Bench Multimeters

45 Dual Display Multimeter


Multifunction Vacuum Fluorescent Dual Display True-RMS Voltage and Current, Including ac + dc RS-232C Interface Standard, GPIB/IEEE-488.2 Optional Frequency Measurement to 1 MHz dB Measurements with Selectable Reference Impedances of 2 to 8000 and Audio Power from 2 to 16 Compare (Hi/Lo/Pass) Function for Quick In-Tolerance Tests
Fluke 45

0.05% dc Current Accuracy for 4-20 mA Current Loop Service Touch Hold, Relative and Min Max Audible Continuity and Diode Test Optional Rechargeable Battery, Carrying Case, Rack Mount Kit & PC Software Package Closed-Case Calibration

The Fluke 45 is a feature-rich 5 digit, 100,000 count DMM with a unique multifunction dual display, allowing measurement of two signal parameters from a single test connection. The Fluke 45 offers high performance and versatility for manufacturing test, depot and field service, and research and development. A standard RS-232C makes it ideal for PC instrument applications.

impedances, the meter automatically calculates and displays audio power in watts.

Closed-Case Calibration
Calibration can be performed via the RS-232C (or optional IEEE-488.2) interface or manually from the front panel. No internal adjustments are required.

Compare Function
The Fluke 45 has a compare function for fast in-tolerance limits testing. Upper- and lower-limits are entered through the front panel. Readouts show both a Hi/Lo/Pass evaluation and measured value.

Standard Equipment
Each Fluke 45 Dual Display Multimeter includes an operators manual, quick reference guide, line cord, and test leads.

Dual Display
The Fluke 45 is the first DMM with a multifunction dual display, allowing the user to select a wide variety of measurement combinations. It is particularly useful in applications requiring two different measurements of the same signal; i.e. power supply testing, where Vdc output can be viewed on the primary display while the Vac ripple is shown on the secondary display.

Touch Hold Relative and Min/Max


Touch Hold captures the measurement, beeps and locks it on the digital display until you are ready to view it. It automatically updates with each new stable measurement. The Relative mode remembers a reading and shows the change (difference) between it and any readings that follow. Min/Max records the highest and lowest values measured. Either can be recalled and displayed at any time.

Specifications
Accuracy specifications are given as ([% of reading] + [number of least significant digits]) at 18C to 28C with relative humidity up to 90%, for a period of one year after calibration. Six months specifications are also provided for dc volts. AC inputs are ac-coupled and True-RMS responding.

Technical Specifications

Standard RS-232C Interface


The RS-232C interface, standard in each instrument, allows measurement data to be filed, manipulated, printed or transmitted by modem. The print mode automatically formats measurement data for printing on an RS-232C printer. Rates for automated printing over RS-232C are adjustable from 1 reading every 70 ms to 1 reading every 5.6 hours. The optional QuickStart 45TM Software Package allows automated communications and filing of measurements with the Fluke 45 and an IBM-PC or compatible via RS-232C.

Optional Battery and Case


An optional rechargeable battery and soft carrying case are available for precision field service applications. These options, coupled with a 30 mA dc current range and 0.05% accuracy, allow calibration of 4-20 mA loops in process control applications. The battery is available as a factory installed option or can be user-installed at a later date.

Display
Dual vacuum fluorescent displays, 99,999 counts each display. Annunciators: m, M, k, V, A, (ohms), Hz, ? + diode test, (audible continuity), REMOTE, EXT TRIG, SMF (reading rates), MAX, MIN, dB, HOLD, REL, AUTO, + (low battery). Resolution and accuracy are dependent on selectable reading rates of 2.5 (slow), 5 (medium) or 20 (fast) readings per second.
Rate Slow Medium Fast Readings Per Second 2.5 5 20 Full Range Display Counts 99,999 30,000 3000

Optional GPIB/IEEE-488.2 and Rack Mount


The Fluke 45 may be used with GPIB/IEEE488.2 systems, including existing IEEE488 implementations. The IEEE- 488.2 option is available as a factory installed option, or can be user installed and does not require removal of the RS-232C interface. A rack mount kit is also available.

dB Measurement
The Fluke 45 provides digital readout of decibels with front panel selection of any of twenty-one reference impedances from 2 to 8000. For 2, 4, 8 and 16

Bench Multimeters
45 Dual Display Multimeter

RS-232C and IEEE-488 Reading Transfer Rates


Internal Trigger Operation (TRIGGER 1) Rate Slow Medium Fast Rate Slow Medium Fast Rate Slow Medium Fast Readings Per Second 2.5 4.5 4.5 Readings Per Second 1.5 2.4 3.8 Readings Per Second 2.5 5.0 13.5

DC Voltage
Range Slow 300 mV 3V 30V 300V 1000V 100 mV 1000 mV 10V 100V 1000V 1 V 10 V 100 V 1 mV 10 mV Resolution Medium 10 V 100 V 1 mV 10 mV 100 mV Fast 100 V 1 mV 10 mV 100 mV 1V Accuracy (6 Month) Accuracy (1 Year)

0.02% + 2

0.025% + 2

External Trigger Operation (TRIGGER 4)

0.02%+ 6

0.025% + 6

Print Mode Operation (Print set at 1)

Maximum Input: To be used in protected, low-energy circuits only, not to exceed 250V or 4800 volt-amps mA: 300 mA dc or ac rms. Protected with a 500 mA, 250V, IEC 127 sheet 1 fast blow fuse A: 10A dc or ac rms continuous or 20A dc or ac rms for 30 seconds maximum. Protected with a 15A, 250V, 10,000A interrupt rating fast blow fuse.

Input Impedance: 10 M in parallel with <100 pF Normal Mode Rejection Ratio: >80 dB at 50 or 60 Hz, slow and medium rates; >54 dB for frequencies between 50 and 440 Hz, slow and medium rates; >60 dB at 50 Hz, fast rate Common Mode Rejection Ratio: >90 dB at dc, 50 or 60 Hz (1 k unbalanced, slow & medium rates) Maximum Input: 1000V dc or peak ac on any range

AC Voltage (True-RMS, ac coupled)


Range Slow 300 mV 3V 30V 300V 750V Slow/Medium 0.01 dB Frequency Slow Linear Resolution Medium 10 V 100 V 1 mV 10 mV 100 mV Fast 0.1 dB Linear Accuracy Medium Fast 7%+2 0.5%+2 0.5%+2 2%+3 5%+6 dB Accuracy Slow/Med 0.15 0.08 0.11 0.29 0.70 Fast 0.72 0.17 0.17 0.34 0.78 2%+10 0.4%+10 1%+10 4%+20 10%+50 Power* Max Input Upper Freq 750V 750V 750V 400V 200V Fast 100 V 1 mV 10 mV 100 mV 1V 100 mV 1000 mV 10V 100V 750V Range Slow 1 V 10 V 100 V 1 mV 10 mV Linear Resolutions Medium Fast

Diode Test/Continuity
Accuracy Slow Medium Fast Maximum Reading 999,99 mV 2.5V 2.5V Resolution 10 V 100 V 1 mV

Decibels Resolution

Test Current: Approximately 0.7 mA when measuring forward bias junction Audible Tone: Continuous tone for continuity. Brief tone for normal forward biased diode or semiconductor junction Open Circuit Voltage: 3.2V maximum Continuity Capture Time: 50 s maximum, 10 s typical Input Protection: 500V dc or ac rms

20-50 Hz 1%+100 1%+10 50 Hz-10 kHz 0.2%+100 0.2%+10 10-20 kHz 0.5%+100 0.5%+10 20-50 kHz 2%+200 2%+20 50-100 kHz 5%+500 5%+50

*Error in power mode will not exceed twice the linear accuracy specification

Accuracy specifications apply within the following limits, based on reading rate: Slow: Between 15,000 counts and full range Medium: Between 1,500 counts and full range Fast: Between 150 counts and full range Input Impedance: 1 M in parallel with <100 pF Maximum Crest Factor: 3.0 Common Mode Rejection Ratio: >60 dB at 50 or 60 Hz (1 k unbalanced medium rate) Maximum Input: 750V rms, 1000V peak 2 107 volt-hertz product on any range, normal mode input 1 106 volt-hertz product on any range, common mode input

AC + DC Voltage (calculated)

Total measurement error will not exceed the sum of the separate ac and dc accuracy specifications, plus one display count.

Bench Multimeters
45 Dual Display Multimeter

Ohms
Range Slow 300 3 k 30 k 300 k 3 M 30 M 300 M 100 1000 10 k 100 k 1000 k 10 M 100 M 1 m 10 m 100 m 1 10 100 100 k Resolution Medium 10 m 100 m 1 10 100 1 k 100 k Fast 100 m 1 10 100 1 k 10 k 1 M 0.05% + 2 + 0.02 0.05% + 2 0.05% + 2 0.05% + 2 0.06% + 2 0.25% + 3 2% 0.05% + 8 + 0.02 0.05% + 8 + 0.02 0.05% + 8 0.05% + 8 0.06% + 8 0.25% + 6 2% + 2 Accuracy Typical Full Scale Voltage 0.25 0.24 0.29 0.29 0.30 2.25 2.90 0.09 0.10 0.11 0.11 0.12 1.50 2.75 Max Current Thru Unknown 1 mA 120 A 14 A 1.5 A 150 nA 320 nA 320 nA 1 mA 120 A 14 A 1.5 A 150 nA 150 nA 320 nA

Frequency Range: 5 Hz to >1 MHz Accuracy: 0.05% + 1, except 1000 Hz range, 0.05% + 2
Range 1000 Hz 10 kHz 100 kHz 1000 kHz 1 MHz* Resolution Slow/Medium 10 mHz 100 mHz 1 Hz 10 Hz 100 Hz Fast 100 mHz 1 Hz 10 Hz 100 Hz 1 kHz

Frequency

* Specified to 1 MHz, usable to 8 MHz when overdriven Sensitivity of AC Voltage Frequency 5 Hz-100 kHz 100 kHz-300 kHz 300 kHz-1 MHz Above 1 MHz Level 30 mV rms sinewave 100 mV rms sinewave 1V rms sinewave Not specified

Open Circuit Voltage: 3.2V max on 100, 300, 30 M, 100 M & 300 M ranges; 1.5V max on all other ranges Input Protection: 500V dc or ac rms on all ranges

DC Current
Range Slow 30 mA 100 mA 10A 10 mA 100 mA 10A 100 nA 1 A 100 A Resolution Medium 1 A 10 A 1 mA Fast 10 A 100 A 10 mA 0.05% + 3 0.05% + 2 0.2% + 5 0.05% + 20 0.05% + 5 0.2% + 7 Accuracy Typical Full Scale Burden Voltage 0.45V 1.4V 0.25V 0.14V 1.4V 0.25V

Sensitivity Level of AC Current Frequency Input Level 5 Hz-20 kHz 100 mA > 3 mA rms sinewave 45 Hz-2 kHz 10A > 3A rms sinewave

General Specifications
Power Requirements
Power: 90 to 264V ac (no switching required), 50 and 60 Hz <15 VA minimum

Maximum Crest Factor: 30 Maximum Input: To be used in protected, low-energy circuits only, not to exceed 250V or 4800 volt-amps mA: 300 mA dc or ac rms. Protected with a 500 mA, 250V, IEC 127 sheet 1 fast blow fuse. A: 10A dc or ac rms continuous or 20A dc or ac rms for 30 seconds maximum. Protected with a 15A, 250V, 10,000A interrupt rating fast blow fuse.

Environmental Data
Maximum Common Mode Voltage: 1000V dc or peak ac from any input to earth Warmup Time: 1 hour to rated specifications Temperature Coefficient: <0.1 times the applicable accuracy specifications per degree C from 0C to 18C and from 28C to 50C, (32F to 64.4F and 82.4F to 122F) Operating Temperature: 0C to 50C (32F to 122F) Storage Temperature: 40C to 70C (40F to 158F). Elevated temperature storage of battery will accelerate battery self-discharge. Maximum storage time before battery must be recharged: 20C to 25C 1000 days 50C 180 days 70C 40 days Relative Humidity (non-condensing): Up to 90%, 0C to 28C (32F to 82.4F); up to 80%, 28C to 35C (82.4 F to 95F); up to 70%, 35C to 50C (95F to 122F), except 70%, 0C to 50C (32F to 122F) for the 1 M, 3 M, 10 M, 30 M, 100 M and 300 M ranges. Altitude: Operating, 0 to 10,000 feet; non-operating, 0 to 40,000 feet Vibration: 3G @ 55 Hz per MIL-T-28800D, Class 3, Style E Shock: Half-sine 40G per MIL-T-28800D, Class 3, Style E. Bench handling.

AC Current (True-RMS, ac coupled)


Range Slow 10 mA 30 mA 100 mA 10A Range mA mA mA A (1A to 10A) A (1A to 10A) A (0.5A to 1A) A (0.5A to 1A) 100 nA 1 A 100 A Frequency Slow 20-50 Hz 50 Hz-10 kHz 10-20 kHz 20-50 Hz 50 Hz-2 kHz 20-50 Hz 50 Hz-2 kHz 2% + 100 0.5% + 100 2% + 200 2% + 100 1% + 100 2% + 300 1% + 300 Resolution Medium 1 A 10 A 1 mA Fast 10 A 100 A 10 mA Accuracy Medium 2% + 10 0.5% + 10 2% + 20 2% + 10 1% + 10 2% + 30 1% + 30 Fast 7% + 2 0.8% + 2 2% + 3 7% + 2 1.3% + 2 7% + 4 1.3% + 4 Typical Full Scale Burden Voltage 0.14V 0.45V 1.4V 0.25V

Accuracy specifications apply within the following limits, based on reading rate: Slow: Between 15,000 counts and full range; Medium: Between 1,500 counts and full range; Fast: Between 150 counts and full range Maximum Crest Factor: 3.0 Maximum Input: To be used in protected, low-energy circuits only, not to exceed 250V or 4800 volt-amps mA: 300 mA dc or ac rms. Protected with a 500 mA, 250V, IEC 127 sheet 1 fast blow fuse. A: 10A dc or ac rms continuous or 20A dc or ac rms for 30 seconds maximum. Protected with a 15A, 250V, 10,000A interrupt rating fast blow fuse.