200.
142. Giger U, Michel JM, Vonlanthen R, Becker K, Kocher T, Krähenbühl L (2005)
Laparoscopic cholecystectomy in acute cholecystitis: indication, technique, risk and outcome.
Langenbeck's Archives of Surgery 390(5):373-80.
143. Peters JH, Ellison EC, Innes JT, Liss JL, Nichols KL, Lomano JM, et al. (1991) Safety
and efficacy of laparoscopic cholecystectomy. A prospective analysis of 100 initial patients.
Annals of Surgery 213(1):3-12.
144. Kandil T, El Nakeeb A, El Hefnawy E (2010) Comparative Study between Clipless
Laparoscopic Cholecystectomy by Harmonic Scalpel Versus Conventional Method: A
Prospective Randomized Study. Journal of Gastrointestinal Surgery 14(2):323-8.
145. Tsimoyiannis EC, Jabarin M, Glantzounis G, Lekkas ET, Siakas P, Stefanaki-Nikou S
(1998) Laparoscopic Cholecystectomy Using Ultrasonically Activated Coagulating Shears.
Surgical Laparoscopy Endoscopy & Percutaneous Techniques 8(6):421-4.
146. Lane GE, Lathrop JC (1993) Comparison of results of KTP/532 laser versus monopolar
electrosurgical dissection in laparoscopic cholecystectomy. J Laparoendosc Surg 3(3):209 -14.
147. Bordelon BM, Hobday KA, Hunter JG (1993 ) Laser vs electrosurgery in laparoscopic
cholecystectomy. A prospective randomized trial. Arch Surg 128(2):233-6.
148. El Nakeeb A, Askar W, El Lithy R, Farid M (2010) Clipless laparoscopic cholecystectomy
using the Harmonic scalpel for cirrhotic patients: a prospective randomized study. Surgical
Endoscopy 24(10):2536-41.
149. Amaral JF (1995) Laparoscopic Cholecystectomy in 200 Consecutive Patients using an
Ultrasonically Activated Scalpel. Surgical Laparoscopy Endoscopy & Percutaneous Techniques
5(4):255-62.
150. (1991) A Prospective Analysis of 1518 Laparoscopic Cholecystectomies. New England
Journal of Medicine 324(16):1073-8 %U
http://www.nejm.org/doi/full/10.56/NEJM199104183241601.
151. Kurauchi N, Kamii N, Kazui K, Saji Y, Uchino J (1998) Laparoscopic cholecystectomy: A
report on the community hospital experience in Hokkaido. Surgery Today 28(7):714-8. doi:
10.1007/bf02484617.
152. Hideki N, Masato N, Norihiro Y, AL. E (2003) Electrosurgical devices for abdominal
surgery: Comparison between monopolar devices and bipolar scissors. Surgical Therapy
89(1):8-13.
153. Mantke R, Halangk W, Habermann A, Peters B, Konrad S, Guenther M, et al. (2010)
Efficacy and safety of 5-mm-diameter bipolar and ultrasonic shears for cutting carotid arteries of
the hybrid pig. Surgical Endoscopy:1-9.
154. Richter S, Kollmar O, Schilling M, Pistorius G, Menger M (2006) Efficacy and quality of
vessel sealing. Surgical Endoscopy 20(6):890-4.
155. Carbonell AM, Joels CS, Kercher KW, Matthews BD, Sing RF, Heniford BT (2003) A
Comparison of Laparoscopic Bipolar Vessel Sealing Devices in the Hemostasis of Small-,
Medium-, and Large-Sized Arteries. Journal of Laparoendoscopic & Advanced Surgical
Techniques 13(6):377-80.
156. Goldstein SL, Harold KL, Lentzner A, Matthews BD, Kercher KW, Sing RF, et al. (2002)
Comparison of Thermal Spread after Ureteral Ligation with the Laparo-Sonicâ„¢ Ultrasonic
Shears and the Ligasureâ„¢ System. Journal of Laparoendoscopic & Advanced Surgical
Techniques 12(1):61-3. doi: doi:10.1089/109264202753486957.
157. Phillips CK, Hruby GW, Durak E, Lehman DS, Humphrey PA, Mansukhani MM, et al.
(2008) Tissue Response to Surgical Energy Devices. Urology 71(4):744-8.
158. Wetter PA, Kavic MS, al. e, editors. Prevention & Management of Laparoendoscopic
Surgical Complications.: Society of Laparoendoscopic Surgeons; 1999.
Table 1. Most commonly used energy devices in minimally invasive surgery
Group 2:
Bipolar
Electrosur
gery
Scissors
(BEC) (40
patients)
Janssen et 200 Group 1: Group 1: Group 1: Group 1:
al. Harmonic 60.0 (median) 1.9 ±0.5 15 (16 %)
[51] Scalpel days
1998-2000 (Ultracisio Group 2: Group 2:
n, UC) 65.0 (median) Group 2: 51 (50 %)
1.4±0.2
(96 days
patients)
Group2:
Electrosur
gery(ES)
(103
patients)
Lane et al. 641 Group1: Group1: Group 1: Group 2: G
[146] 69.8 1.4 ±0.06 4 (0.9 %) roup
1991 (KTP/532) days 1:
laser Group 2: 2
111.8 Group 2: (1%)
(199 1.7
patients) ±0.08 G
Group2: days roup
2:
Monopolar 1
Electrosur 0 (2.3
gery %)
(MES)
(44
2 patients)
Bordelon 103 Group1: Group 1: Group 1: G
et al. 23.6±9.56 15 roup
[147] Nd:YAG 28.84%) 1:
1990-1991 laser Group 2: 2
19.2±8.8 Group 2: (1.92
(52 8 (15.68 %) %)
patients)
Group2: G
roup
Monopolar 2:
Electrosur 1
gery
(MES) (1.96)
(51
patients)
(60
patients)
Electosurg
ery (1077
patients)
Monopolar Electrosurgery
Advantages Disadvantages
Bipolar Electrosurgery
Advantages Disadvantages
1. Passage of current through only the
portion of the tissue that is operated.
1. Operational time is usually longer than
2. Smaller thermal spread ( For example,-
monopolar electrosurgery and not as
EnSeal produced a thermal spread of
effective on small blood vessels.
only 1.84 mm in medium sized (4-5 mm)
2. The amount of thermal spread usually
porcine arteries [153].
depends on the power setting and the
3. Good for coaptive vessel sealing
skill level of the user - Ligasure
4. Bipolar devices produce equal peak
produced thermal spread in the range of
temperatures on different types of
0.6 – 6 mm in porcine organs [26, 38,
tissues with various thickness [10].
154-157]
5. Bipolar devices can be available in
3. Production of smoke.
many forms – scissors, forceps, grips
etc.
Ultrasonic energy
Advantages Disadvantages
1. Produces high blade temperatures and
can damage adjacent tissues or organs
when come in contact immediately after
1. Ultrasonic energy usually produces the swictched OFF [10].
least amount of thermal spread [42, 2. Temperature produced is inversely
165]. proportional to the tissue thickness [10].
2. No smoke and only mist due to 3. Not all devices are the same -the LCS
cavitation effect. device was shown to have caused high
3. Best energy method for sealing small thermal spread – 8.5 mm in porcine
blood vessels (upto 2 mm in diameter) veins [26].
[27, 42, 166] . 4. The high power level settings of
4. Ultrasonic devices (ACE, Wave, ultrasonic devices can cause significant
CS14C) produced the best quality seals thermal spread (upto 25.7 mm) and
at lower power levels [153]. peak temperatures (upto 140°C in
5. Lesser operating time. porcine organs) [44].
5. The ultrasonic devices do not produce
effective sealing for blood vessels over
2mm in diameter [26, 38].
Laser energy
Advantages Disadvantages
1. Can be effective when right laser type
1. Very expensive equipment.
and configuration is chosen.
2. Risk of OR fire.
2. Most widely used in gynecological
3. Increased operating time in general.
procedure because of precise control of
4. Air embolism which can be fatal.
amount and depth of tissues to be
5. Either the precision or efficiency of the
affected,
laser has to be compromised – one of
3. Less scarring compared to other modes
them is usually sacrificed [158].
of energy.