Lab Report 4
Lab Report 4
Walt Williams
Group 2
ABSTRACT:
When a company designs a flow meter, the design will include
the discharge coefficient. This coefficient represents the friction losses
in the flow meter. In lab tests, the coefficients of two different meters
were determined. A Venturi meter had a value of .907 and an orifice
meter had a value of .761. The same factors that contribute to these
coefficients also contribute to energy loss through the meter. Energy
losses were also calculated in the lab. A rotameter had the highest
energy loss, while the Venturi had the lowest.
INTRODUCTION:
Obstruction flow meters are simple mechanisms fitted to the
inside of pipes to measure the flow inside the pipe. This lab tests three
different types of obstruction flow meters; Venturi, orifice, and
rotameter.
Venturi and orifice flow meters act by reducing the cross
sectional area of the pipe. The reduction in area increases velocity,
which leads to a decrease in pressure. The pressure drop can be
measured and used to calculate velocity using Bernoullis equation
(Equation 1).
2
Equation 1:
P1 V 1
P V
z1 g= 2 2 z 2 g
Though the two act using the same principles, their construction is
different. An orifice flow meter is comprised of a plate with a circular
hole in the middle, while a Venturi flow meter is a gradual decrease in
cross sectional area, then a gradual increase back to the original area
once reaching the smallest point.
The only material used for this lab was a FNE18 Flow Meter
Module from Edibon. This device works by pumping water through a
hose that connects to three different flow meters: a venture, an orifice,
and a rotameter. Each flow meter is connected to a set of manometers
that measure differential pressure across the device. This allows
pressure drop and energy losses, to be measured. Velocity of the liquid
in the hose is given by the rotameter, and can also be calculated by
measuring flow rate out of the system.
Methods:
The lab consisted of taking 6 different readings at different
velocities between 250 and 1250 L/h. The Edibon system was set to a
velocity and pressure drop was recorded for each of the three flow
meters. Velocity was also manually tested by recording the liters of
fluid flow per minute. Once velocity and pressure were recorded, the
system was set to a new velocity and the process was repeated.
Calculations:
Over the duration of the lab time, values for velocity and
pressure were recorded. All of the following calculations use those
recorded values and common assumptions (such as density of water
and force of gravity).
Bernoullis equation is used to find energy loss. In this situation,
it can be simplified down to Equation 2. Fluid height in the manometer
is converted into energy per unit mass. The variables
z b and
za
stand for fluid height in the manometer before flow meter, and after
the flow meter, respectively. The difference between the two is energy
loss, calculated in kJ/kg.
z
Equation 2:
g
1 kJ
( bz a)
1000 J
E L =
Pd =g( z bz a )
Equation 4:
Slo pe=C d A2
( [ ])
2
A2
1
A1
C d=0.5959+ 0.0312
d 2.1
d8
0.184
+
D
D
91.71
d 2.5
D
0.75
Venturi meters have a higher Cd value than orifice meters, they have
lower energy losses. Low energy losses make Venturi meters the most
efficient and have the lowest pressure drop of any of the three flow
meters tested in this exercise. Friction is significantly higher in
rotameters than either of the other two flow meters; this causes it to
have much higher energy losses. However, because the fluid moves
around the rotameter float, the energy loss is fairly constant regardless
of flow rate.
The slope of the trend lines in Figure 2 has been plugged into
equation 4, which was solved for the Cd value. The experimental Cd for
Venturi meters was 0.907. This is less than an 8% difference from the
general assumption for Venturi flow meters of 0.98 (Cengel and
Cimbala). The experimental orifice meter Cd was calculated in a similar
manner, and turned out to be 0.761. Orifice Cd was also calculated
using Equation 6. This Cd value turned out to 0.611, a 22% difference
from the experimental value. The differences in the two pairs of Cd
values can be attributed to inaccurate readings of the Edibon system,
and imperfect approximations made for the two calculated values.
TABLES AND FIGURES:
Venturi
Rotameter
Orifice
0
0
f(x) = 0x
0
0
Flow Rate (m^3/s)
Venturi
0
Linear (Venturi)
0
Rotameter
Orifice
Linear (Orifice)
0
0
0
10
20
30
40
50
60