Effect of Clutter Filter in High-Frame-Rate Ultrasonic Backscatter Coefficient Analysis

1
Faculty of Engineering, University of Toyama, Toyama 930-8555, Japan
2
School of Medicine, Kanazawa Medical University, Kanazawa 920-0293, Japan
3
Center for Frontier Medical Engineering, Chiba University, Chiba 263-8522, Japan
*
Authors to whom correspondence should be addressed.
Sensors 2023, 23(5), 2639; https://doi.org/10.3390/s23052639
Submission received: 20 January 2023 / Revised: 15 February 2023 / Accepted: 25 February 2023 / Published: 27 February 2023

Abstract

:
High-frame-rate imaging with a clutter filter can clearly visualize blood flow signals and provide more efficient discrimination with tissue signals. In vitro studies using clutter-less phantom and high-frequency ultrasound suggested a possibility of evaluating the red blood cell (RBC) aggregation by analyzing the frequency dependence of the backscatter coefficient (BSC). However, in in vivo applications, clutter filtering is required to visualize echoes from the RBC. This study initially evaluated the effect of the clutter filter for ultrasonic BSC analysis for in vitro and preliminary in vivo data to characterize hemorheology. Coherently compounded plane wave imaging at a frame rate of 2 kHz was carried out in high-frame-rate imaging. Two samples of RBCs suspended by saline and autologous plasma for in vitro data were circulated in two types of flow phantoms without or with clutter signals. The singular value decomposition was applied to suppress the clutter signal in the flow phantom. The BSC was calculated using the reference phantom method, and it was parametrized by spectral slope and mid-band fit (MBF) between 4–12 MHz. The velocity distribution was estimated by the block matching method, and the shear rate was estimated by the least squares approximation of the slope near the wall. Consequently, the spectral slope of the saline sample was always around four (Rayleigh scattering), independently of the shear rate, because the RBCs did not aggregate in the solution. Conversely, the spectral slope of the plasma sample was lower than four at low shear rates but approached four by increasing the shear rate, because the aggregations were presumably dissolved by the high shear rate. Moreover, the MBF of the plasma sample decreased from 36 to 49 dB in both flow phantoms with increasing shear rates, from approximately 10 to 100 s−1. The variation in the spectral slope and MBF in the saline sample was comparable to the results of in vivo cases in healthy human jugular veins when the tissue and blood flow signals could be separated.

1. Introduction

The viscoelasticity of blood has been noted as a hemodynamic factor for impaired vascular function [1]. The flow field of blood (velocity gradient, i.e., shear rate) is affected by the hemodynamic property of blood [2,3]. The most dominant component of blood is the red blood cell (RBC), and the flow pattern depends on its behavior. The backscattered signal from the RBC is related to the biochemical properties of the blood sample, e.g., the hematocrit and the level of aggregation. The interaction of the RBC with plasma proteins results in the formation of RBC aggregates, which increase the flow resistance and, thus, increase the viscosity of the blood at a low shear rate. Hence, the relationship between the physical properties and flow patterns of RBCs must be comprehended. The viscoelasticity of blood has been noted as a hemodynamic factor for impaired vascular function [1]. The flow field of blood (velocity gradient, i.e., shear rate) is affected by the hemodynamic property of blood [2,3]. The most dominant component of blood is the red blood cell (RBC), and the flow pattern depends on its behavior. The backscattered signal from the RBC is related to the biochemical properties of the blood sample, e.g., the hematocrit and the level of aggregation. The interaction of the RBC with plasma proteins results in the formation of RBC aggregates, which increase the flow resistance and, thus, increase the viscosity of the blood at a low shear rate. Hence, the relationship between the physical properties and flow patterns of RBCs must be comprehended.
High-frame-rate imaging with a clutter filter can clearly visualize blood flow signals and provide more efficient discrimination with tissue signals. The tissue signals are commonly removed using a high-pass finite or infinite impulse response filters [4]. The idea behind the clutter filter is that the low and high temporal frequency components are echoes from slowly moving tissues and blood flow signals, respectively. Singular value decomposition (SVD) is one of the advanced techniques for the clutter filter design [5,6,7,8]. An SVD-based clutter filter improves the extraction of tiny vessels with slow velocities, in addition to the sensitivity in clutter rejection between slow tissues and blood flow signals [9,10].
The degree of the RBC aggregation can be evaluated in high-frequency ultrasound by analyzing the frequency dependence of the ultrasonic backscatter signals, i.e., backscatter coefficient (BSC) [11,12,13,14]. The attenuation coefficient of blood has been measured through preliminary in vivo study in the frequency range of 10–45 MHz [15]. A pilot clinical study of quantitative BSC analysis has been demonstrated for the RBC aggregation measurement [16]. Conventionally focused line-by-line imaging with frame rates of several 10 s to 100 Hz is enough to observe moving RBCs at a low shear rate, and these studies have mainly focused on the hemorheology under peripheral circulation. An in vitro study has revealed the efficiency of high-frame-rate plane wave imaging with a 7.5-MHz ultrasound to characterize static tissue [17] and flowing RBCs in the agar phantom, i.e., in the clutter-less case [18]. Our previous study has also analyzed spatial features such as the contrast of flowing blood in in vitro experiments and in in vivo measurements of human jugular veins [19]. For in vivo imaging in the central vasomotor, such as a carotid artery and jugular vein, the clutter filter is required to visualize echoes from RBCs by suppressing tissue signals. However, the clutter filter has not been considered in the BSC analysis so far.
The novelty of this study is to develop non-invasive hemorheological imaging based on the BSC analysis for in vivo situations. In this study, the performance of the clutter filter was experimentally evaluated to indicate the hemorheological property dependently on a shear rate under physiological conditions. The spectral BSC analysis of porcine blood in a flow phantom without or with clutter signals was analyzed from low to high steady shear rates that were comparable with the human jugular vein [20]. Secondly, in vivo experiments in jugular veins of healthy subjects were carried out to confirm the feasibility under physiological flow conditions. The reproducibility and feasibility of high-frame-rate spectral analysis with clutter filters were compared among inter-subject variance.

2. Materials and Methods

2.1. In Vitro Porcine Blood Measurement

Whole porcine blood (within 24 h of its collection) anticoagulated by sodium citrate was centrifuged and separated into RBCs, plasma, and platelet at room temperature (24 °C), as followed in previous studies [18,21,22]. The RBCs were diluted in phosphate-buffered solution (PBS) or autologous plasma (40% hematocrit) to reproduce dispersed or aggregated RBCs. Each sample was circulated at a steady flow in the cylindrical lumen surrounded by 2% agar (i.e., clutter-less) or 2% agar-7% graphite (i.e., clutter-rich) flow phantoms as illustrated in Figure 1. Acoustic properties of clutter-less and clutter-rich phantoms were as follows: speed of sound (SoS)—1500 and 1520 m/s; attenuation coefficient—0.05 and 0.52 dB/cm/MHz. The blood of 0.8 L was passed through the wall-less lumen (8 mm diameter and 350 mm length) in the flow phantom using a peristaltic pump (Model 07528-10, Masterflex). The distance from the phantom surface to the center lumen was 15 mm. The steady flow was controlled with a flow regulator (CY-1208, ASOH) and monitored with a Coriolis flowmeter (FD-SF8, Keyence). For in vitro experiment, ultrasound data were obtained at selected flow rates from 10 to 600 mL/min.

2.2. In Vivo Jugular Vein Measurement

All procedures were carried out following the ethical standards of the responsible committee on human experimentation and with the Helsinki Declaration of 1964 and later versions. The Ethics Committee of Toyama University Hospital approved the study protocol (IRB# R2015150 and R2019135). Informed consent was acquired from all subjects. The jugular veins of healthy subjects ( n   = 5, 21–49 y.o., male) were scanned along the long axis in vivo. All subjects were in the supine position and breathed normally during the measurement.

2.3. Data Acquisition and Post-Processing

A 7.5-MHz linear array probe (UST-5412, Fujifilm) was placed over the middle of the flow phantom so that the lumen along the long axis of the human jugular vein reached its maximum. Multiple plane waves at 5 angles of −10, −5, 0, 5, and 10 degrees were transmitted in each frame at pulse repetition periods of 96 μs, i.e., frame rate of about 2 kHz for compounded images. Tukey apodization with a coefficient of 0.4 was used in transmission. The radio frequency (RF) channel data were acquired at a sampling frequency of 31.25 MHz for 0.1 s (porcine blood experiment) and 0.96 s (in vivo measurement) using a research platform scanner (RSYS0016, Microsonic). The delay-and-sum method with the dynamic aperture (F-number of 1) was applied to the RF channel data [23]. The beamformed RF data (axial 24 mm × lateral 24 mm, 893 × 242 pixels) was reconstructed on a pixel-by-pixel basis (axial 25 μm × lateral 100 μm) at an SoS of 1540 m/s. The coherently compounded RF data were used for the following analysis.

2.4. Clutter Filter

A spatiotemporal filter based on SVD was applied to the beamformed RF data to emphasize the blood flow signal [9]. Briefly, a spatiotemporal matrix (spatial 216,106 pixels × temporal 200 (porcine blood experiment) or 2000 (in vivo measurement) frames) composed of the coherently compounded beamformed signals in all frames was used for the SVD filter. The spatiotemporal matrix S is decomposed by SVD into a product of three matrices as
S = U ^ Σ ^ V ^ T ,
where U ^ and V ^ are matrices in terms of spatial and temporal singular vectors, respectively, Σ ^ is a diagonal matrix of singular values with descending order, and T is the transpose operation. The clutter-filtered signal S is defined as
S = U ^ Σ ^ V ^ T ,
where Σ ^ is a diagonal matrix in which the components in the matrix of Σ ^ at orders less than the lower threshold and those higher than the higher threshold are replaced with zeros. The low- and high-rank thresholds of singular values for clutter, blood flow, and system noise components were empirically set among the porcine blood experiments (from 67 to 27 dB normalized by the maximum singular value from a reference phantom in Section 2.5) and in vivo (from 67 to 37 dB similar to the normalization criteria in the porcine experiments) measurements, respectively, in reference to the power curve of singular values, as shown in Figure 2. The 1st inflection point was assumed to be the boundary between clutter and blood flow signals, and 2nd inflection point was the boundary between blood flow signal and system noise.

2.5. Backscatter Coefficient Analysis

The BSC of the RBCs was calculated using the reference phantom method [24,25,26] in the frequency domain expressed as
B S C ( f , d ) = P ( f , d ) ¯ P r e f ( f , d ) ¯ B S C r e f ( f ) exp [ 4 d f ( α α r e f ) 8.686 ] ,
where P ¯ and P r e f ¯ are the mean of power spectra obtained from the RBCs and a reference medium, respectively; B S C r e f is the theoretical BSC, d is the distance between the transducer and central position of the analysis window; and α and α r e f are the attenuation coefficients of the RBCs and the reference medium, respectively. The window size for calculation of the mean of the power spectra was 128 × 30 pixels (3.2 × 3.0 mm2, with 10 wavelengths in axial × 5-point spread functions in lateral direction under the SoS of 1540 m/s and center frequency of 7.5 MHz), and then ensemble averaging was performed for 4 frames (1.92 ms). Adjacent analysis windows overlapped by 80% in both the axial and lateral directions.
For in vitro measurement data, the reference medium consisted of a suspension of the RBC in saline at a low hematocrit [27]. In this study, a PBS sample at a hematocrit of 3% flowing in the geometry, as in Figure 1, in the clutter-less and clutter-rich phantoms at a flow rate of 10 mL/min, was used for the reference medium. The reference phantom used for in vivo analysis was made from 96.6 wt% purified water, 2 wt% agar, 0.9 wt% dispersant, and 0.5 wt% polyamide particles with a diameter of 10 μm (Orgasol 2002 EXD NAT1, Arkema). The means of SoS and attenuation coefficient of the reference phantom were 1503 m/s and 0.10 dB/cm/MHz, respectively, in the reflector method [17,28] using a single element plane transducer (V312, Olympus), as well as the backscatter power comparable to human blood [29].
In the attenuation compensation function, the constants α and α r e f were obtained using a typical attenuation coefficient: 0.30 dB/MHz for in vitro study determined by the reflector measurement in addition to considering the literature value [30]; 0.50 dB/cm/MHz in the tissue [31]; 0.15 dB/cm/MHz in the blood [32] for in vivo study. B S C r e f for in vitro measurement was the theoretical backscatter of human blood as defined by 5 × 10 31   f 4   m 1   sr 1 [33]. B S C r e f for in vivo reference phantom was calculated based on the Faran model [34]. Here, the B S C r e f was calculated using the known parameters of the Faran model: particle diameter = 10 μm; volume fractions = 0.5%; SoS in particle’s material = 2300 m/s; SoS in the surrounding medium = 1500 m/s; particle density = 1.03 g/cm3; medium density = 1.0 g/cm3; and Poisson ratio = 0.42.
The frequency dependence of the BSC was also evaluated by fitting a function to the calculated BSC in the frequency range from 4 to 12 MHz. The fitting function is expressed as
10 log 10 [ B S C ( f , d ) ] n log 10 f + 10 log 10 b
Equation (4) could be modeled by a line ( y = n x + B ) having spectral slope n and intercept B in the least-square method, where B = 10 log 10 b . It means that the curve fitting was carried out in log scale. Also, the ideal estimate of n of the RBC has been expected to be 4 dB/MHz. In addition, the mid-band fit (MBF) was computed as the magnitude of the BSC, i.e., 10 log 10 [ B S C ( 7.5   MHz , d ) ] , at the frequency of 7.5 MHz.

2.6. Shear Rate Estimation

A block matching analysis was carried out to calculate the blood flow velocity. The input of the block matching analysis was the amplitude envelope of the coherently compounded beamformed signal. The block size was 60 × 60 pixels (1.4 × 6.0 mm2 for the axial and lateral directions), and the search distance in both directions was 40 pixels in reference to our previous numerical and experimental studies [29,35]. Adjacent macro blocks overlapped each other by 0.5 and 1.0 mm in the axial and lateral directions, respectively. The normalized cross-correlation function was up-sampled using the reconstructive interpolation method to estimate a sub-sample displacement [36,37]. Both the lateral and axial interpolation factors were 4. Also, the ensemble average of the correlation function was performed for 1.92 ms (4 frames).
The slope of flow velocity profile along the axial direction was calculated as the shear rate in the neighborhood around the proximal and distal boundaries between vein wall and lumen, which were manually traced. The gate length was 3 mm from each boundary in each beam line to estimate the slope by the least-squares method.

3. Results

Figure 3a–c show examples of coherently compounded B-mode images in the clutter-less phantom without the clutter filter and clutter-rich phantoms without or with filters in the case of low flow rate (50 mL/min). Ensemble averaging was applied among 4 frames. Note that mean velocity and shear rate were from 3 to 4 cm/s and from 10 to 20 s−1 at the minimum to maximum, respectively. In addition, Figure 3(1)–(2) compare PBS and plasma samples, respectively. The echogenicity of the plasma sample was higher; however, the PBS sample had the same echogenicity that was independent of the shear rate. Conversely, B-mode images in the case of high flow rate (450 mL/min, mean velocity of 12–13 cm/s, and shear rate of 90–100 s−1) were equivalent when compared between PBS and plasma samples, as is shown in Figure 4.
Figure 5 shows the typical BSC of the PBS and plasma samples in the clutter-less and clutter-rich phantoms at selected flow rates of 10, 50, 150, 250, 350, and 450 mL/min. The standard deviation is unmarked for clarity. The magnitude of the BSC in the plasma sample increased with a lower shear rate in both phantoms. In contrast, the magnitude of the BSC in the PBS sample was constant around the theoretical BSC of human blood [33], which was irrelevant to the flow rate. Additionally, as is seen in Figure 6a, the spectral slope of the PBS sample was always around 4 dB/MHz, i.e., Rayleigh scattering was independent of the shear rate, because the RBCs did not aggregate in the PBS solution. On the other hand, the spectral slope of the plasma sample, especially in the clutter-rich phantoms, was lower than four at low shear rates (2.2 dB/MHz at 25 s−1) but approached four by increasing the shear rate, because the aggregations were presumably dissolved by the high shear rate. Also, the MBF of the plasma sample decreased from 36 to 49 dB with an increasing shear rate, as is seen in Figure 6b. Conversely, the MBF of the PBS sample was constant around 52 to 49 dB and was unrelated to the shear rate.
Typical B-mode images of in vivo jugular veins at selected low and high shear rates within intra-subjects are displayed in Figure 7a,b. Ensemble averaging was also applied among four frames to emphasize the signal-to-noise ratio. B-mode images were observed as the horizontal blurring effect of the diffusive suspensions; however, the echogenicity of blood flow could be clearly confirmed. Temporal variations in the spectral slope, MBF, shear rate (mean and standard deviation), and mean velocity are illustrated in Figure 8. Standard deviations in the BSC features and velocity are unmarked for clarity. For five healthy subjects, the spectral slope and MBF in Figure 8a,b were distributed around 2.8 to 5.0 dB/MHz and 53 to 46 dB within the approximate accelerating phase of the shear rate (e.g., increasing from less than 50 s−1 to over 100 s−1) through 0.2 to 0.4 s and the hereafter phase until around 0.8 s. In contrast, the spectral slope rapidly increased to over 5.0 dB/MHz at around 0.1 and 0.9 s (constant phase of the low shear rate and velocity) and exhibited a low MBF of less than 55 dB.

4. Discussion

This study initially evaluated the effect of a clutter filter in high-frame-rate ultrasonic backscatter coefficient analysis for blood characterization. In the porcine blood experiment, the property of the RBC aggregation was visualized in a low shear rate (around < 50 s−1) by comparing PBS and plasma samples, as is shown in Figure 3. The RBC aggregation by force between the cell and protein in the overall plasma would induce increasing echogenicity in accordance with previous studies [18,21,38,39]. In contrast, the RBC disaggregation that was caused by the high shear was confirmed through the lower echogenicity of both samples in Figure 4. These findings in the clutter-rich phantom were consistent with the previous studies on the dependence of ultrasonic backscatter power on the shear rate and hemorheology in the case of a clutter-less geometry [18,21,38,39]. In addition, the BSC features, such as spectral slope and MBF, could be quantitatively evaluated in the different shear rates, as is shown in Figure 6.
For five in vivo subjects within the accelerating and deaccelerating phases of shear rate, such as from 0.2 to 0.8 s, the spectral slope and MBF were comparable to those in in vitro experiments of PBS sample, and low shear plasma sample as can be seen in Figure 6 and Figure 8. The variability in BSC features among inter-subjects would be induced by the assumption of same attenuation and by the level of clutter rejection. Further investigation will be necessary to conduct an extensive clinical study to clarify the findings more explicitly in this preliminary study. Although the spectral slope and MBF in 3 of 5 subjects (#1 to #3) were also steady temporally, those in 2 of 5 subjects (#4 and #5) were rapidly changed in the constant phase of shear rate, such as around 0.1 and 0.9 s. Figure 9 displays typical B-mode images and spatial distributions of the spectral slope and MBF with vector velocity and shear rate. The echogenicity inside the vein was lower in Figure 9a, because the blood signals were suppressed by the clutter filter due to the near velocity or echogenicity between tissue and blood flow.
We further investigated the relationship of frequencies (in the clutter filter and BSC feature) between temporal and axial directions. The main temporal frequency was calculated by the column vectors v ^ n in the temporal singular matrix V ^ = [ v ^ 0   v ^ 1   v ^ 2     v ^ N f r a m e 1 ] , which corresponds to the temporal basis functions rejected by the SVD filter as mentioned in our previous study [40]. Briefly, the main temporal frequency ω ¯ S V D was estimated as
ω ¯ S V D = ω ω | R ^ ( ω ) | ω | R ^ ( ω ) | ,
where R ^ ( ω ) is the frequency spectrum of r ^ ( t ) along the temporal direction t . The vector r ^ of the rejected components is obtained as
r ^ = n = 0 N l o w r a n k σ n v ^ n ,
where σ n is the n-th singular value. While the components of the vector r ^ are temporal samples of the rejected components, they are redefined as r ^ r ^ ( t ) .
Figure 10 shows the mean frequency ω ¯ S V D in the porcine blood data, as is shown in Figure 2a. The components with low frequency under several 100 Hz (in Figure 10) were suppressed by the clutter filter at the cut-off value of approximately 25 to 35 dB (the second and the following plots from a plot at around 0 dB). Furthermore, low frequency components around 4 to 8 MHz (in the axial direction) dominantly decreased with decreasing cut-off values of the SVD filter, as are shown in the frequency spectra P ( f ) of Figure 11. For that reason, a high spectral slope and low MBF occurred in in vivo jugular vein in the constant phase of the shear rate. Such features were also confirmed in the porcine blood data, as are shown in Figure 12, Figure 13 and Figure 14. Typical B-mode images at selected flow rates of 10 and 350 mL/min are visualized in Figure 12 and Figure 13. Some different cut-offs of high singular values were selected in reference to Figure 2. At the flow rate of 350 mL/min (mean velocity and shear rate around 12 cm/s and 100 s−1), the clutter filter presumably distinguished between blood flow and tissue signals. Hence, the spectral slope was constant around 4 dB/MHz, independent of the cut-off property of singular values, as is shown in Figure 14b. However, the change in echogenicity occurred at the flow rate of 10 mL/min (mean velocity and shear rate around 1 cm/s and 10 s−1) for the different singular cut-off values. Additionally, the spectral slopes of the PBS and plasma samples increased from 2.2 to 6.0 dB/MHz with decreasing singular cut-off values in Figure 14a. Hence, the limitation of the condition of the clutter filter should be calibrated if the level of clutter rejection is uncertain. The criteria based on mean velocity, shear rate, and contrast might be necessary to reject the evaluation frame affected by the clutter filter. Another scenario is to measure with more wide-band BSC toward high frequency. Although the relationship between frequency range and penetration depth for this vascular imaging must be proven, high frequency ultrasound should emphasize backscattering power from the RBCs.
While the plasma sample in the low shear case presented the changes in the spectral slope and MBF in Figure 6, the physiological states were not confirmed in healthy subjects in in vivo situations. Diabetes [2], coronary artery disease [41], thrombosis [42], and hyperlipidemia [43] are associated with abnormally high levels of RBC aggregation. Chronic lower shear rates and flow disturbance in aging patients with diabetes might be distinct from the change in the hemorheological property, such as for the RBC aggregation and thrombus [44,45]. The RBC aggregation has been successively measured using line-by-line focused imaging in vivo in diabetic patients [46]. One of the potentials of high-frame-rate backscatter coefficient analysis for clinical applications will be conducted on a non-invasive quantitative analysis for anticoagulant therapy for diabetes subjects in future works.

5. Conclusions

The effect of a clutter filter in high-frame-rate ultrasonic BSC analysis was confirmed in the porcine blood and in in vivo jugular vein measurements. For the porcine blood experiment, the characteristics of the RBCs, such as aggregation and disaggregation, were evaluated as the change in the spectral slope and MBF of the BSC through the surrounding tissues with different clutter levels. The feasibility and reproducibility of this analysis were also compared in in vivo jugular veins in healthy subjects when the tissue and blood flow signals could be separated. High-frame-rate imaging with the clutter filter has the potential to characterize blood by means of the BSC analysis. In future works, the criteria based on flow features and image quality, such as the contrast, will provide feedback regarding the condition of the clutter filter to develop a robust analysis of the BSC.

Author Contributions

Conceptualization, M.O., K.Y. (Kunimasa Yagi), T.Y. and H.H.; methodology, M.O.; software, M.O.; validation, M.O. and K.Y. (Kunimasa Yagi); formal analysis, M.O.; investigation, M.O., K.Y. (Kunimasa Yagi) and H.H.; resources, M.O., K.Y. (Kunimasa Yagi), R.N., K.Y. (Kenji Yoshida), T.Y. and H.H.; data curation, M.O.; writing—original draft preparation, M.O.; writing—review and editing, M.O., K.Y. (Kunimasa Yagi), R.N., K.Y. (Kenji Yoshida), T.Y. and H.H., visualization, M.O.; supervision, K.Y. (Kunimasa Yagi), R.N., T.Y. and H.H.; project administration, M.O., K.Y. (Kunimasa Yagi) and H.H.; funding acquisition, M.O., T.Y. and H.H. All authors have read and agreed to the published version of the manuscript.

Funding

This work was partially supported by the JSPS Moonshot R&D Grant Number JPMJMS2021, the JSPS Core-to-Core Program JPJSCCA20170004, and the KAKENHI Grant Numbers 18KK0110 and 20J01391.

Institutional Review Board Statement

The study was conducted in accordance with the Declaration of Helsinki and approved by the Institutional Ethics Committee of the University of Toyama (IRB# R2015150 and R2019135).

Informed Consent Statement

Informed consent was obtained from all subjects involved in the study.

Data Availability Statement

Not applicable.

Conflicts of Interest

The authors declare no conflict of interest.

References

  1. Thurston, G.B. Viscoelasticity of Human Blood. Biophys. J. 1972, 12, 1205–1217. [Google Scholar] [CrossRef] [Green Version]
  2. Schmid-Schonbein, H.; Volger, E. Red-Cell Aggregation and Red-Cell Deformability in Diabetes. Diabetes 1976, 25, 897–902. [Google Scholar] [PubMed]
  3. Sugaya, K.; Kadekawa, K.; Unten, Y.; Nishijima, S.; Ashitomi, K.; Yamamoto, H. Relationship of Blood Flow in the Common Iliac Vein to Lower Urinary Tract Disease. J. Med. Ultrason. 2019, 46, 223–229. [Google Scholar] [CrossRef] [PubMed]
  4. Bjærum, S.; Torp, H.; Kristoffersen, K. Clutter Filter Design for Ultrasound Color Flow Imaging. IEEE Trans. Ultrason. Ferroelectr. Freq. Control 2002, 49, 204–216. [Google Scholar] [CrossRef] [PubMed]
  5. Ledoux, L.A.F.; Brands, P.J.; Hoeks, A.P.G. Reduction of the Clutter Component in Doppler Ultrasound Signals Based on Singular Value Decomposition: A Simulation Study. Ultrason. Imaging 1997, 19, 1–18. [Google Scholar] [CrossRef] [PubMed]
  6. Baranger, J.; Arnal, B.; Perren, F.; Baud, O.; Tanter, M.; Demene, C. Adaptive Spatiotemporal SVD Clutter Filtering for Ultrafast Doppler Imaging Using Similarity of Spatial Singular Vectors. IEEE Trans. Med. Imaging 2018, 37, 1574–1586. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  7. Zhu, Y.; Kim, M.W.; Hoerig, C.; Insana, M.F. Experimental Validation of Perfusion Imaging with HOSVD Clutter Filters. IEEE Trans. Ultrason. Ferroelectr. Freq. Control 2020, 67, 1830–1838. [Google Scholar] [CrossRef]
  8. Hasegawa, H.; Omura, M.; Nagaoka, R. On the Investigation of Autocorrelation-Based Vector Doppler Method With Plane Wave Imaging. IEEE Trans. Ultrason. Ferroelectr. Freq. Control 2022, 69, 1301–1311. [Google Scholar] [CrossRef]
  9. Demené, C.; Deffieux, T.; Pernot, M.; Osmanski, B.F.; Biran, V.; Gennisson, J.L.; Sieu, L.A.; Bergel, A.; Franqui, S.; Correas, J.M.; et al. Spatiotemporal Clutter Filtering of Ultrafast Ultrasound Data Highly Increases Doppler and FUltrasound Sensitivity. IEEE Trans. Med. Imaging 2015, 34, 2271–2285. [Google Scholar] [CrossRef]
  10. Huang, C.; Song, P.; Trzasko, J.D.; Gong, P.; Lok, U.W.; Tang, S.; Manduca, A.; Chen, S. Simultaneous Noise Suppression and Incoherent Artifact Reduction in Ultrafast Ultrasound Vascular Imaging. IEEE Trans. Ultrason. Ferroelectr. Freq. Control 2021, 68, 2075–2085. [Google Scholar] [CrossRef]
  11. Fukushima, T.; Hasegawa, H.; Kanai, H. Estimation of Scatterer Diameter by Normalized Power Spectrum of High-Frequency Ultrasonic RF Echo for Assessment of Red Blood Cell Aggregation. Jpn. J. Appl. Phys. 2011, 50, 07HF02. [Google Scholar] [CrossRef] [Green Version]
  12. Garcia-Duitama, J.; Chayer, B.; Garcia, D.; Goussard, Y.; Cloutier, G. Protocol for Robust In Vivo Measurements of Erythrocyte Aggregation Using Ultrasound Spectroscopy. Ultrasound Med. Biol. 2017, 43, 2871–2881. [Google Scholar] [CrossRef]
  13. Higashiyama, K.; Mori, S.; Arakawa, M.; Yashiro, S.; Ishigaki, Y.; Kanai, H. Estimation of Aggregate Size of Red Blood Cell by Introducing Reference Power Spectrum Measured for Hemispherical Ultrafine Wire. Jpn. J. Appl. Phys. 2022, 61, SG1046. [Google Scholar] [CrossRef]
  14. Nagasawa, K.; Fukase, A.; Mori, S.; Arakawa, M.; Yashiro, S.; Ishigaki, Y.; Kanai, H. Evaluation Method of the Degree of Red Blood Cell Aggregation Considering Ultrasonic Propagation Attenuation by Analyzing Ultrasonic Backscattering Properties. J. Med. Ultrason. 2021, 48, 3–12. [Google Scholar] [CrossRef] [PubMed]
  15. Arakawa, M.; Higashiyama, K.; Mori, S.; Yashiro, S.; Ishigaki, Y.; Kanai, H. In Vivo Measurement of Attenuation Coefficient of Blood in a Dorsal Hand Vein in a Frequency Range of 10–45 MHz: A Preliminary Study. Front. Phys. 2023, 11, 1077696. [Google Scholar] [CrossRef]
  16. Chayer, B.; Allard, L.; Qin, Z.; Garcia-Duitama, J.; Roger, L.; Destrempes, F.; Cailhier, J.F.; Denault, A.; Cloutier, G. Pilot Clinical Study of Quantitative Ultrasound Spectroscopy Measurements of Erythrocyte Aggregation within Superficial Veins. Clin. Hemorheol. Microcirc. 2020, 74, 109–126. [Google Scholar] [CrossRef] [Green Version]
  17. Omura, M.; Hasegawa, H.; Nagaoka, R.; Yoshida, K.; Yamaguchi, T. Validation of Differences in Backscatter Coefficients among Four Ultrasound Scanners with Different Beamforming Methods. J. Med. Ultrason. 2020, 47, 35–46. [Google Scholar] [CrossRef]
  18. Garcia-Duitama, J.; Chayer, B.; Han, A.; Garcia, D.; Oelze, M.L.; Cloutier, G. Experimental Application of Ultrafast Imaging to Spectral Tissue Characterization. Ultrasound Med. Biol. 2015, 41, 2506–2519. [Google Scholar] [CrossRef]
  19. Omura, M.; Yagi, K.; Nagaoka, R.; Hasegawa, H. Contrast Analysis in Ultrafast Ultrasound Blood Flow Imaging of Jugular Vein. J. Med. Ultrason. 2023; in press. [Google Scholar] [CrossRef]
  20. Samijo, S.K.; Willigers, J.M.; Barkhuysen, R.; Kitslaar, P.J.E.H.M.; Reneman, R.S.; Brands, P.J.; Hoeks, A.P.G. Wall Shear Stress in the Human Common Carotid Artery as Function of Age and Gender. Cardiovasc. Res. 1998, 39, 515–522. [Google Scholar] [CrossRef] [Green Version]
  21. Yuan, Y.W.; Shung, K.K. Ultrasonic Backscatter from Flowing Whole Blood. I: Dependence on Shear Rate and Hematocrit. J. Acoust. Soc. Am. 1988, 84, 52–58. [Google Scholar] [CrossRef] [PubMed]
  22. Yu, F.T.H.; Franceschini, É.; Chayer, B.; Armstrong, J.K.; Meiselman, H.J.; Cloutier, G. Ultrasonic Parametric Imaging of Erythrocyte Aggregation Using the Structure Factor Size Estimator. Biorheology 2009, 46, 343–363. [Google Scholar] [CrossRef] [PubMed]
  23. Rodriguez-Molares, A.; Rindal, O.M.H.; Bernard, O.; Nair, A.; Bell, M.A.L.; Liebgott, H.; Austeng, A.; LØvstakken, L. The UltraSound ToolBox. In Proceedings of the 2017 IEEE International Ultrasonics Symposium (IUS), Washington, DC, USA, 6–9 September 2017; pp. 1–4. [Google Scholar]
  24. Yao, L.X.; Zagzebski, J.A.; Madsen, E.L. Backscatter Coefficient Measurements Using a Reference Phantom to Extract Depth-Dependent Instrumentation Factors. Ultrason. Imaging 1990, 12, 58–70. [Google Scholar] [CrossRef] [PubMed]
  25. Omura, M.; Yoshida, K.; Akita, S.; Yamaguchi, T. High-Frequency Ultrasonic Backscatter Coefficient Analysis Considering Microscopic Acoustic and Histopathological Properties of Lymphedema Dermis. Jpn. J. Appl. Phys. 2020, 59, SKKE15. [Google Scholar] [CrossRef]
  26. Oguri, T.; Omura, M.; Saito, W.; Yoshida, K.; Yamaguchi, T. Stability of Backscattering Coefficient Evaluation with Clinical Ultrasound Scanner in Homogeneous Medium When Sound Field Characteristics Differ from Reference Signal. Jpn. J. Appl. Phys. 2021, 60, SDDE24. [Google Scholar] [CrossRef]
  27. Wang, S.-H.; Shung, K.K. An Approach for Measuring Ultrasonic Backscattering from Biological Tissues with Focused Transducers. IEEE Trans. Biomed. Eng. 1997, 44, 549–554. [Google Scholar] [CrossRef]
  28. Kuc, R.; Schwartz, M. Estimating the Acoustic Attenuation Coefficient Slope for Liver from Reflected Ultrasound Signals. IEEE Trans. Sonics Ultrason. 1979, 26, 353–361. [Google Scholar] [CrossRef]
  29. Omura, M.; Nagaoka, R.; Yagi, K.; Yoshida, K.; Yamaguchi, T.; Hasegawa, H. Characterization of Blood Mimicking Fluid with Ultrafast Ultrasonic and Optical Image Velocimeters. Jpn. J. Appl. Phys. 2022, 61, SG1067. [Google Scholar] [CrossRef]
  30. Yuan, Y.W.; Shung, K.K. Ultrasonic Backscatter from Flowing Whole Blood. II: Dependence on Frequency and Fibrinogen Concentration. J. Acoust. Soc. Am. 1988, 84, 1195–1200. [Google Scholar] [CrossRef]
  31. Insana, M.F.; Wagner, R.F.; Brown, D.G.; Hall, T.J. Describing Small-scale Structure in Random Media Using Pulse-echo Ultrasound. J. Acoust. Soc. Am. 1990, 87, 179–192. [Google Scholar] [CrossRef]
  32. Ramnarine, K.V.; Nassiri, D.K.; Hoskins, P.R.; Lubbers, J. Validation of a New Blood-Mimicking Fluid for Use in Doppler Flow Test Objects. Ultrasound Med. Biol. 1998, 24, 451–459. [Google Scholar] [CrossRef] [PubMed]
  33. International Electrotechnical Commission. Ultrasonics Flow Measure- Ment Systems—Flow Test Object. Draft IEC 2001, 1685. [Google Scholar]
  34. Faran, J.J. Sound Scattering by Solid Cylinders and Spheres. J. Acoust. Soc. Am. 1988, 405, 405–418. [Google Scholar] [CrossRef]
  35. Mozumi, M.; Omura, M.; Nagaoka, R.; Cinthio, M.; Hasegawa, H. Impact of Spacing of Ultrasound Receiving Beams on Estimation of 2D Motion Velocity. Jpn. J. Appl. Phys. 2021, 60, SDDE07. [Google Scholar] [CrossRef]
  36. Céspedes, I.; Huang, Y.; Ophir, J.; Spratt, S. Methods for Estimation of Subsample Time Delays of Digitized Echo Signals. Ultrason. Imaging 1995, 17, 142–171. [Google Scholar] [CrossRef]
  37. Hasegawa, H.; Kanai, H. Simultaneous Imaging of Artery-Wall Strain and Blood Flow by High Frame Rate Acquisition of RF Signals. IEEE Trans. Ultrason. Ferroelectr. Freq. Control 2008, 55, 2626–2639. [Google Scholar] [CrossRef]
  38. van der Heiden, M.S.; de Kroon, M.G.M.; Bom, N.; Borst, C. Ultrasound Backscatter at 30 MHz from Human Blood: Influence of Rouleau Size Affected by Blood Modification and Shear Rate. Ultrasound Med. Biol. 1995, 21, 817–826. [Google Scholar] [CrossRef]
  39. Savéry, D.; Cloutier, G. A Point Process Approach to Assess the Frequency Dependence of Ultrasound Backscattering by Aggregating Red Blood Cells. J. Acoust. Soc. Am. 2001, 110, 3252–3262. [Google Scholar] [CrossRef]
  40. Hasegawa, H.; Nagaoka, R.; Omura, M.; Mozumi, M.; Saito, K. Investigation of Feasibility of Singular Value Decomposition Clutter Filter in Plane Wave Imaging with Packet Transmission Sequence. J. Med. Ultrason. 2021, 48, 13–20. [Google Scholar] [CrossRef]
  41. Neumann, F.J.; Katus, H.A.; Hoberg, E.; Roebruck, P.; Braun, M.; Haupt, H.M.; Tillmanns, H.; Kubler, W. Increased Plasma Viscosity and Erythrocyte Aggregation: Indicators of an Unfavourable Clinical Outcome in Patients with Unstable Angina Pectoris. Br. Heart J. 1991, 66, 425–430. [Google Scholar] [CrossRef] [Green Version]
  42. Chabanel, A.; Horellou, M.H.; Conard, J.; Samama, M.M. Red Blood Cell Aggregability in Patients with a History of Leg Vein Thrombosis: Influence of Post-Thrombotic Treatment. Br. J. Haematol. 1994, 88, 174–179. [Google Scholar] [CrossRef] [PubMed]
  43. Razavian, S.M.; Atger, V.; Giral, P.; Cambillau, M.; Del-Pino, M.; Simon, A.C.; Moatti, N.; Levenson, J. Influence of HDL Subfractions on Erythrocyte Aggregation in Hypercholesterolemic Men. Arterioscler. Thromb. 1994, 14, 361–366. [Google Scholar] [CrossRef] [PubMed] [Green Version]
  44. Fornal, M.; Lekka, M.; Pyka-Fościak, G.; Lebed, K.; Grodzicki, T.; Wizner, B.; Styczeń, J. Erythrocyte Stiffness in Diabetes Mellitus Studied with Atomic Force Microscope. Clin. Hemorheol. Microcirc. 2006, 35, 273–276. [Google Scholar] [PubMed]
  45. Kim, J.; Lee, H.; Shin, S. Advances in the Measurement of Red Blood Cell Deformability: A Brief Review. J. Cell. Biotechnol. 2015, 1, 63–79. [Google Scholar] [CrossRef] [Green Version]
  46. Tripette, J.; Nguyen, L.-C.; Allard, L.; Robillard, P.; Soulez, G.; Cloutier, G. In Vivo Venous Assessment of Red Blood Cell Aggregate Sizes in Diabetic Patients with a Quantitative Cellular Ultrasound Imaging Method: Proof of Concept. PLoS ONE 2015, 10, e0124712. [Google Scholar] [CrossRef] [Green Version]
Figure 1. Illustration of in vitro measurement. Two flow phantoms (*) were compared in this study.
Figure 1. Illustration of in vitro measurement. Two flow phantoms (*) were compared in this study.
Sensors 23 02639 g001
Figure 2. Illustration of the power curve of singular values (a,(b) porcine blood and in vivo jugular vein experiments.
Figure 2. Illustration of the power curve of singular values (a,(b) porcine blood and in vivo jugular vein experiments.
Sensors 23 02639 g002
Figure 3. B-mode images in the porcine blood experiment in case of low shear rate. (a) Clutter-less case. (b,c) Clutter-rich cases without and with filters. (1) and (2) PBS and plasma samples.
Figure 3. B-mode images in the porcine blood experiment in case of low shear rate. (a) Clutter-less case. (b,c) Clutter-rich cases without and with filters. (1) and (2) PBS and plasma samples.
Sensors 23 02639 g003
Figure 4. B-mode images in the porcine blood experiment in case of high shear rate, as well as Figure 3.
Figure 4. B-mode images in the porcine blood experiment in case of high shear rate, as well as Figure 3.
Sensors 23 02639 g004
Figure 5. Frequency dependence of the BSC at selected flow rates. (a,b) PBS and plasma samples in the clutter-less phantom. (c,d) Those in the clutter-rich phantom.
Figure 5. Frequency dependence of the BSC at selected flow rates. (a,b) PBS and plasma samples in the clutter-less phantom. (c,d) Those in the clutter-rich phantom.
Sensors 23 02639 g005
Figure 6. Shear rate dependence of the spectral slope (a) and MBF (b).
Figure 6. Shear rate dependence of the spectral slope (a) and MBF (b).
Sensors 23 02639 g006
Figure 7. In vivo B-mode images in the human jugular vein in subjects #1 (a) to #3 (c) compared between low shear (1) and high shear rates (2) within the intra-subject.
Figure 7. In vivo B-mode images in the human jugular vein in subjects #1 (a) to #3 (c) compared between low shear (1) and high shear rates (2) within the intra-subject.
Sensors 23 02639 g007
Figure 8. Temporal variation in the spectral slope (a), MBF (b), shear rate (c), and mean velocity (d) in each subject.
Figure 8. Temporal variation in the spectral slope (a), MBF (b), shear rate (c), and mean velocity (d) in each subject.
Sensors 23 02639 g008
Figure 9. B-mode image (a) and spatial distribution of the spectral slope (b) and MBF (c) with velocity vector and shear rate in subject #4.
Figure 9. B-mode image (a) and spatial distribution of the spectral slope (b) and MBF (c) with velocity vector and shear rate in subject #4.
Sensors 23 02639 g009
Figure 10. Mean temporal frequency in each cut-off power of singular value. (a,b) PBS and plasma samples.
Figure 10. Mean temporal frequency in each cut-off power of singular value. (a,b) PBS and plasma samples.
Sensors 23 02639 g010
Figure 11. Power spectra in the axial direction. (a,b) PBS and plasma samples in the different low rank singular values at a selected flow rate of 10 mL/min. The spectra were computed in the typical frame at the center of the lumen.
Figure 11. Power spectra in the axial direction. (a,b) PBS and plasma samples in the different low rank singular values at a selected flow rate of 10 mL/min. The spectra were computed in the typical frame at the center of the lumen.
Sensors 23 02639 g011
Figure 12. Typical B-mode images of PBS (a) and plasma (b) samples in the different clutter filter conditions at the flow rate of 10 mL/min.
Figure 12. Typical B-mode images of PBS (a) and plasma (b) samples in the different clutter filter conditions at the flow rate of 10 mL/min.
Sensors 23 02639 g012
Figure 13. The same as in Figure 12, except at the flow rate of 350 mL/min.
Figure 13. The same as in Figure 12, except at the flow rate of 350 mL/min.
Sensors 23 02639 g013
Figure 14. Effect of clutter filter conditions for the spectral slope in the porcine blood experiment at the flow rates of 10 mL/min (a) and 350 mL/min (b).
Figure 14. Effect of clutter filter conditions for the spectral slope in the porcine blood experiment at the flow rates of 10 mL/min (a) and 350 mL/min (b).
Sensors 23 02639 g014
Disclaimer/Publisher’s Note: The statements, opinions and data contained in all publications are solely those of the individual author(s) and contributor(s) and not of MDPI and/or the editor(s). MDPI and/or the editor(s) disclaim responsibility for any injury to people or property resulting from any ideas, methods, instructions or products referred to in the content.

Share and Cite

MDPI and ACS Style

Omura, M.; Yagi, K.; Nagaoka, R.; Yoshida, K.; Yamaguchi, T.; Hasegawa, H. Effect of Clutter Filter in High-Frame-Rate Ultrasonic Backscatter Coefficient Analysis. Sensors 2023, 23, 2639. https://doi.org/10.3390/s23052639

AMA Style

Omura M, Yagi K, Nagaoka R, Yoshida K, Yamaguchi T, Hasegawa H. Effect of Clutter Filter in High-Frame-Rate Ultrasonic Backscatter Coefficient Analysis. Sensors. 2023; 23(5):2639. https://doi.org/10.3390/s23052639

Chicago/Turabian Style

Omura, Masaaki, Kunimasa Yagi, Ryo Nagaoka, Kenji Yoshida, Tadashi Yamaguchi, and Hideyuki Hasegawa. 2023. "Effect of Clutter Filter in High-Frame-Rate Ultrasonic Backscatter Coefficient Analysis" Sensors 23, no. 5: 2639. https://doi.org/10.3390/s23052639

Note that from the first issue of 2016, this journal uses article numbers instead of page numbers. See further details here.

Article Metrics

Back to TopTop