Operating Instructions
Liquisys M CPM223/253
Transmitter for pH and ORP
Liquisys M CPM223/253 Table of contents
Table of contents
Endress+Hauser 3
Table of contents Liquisys M CPM223/253
11 Accessories . . . . . . . . . . . . . . . . . . . . . . 106
11.1 Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . 106
11.1.1 pH glass electrodes . . . . . . . . . . . . . 106
11.1.2 pH ISFET sensors . . . . . . . . . . . . . . . 106
11.1.3 ORP sensors . . . . . . . . . . . . . . . . . . 107
11.1.4 Sensor simulators . . . . . . . . . . . . . . 108
11.2 Connection accessories . . . . . . . . . . . . . . . . . 108
11.3 Installation accessories . . . . . . . . . . . . . . . . 109
11.4 Software and hardware add-ons . . . . . . . . . . 110
11.5 Buffer solutions . . . . . . . . . . . . . . . . . . . . . . 110
13 Appendix . . . . . . . . . . . . . . . . . . . . . . . . 117
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 122
4 Endress+Hauser
Liquisys M CPM223/253 Document information
1 Document information
1.1 Warnings
Structure of information Meaning
Endress+Hauser 5
Document information Liquisys M CPM223/253
Direct current
A0027423 A terminal at which DC is present or through which DC flows.
Alternating current
A0027424 A terminal to which alternating voltage (sine-wave) is applied or through
which alternating current flows.
Ground connection
A terminal which, from the user's point of view, is already grounded via a
A0027426
grounding system.
Class II equipment
Reinforced or double insulation
A0019929
Alarm relay
A0027420
Input
A0027428
Output
A0027429
DC voltage source
A0027430
Temperature sensor
A0027431
6 Endress+Hauser
Liquisys M CPM223/253 Basic safety instructions
Endress+Hauser 7
Basic safety instructions Liquisys M CPM223/253
2.5.2 IT security
We only provide a warranty if the device is installed and used as described in the
Operating Instructions. The device is equipped with security mechanisms to protect it
against any inadvertent changes to the device settings.
IT security measures in line with operators' security standards and designed to provide
additional protection for the device and device data transfer must be implemented by the
operators themselves.
8 Endress+Hauser
Liquisys M CPM223/253 Incoming acceptance and product identification
Endress+Hauser 9
Incoming acceptance and product identification Liquisys M CPM223/253
3.3.1 Nameplate
The nameplate provides you with the following information on your device:
• Manufacturer identification
• Order code
• Extended order code
• Serial number
• Ambient and process conditions
• Input and output values
• Safety information and warnings
Compare the data on the nameplate with your order.
3.4.1 mark
The product meets the requirements of the harmonized European standards. As such, it
complies with the legal specifications of the EC directives. The manufacturer confirms
successful testing of the product by affixing to it the mark.
10 Endress+Hauser
Liquisys M CPM223/253 Installation
4 Installation
Endress+Hauser 11
Installation Liquisys M CPM223/253
1 2 3
4
7
6 5
A0024634
12 Endress+Hauser
Liquisys M CPM223/253 Installation
Pg 13.5
Pg 7
Pg 16
11 (0.43)
M5
157 (6.18)
247 (9.72)
70 (2.76)
Ø 6 (0.24)
154 (6.06)
170 (6.69)
A0024637
There is a hole in the perforation for the cable entry (connection of supply voltage). It
serves as a pressure balance during air shipment. Make sure no moisture penetrates
the inside of the housing before the cable installation. The housing is completely air-
tight after cable installation.
Endress+Hauser 13
Installation Liquisys M CPM223/253
1 4
3 2
A0024640
6 (0.24)
96 (3.78)
92 (3.62)
90 (3.54)
149.5 (5.89)
139 (5.47)
A0024641
14 Endress+Hauser
Liquisys M CPM223/253 Installation
NOTICE
Effect of climatic conditions (rain, snow, direct sunlight etc.)
Impaired operation to complete transmitter failure
‣ When installing outside, always use the weather protection cover (accessory).
Ø 6 (0.24)
2
154 (6.06)
A0024638
Endress+Hauser 15
Installation Liquisys M CPM223/253
Ø max. 60 (2.36)
1
1
3
A0024635
You can also secure the field device to the Flexdip CYH112 bracket in conjunction with the
weather protection cover. These can be acquired as accessories, see the "Accessories"
section.
A0027433
16 Endress+Hauser
Liquisys M CPM223/253 Installation
+0.5 +0.02
92 (3.62 )
92+0.5 (3.62+0.02)
1
2
92 (3.62)
3
90 (3.54)
96 (3.78)
3
3
57 (2.24)
33 (1.30)
max. 45 (1.77)
~ 25 (0.98) 139 (5.47) 6 (0.24)
~ 165 (6.50)*
A0024639
8 Dimensions in mm (inch)
1 Mounting plate
2 Seal
3 Tensioning screws
* Necessary installation depth
Endress+Hauser 17
Electrical connection Liquisys M CPM223/253
5 Electrical connection
LWARNING
Device is live
Incorrect connection may result in injury or death.
‣ The electrical connection may be performed only by an electrical technician.
‣ The electrical technician must have read and understood these Operating Instructions
and must follow the instructions contained therein.
‣ Prior to commencing connection work, ensure that no voltage is present on any cable.
5.1 Wiring
LWARNING
Risk of electric shock!
‣ At the supply point, the power supply must be isolated from dangerous live cables by
double or reinforced insulation in the case of devices with a 24 V power supply.
NOTICE
The device does not have a power switch
‣ The customer must provide a protected circuit breaker in the vicinity of the device.
‣ The circuit breaker must be a switch or power switch, and you must label it as the
circuit breaker for the device.
The electrical connection of the transmitter differs depending on the device version:
• If you are using a device without Memosens functionality, please read the instructions in
the "Electrical connection without Memosens functionality" section.
• If you are using a device with Memosens functionality, please read the instructions in
the "Electrical connection with Memosens functionality" section.
18 Endress+Hauser
Liquisys M CPM223/253 Electrical connection
pH 41
42
A K
Ref. 43
SRC
47
B 48
DRN.
49
L
C S
57
PA/
D PM 58
M
11 59
12
E 51
13
52
N
53
31
mA
F 32 54
pH
55
O
33 56
mA
G 34
23 +
24 -
81 P
mA
H 10-50 V ~ =
82 L1 L+
N L-
Q
93
I 10-50 V
94
85
15 V
J 86
A0026018
Endress+Hauser 19
Electrical connection Liquisys M CPM223/253
pH / Redox (Glas)
REF (Glas/IsFET)
AC: 250V/500VA
DC: 30V/60W
RD WH BN
54 55 56 51 52 53 57 58 59 47 48 49 41 42 43
PA/PM
Mains
Hilfsenergie
DRN 12
NC
NC 31 32 33 34 NC 85 86 93 94 81 82 23 24 NC L1 N (AC)
+ - + - + - + - + - + - + - (DC)
4..20mA
Digital 2
Digital 1
SRC 11 13 S
GN YE
Redox
pH
(opt.)
Temp.
10mA
+15V
12
(IsFET)
1
11 13 131094-4G
A0026019
NOTICE
Non-observance could cause incorrect measurement
‣ Make absolutely sure to protect the cable ends and terminals from moisture.
‣ Terminals marked NC may not be connected.
‣ Unmarked terminals may not be connected.
Please label the sensor terminal block with the sticker provided.
20 Endress+Hauser
Liquisys M CPM223/253 Electrical connection
131084-4C 131081-4E
Sensor
BNC: pH/Redox + REF (Glass)
REL1 47 31 +
pH (
REF IsFET
Glass )
Redox
COM 48 32 - PA/PM BN
59 33 + 13 YE
1 Temp.
(opt.)
REL2 57 34 - 12 WH
COM 58 NC 11 GN
53 85 + SRC (IsFET)
+15V
10mA
REL3 51 86 - DRN RD
COM 52 93 + S
Digital 2
56 94 - 42
REL4 54 81 + 43
Digital 1
COM 55 82 - 41
- 24 NC AC DC
Mains
4..20mA L1 +
+ 23 NC N -
A0026020
NOTICE
Non-observance could cause incorrect measurement
‣ Make absolutely sure to protect the cable ends and terminals from moisture.
‣ Terminals marked NC may not be connected.
‣ Unmarked terminals may not be connected.
Please label the sensor terminal block with the sticker provided.
Electrode without temperature sensor CPK1 VBA / VBM box + CYK71 cable
Electrode with temperature sensor Pt 100 and TOP 68 plug-in CPK9 VBA / VBM box + CYK71 cable
head
ISFET sensor with temperature sensor Pt 100 / Pt 1000 and TOP CPK12 VBA / VBM box + CYK12 cable
68 plug-in head
pH individual electrode with separate reference electrode and CPK2 VBA / VBM box + PMK cable
separate temperature sensor
Endress+Hauser 21
Electrical connection Liquisys M CPM223/253
BN PM
PM BN PM
CPK2
pH or Redox
pH or Redox
RD Redox or Ref.
Ref. or Redox
BU Ref.
Pt 100 GN Temp.
YE Temp.
WH Temp.
PM
Coax BK Ref.
CPK9 pH + Pt 100 / Pt 1000 inner pH
GN Temp.
WH Temp.
YE Temp.
BN PM
PM Coax BK Ref
inner SRC ISFET
CPK12 pH + Pt 100 / Pt 1000
GN Temp.
WH Temp.
YE Temp.
RD DRN
BN PM
A0027564-EN
For further information on the cables and junction boxes, please refer to the
"Accessories" section.
Field device measuring cable connection
NOTICE
Moisture can cause incorrect measurements
‣ Make absolutely sure to protect the connectors, cable ends and terminals from
moisture.
22 Endress+Hauser
Liquisys M CPM223/253 Electrical connection
If you are using glass electrodes with the panel-mounted device, you have to terminate the
measuring cable with a BNC connector. A solder-free BNC connector is supplied with the
device. To do this, proceed as follows:
3
1
A0005744
A0005745
A0005746
A0005747
Endress+Hauser 23
Electrical connection Liquisys M CPM223/253
The following diagrams show the connection of various pH and ORP sensors.
ϑ
PA/ PA/
S PM S S PM
pH/
11 12 13 Redox Ref 11 12 13 DRN SRC Ref SRC Ref
BN innen / BK
Schirm /
GN WH YE BN BNC innen / BK GN WH YE RD BK
Screen
internal internal
PE
PE
)
CPK9 CPK12
A B A B
BN BN
C C
CPS11 CPS471
A0027565
17 Connecting glass electrode CPS11 with CPK9 (left) and ISFET sensor CPS471 with CPK12 (right) to
Liquisys M
A Panel-mounted device
B Field device
C Potential matching PM for symmetrical connection
PA/ PA/
S NC PM S NC PM
Ref Ref
Schirm /
Schirm /
Screen
screen
BN BNC BK BN BNC BK
CPK1 CPK1
A B A B
BN BN
CPS12 CPS12
A0027566
18 Asymmetrical (without PML) and symmetrical (with PML) connection of ORP electrodes
A Panel-mounted device
B Field device
C Potential matching (PM) in the medium for symmetrical connection
24 Endress+Hauser
Liquisys M CPM223/253 Electrical connection
The pH and ORP sensors can be connected both symmetrically and asymmetrically.
Generally, the following applies:
• No potential matching connection present: asymmetrical connection
• Potential matching connection present: symmetrical connection
The decision can also depend on the operating conditions.
Please note the following:
• Liquisys M is pre-programmed for symmetrical measurement with potential matching. If
you want asymmetrical measurement, you have to change the configuration in the A2
field.
• If the "asymmetrical" software setting is selected for a symmetrical connection, the
operating life of the reference electrode is reduced.
In the case of a symmetrical connection, the potential matching pin must be
connected and must always be immersed in the medium.
Advantages of symmetrical vs. asymmetrical:
• Symmetrical measurement:
– No leak current since the reference and the pH/ORP electrode is connected with high
impedance
– Reliable measurement under difficult process conditions (strong flowing and high-
impedance media, partially soiled diaphragm)
• Asymmetrical measurement:
Possible to use assemblies without potential matching
Endress+Hauser 25
Electrical connection Liquisys M CPM223/253
41
A S
42
BN 87
H
43
WH 88
47
48
B GN A 96 49
I
YE B 97 57
58
J
59
31
mA
C 32 51
pH
52
K
53
33
mA
D 34
54
55
L
81 56
E 10-50 V
82
23 +
24 -
93 M
mA
F 10-50 V ~ =
94 L1 L+
N L-
N
85
15 V
G 86
A0026028
The device is approved for protection class II and is generally operated without a
protective ground connection. Do not connect the sensor shielding to the transmitter.
Connecting a field device with Memosens functionality
Guide the measuring cables through the PG glands into the housing. Connect the
measuring cables in accordance with the terminal assignment.
26 Endress+Hauser
Liquisys M CPM223/253 Electrical connection
WH YE 54 55 56 51 52 53 57 58 59 47 48 49 41 42 43
B Mains
Hilfsenergie
NC 88 97 NC
NC
NC 31 32 33 34 NC 85 86 93 94 81 82 23 24 NC L1 N (AC)
+ - + - + - + - + - + - + - (DC)
4..20mA
Digital 2
Digital 1
NC 87 96 S
A
(opt.)
Temp.
10mA
+15V
pH
BN GN
87, 88
136771-0003-4A
A0026033
NOTICE
Non-observance could cause incorrect measurement
‣ Terminals marked NC may not be connected.
‣ Unmarked terminals may not be connected.
Connecting a panel-mounted device with Memosens functionality
136771-0001-4A
Sensor
31 + NC
pH
32 - 97 B YE
33 + 96 A GN
Temp.
(opt.)
34 - 88 - WH
NC 87 + BN
85 + NC
+15V
10mA
86 - NC
93 + S
Digital 2
94 - 42
81 + 43
Digital 1
82 - 41
NC AC DC
Mains
L1 +
NC N -
A0026029
NOTICE
Non-observance could cause incorrect measurement
‣ Terminals marked NC may not be connected.
‣ Unmarked terminals may not be connected.
Please label the sensor terminal block with the TU sticker provided.
Do not use the pH sticker.
Endress+Hauser 27
Electrical connection Liquisys M CPM223/253
YE Com B 97
GN Com A 96
WH - 88
1 BN + 87
A0026030
For further information on the cable, please refer to the "Accessories" section.
Proceed as follows to attach the cable coupling to the sensor plug-in head:
A0026031
28 Endress+Hauser
Liquisys M CPM223/253 Electrical connection
S
87 88 96 97
BN WH GN YE
CYK10
CPS11D
A0026032
Signal transmission between the Memosens electrode and the coupling of the CYK10 cable
is non-contact and takes place via completely potted coils. This offers the following
advantages:
• As the electrode and transmitter are galvanically isolated, the signals are not affected by
secondary potential. As a result, in contrast to sensors without Memosens functionality,
a symmetrically high-impedance connection is not required to guarantee reliable
measurement.
• The Memosens plug-in head and the Memosens coupling are completely water-proof.
• There are no open contacts. Contact corrosion, creepage currents and shunts are ruled
out.
Endress+Hauser 29
Electrical connection Liquisys M CPM223/253
41 41
42 42
43 43
A0006415
Are the devices and cables free from damage on the outside? Visual inspection
Are the power cable and signal cables connected correctly and in
accordance with the wiring diagram?
Are the PE distributor blocks grounded (if present)? Grounding is carried out at the point of
installation.
30 Endress+Hauser
Liquisys M CPM223/253 Operation options
6 Operation options
6.2.1 Display
LED displays
Endress+Hauser 31
Operation options Liquisys M CPM223/253
LC display
1 2 3 4 5
pH
05 R262
14 6
7
13
12 Setpoint 8
11 10 9
A0027223-EN
26 Transmitter LC display
32 Endress+Hauser
Liquisys M CPM223/253 Operation options
7 3
6 5
A0024629-EN
27 Operating elements
1 LC display for displaying the measured values and configuration data
2 Key to switch relays in manual mode and to display the active contact
3 LED for alarm function
4 Changeover switch for automatic/manual mode
5 LEDs for limit contactor relay (switch status)
6 Main operating keys for calibration and device configuration
7 Field for user-defined information
CAL key
When you press the CAL key, the device first prompts you for the
A0027235 calibration access code:
• Code 22 for calibration
• Code 0 or any other code for reading the last calibration data
Use the CAL key to accept the calibration data or to switch from
field to field within the calibration menu.
ENTER key
When you press the ENTER key, the device first prompts you for the
A0027236 setup mode access code:
• Code 22 for setup and configuration
• Code 0 or any other code for reading all the configuration data.
The ENTER key has several functions:
• Calls up the Setup menu from the measuring mode
• Saves (confirms) data entered in the setup mode
• Moves on within function groups
Endress+Hauser 33
Operation options Liquisys M CPM223/253
Escape function
If you press the PLUS and MINUS key simultaneously, you return to
the main menu, or are taken to the end of calibration if calibrating.
If you press the PLUS and MINUS key again, you return to the
measuring mode.
A0027237
A0027238
A0027239
34 Endress+Hauser
Liquisys M CPM223/253 Operation options
A0027242
A0027243
3. To enable the manual mode, enter the code 22 via the PLUS
and MINUS keys and press ENTER to confirm.
The lower LED (manual mode) is lit.
A0027240
A0027240
• The operating mode remains in effect even after a power failure. The relays assume
the quiescent state, however.
• The manual mode has priority over all other automatic functions.
• Hardware locking is not possible in the manual mode.
• The manual settings are kept until they are actively reset.
• Error code E102 is signaled during manual operation.
Endress+Hauser 35
Operation options Liquisys M CPM223/253
Operating modes
Measuring mode:
standard mode of
operation, displaying
current measured values
Code
A0027244-EN
If no key is pressed in the setup mode for approx. 15 min, the device automatically
returns to the measuring mode. Any active hold (hold during setup) is canceled.
Access codes
All device access codes are fixed and cannot be altered. When the device requests the
access code, it distinguishes between different codes.
• CAL key + code 22: access to Calibration and Offset menu
• ENTER key + code 22: access to the menus for the parameters which make
configuration and user-specific settings possible
• PLUS + ENTER keys simultaneously (min. 3 s): lock the keyboard
• CAL + MINUS keys simultaneously (min. 3 s): unlock the keyboard
• CAL or ENTER key + any code: access to read mode, i.e. all the settings can be read but
not modified.
The device continues measuring in the read mode. It does not shift to the "Hold" status.
The current output and the controllers remain active.
36 Endress+Hauser
Liquisys M CPM223/253 Operation options
Menu structure
The configuration and calibration functions are arranged in function groups.
• In the setup mode, select a function group with the PLUS and MINUS keys.
• In the function group itself, switch from function to function with the ENTER key.
• Within the function, select the desired option with the PLUS and MINUS keys or edit the
settings with these keys. Then confirm with the ENTER key and continue.
• Press the PLUS and MINUS keys simultaneously (Escape function) to exit programming
(return to the main menu).
• Press the PLUS and MINUS keys simultaneously again to switch to the measuring mode.
If a modified setting is not confirmed by pressing ENTER, the old setting is retained.
An overview of the menu structure is provided in the Appendix to these Operating
Instructions.
1
-
E E E
E E
A0027245
29 Menu structure
1 Functions (selection of parameters, entry of numbers)
2 Function groups, scroll backwards and forwards with the PLUS and MINUS keys
3 Switch from function to function with the ENTER key
Endress+Hauser 37
Commissioning Liquisys M CPM223/253
7 Commissioning
38 Endress+Hauser
Liquisys M CPM223/253 Commissioning
7.4 Switching on
Familiarize yourself with the operation of the transmitter before it is first switched on. In
particular please read the "Basic safety instructions" and "Operation options" sections. After
power-up, the device performs a self-test and then goes to the measuring mode.
Now calibrate the sensor in accordance with the instructions in the "Calibration" section.
During initial commissioning, the sensor must be calibrated so that the measuring
system can return precise measurement data (does not apply for digital sensors).
Then perform the first configuration in accordance with the instructions in the "Quick
setup" section. The values set by the user are kept even in the event of a power failure.
The following function groups are available in the transmitter (the groups that are only
available in the Plus Package are marked accordingly in the functional description):
Setup mode
• SETUP 1 (A)
• SETUP 2 (B)
• CURRENT INPUT (Z)
• CURRENT OUTPUT (O)
• ALARM (F)
• CHECK (P)
• RELAY (R)
• SERVICE (S)
• E+H SERVICE (E)
• INTERFACE (I)
Calibration and offset mode
• CALIBRATION (C)
• NUMERIC (N)
• OFFSET (V)
A detailed explanation of the function groups available in the transmitter can be
found in the "Device configuration" section.
Function display:
The displayed code indicates
s the function position in the
2 F3
function group.
A0025560-EN
31 Function code
Factory settings
The first time the device is switched on, the factory setting is set for all the functions. The
table below provides an overview of the most important settings.
Endress+Hauser 39
Commissioning Liquisys M CPM223/253
All other factory settings can be found in the description of the individual function groups
in the "System configuration" section (the factory setting is highlighted in bold).
40 Endress+Hauser
Liquisys M CPM223/253 Commissioning
SERVICE
A0008408-EN
5. Select your language in S1, e.g. "ENG" for English. ENG = English
Press ENTER to confirm your entry. GER = German
FRA = French
ITA = Italian
ENG S1
NEL = Dutch
ESP = Spanish
Language
A0008409-EN
7. Press the MINUS key until you get to the "Setup 1"
function group.
Oper.Mode
A0007825-EN
10. In A2, select the type of connection for your sensor. sym = symmetrical
Also refer to the "Sensor connection" section for asym =
this.
Press ENTER to confirm your entry.
asymmetrical
sym A2
Wiring
A0007826-EN
12. In A4, specify the type of sensor that you are using, Glass
e.g. "Glass" for glass electrode. ISFET
Press ENTER to confirm your entry.
Glass A4
Sensor
A0007828-EN
Endress+Hauser 41
Commissioning Liquisys M CPM223/253
14. Press the MINUS key until you get to the "Setup 2"
function group.
Press ENTER to make your settings for "Setup 2".
B
SETUP 2
A0009009-EN
¡C-Cal
A0007833-EN
42 Endress+Hauser
Liquisys M CPM223/253 Commissioning
SETUP 1
A0007824-EN
A1 Select the pH
operating mode ORP (= redox) When the operating mode
is changed, all user settings
auswählen mV
ORP (= redox) %
pH A1
are automatically reset.
Oper.Mode
A0007825-EN
Endress+Hauser 43
Commissioning Liquisys M CPM223/253
SETUP 1
A0007824-EN
Oper.Mode
A0007825-EN
44 Endress+Hauser
Liquisys M CPM223/253 Commissioning
SETUP 2
A0007830-EN
measurement mode:
Off
On
Endress+Hauser 45
Commissioning Liquisys M CPM223/253
connect the current output of an external measured variable (e.g. flowmeter) to the 4 to
20mA input of the transmitter. The following assignment applies:
1
A
1
B
Z2 Z3
0
C
0
E
D F
A0008923
B Relay contacts of PID controller Z2 Delay for controller switch-off, see field Z2
E Flow alarm 1 On
46 Endress+Hauser
Liquisys M CPM223/253 Commissioning
1 2 3 4
ENDRESS+HAUSER
LIQUISYS S
6 5
4 ... 20 mA
A0025124
33 Sample arrangement for feedforward control of the flow in the main stream to the PID controller
1.5
0.5
0 20 40 Z7 60 80 100 X
A0008942
Endress+Hauser 47
Commissioning Liquisys M CPM223/253
Functions marked in italics are not supported by the basic device version.
CUR.INPUT
A0024903-EN
A0024904-EN
A0024934-EN
A0024935-EN
Stop Dir
A0024939-EN
48 Endress+Hauser
Liquisys M CPM223/253 Commissioning
pH
12
10
8
Δ Signal
6
0 4 8 12 16 20
I [mA]
Δ I [mA]
A0025023
1 500 4
2 1000 16 41.66
3 2000 20 250
Endress+Hauser 49
Commissioning Liquisys M CPM223/253
First enter the desired current output configuration into the following blank table with a
pencil. Calculate the resulting signal distance per mA to observe the necessary minimum
slope. Then enter the values in the device.
Value pair [pH; mV; %; °C] Current Distance [pH; mV; %; °C] Current Distance
[mA] per mA [mA] per mA
10
Functions marked in italics are not supported by the basic device version.
Sel.Out
A0025027-EN
A0025028-EN
O3 (1) Enter the Lin = linear (1) The characteristic can have a
characteristic Sim = simulation positive or negative slope for
type (2)
Tab = table (3)
lin O3
the measured value output.
In the case of actuating
Sel.Type variable output (O2 = Contr),
an increasing current
A0025029-EN corresponds to an increasing
actuating variable.
50 Endress+Hauser
Liquisys M CPM223/253 Commissioning
Sel.Range
A0025030-EN
O3 (2) Simulate current Lin = linear (1) Simulation is not ended until
output Sim = simulation O3(1) or O3(3) is selected.
(2)
Tab = table (3)
sim O3
For further characteristics,
see O3 (1), O3 (3).
Sel.Type
A0025039-EN
O3 (3) Enter current Lin = linear (1) Only for Plus Package
output table Sim = simulation Values can also be
(2)
Tab = table (3)
table O3 subsequently added or
modified. The values entered
are automatically sorted by
Sel.Type increasing current value.
A0025041-EN For further characteristics,
see O3 (1), O3 (2).
Endress+Hauser 51
Commissioning Liquisys M CPM223/253
-20.0 to 150.0 °C
52 Endress+Hauser
Liquisys M CPM223/253 Commissioning
7.6.5 Alarm
You can use the "Alarm" function group to define various alarms and configure output
contacts.
Each individual error can be defined to be effective or not (at the contact or as an error
current).
Moreover, the electrode can be checked for glass breakage and leak current (P1, P2, P7).
In the event of an alarm, a cleaning function can also be activated (F8).
Functions marked in italics are not supported by the basic device version.
ALARM
A0025141-EN
momentary
contact Cont.Type
A0025142-EN
Err.Delay
A0025144-EN
F4 Select error
current
22 mA
2.4 mA IfO311,
"0-20 mA" was selected in
"2.4 mA" may not be
22mA F4
used.
Err.Curr
A0025145-EN
F5 Select the error 1 Here you can select all the errors
number 1 to 255 which should trigger an alarm. The
1 F5 errors are selected by the error
numbers. Please refer to the table
in the "System error messages"
Sel.error section for the meaning of the
A0025146-EN individual error numbers. The
factory settings remain in effect
for all errors that are not edited.
Endress+Hauser 53
Commissioning Liquisys M CPM223/253
CleanTrig
A0025149-EN
7.6.6 Check
The CHECK function group is only available for devices with a Plus Package.
In the CHECK function group, you can select different monitoring functions for the
measurement.
SCS electrode monitoring
The Sensor Check System monitors the pH and reference electrode for incorrect
measurement and complete failure.
SCS identifies the following reasons for incorrect measurement:
• Electrode glass breakage
• Fine short circuits in the pH measuring circuit, also e.g. moisture or dirt bridges at
terminal points
• Contamination or clogging of the reference electrode
• Leak current for ISFET sensor
54 Endress+Hauser
Liquisys M CPM223/253 Commissioning
22 mA
A0025195
36 SCS alarm
Do not remove the standard electrode from the process without a hold! Since SCS is
measured against the PML, no contact between the inner conductor and PML triggers
an alarm. In digital sensors, SCS is not measured against the PML.
PCS alarm (process check system)
The AC function is used to check measuring signals for deviations. If the measuring signal
change within an hour is smaller than 0.5% (of full scale value of the selected measuring
range), an alarm (E152) is triggered . The reason for such sensor behavior can be
contamination, cable open circuit or similar.
You can monitor the controller activity with the CC function. A malfunction of the
controller is detected and reported thanks to freely adjustable monitoring times (E154 to
E157).
t
A0025196
Endress+Hauser 55
Commissioning Liquisys M CPM223/253
If the connection between the pH sensor with Memosens functionality and cable is just
connected but not locked, the supply voltage can drop below the required voltage value due
to the poor coupling. This results in incorrect measurement.
With the operating voltage SCS, the supply voltage of a pH sensor with Memosens
functionality is monitored.
If this value drops below the safety limit, the measured value is ignored and the error
E127 is output.
56 Endress+Hauser
Liquisys M CPM223/253 Commissioning
You can use the "Check" function group to monitor the measured value for permissible
upper and lower limits and trigger an alarm.
Functions marked in italics are not supported by the basic device version.
CHECK
A0009045-EN
Leak Curr
A0007909-EN
LoHi!
Endress+Hauser 57
Commissioning Liquisys M CPM223/253
58 Endress+Hauser
Liquisys M CPM223/253 Commissioning
The transmitter has different ways of assigning a relay contact. Switch-on and switch-off
points and pickup and dropout delays can be assigned to the limit contactor. In addition,
you can configure an alarm threshold to output an error message and to start a cleaning
function in conjunction with this.
These functions can be used both for the primary value and for temperature measurement.
Please refer to the switch states in → 38 for a clear illustration of the relay contact
states.
• When the measured values increase (maximum function), the relay contact is closed as
of t2 after the switch-on point (t1) has been exceeded and the pickup delay has elapsed
(t2-t1).
The alarm contact switches if the alarm threshold (t3) is reached and the alarm delay
(t4-t3) has also elapsed (errors E067 to E070).
• When the measured values decrease, the alarm contact is reset when the value falls
below the alarm threshold (t5) again, as is the relay contact (t7) after the dropout delay
(t7-t6).
• If the pickup and dropout delays are set to 0 s, the switch-on and switch-off points are
also switch points of the contacts.
The same settings can also be made for a minimum function in the same way as for the
maximum function.
5
6
1
4
2
A 3
7
t1 t2 t3 t4 t5 t6 t7 t
7
3
2
B 4
6
1 5
t1 t2 t3 t4 t5 t6 t7 t
A0025215
Endress+Hauser 59
Commissioning Liquisys M CPM223/253
P(ID) controller
You can define various controller functions for the transmitter. On the basis of the PID
controller, P, PI, PD and PID controllers can be implemented. For an optimum control
system, use the controller that best suits your application.
• P controller
Used for simple linear control purposes with small system deviations. Where major
changes are to be controlled, overshooting may occur. In addition, a lasting control
deviation must be expected.
• PI controller
Is used for control systems where overshooting is to be avoided and no lasting control
deviation should occur.
• PD controller
Is used for processes that require quick changes and where peaks must be corrected.
• PID controller
Is used for processes where a P, PI or PD controller does not control sufficiently.
Configuration options of the P(ID) controller
The following configuration options are available for a PID controller:
• Change control gain Kp (P influence)
• Set integral action time Tn (I influence)
• Set derivative action time Tv (D influence)
Commissioning
If you do not yet have any experience for setting the control parameters, set the values
that yield the greatest possible stability in the control circuit. Proceed as follows to
optimize the control circuit further:
• Increase the control gain Kp until the controlled variable just starts to overshoot.
• Reduce Kp slightly again and then reduce the integral action time Tn so that the shortest
possible correction time without overshooting is achieved.
• To reduce the response time of the controller, also set the derivative action time Tv.
Control and fine optimization of the set parameters with a recorder
60 Endress+Hauser
Liquisys M CPM223/253 Commissioning
A A
1 2
B B
A A
3 4
B B
B
A0025218
A Actual value
B Time
1 Tn too small
2 Tn too large
3 Kp too large
4 Kp too small
5 Optimum setting
Endress+Hauser 61
Commissioning Liquisys M CPM223/253
A t1 A
t2 t1
1 1
0 0
t [s] t [s]
T T1 T2
A0025219
40 Signal of a pulse-length modulated controller contact (left) and of a pulse frequency-modulated controller
contact (right)
Y
100 %
– XW XW
Y
100 %
– XW XW
A0025222
41 Control characteristic of a proportional controller with direct and inverse control action
A Direct = maximum function
B Inverse = minimum function
XW Control deviation
Y Current output signal = controller actuating variable
62 Endress+Hauser
Liquisys M CPM223/253 Commissioning
1
A
0
t
1
B
0
t
t0 t1 t2 t3 t4
A0025223
42 Connection between cleaning time, pause time and hold dwell period
A Wiper and/or spray cleaning system
B Hold function
0 Inactive
1 active
t0 Normal operation
t1 cleaning start
t2-t1Cleaning time
t3-t2Clean hold dwell period (0 to 999 s)
t4-t3Pause time between two cleaning intervals (1 to 7200 min)
Chemoclean function
Just like the timer function, Chemoclean can also be used to start a cleaning cycle.
However, Chemoclean also gives you the added option of defining various cleaning and
rinsing intervals and of dosing cleaning agent.
Therefore, it is possible to clean irregularly with different repeat cycles and to separately
set the cleaning times with post rinse times.
Please note the following:
• To use the Chemoclean function the transmitter has to be equipped with a designated
relay board (see product structure or chapter "accessories").
• The timer and Chemoclean are mutually dependent. While one of the two functions is
active, the other cannot be started.
• For the Chemoclean function, the relays 3 (water) and 4 (cleaner) are used.
• If the cleaning is prematurely aborted, a post rinse time always follows.
• If the setting is "Economy", cleaning only takes place with water.
Endress+Hauser 63
Commissioning Liquisys M CPM223/253
1
A
0
1
B
0
1
C
0
t0 t1 t2 t3 t4 t5
A0025216
Neutralization controller
During neutralization control, the pH value of a medium is kept constant by dosing acid
and alkali. Two separate actuating signals are required for this task, one for acid and one
for alkali.
The neutralization controller is a controller with two relay contacts and is specially
designed for this task. The P(ID) controller is available as the controller.
The values for the control gain Kp for acid and alkali can be set separately. The integral
action time Tn and derivative action time Tv apply to both controllers (see the "P(ID)
controller" section).
The "neutral zone" is located between the set values 1 and 2. There is no acid or alkali
dosing (Y=0, ) in the "neutral zone" with a controller without an integral component (P,
PD). In the case of a controller with an integral component (PI, PID), there is constant
alkali/acid dosing (Ynew=Yold). The behavior of the I-component within the neutral zone
depends on the process type (inline/batch).
The "neutral zone" can be shifted as desired in the X direction via set point 1 and 2.
Neutralization control is only possible with relays 1 and 2.
64 Endress+Hauser
Liquisys M CPM223/253 Commissioning
Y [%]
100
1
50
0 X
2 3 4
5 6
A0025220
Functions marked in italics are not supported by the basic device version.
RELAIS
A0009058-EN
Function
A0009067-EN
Endress+Hauser 65
Commissioning Liquisys M CPM223/253
R212 Enter the switch- pH 16.00 Never set the switch-on point
on point of the pH -2.00 to 16.00 and the switch-off point to the
pH
contact 1500 mV
-1500 to 15000
16.00 R212
same value! (Only the
operating mode selected in A1
100.0 %
0.0 to 100.0 %
On value is displayed.)
A0008187-EN
On Delay
A0009070-EN
Off Delay
A0009071-EN
Clean=
Chemoclean (5)
Neutra controller
(6)
66 Endress+Hauser
Liquisys M CPM223/253 Commissioning
Function
A0009074-EN
On value
A0008214-EN
On Delay
A0009077-EN
Off Delay
A0009078-EN
Endress+Hauser 67
Commissioning Liquisys M CPM223/253
Timer (4)
Clean=
Chemoclean (5)
Neutra controller
(6)
Function
A0009081-EN
R232 Enter set point pH 16.00 The set point is the value to be
pH -2.00 to 16.00 maintained by the control
pH
1500 mV
-1500 to 15000
16.00 R232 system. Using this control
process, this value is restored
0.0 %
0.0 to 100.0 %
Setpoint when an upwards or
downwards deviation occurs.
A0008226-EN
Time Tv
A0009087-EN
R237 Select pulse Len = pulse Pulse length e.g. for solenoid
length or pulse length valve, pulse frequency e.g. for
frequency Freq = pulse
frequency
len R237
solenoid dosing pump, see the
"Actuating signal outputs"
Curr = current section.
output 2 Oper.Mode Curr = current output 2 can
A0009089-EN only be selected if O2 = Contr.
68 Endress+Hauser
Liquisys M CPM223/253 Commissioning
R239 Enter maximum 120 min-1 This field only appears if pulse
pulse frequency 60 to 180 min-1 frequency is selected in R237.
1/min
of the adjuster
120 R239
If pulse length is selected,
R239 is skipped and entries
continue with R2310.
Max.PFreq
A0009091-EN
R2311 Enter basic load 0% When you select the basic load,
0 to 40 % you enter the desired dosing
%
0 R2311 quantity.
100% basic load would
BasicLoad correspond to:
• Constantly on if R237 = len
A0009083-EN
• Fmax if R237 = freq (field
R239)
• 20 mA if R237 = curr
Function
A0009092-EN
R243 Enter pause time 360 min The pause time is the time
1 to 7200 min between two cleaning cycles
min
360 R243 (see the "Timer for cleaning
function" section).
PauseTime
A0009094-EN
Endress+Hauser 69
Commissioning Liquisys M CPM223/253
Min.Pause
A0009095-EN
R252 Select type of Int = internal The cycle for the "int" function
start pulse (time- is started when the pause time
controlled)
Ext = external
int R252
ends (R257).
No real time clock is available.
(digital input 2) External suppression is
I+ext = internal + CleanTrig required for irregular time
external A0009097-EN intervals (e.g. weekends).
I+stp = internal,
suppressed by
external
70 Endress+Hauser
Liquisys M CPM223/253 Commissioning
R257 Enter pause time 360 min The pause time is the time
1 to 7200 min between two cleaning cycles
min
360 R257 (see the "Chemoclean function"
section).
PauseTime
A0009102-EN
A0009104-EN
Function
A0009105-EN
Endress+Hauser 71
Commissioning Liquisys M CPM223/253
R266 Select pulse Len = pulse Pulse length e.g. for solenoid
length or pulse length valve, pulse frequency e.g. for
frequency Freq = pulse
frequency
len R266
solenoid dosing pump, see the
"Actuating signal outputs"
Curr = current section.
output 2 Oper.Mode Curr = current output 2 can
A0008404-EN only be selected if O2 = Contr.
R268 Enter maximum 120 min-1 This field only appears if pulse
pulse frequency 60 to 180 min-1 frequency is selected in R266.
1/min
of the adjuster
120 R268
If pulse length is selected,
R268 is skipped and entries
continue with R269.
Max.PFreq
A0008406-EN
72 Endress+Hauser
Liquisys M CPM223/253 Commissioning
7.6.8 Service
SERVICE
A0008408-EN
calibration
Setup = hold during Auto HOLD
configuration A0008413-EN
None = no hold
Man. HOLD
A0008414-EN
S4 Enter hold 10 s
dwell period 0 to 999 s
s
10 S4
Cont.Time
A0008415-EN
Endress+Hauser 73
Commissioning Liquisys M CPM223/253
PR0005
A0008418-EN
S8 Serial number is
displayed
SerNo S8
12345678
A0008420-EN
Test
A0008410-EN
74 Endress+Hauser
Liquisys M CPM223/253 Commissioning
E+H SERV
A0007857-EN
(4)
Sens = sensor (5)
E152
HW-Vers.
A0007861-EN
E153
12345678
A0007860-EN
E154
Modul-ID
A0007862-EN
Endress+Hauser 75
Commissioning Liquisys M CPM223/253
7.6.10 Interfaces
INTERFACE
A0007863-EN
7.6.11 Communication
For devices with a communication interface, please also refer to the separate Operating
Instructions BA00208C/07/EN (HART®) or BA00209C/07/DE (PROFIBUS®).
76 Endress+Hauser
Liquisys M CPM223/253 Commissioning
7.7 Calibration
Use the CAL key to access the calibration function group.
Use this function group to calibrate the sensor. The calibration can be performed in a
number of ways:
• By measuring in two calibration solutions with known pH value.
• By entering data for the slope and zero point
• In the case of ORP measurement, by entering the mV value or two different % values
Please note the following:
• During initial commissioning of amperometric sensors, calibration is absolutely essential
(except for sensors with Memosens functionality) so that the measuring system can
return precise measurement data.
• If the calibration is aborted by simultaneously pressing the PLUS and MINUS keys
(return to C19, C25 or C36), or if the calibration is faulty, the original calibration data
are used again. A calibration error is indicated by “ERR” and the sensor symbol flashes
on the display.
Repeat calibration!
• For each calibration, the device automatically switches to hold (factory setting).
• Any offset set is automatically deleted after the calibration is accepted.
• If the slope or zero point are outside the ranges given in C16 and C17, error 32 becomes
active in the case of the slope or error 33 becomes active in the case of the zero point.
The electrode must then be checked and replaced if necessary.
• If precalibrated digital sensors (Memosens functionality) are connected, the calibration
data are automatically transmitted to the transmitter.
Points to note when calibrating ISFET sensors
Switch-on behavior
A closed-control loop is created when the measuring system is switched on. The measured
value adjusts to the real value during this time (approx. 5 to 8 minutes). This settling
behavior occurs every time the liquid film between the pH-sensitive semiconductor and the
reference lead is interrupted (e.g. caused by dry storage or intensive cleaning with
compressed air). The settling time depends on the length of the interruption.
Sensitivity to light
Like all semiconductor components, the ISFET chip is sensitive to light (measured value
variations). However, this only affects the measured value if the sensor is directly exposed
to sunlight. For this reason, avoid direct sunlight when calibrating. Normal ambient light
does not have any effect on the measurement.
Endress+Hauser 77
Commissioning Liquisys M CPM223/253
C12 Enter pH value of the Buffer value of You can edit the
first buffer solution the last displayed value. The
pH
calibration
pH 0.00 to 14.00
7.00 C12 value is specified by
the corresponding
buffer solution.
Buffer 1
A0007838-EN
Insert the electrode into the buffer indicated. In the In the case of
case of ATC operation, the temperature sensor must symmetrical
also be immersed in the buffer solution. Press CAL to measuring operation,
display the current measured value. Then start the the potential
calibration with CAL. matching pin must
also be immersed in
the buffer.
A0025778
Manual continuation: if the value becomes stable, you can accept the calibration for buffer solution 1 with the
CAL key.
Automatic continuation: takes place if the value is stable (difference between measured values ≤ 0.05 and
constant value over 10 s). If the value does not stabilize within 5 min, error 44 is set and the calibration is
aborted.
Buffer 2
A0007841-EN
25.00 to 65.00
mV/pH
ISFET: 59.16
mV/pH
38.00 to 65.00
mV/pH
Press CAL.
78 Endress+Hauser
Liquisys M CPM223/253 Commissioning
C17 Zero point (zero Normal values: For ISFET, the zero
point / U_is) is Glass: pH 7.00 point is displayed in
7.00 C17
displayed pH 5.00 to 9.00 pH mV.
Antimony: pH
1.00
pH -1.00 to 3.00 Zero
ISFET: current A0007843-EN
value
-500 to +500 mV
Press CAL.
Press CAL.
CALIBRAT
A0009141-EN
Endress+Hauser 79
Commissioning Liquisys M CPM223/253
Press CAL.
CALIBRAT
A0009141-EN
For the calibration, a sample of the medium is taken in Default values: The calibration range
two containers. The contents of the first container is 0 % = -1000 mV is ±1500 mV, the
detoxified. The contents of the second container 100 % = +1000 mV minimum difference
remains unchanged. A relative value of 80 % is set with should be 60 mV.
the "toxic" sample. A relative value of 20 % is set with
the "non-toxic" sample.
80 Endress+Hauser
Liquisys M CPM223/253 Commissioning
Press CAL.
Endress+Hauser 81
Commissioning Liquisys M CPM223/253
N
NUMERICAL
CALIBRATION
N
NUMERIC
A0007866-EN
Status
A0007870-EN
Press CAL.
Store
A0007871-EN
7.7.2 Offset
The settings in the OFFSET function group can be used to adjust the measurement to a
reference measurement. This requires a linear shift of all the measured values, i.e. the
adjustment is determined for one measured value, and all others are calculated using the
same adjustment.
82 Endress+Hauser
Liquisys M CPM223/253 Commissioning
Y
A
A0025780
45 Offset
X Time
Y Measured value
A Adjusted value
B Current measured value
A0008423-EN
V3 Calibration o.k.
status is E xxx
displayed
o.k. V3
Status
A0008425-EN
Press CAL.
Endress+Hauser 83
Diagnostics and troubleshooting Liquisys M CPM223/253
84 Endress+Hauser
Liquisys M CPM223/253 Diagnostics and troubleshooting
E033 pH value zero point too and check the device and No No X 80
low or too high measuring cable with a
simulator. X 80
Endress+Hauser 85
Diagnostics and troubleshooting Liquisys M CPM223/253
X 80
86 Endress+Hauser
Liquisys M CPM223/253 Diagnostics and troubleshooting
X 80
X 0C
Endress+Hauser 87
Diagnostics and troubleshooting Liquisys M CPM223/253
Device cannot be Operation locked Press CAL and MINUS keys See the "Key functions"
operated, display simultaneously section
value 9999
Measuring chain Contaminated reference Test with new sensor pH/ORP sensor
zero-point cannot system
be adjusted
Membrane clogged Clean or grind membrane HCl 3 %, use file (only file in
one direction)
Asymmetric sensor Clean junction or test with HCl 3 %, use file (only file in
voltage too high another sensor one direction)
No or slow change Sensor fouled Clean sensor See the "Cleaning pH/ORP
of readings electrodes" section
No internal buffer Check KCl supply (0.8 bar above KCl (CPY 4-x)
medium pressure).
Measuring chain Connection not at high Check cable, plug connector and pH simulator, insulation,
slope not impedance (moisture, junction boxes see the "Checking the
adjustable/slope dirt) connecting lines and boxes"
too small section
88 Endress+Hauser
Liquisys M CPM223/253 Diagnostics and troubleshooting
Device in impermissible Switch off device and switch it EMC problem: if this
operating condition (does on again. persists, check the
not respond when key grounding, shields and line
pressed) routing or have checked by
Endress+Hauser Service.
Conductivity of medium Select pH sensor with liquid KCl. e.g. Ceraliquid CPS41
too low
Sensor fouling or buildup Clean the sensor (see the For heavily contaminated
on sensor "Cleaning pH/ORP sensors" media:
section) Use spray cleaning
Incorrect sensor type Set type of temperature sensor Glass electrode: Pt 100
at the device (field B1) ISFET: Pt 1000
Interference on signal Check cable routing, possibly Route signal output and
output cable route cable separately measuring input lines
separately
Pickup delay setting too Switch off or shorten pickup See fields R2xx
long delay time
Endress+Hauser 89
Diagnostics and troubleshooting Liquisys M CPM223/253
Controller/limit Controller in operating Select "Auto" or "Manual off" Keyboard, REL and AUTO
contact works mode "Manual on" mode keys
continuously
Dropout delay setting too Shorten dropout delay time See fields R2xx
long
No current output Cable disconnected or Disconnect cable and measure mA meter 0–20 mA
signal short-circuited directly at device
Fixed current Current simulation active Switch off simulation. See field O2
output signal
Impermissible operating Disconnect line voltage for Possibly EMC problem: if
state of processor system approx. 10 seconds problem persists, check
grounding and wire routing.
Incorrect current Incorrect current Check current assignment: 0–20 Field O211
output signal assignment mA or 4–20 mA?
Total load in the current Disconnect output and measure mA meter for 0–20 mA DC
loop too high (> 500 Ω) directly at device
No temperature Device does not have a Check version using nameplate, LSCH-x2 module, see the
output signal second current output if necessary replace the LSCH-x1 "Spare parts" section
module
Chemoclean No relay module (LSR1- Install LSR1-4 module. LSR1-4 module, see the
function not x) installed or only Chemoclean is enabled using the "Spare parts" section
available LSR1-2 available release code supplied by the
Additional function not manufacturer in the
enabled Chemoclean retrofit kit.
To check the version, see the
nameplate
Plus Package Plus Package not enabled • For Plus Package retrofit: For a detailed description,
functions not (enable by entering a code is supplied by E+H → see the "Replacement of
available code which depends on enter this code. central module" section.
the serial number and • After replacing a defective
which is supplied by E+H LSCH/LSCP module: first
when a Plus Package is enter device serial number
ordered) manually (see nameplate),
then enter the existing code
number.
No HART or Several devices at the Check addresses and re-enter if No communication possible
PROFIBUS same address necessary if several devices set to the
communication same address
No HART No HART central module Verify using nameplate: HART = Upgrade to LSCH-H1 / -H2
communication -xxx5xx and -xxx6xx
90 Endress+Hauser
Liquisys M CPM223/253 Diagnostics and troubleshooting
Interference on line
No PROFIBUS No PA/DP central Verify using the nameplate: PA Upgrade to LSCP module,
communication module = -xxx3xx /DP = xxx4xx see the "Spare parts" section
Endress+Hauser 91
Diagnostics and troubleshooting Liquisys M CPM223/253
Information on the exact spare part designations and on how to install these parts can be
found in the "Spare parts" section.
Device cannot be Operation locked Press CAL and MINUS keys See the "Key functions"
operated, display simultaneously. section
value 9999
Display dark, no No line voltage Check whether line voltage is Electrical technician / e.g.
light-emitting present multimeter
diode active
Supply voltage Compare actual line voltage and User (data for energy supply
wrong/too low nameplate data company or multimeter)
Device fuse defective Compare line voltage and the Electrical technician/suitable
nameplate data and replace fuse; see exploded drawing
fuse in the the "Spare parts"
section
Power unit defective Replace power unit, note Onsite diagnosis by Endress
version +Hauser Service, test module
necessary
Central module defective Replace central module, note Onsite diagnosis by Endress
version +Hauser Service, test module
necessary
Field device: ribbon cable Check ribbon cable, replace if See the "Spare parts" section
loose or defective necessary
Display dark, light- Central module defective Replace central module, note Onsite diagnosis by Endress
emitting diode (module: LSCH/LSCP) version +Hauser Service, test module
active necessary
Values appear on Device or module in Panel-mounted device: reinstall Perform with the aid of the
display but: device not correctly insert. installation drawings in the
• Display does not mounted Field device: remount display the "Spare parts" section
change and / or module
• Device cannot
be operated Impermissible operating Disconnect line voltage for Possibly Possible EMC
system condition approx. 10 seconds problem: if this persists,
check the installation or have
checked by Endress+Hauser
Service
Device gets hot Voltage wrong/too high Compare line voltage and User, electrical technician
nameplate data
pH/mV measured Transmitter module Measuring input test: If the test is negative: replace
value and/or defective (module: • Connect pH, ref and PM module (note version).
temperature MKIC), please first carry directly at the device with Perform with the aid of the
measured value out tests and take wire jumpers = reading pH 7 exploded drawings in the
incorrect measures as per the • 100 Ω resistance at terminals "Spare parts" section.
"Process-specific errors" 11/ 12 + 13 = display 0 °C
section
Current output, Incorrect adjustment Check with integrated current If simulation value incorrect:
incorrect current simulation, connect mA meter adjustment in factory or new
value Load too large directly to current output. LSCH module required.
Shunt / short to ground If simulation value correct:
in current loop check current loop for load
and shunts.
Incorrect mode of Check whether 0–20 mA or 4–
operation 20 mA is selected.
92 Endress+Hauser
Liquisys M CPM223/253 Maintenance
No current output Current output stage Check with integrated current If test negative:
signal defective (only for LSCH simulation, connect mA meter Replace central module (note
module; LSCP has no directly to current output version)
current output)
Additional relays Field device: ribbon cable Check ribbon cable seating, See the "Spare parts" section
not working loose or defective replace cable if required.
Only 2 additional LSR1-2 relay module Upgrade to LSR1-4 with 4 User or Endress+Hauser
relays can be with 2 relays is installed relays. Service
triggered
Incorrect device serial Check whether serial number The serial number of the
number saved in LSCH/ on the nameplate matches SNR device is definitive for the
LSCP module in LSCH/ LSCP (field S 8). Plus Package.
Additional Replacement modules In the case of LSCH/LSCP with For a detailed description,
functions (Plus LSCH or LSCP have the SNR 0000, a device serial see the "Replacement of
Package and/or device serial number number can be entered once in central module" section.
Chemoclean) 0000 when they leave fields E115 to E117.
missing after the factory. The Plus Then enter the release codes for
LSCH/LSCP Package or Chemoclean the Plus Package and/or
module are not enabled on Chemoclean if necessary.
replacement leaving the factory.
No HART or Incorrect central module HART: LSCH-H1 or H2 module, Replace central module;
PROFIBUS PA/DP PROFIBUS-PA: LSCP-PA User or Endress+Hauser
interface function module, Service.
PROFIBUS-DP: LSCP-DP
module,
see field E112.
9 Maintenance
LWARNING
Process pressure and temperature, contamination, electrical voltage
Risk of serious or fatal injury
‣ If the sensor has to be removed during maintenance work, avoid hazards posed by
pressure, temperature and contamination.
‣ Make sure the device is de-energized before you open it.
‣ Power can be supplied to switching contacts from separate circuits. De-energize these
circuits before working on the terminals.
Take all the necessary precautions in time to ensure the operational safety and reliability
of the entire measuring point.
The maintenance of the measuring point comprises:
• Calibration
• Cleaning the controller, assembly and sensor
• Checking the cables and connections
When performing any work on the device, bear in mind any potential impact this may
have on the process control system or on the process itself.
Endress+Hauser 93
Maintenance Liquisys M CPM223/253
NOTICE
Electrostatic discharge (ESD)
Risk of damaging the electronic components
‣ Take personal protective measures to avoid ESD, such as discharging beforehand at PE
or permanent grounding with a wrist strap.
‣ For your own safety, only use genuine spare parts. With genuine parts, the function,
accuracy and reliability are also ensured after maintenance work.
NOTICE
Prohibited cleaning agents
Damage to the housing surface or housing seal
‣ Never use concentrated mineral acids or alkaline solutions for cleaning.
‣ Never use organic cleaners such as benzyl alcohol, methanol, methylene chloride,
xylene or concentrated glycerol cleaner.
‣ Never use high-pressure steam for cleaning purposes.
LCAUTION
Risk of injury from cleaning agents
‣ When using the following cleaning agents, make sure to protect your hands, eyes and
clothing.
94 Endress+Hauser
Liquisys M CPM223/253 Maintenance
A0027183
Endress+Hauser 95
Maintenance Liquisys M CPM223/253
9.1.5 Assembly
Refer to the assembly operating manual for information on servicing and troubleshooting
the assembly. The assembly operating manual describes the procedure for mounting and
disassembling the assembly, replacing the sensors and seals, and contains information on
the material resistance properties, as well as on spare parts and accessories.
NOTICE
Incorrect measurements caused by moisture in the measuring cable
‣ If there is moisture in the measuring cable, the cable must be replaced!
A shunt in the cable of > 20 MΩ cannot be measured with normal multimeters but is
damaging for the pH measurement. Connect a pH simulator instead of the sensor. The
value that is displayed on the transmitter must match the value set at the simulator. The
value can differ at the second decimal place at the very maximum.
96 Endress+Hauser
Liquisys M CPM223/253 Maintenance
If you do not have a pH simulator, you can test the cable with a commercially available
megohmmeter. Please note the following when performing the test:
• Make sure to disconnect the pH measuring cable from the sensor and device!
• If you are using a junction box, you should check the infeed and outfeed measuring cable
separately.
• Check the cable with 1000 V DC (at least with 500 V DC) testing voltage.
• If the cable is intact, the insulation resistance is > 100 GΩ.
• If the cable is defective (moist), there is flashover.
The cable must be replaced.
You can clean (desalinate) the sensor head and junction box with de-ionized water
and dry them with a hot air dryer.
Endress+Hauser 97
Repair Liquisys M CPM223/253
10 Repair
Assembly is the reverse of the disassembly sequence. Tighten the special screw finger-
tight without using a tool.
98 Endress+Hauser
Liquisys M CPM223/253 Repair
422a
422b
340
400
435
235
230
10/20
40
320
330
310
430
290/300
Pos. 10
Netzteil 100 – 230 VAC
10
Pos.
3030 / Steckbrücke
A
Sicherung
230 V
B 115 V
100 V
A0027295
Endress+Hauser 99
Repair Liquisys M CPM223/253
The exploded drawing contains the components and spare parts of the panel-mounted
device. You can take the spare parts and the corresponding order number from the
following section using the item numbers.
230 pH/mV transmitter MKP3 pH/mV + temperature input, glass electrode 51518244
ab Software-Version 2.55 HART, bzw. 2.33
PROFIBUS
290 Kit for CxM2x3 relay LSR2-DP Relay module + 2 relays 71134732
module Current inp. + DP terminals
PROFIBUS DP Hardware version 2.10 and higher
310, 320, Housing mechanical Rear plate, side panel, end frame, special 51501076
340, 400 parts screw
330, 400 Housing module Housing with front membrane, sensory 51501075
tappets, seal, special screw, tensioning dogs,
connection plates and nameplates
340 End frame Rear frame for PROFIBUS DP, with D- 51502513
submin plug connector
422a, Terminal strip set Complete standard terminal strip set 51501077
422b + HART
100 Endress+Hauser
Liquisys M CPM223/253 Repair
To assemble, carefully push the modules into the guide rails of the electronics box and let
them engage in the side box noses.
It is not possible to mount the modules incorrectly. Modules inserted in the electronics
box incorrectly are not operable since the ribbon cables cannot be connected.
Make sure the cover seals are intact to guarantee IP 65 ingress protection.
Endress+Hauser 101
Repair Liquisys M CPM223/253
A
B
Pos. 10
Netzteil 100 – 230 V AC
10
3030 / Steckbrücke
Pos.
C
Sicherung
410
230 V
370
D 115 V
460b 100 V
40
450b 330
450a
310/320
400
460a
230
430
420
10/20
999
290/300
490
340
450c 480
470
A0027296
102 Endress+Hauser
Liquisys M CPM223/253 Repair
The exploded drawing contains the components and spare parts of the field device. You
can take the spare parts and the corresponding order number from the following section
using the item numbers.
290 Kit for CxM2x3 relay LSR2-DP Relay module + 2 relays 71134732
module Current inp. + DP terminals
PROFIBUS DP Hardware version 2.10 and higher
340, 330, Inner housing fittings Docking module, empty electronics box, 51501073
450 small parts
Endress+Hauser 103
Repair Liquisys M CPM223/253
You can only enter the serial number for new modules with the serial number 0000.
This can only be done once! For this reason, make sure the number entered is correct
before you press ENTER to confirm!
If an incorrect code is entered, the additional functions are not enabled. An incorrect
serial number can only be corrected at the factory!
1. Press ENTER to confirm the serial number or cancel the entry to enter the number
again.
2. If available, enter the release codes for the Plus Package and/or Chemoclean in the
"Service" menu.
3. Check the Plus Package release (e.g. by opening the function group CHECK / Code P)
or the Chemoclean function.
4. Make the customized device settings again.
104 Endress+Hauser
Liquisys M CPM223/253 Repair
10.5 Return
The product must be returned if repairs or a factory calibration are required, or if the
wrong product was ordered or delivered. As an ISO-certified company and also due to legal
regulations, Endress+Hauser is obliged to follow certain procedures when handling any
returned products that have been in contact with medium.
To ensure swift, safe and professional device returns, please read the return procedures
and conditions at www.endress.com/support/return-material.
10.6 Disposal
The device contains electronic components and must therefore be disposed of in
accordance with regulations on the disposal of electronic waste.
Observe the local regulations.
Endress+Hauser 105
Accessories Liquisys M CPM223/253
11 Accessories
11.1 Sensors
106 Endress+Hauser
Liquisys M CPM223/253 Accessories
Endress+Hauser 107
Accessories Liquisys M CPM223/253
108 Endress+Hauser
Liquisys M CPM223/253 Accessories
320 (12.6)
A0024627
49 Dimensions in mm (inch)
Flexdip CYH112
• Modular holder system for sensors and assemblies in open basins, channels and tanks
• For Flexdip CYA112 water and wastewater assemblies
• Can be affixed anywhere: on the ground, on the capstone, on the wall or directly onto
railings.
• Stainless steel version
• Product Configurator on the product page: www.endress.com/cyh112
Technical Information TI00430C
M6
70 (2.76)
90 (3.54)
60 (2.36)
Ø 6 (0.24)
70 (2.76)
90 (3.54)
A0024660
50 Dimensions in mm (inch)
Endress+Hauser 109
Accessories Liquisys M CPM223/253
110 Endress+Hauser
Liquisys M CPM223/253 Technical data
12 Technical data
12.1 Input
Measuring range pH 0 to 14
ORP –1500 to +1500 mV / 0 to 100 %
Temperature
Pt 100 -50 to +150 °C (-60 to +300 °F)
Pt 1000 (CPM2x3-IS) -50 to +150 °C (-60 to +300 °F)
NTC 30K (CPM2x3-IS) -20 to +100 °C (0 to +212 °F)
Input impedance > 1012 Ω (under rated operating conditions) for standard sensors
12.2 Output
Signal encoding Frequency Shift Keying (FSK) + 0.5 mA via current output signal
PROFIBUS PA
PROFIBUS DP
Data transmission rate 9.6 kBd, 19.2 kBd, 93.75 kBd, 187.5 kBd, 500 kBd, 1.5 MBd
Endress+Hauser 111
Technical data Liquisys M CPM223/253
112 Endress+Hauser
Liquisys M CPM223/253 Technical data
PROFIBUS PA
Manufacturer ID 11h
Device type 1516h
Device revision 0001h
Profile version 2.0
GSD files www.endress.com/profibus
GSD version
Output values Primary value, temperature
Input variables PCS display value
Supported features Device lock: The device can be locked
using the hardware or software.
PROFIBUS DP
Manufacturer ID 11h
Device type 1520h
Profile version 2.0
GSD files www.endress.com/profibus
GSD version
Output values Primary value, temperature
Input variables PCS display value
Supported features Device lock: The device can be locked
using the hardware or software.
Endress+Hauser 113
Technical data Liquisys M CPM223/253
PROFIBUS PA
Supply voltage 9 V to 32 V, max. 35 V
Sensitivity to reverse polarity No
FISCO/FNICO compliant according to IEC No
60079-27
PROFIBUS DP
Supply voltage 9 V to 32 V, max. 35 V
Sensitivity to reverse polarity Not applicable
FISCO/FNICO compliant according to IEC No
60079-27
114 Endress+Hauser
Liquisys M CPM223/253 Technical data
Slope adjustment Glass electrode 38.00 to 65.00 mV/pH (nominal 59.16 mV/ pH)
Antimony electrode 25.00 to 65.00 mV/pH (nominal 59.16 mV/ pH)
ISFET sensor 38.00 to 65.00 mV/pH (nominal 59.16 mV/ pH)
Offset pH ±2 pH units
ORP ±120 mV/±50 %
Temperature ±5 °C for adjusting the temperature display
12.5 Environment
Endress+Hauser 115
Technical data Liquisys M CPM223/253
CSA Device versions with CSA General Purpose approval are certified for indoor use.
116 Endress+Hauser
Liquisys M CPM223/253 Appendix
13 Appendix
Endress+Hauser 117
Appendix Liquisys M CPM223/253
Function group Entry of absolute value Current offset is Calibration status Store offset results
OFFSET displayed is displayed
current measured value yes; no; new
-2.00...16 pH 0.00 pH, -2.00...2.00 pH o.k.
-1500...1500 mV 0 mV, -120...120 mV E--
V 0.0...100.0 % 0.0 %, -50.0...50.0 % V2
V1 V3 V4
Function group Enter reference Enter slope Enter zero point Calibration status Store calibration
temperature Glass 59.18 mV/pH Glass 7.00 pH is displayed results
NUMERIC 38.00… 65.00 mV/pH 5.00… 9.00 pH
CALIBRATION 25 °C Antimon 59.18 mV/ pH Antimon 1.00 pH o.k. yes; no; new
-20.0...150.0 °C 25.00 … 65.00 pH –1.00 … 3.00 pH E---
ISFET 59.18 mV/pH ISFET 0 mV
N N1 38.00 ... 65.00 mV/pH N2 -500 ... +500 V N3 N4 N5
Calibration of 80% Calibration Calibration of 20% Calibration Calibration status Store calibration
Function group
value (toxic sample) value (non-toxic sample) is displayed results
CALIBRATION Acceptance when stable at Acceptance when stable at
-1500...1500 mV ≤ ± 5 mV for more than 5 s -1500...1500 mV ≤ ± 5 mV for more than 5 s o.k. yes; no; new
E--
Redox % C31 C32 C33 C34 C35 C36
calibration
Enter value of Calibration Zero point is displayed Calibration status Store calibration
Redox mV redox buffer is displayed results
calibration Acceptance when stable -100...100 mV
current measured value ≤ ± 1 mV for more than 5 s o.k. yes; no; new
-1500 mV ... 1500 mV E--
C21 C22 C23 C24 C25
pH
calibration Enter calibration Enter pH value of Calibration Enter pH value of the Calibration Display of slope
temperature first buffer solution second buffer solution Glass 59.16 mV/pH
(displayed calibration (if B3 = MTC) 38.00...65.00 mV/pH
Acceptance when stable at Acceptance when stable at Antimon 59.16 mV/pH
type options depend C Buffer value of last ≤ ± 0,05 pH for more Buffer value of last ≤ ± 0,05 pH for more
calibration 25.00...65.00 mV/pH
on selection in A1) 25.0°C calibration; than 10 s than 10 s ISFET 59.16 mV/pH
-20.0...150.0°C 0.00...14.00 pH 0.00 pH...14.00 pH 38.00 … 65.00 mV/pH C16
C11 C12 C13 C14 C15
any code
Read mode:
Code 22
Edit mode:
Temperature display Temperature display Measured value display Measured value display Measured value display
MEAS. VALUE DISPLAY in °F suppressed in mV
Current input in % Current input in mA
with
Function group Select operating mode Select connection type Enter measured Select sensor Select temperature
SETUP 1 value damping sensor
pH; ORP (mV); sym = symmetrical Glass (E0 = 7.0) Pt 100
ORP (%) asym = asymmetrical 1 (no damping) Antim = Antimon Pt 1k
A A1 A2 1-60 A3 ISFET A4 NTC 30K A5
Function group Select temperature Enter MTC temperature Select temperature Enter correct Display of
compensation (if B1=MTC and A1=pH) compensation process temperature
SETUP 2 (for the process) (for the calibration)
temperature difference
(if B1=ATC) (Offset)
pH: Redox: 25.0°C ATC; MTC 25.0°C 0.0°C
ATC; on -50.0 ... +150.0°C -50.0°C ... +150.0°C -5.0...5.0°C
B MTC off B1 B2 B3 B4 B5
Function group Controller switch-off Delay of controller Delay of controller Switch-off limit value Switch-off direction Feedforward control to
by current input switch-off by current input switch-on by current input for current input for current input PID controller
CURRENT INPUT
Off; input 0s 0 ... 2000 s 0s 0 ... 2000 s 50% 0 ... 100% Low; high Off; lin = linear
Z Z1 Z2 Z3 Z5 Z6
Z4
Select characteristic Select table options Enter number of value Select table
pairs in table value pair
read;
edit 1 1
Tab = table O3 (3) 1...10 1... number of
table value pairs;
O331 O332 asign O333
current value
sim = simulation
O3 (2) 0... 22,00 mA
O321
Select current output Select measured variable Select current range Enter 0/4 mA value Enter 20 mA value
Function group
for 2nd current output
CURRENT OUTPUT 4-20 mA; 0-20 mA +2.00 pH; -2.00...16.00 pH 12.00 pH; -2.00...16.00 pH
Out 1; Out 2 °C; pH, mV; Contr lin = linear -1500 mV; -1500...1500 mV 1500 mV; -1500...1500 mV
O3 (1) 0.0 %; 0.0...100.0 % 100.0 %; 0.0...100 %
0.0°C; -50...150.0°C 100.0°C; -50...150.0°C
O O1 O2 O311 O312 O313
Function group Select contact type Select alarm delay unit Alarm delay Set error current Select error number Set alarm contact
ALARM to be effective
Latch = latching contact; s; min 0 s (min) 22 mA; 1
Momen = momentary cont. 0 s... 2000 s (min) 2,4 mA 1... 255 yes; no
F F1 F2 (depends on F2) F3 F4 F5 F6
A0027498-EN
118 Endress+Hauser
Liquisys M CPM223/253 Appendix
Feedforward control = 1
at
50% 0 … 100%
Z7
A0027499-EN
Endress+Hauser 119
Appendix Liquisys M CPM223/253
Function group SCS alarm SCS alarm SCS alarm threshold Leakage current is Select process Alarm delay
CHECK Measuring sensor Reference sensor displayed monitoring
(if A2=sym) (ISFET sensors only)
50 kΩ Off; Low; High; 0 min (s)
P off; on P1 off; on P2 1.5...50 kΩ P3 0.0 … 9.9 mA P4 LoHi; Lo!; Hi!; LoHi! P5 0 … 2000 min (s) P6
Limit contactor Function R2 (6) Set point 1 (or 2) Enter control Enter integral action time
configuration switch off or on gain Kp1 (or Kp2) Tn (0.0 = no I component)
6.00 pH
Neutr = R2 (6) off; on -2.00...16.00 pH 1.00 0.0 min
neutralisation controller 0.01...20.00 0.0...999.9 min
(with Rel1 and Rel2
R261 R262 R263 R264
and A1 = pH only)
Function R2 (5) Select start pulse Enter pre-rinse time Enter cleaning time
switch off or on int = internal;
ext = external; 30 s 10 s
off; on i+ext = internal +external; 0...999 s 0...999 s
Clean = i+stp = internal,
Chemoclean R2 (5)
suppressed by ext R254
(with Rel3 only) R251 R252 R253
Function R2 (4) Set rinse time Set pause time Set minimum pause
switch off or on time
30 s 360 min
Timer off; on 0...999 s 1...7200 min 120 min
1...3600 min
Function R2 (3) Enter set point Enter control Enter integral action time
switch off or on gain Kp Tn (0.0 = no I component)
pH 16.00; -2.00...16.00 pH
PID controller off; on; basic; PID+B 1500 mV; -1500...1500 mV 1.00 0.0 min
100.0 %; 0...100.0 % 0.01...20.00 0.0...999.9 min
R231 R232 R233 R234
R2 (3)
Select contact to be LC PV = Function R2 (1) Select contact switch-on Select contact switch-off Enter pick-up delay
Function group
configured Limit contactor switch off or on point point
RELAY pH/Redox
Rel1; Rel2; Rel3; Rel4; off; on 16.00 pH; -2.00...16.00 pH pH 16.00; pH -2.00...16.00 0s
1500 mV; -1500...1500 mV 1500 mV; -1500...1500 mV 0...2000 s
100.0 %; 0...100.0 % 100.0 %; 0...100.0 %
R R1 R2 (1) R211 R212 R213 R214
Function group Select language Hold configuration Manual hold Enter hold Enter release code Enter release code for
SERVICE s+c=during setup and dwell period for SW upgrade SW upgrade
ENG; GER calibration off; on (Plus package) ChemoClean
ITA; FRA CAL=during calibration 10 s
ESP; NEL Setup=during setup 0...999 s 0000 0000
S S1 none=no hold S2 S3 S4 0000...9999 S5 0000...9999
S6
Select module Software version Hardware version Serial number Module name
is displayed is displayed
SW version HW version
Sens = sensor E1(5)
E151 E152 E153 E154
A0027500-EN
120 Endress+Hauser
Liquisys M CPM223/253 Appendix
Set lower alarm Set upper alarm Select process Set max. perm. period Set max. permi. period Set monitoring value
threshold threshold monitoring of lower limit exceeded of upper limit exceeded
Off; AC; CC; AC CC
pH -2.00 pH -2 ... 16 pH 16.00 pH -2 ... 16 AC!; CC!; ACCC! 60 min 0 ... 2000 min 120 min 0 ... 2000 min pH 1.00 pH -2... 16
P7 P8 P9 P10 P11 P12
Enter derivative action time Select Enter pulse interval Enter maximum Enter minimum ON Enter process type
Tv (0.0 = no D componentl) pulse frequency time ton
len = pulse length
0.0 min freq = pulse frequency 10.0 s 120 1/min 0.3 s Batch
0.0...999.9 min curr = current output 2 0.5...999.9 s 60...180 1/min 0.1...5.0 s Inlne
Enter post-rinse time Set number of repeat Set interval between two Enter minimum pause Enter number of cleaning
cycles cleaning cycles time cycles without cleaning
20 s (pause time) agent
0...999 s 0 120 min
0...5 360 min 1...R357 min 0
R256 1...7200 min 0...9
R255 R257 R258 R259
Enter derivative action Select control Select Enter pulse interval Enter maximum Enter minimum ON Enter basic load Enter process type
time Tv characteristic pulse frequency time ton
(0.0 = no D component) Ien = pulse length
dir = direct; freq = pulse frequency 10.0 s 120 1/min 0.3 s 0% Batch
0.0 min inv = inverted; curr = current output 2 0.5...999.9 s 60...180 1/min 0.1...5.0 s 0 … 40% Inlne
0.0...999.9 min
R235 R236 R237 R238 R239 R2310 R2311 R2312
0s 150.0 °C MAX
0...2000 s -20.0...+150.0 °C MIN
Order number Serial number is Reset instrument to Perform instrument test Reference voltage Select AC frequency
is displayed displayed default values is displayed
no; no; display
Sens = sensor data;
Facty = factory settings.
S7 S8 S9 S10 S11 S12
A0027501-EN
Endress+Hauser 121
Index Liquisys M CPM223/253
Index
A Interfaces . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
Access codes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 106 J
Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53 Junction boxes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
Alarm contact . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
Approvals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
K
Key functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Automatic mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
B L
Liquid KCl supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
Buffer solutions . . . . . . . . . . . . . . . . . . . . . . . . . . . . 110
Local operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
C
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77
M
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93
Central module replacement . . . . . . . . . . . . . . . . . . 104
Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
Certificates . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Digital sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
Maintenance of the entire measuring point . . . . . . . . 94
Cleaning
Manual mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94
Measuring system . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94
Mechanical construction . . . . . . . . . . . . . . . . . . . . . 116
Commissioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
Menu structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Digital electrodes . . . . . . . . . . . . . . . . . . . . . . . . . 38
ISFET sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
N
Communication . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
Nameplate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Connecting cables . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
Neutralization controller . . . . . . . . . . . . . . . . . . . . . . 64
Current input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Numeric calibration . . . . . . . . . . . . . . . . . . . . . . . . . . 82
Current outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
O
D Offset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82
Designated use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Operating concept . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Device configuration . . . . . . . . . . . . . . . . . . . . . . . . . 43
Operating elements . . . . . . . . . . . . . . . . . . . . . . . . . . 33
Device-specific errors . . . . . . . . . . . . . . . . . . . . . . . . . 91
Operating modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
Diagnostics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84
Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
Digital sensors
Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 111
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
Disassembling P
Field device . . . . . . . . . . . . . . . . . . . . . . . . . . . . 101 P(ID) controller . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
Panel-mounted device . . . . . . . . . . . . . . . . . . . . . 98 Performance characteristics . . . . . . . . . . . . . . . . . . . 115
Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31 Post mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 105 Post-connection check . . . . . . . . . . . . . . . . . . . . . . . . 30
Post-installation check . . . . . . . . . . . . . . . . . . . . . . . 17
E Power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 114
E+H Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
Process-specific errors . . . . . . . . . . . . . . . . . . . . . . . . 88
Electrical connection . . . . . . . . . . . . . . . . . . . . . . . . . 18
Product identification . . . . . . . . . . . . . . . . . . . . . . . . . 9
Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 115
Q
F Quick Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Feedforward control . . . . . . . . . . . . . . . . . . . . . . . . . 46
R
H Relay configuration . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Hold function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98
I Replacing the central module . . . . . . . . . . . . . . . . . . 104
Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 105
Incoming acceptance . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 111
S
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Safety instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Installation conditions . . . . . . . . . . . . . . . . . . . . . . . . 13
Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Installation instructions . . . . . . . . . . . . . . . . . . . . . . . 15
122 Endress+Hauser
Liquisys M CPM223/253 Index
Sensor connection
With Memosens . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Without Memosens . . . . . . . . . . . . . . . . . . . . . . . 21
Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
Setup 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Setup 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Spare parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98
Switching on . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
System error messages . . . . . . . . . . . . . . . . . . . . . . . 84
T
Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 111
Trouble shooting instructions . . . . . . . . . . . . . . . . . . 84
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84
U
Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
W
Wall mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Wiring diagram
With Memosens . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Without Memosens . . . . . . . . . . . . . . . . . . . . . . . 18