This document provides testing procedures to analyze sterile single-use plastic syringes for various quality attributes. It describes tests for ethylene oxide residues using gas chromatography with a flame ionization detector. A calibration curve is prepared using standard solutions of known ethylene oxide concentrations in dimethylacetamide. Additional tests evaluate syringes for clarity, acidity, absorbance, particles, flow rate, pressure resistance, transparency, residue levels, sterility, and pyrogens. The syringe labeling is also specified.
This document provides testing procedures to analyze sterile single-use plastic syringes for various quality attributes. It describes tests for ethylene oxide residues using gas chromatography with a flame ionization detector. A calibration curve is prepared using standard solutions of known ethylene oxide concentrations in dimethylacetamide. Additional tests evaluate syringes for clarity, acidity, absorbance, particles, flow rate, pressure resistance, transparency, residue levels, sterility, and pyrogens. The syringe labeling is also specified.
This document provides testing procedures to analyze sterile single-use plastic syringes for various quality attributes. It describes tests for ethylene oxide residues using gas chromatography with a flame ionization detector. A calibration curve is prepared using standard solutions of known ethylene oxide concentrations in dimethylacetamide. Additional tests evaluate syringes for clarity, acidity, absorbance, particles, flow rate, pressure resistance, transparency, residue levels, sterility, and pyrogens. The syringe labeling is also specified.
This document provides testing procedures to analyze sterile single-use plastic syringes for various quality attributes. It describes tests for ethylene oxide residues using gas chromatography with a flame ionization detector. A calibration curve is prepared using standard solutions of known ethylene oxide concentrations in dimethylacetamide. Additional tests evaluate syringes for clarity, acidity, absorbance, particles, flow rate, pressure resistance, transparency, residue levels, sterility, and pyrogens. The syringe labeling is also specified.
Carrier gas : helium for chromatography R. Flow rate: 20 mL/min. Temperature: column: 40 C; injection port : 100 C; detector: 150 C. Detection: ame ionisation. Verify the absence of peaks interfering with the ethylene oxide peak by carrying out the test using an unsterilised set or using the same chromatographic system with the following modications. Column: size: l = 3 m, = 3.2 mm; stationary phase: silanised diatomaceous earth for gas chromatography R impregnated with triscyanoethoxypropane R (2 g per 10 g) ; temperature: 60 C. Ethylene oxide solution. Prepare in a fume cupboard. Place 50.0 mL of dimethylacetamide R in a 50 mL vial, stopper, secure the stopper and weigh to the nearest 0.1 mg. Fill a 50 mL polyethylene or polypropylene syringe with gaseous ethylene oxide R, allow the gas to remain in contact with the syringe for about 3 min, empty the syringe and ll again with 50 mL of gaseous ethylene oxide R. Fit a hypodermic needle to the syringe and reduce the volume of gas in the syringe from 50 mL to 25 mL. Inject these 25 mL of ethylene oxide slowly into the vial, shaking gently and avoiding contact between the needle and the liquid. Weigh the vial again: the increase in mass is 45 mg to 60 mg and is used to calculate the exact concentration of the solution (about 1 g/L). Test. Weigh the set after removing the package. Cut the set into pieces of maximum dimension 1 cm and place the pieces in a 250-500 mL vial containing 150 mL of dimethylacetamide R. Close the vial with a suitable stopper and secure the stopper. Place the vial in an oven at 70 1 C for 16 h. Remove 1 mL of the hot gas from the vial and inject it onto the column. From the calibration curve and the height of the peak obtained, calculate the mass of ethylene oxide in the vial. Calibration curve. In a series of 7 vials of the same type as that used for the test and each containing 150 mL of dimethylacetamide R, place respectively 0 mL, 0.05 mL, 0.10 mL, 0.20 mL, 0.50 mL, 1.00 mL and 2.00 mL of the ethylene oxide solution, i.e. about 0 g, 50 g, 100 g, 200 g, 500 g, 1000 g and 2000 g of ethylene oxide. Stopper the vials, secure the stoppers and place the vials in an oven at 70 1 C for 16 h. Inject 1 mL of the hot gas from each vial onto the column and draw a calibration curve from the heights of the peaks and the mass of ethylene oxide in each ask. Limit : if the label states that ethylene oxide has been used for sterilisation: ethylene oxide: maximum 10 ppm. Extraneous particles. Fill the set via the normal inlet with a 0.1 g/L solution of sodium laurilsulfate R, previously ltered through a sintered-glass lter (16) (2.1.2) and heated to 37 C. Collect the liquid via the normal outlet. When examined under suitable conditions of visibility, the liquid is clear and practically free from visible particles and laments (it is assumed that particles and laments with a diameter equal to or greater than 50 m are visible to the naked eye). Flow rate. Pass through a complete set with the ow regulator fully open 50 mL of a solution having a viscosity of 3 mPas (3 cP) (for example a 33 g/L solution of macrogol 4000 R at 20 C) under a static head of 1 m. The time required for passage of 50 mL of the solution is not greater than 90 s. Resistance to pressure. Make tight the extremities of the set and any air-inlet device. Connect the set to a compressed air outlet tted with a pressure regulator. Immerse the set in a tank of water at 20-23 C. Apply progressively an excess pressure of 100 kPa and maintain for 1 min. No air bubble escapes from the set. Transparency. Use as reference suspension the primary opalescent suspension (2.2.1) diluted 1 in 8 for sets having tubing with an external diameter less than 5 mm and diluted 1 in 16 for sets having tubing with an external diameter of 5 mm or greater. Circulate the reference suspension through the set and compare with a set from the same batch lled with water R. The opalescence and presence of bubbles are discernible. Residue on evaporation. Evaporate 50.0 mL of solution S to dryness on a water-bath and dry to constant mass in an oven at 100-105 C. Carry out a blank test using 50.0 mL of water for injections R. The difference between the masses of the residues is not greater than 1.5 mg. Sterility (2.6.1). The sets comply with the test for sterility. If the sets are stated to be sterile only internally, pass 50 mL of buffered sodium chloride-peptone solution pH 7.0 (2.6.12) through the set and use to carry out the test by the membrane ltration method. If the sets are stated to be sterile both internally and externally, open the package with the necessary aseptic precautions and: for the direct inoculation method, place the set or its components in a suitable container containing a sufcient quantity of the culture medium to ensure complete immersion; for the membrane ltration method, place the set or its components in a suitable container containing a sufcient quantity of buffered sodium chloride-peptone solution pH 7.0 (2.6.12) to allow total rinsing for 10 min. Pyrogens (2.6.8). Connect together 5 sets and pass through the assembly at a ow rate not exceeding 10 mL/min 250 mL of a sterile, pyrogen-free 9 g/L solution of sodium chloride R. Collect the solution aseptically in a pyrogen-free container. The solution complies with the test for pyrogens. Inject per kilogram of the rabbits mass, 10 mL of the solution. LABELLING The label states, where applicable, that the set has been sterilised using ethylene oxide. 01/2008:30208 3.2.8. STERILE SINGLE-USE PLASTIC SYRINGES DEFINITION Sterile single-use plastic syringes are medical devices intended for immediate use for the administration of injectable preparations. They are supplied sterile and pyrogen-free and are not to be re-sterilised or re-used. They consist of a syringe barrel and a piston which may have an elastomer sealing ring; they may be tted with a needle which may be non-detachable. Each syringe is presented with individual protection for maintaining sterility. The barrel of the syringe is sufciently transparent to permit dosages to be read without difculty and allow air bubbles and foreign particles to be discerned. The plastics and elastomer materials of which the barrel and piston are made comply with the appropriate specication or with the requirements of the competent authority. The most commonly used materials are polypropylene and polyethylene. The syringes comply with current standards regarding dimensions and performance. Silicone oil (3.1.8) may be applied to the internal wall of the barrel to assist in the smooth operation of the syringe but there remains no excess capable of contaminating the contents at the time of use. The inks, glues and adhesives for the marking General Notices (1) apply to all monographs and other texts 419 3.2.8. Sterile single-use plastic syringes EUROPEAN PHARMACOPOEIA 8.0 on the syringe or on the package and, where necessary, the assembly of the syringe and its package, do not migrate across the walls. TESTS Solution S. Prepare the solution in a manner that avoids contamination by foreign particles. Using a sufcient number of syringes to produce 50 mL of solution, ll the syringes to their nominal volume with water for injections R and maintain at 37 C for 24 h. Combine the contents of the syringes in a suitable borosilicate-glass container. Appearance of solution. Solution S is clear (2.2.1) and colourless (2.2.2, Method II) and is practically free from foreign solid particles. Acidity or alkalinity. To 20 mL of solution S add 0.1 mL of bromothymol blue solution R1. Not more than 0.3 mL of 0.01 M sodium hydroxide or 0.01 M hydrochloric acid is required to change the colour of the indicator. Absorbance (2.2.25) : maximum 0.40, determined between wavelengths of 220 nm and 360 nm on solution S. Ethylene oxide. Gas chromatography (2.2.28). Column: material : stainless steel ; size: l = 1.5 m, = 6.4 mm; stationary phase: silanised diatomaceous earth for gas chromatography R impregnated with macrogol 1500 R (3 g per 10 g). Carrier gas : helium for chromatography R. Flow rate: 20 mL/min. Temperature: Column: 40 C; Injection port : 100 C; Detector: 150 C. Detection: ame ionisation. Verify the absence of peaks interfering with the ethylene oxide peak, either by carrying out the test using an unsterilised syringe or using the same chromatographic system with the following modications : Column: size: l = 3 m, = 3.2 mm; stationary phase: silanised diatomaceous earth for gas chromatography R impregnated with triscyanoethoxypropane R (2 g per 10 g) ; temperature: 60 C. Ethylene oxide solution. Prepare in a fume cupboard. Place 50.0 mL of dimethylacetamide R in a 50 mL vial, stopper, secure the stopper and weigh to the nearest 0.1 mg. Fill a 50 mL polyethylene or polypropylene syringe with gaseous ethylene oxide R, allow the gas to remain in contact with the syringe for about 3 min, empty the syringe and ll again with 50 mL of gaseous ethylene oxide R. Fit a hypodermic needle to the syringe and reduce the volume of gas in the syringe from 50 mL to 25 mL. Inject these 25 mL of ethylene oxide slowly into the vial, shaking gently and avoiding contact between the needle and the liquid. Weigh the vial again: the increase in mass is 45 mg to 60 mg and is used to calculate the exact concentration of the solution (about 1 g/L). Calibration curve. In a series of seven vials of the same type as that used for the test and each containing 150 mL of dimethylacetamide R, place respectively 0 mL, 0.05 mL, 0.10 mL, 0.20 mL, 0.50 mL, 1.00 mL and 2.00 mL of the ethylene oxide solution, i.e. about 0 g, 50 g, 100 g, 200 g, 500 g, 1000 g and 2000 g of ethylene oxide. Stopper the vials, secure the stoppers and place the vials in an oven at 70 1 C for 16 h. Inject 1 mL of the hot gas from each vial onto the column and draw a calibration curve from the heights of the peaks and the mass of ethylene oxide in each ask. Test. Weigh the syringe after removing the package. Cut the syringe into pieces of maximum dimension 1 cm and place the pieces in a 250 mL to 500 mL vial containing 150 mL of dimethylacetamide R. Close the vial with a suitable stopper and secure the stopper. Place the vial in an oven at 70 1 C for 16 h. Remove 1 mL of the hot gas from the vial and inject it onto the column. From the calibration curve and the height of the peak obtained, calculate the mass of ethylene oxide in the vial. Limit : if the label states that ethylene oxide has been used for sterilisation: ethylene oxide: maximum 10 ppm. Silicone oil. Calculate the internal surface area of a syringe in square centimetres using the following expression:
V = nominal volume of the syringe, in cubic centimetres; h = height of the graduation, in centimetres. Take a sufcient number of syringes to give an internal surface area of 100 cm 2 to 200 cm 2 . Aspirate into each syringe a volume of methylene chloride R equal to half the nominal volume and make up to the nominal volume with air. Rinse the internal surface corresponding to the nominal volume with the solvent by inverting the syringe ten times in succession with the needle tting closed by a nger covered by a plastic lm inert to methylene chloride. Expel the extracts into a tared dish and repeat the operation. Evaporate the combined extracts to dryness on a water-bath. Dry at 100-105 C for 1 h. The residue weighs not more than 0.25 mg per square centimetre of internal surface area. Examine the residue by infrared absorption spectrophotometry (2.2.24). It shows absorption bands typical of silicone oil at 805 cm 1 , 1020 cm 1 , 1095 cm 1 , 1260 cm 1 and 2960 cm 1 . Reducing substances. To 20.0 mL of solution S add 2 mL of sulfuric acid R and 20.0 mL of 0.002 M potassium permanganate. Boil for 3 min. Cool immediately. Add 1 g of potassium iodide R and titrate immediately with 0.01 M sodium thiosulfate using 0.25 mL of starch solution R as indicator. Carry out a blank titration using 20.0 mL of water for injections R. The difference between the titration volumes is not greater than 3.0 mL. Transparency. Fill a syringe with water R (blank) and ll another with a 1 in 10 dilution of primary opalescent suspension (2.2.1). Use primary opalescent suspension that has been allowed to stand at 20 2 C for 24 h before use. Compare with the naked eye in diffused light against a dark background. The opalescence of the suspension is detectable when compared with the blank. Sterility (2.6.1). Syringes stated to be sterile comply with the test for sterility carried out as follows. Using aseptic technique, open the package, withdraw the syringe, separate the components and place each in a suitable container containing sufcient culture media to cover the part completely. Use both the recommended media (2.6.1). Syringes stated to be sterile only internally comply with the test for sterility carried out as follows. Use 50 mL of inoculation medium for each test syringe. Using aseptic technique, remove the needle protector and submerge the needle in the culture medium. Flush the syringe ve times by withdrawing the plunger to its fullest extent. Pyrogens (2.6.8). Syringes with a nominal volume equal to or greater than 15 mL comply with the test for pyrogens. Fill a minimum of three syringes to their nominal volume with a pyrogen-free 9 g/L solution of sodium chloride R and maintain at a temperature of 37 C for 2 h. Combine the solutions aseptically in a pyrogen-free container and carry out the test immediately. Inject per kilogram of the rabbits mass 10 mL of the solution. 420 See the information section on general monographs (cover pages) EUROPEAN PHARMACOPOEIA 8.0 3.2.9. Rubber closures for containers LABELLING The label on the package states : the batch number ; a description of the syringe; that the syringe is for single-use only. The label on the outer package states: the method of sterilisation; that the syringe is sterile or that it is sterile only internally; the identity of the manufacturer ; that the syringe is not to be used if the packaging is damaged or the sterility protector is loose. 01/2014:30209 3.2.9. RUBBER CLOSURES FOR CONTAINERS FOR AQUEOUS PARENTERAL PREPARATIONS, FOR POWDERS AND FOR FREEZE-DRIED POWDERS Rubber closures for containers for aqueous parenteral preparations, for powders and for freeze-dried powders are made of materials obtained by vulcanisation (cross-linking), using appropriate additives, of macromolecular organic substances (elastomers). The elastomers are produced from natural or synthetic substances by polymerisation. The choice of the principal components and of the various additives (for example, vulcanisers, accelerators, stabilisers, pigments) depends on the properties required for the nished article. The specications do not apply to closures made from silicone elastomer (which are dealt with in chapter 3.1.9. Silicone elastomer for closures and tubing), to laminated closures or to lacquered closures. Rubber closures may be classied in 2 types: type I closures meet the strictest requirements and are preferred; type II closures have mechanical properties suitable for special uses (for example, multiple piercing) and cannot meet requirements as severe as for type I closures because of their chemical composition. The closures chosen for use with a particular preparation are such that : the components of the preparation in contact with the closures are not adsorbed onto the surface of the closures and do not migrate into or through the closures to an extent sufcient to affect the preparation adversely; the closures do not release substances in quantities sufcient to affect the stability of the preparation or to present a risk of toxicity. The closures are compatible with the preparation for which they are used throughout its period of validity. The manufacturer of the preparation must obtain from the supplier an assurance that the composition of the closure does not vary and that it is identical to that of the closure used during compatibility testing. If the supplier informs the manufacturer of the preparation that changes have been made to the composition, compatibility testing must be repeated, totally or partly, depending on the nature of the changes. The closures are washed and may be sterilised before use. CHARACTERS Rubber closures are elastic. They are translucent or opaque and have no characteristic colour, the latter depending on the additives used. They are practically insoluble in tetrahydrofuran, in which, however, a considerable reversible swelling may occur. They are homogeneous and practically free from ash and adventitious materials (for example, bres, foreign particles, waste rubber). Identification of the type of rubber used for the closures is not within the scope of this specification. The identification tests given below distinguish between closures made from rubber and those made from silicone elastomer and plastic materials but do not differentiate all types of rubber. Other identity tests may be carried out with the aim of detecting differences in a batch compared with the closures used for compatibility testing. One or more of the following analytical methods may be applied for this purpose: determination of relative density, determination of sulfated ash, determination of sulfur content, thin-layer chromatography carried out on an extract, ultraviolet absorption spectrophotometry of an extract, infrared absorption spectrophotometry of a pyrolysate or attenuated total reflectance (ATR). IDENTIFICATION A. Infrared absorption spectrophotometry (2.2.24). Examine by attenuated total reectance (ATR). The spectrum obtained is identical to the spectrum obtained with the type sample. If necessary, cut the sample along an appropriate axis, examine the cut surface and compare the spectrum with that obtained with the type sample prepared in the same way. If direct ATR measurement on the surface is not feasible (mainly rubber closures lled with carbon black), heat 1-2 g in a heat-resistant test-tube over an open ame to dry the sample and continue heating until pyrolysate vapours are condensed near the top edge of the test-tube. Examine the pyrolysate of the sample by ATR and compare the spectrum with that obtained with the pyrolysate of the type sample. B. Total ash (2.4.16). Determine the percentage content of total ash in the sample to be examined and compare with the percentage content of total ash in the type sample (A 0 ). The total ash content falls within the following ranges depending on the total ash content of the type sample. Total ash in the type sample, A 0 (per cent) Limit for total ash in the sample (per cent) A 0 5.0 (A 0 0.75) to (A 0 + 0.75) 5.0 < A 0 10 (A 0 1.0) to (A 0 + 1.0) A 0 > 10 (A 0 2.0) to (A 0 + 2.0) In addition to the use of platinum and silica crucibles described in general chapter 2.4.16, porcelain crucibles may be used. The sample may be ignited using a microwave oven instead of a mufe furnace. TESTS The samples to be analysed may be washed and sterilised before use. Solution S. Place a number of uncut closures with a total surface area of about 100 cm 2 in a suitable glass container, cover with water R, boil for 5 min and rinse 5 times with cold water R. Place the washed closures in a wide-necked ask (type I glass, 3.2.1), add 200 mL of water R and weigh. Cover the mouth of the ask with a borosilicate-glass beaker. Heat in an autoclave so that a temperature of 121 2 C is reached within 20-30 min and maintain at this temperature for 30 min. Cool to room temperature over about 30 min. Make up to the original mass with water R. Shake and decant the solution immediately. Shake solution S before each test. If using a tightly closed ask (type I glass, 3.2.1) with an inert closure instead of a wide-necked ask covered with a borosilicate-glass beaker, it is not necessary to make up to the original mass. Blank solution. Prepare a blank solution in the same manner using 200 mL of water R. General Notices (1) apply to all monographs and other texts 421