Location via proxy:   [ UP ]  
[Report a bug]   [Manage cookies]                
clinical investigations in critical care Variations in the Measurement of Weaning Parameters* A Survey of Respiratory Therapists Guy W. Soo Hoo, MD, MPH, FCCP; and Louis Park, MD Objectives: Respiratory therapists differ in the methods used to obtain weaning parameters. A questionnaire survey was conducted to better characterize those differences. Design: A questionnaire survey was conducted among respiratory therapists from nine hospitals in the Los Angeles area. The four-page, 32-question instrument was self-administered and anonymous. Responses were tabulated for analysis. Setting: Respondents from nine hospitals, three hospitals with residency training programs and six community hospitals without training programs in the Los Angeles area. Participants: One hundred two respiratory therapists. Results: There was no universally acknowledged group of weaning parameters, although four parameters were named by > 90%. There was wide variation in methods used to obtaining weaning parameters. Almost all (91%) obtained measurements with the patients breathing their current fraction of inspired oxygen, but there was great variability in the ventilator mode used to collect these parameters (T-tube, continuous positive airway pressure, pressure support), with an equally wide range of pressures added to each mode (0 to 10 cm H2O). There was great variation in the time (< 1 to > 15 min) before recording weaning parameters. Measurement of parameters was done either with bedside instruments or read from the ventilator display. The maximal inspiratory pressure had great variation in the duration of airway occlusion (< 1 to 20 s), with the most frequent time frame being 2 to 4 s. Differences were noted between therapists from the same hospital as well as between hospitals. Conclusions: There is great variation among respiratory therapists when obtaining weaning parameters. This calls for further standardization of the measurement of weaning parameters. (CHEST 2002; 121:1947–1955) Key words: mechanical ventilation; respiratory therapists; weaning parameters Abbreviations: CPAP ⫽ continuous positive airway pressure; Fb/Vt ⫽ frequency/tidal volume ratio; Fio2 ⫽ fraction of inspired oxygen; MIP ⫽ maximum inspiratory pressure; PS ⫽ pressure support; RR ⫽ respiratory rate; V̇e ⫽ minute ventilation; Vt ⫽ tidal volume timing of extubation or discontinuation of T hemechanical ventilation remains one of the more challenging aspects of critical care management. A *From the Pulmonary and Critical Care Section, West Los Angeles Veterans Affairs Medical Center, and Department of Medicine, UCLA School of Medicine, Los Angeles, CA. Manuscript received April 17, 2001; revision accepted January 4, 2002. Correspondence to: Guy W. Soo Hoo, MD, MPH, FCCP, West Los Angeles VAMC, Pulmonary and Critical Care (111Q), 11301 Wilshire Blvd, Los Angeles, CA 90073; e-mail: Guy.Soohoo@ med.va.gov www.chestjournal.org variety of strategies have been proposed, beginning with measurements collectively referred to as “weaning parameters,” followed by trials of spontaneous breathing, partially supported breathing, or a more gradual reduction in ventilator support.1 These weaning parameters usually consist of measurements of respiratory rate (RR), tidal volume (Vt), vital capacity, minute ventilation (V̇e), and maximum inspiratory pressure (MIP). Other parameters include the frequency/tidal volume ratio (Fb/Vt), dead space ventilation, inspiratory pressure at 0.1 s, work of breathing, compliance, or an index (eg, thoracic CHEST / 121 / 6 / JUNE, 2002 1947 compliance, RR, arterial oxygenation, MIP index) derived from the aforementioned variables.2–9 None of the parameters have perfect positive or negative predictive capability, but in unison provide insight about a patient’s status. They may serve as predictors of the ability to tolerate unassisted ventilation as well as the need for continued ventilator support. It is imperative that these measurements are reproducible and accurate. Uniformity in techniques and data acquisition would be obvious requirements. This is self-evident and has led several investigators to call for the standardization of weaning parameters.10 –12 Current clinical guidelines primarily list threshold values.13 Full descriptions of weaning parameters and techniques are not provided, but are available as references. These references usually refer to experimental and/or clinical studies with conditions and equipment that may not be universally available or applicable. Other sources may provide a reasonably detailed description of the technique,14,15 but are lacking in details that may be considered routine (fraction of inspired oxygen [Fio2], ventilator mode, duration before measurement). The extent of variation in measurement of weaning parameters is unknown. Observation at our institution suggested significant variation, but this may not be reflective of general practice. Better characterization would not only provide support for standardization, but also identify areas in which to focus educational efforts. This led to the following questionnaire survey on the weaning practices of respiratory therapists in the Los Angeles area. Materials and Methods A self-administered four-page, 32-item questionnaire was developed and piloted among respiratory therapists at our institution (see Appendix). The items specifically addressed the methods by which weaning parameters were obtained, and were based on published descriptions of the techniques incorporating input from the therapists. The questionnaires were anonymous and required about 10 min to complete. Some questions allowed multiple responses, and each response was tabulated for analysis. Nine hospitals in the Los Angeles area participated in the survey. Three of the hospitals have active residency training programs in all the major disciplines (medicine, surgery, etc.), while the other six were community hospitals without training programs. All of the hospitals have ⬎ 100 inpatient beds, with two of the teaching hospitals having ⬎ 500 inpatient beds. All of the hospitals have from 20 to 50 ICU beds, except one of the training hospitals with ⬎ 50 ICU beds. After obtaining permission for administering the questionnaire from the director of each respiratory therapy department, the questionnaires were distributed to members at the individual hospitals. When completed, the questionnaires were either retrieved by one of the investigators or mailed back to our institution. The responses were tabulated using an Excel spreadsheet (Microsoft; Redmond, WA). Data were examined both in aggregate pooling all respondents as well as focusing on respondents 1948 from the same hospital. Statistical analysis with comparison of proportions was performed using Primer for Biostatistics software (Version 4.0; McGraw-Hill; New York, NY). Results Of a total of 166 questionnaires distributed, 102 questionnaires (61%) were completed and returned for analysis. The majority of respondents (75%) report ⬎ 5 years of experience, and over half are certified respiratory therapists. Of the 102 respondents, 46 persons (45%) work in teaching institutions. Ventilator management is primarily directed by specialty internists (pulmonary/critical care) [41%], but the therapists also receive directions from others (general internists, surgeons, specialty surgeons, anesthesiologists, and physicians in training). Most of the respondents were responsible for 1 to 9 patients receiving ventilation per shift, with two respondents (2%) responsible for ⬎ 10 patients receiving ventilation. Most of the respondents (69%) noted no specific time for obtaining weaning parameters, but 32 respondents (31%) responded that weaning parameters were obtained only during daytime hours (before 3 pm). The majority of respondents (68%) noted that patients were extubated at any time, with 32% noting that extubation occurred primarily during daytime hours and never after 8 pm. About one half of the therapists (49%) responded that they did not require a specific order for weaning parameters. When analyzed based on the type of institution, 30 therapists (65%) from teaching institutions, compared to 22 therapists (39%) at a nonteaching institution, responded that weaning parameters required an order. This difference (p ⫽ 0.016, z ⫽ 2.414) likely reflects the increased availability of physicians at a teaching institution. The majority of respondents (79%) believed that weaning parameters consisted of a predesignated group of variables. Most of the therapists (95%) required patients to be arousable before performing weaning parameters, but approximately one half have obtained these measurements in unconscious patients (47%). Most of the therapists (91%) obtained weaning parameters with the patient breathing supplemental oxygen (current Fio2 setting), with a minority having the patient breathe room air. Table 1 shows the ventilator mode in which weaning parameters were obtained. Multiple responses were permissible and responses included combinations of T-tube, continuous positive airway pressure (CPAP), and pressure support (PS). Only three respondents exclusively obtained weaning parameters with the patient breathing off the ventilator and through a T-tube. The remainder utilized some combination of T-tube or CPAP (97%), with or Clinical Investigations in Critical Care Table 1—Ventilator Mode in Obtaining Weaning Parameters Ventilator Modes No. (%) T-tube T-tube T-tube, CPAP T-tube, CPAP, PS CPAP CPAP, no PS CPAP/PS CPAP level, cm H2O 0 1–5 6–10 Not specified PS PS, CPAP ⫽ 0 CPAP ⬎ 1 cm H2O, PS PS level, cm H2O 1–5 6–10 Variable Not specified 25 (25) 3 15 7 99 (97) 67 32 77 16 1 5 32 (31) 23 9 5 19 5 3 without added pressure, and with or without added PS. Twenty-three respondents (23%) obtained weaning parameters using PS, but without added CPAP pressure. Nine of the respondents used some combination of CPAP pressure and added support with PS. Figure 1 provides a breakdown by hospital on the modes used to measure weaning parameters. In two hospitals (hospital A and hospital H), weaning parameters were obtained only in the CPAP mode, but even in those two hospitals, there were differences among respondents whether measurements were performed with a CPAP of 0 cm H2O or 5 cm H2O. In the remaining hospitals, CPAP was the most common mode. In hospital F, hospital G, and hospital I, all of the respondents reported no additional pressure in the CPAP mode (0 cm H2O). In the others that reported measurement with added pressure, the pressure was variable, with the most common response of 3 to 5 cm H2O. This variability was also noted when PS was used, with the range of responses spanning 3 to 12 cm H2O. Table 2 provides further information on the frequency, time to recording, and types of weaning parameters. Most respondents (68%) reported obtaining weaning parameters one to five times per week. Two respondents reported that they measured zero weaning parameters in an average week. With respect to the time, a small number (6%) reported values for weaning parameters almost immediately (⬍ 1 min), while others recorded values after ⬎ 15 min (5%). The most frequent time frame was between 1 min and 5 min (68%). Table 2 also lists the most commonly reported weaning parameters. No measurement was named by all respondents. The most frequently reported parameters (⬎ 90%) were the MIP, Vt, RR, and V̇e. The Fb/Vt ratio was specifically reported by ⬍ 20%. The combination of RR and Vt was reported by 94 of respondents (92%), and the Fb/Vt ratio can be easily calculated from these two values. This variable (Fb/Vt) was more often specifically reported by Figure 1. Breakdown by hospital and representing the percentage of responses from each hospital of the ventilator mode used to measure weaning parameters. www.chestjournal.org CHEST / 121 / 6 / JUNE, 2002 1949 therapists from the teaching hospitals (16 of 46 respondents, 35%) as compared to community hospitals (3 of 56 respondents, 5%; z ⫽ 3.62; p ⬍ 0.01). No other differences were noted between therapists based on their hospital of practice in terms of methods by which they measured or reported weaning parameters (T-tube vs CPAP vs PS). Table 3 provides more information on RR, Vt, and V̇e. These parameters can be directly measured using a bedside spirometer, but can also be recorded from the ventilator display or bedside monitor. About 70% reported values using the ventilator display. About 30% reported making direct measurements of these parameters, except for RR, which was directly observed by almost 75% of respondents. This parameter (RR) also had the greatest number of dual responses, with approximately 40% using both methods. This dual response was noted in ⬍ 10% for Vt or V̇e. Table 4 is devoted to the MIP. In contrast to the other variables outlined in Table 3, the MIP was obtained with a manometer by almost 90% of respondents as opposed to using the ventilator software package. Only 25% responded using both the manometer or ventilator software package. There was a wide range of occlusion times before reporting the MIP, with the most frequent time between 2 s and 10 s (88%). Most respondents (68%) obtained Table 2—Weaning Parameters Parameters No. obtained/wk 0 1–5 6–10 11–15 ⱖ 16 Time to recording values, min ⬍1 1–2 3–5 6–10 11–15 ⬎ 15 Variable Weaning parameters reported MIP Vt RR V̇e Vital capacity Oxygen saturation Heart rate BP Fb/Vt Compliance Temperature Others (maximum expiratory pressure) 1950 No. (%) 2 (2) 69 (68) 20 (20) 7 (7) 4 (4) 6 (6) 45 (44) 24 (24) 6 (6) 3 (3) 5 (5) 13 (13) 100 100 95 91 89 69 50 22 19 16 8 1 three values, and most reported the highest value (75%). The occlusion time was not uniform among therapists in any of the hospitals surveyed. Figure 2 depicts the percent of therapists responding for each time frame at each of the surveyed hospitals. In hospital I, almost 90% of the therapists occlude the airway for 2 to 4 s to obtain a MIP. The most diversity can be seen in hospital B, where no single time frame was marked by ⬎ 30% of the respondents. Discussion Despite several decades of use, the constellation of bedside measures collectively referred to as “weaning parameters” have never quite met expectations. Witness the large number of variables that are considered weaning parameters and the continued search for bedside measurements that more accurately predict successful discontinuation of mechanical ventilation.1–9,15 These physiologic markers may provide insight about a patient’s ability to tolerate discontinuation of mechanical ventilation, but they do not address other factors that may prevent successful extubation (secretion control or clearance, upper-airway obstruction, comorbidities). Their predictive value may worsen, as patients may be very different from the original reference group (longer duration of ventilation, age, gender, different measurement methodology, etc.). Trials of spontaneous breathing have gained enthusiasm, but they can also be viewed another type of weaning parameter, albeit more dynamic and more pragmatic,16 but with the same limitations as noted earlier. Irrespective of the measurement or the manner in Table 3—Weaning Practices Practices No. (%) RR Direct observation Ventilator display Both direct observation and ventilator display Bedside monitor reading Either direct observation, ventilator display, or bedside monitor Vt Direct measurement with bedside spirometer Ventilator display Both direct measurement or ventilator display Calculated from ventilator readings (V̇e/frequency) V̇e Direct measurement with bedside spirometer Ventilator display Both direct or vent display 74 (73) 68 (67) 40 (39) 11 (11) 9 (9) 31 (30) 75 (74) 9 (9) 19 (19) 33 (32) 72 (71) 7 (7) Clinical Investigations in Critical Care Table 4 —Weaning Parameters Parameters MIP Direct measurement with manometer Measurement using ventilator software Both direct measurement or ventilator software Duration of airway occlusion during MIP, s ⬍1 2–4 5–10 11–15 16–20 ⬎ 21 Variable MIP measurements, No. One Two Three Four Variable Reported MIP Average of values obtained Highest value obtained Most reproducible No. (%) 89 (87) 39 (38) 25 (25) 7 (7) 57 (56) 22 (22) 5 (5) 5 (5) 0 (0) 6 (6) 3 (3) 27 (26) 69 (68) 2 (2) 1 (1) 24 (24) 77 (75) 1 (1) which the information is used, weaning parameters can only be useful if they accurately reflect the patient’s condition. Respiratory therapists play a crucial role, as they are delegated the task of obtaining and reporting weaning parameters. It follows that accuracy is related to measurement techniques, with uniform techniques likely to produce the most reliable information. Available guidelines are general and lack explicit details. This in turn allows some latitude in these measurements.13,14 The results of this survey confirm the wide diversity in measurement techniques of weaning parameters among respiratory therapists. The differences involve nearly every facet of measurement, and occur among therapists from different hospitals, as well as among therapists in the same hospital. There is no “consensus” as to which measures constitute weaning parameters. No parameter was named by all of the respondents, although four were noted in ⬎ 90% (MIP, Vt, V̇e, and RR) and are often considered the core group of weaning parameters. Although the Fb/Vt ratio has gained support as the single most predictive measurement in weaning,2 it was specifically noted by ⬍ 20%. However, the ratio is derived from RR and Vt, and therefore, if not specifically reported, easily calculated. It was also marked more often by therapists from teaching hospitals, suggesting that the Fb/Vt ratio may not be embraced as widely in community hospitals as teaching institutions. Among the many variations in measurement, two areas are especially noteworthy. First, there is conwww.chestjournal.org siderable variation in the ventilator mode used to record these variables. Most therapists (85%) made measurements in the CPAP mode, with a minority of respondents using a T-tube. In addition, there is no uniformity in the amount of supplemental pressure present during measurement. Added pressure during measurement adds another layer of complexity to this issue. Minimal amounts of pressure are based on the premise that the pressure eliminates the resistance of the endotracheal tube, providing a more accurate assessment of a patient’s status.17 It is unclear if this is a valid assumption. There is no consensus as to the exact pressure required to compensate for the endotracheal tube, and there may not be an attainable consensus.18,19 A patient’s work of breathing can be reduced with small amounts of CPAP or PS.20 It follows that measurement over the range of pressures (1 to 10 cm H2O) reported in this survey could also change a patient’s weaning parameters. The effect of added pressure on one parameter (Fb/Vt ratio) alters its ability to predict successful extubation. The original report identified a threshold value of 105, obtained with patients breathing through a T-tube.2 Other investigators have examined the ratio with patients receiving partial support (CPAP or PS) or combinations of both modes with different levels of pressure or have followed the ratio over time.21–23 They have either reported a decline in its predictive capability or a need to adjust the ratio higher when using added pressure. The time to record values represents the other major area of variation in measurement. The responses (⬍ 1 s to 15 min) in Table 2 outline this variation. A better appreciation of the importance of time can be gained by reviewing the MIP. Early investigations utilized 30 to 50 s of airway occlusion in this measurement.24,25 Marini and colleagues26 demonstrated continued increasing negative pressures that plateau after 20 s. Branson and colleagues27 found maximal pressures required a minimum of 12 s, and recommend 15 s of airway occlusion. It follows that the MIP should be recorded after 15 to 25 s of airway occlusion.14 In this survey, only 5% reported airway occlusion for 16 to 20 s. None responded with ⬎ 20 s. The most frequent response of 2 to 4 s is an inadequate occlusion time, and would likely underestimate this parameter. Almost 40% report using the ventilator software package. This might minimize concerns with standardization, but these packages also underestimate the MIP.28,29 This becomes evident when examining the MIP measurement of one ventilator. In the Puritan Bennett 7200a ventilator (Puritan Bennett; Carlsbad, CA), activation of the maneuver initiates a 10-s time frame for measurement.30 Inspiration CHEST / 121 / 6 / JUNE, 2002 1951 Figure 2. Breakdown by hospital and representing the percentage of responses from each hospital of the occlusion time used when measuring the MIP. sec ⫽ second. closes the proportional solenoid valves of the ventilator and occludes the airway for 3 s. Exhalation ends the maneuver, and the maximum negative pressure generated is reported as the MIP. The 3-s period is much less than the recommended time frame for airway occlusion, and only one breath is reported. Other software packages may have different algorithms, but it is not surprising that this package underestimates the MIP. Based on this survey, the vast majority of patients undergoing MIP measurements do not have the airway occluded for the recommended amount of time. Other areas that may lead to variations in the MIP include issues with reproducibility, maximal effort, and test-to-test variation.31,32 If decision making is based on this parameter, numerous management issues can be affected, including the timing of “weaning trials,” extubation, and tracheostomy. Although there is much debate on the clinical utility of weaning parameters, it is self-evident that they must be obtained in a standardized fashion. Only then can values be considered reflective of a patient’s status and confidently incorporated into clinical decision making. It is acknowledged that the role of weaning parameters has decreased over the past decade. However, one third of therapists in this survey report making measurements more than six times a week, and only two respondents reported zero parameters a week. This suggests that physicians still use these measurements in patient management. Based on responses to this survey, it is conceivable 1952 that weaning parameters obtained in the same hospital on the same day by different therapists on the same patient could be very different because of differences in measurement technique. Patients with borderline weaning parameters would be most affected by this lack of standardization, but all patients risk misclassification. This study was designed to assess the measurement of weaning parameters by the respiratory therapists of one community. Although limited in scope, it represents both private and public hospitals and is likely representative of the practice in other communities. These results also raise questions about other practices, specifically trials of spontaneous breathing. Although this was not specifically addressed by the questionnaire, there is likely a similar degree of variation with respect to the conduct of the trials using T-tube, CPAP, or CPAP plus PS. The results of this survey reinforce the need for standardization of all techniques, as well as continued study of the parameters that may better identify a patient’s ability to tolerate extubation. Appendix Weaning Parameters Survey Questionnaire 1. How many inpatient beds are in your hospital? ⱕ 100 101 to 499 ⱖ 500 Not sure Clinical Investigations in Critical Care 2. How many total ICU beds are in your hospital? ⱕ 10 11 to 24 25 to 49 50 –99 ⱖ 100 3. What types of ICUs are in your hospital? Medical ICU Surgical ICU Pediatric ICU Coronary care unit Combined medical-surgical ICU Step-down unit with ventilators Step-down unit without ventilators Specialized ICU (ie, burn), please specify 4. Does your hospital have a chronic ventilator unit (ward setting) without monitors? Yes No 5. What best describes the hospital that is your primary practice? Community hospital (example, Cedars-Sinai, Long Beach Memorial) County or municipal hospital (example, LAC-USC, Harbor General) University teaching hospital (example, UCLA, UCI Medical Center) Military or Veterans Affairs hospital Specialized care hospital (example, Vencor, Barlow) 6. What best describes the physicians who most often manage ventilator patients in your hospital? General internist, general surgeon Specialty internist (ie, critical care, pulmonary) Specialty surgeon (cardiothoracic surgeon) Anesthesiologist Physicians in training (ie, interns, residents, fel lows) 7. Please mark the best description of your status. Certified respiratory therapist Registered respiratory therapist Director of respiratory therapy unit (nonphysician) Other, please specify 8. How long have you been in respiratory therapy? ⬍ 1 year 1 to 5 years 6 to 19 years 11 to 19 years ⱖ 20 years 9A. Do “weaning parameters” (or extubation criteria) require a physicians order? Yes (if yes, skip to question 10) No (if no, proceed to question 9B) 9B. If physician’s orders are not required for weaning parameters, are the parameters obtained on a regular basis in ventilator patients? Yes, please specify: daily, weekly No Other, please specify 10. How many ventilator patients are you responsible for during an average shift? 1 to 4 5 to 9 ⱖ 10 Other, please specify 11. Is there a specified time in which weaning parameters are obtained? Yes, please specify Only on day shift (7 am to 3 pm) www.chestjournal.org 12. 13. 14. 15. 16. 17. 18. 19. 20. Only on morning shift (7 am to 12 noon) Only during a specified time of the morning (7 am to 10 am) No, can be obtained at any time How many times do you obtain weaning parameters per week? None 1 to 5 6 to 10 11 to 15 ⬎ 16 Do physicians order “weaning parameters” as a general rule, or do they specify specific parameters to be obtained? Weaning parameters as a general order Specify specific parameters to be obtained Either, depending on the physician Other, please specify Which weaning parameters do you routinely report? Heart rate Temperature RR Vt BP V̇e Negative inspiratory pressure or MIP Fb/Vt (rapid shallow breathing index) Oxygen saturation Vital capacity Compliance Other, please specify Do you obtain weaning parameters in unconscious patients? Yes No What minimum level of consciousness is required before proceeding with weaning parameters? Asleep, arousable with light touch Asleep, arousable with voice Awake, able to follow commands but unable to communicate by writing or mouthing words Awake, able to follow commands and communi cate by writing or mouthing words Which mode of ventilation do you use when obtaining weaning parameters? T-tube (off ventilator) CPAP, please specify amount of CPAP PS, please specify amount of PS Other, please specify After changing the vent to obtain weaning parameters, how long do you wait prior to recording the values? ⬍ 1 min 1 to 2 min 3 to 5 min 6 to 10 min 11 to 15 min ⬎ 15 min Variable length of time, please specify reason for wait Are weaning parameters obtained with the patient breathing room air or with their current Fio2? Room air Fio2 Other, please specify How do you measure RR? Direct observation Obtained from the digital readout of the ventilator Obtained from the patient’s ICU monitor readout Other, please specify CHEST / 121 / 6 / JUNE, 2002 1953 21. How do you measure Vt? Direct measurement using a hand-held spirometer or ventilation measurement nodule Obtained from instantaneous digital readout of the ventilator Calculated by dividing V̇e by frequency Other, please specify 22. How do you obtain V̇e? Direct measurement using a bedside spirometer or ventilation measurement module over 1 min Spontaneous digital readout of the ventilator Other, please specify 23. Do you routinely obtain negative inspiratory force or MIP? Yes No 24. What is the most common method of weaning patients at your hospital? Synchronized intermittent mandatory ventilation PS CPAP (no PS) Synchronized intermittent mandatory ventilation plus PS T-tube CPAP plus PS No uniform method, depends on physician Other, please specify 25. When are patients extubated at your hospital? Only during the day shift (before 3 pm) Only during the morning (before 12 noon) At any time, but never in the evening (after 8 pm) At any time Other, please specify 26. How do you measure negative inspiratory force or MIP? Manually, using a bedside spirometer Using the bedside software package Using an esophageal balloon/pulmonary monitor unit (ie, Bicore CP-100) Other, please specify 27. How long do you occlude the airway for the negative inspiratory force or MIP? ⬍1 s 2 to 4 s 5 to 10 s 11 to 15 s 16 to 20 s ⬎ 21 s 28. How many times do you repeat the negative inspiratory force or MIP? Once Twice Three times Four times Other, please specify 29. If you obtain multiple negative inspiratory force values, how do you report the final MIP? Only obtain one negative inspiratory force value, thus not applicable Average the negative inspiratory force values ob tained Use the highest negative inspiratory force value obtained Other, please specify 30. What types of ventilators do you most commonly use? Puritan Bennett 7200 Bear III or II Siemens 900C Hamilton Veolar Other, please specify 31. Do you use continuous ECG monitoring while obtaining the weaning parameters? Always (100%) Most of the time (⬎ 75%) 1954 Infrequently (⬍ 25%) Almost never (⬍ 5%) Other, please specify 32. Do you use continuous pulse oximetric monitoring when obtaining weaning parameters? Always (100%) Most of the time (⬎ 75%) Infrequently (⬍ 25%) Almost never (⬍ 5%) Other, please specify Thank your for your cooperation. We would appreciate any other comments. References 1 Manthous CA, Schmidt GA, Hall JB. Liberation from mechanical ventilation. Chest 1998; 114:886 –901 2 Yang KL, Tobin MJ. A prospective study of indexes predicting the outcome of trials of weaning from mechanical ventilation. N Engl J Med 1991; 324:1445–1450 3 Morganroth ML, Morganroth JL, Nett LM, et al. Criteria for weaning from prolonged mechanical ventilation. Arch Intern Med 1984; 144:1012–1016 4 Sassoon CSH, Te TT, Mahutte CK, et al. Airway occlusion pressure: an important indicator for successful weaning in patients with chronic obstructive pulmonary disease. Am Rev Respir Dis 1987; 135:107–113 5 Murciano D, Boczkowski J, Lecocguic Y, et al. Tracheal occlusion pressure: a simple index to monitor respiratory muscle fatigue during acute respiratory failure in patients with chronic obstructive pulmonary disease. Ann Intern Med 1988; 108:800 – 805 6 Fiastro JF, Habib MP, Shon BY, et al. Comparison of standard weaning parameters and the mechanical work of breathing in mechanically ventilated patients. Chest 1988; 94:232–238 7 Levy MM, Miyaski A, Langston D. Work of breathing as a weaning parameter in mechanically ventilated patients. Chest 1995; 108:1018 –1020 8 Vassilakopoulos T, Zakynthinos S, Roussou C. The tensiontime index and the frequency/tidal volume ratio are the major pathophysiologic determinants of weaning failure and success. Am J Respir Crit Care Med 1998; 158:378 –385 9 Epstein SK. Weaning parameters. Respir Care Clin N Am 2000; 6:253–301 10 Hess D. Measurement of maximal inspiratory pressure: a call for standardization. Respir Care 1989; 34:857– 859 11 Yang KL, Tobin MJ. Measure of minute ventilation in ventilator dependent patients: need for standardization. Crit Care Med 1991; 19:49 –53 12 Yang KL. Reproducibility of weaning parameters: a need for standardization. Chest 1992; 102:1829 –1832 13 American Association for Respiratory Care. Clinical practice guideline: removal of the endotracheal tube. Respir Care 1999; 44:85–90 14 Hess DR. Mechanical ventilation in the adult patient: initiation, management and weaning. In: Burton GC, Hodgkin JE, Ward JJ, eds. Respiratory care: a guide to clinical practice. 4th ed. New York, NY: Lippincott, 1997; 611– 641 15 Sahn SA, Lakshminarayan S. Bedside criteria for discontinuation of mechanical ventilation. Chest 1973; 63:1002–1005 16 Ely EW, Baker AM, Dunagan DP, et al. Effect of the duration of mechanical ventilation on identifying patients capable of breathing spontaneously. N Engl J Med 1996; 335:1864 –1869 17 Brochard L, Rua F, Lorina H, et al. Inspiratory pressure Clinical Investigations in Critical Care 18 19 20 21 22 23 24 support compensates for the additional work of breathing caused by the endotracheal tube. Anesthesiology 1991; 75: 739 –754 Nathan SD, Ishaaya AM, Koerner SK, et al. Prediction of minimal pressure support during weaning from mechanical ventilation. Chest 1993; 103:1215–1219 Straus C, Louis B, Isabey D, et al. Contribution of the endotracheal tube and the upper airway to breathing workload. Am J Respir Crit Care Med 1998; 157:23–30 Sassoon CSH, Light RW, Lodia R, et al. Pressure-time product during continuous positive airway pressure, pressure support ventilation and T-piece during weaning from mechanical ventilation. Am Rev Respir Dis 1991; 143:469 – 475 Lee KH, Hui KP, Chan TB, et al. Rapid shallow breathing (frequency-tidal volume ratio) did not predict extubation outcome. Chest 1994; 105:540 –543 Krieger BP, Isber J, Breitenbucheer A, et al. Serial measurements of the rapid-shallow-breathing index as a predictor of weaning outcome in elderly medical patients. Chest 1997; 112:1029 –1034 Johannigman JA, Davis K, Campbell RS, et al. Use of the rapid/shallow breathing index as an indicator of patient work of breathing during pressure support ventilation. Surgery 1997; 122:737–741 Bendixen HH, Bunker JP. Measurement of inspiratory force www.chestjournal.org in anesthesized dogs. Anesthesiology 1962; 23:315–323 25 Wescott DA, Bendixen HH. Neostigmine as a curare antagonist: a clinical study. Anesthesiology 1962; 23:324 –332 26 Marini JJ, Smith TC, Lamb V. Estimation of inspiratory muscle strength in mechanically ventilated patients: the measurement of maximal inspiratory pressure. J Crit Care 1986; 1:32–38 27 Branson RD, Hurst JM, Davis K, et al. Measurement of maximal inspiratory pressure: a comparison of three methods. Respir Care 1989; 34:789 –794 28 Kacmarek RM, Cycyk-Chapman MC, Young-Palazzo PJ, et al. Determination of maximal inspiratory pressure: a clinical study and literature review. Respir Care 1989; 34:868 – 878 29 Brooks CW, Dennison F, Mishoe SC, et al. A comparison of two different maximal inspiratory pressure procedures [abstract]. Respir Care 1996; 41:916 30 7200 series microprocessor ventilator operator’s manual, respiratory mechanics, option 30/40. Carlsbad, CA: Puritan Bennett Corporation, 1988; 1–21 31 Aldrich TK, Spiro P. Maximal inspiratory pressure: does reproducibility indicate full effort? Thorax 1995; 50:40 – 43 32 Multz AS, Aldrich TK, Prezant DJ, et al. Maximal inspiratory pressure in not a reliable test of inspiratory muscle strength in mechanically ventilated patients. Am Rev Respir Dis 1990; 142:529 –532 CHEST / 121 / 6 / JUNE, 2002 1955