Medical Isolation Transformer

Transformer data
Reference THX/THW4 THX/THW6.3 THX/THW8 THX/THW10
Rated power (kVA) 4000 6300 8000 10000
Rated frequency (Hz) 50..60 Hz 50..60 Hz 50..60 Hz 50..60 Hz
Rated input voltage (Vac) 230 230 230 230
Rated input current (In) 17,4 27,4 34,8 43,5
Rated output voltage (Vac) 115‐0‐115 115‐0‐115 115‐0‐115 115‐0‐115
Rated output current (In) 17,4 27,4 34,8 43,5
Inrush current IE ≤8In ≤8In ≤8In ≤8In
Leakage current <0.5mA <0.5mA <0.5mA <0.5mA
No‐load input current Io <3% <3% <3% <3%
No‐load output voltage Uo <235 <233 <233 <233
Short‐circuit voltage Uk ≤3% ≤3% ≤3% ≤3%
R primary Ω (20ºC) 0,182 0,086 0,070 0,055
R secondary Ω (20ºC) 0,238 0,099 0,092 0,062
Ambient temperature 40ºC 40ºC 40ºC 40ºC
temperature rise 70ºC 70ºC 70ºC 70ºC
Maximum temperature 110ºC 110ºC 110ºC 110ºC
Fe losses (W) 110ºC 31 46 46 47
Winding losses (W) 110ºC 167 194 262 310
Efficiency 95% 96% 96% 97%
Noise level <35dB <35dB <35dB <35dB

Transformer Enclosure

WEIGHTS AND DIMENSIONS


Rating A (mm) B (mm) C (mm) D (mm) E (mm) Weight (kg)
4 kVA 440 430 535 390 180 42
6.3 kVA 440 430 535 390 180 53
8 kVA 440 430 535 390 180 70
10 kVA 440 430 535 390 180 82
Medical Isolation Transformer
Starkstrom transformers are manufactured in accordance with IEC 61558‐2‐15, with the following additional
requirements:
• The leakage current of the output winding to earth and the leakage current of the enclosure, when measured in
no‐load condition and the transformer supplied at rated voltage and rated frequency, do not exceed 0.5 mA.
• Single‐phase transformers are used to form the medical IT systems for portable and fixed equipment and the
rated output is not less than 0.5 kVA and does not exceed 10 kVA.
• The medical isolation transformers are installed into the IPS panels at the manufacturer’s facility. The IPS system
is then tested, shipped, installed and commissioned as a complete system without any subsequent transformer
removal and re‐installation.

Heat Losses per Transformer


Starkstrom has invested in a range of transformers that
are more efficient over their entire working range, and
especially in the 0‐50% loading area where the majority
of the transformers operating life are spent.
Starkstrom transformers can generate energy savings of
up to 63% compared to other suppliers of medical
isolated transformers (see 10KVA comparison graph).
This has the benefits of reducing direct energy costs and
reducing the cost of removing the excess heat.

SIZE NO LOAD HEAT LOSSES FULL LOAD HEAT LOSSES EFFICIENCY


4 kVA 37W 146W 96%
6.3 kVA 53W 254W 96%
8 kVA 50W 270W 97%
10 kVA 56W 345W 97%

Starkstrom medical isolation transformers have been engineered to give a maximum of 8 times “in rush”. We would
suggest considering the following MCB’s to allow adequate circuit protection. However, this must be confirmed with
your project electrical designer.
• 4kVA Transformer 40 Amp ‘C’ Curve
• 6.3kVA Transformer 50 Amp ‘C’ Curve
• 8kVA Transformer 63 Amp ‘C’ Curve
• 10kVA Transformer 63 Amp ‘D’ Curve