29193223
29193223
29193223
Operating Instructions
Original instructions
cytiva.com
Table of Contents
Table of Contents
1 Introduction ........................................................................................................ 4
1.1 About this manual ......................................................................................................................................... 5
1.2 Important user information ....................................................................................................................... 6
4 Installation .......................................................................................................... 33
4.1 Site requirements .......................................................................................................................................... 34
4.2 Transport .......................................................................................................................................................... 35
4.3 Unpacking ......................................................................................................................................................... 36
4.4 How to level the column .............................................................................................................................. 38
4.5 Grounding kit ................................................................................................................................................... 40
4.6 Setup ................................................................................................................................................................... 42
5 Operation ............................................................................................................ 45
5.1 Preparation ...................................................................................................................................................... 47
5.1.1 Column flow packing method ................................................................................................................. 49
5.1.2 Slurry preparation ........................................................................................................................................ 50
5.2 Column packing .............................................................................................................................................. 53
5.2.1 Preparing the column ................................................................................................................................. 54
5.2.2 Preparing the adapter ................................................................................................................................ 57
5.2.3 Packing the column - first step ............................................................................................................... 62
5.2.4 Packing the column - second step ........................................................................................................ 67
5.2.5 Storage and unpacking of packed columns ...................................................................................... 70
5.2.6 Pressure flow curve ..................................................................................................................................... 71
5.3 Performance evaluation of the column ................................................................................................. 72
6 Maintenance ....................................................................................................... 75
6.1 Service and preventive maintenance .................................................................................................... 77
6.2 Replacements of nets and O-rings .......................................................................................................... 79
6.2.1 Replacements in the adapter .................................................................................................................. 81
6.2.2 Replacements in the column ................................................................................................................... 87
6.3 Replacement sealing washer .................................................................................................................... 98
6.4 Fitting a packing extension tube ............................................................................................................. 99
6.5 Disassemble and assemble the adapter ............................................................................................... 101
6.6 Leakage test .................................................................................................................................................... 113
6.7 Cleaning ............................................................................................................................................................ 115
1 Introduction
In this chapter
All users must read the entire Operating Instructions before installing, oper-
ating or maintaining the product.
Always keep the Operating Instructions at hand when operating the product.
Do not install, operate, or perform maintenance on the product in any other way than
described in the user documentation. If you do, you may be exposed or expose others
to hazards that can lead to personal injury and you may cause damage to the equip-
ment.
Prerequisites
In order to operate the equipment in the way it is intended, the following prerequisites
must be fulfilled:
• You must have read and understood the safety instructions in the user documenta-
tion.
• You must have experience of working with pressure vessels.
• You must have knowledge of the entire system and process that the column is part
of.
• All operations should be performed by qualified personnel who are adequately
trained.
Definitions
This user documentation contains safety notices (WARNING, CAUTION, and NOTICE)
concerning the safe use of the product. See definitions below.
WARNING
WARNING indicates a hazardous situation which, if not avoided,
could result in death or serious injury. It is important not to proceed
until all stated conditions are met and clearly understood.
CAUTION
CAUTION indicates a hazardous situation which, if not avoided,
could result in minor or moderate injury. It is important not to
proceed until all stated conditions are met and clearly understood.
NOTICE
NOTICE indicates instructions that must be followed to avoid
damage to the product or other equipment.
2 Safety instructions
Important
WARNING
Before installing, operating or maintaining the product, all
users must read and understand the entire contents of this
chapter to become aware of the hazards involved.
In this chapter
General precautions
WARNING
Do not operate the product in any other way than described in the
user documentation.
WARNING
Only properly trained personnel may operate and maintain the
product.
WARNING
Do not use any accessories not supplied or recommended by
Cytiva.
WARNING
Pressure may remain in the column and tubing for an extended
period after shutdown. Make sure the system is depressurized
before opening or disconnecting.
WARNING
EXPLOSION HAZARD! If cleaning the column in a potentially
explosive atmosphere, prevent static electric charges building up.
Do not rub excessively using a dry cloth and over large areas at a
time. Instead, clean the column using a damp cloth with slow
movements across small areas at a time.
WARNING
EXPLOSION HAZARD! Some of the chemicals used with BPG
columns may be flammable under certain conditions. Make sure to
use chemicals only under conditions where they are not flam-
mable. Refer to local and/or national classifications of flammable
liquids.
WARNING
EXPLOSION HAZARD! When the BPG column is operated or
maintained in a potentially explosive atmosphere, it MUST be
properly grounded to avoid static discharge.
WARNING
EXPLOSION HAZARD! When connecting the grounding cable,
make sure that there is no explosive atmosphere present. Static
electricity discharge could occur when connecting the grounding
cable.
Personal protection
WARNING
For personal safety during the transportation, installation, opera-
tion, maintenance and service of columns, use protective glasses
and other personal protective equipment appropriate for the
current application at all times. The following personal protective
equipment should always be available:
• Protective glasses
• Working gloves to protect against sharp edges
• Protective footwear, preferably with steel toe-cap
• Disposable gloves
Always use clean disposable gloves when manually handling parts.
WARNING
When using hazardous chemical and biological agents, take all
suitable protective measures, such as wearing protective glasses
and gloves resistant to the substances used. Follow local and/or
national regulations for safe operation and maintenance of the
BPG columns.
WARNING
Only personnel with appropriate training may move and unpack
the delivery crates. All moving and unpacking must be performed
according to local regulations.
Even if the safety instructions in the user documentation are
followed, it is the customers' responsibility to guarantee the safety
of the personnel working with BPG columns.
WARNING
Move transport crates. Make sure that the lifting equipment has
the capacity to safely lift the crate weight. Make sure that the crate
is properly balanced so that it will not accidentally tip when moved.
WARNING
Have a minimum of two people on hand, as well as the appropriate
lifting equipment for the application, such as a hoist, when moving
or lifting a heavier BPG column, i.e., BPG 300 or 450.
WARNING
Be careful not to unscrew the wheels too far. If they are unscrewed
totally the column may fall and cause injury or damage. Make sure
that the wheel support is fully engaged into the stand before start
leveling.
WARNING
Make sure that the wheel support is fully engaged into the stand
before start moving.
CAUTION
After moving the column to its resting position, lock the wheel
brakes and level the column. Place the column on an even floor
area.
CAUTION
Make sure that the safety equipment's (for example pressure relief
valve or rupture disc) connecting tubing has an unrestricted flow
path and connected to waste.
Operation
WARNING
The working pressure of the column should never exceed its design
pressure, otherwise there is a risk of personal injury and damage to
the column. Always use appropriate safety equipment, for example
pressure relief valve or rupture discs.
WARNING
There should NEVER be air or gas under pressure in the column.
WARNING
Pay extra attention when fitting the adapter to the column as there
is a risk of crushing your hands.
CAUTION
Do not use chemicals with temperatures above the specified limits.
Maintenance
WARNING
Thoroughly check the glass column tube to make sure that it is not
cracked or has been severely scratched. Pay extra attention to the
end parts of the column tube. The pressure specifications are only
valid if the glass tube is undamaged.
WARNING
Decontaminate before maintenance. To avoid personnel being
exposed to potentially hazardous substances, make sure that the
column is properly decontaminated and sanitized before mainte-
nance or service.
CAUTION
For continued protection against injury risks due to fluid jets, burst
pipes or explosive atmosphere, the piping system must be tested
for leakage at maximum pressure when the following situations
apply:
• After assembly or maintenance
• Before operation or Cleaning-In-Place (CIP)
CAUTION
After assembly, check that the column is tight and not leaking.
CAUTION
Removing the top plate and adapter often causes spillages and
splashing. Remove any spillage on the floor immediately to mini-
mize the risk for slipping accidents.
System label
The system label is located on the back of the equipment. The system label identifies
the equipment and shows electrical data, regulatory compliance, and warning
symbols.
WARNING
Pressure may remain in the column and tubing for an extended
period after shutdown. Make sure the system is depressurized
before opening or disconnecting.
Emergency shutdown
The operation of BPG columns is controlled by the equipment that is connected to the
column. All operations are therefore stopped by stopping the operation of the
connected equipment.
If the column is connected to a system, the procedure for the emergency shutdown of
the system should be followed. This may involve pressing the EMERGENCY STOP
button to stop the operation of the equipment controlling the pumping system
connected to the column. Alternatively, the process system may require to be paused
or shutdown. Follow the manufacturer's instructions for the system.
Restart
To restart after an emergency shutdown:
Step Action
1 Make sure that the condition that was the root cause of the emergency stop
has been corrected.
3 Column descriptions
In this chapter
3.1 Overview 19
3.5 Accessories 31
3.1 Overview
Introduction
The BPG column can be used conjunction with a BioProcess system or as a stand-alone
unit connected to a pump. There are five sizes of the BPG columns.
The BPG columns are listed in the table below with the internal diameters, specific
design pressure and the various heights.
Characteristics BPG 100 BPG 140 BPG 200 BPG 300 BPG 450
Internal diameter 100 140 200 296 446
(mm)
Design pressure 8 6 6 4 2.5
(bar)
Column height 500, 750 , 500, 750 , 500, 750 , 500, 750 , 500, 750 ,
(mm) 950 950 950 950 1000
For detailed specifications of the BPG columns, refer to Column specifications, on page
125.
All BPG columns are equipped with wheels with brakes, except the BPG 100 column, for
which wheel with brakes are an option.
Column material
The materials used in the BPG columns are compatible with the solvents most
commonly used in chromatography separations of biomolecules, column mainte-
nance and cleaning processes. All polymeric materials in contact with the process
stream meet the requirements for USP class VI, described in USP <88> Biological Reac-
tivity Tests, In Vivo.
Nets
The column is delivered with 23 μm (PolyPropylene) nets. For media with an average
particle diameter <70 μm, change to 10 μm (Polyamide), or 12 μm (PEEK) in both
adapters and end-pieces. For big beads use 54 μm nets, see Recommended net sizes,
on page 48.
Connections
BPG columns can be connected to different types of tubing and systems. To protect
the packed bed it is advisable to have isolation valves on both the inlets and outlets of
the column.
Explosive atmosphere
Perform a risk assessment for any risks due to the process or process environment.
Evaluate the effects the use of the product and the operational processes may have on
the classification of the hazardous area. The process might cause the area to increase
or the zone classification to change. Implement the risk reduction measures needed,
including use of personal protective equipment.
If potentially explosive liquids will be used with BPG columns consult local authorities
regarding local rules and regulations before installing or operating.
The column can be placed in an area with explosive atmosphere according to EU ATEX
workplace directive, Zone 2.
The column does not fall under the ATEX equipment directive, and as a result it is not
marked with EX labels, but it fulfills the explosion safety requirements in the directive’s
harmonized standards, for example EN 13463-1 or EN ISO 80079-36.
This is valid for the BPG column in its standard execution and with a grounding cable,
see Grounding kit, on page 32. Any additional accessory or equipment must carry EX-
marking or be assessed to be in conformity with the EU ATEX workplace directive.
Part Description
1 Seal adjuster knob
2 Adapter tube, outer
3 Adapter height adjuster
4 Domed nut
5 Adapter top plate
6 Flange
7 Column glass tube
8 Support rod
9 Adapter plate
10 O-ring adapter
11 Net (adapter)
12 O-ring bottom
13 Net (column)
14 End-piece
15 Stand
Part Description
1 Column
2 Adapter
Part Description
1 Seal adjuster knob
2 Adapter tube, outer
3 Adapter height adjuster
4 Lifting grip
5 Domed nut
6 Adapter top plate
7 Flange
8 Column glass tube
9 Support rod
10 Adapter plate
11 O-ring adapter
12 Net (adapter)
13 O-ring bottom
14 Net (column)
15 End-piece
16 Stand
2
1
Part Description
1 Column
2 Adapter
Part Description
1 Seal adjuster knob
2 Adapter tube, outer
3 Adapter height adjuster
Part Description
4 Eyebolt
5 Adapter top plate
6 Flange
7 Column glass tube
8 Support rod
9 Adapter plate
10 O-ring adapter
11 Net (adapter)
12 O-ring bottom
13 Net (column)
14 End-piece
15 Stand
Part Description
1 Column
2 Adapter
3.5 Accessories
Packing extension tube
Packing extension tubes are available for all BPG columns. The packing tubes are made
of glass for BPG 100 to 300, and of stainless steel for BPG 450. The illustration below
shows two examples of extension tubes.
Air trap
The air trap kit includes the air trap, bracket, steel valves, clamps and gaskets.
For air traps for BPG 100, 140 and 200, tubing is included. A manually operated valve is
recommended at the top of the air trap as an air outlet control.
Grounding kit
To avoid discharge from static electricity, the column can be properly grounded
(earthed). For this purpose, a suitable grounding kit is available as an accessory and a
grounding connection terminal is provided. The equipment must be grounded when it
is to be used in a potentially explosive atmosphere.
4 Installation
In this chapter
4.2 Transport 35
4.3 Unpacking 36
4.6 Setup 42
Equipment requirements
When transporting and maneuvering BPG columns and adapters the following equip-
ment is recommended:
• Suitable lifting equipment (crane or hoist) for the larger BPG columns, see Moving
the adapter BPG 300 and BPG 450, on page 76.
• Lifting eyebolts.
Ambient environment
The following should be avoided:
• Direct sunlight
• Vibrations
• Corrosive gas
• Dust
4.2 Transport
Check the delivery
On receipt of the crate check to see if there is any apparent damage to the crate and
the equipment. When unpacking BPG columns check that all equipment is enclosed in
the transport crate according to the packing list.
If any damage is observed do not continue with the installation of the column, record
this on the receiving documents. Contact your Cytiva representative for advice and
further instruction.
Transport in crate
WARNING
Move transport crates. Make sure that the lifting equipment has
the capacity to safely lift the crate weight. Make sure that the crate
is properly balanced so that it will not accidentally tip when moved.
Use proper lifting device with a minimum capacity to match the empty weight of the
column plus the crate. If maneuvering is performed with a crane or hoist, keep lifting
heights to the minimum necessary to make sure that there is sufficient clearance
space to allow passage of the column when lifted from the floor. Refer to the product
documentation for information about column weight and dimensions for each column.
WARNING
Make sure that the wheel support is fully engaged into the stand
before start moving.
Be careful not to unscrew the wheels too far before moving. If they are unscrewed
totally the column may fall and cause injury or damage. The wheels should be fully
engaged into stand before moving the column.
4.3 Unpacking
Tools required for crate unpacking
The following tools are recommended when removing larger BPG columns from the
transport crate:
• Suitable lifting equipment, such as a crane or hoist
• Round slings or lifting chains with the capacity to safely lift the weight of the column
• Lifting eyebolts
• Screwdriver
• Suitable wrench or ratchet wrench
Crate unpacking
WARNING
Only personnel with appropriate training may move and unpack
the delivery crates. All moving and unpacking must be performed
according to local regulations.
Even if the safety instructions in the user documentation are
followed, it is the customers' responsibility to guarantee the safety
of the personnel working with BPG columns.
Unpack the BPG 300 and 450 columns by following the unpacking instructions
attached to the outside of the transport crate. In the absence of unpacking instruc-
tions, unpack BPG column from the transport crate as follows:
Step Action
1 Move the crate to a protected indoor location. Make sure that the transport
crate is located on a level and even floor surface.
2 Remove the screws marked with black paint from the side panel of the trans-
port crate, and carefully remove that side panel from the crate.
4 Remove the enclosed accessories that are packed inside the crate.
5 Place the ramp that is provided inside the crate on the floor surface.
6 Secure the ramp tightly against the open side of the crate. Make sure there is
a smooth transition between the bottom of the crate and the floor.
7 Release the column wheel brakes and carefully roll the column out of the
crate and onto the floor.
WARNING
Have a minimum of two people on hand, as well as the appropriate
lifting equipment for the application, such as a hoist, when moving
or lifting a heavier BPG column, i.e., BPG 300 or 450.
The BPG column can be rolled to the intended operational location on its own wheels,
or the larger columns can be lifted into position using a crane or hoist.
Tip: It is recommended to wrap PTFE tape around the threads of the eyebolts
used with BPG column.
WARNING
Be careful not to unscrew the wheels too far. If they are unscrewed
totally the column may fall and cause injury or damage. Make sure
that the wheel support is fully engaged into the stand before start
leveling.
CAUTION
After moving the column to its resting position, lock the wheel
brakes and level the column. Place the column on an even floor
area.
NOTICE
Make sure that the column is located on a level even surface.
WARNING
EXPLOSION HAZARD! When the BPG column is operated or
maintained in a potentially explosive atmosphere, it MUST be
properly grounded to avoid static discharge.
WARNING
EXPLOSION HAZARD! When connecting the grounding cable,
make sure that there is no explosive atmosphere present. Static
electricity discharge could occur when connecting the grounding
cable.
Illustration
A grounding kit is available as an accessory for BPG columns and a grounding connec-
tion terminal is provided on the column stand as shown below.
Part Function
1 Grounding terminal/cable on BPG 100, 140, 200 and 450 columns.
2 Grounding terminal on BPG 300 column.
4.6 Setup
Precautions
WARNING
The working pressure of the column should never exceed its design
pressure, otherwise there is a risk of personal injury and damage to
the column. Always use appropriate safety equipment, for example
pressure relief valve or rupture discs.
CAUTION
Make sure that the safety equipment's (for example pressure relief
valve or rupture disc) connecting tubing has an unrestricted flow
path and connected to waste.
NOTICE
The bottom outlet should be above the top of the column to elimi-
nate any chance of a siphon.
NOTICE
A pressure relief valve or rupture disc shall be connected between
the pump and column inlet.
Column setup
The column is fitted with a 4-port 4-way valve on the adapter inlet (B), and a 4-port 2-
way valve on the bottom outlet.
• One port (A) on the top has a pressure gauge followed by tubing connected to a
pump. Another port (C) is directed to the waste for purging the pump and column.
• The bottom valve is equipped with tubing on a port perpendicular to the column
outlet that will be directing flow to waste, or recycling water back to the tank.
3
1 A C
2
B
6
4
Flow Description
1 From pump
2 Pressure gauge
3 4-port, 4-way valve
4 Drain
5 4-port, 2-way valve
6 Drain
7 Return
B
A C
B to C A C Column to waste
5 Operation
In this chapter
5.1 Preparation 47
Precautions
WARNING
Only properly trained personnel may operate and maintain the
product.
WARNING
Thoroughly check the glass column tube to make sure that it is not
cracked or has been severely scratched. Pay extra attention to the
end parts of the column tube. The pressure specifications are only
valid if the glass tube is undamaged.
WARNING
There should NEVER be air or gas under pressure in the column.
WARNING
The working pressure of the column should never exceed its design
pressure, otherwise there is a risk of personal injury and damage to
the column. Always use appropriate safety equipment, for example
pressure relief valve or rupture discs.
CAUTION
Do not use chemicals with temperatures above the specified limits.
NOTICE
Make sure that any water is WFI, since particles can block and
damage the nets.
NOTICE
Use a harmless fluid in the beginning of the process. This will make
it possible to detect leakage with minimized consequences and the
risk for potential leakage of hazardous fluids is avoided.
5.1 Preparation
About this section
This section contains information how to prepare the BPG columns before column
packing. Information and support related to optimal column packing can be obtained
from your Cytiva representative.
In this section
Location
Take care in selecting the position of the column before proceeding. System, tanks,
drains and handling devices for the larger columns all require space and easy, safe
access during all operations of the column. Take care to route the hoses carefully to
avoid tripping risks and always try to keep hose lengths to a minimum.
Arrange connections
NOTICE
Only use tubing and clamps approved for the maximum pressure of
the column and resistant to the chemicals intended to be used.
Make sure the tubing inner diameter is constant throughout your system setup to
minimize unwanted volume and keep a consistent flow and pressure in the process
stream. The valves, tubing and gaskets used should have the same inner diameter as
the column outlets or slightly larger
Materials
The following materials are suitable for the column packing:
• BPG column with suitable net for the media
• Media about 50% slurry in water
• WFI for packing
Method description
Constant flow packing is a two step method that is applicable to all types of media.
Assuming a suitable liquid delivery system is available, the main requirements are that
the column must have a moveable adapter and a pressure rating which allows the
optimal packing flow rate to be obtained.
1. In the first step, a low flow rate is used to form a loosely packed bed from the slurry.
This allows the adapter to be lowered the majority of the distance towards its final
packed position with little or no expansion of the bed during travel time.
2. In the second step, the flow rate is increased to a predetermined optimal value to
obtain the desired packing density. When flow is stopped, the proximity of the
adapter to the bed minimizes the time necessary to successfully reposition the
adapter to stabilize the packed bed.
Slurry concentration
Mix the packing buffer with the media to form a 50% slurry, i.e., sediment volume/total
slurry volume = 0.50. This slurry volume is appropriate for BioProcess media.
Make sure that the correct amount of slurry is used to end up with the target bed
height and optimal bed compression.
More information regarding how to determine slurry concentration using the Slurry
Concentration Kit can be found in Instruction 29112530.
Compression factor
The media needs to be compressed slightly to maintain a stable packed bed. The multi-
plication factor to convert packed bed volume to settled media volume is the compres-
sion factor.
Compression Factor CF = (gravity settled bed height)/(packed bed height)
Compression factor should read for most Cytiva media, between 1.10 and 1.15.
Sephadex G-25
The Sephadex™ G25 media swell 4 to 5 times (1 g = 4-5 ml of media). The table shows
the amount of media per centimeter of packed bed.
Note: The figures in the table below are based on four-fold swelling and that there
can be batch-to-batch variation.
Step Action
2 Pour the slurry into a suitable container for pouring into the column.
In this section
Column preparation
Operation of the adapter, integrity of the adapter O-ring and porosity of the support
nets should be checked before starting. The column tube and all internal surfaces must
be clean.
Step Action
1 Make sure that the column is level, see Section 4.4 How to level the column,
on page 38.
2 Make sure that the column is grounded, see Section 4.5 Grounding kit, on
page 40.
3 Wet the column net with 20% ethanol and pour 2 to 3 cm of packing buffer
into the column with the bottom valve closed and the outlet to drain.
Step Action
4 Open the outlet valve at the bottom of the column and allow liquid and air to
drain from the bottom distributor plate.
6 Connect a pipe to a pump, or a vacuum line, and lower the pipe down onto
the surface of the bottom net.
7 Start the suction to remove any remaining air from under or on the surface
of the bottom net.
Step Action
Result:
The column tube is now ready for packing.
2 Position the valve from pump to waste (A to C) to purge air and start the
pump.
A C
A C
Step Action
5 Carefully pour the homogenous slurry into the column. Make sure that no air
bubbles are trapped within the slurry and the bottom outlet is higher than
the column.
6 Measure the height (H) from the top of the column to the surface of the
slurry.
H cm
Step Action
A C
H - (0.5 - 1.0) cm
8 Insert the adapter in the column just above the slurry level in the column
with the seal fully relaxed.
A C
H - (0.5 - 1.0) cm
WARNING
Pay extra attention when fitting the adapter to the
column as there is a risk of crushing your hands.
Step Action
10 Seal the adapter O-ring by turning the seal adjuster knob or handle on the
top of the adapter, clockwise.
Step Action
A C
13 Result:
Air, and then liquid, is pushed up through the adapter and out to waste as the
adapter is pushed downwards.
A C
First step
Step Action
1 Remove any air in the hoses by positioning the valve from system to waste (A
to C) and start the pump.
A C
Step Action
2 Position the valve to flow from the pump to the column (A to B) through the
inlet on the adapter, and at the same time open the bottom valve to waste.
A C
4 Stop the pump when the bed has settled, normally one to two column
volumes.
A C
Step Action
5 Close the bottom valve when the pressure in the column has returned to
zero and set the top valve to waste (B to C).
A C
6 Loosen the adapter O-ring seal by turning the adjuster knob, or the seal
adjuster handle, on the top of the column counter-clockwise.
A C
Step Action
7 Lower the adapter quickly by turning the height adjuster handle clockwise.
The buffer between the sedimented bed and the adapter passes between
the adapter O-ring and the column wall. During this operation the bed will
start to rise. Stop lowering the adapter when it is 0.5 to 1.0 cm above the bed.
A C
0.5 - 1.0 cm
8 Seal the adapter O-ring by turning the knob, or seal adjuster handle, clock-
wise and lower the adapter to 0.1 or 0.2 cm above the bed.
A C
0.1 - 0.2 cm
Step Action
9 Result:
The buffer is passing out the top valve to waste (B to C).
A C
Second step
Step Action
1 Open the bottom valve and put the top valve from pump to column (A to B)
and start the pump. This will move the bed down and once stabilized the
level should be marked on the tube.
A C
Step Action
3 When the pressure reaches zero, position the top valve to waste (B to C), and
move the adapter down, pushing into the media until the adapter reaches
the mark on the tube.
A C
5 Result:
The column is now packed.
6 Test the packed column, see Section 5.3 Performance evaluation of the
column, on page 72.
Step Action
A C
2 Wash, if necessary.
3 Repeat.
Storage
Both outlets of the column should be sealed after equilibrating the column in storage
buffer. The column must be stored in a temperature-controlled environment with
water or 20% ethanol above the adapter O-ring to prevent it from drying out and
cracking. Covering the column with a clean plastic bag helps prevent dust and debris
from accumulating on exterior surfaces.
Unpacking
Place several centimeters of buffer or water above the adapter O-ring of the BPG
column and fully loosen the wing nut which seals the adapter O-ring. Make sure the
adapter outlet is open to waste. If the adapter is not easily raised, try slightly lowering it
to break the O-ring seal on the glass tube. If the O-ring rolls, tighten the wing nut a bit
to prevent this during adapter ascent. A firm twist on the threaded adapter shaft is also
useful when attempting to unseal the O-ring.
Raise and remove the adapter. If the packed bed is difficult to reslurry, pumping buffer
up through the bottom valve will help to lift and disperse the media more rapidly and
clear the bottom net.
70% 100%
b c
Value Meaning
a Packing flow rate
b Packing pressure
c Max pressure
Test chromatogram
The following test chromatogram shows the parameters used to calculate peak broad-
ening and peak symmetry.
Ve
y
WVhh
50%
a b
10%
x
HETP calculation
Calculate the height equivalent of a theoretical plate (HETP) from the UV curve or
conductivity curve if sodium chloride is used as sample, as follows:
HETP = L/N
N=5.54 (Ve/Wh)2
Variable Meaning
L Bed height
N Number of theoretical plates
Ve Elution volume
Wh Peak width at half peak height
Asymmetry calculation
Calculate the asymmetry factor (AS) from the UV curve or conductivity curve if sodium
chloride is used as sample, as follows:
The peak should be symmetrical and the asymmetry factor as close to 1 as possible
(values between 0.8 and 1.8 are usually acceptable). A change in the shape of the peak
is usually the first indication of bed deterioration due to use.
The asymmetry factor As describes the deviation from an ideal Gaussian peak shape
and is calculated from the peak width at 10% of peak height.
As = b/a
6 Maintenance
In this chapter
Precautions
WARNING
Pressure may remain in the column and tubing for an extended
period after shutdown. Make sure the system is depressurized
before opening or disconnecting.
WARNING
Decontaminate before maintenance. To avoid personnel being
exposed to potentially hazardous substances, make sure that the
column is properly decontaminated and sanitized before mainte-
nance or service.
WARNING
Have a minimum of two people on hand, as well as the appropriate
lifting equipment for the application, such as a hoist, when moving
or lifting a heavier BPG column, i.e., BPG 300 or 450.
The adapters on BPG 300 and 450 columns are heavy. For safety reasons, it is recom-
mended that the adapter is lifted with an adapter handling unit or ceiling hoist
connected to the eyebolts on the adapter top plate.
WARNING
Only properly trained personnel may operate and maintain the
product.
WARNING
Do not use any accessories not supplied or recommended by
Cytiva.
NOTICE
If the column tube or the rods are replaced, the column assembly
must be checked in accordance with local authority instructions.
Service frequency
Regular service and maintenance of the column is necessary to maintain the column in
optimal condition and extend the operational lifetime of column components. Nets
and and worn O-ring seals should be checked regularly for wear.
The requirements for frequency of service and preventive maintenance depend on the
frequency of use of the column and the specific applications performed.
A general recommendation is one service and preventive maintenance visit every two
years, but if columns are in continuous operation an annual visit is recommended.
Contact your local Cytiva representative for information about frequency of service
requirements to suit individual application needs.
Note: Note that the recommendations may not apply to your specific use of the
system. The system owner is solely responsible for establishing applicable
routines for periodic maintenance.
Please complete the checklist in the On Site Service Health and Safety Declaration
Form or the Health and Safety Declaration Form for Product Return or Servicing,
depending on whether the instrument is going to be serviced on site or returned for
service, respectively.
Passivation procedure
It is recommended to include a passivation procedure into a service and preventive
maintenance program. Contact your Cytiva representative for additional information.
Nets
Blockage of the nets causes malfunction of the columns and an over-used net can
affect distribution. To prevent blockage follow the instructions listed below.
• When columns are to be re-packed, always check the nets and wash them in a
detergent solution or replace them.
• It is important to select the appropriate adapter and end-piece nets, see Recom-
mended net sizes, on page 48.
• Check the HETP and As regularly to prevent poor performance due to old nets.
Illustration Part
Hex keys
Spanners
Screwdrivers
Adapter items
Step Action
1 Turn the handles on the adjusting handle (or knob) several turns to loosen
the adapter O-ring seal, counter-clockwise.
Tip:
Move the adapter down a few millimeters to loosen the O-ring from the glass
wall. Then move it up as far as it will go.
NOTICE
If the adapter O-ring seal is dry, it may be damaged by
forced movement of the adapter. Wet the O-ring with
water or 20% ethanol solution. Do NOT grease the O-
ring.
2 Remove the domed nuts and washers from the adapter top plate with help of
a spanner. Loosen the screws cross-wise according to the illustration below.
Step Action
6
7
3
1
2
4
8
NOTICE
Do not loosen each nut completely in one movement.
Three to four turns per nut is enough to loosen and
remove the nuts.
3
CAUTION
Removing the top plate and adapter often causes spill-
ages and splashing. Remove any spillage on the floor
immediately to minimize the risk for slipping accidents.
Loosen the adapter top plate and lift out the adapter. Remove the adapter
without letting the stainless steel touch the glass.
Step Action
Step Action
Note:
BPG 450: Remove the net by twisting a spatula or similar stainless blunt
blade in the slot.
3 Remove the support net. Be careful not to scratch the adapter as this may
cause leakage.
4 Fit new net items according the order in the table Adapter items, on page 81.
Step Action
5 Wet the net with 20% ethanol to remove all air bubbles trapped in the net. If
the net is tight, warm it in 50°C water.
NOTICE
Wet the net with 20% ethanol to eliminate trapped air.
NOTICE
Make sure not to scratch any steel surfaces of the column when
removing seals and O-rings. If scratched, the sealing of the column
may be compromised and leakage may result.
Step Action
End-piece items
Torque wrench
Use the torque wrench listed in the following table when securing the column tube to
the end-piece. The torque apply for both ends of rods for BPG 450. For BPG 100 to 300
the rod is threaded into the bottom flange and the same torque apply as for the top.
Tighten the bolts in crosswise sequence according to figures below.
NOTICE
Tighten the nuts and bolts close to the given torque. But do not
exceed the given torque, this could damage the column. Always
use a torque wrench.
4
2
4 5
4
8
4 2
6
End-piece N/A N/A N/A N/A 8
7
domed nuts Nm
1 3
4 2
5 8
5 8
Step Action
1 Loosen the nuts on the flange cross-wise in two steps to avoid damage to
the column tube.
1 3
4
2
NOTICE
When the flange is removed, the column tube is unsup-
ported. Make sure that the column tube does not tip
over.
Step Action
4 Lift off the glass tube. Carefully place the glass tube on a secure surface.
5 Remove the bolts and washers securing the end-piece to the stand using a
spanner.
Step Action
NOTICE
Make sure not to scratch any steel surfaces of the
column when removing seals and O-rings. If scratched,
the sealing of the column may be compromised and
leakage may result.
Step Action
1 Fit new end-piece items according the order in table End-piece items, on
page 87.
Note:
Wet the net with 20% ethanol to eliminate trapped air.
Note:
BPG 450: Carefully place the column tube on the U-shaped seal. Make sure
the flange is centered on the U-shaped seal.
2 Secure the end-piece to the stand using the bolts and washers.
Step Action
Step Action
1 Carefully place the glass tube on the O-ring in the end-piece. Make sure that
the glass tube does not touch the stainless steel. Align the glass tube with
the bottom net.
Note:
BPG 450: Carefully place the column tube on the U-shaped seal. Make sure
the flange is centered on the U-shaped seal.
NOTICE
The glass tube can crack if touching any stainless steel.
2 Screw the support rods into the end-piece, see Torque wrench, on page 88.
4 Position the flange so that the support rods fit into the appropriate holes.
Step Action
5 Make sure that the O-ring on the flange is aligned with the glass tube.
NOTICE
Twisted or deformed O-rings will cause leakage.
6 Align the inner surface of the glass tube with the inner surface of the flange.
7 Fit the washers and the nuts on the support rods and fingertighten the nuts.
NOTICE
Do not exceed the given torque, this could damage the
column.
CAUTION
After assembly, check that the column is tight and not
leaking.
Step Action
NOTICE
Failure to use cross and stepwise tightening can damage
the column tube.
Step Action
1 Unscrew the retaining nut at the bottom of the valve with help of the handle.
3 Insert the positioning disc in the square of the tapered plug (the PTFE sealing
washer has to touch the body).
4 Insert the dished washer (the larger diameter has to touch the positioning
disc).
7 Stabilize the valve for 24 hours, minimum (as PTFE is a floating material).
8 Screw the retaining nut off 3/8 turn (the tightening torque of the retaining
nut is 1.5 Nm).
NOTICE
Do not pack the column to bed heights higher than the length of
the column, i.e., the adapter should seal against the column wall,
NOT against the extension tube wall. A tube length of 10 cm is
required when installing the adapter onto the column tube if the
packing extension is removed.
Step Action
1 Remove the adapter from the column and place the extension tube onto the
upper flange. Note that the column rods fit into the open holes in the flange
on the extension tube.
2 Make sure that the thin O-ring is positioned properly in the flange.
3 Use the washers and domed nuts from the column lid to secure the exten-
sion tube to the column.
4 The extended column is now ready for the first packing step in two-step
packing procedures.
Step Action
1 Remove the adapter from the column and place the extension tube onto the
upper flange.
2 Make sure that the thin O-ring is properly positioned in the flange.
3 Use the bolts delivered with the extension tube to secure the extension tube
to the column.
4 The extended column is now ready for the first packing step in two-step
packing procedures.
Step Action
3 Lift off the adapter top plate and adapter screws. Take care not to bend the
inlet tube.
Step Action
5 Unscrew and remove the cap screws that secure the stopper on the adapter
top plate with the hex key.
Step Action
7 Remove the stop screw and locking nut from the top plate.
8 Unscrew the bolts on the upper ring of the sealing unit. This will release the
sealing unit and the inner adapter tube.
Step Action
9 Carefully unscrew and remove the adjusting knob. This releases the tension
on the spring situated under the inner adapter tube.
Step Action
Step Action
1 Position the bushing ring(s) in the recess on the adapter top plate.
Note:
This image shows BPG 140 column but is applicable to other BPG columns.
2 Using the screws secure the adjuster nut insert to adjuster nut body through
the adapter top plate so that the adjuster nut body turns on the bushing ring.
Note:
This image shows BPG 140 column but is applicable to other BPG columns.
3 Slide the spring (c), inner adapter tube (b) and outer adapter tube (a) onto the
adapter plate.
Step Action
4 Fasten the bolts by hand to assemble the outer adapter tube to the adapter
plate.
Step Action
Torque:
BPG 100 4Nm
BPG 200 4Nm
BPG 300 8Nm
BPG 450 8Nm
Step Action
8 Turn the adjusting knob clockwise onto the top of the adapter until the holes
in the inner adjusting tube are level with the holes in the outer adapter tube.
(See Chapter 3 Column descriptions, on page 18 for respective size of the
adjusting knob.)
9 Use a flashlight and a screwdriver through one hole to lock the inner tube in
position. Then two of the adapter head screws can fasten the sealing unit in
place.
Step Action
10 Level the hole in the outer adapter tube with the inner adapter tube with the
help of a fine screwdriver.
Step Action
13 Locate the stop screw and locking nut in the top plate.
Step Action
14 Turn the stop screw until it engages the groove in the outer adaptor tube,
preventing sideways movement of the tube without preventing vertical
movement.
CAUTION
For continued protection against injury risks due to fluid jets, burst
pipes or explosive atmosphere, the piping system must be tested
for leakage at maximum pressure when the following situations
apply:
• After assembly or maintenance
• Before operation or Cleaning-In-Place (CIP)
Step Action
2 Position the adapter in the column to the same height as you intend to run
the column.
NOTICE
If the adapter O-ring seal is dry, it may be damaged by
forced movement of the adapter. Wet the O-ring with
water or 20% ethanol solution. Do NOT grease the O-
ring.
Step Action
6 Raise the pressure to maximum working pressure. Note that pressure spikes
may cause poor packing (cracking).
A pressure relief valve should be placed between the pump and column.
WARNING
The working pressure of the column should never
exceed its design pressure, otherwise there is a risk of
personal injury and damage to the column. Always use
appropriate safety equipment, for example pressure
relief valve or rupture discs.
8 Check the pressure drop on the pressure gauge. The pressure drop should
not exceed 0.1 bar during 5 minutes. A recorder linked to a pressure
controller gives more precise figures.
6.7 Cleaning
Regular cleaning
Regular cleaning of BPG columns is essential for reliable results. It is important that the
column is kept free from contamination such as microbiological growth and adsorbed
proteins.
Precautions
WARNING
EXPLOSION HAZARD! If cleaning the column in a potentially
explosive atmosphere, prevent static electric charges building up.
Do not rub excessively using a dry cloth and over large areas at a
time. Instead, clean the column using a damp cloth with slow
movements across small areas at a time.
NOTICE
Cleaning. Keep the exterior of the instrument dry and clean. Wipe
regularly with a soft damp tissue and, if necessary, a mild cleaning
agent. Let the instrument dry completely before use.
NOTICE
Make sure that any water is WFI, since particles can block and
damage the nets.
NOTICE
Do not leave O-rings or seals in hot water for longer than the time
period recommended in Cytiva product documentation.
Cleaning agents
All column parts can be cleaned with the most commonly used agents, such as deter-
gents, ethanol, weak acids, sodium hydroxide and high salt concentrations. For special
cleaning agents, please refer to Section 8.2 Chemical resistance, on page 132.
In time, even electropolished 316L stainless steel may show signs of corrosion, espe-
cially if it is in frequent contact with high salt concentrations. Should this occur, repas-
sivation of steel components is recommended and should correct the problem.
Cleaning-in-place (CIP)
One of the most important aspects of production chromatography is the maintenance
of the packed column by cleaning-in-place (CIP) procedures. Regular CIP of the packed
column between production batches is needed to ensure the proper product quality
and the expected media and equipment life. CIP removes precipitated material,
strongly bound substances, and other contaminants from the column bed without
dismantling the column.
After cleaning
After cleaning, restore the column to working conditions using the follow-up proce-
dure. All conditions refer to room temperature unless otherwise stated. If special main-
tenance routines are required, contact your local Cytiva representative for expert
advice. For affinity media, please contact your local Cytiva representative.
Autoclaving
To ensure effective autoclaving it is recommended that the column is completely
disassembled. All plastic components, support net, net, snap plug (BPG 450) and distri-
bution plate (BPG 300 and 450) as well as the wheels and plastic handle caps of the
adjuster nut must be disassembled from the column.
After autoclaving it is recommended that all process contact parts that have not been
autoclaved are replaced with new parts. Essentially seals and nets. Any plastic process
contact parts that are to be reused must be appropriately cleaned prior to reassembly
of the column.
NOTICE
The accessory valves have bodies in PTFE, which can shrink, and
thereby cause leakage after repeated autoclaving. Therefore,
exclude the valves if autoclaving the column.
6.8 Storage
Bacteriostatic solution
BPG columns packed with media should be stored in the bacteriostatic solution
recommended for the media. Refer to your media instructions for media-specific
recommendations. Refer also to Section 8.2 Chemical resistance, on page 132. Avoid
any solution that may compromise the quality of the media.
Follow the instructions below to apply bacteriostatic solution:
Step Action
1 Clean the column before applying bacteriostatic solution, for example with
0.5M NaOH, or other CIP solution recommended for the media.
3 Rinse with purified water after the CIP for at least five column volumes (CV).
4 Apply the bacteriostatic solution and the column is ready to be stored. See
the instructions below for storage of the BPG column.
Step Action
5 Seal all inlets and outlets as indicated in wet storage protocol above.
After storage
When the BPG column is going to be taken into operation after storage, follow the
instructions below:
Step Action
7 Troubleshooting
Troubleshooting guide
The column packing has not been Check the column packing.
checked.
The column has not been correctly equili- Equilibrate the column.
brated.
Old sample mixed with new. Check sample and sample
handling.
8 Reference information
In this chapter
8.1 Specifications
In this section
Column specifications
Bed volumes and bed heights are based on a slurry concentration of 75% and a packing
compression of 15%. The compression is the difference in volume between a sedi-
mented bed and a pressurized bed.
Note: The optimal slurry concentration is close to 50%.
Height
Height (H) used in this section refers to a BPG column with the adapter in its most
upper position.
Chemicals to be avoided
In general, the use of the following chemicals in BPG columns should be avoided:
• Extreme oxidizers (such as peroxides)
• Halogenated compounds
• Chlorinated solvents (such as methylene chloride)
• Esters (such as acetates)
• Aromatic hydrocarbons (such as toluene)
• High concentration of strong acids
• Chloroform
• THF
• Toluene
• Methylenechloride
NOTICE
Only use chemicals listed in the Chemical Resistance information.
The wetted parts of the product may be damaged by chemicals not
listed in the Chemical Resistance information. Contact your Cytiva
representative before using chemicals that are not listed.
Note: Avoid salt in combination with pH below 4.5. Always wash the column with at
least two column volumes of pH-neutral solution between the use of salt
and low pH buffers.
The table below is intended as a guide to the solvent resistance of the materials used in
BPG columns. The information has been compiled from published material from
several sources, not from individual tests on the components of the column. Conse-
quently, the user should use this list only as a guide to the level of chemical resistance
exhibited by each material. It should be noted that the effects of a chemical will be
more severe at higher temperatures and pressures, and that combined effects have
not been taken into consideration.
8.3 Abbreviations
The abbreviations used in this manual are defined as follows:
WARNING
Before decommissioning work is performed on the column, make
sure that:
• the column is empty and depressurized.
• relevant energy sources and process feed are disconnected.
• all process wetted areas are clean and decontaminated.
• the column cannot be accidentally re-energized, refilled or re-
contaminated during maintenance.
WARNING
Decommissioning. Decontaminate the equipment before
decommissioning to ensure that hazardous residues are removed.
Disposal
When taking BPG columns out of service, the different materials must be separated
and recycled according to national and local environmental regulations.
In this section
Manufacturing information
The table below summarizes the required manufacturing information.
Requirement Information
Name and address of manufacturer Cytiva Sweden AB
Björkgatan 30
SE 751 84 Uppsala
Sweden
Telephone number of manufacturer + 46 771 400 600
Introduction
This section describes regulatory information for the European Union and European
Economic Area that applies to the equipment.
CE marking
The CE marking and the corresponding EU Declaration of Conformity is valid for the
instrument when it is:
• used according to the Operating Instructions or user manuals, and
• used in the same state as it was delivered, except for alterations described in the
Operating Instructions or user manuals.
Service Ticket #:
To make the mutual protection and safety of Cytiva service personnel and our customers, all equipment and work areas must be
clean and free of any hazardous contaminants before a Service Engineer starts a repair. To avoid delays in the servicing of your
equipment, complete this checklist and present it to the Service Engineer upon arrival. Equipment and/or work areas
not sufficiently cleaned, accessible and safe for an engineer may lead to delays in servicing the equipment and could be subject to
additional charges.
Adequate space and clearance is provided to allow safe access for instrument service, repair or
installation. In some cases this may require customer to move equipment from normal operating location
prior to Cytiva arrival.
Consumables, such as columns or gels, have been removed or isolated from the instrument and from
any area that may impede access to the instrument .
Provide
explanation
for any “No”
answers here:
I hereby confirm that the equipment specified above has been cleaned to remove any hazardous substances and that the area
has been made safe and accessible.
Position or
Date (YYYY/MM/DD):
job title:
Signed:
Cytiva and the Drop logo are trademarks of Global Life Sciences IP Holdco LLC or an affiliate.
© 2020 Cytiva.
All goods and services are sold subject to the terms and conditions of sale of the supplying company operating
within the Cytiva business. A copy of those terms and conditions is available on request. Contact your local Cytiva
representative for the most current information.
To make sure the mutual protection and safety of Cytiva personnel, our customers, transportation personnel and our environment,
all equipment must be clean and free of any hazardous contaminants before shipping to Cytiva. To avoid delays in the processing of
your equipment, complete this checklist and include it with your return.
1. Note that items will NOT be accepted for servicing or return without this form
2. Equipment which is not sufficiently cleaned prior to return to Cytiva may lead to delays in servicing the equipment and
could be subject to additional charges
3. Visible contamination will be assumed hazardous and additional cleaning and decontamination charges will be applied
Yes No Specify if the equipment has been in contact with any of the following:
Radioactivity (specify)
Telephone No:
Ethanol
None, empty
Liquid Nitrogen
Other, specify
I hereby confirm that the equipment specified above has been cleaned to remove any hazardous substances and that
the area has been made safe and accessible.
Company or
Name:
institution:
Signed:
Cytiva and the Drop logo are trademarks of Global Life Sciences IP Holdco LLC or an affiliate. To receive a return authorization number
© 2020 Cytiva. or service number, call local
All goods and services are sold subject to the terms and conditions of sale of the supplying company operating technical support or customer service.
within the Cytiva business. A copy of those terms and conditions is available on request. Contact your local Cytiva
representative for the most current information.
Index
A Cleaning, 37, 115
agents, 115
Adapter, 76, 81, 82, 101, 106 Cleaning-in-place, 116
assembly, 106 Column, 20, 37, 39, 59, 72, 79, 87,
disassembly, 101 90, 94, 119, 123
moving, 76 assembly, 59
net, 81 blockage, 79
remove, 82 change O-ring, 94
Air, 54, 121 contaminated, 123
remove from column, 121 disassembly, 90
remove from top of adapter, leveling, 39
54 material, 20
Air trapped, 123 moving, 37
As, 72 nets, 87
ATEX directive, 21 performance evaluation, 72
Autoclaving, 118 storage, 119
Column leakage, 122
B Columns, 20, 34
floor load, 34
Bacteriostatic solution, 119 net, 20
Bed height, 50 space requirements, 34
BPG 100-140-200 columns illus- Compression factor, 50
trations, 22 Connections, 47
BPG 300 columns illustrations, 25 arrange, 47
BPG 450 columns illustrations, 28
BPG columns, 6, 19, 20, 31, 125 D
accessories, 31
design pressure, 19 Decontamination, 136
intended use, 6 Disposal, 136
material, 20 general instructions, 136
specifications, 125 Drain, 43
tubing, 20 symbol, 43
volumes, 19
E
C
Emergency, 17
Calculation, 73, 74 restart, 17
assymetry, 74 shutdown, 17
HETP, 73 Emergency stop, 17
CE, 139 End-piece items, 87
conformity, 139 Environment, 34
marking, 139 Evaluation, 72
Chemical resistance, 132 column performance, 72
Chromatogram, 73, 122 Explosive atmosphere, 21
CIP, 116 Explosive environment, 10
protocol, 116 precautions, 10
T
Tools, 36, 80
replacement, 80
unpacking, 36
Torque wrench, 88
Transport, 35
crate, 35
Troubleshooting, 121
U
Unpacking, 36
instructions, 36
W
Wheel, 38
lock, 38