GOM-804_GOM-805-m

Download as pdf or txt
Download as pdf or txt
You are on page 1of 158

D.C.

Milli-Ohm Meter
GOM-804 & GOM-805

USER MANUAL
GW INSTEK PART NO. 82OM-80500E01

ISO-9001 CERTIFIED MANUFACTURER


This manual contains proprietary information, which is protected by
copyright. All rights are reserved. No part of this manual may be
photocopied, reproduced or translated to another language without
prior written consent of the Good Will company.
The information in this manual was correct at the time of printing.
However, Good Will continues to improve products and reserves
the right to change specifications, equipment, and maintenance
procedures at any time without notice.

Good Will Instrument Co., Ltd.


No. 7-1, Jhongsing Rd., Tucheng Dist., New Taipei City 236, Taiwan.
Table of Contents

Table of Contents
SAFETY INSTRUCTIONS .................................................... 5
Safety Symbols .................................................... 5
Safety Guidelines ................................................ 6

GETTING STARTED ............................................................ 9


GOM-804/805 Characteristics........................... 10
Key Features ...................................................... 13
Model Lineup .................................................... 14
Front Panel Overview ........................................ 15
TFT-LCD Overview ............................................ 19
Rear Panel Overview ......................................... 21
Set Up ............................................................... 23

MEASUREMENT ............................................................... 27
Resistance Measurement .................................. 29
Compare Function ............................................. 41
Binning Function ............................................... 46
Temperature Measurement ............................... 50
Temperature Compensation .............................. 52
Temperature Conversion ................................... 56
Measurement Settings ...................................... 60
System Settings ................................................ 69

HANDLER/SCAN INTERFACE .......................................... 77


Handler Overview ............................................. 78
Pin Definitions for the Handler Interface .......... 80
Scan Overview ................................................... 82
Configure Interface ........................................... 90

SAVE/RECALL ................................................................... 99

COMMAND OVERVIEW ................................................. 102


Command Syntax ............................................ 102
Command List ................................................. 105

3
GOM-804 & GOM-805 User Manual

BINNing Commands ....................................... 108


Calculate Commands ...................................... 113
Memory Commands ........................................ 120
Sense Commands ............................................ 122
Source Commands .......................................... 126
Status Commands ........................................... 127
System Commands.......................................... 128
Temperature Commands ................................. 133
Trigger Commands .......................................... 138
Userdefine Commands .................................... 141
IEEE 488.2 Common Commands ..................... 143
Status system .................................................. 146

FAQ ................................................................................ 147

APPENDIX ...................................................................... 148


Temperature Measurement ............................. 149
Specifications .................................................. 152
Dimensions ..................................................... 155
Declaration of Conformity ............................... 156

INDEX ............................................................................ 157

4
SAFETY INSTRUCTIONS

SAFETY INSTRUCTIONS
This chapter contains important safety instructions that you must follow
when operating the GOM-804/805 or when keeping it in storage. Read the
following before any operation to insure your safety and to keep the
GOM-804/805 in the best possible condition.

Safety Symbols
These safety symbols may appear in this manual or on the GOM-804/805.

Warning: Identifies conditions or practices that could


WARNING result in injury or loss of life.
Caution: Identifies conditions or practices that could
CAUTION result in damage to the instrument or to other properties.

DANGER High Voltage

Attention Refer to the Manual

Protective Conductor Terminal

Earth (ground) Terminal

Do not dispose electronic equipment as unsorted


municipal waste. Please use a separate collection facility
or contact the supplier from which this instrument was
purchased.

5
GOM-804 & GOM-805 User Manual

Safety Guidelines
General Guideline  Do not place any heavy objects on the instrument.
 Avoid severe impact or rough handling that leads to
CAUTION damaging the instrument.
 Do not discharge static electricity to the instrument.
 Use only mating connectors, not bare wires, for the
terminals.
 Do not disassemble the instrument unless you are
qualified as service personnel.
(Note) EN 61010-1:2010 specifies the measurement categories and
their requirements as follows. The GOM-804/805 doesn’t fall
under category II, III or IV.
 Measurement category IV is for measurements performed at the
source of low-voltage installation.
 Measurement category III is for measurements performed in the
building installation.
 Measurement category II is for measurements performed on the
circuits directly connected to the low voltage installation.
Power Supply  AC Input voltage: 100 - 240 V AC, 50 - 60Hz, 25VA
 The power supply voltage should not fluctuate more
WARNING than 10%.
 Connect the protective grounding conductor of the AC
power cord to an earth ground, to avoid electrical
shock.
Cleaning the  Disconnect the power cord before cleaning.
GOM-804/805  Use a soft cloth dampened in a solution of mild
detergent and water. Do not spray any liquid into the
instrument.
 Do not use chemicals or cleaners containing harsh
material such as benzene, toluene, xylene, and acetone.
Operation  Location: Indoor, no direct sunlight, dust free, almost
Environment non-conductive pollution (Note below)
 Relative Humidity: < 80%
 Altitude: < 2000m
 Temperature: 0°C to 40°C (operation)

6
SAFETY INSTRUCTIONS

(Note) EN 61010-1:2010 specifies the pollution degrees and their


requirements as follows. The GOM-804/805 falls under degree 2.
Pollution refers to “addition of foreign matter, solid, liquid, or
gaseous (ionized gases), that may produce a reduction of dielectric
strength or surface resistivity”.
 Pollution degree 1: No pollution or only dry, non-conductive pollution
occurs. The pollution has no influence.
 Pollution degree 2: Normally only non-conductive pollution occurs.
Occasionally, however, a temporary conductivity caused by
condensation must be expected.
 Pollution degree 3: Conductive pollution occurs, or dry,
non-conductive pollution occurs which becomes conductive due to
condensation which is expected. In such conditions, equipment is
normally protected against exposure to direct sunlight, precipitation,
and full wind pressure, but neither temperature nor humidity is
controlled.
Storage  Location: Indoor
Environment  Temperature: −10°C to 70°C
Disposal Do not dispose this instrument as unsorted municipal
waste. Please use a separate collection facility or contact
the supplier from which this instrument was purchased.
Please make sure discarded electrical waste is properly
recycled to reduce environmental impact.

7
GOM-804 & GOM-805 User Manual

Power cord for the United Kingdom


When using the instrument in the United Kingdom, make sure the power
cord meets the following safety instructions.

NOTE: This lead / appliance must only be wired by competent persons

WARNING: THIS APPLIANCE MUST BE EARTHED


IMPORTANT: The wires in this lead are coloured in accordance with the
following code:
Green/ Yellow: Earth
Blue: Neutral
Brown: Live (Phase)
As the colours of the wires in main leads may not correspond with the
coloured marking identified in your plug/appliance, proceed as follows:
The wire which is coloured Green & Yellow must be connected to the Earth
terminal marked with either the letter E, the earth symbol or coloured
Green/Green & Yellow.
The wire which is coloured Blue must be connected to the terminal which is
marked with the letter N or coloured Blue or Black.
The wire which is coloured Brown must be connected to the terminal marked
with the letter L or P or coloured Brown or Red.
If in doubt, consult the instructions provided with the equipment or contact
the supplier.
This cable/appliance should be protected by a suitably rated and approved
HBC mains fuse: refer to the rating information on the equipment and/or
user instructions for details. As a guide, a cable of 0.75mm2 should be
protected by a 3A or 5A fuse. Larger conductors would normally require 13A
types, depending on the connection method used.
Any exposed wiring from a cable, plug or connection that is engaged in a live
socket is extremely hazardous. If a cable or plug is deemed hazardous, turn
off the mains power and remove the cable, any fuses and fuse assemblies. All
hazardous wiring must be immediately destroyed and replaced in accordance
to the above standard.

8
GETTING STARTED

GETTING STARTED
This chapter describes the GOM-804/805 in a nutshell, including its main
features as well as its front and rear panels. After going through the panel
overview, follow the Power-up sequence before attempting to use the
instrument.
Please note the information in this manual was correct at the time of printing.
However as GW Instek continues to improve its products, changes can occur
at any time without notice. Please see the GW Instek website for the latest
information and content.

Characteristics GOM-804/805 Characteristics ................................ 10


Key Features ........................................................... 13
Model Lineup ......................................................... 14

Panel Overview Front Panel Overview ............................................. 15


TFT-LCD Overview ................................................. 19
Rear Panel Overview ............................................... 21

Setup Tilt Stand ................................................................ 23


Power Up ................................................................ 24
4 Wire Kelvin Connection ....................................... 25
Zeroing (Relative Function) .................................... 26

9
GOM-804 & GOM-805 User Manual

GOM-804/805 Characteristics
GOM-804 and GOM-805 are modern high precision programmable DC
Milli-ohm meters suitable for low resistance measurements of switches, relays,
connectors, PCB tracks and a variety of other devices. The meters feature a
color TFT-LCD screen with easy-to-read measurement results. With the
easy-to-use features, superior performance and automatic test interfaces, these
meters are dependable instruments for resistance measurements.

Easy to Use Each test function on the GOM-804/805 can be easily


Features activated by pressing a single front panel key. All the
settings and measurement results are displayed and set on
the TFT-LCD panel at the same time making each
function naturally intuitive to use.
Each primary and secondary measurement result is
displayed prominently on the display along with any
corresponding settings. For sequential measurement
results, such as those from the scan or binning function,
are tabulated in an intuitive and easy-to-read format.
In addition, the meters can recall previously used settings
upon startup, allowing the meter to be ready the next
time it used in a matter of moments. The meters can also
save or recall up to 20 sets of function settings.

Performance The GOM-804/805 has nine selectable measurement


ranges from 50mΩ to 5MΩ, a constant current source of
1uA to 1A, an accuracy of up to 0.05%, a 1uΩ resolution
and performs measurements using four wire Kelvin
connections for accurate, consistent measurements.
The ability to choose between high accuracy
measurements at 10 samples/sec (full scale at 50000
counts) or high speed measurements at 60 samples/sec
(full scale at 50000 counts), allows the GOM-804/805 the
flexibility to fulfill a number of different measurement
roles.

10
GETTING STARTED

Advanced The GOM-804/805 has a number of advanced


Temperature temperature functions that can be used with the optional
Measurements temperature probe, PT-100.
The temperature compensation function can extrapolate
what the resistance of a DUT will be at a desired
temperature, if the temperature coefficient of the DUT
and the resistance of the DUT at ambient temperature
are known.
The temperature conversion function can be used to
extrapolate what the temperature rise of a DUT will be at
specified resistance if the initial resistance, initial
temperature and the constant for the DUT are known.

Drive Signals The GOM-805 can select a number of different drive


signals to suit a number of different measurement
scenarios, for example the Pulse setting can be used to
cancel the effects of thermoelectric EMF on the
measurement results.

Dry Circuit Dry circuit testing allows the GOM-805 to measure the
Testing contact resistance of switches and connectors according
to the DIN IEC 512 and ASTM B539 standards. The
open circuit voltage will not exceed 20mV in this mode
to prevent the oxidization layer on metal switches and
connector points from breakdown. GOM-805 only.

Automatic For automatic testing The GOM-804/805 has a handler


Testing interface designed for automatic testing. The handler
interface outputs the status of PASS, FAIL, HI, LO,
READY and EOT signals and inputs a trigger control
signal. Automatic testing is used with the binning,
compare and scan functions.
For computer control applications, RS-232 and USB are
standard remote interfaces, with GPIB as standard only
for the GOM-805 and GOM-804G.

11
GOM-804 & GOM-805 User Manual

Applications  Production testing for contact resistance of switches,


relays, connectors, cables and printed circuit boards
and other low resistance devices.
 Component testing of resistors, motors, fuses and
heating elements.
 Incoming inspection and quality assurance testing.
 Conductivity evaluation for product design.

12
GETTING STARTED

Key Features
 50,000 counts
 Measurement Range: 50mΩ~5MΩ
 Accuracy of up to 0.05%
 Compare function
 Binning function
 Manual or Auto-ranging
 Continuous or Triggered measurement modes
 Temperature measurement, temperature
compensation and temperature conversion
 Four-wire Kelvin measurement method
 Selectable power-on settings
 Diode test
 Alarm settings for function-specific PASS/FAIL test
results
 Sampling rate: 10 or 60 sampling/sec
 Standard interfaces:
USB/RS232/Scan/Handler/GPIB(GOM-805,
GOM-804G)
 Save/Recall settings: 20 memory sets
 External I/O logic function

13
GOM-804 & GOM-805 User Manual

Model Lineup
Feature / Model GOM-804 GOM-804G* GOM-805
Ohm Measurement ✔ ✔ ✔
Compare Function ✔ ✔ ✔
Diode Measurement ✔ ✔ ✔
Temp. Compensation ✔ ✔ ✔
Temp. Conversion ✔ ✔ ✔
Temp Measurement ✔ ✔ ✔
Dry Circuit ✘ ✘ ✔
Drive Selection ✘ ✘ ✔
Binning Function ✘ ✘ ✔
GPIB Interface ✘ ✔ ✔

* The GOM-804G is simply the GOM-804 with the factory-installed


GPIB option. Please note that the GPIB option cannot be
user-installed on the GOM-804. The option must be ordered prior to
purchase.

14
GETTING STARTED

Front Panel Overview


Arrow keys,
LCD display Function key Enter keys ESC key
GOM-805 DC Milliohm Meter
Ohm Compare Binning
GND
GND
TC TCONV TEMP Enter
terminal
Speed REL RT

Scan Dry Trigger ESC


GUARD
GUARD

Display Local Diode Drive Range terminal


SOURCE SENSE SENSE SOURCE POWER

Rx
GOM-805 DC Milliohm Meter
GND
Ohm Compare Binning

TC TCONV TEMP Enter

Speed REL RT

Scan Dry
Sense-, Source-
Trigger ESC

GUARD
Sense+, Source+ Power key
Display Local Diode Drive Range

Turns On or Off the main


POWER

Power Switch
SOURCE SENSE SENSE SOURCE

Rx
power. For details about the power up
sequence, see page 24.

Measurement Terminals

Source, Sense SOURCE SENSE SENSE SOURCE

Terminals
Rx

Sense + and Sense - terminals.

Current source terminals: Source + and


Source -.

When measuring components


CAUTION with polarity, connect Source+
to the positive potential and
connect Source- to the negative
potential of the component.

Discharge any DUT before


WARNING measurement to avoid
damaging the GOM-804/805.

15
GOM-804 & GOM-805 User Manual

GND Terminal GND Connect the GND (ground) terminal to


the earth ground.

GUARD Terminal The GUARD terminal has the same


potential as earth, but cannot be
substituted for it. Connect the GUARD
terminal to the cable shield layer of the
GUARD test leads to help reduce noise.
Function Keys Ohm The Ohm key activates the resistance
measurement function.

Compare The Compare key activates the


comparator function.

Binning The Binning key activates the binning


function to grade the DUTs into eight
bins according to the tolerance settings.
GOM-805 only.

TC The TC key activates the TC


(temperature compensation) function
which calculates the resistance of a
DUT at a specified temperature given
the resistance of the DUT at the
ambient temperature and the
temperature coefficient of the DUT is
known.

TCONV The TCONV (Temperature


Conversion) function calculates the
temperature of a DUT given an initial
temperature, initial resistance, measured
resistance and a constant (inferred zero
resistance temperature) for the DUT.

TEMP The TEMP key activates the


temperature measurement function.

16
GETTING STARTED

Speed The Speed key toggles between 10


samples per second and 60 samples per
second (Slow rate and Fast rate).

REL The REL key is used to perform a zero


adjustment to the test leads or a DUT.

RT The RT key is used to display the


real-time (not averaged) measured
resistance value.

Scan The Scan key is used to turn on the


Scan function.

Dry The Dry key is used to turn on the dry


circuit measurement mode which allows
the GOM-805 to measure the contact
resistance of switches and connectors
according to DIN IEC 512 and ASTM
B539 standards. GOM-805 only.

Trigger When in the internal trigger mode,


pressing the Trigger key will turn on the
external trigger mode. When in the
external trigger mode, pressing the
Trigger key will perform a manual
trigger.
A long press of the Trigger key when in
external trigger mode will reset the
trigger mode back to the internal trigger
mode.

Display The Display key toggles between the


standard display mode and the
simplified display mode (sans menus
and display icons).

Local The LOCAL key will switch the


milliohm meter between local and
remote mode.

Diode The Diode key is used to turn on the


Diode measurement function.

17
GOM-804 & GOM-805 User Manual

Drive The Drive key in conjunction with the


+ up/down arrow keys is used to select
the measuring signal: DC+, DC-, Pulse,
PWM, Zero. In particular, the Zero
Enter setting can be used as a +/-10mV DC
voltmeter to measure the EMF of
passive components. See page 33 for
details. GOM-805 only. The drive signal
is fixed to DC+ on the GOM-804.

Range Long pressing the Range key will


activate the auto ranging mode.

Range The Range key in conjunction with the


+ up/down arrow keys is used to select
the resistance measurement range.
When in auto ranging mode, pressing
Enter
the Range key will activate the manual
ranging mode.

ESC The ESC key cancels the current setting


and returns the cursor to its default
location or returns to the previous
menu, depending on the circumstances.

Arrow Keys, The arrow keys and Enter key are


Enter Key used to edit parameters, to navigate
the menu system and to select
Enter parameter ranges.

18
GETTING STARTED

TFT-LCD Overview
Function mode Range Trigger mode Rate Remote mode
Average value Dry circuit Remote error Drive signal

Memory
REL value
number
Function control
Main
indicators
measurement
display

Function
mode
settings

Secondary menus

Function Control The function control indicators show all the currently
Indicators active settings for the selected function mode:
Func Currently selected function mode
Range The measurement range. Auto
indicates that auto ranging is
active
Trigger mode Int/Ext
Rate Slow/Fast
Drive: DC+, DC-, Pulse, PWM, Zero
Rel Shows the relative (nominal)
reference value
Avg Number of samples used for the
Average function.
Dry Indicates that the dry circuit
function is active
Err Indicates a remote command
error

19
GOM-804 & GOM-805 User Manual

RMT Indicates that the unit is in


remote control mode
Mem No. Indicates which memory setting
has been recalled

Main Shows all measurement results for the selected function


Measurement mode.
Display

Function Mode Shows any function mode-specific settings.


Settings

Secondary The secondary menus show global menus (Meas. Setup),


Menus System, Memory) as well as function-specific secondary
menus.
Meas. Setup Goes to the global Measurement Setup
menu.
System Goes to the global System menu
Memory Allows you to save, recall and clear
memory settings.
View Shows the all results for all the channels
when a scan has finished.
Clear Clears the measurement results in the
Binning function when the display
mode is set to Count.

20
GETTING STARTED

Rear Panel Overview


GPIB port RS232 port Handler/Scan/Ext I/O

GPIB

RS232 HANDLER / SCAN / EXT I/O

AC 100 240V , 50 60Hz


25VA MAX
TC SENSOR
WARNING
TO AVOID SHOCK,
REMOVE INPUTS BEFORE OPENING. USB B
port
SER.NO. LABEL

AC power input Temperature sensor port

AC Input AC 100 240V , 50


25VA MAX
60Hz Accepts the power cord. AC 100 -
240Vac; 50 - 60Hz.
For the power up sequence, see page
24.

RS-232 Port RS232


Accepts an RS-232C cable for remote
control; DB-9 male connector.
For remote control details, see page 92.

GPIB Port GPIB


Accepts a GPIB cable for remote
control. See page 93 for details.

USB Device Port USB device port for remote control.


See page 90 for details.

Handler / Scan / HANDLER / SCAN / EXT I/O


The Handler / Scan / EXT I/O
EXT I/O Port port is used to output
pass/fail/high/low comparison
results. This port is also used for
the user-programmable EXT I/O
pins.

21
GOM-804 & GOM-805 User Manual

Temperature TC SENSOR
The temperature sensor input is for the
Sensor Port optional PT-100 temperature probe.

22
GETTING STARTED

Set Up
Tilt Stand

Tilt To tilt, pull the legs forward, as shown below.

Stand Upright To stand the unit upright, push the legs back under the
casing as shown below.

23
GOM-804 & GOM-805 User Manual

Power Up

1. Connection Ensure that the input AC power voltage is within the


range of 100~240 V.
Connect the power cord to the AC Voltage input.

Ensure the ground connector of the power cord is


CAUTION connected to a safety ground. This will affect the
measurement accuracy.

1. Power up Press the main power switch on the front panel.

The display will light up and show the last setting used
before the last shut down.
Example:
Resistance
measurement
mode

24
GETTING STARTED

4 Wire Kelvin Connection

Background The GOM-804/805 uses 4 wire Kelvin connections for


accurate measurements.

Connection SOURCE SENSE SENSE SOURCE

Diagram

GUARD
shielding

-
Description Source + The Source + terminal carries the
measuring current source. It is
connected to the + side of the DUT.
Source - The Source - terminal accepts the signal
return current and connects to the –
side of the DUT.
Sense + Monitors the positive (+) potential.
Sense - Monitors the negative (-) potential.
Guard Grounds the shielding layer of the test
lead cables to reduce noise.
GND Provides a reference ground for the
GOM-804/805.

25
GOM-804 & GOM-805 User Manual

Zeroing (Relative Function)

Background The Relative function is used to perform a zero


adjustment on the test leads.
After the Relative value is pre-set, each measurement that
is displayed is equal to the actual value minus the relative
preset value.

1. Short the Short the test cables together as shown in the diagram
cables below:
SOURCE SENSE SENSE SOURCE

Source+ Source-
/Sense+ /Sense-

2. Set the Press the REL key.


Reference value

3. Relative mode Before REL


display appears

After REL

Rel: Indicates the Relative function is active

26
MEASUREMENT

MEASUREMENT
Ohm Compare Binning

TC TCONV TEMP Enter

Speed REL RT

Scan Dry Trigger ESC

Display Local Diode Drive Range

Resistance Resistance Measurement ....................................... 29


Select the Resistance Range ................................... 30

Drive Signal Measuring Signal (Drive) Overview ........................ 31


Select Measuring Signal (Drive) ............................. 33

Rate Select Measurement Rate ....................................... 34

Display Mode Display Mode ......................................................... 35

Real-Time View Real-Time Measurement ................................ 36

Dry-Circuit Dry-Circuit Measurement ....................................... 37

Trigger Using the Trigger Function ..................................... 38

Diode Diode Function ....................................................... 40

Compare Compare Function .................................................. 41


Function

Binning Binning Function .................................................... 46


Function

27
GOM-804 & GOM-805 User Manual

Temperature Temperature Measurement .................................... 50


Measurement

Temperature Temperature Compensation ................................... 52


Compensation

Temperature Temperature Conversion ........................................ 56


Conversion

Measurement Average Function ................................................... 60


Settings Measure Delay ....................................................... 61
Trigger Delay .......................................................... 63
Trigger Edge ........................................................... 64
Temperature Unit ................................................... 65
Ambient Temperature ............................................ 66
Line Frequency ....................................................... 67
PWM Setting .......................................................... 68

System Settings System Information ............................................... 69


Power On Status Setup .......................................... 70
Interface................................................................. 71
Brightness .............................................................. 72
User Define Pins .................................................... 73
Handler Mode ........................................................ 74
Beep ....................................................................... 76

28
MEASUREMENT

Resistance Measurement
1. Select the Press Ohm to access the Resistance measurement
Resistance mode.
function.

2. Resistance Ohm measurement Resistance range


mode display function indicator and mode
appears.

Ohm measurement

3. Connect the 4-wire resistance:


test lead and Use the SOURCE + and the SOURCE - terminal for
measure measurement, and the SENSE +, and SENSE - terminal
for sensing.
SOURCE SENSE SENSE SOURCE

GUARD
shielding

When switching between measurement ranges, please


Note allow a moment for the circuits to settle before
measuring.

29
Ohm rangeGOM-804 & GOM-805 User Manual

Select the Resistance Range

Background The resistance range can be used with normal resistance


measurement as well as the temperature compensation
function.

Manual Press the RangeOhm keymeasurement


and use the up and down arrow
keys to manually select the resistance range.

Set range
Range select
indicator

Auto Range Long press the Range key to turn on automatic


ranging.
Range, Auto range

Selection List Range Resolution


50mΩ 1uΩ
500mΩ 10uΩ
5Ω 100uΩ
50Ω 1mΩ
500Ω 10mΩ
5kΩ 100mΩ
50kΩ 1Ω
500kΩ 10Ω
5MΩ 100Ω

For detailed specifications, please see the specifications


Note on page 152.

30
MEASUREMENT

Measuring Signal (Drive) Overview

Background Resistance measurement has 5 different measuring signals


that can be applied to obtain a resistance measurement:
DC+, DC-, Pulse, PWM, Zero.
These 5 signals are described in below.
The drive function is only applicable to the GOM-805.
Note The drive signal for the GOM-804 is fixed to DC+.
V
DC+ Open circuit Default drive
~ +6.5V signal.
voltage
0V
t

DC- V
Negative drive
t signal.
0V
Open circuit
~ -6.5V
voltage
Pulse V
50ms This mode can be used to
~+6.5V eliminate the thermoelectric
0V t EMF formed on the contact
~-6.5V between a test lead and a
50ms
DUT.
PWM V ON duty
This mode can be used to
~+6.5V avoid heating up the DUT
0V t and thus avoid having the
measurement accuracy
compromised on
temperature-sensitive DUTs.
Zero V In this mode, GOM-805
outputs no measuring signal
0V t on the Source loop; therefore,
the Sense loop can be used as
a voltage meter which can
measure up to +/-10mV for
thermoelectric EMF
measurement. This function
is useful for measuring the
Vemf of thermocouple wires.

31
GOM-804 & GOM-805 User Manual

A note about Thermoelectric EMF


When making low resistance measurements, thermoelectric electromotive
force (Vemf) can affect measurement accuracy. Vemf is created at the junction
of two dissimilar metals, such as the contact point of a test lead and the pin
of a DUT. Vemf adds a small but measurable voltage to the measurement.

There are primarily two different methods to compensate for Vemf in low
resistance measurements: Offset Compensation and Vemf Cancelling. The
GOM-805 uses Vemf Cancelling with the pulse drive signal setting (see page
33).
The Pulse drive mode supplies a positive and a negative measurement current
source.

R
I V
Vemf

This produces a positive and negative measurement voltage across the DUT,
which also includes the Vemf (V1+Vemf & V2+Vemf).

V1

0V Vemf
t
V2

To cancel the Vemf, V2 is deducted from V1 and divided by 2 to get the


average measurement, as shown in the formula below:
(V 1  Vemf )  (V 2  Vemf )
Vx 
2
Where Vx = measured voltage sans Vemf.

32
MEASUREMENT

Select Measuring Signal (Drive)

Background Resistance measurement has 5 different measuring signals


that can be applied to obtain a resistance measurement:
DC+, DC-, Pulse, PWM, Zero.
The drive function is only applicable to the GOM-805.
Note The drive signal for the GOM-804 is fixed to DC+.

1. Select Drive Press the Drive key and use the up and down arrow
keys to select a drive signal.
Drive mode

Set drive signal


Drive selection
indicator
Drive Range DC+, DC-, Pulse, PWM, Zero

33
GOM-804 & GOM-805 User Manual

Select Measurement Rate

Background The resistance measurement speed has 2 ranges: slow and


fast. Slow speed is the most accurate with 10
measurements/second. Fast speed has 60
measurements/second. Both have the same measurement
resolution.

The rate selection function is not applicable in Diode


measurement mode. When the PWM drive signal is used
or when the Scan function is activated, the only available
rate setting is fast.

1. Select Rate Press the Speed key to toggle between the Slow and Fast
rates.
Measurement rate

34
MEASUREMENT

Display Mode

Background The Display key can be used to toggle between the


normal and the simplified display mode. The simplified
display mode clears all text, menus and function
indicators from the screen except for the measurement
and measurement mode indicators.

1. Toggle Display Press the Display key to toggle the display between
mode normal and simplified. The display will change
accordingly.
Simplified Measurement mode
Display Mode
Example

Measurement

35
GOM-804 & GOM-805 User Manual

View Real-Time Measurement

Background When measurements are smoothed using the averaging


function, the RT key can be used to view the real-time
results in addition to the averaged results.

See page 60 for Average configuration.

1. Toggle Press the RT key to toggle the real-time display on


Real-Time or off.
display
The real-time measurement will appear in the bottom
left-hand corner.

Real-time
measurement

36
MEASUREMENT

Dry-Circuit Measurement

Background The Dry Circuit measurement function is used where the


maximum open-circuit voltage must be kept to a
minimum for applications such as measuring the contact
resistance of switches, relays and connectors. The
GOM-805 provides a maximum of up to 20mV in this
mode.
Dry circuit testing is for switch and connector contact
Note resistance. Switch and connector contact resistance
measurement is in accordance with DIN IEC 512 and
ASTM B539 which requires that the open circuit voltage
of the measuring device should not exceed 20mV DC.
Voltage at such low levels avoids the breakdown of any
oxides that may be present on the contacts. In this mode
the open circuit measuring voltage is limited <20mV,
while modes like DC+ or pulse mode can have an open
circuit measuring voltage as high as 6.5V.
Dry Limitations When the Dry Circuit measurement function is turned
on, the measurement range is reduced. See the
specifications for more details.
Range Dry Mode Rate
50mΩ ✘
500mΩ ✔ Slow/Fast
5Ω ✔ Slow/Fast
50Ω ✔ Slow/Fast
500Ω ✘
5kΩ ✘
50kΩ ✘
500kΩ ✘
5MΩ ✘

37
GOM-804 & GOM-805 User Manual

1. Toggle Dry Press the Dry key to toggle the dry circuit
mode on or off measurement mode on or off.
The DRY function indicator will appear in the middle of
the display when active.
Dry Circuit measurement
mode indicator

Using the Trigger Function

Background The GOM-804/805 can use internal or manual triggering


for the Resistance, Temperature, Temperature
Compensation, Temperature Conversion, Binning,
Handler and Scan modes.
By default the GOM-804/805 is set to internal triggering
mode.

1. Select Manual Short press Trigger to switch to manual triggering mode.


Trigger
The Ext indicator will be shown on the display
when the manual trigger is active.
Trigger source

2. Manually Short press the Trigger key each time you want to start a
Triggering single measurement (when in the manual mode).
Measurements

38
MEASUREMENT

3. Internal Long press Trigger to return the triggering mode back to


Triggering internal mode.
The Int indicator will be shown on the display.
Internal trigger source

39
GOM-804 & GOM-805 User Manual

Diode Function

Background The Diode function can be used to measure the forward


bias voltage of a diode under test.
1. Select the Press Diode to access the Diode measurement mode.
Diode function.

2. Diode mode Diode function


appears. indicator

Forward bias voltage

3. Connect the Connect the Sense+, Source+ to the anode.


test lead and Connect the Sense-, Source- to the cathode.
measure
SOURCE SENSE SENSE SOURCE

40
MEASUREMENT

Compare Function
Background The compare function compares a measured value to a
“Reference” value that has an upper (HI) and lower (LO)
limit. If the measured value is within the upper and lower
limit, then the measured value is judged as IN.
There are three compare modes that can be used to make
a judgment: ABS, △% and % modes.
The ABS mode displays the absolute difference between
the measured and the reference value (shown as △) and
compares the measured value to the upper (HI) and
lower (LO) limit. The upper and lower limits are set as
absolute resistance values.
Compare function
Measured value
indicator
Absolute
difference of
measured value
from Reference
value

Pass/Fail
judgment

Reference, limits, compare


mode and beep mode
A measured value that falls within the upper and lower
limits is considered IN (pass), a value that falls below the
lower limits is considered LO, and a value that falls over
the upper limit is a HI.

Lower Reference Upper


limit value limit

LO IN HI
[Note that the reference value in the ABS mode is only
for reference purposes and is not used to make a
judgment.]

41
GOM-804 & GOM-805 User Manual

The △% compare function displays the deviation of the


measured value from the reference value as a percentage.
{ [(Measured Value-Reference)/Reference]%}.
Compare function
Measured value
indicator

Deviation of the
measured value from
the reference value as
a percentage

Pass/Fail judgment

Reference, limits, compare


mode and beep mode

The upper (HI) and low (LO) limits are set as a


percentage from the reference value. (Identical to the %
compare mode)
A measured value that falls within the upper and lower
limits is considered IN (pass), a value that falls below the
lower limits is considered LO, and a value that falls over
the upper limit is a HI.
Lower Reference Upper
limit value limit
%
%

LO IN HI

The % compare mode displays the measured value as a


percentage of the reference value [(Measured
Value/Reference Value)%].
The upper (HI) and low (LO) limits are set as a
percentage from the reference value. (Identical to the △
% compare mode)

42
MEASUREMENT

Compare function
Measured value
indicator

Measured value as a
percentage of the
reference value

Pass/Fail judgment

Reference, limits, compare


mode and beep mode

A measured value that falls within the upper and lower


limits is considered IN (pass), a value that falls below the
lower limits is considered LO, and a value that falls over
the upper limit is a HI.
Lower Reference Upper
limit value limit
%
%

LO IN HI
For all the compare modes, IN, HI or LO will be shown
on the display for each judgment.

1. Select the Press Compare to access the compare mode, as shown


compare function above.

2. Select the Use the arrow keys to navigate to the Mode setting. Press
compare mode the Enter key to toggle the compare mode.

Move

Enter Toggle
Mode

Range Abs,△ %, %

43
GOM-804 & GOM-805 User Manual

3. Reference Use the arrow keys to navigate to the Reference setting


value setting and press Enter.
Use the left and right arrow keys to select a digit. Use the
up and down arrow keys to edit the value of the selected
digit and the unit. Press Enter to confirm the setting.

Move
and edit

Enter
Select and
confirm Reference

Range: 000.0001~ 999.9999


(mΩ/Ω/kΩ/MΩ)

After setting the Reference value, the displayed △, % or


Note △% values will be changed to reflect the new Reference
value setting.

4. Upper & lower Use the arrow keys to navigate to the Upper or Lower
limit setting limit setting and press Enter.
Use the left and right arrow keys to select a digit. Use the
up and down arrow keys to edit the value of the selected
digit. Press Enter to confirm the setting.
Repeat for the other limit (Upper or Lower).

Move
and edit

Enter
Select and
confirm Upper, Lower reference

Setting Range: ABS mode: 000.0000~999.9999


(mΩ/Ω/kΩ/MΩ)
△ % and % mode:
-999.99 ~ +999.99

The upper limit must be higher than the lower limit. Not
Note setting the upper limit higher than the lower limit is not
allowed. Likewise the lower limit cannot be set higher
than the upper limit.

44
MEASUREMENT

5. Beep setting Use the arrow keys to navigate to the Beep setting.
Press Enter to toggle the beep setting.

Move

Enter Toggle
Beep setting

Beep Setting: Off, Pass, Fail

The Beep setting can also be set from the


Note System>Utility>Beep>Compare menu.

45
GOM-804 & GOM-805 User Manual

Binning Function
Background The Binning function is used to grade DUTs into eight
different bins according to 8 sets of upper and lower
limits. Two compare modes can be used in this function,
ABS and △% modes.
Binning function
Grading results
indicator

Upper and lower


limits for the 8
bins

Reference, compare mode, beep


mode and display mode

1. Select the Press the Binning key to access this function.


Binning function

2. Select the Use the arrow keys to go to the Mode setting.


compare mode
Press Enter to toggle between ABS or △% compare
modes.

Move

Enter Toggle
Mode setting

ABS Mode The ABS mode allows you to set


the upper and lower limits of each
bin as absolute resistance values.
△% The Delta % mode allows you to
set the upper and lower limits of
each bin as percentage value from
the reference value.

46
MEASUREMENT

For further details on the ABS or △% compare modes,


Note see the description in the Compare section, page 41.

3. Reference Although the 8 bins have their own upper and lower
value setting limits, they still share a common reference value.
Use the arrow keys to go to the Reference setting and
press Enter.
Use the left and right arrow keys to select a digit. Use the
up and down arrow keys to edit the value of the selected
digit and the unit. Press Enter to confirm the setting.

Move
and edit

Enter
Select and
confirm Reference

Range 000.0001~
999.9999(mΩ/Ω/kΩ/MΩ)

4. Upper & lower Use the arrow keys to go to the upper limit of the first
limit settings bin and press Enter.
Use the Left and Right arrow keys to select a digit. Use
the Up and Down arrow keys to edit the value of the
selected digit and unit. Press the Enter key to confirm the
setting.
Repeat for the lower setting.
Repeat for the remaining bins.
Upper, lower
reference

Move
and edit

Enter
Select and
confirm

Setting range ABS mode: 000.0000~999.9999


(mΩ/Ω/kΩ/MΩ)
△ % mode: -999.99 ~ +999.99

47
GOM-804 & GOM-805 User Manual

The upper limit must be higher than the lower limit. Not
Note setting the upper limit higher than the lower limit is not
allowed. Likewise the lower limit cannot be set higher
than the upper limit.

5. Beep setting Use the arrow keys to navigate to the Beep setting.
Press Enter to toggle the beep setting.

Move

Enter Toggle
Beep setting
Beep Setting: Off, Pass, Fail

The Beep setting can also be set from the


Note System>Utility>Beep>Binning menu.

6. To start The binning function starts automatically if you are in


binning internal trigger mode.
If you are using the manual triggering mode, press the
Trigger button or apply a pulse on the trigger pin of the
Handler interface to start binning.
See page 38 to set the triggering modes.

7. Display the There are two different display modes to view results.
binning results The Comp (Compare) display mode is the default display
mode. This mode will display the currently measured
value and displays which of the bins (if any) the
measured value is graded as.
Grading results:
Green = IN
Red = OUT

Measurement

48
MEASUREMENT

The Count display mode tabulates the results on the


right-hand side of the display and shows the bin settings
on the left.
Tabulated result
of each bin

Overall
results

Clear
results
Upper and lower limits of Bin 1~8

To toggle the display mode, go to the Disp setting and


press Enter.

Move

Enter Toggle
Disp setting

8. How to clear When in the Count display mode, press the ESC key.
the result count Go to the Clear setting and press Enter. The accumulated
results will be cleared from the display.

Move

Clear setting Enter Clear results

49
GOM-804 & GOM-805 User Manual

Temperature Measurement
Background The temperature measurement function uses the optional
PT-100 temperature probe. The measured temperature is
displayed on the display. For more information on the
optional PT-100 sensor, see the appendix on page 149.
There is only one range for the temperature function.
However the resistance measurement range can still be
changed when in the temperature function.
Note: The temperature measurement function is used in
conjunction with the Ohm measurement function. The
two measurements share the same display, so the Ohm
readings stay on the display even after the temperature
measurement function is activated. Thus when the
Temperature function is selected, “Ohm+T” is shown as
the selected function.

1. Select the Press TEMP to enter the temperature measurement


Temperature function.
function
Temperature + Ohm
Resistance measurement
function indicator

(Ambient)
temperature
source

Ambient
temperature

The temperature is displayed on the Ohm display.

2. Select the From the bottom menu, go to Meas. Setup>Temperature


temperature Unit and select ºC or ºF.
units See page 65 for setting details.

50
MEASUREMENT

3. Ambient The Ambient temperature setting should be turned off


Temperature when using the temperature function.
From the bottom menu go to Meas. Setup > Ambient
Temperature and turn the Ambient Temperature setting
off.
See page 66 for setting details.

4. Temperature The temperature sensor uses the rear panel TC Sensor


mode port for input.
connection GPIB

RS232 HANDLER / SCAN / EXT I/O

AC 100 240V , 50 60Hz


PT-100 temperature
25VA MAX

WARNING
TO AVOID SHOCK,
REMOVE INPUTS BEFORE OPENING.
TC SENSOR

sensor
SER.NO. LABEL

51
GOM-804 & GOM-805 User Manual

Temperature Compensation
Background If the resistance of a DUT at a particular temperature is
needed, the compensation function can be used. This
function can simulate the resistance of a DUT at a
desired temperature. If the ambient temperature and the
temperature coefficient of the DUT are known, it is
possible to determine the resistance of a DUT at any
temperature.
The Temperature Compensation works on the following
formula:
Rt
Rt 0 
1  αt 0(t-t0)
Where:
Rt = Measured resistance value (Ω)
Rt0 = Corrected resistance value (Ω)
T0 = Inferred absolute temperature
t0 = Corrected temperature (ºC)
t = Current ambient temperature (ºC)
αto = Temperature coefficient of resistance at the correct
1
temperature. ato  .
| T 0 | t 0

52
MEASUREMENT

1. Select the Press TC to access the Temperature Compensation


Temperature function.
Compensation The temperature-compensated resistance measurement
mode will appear on the display.
Temperature Extrapolated resistance
compensation measurement at the desired
function indicator (“correct”) temperature

Ambient
temperature
source

Ambient
temperature

Correct Temperature,
Temperature Coefficient settings

2. Ambient The ambient temperature can be either measured with


Temperature the PT-100 sensor or be set manually.
If using the PT-100 sensor the Ambient temperature
setting should be turned off. If the PT-100 probe is not
used, then the ambient temperature needs to be manually
set.
From the bottom menu, go to Meas. Setup > Ambient
Temperature and set the ambient temperature.
See page 66 for setting details.
Range Off, -50.0 ºC ~ 399.9ºC

53
GOM-804 & GOM-805 User Manual

3. Temperature Use arrow keys to go to Correct Temperature or to


compensation Temperature Coefficient and press Enter to select the
setting.
To edit the setting values use the left and right arrow keys
to select a digit and use the up and down arrow keys to
edit the digit. Press Enter to confirm the setting.

Move
and edit

Select and
Enter Correct temperature, temperature
confirm
coefficient settings

Desired Temperature range -50.0 ~ +399.9 ˚C


Temperature Coefficient range -9999 ~ +9999 ppm

Below are the inferred zero resistance temperatures of


some common conductors:
Material Inferred Absolute Temperatures
Silver -243
Copper -234.5
Gold -274
Aluminium -236
Tungsten -204
Nickel -147
Iron -162

54
MEASUREMENT

3. Temperature Sensor Connection:


compensation GPIB

connection
RS232 HANDLER / SCAN / EXT I/O

AC 100 240V , 50 60Hz


PT-100 temperature
25VA MAX

WARNING
TO AVOID SHOCK,
REMOVE INPUTS BEFORE OPENING.
TC SENSOR

sensor
SER.NO. LABEL

Note: If the sensor is not connected, then the Ambient


temperature needs to be manually set.
DUT connection:
4 wire Kelvin:
SOURCE SENSE SENSE SOURCE

GUARD
shielding

55
GOM-804 & GOM-805 User Manual

Temperature Conversion
Background The Temperature Conversion function allows you to
determine the temperature change of a DUT at any given
resistance, if the initial temperature, the inferred zero
resistance temperature for the DUT and the initial
resistance of the DUT are known. The displayed result
can also be the extrapolated to calculate the final
temperature (T) or the extrapolated temperature
difference (△T)*.
Temperature Conversion function works on the following
formula:
R2 t 0  t 2

R1 t 0  t1
Where:
R2 = resistance @ temperature t2
R1 = resistance @ temperature t1
t0 = inferred zero resistance temperature in ºC**
t1 = temperature at R1
t2 =temperature at R2
The temperature conversion function is can be used to
determine the temperature of transformer windings,
electric motors, or other materials where it may not be
practical to embed a temperature sensor.

*(T) Final temperature = t2 = △T +TA


(TA) Ambient temperature = Ambient temperature when
R2 is measured. TA can either by manually measured with
the PT-100 sensor or it can be manually set.
(△T) Extrapolated temperature difference = T - TA

**“Constant” setting on the panel display is equivalent to


the absolute value of the inferred zero resistance
temperature.

56
MEASUREMENT

Common inferred Metallic conductors show increased resistivity when


zero resistance temperature is increased, and likewise show reduced
temperatures resistivity when temperature is reduced. Inferred zero
resistance temperature is simply the inferred temperature
at which the material will have no resistance. This value is
derived from the temperature coefficient of the material.
Note: the inferred zero resistance temperature is an ideal
value, and not a real-world value.

Material Inferred zero resistance temp. in ºC


Silver -243
Copper -234.5
Gold -274
Aluminium -236
Tungsten -204
Nickel -147
Iron -162

1. Select the Press TCONV to access the temperature compensation


Temperature function.
compensation The temperature-converted measurement will appear on
mode. the display.
Temperature
conversion function Resistance measurement
indicator

(Ambient)
temperature
source

Extrapolated
temperature
difference or final
temperature

57
GOM-804 & GOM-805 User Manual

2. Initial Use the arrows keys to go to Initial Resistance, Initial


Resistance, Temperature or Constant (inferred initial resistance
Initial temperature) and press Enter.
Temperature and Use the left and right arrow keys to select a digit and use
Constant settings the up and down arrow keys to edit the digit. Press Enter
to confirm the edit.

Move
and edit

Select and
Enter Initial Resistance, Initial Temperature
confirm
and Constant settings

Initial Resistance 000.0001~999.9999 mΩ, Ω, kΩ,



Initial Temperature -50.0 ~ +399.9 ℃
Constant 000.0~999.9

3. Display mode Use the arrow keys to go to Disp. Press Enter to toggle
between the T and △T modes.

Move

Enter Toggle
Disp setting

T displays the extrapolated temperature at the measured


resistance of the DUT.

△T displays the difference from the extrapolated


temperature at the measured resistance of the DUT and
the ambient temperature. Please refer to page 56 for
further details.

3. Temperature Sensor Connection:


compensation GPIB

connection.
RS232 HANDLER / SCAN / EXT I/O

AC 100 240V , 50 60Hz


PT-100 temperature
25VA MAX

WARNING
TO AVOID SHOCK,
REMOVE INPUTS BEFORE OPENING.
TC SENSOR

sensor
SER.NO. LABEL

58
MEASUREMENT

DUT connection
4 wire Kelvin:
SOURCE SENSE SENSE SOURCE

GUARD
shielding

59
GOM-804 & GOM-805 User Manual

Measurement Settings
Background The following measurement settings are used to
configure the various measurement modes.

Average Function

Background The average function smoothes measurements using a


moving average. The average function sets the number of
samples used for the moving average; a higher number
results in smoother measurement results. The average
function is turned off by default.

1. Select Average From one of the main screens, press Meas. Setup
setting the ESC key so that the menu menu icon
system at the bottom of the display
has focus.
Go to Meas. Setup and press Enter. Move

Go to Average and press Enter.


Enter
Select menu
or setting

2. Average Use the arrow keys to turn Average on and set the
setting appears average number. Press Enter to confirm the setting.
Average settings

Average OFF, ON: 2~10

Pressing ESC before pressing ENTER will exit the Average


Note function settings.

60
MEASUREMENT

Measure Delay

Background The Measure Delay setting inserts a delay time between


each measurement. Measure delay is turned off by
default.
Measurement start with
Measure delay time
Test signal

Measure delay time

Default Measurement
start time
The measure delay setting is useful for measuring
components that need some time to charge if the default
measurement start time is not adequate. An adequate
delay time allows the meter to avoid the effects of
transient disturbances that are usually seen when
measuring reactive DUTs with a current source.

1. Select From one of the main screens, press Meas. Setup


Measure Delay the ESC key so that the menu menu icon
setting system at the bottom of the display
has focus.
Go to Meas. Setup and press Enter. Move

Go to Measure Delay and press


Select menu
Enter. Enter
or setting

2. Measure Delay Use the arrow keys to turn Measure Delay on and set the
setting appears delay time. Press Enter to confirm the setting.

Measure delay
setting

Measure Delay* OFF, ON: 000.000 ~ 100.000s


* When the set value is > 0.1s, the resolution is 0.1s.
When the set value is < 0.1S, the resolution is 1mS.

61
GOM-804 & GOM-805 User Manual

Pressing ESC before pressing ENTER will exit the Measure


Note Delay settings.

62
MEASUREMENT

Trigger Delay

Background The Trigger Delay setting adds a delay to when an


external trigger signal is recognized. Normally the
external trigger is recognized when there is no contact
bounce in the signal for a fixed length of time, this time
is known as the bounce monitoring window. This ensures
that the external trigger signal is stable before it is
recognized. The Trigger Delay time starts right after the
bounce monitoring window ends.
Bounce monitoring
window Start measurement
Contact period
bounce

External
Trigger signal
Trigger Measurement
delay time period

Measurement
process
Measurement Measurement
delay time time

The Trigger Delay setting is turned off by default.


Pin 2 of the Handler/Scan/Ext I/O interface is used for
Note external triggering, See page 77 for pinout details.

1. Select Trigger From one of the main screens, press Meas. Setup
Delay setting the ESC key so that the menu menu icon
system at the bottom of the display
has focus.
Go to Meas. Setup and press Enter. Move

Go to Trigger Delay and press


Select menu
Enter. Enter
or setting

63
GOM-804 & GOM-805 User Manual

2. Trigger Delay Use the arrow keys to turn Trigger Delay on and set the
setting appears delay time. Press Enter to confirm the settings.

Trigger Delay
setting

Trigger Delay OFF, ON: 0 ~ 1000ms

Pressing ESC before pressing ENTER will exit the Trigger


Note Delay settings.

Trigger Edge

Background The Trigger Edge setting sets the external trigger edge as
rising or falling. By default the trigger edge is set to rising.

1. Select Trigger From one of the main screens, press Meas. Setup
Edge setting the ESC key so that the menu menu icon
system at the bottom of the display
has focus.
Go to Meas. Setup and press Enter. Move

Go to Trigger Edge and press Enter.


Enter
Select menu
or setting

2. Trigger Edge Use the arrow keys to set the Trigger Edge. Press Enter
setting appears to confirm the setting.

Trigger Edge
setting

Trigger Edge Rising, Falling

Pressing ESC before pressing ENTER will exit the Trigger


Note Edge settings.

64
MEASUREMENT

Temperature Unit

Background Temperature units can be set to Fahrenheit or Celsius for


all temperature measurements.

1. Select From one of the main screens, press Meas. Setup


Temperature Unit the ESC key so that the menu menu icon
setting system at the bottom of the display
has focus.
Go to Meas. Setup and press Enter. Move

Go to Temperature Unit and press


Select menu
Enter. Enter
or setting

2.Temperature Use the arrow keys to set the Temperature Unit. Press
Unit setting Enter to confirm the setting.
appears
Temperature
Unit

Temperature Unit Fahrenheit, Celsius

Pressing ESC before pressing ENTER will exit the


Note Temperature Unit setting.

65
GOM-804 & GOM-805 User Manual

Ambient Temperature

Background The Ambient Temperature setting is used to set the


ambient (room temperature) for the Temperature
Compensation or Temperature Conversion function in
the absence of the PT-100 temperature sensor. See page
52 and 56 respectively for details.

1. Select Ambient From one of the main screens, press Meas. Setup
Temperature the ESC key so that the menu menu icon
setting system at the bottom of the display
has focus.
Go to Meas. Setup and press Enter. Move

Go to Ambient Temperature and


Select menu
press Enter. Enter
or setting

2.Ambient Use the arrow keys to set the Ambient Temperature.


Temperature Press Enter to confirm the setting.
setting appears
Ambient
Temperature

Ambient Temperature Off, On: -50ºC ~ 399.9ºC

Pressing ESC before pressing ENTER will exit the Ambient


Note Temperature setting.

66
MEASUREMENT

Line Frequency

Background The Line Frequency setting selects the appropriate line


filter to reduce the influence of the AC line frequency on
the milliohm measurements. This setting is set to AUTO
by default.

1. Select Line From one of the main screens, press Meas. Setup
Frequency the ESC key so that the menu menu icon
setting system at the bottom of the display
has focus.
Go to Meas. Setup and press Enter. Move

Go to Line Frequency and press


Select menu
Enter. Enter
or setting

2.Line Frequency Use the arrow keys to set the Line Frequency. Press Enter
setting appears to confirm the setting.

Line Frequency

Line Frequency Auto, 50Hz, 60Hz

Pressing ESC before pressing ENTER will exit the Line


Note Frequency setting.

67
GOM-804 & GOM-805 User Manual

PWM Setting

Background The PWM setting will set the duty of the PWM Drive
setting. The duty is set with ON and OFF times for the
waveform.

OFF time
ON
time

See page 31 for Drive setting details.

1. Select PWM From one of the main screens, press Meas. Setup
setting the ESC key so that the menu menu icon
system at the bottom of the display
has focus.
Go to Meas. Setup and press Enter. Move

Go to PWM and press Enter.


Enter
Select menu
or setting

2.PWM setting Use the arrow keys to set the ON and OFF time for the
appears duty. Press Enter to confirm the setting.

PWM

ON 03 ~ 99 time units*
OFF 0100 ~ 9999 ms
*The ON time setting is set in “time units”, not
milliseconds. The amount of time in a time unit depends
on the line frequency settings (see page 67).
Line frequency 1 Time Unit
60Hz 16.6mS
50Hz 20mS

Pressing ESC before pressing ENTER will exit the PWM


Note setting.

68
MEASUREMENT

System Settings
Background The System settings are used to view the system
information, set the power on state, the remote interface,
screen brightness, external interface and beep settings as
well as access the calibration menu.

System Information

Background The System Information will show the manufacturer,


model, software version and serial number of the unit.
The system information is the equivalent of the return
string from the *idn? query (page 144).

1. View System From one of the main screens, press System


Information the ESC key so that the menu menu icon
system at the bottom of the display
has focus.
Go to System and press Enter. Move

System information will be displayed


Select menu
at the top of the System menu. Enter
or setting

System Information

Pressing ESC will exit from the System menu.


Note

69
GOM-804 & GOM-805 User Manual

Power On Status Setup

Background The Power On Status Setup allows you to either load the
previous settings or the default settings on startup.

1. Select Power From one of the main screens, press System


On Status the ESC key so that the menu menu icon
setting system at the bottom of the display
has focus.
Go to System and press Enter. Move

Go to Power On Status Setup and


Select menu
press Enter. Enter
or setting

2. Power On Use the arrow keys to set Power ON Status Setup. Press
Status Setup Enter to confirm the setting.
appears

Power On
Status Setup

Power On Status Recall Previous Settings,


Load Default

Pressing ESC before pressing ENTER will exit the Power


Note On Status Setup.

70
MEASUREMENT

Interface

Background The remote interface can be set to RS232, GPIB or USB.


The GPIB interface is only available on the GOM-804G
Note and the GOM-805.

1. Select From one of the main screens, press System


Interface setting the ESC key so that the menu menu icon
system at the bottom of the display
has focus.
Go to System and press Enter. Move

Go to Utility and press Enter.


Enter
Select menu
Go to Interface and press Enter. or setting

2. Interface Use the arrow keys to choose an interface and to set the
setting appears baud rate (RS232) or primary address (GPIB). Press
Enter to confirm the setting.

Interface

Interface GPIB, Primary Address (1 ~ 30)


RS232, Baud Rate (1200, 2400, 4800,
9600, 19200, 38400, 57600, 115200)
USB

Pressing ESC before pressing ENTER will exit from the


Note Interface settings.

71
GOM-804 & GOM-805 User Manual

Brightness

Background The Brightness setting sets the backlight brightness of


the TFT-LCD panel.

1. Select From one of the main screens, press System


Brightness the ESC key so that the menu menu icon
setting system at the bottom of the display
has focus.
Go to System and press Enter. Move

Go to Utility and press Enter.


Enter
Select menu
Go to Brightness and press Enter. or setting

2. Brightness Use the arrow keys to set the brightness level. Press
setting appears Enter to confirm the setting.

Brightness

Brightness 01 (dim) ~ 05 (bright)

Pressing ESC before pressing ENTER will exit from the


Note Brightness settings.

72
MEASUREMENT

User Define Pins


Background The External I/O User Define Pin settings set the logic
and the active level for the Define 1 and Define 2 pins on
the Handler/Scan/EXT I/O port on the rear panel. The
External I/O pins are used with the compare or bin
functions. The logic settings can be based on the pass,
fail, high, low or bin grade results of the selected
function.
1. Select External From one of the main screens, press System
I/O Setting the ESC key so that the menu menu icon
system at the bottom of the display
has focus.
Go to System and press Enter. Move

Go to Utility and press Enter.


Enter
Select menu
Go to External I/O and press or setting
Enter.
2. External I/O Use the arrow keys to choose either User Define 1 or
Menu Appears User Define 2 and press Enter.
Use the arrow keys to set the active level of the pin when
the logic conditions are true and to set the logic settings.
Press Enter to confirm the settings.

User Define 1

User Define 2

User Define 1/2: Pin Active: High, Low


Logic:
Operand1 Operator Operand2
Fail Fail
Pass Logical OR, Pass
Low Logical Low
High AND, High
Bin O** OFF* Bin O**
Bin 1 ~ 8 Bin 1 ~ 8

73
GOM-804 & GOM-805 User Manual

*The OFF operator sets the Logic as


true when Operand1 is true.
** Bin O is defined as outside bin 1~ 8.
The Bin logic settings are not available for the GOM-804.
Note
Pressing ESC before pressing ENTER will exit from the
selected External I/O setting.

Handler Mode

Background The Handler Mode setting determines the behavior of


the result signals from the handler interface. There are
two settings, Clear and Hold. The Clear setting will clear
the results of the previous test before starting the
succeeding one and the Hold setting will keep the test
result of the previous test until the succeeding test has
completed.
The timing diagrams below are used as examples. All the
result signals in the examples are active high.
Clear example Clear: All result signals (PASS, Fail, High and Low) are
cleared at the falling edge of EOT and the results from
the current test are output at the rising edge of the EOT
signal.
EOT falling EOT rising
edge edge

Trigger
Ready
EOT
Pass
Fail
High
Low
Previous
Previous results New
results cleared results

74
MEASUREMENT

Hold example Hold: The results of the previous tests are held until the
current test has completed.
EOT falling EOT rising
edge edge

Trigger
Ready
EOT
Pass
Fail
High
Low

Previous Previous results New results


results held

1. Select External From one of the main screens, press System


I/O setting the ESC key so that the menu menu icon
system at the bottom of the display
has focus.
Go to System and press Enter. Move

Go to Utility and press Enter.


Enter
Select menu
Go to External I/O and press or setting
Enter.
2. External I/O Use the arrow keys to choose Handler Mode and press
menu appears Enter.
Use the arrow keys to set the handler mode. Press Enter
to confirm the setting.

Handler Mode

Handler Mode HOLD, CLEAR

Pressing ESC before pressing ENTER will exit from the


Note Handler Mode setting.

75
GOM-804 & GOM-805 User Manual

Beep

Background The Beep setting will configure the beeper sound for the
key presses, the Compare function and the Binning
function.
For the Compare and Binning function the beep can be
configured to beep on a pass or fail judgment.

1. Select Beep From one of the main screens, press System


setting the ESC key so that the menu menu icon
system at the bottom of the display
has focus.
Go to System and press Enter. Move

Go to Utility and press Enter.


Enter
Select menu
Go to Beep and press Enter. or setting

2. Beep menu Use the arrow keys to choose a beep setting and press
appears Enter.
Use the arrow keys to set the selected setting and press
Enter to confirm.

Key Click Setting

Compare Setting

Binning Setting

Beep Settings: Key Click On, Off


Compare Off. Pass, Fail
Binning Off. Pass, Fail

Pressing ESC before pressing ENTER will exit from the


Note selected Beep setting.

76
HANDLER/SCAN INTERFACE

HANDLER/SCAN
INTERFACE

Handler Handler Overview ................................................... 78


Pin Definitions for the Handler Interface ............... 80
Handler Interface for Binning and Compare
Functions ............................................................... 80

Scan Scan Overview ........................................................ 82


Pin Definitions for the SCAN Interface ................... 83
Scan Interface ........................................................ 83
Scan Setup ............................................................. 84
Scan Output ........................................................... 88

GOM-802 GOM-802 Compatibility for Scan and Handler


Compatibility Interfaces ............................................................... 89
GOM-805 to GOM-802 Handler/Scan Interface ..... 89

Remote Interface Configure USB Interface ......................................... 90


Install USB Driver .................................................. 91
Configure RS-232 Interface ..................................... 92
Configure GPIB Interface ....................................... 93
RS232/USB Function Check ................................... 93
Using Realterm to Establish a Remote Connection
............................................................................... 94
GPIB Function ........................................................ 96

77
GOM-804 & GOM-805 User Manual

Handler Overview
Background The Handler interface is used to help grade components
based on the Compare or Binning function test results.
The appropriate pins on the handler interface are active
when the Compare or Binning function is used.
There are 17 TTL outputs and 1 TTL inputs. The
Handler interface is only applicable with the Binning
function or Compare measurement modes.

Please see following pages for related functions and


Note settings:
Compare function: 41
Binning function: 46
Ext I/O settings: 73
Handler mode settings 74
Interface and pin 25-Pin D-SUB HANDLER / SCAN / EXT I/O

assignment (Female)

Pin assignment TRIGGER Starts the trigger for a single


measurement.
READY High when the measurement has
finished. The instrument is ready for
the next trigger.
EOT High when the AD conversion has
completed. The DUT is ready to be
changed.
BIN 1~8 High when the sorting result is in one
of the eight bin grades. Bin1~8 (pass).
BIN OUT High when the sorting result is out of
all the eight bin grades (Bin1~8). The
status of this pin reflects either a HI or
LO result (fail).
LOW High when the compare result is
deemed LO.
HIGH High when the compare result is
deemed HI.

78
HANDLER/SCAN INTERFACE

FAIL High when the compare result is either


HI or LO (fail).
PASS High when the compare result is IN
(pass).
For the full pin definition, please refer to the table listed below.

The output current from all the pins and the VINT(+5V)
Note pin cannot exceed 60mA.

79
GOM-804 & GOM-805 User Manual

Pin Definitions for the Handler


Interface
As this interface is used for the handler and scan functions, the interface
pinout depends on the function mode. The following pinout is only applicable
when using the Binning or Compare function.

HANDLER / SCAN / EXT I/O


13 1

25 14

Handler Interface for Binning and Compare Functions


Pin Name Description Active In/
modes Out
1, 17 Reserved
2 Trigger Trigger for a single measurement. All In
3, 14, GND Ground.
18
4 Fail High when the compare result is Compare Out
either HI or LO (fail).
5 High High when the compare result is Compare Out
deemed HI.
6 Pass High when the compare result is IN Compare Out
(pass).
7 EOT High when the AD conversion has Ext Out
completed. The DUT is ready to be trigger
changed. mode
8 VINT Internal DC Voltage +5V. Out
9 Bin1 High when the binning sorting result Binning Out
is within the bin1 setting range.
10 Bin2 High when the binning sorting result Binning Out
is within the bin2 setting range.
11 Bin3 High when the binning sorting result Binning Out
is within the bin3 setting range.
12 Bin4 High when the binning sorting result Binning Out
is within the bin4 setting range.

80
HANDLER/SCAN INTERFACE

13 Bin5 High when the binning sorting result Binning Out


is within the bin5 setting range.
15 Userdefine2 High or low when the user define2 Compare, Out
logic conditions are met. Binning
16 Userdefine1 High or low when the user define1 Compare, Out
logic conditions are met. Binning
19 VEXT External DC Voltage, acceptable In
range is +5V.
20 Ready High when the measurement has Ext Out
finished. The instrument is ready for trigger
the next trigger. mode
21 Bin6 High when the binning sorting result Binning Out
is within the bin6 setting range.
22 Low High when the compare result is Compare Out
deemed LO.
23 Bin7 High when the binning sorting result Binning Out
is within the bin7 setting range.
24 Bin8 High when the binning sorting result Binning Out
is within the bin8 setting range.
25 Bin Out High when the binning sorting result Binning Out
is out of all the bin setting ranges.
For backwards compatibility with the GOM-802 handler interface, please see
page 89.

81
GOM-804 & GOM-805 User Manual

Scan Overview
Background The Scan function is used to automatically bin groups of
up to 100 components. The associated pins in the handler
interface are active when the Scan function is activated.
There are a total of 6 outputs, 3 inputs as well as a GND
and power (+5V) pin.
Interface and pin 25Pin D-SHELL HANDLER / SCAN / EXT I/O

assignment (Female)

Pin Assignment Relay Controls the relay output.


Pass Pass signal. Indicates the compare result
is IN(pass).
Low Low signal. Indicates a LO compare
result.
High High signal. Indicates a HI compare
result.
Clock The clock signal will pulse high when
each group of output signals (Relay,
Pass, Low, High) are ready. There are up
to 100 groups of output signals
STRB After all (100) output groups are ready,
the STRB signal will pulse high.

82
HANDLER/SCAN INTERFACE

Pin Definitions for the SCAN Interface


As this interface is used for the handler and scan functions, the interface
pinout depends on the function mode. The following pinout is only applicable
when using the Scan function.

HANDLER / SCAN / EXT I/O


13 1

25 14

Scan Interface
Pin Name Description In/Out
1,9-13,15-17,21,23 Reserved
-25
2 Trigger Start for Scan measurement. In
3,14,18 GND Ground.
4 High High signal. Indicates a HI compare Out
result.
5 Clock The clock signal will pulse high when Out
each group of output signals (Relay,
Pass, Low, High) are ready. There are
up to 100 groups of output signals.
6 Low Low signal. Indicates a LO compare Out
result.
7 Pass Pass signal. Indicates an IN compare Out
result (pass).
8 VINT Internal DC Voltage +5V. Out
19 VEXT External DC Voltage, acceptable In
range is +5V.
20 Relay Controls the relay output. Out
22 STRB After all (up to 100) output groups Out
are ready, the STRB signal will pulse
high.
For backwards compatibility with the GOM-802 scanner interface, please see
page 89.

83
GOM-804 & GOM-805 User Manual

Scan Setup

Background The Scan function sequentially scans up to 100 channels


and grades the resistance of the DUT on each channel to
a reference value. An automated handler or test fixture is
required to interface the DUTs to the measurement
terminals and the scan interface that controls the timing
of each scan.

Scan I/O

GOM-805

Measurement terminals

Source- S- S+ Source+
Automated Handler/Test Fixture
Channel 1 2 3 4 5 6 ····· N

DUT 1~N

Note: The automated handler/test fixture is


user-supplied. Please see your distributor for support and
technical details.
Grading of each DUT is essentially the same as the
compare function (page 41), the difference being the
Scan function will compare up to 100 DUTs sequentially,
whereas the Compare function will compare only one
DUT at a time.
The scan function compares a measured value to a
“Reference” value that has an upper (HI) and lower (LO)
limit. If the measured value is within the upper and lower
limit, then the measured value is judged as IN.
There are two modes that can be used to make a
judgment: ABS and △% modes.
The ABS mode compares the measured value to the
upper (HI) and lower (LO) limits. The upper and lower
limits are set as absolute resistance values.
The △% compare function compares the deviation of

84
HANDLER/SCAN INTERFACE

the measured value from the reference value as a


percentage.
{ [(Measured Value-Reference)/Reference]%}.
A measured value that falls within the upper and lower
limits is considered IN (pass), a value that falls below the
lower limits is considered LO, and a value that falls over
the upper limit is a HI.
Lower Reference Upper
limit value limit

LO IN HI
For both scan modes, the IN, HI or LO will be shown on
the display for each judgment (if the time between each
judgment is not too fast).
Display Overview Scan function
Ready to start scan message
indicator

Change
display view

Reference, limits, scan mode, current


channel, measurement delay

1. Select the Press Scan Scan to access the scan mode, as shown
Scan function above.

2. Select the Use the arrow keys to navigate to the Mode setting. Press
compare mode the Enter key to toggle the compare mode.

Move

Enter Toggle
Mode

Range Abs, △ %

85
GOM-804 & GOM-805 User Manual

3. Channel The Channel setting sets the number of DUT channels


setting that are used.
Use the arrow keys to navigate to the Channel setting and
press Enter.
Use the left and right arrow keys to select a digit. Use the
up and down arrow keys to edit the value of the selected
digit. Press Enter to confirm the setting.

Move
and edit

Enter
Select and
Channel setting confirm

Channel Range: 01 ~100

4. Delay setting The Delay setting adds a pause between each channel
measurement.
The Use the arrow keys to navigate to the Delay setting
and press Enter.
Use the left and right arrow keys to select a digit. Use the
up and down arrow keys to edit the value of the selected
digit. Press Enter to confirm the setting.

Move
and edit

Enter
Select and
Delay setting confirm

Delay Range: 400ms ~ 30000ms

5. Start the scan. Press the key or input a pulse signal on the
Trigger

Trigger pin of the SCAN interface port to start a scan


test.

See page 64 to set the external trigger edge as a rising or


Note falling leading edge.

The results will be displayed on the screen as each test is


performed. The results will also be output through the
scan port until the scan has finished.

86
HANDLER/SCAN INTERFACE

Scan function
Measured value
indicator

Pass/Fail
judgment

Change
display view

Reference, limits, scan mode, current


channel, measurement delay

6. View Results After the last SCAN test has


finished, press the ESC key
so that the menu system at the View display
bottom of the display has focus. results

Go to View and press Enter to Move


view the results of each channel.
Enter Select

Use the Previous and Next soft-keys to view each page.


Use the Back soft-key to return to the previous window.
Measurement
Channel number
results

Pass/High/Low
(P/H/L)
judgment results

Return to
the
previous
window
Previous Next results
results page page

87
GOM-804 & GOM-805 User Manual

Scan Output

Background The timing diagrams for the scan output under different
conditions are shown below.
Ready message displayed… After the manual trigger key is
pressed….
Relay Relay
Pass Pass
Low Low
High High
Clock Clock
100 99 98 97 2 1 100 99 2 1
STRB STRB

Scan channel 1. Delay time has Scan channel n. Delay time has
elapsed. elapsed.
Relay Relay
Pass Pass
Low Low
High High
Clock Clock
100 99 2 1 100 99 98 n+1 n 1
STRB STRB

Scan Channel 100. Delay time has Scan output signal timing.
elapsed.
Relay Data
Pass Pass
Low STRB 20us
High 20us
168us 38us
Clock
100 99 98 97 2 1
STRB

88
HANDLER/SCAN INTERFACE

GOM-802 Compatibility for Scan and Handler Interfaces


As the handler interface on GOM-802 is a 9-pin D-sub and the GOM-805 is a
25-pin D-sub, the GOM-805 handler interface cannot be used with existing
GOM-802 ATE equipment or environments without modification.
For backwards compatibility with the GOM-802 handler interface, please refer
to the chart below:

GOM-805 to GOM-802 Handler/Scan Interface


GOM-805 Handler Interface GOM-802 Handler Interface
Pin Handler Scan Pin Handler Scan
1, 17 Reserved Reserved
2 Trigger Trigger → 3 Start NC
3, 14, GND GND → 2 GND GND
18
4 Fail High → 7 Fail High
5 High Clock → 8 High Clock
6 Pass Low → 6 Pass Low
7 EOT Pass → 5 EOT Pass
8 VINT +5V → 1 +5V +5V
9 Bin1
10 Bin2
11 Bin3
12 Bin4
13 Bin5
15 Userdefine2
16 Userdefine1
19 VEXT VEXT
20 Ready Relay → 4 Ready Relay
21 Bin6
22 Low STRB → 9 Low STRB
23 Bin7
24 Bin8
25 Bin Out

89
GOM-804 & GOM-805 User Manual

Configure Interface
Overview The RS-232 and USB interfaces are standard for all
models, however the GPIB interface is only applicable
for the GOM-804G and GOM-805. The remote control
interfaces allow the GOM-804/805 to be programmed
for automatic testing.
For more information on remote control programming,
please see the Command Overview chapter on page 102.

Interface USB USB HOST


RS-232 DB-9 male port
GPIB 24 pin female GPIB port (GOM-804G,
GOM-805 only)

Configure USB Interface

Background The Type B USB port on the rear panel is used for
remote control. This interface creates a virtual COM port
when connected to a PC.
The USB interface requires the USB driver to be
Note installed. See page 91 to install the USB driver.

1. Connect and Configure the interface to USB in Page 71


configure to System>Utility>Interface menu.
USB.
Connect the Type A-B USB cable to the
rear panel USB B port on the
GOM-804/805.

Connect the other end to the Type A port


on the PC.

90
HANDLER/SCAN INTERFACE

Install USB Driver

Background The USB driver needs to be installed when using the USB
port for remote control. The USB interface creates a
virtual COM port when connected to a PC.

1. Select the USB Configure the interface to USB in Page 71


driver. System>Utility>Interface menu.
Connect the Type A-B USB cable to the
rear panel USB B port on the
GOM-804/805. Connect the other end to
the Type A port on the PC.
Go to the Windows Device Manager.
For Windows 7 go to:
Start Menu > Control Panel > Hardware and Sound >
Device Manager
The GOM-804/805 will appear as an unknown Virtual
Com Port under “Other Devices”.

Right-click Other Devices and select “Update Driver


Software”.
Select “Browse my computer for driver software” and
select the driver on the User Manual CD.
The GOM-805 and the COM port that it is assigned to
will now appear in under the Ports (COM & LPT) node.

91
GOM-804 & GOM-805 User Manual

Configure RS-232 Interface

Background The GOM-804/805 can also use an RS-232C connection


for remote control. When connecting to a PC ensure the
correct baud rate, parity, data bits, stop bit and data
control settings are used.
Settings Baud rate 1200, 2400, 4800, 9600, 19200, 38400,
57600, 115200
Parity None
Data bits 8
Stop bit 1
Data flow None
control

1. Select the Configure the interface to RS232 and Page 71


RS-232 baud rate set the baud rate in
System>Utility>Interface menu.
Connect the RS-232C cable to the rear RS232

panel RS232 port.

RS-232 pin Pin 2: RxD 1 5


assignment Pin 3: TxD
Pin 5: GND 6 9
Pin 1, 4, 6 ~ 9: No Connection

PC – GOM The RS232C connection uses a Null-modem connection,


RS-232C in which transmit (TxD) and receive (RxD) lines are
connection cross-linked.
GOM PC

Pin2 RxD RxD Pin2


Pin3 TxD TxD Pin3

Pin5 GND GND Pin5

92
HANDLER/SCAN INTERFACE

Configure GPIB Interface

Background The GPIB interface is SCPI-1994, IEEE488.1 and


IEEE488.2 compliant.
The GPIB interface is only available on the GOM-804G
Note and GOM-805.

1. Select the Configure the interface to GPIB and set Page 71


GPIB address the GPIB address in
System>Utility>Interface menu.
Connect one end of the GPIB

GPIB cable to the computer


and the other end to the
GPIB port on the
GOM-805.

RS232/USB Function Check

Operation Invoke a terminal application such as Realterm.


For RS-232, set the COM port, baud rate, stop bit, data
bit and parity accordingly.
To check the COM settings in Windows, see the Device
Manager. For example, in WinXP go to the Control panel
→ System → Hardware tab.
Run this query from the terminal.
*idn?
This should return the Manufacturer, Model number, and
Firmware version.
GWINSTEK,GOM805,GXXXXXXXX,V1.00
If you are not familiar with using a terminal application
Note to send/receive remote commands from the serial port or
via a USB connection, please page 94 (Using Realterm to
Establish a Remote Connection) for more information.

93
GOM-804 & GOM-805 User Manual

Using Realterm to Establish a Remote Connection

Background Realterm is a terminal program that can be used to


communicate with a device attached to the serial port of
a PC or via an emulated serial port via USB.
The following instructions apply to version 2.0.0.70.
Even though Realterm is used as an example to establish
a remote connection, any terminal program can be used
that has similar functionality.
Realterm can be downloaded on Sourceforge.net free of
Note charge.
For more information please see
http://realterm.sourceforge.net/

1. Install Download Realterm and install according to the


Realterm instructions on the Realterm website.

2. Configure Connect the GOM-804/805 via USB (page 90) or via


connection RS232 (page 92).
If using RS232, make note of the configured baud rate.
Go to the Windows device manager and find the COM
port number for the connection.
For example in Windows 7, go to the Start menu >
Control Panel > Hardware and Sound > Device Manager
Double click the Ports icon to reveal the connected serial
port devices and the COM port for each connected
device.

If using USB, the baud rate, stop bit and parity settings
can be viewed by right-clicking connected device and
selecting the Properties option.

94
HANDLER/SCAN INTERFACE

2. Run Realterm Start Realterm on the PC as an administrator.


Click:
Start menu>All Programs>RealTerm>realterm
Tip: to run as an administrator, you can right click the
Realterm icon in the Windows Start menu and select the
Run as Administrator option.
After Realterm has started, click on the Port tab.
Enter the Baud, Parity, Data bits, Stop bits and Port
number configuration for the connection.
The Hardware Flow Control and Software Flow Control
options can be left at the default settings.
Press Open to connect to the GOM-804/805.

95
GOM-804 & GOM-805 User Manual

3. Test remote Click on the Send tab.


command In the EOL configuration, check on the +CR and +LF
check boxes.
Enter the query:
*idn?
Click on Send ASCII.

The terminal display will return the following:


GWINSTEK,GOM805,GXXXXXXXX,V1.00
(manufacturer, model, serial number, version)

4. Errors or If Realterm fails to connect to the GOM-804/805, please


Problems check all the cables and settings and try again.

GPIB Function

Background Please use the National Instruments Measurement &


Automation Controller software to confirm GPIB/LAN
functionality.
See the National Instrument website, http://www.ni.com
for details.

1. Operation Start the NI Measurement and


Automation Explorer (MAX) program.
Using Windows, press:

96
HANDLER/SCAN INTERFACE

Start>All Programs>National
Instruments>Measurement & Automation

Step a. From the Configuration panel access;


My System>Devices and Interfaces>GPIB0
Step b. Press the Scan for Instruments button.

Step c. In the Connected Instruments panel the GOM-804/805


should be detected as Instrument 0 with the address the
same as that configured on the unit.
Step d. Double click the Instrument 0 icon.

b
a

c
d

Step e. Click on the Attributes tab at the bottom.

Step f. Click on Communicate with Instrument.

97
GOM-804 & GOM-805 User Manual

Step g. In the NI-488.2 Communicator window, ensure *IND? is


written in the Send String: text box.
Click on the Query button to send the *IDN? query to
the instrument.
Step h. The String Received text box will display the query
return:
GWINSTEK,GOM805,GXXXXXXXX,V1.00
(manufacturer, model, serial number, version)

g
h

e
The function check is complete.

98
SAVE/RECALL

SAVE/RECALL
The settings for all the major functions can be saved and recalled from 20
memory slots.

Settings can saved/recalled for the following functions:


Ohm, Compare, Binning, TC, TCONV, TEMP, Scan, Diode.

Save/Recall Settings

Background The save function saves the current function as well the
settings related to that function.
There are 20 memory slots that can be used to save and
recall settings on the GOM-804/805.

1. Enter the When you are in the desired function mode, press the
Memory menu ESC key (if necessary) to so that the menu system at
the bottom of the display has focus.
Use the arrow keys to navigate to the Memory setting
and press Enter.

Function mode

Memory setting
Move

Enter
Select menu
or setting

99
GOM-804 & GOM-805 User Manual

The Recall/Save Setup menu will appear.

2. Save/ The No. setting should be already highlighted when


Recall/Clear entering the Recall/Save Setup menu. If not, use the
Memory Left/Right arrow keys to highlight the No. setting.
No. setting Recall, Save, Clear settings

Move Enter
Select and
and edit confirm

Use the up and down arrow keys to select a memory


space.
Range 01~20
*If a memory space has been used before, the
settings for that memory slot will also be shown
on the display.

To Save:
Use the arrow keys to go to Save
and press Enter.

To Recall:
Use the arrow keys to go to Recall
and press Enter.

To Clear:
Use the arrow keys to go to Clear
and press Enter.

100
SAVE/RECALL

Press Enter again when asked to confirm the selected


operation.
After saving the settings, press ESC to return to the
current function mode.
After recalling settings, the unit will automatically go to
the recalled setting function.

Pressing ESC before pressing Enter will exit the


Note Save/Recall/Clear operation.

View memory Press the Enter key when the No. setting is highlighted to
slot availability see which memory slots are empty.
The status of memory slots 01 ~ 20 are shown at the
bottom of the display.
Memory slots in red are empty slots while those in black
have already been used.
Press Enter again to exit from this view.

Enter Press Enter Settings in


selected
memory slot

Available memory slots in red.


Used memory slots in black.

The memory number can also be selected when in the


Note above view using the arrow keys.

101
GOM-804 & GOM-805 User Manual

COMMAND OVERVIEW
The Command overview chapter lists all the programming commands in
alphabetical order. The command syntax section shows you the basic syntax
rules you have to apply when using commands.

Command Syntax
Compatible IEEE488.2 Partial compatibility
Standard
SCPI, 1994 Partial compatibility

Command SCPI (Standard Commands for Programmable


Structure Instruments) commands follow a tree-like structure,
organized into nodes. Each level of the command tree is
a node. Each keyword in an SCPI command represents
each node in the command tree. Each keyword (node) of
an SCPI command is separated by a colon (:).
For example, the diagram below shows an SCPI
sub-structure and a command example.

Command Types There are a number of different instrument commands


and queries. A command sends instructions or data to the
unit and a query receives data or status information from
the unit.
Command Types

Simple A single command with/without a


parameter
Example SENSe:FUNCtion OHM

102
COMMAND OVERVIEW

Query A query is a simple or compound


command followed by a question mark
(?). A parameter (data) is returned.
Example SENSe:RANGe?

Command Forms Commands and queries have two different forms, long
and short. The command syntax is written with the short
form of the command in capitals and the remainder
(long form) in lower case.
The commands can be written either in capitals or
lower-case, just so long as the short or long forms are
complete. An incomplete command will not be
recognized.
Below are examples of correctly written commands.

Long form CALCulate:COMPare:BEEPer


CACLULATE:COMPARE:BEEPER
calculate:compare:beeper

Short form CALC:COMP:BEEP


calc:comp:beep

Command
Format

1. Command header
2. Space
3. Parameter

Common Input Type Description Example


Parameters
<Boolean> boolean logic 0,1
<NR1> integers 0,1,2,3
<NR2> decimal numbers 0.1,3.14,8.5
<NR3> floating point with 4.5e-1,8.25e+1
exponent

103
GOM-804 & GOM-805 User Manual

<NRf> Any of NR1,2,3 1,1.5,4.5e-1


<string> ASCII text string TEST_NAME

Message Remote Marks the end of a command line. The


Terminator (EOL) Command following messages are in accordance
with IEEE488.2 standard.
LF, CR, CR+LF, The most common
LF+CR EOL character is
CR+LF

104
COMMAND OVERVIEW

Command List

Binning Commands
BINNing:COUNt:CLEar ....................................................................... 108
BINNing:COUNt:TOTal ....................................................................... 108
BINNing:COUNt:OUT ......................................................................... 108
BINNing<X>:COUNt:RESult ................................................................ 108
BINNing<X>:LIMit:LOWer ................................................................... 109
BINNing<X>:LIMit:UPPer .................................................................... 109
BINNing<X>:PERCent:LOWer .............................................................. 110
BINNing<X>:PERCent:UPPer ............................................................... 110
BINNing:LIMit:BEEPer ........................................................................ 111
BINNing:LIMit:DISPlay ....................................................................... 111
BINNing:LIMit:MODE ......................................................................... 111
BINNing:LIMit:REFerence ................................................................... 112
BINNing:LIMit:RESult ......................................................................... 112

Calculate Commands
CALCulate:COMPare:BEEPer ............................................................... 113
CALCulate:COMPare:LIMit:LOWer ...................................................... 113
CALCulate:COMPare:LIMit:MODE ....................................................... 114
CALCulate:COMPare:LIMit:REFerence ................................................. 114
CALCulate:COMPare:LIMit:RESult ....................................................... 115
CALCulate:COMPare:LIMit:UPPer ....................................................... 115
CALCulate:COMPare:MATH:DATa ........................................................ 115
CALCulate:COMPare:PERCent:LOWer .................................................. 116
CALCulate:COMPare:PERCent:UPPer .................................................. 116
CALCulate:SCAN:CHANnel.................................................................. 116
CALCulate:SCAN:DELay ...................................................................... 117
CALCulate:SCAN:LIMit:LOWer ............................................................ 117
CALCulate:SCAN:LIMit:MODE ............................................................ 117
CALCulate:SCAN:LIMit:REFerence ....................................................... 118
CALCulate:SCAN:LIMit:UPPer ............................................................. 118
CALCulate:SCAN:PERCent:LOWer ....................................................... 119
CALCulate:SCAN:PERCent:UPPer ........................................................ 119

Memory Commands
MEMory:CLEar .................................................................................... 120
MEMory:RECall ................................................................................... 120
MEMory:SAVe ...................................................................................... 120
MEMory:STATe .................................................................................... 120

105
GOM-804 & GOM-805 User Manual

Sense Commands
SENSe:AUTo ....................................................................................... 122
SENSe:DISPlay .................................................................................... 122
SENSe:FUNCtion ................................................................................ 122
SENSe:RANGe .................................................................................... 123
SENSe:SPEed ...................................................................................... 123
SENSe:REL:DATa ................................................................................. 124
SENSe:REL:STATe ................................................................................ 124
SENSe:REALtime:STATe ....................................................................... 124

Source Commands
SOURce:DRY ....................................................................................... 126
SOURce:DRIVe .................................................................................... 126

Status Commands
STATus:PRESet .................................................................................... 127
STATus:QUEStionable:ENABle ............................................................. 127
STATus:QUEStionable:EVENt .............................................................. 127

System Commands
SYSTem:AVERage:DATa ....................................................................... 128
SYSTem:AVERage:STATe ...................................................................... 128
SYSTem:BRIGhtness ........................................................................... 128
SYSTem:ERRor .................................................................................... 129
SYSTem:HANDler ............................................................................... 129
SYSTem:KEYClick:BEEPer .................................................................... 129
SYSTem:LFRequency ........................................................................... 130
SYSTem:LOCal .................................................................................... 130
SYSTem:MDELay:DATa ........................................................................ 130
SYSTem:MDELay:STATe ....................................................................... 131
SYSTem:PWM:ON ............................................................................... 131
SYSTem:PWM:OFF .............................................................................. 132
SYSTem:SERial .................................................................................... 132
SYSTem:VERSion ................................................................................. 132

Temperature Commands
TEMPerature:AMBient:DATa ................................................................ 133
TEMPerature:AMBient:STATe .............................................................. 133
TEMPerature:COMPensate:COEFficient .............................................. 134
TEMPerature:COMPensate:CORRect ................................................... 134
TEMPerature:CONVersion:CONStant .................................................. 134
TEMPerature:CONVersion:DISPlay ...................................................... 135
TEMPerature:CONVersion:MATH:DATa ............................................... 135

106
COMMAND OVERVIEW

TEMPerature:CONVersion:RESistance ................................................. 135


TEMPerature:CONVersion:TEMPerature .............................................. 136
TEMPerature:DATa .............................................................................. 136
TEMPerature:STATe ............................................................................. 136
TEMPerature:UNIT .............................................................................. 137

Trigger Commands
READ................................................................................................... 138
MEASure<X> ....................................................................................... 138
SHOW ................................................................................................. 138
TRIGger:EDGE..................................................................................... 139
TRIGger:DELay:DATa ........................................................................... 139
TRIGger:DELay:STATe .......................................................................... 139
TRIGger:SOURce ................................................................................. 140

Userdefine Commands
USERdefine<X>:ACTive ....................................................................... 141
USERdefine<X>:FIRStdata ................................................................... 141
USERdefine<X>:LOGic ........................................................................ 141
USERdefine<X>:SEConddata ............................................................... 142

Common Commands
*CLS.................................................................................................... 143
*ESE .................................................................................................... 143
*ESR ................................................................................................... 143
*IDN ................................................................................................... 144
*OPC ................................................................................................... 144
*RST ................................................................................................... 144
*SRE ................................................................................................... 144
*STB ................................................................................................... 145
*TRG ................................................................................................... 145

107
GOM-804 & GOM-805 User Manual

BINNing Commands

BINNing:COUNt:CLEar Set
Query
Description Clear all bin sorting function test result counts.
Syntax BINNing:COUNt:CLEar
Parameter/ <None>

BINNing:COUNt:TOTal Query

Description Returns the total number (count total) of test bin results.
Query Syntax BINNing:COUNt:TOTal?
Return parameter <NR1> 0~999999999
Example BINN:COUN:TOT?
>150
Indicates that the total number (count total) of test
results (pass and fail) is 150.

BINNing:COUNt:OUT Query

Description Returns the number of failed (judged OUT) test results


for the bin sorting function test.
Query Syntax BINNing:COUNt:OUT?
Return parameter <NR1> 0~99999999
Example BINN:COUN:OUT?
>50
Indicates that the number of failed test results is 50.

BINNing<X>:COUNt:RESult Query

Description Returns the number of passed (judged IN) test results for
the selected bin.
Query Syntax BINNing<X>:COUNt:RESult?
Parameter <X> 1~8

108
COMMAND OVERVIEW

Return parameter <NR1> 0~99999999


Example BINN1:COUN:RES?
>100
Indicates that bin1 has a pass count of 100.
Set
BINNing<X>:LIMit:LOWer Query

Description Sets or returns the lower limit value (absolute value) for
the selected bin.
Syntax BINNing<X>:LIMit:LOWer {<NRf>[,<String>]}
Query Syntax BINNing<X>:LIMit:LOWer?
Parameter <X> 1~8
<NRf> 000.0000~999.9999
<String> mohm/ohm/kohm/maohm,unit
If the unit is not set, the unit will be
automatically set by the present range.
Return parameter <NR3> 000.0000~999.9999E± X
Example BINN1:LIM:LOW 23.8,kohm
Sets the bin1 lower limit value to 23.8kΩ.
BINN1:LIM:LOW?
>23.8000E+3
Returns the lower limit as 23.8kΩ.
Set
BINNing<X>:LIMit:UPPer Query

Description Sets or returns the upper limit value (absolute value) for
the selected bin.
Syntax BINNing<X>:LIMit:UPPer {<NRf>[,<String>]}
Query Syntax BINNing<X>:LIMit:UPPer?
Parameter <X> 1~8
<NRf> 000.0000~999.9999
<String> mohm/ohm/kohm/maohm,unit
If the unit is not set, the unit will be
automatically set by the present range.
Return parameter <NR3> 000.0000~999.9999E± X

109
GOM-804 & GOM-805 User Manual

Example BINN1:LIM:UPP 0.95,maohm


Sets bin1 upper limit value to 0.95MΩ.
BINN1:LIM:UPP?
>0.9500E+6
Returns the upper limt as 0.95MΩ.
Set
BINNing<X>:PERCent:LOWer Query

Description Sets or returns the lower value percentage value for the
selected bin. The value is a percentage offset from the
reference value.
Syntax BINNing<X>:PERCent:LOWer <NRf>
Query Syntax BINNing<X>:PERCent:LOWer?
Parameter <X> 1~8
<NRf> 000.00~999.99
Return parameter <NR2> 000.00~999.99
Example BINN1:PERC:LOW 10.15
Sets the bin1 lower limit percent value to -10.15%.
BINN1: PERC:LOW?
>10.15
Returns the lower limit percentage value as -10.15%.
Set
BINNing<X>:PERCent:UPPer Query

Description Sets or returns the upper value percentage value for the
selected bin. The value is a percentage offset from the
reference value.
Syntax BINNing<X>:PERCent:UPPer <NRf>
Query Syntax BINNing<X>:PERCent:UPPer?
Parameter <X> 1~8
<NRf> 000.00~999.99
Return parameter <NR2> 000.00~999.99
Example BINN1:PERC:UPP 150.95
Sets the bin1 upper limit percent value to +150.95%.
BINN1:LIM:UPP?
>150.95
Returns the upper limit percentage value as +150.95%.

110
COMMAND OVERVIEW

Set
BINNing:LIMit:BEEPer Query

Description Sets or returns beeper mode for the bin sorting function.
Syntax BINNing:LIMit:BEEPer {OFF|PASS|FAIL}
Query Syntax BINNing:LIMit:BEEPer?
Parameter/ OFF Turns the beeper off.
Return parameter PASS The beeper will sound on a pass test
result.
FAIL The beeper will sound on a fail test
result.
Example BINN:LIM:BEEP OFF
Turns the beeper off.
Set
BINNing:LIMit:DISPlay Query

Description Sets or returns the bin sorting function display mode.


Syntax BINNing:LIMit:DISPlay {COMP|COUNT}
Query Syntax BINNing:LIMit:DISPlay?
Parameter/ COMP The display is set to compare mode.
Return parameter COUNT The display is set to count mode.
Example BINN:LIM:DISP COMP
Sets the bin sorting function display mode to compare.
Set
BINNing:LIMit:MODE Query

Description Sets or returns the setting mode for upper and lower
limits (absolute or Δ%).
Syntax BINNing:LIMit:MODE {ABS|DPER}
Query Syntax BINNing:LIMit:DISP?
Parameter/ ABS The test results are judged from
Return parameter absolute values.
DPER The test results are judged from a
reference value ± a percentage offset.
(delta percent)

111
GOM-804 & GOM-805 User Manual

Example BINN:LIM:DISP DPER


Sets the mode toΔ %.
Set
BINNing:LIMit:REFerence Query

Description Sets or returns the limit reference value for the bin
sorting function.
Syntax BINNing<X>:LIMit:REFerence {<NRf>[,<String>]}
Query Syntax BINNing<X>:LIMit:REFerence?
Parameter <NRf> 000.0001~999.9999
<String> mohm/ohm/kohm/maohm,unit
If the unit is not set, the unit will be
automatically set by the present range.
Return parameter <NR3> 000.0001~999.9999E± X
Example BINN:LIM:REF 100
Sets the limit reference value to 100Ω.
BINN:LIM:REF?
>100.0000E+0
Returns the reference as 100Ω.

BINNing:LIMit:RESult Query

Description Returns the bin sorting function test result.


Query Syntax BINNing:LIMit:RESult?
Return parameter <NR1> 1~8: Bin1~Bin8
9: Bin Out
Example BINN:LIMit:RES?
>1
Indicates a pass for bin1.

112
COMMAND OVERVIEW

Calculate Commands
Set
CALCulate:COMPare:BEEPer Query

Description Sets or returns the compare function beeper mode.


Syntax CALCulate:COMPare:BEEPer {OFF|PASS|FAIL}
Query Syntax CALCulate:COMPare:BEEPer?
Parameter/ OFF Turns the beeper off.
Return parameter PASS The beeper will sound on a pass test
result.
FAIL The beeper will sound on a fail test
result.
Example CALC:COMP:BEEP FAIL
Sets the beeper on when the test result is a fail.
Set
CALCulate:COMPare:LIMit:LOWer Query

Description Sets or returns the lower limit value for the compare
function.
Syntax CALCulate:COMPare:LIMit:LOWer {<NRf>[,<String>]}
Query Syntax CALCulate:COMPare:LIMit:LOWer?
Parameter <NRf> 000.0000~999.9999
<String> mohm/ohm/kohm/maohm,unit
If the unit is not set, the unit will be
automatically set by the present range.
Return parameter <NR3> 000.0000~999.9999E± X
Example CALC:COMP:LIM:LOW 0.123,maohm
Sets the lower limit value to 0.123MΩ.
CALC:COMP:LIM:LOW?
>0.1230E+6
Returns the lower limit as 0.123MΩ.

113
GOM-804 & GOM-805 User Manual

Set
CALCulate:COMPare:LIMit:MODE Query

Description Sets or returns the compare mode for the compare


function.
Syntax CALCuate:COMPare:LIMit:MODE {ABS|DPER|PER}
Query Syntax CALCulate:COMPare:LIMit:MODE?
Parameter/ ABS The test results are judged from
Return parameter absolute values.
DPER The test results are judged from a
reference value ± a percentage offset.
(delta percentage)
PER The test results are displayed as a
percentage of the reference value.
Example CALC:COMP:LIM:MODE ABS
Sets test results as absolute values for the compare
function.
Set
CALCulate:COMPare:LIMit:REFerence Query

Description Sets or returns the limit reference value for the compare
function.
Syntax CALCulate:COMPare:LIMit:REF {<NRf>[,<String>]}
Query Syntax CALCulate:COMPare:LIMit:REF?
parameter <NRf> 000.0001~999.9999
<String> mohm/ohm/kohm/maohm,unit
If unit is not set, the unit will be
automatically set by the present range.
Return parameter <NR3> 000.0001~999.9999E± X
Example CALC:COMP:LIM:REF 10.00,mohm
Sets the limit reference value to 10.00mΩ.
CALC:COMP:LIM:REF?
>10.0000E-3
Returns the limit as 10.00mΩ.

114
COMMAND OVERVIEW

CALCulate:COMPare:LIMit:RESult Query

Description Returns the compare function test result.


Query Syntax CALCulate:COMPare:LIMit:RESult?
Return parameter <NR1> 0: LO
1: IN
2: HI
Example BINN:LIMit:RES?
>2
Indicates that the test result is HI.
Set
CALCulate:COMPare:LIMit:UPPer Query

Description Sets or returns the upper limit value for the compare
function.
Syntax CALCulate:COMPare:LIMit:UPPer {<NRf>[,<String>]}
Query Syntax CALCulate:COMPare:LIMit:UPPer?
Parameter <NRf> 000.0000~999.9999
<String> mohm/ohm/kohm/maohm,unit
If unit is not set, the unit will be
automatically set by the present range.
Return parameter <NR3> 000.0000~999.9999E± X
Example CALC:COMP:LIM:UPP 0.95,kohm
Sets the upper limit value to 0.95kΩ.
CALC:COMP:LIM:UPP?
>0.9500E+3
Returns the upper limit as 0.95kΩ.

CALCulate:COMPare:MATH:DATa Query

Description Returns the deviation value for the compare function.


Query Syntax CALCulate:COMPare:MATH:DATa?
Return parameter <NR3> ±0.0000~9.9999E± X.
Example CALC:COMP:MATH:DAT?
>+0.3658E+2
Returns the deviation as 36.58%.

115
GOM-804 & GOM-805 User Manual

Set
CALCulate:COMPare:PERCent:LOWer Query

Description Sets or returns the lower limit percent value for the
compare function.
Syntax CALCulate:COMPare:PERCent:LOWer <NRf>
Query Syntax CALCulate:COMPare:PERCent:LOWer?
Parameter <NRf> 000.00~999.99
Return parameter <NR2> 000.00~999.99
Example CALC:COMP:PERC:LOW 10.00
Sets the lower limit percent value to -10.00%.
CALC:COMP:PERC:LOW?
>10.00
Returns the lower limit as -10.00%.
Set
CALCulate:COMPare:PERCent:UPPer Query

Description Sets or returns the upper limit percent value for the
compare function.
Syntax CALCulate:COMPare:PERCent:UPPer <NRf>
Query Syntax CALCulate:COMPare:PERCent:UPPer?
Parameter <NRf> 000.00~999.99
Return parameter <NR2> 000.00~999.99
Example CALC:COMP:PERC:UPP 90.00
Sets the upper limit percent value to +90.00%.
CALC:COMP:PERC:UPP?
>90.00
Returns the upper limit as +90.00%.
Set
CALCulate:SCAN:CHANnel Query

Description Sets or returns the channel for the scan function.


Syntax CALCulate:SCAN:CHANnel <NR1>
Query Syntax CALCulate:SCAN:CHANnel?
Parameter/ <NR1> 1~100
Return parameter

116
COMMAND OVERVIEW

Example CALC:SCAN:CHAN 5
Sets the channel to 5.
Set
CALCulate:SCAN:DELay Query

Description Sets or returns the interval delay for the scan function.
Syntax CALCulate:SCAN:DELay <NR1>
Query Syntax CALCulate:SCAN:DELay?
Parameter/ <NR1> 400~30000
Return parameter Unit:ms
Example CALC:SCAN:DEL 500
Sets interval delay of the scan to 500ms.
Set
CALCulate:SCAN:LIMit:LOWer Query

Description Sets or returns the lower limit value for the scan function.
Syntax CALCulate:SCAN:LIMit:LOWer {<NRf>[,<String>]}
Query Syntax CALCulate:SCAN:LIMit:LOWer?
Parameter <NRf> 000.0000~999.9999
<String> mohm/ohm/kohm/maohm,unit
If unit is not set, the unit will be
automatically set by the present range.
Return parameter <NR3> 000.0000~999.9999E± X
Example CALC:SCAN:LIM:LOW 0.123,maohm
Sets the lower limit value to 0.123MΩ.
CALC:SCAN:LIM:LOW?
>0.1230E+6
Returns the lower limit as 0.123MΩ.
Set
CALCulate:SCAN:LIMit:MODE Query

Description Sets or returns the scan function compare mode.


Syntax CALCulate:SCAN:LIMit:MODE {ABS|DPER}
Query Syntax CALCulate:SCAN:LIMit:MODE?
Parameter/ ABS The test results are judged from
Return parameter absolute values.

117
GOM-804 & GOM-805 User Manual

DPER The test results are judged from a


reference value ± a percentage offset.
(delta percent)
Example CALC:SCAN:LIM:MODE ABS
Sets compare mode to absolute values.
Set
CALCulate:SCAN:LIMit:REFerence Query

Description Sets or returns the reference limit for the scan function.
Syntax CALCulate:SCAN:LIMit:REFerence {<NRf>[,<String>]}
Query Syntax CALCulate:SCAN:LIMit:REFerence?
Parameter <NRf> 000.0001~999.9999
<String> mohm/ohm/kohm/maohm,unit
If unit is not set,the unit will be
automatically set by the present range.
Return parameter <NR3> 000.0001~999.9999E± X
Example CALC:SCAN:LIM:REF 10.00,mohm
Sets the reference limit to 10.00mΩ.
CALC:SCAN:LIM:REF?
>10.0000E-3
Returns the reference limit as 10.00mΩ.
Set
CALCulate:SCAN:LIMit:UPPer Query

Description Sets or returns upper limit of the scan function.


Syntax CALCulate:SCAN:LIMit:UPPer {<NRf>[,<String>]}
Query Syntax CALCulate:SCAN:LIMit:UPPer?
Parameter <NRf> 000.0000~999.9999
<String> mohm/ohm/kohm/maohm,unit
If unit is not set, the unit will be
automatically set by the present range.
Return parameter <NR3> 000.0000~999.9999E± X
Example CALC:SCAN:LIM:UPP 1.37,kohm
Sets the upper limit to 1.37kΩ.
CALC:SCAN:LIM:UPP?
>1.3700E+3
Returns the upper limit as 1.37kΩ.

118
COMMAND OVERVIEW

Set
CALCulate:SCAN:PERCent:LOWer Query

Description Sets or returns lower limit percent value for the scan
function.
Syntax CALCulate:SCAN:PERCent:LOWer <NRf>
Query Syntax CALCulate:SCAN:PERCent:LOWer?
Parameter <NRf> 000.00~999.99
Return parameter <NR2> 000.00~999.99
Example CALC:SCAN:PERC:LOW 10.00
Sets the lower limit percent value to -10.00%.
CALC:SCAN:PERC:LOW?
>10.00
Returns the lower limit as -10.00%.
Set
CALCulate:SCAN:PERCent:UPPer Query

Description Sets or returns the upper limit percent value for the scan
function.
Syntax CALCulate:SCAN:PERCent:UPPer <NRf>
Query Syntax CALCulate:SCAN:PERCent:UPPer?
Parameter <NRf> 000.00~999.99
Return parameter <NR2> 000.00~999.99
Example CALC:SCAN:PERC:UPP 90.00
Sets the upper limit percent value to +90.00%.
CALC:SCAN:PERC:UPP?
>90.00
Returns the upper limit as +90.00%.

119
GOM-804 & GOM-805 User Manual

Memory Commands

MEMory:CLEar Set
Query
Description Clears the data from the selected memory slot.
Syntax MEMory:CLEar <NR1>
Parameter <NR1> 1~20
Example MEM:CLE 1
Clear data from memory slot 1.

MEMory:RECall Set
Query
Description Recalls the settings from the selected memory slot.
Syntax MEMory:RECall <NR1>
Parameter <NR1> 1~20
Example MEM:REC 1
Recall the settings from memory slot 1.

MEMory:SAVe Set
Query
Description Saves the settings to the selected memory slot.
Syntax MEMory:SAVe <NR1>
Parameter <NR1> 1~20
Example MEM:SAV 1
Saves the settings to memory slot 1.

MEMory:STATe Query

Description Returns the status of all the memory slots.


Query Syntax MEMory:STATe?
Return parameter <String> 23 Characters composed of “N” or
“F”, where “N” indicates “Not used”
and “F” indicates “Full”.

120
COMMAND OVERVIEW

Example MEM:STAT?
> NFFNN-NNNNN-NNNNN-NNNNN
Indicates that memory slots 2 and 3 have data and that all
other memory slots are empty.

121
GOM-804 & GOM-805 User Manual

Sense Commands
Set
SENSe:AUTo Query

Description Sets or returns the auto-range state.


Syntax SENSe:AUTo <NR1> | {OFF|ON}
Query Syntax SENSe:AUTo?
Parameter/ <NR1> 0:OFF.
Return parameter 1:ON.
OFF Auto-Range is off.
ON Auto-Range is on.
Example SENS:AUT ON
Sets auto-range mode on.
Set
SENSe:DISPlay Query

Description Sets or returns the display mode. There are two display
modes, normal and simple.
Syntax SENSe:DISPlay <NR1> | {OFF|ON}
Query Syntax SENSe:DISPlay?
Parameter/ <NR1> 0:OFF.
Return parameter 1:ON.
OFF Display mode is normal.
ON Display mode is simple.
Example SENS:DISP OFF
Sets the display mode to normal.
Set
SENSe:FUNCtion Query

Description Sets or returns the function mode.


Syntax SENSe:FUNCtion
{OHM|COMP|BIN|TC|TCONV|SCAN|DIODE}
Query Syntax SENSe:FUNCtion?

122
COMMAND OVERVIEW

Parameter/ OHM OHM MODE


Return parameter COMP COMP MODE
BIN BIN MODE
TC TC MODE
TCONV TCONV MODE
SCAN SCAN MODE
DIODE DIODE MODE
Example SENS:FUNC OHM
Sets ohm mode on.
Set
SENSe:RANGe Query

Description Sets or returns the range of the present function.


Syntax SENSe:RANGe <NRf>
Query Syntax SENSe:RANGe?
Parameter <NRf> 5E-2 ~ 5E+6
Return parameter <NR3> 5E-2 ~ 5E+6
Example SENS:RANG 0.05
Sets range to 50mΩ.
SENS:RANG?
>5.0000E-2
Returns the range as 50mΩ.
Set
SENSe:SPEed Query

Description Sets or returns the measurement speed.


Syntax SENSe:SPEed {SLOW|FAST}
Query Syntax SENSe:SPEed?
Parameter/ SLOW Measurement speed is slow.
Return parameter FAST Measurement speed is fast.
Example SENS:SPE FAST
Sets measurement speed to the fast rate.

123
GOM-804 & GOM-805 User Manual

Set
SENSe:REL:DATa Query

Description Sets or returns the relative value for the relative function.
Syntax SENSe:REL:DATa <NRf>
Query Syntax SENSe:REL:DATa?
Parameter <NRf> 0.0000~500.00
The unit will be auto set by the present
range.
Return parameter <NR3> ± 0.0000~5.1000E± X
Example SENS:REL:DAT 490.32
Sets the relative function value to 490.32Ω.
SENS:REL:DAT?
>4.9032E+2
Returns the relative value (490.32Ω).
Set
SENSe:REL:STATe Query

Description Sets or returns the relative function state.


Syntax SENSe:REL:STATe <NR1> | {OFF|ON}
Query Syntax SENSe:REL:STATe?
Parameter/ <NR1> 0:OFF.
Return parameter 1:ON.
OFF Turn the relative function off.
ON Turn the relative function on.
Example SENS:REL:STAT OFF
Sets the relative function off.
Set
SENSe:REALtime:STATe Query

Description Sets or returns the real time function state.


Syntax SENSe:REALtime:STATe <NR1> | {OFF|ON}
Query Syntax SENSe:REALtime:STATe?

124
COMMAND OVERVIEW

Parameter/ <NR1> 0:OFF.


Return parameter 1:ON.
OFF Turn the real time function off.
ON Turn the real time function on.
Example SENS:REAL:STAT ON
Turns the real time function on.

125
GOM-804 & GOM-805 User Manual

Source Commands
Set
SOURce:DRY Query

Description Sets or returns the dry circuit test mode.


Syntax SOURce:DRY {<NR1> | {OFF|ON}
Query Syntax SOURce:DRY?
Parameter/ <NR1> 0:OFF.
Return parameter 1:ON.
OFF Turn dry circuit test mode off.
ON Turn dry circuit test mode on.
Example SOUR:DRY On
Turns the dry circuit test mode on.
Set
SOURce:DRIVe Query

Description Sets or returns the drive mode.


Syntax SOURce:DRIVe <NR1>
Query Syntax SOURce:DRIVe?
Parameter/ <NR1> 1: the DC+ mode.
Return parameter 2: the DC- mode.
3: the PULSE mode.
4: the PWM mode.
5: the ZERO mode.
Example SOURce:DRIVe 3
Sets the drive mode to pulse.

126
COMMAND OVERVIEW

Status Commands

STATus:PRESet Set
Query
Description Sets the QUESTionable enable register to zero.
Syntax STATus:PRESet <NONE>
Parameter <None>
Set
STATus:QUEStionable:ENABle Query

Description Sets or returns the Questionable Data Enable register.


Syntax STATus:QUEStionable:ENABle <NR1>
Query Syntax STATus:QUEStionable:ENABle?
Parameter/ <NR1> 0~32767.
Return parameter
Example STAT:QUES:ENAB 2560
Sets the Questionable Data Enable register to
000101000000000.

STATus:QUEStionable:EVENt Query

Description Returns the contents of the Questionable Data Event


register.
Query Syntax STATus:QUEStionable:EVENt?
Return parameter <NR1> 0~32767
Example STAT:QUES:EVEN?
>512
512 indicates that the Questionable Data Event
register=0000001000000000.

127
GOM-804 & GOM-805 User Manual

System Commands
Set
SYSTem:AVERage:DATa Query

Description Sets or returns the number of measurements used for the


average function.
Syntax SYSTem:AVERage:DATa <NR1>
Query Syntax SYSTem:AVERage:DATa?
Parameter/ <NR1> 2~10
Return parameter
Example SYST:AVER:DAT 5
5 measurements are used to perform the average
function.
Set
SYSTem:AVERage:STATe Query

Description Sets or returns the average function state.


Syntax SYSTem:AVERage:STATe <NR1> | {OFF|ON}
Query Syntax SYSTem:AVERage:STATe?
Parameter/ <NR1> 0:OFF.
Return parameter 1:ON.
OFF Turn the average function off.
ON Turn the average function on.
Example SYST:AVER:STAT OFF
Turns the average function off.
Set
SYSTem:BRIGhtness Query

Description Sets or returns the brightness level.


Syntax SYSTem:BRIGhtness <NR1>
Query Syntax SYSTem:BRIGhtness?
Parameter/ <NR1> 1(dim)~5(bright)
Return parameter

128
COMMAND OVERVIEW

Example SYST:BRIG 4
Turns the brightness level to 4.

SYSTem:ERRor Query

Description Returns the current system error, if any.


Query Syntax SYSTem:ERRor?
Return parameter <String> Error number,”Error message”
Example SYST:ERR?
>0,”No error”.
Indicates that there is no error message.
Set
SYSTem:HANDler Query

Description Sets or returns the handler state.


Syntax SYSTem:HANDler {CLEAR | HOLD}
Query Syntax SYSTem:HANDler?
Parameter/ Clear It clears the last result before executing
Return parameter measurement.
HOLD It holds the test result and changes
when a different result appears.
Example SYST:HAND HOLD
Sets the test result to the hold state.
Set
SYSTem:KEYClick:BEEPer Query

Description Sets or returns the keyclick beeper state.


Syntax SYSTem:KEYClick:BEEPer <NR1> | {OFF | ON}
Query Syntax SYSTem:KEYClick:BEEPer?
Parameter/ <NR1> 0:OFF.
Return parameter 1:ON.
OFF Turn the keyclick beeper off.
ON Turn the keyclick beeper on.
Example SYST:KEYC:BEEP OFF
Sets the keyclick beeper off.

129
GOM-804 & GOM-805 User Manual

Set
SYSTem:LFRequency Query

Description Sets or returns the frequency setting for the line filter.
Syntax SYSTem:LFRequency {AUTO | 50 | 60}
Query Syntax SYSTem:LFRequency?
Parameter/ AUTO The frequency setting for the line filter
Return parameter is automatically detected.
50 The frequency is 50Hz.
60 The frequency is 60Hz.
Example SYST:LFR 60
Sets the line frequency to 60Hz.
SYST:LFR?
>60Hz
Returns the line frequency as 60Hz.

SYSTem:LOCal Set
Query
Description Enables local control (front panel control) and disables
remote control.
Syntax SYSTem:LOCal
Parameter <None>
Set
SYSTem:MDELay:DATa Query

Description Sets or returns the measurement delay time.


Syntax SYSTem:MDELay:DATa <NRf>
Query Syntax SYSTem:MDELay:DATa?
Parameter/ <NRf> 0.000~100.000
Return parameter Unit:ms
For values under 1s, the unit resolution
is 1ms.
For values above 1s, the unit
resolution is 0.1s.

130
COMMAND OVERVIEW

Example SYST:MDEL:DAT 1.105


Sets the delay time of measure is 1.1s.
SYST:MDEL:DAT?
>001.100
Returns the measurement delay as 1.1s.
Set
SYSTem:MDELay:STATe Query

Description Sets or returns the measurement delay function state.


Syntax SYSTem:MDELay:STATe <NR1> | {OFF|ON}
Query Syntax SYSTem:MDELay:STATe?
Parameter/ <NR1> 0:OFF.
Return parameter 1:ON.
OFF Turn the measurement delay off.
ON Turn the measurement delay on.
Example SYST:MDEL:STAT OFF
Turns the measurement delay function off.
Set
SYSTem:PWM:ON Query

Description Sets or returns the duty ON period for the PWM drive
mode.
PWM drive mode is only available for the GOM-805.
Note
Syntax SYSTem:PWM:ON <NR1>
Query Syntax SYSTem:PWM:ON?
Parameter/ <NR1> 3~99
Return parameter Unit: time units. For 60Hz LF, each
unit is equal 16.6ms. For 50Hz LF,
each unit is equal to 20.0ms.
Example SYST:PWM:ON 5
Sets the duty ON time to 5 adc units.

131
GOM-804 & GOM-805 User Manual

Set
SYSTem:PWM:OFF Query

Description Sets or returns the duty OFF period for the PWM drive
mode.
Syntax SYSTem:PWM:OFF <NR1>
Query Syntax SYSTem:PWM:OFF?
Parameter/ <NR1> 100~9999
Return parameter Unit:ms
Example SYST:PWM:OFF 200
Sets the duty OFF period to 200 ms.

SYSTem:SERial Query

Description Returns the serial number.


Query Syntax SYSTem:SERial?
Return parameter <String> 9 characters
Example SYST:SER?
> GXXXXXXXX

SYSTem:VERSion Query

Description Returns the SCPI version of the device.


Query Syntax SYSTem:VERSion?
Return parameter <String> 10 characters
Example SYST:VERS?
>SCPI1994.0.
SCPI version: 1994

132
COMMAND OVERVIEW

Temperature Commands
Set
TEMPerature:AMBient:DATa Query

Description Sets or returns the user-set ambient temperature value for


the temperature compensation and the temperature
conversion function.
Syntax TEMPerature:AMBient:DATa <NRf>
Query Syntax TEMPerature:AMBient:DATa?
Parameter <NRf> -50.0~399.9 (Unit: ºC)
Return parameter <NR2> -50.0~399.9 (Unit: ºC)
Example TEMP:AMB:DAT 25.6
Sets the user ambient temperature value to +25.6ºC.
TEMP:AMB:DAT?
>25.6
Returns the set ambient temperature as 25.6ºC.
Set
TEMPerature:AMBient:STATe Query

Description Sets or returns the state of the user-set ambient


temperature.
Syntax TEMPerature:AMBient:STATe <NR1> | {OFF|ON}
Query Syntax TEMPerature:AMBient:STATe?
Parameter/ <NR1> 0:OFF.
Return parameter 1:ON.
OFF Disables the user-set ambient
temperature.
ON Enables the user-set ambient
temperature.
Example TEMP:AMB:STAT OFF
Disables the user-set ambient temperature.

133
GOM-804 & GOM-805 User Manual

Set
TEMPerature:COMPensate:COEFficient Query

Description Sets or returns the temperature coefficient for


temperature compensation function.
Syntax TEMPerature:COMPensate:COEFficient <NR1>
Query Syntax TEMPerature:COMPensate:COEFficient?
Parameter/ <NR1> -9999~+9999
Return parameter
Example TEMP:COMP:COEF 3930
Sets the temperature coefficient to 3930ppm.
Set
TEMPerature:COMPensate:CORRect Query

Description Sets or returns the reference temperature for the


temperature compensation function.
Syntax TEMPerature:COMPensate:CORRect <NRf>
Query Syntax TEMPerature:COMPensate:CORRect?
Parameter <NRf> -50.0~399.9 (Unit: ºC)
Return parameter <NR2> -50.0~399.9 (Unit: ºC)
Example TEMP:COMP:CORR 25.5
Sets the reference temperature to 25.5ºC.
Set
TEMPerature:CONVersion:CONStant Query

Description Sets or returns the temperature constant for the


temperature conversion function.
Syntax TEMPerature:CONVersion:CONStant <NRf>
Query Syntax TEMPerature:CONVersion:CONStant?
Parameter <NRf> 0.0~999.9
Return parameter <NR2> 0.0~999.9
Example TEMP:CONV:CONS 235
Sets the temperature constant to 235.

134
COMMAND OVERVIEW

Set
TEMPerature:CONVersion:DISPlay Query

Description Sets or returns the temperature display mode for the


temperature conversion function.
Syntax TEMPerature:CONVersion:DISPlay <NR1>
Query Syntax TEMPerature:CONVersion:DISPlay?
Parameter/ <NR1> 1:ΔT
Return parameter 2:T
Example TEMP:CONV:DISP 1
Sets the temperature display mode for the temperature
conversion function is ΔT.

TEMPerature:CONVersion:MATH:DATa Query

Description Returns conversion function deviation value.


Query Syntax TEMPerature:CONVersion:MATH:DATa?
Return parameter <NR3> ±0.000~9.999E± X
Example TEMP:CONV:MATH:DAT?
Returns 1.250E+2.
Set
TEMPerature:CONVersion:RESistance Query

Description Sets or returns the initial resistance for the temperature


conversion function.
Syntax TEMPerature:CONVersion:RESistance {<NRf>[,<String>]}
Query Syntax TEMPerature:CONVersion:RESistance?
Parameter <NRf> 000.0001~999.9999
<String> mohm/ohm/kohm/maohm,unit
If the unit is not set, the unit will be
automatically set by the present range.
Return parameter <NR3> 000.0001~999.9999E± X

135
GOM-804 & GOM-805 User Manual

Example TEMP:CONV:RES 10.00,maohm


Sets initial resistance value to 10.00MΩ.
TEMP:CONV:RES?
>10.0000E+6
Returns the initial resistance as 10.00MΩ.
Set
TEMPerature:CONVersion:TEMPerature Query

Description Sets or returns the initial temperature for the temperature


conversion function.
Syntax TEMPerature:CONVersion:TEMPerature <NRf>
Query Syntax TEMPerature:CONVersion:TEMPerature?
Parameter <NRf> -50.0~399.9 (Unit: ºC)
Return parameter <NR2> -50.0~399.9 (Unit: ºC)
Example TEMP:CONV:TEMP 25.6
Sets the initial temperature to +25.6ºC.

TEMPerature:DATa Query

Description Returns the PT-100 sensor temperature measurement in


degrees Celsius.
Query Syntax TEMPerature:DATa?
Return parameter <NR3> -50.0~399.9
Example TEMP:DAT?
>0.250E+2
Returns the temperature as 25ºC.
Set
TEMPerature:STATe Query

Description Sets or returns the temperature function state.


Syntax TEMPerature:STATe {<NR1>|OFF|ON}
Query Syntax TEMPerature:STATe?
Parameter/ <NR1> 0:OFF
Return parameter 1:ON
OFF Turn the temp function off.
ON Turn the temp function on.

136
COMMAND OVERVIEW

Example TEMP:STAT ON
Sets the temp function on.
Set
TEMPerature:UNIT Query

Description Sets or returns the temperature unit. (Only used for the
display readback.)
Syntax TEMPerature:UNIT {DEGC|DEGF}
Query Syntax TEMPerature:UNIT?
Parameter/ DEGC ºC
Return parameter
DEGF ºF
Example TEMP:UNIT DEGF
Sets temperature unit to ºF (Fahrenheit).

137
GOM-804 & GOM-805 User Manual

Trigger Commands

READ Query

Description Returns the measurement value.


Query Syntax READ?
Return parameter <NR3> ± 0.0000~5.1000E± X
Example READ?
>+2.2012E+0
Returns the measurement.

MEASure<X> Query

Description Returns the results of the selected channel in the scan


mode, including HI/LO/IN and value.
Query Syntax MEASure<X>?
Parameter <X> Channel 1~100
Return parameter 0|1|2,<NR3> 0:LO
1:IN
2:HI
<NR3>: Measurement result.
Example MEAS1?
>1,+0.9978E+1
Returns channel 1 as 9.978Ω.

SHOW Query

Description Returns the judgments of all (up to 100) channels in the


scan mode.
Query Syntax SHOW?
Return parameter <String> 100 characters
0:LO
1:IN
2:HI
_:Channel not active

138
COMMAND OVERVIEW

Example SHOW?
Returns
1111111111______________________________________
________________________________________________
____.
Set
TRIGger:EDGE Query

Description Sets or returns the trigger edge (falling or rising edge).


Syntax TRIGger:EDGE {RISING|FALLING}
Query Syntax TRIGger:EDGE?
Parameter/ RISING Select rising trigger.
Return parameter FALLING Select falling trigger.
Example TRIG:EDGE FALLING
Sets the trigger to falling edge.
Set
TRIGger:DELay:DATa Query

Description Sets or returns the trigger delay time.


Syntax TRIGger:DELay:DATa <NR1>
Query Syntax TRIGger:DELay:DATa?
Parameter/ <NR1> 0~1000
Return parameter Unit:ms
Example TRIG:DEL:DAT 100
Sets the trigger delay time to 100ms.
Set
TRIGger:DELay:STATe Query

Description Sets or returns the trigger delay function state.


Syntax TRIGger:DELay:STATe <NR1> | {OFF|ON}
Query Syntax TRIGger:DELay:STATe?
Parameter/ <NR1> 0:ON
Return parameter 1:OFF
OFF Turn the trigger delay function off.
ON Turn the trigger delay function on.

139
GOM-804 & GOM-805 User Manual

Example TRIG:DEL:STAT OFF


Turns the trigger delay function off.
Set
TRIGger:SOURce Query

Description Sets or returns current trigger source.


Syntax TRIGger:SOURce {INT|EXT}
Query Syntax TRIGger:SOURce?
Parameter/ INT Internal trigger mode.
Return parameter EXT External trigger mode.
Example TRIG:SOUR EXT
Sets the current trigger source to external trigger.

140
COMMAND OVERVIEW

Userdefine Commands
Set
USERdefine<X>:ACTive Query

Description Sets or returns the active output state of the selected


Userdefine pin.
Syntax USERdefine<X>:ACTive <NR1>
Query Syntax USERdefine<X>:ACTive?
Parameter/ <X> Userdefine pin 1~2
Return parameter <NR1> 1:active low state
2:active high state
Example USER1:ACT 1
Sets the userdefine1 pin IO to active low state.
Set
USERdefine<X>:FIRStdata Query

Description Sets or returns the first operand for the selected user
define pin.
Syntax USERdefine<X>:FIRStdata <NR1>
Query Syntax USERdefine<X>:FIRStdata?
Parameter/ Return <X> Userdefine pin 1~2
parameter <NR1> 1~8:bin1~bin8 state
9:bin out state
10:hi state
11:low state
12:pass state
13:fail state
Example USER1:FIRS 12
Sets first operand of userdefine1 as pass state.
Set
USERdefine<X>:LOGic Query

Description Sets or returns operator for the selected user define pin.

141
GOM-804 & GOM-805 User Manual

Syntax USERdefine<X>:LOGic <NR1>


Query Syntax USERdefine<X>:LOGic?
Parameter/ <X> Userdefine pin 1~2
Return parameter <NR1> 1:off(only judge first data)
2:logical and.
3:logical or.
Example USER1:LOG 1
Sets the operator of userdefine1 to off. (I.e., only the first
operand determines the output of userdefine1.)
Set
USERdefine<X>:SEConddata Query

Description Sets or returns the second operand for the selected user
define pin.
Syntax USERdefine<X>:SECondata <NR1>
Query Syntax USERdefine<X>:SECondata?
Parameter/ <X> 1~2
Return parameter <NR1> 1~8:bin1~bin8 state
9:bin out state
10:hi state
11:low state
12:pass state
13:fail state
Example USER1:SEC 3
Sets the last operand of userdefine1 as the state of the
bin3 result.

142
COMMAND OVERVIEW

IEEE 488.2 Common Commands

*CLS Set
Query
Description Clears the Event Status register (Output Queue,
Operation Event Status, Questionable Event Status,
Standard Event Status).
Syntax *CLS
Parameter <None>
Set
*ESE Query

Description Sets or returns the ESER (Event Status Enable Register)


contents.
Syntax *ESE <NR1>
Query Syntax *ESE?
Parameter/ <NR1> 0~255
Return parameter
Example *ESE 65
Sets the ESER to 01000001
*ESE?
>130
ESER=10000010

*ESR Query

Description Returns SESR (Standard Event Status Register) contents.


Syntax *ESR?
Query Syntax
Return parameter <NR1> 0~255
Example *ESR?
>198
SESR=11000110

143
GOM-804 & GOM-805 User Manual

*IDN Query

Description Returns the manufacturer, model No., serial number and


system version number.
Query Syntax *IDN?
Return parameter <String> 31 characters
Example *IDN?
>GWINSTEK,GOM805,GXXXXXXXX,V1.00.
Set
*OPC Query

Description Sets or returns the operation complete bit (bit0) in SERS


(Standard Event Status Register) when all pending
operations are completed.
Syntax *OPC
Query Syntax *OPC?
Parameter <None>

Return parameter <NR1> 0:operation not complete


1:operation complete
Example *OPC?
Returns 1.

*RST Set
Query
Description Recalls default panel setup.
Syntax *RST
Parameter <None>
Set
*SRE Query

Description Sets or returns the SRER (Service Request Enable


Register) contents.

144
COMMAND OVERVIEW

Syntax *SRE <NR1>


Query Syntax *SRE?
Parameter/ <NR1> 0~255
Return parameter
Example *SRE 7
Sets the SRER to 00000111
*SRE?
>3
SRER=00000011

*STB Query

Description Returns the SBR (Status Byte Register) contents.


Query Syntax *STB?
Return parameter <NR1> 0~255
Example *STB?
>81
SESR=01010001

*TRG Set
Query
Description Manually triggers the instrument.
Syntax *TRG
Parameter <None>

145
GOM-804 & GOM-805 User Manual

Status system
The diagram below is a description of the status system.
Questionable Data Status
Register
Event Enable
Voltage Overload 0 0
Not Used 1 1
Not Used 2 2
Not Used 3 3
Output
Not Used 4 4
Buffer
Temp Overload 5 5
:
Not Used 6 6
:
Not Used 7 7
:
Not Used 8 8
Ohms Overload 9 9
Not Used 10 10
Limit Test Fail LO 11 11
Limit Test Fail HI 12 12
Not Used 13 13
Not Used 14 14
Not Used 15 15
Status Byte Register
Standard Event Status
Summary Enable
Register
Not Used 0 0
Event Enable
Not Used 1 1
OPC (Operation Complete) 0 0
Not Used 2 2
Not Used 1 1
QUES(Questionable Data) 3 3
QUE (Query Error) 2 2
MAV(Message Available) 4 4
Device Error 3 3
ESB(Standard Event) 5 5
EXE (Execution Error) 4 4
MSS(Request Service) 6 6
CME (Command Error) 5 5
Not Used 7 7
Not Used 6 6
PON (Power On) 7 7

For the following command sets, please refer to the diagram above:
STAT: QUES: EVEN?
STAT: QUES: ENAB
STAT: QUES: ENAB?
*ESR?
*ESE
*ESE?
*STB?
*SRE
*SRE?

146
FAQ

FAQ
● What are the different measurement speeds?
● The GOM-804/805 performance does not match the specifications.

What are the different measurement speeds?


There are two measurement speeds for both resistance and temperature
measurement. At the slow measurement rate, the measurement speed is 10
samples/s and at the fast measurement rate the measurement speed is at 60
samples/s.
The GOM-804/805 performance does not match the specifications.
Make sure the device is powered on for at least 30 minutes, is operated at the
slow measurement rate and is within +18°C~+28°C with a humidity not
exceeding 80%. This is necessary to stabilize the unit to match the
specifications.

If there is still a problem, please contact your local dealer or GWInstek at


marketing@goodwill.com.tw.

147
GOM-804 & GOM-805 User Manual

APPENDIX
Temperature Reference Temperature Table ............................... 149
Measurement RTD SensorsTemperature .................................. 150
Optional Platinum Sensor.................................... 150

Specifications Resistance Measurement ..................................... 152


Dry Resistance Measurement .............................. 153
Temperature Measurement .................................. 153
Temperature Correction Function ........................ 153
Interface............................................................... 154
Environmental ..................................................... 154
General ................................................................ 154
Dimensions ......................................................... 155

CE Declaration Declaration of Conformity .................................... 156

148
APPENDIX

Temperature Measurement
Reference Temperature Table

Overview The International Temperature Scale (ITS) is based on the


Background following table. The table has 17fixed calibration points as
of 1990.
Temperature
Element Type ˚K ˚C
(H 2 ) Hydrogen Triple point 13.8033 -259.3467
(Ne) Neon Triple point 24.5561 -248.5939
(O 2 ) Oxygen Triple point 54.3584 -218.7916
(Ar) Argon Triple point 83.8058 -189.3442
(Hg) Mercury Triple point 234.325 -38.8344
(H 2 O) Water Triple point 273.16 +0.01
(Ga) Gallium Melting point 302.9146 29.7646
(In) Indium Freezing point 429.7485 156.5985
(Sn) Tin Freezing point 505.078 231.928
(Zn) Zinc Freezing point 692.677 419.527
(Al) Aluminum Freezing point 933.473 660.323
(Ag) Silver Freezing point 1234.93 961.78
(Au) Gold Freezing point 1337.33 1064.18

149
GOM-804 & GOM-805 User Manual

RTD Sensors

Overview Resistive Thermal Devices (RTDs) are commonly used as


temperature sensors. RTDs change resistance linearly over
a specific range of temperature. The table below shows
some of the inherent features of RTDs compared to
thermocouples.

Feature Description
Accuracy Higher accuracy
Resolution 0.1~1.0˚C, higher resolution
Speed of response Slower
Self-heating Yes
Long term stability Good
Output characteristics Approx. 0.4ohm/˚C, near
linear

Optional Platinum Sensor


Introduction The optional platinum sensor is a PT-100 sensor.
The PT-100 sensor meets the German DIN43760:
1968 3 wire measurement specification.
These sensors are one of the most common
temperature sensors used in industry. These sensors
have a nominal resistance of 100Ω at 0˚C.
The relationship between temperature and resistance
for the PT-100 sensor can be described with the
Gallendarvan Dusen equation shown below:

R RTD =R 0 [1+AT+BT 2 +CT 3 (T-100)]

Where: R RTD is the calculated resistance of the


RTD.
R 0 is the known RTD resistance at 0˚C.
T is the temperature in˚C
A=alpha [I+(delta/100)]
B=-I(alpha)(delta)(Ie-4)
C=-I(alpha)(beta)(Ie-8)
The Alpha (A), Beta (B), Delta (D) values for the

150
APPENDIX

PT-100 sensor are listed below:


Type Standard Alpha Beta Delta Ω @ 0˚C
PT-100 ITS90 0.003850 0.10863 1.49990 100Ω
Temperature Example—Calculating the resistance of a PT-100
Calculation Example RTD at 100˚C (T). The following R0 (Ω at 0˚C),
alpha, beta, and delta values are used for the
PT-100 RTD:
T=100˚C
R 0 ( Ω at 0˚C) = 100 Ω
Alpha=0.003850
Beta=0.10863
Delta=1.49990
A, B, and C are calculated according to equations
listed above:
A=0.00391
B=5.77e-7
C=4.18e-12
The resistance of the RTD at 100˚ C (R100) is then
calculated as follows:
R 100 : =R 0 [1+AT=BT 2 +CT 3 (T-100)]
=100{1+[(0.00391)(100)]+[(-5.77e-7)(100 2 )
+[(-4.18E-12)(100 3 )(100-100)]]}
=138.5 Ω

151
GOM-804 & GOM-805 User Manual

Specifications
Conditions The specifications are applicable under the following
Background conditions:
 A 1-year calibration cycle.
 An operating temperature of 18 to 28 ˚C (64.4 to
82.4˚F).
 Relative humidity not exceeding 80%.
 Accuracy is expressed as ± (percentage of reading +
digits).
 The instrument requires 30 minutes warm-up time
and must be operated at the slow measurement rate to
achieve rated accuracy.
 The power cord protective grounding conductor must
be connected to ground.

Resistance Measurement
50000 counts
Measuring Open-Termin
Range Resolution Current Accuracy al Voltage
50mΩ 1μΩ 1A ±(0.1%+0.02%) ~6.5V
500mΩ 10μΩ 100mA ±(0.05%+0.02%) ~6.5V
5Ω 100μΩ 10mA ±(0.05%+0.02%) ~6.5V
50Ω 1mΩ 1mA ±(0.05%+0.02%) ~6.5V
500Ω 10mΩ 1mA ±(0.05%+0.008%) ~6.5V
5kΩ 100mΩ 1mA ±(0.05%+0.008%) ~6.5V
50kΩ 1Ω 100μA ±(0.05%+0.008%) ~6.5V
500kΩ 10Ω 10μA ±(0.05%+0.008%) ~6.5V
5MΩ 100Ω 1μA ±(0.2%+0.008%) ~6.5V
*When the instrument is set to 50mΩ or 500mΩ ranges, the resistance value
will be changed while connecting or disconnecting the test lead to the panel due
to the different temperature between internal and external parts of the
instrument. Therefore, please wait 1 minute in order to obtain an accurate value
after the test leads have been connected or disconnected.
* When Kelvin clips are used to resume testing after a long period of time,
please wait for a short time to stabilize the measurement.
*Fast and Slow measurement rates have the same specifications. However, the
Slow rate is more accurate as it will correct for any errors associated with
temperature drift that occurs from the difference between the measurement

152
APPENDIX

temperature and the calibration temperature.

Measurement Four-terminal method.


Auto-ranging Provided.
Over input range “------” indicates over range
Comparator 20 sets of comparator status can be selected.
Buzzer mode OFF, PASS, FAIL
switchable

Dry Resistance Measurement


Range Measuring Current Accuracy
500mΩ 100mA ±(0.3%+0.05%)
5Ω 10mA ±(0.3%+0.05%)
50Ω 1mA ±(0.3%+0.05%)

Temperature Measurement
Temperature sensor Platinum resistor.
(option) Lead length: 1.5m approx.
-10˚C ~40˚C 0.3%±0.5˚C
Other 0.3%±1.0˚C

Temperature Correction Function


Reference temperature -50.0˚C~399.9˚C
range
Thermal coefficient ±9999 ppm
range
Temperature range Accuracy of temperature compensation for 3930
ppm/Cu wire.*
-10˚C~40.0˚C 0.3%+resistance measurement accuracy.
Other 0.6%+resistance measurement accuracy.
*The temperature coefficient for the other settings must be calculated
individually according to different conditions.
*If the temperature coefficient or the difference between the environmental
temperature and the required temperature exceeds normal operation, after
calculating the compensation, the variation to the reading value will be
significant.
*When using the PT-100 temperature sensor for temperature measurements,
the accuracy of the sensor (typical accuracy of <±0.5ºC) should also be taken
into account and calculated for.

153
GOM-804 & GOM-805 User Manual

Interface
Handler interface* Signal: Trigger: TTL input
Signal: LOW, HIGH, FAIL, PASS, EOT, READY, BIN 1~8,
BIN OUT: total 15 TTL outputs.
Scan* Signal: RELAY, PASS, LOW, HIGH, CLOCK, STRB total 6
TTL outputs.
GOM-804: USB/RS-232
Communication
GOM-804G: USB/RS-232/GPIB
Interfaces
GOM-805: USB/RS-232/GPIB
*The Scan and Handler interface use the same
connector.

Environmental
Operation Indoor use, altitude up to 2000m.
Environment Ambient Temperature 0˚C to 40˚C.
Relative Humidity 80% (Maximum).
Pollution Degree 2
Storage temperature -10˚C to 70˚C.

General
Power source AC 100-240V±10%, 50-60Hz, 25VA
Accessories Power cord x1
Test lead: GTL-308 x1
User manual x1 (CD)
Safety instruction sheet x1
USB cable (option): GTL-246
Temperature sensor (option): PT-100
Dimension 223(W)×102(H)×283(D) mm
Weigh Approx. 3 kg

154
APPENDIX

Dimensions

GOM-805 DC Milliohm Meter


GND
Ohm Compare Binning

TC TCONV TEMP Enter

Speed REL RT

85.9

101.5
Scan Dry Trigger ESC

GUARD

Display Local Diode Drive Range

SOURCE SENSE SENSE SOURCE POWER

Rx

275.5 214.0

282.9 222.8

155
GOM-804 & GOM-805 User Manual

Declaration of Conformity
We
GOOD WILL INSTRUMENT CO., LTD.
No.7-1, Jhongsing Rd., Tucheng Dist., New Taipei City, Taiwan
GOOD WILL INSTRUMENT (SUZHOU) CO., LTD.
No. 69, Lu San Road, Suzhou New District, Jiangsu, China
declare, that the below mentioned product
Type of Product: DC Milliohm Meter
Model Number: GOM-804, GOM-805

are herewith confirmed to comply with the requirements set out in the
Council Directive on the Approximation of the Law of Member States
relating to Electromagnetic Compatibility (2004/108/EC) & (2014/30/EU)
and Low Voltage Directive (2006/95/EC) & (2014/35/EU).
For the evaluation regarding the Electromagnetic Compatibility and Low
Voltage Equipment Directive, the following standards were applied:

◎ EMC
EN 61326-1: Electrical equipment for measurement, control and
EN 61326-2-1: laboratory use -- EMC requirements (2013)
EN 61326-2-2:
Conducted and Radiated Emission Electrostatic Discharge
EN 55011: 2009+A1:2010 EN 61000-4-2: 2009
Current Harmonics Radiated Immunity
EN 61000-3-2: EN 61000-4-3 :
2006+A1:2009+A2:2014 2006+A1 :2008+A2 :2010
Voltage Fluctuation Electrical Fast Transients
EN 61000-3-3 :2013 EN 61000-4-4: 2012
Surge Immunity
-------------------------
EN 61000-4-5 :2006
Conducted Susceptibility
-------------------------
EN 61000-4-6 : 2014
Power Frequency Magnetic Field
-------------------------
EN 61000-4-8: 2010
Voltage Dip/ Interruption
-------------------------
EN 61000-4-11: 2004

Low Voltage Equipment Directive 2006/95/EC & 2014/35/EU


Safety Requirements EN 61010-1: 2010
EN 61010-2-030: 2010

156
Index

INDEX
Binning function measure delay ................................................. 61
setting ............................................................... 46 PWM duty ....................................................... 68
Characteristics .................................................. 10 setting ............................................................... 60
Compare function temperature unit ............................................. 65
setting ............................................................... 41 trigger delay .................................................... 63
Declaration of conformity ............................. 156 trigger edge ..................................................... 64
Dimensions ..................................................... 155 Power supply safety instructions..................... 6
Diode ................................................................. 40 Power up ........................................................... 24
Display mode.................................................... 35 PT-100 sensor temperature calculation ....... 150
Disposal instructions ......................................... 7 PWM duty ......................................................... 68
Drive overview ................................................. 31 Range ................................................................. 30
Drive setting ..................................................... 33 Rate
Dry circuit ......................................................... 37 setting ............................................................... 34
EN 61010 Real time display .............................................. 36
measurement category ..................................... 6 Rear panel overview ........................................ 21
pollution degree ................................................ 7 Recall settings ................................................... 99
Environment Reference temperature table ......................... 149
operation ............................................................ 6 Remote control
storage ................................................................ 7 binning commands ....................................... 108
External IO ........................................................ 73 calculate commands ..................................... 113
FAQ .................................................................. 147 Command list ................................................ 105
Front panel overview....................................... 15 command syntax ........................................... 102
Getting Started chapter ..................................... 9 common commands ..................................... 143
Handler memory commands ...................................... 120
compatibility ................................................... 89 sense commands ........................................... 122
overview .......................................................... 78 source commands ......................................... 126
pinout ............................................................... 80 status commands .......................................... 127
system commands ........................................ 128
Handler mode................................................... 74
temperature commands ............................... 133
Interface
trigger commands......................................... 138
GPIB
userdefine commands .................................. 141
function check ................................. 96 Resistance
range ................................................................. 30
setting ............................................. 93
setting ......................................................... 29, 40
overview .......................................................... 90 Resistance measurement
RS232 connection........................................................ 25
function check ................................. 93 RT display ......................................................... 36
Safety instruction
Realterm example .......................... 94
Guidelines .......................................................... 6
setting ............................................. 92 Safety instructions
USB power supply .................................................... 6
symbol ................................................................ 5
driver ............................................... 91 Save settings ..................................................... 99
function check ................................. 93 Scan
GOM-802 compatibility ................................. 89
Realterm example .......................... 94 output ............................................................... 88
setting ............................................. 90 overview .......................................................... 82
pinout ............................................................... 83
Measurement settings setup ................................................................. 84
ambient temperature ...................................... 66 Service contact ................................................ 147
average ............................................................. 60
Specifications .................................................. 152
line frequency .................................................. 67

157
GOM-804 & GOM-805 User Manual

Status system .................................................. 146 setting ............................................................... 52


System settings Temperature conversion
beep ................................................................... 76 setting ............................................................... 56
brightness ......................................................... 72 Temperature measurement
external IO ........................................................ 73 reference ......................................................... 149
handler mode ................................................... 74 TFT-LCD overview .......................................... 19
interface ............................................................ 71 Tilt stand ........................................................... 23
power on settings ............................................ 70 Trigger
system information ......................................... 69 setting ............................................................... 38